Visual Field Interpretation - EyeSuite - PeriTrend PDF
Visual Field Interpretation - EyeSuite - PeriTrend PDF
c Correct Patient & age d Correct refraction e Pupil size > 3mm ok 2.5 .. 3 mm borderline < 2.5mm critical f Catch trials < 15..20% false pos < % of severely depressed field false neg g Method, Strategy, Test duration Result h Greyscale of comparisons: "all white" or max. 3 small light yellow non-cluster dots Comparisons: < 4 numbers with 5..8dB in case of cataract evaluate "Corrected comparison" i Defect (Bebie) curve Within 5..95% bandwidth g overshoot left: happy trigger parallel lower: diffuse defect f drop on the right side: d local defect
h i
PeriTrend LV Deviation
j Global indices MD < 2dB "normal" & sLV < 2.5 dB "normal" MD 2..6dB "early" MD 6..12dB "moderate" MD > 12dB "advanced"
Octopus Visual Field Interpretation on screen with EyeSuite Perimetry Both eyes display Reliability c
c Correct age d Examination time
d i
e Pupil size > 3mm ok 2.5 .. 3 mm borderline < 2.5mm critical f Catch trials < 15..20% false pos < % of severely depressed field false neg
g Method, Strategy, Test duration Result h Greyscale of comparisons: "all white" or max. 3 small red non-cluster dots Comparisons: < 4 numbers with 5..8dB in case of cataract evaluate "Corrected comparison" i Qauadrant values & left/right comparison MD < 1.3dB sLV < 0.6dB
e g f
j Global indices MD < 2dB "normal" & sLV < 2.5 dB "normal" MD 2..6dB "early" MD 6..12dB "moderate" MD > 12dB "advanced"