100% found this document useful (1 vote)
469 views

Visual Field Interpretation - EyeSuite - PeriTrend PDF

The document provides guidance on interpreting visual field tests from two perimetry systems: PeriTrend & EyeSuite and Octopus. It lists factors to check for reliability including correct patient information, refractive correction, pupil size, and number of false positive and false negative catch trials. It also provides details on how to interpret the greyscale comparisons, defect curves, quadrant values, and global indices for mean deviation and loss variance to determine if results are within normal limits or indicate early, moderate, or advanced defects.

Uploaded by

mdmmm
Copyright
© Attribution Non-Commercial (BY-NC)
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
100% found this document useful (1 vote)
469 views

Visual Field Interpretation - EyeSuite - PeriTrend PDF

The document provides guidance on interpreting visual field tests from two perimetry systems: PeriTrend & EyeSuite and Octopus. It lists factors to check for reliability including correct patient information, refractive correction, pupil size, and number of false positive and false negative catch trials. It also provides details on how to interpret the greyscale comparisons, defect curves, quadrant values, and global indices for mean deviation and loss variance to determine if results are within normal limits or indicate early, moderate, or advanced defects.

Uploaded by

mdmmm
Copyright
© Attribution Non-Commercial (BY-NC)
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

PeriTrend & EyeSuite Perimetry Visual Field Interpretation on the 7-in-1 Printout Reliability c d e g f

c Correct Patient & age d Correct refraction e Pupil size > 3mm ok 2.5 .. 3 mm borderline < 2.5mm critical f Catch trials < 15..20% false pos < % of severely depressed field false neg g Method, Strategy, Test duration Result h Greyscale of comparisons: "all white" or max. 3 small light yellow non-cluster dots Comparisons: < 4 numbers with 5..8dB in case of cataract evaluate "Corrected comparison" i Defect (Bebie) curve Within 5..95% bandwidth g overshoot left: happy trigger parallel lower: diffuse defect f drop on the right side: d local defect

h i

PeriTrend LV Deviation

EyeSuite Perimetry sLV square root of Loss Variance DD Diffuse Defect

j Global indices MD < 2dB "normal" & sLV < 2.5 dB "normal" MD 2..6dB "early" MD 6..12dB "moderate" MD > 12dB "advanced"

Octopus Visual Field Interpretation on screen with EyeSuite Perimetry Both eyes display Reliability c
c Correct age d Examination time

d i

e Pupil size > 3mm ok 2.5 .. 3 mm borderline < 2.5mm critical f Catch trials < 15..20% false pos < % of severely depressed field false neg

g Method, Strategy, Test duration Result h Greyscale of comparisons: "all white" or max. 3 small red non-cluster dots Comparisons: < 4 numbers with 5..8dB in case of cataract evaluate "Corrected comparison" i Qauadrant values & left/right comparison MD < 1.3dB sLV < 0.6dB

e g f

j Global indices MD < 2dB "normal" & sLV < 2.5 dB "normal" MD 2..6dB "early" MD 6..12dB "moderate" MD > 12dB "advanced"