Equivalent Time Sampling (ETS)
With ETS
Without ETS
Average
Smoothing function
Envelope
With smoothing function
With averaging
Ohne Envelope
Mit Envelope
Without averaging
Without smoothing function
What modes of operation are
usually available to a DSO user?
For repetitive signals, time shifted samp-
ling of the waveform allows a higher
sample rate than the maximum specified
for the A/D converter.
In this mode of operation, all waveforms
remain visible on the display (similar to the
persistence display on analogue storage
oscilloscopes)
The averaging function suppresses ran-
dom signal components i.e. the signal can
be displayed without the noise.
This function corresponds to digital filte-
In this mode, the waveform is sampled at
the maximum rate of the A/D converter
irrespective of memory length or timebase
setting. The display is a Max-Min represen-
tation.
Without Envelope
With Envelope
Accumulate (Persistance)
Equivalent Time Sampling (ETS)
Smoothing function
Average
Envelope
S
C
S
I
I
E
E
E
-
4
8
8
R
S
-
2
3
2
-
C
ASCII data
123.456 123.456 123.456 123.456
123.456 123.456 123.456 123.456
123.456 123.456 123.456 123.456
123.456 123.456 123.456 123.456
123.456 123.456 123.456 123.456
123.456 123.456 123.456 123.456
123.456 123.456 123.456 123.456
123.456 123.456 123.456 123.456
Graphics file:
TIFF/BMP/HPGL/PostScript
How can the results
obtained by a modern
DSO be exported?
The measurement is finished - what next?
I
t should be possible to obtain a recording of the measurement results in various forms. A
paper record can be rapidly produced on site by using the built-in printer. If the results are
required for use by a pc or for later interpretation on the DSO, they can be stored on a Floppy
disk, either in a graphics file or in numerical form. Direct saving of the measurement data is the-
refore also possible in ASCII format. All file formats can be read and processed by most
Windows programs. If a DSO has a larger acquisition memory, it is not feasible to save all the
measurement data on a floppy disk, due to the size limitation. It is therefore preferable to use
a SCSI device for storing hundreds of kilobytes or even megabytes of data.
Instant documentation
with built-in printer