The document discusses converting differential pressure measurements from a transmitter to an air flow rate in m3/hr. It notes that the conversion requires a constant that depends on the instrument installation, which can be provided by the manufacturer. It also explains that the differential pressure transmitter assumes the use of a primary flow element like an orifice plate, and that the element manufacturer provides sizing information relating differential pressure to flow rate.

Uploaded by

sanjay975
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
161 views1 page

DP Transmitter For Flow

The document discusses converting differential pressure measurements from a transmitter to an air flow rate in m3/hr. It notes that the conversion requires a constant that depends on the instrument installation, which can be provided by the manufacturer. It also explains that the differential pressure transmitter assumes the use of a primary flow element like an orifice plate, and that the element manufacturer provides sizing information relating differential pressure to flow rate.

Uploaded by

sanjay975
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
You are on page 1/ 1

I'm using ABB differential Pressure Transmitter to calculate air flow in a circular pipe.

The Pressure transmitter is


calibrated using HART communicator in units of mmH2o at 20degcelsius.
Min-0 mmH2o (4mA)
Max- 78 mmH2o (20mA)
I would like to convert mmH2o to m3/hr. Any functional block or formula/Standard chart available for this conversion?

The formula is Constant*(Square root of the Diff Press) where constant depends on your instrument installation. The
manufacturer of the instrument should be able to tell you the constant.
To get flow rate from a DP transmitter assumes that you have a primary flow element mounted in a pipe or duct,
something like an orifice plate, an averaging pitot tube (Annubar), or one of several other specialized primary flow
elements.
Primary flow elements are designed to produce a certain differential pressure at a certain flow rate as calculated from
design information including medium viscosity, density, temperature, upstream static working pressure, and a dozen
others.
The manufacturer of the primary flow element provides a "sizing sheet" with the primary flow element that includes all
the data used to calculate the flow rate at a given DP. You need that sizing sheet, because that sizing sheet tells you
what flow rate 100"W.C. equals for your specific primary flow element.
And you need to know whether the DP transmitter is configured to output the 'raw' DP or whether it extracts the
square root of the DP.
And you should be aware that you cannot get SCFM (standard cubic feet per minute, a mass flow measurement)
from just a primary flow element and a DP transmitter, unless your static working pressure and temperature never
vary from the design standard, which never happens in the real world. 'Standard' cubic per minute is referenced back
to STP. Your DP transmitter does not measure gas temperature and the static pressure and compensate for those
changes, so your engineering units are CFM (a volumetric measurement), not SCFM (a mass flow measurement).