CMXDL Operation Manual
CMXDL Operation Manual
DAKOTA ULTRASONICS
CMX DL
Material & Coating Thickness Gauge
P/N P-171-0002
CHAPTER ONE
INTRODUCTION
The Dakota Ultrasonics model CMX DL is an ultrasonic thickness gauge that
measures with extreme versatility. It has the ability to measure simultaneously
measure coatings and material thicknesses while maintaining the ability to locate pits,
flaws and defects in the material. Based on the same operating principles as
SONAR, the CMX DL is capable of measuring the thickness of various materials with
accuracy as high as 0.001 inches, or 0.01 millimeters. The principle advantage of
ultrasonic measurement over traditional methods is that ultrasonic measurements
can be performed with access to only one side of the material being measured.
Dakota Ultrasonics maintains a customer support resource in order to assist users
with questions or difficulties not covered in this manual. Customer support may be
reached at any of the following:
CHAPTER TWO
QUICK STARTUP GUIDE
Turn the CMX DL on and off using the switch located on the bottom right corner of the
keypad. When CMX DL is initially turned on, a flash logo and blinking lights will be
displayed, followed by attempting to identify the transducer(probe) currently plugged
into the gauge. The CMX DL is equipped with an Auto Probe Recognition feature
that attempts to identify special transducers with this built in feature. If the CMX DL
doesnt find a transducer equipped with this feature, the user will be advanced to a
list of transducers requiring the user to select a specific transducer type. The
following sections outline each scenario. Note: This section is primarily written as a
basic startup guide only.
In order to understand how to operate the CMX DL, its best to start off with an
understanding of what it is were looking at exactly. The CMX DL has a lot of great
features and tools that will prove to be a huge benefit for the variety of applications
youre constantly facing on a continual basis. Lets have a brief look at the screens
youll be looking at most often:
A. Repeatability/Stability Indicator This indicator should be commonly used
in conjunction with the digital thickness values displayed. When all the vertical
bars are fully illuminated and the last digit on the digital thickness value is
stable, the CMX DL is reliably measuring the same value 3 to 200 times per
second, depending on which measurement mode and features are enabled.
Dakota Ultrasonics
1) Press the OK key once to use the identified probe, or ESC to display a list of
DL
the user should always select OK to use the identified probe. The only time
an alternative probe should be selected from a list is if the user switched
probes following initial power up and recognition.
DL
the CMX
3) Press the UP and DOWN arrow keys to toggle the coating option on/off.
4) Wipe all couplant from the transducer face and advance to the Probe Zero
& Calibration section outlined below.
Dakota Ultrasonics
1) Press the OK or ESC keys to display the factory list of transducer types (by
diameter and frequency).
2) Press the UP and DOWN arrow keys to scroll through the transducer list
until the appropriate type is highlighted.
3) Press the ENTER key to select the transducer type and display overwrite
existing probe screen.
4) Press the OK key to overwrite the existing probe type with the newly
selected probe type. The zero probe screen will be displayed. Proceed to
the zero probe section that follows.
Dakota Ultrasonics
1) Be sure all couplant has been removed from the face of the transducer.
2) Press the OK key to perform the automatic probe zero, or ESC key to
cancel the zero operation.
3) The screens illustrated above will be briefly displayed followed by the main
measurement screen. The CMX DL is ready to be calibrated.
Note: When the zero probe option is set to manual, the probe zero disk
(battery cap) located on the top of the gauge, will be used as a zero standard
and the warning screen illustrated above will be displayed.
1) Press the OK or ESC keys to enter the main measurement screen and
begin the manual zero process.
2) Apply a drop of couplant on the transducer and place the transducer in
steady contact with the probe zero disk, and obtain a steady reading.
3) Press the MENU key once to activate the menu items tab. Press the MENU
key multiple times to tab right and the ESC key multiple times to tab left until
the PRB menu is highlighted and displaying the submenu items.
4) Press the UP and DOWN arrow keys to scroll through the sub menu items
until ZERO PROBE is highlighted.
Dakota Ultrasonics
10
Note: Be sure that the probe zero procedure has been performed prior to
performing this calibration procedure.
1) Physically measure an exact sample of the material or a location directly on
the material to be measured using a set of calipers or a digital micrometer.
2) Apply a drop of couplant on the transducer and place the transducer in
steady contact with the sample or actual test material. Be sure that the
reading is stable and the repeatability indicator, in the top left corner of the
display, is fully lit and stable. Press the MENU key once to activate the
menu items tab. Press the MENU key multiple times to tab right and the
ESC key multiple times to tab left until the CAL menu is highlighted and
displaying the submenu items.
3) Use the UP and DOWN arrow keys to scroll through the sub menu items
until MATL 1PT is highlighted.
11
Dakota Ultrasonics
12
1) Press the MULTI MODE key once to activate the measurement mode
options.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items
until Coating Only (CT) mode is highlighted.
3) Press the ENTER key to select the measurement mode and return to the
measurement screen.
4) Apply a drop of couplant on the transducer and place the transducer in
steady contact with the probe zero disk (battery cover) and obtain a steady
reading.
Note: The coating measurement displayed will potentially be a value greater or
less than 0.
5) Press the MENU key once to activate the menu items tab. Press the MENU
key multiple times to tab right and the ESC key multiple times to tab left until
the PRB menu is highlighted and displaying the submenu items.
13
Dakota Ultrasonics
6) Use the UP and DOWN arrow keys to scroll through the sub menu items
until ZERO COATING is highlighted.
7) Press the ENTER key to display the confirmation screen.
8) Press the OK key to zero the coating and return to the PRB menu, or ESC
to cancel the coating zero process.
9) Press the MULTI MODE key once to activate the measurement mode
options.
10) Use the UP and DOWN arrow keys to scroll through the sub menu items
until Coating On (PECT) is highlighted.
11) Press the ENTER key to select the measurement mode and return to the
measurement screen, and begin taking readings.
14
12) Press the MENU key once to activate the menu items tab. Press the MENU
key multiple times to tab right and the ESC key multiple times to tab left until
the CAL menu is highlighted and displaying the submenu items.
13) Use the UP and DOWN arrow keys to scroll through the sub menu items
until COATING VEL is highlighted.
14) Press the ENTER key to display the Digits Edit Box.
15) Press the UP and DOWN arrow keys to scroll the highlighted value.
16) Press the LEFT and RIGHT arrow keys to scroll the digit locations.
15
Dakota Ultrasonics
17) Repeat steps 4 & 5 until the velocity number is correctly displayed.
18) Press the OK key to set the coating velocity and return to the menu screen,
or ESC to cancel entering the coating velocity.
19) Finally, press the MEAS key to return to the measurement screen and
begin taking readings.
16
Important Note: In PECT (pulse-echo coating) mode, the coating sample must
be coupled to metal in order to calibrate successfully. Simply place a drop of
couplant on a piece of metal, lay the coating sample over the couplant on the
metal and proceed to step 2.
2) Apply a drop of couplant on the transducer and place the transducer in
steady contact with the coating (on metal) sample or actual test material. Be
sure that the reading is stable and the repeatability indicator, in the top left
corner of the display, is fully lit and stable. Press the MENU key once to
activate the menu items tab. Press the MENU key multiple times to tab right
and the ESC key multiple times to tab left until the CAL menu is highlighted
and displaying the submenu items.
3) Use the UP and DOWN arrow keys to scroll through the sub menu items
until COATING 1PT is highlighted.
17
Dakota Ultrasonics
8) Press the OK key to calculate the coating velocity and return to the menu
screen, or ESC to cancel the one point calibration.
9) Finally, press the MEAS key to return to the measurement screen and begin
taking readings.
Note: CHECK YOUR CALIBRATION! Place the transducer back on the
calibration point. The coating thickness reading should now match the known
coating thickness sample. If the thickness is not correct, repeat the steps
above.
2.7 Measure
The CMX DL is now ready to measure. There are two different measurement view
options, each with a specific purpose Digits & B-Scan. The steps below outline
how to toggle between the different view mode options:
1) Press the MENU key once to activate the menu items tab. Press the MENU
key multiple times to tab right and the ESC key multiple times to tab left until
the DISP menu is highlighted and displaying the sub menu items.
18
2) Use the UP and DOWN arrow keys to scroll through the sub menu items
until VIEW is highlighted.
3) Use the LEFT and RIGHT arrow keys to scroll the view options.
4) Once the view is displayed, press the MEAS key to return to measurement
mode.
DIGITS: Displays the digital thickness value using a large font size. This view is
useful when the CMX DL is being used as a basic thickness gauge.
BSCAN: The Time Based B-Scan provides the user with a cross sectional view of
the material being tested. This mode is useful when there is concern regarding the
profile of the blind surface. This can also be a useful view when scanning for pits and
flaws.
Once the view has been selected according to the application requirements, the BSTART and B-DEPTH of the screen will potentially need to be adjusted if the view
has been set to BSCAN. Use the following steps to adjust these as follows:
Note: The B-Start and B-Depth are also used to adjust the parameters of Scan
Bar.
19
Dakota Ultrasonics
1) Press the MENU key once to activate the menu items tab. Press the MENU
key multiple times to tab right and the ESC key multiple times to tab left until
the DISP menu is highlighted and displaying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items
until B-START or B-DEPTH is highlighted.
3) Use the LEFT or RIGHT arrow keys to increase/decrease the start or depth
values in coarse increments/decrements.
Alternatively, the B-START and B-DEPTH values can be changed using the
Digit Edit Box as follows:
1) Use the UP and DOWN arrow keys to scroll through the sub menu items
until B-START or B-DEPTH is highlighted.
2) Press the ENTER key to display the digits edit box.
3) Press the UP and DOWN arrow keys to scroll the highlighted value.
4) Press the LEFT and RIGHT arrow keys to scroll the digit locations.
20
7) Finally, press the MEAS key to return to the measurement screen and begin
taking readings.
DIGITS
B-SCAN
In the upper left corner of each of the display photos above, is the repeatability
indicator. The repeatability indicator is represented by five vertical bars and
represents how repeatable the measurements are. In regular measurement mode,
the CMX DL makes 8 measurements a second. In scan mode, the CMX DL makes 200
measurements a second. If the coating mode option is activated, the CMX DL makes
3 measurements a second in regular measurement mode and 65 measurements a
second in scan mode. When the CMX DL is idle, only the left vertical bar will be
displayed. However, when the CMX DL is making a measurement, four or five of the
bars should be displayed on the repeatability indicator. If fewer than four bars are
showing, the CMX DL is having difficulty achieving a stable measurement and the
thickness value displayed is potentially unstable.
21
CHAPTER THREE
KEYBOARD, MENU, & CONNECTOR REFERENCE
22
1) Press the MENU key once to activate the menu items tab. Press the MENU
key multiple times to tab right, and the ESC key multiple times to tab left
until the desired tab group is highlighted and displaying the submenu items.
The tab groups are illustrated above (A).
Now that youre familiar with activating and moving amongst the tab groups, lets
have a look at how to move around in the sub menu items as follows:
The sections to follow will provide the user with an explanation of the sub menu
functions:
23
Dakota Ultrasonics
MAT: Select the material velocity from a chart of basic material types when a known
sample thickness, or material velocity cannot be obtained. Refer to page 52 for
further info.
MATL 1PT: Performs a single point calibration. This option allows the user to
automatically calculate the velocity by entering a known sample thickness. Refer to
page 49 for further info.
MATL 2PT: Performs a two-point calibration. This option allows the user to
automatically calculate the velocity by entering a second known sample thickness.
Refer to page 51 for further info.
VELOCITY: Function to calibrate the CMX DL by setting the velocity to a known
material velocity. Refer to page 46 for further info.
24
25
Dakota Ultrasonics
26
27
Dakota Ultrasonics
RESTORE SETUPS: Enables the user the ability to restore the setups currently
saved on a PC to an CMX DL via RS232 port. Refer the help section of the CMX DL
DakView software for a complete electronic manual.
BACKUP DATA: Enables the user the ability to backup grids or sequential log files
currently stored in the CMX DL to a PC via RS232 port. Refer the help section of the
CMX DL DakView software for a complete electronic manual.
RESTORE DATA: Enables the user the ability to restore grids or sequential log files
currently saved on a PC to an CMX DL via RS232 port. Refer the help section of the
CMX DL DakView software for a complete electronic manual.
ABOUT: Provides the user with Dakota Ultrasonics contact information and the
CMXDL software version. Refer the Dakota Ultrasonics web site for information on
the latest firmware versions available for download.
3.12 OK Key
The primary function of the OK key is confirmation of a change or selection. If the
CMXDL is displaying a grid log, the OK key toggles an advance to row number option.
28
29
The top & bottom end panels are where all connections are made to the CMX DL. The
diagram above shows the layout and description of the connectors:
Transducer Connectors
Refer to Diagram: The transducer connectors, and battery cover/probe zero disk are
located on the CMX DLs top end cap. The transducer connectors are of type Lemo
00. Note: There is no polarity associated with connecting the transducer to the
CMXDL.
Probe Zero Disk & Battery Cover
Refer to Diagram: The Battery cover is the large round disk shown in the diagram.
Note: This same disk is also used as a probe zero disk. Simply remove the cover
when replacing the batteries (3 AA cells). When performing a manual probe zero
function, simply place the transducer on disk making firm contact. Important: Be
sure to follow the polarity labels located on the back label of the CMX DL. Note:
Rechargeable batteries can be used, however they must be recharged outside of the
unit in a stand alone battery charger.
RS-232 Connector
Refer to Diagram: The RS-232 connector, located on the bottom end cap, is a 2 pin
female Lemo connector. It is designed to connect directly from the CMX DL to a
standard AT serial port on a PC. The cable supplied with the CMX DL is a Lemo to 9
pin serial cable. Note: This connector is also used to upgrade the CMX DL with the
latest version of firmware.
USB Serial to USB Converter Cable
A converter cable can be attached to the 9 pin serial cable in needed (part no. N-4020510).
30
CHAPTER FOUR
PRINCIPALS OF ULTRASONIC MEASUREMENT
4.1 Time versus thickness relationship
Ultrasonic thickness measurements depend on measuring the length of time it takes
for sound to travel through the material being tested. The ratio of the thickness
versus the time is known as the sound velocity. In order to make accurate
measurements, a sound velocity must be determined and entered into the
instrument.
The accuracy of a thickness measurement therefore depends on having a consistent
sound velocity. Some materials are not as consistent as others and accuracy will be
marginal. For example, some cast materials are very granular and porous and as a
result have inconsistent sound velocities.
While there are many different ultrasonic techniques to measure thickness, which will
be discussed below, all of them rely on using the sound velocity to convert from time
to thickness.
4.4 Couplant
All ultrasonic applications require some medium to couple the sound from the
transducer to the test piece. Typically a high viscosity liquid is used as the medium.
The sound frequencies used in ultrasonic thickness measurement do not travel
31
Dakota Ultrasonics
through air efficiently. By using a liquid couplant between the transducer and test
piece the amount of ultrasound entering the test piece is much greater.
4.5 Temperature
Temperature has an effect on sound velocity. The higher the temperature, the slower
sound travels in a material. High temperatures can also damage transducers and
present a problem for various liquid couplants.
Since the sound velocity varies with temperature it is important to calibrate at the
same temperature as the material being measured.
Normal temperature range
Most standard transducers will operate from 0F to 180F.
High temperature measurements
Special transducers and couplants are available for temperatures above 180F up to
650F with intermittent contact. It is necessary to cool the transducer, by submerging
the transducer in water between readings, when measuring high temperatures.
Modes and temperature errors
In addition to errors caused by velocity changing with temperature, some modes
(measurement techniques) are affected more than others. For e xample, dual
element pulse-echo mode has larger errors due to changes in the temperature of the
delay line. However, multi-echo techniques offer temperature compensation help to
minimize these errors.
32
33
Dakota Ultrasonics
34
35
CHAPTER FIVE
SELECTING THE MEASUREMENT MODE
5.1 The setup library
The CMX DL contains 64 user configurable preset locations to store custom setups for
easy recall. These setups can be optimized for the users specific application needs
and can also be stored on a PC and transferred bi-directionally using Dakotas PC
interface software included with the instrument.
The setups supplied with the instrument cover some of the more typical applications
commonly used with this type of instrument. These setups can be recalled, modified,
and overwritten to one of 64 setup locations. Therefore, these factory setups can
also be considered a good starting point to be modified for custom applications. The
PC software includes a default setup file that can be uploaded to the gauge at any
time to restore factory settings. However, it is recommended that the user consider
saving modified setups to an empty location rather than overwriting the factory setups
in the CMX DL. Once again, these factory settings are excellent starting points for
custom setups.
36
37
Dakota Ultrasonics
Restricted access
Measuring materials with extreme curvatures or restricted access, higher frequencies
with smaller diameters should be considered. The smallest diameter uses 3/16
crystals with a contact area of .250. Custom transducers are available on request.
Name
3
4
Comment 1
38
Gn/AGC
Velocity
CHAPTER SIX
MAKING MEASUREMENTS
The steps involved in making measurements are detailed in this section. The
following sections outline how to setup and prepare your CMX DL for field use.
An automatic or manual zero must always be performed. The auto zero is an off
block electronic zero that does not require a zero reference block. This will most
always be the zero option of choice, as it makes the zeroing process very easy and
convenient to perform. However, If the manual zero option is enabled, the probe
zero must be measured on the reference disk (battery disk) attached to the top of the
instrument. The zero compensates for variations in the transducer. In all modes the
sound velocity must be determined. The sound velocity is used to convert the transit
time to a physical length. The sound velocity can be selected from a material chart in
the manual, selected from a material list in the CMX DL, or for greater precision
determined from a sample of the test material that has been mechanically measured.
To enter the velocity from a table, look up the material on the chart in the appendix of
this manual and refer to the section below on Calibration to a Known Velocity. To
determine the velocity of a single sample, refer to the Material Calibration section on
page 49.
When measuring curved materials, it is more accurate to calibrate from two test
points, one at the minimum limit of the target thickness and one at the maximum limit.
In this case the reference disk mounted to the CMX DL is not used. This is called two point calibration and is described on page 51.
39
Dakota Ultrasonics
1) Press the OK key once to use the identified probe, or ESC to display a list of
DL
the user should always select OK to use the identified probe. The only time
an alternative probe should be selected from a list is if the user switched
DL
has
the CMX
40
3) Press the UP and DOWN arrow keys to toggle the coating option on/off.
4) Wipe all couplant from the transducer face and proceed to the Probe Zero
section that follows .
In this second example the transducer was not identified and will force the user to
select the transducer type from a predefined list of transducers:
5) Press the OK or ESC keys to display the factory list of transducer types (by
diameter and frequency).
41
Dakota Ultrasonics
6) Press the UP and DOWN arrow keys to scroll through the transducer list
until the appropriate type is highlighted.
7) Press the ENTER key to select the transducer type and display overwrite
existing probe screen.
8) Press the OK key to overwrite the existing probe type with the newly
selected probe type. The zero probe screen will be displayed. Proceed to
the zero probe section that follows.
42
1) Be sure all couplant has been removed from the face of the transducer.
43
Dakota Ultrasonics
2) Press the OK key to perform the automatic probe zero, or ESC key to
cancel the zero operation.
3) The screens illustrated above will be briefly displayed followed by the main
measurement screen. The CMX DL is ready to be calibrated.
44
Note: When the zero probe option is set to manual, the probe zero disk
(battery cap) located on the top of the gauge will be used as a zero standard
and the warning screen illustrated above will be displayed.
1) Press the OK or ESC keys to enter the main measurement screen and
begin the manual zero process.
2) Apply a drop of couplant on the transducer and place the transducer in
steady contact with the probe zero disk and obtain a steady reading.
3) Press the MENU key once to activate the menu items tab. Press the MENU
key multiple times to tab right and the ESC key multiple times to tab left until
the PRB menu is highlighted and displaying the submenu items.
4) Press the UP and DOWN arrow keys to scroll through the sub menu items
until ZERO PROBE is highlighted.
45
Dakota Ultrasonics
7) Press the OK key to complete the probe zero function, or ESC key to cancel
the probe zero function.
8) Remove the transducer from the probe zero disk, and proceed to the
calibration section.
Note: The value that is displayed will change depending on the current velocity
setting in the CMX. Disregard the number that is displayed. It is not
important. What is important is accurately performing the steps outlined above
to insure reliability of the probe zero calculation.
Known Velocity
If the material velocity is known, the user may wish to simply enter the velocity
number into the CMX DL, rather than have the CMX DL calculate the velocity value
using a known thickness on a material sample. The steps for entering the velocity
are outlined below:
46
1) Press the MENU key once to activate the menu items tab. Press the MENU
key multiple times to tab right and the ESC key multiple times to tab left until
the CAL menu is highlighted and displaying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items
until VELOCITY is highlighted.
47
Dakota Ultrasonics
7) Press the OK key to set the velocity and return to the menu screen, or ESC
to cancel entering the velocity.
8) Finally, press the MEAS key to return to the measurement screen and begin
taking readings.
48
Known Thickness
Sometimes the sound velocity of a material is unknown. In this case a sample with
one or two known thicknesses can be used to determine the sound velocity. As
previously discussed, the CMX DL has a one or two point calibration option. The one
point calibration option is most suited for linearity o ver large ranges, as noted above.
The user should also consider calibrating on high side of the intended measurement
range, when using the one point option, minimize overall error. For example, if the
measurement range is .100 (2.54mm) to 1.0 (25.4mm), the user should calibrate on
a known thickness sample close to 1.0 (25.4mm). Note: Its always handy to carry a
set of mechanical calipers to use in conjunction with the CMX DL for calibration in the
field:
Note: Be sure that the probe zero procedure has been performed prior to
performing this calibration procedure.
1) Physically measure an exact sample of the material or a location directly on
the material to be measured using a set of calipers or a digital micrometer.
2) Apply a drop of couplant on the transducer and place the transducer in
steady contact with the sample or actual test material. Be sure that the
reading is stable and the repeatability indicator, in the top left corner of the
display, is fully lit and stable. Press the MENU key once to activate the
menu items tab. Press the MENU key multiple times to tab right and the
49
Dakota Ultrasonics
ESC key multiple times to tab left until the CAL menu is highlighted and
displaying the submenu items.
3) Use the UP and DOWN arrow keys to scroll through the sub menu items
until MATL 1PT is highlighted.
50
51
Dakota Ultrasonics
52
1) Press the MENU key once to activate the menu items tab. Press the MENU
key multiple times to tab right and the ESC key multiple times to tab left until
the CAL menu is highlighted and displaying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items
until MAT is highlighted.
53
Dakota Ultrasonics
4) Press the UP and DOWN arrow keys to scroll through the material list until
the appropriate material is highlighted.
5) Press the ENTER key to overwrite the material type and display the menu
items with the new material type selected.
6) Finally, press the MEAS key to return to the measurement screen and begin
taking readings.
To calibrate the CMX DL for a specific type of coating using samples with known
thicknesses, please refer to the chapter 9 Pulse-Echo Coating (PECT) or
Coating (CT) for a complete explanation of using the CMX DL for coating
measurements.
54
CHAPTER SEVEN
USING THE DIGITS & B-SCAN DISPLAYS
A key feature of the CMX DL is the ability to toggle between two different display
options, Digits and B-Scan. The Digits view provides the user with a large digital
readout of the thickness. Both views also include a scan bar features that uses a bar
graph to indicate thickness. The Scan bar can be very handy while scanning a tank
or pipe wall by graphically displaying deflections from pits or internal flaws , rather
than having to constantly watch for changes in the digital readout. Its just another
visual aid to alarm the user when something out of the ordinary has been detected.
The B-Scan display is also very useful when scanning surfaces and viewing the cross
section of the test material. It provides a convenient way of visually profiling, or
drawing a picture of, the blind surfaces during a scan. The B-Scan display is also
equipped with a scan bar representing the overall thickness. Once again, the scan
bar gives the user a visual indication when a flaw or defect has been passed over
during the scan process. The scan bar will deflect off of the defect and return back to
the overall thickness. In this chapter well outline some of the fine adjustment
features of the CMX DL. Well take a better look at these options in this chapter.
Note: In order to recall and use the new adjustments made to the CMX DL at a later
time, the user must save the modified settings in one of the setup locations . Refer
page 112 for more information on setups.
55
Dakota Ultrasonics
DIGITS VIEW
DIGITS
The Digits view is a basic digital thickness gauge look and feel. The larger digits
make it much easier for the operator to monitor the thickness readings. The Scan
Bar has also been added to the Digits view to provide the user with yet another visual
tool for easily monitoring changes in thickness readings due to internal flaws or
defects.
The following is a list of the viewable features on the display:
A. Repeatability/Stability Indicator This indicator should be commonly used
in conjunction with the digital thickness values displayed. When all the vertical
bars are fully illuminated and the last digit on the digital thickness value is
stable, the CMX DL is reliably measuring the same value 3 to 200 times per
second, depending on which measurement mode and features are enabled.
B. Battery Icon Indicates the amount of battery life the CMX DL has remaining.
C. Velocity The material velocity value the CMX DL is currently using or
calibrated for. Displayed in English or Metric units, depending on the what
units the gauge is set for.
D. Feature Status Bar Indicates the features currently enabled and in use in
the following order:
Measurement Mode
Differential Mode
High Speed Scan Mode
56
E.
F.
G.
H.
I.
J.
Alarm Mode
Gain Setting
Digital Material Thickness Value Extra large font size for viewing ease.
Scan Bar Another view of material thickness in a deflection style horizontal
bar. This is another visual tool that would enable the user the ability to see
thickness changes during high speed scans from flaws and pits.
Units The current measurement units being used (English, Metric).
Coating Thickness Value Displays the actual thickness of any coating
adhered to a metallic material surface (PECT Mode), or a coating adhered to a
non-metallic surface (CT Mode).
Minimum Material Thickness Part of the Alarm feature. Displays the
minimum thickness value found during a scan.
Maximum Material Thickness Part of the Alarm feature. Displays the
maximum thickness value found during a scan.
B-Scan View
B-Scan
The B-Scan displays a time based cross section view of test material. This view is
commonly used to display the contour of the blind, or underside, surface of a pipe or
tank application. It is very similar to a fish finder. If a flaw or pit is located during a
scan, the B-Scan will draw the pit on the screen. The solid black rectangle in the
diagram at location K represents the cross section, or side view of the material. The
B-Scan view draws at a rate of 7 seconds per screen from right to left. Also notice at
location K, the pits and corroded bottom surface of the material.
Its important to note that the measurement range on the display be set wide enough,
so that the maximum thickness of the material can be viewed on the display. Using
the diagram above, if the material thickness was actually 1.75, the underside of the
material would not be viewable according to the current range at 0.00 1.00. All
57
Dakota Ultrasonics
the user would see is a black screen from 0.00 1.00 with no view of the bottom
contour at 1.75.
The following is a list of the viewable features on the display:
A. Repeatability/Stability Indicator This indicator should be commonly used
in conjunction with the digital thickness values displayed. When all the vertical
bars are fully illuminated and the last digit on the digital thickness value is
stable, the CMX DL is reliably measuring the same value 3 to 200 times per
second, depending on which measurement mode and features are enabled.
B. Battery Icon Indicates the amount of battery life the CMX DL has remaining.
C. Velocity The material velocity value the CMX DL is currently using or
calibrated for. Displayed in either English or Metric units, depending on the
what units the gauge is set for.
D. Feature Status Bar Indicates the features currently enabled and in use in
the following order:
Measurement Mode (P-E, PECT, PETP, E-E, COAT)
Differential Mode (ON/OFF)
High Speed Scan Mode (ON/OFF)
Alarm Mode (ON/OFF/AUDIBLE)
Gain Setting (VLOW, LOW, MED, HI, VHI)
E. Digital Material Thickness Value Smaller font size when the B-Scan
display view is enabled.
F. Scan Bar Another view of material thickness in a deflection style horizontal
bar. This is another visual tool that would e nable the user the ability to see
thickness changes during high speed scans from flaws and pits.
G. Units The current measurement units being used (English, Metric).
H. Coating Thickness Value Displays the actual thickness of any coating
adhered to a metallic material surface (PECT Mode), or a coating adhered to a
non-metallic surface (CT Mode).
I. Minimum Material Thickness Part of the Alarm feature. Displays the
minimum thickness value found during a scan.
J. Maximum Material Thickness Part of the Alarm feature. Displays the
maximum thickness value found during a scan.
K. B-Scan Display Cross section view of the material. Provides the user with
graphical view of the opposite/blind surface (i.e. inside pipe wall surface), to
give the user some idea of the condition, or integrity of the material being
tested.
58
1) Press the MENU key once to activate the menu items tab. Press the MENU
key multiple times to tab right, and the ESC key multiple times to tab left,
until the DISP menu is highlighted and displaying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items
until VIEW is highlighted.
3) Press the LEFT and RIGHT arrow keys to toggle DIGITS or BSCAN. When
the correct view is being displayed continue on to setting the B-START or BDEPTH values, or press the MEAS key to return to the measurement screen
and begin taking readings.
59
Dakota Ultrasonics
60
1) Press the MENU key once to activate the menu items tab. Press the MENU
key multiple times to tab right, and the ESC key multiple times to tab left,
until the DISP menu is highlighted and displaying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items
until B-START is highlighted.
3) Press the LEFT and RIGHT arrow keys to scroll the value. When the
correct B-Start value is being displayed, proceed to step 8.
4) Alternatively , press the ENTER key to display the Digits Edit Box.
61
Dakota Ultrasonics
5) Press the UP and DOWN arrow keys to scroll the highlighted value.
6) Press the LEFT and RIGHT arrow keys to scroll the digit locations.
7) Repeat steps 5 & 6 until the B-Start value is correctly displayed.
8) Press the OK key to set the B-Start value and return to the menu screen, or
ESC to cancel entering the B-Start.
9) Finally, press the MEAS key to return to the measurement screen and begin
taking readings.
1) Press the MENU key once to activate the menu items tab. Press the MENU
key multiple times to tab right, and the ESC key multiple times to tab left,
until the DISP menu is highlighted and displaying the submenu items.
62
2) Use the UP and DOWN arrow keys to scroll through the sub menu items
until B-DEPTH is highlighted.
3) Press the LEFT and RIGHT arrow keys to scroll the value. When the
correct width is being displayed, proceed to step 8.
4) Alternatively , press the ENTER key to display the Digits Edit Box.
5) Press the UP and DOWN arrow keys to scroll the highlighted value.
6) Press the LEFT and RIGHT arrow keys to scroll the digit locations.
7) Repeat steps 5 & 6 until the B-Depth value is correctly displayed.
8) Press the OK key to set the B-Depth and return to the menu screen, or ESC
to cancel entering the B-Depth.
9) Finally, press the MEAS key to return to the measurement screen and begin
taking readings.
63
Dakota Ultrasonics
7.4 Gain
The gain, or amplification of the return echoes, can be adjusted in the CMX DL to
accommodate a variety of applications. The setting of the gain is crucial in order to
obtain valid readings during the measurement process. Too much gain may result in
erroneous measurements, by detecting on noise rather than the actual material back
wall itself. Not enough gain may result in intermittent detection. It may also result in
lack of detection on internal flaws, pits, or porosity. The gain can easily be compared
to the volume control of a home stereo system. If you turn it up too much, you cant
hear the music clearly. If its turned down too much, you cant hear it at all.
The CMX DL has three gain settings (VLOW, LOW, MED, HI, VHI). The gain range is
42dB 50dB in 2dB increments. Note: The CMX DL has been optimized for the MED
gain setting for all common applications. It should be operated in this mode as
standard. However, some applications may require the lower or higher gain settings.
When? The low settings may be necessary for noisy or granular cast materials. How
do I know when to lower the gain? If the reading becomes sporadic and wont settle
down or resolve on a thickness value, the user can assume that the material is either
very noisy aluminum, or granular cast iron. This would be a good time to change the
CMXDL to lower gain setting and see if the reading settles down and become stable.
How do I know when to increase the gain? Often times the user will be trying to
measure a material that is hard to penetrate or pass sound through. This may be
due to the material type, or overall thickness of the material. When a material is hard
to pass sound through because of the thickness or general make-up, it would be a
good time to consider switching the CMX DL into a higher gain setting. Another
example might be the need to increase overall sensitivity for locating fine pits or
flaws. In any case, the selectable gain settings offer the user some additional options
to resolve and overcome application issues.
Note: When the echo-echo thru-paint measurement mode is selected, the manual
gain feature is disabled and grayed out in the menu items. In this mode, the CMX DL
switches to an automatic gain mode (AGC) that optimizes the gain setting
automatically in the hardware of the CMX DL.
The procedures to adjust the Gain are outlined below:
64
1) Press the MENU key once to activate the menu items tab. Press the
MENU key multiple times to tab right, and the ESC key multiple times to tab
left, until the TUNE menu is highlighted and displaying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items
until GAIN is highlighted.
3) Press the LEFT and RIGHT arrow keys to scroll the value. When the
correct Gain is being displayed, press the MEAS key to return to the
measurement screen and begin taking readings.
65
CHAPTER EIGHT
THRU PAINT MEASUREMENT TECHNIQUE
8.1 Introduction to Thru Paint Measurement
The principle behind thru paint measurement is by measuring the time between two
backwall echoes returning from the test material. Since both of these backwall
echoes travel the same path through the paint or coating, the thickness of the coating
is subtracted out of the measurement so that only the actual material thickness can
be measured. This feature saves the user a great deal of time scraping and
removing the coating from tanks and pipes during the inspection process.
The primary purpose of thru paint measurement is to determine actual or overall
material thickness by eliminating the coating thickness. Thru paint mode cannot be
used for flaw or pit detection. Therefore, inspectors may need to use echo-echo thru
paint mode in conjunction with the standard pulse-echo flaw detection mode for some
applications. Chassis tubing inspectors and sanctioning bodies will use the echo echo mode for tubing with powder coatings, and pulse-echo mode for tubing without
coating. The combination of using both modes is ideal for the advanced inspectors
needs.
CHAPTER NINE
PULSE-ECHO COATING & COATING TECHNIQUES
9.1 Introduction to Pulse-Echo Coating Measurement (PECT)
In the previous sections weve discussed the need for detecting pits and flaws (pulseecho) in materials, along with the requirement to measure through and eliminate
errors caused by coated materials (echo-echo). Until now, both modes were needed
in order accomplish both tasks. Pulse-echo mode was used for flaw detection, and
echo-echo mode was used to eliminate the coating thickness and provide a nominal
material thickness only. With this in mind, it often became cumbersome toggling
between both modes respectively.
In a majority of applications involving coated materials, inspectors are often
interested in monitoring both the thickness of the material, as well as the thickness of
the coating . Therefore, the CMX DL has been designed to provide the user with the
ability to measure the material and coating thicknesses simultaneously, while
maintaining the ability to detect flaws and pits all in a single mode called Pulse-Echo
Coating (PECT). This is accomplished by using a custom hybrid combination mode
utilizing properties from the basic modes along with a group of special techniques
and theoretical wave phenomenas.
The CMX DL is preset to a coating velocity of 0.0850 in/sec (2159 m/sec) from the
factory. This velocity is a very close approximation of the common coating velocities
found in the field. However, the CMX DL also has a built-in one and two point
calibration options to calibrate to coating samples with a known thicknesses.
There are two coating modes built into the CMX DL:
1) Pulse-Echo Coating (PECT) used during inspections where both the
material and coating thicknesses are required.
2) Coating Only (CT) used during inspections where only the coating thickness
is required.
67
Dakota Ultrasonics
3) From the tabbed menus under TUNE, MEASURE MODE.
The steps that follow will demonstrate all three methods in the order listed above:
1) Press the OK key once to use the identified probe, or ESC to display a list of
DL
the user should always select OK to use the identified probe. The only time
an alternative probe should be selected from a list is if the user switched
probes following initial power up and recognition.
DL
the CMX
68
3) Press the UP and DOWN arrow keys to toggle the coating option on/off.
Applied to Metals
1) Press the MULTI MODE key located on bottom left of the keypad to display
the MEASURE MODE options menu.
2) Press the UP and DOWN arrow keys to highlight the COATING ON (PECT)
or COATING ONLY (CT) menu option.
69
Dakota Ultrasonics
3) Press the ENTER key to enable the coating option, or ESC to cancel
changing the measure mode, and return to the main measurement screen.
Applied to Metals
1) Press the MENU key once to activate the menu items tab. Press the MENU
key multiple times to tab right, and the ESC key multiple times to tab left,
until the TUNE menu is highlighted and displaying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items
until MEASURE MODE is highlighted.
3) Press the LEFT and RIGHT arrow keys to scroll the modes.
4) When the correct MODE is being displayed, press the MEAS key to return
to the measurement screen and begin taking readings.
70
1) Press the MULTI MODE key once to activate the measurement mode
options.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items
until Coating Only (CT) mode is highlighted.
3) Press the ENTER key to select the measurement mode and return to the
measurement screen.
4) Apply a drop of couplant on the transducer and place the transducer in
steady contact with the probe zero disk (battery cover) and obtain a steady
reading.
Note: The coating measurement displayed will potentially be a value greater or
less than 0.
71
Dakota Ultrasonics
5) Press the MENU key once to activate the menu items tab. Press the MENU
key multiple times to tab right and the ESC key multiple times to tab left until
the PRB menu is highlighted and displaying the submenu items.
6) Use the UP and DOWN arrow keys to scroll through the sub menu items
until ZERO COATING is highlighted.
7) Press the ENTER key to display the confirmation screen.
8) Press the OK key to zero the coating and return to the PRB menu, or ESC
to cancel the coating zero process.
9) Press the MULTI MODE key once to activate the measurement mode
options.
10) Use the UP and DOWN arrow keys to scroll through the sub menu items
until Coating On (PECT) is highlighted.
11) Press the ENTER key to select the measurement mode and return to the
measurement screen, and begin taking readings. Proceed to the coating
calibration section.
72
1) Press the MENU key once to activate the menu items tab. Press the MENU
key multiple times to tab right and the ESC key multiple times to tab left until
the CAL menu is highlighted and displaying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items
until COATING VEL is highlighted.
73
Dakota Ultrasonics
74
Known Thickness
Sometimes the sound velocity of a coating material is unknown. In this case a
sample with a known thickness can be used to determine the sound velocity of the
coating. As previously discussed, the CMX DL offers a one point calibration option for
coating in PECT measurement mode. It is also import to reiterate that the coating
sample must be coupled to a metallic material in order to perform the calibration.
The calibration should also consider calibrating on high side of the intended
measurement range, when using the one point option, minimize overall error. For
example, if the measurement range of the coating will be from .040 - .080 (1.0mm
2.3mm), a coating closest to .080 (2.0mm) should be used. Note: Its always handy
to carry a set of mechanical calipers to use in conjunction with the CMX DL for
calibration in the field:
Note: Be sure that the probe zero procedure has been performed prior to
performing this calibration procedure.
1) Physically measure an exact sample of the coating using a set of calipers or
a digital micrometer.
Note: The coating sample must be coupled to a metallic material in order to be
calibrated.
2) Apply a drop of couplant on a piece of metal and place the coating sample
on the metal over the couplant. Apply a drop of couplant on the transducer
75
Dakota Ultrasonics
and place the transducer in steady contact with the coating and sample or
actual test material. Be sure that the reading is stable and the repeatability
indicator, in the top left corner of the display, is fully lit and stable. Press the
MENU key once to activate the menu items tab. Press the MENU key
multiple times to tab right and the ESC key multiple times to tab left until the
CAL menu is highlighted and displaying the submenu items.
3) Use the UP and DOWN arrow keys to scroll through the sub menu items
until COATING 1PT is highlighted.
76
77
Dakota Ultrasonics
Note: Use the maximum coating sample for the one point calibration first.
1) Physically measure the thicker of the two samples of coating, as close as
possible to the maximum expected coating measurement range, using a set
of calipers or a digital micrometer.
Very Important: If coating measurements will be made with the coating
applied to a metal surface, the calibration must be done in the same manner,
with the samples coupled to a metal surface. However, if the coating will be
measured as a stand alone material, the calibration must be performed the
same way.
2) Apply a drop of couplant on the transducer and place the transducer in
steady contact with the thicker of the two coating and samples fist. Be sure
that the reading is stable and the repeatability indicator, in the top left corner
of the display, is fully lit and stable. Press the MENU key once to activate
the menu items tab. Press the MENU key multiple times to tab right and the
ESC key multiple times to tab left until the CAL menu is highlighted and
displaying the submenu items.
78
3) Use the UP and DOWN arrow keys to scroll through the sub menu items
until COATING 1PT is highlighted.
79
Dakota Ultrasonics
Note: Use the minimum coating sample for the two point calibration.
1) Physically measure the thinner of the two samples of the coating, as close
as possible to the minimum expected coating measurement range, using a
set of calipers or a digital micrometer.
Very Important: If coating measurements will be made with the coating
applied to a metal surface, the calibration must be done in the same manner,
with the samples coupled to a metal surface. However, if the coating will be
measured as a stand alone material, the calibration must be performed the
same way.
2) Apply a drop of couplant on the transducer and place the transducer in
steady contact with the thinner of the two coating and samples. Be sure that
the reading is stable and the repeatability indicator, in the top left corner of
the display, is fully lit and stable. Press the MENU key once to activate the
menu items tab. Press the MENU key multiple times to tab right and the
ESC key multiple times to tab left until the CAL menu is highlighted and
displaying the submenu items.
80
3) Use the UP and DOWN arrow keys to scroll through the sub menu items
until COATING 2PT is highlighted.
81
CHAPTER TEN
ADDITIONAL FEATURES OF THE CMXDL
10.1 High Speed Scan
The High Speed Scan feature of the CMX DL increases the overall repetition rate to a
maximum of 140Hz with a high speed screen refresh rate of 25 times a second. This
feature enables a user to make scanned passes over an arbitrary length of the test
material, while still maintaining a reasonable representation of thickness over the
scanned area or region. This feature can be used in conjunction with High and Low
alarm limits features to dynamically keep track of both values. The feature is typically
used to provide a better representation of the area scanned, by taking more readings
at a faster rate repetition rate.
The procedure to use the SCAN MODE feature is outlined below:
1) Press the MENU key once to activate the menu items tab. Press the MENU
key multiple times to tab right and the ESC key multiple times to tab left until
the UTIL menu is highlighted and displaying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items
until SCAN MODE is highlighted.
3) Use the LEFT and RIGHT arrow keys to toggle the SCAN MODE on/off.
82
1) Press the MENU key once to activate the menu items tab. Press the MENU
key multiple times to tab right, and the ESC key multiple times to tab left,
until the UTIL menu is highlighted and displaying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items
until ALARM is highlighted.
3) Use the LEFT and RIGHT arrow keys to toggle the ALARM on/off/audible.
4) Continue on to the next section Setting the Alarm Low Limit.
83
Dakota Ultrasonics
1) Assuming the ALARM is ON, use the UP and DOWN arrow keys to scroll
through the sub menu items until ALARM LOW is highlighted.
2) Press the LEFT and RIGHT arrow keys to scroll the value. When the
correct alarm value is being displayed, proceed to step 7 .
3) Alternatively , press the ENTER key to display the Digits Edit Box.
4) Press the UP and DOWN arrow keys to scroll the highlighted value.
5) Press the LEFT and RIGHT arrow keys to scroll the digit locations.
84
6) Repeat steps 4 & 5 until the ALARM LOW value is correctly displayed.
7) If only one limit will be used, press the MEAS key to return to the
measurement screen and begin taking readings. Otherwise, continue on to
set the ALARM HIGH limit.
1) Press the MENU key once to activate the menu items tab. Press the MENU
key multiple times to tab right, and the ESC key multiple times to tab left,
until the UTIL menu is highlighted and displaying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items
until DIFFERENTIAL is highlighted.
85
Dakota Ultrasonics
3) Use the LEFT and RIGHT arrow keys to toggle the DIFFERENTIAL on. A
value will appear to the right of DIFFERENTIAL.
3) Press the LEFT and RIGHT arrow keys to scroll the digit locations.
4) Repeat steps 2 & 3 until the DIFFERENTIAL value is correctly displayed.
5) Press the OK key to set the DIFFERENTIAL value and return to the menu
screen.
6) Finally, press the MEAS key to return to the measurement screen and begin
taking readings.
86
CHAPTER ELEVEN
DATA STORAGE SETUP, EDIT, & VIEW FILES
11.1 Introduction to Grid and Sequential file formats
The CMX DL is equipped with two data file format options, GRID LOG and SEQ LOG.
The GRID file format is very similar to a spreadsheet format found in popular
software programs like Excel. A GRID is simply a table of readings. A location in a
grid is specified by giving a row and column coordinate. The rows are numbered
from 1 to 999 and the columns are labeled from A to ZZ (999 Rows & 52 Columns).
The sequential file format can be viewed as a file as a single column of up to 512
possible rows (readings), and a column of corresponding identifiers associated with
each individual reading. The identifier can be a combination of up to 10 numeric,
alpha, or special characters listed above, while the file name can consist of a
combination of up to 20 of the same character set. Note: The identifier cannot start
or end with a special character. Once a start and end ID are entered into the CMX DL
and the log created, the CMX DL will automatically generate all the identifiers within
that range.
The following character set listed below are all the allowable characters that will be
used for both file formats: GRID & SEQ LOG. Any combination of these characters
can be used for creating a Name and Note regardless of the selected format. The
allowable characters are as follows:
Numeric characters: 0 9 Alpha Characters: A Z
Special Characters: ! _ # space / . ( )
Multiple grids can be created and stored until the CMX DLs memory is full. If the user
attempts to store a new file in the CMX DL and the size of the file exceeds the capacity
of memory, the CMX DL will respond with an error message indicating that the memory
is unable to store the new file.
The CMX DL can store a total of 16,000+ readings with a corresponding screenshot of
the DIGITS or B-SCAN views with every reading , as well as all the CMX DL settings
for every individual reading . If the graphics option is disabled, not saving
screens hots, the CMX DL can store a total of 210,000+ readings. The screenshot
stored with the reading will depend on what view the CMX DL was in at the time it was
saved. For example, if the view was in B-SCAN, then a B-SCAN will be saved with
the reading.
In the sections that follow, the procedures for creating, using, and editing GRIDs and
SEQ LOGs have been combined together for the purpose of similarity in overall
87
Dakota Ultrasonics
functionality and structure. The illustrations below are snapshots of typical GRID and
SEQ LOG file formats:
Grid File Formats
Important Note: For the duration of this chapter, all references to GRIDS and SEQ
LOGS should be considered synonymous with references to FILES.
88
Creating a Name
Grid/Seq Log Name : Can contain a combination of up to 20 numeric, alpha,
or special characters listed in the first section of this chapter.
Grid Log
Sequential Log
1) Press the MENU key once to activate the menu items tab. Press the MENU
key multiple times to tab right, and the ESC key multiple times to tab left,
until the DATA menu is highlighted and displaying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items
until NEW is highlighted.
3) Press the LEFT & RIGHT arrow keys to toggle the data formats GRIG
LOG & SEQ LOG.
89
Dakota Ultrasonics
Grid Log
Sequential Log
4) Press the ENTER key to display the new Grid or Seq Edit Box.
5) Use the UP and DOWN arrow keys to scroll through the new Grid or Seq
List Items until NAME is highlighted.
90
Creating a Note
Grid/Seq Log Note: Can contain a combination of up to 20 numeric, alpha, or
special characters listed in the first section of this chapter.
Grid Log
Sequential Log
1) Assuming the Grid or Seq List Items are displayed, use the UP and DOWN
arrow keys to scroll through the new Grid or Seq List Items until NOTE is
highlighted.
2) Press the ENTER key to activate the Alpha Edit Box.
91
Dakota Ultrasonics
3) Use the UP, DOWN, LEFT, & RIGHT arrow keys to highlight the appropriate
alpha characters.
4) Press the ENTER key to select a character and advance to the next field of
the Grid or Seq Note.
5) Use the CLR key to backspace if necessary.
6) Repeat steps 3 through 5 until the Grid or Seq Note is completed.
7) Press the OK key to save the Grid or Seq Note and return to the Grid or
Seq List Items menu, or ESC to cancel entering the Grid or Seq Note.
Grid Log
Sequential Log
1) Use the UP and DOWN arrow keys to scroll through the new Grid or Seq
List Items until TOP LEFT or START ID is highlighted.
2) Press the ENTER key to activate the Coordinate or Start ID Edit Box.
93
Dakota Ultrasonics
Grid Log
Sequential Log
3) GRID LOG: Use the LEFT, & RIGHT arrow keys to scroll the Columns,
and the UP, DOWN arrow keys to scroll the Rows.
SEQ LOG: Use the UP, DOWN, LEFT, & RIGHT arrow keys to highlight the
appropriate alpha characters. Press the ENTER key to select a character
and advance to the next character field, in conjunction with using the CLR
key to backspace if necessary.
4) Press the OK key to select the coordinate or start ID and return to the Grid
or Seq List Items screen, or ESC to cancel the selection and return to the
Grid or Seq List Items menu.
94
Grid Log
Sequential Log
1) Use the UP and DOWN arrow keys to scroll through the new Grid or Seq
List Items until LOWER RIGHT or END ID is highlighted.
2) Press the ENTER key to activate the Coordinate or End ID Edit Box.
Grid Log
Sequential Log
3) GRID LOG: Use the LEFT, & RIGHT arrow keys to scroll the Columns,
and the UP, DOWN arrow keys to scroll the Rows.
SEQ LOG: Use the UP, DOWN, LEFT, & RIGHT arrow keys to highlight the
appropriate alpha characters. Press the ENTER key to select a character
95
Dakota Ultrasonics
and advance to the next character field, in conjunction with using the CLR
key to backspace if necessary.
4) Press the OK key to select the coordinate or end ID and return to the Grid
or Seq List Items screen, or ESC to cancel the selection and return to the
Grid or Seq List Items menu.
Note: If there is not enough memory available to create the grid or sequential
log, an error message box NOT ENOUGH MEMORY will be displayed. Press
the OK or ESC key to return to the Grid or Seq List Items menu. It may be
necessary to free some memory in the CMX DL at this time. Refer to page 103
for more information on Deleting a File.
96
Grid Log
Sequential Log
5) Use the UP and DOWN arrow keys to scroll through the new Grid or Seq
List Items until INCR. DIR or DIRECTION is highlighted.
6) Use the LEFT & RIGHT arrow keys to toggle the Increment direction
NONE, NORTH, EAST, SOUTH, or WEST for a grid, or INC, DEC for a
sequential log.
7) When the correct Increment direction is displayed, continue on to the next
section Graphics.
97
Dakota Ultrasonics
Saving Graphics
The CMX DL provides the user with the ability to save a snapshot of the display
screen and all the current settings of the CMX DL with every reading, or just save
the reading only. Saving the graphics might be advantageous to the user when
the B-Scan view will be used to graphically save a picture of the scanned areas
for reporting purposes. It might also come in handy when user requires backup
of all the current CMX DL parameters for each reading. However, if neither of
these requirements are necessary, this option should be disabled in order make
more efficient use to the CMX DL storage capacity.
Grid Log
Sequential Log
1) Use the UP and DOWN arrow keys to scroll through the new Grid or Seq
List Items until SAVE GRAPHICS is highlighted.
2) Use the LEFT & RIGHT arrow keys to toggle the Increment direction to
YES or NO.
3) Continue on to the next section Saving a Grid or Seq File.
98
Saving a Grid
Once all the parameters are set, the user has the option of saving or canceling
the new grid.
Grid Log
Sequential Log
4) Use the UP and DOWN arrow keys to scroll through the new Grid or Seq
List Items until CREATE GRID or LOG? is highlighted.
5) Press the ENTER key to accept the grid or seq log parameters and activate
the confirmation screen.
99
Dakota Ultrasonics
Grid Log
Sequential Log
6) Press the OK key to save the New Grid or Seq Log, or the ESC key to
cancel the New Grid or Seq Log setup and return to the DATA menu.
7) Press the MEAS key to return to the measurement screen and begin storing
readings.
Storing a Reading
Note: Once the gird or sequential log has been created it will automatically be
displayed following the create confirmation screen. It is located at the bottom of
the measurement screen, and can be left open for the duration of the inspection
process. This section assumes the grid or sequential log was just created.
Refer to the section on opening a grid or seq log if the file was created at
another time.
DL
key will initiate an advance to row number option. Use the Left, Right, UP, and
DOWN arrow keys to enter the row number. Once the row number is correctly
displayed, press the OK key to advance directly to that row number in the grid
log.
DL
key multiple times, will toggle throug h the display options: Digits, B-SCAN
views.
100
Grid Log
Sequential Log
1) Press the UP, DOWN, LEFT, and RIGHT arrow keys to scroll the target cell
cursor to the desired storage location.
2) Press the ENTER key to save the current reading in the highlighted cell
location. Its as simple as that!
3) Press the MEAS key to hide the GRID/SEQ View Box, or ENTER to display
it as necessary.
Note: Once the file is open, it will remain open until its closed or another file is
DL
will automatically
open the file when powered on. Simply press the ENTER key to display the file
from the measurement screen.
101
Dakota Ultrasonics
Grid Log
Sequential Log
1) Assuming the Grid/Seq View Box is currently displayed, press the UP,
DOWN, LEFT, and RIGHT arrow keys to scroll the stored readings and
corresponding display view. Notice as the cursor is moved to a different
cell, the display will be updated with the display view saved with the reading.
Readings stored in memory are indicated by displaying a MEM in the top left
corner of the measurement screen.
2) The user may opt to clear a specific reading and save a new one at any
time. Press the CLR key in the appropriate cell location to clear the reading,
take a new measurement, and press the ENTER key to save the new
reading.
3) Abort the Grid/Seq Log View Box by pressing the MEAS key at any time.
102
1) Press the MENU key once to activate the menu items tab. Press the MENU
key multiple times to tab right and the ESC key multiple times to tab left until
the DATA menu is highlighted and displaying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items
until DELETE ONE FILE is highlighted.
103
Dakota Ultrasonics
2) Use the UP and DOWN arrow keys to scroll through the sub menu items
until DELETE ALL DATA is highlighted.
104
Editing a Grid
105
Dakota Ultrasonics
1) Press the MENU key once to activate the menu items tab. Press the MENU
key multiple times to tab right and the ESC key multiple times to tab left until
the DATA menu is highlighted and displaying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items
until EDIT is highlighted.
3) Press the ENTER key to display the Edit List Box.
4) Use the UP and DOWN arrow keys to scroll through the Edit options until
NOTE or INCR. DIR is highlighted.
Note: If editing the INCR. DIR, simply use the LEFT or RIGHT arrow keys to
scroll NONE, NORTH, EAST, SOUTH, WEST for a Grid, or INC, DEC for a Seq
Log. Proceed to step 10.
5) Press the ENTER key to activate the Alpha Edit box Only used when
editing the NOTE.
6) Use the UP, DOWN, LEFT, & RIGHT arrow keys to highlight the appropriate
alpha characters.
7) Press the ENTER key to select a character and advance to the next field of
the Comments.
106
10) Press the UP or DOWN arrow key to highlight SAVE CHANGES, and the
OK key to activate the confirmation screen.
11) Press the OK key to save the changes or the ESC key to cancel editing the
file parameters.
12) Press the MEAS key to return to the measurement screen.
107
Dakota Ultrasonics
Grid Log
Sequential Log
1) Press the MENU key once to activate the menu items tab. Press the MENU
key multiple times to tab right and the ESC key multiple times to tab left until
the DATA menu is highlighted and displaying the submenu items.
108
2) Use the UP and DOWN arrow keys to scroll through the sub menu items
until OPEN is highlighted.
109
Dakota Ultrasonics
1) Press the MENU key once to activate the menu items tab. Press the MENU
key multiple times to tab right and the ESC key multiple times to tab left until
the DATA menu is highlighted and displaying the submenu items.
110
2) Use the UP and DOWN arrow keys to scroll through the sub menu items
until CLOSE is highlighted.
3) Press the ENTER key to close the active file. Note: Following the key
press, the CLOSE text will be grayed out indicating the file has been close
and is no longer active.
111
CHAPTER TWELVE
SETUPS CREATE, STORE, EDIT, & RECALL
12.1 Introduction to Setups
Often times, users are faced with a variety of tasks and applications that are
sometimes similar, but often times very different. With a standard thickness gauge,
the user would have to recalibrate for each individual application respectively. With
all the features of the CMX DL, the number of potential applications also increases
based on ability alone. This is primarily in reference to the addition of those very
difficult applications, where a standard thickness gauge would not be feasible, or
capable of accomplishing the tasks.
The increased number of features and parameters also adds to the overall setup
time, or amount of time to set up marginal applications with perfection and
understanding. Because of the additional time involved, the CMX DL has been
equipped with the ability to save these setups to memory and be recalled at any time.
The CMX DL can store up to 64 custom setups. These setups can be bi-directionally
transferred to and from a PC. Therefore, the user can save as many setups as
necessary for all their individual applications requirements. This saves a great deal
of time and knowledge for future inspections of the same job or project. This feature
also eliminates error between two or more users during the setup and calibration
process.
112
Opening a Setup
1) Press the MENU key once to activate the menu items tab. Press the MENU
key multiple times to tab right and the ESC key multiple times to tab left until
the SETUP menu is highlighted and displaying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items
until OPEN is highlighted.
113
Dakota Ultrasonics
4) Use the UP and DOWN arrow keys to scroll through the setups until the
target setup is highlighted.
114
Saving a Setup
1) Press the MENU key once to activate the menu items tab. Press the MENU
key multiple times to tab right and the ESC key multiple times to tab left until
the SETUP menu is highlighted and displaying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items
until SAVE is highlighted.
3) Press the ENTER key to display the Save Setup Parameters List Box.
4) Press the UP and DOWN arrow keys to scroll the Name and Note
parameters.
115
Dakota Ultrasonics
5) When the parameter to edit is highlighted, press the ENTER key to activate
the Alpha Edit Box.
6) Use the UP, DOWN, LEFT, and RIGHT arrow keys to scroll through the
characters, the ENTER key to select characters, and the CLR key to
backspace through the characters, until the Name or Note fields have been
edited.
7) Press the OK key to return to the Save Setup Parameters List Box.
8) If both parameters will be edited, repeat steps 4 7.
9) Use the UP and DOWN arrow keys to scroll to and highlight SAVE SETUP.
116
10) Press the ENTER key to activate the Setup List Box.
11) Use the UP and DOWN arrow keys to scroll through the setups until the
target location to save the Setup is highlighted.
117
Dakota Ultrasonics
Deleting a Setup
1) Press the MENU key once to activate the menu items tab. Press the MENU
key multiple times to tab right and the ESC key multiple times to tab left until
the SETUP menu is highlighted and displaying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items
until DELETE is highlighted.
118
119
Dakota Ultrasonics
1) Press the MENU key once to activate the menu items tab. Press the MENU
key multiple times to tab right and the ESC key multiple times to tab left until
the SETUP menu is highlighted and displaying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items
until DEFAULT SETUP is highlighted.
120
Selecting a Language
1) Press the MENU key once to activate the menu items tab. Press the MENU
key multiple times to tab right and the ESC key multiple times to tab left until
the SETUP menu is highlighted and displaying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items
until LANGUAGE is highlighted.
3) Press the LEFT and RIGHT arrow keys to toggle the language options.
4) Once the desired language is displayed, press the MEAS key to return to
the measurement screen.
121
CHAPTER THIRTEEN
USING THE UTILITY SOFTWARE
13.1 Computer System Requirements
DakView will run on many different operating systems: Windows 98 (1st or 2nd
edition), Windows NT 4.0 with Service Pack 5, Windows ME, Windows XP, Windows
2000 Professional, Windows 2000 Server, or Windows 2000 Advanced Server
operating systems running on Intel or AMD hardware.
A Pentium 166MHz or faster processor with at least 32 megabytes of physical RAM
is required. You should have 40 megabytes of free disk space before attempting to
install DakView.
DakView requires an available communications port in order to transfer data to and
from the CMX DL. DakView supports COM1, COM2, COM3, and COM4.
USB to Serial Converter
Some newer laptop computers do not have standard serial ports. In this case it is
possible to use a USB to Serial converter. If a serial to USB cable is needed,
Contact Dakota Ultrasonics (pt# N-402-0510).
122
1) Press the MENU key once to activate the menu items tab. Press the MENU
key multiple times to tab right, and the ESC key multiple times to tab left,
until the XFER menu is highlighted and displaying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items
until the desire option is highlighted.
3) Press the ENTER key to activate the option selected.
4) Once the Backup or Restore function has been completed, press the
MEAS key once to return to the menu items, or twice to return to the
measurement mode.
123
sound velocity
in/us
sound velocity
m/s
Aluminum
Beryllium
Brass
Bronze
Cadmium
Columbium
Copper
Glass (plate)
Glycerine
Gold
Inconel
Iron
Cast Iron
Lead
Magnesium
Mercury
Molybdenum
Monel
Nickel
Nylon
Platinum
Plexiglas
Polystyrene
PVC
Quartz glass
Rubber vulcanized
Silver
Steel (1020)
Steel (4340)
Steel Stainless"
Teflon
0.2510
0.5080
0.1730
0.1390
0.1090
0.1940
0.1830
0.2270
0.0760
0.1280
0.2290
0.2320
0.1800
0.0850
0.2300
0.0570
0.2460
0.2110
0.2220
0.1060
0.1560
0.1060
0.0920
0.0940
0.2260
0.0910
0.1420
0.2320
0.2330
0.2230
0.0540
6375
12903
4394
3531
2769
4928
4648
5766
1930
3251
5817
5893
4572
2159
5842
1448
6248
5359
5639
2692
3962
2692
2337
2388
5740
2311
3607
5893
5918
5664
1372
(approx)
(approx)
124
0.1310
0.2400
0.2040
0.1330
0.0580
0.1660
0.1830
3327
6096
5182
3378
1473
4216
4648
125
Name
3
4
Comment 1
126
Gn/AGC
Velocity
WARRANTY INFORMATION
Warranty Statement
Dakota Ultrasonics warrants the CMX DL against defects in materials and
workmanship for a period of two years from receipt by the end user. Additionally,
Dakota Ultrasonics warrants transducers and accessories against such defects for a
period of 90 days from receipt by the end user. If Dakota Ultrasonics receives notice
of such defects during the warranty period, Dakota Ultrasonics will either, at its
option, repair or replace products that prove to be defective.
Should Dakota Ultrasonics be unable to repair or replace the product within a
reasonable amount of time, the customer's alternative exclusive remedy shall be
refund of the purchase price upon return of the product.
Exclusions
The above warranty shall not apply to defects resulting from: improper or
inadequate maintenance by the customer; unauthorized modification or misuse; or
operation outside the environmental specifications for the product.
Dakota Ultrasonics makes no other warranty, either express or implied, with
respect to this product. Dakota Ultrasonics specifically disclaims any implied
warranties of merchantability or fitness for a particular purpose. Some states or
provinces do not allow limitations on the duration of an implied warranty, so the
above limitation or exclusion may not apply to you. However, any implied warranty of
merchantability or fitness is limited to the five-year duration of this written warranty.
This warranty gives you specific legal rights, and you may also have other rights
which may vary from state to state or province to province.
Obtaining Service During Warranty Period
If your hardware should fail during the warranty period, contact Dakota
Ultrasonics and arrange for servicing of the product. Retain proof of purchase in
order to obtain warranty service.
For products that require servicing, Dakota Ultrasonics may use one of the
following methods:
- Repair the product
- Replace the product with a re-manufactured unit
- Replace the product with a product of equal or greater performance
- Refund the purchase price.
After the Warranty Period
If your hardware should fail after the wa rranty period, contact Dakota Ultrasonics
for details of the services available, and to arrange for non-warranty service.
127
Routes of Entry:
Skin: not likely
Ingestion: not normally
Eyes: not normally
Inhalation: no
Effects of Overexposure: Acute: May cause temporary
eye irritation
Chronic: none expected
SONOTECH, INC.
Toll Free: 1-800-458-4254