Axon Guide 3rdedition
Axon Guide 3rdedition
Copyright
Third Edition © Copyright 2008, MDS Analytical Technologies. All rights reserved. No part of this
publication may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated
into any language or computer language, in any form or by any means, electronic, mechanical,
magnetic, optical, chemical, manual, or otherwise, without the prior written permission of MDS
Analytical Technologies, 1311 Orleans Drive, Sunnyvale, California, 94089, United States of
America.
Trademarks
CYBERAMP, DIGIDATA, GENECLAMP and PCLAMP are registered trademarks and
AXOPATCH, AXOCLAMP and MULTICLAMP are trademarks of MDS Analytical
Technologies.
These trademarks are not to be used in any type of promotion or advertising without written
permission from MDS Analytical Technologies.
All other trademarks or registered trademarks are the property of their respective owners.
Disclaimer
MDS Analytical Technologies reserves the right to change its products and services at any time to
incorporate technological developments. This publication is subject to change without notice.
Although this manual has been prepared with every precaution to ensure accuracy, MDS Analytical
Technologies assumes no liability for any errors or omissions, nor for any damages resulting from
the application or use of this information.
Questions?
Phone: +1-800-635-5577
+1-408-747-1700
Fax: +1-408-747-3603
Web: www.moleculardevices.com/support
For additional offices, please see the contact information on the back cover of this publication.
Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xv
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xvii
Editorial . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .xix
Contributors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .xxi
1. Bioelectricity
Electrical Potentials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Electrical Currents. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Resistors and Conductors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Ohm’s Law . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
The Voltage Divider . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Perfect and Real Electrical Instruments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Ions in Solutions and Electrodes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Capacitors and their Electrical Fields . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Currents Through Capacitors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Current Clamp and Voltage Clamp . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Glass Microelectrodes and Tight Seals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Further Reading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
7. Transducers
Temperature Transducers for Physiological Temperature Measurement . . . . . . . . . 167
Thermistors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 167
IC Temperature Transducers that Produce an Output Current Proportional
to Absolute Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 168
IC Temperature Transducers that Produce an Output Voltage Proportional
to Absolute Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 169
Temperature Transducers for Extended Temperature Ranges . . . . . . . . . . . . . . . . . . 169
Thermocouples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 169
Resistance Temperature Detectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 170
Electrode Resistance and Cabling Affect Noise and Bandwidth . . . . . . . . . . . . . . . 171
High Electrode Impedance Can Produce Signal Attenuation . . . . . . . . . . . . . . . . . 171
Unmatched Electrode Impedances Increase Background Noise and Crosstalk . . . . . 172
High Electrode Impedance Contributes to the Thermal Noise of the System . . . . . 173
Cable Capacitance Filters Out the High-Frequency Component of the Signal. . . . . 174
EMG, EEG, EKG and Neural Recording . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 174
EMG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 174
EKG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 175
EEG. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 175
Nerve Cuffs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 175
9. Acquisition Hardware
Fundamentals of Data Conversion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 189
Quantization Error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 190
Choosing the Sampling Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 191
Converter Buzzwords . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 192
Gain Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 192
Linearity Error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 192
Differential Nonlinearity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 192
Least Significant Bit (LSB) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193
Monotonicity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193
Deglitched DAC Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193
Timers. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 194
Digital I/O . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 194
Index. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 253
The Axon Guide third edition is a tool benefitting both the novice and the expert electro-
physiologist. Newcomers to electrophysiology will gain an appreciation of the intricacies
of electrophysiological measurements and the requirements for setting up a complete
recording and analysis system. For experienced electrophysiologists, we include in-depth
discussions of selected topics, such as advanced methods in electrophysiology and noise.
This edition is the first major revision of the Axon Guide since the original edition was
published in 1993. While the fundamentals of electrophysiology have not changed in that
time, changes in instrumentation and computer technology made a number of the origi-
nal chapters interesting historical documents rather than helpful guides. This third edi-
tion is up-to-date with current developments in technology and instrumentation.
This guide was the product of a collaborative effort of many researchers in the field of
electrophysiology and of MDS Analytical Technologies’ staff. We are deeply grateful to
these individuals for sharing their knowledge and devoting significant time and effort to
this endeavor.
David Yamane
Director, Electrophysiology Marketing
MDS Analytical Technologies
November 2007
Editorial Committee
Michael J. Delay, Ph.D.
Artwork
Elizabeth Brzeski
Warburton H. Maertz
1.Lester is at the California Institute of Technology, Pasadena, CA. The other editorial staff are from MDS Analytical Technologies.
Eveline Pajares
Peter Valenzuela
Acknowledgements
The valuable inputs and insightful comments of Dr. Bertil Hille, Dr. Joe Immel and
Dr. Stephen Redman are much appreciated.
David Gallegos.
Two handy rules about currents often help to understand electrophysiological phenom-
ena: 1) current is conserved at a branch point (Figure 1.1); and 2) current always flows in
a complete circuit (Figure 1.2). In electrophysiological measurements, currents can flow
through capacitors, resistors, ion channels, amplifiers, electrodes and other entities, but
they always flow in complete circuits.
I total
I1 I2
I total = I 1 + I 2
I
I
Microelectrode
Battery
I = I 1+ I 2
Cell
I1 I2
Capacitor
I
ELECTRONIC
INSTRUMENT
I
Example of an electrical circuit with various parts. Current always flows in a complete
circuit.
G G Gtotal = 2G
γ γ
G total = 2 γ
ion lipid
channel bilayer
A more accurate representation of an ion channel is a conductor in series with two addi-
tional circuit elements (Figure 1.4):
1 A switch that represents the gate of the channel, which would be in its conducting
position when the gate is open, and
2 A battery that represents the reversal potential of the ionic current for that channel.
The reversal potential is defined operationally as the voltage at which the current
changes its direction. For a perfectly selective channel (i.e., a channel through which
only a single type of ion can pass), the reversal potential equals the Nernst potential
for the permeant ion. The Nernst potential for ion A, Ea, can be calculated by the
Nernst equation:
where R is the gas constant (8.314 V C K-1 mol-1), T is the absolute temperature
(T = 273° + °C), zA is the charge of ion A, F is Faraday’s constant (9.648x104 C mol-1),
and [A]o and [A]i are the concentrations of ion A outside the cell and inside the cell,
respectively. At 20 °C (“room temperature”), 2.303(RT/zAF) log10{[A]o/[A]i} =
58 mV log10{[A]o/[A]i}for a univalent ion.
E reversal
For instance, at room temperature, a Na+ channel facing intracellular Na+ concentration
that is ten-fold lower than the extracellular concentration of this ion would be represented
by a battery of +58 mV. A K+ channel, for which the concentration gradient is usually
reversed, would be represented by a battery of -58 mV.
Reversal potentials are not easily predicted for channels that are permeable to more than
one ion. Nonspecific cation channels, such as nicotinic acetylcholine receptors, usually
have reversal potentials near zero millivolts. Furthermore, many open channels have a
nonlinear relation between current and voltage. Consequently, representing channels as
resistors is only an approximation. Considerable biophysical research has been devoted to
understanding the current-voltage relations of ion channels and how they are affected by
the properties and concentrations of permeant ions.
The transmembrane potential is defined as the potential at the inner side of the mem-
brane relative to the potential at the outer side of the membrane. The resting membrane
potential (Erp) describes a steady-state condition with no net flow of electrical current
across the membrane. The resting membrane potential is determined by the intracellular
and extracellular concentrations of ions to which the membrane is permeable and on their
permeabilities. If one ionic conductance is dominant, the resting potential is near the
Nernst potential for that ion. Since a typical cell membrane at rest has a much higher per-
meability to potassium (PK) than to sodium, calcium or chloride (PNa, PCa and PCl,
respectively), the resting membrane potential is very close to EK, the potassium reversal
potential.
G = I/R Δ V=IR
I
V
R1 R1
V1 = E
E R 1+ R 2
R2 R2
V2 = E
R 1+ R 2
The total potential difference provided by the battery is E; a portion of this voltage
appears across each resistor:
R1 R2
ΔV1 = E ; ΔV 2 = E (2a)
R1 + R 2 R1 + R 2
When two resistors are connected in series, the same current passes through each of them.
Therefore the circuit is described by:
ΔV1 + ΔV 2 = E (2b)
where E is the value of the battery, which equals the total potential difference across both
resistors. As a result, the potential difference is divided in proportion to the two resistance
values.
The best way to measure an electrical potential difference is to use a voltmeter with infi-
nite resistance. To illustrate this point, consider the arrangement of Figure 1.8 A, which
can be reduced to the equivalent circuit of Figure 1.8 B.
MEASURING
CIRCUIT
Rm Em
Membrane Resistance
Bath Electrode
B EQUIVALENT CIRCUIT
Re
Resting
Potential
Em Rm Rin
V
R in
V = ER + R
in e
Instruments used to measure potentials must have a very high input resistance Rin.
Before making the measurement, the cell has a resting potential of Erp, which is to be
measured with an intracellular electrode of resistance Re. To understand the effect of the
measuring circuit on the measured parameter, we will pretend that our instrument is a
“perfect” voltmeter (i.e., with an infinite resistance) in parallel with a finite resistance Rin,
which represents the resistance of a real voltmeter or the measuring circuit. The combina-
tion Re and Rin forms a voltage divider, so that only a fraction of Erp appears at the input
of the “perfect” voltmeter; this fraction equals ErpRin/(Rin + Re). The larger Rin, the closer
V is to Erp. Clearly the problem gets more serious as the electrode resistance Re increases,
but the best solution is to make Rin as large as possible.
On the other hand, the best way to measure current is to open the path and insert an
ammeter. If the ammeter has zero resistance, it will not perturb the circuit since there is no
IR-drop across it.
1 The current is carried by at least two types of ions (one anion and one cation) and
often by many more. For each ion, current flow in the bulk solution is proportional to
the potential difference. For a first approximation, the conductance of the whole
solution is simply the sum of the conductances contributed by each ionic species.
When the current flows through ion channels, it is carried selectively by only a subset
of the ions in the solution.
2 At the electrodes, current must be transformed smoothly from a flow of electrons in
the copper wire to a flow of ions in solution. Many sources of errors (artifacts) are
possible. Several types of electrodes are used in electrophysiological measurements; the
most common is a silver/silver chloride (Ag/AgCl) interface, which is a silver wire
coated with silver chloride (Figure 1.9). If electrons flow from the copper wire
through the silver wire to the electrode AgCl pellet, they convert the AgCl to Ag
atoms and the Cl- ions become hydrated and enter the solution. If electrons flow in
the reverse direction, Ag atoms in the silver wire that is coated with AgCl give up their
electrons (one electron per atom) and combine with Cl- ions that are in the solution
to make insoluble AgCl. This is, therefore, a reversible electrode, i.e., current can flow
in both directions. There are several points to remember about Ag/AgCl electrodes:
a The Ag/AgCl electrode performs well only in solutions containing chloride ions.
b Because current must flow in a complete circuit, two electrodes are needed. If the
two electrodes face different Cl- concentrations (for instance, 3 M KCl inside a
micropipette1 and 120 mM NaCl in a bathing solution surrounding the cell), there
will be a difference in the half-cell potentials (the potential difference between the
solution and the electrode) at the two electrodes, resulting in a large steady
potential difference in the two wires attached to the electrodes. This steady
potential difference, termed liquid junction potential, can be subtracted
electronically and poses few problems as long as the electrode is used within its
reversible limits.
c If the AgCl is exhausted by the current flow, bare silver could come in contact with
the solution. Silver ions leaking from the wire can poison many proteins. Also, the
1. A micropipette is a pulled capillary glass into which the Ag/AgC1 electrode is inserted (see Chapter 4).
_
Cl AgCl Coating
Ag
Electrode reaction:
Ag + Cl- AgCl + electron (e- )
This reaction can also be presented by:
Ag + Cl-
+
AgCl
+e- -e-
Ag
Another type of electrode, made of platinum (Pt) (Figure 1.10), is irreversible but not
exhaustible. At its surface, Pt catalyzes the electrolysis of water. The gaseous H2 or O2
produced, depending on the direction of current flow, leaves the surface of the electrode.
If both electrodes are Pt electrodes, the hydroxyl ions and protons are produced in equal
numbers; however, local pH changes can still occur.
Pt
H 2 or O2 bubbles
1
e - + H 2O OH - + 2 H 2 or H 2O
+ 1
2H + 2 O2 + e-
The concept of the electrical field is important for understanding membrane function.
Biological membranes are typically less than 10 nm thick. Consequently, a transmem-
brane resting potential of about 100 mV produces a very sizable electrical field in the
membrane of about 105 V/cm. This is close to the value at which most insulators break
down irreversibly because their atoms become ionized. Of course, typical electrophysio-
logical equipment cannot measure these fields directly. However, changes in these fields
are presumably sensed by the gating domains of voltage-sensitive ion channels, which
determine the opening and closing of channels, and so the electrical fields underlie the
electrical excitability of membranes.
Q = CΔV (4)
The formal symbol for a capacitor is two parallel lines (Figure 1.12). This symbol arose
because the most effective capacitors are parallel conducting plates of large area separated
by a thin sheet of insulator (Figure 1.11) an excellent approximation of the lipid bilayer.
The capacitance C is proportional to the area and inversely proportional to the distance
separating the two conducting sheets.
+ + + +
C ΔV
- - - -
Figure 1.11: Capacitance. A charge Q is stored in a capacitor of value C held at a potential ΔV.
C C Ctotal = 2C
C C 2C
ΔV
I =C (5)
Δt
Until now, we have been discussing circuits whose properties do not change with time. As
long as the voltage across a membrane remains constant, one can ignore the effect of the
membrane capacitance on the currents flowing across the membrane through ion chan-
nels. While the voltage changes, there are transient capacitive currents in addition to the
steady-state currents through conductive channels. These capacitive currents constitute
one of the two major influences on the time-dependent electrical properties of cells (the
other is the kinetics of channel gating). On Axon Cellular Neuroscience voltage- or patch-
clamp amplifiers, several controls are devoted to handle these capacitive currents. There-
fore it is worth obtaining some intuitive “feel” for their behavior.
The stored charge on the membrane capacitance accompanies the resting potential, and
any change in the voltage across the membrane is accompanied by a change in this stored
charge. Indeed, if a current is applied to the membrane, either by channels elsewhere in
the cell or by current from the electrode, this current first satisfies the requirement for
charging the membrane capacitance, then it changes the membrane voltage. Formally, this
can be shown by representing the membrane as a resistor of value R in parallel with capac-
itance C (Figure 1.13).
C R = I/G
Capacitance Conductance
Now, if we apply a pulse of current to the circuit, the current first charges up the capaci-
tance, then changes the voltage (Figure 1.14).
I
time
The voltage V(t) approaches steady state along an exponential time course:
V (t ) = Vin f (1 − e −t / τ )
The steady-state value Vinƒ (also called the infinite-time or equilibrium value) does not
depend on the capacitance; it is simply determined by the current I and the membrane
resistance R:
V inf = IR (6)
This is just Ohm’s law, of course; but when the membrane capacitance is in the circuit, the
voltage is not reached immediately. Instead, it is approached with the time constant τ,
given by:
τ = RC (7)
Thus, the charging time constant increases when either the membrane capacitance or the
resistance increases. Consequently, large cells, such as Xenopus oocytes that are frequently
used for expression of genes encoding ion-channel proteins, and cells with extensive mem-
brane invigorations, such as the T-system in skeletal muscle, have a long charging phase.
In a voltage clamp experiment one controls the membrane voltage and measures the trans-
membrane current required to maintain that voltage. Despite the fact that voltage clamp
does not mimic a process found in nature, there are three reasons to do such an experi-
ment:
1 Clamping the voltage eliminates the capacitive current, except for a brief time
following a step to a new voltage (Figure 1.15). The brevity of the capacitive current
depends on many factors that are discussed in following chapters.
2 Except for the brief charging time, the currents that flow are proportional only to the
membrane conductance, i.e., to the number of open channels.
3 If channel gating is determined by the transmembrane voltage alone (and is insensitive
to other parameters such as the current and the history of the voltage), voltage clamp
offers control over the key variable that determines the opening and closing of ion
channels.
V1
capacitive transient
Idt = CV1
I = V1 /R
The patch clamp is a special voltage clamp that allows one to resolve currents flowing
through single ion channels. It also simplified the measurement of currents flowing
through the whole-cell membrane, particularly in small cells that cannot be easily pene-
trated with electrodes. The characteristics of a patch clamp are dictated by two facts:
1 The currents measured are very small, on the order of picoamperes in single-channel
recording and usually up to several nanoamperes in whole-cell recording. Due to the
small currents, particularly in single-channel recording, the electrode polarizations
and nonlinearities are negligible and the Ag/AgCl electrode can record voltage
accurately even while passing current.
2 The electronic ammeter must be carefully designed to avoid adding appreciable noise
to the currents it measures.
Requirement 2) above (see Section 1.6., “Perfect and Real Electrical Instruments”) states
that the quality of the measurement depends on minimizing perturbation of the cells. For
the case of voltage recording, this point can be presented with the voltage divider circuit
(Figure 1.16).
seal resistance R S
EQUIVALENT CIRCUIT
V intracellular
RS
V = EK
EK RS + RK
RS V
RK
extracellular
This intracellular electrode is measuring the resting potential of a cell whose membrane
contains only open K+ channels. As the seal resistance Rs increases, the measurement
approaches the value of EK.
For the case of patch recording, currents through the seal do not distort the measured
voltage or current, but they do add to the current noise. Current noise can be analyzed
either in terms of the Johnson noise of a conductor, which is the thermal noise that
increases with the conductance (see Chapter 7 and Chapter 11), or in terms of simple sta-
tistics. The latter goes as follows: If a current of N ions/ms passes through an open chan-
nel, then the current will fluctuate from one millisecond to the next with a standard
deviation of √N. These fluctuations produce noise on the single-channel recorded traces.
If an additional current is flowing in parallel through the seal (Figure 1.17), it causes an
increase in the standard deviations. For instance, if the current through the seal is ten-fold
larger than through the channel, then the statistical fluctuations in current flow produced
by the seal are √10 (316%) larger than they would be for a “perfect” seal.
AMPLIFIER
Iseal
Ichannel
I=0
good seal
bad seal
single-channel event
In a patch recording, currents through the seal also flow through the measuring cir-
cuit, increasing the noise on the measured current.
To illustrate the practical implications of these requirements, two kinds of “typical” setups
are briefly described, one for in vitro extracellular recording, the other for single-channel
patch clamping.
done in an unperfused culture dish at room temperature. On the other hand, the optical
and mechanical requirements are dictated by the need to accurately place a patch elec-
trode on a particular 10 or 20 μm cell. The microscope should magnify up to 300 or 400
fold and be equipped with some kind of contrast enhancement (Nomarski, Phase or Hoff-
man). Nomarski (or Differential Interference Contrast) is best for critical placement of
the electrode because it gives a very crisp image with a narrow depth of field. Phase con-
trast is acceptable for less critical applications and provides better contrast for fine pro-
cesses. Hoffman presently ranks as a less expensive, slightly degraded version of Nomarski.
Regardless of the contrast method selected, an inverted microscope is preferable for two
reasons: 1) it usually allows easier top access for the electrode since the objective lens is
underneath the chamber, and 2) it usually provides a larger, more solid platform upon
which to bolt the micromanipulator. If a top-focusing microscope is the only option, one
should ensure that the focus mechanism moves the objective, not the stage.
Finally, the electronic requirements for single-channel recording are more complex than
for extracellular recording. However, excellent patch clamp amplifiers, such as those of the
Axopatch™ amplifier series from MDS Analytical Technologies, are commercially avail-
able.
A recent extension of patch clamping, the patched slice technique, requires a setup that
borrows features from both in vitro extracellular and conventional patch clamping config-
urations. For example, this technique may require a chamber that continuously perfuses
and oxygenates the slice. In most other respects, the setup is similar to the conventional
patch clamping setup, except that the optical requirements depend upon whether one is
using the thick-slice or thin-slice approach (see Further Reading at the end of this).
Whereas a simple dissecting microscope suffices for the thick-slice method, the thin-slice
approach requires a microscope that provides 300- to 400-fold magnification, preferably
top-focusing with contrast enhancement.
tor should be attached close to the chamber; preferably bolted directly to the microscope
stage. The headstage of the recording amplifier should, in turn, be bolted directly to the
manipulator (not suspended on a rod), and the electrode should be short.
For most fine work, such as patch clamping, it is preferable to use remote-controlled
micromanipulators to eliminate hand vibration (although a fine mechanical manipulator,
coupled with a steady hand, may sometimes be adequate). Currently, there are three main
types of remote-controlled micromanipulators available: motorized, hydraulic/pneumatic
and piezoelectric. Motorized manipulators tend to be solid and compact and have excel-
lent long-term stability. However, they are often slow and clumsy to move into position,
and may exhibit backlash when changing direction. Hydraulic drives are fast, convenient
and generally backlash-free, but some models may exhibit slow drift when used in certain
configurations. Piezoelectric manipulators have properties similar to motorized drives,
except for their stepwise advancement.
Anti-vibration tables usually comprise a heavy slab on pneumatic supports. Tables of vary-
ing cost and complexity are commercially available. However, a homemade table, consist-
ing of a slab resting on partially-inflated inner tubes, may be adequate, especially if high-
quality micromanipulators are used.
fluid. Never directly ground the solution other than at the ground wire in the chamber,
which is the reference ground for the amplifier and which may not be the same as the sig-
nal ground used for shielding purposes. Further suggestions are given in Section 6.6.,
“Line-Frequency Pick-Up (Hum)”.
The logical approach to reducing noise in the setup is to start with all equipment switched
off and disconnected; only an oscilloscope should be connected to the microelectrode
amplifier. First, measure the noise when the headstage is wrapped in grounded metal foil;
microelectrode headstages should be grounded through a low resistance (for instance,
1 MΩ), whereas patch-clamp headstages should be left open circuit. This provides a refer-
ence value for the minimum attainable radiative noise. Next, connect additional pieces of
electronic apparatuses while watching for the appearance of ground loops. Last, install an
electrode and add shielding to minimize radiative pickup. Finally, it should be admitted
that one always begins noise reduction in a mood of optimistic rationalism, but invariably
descends into frustrating empiricism.
Choice and placement of electronics is again a matter of personal preference. There are
minimalists who make do with just an amplifier and a computer, and who look forward to
the day when even those two will coalesce. Others insist on a loaded instrument rack. An
oscilloscope is important, because the computer is often insufficiently flexible. Further-
more, an oscilloscope often reveals unexpected subtleties in the signal that were not appar-
ent on the computer screen because the sample interval happened not to have been set
exactly right. The oscilloscope should be at eye level. Directly above or below should be
the microelectrode amplifier so that adjustments are easily made and monitored. Last, the
computer should be placed as far as possible—but still within a long arm’s reach—from
the microscope. This is necessary both to reduce the radiative noise from the monitor and
to ensure that one’s elbows will not bump the microscope when hurriedly typing at the
keyboard while recording from the best cell all week.
A final, general piece of advice is perhaps the most difficult to heed: resist the temptation
to mess eternally with getting the setup just right. As soon as it is halfway possible, do an
experiment. Not only will this provide personal satisfaction, it may also highlight specific
problems with the setup that need to be corrected or, better, indicate that an anticipated
problem is not so pressing after all.
Micromanipulators
hydraulic Narishige International USA
motorized Newport
piezoelectric EXFO (Burleigh)
Sutter Instrument
Oscilloscopes Tektronix
Pipette fabrication
Patch-Slice Setup
Item Suggested Manufacturers
Vibratome Vibratome
(other requirements as for a traditional patch-clamp setup)
Extra/Intracellular Microelectrode Setup
Item Suggested Manufacturers
Oscilloscopes Tektronix
Electrode fabrication
glass Garner Glass
Friedrich & Dimmock
Sutter Instrument
pullers David Kopf
Sutter Instrument
GlasswoRx
Smith, T. G., Jr., Lecar, H., Redman, S. J., Gage, P. W. Ed. Voltage and Patch Clamping
with Microelectrodes. American Physiological Society, Bethesda, MD, 1985.
Standen, N. B., Gray, P. T. A., Whitaker, M. J. Ed. Microelectrode Techniques. The Com-
pany of Biologists Limited, Cambridge, UK, 1987.
General patch-clamp recording
Hamill, O. P., Marty, A., Neher, E., Sakmann, B., Sigworth, F. J. Improved patch-clamp
techniques for high-resolution current from cells and cell-free membrane patches. Pflügers
Arch. 391: 85–100, 1981.
Sakmann, B. and Neher, E. Ed. Single-Channel Recording. Plenum Press, New York, NY,
1983.
Blanton, M. G., Lo Turco, J. J., Kriegstein, A. Whole cell recording from neurons in slices
of reptilian and mammalian cerebral cortex. J. Neurosci. Meth. 30: 203–210, 1989.
Vibration isolation methods
Newport Catalog. Newport Corporation, 2006.
Electrical isolation methods
Horowitz, P., Hill, W. The Art of Electronics. Cambridge, 1988.
The most important design criterion for extracellular amplifiers is low instrumentation
noise. Noise of less than 10 μV peak-to-peak (μVp-p) is desirable in the 10 kHz band-
width. At the same time, the input bias current3 of the amplifier should be low (< 1 nA)
1. In this chapter “Pipette” has been used for patch-clamp electrodes. “Micropipette” has been used for intracellular electrodes, except
where it is unconventional, such as single-electrode voltage clamp. “Electrode” has been used for bath electrodes. “Microelectrode” has
been used for extracellular electrodes.
2. Some level of capacitance exists between all conductive elements in a circuit. Where this capacitance is unintended in the circuit design,
it is referred to as stray capacitance. A typical example is the capacitance between the micropipette and its connecting wire to the
headstage enclosure and other proximal metal objects, such as the microscope objective. Stray capacitances are often of the order of a
few picofarads, but much larger or smaller values are possible. When the stray capacitances couple into high impedance points of the
circuit, such as the micropipette input, they can severely affect the circuit operation.
3. In the ideal operational amplifier (op amp), no current flows into the inputs. Similarly, in the ideal transistor, no current flows into
the gate. In practice, however, amplifying devices always have an input current. This current is commonly known as the input bias
current.
so that electrodes do not polarize. Ideally, the amplifier will have built-in high-pass and
low-pass filters so that the experimenter can focus on the useful signal bandwidth.
While not specifically targeted at extracellular recording, the Axoclamp™ 900A and the
MultiClamp™ 700B amplifiers are particularly suitable for single-cell extracellular record-
ing if the extracellular electrode is a microelectrode of several megohms or more. The
input leakage current of these amplifiers is very low and headstages are designed to
directly accommodate the micropipette holder. If necessary, capacitance compensation
can be used to speed up the response. The x100 AC-coupled output signal of the
MultiClamp 700B amplifier, in particular, is useful for measuring small extracellular sig-
nals, because up to 2000x further amplification can be applied to the signal with the built-
in output gain. Thus the total amplification is 200,000x, allowing for easier measurement
of extracellular potentials of less than 1 mV.
The Axon CyberAmp® 380 programmable signal conditioner amplifier has special ultra
low-noise probes suitable for extracellular recording. These probes do not have the special
features of the Axoclamp 900A and the MultiClamp 700B amplifiers, but for low-resis-
tance electrodes, from tens of ohms up to a few hundred kilohms, these ultra low-noise
probes have superior noise characteristics. The AI 402 x50 probe for the CyberAmp 380
amplifier contributes less noise than the thermal noise of a 250 Ω resistor. Electrodes can
be connected directly to the CyberAmp 380 amplifier without using a separate low-noise
probe. In this case, the additional noise due to the amplifier will still be very low—less
than the thermal noise of a 5 kΩ resistor. If electrodes are directly connected to the main
instrument, there is always a risk of picking up line-frequency noise or noise from other
equipment. Using a probe located very close to the electrode greatly reduces the probabil-
ity that it will pick up extraneous noise.
+1 A1
+1 A1
-
If a high-quality current injection circuit (current source) is connected to the input node,
all of the injected current flows down the micropipette and into the cell (see Figure 3.2).
The current source can be used to inject a pulse of current to stimulate the cell, a DC cur-
rent to depolarize or hyperpolarize the cell, or any variable waveform that the user intro-
duces into the control input of the current source.
Vp + Vcmd +1 Vcmd
A2
+1
+
Ro Vcmd
I -
+1 A1 Vp
Vp
I Vm
A high-quality current source can be made by adding a second amplifier to the buffer
amplifier circuit. The inputs to A2 are a command voltage, Vcmd, and the pipette voltage
(Vp) buffered by A1. The voltage across the output resistor, Ro, is equal to Vcmd regardless
of Vp. Thus the current through Ro is given exactly by I = Vcmd/Ro. If stray capacitances
are ignored, all of this current flows through the pipette into the cell, then out through the
cell membrane into the bath grounding electrode.
A
Vp +1
Vm
I ∝
( Vp = Vm+ IR p )
B
Vp
Vm
This technique is used to separate the membrane potential (Vm) from the total potential
(Vp) recorded by the micropipette. The technique is schematically represented in A. A dif-
ferential amplifier is used to subtract a scaled fraction of the current (I) from Vp. The scal-
ing factor is the micropipette resistance (Rp). The traces in B illustrate the operation of
the bridge circuit. When the current is stepped to a new value, there is a rapid voltage step
on Vp due to the ohmic voltage drop across the micropipette. Since the micropipette is
intracellular, changes in Vm are included in Vp. Thus the Vp trace shows an exponential
rise to a new potential followed by some membrane potential activity. The bridge ampli-
fier removes the instantaneous voltage step, leaving the Vm trace shown.
There are several ways to set the bridge balance. A commonly used technique is to apply
brief repetitive pulses of current to the micropipette while it is immersed in the prepara-
tion bath. The Bridge Balance control is advanced until the steady-state pulse response is
eliminated (Figure 3.4). At this point, the circuit is balanced and the micropipette resis-
tance can be read from the calibrated readout of the Bridge Balance control. The same
technique can even be used after the micropipette has penetrated the cell. Details of how
to achieve this are described in the Axoclamp 900A and MultiClamp 700B manuals.
50 mV
1 ms
All traces show the Vm output from the bridge-balance circuit. A 5 nA, 2 ms pulse is
applied to a 50 MΩ electrode. No bridge balance is used in the left trace, revealing the full
voltage drop across the micropipette. In the middle trace, optimum bridge balance is used
and the voltage drop across the micropipette is eliminated from the record. The transients
at the onset and finish of the step result from the finite bandwidth of the headstage ampli-
fier. In the right trace, the bridge-balance control is advanced too far; the voltage drop
across the micropipette is overcompensated and a reverse-polarity step appears.
3.2.4. TRACK
A special form of automatic offset compensation is available in many patch clamp ampli-
fiers, including the Axopatch™-1 and the Axopatch 200 amplifier series. In Axopatch
amplifiers this technique is called “Track.” In some instruments of other manufacturers it
is called “Search.” The Track circuit is used only during seal formation. During seal for-
mation the tip comes in and out of contact with the membrane and rapid shifts in the tip
potential of the headstage sometimes occur. These potential shifts lead to equivalent rapid
shifts in the current record, which may exceed the dynamic range of the headstage. Since
the headstage itself saturates, AC coupling at the oscilloscope or any other form of exter-
nal DC offset removal will not be useful. Instead, the Track circuit dynamically adjusts the
offset potential of the micropipette itself, so that the current is forced to be zero on aver-
age. It is important that the Track circuit be switched off before recording of ionic cur-
rents commences since the Track circuit introduces the same type of distortion produced
by AC coupling.
+1 Vcmd
A2
+1
Ro +1
A3 VI
+1
+1 A1 Vp
C in
Cw
Im
The circuit in Figure 3.2 is extended by the addition of a differential amplifier (A3) to
measure the voltage drop across the current setting resistor, Ro. This voltage drop is pro-
portional to the total current (I) flowing through Ro. For high-frequency currents
(e.g., the transient current during a step change in the command potential), the current
through Ro is not quite identical to the membrane current (Im). There is an additional
current that flows through the capacitance (Cw) of the wall of the micropipette and
another current that flows into the input capacitance (Cin) of the amplifier. Nevertheless,
if precautions are taken to minimize Cw (e.g., shallow bath immersion, Sylgard or mineral
oil coating) and Cin (e.g., bootstrapped buffer amplifier, capacitance neutralization), the
error current in these stray capacitances can be minimized and the output signal, VI, is a
good representation of the membrane current.
Another way to measure the true micropipette current is to use a separate circuit called a
“virtual ground” (Figure 3.6). Instead of connecting the bath ground electrode directly to
ground, it is connected to the virtual ground formed at the negative input of an inverting
operational amplifier. The negative input in this configuration is called a “virtual” ground
because the operational amplifier endeavors to keep the negative input equal to the
ground potential at the positive input. To achieve this goal, the operational amplifier out-
put must continuously pass a current through the feedback resistor (into the node at the
negative input) that is equal and opposite to the bath ground current. Thus the voltage at
the output of the operational amplifier is directly proportional to the bath current.
+1
Vcmd
A2
+1
+1 A1 Vp
I in
C in
Cw
Iw
Rf If
Im
-
A3 Vp
+
The bath grounding electrode is connected to the negative input of operational amplifier
A3. The output (VI) continually adjusts to pass current through the feedback resistor (Rf )
equal and opposite to the total current flowing into the junction at the negative input of
A3. The potential at the negative input is “virtually” equal to ground at all times. Most of
the current measured by A3 is the membrane current, Im. However, during rapid potential
changes, transient current flows through the capacitance (Cw) of the wall of the micropi-
pette and the input capacitance (Cin) of the amplifier. The current (Iw) through Cw flows
into the bath and is recorded by the virtual-ground circuit. The current (Iin) through Cin
flows into the system ground and is not recorded by the virtual-ground circuit.
1 The input lead to the virtual ground is very sensitive and easily picks up line-
frequency and other interference.
2 It is extremely difficult to use two microelectrode amplifiers in the same preparation
bath if one or both of them uses a virtual-ground current-measurement circuit. If one
amplifier uses a virtual ground, the individual micropipette currents will be difficult
to identify because the virtual-ground current monitor measures the summed current
from all sources. If two amplifiers use a virtual ground, large DC-error currents will
flow between the virtual-ground circuits because of the small differences in offset
voltages between the two circuits. This is equivalent to the problem that would exist if
two voltage sources were connected in parallel.
3 The virtual-ground probe is just one more box of electronics that has to be mounted
in the crowded space around the preparation bath.
Nevertheless, the virtual ground technique is still used in some circumstances, most com-
monly with high-voltage two-electrode voltage-clamp amplifiers, because of the technical
difficulty of making a series current-measurement circuit operate at high voltages.
The Bath Error Potentials section below describes an important variation on the virtual-
ground design to eliminate the voltage error due to current flow through the grounding
electrode.
sources: the capacitance across the glass wall (transmural capacitance) of the immersed
part of the micropipette; the stray capacitance from the rest of the micropipette to nearby
grounded surfaces; the stray capacitance from the micropipette holder; and the capaci-
tance to ground at the input of the buffer operational amplifier.
The input capacitance and the micropipette form a simple low-pass filter. That is, high-
frequency signals at the tip of the micropipette are shunted to ground by the input capac-
itance. There are several ways to improve the bandwidth of the recording system.
The best is to minimize the physical magnitude of the various elements contributing to
the input capacitance. The transmural capacitance of the immersed part of the micropi-
pette can be quite large. Typically, it is 1 pF or more per millimeter of immersion depth.
An effective way to reduce this capacitance is to thicken the wall of the micropipette. This
can be done by coating the micropipette with Sylgard or an equivalent material (see Chap-
ter 4). Other materials have been used, and in extreme cases, researchers have used a wide
micropipette to form a jacket over the narrower recording micropipette.
The other obvious way to reduce the transmural capacitance is to reduce the fluid level so
that the immersion depth is minimal. This is not always as effective as might be expected,
because surface tension causes the bath solution to creep up the surface of the micropi-
pette. This generates significant capacitance between the inside of the micropipette and
the film of solution on the outside. Forming this film can be prevented by making the sur-
face of the micropipette hydrophobic by dipping the filled micropipette into mineral oil
(or silane) immediately before using it. Note that it is essential to fill the micropipette first
with the aqueous electrolyte pipette solution so that the aqueous solution prevents the
mineral oil from entering the tip. Finally, one should try not to place the holder or the
microelectrode too close to grounded surfaces (such as the microscope objective).
Once the physical magnitude of the stray capacitance has been minimized, electrical tech-
niques can be used to reduce the effective magnitude. There are three such techniques:
1 Some researchers surround the micropipette with a metal shield that is connected to
the unity-gain output of the buffer amplifier. The principle here is to make sure that
the stray capacitance does not have a signal voltage across it. As long as there is no
signal voltage across the capacitance, no high-frequency current is drawn and the
bandwidth is increased. Unfortunately, this technique can significantly compromise
the noise performance of the amplifier because random noise generated in the buffer
amplifier is coupled back into its input via the shield capacitance. For this reason,
MDS Analytical Technologies does not recommend the use of a driven shield.
2 Unity-gain feedback can be used to reduce the component of stray capacitance that
exists between the amplifier input and its power supplies and case (Figure 3.7). In a
technique known as “bootstrapping,” sophisticated circuitry is used to superimpose
the unity-gain output of the buffer amplifier back onto its own power supplies and
the headstage case. This eliminates the high-frequency current loss through the power
supply capacitance and consequently increases the bandwidth. Since the power supply
Vp + 10V
Vp +1 C+
A1 Vp
C-
Cext Vp - 10V
+1 A2 Vp
Bootstrapped power supplies are used to increase the bandwidth by minimizing the
high-frequency current loss through the power-supply capacitance. The power-supply
capacitance is made up of C+ and C- from the positive input of A1 to the positive and
negative power supplies, respectively. All of the stray capacitances other than C+ and
C- are lumped together and called Cext in this figure. The power supplies of A1 are
batteries. The center point of the batteries is connected to a buffered version of Vp. If
the batteries are 10 V, the power supplies for A1 are (Vp + 10 V) and (Vp – 10 V). The
voltages across C+ and C- are constant at +10 V and -10 V, respectively. Hence, there
are no transient currents in neither C+ nor C-, except at very high frequencies where
these simple amplifier models are not appropriate. To ensure stability, the bandwidth
of A2 must be half or less of the bandwidth of A1. In practical implementations such
as the MultiClamp 700B and Axoclamp 900A microelectrode amplifiers, the batteries
are simulated by electronics.
3 Finally, but not least, the technique known as “capacitance compensation” or
“negative capacitance” can be used to reduce the effective value of whatever
capacitance remains. An amplifier at the output of the unity-gain buffer drives a
current-injection capacitor connected to the input (Figure 3.8). At the ideal setting of
the amplifier gain, the current injected by the injection capacitor is exactly equal to
the current that passes through the stray capacitance to ground. If the amplifier gain is
increased past the ideal setting, the current injected back into the input will cause the
input signal to overshoot. As the gain is increased past a certain setting, the circuit will
oscillate with potentially disastrous consequences for the cell.
knVp
In A2 +k n
Cn
Ip
Vp
+1 A1 Vp
I in
C in
Rp
Vm
Amplifier A2 places a voltage proportional to the pipette potential (Vp) across the capaci-
tance-neutralization capacitor (Cn). At a particular setting of the gain (kn) of A2, the cur-
rent (In) through Cn is exactly equal to the current (Iin) through the stray capacitance
(Cin) at the input of A1. By conservation of current, the pipette current (Ip) is zero. In the
ideal case, the effective capacitance loading the pipette resistance (Rp) is zero and the
bandwidth is therefore very high. In practice, limitations due to the finite bandwidth of
the amplifiers, the distributed nature of Cin and other second-order effects limit the
recording bandwidth.
With the Axoclamp 900A amplifier, the HS-9A headstage series was introduced. These
headstages have an increased capacitance neutralization range (-10 to 35.5 pF), indepen-
dent of the gain of the headstage. These headstages can be used equally effectively as volt-
age-recording or current-passing electrodes with the Axoclamp 900A amplifier. The
choice of HS-9A headstages for TEVC depends on the size of the typical current to be
measured.
(e.g., 2), the amplifier may saturate during large signals. In many experiments, large inputs
would be unusual, and even if they occurred the speed penalty incurred would not be
disastrous. However, in two-electrode voltage clamp experiments, large signal excursions
are common and speed is crucial. To eliminate the saturation of the capacitance compen-
sation circuit, the Axoclamp 900A amplifier uses ±180 V amplifiers in the capacitance
neutralization circuit in two-electrode voltage clamp mode.
If everything were ideal, the bandwidth of the recording system would approach the band-
width of the buffer amplifier alone (typically a few megahertz). However, in practice the
bandwidth is limited by the phase delays in the capacitance compensation pathway, and
by the false assumption that the stray capacitance can be represented by a lumped capaci-
tor at the input of the amplifier. The capacitance is distributed along the length of the
micropipette. Nevertheless, the techniques described above can be used to substantially
improve the bandwidth. It is not unusual to achieve bandwidths of 30 kHz with micropi-
pettes that would have a bandwidth of just one or two kilohertz if the above techniques
were not implemented.
There are two ways to ensure low-leakage currents for these microelectrodes: First, use
special operational amplifiers that have bias currents of just a few tens of femtoamps. Sec-
ond, make the value of Ro very large (100 GΩ) so that the current induced by small offset
voltages across Ro is small.
Depending upon the grounding method, the resistance of the bath grounding electrode
(Rb) could be as much as 10 kΩ, although with care it is not difficult to achieve Rb values
less than 1 kΩ.
In a simple two-electrode voltage clamp (TEVC) setup, the voltage drop across Rb is
indistinguishable from the membrane potential. That is, the potential (V1) recorded by
the voltage-recording micropipette is the sum of the transmembrane potential (Vm) and
the bath potential (Vb). Problems arise if the product of the clamp current (I2) and Rb is
significant. For example, for I2 = 5 μA and Rb = 2 kΩ, the error voltage is 10 mV. In some
experiments, a worst-case error of this magnitude might be tolerable; but if the error were
to be much greater, the position of the peak of I-V curves and other responses would be
seriously affected.
To faithfully record Vm, either Vb must be made equal to or nearly equal to zero, or the
value of Vb must be independently measured and subtracted from the potential recorded
by ME1. There are several methods to choose from:
Minimizing Rb
There are three main contributors to Rb:
1 The cell access resistance from the membrane surface to the bath.
2 The resistance of an agar bridge, if used.
3 The resistance of the grounding wire.
Cell Access Resistance
The access resistance to a sphere is given by:
ρ
Ra = (1)
4πr
Ra = 127 Ω
ρ × length (cm)
Ragar = (2)
area (cm 2 )
where length refers to the length of the agar bridge and area is the cross-sectional area
of the agar bridge.
For a 1 cm-long agar bridge with a 2 mm internal diameter (ID) filled with Ringer’s
solution:
Ragar = 2.6 kΩ
Ragar = 10.2 kΩ
The resistivity of 3 M KCl is approximately 20 times lower than for Ringer’s solution.
Thus, for an agar bridge filled with 3 M KCl, having the same length as above:
To minimize Rb, it would be best to eliminate the agar bridge and ground the prepa-
ration directly with an Ag/AgCl pellet. The pellet should be as large as practical, and
the area of contact between the pellet and the solution should be maximized.
However, if the bathing solution is changed during the experiment, the DC offset of
the Ag/AgCl pellet will change with the chloride activity. Thus, in this case it is essen-
tial to either use an agar bridge to prevent the DC offset of the bath from changing,
or to use a bath-potential recording headstage with a 3 M KCl electrode to follow and
remove the shift. Another advantage of an agar bridge is that it prevents metal ions
from the grounding electrode from entering the bathing solution.
If an agar bridge is used, it is best to fill it with 3 M KCl instead of Ringer’s solution
in order to minimize Rb. When the agar bridge is filled with 3 M KCl, the sum of all
components of Rb will be approximately 1–2 kΩ. If leakage of KCl from the agar
bridge is a problem, it may be necessary to fill the agar bridge with Ringer’s solution.
In this case, Rb will be several kilohms or more.
Series-Resistance Compensation
This sophisticated technique is used to minimize the effects of the electrode resistance in a
whole-cell voltage clamp. It is described later in this chapter and is of paramount impor-
tance when there is no alternative available. However, series-resistance compensation is
never completely effective. The other methods discussed in this section are more reliable
and easier to apply when it comes to eliminating the voltage error across Rb.
In this method, the bath is grounded by any preferred method. The bath potential is
allowed to shift because of solution changes, temperature changes or current flow. The
changing values of the bath potential (Vb) are continuously monitored by a 3 M KCl-
filled electrode placed near the cell. The measured value of Vb is subtracted from the
potentials measured by the intracellular microelectrodes. For microelectrodes being used
for voltage recording, this subtraction yields the true transmembrane potential (Vm). The
signal provided to all subsequent circuits in the amplifier is Vm.
This method operates well. The minor inconveniences are the more complex electronics,
the noise of the bath electrode that is added to the noise of the intracellular recording
micropipette, and some (minor) potential for instability in a two-electrode voltage clamp
if the bandwidth of the bath potential headstage is greater than the bandwidth of the
intracellular recording micropipette.
Clamp Vb Using a Bath Clamp
Another means to eliminate the effect of the voltage drop across Rb is to actively control
the bath potential measured near the outside surface of the cell. This is achieved using a
two-electrode virtual-ground circuit (Figure 3.9). One electrode (SENSE) is a voltage-
sensing electrode. This electrode is placed in the preparation bath near the cell surface. It
is connected to the virtual-ground circuit by an agar bridge or other means, of resistance
Rb2. Since there is no current flow through this electrode, there is no voltage drop across
Rb2. The other electrode (IBATH) is placed in the preparation bath. This electrode carries
the ionic current. The feedback action of the operational amplifier ensures that the poten-
tial at the SENSE electrode is equal to the potential at the negative amplifier input (ignor-
ing the offset voltage at the inputs to the operational amplifier).
SOLUTION
CELL
R b1 I BATH Rf
I BATH
SENSE
R b2
SENSE
- V1
+
VG-2A x100
In this implementation, separate electrodes are used to pass the bath current and control
the bath potential (unlike Figure 3.6 above). The basic operation is the same as for the vir-
tual-ground circuit described earlier, except that instead of connecting the feedback resis-
tor directly to the negative input of the operational amplifier, it is connected indirectly via
the bath solution. The bath potential near the surface of the cell is recorded by the SENSE
electrode (resistance Rb2) and forced by the feedback action of the operational amplifier to
be near ground. The bath current required to achieve this is passed by the operational
amplifier through the feedback resistor (Rf ) and through the bath current electrode
(IBATH; resistance Rb1).
Clamp Vb Using a Virtual-Ground Current Monitor
This technique is identical to the bath clamp described above, except that in this case
instead of being ignored, the output of the circuit is used to monitor the current flowing
from the micropipettes into the bath. The output (VI) of this circuit is proportional to the
bath current.
There are several problems with this technique. First, if the output is used to record the
bath current, it is usual to ignore the Rb1 term (because its value is unknown) and to
assume that
V I = R f I BATH (4)
Thus, there is a small error equal to Rb1/(Rb1 + Rf ). Second, if there are fluid-filled tubes
connected to the bath, and if some of these run outside of the Faraday cage, they will act
as antennas and conduct a lot of hum current into the bath that will be recorded by the
virtual-ground circuit. Third, if it is desired to use more than one amplifier with the same
preparation, it is essential that no more than one of them uses a virtual-ground circuit to
clamp the bath potential. This is not a serious problem, since it is easy to make one vir-
tual-ground serve the needs of both amplifiers. The problem is serious, however, if one of
the amplifiers introduces command signals via the virtual-ground bath clamp. This is a
practice employed by some manufacturers because in some circumstances it is easier to
design the electronics for this mode of operation. However, if command potentials are
introduced via the bath electrode, it is extremely difficult to perform experiments using
two amplifiers, e.g., whole-cell patch clamp of two fused cells.
When used with consideration for the measurement error, and if hum pick up is avoided,
a virtual-ground circuit to measure IBATH offers excellent performance.
Summary
As a first priority, we recommend that the value of Rb be minimized by the techniques
described above. If it is still considered to be important to compensate for the IR voltage
drop, a virtual-ground headstage (such as the VG-9A series) can be used to clamp the bath
potential, or a unity-gain voltage follower headstage (such as the HS-9A and HS-2A
series) can be used to record and subtract the bath potential.
The second electronic alternative is to drive a large positive or negative current step into
the cell.
In the Axoclamp 900A amplifier, both the duration and height of the current pulse are
adjustable. Again, the mechanism of penetration is not known. One cannot even predict
a priori whether a positive or a negative pulse will be more effective.
None of the three methods described here can be described as “best.” For a particular cell
type, the researcher must experiment to find the method that works best with those cells.
then instead of a command of zero generating 0 nA, it would generate 3 nA. In the second
case, the amplifier passes 1 nA for each 100 mV from the computer. This time, the 2 nA
of noise from the D/A converter only generates 20 pA of noise in the recording, and the
3 mV offset error only generates a 30 pA current error. The full ±100 nA range of the
amplifier can still be controlled, since most D/A converters produce up to ±10 V.
All of the Axon Cellular Neuroscience current-clamp and voltage-clamp instruments from
MDS Analytical Technologies use the maximum possible attenuation so that the
±10.24 V output from the D/A converter corresponds to the normal maximum operating
limits of the instrument.
Another important consideration when using externally generated commands is the pres-
ence or absence of polarity inversion. To minimize the risk of making mistakes during an
experiment, it is best that the instrument does not invert the command signal. Thus the
user should always know that a positive command from the computer will generate a pos-
itive current or voltage in the instrument.
The so-called “whole-cell patch” voltage clamp is a technique wherein only one micropi-
pette is used full-time for both voltage recording and current passing. A method known as
series-resistance compensation (see Section 3.3.7., “Series-Resistance Compensation”)
attempts to eliminate the aberrations arising when one micropipette is made to perform
the work of two. For series-resistance compensation to work, the micropipette resistance
must be reasonably low compared to the impedance of the cell. Since the resistance of
intracellular micropipettes generally is too high, researchers use very blunt micropipettes
that can be tightly sealed to a patch of membrane. The patch of membrane is subse-
quently ruptured so that the micropipette provides a low-resistance access to the whole
cell, hence the term “whole-cell patch” clamp.
Vm
Vcmd Rm Cm
The ideal voltage clamp simply consists of a battery, a switch, a wire, the cell and an
ammeter. When the switch closes, the membrane potential steps instantly to the battery
voltage. (For simplicity, it is assumed that Vm = 0 before the step.) There is an impulse of
current injecting a charge Q = CmVcmd; the steady-state current is Vcmd/Rm.
In this experiment, the voltage is the independent variable. Its value is controlled and
equal to the battery value. The current is the dependent variable; its value is measured by
an ammeter. For this reason, the voltage-clamp circuit is sometimes called a “current fol-
lower.”
A conventional two-electrode voltage clamp is shown in Figure 3.11. This figure has been
simplified by ignoring several frequency-dependent components: the input capacitance at
the input of micropipette ME1, the coupling capacitance between the micropipettes, and
the special phase lead and lag circuitry that is often introduced deliberately into the elec-
tronics. Furthermore, series resistance in the membrane and the bathing solution has been
ignored. In the figure, the sole frequency-dependent component is the membrane capaci-
tance.
A Vcmd
+μ
ε A2
+1 A1 -μ
R p1 R p2
ME1 ME2
B
Vm
Vm
Rm Cm R p2 Cm
τ= μ
Im
A. The membrane potential (Vm) is recorded by a unity-gain buffer amplifier (A1) con-
nected to the voltage-recording microelectrode (ME1). Vm is compared to the
command potential (Vcmd) in a high-gain differential amplifier (A2; gain = μ). The
output of A2 is proportional to the difference (ε) between Vm and Vcmd. The volt-
age at the output of A2 forces current to flow through the current-passing micro-
electrode (ME2) into the cell. The polarity of the gain in A2 is such that the current
in ME2 reduces ε.
B. Unlike the ideal case, there is a finite time required to charge the cell capacitance.
The time constant for the current and potential transients is τ = Rp2Cm/μ, where
Rp2 is the resistance of ME2. If was infinite, or if Rp2 was zero, the response would
approach the ideal case.
Error
The steady-state membrane potential (Vm) after a step change in the command voltage
(Vcmd) is:
μK
Vm = Vcmd (5)
μK + 1
where μ is the gain of the clamp amplifier and K is the attenuation of the clamp amplifier
caused by the cell membrane resistance (Rm) and the resistance (Rp2) of the output
micropipette (ME2):
Rm
K= (6)
Rm + R p 2
As the product μK becomes very large, the difference between Vm and Vcmd becomes very
small. Ideally, the error will be very low: just a fraction of one percent. The gain of the
clamp μ is typically set by a front-panel gain control in the range of about 100 to 10,000.
If K were unity, the error would vary from 1 percent down to 0.01 percent. However, if K
is less than unity (as it always is), the error will be worse. If the output micropipette resis-
tance is 90 MΩ and the membrane resistance is 10 MΩ, K is 0.1 and the error will be ten
times worse than if K were unity. If the two resistances are equal, K will be 0.5. Thus, as a
rule of thumb it is desirable to use an output micropipette whose resistance is as low as
possible, ideally about the same size or smaller than the membrane resistance.
Step Response and Bandwidth
After a step command, the membrane potential relaxes exponentially towards its new
value. For μK >> 1, the time constant for the relaxation is:
R p 2C m (7)
τ=
μ
Increasing the clamp gain decreases the time constant for the step response. For example,
if Rp2 = 10 MΩ, Cm = 1000 pF and μ = 100, the time constant is 100 μs.
Stated differently, increasing the clamp gain also increases the bandwidth with which Vm
can follow changes in Vcmd. The -3 dB frequency of the bandwidth is:
μ
f −3 = (8)
2πR p 2C m
Stability
The voltage clamp circuit shown in Figure 3.12 is unconditionally stable. The membrane
capacitance provides a 90° phase shift, which is required for stability in all negative feed-
back circuits. Unfortunately, other factors combine to make the circuit unstable at high
clamp gains.
The coupling capacitance (Cx) between the micropipettes is extremely destabilizing. Val-
ues as small as 0.01 pF can lead to oscillation if μ has a magnitude of several hundred or
more. There are several ways to minimize this coupling capacitance. The two best ways
are by:
1 Introducing the two micropipettes into the preparation at a wide angle, preferably
greater than 90°.
2 Placing a grounded metal shield between the two micropipettes. This shield should be
large enough to block all line-of-sight pathways between the two micropipettes and
their holders.
Another destabilizing factor is the non-ideal nature of the membrane. In Figure 3.11, the
membrane is simply modeled as a parallel resistor and capacitor. In practice, a distributed
model applies. The capacitance elements are themselves non-ideal; they should be mod-
eled by an ideal capacitor with a series-resistance component. For real membranes, the
phase shift at high frequencies is less than 90°. In the Axoclamp 900A amplifier, a phase-
shift control is included to allow the user to empirically introduce a phase lag to the cir-
cuit to build the total high-frequency phase shift up to 90°.
The input capacitance of the voltage-recording micropipette (ME1) adds another fre-
quency-dependent variable into the system that also tends to decrease the stability. The
effect of this input capacitance is usually minimized by carefully adjusting the capacitance
neutralization control to maximize the bandwidth of ME1.
Small instabilities in the voltage clamp usually show up as an overshoot in the current and
voltage responses. Full instability shows up as a continuous oscillation that can destroy the
cell. In some voltage clamps, special circuits are used to detect oscillations and take appro-
priate action to avoid damage to the cell. The Axoclamp 2 series and the GeneClamp®
500 amplifiers do not include an oscillation guard circuit. The Axoclamp 900A amplifier
includes an oscillation detection feature that automatically reduces the voltage clamp gain,
protecting the cells from harm.
Membrane Conductance Changes
It can be shown that the current response to a step change in the membrane conductance
is identical to the membrane potential response to a step change in the command voltage.
This is fortunate, because it means that if the voltage clamp is set up optimally by observ-
ing the response to a repetitive command step, it is also optimally set up for measuring
conductance changes. In order for this equivalence to be maintained, it is essential that
the experimenter not use tricks such as filtering the command voltage to eliminate an
overshoot in the response. This is a bad practice because it disguises the dynamic perfor-
mance of the clamp circuit.
Noise
There are several sources of noise in a two-electrode voltage-clamp circuit, but the most
important is the thermal and excess noise of the input micropipette, ME1. To the voltage-
clamp circuit, this noise appears as an extra command signal. It is imposed across the
membrane and leads to a noise current. Because of the membrane capacitance, this noise
current increases directly with frequency in the bandwidths of interest. The magnitude of
the noise is worst in cells with large Cm values.
Since the current noise increases with frequency, a single-pole filter is inadequate. An
active two-pole filter is required at minimum; while a four-pole filter is preferred. The
Axoclamp 900A amplifier includes a four-pole filter with a corner frequency up to 30 kHz
on each output. The user can choose between a Bessel and a Butterworth filter.
Micropipette Selection
Ideally, both micropipettes should have very small resistances. The resistance of ME1
should be small to minimize the micropipette noise, and the resistance of ME2 should be
small to maximize the product μK. However, low-resistance micropipettes are more blunt
than high-resistance micropipettes and do more damage to the cell. In general, it is more
important to use a low resistance micropipette for ME2. If its value is high, the amplifier’s
power supply will saturate at a lower current, causing a voltage-clamp error because the
ME2 headstage cannot inject enough current to hold the membrane voltage “clamped.”
As a result, the recorded data will be of dubious merit. A relatively high-resistance
micropipette for ME1 has its problems—increased noise and reduced bandwidth—but at
least it does not introduce an error.
Input vs. Output Offset Removal
Certain offset-removal circuits act only on the output signal delivered to the oscilloscope
or data acquisition system. Examples of these are the AC input of the oscilloscope itself,
the DC-offset removal in the CyberAmp 380 amplifier, and the output offset controls in
the Axoclamp 900A. Adjusting these controls does not affect the membrane currents
injected through the micropipette.
On the other hand, some offset-removal circuits act on the input signal seen by the volt-
age-clamp circuit. For example, the micropipette-offset controls alter the potential that is
clamped by the voltage-clamp circuit. Thus the actual membrane potential is affected
when these controls are altered and, consequently, the membrane current is affected. Care
should be taken not to alter the setting of an input-offset control once the experiment has
begun.
A block diagram and a timing diagram illustrating the technique are shown in
Figure 3.12. A single micropipette (ME1) penetrates the cell and the voltage recorded
(Vp) is buffered by a unity-gain headstage (A1). To begin the discussion, assume that Vp is
exactly equal to the instantaneous membrane potential (Vm). A sample-and-hold circuit
(SH1) samples Vm and holds the recorded value (Vms) for the rest of the cycle.
S1
voltage recording
0 volts
GT
CCS
Io current passing
Vms
+1 A1 SH1 - S2
Vp ε A2
ME1 + voltage current
clamp clamp
I cmd
Vcmd
Vm
T
current passing
S1 voltage recording
τp
Vp
sample
sample
Vm
Vcmd
Vms
Vms (average)
Vms is compared with a command voltage (Vcmd) in a differential amplifier (A2). The
output of this amplifier becomes the input of a controlled-current source (CCS) if the
switch S1 is in the current-passing position.
The CCS injects a current into the micropipette that is directly proportional to the volt-
age at the input of the CCS irrespective of the resistance of the micropipette. The gain of
this transconductance circuit is GT.
The period of current injection is illustrated at the start of the timing waveform. S1 is
shown in the current-passing position during which a square pulse of current is injected
into the micropipette, causing a rise in Vp.
The rate of rise is limited by the parasitic effects of the capacitance through the wall of the
glass micropipette to the solution and the capacitance at the input of the buffer amplifier.
The final value of Vp mostly consists of the IR voltage drop across the micropipette due to
the passage of current Io through the micropipette resistance Rp. Only a tiny fraction of
Vp consists of the membrane potential (Vm) recorded at the tip.
S1 then switches to the voltage-recording position. When the input of the CCS is 0 volts,
its output current is zero and Vp passively decays. During the voltage-recording period Vp
decays asymptotically towards Vm. Sufficient time must be allowed for Vp to reach within
a millivolt or less of Vm. This requires a period of up to nine micropipette time constants
(τp). At the end of the voltage-recording period, a new sample of Vm is taken and a new
cycle begins.
The actual voltage used for recording purposes is Vms. As illustrated in the bottom timing
waveform, Vms moves in small increments about the average value. The difference
between Vms,avg and Vcmd is the steady-state error (ε) of the clamp that arises because the
gain (GT) of the CCS is finite. The error becomes progressively smaller as GT is increased.
Minimum Sampling Rate and Maximum Gain
If the sampling rate (fs) is too slow, the dSEVC will become unstable. This is because the
resultant long current-passing period (Ti) allows the membrane potential to charge right
through and past the desired potential before the clamp has had an opportunity to take a
new sample of potential and adjust the current accordingly. The larger the cell membrane
capacitance (Cm) the larger the value of Ti (and hence the slower the sampling rate) that
can be used for a given average gain (GT). The stability criterion is:
GT Ti
0< <2 (9)
Cm
GT Ti
=1 (10)
Cm
Thus for a given GT, if Cm is small, Ti must be small (i.e., fs must be large).
For example, if GT = 1 nA/mV and Cm = 100 pF, then Ti must be 100 μs for optimum
response. If Ti is greater than 100 μs, the step response will overshoot and at 200 μs the
clamp will oscillate. (Note that in Axoclamp amplifiers, Ti = 100 μs corresponds to a sam-
pling rate of 3 kHz because the duty cycle (discussed later) is 30%.)
If Ti in this example cannot be as short as 100 μs because the micropipette response is too
slow, then a lower value of GT will have to be used to maintain optimum response.
Fastest Step Response
In a critically damped system, that is, where the gain has been turned up to achieve the
fastest possible response with no overshoot, the clamp can settle to its final value in as lit-
tle as one or two periods.
Error Number One—Clamping the Micropipette
If the micropipette voltage does not have sufficient time to decay to zero before a new
sample of Vm is taken, the sample will include a micropipette artifact voltage. The clamp
circuit will be incapable of distinguishing this micropipette artifact voltage from the mem-
brane potential and thus it will voltage clamp the sum of the two potentials. This leads to
inaccurate and bizarre data collection. For example, cells with known rectifiers will gener-
ate I-V curves that are almost linear.
This error is easily avoided by carefully observing the monitor output on the Axoclamp
900A amplifier. It is essential that the micropipette voltage trace on this output be given
sufficient time to decay to zero before the sample is taken. The user can achieve this by
adjusting the sampling frequency and by optimizing the micropipette capacitance com-
pensation. With the Axoclamp 900A amplifier, the micropipette voltage monitor is trans-
mitted to the computer through a USB 2.0 connection. When dSEVC is activated, a
window displaying the micropipette voltage is automatically displayed. This facilitates the
experimenter in observing the micropipette voltage and avoiding clamp errors.
Error Number Two—Steady-State Clamp Error
Even if the sampling is such that there is negligible voltage drop across the micropipette,
there will still be a steady-state error due to the finite gain, GT, of the clamp. To minimize
the error, it is necessary to use as high a gain setting as possible, consistent with stable
operation.
The anti-alias filter is a special filter used in the Axoclamp 2-series amplifiers to reduce the
bandwidth of the micropipette. Ordinarily, this is not desirable, because the sampling rate
has to be reduced to suit. However, some micropipettes have a biphasic decay after a cur-
rent pulse (Figure 3.13). There is an initial rapid decay as the micropipette capacitance is
discharged, followed by a smaller, slower decay resulting from ion redistribution in the tip.
For these micropipettes, the maximum sampling rate is severely restricted by the slow
phase of the micropipette decay. However, the wide-bandwidth noise of the micropipette
is related to the initial fast phase. For these micropipettes, it makes sense to slightly filter
the micropipette response to try to remove some of the wideband noise as long as this can
be achieved without increasing the ultimate settling time (which is limited by the slow
decay). This filtering is provided by a variable single-pole filter called the Anti-Alias Filter.
With the Axoclamp 900A amplifier, the anti-alias filter is applied along with the lag filter.
This implementation leads to increased clamp stability and speed.
Fast phase
Slow phase
The voltage response of a high-resistance micropipette is often biphasic. The initial fast
phase reflects the time constant formed by the micropipette resistance and the stray capac-
itance. This time constant sets the cutoff frequency for thermal noise. The slow phase is
thought to be due to ion redistribution in the tip. The maximum sampling rate during
dSEVC is limited by the time taken for the slow phase to settle.
Capacitance Neutralization Noise
Another source of noise in discontinuous single-electrode voltage clamps arises from the
capacitance neutralization circuitry. A fundamental property of all capacitance neutraliza-
tion circuits is that they introduce noise in excess of what is contributed by the thermal
noise of the recording micropipette and the input noise of the buffer amplifier. The excess
noise becomes progressively larger as the amount of capacitance neutralization is increased
to reduce the micropipette time constant. In discontinuous systems the micropipette time
constant must be reduced more than in continuous systems because of the need for the
micropipette voltage to decay to Vm within the time allotted for passive recording. Com-
pared to a continuous two-electrode voltage clamp, the dSEVC will always be a factor of
three or more times noisier at similar gains.
Selecting the Duty Cycle
The duty cycle is the fraction of each period spent in current passing. As the duty cycle
approaches unity, the clamp fails to operate properly because insufficient time is left for
the micropipette artifact to decay, unless the sampling period is made unreasonably slow.
On the other hand, if the duty cycle is made very small, the magnitude of the current
pulse becomes very large (to maintain a given average current). Since there are conflicting
requirements on the selection of the duty cycle, a compromise must be found. It has been
shown that the best compromise is a duty cycle equal to 0.3. This is the value used in the
Axoclamp amplifiers.
Useful Signal Bandwidth
As a general rule, the useful signal bandwidth is about one tenth of the switching rate. If
the dSEVC is switching at 20 kHz, it is reasonable to expect to record membrane currents
to within a bandwidth of 2 kHz.
Current and Voltage Measurement
Since the system is injecting current through the micropipette during every cycle, there is
no place that a continuous record of the membrane current and voltage exists. Instead,
sample-and-hold amplifiers must be used to store the values measured at the appropriate
times.
For the Vm output, the appropriate time to measure the voltage is at the end of the period
of passive decay, just before the next current pulse begins. This sampled value is held for
the duration of the cycle.
For the Im output, the current is sampled during the middle of the current pulse. The
value is held until the next current pulse. Since the user is interested in the average current
entering the cell, the output of the sample and hold is multiplied by the duty cycle before
being presented on the output.
Conditions For Good dSEVC
1 The micropipette resistance should be as small as possible.
2 The micropipette capacitance should be minimized.
3 The membrane time constant must be at least ten times the micropipette time
constant.
Comparison with Two-Electrode Voltage Clamp
The TEVC is generally superior to the dSEVC. If the gains are adjusted for a similar
steady-state error, the TEVC will generally have just one third of the noise of the dSEVC.
If the gain of the TEVC is increased so that the noise of the two clamps is similar, the
TEVC will generally settle three times faster.
It is clear that a dSEVC should not be used in preference to a TEVC. It should only be
used in those cases where it is impractical to implement a TEVC, either because the cells
are not individually discernible or because they are too small.
The main use of the discontinuous current clamp technique is to aid in the setup of the
instrument before selecting dSEVC. DCC-mode recording has been used in some experi-
ments in place of Bridge-mode recording to eliminate the possibility of a badly balanced
Bridge introducing an error in the voltage measurement. However, in general, the advan-
tage of DCC mode is small because in DCC mode the capacitance neutralization control
plays a similar role to the Bridge Balance control in continuous current clamping. If the
micropipette response is too slow, the transient after the current pulses will not have
decayed to baseline before the next sample is taken, and a current-dependent error voltage
will be measured. Thus, DCC mode just replaces one problem with another. At the same
time, DCC mode has more noise and a lower signal-recording bandwidth, so its use as a
substitute for Bridge mode should be undertaken with caution.
The voltage at the top of the pipette is controlled by a voltage-clamp circuit. It is impor-
tant to realize that this is quite different from the situation in TEVC or dSEVC. In both
of the latter cases, the voltage at the tip of the voltage-recording micropipette is controlled
(remember, in dSEVC, time-division multiplexing effectively yields two micropipettes:
the voltage-recording micropipette and the current-passing micropipette). In cSEVC, the
same electrode is used simultaneously for voltage recording and for current passing. The
voltage recorded at the top of the pipette is the sum of the membrane potential Vm, which
the experimenter wishes to control, and the current-induced voltage drop across the
pipette.
The current through the series resistance of the pipette and the residual resistance of the
ruptured patch is often sufficiently large to introduce significant voltage errors. Tech-
niques exist for compensating these errors. To get a feeling for the magnitude of the errors,
assume that the maximum compensation is 90%, beyond which the system oscillates and
destroys the cell. Further assume that the access resistance (Ra; the sum of the pipette
resistance and the residual resistance of the ruptured patch) is 5 MΩ. After compensation,
the effective value of Ra (Ra,eff ) would be just 0.5–1 MΩ. In this case, a 10 nA current
would cause a 50 mV uncompensated voltage error, reduced to 5 mV by the compensa-
tion. Clearly, the cSEVC technique cannot be used to record large currents. Even for
modest whole-cell currents, care must be taken to compensate for the series resistance and
then correctly interpret the residual error. The dSEVC technique should be considered as
an alternative to the cSEVC technique when the access resistance is too large.
The temporal resolution of the whole-cell patch clamp is also affected by Ra,eff. The time
constant for resolving currents is the product of Ra,eff (assuming that the membrane resis-
tance is much greater) and the membrane capacitance (τ = Ra,effCm). Thus the technique
is also limited to small cells where this product is small enough for the desired time resolu-
tion to be achieved. Figure 3.14 illustrates the voltage and temporal errors caused by the
presence of Ra.
A R a,eff
Im
Vm
Vcmd = Vp Rm Cm
V1
B
Vcmd
V1
Vm1
τ τ
Vm
τ
I1
IR
f
τ
Vm1 = V1 - I 1 R a,eff
V1 R m
= R +R
m a,eff
Im= V1
R a,eff + R m
Figure 3.14: Voltage and temporal errors caused by the presence of Ra.
A. The cSEVC circuit is simply illustrated as a voltage source (Vcmd) in series with the
effective access resistance (Ra,eff) and the membrane (Rm, Cm). The cSEVC circuit
ensures that the pipette potential (Vp) is equal to Vcmd.
B. After Vcmd steps to V1, a steady-state current, Im, flows in the circuit. The mem-
brane potential is equal to Vcmd - ImRa,eff. After the step change in the command
potential, Im and Vm settle exponentially to their steady-state values with τ =
[Ra,effRm/(Ra,eff + Rm)]Cm, but since in general Rm >> Ra,eff, a good approximation
is τ ≈ Ra,effCm.
tional to the measured current is used to increase the command potential. This increased
command potential compensates in part for the potential drop across the micropipette
(Figure 3.15). The amount of compensation achievable is limited by two considerations.
First, as the compensation level (α) approaches 100%, the increase in the command
potential hyperbolically approaches infinity. For example, at 90% compensation, the
command potential is transiently increased by a factor of ten (Vcmd/(1 – α)). Thus at
large compensation levels the electronic circuits approach saturation. Second, the current
feedback is positive; therefore, the stability of the circuit is degraded by the feedback and
at 100% compensation the circuit becomes an oscillator. In practice, the oscillation point
is much lower than 100% because of non-ideal phase shifts in the micropipette and the
cell membrane.
Rf
I
-
Vp
A1 +1
+ A2 VI
-1
Vm
Rm Cm
+1
A3
Vp +1
Vcmd
In this figure, a single pipette is used to voltage-clamp the cell. Operational amplifier A1 is
configured as a current-to-voltage converter. Differential amplifier A2 subtracts the
pipette potential (Vp) to generate the current output (VI). A fraction (α) of VI is summed
with the command voltage (Vcmd) used to control Vp. This causes both a transient and a
steady-state increase in Vp compared with Vcmd. As a result, the membrane charges faster,
the voltage drop across the electrode resistance is compensated and the bandwidth is
increased.
The first problem, saturation of the electronics, could in principle be reduced by using
high-voltage (e.g., ±120 V) operational amplifiers. However, this approach has not been
pursued because these types of operational amplifiers have more noise and worse drift
than good conventional operational amplifiers. The second problem, stability, has been
partially reduced by adding a variable low-pass filter in the current-feedback loop (e.g., the
“lag” control of the Axopatch-1D and the Axopatch 200 series amplifiers, and “band-
width” in the MultiClamp 700 series amplifiers). By empirically setting the low-pass filter
cutoff frequency, large percentage compensations can be used, but these only apply to the
currents at bandwidths below that of the filter cutoff. Thus the DC, low- and mid-fre-
quency series resistance errors can be substantially reduced while the high-frequency errors
remain large.
Supercharging (“prediction”)
Another technique for speeding up the response to a command step is the “supercharging”
technique, also known as “prediction.” In contrast to the “correction” method of series-
resistance compensation, “prediction” is an open-loop method. That is, there is no feed-
back and there is little risk of oscillations.
Supercharging is accomplished by adding a brief “charging” pulse at the start and the end
of the command voltage pulse. This means that initially the membrane is charging
towards a larger final value than expected (Figure 3.16). In its crudest form, the charging-
pulse amplitude or duration are adjusted empirically by the investigator so that the mem-
brane potential does not overshoot. Since the membrane potential is not directly observ-
able by the user, this is accomplished by adjusting the controls until the current transient
is as brief as possible. Once the optimum setting for one step size has been found, the size
of the supercharging pulse for all other step sizes can be calculated by the computer. Since
the large, transient supercharging current has to be carried by the feedback resistor in the
headstage, the amount of supercharging that can be used is limited by saturation of the
current-to-voltage converter in the headstage.
A Rf
-
Vp A1 +1 VI
+ A2
-1
Vcmd
Vm
B
Vpre
Vstep
Vcmd
Vpre
τm
Vstep
Vm
time
A. A brief charging pulse is added to the start and the end of Vcmd. The size and dura-
tion of the charging pulse is empirically determined to give the fastest charging
current. Vm charges rapidly to its final value. Vcmd and the supercharging pulses
are not part of a positive feedback circuit.
B. An expanded view of Vcmd and Vm. Initially, Vm increases towards the amplitude
(Vpre) of the pre-pulse. However, the pre-pulse is terminated early, just when Vm
reaches the desired command voltage, Vstep.
The Axopatch 200 amplifier pioneered an alternative technique for generating the super-
charging current. This technique takes advantage of having full knowledge of the Ra and
Cm values. Fortunately, this information is generally available on modern whole-cell
patch-clamp amplifiers as part of the technique used to unburden the feedback resistor
(Rf ) from the necessity of passing the charging current into the cell (i.e., whole-cell capac-
itance compensation). Once these parameters are determined, it is possible to automati-
cally boost changes in the command voltage to supercharge the cell (Figure 3.17). This
technique has the significant advantage that no empirical determination of the charging-
pulse amplitude or time is needed and that it works with any command-voltage shape
without requiring a computer to calculate the supercharging pulse.
Rf
-
Vp
A1 +1
+ A2
-1
Vcmd
+1
A3
Vm +1
SUPERCHARGING
PROCESSOR
The supercharging pulse added to Vcmd is a fraction (α) of a shaped version of Vcmd itself.
The shaped version, Vcmd(s), is generated in the whole-cell capacitance compensation cir-
cuit described later. Since the calculation of the supercharging pulse is determined by the
parameters of the cell and Vcmd itself, the system works with all commands (step, triangle,
sine, etc.) without the need for an empirical determination of the shape and amplitude for
each command shape and size.
There are two significant shortcomings that the investigator should be aware of when
using the supercharging technique:
1 The technique does not correct for the voltage error that occurs when current flows
through the series resistance of the pipette.
2 The dynamic response of the circuit is not improved. That is, there is no
improvement in the speed with which changes in current (and hence, changes in
conductance) can be resolved. Membrane current changes are still filtered by the time
constant of the access resistance and the membrane capacitance. The investigator
should not be misled by the rapid settling of the response to a command step into
thinking that this settling time represents the time resolution of the recording system.
Neither of these problems occurs with the correction method of series-resistance compen-
sation. The relative merits of correction versus prediction are shown in Table 3.1.
Table 3.1: Correction vs. Prediction.
Compensating the pipette capacitance in a patch clamp has three purposes. First, many
researchers want to remove the transient from the records for “cosmetic” reasons. Second,
during the transient the potential at the top of the pipette is changing slowly while the
pipette capacitance charges. By rapidly charging the pipette capacitance through the com-
pensation circuitry, the potential at the top of the pipette is stepped more rapidly, reduc-
ing the likelihood that rapid-onset ionic currents will be distorted. Third, uncompensated
pipette capacitance has a detrimental effect on the stability of the series-resistance correc-
tion circuitry. The component of the current that flows into the pipette capacitance is not
in series with any resistance. Thus the series-resistance correction circuit exceeds 100%
compensation for this component of the current as soon as the circuit is switched in.
In the Axopatch and MultiClamp amplifiers, two pairs of pipette capacitance compensa-
tion controls are available. With these controls, it is often possible to reduce the pipette
transients to extremely low levels. The bulk of the transient is reduced by using the fast
magnitude and time constant (τ) controls. The magnitude control compensates the net
charge. The τ control adjusts the time constant of the charge compensation to match the
time constant of the command pathway and to compensate for small non-idealities in the
frequency response of the pipette and electronics.
The residual slow component seen in many pipettes is reduced by using the slow magni-
tude and τ controls. A simplified circuit of the fast and slow compensation circuitry is
shown in Figure 3.18.
I C1
C1 = 1 pF
Rf
Rp
OPEN
CIRCUIT -
Cp I
+
Ip
Vcmd
FAST τ
10 V cmd +
FAST MAG
SLOW τ +
+
+
SLOW MAG
When the command potential (Vp) changes, current Ip flows into Cp to charge it to the
new potential. If no compensation is used, Ip is supplied by the feedback element (Rf ),
resulting in a large transient signal on the output (I).
By properly setting the fast and slow magnitude and τ controls, a current (IC1) can be
induced in capacitor C1 (connected to the headstage input) to exactly equal Ip. In this
case no current needs to be supplied by Rf, and there is no transient on the output.
It is important to understand that the transient is not suppressed in the belief that some-
how the data during this period is more meaningful in a patch-clamp amplifier than it is
in a two-electrode voltage clamp. The transient is suppressed for technological reasons. In
a patch-clamp amplifier the value of the feedback resistor for whole-cell clamping is typi-
cally 500 MΩ. In an instrument driven from ±15 volt power supplies, the maximum cur-
rent that can be passed through a 500 MΩ resistor is less than 30 nA. The transient
current required to apply a 100 mV step to a cell that is clamped through a 1 MΩ resistor
is typically 100 nA. If series-resistance compensation is used, much larger currents are
required. Since the 500 MΩ resistor cannot pass the required current, the system would
saturate and the time to effect a step would be prolonged significantly. In order to prevent
saturation of the system, the transient current is injected through a capacitor. Since the
current monitor output on patch clamps is only proportional to the current through the
feedback resistor, the transient current is not seen, even though in reality it is still being
passed through the pipette into the cell.
The settings of the whole-cell capacitance controls to eliminate the transient are unique to
the cell being clamped. The values of the cell membrane capacitance and the access resis-
tance can be directly read from the controls. Figure 3.19 is a simplified circuit illustrating
how the whole-cell capacitance controls operate.
C2 = 5 pF (50 pF)
Rf
Im Rp
-
I
+
Rm Cm
Vcmd
SERIES RESISTANCE
10 Vcmd +
WHOLE +
CELL
CAP
+
Vp
Assume that the fast and slow electrode compensation controls have already been set to
compensate for Cp. By appropriately adjusting the series-resistance and whole-cell capaci-
tance controls, the current injected through C2 will supply the transient membrane cur-
rent (Im). These adjustments do not alter the time constant for charging the membrane.
Their function is to offload the burden of this task from the feedback resistor, Rf. In many
cells, even a small command voltage of a few tens of millivolts can require such a large cur-
rent to charge the membrane that it cannot be supplied by Rf. The headstage output satu-
rates for a few hundred microseconds or a few milliseconds, thus extending the total time
necessary to charge the membrane. This saturation problem is eliminated by the appropri-
ate adjustment of the series-resistance and whole-cell capacitance controls. This adjust-
ment is particularly important during series-resistance correction since series-resistance
correction increases the current-passing demands on Rf. By moving the pathway for
charging the membrane capacitance from Rf to C2, the series-resistance circuitry can
operate without causing the headstage input to saturate. The effect of transferring the cur-
rent-passing burden from Rf to C2 is illustrated in Figure 3.20.
Vcmd
Vm
I Rf
I C2
Before After
Whole-cell Compensation Whole-cell Compensation
Figure 3.20: Using the injection capacitor to charge the membrane capacitance.
Vcmd is the patch-clamp voltage command; Vm is the voltage across the cell membrane;
IRf is the current across the patch-clamp feedback resistor; IC2 is the current injected
across the patch-clamp compensation capacitor.
The traces in this figure were recorded using an Axopatch 200 amplifier and a model cell.
After perfect whole-cell compensation is applied, the current to charge the membrane
capacitor is removed from the IRf trace and only the steady-state current remains. All of
the transient current appears in the IC2 trace. (The IC2 trace in the figure was recorded
using an oscilloscope probe connected to the internal circuitry). The I and Vm outputs on
the Axopatch 200 amplifier show the IRf and Vcmd trace illustrated in Figure 3.20. It is
easy to mistakenly think that the time course for charging the membrane is very fast, but
this is clearly not the case. Use of an independent electrode in the cell would show that the
cell charging rate is not affected by these adjustments.
Absolute Value
The absolute value of the membrane capacitance is displayed on the whole-cell capaci-
tance dial after the whole-cell current transient has been eliminated. This value may be
used to estimate the surface area of the cell assuming that the membrane capacitance per
unit area is 1 μF/cm2.
“Zap” is another alternative for going whole-cell. In this technique, a large, brief hyperpo-
larizing voltage pulse (about 1.5 V) is applied to the cell. This pulse initiates dielectric
breakdown of the membrane patch and allows access to the interior of the cell. It is impor-
tant to use the briefest pulse consistent with low-access resistance. If the pulse is too long,
the seal might be lost. Both the Axopatch 200B and MultiClamp 700B amplifier allow
the Zap pulse to be applied for 0.5 ms or less at the minimum.
With some cells it is difficult to go whole cell without losing the seal. An alternative is the
perforated patch technique, which is discussed in Chapter 5.
controlled, the axonal or dendritic membrane potential may be very poorly controlled. In
these cases, the time course of synaptic currents, regenerative currents and measurements
of reversal potentials may be grossly distorted.
As a general rule, the voltage clamp is considered to be acceptable if the length of the
attached axon or dendrites is no more than 1/10 of the length constants. Even this short
length will cause significant distortion of fast currents (see Figure 7 in Rall and Segev,
1985). Calculation of the length constant for a cell is complicated since it depends on the
geometry of the particular cell under investigation. Some of the common ways to avoid
the problems of poor space clamping are as follows:
Like the whole-cell patch technique, a blunt pipette is sealed onto a patch of membrane.
If single-channel recording is intended, the membrane at the tip of the pipette is preserved
(i.e., not ruptured). The current recorded is then simply the current that flows through
the membrane at the tip of the pipette. Since this membrane area is very small, there is a
5. Parts of this section have been reprinted with permission from Finkel, A. S., Progress in Instrumentation Technology for Recording
from Single Channels and Small Cells, in Cellular and Molecular Neurobiology: A Practical Approach, Oxford University Press, 1991.
good chance that just one or a a few ion channels are located in the patched membrane.
Individual ion-channel currents can thus be recorded. Previously, the only way to estimate
kinetics or conductance of these channels was the technique of “noise” or “fluctuation”
analysis. Noise analysis yields an estimate of the mean lifetimes of the channels and the
mean amplitudes, but no information about actual opening and closing transitions nor
about the shape of the conductance changes.
In single-channel recording, the current through the series resistance of the pipette is neg-
ligible, perhaps only a few picoamps flowing through a series resistance of just 10 MΩ.
The resulting voltage error is just a few tens of microvolts and is always ignored.
Current leakage though the seal resistance has a crucial bearing on the quality of the patch
current recording. Firstly, depending on the size of the seal resistance, a fraction of the
current passing through the membrane patch will leak out through the seal and will not
be measured. The lower the seal resistance the larger the fraction of undetected current.
Secondly, thermal movement of the charges in the conducting pathways of the seal consti-
tute the major source of noise in the recording unless the seal resistance is very high (sev-
eral gigohms or more). A high seal resistance is a prerequisite of low-noise recordings.
High-resistance seals, often called “gigaseals,” require that very clean pipettes be used and
that they only be used once. If this is the case, gigaseals will routinely form after the appli-
cation of gentle suction. When good gigaseals are formed, the noise due to the leakage
current is virtually eliminated and other sources of noise remain the dominant limitations
in the resolution of the recording technique. These are the noise of the electronics, the
pipette glass, the input capacitance and the feedback resistor. Commercial patch clamp
amplifiers such as the Axopatch 200B amplifier have minimized the noise of the electron-
ics and the feedback element in the headstage. Chapter 4 describes the attributes of desir-
able glasses and the ways to fabricate low-noise pipettes. It also discusses the use of Sylgard
coatings to minimize pipette capacitance.
through Rf. This current is proportional to the voltage across Rf which is measured at the
output of the differential amplifier (A2).
C Rf
If Rf
I
-
A1 +1
+ A2 BOOST
-1 VI
Vcmd
large, this source of noise becomes dominant; it is proportional to the product of the
input capacitance and the noise of the input FETs.
2 The minimum theoretically achievable current noise is lower for larger Rf values;
therefore it is important to choose large Rf values. The largest value typically used is
50 GΩ, since it allows a reasonable maximum current of more than 200 pA to be
measured. Unfortunately, for reasons that are not well understood, high-value resistors
are several times noisier than predicted by thermal-noise theory. Since the noise of
these resistors cannot be predicted, various brands of resistors must be tested until the
best one is found.
3 The inherent bandwidth of a 50 GΩ resistor is limited by the stray capacitance across
the resistor (CRf in Figure 3.21). For example, a 50 GΩ resistor with 0.1 pF stray
capacitance has a 5 ms time constant, corresponding to a bandwidth of only 32 Hz.
This poor time resolution is unacceptable for measuring ionic currents. Therefore, the
high-frequency components of the headstage output signal must be boosted. This is
typically achieved by an analog frequency compensation circuit. This circuit is made
complicated by the fact that Rf cannot be considered as an ideal capacitor in parallel
with an ideal resistor; thus a simple high-pass filter cannot be used. Complex circuits
consisting of up to four poles and three zeros in the transfer function are commonly
used. The placement of the poles and zeros must be carefully set for the particular
resistor.
4 The current required to charge the input capacitance during a step voltage command
can easily exceed the maximum current that can be passed by Rf from the typical 13 V
swing of the operational amplifier. For example, to linearly charge 5 pF of input
capacitance to 100 mV in 10 μs would require a charging current of 50 nA. This is
well beyond the 260 pA that can be passed by a 50 GΩ resistor driven from ±13 V.
Thus, special circuits are typically added to inject the required charging current
through a capacitor.
5 The best high-value resistors seem to be available only in a miniature “surface mount”
form. This means that the headstage electronics have to be manufactured in a hybrid.
This manufacturing technique has some advantages. First, the U430 input FETs and
other FETs typically used to switch between different values of Rf can be used in an
unpackaged form. This is to be preferred, since the sealing glasses used in packaged
transistors typically have a leakage resistance that increases the noise. Second, it means
that the sensitive input components can be maintained in a hermetically sealed
environment that reduces the rate at which they age due to environmental
contamination.
If the Probe Output is Slow, How Can Voltage Clamping be Fast?
In resistive-feedback headstages (but not capacitive-feedback headstages, discussed below)
the current output of the current-to-voltage (I-V) converter in the probe is slow. The
high-frequency boost occurs afterwards and cannot influence the events at the pipette.
Thus, one might conclude that the voltage clamp of the pipette must also be slow.
In fact, despite the slow current output of the I-V converter, the voltage clamp of the
pipette is rapid. The pipette is connected to the negative input (summing junction) of the
op amp. The command potential is connected to the positive input of the op amp. The
operation of the op amp in this configuration is to force the potential at the summing
junction to rapidly follow the potential at the positive input. If the command potential is
a step, the potential at the summing junction (and hence the pipette) is also a step. The
current required to achieve this step is passed through the feedback resistor (Rf ) and the
associated stray feedback capacitance (CRf ) of the I-V converter. The output of the I-V
converter settles to the final value with time constant RfCRf. This relatively slow settling
occurs despite the fact that the step at the summing junction is fast.
In this discussion, we have carefully referred to the fact that it is the pipette that is rapidly
voltage clamped. The membrane potential is voltage clamped to its final value much more
slowly. To a reasonable approximation, the time constant for voltage clamping the mem-
brane is RpCm, where Rp is the pipette resistance and Cm is the membrane capacitance.
In this technique, the feedback resistor in the headstage is replaced by a capacitor (Cf in
Figure 3.22). When a capacitor is used instead of a resistor, the headstage acts as an inte-
grator. That is, for a constant input current, the output of the headstage is a ramp. The
slope of the ramp is proportional to the current. To recover a voltage that is simply pro-
portional to the input current, the integrator must be followed by a differentiator (A2).
Cf
Rd
I
- Cd
A1 - VI
+ A2
+
Vcmd
The feedback element around operational amplifier A1 is capacitor (Cf ). Thus the output
of A1 is the integral of the pipette current. The actual current is recovered from the inte-
gral by differentiating it in the differentiator formed by Cd, Rd and A2.
Cd (10)
RT = Rd
Cf
Because the noise is theoretically independent of the Cf value and the gain of the
differentiator, RT, may be kept quite low, e.g., 100 MΩ. At this gain, the maximum
current that can be recorded is 500 times greater than for an RIV using a 50 GΩ
feedback resistor. Thus the CIV potentially has vastly improved dynamic range.
3 In the section on Resistor Feedback Technology, it was pointed out that in order to
achieve an acceptable bandwidth of, e.g., 20 kHz for single-channel recording, a
complex boost circuit has to be used to correct the frequency response. Even if the
boost box has as many as four poles and three zeros, the frequency response of Rf is
not perfectly corrected. The imperfections of the correction are most easily seen by
observing the step response of the headstage. Frequently, there will be overshoots and
ripples that can amount to as much as 2% of the response. However, because excellent
capacitors are available, the step response is inherently square with a CIV.
4 The bandwidth of a CIV can be very wide. It is maximized by using a small value of
Cf and a small value of Rd. For practical values (Cf = 1 pF and Rd = 150 kΩ), the
bandwidth is of the order of 70 kHz or more. This is considerably better than the
20 kHz of a good RIV using a 50 GΩ feedback resistor.
Whether it is practical to use this increased bandwidth or not is another matter. The
noise of both the RIV and the CIV headstages increases dramatically with frequency.
Problems
There are two major problems that make the capacitor feedback technology difficult to
implement. The first problem is that after a sustained net DC input current the integrator
voltage ramps towards one of the power-supply limits. When this happens, the integrator
must be reset by discharging the capacitor. The frequency of the resets depends on the size
of Cf and the average input current. For example, if Cf is 1 pF and the input current is
2 pA, the output of the integrator will ramp at 2 V/s. If the resetting circuitry is triggered
when the integrator voltage exceeds 10 V, resets will occur every five seconds.
The reset itself lasts approximately 50 μs. During the reset, sample-and-hold circuits are
used to maintain the current output at its value immediately prior to the start of the reset.
If it is acceptable to lose 0.1% of the data, resets as frequent as every 50 ms would be
acceptable. In the above example, this corresponds to an input current of 200 pA. For sin-
gle-channel recording, this is more than adequate.
Figure 3.23 shows the signal pathway for the capacitor-feedback headstage. The output
current (I) of the capacitor-feedback headstage is normally connected through a switch to
the output pre-filter amplifier, then to the low-pass filter, and finally to the post-filter
amplifier. The signal also goes through a low-pass filter to a sample-and-hold amplifier.
During reset, the switch shifts to the RESET position. Simultaneously, the sample-and-
hold amplifier is switched to the hold mode so that the signal immediately before the reset
transient occurs is presented to the output amplifiers.
50 μs RESET
DATA
NOT VALID
A capacitor feedback is not practical for whole-cell recordings. Due to the large whole-cell
currents the resets would be too frequent unless the value of Cf was increased. In practice,
it is hardly worth it. It is simpler to switch to a modest-sized resistor (e.g., 500 MΩ) and
avoid the problem of resets altogether. Reverting to resistor-feedback methods for whole-
cell recordings does not represent a setback in terms of noise because the 500 MΩ resistor
is not the dominant source of noise.
The second problem is that during resets transients are injected into the headstage input
by the reset circuitry. If left uncompensated, these would cause unwanted and possibly
damaging current pulses to be injected down the micropipette into the patch. Special
compensation circuitry needs to be included in the design to exactly balance out these
injected currents.
There are other transient problems that occur during and after the reset. These include
dielectric absorption in the differentiator and feedback capacitors, and other ill-defined
transients that work their way into the system. Compensation circuitry must also be
included to compensate for these transients.
Overall, the need to reset and to compensate for the many transients that occur during
reset make the design of an integrating patch clamp challenging. However, in a well-
designed system, the reset transients measured at the output will be less than 1 pA mea-
sured in a 1 kHz bandwidth.
For bilayer recording done at a low bandwidth (<1 kHz), the enCin noise has not yet
become the major contributor to the overall noise (where en is the voltage noise of the
probe input FETs and Cin is the capacitance at the input of the headstage primarily due to
the bilayer membrane capacitance). If the access resistance is large enough, it becomes the
dominant noise contributor at low bandwidths.
Figure 3.24 shows the dependence of peak-to-peak noise on access resistance in a 1 kHz
bandwidth for a given bilayer membrane capacitance. It can be seen that the lower the
access resistance, the lower the noise. It is therefore important that the patch-clamp
instrument be capable of operating without oscillation even when extremely small, or
zero, access resistances are used. The CV-203BU headstage for the Axopatch 200B ampli-
fier and the CV-7B/BL headstage for the MultiClamp 700B amplifier have been designed
for complete stability even with extremely low access resistances.
10 MEMBRANE
CAPACITANCE
BW =1 kHz
300 pF
(pA peak-to-peak)
200 pF
TOTAL NOISE
100 pF
1
0.1
100 1k 10 k
SERIES RESISTANCE (ohms)
cient to resolve the 10–20 fF capacitance increases occurring during fusion of single gran-
ules with the membrane.
A much more sensitive technique involves the use of a lock-in amplifier (also known as a
phase detector). In this technique, a sinusoidal command voltage is applied to the cell.
The magnitude of this voltage must be small enough so that the cell’s properties are essen-
tially linear. One measures two outputs that are proportional to the sinusoidal current
response at two orthogonal frequencies. It can be shown that the magnitudes of these two
responses are a function of the access resistance, the membrane resistance and the cell
capacitance. With a third measurement of the DC current, and assuming that the reversal
potential of the DC currents is known, there is sufficient information to calculate the
value of all three parameters. The resolution of this technique can be as good as 1 fF, with
measurements being made up to one hundred times per second.
Joshi and Fernandez (1988) have described a similar technique where all of the phase-sen-
sitive measurements and sinusoidal stimulations are performed by software. This has the
great advantage of not requiring any special equipment other than the patch clamp.
In patch clamps, the measurement of the combined pipette and seal resistance is not quite
as straight forward because the patch clamp is normally used in voltage-clamp mode
rather than current-clamp mode. In this case, a voltage step is applied to the pipette. The
combination of the pipette and seal resistance is calculated from the current response
using the formula:
R = V/I
For example, if the test pulse is 1 mV and the current response is 67 pA, the resistance is
15 MΩ.
With the MultiClamp 700B and Axoclamp 900A amplifiers, the Commander software
controlling the amplifier can apply a pulse and automatically calculate the electrode resis-
tance, whether in current or voltage clamp mode.
Clearly, a low-drift micromanipulator is required (as discussed in Chapter 2). But if the
holder is loose in the headstage connector or if the pipette does not seat properly, drift will
occur despite the high quality of the micromanipulator. One common, exasperating prob-
lem associated with loose holders is movement of the pipette when suction is applied,
which either prevents seal formation or damages the cell. The HL-U holders from MDS
Analytical Technologies fit all recent Axon Cellular Neuroscience headstages and have
been designed to fit snugly into the headstage connector and to firmly grip the pipette to
prevent drift. Second, the holder must not introduce noise. To assure this:
1 Select the right holder material. When a holder is connected to the headstage, even
before a pipette is connected, the open-circuit noise goes up slightly. The reasons are
not understood. Empirical tests have shown that when used with glass micropipettes,
polycarbonate adds the least noise. The HL-U holders use polycarbonate for the
pipette cap and the body of the holder. The portion that plugs into the headstage is
made of Teflon. Teflon in this position does not add noise, and it facilitates smooth
insertion and removal.
2 Use a small pin for the electrical connection. Large pins have more capacitance
contributed by the surrounding grounded surfaces.
3 Do not use a shield. Surrounding the holder with a driven or grounded shield adds to
the input capacitance and thus increases the noise.
3.5.1. DEFINITIONS
Positive Current
The flow of positive ions out of the headstage into the microelectrode and out of the
microelectrode tip into the preparation is termed positive current.
Inward Current
Current that flows across the membrane, from the outside surface to the inside surface, is
termed inward current.
Outward Current
Current that flows across the membrane, from the inside surface to the outside surface, is
termed outward current.
Positive Potential
The term positive potential means a positive voltage at the headstage input with respect to
ground.
Transmembrane Potential
The transmembrane potential (Vm) is the potential at the inside of the cell minus the
potential at the outside. This term is applied equally to the whole-cell membrane and to
membrane patches.
Depolarizing / Hyperpolarizing
The resting Vm value of most cells is negative. If a positive current flows into the cell, Vm
initially becomes less negative. For example, Vm might shift from an initial resting value
of -70 mV to a new value of -20 mV. Since the absolute magnitude of Vm is smaller, the
current is said to depolarize the cell (i.e., it reduces the “polarizing” voltage across the
membrane). This convention is adhered to even if the current is so large that the absolute
magnitude of Vm becomes larger. For example, a current that causes Vm to shift from
-70 mV to +90 mV is still said to depolarize the cell. Stated simply, depolarization is a pos-
itive shift in Vm. Conversely, hyperpolarization is a negative shift in Vm.
An inward sodium current flows in some cells after a depolarizing voltage step. When the
cell is voltage clamped, the sodium current is canceled by an equal and opposite current
flowing into the headstage via the microelectrode. Thus it is a negative current. When
two-electrode voltage clamping was first used in the early 1950’s, the investigators chose
to call the negative current that they measured a depolarizing current because it corre-
sponded to the depolarizing sodium current. This choice, while based on sound logic, was
unfortunate because it means that from the recording instrument’s point of view, a nega-
To prevent confusion, Axon Cellular Neuroscience instruments always use current and
voltage conventions based on the instrument’s perspective. That is, the current is always
unambiguously defined with respect to the direction of flow into or out of the headstage.
Some instrument designers have put switches into the instruments to reverse the current
and even the command voltage polarities so that the researcher can switch the polarities
depending on the type of experiment. This approach has been rejected by MDS
Analytical Technologies because of the real danger that if the researcher forgets to move
the switch to the preferred position, the data recorded on the computer could be wrongly
interpreted. MDS Analytical Technologies believes that the data should be recorded
unambiguously.
From the examples it can be seen that the transmembrane patch potential is inversely pro-
portional to the command potential, and shifted by the resting membrane potential
(RMP) of the cell. A positive pipette current flows through the pipette across the patch
membrane into the cell. Therefore a positive current is inward.
Inside-Out Patch
The membrane patch is detached from the cell. The surface that was originally the inside
surface is exposed to the bath solution. Now the potential on the inside surface is 0 mV
(bath potential). The pipette is still connected to the outside surface of the membrane. A
positive command voltage causes the transmembrane potential to become more negative,
therefore it is hyperpolarizing. For example, to approximate resting membrane conditions
of Vm = -70 mV, the potential inside the pipette must be adjusted to +70 mV. If the
potential inside the pipette is increased from +70 mV to +90 mV, the transmembrane
potential of the patch hyperpolarizes from -70 mV to -90 mV.
From the example it can be seen that the transmembrane patch potential is inversely pro-
portional to the command potential. A positive pipette current flows through the pipette
across the patch membrane from the outside surface to the inside surface. Therefore a pos-
itive current is inward.
Outside-Out Patch
The membrane patch is detached from the cell in such a way that the surface that was
originally the outside surface remains exposed to the bath solution. The potential on the
outside surface is 0 mV (bath potential). The pipette interior is connected to what was
originally the inside surface of the membrane. A positive command voltage causes the
transmembrane potential to become less negative, therefore it is depolarizing. For exam-
ple, to approximate resting membrane conditions, assuming that Vm = -70 mV, the
potential inside the pipette must be adjusted to -70 mV. If the potential inside the pipette
is then increased from -70 mV to -50 mV, the transmembrane potential of the patch
depolarizes from -70 mV to -50 mV.
3.5.4. SUMMARY
1 Positive current corresponds to:
Cell-attached patch patch inward current
Inside-out patch patch inward current
Outside-out patch patch outward current
Whole-cell voltage clamp outward membrane current
Whole-cell current clamp outward membrane current
3 The correspondence between the command potential (Vcmd) and the transmembrane
potential (Vm) is:
Cell-attached patch Vm = RMP – Vcmd
Inside-out patch Vm = – Vcmd
Outside-out patch Vm = Vcmd
Whole-cell voltage clamp Vm = Vcmd
3.6. REFERENCES
3.6.1. PATCH CLAMP
Rae, J.L. and Levis, R.A. Patch voltage clamp of lens epitheleal cells: theory and practice.
Molec. Physiol, 6, 115–162, 1984.
Hamill, O. P., Marty, A., Neher, E., Sakmann, B., and Sigworth, F.J. Improved patch-
clamp techniques for high-resolution current recording from cells and cell-free membrane
patches. Pflügers Archiv., 391, 85–100, 1981.
Sakmann, B. and Neher, E., Eds. Single-Channel Recording. New York: Plenum Press,
1983.
Finkel, A. S., Progress in instrumentation technology for recording from single channels
and small cells, in Cellular and Molecular Neurobiology: A Practical Approach. Oxford
University Press, 1991.
3.6.5. OTHER
Hille, B. Ionic Channels of Excitable Membranes. Sunderland, Massachusetts: Sinauer
Associates. 1984.
C1 - + Ag AgC1 + e−
The electrical potential at one of these electrodes is equal to the standard electrochemical
potential for Ag/AgCl plus RT/F ln(aCl-), where R is the gas constant (8.314 V C K-1
mol-1), T is the absolute temperature on the Kelvin scale, F is Faraday’s constant (9.648 x
104 C mol-1), and aCl- is the activity (i.e., the effective concentration) of Cl- in solution
near the electrode solution interface. The potential difference between a pair of these elec-
trodes should be zero if they are immersed in connected compartments with equal chlo-
ride concentrations. These concentrations must be kept constant during the course of an
experiment or an offset voltage will occur. An offset may be encountered when a Ag/AgCl
electrode is used as a bath electrode and the bath Cl- concentration is changed. To prevent
this type of voltage offset, many investigators connect their bath electrode through an agar
bridge that maintains a constant Cl- concentration in the immediate vicinity of the
Ag/AgCl electrode.
The composition of the pipette solution depends on the type of measurement being
made. Standard intracellular microelectrodes that are used to penetrate the cell membrane
must be filled with a highly concentrated electrolyte solution, usually 2–4 M salt. The
high electrolyte concentration reduces the electrode resistance. This minimizes the rectify-
ing current flow, lowers voltage noise and provides a wider recording bandwidth. A solu-
tion this concentrated dominates the liquid junction potential formed between the pipette
solution and the cell cytoplasm and results in a final junction potential that depends pri-
marily on the relative mobilities of the anions and cations in the filling solution. This is
important for the measurement of DC voltages because the cytoplasm anions are, pre-
dominantly, low mobility ions (mostly negative charges on large proteins), whereas the
cations are small, high-mobility charges. The difference in the mobilities of the cellular
anions and cations may result in a positive liquid junction potential. This positive liquid
junction potential would lead to the measurement of an artificially depolarized membrane
resting potential. The concentrated pipette solution is important because it negates the
development of the liquid junction potential, thus preventing an erroneous measurement
of the resting potential. However, there is a disadvantage in using the concentrated filling
solution since it can enter the cell and produce a hyperosmotic load that would cause the
cell to swell and alter its normal anion and cation content. While pipettes with very small
tip diameters can minimize or prevent concentrated solution from entering the cell, they
do so at the expense of noise, diminishing current passing ability, and limiting recording
bandwidth. These limitations result from the high resistance of small-tip pipettes. In prac-
tice, the choice of the pipette tip size reflects a compromise between its ability to prevent
the concentrated solution from entering the cell and the pipette’s resistance.
Patch micropipettes are usually filled with a solution that is iso-osmotic to cell cytoplasm
to allow current measurements in the absence of bulk water flow. Because of the much
larger diameter of the patch pipette tip (1–2 μm) as compared to the tip diameter of an
intracellular microelectrode (0.01–0.1 μm), the lower ionic-strength filling solution
results in an electrode whose resistance is as low as a few megohms. The specific choice of
the filling solution is dictated by the specific experiment to be done. Care must be taken
to minimize junction-potential artifacts. When the pipette tip is immersed in the bath, a
junction potential is generated. Its magnitude depends on the concentrations and mobili-
ties of the ions in both the pipette solution and the bath (see discussions of the Henderson
equation in electrochemistry textbooks). The junction potential is usually nulled while
the pipette tip is immersed in the bath. When the seal forms, the junction potential disap-
pears or changes its magnitude, and, consequently, the nulling voltage is no longer correct.
The junction potential also changes upon going whole cell. These effects must be taken
into account or offset errors will exist in either single-channel or whole-cell current
recordings.
The filling solution in a patch pipette can be changed during the course of an experiment
by inserting a fine tube into the back of the pipette through a special post in the electrode
holder. Different solutions can be injected into the tube. Since the tube can be placed
quite close to the pipette tip, the new solution can reach the vicinity of the pipette tip
after a short diffusion delay. The procedure of draining excess solution from the pipette
depends on the specific setup. This technique cannot be applied to standard intracellular
microelectrodes; due to their small tip size, the tube cannot be placed close enough to the
tip to allow changing the tip solution within a reasonable time.
The pipette is mounted in a holder that contains a Ag/AgCl electrode and a proper con-
nector to plug into the headstage of the amplifier. The holder must be constructed of a
low-noise, inert material, which is often chosen empirically. Axon Cellular Neuroscience
holders are constructed from polycarbonate, which yields low noise with several patch-
clamp pipettes. HL-U headstages have an electrode well to accommodate pipettes with an
outside diameter 1.0–1.7 mm. The HL-1-17 (for the non-U-type CV-type headstages)
and HL-2-17 holders (for the HS-type and VG-type headstages) have an electrode well
that can accommodate pipettes with an outside diameter of 1.5–1.7 mm. Pipettes of this
size have an internal bore that allows insertion of a 1 mm diameter Ag/AgCl pellet elec-
trode. The HL-1-12 (for the non-U-type CV-type headstages) and HL-2-12 (for the HS-
type and VG-type headstages) holders have a pipette well that can accommodate pipette
sizes of 1.0–1.2 mm diameter. This pipette size requires an electrode made of small-gauge
Ag wire coated with AgCl. A simple yet effective coating procedure is to clean the silver
wire down to bare metal using fine sand paper and immerse the cleaned wire in bleach
(Clorox) for 20–30 minutes, until the wire is uniformly blackened by the Ag/AgCl coat-
ing. Coating of the silver wire must be repeated whenever the AgCl becomes scraped from
the wire. Disruption of the coat can be minimized either by heat polishing the back of
each pipette before insertion into the holder or by encapsulating the coated portion of the
wire in a perforated Teflon sleeve.
The selection of the glass used for making the electrodes is more important for patch
pipettes than for intracellular micropipettes. Most intracellular micropipettes are made
from Corning #7740 Pyrex glass (Corning Inc., Corning, NY). However, when used for
tiny current measurements as in patch recordings, this glass is too noisy. The best patch
pipettes are made from glasses with low loss factors (see Table 4.1). These glasses generate
lower noise, and often have lower capacitances and fewer time constants to compensate.
If not pre-cleaned, the glass tubing should be cleaned prior to the preparation of the patch
pipette. Sonicating the glass in 100% ethanol in an ultrasonic cleaner is an effective clean-
ing procedure. After cleaning, the glass should be placed in an oven at 200 °C for
10–30 minutes to achieve complete drying. This heat treatment is necessary for low-noise
recording in environments with very high humidity.
Patch pipettes require much blunter tips than standard intracellular micropipettes and it is
usually impossible to pull them adequately on single-stage electrode pullers. Many labora-
tories have modified standard vertical pipette pullers to operate as multiple-stage pipette
pullers. Pulling pipettes, however, has become remarkably easier with the advent of micro-
processor-driven microelectrode pullers like those available from the Sutter Instrument
Company. With these pullers, it is possible to implement very complicated multistage
pulls of glass pipettes and to store the parameters required for the desired operation in the
instrument’s memory.
For the lowest noise recordings, patch pipettes must be coated with a hydrophobic mate-
rial to prevent the bathing solution from creeping up the wall of the pipette and, thus,
limit a substantial noise source. A commonly used compound is Sylgard #184 from Dow
Corning (Midland, MI). Coating the glass with Sylgard also improves the electrical prop-
erties of the glass. After preparing the Sylgard solution per the manufacturer’s instructions,
it can be stored in the freezer in small, well-capped centrifuge tubes for several weeks.
When brought to room temperature for use in painting electrodes, this freezer-stored Syl-
gard will last for several hours before it begins to polymerize. Sylgard is painted on the
pipette tip using a small utensil, such as a piece of capillary tubing pulled over flame to a
reasonably fine tip. Sylgard painting can be done using the magnifications available with
standard dissecting microscopes. The pipette is coated to within 100 μm or less from its
tip. It is important that the Sylgard be directed away from the tip by gravity at all times
during the painting procedure. Otherwise, it will flow into the tip and make firepolishing
and/or sealing impossible. After painting, Sylgard can be cured by holding the tip for
5–10 seconds in the hot air stream coming from a standard heat gun of the type used in
electronics fabrication.
To promote gigaohm seals and to reduce the possibility of tip penetration into the cell
during seal formation, pipette tips should be firepolished. Firepolishing is accomplished
by connecting to a micromanipulator a rod of inert material to which a short loop of plat-
inum iridium wire has been fastened. The ends of this wire must be soldered to two other
pieces of wire that can be connected to a voltage or current source to allow current passing
through the platinum wire. The platinum loop is generally bent into a very fine hairpin so
that it can be brought to within a few microns of the electrode tip under direct observa-
tion using a compound microscope. Magnifications of 600x–1500x are required for ade-
quate visibility. The platinum wire is usually coated with a glass, such as Pyrex or Corning
#7056, to prevent sputtering. After positioning the electrode tip near the glass-coated
wire, current is passed through the wire to heat it. The amount of current required
depends on the softening temperature of the glass from which the pipette is constructed.
The hot platinum wire heats the electrode tip and causes sharp corners to round and
rough surfaces to smooth. This polishing is best done under direct observation at a magni-
fication of 600x–1500x.
Log10
Loss Dielectric Softening
Glass Volume Description
Factor Constant Temp. °C
Resistivity
Log10
Loss Dielectric Softening
Glass Volume Description
Factor Constant Temp. °C
Resistivity
Note that the electrical properties and the thermal properties of the glasses are not neces-
sarily related to each other. Note also that Sylgard has better electrical properties than
most glasses shown in the table. It is, therefore, not surprising that heavy Sylgard coating
of pipettes fabricated from many types of glasses improves their electrical properties.
Table 4.2 shows the chemical constituents of many of the same glasses listed in Table 4.1.
This table may be useful in deciding which glasses are likely to contain leachable compo-
nents that might affect channel currents.
Chemical Constituent
Glass SiO2 B2O3 Al2O3 Fe2O3 PbO BaO CaO MgO Na2O K2O Li2O As2O3 Sb2O3 SO3
7070 70.7 24.6 1.9 ___ ___ 0.2 0.8 0.8 ___ ___ 0.56 ___ ___ ___
8161 38.7 ___ 0.2 ___ 51.4 2.0 0.3 0.04 0.2 6.6 ___ 0.04 0.38 ___
7059 50.3 13.9 10.4 ___ ___ 25 ___ ___ 0.08 ___ ___ ___ ___ ___
7760 78.4 14.5 1.7 ___ ___ ___ 0.1 0.1 2.7 1.5 ___ 0.18 ___ ___
7040 66.1 23.8 2.9 ___ ___ ___ 0.1 0.1 4.1 2.7 ___ 0.1 ___ ___
KG-12 56.5 ___ 1.5 ___ 28.95 ___ 0.1 0.1 3.7 8.6 ___ 0.4 0.25 ___
1723 57.0 4.0 16.0 ___ ___ 6.0 10.0 7.0 ___ ___ ___ ___ ___ ___
0010 61.1 ___ ___ ___ 22.5 ___ 0.3 0.1 7.2 7.3 ___ ___ ___ ___
EN-1 65.0 18.0 7.6 ___ 0.01 2.7 0.1 0.1 2.3 3.2 0.6 ___ ___ ___
7720 71.4 15.2 2.0 ___ 6.1 0.3 0.2 0.1 3.7 0.3 ___ ___ 0.5 ___
7056 69.0 17.3 3.9 ___ ___ ___ 0.12 0.91 7.5 0.68 0.48 ___ ___
3320 75.3 14.3 ___ ___ ___ ___ 0.1 0.1 4.0 ___ ___ ___ 0.8 ___
7050 67.6 23.0 3.2 ___ ___ 0.1 0.1 0.1 5.1 0.2 ___ ___ ___ ___
KG-33 80.4 12.9 2.6 ___ 0.005 ___ 0.05 4.0 0.05 ___ ___ ___ ___
7740 80.4 13.0 2.1 ___ ___ ___ 0.1 0.1 4.1 ___ ___ ___ ___ ___
1720 62.0 5.3 17.0 ___ ___ ___ 8.0 7.0 1.0 ___ ___ ___ ___ ___
N51-A 72.3 9.9 7.3 ___ 0.02 ___ 0.9 0.05 6.5 0.7 ___ 0.02 ___ ___
R-6 67.7 1.5 2.8 ___ ___ 2.0 5.7 3.9 15.6 0.6 ___ ___ ___ 0.2
0080 73.0 0.04 ___ ___ ___ 0.1 4.8 3.2 16.8 0.4 ___ ___ ___ 0.22
resistance (Ra) for whole-cell recording. Furthermore, the blunt-tapered pipettes are less
likely to penetrate into the cell when pressed against the cell membrane during seal forma-
tion. High-lead glasses pull at the lowest temperatures and are remarkably amenable to
firepolishing. It is possible to pull pipettes at such a low temperature that their tips are
broken and jagged, forming tip openings with diameters in excess of 50 μm, and yet the
pipettes are easily firepolished into usable patch pipettes. Although the resulting tips are
exceedingly blunt, these pipettes are effective in forming a tight seal with the cell even
when the final pipette resistance is < 0.5 MΩ.
Blunt tips are very beneficial for perforated-patch recording because they are capable of
drawing in large omega-shaped pieces of membrane when suction is applied. The large
surface area of the omega-shaped membrane maximizes the number of parallel channels
created by amphotericin B or nystatin, thus minimizing the final Ra that is achievable (see
Chapter 5).
Most of the pipettes used for patch-clamp recording are fabricated from borosilicate
glasses. These glasses soften at temperatures in the 700–850 °C range (see Table 4.1).
Although those at the low end of the range are quite easily pulled and firepolished, they
are clearly not as good as the high-lead glasses in this regard. Compared to high-lead
glasses, the success of firepolishing borosilicate glasses depends more on the shape of the
tip obtained after pulling. However, with the advent of multi-stage, computerized pipette
pullers, one can routinely make both patch and whole-cell pipettes from almost any glass.
Aluminosilicate glasses are very hard glasses with high softening temperatures. They pro-
duce low-noise single-channel recordings. However, their low noise comes at a high price.
Glasses in this class soften at temperatures above 900 °C and, therefore, pulling them
wears out the coils and filaments of pipette pullers. Consequently, the coils change their
properties with time and must be replaced or readjusted frequently. Moreover, pipettes
fabricated from aluminosilicate glasses produce undesirably thin-walled tips after pulling.
This, along with their high softening temperature, makes them much more difficult to
firepolish than softer glasses.
8161
7740
10
1720
R-6
30 0080
50
0.0 0.1 0.2 0.3 0.4 0.5 0.6
The dependence of the rms noise of a patch pipette on the loss factor of the glass.
As seen in Figure 4.1, glasses with the lowest loss factor exhibit the lowest noise. Further-
more, noise decreases as the wall thickness increases (compare 7052 with 7052TW, thin
walled). Even greater improvement in noise can be achieved using quartz (fused silica)
pipettes. The noise of pipettes fabricated from Corning 7070 (low loss electrical) or Corn-
ing 1724 (aluminosilicate) glasses could also be improved if proper techniques of fabricat-
ing patch pipettes from these glasses would be developed. The noise values shown in
Figure 4.1 are higher than would be obtained with the Axopatch™ 200B amplifier in sin-
gle-channel recording mode since its capacitive headstage produces lower noise than is
produced by standard resistive headstages. It has been observed that as the noise of the
headstage electronics decreases, the noise of other components, such as holder and glass, is
reduced as well.
Sakmann, B., Neher, E. Geometric parameters of pipettes and membrane patches. Single-
Channel Recording. Sakmann, B., Neher, E., Eds. pp. 37–51. Plenum Press. New York
and London, 1983.
Hammill, O.P., Marty, A., Neher, E., Sakmann, B., Sigworth, F.J. Improved patch-clamp
techniques for high resolution current-recording from cells and cell-free membrane
patches. Pflügers Arch. 391, 85–100, 1981.
Rae, J. L., Levis, R. A. Patch-clamp recordings from the epithelium of the lens obtained
using glasses selected for low noise and improved sealing properties. Biophys. J. 45,
144–146, 1984.
Rae, J. L., Levis, R. A. Patch voltage clamp of lens epithelial cells: theory and practice.
Molec. Physiol. 6, 115–162, 1984.
Rae, J. L., Levis, R. A., Eisenberg, R. S. Ion Channels. Narahashi, T. Ed. pp. 283–327.
Plenum Press, New York and London, 1988.
Furman, R. E., Tanaka, J. C. Patch electrode glass composition effects on ion channel cur-
rents. Biophys J. 53, 287–292, 1988.
Rojas, L., Zuazaga, C. Influence of the patch pipette glass on single acetylcholine channels
recorded from Xenopus myocytes. Neurosci. Lett. 88: 39–44, 1988.
Since these initial experiments, a number of workers have adopted this expression system
to study various aspects of the structure and function of ion channels and receptors. Some
of the more common types of studies that have employed expression of exogenous ion
channels and receptors in oocytes include:
Vitelline Vegetal
Membrane Pole
The oocyte is a large cell, with a diameter of approximately 1–1.2 mm. It is surrounded by
the vitelline membrane, which is a glycoprotein matrix that gives the oocyte some struc-
tural rigidity and helps it maintain a spherical shape. The mesh formed by this matrix is
rather large; consequently, small molecules and even small proteins such as α-bungaro-
toxin (molecular weight 8,000 dalton) can interact with proteins on the oocyte surface.
Since devitellinized oocytes are extremely fragile, the vitelline membrane is usually
removed only for single-channel recording, where the surface of the oocyte must be suffi-
ciently clean to allow the formation of a gigohm seal. A layer of follicle cells surrounds the
vitelline membrane. The individual cells in the follicle cell layer are electrically coupled to
each other and to the oocyte through gap junctions, so that electrical events taking place
in the follicle cells can be detected in the oocyte.
The follicle cell layer is a potential source for many problems. Therefore, most researchers
remove it prior to mRNA injection or recording by one of two ways: 1) an extensive colla-
genase treatment to completely strip off the layer by immersing the oocyte for 1–3 hours
in a calcium-free saline containing 2 mg/mL collagenase Type IA (Sigma Chemical Com-
pany, St. Louis, MO) until about half of the oocytes have been released from the ovarian
lobe; or 2) a less extensive collagenase treatment followed by manual removal of the follic-
ular layer with watchmaker’s forceps. The latter is preferred, since oocytes that undergo
the extensive collagenase treatment sometimes do not survive as long as those treated less
extensively. Some of the problems introduced by the follicle cell layer are rather trivial. For
instance, it is more difficult to impale a follicle-encased oocyte with an injection needle or
a microelectrode. Other complications, however, can create major problems. The electri-
cal coupling between the oocyte proper and the follicle cells means that one records from
both. Numerous examples of oocyte “endogenous” receptors that turned out to be in the
follicle cell layer were published. In fact, the Xenopus oocyte is a nearly ideal expression
system for ion channels and receptors since it has very few types of endogenous channels
and receptors, an advantage that is defeated if the follicle cells are left on the oocyte.
One of the striking features of a Xenopus oocyte is its two-toned color scheme: the animal
pole hemisphere is dark colored, while the vegetal pole is light colored. This polarity is
maintained inside the oocyte: the nucleus is found in the animal pole, and different popu-
lations of mRNAs exist in the hemispheres. In addition, a standing Cl- current flows from
one pole to the other, indicating that the distribution of channels in the oocyte membrane
is not homogeneous. Published reports showed that ACh receptors expressed from exoge-
nous mRNAs are also unevenly distributed; although receptors were expressed all over the
oocyte, more receptors were expressed on the vegetal pole. It is not clear if there is any
relation between the site of RNA injection and the location of the expressed channels on
the oocyte membrane. This non-homogeneity is not important for whole-cell recording,
but could be important for single-channel recording.
The response time (τ) of a voltage clamp to a step voltage change is:
RI C m (1)
τ=
A
The other parameter that one can control in order to maximize the speed of the voltage
clamp is the command amplifier gain. As with other types of cell-electrode combinations,
one must usually introduce a frequency response compensation network to ensure clamp
stability at high gains. Therefore, the absolute gain of the clamp and the command voltage
of the amplifier are less important than the maximum gain reached before the clamp
becomes unstable. The LAG and (AC) GAIN controls of the Axoclamp™ 900A micro-
electrode amplifier allow one to fine-tune the clamp. For oocytes, start with the LAG at
about 0.2 ms and the GAIN set to its lowest position, and gradually increase it as much as
possible until the clamp starts oscillating. The lower the LAG setting, the faster the clamp.
Axoclamp 900-series amplifiers include an Oscillation Detection circuit that automati-
cally reduces the voltage clamp gain when oscillation in the circuit is detected.
The goal is to achieve a fast voltage clamp. But, how fast is fast enough? The actual speed
required depends on the type of signal being measured. For example, when studying a
ligand-activated ion channel that is activated by bath application of an agonist, it makes
no difference if the clamp settles in 2 ms or 20 ms, since the exchange time of the cham-
ber is the limiting factor in the rate of activation. On the other hand, when studying volt-
age-gated channels, one wants the fastest clamp attainable. A well-tuned, two-
microelectrode voltage clamp can clamp an oocyte fast enough to study the gating kinetics
of most voltage-gated channels. However, it is probably impossible to clamp an oocyte fast
enough to study the activation kinetics of fast-gating voltage-gated Na+ channels in which
the rising phase of the current can be on the order of 1 ms or shorter. It is still possible to
study slower processes, such as inactivation or pharmacological properties, even if the
clamp is not fast enough to study the activation process. A faster clamp can be achieved
either by recording from smaller oocytes (stage II or III) whose capacitances are approxi-
mately one-fifth those of the larger stage V and VI oocytes, or applying the more common
“macropatch” technique. Either of these two approaches provides a sufficiently fast clamp
for studying the kinetics of Na+ channel activation.
Electrode penetration is most easily achieved simply by advancing the electrode into the
oocyte until it dimples the membrane and then visibly pops into the cell. Due to the large
size of the oocyte, a coarse, inexpensive manipulator provides adequate control of move-
ment. Typical resting potentials in a physiological saline solution (100 mM NaCl,
2 mM KCl, 1.8 mM CaCl2, 1 mM MgCl2, 5 mM HEPES, pH 7.6) are approximately -
40 mV or more negative. The voltage clamp is tuned by applying a small (2 mV) square
wave at the resting potential and then gradually increasing the gain of the clamp. As the
gain increases, the voltage trace “squares up” and the capacitive transient in the current
trace gets faster and faster. Instability in the clamp will first appear as small oscillations in
the traces; the oscillations will increase as the gain increases. If oscillations do appear, the
TIME CONSTANT control should be adjusted until they disappear. This process should
be continued until one reaches the highest gain possible without any apparent oscillations.
In general, if the electrodes are shielded as described above, full gain can be achieved.
Once the vitelline membrane has been removed, gigohm seals can be obtained with high
success rates. Oocytes, like most cells, contain stretch-activated channels that can interfere
with recording “interesting” channels in the cell-attached mode. Therefore, most research-
ers prefer to use excised patches for single-channel work, where the incidence of stretch-
activated channel currents is much lower. The actual mechanics of obtaining excised
patches is the same as for any cell type, with the exception that some researchers prefer to
rupture the membrane by applying positive pressure rather than suction, since suction can
sometimes cause yolk platelets to clog the pipette tip.
Gurdon, J.B., Lane, C.D., Woodland, H.R., Marbaix, G. Use of frog eggs and oocytes for
the study of messenger RNA and its translation in living cells. Nature. 233:177–182,
1971.
Krafte D., Lester, H.A. Expression of functional sodium channels in stage II–III Xenopus
oocytes. J. Neurosci. Meth. 26:211–215, 1989.
Methfessel, C., Witzemann, V., Takahashi, T., Mishina, M., Numa, S., Sakmann, B.
Patch clamp measurements on Xenopus laevis oocytes: currents through endogenous
channels and implanted acetylcholine receptor and sodium channels. Pflügers Arch.
407:577–588, 1986.
Snutch, T.P. The use of Xenopus oocytes to probe synaptic communication. Trend Neu-
rosci. 11:250–256, 1988.
Stühmer, W., Methfessel, C., Sakmann, B., Noda, M. Y., Numa, S. Patch clamp charac-
terization of sodium channels expressed from rat brain cDNA. Eur. Biophys. J. 14:131–
138, 1987.
Stühmer, W., and Parekh, A.B., Electrophysiological Recordings from Xenopus Oocytes
in Single-Channel Recording, 2nd Ed., Sakmann, B., and Neher, E., eds., pp. 341–355,
1995.
1 Cells in tissue slices are likely to be much closer to their original state than cells
subjected to the above-mentioned treatments. No disruption of the normal cellular
environment need take place until the preparation of slices and disruption is limited
to the surface of the slice.
2 The increased signal-to-noise ratio and improved voltage-clamp quality, compared
with conventional intracellular recording using sharp microelectrodes, provide
substantial benefits for cellular electrophysiology. These benefits are particularly
helpful when measuring small or rapid events.
3 The ability to control the intracellular environment is greatly enhanced by whole-cell
recording, thus facilitating the study of interactions between intracellular
biochemistry and electrophysiology.
It should be noted however, that not all problems are solved by applying whole-cell tech-
niques. “Washout” of intracellular biochemical machinery may affect electrophysiological
properties, and control of membrane potential in branched or elongated neurons is often
inadequate.
Two strategies have been applied in obtaining patch-clamp recordings from tissue slices.
One approach, the “cleaning” technique, involves removing overlying neuropil or cellular
debris from a visually-identified cell using a relatively large pipette filled with the same
solution used to superfuse the slice. Recording is then possible in a manner similar to
standard patch recording. In the other approach, called the “blind” technique, the record-
ing pipette is lowered into a tissue slice without high-magnification visual guidance; seal
formation is monitored using electrical measurements.
The only major piece of equipment required for this technique, in addition to a standard
recording setup, is an upright microscope equipped with a water-immersion objective,
most commonly 40x magnification. Use of an inverted microscope is not feasible since,
even if very thin slices are used, visibility through the tissue to a recording pipette is signif-
The recording chamber must allow continuous superfusion of the slice; it must also per-
mit immobilization of the tissue despite the fluid flow, and do so using a device that does
not interfere with the objective or the pipettes. One solution to this problem is a net con-
sisting of a U-shaped piece of flattened platinum wire with a parallel array of very fine
nylon monofilaments, such as stocking fibers, glued across the arms of the U-shaped wire
(Edwards et al., 1989; Sakmann et al., 1989). The net can be placed on the slice and is
usually heavy enough to prevent movement. Although the filaments may cause some
damage to the surface of the tissue, they may be placed sufficiently far apart to leave con-
siderable working space (the precise distance depends on the cellular architecture in the
slice). Another solution is to attach the slice to the bottom of the recording chamber using
a plasma clot (Blanton et al., 1989).
The pipette used for cleaning should have a tip much larger in diameter than a recording
pipette, usually in the range of 5–20 μm. The optimal tip size depends on several factors;
in general, larger tips are preferable for larger cells and for deeper cleaning. However, the
consistency of the tissue and cell density may also affect the choice of tip diameter. For
example, with particularly “sticky” tissue, such as neocortex, small cleaning pipettes may
tend to become clogged with membranous debris. For a given preparation, it is best to
arrive empirically at an appropriate size. Additionally, since debris tends to accumulate on
the jagged edges of broken pipettes, it is usually best to pull unbroken pipettes to the
appropriate size. To minimize any damage that the process might inflict on the cell, the
cleaning pipette should be filled with the same solution used for the bathing medium.
Note that some researchers, instead of using a separate cleaning electrode, perform the
cleaning with the same recording electrode for the sake of expediency.
The first step in establishing a recording is to locate a healthy cell; it is usually possible to
distinguish cells in good condition by their appearance of solidity or opacity. In such neu-
rons, it is often possible to follow dendritic processes for some distance. Some healthy cells
may be observed whose somata are on the surface of the slice and which display a “clean”
membrane suitable for immediate patching with no further cleaning. Such neurons often
have dendrites that can be seen projecting into the depths of the slice. It seems that cells
that have lost most of their dendrites due to the slicing procedure are often “sick,” as char-
acterized by a granular and/or transparent appearance of the soma.
Once a healthy cell has been located, the next step is to remove the overlying neuropil.
Mouth control of the pressure applied to the back of the pipette permits rapid, highly pre-
cise application of pressure and suction. It is often best to begin by applying positive pres-
sure gently with the pipette placed just at the surface of the slice and directly above the
targeted cell. Blowing accomplishes two goals. First, it provides information about the
consistency of the tissue, which may be rather variable from slice to slice and which differs
dramatically from one brain region to another. Second, application of positive pressure
begins the process of disrupting the tissue overlying the desired cell. When tissue has been
visibly disrupted, slight suction helps remove the loosened debris (Figure 5.2). Repeated
application of this blowing and sucking cycle, with appropriate movement of the pipette
to remove specific chunks of tissue, eventually results in a clean cell.
cleaning
pipette
brain tissue
slice
positive pressure
(blowing)
negative pressure
(sucking)
Use of a cleaning pipette to remove tissue overlying a cell in tissue slices. Repro-
duced with permission according to Edwards et al., (1989).
At first, this procedure may take a long time. It is almost inevitable that an inexperienced
experimenter, who has not yet developed a feel for the tissue, will kill cells by applying
pressure or suction that is too vigorous. With practice, however, one can clean even deep
cells quite rapidly. Careful and repeated adjustment of the focus aids greatly in determin-
ing the degree of cleaning required at each stage. Depending on the particular type of tis-
sue, a few tricks to facilitate and accelerate the process of producing clean cells may be
applied. For example, in regions of densely packed somata, such as the CA1 pyramidal cell
layer of the hippocampus, it is possible to remove tissue at the slice surface over a relatively
large length of the cell layer by applying continuous suction and moving the pipette along
the layer as if using a vacuum cleaner to remove dead tissue. Inspection of the underlying
neurons often reveals at least one clean pyramidal cell.
After completing the cleaning procedure, patch-clamp recordings of cleaned cells in slices
are carried out following essentially the same procedures that would be applied to cells in
culture. All four major patch-recording configurations—cell-attached, inside-out, out-
side-out and whole-cell—are possible, as well as the nystatin-perforated patch technique
(see Section 5.8., “Recording from Perforated Patches and Perforated Vesicles”).
The procedure for establishing gigohm seals is very simple (Figure 5.3). It is possible to
use either the current-clamp or voltage-clamp recording mode. A small repetitive current
or voltage pulse is applied to the electrode at relatively high frequency (e.g., 10 Hz) and
the voltage or current response is monitored with an oscilloscope. Continuous, slight pres-
sure is applied to the back of the pipette, for example, by blowing gently into the tubing
connected to the pipette and closing a stopcock to retain the pressure. The pipette is then
slowly lowered into the tissue in the region where the desired cells are found. It is
advanced slowly—1–5 steps per second, each step of 1–2 μm—into the tissue until a sud-
den increase in resistance is detected at the pipette tip, indicating contact with a cell.
Often a further advance of 1–2 μm causes a further increase in the resistance. Release of
the positive pressure should result in another approximate doubling of the resistance if the
electrode is indeed contacting a cell. Although gigohm seals sometimes form spontane-
ously at this stage, the next step is to apply light suction to cause the seal to form. This
may occur rapidly, in less than one second, or slowly, up to several seconds, requiring con-
stant, gentle suction. The patch is then ruptured by additional suction creating electrical
continuity between the pipette and cell interior.
Approach neuron
Light suction
Gigohm seal
Rupture patch
Obtaining a recording using the blind technique by advancing the patch pipette
through a tissue slice. According to the method of Blanton et al., (1989).
Although this technique can be applied successfully, achieving a good recording on nearly
every attempt, it is almost inevitable that at some point one will encounter some difficulty
either in generating seals or establishing whole-cell recordings. Through experience, it is
possible to identify the problem-causing aspects of the procedure. Often, changing the
pipette size helps. A general rule is to use smaller tips to improve sealing, and larger ones
to facilitate the transition to the whole-cell mode. One should be prepared to use a num-
ber of pipettes, since the probability of obtaining a recording on the second attempt to
apply suction to the same pipette is very small. Another critical variable, which applies to
all patch-clamp recording, is the intracellular solution whose osmolarity should be verified
routinely, particularly when difficulties are encountered.
• Need upright microscope and water- • Need no special equipment beyond that
immersion objective. for intracellular recording in slices.
• Once a cell is located, additional time is • Less preparation time for each
invested in cleaning. penetration, but no certainty of cell
health.
• Can see neighboring cells and apply • Visually-guided local stimulation is not
local stimuli. possible.
Coleman, P.A., Miller, R.F., Measurements of passive membrane parameters with whole-
cell recording from neurons in the intact amphibian retina. J. Neurophysiol. 61:218–230,
1989.
Edwards, F.A., Konnerth, A., Sakmann, B., Takahashi, T., A thin slice preparation for
patch clamp recordings from neurones of the mammalian central nervous system. Pflügers
Arch. 414:600–612, 1989.
Hestrin, S., Nicoll, R.A., Perkel, D.J., Sah, P., Analysis of excitatory synaptic action in the
rat hippocampus using whole cell recording from thin slices. J. Physiol., 422, 203–225,
1990.
Rall, W., Segev, I., Space clamp problems when voltage clamping branched neurons with
intracellular microelectrodes. T.G.J. Smith, H. Lecar, S.J. Redman, P.W. Gage (Eds.),
Voltage and Patch Clamping With Microelectrodes. Bethesda: American Physiological
Society, pp. 191–215, 1985.
Sakmann, B., Edwards, F., Konnerth, A., Takahashi, T., Patch clamp techniques used for
studying synaptic transmission in slices of mammalian brain. Q J Exp Physiol 74: 1107–
18, 1989.
Sakmann, B., and Stuart, G., Patch-pipette Recordings from the Soma, Dendrites, and
Axon of Neurons in Brain Slices in Single-Channel Recording, 2nd Ed., Sakmann, B.,
and Neher, E., eds. pp. 199–211, 1995.
Two approaches are used to record ionic currents from patches of membrane containing
tens to hundreds of ion channels. The first applies standard gigohm-seal (gigaseal) patch-
clamp methods employing large-diameter patch pipettes (5–10 μm tip diameter) and is
used to sample currents from much larger areas than are commonly used in single-channel
recording (which is done with 0.5–2 μm tip diameter pipettes). The second method of
macropatch recording has been termed loose-patch recording. This method, which
enables one to gather data that could not be obtained with the gigaseal method, differs
from single-channel recording in three ways:
1 The tip of the loose-patch pipette is usually much larger than the conventional single-
channel patch pipette;
2 The loose-patch pipette does not form a gigaseal with the membrane, whereas gigohm
seals are required for single-channel recording; and
3 The loose-patch pipette can be reused and repositioned to map the spatial distribution
of channel subtypes on individual cells (Thompson and Coombs, 1988).
There are a few problems associated with this method that have to be considered in the
experimental design. The patch-clamp amplifier must be able to fully compensate for the
greater capacitance of larger patches. When voltage pulses are applied to macropatches in
the cell-attached configuration, current flowing in the patch will change the cell voltage;
this change can be significant. For example, with a macropatch current of 100 pA and a
cell input resistance of 100 MΩ, the cell voltage will change by 10 mV. Dynamic current-
dependent errors of this type are difficult to correct and are most severe in small cells. In
some preparations it may be impossible to use cell-attached macropatches to obtain accu-
rate I/V curves, especially for rapidly changing currents. One can eliminate this source of
error by studying excised patches.
Milton & Caldwell (1990) identified a more subtle error associated with macropatch
recording. They observed that the suction used to form the seal between the patch pipette
and the membrane can cause a change in ion channel density in the patch. With large
patch pipettes, a noticeable membrane bleb may form in the orifice of the pipette during
suction. This membrane bleb can be visualized under a microscope. It appears that the
ion-channel density decreases during bleb formation, possibly because of a flow of lipid
into the bleb causing membrane rearrangement and changes in channel distribution.
Therefore, the gigaseal-macropatch method may be unsuitable for mapping ion-channel
densities because the procedure modifies the channel density before the measurement is
made.
1 The pipette can be repositioned to sample currents from a number of patches on the
same cell; and
2 The loose-patch method can be used to measure local currents in voltage-clamped
preparations where establishing gigaseals is impossible due to the presence of adherent
connective tissue, basement membrane or glial cells. Note that loose-patch recording
is done only in the cell-attached configuration.
The low seal resistance between the pipette and the membrane presents the principal tech-
nical difficulty one must overcome in applying loose-patch methods. Low seal resistance
creates two types of problems. First, the current flowing through the seal resistance and
capacitance introduces noise in the current record, thus limiting the resolution (see Chap-
ter 3, “The Importance of a Good Seal”). A more significant problem arises from the fact
that membrane current flowing through the seal resistance is lost.
The following discussion will focus on the errors introduced by low seal resistance and the
procedures used to limit errors to acceptable levels. Three designs are available to combat
the problems associated with low seal resistance in loose-patch recording. The choice of
the most appropriate method depends on the nature of the preparation. The original liter-
ature should be consulted for further details.
Combining Whole-Cell Clamp with Focal Current Recording
The first method to be introduced combines whole-cell voltage clamping with loose-patch
current recording. This is the preferred method for many applications. The cell is voltage
clamped by a microelectrode or a whole-cell patch pipette using standard microelectrode
voltage-clamp or patch-clamp instrumentation, while current is recorded from a restricted
area of membrane with a loose-patch pipette using patch-clamp instrumentation. Voltage-
clamp pulses are applied via the whole-cell clamp while the loose-patch pipette is main-
tained at the bath potential to ensure that no current flows across the seal resistance. This
method has been used to record local currents in large molluscan neurons (Neher and
Lux, 1969; Johnson and Thompson, 1989; Thompson and Coombs 1989; Premack et
al., 1990) and in muscle cells (Almers et al., 1983). The method is particularly well suited
for measuring the kinetics and current/voltage relationships of macroscopic currents
expressed in Xenopus oocytes where a whole-cell voltage clamp is severely limited by the
large capacitance of the cell and high intracellular series resistance. The electrode arrange-
ment is shown in Figure 5.4.
Vc Vo
A
Patch Clamp
Im Amplifier
Ip
Cc
Rp
R ei R sl
R sr
Bath Clamp
C mp Amplifier
Figure 5.4: Combining whole-cell voltage clamping with loose-patch current recording.
Schematic diagram of a cell with voltage clamp electrodes and the large patch pipette in
place. The symbols used in the figure are as follows: A, voltage clamp open loop gain; Cc,
coupling capacitance between voltage clamp current electrode and patch electrode; Cmp,
capacitance of membrane patch; Ic, current error; Im, membrane current supplied by volt-
age clamp; Ip, patch membrane current supplied by patch clamp amplifier; Rei, resistance
of voltage clamp current electrode; Rp, axial resistance of patch electrode; Rsl, resistance of
seal between patch pipette and cell membrane; Rsr, series resistance between the outside of
the patch electrode and the bath clamp amplifier; Vc, voltage clamp command voltage;
Vo, voltage clamp output voltage. Reproduced with permission according to Johnson and
Thompson (1989).
This method allows one to simultaneously measure whole-cell and patch currents as well
as whole-cell and patch capacitances. The areas of the cell and of the patch can be esti-
mated by measuring the whole-cell and patch capacitances. A convenient way to do this is
to apply a triangle-wave voltage command to the whole-cell voltage clamp and record the
resulting whole-cell and patch current waveforms. The patch capacitance (c) is calculated
from the change in the time derivative of the voltage (dV/dt) at a vertex of the triangle
wave and the corresponding jump in patch current (I), using the following equation
(Neher, 1971; Palti, 1971):
I
c= (2)
dV dt
Because the voltage of the patch pipette is equal to the bath voltage and only the voltage
across the membrane patch covered by the pipette is changing, the capacitance of the
patch pipette itself is rejected. The capacitance of the whole cell is measured in a similar
fashion. However, the interpretation is complicated by current contributed from poorly
clamped regions (see Johnson and Thompson (1989) for further discussion).
There are several issues to consider when applying this combined voltage-clamp method.
The most important requirement is that the patch current must be measured from a well-
clamped region of membrane, i.e., from a region that is “space clamped” by the whole-cell
voltage clamp amplifier. In order to accurately measure current-voltage relationships and
limit the capacitive current flowing across the patch during voltage steps, the loose-patch
pipette should be positioned close to the point where the whole-cell voltage is controlled.
Cell geometry and electrode positioning should be considered when designing the physi-
cal layout of the experiment. These considerations limit the kinds of preparations that can
be studied with the method. It is reasonable to use the loose-patch pipette to record local
current densities at different points on a well-clamped spherical cell body. However, it
would be inappropriate to record patch currents from a thin dendrite when the voltage
clamp is applied to the soma.
The most significant error results from current flowing in the seal resistance between the
loose-patch pipette and the bath. This error current (Ie), which is not measured by the
patch-clamp amplifier, is given by:
I p Rp
Ie = (3)
R p + Rsl
where Rp is the resistance of the loose-patch pipette and Rsl is the seal resistance. If
(Rp + Rsl) is ten times larger than Rp the error in the measured current is 10%. To mini-
mize the error, one can reduce Rp by using a patch pipette with a blunt taper and increase
Rsl by using suction to improve the seal with the membrane. Enzyme treatment with dis-
pase, pronase or trypsin can be used to clean the cell surface and improve the seal resis-
tance.
1 The voltage drop that results from macroscopic patch currents flowing through the
axial resistance of the patch pipette (Rp) will produce an insignificant error in most
cases. For example, if the value of Rp is 300 kΩ and the maximal patch current is
500 pA, the resulting voltage error is 0.15 mV.
2 A more troublesome error occurs when the bath voltage outside the patch pipette
changes due to large whole-cell currents flowing in the series resistance (Rsr) between
the cell membrane and the bath ground or virtual ground.
The error current is given by:
I m Rsr
Ie = (4)
Rsl
where Rsr is the sum of the series resistances of the bath, the agar bridges and the liquid-
silver chloride junction at the bath electrode; it typically has a value of several kilohms.
Bath series resistance creates an insidious problem when membrane currents are large.
However, the error can be reduced by measuring the bath voltage with a separate electrode
and using it as the reference voltage at the patch-clamp amplifier. Good practice dictates
that the bath voltage be used to correct the membrane voltage signal applied to the whole-
cell clamp amplifier as well. Instability can result, however, if the bath voltage is measured
with a greater bandwidth than the membrane voltage. To remedy this, one can limit the
frequency response of the bath voltage follower, but this degrades performance and causes
other errors. A better solution is to eliminate the need to subtract the bath voltage at the
other circuits by voltage clamping the bath to ground potential. This can be accomplished
using an independent voltage-clamp amplifier operating at a high gain and bandwidth.
Two separate agar bridges and silver-chloride electrodes can be used to measure the bath
voltage and pass current to the bath. Because of the low resistance of these electrodes, a
simple voltage-clamp circuit employing a single operational amplifier is adequate for
clamping the bath to ground potential. This arrangement ensures that the bath voltage
does not change during clamp steps. The output of the circuit provides the record of
whole-cell current.
Other problems can be overcome by a careful design of the experimental layout. Since this
method uses several voltage-clamp amplifiers working simultaneously, it is necessary to
limit capacitive coupling between electrodes. This can be achieved by inserting a
grounded shield between the whole-cell current electrode and the macropatch pipette,
and by lowering the volume of saline to the minimum required for covering the cell. To
reduce high frequency noise in the patch recording, the submerged tip of the pipette
should be coated with Sylgard (Dow Corning, Midland, MI) and only the tip should be
filled with saline. Johnson and Thompson (1989) provide a thorough analysis of the prob-
lems and the approaches used to reduce errors to acceptable levels.
Although extra effort is required with this method, it may be the best approach for some
preparations. It is important to remember that current flowing across a macropatch can
cause a significant change in cell voltage. In its simplest form, this method uses a large
diameter patch pipette in a cell-attached configuration and a conventional patch-clamp
amplifier (Almers et al., 1983a,b; Stühmer et al., 1983). In this configuration, the axial
resistance of the pipette (Rp) and the seal resistance (Rsl) act as a voltage divider. When a
voltage command pulse is applied to the pipette (Vp), it will be attenuated at the pipette
tip. The voltage at the pipette tip (Vt) is given by (Stühmer et al., 1983):
Rsl
Vt = V p (5)
R p + Rsl
When Rsl is large, as is the case in gigaseal recordings, the error is insignificant. In loose-
patch applications, however, Rp and Rsl are approximately equal and a large fraction of the
current delivered to the pipette will flow through the seal resistance. It is necessary to cor-
rect the amplitude of command-voltage steps by a factor equal to Rsl/(Rp+Rsl). To do so,
Rp is measured before the pipette contacts the cell and Rp + Rsl is monitored continually
throughout the experiment. Almers et al., (1983a,b) used a computer to calculate a cor-
rection factor A = Rsl/(Rp+Rsl) before each voltage step. The correction factor was used to
scale command-voltage steps applied by the computer by 1/A to ensure that the intended
voltage appeared at the pipette tip. Recorded currents were divided by A to correct for
current (often termed leakage current) flowing through the seal resistance. This digital
approach to seal resistance compensation is technically sound. To implement it, the user
has to develop a software program for continuously updating the parameters needed for
scaling the amplitude of the command pulses and correct for leakage current; an extra
effort that is well justified in some applications.
Seal resistance compensation can also be achieved with an analog circuit. Stühmer et al.,
(1983) pioneered a method that uses an active voltage divider and potentiometers to com-
pensate for measured values of Rp and Rsl. Their circuit allows one to null Rp and limit
the series resistance errors to acceptable values. When correctly adjusted, their design
ensures that the input impedance of the patch pipette is near zero so that there is minimal
loss of membrane current across the seal. A disadvantage of this method is that it requires
the user to build a specialized patch-clamp headstage and other circuitry following the
example provided by Stühmer et al., (1983).
With both digital and analog seal-resistance compensation, accuracy depends on one’s
ability to measure Rp and (Rsl + Rp) and on the assumption that these resistances remain
constant during voltage steps that activate conductances in the patch. The error increases
as the amplitude of the voltage step and the conductance of the membrane patch increase.
It is not certain that Rp and Rsl are independent of the current amplitude. In practice, one
often observes an unexpected flattening of the current-voltage relationship for inward or
outward current activation when the voltage step or the patch current become large.
Because most of the uncertainty in current measurement depends on Rsl, one should use
suction and/or enzyme treatment to maximize the seal. It also helps to limit the diameter
of the patch pipette to about 20 μm or less. Rp is minimized by using pipettes with a
blunt taper.
Loose-Patch Clamp with Concentric Patch Pipettes
Almers et al., (1984) introduced a modification of the loose-patch method that employs a
two-chambered concentric patch electrode. The electrode is fashioned from two glass cap-
illaries, one inside the other, with the inner capillary held in place by support rods during
the pull. The concentric electrode is pushed against the cell surface where it separates a
central membrane patch from an annular surround. The voltage-clamp amplifier is
designed to clamp both the center patch and the surround to the command potential.
Membrane current is measured in the central barrel of the electrode while the outer annu-
lus acts as a guard. Since the inner and outer barrels are at the same potential during the
command steps, current loss from the central membrane region to ground is minimized
and Rsl takes a value that is effectively infinite. This approach is historically related to
vaseline-gap and sucrose-gap voltage-clamp methods. It has been used successfully to volt-
age clamp mammalian skeletal muscle. However, the implementation is somewhat diffi-
cult since it requires the construction of an elaborate concentric electrode and some
specialized instrumentation. It may still be the best approach in some cases and the origi-
nal literature should be consulted for further details.
References
Almers, W., Stanfield, P.R., Stühmer, W., Lateral distribution of sodium and potassium
channels in frog skeletal muscle: measurements with a patch-clamp technique. J.Physiol.
336, 261–284, 1983a.
Almers, W., Stanfield, P.R., Stühmer, W., Slow changes in currents through sodium chan-
nels in frog muscle membrane. J. Physiol. 339, 253–271, 1983b.
Almers, W., Roberts, W.M., Ruff, R.L., Voltage clamp of rat and human skeletal muscle:
measurements with an improved loose-patch technique. J.Physiol. 347, 751–768, 1984.
Hoshi, T., Zagotta, W.N., Aldrich R.W., Biophysical and molecular mechanisms of
Shaker potassium channel inactivation. Science 250, 533–538, 1990.
Johnson, J.W, Thompson, S., Measurement of non-uniform current density and current
kinetics in Aplysia neurons using a large patch method. Biophys. J. 55, 299–308, 1989.
Milton, R.L., Caldwell, J.H., Na current in membrane blebs: Implications for channel
mobility and patch clamp recording. J. Neurosci. 10, 885–893, 1990.
Neher, E., Two fast transient current components during voltage clamp on snail neurons.
J.Gen.Physiol. 58, 36–53, 1971.
Neher, E., Lux, H.D., Properties of somatic membrane patches of snail neurons under
voltage clamp. Pflügers Arch. 322, 35–38, 1971.
Palti, Y., Varying potential control voltage clamp on axons. in Biophysics and Physiology
of Excitable Membranes. W.F. Adelman, Jr., Ed. Von Nostrand Reinhold Co. pp. 194–
205, 1971.
Premack, B.A., Thompson, S., Coombs-Hahn, J., Clustered distribution and variability
in kinetics of transient K channels in molluscan neuron cell bodies. J. Neurosci. 9, 4089–
4099, 1990.
Stühmer, W., Roberts, W.M., Almers, W., The loose patch clamp. in Single-Channel
Recording. Sakmann, B., Neher, E., Eds. pp. 123–132, 1983.
The giant excised membrane patch technique enables one to achieve 1–10 gigohm seals
routinely, using pipette tips with inner diameters of 12–40 μm. The giant membrane
patch produced by this method has 2–15 pF membrane capacitance, representing a 100-
fold increase of membrane area over the area of conventional, single-channel patch mem-
brane. Pipette diameters of > 50% of the cell diameter can be employed when recording
from small, spherical cells.
The giant patch method has proven to be advantageous for fast macroscopic current
recording whenever free access to a large surface area of the cytoplasmic membrane is
desired. To date, the giant patch method has been applied successfully to cardiac, skeletal
and tracheal myocytes, pancreas acinus cells, a Jurkat human T-cell line, Sf9 cells and
Xenopus oocytes. The method should be useful for a wide range of studies of electrogenic
mechanisms involving native proteins and proteins expressed from cloned genes. Promis-
ing applications include studies of macroscopic current modulation by intracellular com-
pound (e.g., enzymes or ions acting from the cytoplasmic side of the membrane), studies
of charge movement generated during channel gating, and studies of partial reactions of
transporters.
Alternatively, pipettes can be pulled to smaller diameters and then scratched gently on the
edge of a soft glass bead to the desired diameter, leaving a rough edge (pipette B in
Figure 5.5, 17 μm diameter). When such electrodes are employed without fire polishing,
the membrane usually remains localized at the pipette opening during sealing, and vesicles
are obtained in most cases upon membrane excision. For very large cells, particularly for
oocytes, a different procedure is employed. Pipettes are first scratched to a 30–50 μm
diameter (pipette C in Figure 5.5) and then melted to a diameter of 18–35 μm (pipette D
Figure 5.5) close to a glass bead placed on a microforge. The resulting bullet-shaped
pipette tips are advantageous for fast voltage clamping, and the oocyte membrane remains
localized to the rim of the pipettes. We have had good success preparing electrodes by
melting the tip into a bead of soft glass, followed by cooling and “cutting” by a small pull.
smeared with a plastic rod on a clean surface, and a thin film of the mixture is drawn into
air by lifting the rod. The pipette tip is then passed quickly through the film, and the pro-
cedure is repeated once more. In order to reduce the pipette capacitance, a thick mixture is
applied as close as possible to the tip. The pipette is then backfilled and a brief, strong,
positive-pressure pulse is applied with a syringe to clear the tip. The pipette tip usually
appears clean and is visually indistinguishable from that of an untreated electrode.
To date, the best success with oocytes has been obtained by shrinking oocytes with
200 mM K-aspartate rather than sucrose. Patches from oocytes are excised by gently mov-
ing the pipette tip from side to side in gradually increasing distances. Dark, granular cyto-
plasmic material is often excised with the oocyte patches, and vesicles are obtained only
rarely. The cytoplasmic material can be washed away by rapid pulses of solution directed
at the patch.
Hilgemann, D. W., Giant excised cardiac sarcolemmal membrane patches: sodium and
sodium-calcium exchange currents. Pflügers Archiv. 415, 247–249, 1989.
Hilgemann, D.W., The Giant Membrane Patch in Single-Channel Recording, 2nd Ed.,
Sakmann, B., Neher, E., eds. pp. 307–326, 1995.
Standen, N. B., Stanfield, P. R., Ward, T. A., Wilson, S. W. A new preparation for record-
ing single-channel currents from skeletal muscle. Proceedings of the Royal Society of Lon-
don (Biology) B221, 455–464, 1984.
hours employing the perforated-patch approach, it is advisable to use stock and antibiotic-
containing filling solutions within one hour of preparation.
Both amphotericin and nystatin seem to interfere with seal formation. Although it is pos-
sible to seal some cells when either compound is present in the pipette filling solution, the
success rate is lower than when an antibiotic-free filling solution is used with the same
cells. Therefore, it is best to fill the tip of the pipette with an antibiotic-free solution and
then to backfill with the antibiotic-containing solution. The tip-filling process can be
accomplished by simply dipping the tip into the desired solution. With borosilicate
glasses, 1–5 seconds of tip immersion suffices, whereas with the more hydrophobic lead-
containing glasses, 30–60 seconds may be required. With high-lead glasses, gentle suction
for a second or so may be beneficial. With any glass, it is important that the antibiotic-free
solution not exceed 500 μm distance from the tip; otherwise, the time required for the
diffusion of the antibiotic from the backfilling solution to the tip will be excessive. For
example, if the pipette is filled up to 1.0 mm from the tip with an antibiotic-free solution,
more than an hour will be required for the antibiotic to reach a tip-concentration compa-
rable to that in the backfilling solution. Ideally, the tip should be filled to a distance that
allows the antibiotic to reach the tip within a few seconds following a gigohm seal forma-
tion. Since the time required for seal formation depends on the cells and the setup, tip fill-
ing distances should be optimized by each investigator. Consistently optimal tip filling is
routinely possible by first overfilling the tip and then forcing the excess solution out by
applying positive pressure to the back of the pipette while observing it under a dissecting
microscope.
In a real experiment, one can observe the partitioning by applying a voltage pulse to the
pipette at frequent intervals. As perforation occurs, the size of the capacity transient
increases and the time constant of the transient decreases. The time constant to charge the
cell capacitance is approximately RaCm, where Ra is the access resistance and Cm is the cell
capacitance. When Ra is large, a brief voltage pulse will not charge the capacitance fully
before the pulse turns off. Therefore, as partitioning continues and Ra decreases, the
capacity transient gets larger since an increasing fraction of the cell capacity gets charged
during the brief pulse. The decaying phase of the transient gets shorter as Ra decreases. In
most cells, a properly filled pipette will produce minimized Ra by 20–30 minutes after seal
formation. Ra of less than 20 MΩ is often achieved in about 5 minutes. Some investiga-
tors begin recording at this time. However, Ra continues to fall and finally becomes stable
after 20–30 minutes. Once Ra stabilizes, it remains stable for 2.5–3 hours before begin-
ning to increase (presumably due to loss of activity of the antibiotic). The remarkable sta-
bility of Ra is one of the main virtues of the technique. For example, using amphotericin,
one investigator reported Ra of 3.4 MΩ that did not change by more than 100 kΩ for
three hours.
1 The perforated-patch approach does not allow one to dialyze the cell’s cytoplasm and
replace its content with compounds other than the small ions included in the filling
1 The largest pipette tips compatible with seal formation should be used to minimize
the pipette resistance.
2 Pipette geometry should maximize the size of the omega-shaped membrane patch that
is drawn into the tip during seal formation. This maximizes the surface area available
for channel insertion.
3 The perforating compound that produces the highest total conductance should be
used. It may not be simple to predict the conductance-producing ability of a
compound since it requires both the partitioning (number of channels formed) and
single-channel conductance to be maximal. To date, amphotericin B has produced the
lowest Ra. This can be anticipated since the single-channel conductance of
amphotericin channels is twice that of nystatin, at least in measurements made in salts
at molar concentration. However, the relative conductance of the two types of
channels in lower salt concentrations (e.g., 150 mM) is unknown. In addition, there
are no relevant studies of the relative partitioning of the two compounds when
presented to the membrane in saturating concentrations. To date, the lowest patch
resistance obtained by perforating with amphotericin is about 2 MΩ. The lowest
resistance pipette that is capable of forming gigohm seals is approximately 0.5 MΩ.
Thus, at present, an Ra value of approximately 2.5 MΩ appears to be the lowest one
can achieve. Lower Ra values will require either perforating compounds with higher
single-channel conductance or finding ways to enhance the partitioning of the
existing antibiotics. It is unlikely that larger pipettes than those presently used would
be successful in forming seals.
1 The extent to which the cell’s voltage is altered by current flow through the shunt
resistance along the electrode’s outer surface at either the penetration or seal sites; and
2 The extent to which the pipette-filling solution alters the electrolyte concentrations in
the cell cytoplasm. High shunt resistances and minimal alteration of the cell content
are desirable.
Patch electrodes, with their tens to hundreds of gigohm seal resistances, clearly have much
higher shunt resistances than do intracellular microelectrodes. On the other hand, patch
electrodes sealed to cells cause rapid changes in the concentrations of the cellular electro-
lytes if the pipette-filling solution contains different electrolyte concentrations than the
cytoplasm. Intracellular microelectrodes, which are usually filled with molar concentra-
tions of potassium salts, also alter cellular electrolytes but not as rapidly as do patch elec-
trodes. Therefore, the perforated-patch technique provides both the advantage of the
higher shunt resistance than intracellular microelectrodes without the significant change
in cytoplasmic content caused by conventional patch electrodes.
Single-Channel Recording in Outside-Out Vesicles
The outside-out patch configuration has been particularly useful because it allows direct
application of agents to the extracellular side of the membrane while maintaining com-
plete voltage control during single channel recording. These patches have been required
for studying ligand-gated ion channels such as the nicotinic acetylcholine, GABAa, gly-
cine and glutamate receptors. Likewise, transmitters that regulate channels through non-
diffusible second messengers must be examined with outside-out patches. For example,
activation of atrial potassium channels is not seen with cell-attached patches when musca-
rinic agonists are applied to the bath, but is seen with direct application of transmitter to
an outside-out patch.
Similar to the protocol described above, nystatin stock solution (25 mg/mL) is prepared
by dissolving 2 mg nystatin in 80 μL DMSO. Since this pore-forming agent is susceptible
to oxidation, the stock solution can be frozen and used within two days. Simple pipette
solutions can be used (e.g., 150 mM KCl, 5 mM MgCl2, 10 mM KHepes, pH 7.1)
because nucleotides, proteins and calcium buffers cannot permeate through nystatin per-
forations. Four (4) μL of nystatin stock solution is pipetted into a glass tube and then
1 mL of filtered pipette solution is added. After covering the tube and vortexing, the
slightly cloudy solution is vigorously sonicated for 30 seconds in a cylindrical bath sonica-
tor to completely dissolve the nystatin, giving a final concentration of 100 μg/mL. This
aqueous solution can be used for only 3 hours.
Conventional patch pipettes are fabricated. When filled with pipette solution, their resis-
tance is 2–4 MΩ. Smaller pipettes minimize the surface area of the perforated patch and,
thus, give larger series resistances (Rs). On the other hand, larger pipette tips were found
to form fragile patches that could spontaneously break forming the conventional whole-
cell configuration. Pipettes are coated with Sylgard to reduce their capacitance. This not
only lowers noise for single-channel recording, but also permits accurate series resistance
measurements.
Pipettes are momentarily dipped in nystatin-free pipette solution and then backfilled with
nystatin-containing solution. A cell-attached seal is then formed. After compensating,
electrode capacitance (e.g., 3 pF), Rs is monitored by examining the unfiltered current
response to a 10 mV voltage step from -70 mV (holding potential). The height of the
capacitance transient is equal to the voltage step divided by Rs. Series resistance is moni-
tored for 5–20 minutes to ensure that it decreases smoothly to a value less than 50 MΩ.
Cells that show an abrupt decrease in Rs should be discarded.
Nystatin Pipette
perforations
Receptor
withdraw
pipette
Ion channel
Cytoplasm
Perforated
vesicle
Cell-attached
perforated patch
The formation of a perforated vesicle can be verified by many methods. In some cases,
vesicles can be seen with 400x magnification. The presence of cytoplasm can be con-
firmed by preloading the cells with a fluorescent dye. For example, cells may be treated
with 40 μM carboxyfluorescein diacetate, a membrane-permeant nonfluorescent com-
pound. After crossing the membrane, this agent is hydrolyzed to yield water-soluble car-
boxyfluorescein. The presence of organelles can also be verified with specific dyes.
Rhodamine 123 specifically partitions into functioning mitochondria and produces a
characteristic punctate staining that can be detected in GH3 cells and perforated vesicles.
Channel activity can also verify the presence of nystatin perforations. With normal vesi-
cles, channel currents appear attenuated and rounded. Furthermore, channels in the inner
face of the vesicle are not accessible to bath-applied hydrophilic inhibitors and have oppo-
site voltage dependence to channels in the vesicle surface that is in contact with the bath.
In contrast, perforated-vesicle channel activity can be recorded only from the portion of
the vesicle in contact with the bath. Through the use of fluorescent dyes and channel
recordings, the perforated vesicle can be differentiated from normal excised vesicles and
outside-out patches.
The nystatin-perforated vesicle method allows the study of channel activity and modula-
tion that is lost with conventional outside-out patches. Initial studies indicate that cyclic
AMP-dependent protein kinase, phosphodiesterase, phosphatase, G-proteins, phospholi-
pase C and intracellular calcium stores can modulate ion channels in perforated vesicles
excised from GH3 cells. Two channel types that quickly “wash out” of conventional
patches from these cells, L-type calcium channels and voltage-gated potassium channels,
survive for extended periods with perforated vesicles. Perforated vesicles were excised suc-
cessfully from a variety of cell lines and primary cultured pituitary cells. Thus, it appears
likely that this patch clamp configuration will aid the study of ion channel modulation in
many systems.
Finkelstein, A., Water movement through lipid bilayers, pores, and plasma membranes:
theory and reality., Vol. 4, Wiley-Interscience, New York, p. 127, 1987.
Holz, R., Finkelstein, A., The water and nonelectrolyte permeability induced in thin lipid
membranes by the polyene antibiotics nystatin and amphotericin B. J. Gen. Physiol., 56:
125–145, 1970.
Horn, R., Marty, A., Muscarinic activation of ionic currents measured by a new whole-cell
recording method. J. Gen. Physiol., 92: 145–159, 1988.
Korn, S.J., Bolden, A., Horn, R., Control of action potential duration and calcium influx
by the calcium-dependent chloride current in AtT-20 pituitary cells. J. Physiol. (Lond),
439: 423–437, 1991.
Korn, S.J., Horn, R., Influence of sodium-calcium exchange on calcium current rundown
and the duration of calcium-dependent chloride currents in pituitary cells, studied with
whole cell and perforated patch recording. J. Gen. Physiol., 94: 789–812, 1989.
Korn, S.J., Marty, A., Connor, J.A., Horn, R., Perforated patch recording. Methods in
Neuroscience 4: 264–373, 1991.
Kurachi, Y., Asano, Y., Takikawa, R., Sugimoto, T., Cardiac Ca current does not run
down and is very sensitive to isoprenaline in the nystatin-method of whole cell recording.
Archiv. Pharm., 340: 219–222, 1990.
Levitan, E.S., Kramer, R.H., Neuropeptide modulation of single calcium and potassium
channels detected with a new patch clamp configuration. Nature, 348: 545–547, 1990.
Lindau, M., Fernandez, J.M., IgE-mediated degranulation of mast cells does not require
opening of ion channels. Nature, 319: 150–153, 1986.
Rae, J., Cooper, K., Gates, G., Watsky, M., Low access resistance perforated patch record-
ings using amphotericin B. J. Neurosci. Methods., 37: 15–26, 1991.
en2
S i2(f ) = (6)
X c2
where en is the rms noise voltage (V/√Hz) and Xc is the capacitive reactance of the bilayer.
The capacitive reactance is 1/(2αCbf ), where f is the frequency and Cb is the bilayer
capacitance. Therefore:
(7)
S i2( f ) = en2 (2πCb f ) 2
The spectral density of the current noise produced by voltage clamping the thermal volt-
age noise in the access resistance Ra across the bilayer capacitance is:
(8)
S i2(f ) = 4kT Re{Y (f )}
where Re{Y(f )} is the real part of the admittance of Ra in series with Cb, and is given by
α2
Re{Y (f )} = (9)
Ra (1 + α 2 )
with α = 2πfRaCb. For a constant bandwidth, at the lower limit of Ra (small α),
Re{Y(f )} = Ra(2πCbf )2, and equations (7) and (8) have a similar form. At the upper limit
of Ra (large α), Re{Y(f )} = 1/Ra. The transition from a linear to a reciprocal dependence
on Ra produces maxima in plots of this noise component against Ra (see Figure 2A of
Wonderlin et al., 1990).
Strategies for Minimizing the Background Current Noise
Reduce the Bilayer Capacitance. The single most effective approach to reducing the
background current noise is to reduce the bilayer diameter and hence its capacitance.
This assumes that the access resistance is not increased (see Wonderlin et al., 1990
and “Choosing a Recording Configuration” on page 137) for a discussion of open
chamber versus pipette/bilayer configurations). There is, however, a practical limit to
the extent of reduction as the bilayer diameter must be large enough both to allow
easy visualization and physical manipulation and to ensure an acceptable rate of
channel incorporation. Bilayer diameters in the 40–60 μm range (having 10–25 pF
Cb at 0.8 μF/cm2) represent a reasonable compromise between the competing needs
of reducing the capacitance versus ease of channel incorporation and manipulation. A
very good method for making these small bilayers is the shaved-aperture method,
which is described in detail below (Assembling a Bilayer Setup for High-Resolution
Recording: Making a Small Aperture).
Use an Amplifier with Low Internal Noise Voltage. Noise voltages are present in the
FET input of the headstage and in the circuitry with which voltage command inputs
are buffered and applied to the bilayer. Reduction of the noise voltage in these inter-
nal sources is difficult for the investigator to control without being involved in the
design and construction of the amplifier. However, when selecting an amplifier, the
magnitude of the internal noise voltage sources can be quickly but qualitatively deter-
mined by comparing the noise in the current output before and after connection of
the amplifier input through a bilayer-sized capacitor to ground. This test should be
done without the addition of any resistance in series with the test capacitor to avoid
confusing an internal noise voltage source with voltage noise in the access resistance.
It should be noted that amplifiers which exhibit low noise when used for patch
recording (where the input capacitance is small and the background noise is domi-
nated by thermal noise currents in the feedback resistor and seal resistance) can per-
form poorly when the input capacitance is increased from patch to bilayer values,
because of high internal voltage noise.
Minimize the Access Resistance. Even with perfect, noise-free electronics, considerable
background current noise can be generated when the thermal voltage noise in the
access resistance is voltage clamped across the bilayer capacitance. This source can
account for as much as one half the noise under many bilayer recording conditions
(Wonderlin et al., 1990). The access resistance responsible for generating the voltage
noise is composed primarily of the resistance of the solution within the aperture in
which the bilayer is formed and the convergence resistances between each end of the
aperture and the bulk solutions (see Wonderlin et al., 1990). Because the resistance of
these solutions is inversely related to the salt concentration, it may be possible to
reduce the intra-aperture and convergence resistances by increasing the salt concen-
tration, within the limits of the experimental design. Furthermore, the shaved-aper-
ture method can be used to minimize the length of the aperture and, therefore, its
resistance. The resistance of the bulk solution should not contribute appreciably to
the access resistance when an open chamber design is used; but it can be an impor-
tant factor if large bilayers are made on the tips of glass pipettes (Wonderlin et al.,
1990). Finally, if salt bridges are used to connect the Ag/AgCl pellets to each bath,
they should be designed with a low resistance in mind, such as by using relatively
large diameter (e.g., 2 mm) glass capillaries filled with a high-salt solution such as
3 M KCl.
Type of Solvent in a Painted Bilayer. We have found that the type of solvent present in
a painted bilayer can affect the background current noise. If a bilayer is painted from
phospholipids dissolved in decane, the background noise can often be reduced by
rewiping the bilayer with hexadecane alone. The reason for the improvement is not
known and appears somewhat paradoxical, since it is usually associated with a slightly
larger bilayer capacitance.
Maximizing the Bandwidth
Minimization of the background noise currents by one or more of the methods described
above may help increase the usable bandwidth by decreasing the amount of low-pass fil-
tering required to clearly discern single channel gating. If the low-pass filter cutoff (-3 dB)
frequency is increased to several kilohertz or more, a second, less obvious source of band-
width limitation may come into play. This limitation is the decreased efficiency of the
high-frequency boost circuit used in resistive-headstage amplifiers due to loading the
headstage input with large bilayer capacitance. The boost circuit is designed to restore the
bandwidth that is lost because of low-pass filtering (< 100 Hz) by the parallel combina-
tion of the feedback resistor and its stray capacitance. The high-frequency boost circuit
adds a scaled derivative of the headstage current to the raw headstage current, thereby
increasing the headstage bandwidth. For patch recording, a boost circuit can work well
because of the dominant role of fixed electrical elements, such as the feedback resistance
and stray feedback capacitance, whose properties can be optimally compensated in the
selection of circuit components. When a resistive headstage amplifier is used for bilayer
recordings, however, the loss of bandwidth due to the variable, and frequently large,
bilayer capacitance must also be restored. This is much more difficult because the optimal
amount of high-frequency boost that is required will change in parallel with any change in
bilayer capacitance. Therefore, it is difficult to select components in the design of the
amplifier that enable optimal boosting across a large range of bilayer capacitance values.
Furthermore, because the bilayer capacitance will often vary by 50% within a day’s exper-
iments, it may be very difficult to maintain optimal boost adjustment.
Strategies for Maximizing the Bandwidth
Use a Capacitive-Headstage Patch Amplifier. An integrating-headstage patch-clamp
amplifier (e.g., the Axopatch 200B) measures the patch current as the integral of the
current flow across a feedback capacitor, rather than as the voltage drop across a feed-
back resistor. A boost circuit is not required for an integrating-headstage amplifier
because the feedback capacitor does not produce any significant low-pass filtering of
the patch current. Integrating headstage amplifiers therefore provide a broader
recording bandwidth that is also independent of the size of the bilayer capacitance.
Qb = V step C b (10)
Qb R f
Charging time = (11)
V max
where Vmax is the maximum voltage that can be applied across the feedback resistor Rf.
For example, a headstage with a 50 GΩ feedback resistor and a Vmax of 10 V can inject
charge at a maximum rate of 200 pA, requiring 25 ms to change the voltage across a
100 pF bilayer by 50 mV. During this 25 ms period, the output of the amplifier will be
saturated (at Vmax) and the bilayer will not be voltage clamped. This example illustrates
that the primary difficulty associated with applying voltage steps to a bilayer is the
requirement for rapid movement of a large amount of charge across the bilayer. For tradi-
tional resistive-headstage patch amplifiers this is a nearly insurmountable problem because
of the very slow rate at which charge can be delivered to the bilayer capacitance through
the large feedback resistance.
A secondary problem is that planar bilayers do not behave as ideal capacitors. The applica-
tion of a voltage step to a planar bilayer compresses the bilayer (i.e., causes electrostric-
tion). This electrostrictive force can change the relative areas of the bilayer and
surrounding annulus, producing a concomitant change in the capacitance of the bilayer.
During the period of time that the bilayer capacitance is changing, a capacitive current
will flow across the bilayer. Electrostrictive changes in bilayer capacitance are relatively
slow (milliseconds to seconds) and rarely large enough to produce capacitive currents that
obscure single-channel gating. However, the electrostrictive capacitive current tends to be
variable in magnitude and kinetics and it can change significantly as a bilayer “ages” dur-
ing the course of an experiment. This variability can make it difficult to subtract the elec-
trostrictive component of the capacitive current from the current record.
Strategies for Resolving Voltage Steps Across Bilayers
Reduce the Size of the Bilayer and Annulus. An essential first step is to make small
bilayers, because reducing the bilayer capacitance proportionally decreases the charg-
ing time. Unfortunately, reducing the bilayer diameter into the 40–60 μm range
(10–25 pF) may still result in an excessive saturation time (several milliseconds).
Because the magnitude of the electrostrictive change in capacitance depends on the
area of the annulus, decreasing the bilayer diameter also reduces this component. The
width of the annulus (i.e., in the plane of the bilayer) is also roughly proportional to
the thickness of the margin of the aperture. A substantial, further reduction in the
electrostrictive component can be achieved by using the shaved-aperture method,
because it produces a margin of the aperture that is very thin (a few micrometers).
Other techniques, such as drilling or punching apertures, produce a margin that is
much thicker. Finally, any manipulation of the bilayer that produces a “bulkier”
annulus, such as addition of excess solvent or repeated wiping, can greatly increase
the electrostrictive component and should be avoided.
Decrease the Feedback Resistance. The charging time can be reduced by decreasing the
value of Rf, thereby increasing the rate at which charge can be delivered to the bilayer.
However, the price for this improvement is a proportional increase in the background
thermal noise currents in Rf and, therefore, a concomitant decrease in usable band-
width. Although this may work for large conductance channels, it is hardly a general
solution. A better solution is to include both a large (50 GΩ) and a small (50 MΩ)
feedback resistor in the headstage with logic-controlled circuitry that can switch
between the two. This approach was taken in the CV-4B headstage for the Axopatch
1 series of Axon Cellular Neuroscience amplifiers. During a voltage step, the bilayer
capacitance is rapidly charged through the small Rf. Immediately after the bilayer
capacitance is charged, the feedback pathway is switched to the large Rf, enabling
high resolution recording of single-channel activity. This approach was used success-
fully to study voltage-dependent activation of K+ channels from squid (Wonderlin et
al., 1990; Wonderlin et al., 1991).
the voltage across a 100 pF bilayer by 100 mV without saturation of the headstage
output. Because integration of a constant patch current will eventually saturate Cf, an
integrating headstage requires a reset circuit that periodically discharges the feedback
capacitor; this is usually a logic-controlled switch that shunts the feedback capacitor.
In some designs, this same reset switch can be activated during a voltage step. Such a
“forced reset” allows rapid charging of the bilayer capacitance.
How should one choose among these amplifiers? The ideal amplifier for high-resolution
bilayer recordings should have:
When all of the requirements for an amplifier are considered, it is clear that anyone
attempting high-resolution recording from bilayers with voltage steps should seriously
consider using a capacitive-headstage amplifier.
1 A bilayer formed over an aperture in a partition separating two open bath chambers
(open-chamber configuration); and
2 A bilayer formed on the tip of a polished glass pipette (pipette/bilayer configuration),
either by “punching” a preformed planar bilayer or by dipping the tip through a lipid
monolayer.
The relative merits of these configurations have been discussed in (Wonderlin et al.,
1990). Briefly, for bilayer capacitance values larger than a few picofarads, the open-cham-
ber configuration with a shaved aperture will generate less background current noise than
the bilayer/pipette configuration, due to the much higher access resistance of the interior
of the pipette compared to the open chamber. Therefore, if bilayers must be made large
enough to permit incorporation of channels by fusion of vesicles with the bilayer, the
open chamber configuration should be used. If channels can be incorporated during the
formation of the bilayer (e.g., “tip-dip”) so that bilayers with a very small area (less than a
few picofarads capacitance) can be used, then the pipette/bilayer configuration can be
used. Finally, the open chamber configuration offers greater flexibility with regard to
manipulation of the bilayer and bath solutions.
Making Small Apertures
It is fundamentally important to be able to work with small bilayers, usually with diame-
ters in the range of about 40–60 μm. This size range represents a trade-off between the
need to reduce the area (and capacitance) while maintaining a large enough area to ensure
adequate incorporation by fusion of vesicles. Among the many techniques available for
making apertures in plastic partitions, the shaved-aperture method is especially well suited
for making small apertures. This method has been previously described in detail for mak-
ing small apertures in plastic cups (Wonderlin et al., 1990).
A. A metal stylus is warmed and pressed against the inner surface of the cup, pro-
ducing a cone-shaped depression.
B. After cooling, the stylus is retracted and the plastic shaved from the outer surface,
leaving a thin-edged aperture where the cut intersects the depression in the wall
of the cup, as in C. See text for additional details. Reproduced with permission
from Wonderlin et al., 1990.
Briefly, a conical metal stylus is warmed and pressed against the inner surface of a plastic
cup, forming a cone-shaped depression extending part way across the cup (Figure 5.7).
The stylus is then removed and the plastic shaved away from the outer surface, using a dis-
posable microtome blade, until the depression is intersected. The size of the aperture is
controlled by varying the depth of shaving. Apertures made using this method have a thin
margin beyond which the plastic rapidly increases in thickness, thereby decreasing the
stray capacitance across the partition and providing good mechanical strength. The thin
margin of the shaved aperture is very important in reducing the size of the annulus and,
therefore, the electrostrictive change in capacitance following voltage steps. Following are
the essential steps in implementing the method:
1 Making the stylus. The key to successfully using the shaved-aperture method is the
manufacture of a high-quality stylus. Using a lathe and dissecting microscope,
stainless steel darning needles can be ground to a very fine tip (< 5 μm diameter) and
polished to a very smooth finish using ultra-fine abrasive paper (Thomas Scientific,
Swedesboro, NJ). Softer metals can be substituted, but it is more difficult to produce
the highly polished tip. The stylus should be examined under high magnification
(400x) to ensure that the tip is very smooth.
2 Selecting the plastic. The original description of the shaved-aperture technique
(Wonderlin et al., 1990) recommended polystyrene as a good plastic for recording
cups. Since that time, repeated difficulties have been encountered with crazing of the
margin of apertures made in polystyrene cups, which sometimes appears to make the
cups electrically leaky. More recently, shaved apertures in cups were made from Ultra-
Clear centrifuge tubes (Beckman, Colombia, MD). Also, if a flat partition rather than
a cup is preferred, shaved apertures have been formed in plastic coverslips (Fisher
Scientific, Pittsburgh, PA). The margin of these apertures does not craze and bilayers
formed on apertures shaved in these plastics are very stable, lasting several hours.
Other plastics can probably be substituted, with the basic requirement being that they
cut cleanly so that the margin of the aperture is very smooth.
3 Melting the plastic. There are many choices of mechanical apparatus for manipulating
and heating the stylus. A rather simple method is to mount the stylus in a hole in the
tip of a variable temperature soldering iron, and then to mount the soldering iron on
a manipulator so that the tip can be carefully pushed into the plastic. Control of the
heat is important to ensure replicability and to avoid overheating, which may damage
the plastic and, perhaps, the stylus.
4 Shaving the aperture. It is easiest to shave the apertures while observing with a
dissecting microscope. The shaved apertures should be examined under high
magnification (400x) before use to ensure they are free from deformations that might
interfere with bilayer stability.
Figure 5.8: Voltage-step activation of a K-channel from squid giant axon axoplasm.
Wonderlin, W.F., French, R.J., Ion channels in transit: voltage-gated Na and K channels
in axoplasmic organelles of the squid Loligo pealei. Proc. Natl. Acad. Sci. USA, 88: 4391–
4395, 1991.
Hamill, O.P., Marty, A., Neher, E., Sakmann, B., Sigworth, F.J. Improved patch-clamp
techniques for high-resolution current recording from cells and cell-free membrane
patches. Pflügers Arch. Eur. J. Physiol. 391:85–100, 1981.
Sigworth, F.J., Electronic design of the patch clamp. in Single-Channel Recording, 2nd
Edition, pp. 95–126, Sakmann, B., Neher, E. Eds. Plenum Press, New York, 1995.
High-pass filtering is required when the main source of noise is below the frequency range
of the signals of interest. This is most commonly encountered when making intracellular
recordings from nerve cells in the central nervous system. There are low-frequency fluctu-
ations in the membrane potential due to a variety of mechanisms, including the summa-
tion of synaptic inputs. The small, excitatory synaptic potentials that the user might be
interested in are often smaller than the low-frequency fluctuations. Since excitatory synap-
tic potentials are often quite brief, the low-frequency fluctuations can be safely eliminated
by using a high-pass filter. High-pass filtering is also referred to as AC coupling.
Another type of filter that is often used in biological recording is the notch filter. This is a
special filter designed to eliminate one fundamental frequency and very little else. Notch
filters are most commonly used at 50 or 60 Hz to eliminate line-frequency pickup.
6.2.1. -3 DB FREQUENCY
The -3 dB frequency (f-3) is the frequency at which the signal voltage at the output of the
filter falls to √1/2, i.e., 0.7071, of the amplitude of the input signal. Equivalently, f-3 is the
frequency at which the signal power at the output of the filter falls to half of the power of
the input signal. (See Section 6.3.9., “Decibels (dB)”.)
6.2.3. ORDER
A simple filter made from one resistor and one capacitor is known as a first-order filter.
Electrical engineers call it a single-pole filter. Each capacitor in an active filter is usually
associated with one pole. The higher the order of a filter, the more completely out-of-
band signals are rejected. In a first-order filter, the attenuation of signals above f-3
increases at 6 dB/octave, which is equal to 20 dB/decade. In linear terminology, this atten-
uation rate can be re-stated as a voltage attenuation increasing by 2 for each doubling of
frequency, or by 10 for each ten-fold increase in frequency.
be present when the filtering begins, some digital filters use data that arrive after the cur-
rent data. This is clearly impossible in an analog filter because the future cannot be pre-
dicted. Typical digital filters are the box-car smoothing filter and the Gaussian filter.
10 BESSEL
20
BUTTERWORTH
30
40
50
60
70
0.1 1 10
Normalized Frequency, f/f -3dB
The spectra have been normalized so that the signal magnitude in the pass band is 0 dB.
The -3 dB frequency has been normalized to unity.
6.3.1. - 3 DB FREQUENCY
f-3, defined previously, is sometimes called the cutoff frequency or the corner frequency.
While most engineers and physiologists specify a filter’s bandwidth in terms of the -3 dB
frequency, for obscure reasons some manufacturers label the filter frequency on the front
panel of their instruments based on a frequency calculated from the intersection fre-
quency of the pass-band and the stop-band asymptotes. This is very confusing. To make
sure that the filter is calibrated in terms of the -3 dB frequency, a sine wave generator can
be used to find the -3 dB frequency.
6.3.2. ATTENUATION
Attenuation is the reciprocal of gain. For example, if the gain is 0.1 the attenuation is 10.
The term attenuation is often preferred to gain when describing the amplitude response of
filters since many filters have a maximum gain of unity. (For accurate measurements, note
that even filters with a stated gain of unity can differ from 1.00 by a few percent.)
6.3.6. OVERSHOOT
When the phase shift in the pass band is not linearly dependent on the frequency of the
sinusoidal component, the filtered signal generally exhibits overshoot. That is, the
response to a step transiently exceeds the final value.
6.3.7. OCTAVE
An octave is a range of frequencies where the largest frequency is double the lowest fre-
quency.
6.3.8. DECADE
A decade is a range of frequencies where the largest frequency is ten times the lowest fre-
quency.
5
PASS BAND STOP BAND
0 Attenuation:
-3 -6 dB/octave
-20 dB/decade
-5
20 log Vout
Vin
-10
(dB)
-15
-20
-25
decade
octave
30
0.1 1 2 10 100
Frequency, f -3dB ( Hz )
A number of the terms used to describe a filter are illustrated in the context of a single-pole,
low-pass filter.
Voltage:
Vout
dB = 20 log (1)
Vin
Power:
Pout
dB = 10log (2)
Pin
3 dB 1.414:1 2:1
6 dB 2:1 4:1
20 dB 10:1 100:1
40 dB 100:1 10,000:1
60 dB 1,000:1 1,000,000:1
66 dB 2,000:1 4,000,000:1
72 dB 4,000:1 16,000,000:1
80 dB 10,000:1 108:1
6.3.10. ORDER
As mentioned above, the order of a filter describes the number of poles. The order is often
described as the slope of the attenuation in the stop band, well above f-3, so that the slope
of the attenuation has approached its asymptotic value (see Figure 6.2). Each row in the
following table contains different descriptions of the same order filter.
Typically, the higher the order of the filter, the less the attenuation in the pass band. That
is, the slope of the filter in the pass band is flatter for higher order filters (Figure 6.3).
(3)
t 10 -90 ≈ 0.3/f -3
As a general rule, when a signal with t10-90 = ts is passed through a filter with t10-90 = tf ,
the rise time of the filtered signal is approximately:
2 (4)
tr = t s2 + t f
have a filter that was very effective at eliminating high-frequency noise if it caused 15%
overshoot in the pulses. Yet this is what many kinds of filters do.
In general, the best filters to use for time-domain analysis are Bessel filters because they
add less than 1% overshoot to pulses. The Bessel filter is sometimes called a “linear-phase”
or “constant delay” filter. All filters alter the phase of sinusoidal components of the signal.
In a Bessel filter, the change of phase with respect to frequency is linear. Put differently,
the amount of signal delay is constant in the pass band. This means that pulses are mini-
mally distorted. Butterworth filters add considerable overshoot: 10.8% for a fourth-order
filter; 16.3% for an eighth-order filter. The step response of the Bessel and Butterworth
filters is compared in Figure 6.4.
Butterworth
100
mV
Bessel
50
Input
Step
0
0 20 40 ms
Figure 6.4: Step response comparison between bessel and butterworth filters.
In many experiments in biology and physiology (e.g., voltage- and patch-clamp experi-
ments), the signal noise increases rapidly with bandwidth. Therefore, a single-pole filter is
inadequate. A four-pole Bessel filter is usually sufficient, but eight-pole filters are not
uncommon. In experiments where the noise power spectral density is constant with band-
width (e.g., recording from a strain gauge), a single-pole filter is sometimes considered to
be adequate.
In the time domain, a notch filter must be used with caution. If the recording bandwidth
encompasses the notch filter frequency, signals that include a sinusoidal component at the
notch frequency will be grossly distorted, as shown in Figure 6.5. On the other hand, if
the notch filter is in series with a low-pass or high-pass filter that excludes the notch fre-
quency, distortions will be prevented. For example, notch filters are often used in elec-
tromyogram (EMG) recording in which the line-frequency pickup is sometimes much
larger than the signal. The 50 or 60 Hz notch filter is typically followed by a 300 Hz high-
pass filter. The notch filter is required because the high-pass filter does not adequately
reject the 50 or 60 Hz hum (see Figure 6.5).
Voltage
Time
Voltage
Time
A. Shows an inappropriate use of the notch filter. The notch filter is tuned for 50 Hz. The input to the
notch filter is a 10 ms wide pulse. This pulse has a strong component at 50 Hz that is almost elimi-
nated by the notch filter. Thus, the output is grossly distorted.
B. Shows an appropriate use of the notch filter. An EKG signal is corrupted by a large 60 Hz compo-
nent that is completely eliminated by the notch filter.
The simplest and most commonly used filter for frequency-domain analysis in biological
applications is the Butterworth filter. This filter type has “maximally flat amplitude” char-
acteristics in the pass band. All low-pass filters progressively attenuate sinusoidal compo-
nents of the signal as the -3 dB frequency is approached from DC. In a Butterworth filter,
the attenuation in the pass band is as flat as possible without having pass-band ripple.
This means that the frequency spectrum is minimally distorted.
Usually, notch filters can be safely used in conjunction with frequency-domain analysis
since they simply remove a narrow section of the power spectrum. Nevertheless, they are
not universally used this way because many experimenters prefer to record the data “as is”
and then remove the offending frequency component from the power spectrum digitally.
If you wish to make peak-to-peak measurements of your data for high-frequency signals,
you must consider the sampling rate closely. The largest errors occur when samples are
equally spaced around the true peak in the record. This gives the worst estimate of the
peak value. To illustrate the magnitude of this problem, assume that the signal is a sine
wave and that the following number of samples are taken per cycle of the sine wave. The
maximum errors that one could get are then as follows:
The sampling of the peak values varies between no error and the maximum as stated
above.
If one intends to transform the data into the frequency domain, Butterworth or Elliptic
filters are more suitable than the Bessel filter. These filters have a sharper cutoff near the
-3 dB frequency than the Bessel filter, and thus better prevent the phenomenon known as
aliasing. With a fourth-order or higher Butterworth filter, it is usual to set f-3 to about
40% of the sampling rate. Frequently, researchers do not have a Butterworth filter handy.
If a Bessel filter is available, it can be used instead, but f-3 would normally be set to about
25–30% of the sampling rate. This is because the Bessel filter attenuation is not as sharp
near the -3 dB frequency as that of a Butterworth filter.
The table shows that the main reduction in noise is gained by using a Butterworth filter
instead of a Bessel filter. The improvement achieved by going from a 4-pole filter to an
8-pole filter of the same kind is small.
However, for the reasons previously discussed above, the Butterworth filter cannot be used
for time-domain analysis. Since this is the most common kind of analysis performed on
patch-clamp data, Bessel filters are almost invariably used.
x n −1 + xn + xn +1
yn = (5)
3
where yn and xn are the output and input samples at sample interval n.
Nonrecursive filters are also known as “finite impulse response” filters (FIR) because their
response to a single impulse endures only as long as the newest sample included in the for-
mula (i.e., xn+1 in the smoothing-by-3’s filter).
Another example of a nonrecursive digital filter is the Gaussian filter. It has a similar form
to the smoothing-by-3’s filter described above, except that typically there are more terms
and the magnitudes of the coefficients lie on the bell-shaped Gaussian curve.
These types of filters have the advantage of not altering the phase of the signal. That is,
the mid-point for the rise time of a step occurs at the same time both before and after fil-
tering. In contrast, analog filters always introduce a delay into the filtered signal.
A problem with digital filters is that values near the beginning and end of the data cannot
be properly computed. For example, in the formula above, if the sample is the first point
in the data, xn-1 does not exist. This may not be a problem for a long sequence of data
points; however, the end effects can be serious for a short sequence. There is no good solu-
tion other than to use short filters (i.e., few terms). Adding values outside the sequence of
data is arbitrary and can lead to misleading results.
Recursive Filter
The output of a recursive filter depends not only on the inputs, but on the previous out-
puts as well. That is, the filter has some time-dependent “memory.” Recursive filters are
also known as “infinite impulse response” filters (IIR) because their response to a single
impulse extends indefinitely into the future (subject to computer processing limitations).
Digital-filter implementations of analog filters such as the Bessel, Butterworth and RC fil-
ters are recursive.
Correcting for Filter Delay
The delay introduced by analog filters necessarily makes recorded events appear to occur
later than they actually occurred. If it is not accounted for, this added delay can introduce
an error in subsequent data analysis. The effect of the delay can be illustrated by consider-
ing two common questions: How long after a stimulus did an event occur (latency mea-
surement)? And, what was the initial value of an exponentially decaying process
(extrapolation back to the zero time of a fitted curve)? If the computer program was
instructed to record 50 points of baseline and then apply a stimulus, it would be natural,
but incorrect, to assume in the subsequent analysis that zero time begins somewhere
between points 50 and 51 of the recorded data, since zero time may actually be at point
52 or later.
The delay can be seen in Figure 6.4. The times to the 50% rise or fall point of a step signal
(delay) are approximately 0.33/f-3 for a fourth-order low-pass Bessel filter, and 0.51/f-3 for
an eighth-order Bessel filter. In practice, when using a fourth-order Bessel filter with
f-3 = 1 kHz and sampling the trace at 6 kHz, the filter delay is 330 μs. So, the whole
record will have to be shifted by 2.0 points with respect to the stimulus events that were
programmed.
The main advantage of locating a PGA after the multiplexor on the A/D board is that it is
inexpensive. However, there are significant disadvantages:
1 The PGA has to have extremely wide bandwidth since it must be able to settle within
the resolution of the A/D converter at the multiplexing rate. Depending on the
number of channels being sampled, this could mean that the bandwidth of the PGA
has to be ten to several hundred times greater than the bandwidth of the analog
signals. Such fast amplifiers are difficult to design, but in many cases the required
speed can be achieved. The less obvious problem is that every amplifier introduces
intrinsic noise, and the amount of noise observed on the amplifier output increases
with bandwidth. Since the PGA may have several hundred times more bandwidth
than the analog signal, it is likely to contribute more noise to the recording than is
inherent in the signal. This problem cannot be eliminated or even reduced by filtering
because filtering would lengthen the settling time of the PGA. This serious problem
limits the usefulness of a shared PGA. This problem does not exist in the previously
discussed systems in which there is an amplifier for each channel.
2 If the PGA is located on the A/D board, the low-level signals must be brought by
cables to the A/D board. Typically, these cables are a couple of meters or more in
length and provide ample opportunity for pick up of hum, radio-frequency
interference and cross-talk between signals. With careful attention to shielding and
grounding, these undesirable effects can be minimized. In the alternative approaches,
in which the signals are amplified before they are sent to the A/D converter, the
relative impact of these undesirable interferences are reduced in proportion to the
amount of early amplification.
Systems that include a PGA on the A/D board are useful when the signals require only a
small amount of amplification (ten fold or less) or when the user cannot afford or has not
invested in external amplifiers.
If the only rule was to maximize the early gain, all of the gain would be implemented
before any low-pass filter, notch filter or offset stages. However, with some signals, too
much gain in front of the low-pass filter can introduce a different problem. This occurs
when the signal is much smaller than the out-of-band noise.
This problem is illustrated by the signal in Figure 6.6. Panel A shows an input signal con-
sisting of a pair of 1 pA current pulses significantly corrupted by instrumentation noise.
When the noisy signal is first amplified by x100 and then low-pass filtered, the amplified
noisy signal saturates the electronics and the noise is clipped. (Note that for clarity, these
traces are not drawn to scale.) After low-pass filtering, the signal looks clean, but its ampli-
tude is less than x100 the original signal because the low-pass filter extracts a signal that is
the average of the non-symmetrical, clipped noise. When the noisy signal is first amplified
by x10, then low-pass filtered before a further amplification by x10, saturation is avoided
and the amplitude of the filtered signal is x100 the original input. Panel B shows the two
outputs superimposed to illustrate the loss of magnitude in the signal that was amplified
too much before it was low-pass filtered.
A.
Original signal
+ noise
1 kHz
x100 low-pass
x10 1 kHz
low-pass
x10
B.
When a signal is AC-coupled, the DC component of the signal is eliminated and the low-
frequency content is filtered out. This causes significant distortion of the signal, as shown
in Figure 6.7.
0.1 Hz
1 Hz
10 Hz
Input
500 ms
Since there is less distortion when the AC-coupling frequency is lower, it is tempting to
suggest that the AC coupling should always be set to very low frequencies, such as 0.1 Hz.
However, this is often unacceptable because some shifts in the baseline are relatively rapid
and need to be eliminated quickly.
A significant difference between using an AC-coupling circuit and setting a fixed DC off-
set is the way the two circuits handle ongoing drift in the signal. With fixed DC offset
removal, the ongoing drift in the signal is recorded. With AC coupling, the drift is
removed continuously. Whether this is an advantage or a disadvantage depends on
whether the drift has meaning.
For very slow signals, even the lowest AC-coupling frequency causes significant distortion
of the signal. For these signals, an alternative technique known as autozeroing can be used.
This technique is available in some signal conditioners, such as the CyberAmp 380 ampli-
fier, as well as in some amplifiers, such as the Axopatch-1 amplifier. In this technique, the
signal is DC-coupled and a sample of the signal is taken during a baseline period. This
DC sample value is stored and continuously subtracted by the instrument from the signal
until the next sample is taken. In early instruments, the sample was taken using an analog
sample-and-hold circuit. These circuits exhibit the problem known as “droop.” That is,
the sampled value drifts with time. Later systems, such as the ones used in the CyberAmp
380 amplifier, use droop-free, digital sample-and-hold circuits.
DC
10 Hz
1 Hz
0.1 Hz
DC with
Autozero
Autozero
The top trace shows an EKG signal that is sitting on a 5 mV offset resulting from elec-
trode junction potentials. At AC-coupling frequencies of 10 Hz and 1 Hz, the signal is
distorted. In the bottom trace, an Autozero command is issued to the signal conditioner
(the CyberAmp 380 amplifier) at the time indicated by the arrow. The DC component is
immediately removed, but the transients are unaffected.
1
τ= (6)
2π f−3
100.0 1.6
30.0 5.3
10.0 16.0
3.0 53.0
1.0 160.0
0.1 1,600.0
In one time constant, the signal decays to approximately 37% of its initial value. It takes
approximately 2.3 time constants for the signal to decay to 10% of the initial value.
6.4.6. SATURATION
The AC-coupling circuit is the first circuit in most signal conditioning pathways. If a large
step is applied to the AC-coupled inputs, the AC coupling rejects the step voltage with a
time constant determined by the AC-coupling frequency. If the amplifiers are set for high
gain, the output might be saturated for a considerable time. For example, if the gain is
x100, the AC coupling is 1 Hz and the step amplitude is 1 V, the output will be saturated
until the voltage at the output of the AC-coupling circuit falls to 100 mV from its initial
peak of 1 V. This will take about 2.3 time constants. Since the time constant is 160 ms,
the output will be saturated for at least 370 ms. For the next several time constants, the
output will settle towards zero.
providing this headroom since the internal amplifiers in most instruments operate linearly
for signal levels up to about ±12 V.
Normally, the overload circuitry simply indicates the overload condition by flashing a
light on the front panel. In more sophisticated instruments such as the CyberAmp ampli-
fier, the host computer can interrogate the CyberAmp amplifier to determine if an over-
load has occurred.
6.5. AVERAGING
Averaging is a way to increase the signal-to-noise ratio in those cases where the frequency
spectrum of the noise and the signal overlap. In these cases, conventional filtering does
not help because if the -3 dB frequency of the filter is set to reject the noise, it also rejects
the signal.
Averaging is applicable only to repetitive signals, when many sweeps of data are collected
along with precise timing information to keep track of the exact moment that the signal
commences or crosses a threshold. All of these sweeps are summed, then divided by the
total number of sweeps (N) to form the average. Before the final division, the amplitude
of the signal in the accumulated total will have increased by N. Because the noise in each
sweep is uncorrelated with the noise in any of the other sweeps, the amplitude of the noise
in the accumulated signal will only have increased by √N. After the division, the signal
will have a magnitude of unity, whereas the noise will have a magnitude of 1/√N. Thus,
the signal-to-noise ratio increases by √N.
Hum can be further minimized by using a notch filter, as discussed above, or by differen-
tial amplification. To implement the latter technique, the input amplifier of the data
acquisition system is configured as a differential amplifier. The signal from the measure-
ment instrument is connected to the positive input of the differential amplifier, while the
ground from the measurement instrument is connected to the negative input. If, as is
often the case, the hum signal has corrupted the ground and the signal equally, the hum
signal will be eliminated by the differential measurement.
arises from the fact that this corresponds directly to what they see on the oscilloscope
screen or data acquisition monitor.
Engineers prefer to quote rms values because these can be measured consistently. The rms
is a parameter that can be evaluated easily. In statistical terms, it is the standard deviation
of the noise. True rms meters and measurement software are commonly available and the
values measured are completely consistent.
On the other hand, peak-to-peak measurements are poorly defined and no instruments or
measurement software are available for their determination. Depending on the interpreter,
estimates of the peak-to-peak value of Gaussian noise range from four to eight times the
rms value. This is because some observers focus on the “extremes” of the noise excursions
(hence, the x8 factor), while others focus on the “reasonable” excursions (x6 factor) or the
“bulk” of the noise (x4 factor).
MDS Analytical Technologies has developed the pCLAMP® software to measure the rms
and the peak-to-peak noise simultaneously. The peak-to-peak noise is calculated as the
threshold level that would encompass a certain percentage of all of the acquired data. For
white noise, the corresponding values are:
In this Guide and various Axon Cellular Neuroscience product specifications, noise mea-
surements are usually quoted in both rms and peak-to-peak values. Since there is no com-
monly accepted definition of peak-to-peak values, MDS Analytical Technologies usually
uses a factor of about 6 to calculate the peak-to-peak values from the measured rms values.
In an article from Bell Telephone Laboratories (1970), the authors define the peak factor
as “the ratio of the value exceeded by the noise a certain percentage of the time to the rms
noise value. This percentage of time is commonly chosen to be 0.01 per cent.” This ratio,
from tables listing the area under the Gaussian distribution, turns out to be 7.78 times the
rms value. Thus, according to this article, the appropriate factor to calculate the peak-to-
peak values from the measured rms values is closer to x8.
6.8. BLANKING
In certain experiments, relatively huge transients are superimposed on the signal and cor-
rupt the recording. This problem commonly occurs during extracellular recording of
nerve impulses evoked by a high-voltage stimulator. If the isolation of the stimulator is
not perfect (it never is), there is some coupling of the stimulus into the micropipette
input. This relatively large artifact can sometimes cause the coupling capacitors in subse-
quent AC amplifiers to saturate. There might be some time lost while these capacitors
recover from saturation, and thus valuable data might be wasted.
If it is not possible to prevent the stimulus coupling, the next best thing to do is to sup-
press the artifact before it feeds into the AC-coupled amplifiers. This is made possible in
the Axoclamp 900A amplifier by providing sample-and-hold amplifiers early in the signal
pathway. The user provides a logic-level pulse that encompasses the period of the tran-
sient. This logic-level pulse forces the sample-and-hold amplifiers into the “hold” mode.
In this mode, signals at the input of the sample-and-hold amplifier are ignored. Instead,
the output of the sample-and-hold amplifier is kept equal to the signal that existed at the
moment the logic-level pulse was applied.
Two types of audio monitor are used. The first is a power amplifier that applies the signal
directly to the speaker, thereby allowing signals in the audio bandwidth to be heard. This
type of audio monitor is frequently used in EMG monitoring or in central nervous system
recording. Each spike is heard as an audible click, and the rate and volume of clicking is a
good indicator of the muscle or nerve activity. This type of audio monitor is either called
an AM (amplitude modulated) audio monitor, or it is said to be operating in “click”
mode.
The second type of audio monitor is a tone generator whose frequency depends on the
amplitude of the input signal. Usually, the oscillator frequency increases with increasingly
positive signals. This type of audio monitor is often used for intracellular recording. It
provides an extremely sensitive measure of the DC level in the cell. This type of audio
monitor is either called an FM (frequency modulated) audio monitor, or it is said to be
operating in “tone” mode.
Most researchers have a very strong opinion about audio monitors; they either love them
or hate them. Those who love audio monitors appreciate the supplementary “view” of the
data. Those who hate audio monitors have probably suffered the annoyance of having to
listen for interminable lengths of time to the audio output from another rig in the same
room. A constant tone can be irritating to listen to, especially if it is high pitched, unless it
has an important message for you, such as “my cell is maintaining its potential admirably.”
Audio monitors are standard features in Axon Cellular Neuroscience computer controlled
amplifiers. To minimize the potential for aggravation, we include a headphone jack so
that users can listen to the audio output without testing the patience of their colleagues.
Some transducer amplifiers allow the electrode impedance to be easily measured. For
example, the CyberAmp amplifiers have an Electrode Test facility. When this is activated,
an approximate 1 μAp-p, 10 Hz square wave is connected to every input via individual
1 MΩ resistors. The electrode resistance can be directly determined from the amplitude of
the voltage response.
To prevent the common-mode signal from swamping the much smaller differential signal,
the positive and negative gains of the amplifier must be nearly identical. In the above
EMG example, if the amplifier inputs are exactly unity, the 10 mV of hum that appears
equally on both electrodes does not show up at all on the amplifier output. The only sig-
nal to appear on the output is the small signal proportional to the EMG potential gener-
ated between the two electrodes.
In practice, the positive and negative inputs of the amplifier are never exactly equal. The
quality of their matching is measured by the common-mode rejection ratio (CMRR).
This is normally quoted in dB, where 20 dB corresponds to a factor of ten. Returning to
the EMG example, if the amplifier operates at unity gain with a CMRR of 60 dB (i.e.,
one part in a thousand), the 10 mV of common-mode hum results in 10 μV of hum
appearing on the amplifier output. This is small, but still significant compared with the
smallest EMG signals, so an amplifier with higher CMRR, e.g., 80 dB, may be desirable.
The CMRR of an amplifier varies with frequency. It is best at very low frequencies, while
above a certain frequency it diminishes steadily as the frequency of the common-mode
signal increases. It is therefore important to verify the CMRR of the amplifier at a fre-
quency that exceeds the expected frequency of the common-mode signal.
The CMRR of the recording system is adversely affected by imbalances in the source resis-
tances of the recording electrodes. This is because the source resistance of each electrode
forms a voltage divider with the input resistance of the amplifier. If the source resistances
of the two electrodes are not identical, the voltage dividers at the positive and negative
inputs of the amplifier are not equal. Returning to the EMG example, if the resistance of
one electrode is 9 kΩ, the resistance of the other is 10 kΩ, and the amplifier input resis-
tances total 1 MΩ, then the gain for one electrode is 0.9901 instead of unity, while the
gain for the other electrode is 0.9911. The difference is 0.001. Thus, even though the
amplifier may have a CMRR of 80 dB or more, the system CMRR is only 60 dB. In some
cases, 60 dB is acceptable, but in others it is not. The solution to this problem is to use an
amplifier that has very high input resistances of 100 MΩ or more.
If there is a large common-mode signal and a source imbalance of more than a few
kilohms, a high input resistance amplifier probe should be used. Several AI 400 series
probes are available from MDS Analytical Technologies that have input resistances of
10 gigohms (1010 Ω) or more. These probes are distinguished on the basis of noise, cost
and size.
6.12. REFERENCES
Bell Telephone Laboratories. Transmission Systems for Communications. By the mem-
bers of the technical staff at Bell Telephone Laboratories. Western Electric Company, Inc.,
Technical Publications. Winston-Salem, North Carolina, 1970.
Tietze, U., Schenk, Ch., Advanced Electronic Circuits. Springer-Verlag, Berlin, 1978.
There are many different types of transducers available for physiological recording and
they range in their requirements for excitation voltages, amplification and recording tech-
niques as well as mechanical connections. In the past this has required a different ampli-
fier for each transducer type. The introduction of the CyberAmp® 380 amplifier
(MDS Analytical Technologies), a computer-controlled signal-conditioning amplifier,
heralded a new approach to interfacing transducers. A small 15-pin socket on each chan-
nel of the CyberAmp 380 amplifier provides the excitation voltage, the differential inputs,
the offset correction and the filtering required by a wide variety of transducers. This elim-
inates the need for a dedicated amplifier to suit the transducer type. The AI 400 series
probes and adapters can plug into any socket on the CyberAmp 380 amplifier. Within the
15-pin probe connector is a small memory device called an EEPROM (electrically eras-
able programmable read only memory) that stores such information as the transducer
type, its scale factor and the offset. When the transducer is plugged into the
CyberAmp 380 amplifier, this information may be automatically loaded by the acquisi-
tion software, thereby making the transducer ready for immediate use.
7.1.1. THERMISTORS
Thermistors are resistors whose resistance drops significantly as temperature increases.
They are composed of a compressed and sintered mixture of metallic oxides of manga-
nese, nickel, cobalt, copper, magnesium, titanium and other metals. Thermistors exhibit
temperature coefficients many times greater than those of pure metals. For most of them,
the resistance falls by 4–6% per °C with increasing temperature.
The response is exponential with temperature, but can be considered to be linear over a
range of a few tens of degrees Celsius. By combining multiple thermistors and some resis-
tors, the useful linear range can be extended to span more than 100 °C. The Omega Engi-
neering, Inc. (Stamford, CT) thermistor composites 44018 and 44019 each contain a
matched pair of thermistors. When connected by precision resistors to an accurate DC
excitation voltage, these probes are interchangeable within ±0.15 °C over the recom-
mended range of -30 to +100 °C. These probes come in a wide range of shapes and sizes
for different applications such as esophageal, rectal and surface body temperature mea-
surements, or for immersion in hostile chemical environments.
Omega Engineering Inc. produces many other thermistor types as do other manufactur-
ers. These other types can also be interfaced to the CyberAmp 380 amplifier via the AI
490 Connector and AI 491 Cable Kits. Several different interface circuits are suggested in
the CyberAmp 380 manual.
The National Semiconductor temperature probes can be interfaced to the CyberAmp 380
amplifier via the AI 490 Connector and AI 491 Cable Kits with the user providing the
interface design using the range of application circuits provided by National Semiconduc-
tor and in the manual for the CyberAmp 380 amplifier.
7.2.1. THERMOCOUPLES
Thermocouples are economical and rugged transducers, having the advantage of small size
with a very fast response time and a wide temperature range. Thermocouples consist of
two dissimilar metals in contact. The offset potential between the metals is proportional
to temperature (see Table 7.1). The low sensitivity and broad operating range generally
make thermocouples more suitable for industrial applications than for physiological ones.
Because of their greater sensitivity, thermistors are more popular than thermocouples for
most physiological applications. On the other hand, thermocouples can be fabricated in
remarkably small sizes, creating the opportunity for some unusual biological applications.
For example, micron-dimension thermocouples that can be inserted into single living cells
have been described (Cain & Welch, 1974). Commercially available thermocouples can
be obtained with diameters as low as 25 μm and thermal time constants as short as 2 ms
(Omega Engineering, Inc., Stamford, CT).
ANSI Sensitivity at
Min Value °C Max Value °C Material
Type 20 °C μV/°C
E -200 900 60.48 chromel/constantan
J -200 750 51.45 iron/constantan
K -200 1250 40.28 chromel/alumel
R 0 1450 5.80 platinum/Pt-13%
rhodium
S 0 1450 5.88 platinum/Pt-10%
rhodium
T -200 350 40.28 copper-constantan
Thermocouples require a reference temperature that traditionally was an ice bath but is
more commonly provided by a compensation circuit. Complete signal-conditioning mod-
ules, such as the Analog Devices AD594 and 3B47 signal conditioners, exist. They con-
tain the differential amplifiers and the temperature compensation circuitry. Alternatively,
voltage-output temperature transducers, such as the LM35A, can be used to derive a com-
pensation voltage to apply to the negative input of a CyberAmp 380 amplifier. To inter-
face thermocouples to the CyberAmp 380 amplifier, the user must provide an interface
circuit using a temperature transducer for compensation or using the Analog Devices or
equivalent conditioning modules. Circuit examples are given in the CyberAmp manuals.
The resistance of most metals increases with increasing temperature. The sensitivity is
small, less than 0.4% per °C. The most commonly used metal is platinum because of its
wide linear resistance-to-temperature relationship.
Platinum RTD’s offer better stability and linearity than thermocouples, but are limited to
temperatures below 850 °C. Most platinum RTD’s have a resistance of 100 Ω at 0 °C and
a positive temperature coefficient of 0.385% per °C at 0 °C.
The simplest way to configure a platinum RTD is to excite it with a small, accurate DC
current. Currents of 1 mA or less are used so as to minimize self-heating. A differential
amplifier is used to measure the voltage across the RTD. If the excitation current is 1 mA,
the sensitivity is 385 μV per °C. To interface RTD’s to the CyberAmp 380 amplifier, the
user must provide an interface circuit. Circuit examples are given in the CyberAmp man-
uals.
R E+
VS+ Cc +
RZ
VO
RE -
VS- Cc -
RZ
The electrode impedances (resistances) RE combine with the amplifier input impedances
RZ to produce resistive dividers. RZ also combines with the capacitance of the cable to the
electrodes CC to produce low-pass filters. VS and VO are the source and output voltages.
1MΩ
VI = Vs (1)
1kΩ + 1MΩ
However, if the electrode resistance is 100 kΩ, the signal would be reduced by about
9.1%, since:
1MΩ
VI = Vs = 0.909Vs (2)
100kΩ + 1MΩ
VS
RE
VI
RZ
RZ
VI =
RE + RZ
Figure 7.2: High electrode resistance can attenuate a signal and increase crosstalk.
The resistance of each electrode in series with the amplifier input impedance acts as a pair
of voltage dividers. This can reduce the recorded amplitude of the signal. If the electrode
impedances are not matched, it will reduce the CMRR leading to amplification of com-
mon background signals. RE and RZ are the electrode and amplifier impedances. VS and
VI are the source voltage and input voltage to the amplifier.
For a differential amplifier, the CMRR is the ratio of the response for a signal at just one
input to the response for a common mode signal (applied to both inputs) of the same
amplitude. The CMRR of the CyberAmp 380 amplifier is 100 dB (at high gain). That is,
the output voltage with a signal applied to just one input is 100,000 times greater than
the output voltage with the same signal applied to both inputs. (The ratio of amplitudes is
measured in decibels where dB = 20 log10(Vout/Vin). A ten-fold increase in the ratio
equals a 20 dB increase (see Chapter 6).
1MΩ
VI = V s = 0.999V s
1kΩ + 1MΩ
1MΩ
VI = V s = 0.990V s
10 kΩ + 1MΩ
As for the problem of the signal attenuation, the solution to a mismatch of electrode resis-
tances is to use an amplifier probe with very high input resistance such as the AI 401 and
the AI 402 probes. With each of these probes, the electrode impedances used in the exam-
ple are too small to have any influence on the CMRR of the probes.
If high-input impedance probes are not available, the CyberAmp 380 amplifier has a
built-in capacity to measure electrode impedances and this should be carried out prior to
each recording session to ensure a match. This is a good idea anyway as it can reveal the
development of faults such as breaks in the electrodes.
Crosstalk from other body signals can occur for additional reasons, although infrequently.
For example, if invasive EMG leads are routed subcutaneously near the heart, the EKG
can capacitively couple into the EMG leads. This problem is eliminated by recording dif-
ferentially using lead pairs that are twisted together, or by not routing the EMG leads near
the heart.
Thermal noise, also called Johnson noise, is a voltage produced across the terminals of all
resistive elements (including electrodes) due to the random motion of charge carriers
within the element. Thermal noise is proportional to the resistance (R) and absolute tem-
perature (T) of the resistive element (see Chapter 11). To minimize the thermal noise con-
The noise contribution to the signal from the amplifiers is negligible if the amplifier noise
is less than the thermal noise of the electrodes. All Axon Cellular Neuroscience amplifiers
have very low noise levels. The lowest noise amplifier, the AI 402 x50 differential ampli-
fier probe, has extremely low noise of just 1.1 μVp-p in the DC-10 kHz bandwidth
(0.18 μV rms). This is approximately equivalent to the thermal noise of a 250 Ω resistor.
This means that for electrodes whose resistance exceeds 250 Ω, the thermal noise of the
electrode exceeds the noise of the amplifier.
Implanted electrodes must be capable of remaining in the same location and must be of
an appropriate size and separation. Bipolar electrodes should be placed in parallel with the
muscle fibers to record the maximum signal with an electrode spacing of 2–10 mm
appropriate for most mammalian muscles. This close spacing of electrodes reduces
crosstalk from other muscle sites and is therefore appropriate for selective recording from
local areas. Low-frequency components of electrical signals propagate for larger distances
in body tissues than do high-frequency components. The bipolar electrode configuration
acts as a high-pass filter whose cut-off frequency is determined by electrode spacing. Close
spacing results in a high cut-off frequency, thus filtering out some of the remaining low-
frequency components from distant muscle activity. However, close electrode spacing also
reduces the amplitude of the signal in a non-linear fashion. Larger electrodes reduce the
impedance (see “Noise Arising From Distributed Pipette Resistance and Capacitance” on
page 227) but are less selective regarding the site of muscle activity.
The frequency range of the EMG is 10–2,000 Hz, although the electrode configuration
and separation will have considerable influence on what frequencies are recorded. The sig-
nal size ranges from 5 μV–20 mV for surface recording and from 50–1,000 μV for inva-
sive recording.
Several probes are available for recording EMG with the CyberAmp 380 amplifier. These
include the AI 401, 402, and 405 active amplifier probes, the AI 417 passive adapter and
direct user connection.
7.8.2. EKG
There are probably more books available on electrocardiogram (EKG) recording and anal-
ysis than any other electrophysiological topic. The signals are usually of large amplitudes
and readily recorded without the need for any amplifier probes. As electrode spacing is
reduced, there is an increasing possibility of recording unwanted EMG signals from
neighboring muscles. Consequently, the traditional electrode sites are worth considering.
The AI 417 passive 2 mm adapter provides a single differential EKG channel and plugs
directly into the CyberAmp 380 amplifier.
The normal frequency range of the mammalian EKG is 0.2–100 Hz; its amplitude size is
up to 2–3 mV.
7.8.3. EEG
The electroencephalogram (EEG) ranges from 10–300 μV in amplitude and has a fre-
quency range from 0.2–50 Hz. A single differential EEG channel is best recorded with the
assistance of one of the low-noise AI 400 series active probes.
The nerve cuffs themselves are very simple to make by running bare stainless steel fine
wires through a small length of silicone tubing split longitudinally. Any bared wires that
are in contact with the outer surface of the cuff can be insulated with silicone adhesive.
The Wheatstone bridge should be considered as two pairs of resistive dividers with the
output voltage equaling the difference between the voltages at (A) and (B). The bridge cir-
cuit is good for detecting small changes in resistance, and the output is linear in the region
of balance, i.e., when the voltage difference is small.
VDC
R1 R3
Vo
A B
Differential
R2 R4 Amplifier
The largest output comes from having strain gauges for all four resistive elements,
although physical constraints may require the user to have only two or even one “active”
element. For applications where only two gauges can be used, they should be placed in
positions R1 and R2 and the gauges should be located such that one increases in resistance
and the other decreases in resistance during the applied pressure or force. This is usually
achieved by placing the elements on opposites sides of the beam under strain. When
choosing strain gauges, keep in mind that semiconductor types have outputs ten times
higher than metal-film types.
The sensitivity of a bridge circuit decreases with increasing temperature and some applica-
tions may therefore require the addition of a temperature compensation circuit that
should be placed as close as possible to the bridge in order to experience the same temper-
ature changes.
has greatly reduced the cost, decreased the temperature sensitivity and enhanced the range
of transducer devices available.
An important point when measuring very low pressures is to set the height of the pressure
transducer to that of the sense location in order to avoid hydrostatic errors introduced by
fluid-filled catheters. Pressure transducers should also be kept out of the range of heat
sources such as heat lamps to avoid temperature-induced increases in the signal output.
For user-designed force transducers, the two main choices are between resistive or semi-
conductor strain gauges. Resistive gauges are cheaper and generally more rugged while
semiconductor strain gauges are smaller and about 10 times more sensitive. Strain gauges
are available from Measurement Specialties, Inc. (Hampton, VA).
User-designed force transducers can be interfaced to the CyberAmp 380 amplifier via the
AI 490 Connector and AI 491 Cable Kits.
many difficulties and may not be much better than cinematography techniques using
markers on the limb joints for measuring muscle lengths.
the adhesive used to join the two insulations. In practice it is often faster and better to use
PVC-insulated wires because it is much easier to make a strong adhesion, even though
PVC is more permeable to water over the long term. Also, Araldite AV138M (CIBA-
GEIGY) and Epoxylite #6001 (Epoxylite Corporation, Anaheim, CA) both provide excel-
lent water-resistant insulation for rigid applications.
Special attention should be paid to providing strain relief where wires, and particularly
connections, are repeatedly flexed. Sliding a length of silicon tubing over the stress region
often reduces this problem.
Cooke, I.R., Brodecky, V., Becker, P.J., Easily implantable electrodes for chronic recording
of electromyogram activity in small fetuses. J. Neurosci. Meth. 33, 51–54, 1990.
Horowitz, P., Hill, W., Measurements and signal processing. The Art of Electronics, 14.
Cambridge University Press, Cambridge, 1986.
Lemon, R., Prochazka, A. Eds. Methods for neuronal recording in conscious animals.
IBRO Handbook Series: Methods in the Neurosciences, Vol. 4. John Wiley & Sons,
Chichester, 1984.
8.2.2. MEMORY
RAM (Random Access Memory) is the physical device used to store programs and data
while the computer is running. When the computer power is off, RAM cannot hold any
information, so it is referred to as “volatile” memory. It is used because instructions and
data can be read from, and written to, RAM much faster than any other type of memory
device. How much RAM you need and can use depends on the type of computer you buy
as well as the capabilities of the programs you use.
The usefulness of extra memory depends on the type of applications that you will be run-
ning. Acquiring and analyzing very large datasets in general requires more memory than
acquiring and analyzing smaller amounts of data.
For traditional electrophysiology experiments, these hard drives provide more than
enough storage. For applications that include live cell video at high frame rates, a terabyte
of storage is not enough for long-term archiving of data. The good news is that hard
drives using USB, Firewire or SATA can be stacked externally, and are cheap enough to be
able to keep up with the increasing amounts of data.
For data acquisition, hard disk drive performance may be important. To record long seg-
ments of unbroken data, direct acquisition to disk (rather than memory) is necessary, and
the speed of acquisition is limited by the drive performance. High hard drive speeds of
5,400 RPM are standard, and for a small price premium 7,200 RPM drives are readily
available.
There are two common methods of backing up data: you can backup to another hard
drive, or back up to an optical medium such as CD or DVD. With the advent of rewrit-
able optical disk drives, it is possible to store 680 MB of data on a CD, and 4.4 GB on a
single-layer DVD.
Although outstripped by hard drives for storage capacity, optical storage has the advantage
of being compact for small amounts of data and electrically stable. Another clear advan-
tage of optical storage is that once a CD/DVD has been closed, there is a very low chance
of overwriting data. With a hard drive, there is a much higher chance of losing data
through accidental overwrite unless special precautions are taken.
On the other hand, hard drives allow data to be recovered and reorganized relatively easily.
Optical disks can be almost impossible to recover once they become scratched, and reor-
ganizing data requires data transfer and re-burning.
8.5. GLOSSARY
A/D converter
An Analog-to-Digital converter (ADC) maps analog measurements to digital numbers
that the computer can store and use in calculations. See analog and digital.
Analog
Real signals in the physical world are described as analog, meaning that the values change
in a continuous manner. A mercury thermometer is an analog measurement device; the
mercury rises smoothly in response to changing temperature. However, the computer
requires discrete digital numbers to describe measurements of the data. See A/D converter
and digital.
ASCII
American Standard Code for Information Interchange. A standardized 7-bit code for
exchanging information between computer systems. The 128-character set consists of the
alpha-numeric and punctuation characters, as well as control codes such as Escape and
Carriage Return. Files on a computer are often stored in a simple ASCII format, so that
they can be easily transferred and read by different systems and programs.
Bit
Information is stored in computer memory as a series of bits, which are binary switches
(numbers) that can either be on (1) or off (0). A group of eight bits is called a byte, which
is a common unit for addressing and storing information.
Boot
To boot a computer refers to the sequence of events that occurs when a computer is
switched on. When the computer is turned on, its built-in software instructs the com-
puter to load the operating system from disk into memory, and turns over control of the
computer to the operating system.
Byte
A byte is a series of eight bits of data. Since a bit can have two values (zero and one), a byte
can have 28 (256) values. A character on an IBM text screen is stored in a byte of memory,
so there are 256 distinct characters available. A byte is a typical unit for storing data and
computer instructions. See word.
Cache
A cache is a place to store often-accessed data or computer instructions, and is designed to
provide quicker access for increased performance. A memory cache is faster-than-usual
RAM for storing data that the microprocessor requires often from the normal RAM stor-
age area. A disk cache is a RAM storage area for data that is stored on disk but is being
read frequently.
CPU
Central Processing Unit. See Microprocessor.
D/A converter
A Digital-to-Analog converter (DAC) transforms digital information from the computer
into analog voltages to drive real world devices. See Analog and Digital.
Data acquisition
In the context of computers, data acquisition refers to the analog-to-digital conversion of
measurements for storage and analysis by the computer.
Device driver
Software that acts as an extension to the operating system, usually for supporting a hard-
ware device. Mouse pointing devices and network cards are common devices that require a
device driver to be loaded when booting the system up. On PC computers, device drivers
are files that traditionally have a .SYS filename extension.
Digital
Data represented in discrete, discontinuous form is digital, as opposed to the smooth rep-
resentation of data as measured by an analog device, such as a pen chart recorder. Com-
puters require discrete digital numbers for storage and processing. See A/D and analog.
DPI
Dots Per Inch is a measure of resolution on a printed page or computer screen. The more
dots per inch, the sharper that text or a graphic image will appear. A laser printer typically
has a resolution of 300 DPI.
EEPROM
Electrically Erasable Programmable Read-Only Memory. Allows ROM memory chips to
be reprogrammed with new information by users in the field. See ROM.
I/O Interfaces
An I/O interface defines the physical interface and communication method for devices
attached to the computer, such as the display, printer, mouse and disk storage device. For
devices attached internally, such as the display adapter, the physical I/O interface is the
computer’s internal bus. For external devices, e.g., the mouse or printer, there are several
interfaces implemented on Windows and Macintosh computers.
PARALLEL PORT
The parallel port interface, which exists primarily in the Windows world, allows eight
bits of data to be transferred at a time (“in parallel”). It offers high speed, but its
design allows mostly one-way communication; the external device is very limited in
the signals it can send back to the computer. As a result the only devices designed to
use the parallel port are those that mostly receive data, the most common example
being the printer. While most consumer computers implement this port as a 25-pin
connector, commercial printers often use a 36-pin Centronics connector.
Parallel ports are becoming uncommon with the advent of USB printers.
SERIAL PORT
The RS-232 serial port is a general communication interface that allows full, two-
way communication, thereby making it much more versatile than the parallel port. It
exists on many types of computer systems. Common devices which attach to a serial
port include the mouse and the keyboard, but which used dedicated 6-pin PS/2 con-
nectors vs. the general-use 9-pin DB connector.
Serial ports have already been phased out for consumer devices, but because commu-
nication using the serial interface is relatively straightforward, serial ports remain use-
ful for low-level device communication. For device development, therefore, one or
two serial ports is essential.
When buying a new PC computer for use in a laboratory, you should ensure that it
has a serial port.
USB PORT
In just a few years, USB has lived up to its name and become the Universal Serial
Bus. Devices such as cameras, phones, printers, scanners, external hard drives, flash
drives, keyboards, mice and other consumer products now come with USB as the
standard interface. USB is much faster than older protocols such as serial, and so it is
suitable for applications that require large amounts of data to be transferred at high
speed. Modern scientific instruments such as the MultiClamp™ 700B and
Axoclamp™ 900A amplifiers are now being made with USB interfaces, and this
trend is likely to continue.
KB
A kilobyte is approximately 1,000 bytes. As a computer’s natural system of numbers is in
base two, a kilobyte actually refers to 210 (1,024) bytes. It is abbreviated with a capital K
to distinguish it from the abbreviation for 1,000, as in kHz.
MB
A megabyte is 1,024 kilobytes, or 1,048,576 bytes. See KB.
Memory
See RAM and ROM.
Microprocessor
A microprocessor is a single integrated circuit chip that combines functions for doing log-
ical and mathematical operations, managing memory and communicating with external
devices. It is the main control center of a microcomputer.
Operating system
An operating system manages the hardware in a computer and provides a set of services
for an application program, such as writing data to a disk drive.
Parallel port
A type of communication interface that transfers data across several parallel lines. Com-
munication is mostly one-way, and the length of the connecting cable is limited. In the
PC, the parallel port is used mainly for connecting printers.
PC
Originally referred to the IBM PC, but now refers to any computer compatible with the
IBM PC designs based upon Intel CPUs.
RAM
Random Access Memory is composed of specialized integrated circuits designed for stor-
ing programs and data. RAM can be quickly read from and written to, so programs exe-
cute quickly.
However, when the computer power is off, RAM cannot hold any information.
Resolution
The “sharpness” of images and characters on the computer screen or on printer output.
Most accurately described in units of dots per inch.
ROM
Read Only Memory resides on integrated circuits that hold information which is not lost
when power is turned off. The contents of the ROM are usually loaded by the manufac-
turer and are not alterable by the user.
RS-232
A common, standard protocol for communicating over serial ports. It is a general two-way
communication interface that transmits data on a single line and receives data on a single
line.
SCSI
Small Computer System Interface. This is a type of parallel interface that provides for
higher speeds than were possible in the older serial or parallel interfaces, while allowing
multiple devices to be connected to a single port.
Serial port
A legacy standard connection, which provides a general two-way communication interface
that transmits data on a single line and receives data on a single line. The serial interface
no longer exists in most types of computers. Also known as a
RS-232 serial port.
Throughput
A measure of performance describing how much data per unit of time can be transferred
from device to device in a computer system.
Video display card
A specialize circuit card to handle the complexity of graphics display. High-end graphics
cards can be added to computers to upgrade their standard display speeds, resolution and
number of colors.
20 20
18 18
16 16
14 14
12 12
10 10
8 8
6 6
4 4
6 7 8 9 10 11 12 1 2 3 6 7 8 9 10 11 12 1 2 3
Time Time
This figure illustrates the analog and the digital representations of the temperature data
presented in Table 9.1.
A much easier way to accomplish the same task would be to use a data acquisition board
in the computer to directly record the temperature every 30 minutes. In this case the ther-
mometer must report the temperature in a form that the analog-to-digital converter
(ADC) can read. Nearly all ADCs require that the data be presented as a voltage. To do so,
one uses a different type of thermometer, such as a temperature-dependent resistor in an
appropriate circuit, that generates an analog voltage proportional to the analog tempera-
ture. The computer is instructed to initiate an analog-to-digital conversion every 30 min-
utes and store the results on disk.
Sometimes the computer is required to generate an analog waveform that will be propor-
tional to a list of numbers that are held in memory or on disk. In this case, a digital-to-
analog (D/A) converter (DAC) is used. The principles of operation of a DAC are similar
to those of an ADC, but the operation is the reverse.
In an ADC, the total measurement range (e.g., 0–100 °C) is divided into a fixed number
of possible values. The number of values is a power of two, often referred to as the number
of “bits.” Commonly, these values are:
To illustrate the impact on the resolution of using an 8-bit, 12-bit or 16-bit ADC, con-
sider the temperature-measurement example where the electronic thermometer circuit
generates an analog output from -10 V to +10 V for temperatures in the range -100 °C to
+100 °C. In this case, the resolutions are:
The resolution of a 12-bit ADC with an input range of ±10 V is 4.88 mV. Although this
resolution does not represent a difficult number for a computer-based system to use, some
researchers preferred a round number such as 5.00 mV. This could easily be achieved by
setting the span of the ADC to ±10.24 V instead of ±10 V. Now, many ADC systems are
now designed with the ±10.24 V span (e.g., the Axon Digidata® 1440A data acquisition
system).
In practice, it is common to sample at a rate significantly faster than the minimum rate
specified by the sampling theorem. This is known as oversampling. Exactly how much
oversampling should be used depends upon the type of experiment.
For experiments where the data are analyzed in the frequency domain (e.g., noise analysis,
impedance analysis), it is common to oversample only modestly. The main concern is to
prevent aliasing. An anti-aliasing filter is introduced between the signal source and the
analog-to-digital converter to control the bandwidth of the data.
The factor of twice the analog bandwidth required by the sampling theorem is only appli-
cable if the anti-aliasing filter is ideal, i.e., the gain is unity in the pass band and it abruptly
changes to zero in the stop band. Ideal filters cannot be realized, although they can be
closely approximated. For frequency-domain analysis it is common to use sharp cutoff fil-
ters such as Butterworth or Chebyshev realizations. Sampling is typically performed at 2.5
times the filter bandwidth. For example, if the data are filtered at 10 kHz they should be
sampled at about 25 kHz. Slower sampling rates are unacceptable. Faster sampling rates
are acceptable, but offer little advantage and increase the storage and analysis require-
ments.
For experiments where the data are analyzed in the time domain (e.g., pulse analysis, I-V
curves), greater oversampling is required because reconstructing the analog signal requires
not only an ideal anti-aliasing filter but also an ideal reconstruction filter. The simplest
and most common reconstruction filter is to join each sample by a straight line. Other
techniques, such as cubic-spline interpolation, can be used, but due to their much more
demanding computational requirements they are used infrequently.
The reconstruction problem is discussed in the Sampling Rate section in Chapter 11.
There is no commonly accepted rule regarding the sampling rate of data for time-domain
analysis. In general, five times the data bandwidth is a common sampling rate, and ten
times is considered good. Sampling at 20 times is excessive and is rarely used.
9.4.5. MONOTONICITY
Monotonic behavior means that for every increase in the digital input to the D/A con-
verter there will be an increase in the analog output. This requires that the differential
nonlinearity error be less than 1 LSB.
Glitches occur when several bits in the control word change. Each bit is connected to a set
of internal digital logic and analog switches. There is variability in the response time of
each part of the circuit, and therefore the response to the new control word does not hap-
pen at the same instant for each bit. In addition, charge is injected from the digital logic
circuitry into the analog switches. Glitches are worst when a large number of the bits of
the digital input to the D/A are changed. For example, when the analog output is required
to go from negative to positive, the digital input bits can change from 1111 1111 1111 to
0000 0000 0000. That is, every single bit changes. This change generates the worst glitch.
In modern integrated-circuit D/A converters, efforts are made in the design of the chip to
match the propagation and activation delays of each bit in the circuit and to minimize the
charge injection so that glitches are small in the first place. It is thus becoming increas-
ingly less common for a deglitching circuit to be included. On the other hand, the high-
level of integration in modern D/A converters sometimes introduces another problem,
called feedthrough noise. This problem occurs when the digital latches that contain the
word for the D/A converter are integrated into the same chip. It may happen that because
of the proximity of the digital circuits to the D/A converter, digital noise couples into the
D/A converter circuit. The coupled signal manifests itself as a pulse on the output of the
D/A converter circuit each time the digital word is updated. Strangely, this feedthrough
noise often appears even if the D/A value is being held constant. This is because, for sim-
plicity, most software that simultaneously performs A/D and D/A conversions updates the
D/A word once per A/D sample, even if the D/A value is not changing. The best solution
is to eliminate the feedthrough noise at its source by using separate integrated circuits for
the D/A converter latches. This is the approach used in the Digidata 1440A digitizer.
9.6. TIMERS
Most ADC systems have several timers available for a multitude of tasks. The most essen-
tial is the provision of a regular clock signal to initiate each A/D conversion. In most sys-
tems, one D/A conversion is performed for each A/D conversion, but this is not required
and some systems provide for running the D/A converter at a different clock rate to the
A/D converter.
Additional timers are used to implement “gear shifting.” This is a technique wherein the
acquisition rate is rapidly changed (gear shifted) during acquisition from a low to a high
rate or vice versa, without stopping the acquisition.
Uncommitted timers are generally provided for use in frequency measurement, event
counting or interval timing. For example, the frequency of nerve spikes can be counted if
they are first detected by an event detector that puts out a digital pulse for every spike. An
example of event counting is the measurement of the number of photons collected by a
photomultiplier tube in a fixed interval. An example of interval timing is measurement of
the period of a spinning optical filter wheel so that the optical data collected can be nor-
malized against fluctuations in the rotational speed of the wheel.
Digital inputs are used routinely in industrial control applications for monitoring the state
of solenoids and other apparatuses, but they are rarely used in electrophysiological experi-
ments. The main application for digital inputs in electrophysiology is for triggering acqui-
sitions and for indicating that a tag should be attached to the data.
Nevertheless, it is usually possible to design data acquisition systems that do not suffer
from extraneous noise pickup. This is achieved by careful layout and good grounding.
Grounding and shielding, if applied with great skill and diligence, can sometimes be made
to work acceptably for the A/D converter, especially when combined with differential
recording. However, it is very difficult to achieve acceptable noise levels on the DAC out-
puts. The following approach is often invoked for 16-bit systems.
First, move the A/D converter off the plug-in board in the computer to a separate board
housed in an external box. This eliminates the problems due to direct pickup of noise gen-
erated inside the computer housing. However, it does not eliminate the noise introduced
through the system ground. The noise in the system ground can be eliminated by using
optical isolation. In this technique, the digital signals to and from the computer are not
connected directly. Instead, an optical coupler is interposed in each digital line. A separate
power supply is provided for the A/D and D/A side of the optical couplers. This tech-
nique allows a low-noise ground to be maintained for the A/D and D/A converters and
for the experimental setup.
Interrupt handling represents an acute problem for data acquisition in multi-tasking oper-
ating systems. Usually, when an interrupt occurs an immediate response is required. If the
multi-tasking operating system is busy updating a screen for a word processor, an immedi-
ate response will clearly not be forthcoming. Therefore, high performance clearly requires
that the data acquisition task runs as the foreground application. Another problem is that
modern computers generally have long interrupt latencies of several hundred microsec-
onds or more. One way to minimize the problem of excessive interrupt latency is to
ensure that the data acquisition routine has the highest priority and, in many cases, exclu-
sive control over the computer. Another way to minimize excessive interrupt latency is to
provide hardware support in the acquisition system. For example, a memory-buffered data
acquisition system can be configured by the acquisition software and left in a “primed”
state. When it “sees” an external trigger it can generate an interrupt, then commence the
acquisition before it even receives a response from the host.
When choosing hardware and software for a data acquisition system, we recommend that
you choose the software first and then purchase the hardware recommended by the soft-
ware applications that you have selected.
a voltage-clamp pulse. The usual rule for time-domain signals is that each channel should
be sampled at a frequency between 5 and 10 times its data bandwidth. Knowing the value
of the data bandwidth is required in order to set the filter cut-off frequency during acqui-
sition and analysis.
For a sinusoidal waveform, the data bandwidth is the frequency of the sine itself. For most
biological signals, the data bandwidth is the highest frequency of biological information of
interest present in the recorded signal. This can be determined directly by examining a
power spectrum of rapidly sampled unfiltered data, though this is rarely done. Alterna-
tively, one can estimate the number of points per time interval required to give a data
record whose points can be easily “connected” by eye and calculate the sampling rate
directly. The data bandwidth and filter frequency can then be calculated from the sam-
pling rate. For example, if a fast action potential (1 ms to peak) is to be recorded, 25 sam-
ples on the rising phase would yield a reasonably good 40 μs resolution, requiring a
sampling rate of 25 kHz and an approximate data bandwidth of 5 kHz.
The rules are more straightforward in some special cases. Single-channel recording is dis-
cussed below. For signals with exponential relaxation phases, the sampling rate needed to
estimate a time constant depends on the amount of noise present; for moderately noise-
free data, at least 15 points should be taken per time constant over a period of 4 to 5 time
constants. Many fitting routines will fail if sampling is performed over only 3 time con-
stants, since the waveform does not relax sufficiently far towards the baseline. For a sum of
multiple exponentials, the sampling rate is determined in this way from the fastest phase;
sampling must extend to 4 time constants of the slowest phase. If this would result in too
many samples, a split clock (as in the Clampex program of the Axon pCLAMP® suite) or
other methods of slowing the acquisition rate during the acquisition, could be employed
as long as at least 15 points are taken over each time constant.
When a set of several channels is recorded (e.g., channels 0 through 3), most data acquisi-
tion systems sample the channels sequentially rather than simultaneously. This is because
the system usually has only one analog-to-digital converter circuit that must be shared
among the channels in the set. For example, if four channels are sampled at 10 kHz per
channel, one might expect that they would be sampled simultaneously at 0 μs, 100 μs,
200 μs, etc. Instead, channel 0 is sampled at 0 μs, channel 1 at 25 μs, channel 2 at 50 μs,
channel 3 at 75 μs, channel 0 again at 100 μs, channel 1 again at 125 μs, etc. There is
therefore a small time skew between the channels; if this causes difficulties in analysis or
interpretation, a higher sampling rate can be used to minimize the time skew (but this
may cause problems associated with high sampling rates, as mentioned above).
An additional consideration arises from the fact that on many data acquisition systems,
including the Digidata® 1440A digitizer from MDS Analytical Technologies, the digital-
to-analog converter (DAC) is updated whenever the ADC is read, even if there is no
change in the DAC output. This means that the DAC is updated only at the sample rate
over all channels. For example, if a stimulus is a 0 to 150 mV ramp and 50 samples are
acquired from one channel at a sampling interval of 25 μs, the DAC output will appear as
a series of steps each 25 μs long followed by an upward jump of 150 mV/50 = 3 mV,
which may be too large for some electrophysiological applications. Therefore, if a rapidly
changing continuous waveform is applied while acquiring slowly, the output waveform
should be checked with an oscilloscope and, if necessary, the sampling interval should be
increased. The computer preview of a waveform cannot be relied upon for this purpose
because it does not account for the effect of sampling. Note, however, that since most
users acquire significantly more samples per sweep than 50, this problem will not occur
except in very unusual situations.
10.1.3. FILTERING
The signal should be filtered using an analog filter device before it arrives at the ADC. As
discussed in Chapter 6 and in Colquhoun and Sigworth, 1983 and Ogden, 1987, this is
done to prevent aliasing (folding) of high-frequency signal and noise components to the
lower frequencies of biological relevance.
Acquisition-time filtering of time-domain signals is usually performed using a Bessel filter
with the cut-off frequency (-3 dB point; see Chapter 6) set to the desired value of the data
bandwidth. A 4-pole filter is usually sufficient unless excessive higher frequency noise
requires the 6- or 8-pole version. The Bessel filter minimizes both the overshoot (ringing)
and the dependence of response lag on frequency. The latter two effects are properties of
the Chebyshev and Butterworth filters (see Chapter 6 and Chapter 11, or Ogden, 1987),
which are less appropriate for time-domain analysis.
frequency, and (ii) the effect of the filter may be compensated for by a mathematical pro-
cedure analogous to a high-frequency boost circuit in a voltage-clamp amplifier (Sachs,
1983). These advantages are important if the frequency properties must be known
throughout the analysis. If not, the smoothing filter is much faster to execute, especially
on long data records, and easier to implement if one writes one’s own software.
The current from a single ion channel is idealized as a rectangular waveform, with a base-
line current of zero (closed-channel), an open-channel current dependent on the conduc-
tance and membrane potential, and rapid transitions between these current levels. In
practice, this idealized current is distorted by the limited bandwidth of the apparatus and
contaminated by noise. Shifts in the baseline may occur during the recordings, and capac-
itative current spikes are present in sweeps in which voltage changes are applied across the
membrane. The current signal is further altered by filtering and by periodic sampling.
These effects can impede making confident inferences from data regarding channel behav-
ior.
100 f c* (1)
fc = *
d max FTC
Here FTC* is the observed rate of false threshold crossings measured using recordings
made with an arbitrary cut-off frequency fc*. FTC* can be measured from idealized single-
channel records generated using a threshold set on the side of the baseline opposite to
where transitions are observed. This analysis can be achieved using pCLAMP software.
10.4.12. HISTOGRAMS
The simplest histogram is one in which a series of bins are defined, each of which has an
associated upper and lower limit for the quantity of interest, e.g., dwell time. Each dwell
time will then fall into one of the bins, and the count in the appropriate bin is incre-
mented by one. In the cumulative histogram, a bin contains the number of observations
whose values are less than or equal to the upper limit for the bin.
(Figure 10.1 B). The main difficulty with this scaling is that the bins do not have the same
weight during the fit process (Section 10.4.17., “Fitting to Histograms”), because the vari-
ance of the bin depends on the number of counts it contains. This can be compensated
for using the transformation of Sigworth and Sine (1987). In this transformation the bin
width is constant and the square root of the number of counts per bin is plotted on the
ordinate, with time plotted logarithmically on the abscissa (Figure 10.1 C).
1 ms
sample
times
3 ms
event
5 ms
event
Figure 10.2 illustrates how a 1 ms sampling rate can result in an apparent 4 ms open time
for events of between 3 and 5 ms. The dots in the top part of the figure represent times
when the signal is sampled. The lower part of the figure shows two events whose wave-
forms are high at 4 sample times.
A third problem is termed binning promotion error. In an exponentially falling dwell time
distribution, a bin is likely to contain more events whose true dwell times are at the left
side of the bin than are at the right side. The average dwell time of the events in that bin is
therefore less than the midpoint time t. The error occurs when a fit procedure assumes
that all the bin events are concentrated at a single point located at the midpoint t, instead
of at the average, which is less than t. Binning promotion error can occur in addition to
sampling promotion error because binning takes place in a different step. Both of these
errors are due to the asymmetric distribution of true dwell times about the bin midpoint.
A correction procedure has been proposed by McManus et al., (1987).
The degree of these bin-width-related errors may be reduced independently of the correc-
tive procedures mentioned above, if the fit procedure explicitly uses the fact that a bin
actually represents an area instead of an amplitude. Some errors may be eliminated if the
individual data points are used without ever using a histogram, as in the maximum likeli-
hood fitting method.
Lastly, as discussed in the section on missed events, short events may not be detected by
the 50% threshold criterion. This can give rise to a number of errors in the extracted fit
parameters, which relate specifically to state models. For further details, consult the refer-
ences cited in French et al., 1990.
Histogram bins containing zero counts should usually be excluded from a fit because the
chi-square function (see below) is not defined when a bin i contains Ni = 0 counts and
therefore has σi = 0. Alternatively, adjacent bins can be combined to yield a nonzero con-
tent.
10.5. FITTING
10.5.1. REASONS FOR FITTING
Fitting a function to a set of data points, such as a histogram or a time series, may be done
for any of the following reasons:
1 A function could be fitted to a data set in order to describe its shape or behavior,
without ascribing any “biophysical” meaning to the function or its parameters. This is
done when a smooth curve is useful to guide the eye through the data or if a function
is required to find the behavior of some data in the presence of noise.
2 A theoretical function may be known to describe the data, such as a probability
density function consisting of an exponential, and the fit is made only to extract the
parameters, (e.g., a time constant). Estimates of the confidence limits on the derived
time constant may be needed in order to compare data sets.
3 One or more hypothetical functions might be tested with respect to the data, e.g., to
decide how well the data were followed by the best fit function.
The fitting procedure begins by choosing a suitable function to describe the data. This
function has a number of free parameters whose values are chosen in order to optimize the
fit between the function and the data points. The set of parameters that gives the best fit is
said to describe the data, as long as the final fit function adequately describes the behavior
of the data. Fitting is best performed by software programs; the software follows an itera-
tive procedure to successively refine the parameter estimates until no further improvement
is found and the procedure is terminated. Feedback about the quality of the fit allows the
model or initial parameter estimates to be adjusted manually before restarting the iterative
procedure. Fitting by pure manual adjustment of the parameters (the so-called “chi by
eye”) may be effective in simple cases but is usually difficult and untrustworthy in more
complex situations.
The two following topics will be briefly discussed below: statistics, i.e., how good is the fit
and how confident is the knowledge of the parameters, and optimization, i.e., how to find
the best fit parameters. The statistical aspects are well discussed in Eadie et al., (1971);
Colquhoun and Sigworth (1983) provide examples relevant to the electrophysiologist. A
number of aspects of optimization are presented in Press et al., (1988).
ln L
τ∗
τ
(2)
The logarithm of the likelihood presented as a function of variable τ; the maximum likeli-
hood of the function occurs at τ*. For large numbers of samples, the shape of this curve
approaches a Gaussian.
It turns out that τ* reliably converges to the true time constant if the number of events N
is sufficiently large. For this reason it is important to either collect a large enough number
of data points or repeat the experiment several times so as to reduce the variation of the
parameters obtained over the data sets to an acceptable level. If the noise is large or there
are sources of significant variability in the signal, the data may be useless except in a qual-
itative way because of large variations of the best fit parameters between the runs. If one
long run is taken instead of several smaller ones, the run can be broken up into segments
and the analysis results of the segments compared with each other to assure that conver-
gence is near.
Although the maximum likelihood method is the most reliable, the time requested for the
calculations may be prohibitively long. The chi-square method, described below, is an
alternative that requires less time.
The Chi-Square Function (Least-Squares Method)
Suppose that a set of p measurements are made at the times x1, x2, ..., xp, and that the val-
ues measured are y1, ..., yp. If each yi is measured with a measurement error distributed as
a Gaussian with standard deviation σ1, ..., σp, the maximum likelihood method is equiv-
alent to minimizing the chi-square function:
p
( y i − y *i ) 2
χ2 = ∑
i =1 σ i2
where y*i is the fit value corresponding to yi. Minimizing chi-square is also called the
least-squares method. If the fit is made to a data sweep, each yi is the value measured at the
time xi, and each σi is the standard deviation or uncertainty of that value. In the typical
case when all the σi’s are equal (i.e., the uncertainty in the data does not depend on the
time), the σi’s can be ignored while performing the search for the best fit parameters, but
must be specified if the goodness of fit is to be calculated. If the fit is made to a histogram,
each yi is the numbers of events Ni in bin i, and each σi is √Ni.
It is much easier to maximize the chi-square function than to minimize the likelihood
function, whether for data or for many mathematical functions used as models.
Since the use of the chi-square function is equivalent to the use of the likelihood function
only if the uncertainties in the data (e.g., noise) are distributed as Gaussians, the correct-
ness of a least-squares fit can depend on the characteristics of this uncertainty.
b*+σb
b*-σb
a*- σa a*+σa
that vicinity. The Levenberg-Marquardt method is more easily trapped in local minima of
the function, but it can provide better fits than the Simplex because it uses the mathemat-
ical characteristics of the function being minimized to find the precise location of the
minimum or maximum, within the numerical resolution of the computer. This method
also provides statistical information sufficient to find the confidence limits.
These methods are iterative, i.e., they continue refining parameter values until the func-
tion stops changing within a certain convergence criterion. They also require reasonable
starting estimates for the parameters, so that the function to be minimized or maximized
is not too far away from its optimum value; a poor starting set can lead some fit programs
to a dead end in a shallow local minimum.
For Boltzmann and exponential fits, the Axon Cellular Neuroscience analysis program
(Clampfit in the pCLAMP software suite) provides a non-iterative method, in which the
data points and function to be fit are transformed using a set of orthogonal Chebyshev
polynomials, and the fit function coefficients are quickly calculated using these trans-
formed numbers in a linear regression. This method is very fast and requires no initial
guesses, though the parameters may be slightly different than those found by the methods
listed above because the underlying algorithm minimizes a quantity other than the sum of
squared differences between fit and data.
10.6. REFERENCES
Colquhoun, D. and Sigworth, F.J. Fitting and Statistical Analysis of Single-Channel
Records. in Single-Channel Recording. Sakmann, B. and Neher, E., Eds. Plenum Press,
New York, 1983.
Colquhoun, D. and Hawkes, A.G. The Principles of the Stochastic Interpretation of Ion-
Channel Mechanisms. in Single-Channel Recording. Sakmann, B. and Neher, E., Eds.
Plenum Press, New York, 1983.
Dempster, J. Computer Analysis of Electrophysiological Signals. Academic Press, London,
1993.
Eadie, W.T., Drijard, D., James, F.E., Roos, M, Sadoulet, B. Statistical Methods in Exper-
imental Physics. North-Holland Publishing Co., Amsterdam, 1971.
Horn, R. Statistical methods for model discrimination. Biophysical Journal. 51:255–263,
1987.
McManus, O.B., Blatz, A.L. and Magleby, K.L. Sampling, log binning, fitting, and plot-
ting distributions of open and shut intervals from single channels and the effects of noise.
Pflügers Arch, 410:530–553, 1987.
Ogden, D.C. Microelectrode electronics. in Microelectrode Techniques. The Plymouth
Workshop Handbook. Standen, N.B., Gray, P.T.A. and Whitaker, M.J., eds.
The Company of Biologists Ltd., Cambridge, 1987.
Press, W.H., Flannery, B.P., Teukolsky, S.A. and Vetterling, W.T. Numerical Recipes in C.
Cambridge University Press, Cambridge, 1988.
Sachs, F. Automated Analysis of Single-Channel Records. in Single-Channel Recording.
Sakmann, B. and Neher, E., Eds. Plenum Press, New York, 1983.
Sigworth, F.J. and Sine, S.M. Data transformations for improved display and fitting of
single-channel dwell time histograms. Biophysical Journal, 52:1047–1054, 1987.
Sine, S.M. and Steinbach, J.H. Activation of acetylcholine receptors on clonal mamma-
lian BC3H-1 cells by low concentrations of agonist. Journal of Physiology (London),
373:129–162, 1986.
Wonderlin, W.F., French, R.J. and Arispe, N.J. Recording and analysis of currents from
single ion channels. in Neurophysiological Techniques. Basic Methods and Concepts.
Boulton, A.A., Baker, G.B. and Vanderwolf, C.H., Eds. Humana Press, Clifton, N.J.,
1990.
We will begin by discussing the basic noise mechanisms that arise from the physics of the
materials and devices that comprise the electrical system. Interference from external
sources and noise associated with digitization will be considered later.
The main fundamental types of noise are: thermal noise, shot noise, dielectric noise, and
“excess” noise (e.g., 1/f noise). These types of noise form the basis for a discussion of
amplifier noise and electrode noise.
All the fundamental types of noise are completely random in nature. Their average prop-
erties can be measured, but their actual values at any particular instant in time cannot be
predicted. The most convenient measure of the amplitude of noise is its root-mean-square
(rms) value. Many noise processes have a Gaussian distribution of instantaneous ampli-
tudes versus time. The area under the Gaussian distribution represents the probability that
a noise event of a particular amplitude will occur (the total area is unity). The probability
that a noise peak will exceed plus or minus one times its rms value is 0.32; the probability
of a particular noise peak exceeding plus or minus three times its rms value is 0.003. It is
common engineering practice to define peak-to-peak noise as 6 times the rms value; if the
noise process has Gaussian distribution, the noise will remain within this bound 99.7% of
the time. However, peak-to-peak noise is not as clearly defined as is rms noise, and it is
not uncommon to find peak-to-peak noise operationally defined as anything from 5 times
to 8 times rms.
When considering noise in the time domain, it is important to know the bandwidth over
which the noise process is observed. Noise is made up of many frequency components,
frequently extending from DC to many megahertz or gigahertz. Some noise processes are
naturally restricted in bandwidth, but most require the appropriate use of filtering to
restrict the bandwidth of the noise while allowing adequate resolution of the signal. When
a noise amplitude (rms or peak-to-peak) is discussed, it is appropriate to also note the
bandwidth over which the noise is observed and the type of filter that has been used to
restrict the bandwidth.
Due to the random nature of noise, uncorrelated noise sources will add in an rms fashion.
Thus if E1, E2, and E3 are the rms values of three uncorrelated noise sources, the total rms
noise, ET, of these sources together is given by:
(1)
ET = E12 + E 22 + E32
Because of this relationship, the largest individual source of noise will tend to dominate
the total noise. For example, if two noise sources with rms values of 100 μV and 20 μV are
added together, the resulting noise will have an amplitude of {(100 μV)2 + (20 μV)2}1/2 =
102 μV rms.
R (noiseless)
1/2
R (noiseless) Ith= (4kTB/R)
Eth
1/2
Eth = (4kTRB)
The power spectral density (PSD) of thermal noise is white, i.e., it does not vary with fre-
quency. Its value, S2thV, is given as a voltage noise by:
2
S thV = 4 kTR ( units: Volt 2/Hz ) (2)
2 4kT
SthI = (units: Amp2/Hz ) (3)
R
The variance (also called the noise power) over a bandwidth B (units: Hz) is then:
or
4kTB
I th2 = (units: Amp2) (5)
R
or:
4kTB
Ith = (units: Amp rms) (7)
R
Spectral densities, rms current and voltage noise for bandwidths of 1 and 10 kHz are listed
in Table 11.1 for resistances of 100 Ω to 100 GΩ. Note that while thermal voltage noise
increases with increasing resistance, thermal current noise decreases as the resistor’s value
increases.
where q is the charge of the elementary charge carrier (for electrons q = 1.6 x 10-19 cou-
lomb) and I is the DC current in Amps.
An important example of shot noise is the equivalent input current noise of an operational
amplifier that arises from its input bias current Ib (also referred to as gate current for FET1
input devices). For bipolar transistor input devices, Ib is typically in the range of 1 nA to
1 μA; for FET input operational amplifiers, Ib (at room temperature) is typically in the
range of 1 to 10 pA, and can be less than 0.5 pA for special devices. The rms value of shot
noise for operational amplifiers with Ib ranging from 0.1 pA to 1 μA in a 1 and 10 kHz
bandwidth are listed in Table 11.2. In capacitive-feedback patch voltage-clamp amplifiers
(e.g., the Axopatch™ 200 series amplifiers), shot noise current from the headstage ampli-
fier sets the noise floor at low-to-moderate frequencies. By design these devices display
very low levels of shot noise; selected units can have gate currents as low as 0.2 pA, result-
ing in low-frequency current noise less than that of a 250 GΩ resistor.
Where f is the frequency in Hz. The PSD of dielectric noise characteristically rises linearly
with increasing frequency. The dissipation factor is often listed as tan δe where δe is the
loss angle (note that for small δe, tan δe ≈ δe); the “quality factor” Q is the inverse of
D (Q = 1/D). For a bandwidth extending from DC to an uppermost cut-off frequency B,
the rms noise current ID, resulting from a lossy dielectric, is given by:
For the best solid dielectrics (e.g., quartz, sapphire and some ceramics), D is on the order
of 10-5 to 10-4. For poorer (lossier) dielectrics D can be much higher; e.g., some thermo-
setting plastics have D in the range of 0.01 to 0.1. For some glasses used to fabricate patch
pipettes, D is at least 0.01. The dissipation factor has some frequency dependence,
although in the important range from about 1 kHz to 100 kHz it is usually reasonable to
approximate it as a constant. D also shows some temperature dependence (typically
decreasing with lower temperatures), although again in the usual range of temperatures it
may be considered constant.
Dielectric noise is one consideration in the selection of feedback and compensation capac-
itors in capacitive-feedback headstages for patch clamping. It must also be considered in
the packaging materials, electrode holders, etc. for any high-sensitivity current or charge
High-value (gigohm range) resistors, such as those used as the feedback element in resis-
tive-feedback patch-clamp headstages, also display a form of excess noise that rises with
increasing frequency. Generally, such resistors only achieve their thermal noise level
(which is the minimum possible) in the frequency range from about 10 Hz to 1 kHz. At
low frequencies, 1/f noise is observed, and at high frequencies the noise PSD typically
rises as f α, where α is usually in the range of 1 to 2. This noise can result in rms noise
from a resistor having several times the predicted thermal noise value in a 10 kHz band-
width. The elimination of this noise source is one of the motivations behind using capaci-
tive-feedback in patch clamps such as the Axopatch 200 series amplifiers.
En
In
ef
Rf
CURRENT en
INPUT
Cg in
When analyzing the noise of an operational circuit it is convenient to consider the noise
sources in terms of their power spectral densities. In this case the noise sources will be
denoted by lower case symbols, e.g., en(f ) (units: V/√Hz) or en2(f ) (units: V2/Hz). As a
useful example of noise analysis in an operational circuit, consider the simplified current-
to-voltage converter illustrated in Figure 11.3. In this circuit en and in are the input volt-
age and current noise PSDs, respectively, and ef is the PSD of the noise (thermal and
excess) of the feedback resistor Rf. The current to be measured is introduced at the termi-
nal labeled “input.” For simplicity, the positive (+) input is shown to be grounded. Cg is
any capacitance between the negative (-) input and ground; this includes the input capac-
itance of the amplifier plus strays and is usually about 15 pF. The noise PSD at the output
of the IV converter, Sout2(f ), can be shown to be:
2
S out ( f ) = in2 R f2 + e 2f + e n2 ( 1 + 4π 2 f 2 R f2 C g2 ) (units : V 2/Hz) (11)
It is useful to present this result in terms of current so it can be compared directly with the
current signal being measured. This is accomplished by dividing the above expression by
Rf2, thus referring the noise to the input. The input referred PSD, Sin2(f ), is then given
by:
e 2f 1
Sin2 ( f ) = in2 + + en2 ( + 4π 2 f 2C g2 ) (units : Amp2/Hz) (12)
R f2 R f2
In general, in, en and ef are all functions of frequency: in is the shot noise of the input gate
current of the amplifier but usually displays some 1/f behavior at low frequencies; ef is the
thermal and excess noise of Rf; and en also displays 1/f behavior at low-to-moderate fre-
quencies. Nevertheless, it is convenient to assume that each of these terms is independent
of frequency so that equation (12) can be easily integrated over a frequency band (for
instance, DC to some uppermost bandwidth denoted by B). The square root of this result
is then the rms noise for the bandwidth B:
e 2f B 4
I in = in2 B + B + en2 + π 2C g2 B 3 (13)
R 2f R 2f 3
It should be noted that if it is assumed that ef is only the thermal noise of the feedback
resistor, then the term ef2/Rf2 is simply 4kT/Rf (see equation (3) above).
It is instructive to consider the relative magnitudes of the various terms in equation (12)
or (13). In the case of a patch voltage clamp, the value of Rf is selected to be very high,
both to provide adequate gain to resolve tiny (pA) currents and to minimize the noise
contribution of this resistor; 50 GΩ is common for single-channel recording. The PSD of
the thermal noise of an ideal 50 GΩ resistor is 3.2 x 10-31 Amp2/Hz and the rms noise
associated with such a component over a bandwidth of 10 kHz is 0.057 pA rms (although
as already noted, excess high-frequency noise can increase this value several fold). The shot
noise, which accounts for in, associated with an input gate current of 0.2 pA would, by
itself, produce noise of about 0.025 pA rms in a 10 kHz bandwidth. For a typical high-
quality patch-clamp headstage, en is about 2–3 nV/√Hz at frequencies above about
1 kHz; Cg is usually about 15–20 pF. With these values, the term ( 43 π 2en2C g2 B3 )1 / 2
amounts to roughly 0.15 pA rms in a 10 kHz bandwidth. These three terms should be
uncorrelated; therefore, they sum in an rms fashion so that the total predicted noise is
(0.0572 + 0.0252 + 0.152)1/2 = 0.162 pA rms. For a resistive-feedback headstage, actual
noise in a 10 kHz bandwidth is normally substantially higher (about 0.25–0.30 pA rms)
both because of excess noise from the feedback resistor and because of the characteristics
of the low-pass filters that do not abruptly cut off noise at the -3dB bandwidth (Section
11.12., “Filtering”). Nevertheless, it is important to note that for bandwidths above a few
kilohertz, the term arising from en and Cg dominates total noise. To intuitively under-
stand the origin of this term, it is only required to remember that the current through a
capacitor is directly proportional to the rate of change of voltage across the capacitor
(I = C(dV/dt) ). Therefore, the voltage noise at the negative (-) input of the amplifier pro-
duces current noise through the capacitor Cg. This noise increases with increasing fre-
quency simply because higher frequency components of the voltage noise involve more
rapidly changing voltages. The noise current through the capacitor is supplied by the feed-
back resistor and therefore appears as noise at the amplifier output.
headstage amplifier to produce current noise with a PSD that rises with frequency in
essentially the same fashion as the PSD of the open-circuit headstage. In fact, this noise is
perfectly correlated with the noise resulting from the intrinsic capacitance, Cin, associated
with the headstage input. Note that Cin consists of the input capacitance of the
JFET (Ciss = Cgs + Cgd, where Cgs is the gate-source capacitance and Cgd is the gate-drain
capacitance), plus any compensation capacitors connected to the input, plus, for a capaci-
tive-feedback headstage, the feedback capacitor (1 pF for the Axopatch 200 amplifier) and
capacitance associated with the reset switch, plus about 1–2 pF of stray capacitance: in
total, Cin is usually about 15 pF. Denoting the holder plus electrode capacitance as Che,
the PSD of the current noise arising from en is given by 4π2en2(Cin+Che)f2. The capaci-
tance of the holder can vary from about 1–5 pF; larger capacitances are associated with
holders with metallic shielding. The pipette capacitance depends on the depth of immer-
sion of the pipette, the type of glass used, the thickness of the pipette wall, and the use of
Sylgard coating. The total capacitance of the pipette will generally be in the range of
1–5 pF. Thus, Che can range from 2–10 pF. With Che = 2 pF, the increment in wideband
noise from this mechanism above that of the open-circuited headstage should not be
much more than 10%. However, with Che = 10 pF the noise increment could be more
than 50% from this mechanism alone. Obviously, from the point of view of noise, it is
important to minimize the capacitance of the holder and pipette.
Dielectric Noise
The basic characteristics of dielectric noise have already been described above (see equa-
tions (9) and (10)). Dielectric noise of the pipette can be a major contributor to total
noise in patch voltage clamping; in some situations it can be the dominant noise source.
The dielectric noise arising from the pipette depends on the dissipation factor D of the
glass used to fabricate the pipette, on the pipette capacitance (CD in equations (9) and
(10)), and on the presence of Sylgard coating. The dissipation factor D of glasses other
than quartz that have been successfully used for patch pipettes generally falls in the ranges
of 0.001–0.01. The dissipation factor for quartz is variously reported to be 10-5 to as
much as 4 x 10-4. For Amersil T-08 quartz, which has been used in all of the preliminary
tests of quartz pipettes described below, the reported dissipation factor is 10-4. For
uncoated pipettes the value of CD is determined by the dielectric constant ε of the glass,
the ratio of the outer diameter (OD) to the inner diameter (ID) of the pipette, the depth
of immersion of the pipette into the bath, and to some extent by the pulling characteris-
tics of the glass and the geometry near its tip. The dielectric constant ranges from as little
as 3.8 for quartz to more than 9 for some high-lead glasses. Typical glass capillaries used in
the fabrication of patch pipettes have an OD of 1.65 mm and an ID of 1.15 mm; thus
OD/ID ≈ 1.43. If it is assumed that these proportions (OD/ID = 1.43) are maintained as
the glass is drawn out in pulling, then for uncoated pipettes the value of CD will be
approximately 0.15 ε pF per mm of immersion into the bath. For thick-walled glass with
OD/ID = 2, this value would fall to 0.08 pF per mm of immersion; while for thin-walled
glass with OD/ID = 1.2 the capacitance would increase to about 0.30 ε pF per mm.
Despite this precautionary note, it is clear that, all else being equal, the value of CD varies
linearly with the dielectric constant ε of the glass. Equation (10) indicates that if the depth
of immersion and the OD/ID ratio are constant, the rms noise for a given bandwidth aris-
ing from the lossy dielectric characteristics of pipettes fabricated from various glasses will
be proportional to (Dε)1/2. Thus the lowest noise glasses are expected to be those that
minimize the product of the dissipation factor and the dielectric constant. This relation-
ship has been born out experimentally in tests of some 20 different types of glass used to
fabricate patch pipettes (Rae and Levis, Personal Communication).
To illustrate the range of results expected for different glass types, we will consider three
pipettes with identical geometry fabricated from three different glasses. It will be assumed
that OD/ID ≈ 1.4, CD = 0.2 ε pF per mm of immersion, the immersion depth is 2 mm,
the pipettes are uncoated or are coated very lightly with Sylgard (so that the coating does
not reduce the dielectric noise significantly), and the rms noise is measured at a 10 kHz
bandwidth (-3 dB, 8-pole Bessel filter). Corning 7052 glass (D ≈ 0.003 and ε ≈ 5) repre-
sents reasonably low-loss glasses used to fabricate patch pipettes. Under the above stated
conditions, the CD value of this pipette is 2 pF and the dielectric noise contribution pre-
dicted by equation 10 is about 0.17 pA rms, or about 0.2 pA rms if the characteristics of
an 8-pole Bessel filter are taken into account. On the other hand, 0080 soda lime glass
(D ≈ 0.01, ε ≈ 7) represents high-loss glasses, which were commonly used in the early
days of patch clamping. Its CD = 2.8 pF and dielectric noise of about 0.44 pA rms is pre-
dicted. Finally, Corning 7760 is a very low-loss glass with D ≈ 0.0017 and ε ≈ 4.5. With
CD = 1.8 pF, a dielectric noise of slightly less than 0.15 pA rms is predicted. These figures
are in reasonable agreement with experimental findings that have attempted to separate
the components of total noise arising from the dielectric noise of pipettes fabricated from
various glasses.
effective pipette immersion depth was introduced by Rae and Levis (1992) and is
described below.
The above discussion dealt with expected behavior for uncoated pipettes. However, it is
common practice (and highly recommended for low-noise measurements) to apply a coat-
ing of Sylgard #184 covering the entire tapered region of the pipettes (i.e., approx.
5–10 mm) and extending to within roughly 100 μm of the tip (see Chapter 4). Coating
with a hydrophobic substance such as Sylgard is necessary to prevent the formation of a
thin film of solution that will creep up the outer wall of an uncoated pipette. Such a film
can produce very large amounts of noise in uncoated pipettes. Sylgard coating not only
virtually eliminates this noise source but also thickens the wall of the pipette thereby
reducing its capacitance. The dielectric constant of Sylgard #184 is 2.9 and its dissipation
factor is ≈ 0.002, which is lower than that of most glasses that have been used for patch
pipette fabrication. Thus, coating with Sylgard will reduce dielectric noise of patch
pipettes. It is expected that the improvement in noise associated with Sylgard coating will
be greatest for glasses with a high D product (e.g., soda lime glass); this has been con-
firmed experimentally. Improvement of noise should be less for glasses with very low val-
ues of D, but coating with Sylgard will reduce the dielectric noise of all glasses. The effects
of Sylgard coating on noise are, however, difficult to quantify theoretically primarily
because the thickness of the coating is usually not uniform. In particular, it is difficult to
achieve a very thick coating very near the tip.
Experimental tests of the noise properties of patch pipettes fabricated from 19 different
kinds of glass (see Chapter 4) have confirmed the general conclusions described above.
With few exceptions, the noise attributable to the pipette is inversely correlated with the
D product of the glass. In addition, thicker-walled pipettes and shallow depths of immer-
sion reduce noise for any particular glass type. Sylgard coating has its greatest effect on the
glasses with the poorest inherent electrical properties, but it is important to remember
that such coating is necessary for all types of glass.
It has been obvious for some time that pipettes fabricated from quartz should produce
only very small amounts of dielectric noise due to the low dielectric constant of quartz
(ε = 3.8) and, more importantly, its extremely low dissipation factor (D ≈ 10-5 – 10-4).
However, due to the high melting point of quartz (≈ 1600 °C), only with the advent of
automated laser pullers has it become practical to pull patch pipettes from quartz tubing.
A quartz pipette with D = 10-4 that is immersed to a depth of 2 mm (again assuming
0.2 ε pF per mm of immersion) would be predicted to produce only about 0.03 pA rms of
dielectric noise in a bandwidth of 10 kHz (-3 dB, 8-pole Bessel filter); for D = 10-5 this
value would fall to 0.01 pA rms. Preliminary measurements using Amersil T-08 quartz
suggest that the amount of dielectric noise in this situation is closer to 0.04–0.05 pA rms.
A more detailed discussion of preliminary estimates of the noise properties of quartz
pipettes is provided below.
Dielectric noise is probably the largest source of noise for pipettes fabricated from all
glasses other than quartz. For pipettes fabricated from quartz, due to its very low dissipa-
tion factor, sources of noise other than dielectric noise are expected to dominate total
pipette noise (“Noise Properties of Quartz Patch Pipettes” on page 230).
To summarize, dielectric noise can be minimized by using thick-walled glasses with low
values of D and coating the pipette with Sylgard #184. The effects of Sylgard coating are
greatest for glasses with relatively poor electrical properties. For excised patches, dielectric
noise can be minimized by withdrawing the tip of the pipette as close as possible to the
surface of the bath.
It should be noted that dielectric noise will also contribute to the noise associated with the
holder. For an Axopatch 200A amplifier with an open circuit noise of 0.06 pA rms in a
5 kHz bandwidth, total noise should not increase to more than about 0.07 pA rms in this
bandwidth when the Axon-supplied polycarbonate holder is attached. Part of this noise
increment is due to the fact that the holder adds about 1–1.5 pF of capacitance to the
headstage input. The rest of the increment in noise associated with the holder is presum-
ably dielectric noise, which can be estimated to account for roughly 0.03 pA rms in a
5 kHz bandwidth.
Noise Arising From Distributed Pipette Resistance and Capacitance
Most of the resistance of a patch pipette resides at or very near its tip. On the other hand,
the capacitance of an uncoated pipette can be expected to vary linearly with its depth of
immersion into the bath. Therefore, it has sometimes been assumed that current noise
arising from the thermal voltage noise of the pipette resistance in conjunction with the
pipette capacitance can be assumed to be negligible in comparison with other noise
sources. However, significant resistance resides in regions of the pipette that are further
removed from the tip. The thermal voltage noise of this resistance will greatly exceed the
input voltage noise of the headstage itself, with which it is in series. The actual situation is
complicated because both the pipette resistance and capacitance are distributed. In this
section we will consider the pipette capacitance to be lossless, and initially the effects of
Sylgard coating will not be considered. In order to estimate the noise arising from the dis-
tributed pipette resistance and capacitance, we will consider rather idealized pipette geom-
etry.
The pipette has been modeled as a shank and a conical region approaching the tip; the
angle of the cone is 11.4° and the tip opening is 1 μm in diameter. With this cone angle
the ID of the pipette increases to 100 μm at a distance of 0.5 mm from the tip, 200 μm at
a distance of 1 mm from the tip, 300 μm at a distance 1.5 mm from the tip, etc. When
filled with a solution with a resistivity of 50 Ω cm the pipette will have a total resistance of
about 3.2 MΩ. About 2.1 MΩ of this total resistance resides in the first 10 μm from the
tip; slightly more than 3 MΩ occurs in the first 100 μm from the tip. However, about
80 kΩ resides in the region from 100–200 μm from the tip, an additional 27 kΩ resides
in the region from 200–300 μm; and about 37 kΩ occurs in the region from 300 μm to
1 mm from the tip. The region from 1–4 mm from the tip adds another 12 kΩ. It has
been assumed that the capacitance is uniformly distributed along the pipette with a value
of 1 pF/mm of immersion. An Ag/AgCl wire extends into the pipette coming to within
4 mm of the tip and it is assumed that the resistance of the pipette is negligible beyond the
tip of the wire (due to the shunting effect of the wire). A noise equivalent circuit of the
pipette can be created by lumping the resistance and capacitance of each small segment
(e.g., 20–50 μm) of pipette. The approximate circuit is then a ladder network formed of a
series of resistors and their equivalent thermal voltage noise sources with a capacitor to
ground at each node representing a portion of the pipette immersed in the bath. Rough
calculations with such an equivalent circuit indicate that for a depth of immersion of
2 mm the noise arising from this mechanism will be about 0.13 pA rms in a 10 kHz band-
width. For a 1 mm depth of immersion this value would fall to about 0.10 pA rms. For an
idealized pipette identical to the one described here but with a cone angle of 22.6° (total
resistance is about 1.6 MΩ) these values fall to about 0.07 pA rms for a 2 mm depth of
immersion and about 0.05 pA rms for a 1 mm depth of immersion, both for a 10 kHz
bandwidth. Over the frequency range of interest to patch voltage clamping, the PSD
(Amp2/Hz) of this noise should rise with increasing frequency as f 2.
These calculated values are highly approximate and the assumed geometry is obviously an
over-simplification. A variety of factors could increase the noise arising from this mecha-
nism. For example, the noise should increase if there is an extended region behind the tip
where the angle of taper is quite shallow, resulting in increased resistance in this region.
Noise will also increase if the wire does not protrude as close to the tip as assumed above.
In addition, some glasses tend to thin near the pipette tip, resulting in increased capaci-
tance in this region. It should also be noted that shallow depths of immersion would not
decrease this noise quite as much as might be expected since this would decrease the
pipette capacitance, but not its resistance.
The noise from this mechanism can be reduced in several ways. From the above estimates
it seems that noise arising from the distributed resistance and capacitance of the pipette
should be smaller than the dielectric noise of pipettes fabricated from even the best known
glasses. Nevertheless, with better (lower loss) glasses, particularly quartz, this mechanism
could be the dominant source of noise from the pipette itself, and its reduction may
become important. First, it is obvious that the geometry of the first few millimeters of the
pipette will be an important determinant of the magnitude of this noise; therefore, when
possible, pipettes should be pulled to minimize the resistance distal to the tip. Anything
that reduces the pipette capacitance per unit length will also reduce this noise. Thus,
thick-walled pipettes and glasses with low dielectric constants should provide the best
results in terms of noise. Perhaps of more practical importance, coating the pipette with
Sylgard #184 can significantly reduce the pipette capacitance, and consequently reduce
noise of the type considered here. However, as already noted, it is more difficult to build
up a thick coat of Sylgard in the region within the first few hundred micrometers from the
tip than in more distant regions. As was the case with dielectric noise, shallow depths of
immersion will also reduce the noise arising from pipette resistance and capacitance.
Finally, it should also be possible to reduce this noise regardless of the immersion depth by
using a fine wire (Ag/AgCl or platinized Ag/AgCl possibly sharpened at the tip) that pro-
trudes as far as possible toward the tip of the pipette. Such a wire will, in effect, short out
(in a frequency-dependent fashion) the resistance of the pipette-filling solution in the
region into which it extends; thus, while the pipette capacitance will be unaltered for any
given depth of immersion, its resistance up to the end of the wire would be significantly
reduced.
In 1992, Rae and Levis have introduced a technique that has been shown to minimize
pipette noise arising from all of the mechanisms discussed above. In this technique, the
surface of the bathing solution was covered with a layer of Silicone oil (#200 Fluid from
Dow Corning, Midland, MI). With excised patches it was then possible to raise the tip of
the electrode to within a few micrometers of the saline-oil interface; a sharp line of demar-
cation could be observed along the pipette at this interface. The electrical characteristics
of this oil are apparently quite good2, so that only the tip of the pipette that remains in
saline appears to contribute to total noise. If 10 μm of the pipette tip remain in saline, it
would be expected that the capacitance associated with this length should be roughly
0.01 pF. Dielectric noise associated with the immersed portion of the pipette should be
about one tenth of the values predicted for a 1 mm depth of immersion. It should be rea-
sonable to approximate distributed R-C noise in this situation by a lumped resistance
equal to the resistance of the pipette other than that arising from the tip itself is series with
approximately 0.01 pF. If this resistance is taken to be 2.5 MΩ, the expected noise is only
about 7 fA rms in a 10 kHz bandwidth; even for 10 MΩ it is less that 15 fA rms in this
bandwidth. The noise increment associated with the pipette capacitance and the head-
stage input voltage noise is also minimized in this arrangement since the capacitance
remains essentially the same as that with the electrode in air. It should also be noted that
even when patches cannot be excised, this approach can still be effective if the bath is
designed so that the solution level can be lowered, thereby bringing the layer of silicone
fluid very close to the cell surface.
Using quartz pipettes with this technique has resulted in noise levels in actual recording
situations as low as 0.08 pA rms in a 5 kHz bandwidth (4-pole Butterworth filter). For
these experiments, the Axopatch 200A amplifier was used with an open-circuit headstage
noise of 0.057 pA rms in a 5 kHz bandwidth; with the holder and electrode attached, but
with the electrode tip in air, noise typically increased to about 0.07 pA rms in this band-
width. In several experiments it was shown that the total noise closely approximated the
rms sum of the noise of the headstage plus holder/electrode (electrode in air) and the ther-
mal current noise of the measured DC seal resistance. This implies that dielectric noise
and “distributed R-C noise” of the pipette contributed only a negligible amount to total
noise in this situation. Rae and Levis (1993) showed that quartz pipettes can be fabricated
with the Sutter Instrument P-200 laser-based puller to yield lower noise than they previ-
ously achieved, obviating the need for Silicone oil.
Preliminary tests have also been made using this technique with pipettes fabricated from
two other types of glass (Corning 7760, Kimble KG-12). Although the noise reduction
2. Although the dielectric characteristics of the specific Silicone fluid used in these experiments is not available, Table 2.5 of Electronics
Designers' Handbook (1977, L. Giacoleto ed., McGraw-Hill) lists a dissipation factor (D) of 8.5x10-5 at 1 kHz and 2x10-5 at 100 kHz
for Silicone fluid (methyl- or ethyl-siloxane polymer); the dielectric constant ε is 2.68.
was significant, results were never as good as those achieved with quartz. For example, the
best results achieved with KG-12 were slightly more that 0.10 pA rms in a 5 kHz band-
width, even when seal resistances in the range of 50–100 GΩ were obtained. For a total
noise of 0.105 pA rms, and assuming that the headstage plus holder/electrode in air pro-
duced 0.075 pA rms and that a 50 GΩ seal produces 0.04 pA rms, all in a 5 kHz band-
width, the pipette would be estimated to contribute about 0.06 pA rms in this bandwidth.
On the basis of the above discussion, this is several times the amount of noise expected for
a pipette fabricated from this glass with 10 μm of its tip in saline. One possible explana-
tion of at least part of this discrepancy would be to assume that the glass had thinned
excessively near the tip. It is also possible that the noise attributed to the seal in arriving at
this estimate is too low. More work will be needed to clarify this issue.
Noise Properties of Quartz Patch Pipettes
Since 1991, it is now possible to routinely fabricate patch pipettes from quartz using the
Model P-2000 puller from Sutter Instrument Company (Novato, CA). This puller uses a
laser to overcome the extremely high melting temperatures needed to draw pipettes from
quartz tubing. Preliminary results were made using Amersil T-08 quartz pipettes
although, it was difficult to pull small-tipped blunt pipettes with 1.65 mm OD 1.15 mm
ID tubing. Pipettes fabricated from this tubing typically had a rather long and relatively
narrow shank. This geometry is not ideal for the lowest possible noise. Even so, these
pipettes produced significantly less noise than pipettes fabricated from the best glasses
used previously (e.g., 7760, 8161). In actual recording situations with pipettes immersed
about 3 mm into the bath and with seal resistances ≥ 50 GΩ, rms noise in a 5 kHz band-
width was typically about 0.12–0.13 pA rms using an Axopatch 200A amplifier. Estimates
of the contribution of the pipette to this noise were in the range of 0.075–0.09 pA rms in
this bandwidth. With the new technique described above (Rae & Levis, 1992), which
allows the pipette tip to be positioned within a few micrometers of a layer of Silicone oil
covering the surface of the bath, background noise levels as low as 0.08 pA rms in a 5 kHz
bandwidth have been achieved in actual single-channel recording situations with quartz
pipettes. In such cases the contribution of the dielectric noise and distributed R-C noise
of the pipette appears to be negligible in comparison to other noise sources.
More accurate estimates of the noise attributable to the pipette could be made by sealing
quartz pipettes to Sylgard. A typical result for approximately a 3 mm depth of immersion
is 0.115 pA rms in a 5 kHz bandwidth for pipettes coated with a thin layer of Sylgard to
within about 100–200 μm of the tip. The power spectral density (PSD) of this noise was
also measured. An estimate of the PSD attributable to the dielectric noise and distributed
R-C noise of the pipette was obtained by subtracting from the total PSD the PSD of the
headstage with the electrode in air with a correction made for the effects of the additional
capacitance of the immersed pipette (about 2.6 pF) in conjunction with the headstage
input voltage noise. The thermal current noise level associated with the DC seal was also
subtracted. The resulting PSD was taken to be the best estimate of the noise attributable
to the pipette per se. This PSD (Amp2/Hz) increased with frequency with a slope of
approximately f 1.85 in the frequency range from 2–20 kHz. From the PSD it could be
estimated that the pipette accounted for about 0.07 pA rms of noise in a 5 kHz band-
width and about 0.17 pA rms in a 10 kHz bandwidth. If it is assumed that the slope of
f 1.85 was composed of a component with a slope of f attributable to the dielectric noise of
the pipette and a component with a slope of f 2, attributed to distributed resistance-capac-
itance noise of the pipette, it can be estimated that dielectric noise would account for
slightly less than 0.06 pA rms in a 10 kHz bandwidth. Distributed R-C noise would then
account for about 0.16 pA rms at this bandwidth. The estimated value of dielectric noise
is somewhat higher than would be expected from the value of D listed for Amersil
TO-8 quartz (0.0001 at 1 MHz) and the measured value of CD (2.6 pF); with these val-
ues, equation 10 predicts ≈ 0.04 pA rms of dielectric noise in a 10 kHz bandwidth. If the
above parsing of the noise components is correct, then it can be estimated that the actual
value of D for the quartz used was closer to 0.00025. Even if this value is correct, the Dε
product for this quartz is still about an order of magnitude lower than that of any other
glass used to date for the fabrication of patch pipettes. At the time of this writing, neither
different grades (i.e., higher purity) of quartz from the same supplier nor quartz from
other manufacturers have been investigated. It should also be noted that the noise compo-
nent attributed to distributed R-C noise probably could have been reduced if the pipettes
tested had a more favorable geometry.
When the same procedure was used to estimate the pipette noise when the depth of
immersion was decreased such that the increment in capacitance was about 0.7 pF above
that measured with the pipette tip just above the solution, it was found that the noise
attributable to the pipette was roughly a factor of two lower than the figures reported
above. In this case the estimated pipette noise PSD increased with increasing frequency
approximately as f 1.9 in the range from 2–20 kHz. This is consistent with the prediction
that the relative decrease in dielectric noise will be somewhat greater than that of distrib-
uted R-C noise as the depth of immersion decreases.
It should be noted that the results for quartz pipettes are qualitatively as well as quantita-
tively different from those that have been obtained previously from pipettes fabricated
with other types of glass. Estimates of pipette noise based on procedures like those
described above have been performed for many types of glass (i.e., Figure 11.4). Even for
the lowest Dε product glasses (e.g., 7760, 8161, 7740), the estimated pipette noise PSD
rises with increasing frequency as f 1.1– f 1.3 (frequency range 2–20 kHz, immersion depth
≈ 2 mm). This indicates that dielectric noise dominates total pipette noise for these
glasses. However, for quartz not only is the pipette noise significantly less for a given
depth of immersion, but the estimated pipette noise PSD rises much more steeply with
increasing frequency (f 1.8 – f 1.9 in the 2–20 kHz frequency range)3. This indicates that
dielectric noise is no longer the dominant source of noise for patch pipettes fabricated
3. Note, however, that the PSD for quartz pipettes is lower than that of pipettes made from other glasses at all frequencies measured. At
sufficiently high frequencies distributed R-C noise should dominate the total noise for pipettes fabricated from any type of glass.
However, for quartz pipettes the PSD of distributed R-C noise should exceed that of dielectric noise by 1 kHz; for pipettes made from
7760, distributed R-C noise should not exceed dielectric noise up to 10–20 kHz, and for soda lime (e.g., 0080) pipettes this frequency
might be closer to 100 kHz or more. Distributed R-C noise depends on the pipette geometry and a variety of other factors already
described; the only influence of the glass type itself—accept in so far as it effects the geometry which can be pulled—is its dielectric
constant.
from quartz. The data described above are consistent with the interpretation that for
quartz pipettes “distributed R-C noise” has become the dominant noise mechanism due
to the greatly reduced contribution of dielectric noise. As more data from quartz pipettes
becomes available these estimates and conclusions will doubtlessly be refined.
Noise Arising From Pipette Resistance and Patch Capacitance
The capacitance, Cp, of the patch is in series with the entire pipette resistance, Re. The
thermal voltage noise of Re will produce current noise in conjunction with the patch
capacitance. The PSD of this noise, S2p (f ), is given by:
S p2 ( f ) = 4π 2ee2C p2 f 2 (14)
where e2e = 4kTRe is the thermal voltage noise of the pipette. The rms noise arising from
this mechanism from DC to a bandwidth B is then ( 4 π 2ee2C p2B3 )1 / 2 . In general, this
3
noise should be quite small, but it can become significant when the patch area is large;
e.g., when a large “bleb” of membrane is sucked into the pipette tip. Sakmann and Neher
(1983, in Single-Channel Recording, Sakmann, B. and Neher E. eds. pp. 37–51) mea-
sured patch capacitance for a large number of pipettes with resistances ranging from about
1–10 MΩ. They found that the value of Cp varied from as little as 0.01 pF to as much as
0.25 pF. Despite a very large amount of scatter, they found that Cp and Re were corre-
lated, with Cp increasing as Re decreases. The best fit to their data was Cp = 0.126 pF(1/R
+ 0.018), where R is the pipette resistance in MΩ. Using this average relationship it can be
predicted that for a “typical” 10 MΩ pipette (Cp = 0.015 pF) the noise in a 10 kHz band-
width (8-pole Bessel filter) arising from this mechanism will be about 0.03 pA rms, while
for a 1 MΩ pipette (typical Cp = 0.128 pF) this value will increase to 0.08 pA rms. Sak-
mann and Neher’s results indicate that in the most favorable situations in terms of
“Re-Cp” noise, the rms noise from this mechanism can be as low as 0.01 pA rms in a
10 kHz bandwidth. However, in the least favorable situation (Re ≈ 2 MΩ, Cp ≈ 0.25 pF)
the noise from this mechanism can be as large as 0.23 pA rms in a 10 kHz bandwidth
(8-pole Bessel filter).
2 (15)
S sh ( f ) = 4kT Re{Ysh }
where Re{Ysh} is the real part of the seal admittance Ysh. The minimum estimate of seal
noise results from the assumption that Re{Ysh} = 1/Rsh, where Rsh is the DC seal resis-
tance. If this assumption is correct, then for a 10 kHz bandwidth, seal resistances of 1, 10
and 100 GΩ would produce noise of 0.4, 0.13 and 0.04 pA rms, respectively. Since values
of Rsh in the range of 100–200 GΩ are not uncommon, this would imply that the noise
associated with a very tight seal would be small in comparison with other sources of
patch-clamp noise. However, it is possible that the PSD of seal noise may rise above the
minimum level given by 4kT/Rsh as frequency increases (i.e., the real part of the seal
admittance may increase with frequency) due to the capacitance of the glass and the mem-
brane which make up the wall of the seal. Unfortunately, we have no good theoretical
basis upon which to estimate Ysh since the precise nature of the membrane-glass seal is not
known.
It is also very difficult to empirically dissect out the noise associated with the seal from
total patch clamp noise. We believe that previous attempts to do this have overestimated
this noise. For example, as shown in Figure 1-11 of Sigworth (1983), data from F. Sachs
and E. Neher indicate a frequency-dependent seal noise PSD (Rsh ≈ 20 GΩ) that would
amount to at least 0.13 pA rms in a bandwidth from DC to 5 kHz. However, with an
integrating headstage (e.g., the CV 203BU of the Axopatch 200B amplifier), total noise
levels (i.e., including the noise of the headstage, holder, pipette, seal, etc.) lower than this
value have often been achieved in the same 5 kHz bandwidth in actual recording situa-
tions.
Rae and Levis have reported in 1992 measurements using an Axopatch 200A amplifier
with quartz pipettes and a novel technique involving placing a layer of Silicone oil on the
surface of the bathing solution containing the cells to be patched. This technique allows
excised patches to be brought within a few micrometers of the oil-water interface, thereby
minimizing the noise contribution of the pipette. With this approach, total noise levels as
low as ≈ 0.08 pA rms in a 5 kHz bandwidth were achieved. In several experiments with
excised patches, the DC seal resistance was measured (range 40–60 GΩ) and the noise was
measured with the tip just beneath the layer of oil and again with the tip withdrawn into
the oil. The rms difference of these two noise measurements should be dominated by the
seal noise plus the small amount of noise arising from the pipette resistance in series with
the patch capacitance. In all cases, it was found that this rms difference for a bandwidth of
5 kHz was close to the thermal current noise predicted from a measured seal resistance
(i.e., (4kTB/Rsh)1/2). Nevertheless, it is well known to anyone who has struggled to
achieve the lowest possible noise in patch-clamp measurements that there is considerable
variability in the total noise even among patches with very high seal resistances and with
all other conditions seemingly identical. Despite the conclusions drawn above, it seems
reasonable to guess that some of this variability arises from the seal. Noise associated with
the membrane-glass seal represents one of the fundamental limitations of the patch clamp
technique, but it now seems clear that under favorable conditions such noise can be as low
as 0.03 pA rms in a 5 kHz bandwidth. Rae and Levis (1993) achieved ultra low-noise
recording without silicone oil by improving the fabrication of the quartz pipettes.
2 4π 2 f 2 e s2 C m2
S wc ( f )= (units: Amp 2/Hz) (16)
1 + 4π 2 f 2τ sr2
where e2s = 4kTRs and τsr = RsrCm and Rsr is the residual (i.e., uncompensated) series
resistance, e.g., if Rs = 10 MΩ then for series-resistance compensation levels of 50%, 70%
and 90%, Rsr will be 5 MΩ, 3 MΩ and 1 MΩ, respectively. It should be noted that for
100% series-resistance compensation, equation (16) reduces to 4π 2 f 2es2Cm 2 . This
emphasizes that series-resistance compensation (if it is properly designed) only restores the
noise to the level that would have resulted from the thermal voltage noise of Rs in series
with Cm in the absence of the filtering effect of Rs mentioned above. The rms noise aris-
ing from Rs and Cm over a bandwidth from DC to B Hz can be obtained by integrating
equation (16) over f from 0 to B.
Patch clamps commonly use (switch in) a 500 MΩ feedback resistor for whole-cell voltage
clamping. The noise of this resistor dominates open-circuit headstage noise up to band-
widths of about 10 kHz. However, for typical cells the noise arising from Rs and Cm will
be larger than that of the headstage for all bandwidths above a few hundred Hz. Even for
relatively ideal situations in terms of noise (e.g., a small cell with Cm = 5 pF voltage
clamped through Rs = 5 MΩ), the noise of a 500 MΩ feedback resistor will not dominate
total noise for bandwidths above about 1 kHz. Figure 11.4 shows the noise PSD and rms
noise for a typical whole-cell voltage clamp situation with Rs = 10 MΩ and Cm = 33 pF
(these are the values used in the Model Cell provided with the Axopatch-1 and the
Axopatch 200 series of patch clamp amplifiers). The top panel shows the noise PSD (as
computed from equation 16 plus headstage noise) for series-resistance compensation lev-
els of 0%, 50%, 70% and 90%; the noise PSD of an open-circuit headstage alone is
shown for comparison. It is instructive to derive an expression for the high frequency pla-
teau of these PSDs (from equation 16). If the fraction of series-resistance compensation is
defined as α and β = 1 – α (e.g., for 80% compensation β = 0.2; also note that Rsr = βRs),
A 10
-24 Series Resistance
% Correction
-25 90
10
70
-26
10 50
Spectral Density
0
( A2 / Hz ) -27
10
-28
10 system noise
Whole Cell mode
-29
10
10 100 1k 10 k
Frequency ( Hz )
Series Resistance
B 100 % Correction
90
70
10 50
0
0.1
0.01
10 100 1k 10 k
Bandwidth ( Hz )
The lower panel of Figure 11.4 shows the total rms noise as a function of bandwidth for
the same cell parameters with Rs compensation levels of 0%, 50%, 70% and 90%; the
rms noise of the open circuit headstage is also shown. Note that by a bandwidth of only
about 200 Hz the total noise is twice that of the headstage alone; therefore, even if a head-
stage with negligible noise had been used, at this bandwidth total noise would only
decline to about 70% of the value shown. As the bandwidth of current measurement
increases, the noise of the headstage itself becomes progressively more negligible; with
90% compensation at a bandwidth of 1 kHz, a headstage with no noise would have
reduced total noise by less than 1% (recall that the noise of the headstage and the Rs-Cm
noise considered here are not correlated and, therefore, add in an rms fashion). It should
also be pointed out that the bandwidths in this figure refer to the setting of an external fil-
ter (a perfect “brick-wall” filter has been assumed, see Section 11.12., “Filtering”). But it is
important to realize that the actual bandwidth of current measurement is limited to
1/2πRsrCm (1 pole RC filter). Without series-resistance compensation, in this example
the bandwidth is only about 480 Hz and the use of external filter bandwidths much above
this will only add noise, not new signal information. The effective bandwidth increases
with increasing series-resistance compensation, reaching nearly 5 kHz with 90% compen-
sation.
In most cases such noise sources can be controlled by careful grounding, shielding and fil-
tering. (For a detailed discussion of these techniques see Chapter 2). In some situations,
however, shielding can actually increase noise. An example is metal shielding of the
pipette holder used in the patch clamp. Such shielding inevitably increases the capacitance
at the input of the amplifier (Cg in equations (11) – (13) above) by several picofarads.
Even if the mean voltage on the shield is precisely the same as that of the negative input of
the amplifier, the noise voltages will differ and lead to increased high frequency noise. For
this reason MDS Analytical Technologies does not offer or recommend such shielded
holders; it is our experience that grounding of nearby metal objects, such as the micro-
scope, usually provides adequate shielding. Vibration, either transmitted through the floor
or through the air, is another source of external interference and adequate vibration isola-
tion is almost always required for sensitive electrophysiological measurements.
δ2
12
δ2
12
For a 12-bit ADC with a 20 V FSR this noise value is 1.41 mV rms or about 8.5 mV
peak-to-peak.
It is obvious that the quantizing step δ should be small relative to the signal being mea-
sured. This is easily accomplished in most situations by the use of appropriate preamplifi-
cation to scale the desired signal such that it fills a reasonable portion of the dynamic
range of the ADC. Difficulties can arise, however, if you need to measure small changes
embedded in large signals. Analog instruments can often have a dynamic range that con-
siderably exceeds that of ADCs. Again, using the capacitive-feedback patch clamp as an
example, noise levels as low as 0.02 pA rms can be achieved at a bandwidth of 1 kHz;
moreover, such an instrument can achieve noise this low even with a gain as low as
100 μV/pA. This would amount to an rms noise of only 2 μV. In order to utilize the full
dynamic range of such an instrument at this bandwidth, an ADC with 22-bit resolution
(and capable of sampling at 2–5 kHz) would be required. To the best of our knowledge,
this much dynamic range is not required for electrophysiological measurements and 12-
to 16-bit resolution in conjunction with a variable amount of preamplification is quite
adequate (see Chapter 9 for further discussion).
11.11. ALIASING
A signal can be determined completely by a set of regularly spaced samples at intervals
T = 1/fs only if it does not contain components with a frequency equal to or greater than
fs/2. This is basically a statement of the sampling theorem; fs is just the sampling fre-
quency mentioned above. The term fs/2 will be denoted by fn and is often called the
Nyquist frequency or folding frequency for reasons that will be described below. Another
way of putting this is that for sampled data, frequency is only defined over the range from
0 to fn; components with frequencies higher than fn cannot be described (at least 2 points
per cycle are needed to uniquely define a sine wave). Obviously there is nothing (other
than good sense) that will stop one from digitizing a signal with frequency components
extending many times beyond fn. However, digitizing frequency components of the signal
that lie above fn will result in “folding back” of these higher frequency components into
the frequency range from 0 to fn, consequently producing aliases.
The term fn is referred to as the folding frequency because the frequency axis of the power
spectral density will fold around fn in a manner similar to folding a road map or a carpen-
ter’s scale. This folding effect is illustrated in Figure 11.5; frequency components above fn
are shifted to lower frequencies (in the range 0 to fn). If fx is the frequency of a signal com-
ponent (desired or noise) above fn, then the frequency of its alias, fa, is given by:
f a = f x − kf s (17)
where the brackets, | |, indicate absolute value, fs is the sampling frequency, and k is a pos-
itive integer taking on whatever value is required so that fa falls in the range 0 ≤ fa ≤ fn.
For example, with fs = 20 kHz (fn = 10 kHz), a frequency component at 18 kHz will alias
to a component at 2 kHz (fa = | 18 kHz – 1 x 20 kHz | = 2 kHz) in the digitized wave-
form. Similarly, as shown in Figure 11.5, frequency components at 22 kHz, 38 kHz,
58 kHz, 62 kHz, etc., will all alias to 2 kHz in the sampled data.
18 kHz
20 kHz 10 kHz
22 kHz 30 kHz
20 kHz
38 kHz
40 kHz 30 kHz
42 kHz 50 kHz
40 kHz
58 kHz
60 kHz 50 kHz
62 kHz 70 kHz
60 kHz
In this folding of the frequency axis for fs = 20 kHz (fn = 10 kHz), note that all of the
round points shown (18 kHz, 22 kHz, ...) alias to 2 kHz.
As an example of aliasing and the problems it can produce, consider white noise extending
from DC to 10 MHz. To be specific we will assume that the PSD of this noise is
10-14 V2/Hz (100 nV/Hz); the total noise in the 10 MHz bandwidth is then 316 μV rms
(about 2 mV peak-to-peak). If this noise is sampled at fs = 20 kHz without the use of an
anti-aliasing filter, it should be obvious that the rms value of the sampled points will be
the same as that of the original data, i.e., 316 μV. However, the sampled data cannot
describe any frequency component greater than fn, here 10 kHz. If a smooth curve is fit-
ted through the sampled points (e.g., using a cubic spline), you will find that the noise
appears to be bandlimited from DC to 10 kHz and that its amplitude is the same as the
original data; its power spectral density will be 10-11 V2/Hz, i.e., 1,000 times higher than
that of the original data because the sampling has folded over the original spectrum 1,000
times. The frequency components above 10 kHz have all been aliased into the frequency
band extending from DC to fn. Clearly, aliasing has not increased the total amount of
noise, but it has shifted it from higher to lower frequencies. It is worth considering what
will happen if the sampled data is subsequently digitally filtered with a cutoff frequency of
1 kHz. This will result in reducing the noise from 316 μV rms to 100 μV rms. However,
if the original noise signal had been passed through an analog filter with the same cutoff
frequency (1 kHz), the noise amplitude would have been reduced to only 3.16 μV rms.
Once aliasing has occurred it cannot be undone by any digital operation. The solution
here is either to sample much faster (> 20 MHz in this example) or, if a 20 kHz sample
rate is required, to use an analog anti-aliasing filter to adequately reduce the amplitude of
all frequency components above 10 kHz.
If the PSD of the noise is not white but instead rises as f 2 with increasing frequency—as is
the case for high frequency noise from a patch voltage clamp—the consequences of alias-
ing can be even more severe. As an extreme example, consider a voltage noise of
100 nV/√Hz in series with a 10 pF capacitor (see equation (13) and related discussion); in
a 10 MHz bandwidth this would produce a current noise of 115 nA rms. Again, assuming
a digitization rate of 20 kHz with no anti-aliasing filter, all of this noise would be aliased
into the frequency band from DC to 10 kHz, even though the noise in a bandwidth of
10 kHz would have only been about 3.6 pA rms.
Moreover, subsequent digital filtering of the digitized noise with a cut-off frequency of
1 kHz, as in the previous example, would only reduce the noise amplitude to about 35 nA
rms, whereas an analog filter set to a 1 kHz bandwidth would have reduced the noise to
only 0.115 pA rms—300,000 times less than achieved by digitally filtering the aliased
noise. Of course patch clamps do not have a bandwidth of 10 MHz; even with more real-
istic bandwidths, failure to use proper anti-aliasing filters can greatly increase noise
beyond that existing in the bandwidth resolved by the digitization process (i.e., fn) and
can reduce the effectiveness of subsequent digital filtering of the data.
It should be noted that in the above examples it has been assumed that the filter used had
an abrupt cut-off at its -3 dB bandwidth. This will normally not be the case when measur-
ing signals in the time domain. Filters with Gaussian or Bessel characteristics (as are used
most frequently for electrophysiological measurements) roll off quite gradually beyond
their -3 dB bandwidth (fc) and it is therefore not appropriate when using such filters to
eliminate aliasing to set fc = fn. Recall that the requirement to avoid significant aliasing is
that all frequency components above fn must be adequately attenuated. When using a
Bessel filter (typically 4- or 8-pole) for anti-aliasing, the choice of the cut-off frequency
relative to fn depends on the characteristics of the noise and how much aliasing can be tol-
erated. We advise the use of fc ≤ 0.4–0.5fn (0.2–0.25fs) as a useful and generally reason-
able practice.
11.12. FILTERING
The above discussion naturally leads to a brief discussion of filtering. In general, the band-
width of a filter is selected to reduce noise to acceptable levels so that the desired signal
can be adequately observed. As described above, filtering prior to digitization is also neces-
sary to prevent aliasing. If the signal to be measured is large in comparison with the back-
ground noise, then the filter bandwidth—and the appropriate digitization rate—can be
chosen on the basis of the desired time resolution; wider bandwidths allow more rapid
events to be resolved. However, in many electrophysiological measurements very wide
bandwidths will result in background noise levels that will obscure the signals of interest.
In such situations it is necessary to make compromises between the amount of noise that
can be tolerated vs. the time resolution that is achieved after filtering.
Among commonly used filters, those that provide the best resolution with little or no
overshoot and ringing of the step response are the Gaussian and Bessel filters (as the order
of a Bessel filter increases it more closely approximates a Gaussian filter). A true Gaussian
filter is easy to implement digitally, but is not often produced as an analog filter (although
passive filters that are a good approximation of this type can be constructed). Bessel filters
are more commonly used in analog applications. The basic characteristics of both filter
types are quite similar. A Gaussian filter has an impulse response with the shape of the
Gaussian distribution; its step response has no overshoot. The 10–90% rise time of the
step response of a Gaussian filter is approximately 0.34/fc, where fc is the -3 dB band-
width in hertz. However, in the frequency domain the roll-off of this filter is quite grad-
ual. Denoting the transfer function by H(f ), H(fc) = 0.707 (-3 dB), H(2fc) = 0.25
(-12 dB), and H(3fc) = 0.044 (-27 dB). An 8th-order Bessel filter closely approximates
this response in both the time domain and the frequency domain. Clearly, in terms of
noise reduction, a filter whose transfer function rolls off much more quickly after it
reaches fc would appear to be desirable. Analog filters with such characteristics are readily
available (e.g., Elliptical and Chebyshev types); in fact, you can buy sharp cut-off filters
with H(f ) = 0.01 (-40 dB) when f = 1.06fc. Digital filters can achieve even sharper cut-
offs. Unfortunately, however, sharp cut-off filters are characterized by severe overshoot and
prolonged ringing in their step responses. Additionally, for a given value of fc, their rise
times can be nearly twice that of a Gaussian or Bessel filter. Because of this, very sharp cut-
off filters are desirable for frequency domain measurements but quite undesirable for mea-
surements in the time domain. In fact, in order to achieve the same time resolution with a
sharp cut-off filter that is achieved with a Gaussian or Bessel filter it is necessary to use a
higher value of fc. In this case, the sharp cut-off filter (with its higher fc) will pass as much
or even more noise (depending on the spectral characteristics of the noise) as the more
gradual roll-off Gaussian or Bessel filter when the two have been set to provide essentially
equivalent time resolution (as judged by step response rise time; see next paragraph).
Some “exotic” filter types can produce the same rise time as the Gaussian filter with mini-
mal overshoot and reduce the noise by a small amount; however, unless the noise PSD
rises very steeply with increasing frequency, the improvement is only a few percent.
able to relate the time resolutions of the filter to its 10–90% rise time. For single-channel
measurements time resolution is often thought of in terms of the minimum detectable
event duration. To some extent, of course, such a minimum duration is dependent on the
detection algorithm used. Even so, it is reasonable to approximate the minimum detect-
able duration in terms of the filter bandwidth as 1/Tr, where Tr is the 10–90% rise time of
the filter. With such an operational definition of time resolution—or minimum detect-
able duration—it is possible to compare the performances of different filter types. As
already noted, for a Gaussian or Bessel filter (8th order) Tr ≈ 0.34/fc, where fc is the -3 dB
bandwidth in hertz. A 10th-order Chebyshev filter (0.1 dB pass-band ripple) is a reason-
able selection to approximate the “brick wall” characteristic mentioned above; for this fil-
ter H(1.09fc) = 0.1 (-20 dB), H(1.22fc) = 0.01 (-40 dB), and H(2fc) = -95 dB. However,
for this filter Tr ≈ 0.58/fc, i.e., approximately 1.7 times the rise time of a Gaussian or
Bessel filter with the same fc. Moreover the step response is characterized by a peak over-
shoot of about 20% and sustained ringing which is noticeable up to about 10/fc. In order
to achieve the same 10–90% rise time with the 10th-order Chebyshev filter that is
achieved with a Gaussian or 8th-order Bessel filter, it is necessary to select the -3 dB band-
width of the Chebyshev filter to be about 1.7 times higher than the -3 dB bandwidth of
the Gaussian or Bessel filter. For example, the rise time of a Chebyshev filter with fc = 17
kHz will be approximately the same as that of a Gaussian or Bessel filter with fc = 10 kHz;
it should be noted that the step response of the Chebyshev filter will ring severely. Even if
the ringing of its step and impulse response is ignored so that it is assumed that the 10th-
order Chebyshev filter and the Gaussian or Bessel filter have the same time resolution pro-
vided that fc (Chebyshev) ≈ 1.7fc (Gaussian), it is found that the Gaussian or Bessel filter
significantly out-performs the Chebyshev filter in terms of noise in the filtered signal.
For white noise it would be found that the Chebyshev filter would pass approximately
1.3x the noise passed by the Gaussian or Bessel filter, provided, of course, that both filters
had the same 10–90% rise time. For noise with a PSD that rises as f 2, the Chebyshev filter
would actually pass about 1.5x the rms noise passed by the Gaussian or Bessel filter, again
provided that the filters’ corner frequencies were set to produce the same rise time. Obvi-
ously, when time resolution is considered, extremely sharp cut-off filters are not the best
selection. Sharp cut-off filters should only be used when frequency domain data (e.g.,
power spectral density) is being collected.
Finally, as an illustration of the distinction between the -3 dB bandwidth and the upper-
most frequency component of noise that can pass through a Gaussian or Bessel filter, con-
sider a noise process (e2(f )) whose PSD (V2/Hz) rises as f 2 with increasing frequency. On
linear scales such noise is shown in Figure 11.6 along with the transfer function of a Gaus-
sian filter (squared, H2(f )) and the noise PSD that will be observed at the filter output
(i.e., H2(f )e2(f )). Note that the square root of the integral of the filtered noise PSD (i.e.,
the square root of the area under the curve) will be the rms noise. The PSD of the filtered
noise does not fall sharply at fc. In fact, the filtered PSD reaches its peak at about 1.2 fc
and does not fall to negligible levels until more than 3 fc. This must be remembered when
selecting a digitizing rate for any particular -3 dB filter setting if aliasing is to be avoided.
In comparison with a filter with a “brick wall” roll-off at the same fc, the Gaussian filter
will have somewhat more than 40% more noise at its output. But recall that the sharp cut-
off filter will have a much poorer time-domain response. In fact, as described above, if the
-3 dB bandwidth of the sharp cut-off filter is selected to produce essentially the same time
resolution as the Gaussian or Bessel filter, it will pass even more noise in this situation.
4.0
2
3.0
2.5
2
2.0
e2( f )
1.5
Transfer Function
(Squared H (f))
2
1.0 H ( f ) 1.0
2
2
0.5 e2 ( f )H ( f ) 0.5
0 1 2 3 4
Normalized Frequency (f/fc )
Illustrated are squared transfer function of a Gaussian filter (H2(f )), noise process with
PSD that arises as f 2 with increasing frequency (e2(f )), and the same noise process PSD
after passing through the Gaussian filter (e2(f )H2(f )). The scales are linear; frequency is
normalized to the -3 dB bandwidth (fc) of the filter; PSD scale is arbitrary. Note that due
to the gradual roll-off of the Gaussian filter, the filtered noise still has significant power
well beyond fc.
A useful alternative to adjusting the analog filter bandwidth to attempt to achieve an opti-
mum signal-to-noise ratio is to digitize the data at a rapid rate through a Bessel filter with
its -3 dB bandwidth set to prevent aliasing. The wideband digitized data may then be fil-
tered digitally to achieve an acceptable signal to noise level. The digital filter should once
again generally be of the Gaussian or Bessel type. In such situations more than one filter
are used in series; if these are all of the Gaussian or Bessel type, then the final -3 dB band-
width, fcF, is approximately given by:
1
f cF =
1 1 1
2
+ 2 + 2 + ...
f1 f2 f3
where f1, f2, f3 ... are the -3 dB bandwidths of the individual filters in series. Of course,
the same expression holds for a series combination of analog filters. The disadvantage to
this general approach is that initial data storage will be increased due to the high digitiza-
tion rate; once the data has been digitally filtered it can also be decimated if desired.
11.13.2. HOLDER
The pipette holder contributes noise by increasing the capacitance at the headstage input;
some dielectric noise must also be expected. The unshielded polycarbonate holders pro-
vided by MDS Analytical Technologies by themselves will only increase the headstage
noise by about 10% above its open circuit value even for the lowest noise capacitive-feed-
back devices. For example, for the Axopatch 200B amplifier with an open-circuit noise of
0.060 pA rms in a 5 kHz bandwidth (as shown on the panel meter), the total noise with
the holder attached should not increase beyond 0.070 pA rms (note that larger noise
increments probably mean that the holder needs to be cleaned), and will often be as low as
0.065 pA rms. Inserting a saline-filled electrode into the holder will further increase noise
even while the electrode tip is in air. With a saline-filled electrode, the noise of the
Axopatch 200B with an open-circuit noise of 0.060 pA in a 5 kHz bandwidth will gener-
ally increase to about 0.075 pA rms.
11.13.3. ELECTRODE
The noise associated with the patch pipette is determined by the type of glass used, the
wall thickness of the glass, the pipette geometry, the use of Sylgard coating, and the depth
of immersion into the bath. Dielectric noise is probably the dominant source of noise
associated with the electrode for all glasses except quartz. For glasses with lower Dε prod-
ucts—most notably quartz—it can be expected to become a much smaller component of
overall noise. Additional noise will arise from the distributed resistance and capacitance of
the pipette. For the best glasses presently in common use, this source of noise is probably
somewhat less than the dielectric noise of the glass; for very low-loss glasses, such as
quartz, it could be the dominant source of pipette noise. A small amount of noise will also
result from the thermal noise of the pipette resistance in series with the patch capacitance;
only under certain situations (a large patch area) will this “Re-Cp” noise become signifi-
cant.
Theoretical and experimental results indicate that a pipette fabricated from the lowest-
noise glasses (other than quartz) used to date (Corning #7760, #8161) with a moderate
coating of Sylgard will produce noise of about 0.2 pA rms in a 10 kHz bandwidth (8-pole
Bessel filter) for an immersion depth of about 2 mm. Under favorable circumstances, this
value can be cut at least in half when the tip of the pipette is withdrawn to within 100–
200 μm of the surface of the bath. For pipettes fabricated from quartz, preliminary results
indicate that for a 1 mm depth of immersion, noise of somewhat less than 0.1 pA rms can
be expected in a bandwidth of 10 kHz; with the very small immersion depth possible
(10 μm or less) with the Silicone-fluid technique described above, it can be estimated that
the noise contributed by a quartz pipette falls to less than half of this value. Results
obtained for quartz have involved pipettes with relatively long narrow shanks and resis-
tances of roughly 10 MΩ. Such a geometry is not ideal for achieving the lowest possible
noise. As techniques for pulling quartz pipettes improve, and as other grades and suppliers
of quartz are investigated, further improvements are likely.
11.13.4. SEAL
The noise associated with the seal is less easily predicted. Certainly a minimum estimate
of this noise in a bandwidth B is given by (4kTB/Rsh)1/2, i.e., the thermal current noise of
the DC seal resistance, Rsh. For excellent seals in the range of 50–200 GΩ this would
mean that the minimum noise attributable to the seal is in the range of 0.03–0.06 pA rms
in a 10 kHz bandwidth. Recent data suggests that under favorable circumstances seal
noise may be as low as these predicted values. However, experience indicates that there is a
good deal of variability in the noise of patches even when the seal resistances are very high
and all other precautions necessary to minimize noise have been strictly observed. Some of
this variability may well arise from the seal.
Rae, J.L., Levis, R.A., Quartz pipettes for single-channel recording. Biophys. 64, A201
(P394), 1993.
Sakmann, B. and Neher, E., Eds. Single-Channel Recording. New York: Plenum Press,
1983.
en2 = 4kTRB
T = temperature in °K
B = bandwidth in Hertz
The noise specified by this equation has a constant power spectral density within the
bandwidth and is zero outside the bandwidth. When noise is measured through a low-
pass filter having -3 dB bandwidth f-3, the noise will be greater than the amount predicted
because real filters continue to pass signals at frequencies > f-3.
The noise of a 10 MΩ resistor measured through various filters is shown in Table A.1.
Table A.1: Noise of a 10 MΩ Resistor in Microvolts rms.
2-Pole 4-Pole
Bandwidth 1-Pole RC Infinitely Sharp
Butterworth Butterworth
DC-100 Hz 5.07 4.27 4.11 4.06
DC-1 kHz 16.0 13.5 13.0 12.8
DC-10 kHz 50.7 42.8 41.1 40.6
DC-100 kHz 160 135 130 128
1 1.1
f −3 = ≈
2πτ π t10 −90
t 10 -90 ≈ 2.2τ
A.1.4. FILTERS
Low-pass filters are commonly used in electrophysiology to reduce noise. The important
parameters of a low-pass filter are the -3 dB frequency, the type and the order.
The -3 dB frequency (f-3) is the frequency where the voltage response falls to √2 (0.707).
Some manufacturers specify the “cutoff ” or “bandwidth” based on the phase response or
an asymptotic approximation to the high and low frequency amplitude response. There-
fore when using a new filter you should check that the specified settings refer to the -3 dB
frequency.
The three types of low-pass filters commonly used in electrophysiology are the Bessel,
Butterworth and multiple coincident pole (RC) types. The most important differences
are:
1 When driven by a step voltage, the RC filter has no overshoot, the Bessel filter has
< 1% overshoot and the Butterworth filter has around 10% overshoot.
2 When driven by a noisy source, RC filters show the least rejection of the noise at
frequencies above f-3, Bessel filters show moderate rejection and Butterworth filters
show the most.
The order of a filter refers to the number of poles. Each resistor-capacitor section contrib-
utes one pole (inductors are rarely used for bandwidths < 100 kHz). Thus 4th-order,
4-pole and 4 RC sections all refer to the same thing.
The attenuation of high-frequency noise increases with order. At frequencies well above
f-3, the attenuation increases at 20 dB/decade (6 dB/octave) for each pole. Thus a 4-pole
filter rolls off at 80 dB/decade.
Note: 20 dB attenuation corresponds to a drop in voltage amplitude to one tenth.
1 -3 dB bandwidth and order of the filter used in the measurement circuit. Typically the
noise will look 20–30% better if a fourth-order low-pass filter is used instead of a first-
order low-pass filter.
2 Microelectrode bandwidth. The capacitance neutralization should be adjusted so that
the -3 dB bandwidth of the microelectrode is the same as the -3 dB bandwidth of the
measurement circuit.
If a voltage step is applied to a microelectrode headstage via a resistor, the resistor should
have a very low stray capacitance. Otherwise the stray capacitance couples the step directly
into the headstage input and artificially fast rise times are measured. A better estimate of
the rise time can be made by passing a current step out of the headstage into a load resis-
tor.
For consistent comparisons, rise times should be measured with the capacitance neutral-
ization adjusted for zero overshoot.
For an electrode with resistance R and an exponential response to a step input, the effec-
tive input capacitance can be estimated from:
The bandwidth during an experiment is usually much lower. It is typically limited by the
electrode resistance and membrane capacitance, as well as by stray capacitances. For exam-
ple, if a 5 MΩ electrode is used to clamp a whole cell having 20 pF of membrane capaci-
tance, the time constant for measuring membrane currents is 100 μs (5 MΩ x 20 pF).
This corresponds to a -3 dB bandwidth of 1.6 kHz, far below the 100 kHz that may be
specified. 80% series-resistance compensation would improve this to 8 kHz. Nevertheless,
it is still better to have a headstage with very wide bandwidth because in some cases series-
resistance compensation may work better.
A.3.2. NOISE
The ideal RMS current noise in selected bandwidths is shown in Table A.2 for various
feedback resistors. These figures assume ideal resistors (which only have thermal noise)
and noiseless electronics.
Table A.2: rms Current in Picoamps.
BANDWIDTH 500 MΩ 1 GΩ 5 GΩ 10 GΩ 50 GΩ
DC-1 kHz 0.182 0.128 0.057 0.041 0.018
DC-3 kHz 0.331 0.234 0.099 0.074 0.033
DC-10 kHz 0.574 0.406 0.181 0.128 0.057
No patch-clamp amplifier can meet these ideal specifications. The difference between a
value in the table and the actual value is the excess noise. The excess noise is always pro-
portionately worse in the wider bandwidths.
ã=WARNING: The specified noise in a 10 kHz bandwidth will be misleadingly low if the
headstage does not have a 10 kHz bandwidth. If the manufacturer only specifies the
10–90% rise time or the time constant, use the conversion factors given in the General
section to verify that the bandwidth is sufficient to make the noise specification meaning-
ful.
C D
Cable capacitance Data acquisition parameters 197
filtering capabilities 174 D/A update rate 198
Cabling data bandwidth 198
affect on noise and bandwidth 171 filtering 199, 201
Capacitance gain 197
input 249 offset 197
X
Xenopus oocytes
definition 98
patch clamping of 101
recording from 97
two-electrode voltage clamping of 99
Z
Zap 69