Webster User's Manual
Webster User's Manual
1. Instrument Structure
The structure of Webster hardness testers is shown as Fig. 1:
2Pivot screw 4Reset key 6Dial head 8Indenter cylinder 10Indenter 12Return spring
The indenters of each type of Webster hardness testers are different. The shapes of three types of indenters are shown as Fig.2.
Fig. 2
Indenter Shapes
63-105 HRF
18-100 HRE
W-B92
50-92 HRB
3. Operating Method
Put the specimen between the anvil and the indenter and press down the handle until the bottom is felt. At the time the dial indicator will point at a reading which is the hardness value obtained. Excessive pressure beyond this limit will not damage the tester, but it is unnecessary. Hold tight the handle when reading the value. Any torsion or movement will make the reading incorrect during the test.
spring. Calibrationof the Full Scale is actually to set a benchmark for the scale 20. Calibration of the load spring is actually to provide with a reference point for the instrument, and this reference point is the known hardness value of the standard hardness block. After calibrating, the full scale should be verified again and its reading should conform to 4.1.1. 4.2.1 Calibration of Full Scale Hold tight the handle directly to the bottom without putting the specimen in the opening of the hardness tester. As shown in Fig. 3, regulate the full-scale adjusting screw with a small screwdriver to bring the indicator at Scale 20. If the pointer pointing at a smaller value than 20, the adjusting screw should be regulated clockwise; if the indicated value is greater than 20, the adjusting screw should be regulated anticlockwise. If the pointer can not be set to the full scale 20 by regulating the adjusting screw, replace the indenter according to the steps in Section 5.
4.2.2 Calibration of the Load Spring Screw off the pivot screw, and take out the lower handle from the main frame, then remove the dial head from the indenter cylinder, while leaving the cylinder in the frame. Then the adjusting screw with grooves can be seen in the indenter cylinder. As shown in Fig. 4, rotate the adjusting screw with the special wrench equipped with the instrument. If the reading of the hardness block is smaller, rotate the adjusting nut anticlockwise; otherwise, rotate the adjusting nut clockwise. When the nut is rotated 1/4 circle, the scale reading will change 2-3 units. Reassemble the tester and verify the hardness block after regulating it. Repeat the above steps until the full scale and the reading of the hardness block are both according with the requirements in Section 4.1. DO NOT rotate the adjusting screw at will except normal calibration. DO NOT calibrate the readings of the testing standard hardness blocks by regulating the adjusting screw. 4.2.3 Calibration of the B-1 Webster Hardness Tester Put the calibrating sheath on the anvil first; then calibrate the instrument according to the calibrating procedure in 4.2.
the dial head from the indenter cylinder, but the cylinder still stays in the frame. Then the adjusting screw with grooves can be seen in the indenter cylinder. As shown in Fig. 4, rotate the adjusting screw with the calibration wrench equipped with the instrument, then take out the indenter and replace with a new one. Calibrate the instrument again according to the calibrating procedure in 4.2. After the indenter is replaced, adjust the pressure of the load spring with the adjusting nut, when the resistance of the load spring is sensed, only one tight rotating of the nut is enough. The indenter tip could be damaged by the over pressure of the load spring at first adjustment.
6. Verification Report
The verification report should include the following information: Material brand and name Type and serial number of the hardness tester Reading of each test (accurate to 0.5 scale) Average hardness value (accurate to 0.5 scale) No. of standard applied Inspector and date
7. Standard Package
1 Tester 1 standard hardness block (attached with measurement and inspection report) 1 spare indenter 1 wrench 1 small screwdriver 1 instruction manual 1 carrying case
8. Optional Accessories
spare indenter standard hardness block spare dial glass
9. Maintenance
The testers of this series are precision instruments, of which the service life mainly depends on the correct method of use and timely and proper maintenance. The following items should be taken into account: 9.1 Prevent contamination. This instrument should often be cleaned with the soft cloth to remove the dust and dirt on the surface of it, especially on the anvil and supporting surface of the cylinder. Specially prevent the dirty things to get in from the dial seams or indenters, otherwise, the filthy things from those parts might influence on the testing precision or cause malfunction. 9.2 Prevent rust. Although all the parts of instrument have been treated to prevent rust, the improper maintenance or long time of use on the production spot without paying attention to maintenance may also cause the rust to some components. When using it, pay special attention to its waterproof performance except clean it frequently, because many parts will bear rust with water, the water in the dial may cause deadly rust to make the instrument useless. 9.3 Prevent falling off. This instrument is made up of many precision components, falling off of it will result in the permanent damage to some parts.
9.4 Disassembly forbidden. The testing accuracy of this instrument is guaranteed by the good collaboration of some of the precision parts. This collaboration is not easy for the nonprofessional people to master. Consequently, disassembly of the parts is not allowed except for those allowed in calibration, otherwise this will bring you trouble and the warranty of this instrument will terminate.
Table 2 Hardness Conversion Table for W-20 Series Hardness Testers Webster HW 18 17 16 15 14 13 12 11 10 9 8 7 6 5 4 Rockwell E HRE 101 97 92.5 88 84 79.5 75 71 67 62.5 58 54 49.5 45 41 Rockwell F HRF 98.5 95 91 87.2 83 78 74 70 66 62.5 58 54 50 46.5 Vickers HV 131 119 108 99 91 83 78 73 69 65 61 58