G.8262 SyncE Conformance Testing
G.8262 SyncE Conformance Testing
1. 2. 3. 4. 5. 6. 7. 8. 9.
Hardware and Software required ............................................................................................. 2 Connecting an EEC to the Paragon-X ..................................................................................... 3 How to Configure the Paragon-X for ALL G.8262 Tests .......................................................... 5 Measuring Frequency Accuracy G.8262 Section 6 ............................................................... 9 Pull-in, Hold-in and Pull-out ranges G.8262 Section 7 ........................................................ 11 Wander (Noise) Generation G.8262 Section 8 ................................................................... 15 Jitter Generation G.8262 Section 8.3.................................................................................. 19 Wander (Noise) Tolerance G.8262 Section 9 ..................................................................... 23 Jitter Tolerance G.8262 Section 9.2 ................................................................................... 29
10. Wander (Noise) Transfer G.8262 Section 10 ...................................................................... 35 11. Phase Transient Response G.8262 Section 11 .................................................................. 44 12. Appendix 1 - G.8262 Testing; Practical interpretation guidance............................................. 54
Page 1 of 58
Page 2 of 58
The front panel of the Paragon-X provides the following interfaces for testing: 1. 100MbE Electrical or Optical (SGMII SFP) 2. 1GbE Electrical or Optical (SFP) with option 110 fitted 3. 10GbE Optical (XFP or SFP+) with option 111 fitted
10GbE Port 1
100MbE/1GbE Port 1
100MbE/1GbE Port 2
PC Controller Port
Page 3 of 58
The Paragon-X accepts the following reference clocks which should be applied to one of the Reference Inputs on the back panel: 1. 2.048MHz 2. 10MHz 3. E1 (2.048Mb/s) 4. DS1 (T1) (1.544Mb/s)
There are two reference inputs BNC with 75 impedance (the connector on the left in the figure). Bantam connector with 100 impedance. o o o 3 pole bantam jack 4.39mm Diameter Tip; Positive, Ring; Negative, Sleeve; Ground
Page 4 of 58
Page 5 of 58
g) Verify that the SyncE Clock Rx > Tx is not checked h) Select the reference clock source. i) Select the line rate and Electrical/Optical (100M/1GbE/10GbE). Note: If using SFPs, XFPs or SFP+s, BOTH Port 1 and Port 2 optical transceivers must be inserted into the Paragon-X. j) Select the Close button to close the Physical Settings window, and then select Close again to close the setup interface window. Note: once you have selected Sync-E mode, the Paragon-X GUI will display SYNCE LOCK as grey (This verifies that the SyncE Rx > Tx box is unchecked, meaning that the Paragon-x is using its internal or external clock. If the SyncE Rx>Tx box were checked in the required Tx+Rx mode, there would be no clock).
Page 6 of 58
ESMC Generation
Ethernet Synchronisation Message Channel (ESMC) is a Point to Point Protocol in which each EEC generates and terminates an ESMC PDU. This ESMC PDU is normally transmitted once per second on the Slow Protocol Channel and contains the Quality Level (QL) of the associated clock . The receiving EEC uses this QL to select the best quality clock. The EEC then regenerates the ESMC PDU with the appropriate QL TLV value to indicate the quality of the clock being transmitted to the next EEC. If the EEC supports ESMC, it is possible for the Paragon-X to use this functionality to indicate if an EEC switches clock reference or goes into holdover. The Paragon-X can generate ESMC messages on both Ports 1 and 2. For G.8262 conformance testing, ESMC messages are generated on Port 2 with a user- defined Quality Level (QL) such as PRC.
b) Check the box Enable TX+Rx Mode. This sets the instrument in terminate mode.
Page 7 of 58
c) Select Port 2 tab (the output from the Paragon and input to EEC) d) Select SSM code QL-PRC, which tells the EEC that the clock from the Paragon-X is of PRC quality. e) Set IPG (inter-packet gap) rate to 1000mS. This makes the instrument send ESMC packets once per second. f) Click and then
g) The ESMC Generation button will go red to indicate the Paragon-X is generating ESMC packets. After any setup changes to the Paragon-X, please ensure the EEC has had time to settle before making any measurements
Page 8 of 58
Input Stimulus
EEC Option 1 Free run EEC Option 2 Hold over
Pass/Fail Criteria
+/- 4.6 ppm
Notes
Recommend to test for an hour, longer if close to limits Recommend to test for an hour, longer if close to limits
Measurement Setup
a) Connect the EEC to Paragon-X as shown above b) Set up the Paragon-X GUI as per Section 3.
Measurement Process
a) Select the Start Capture button to start measurement.
b) To stop the measurement after a pre-defined period for running the test, select Stop Capture button.
Page 9 of 58
c) Paragon-X will provide the ppm frequency accuracy at the bottom of the graph in red text (as shown in the screen shot at the top of the next page).
Page 10 of 58
Input Stimulus
EEC Option 1 and EEC Option 2 Apply a large Frequency offset ensuring EEC is in holdover. Reduce offset until EEC locks.
Pass/Fail Criteria
EEC starts unlocked with large offset applied EEC locks before offset reaches +/- 4.6ppm
Notes
Lock can also be monitored by using ESMC (if supported)
Input Stimulus
EEC Option 1 EEC Option 2 EEC is locked to the clock from the Paragon-X. The Frequency is then offset to +/-4.6ppm
Pass/Fail Criteria
Not Applicable
Notes
Page 11 of 58
Input Stimulus
EEC Option 1 EEC is locked to the clock from the Paragon-X. The Frequency is then offset until the EEC unlocks
Pass/Fail Criteria
EEC should remain locked at an offset at +/-4.6ppm but lock should extend beyond this. Not Applicable
Notes
G.8262 states this is for further study
EEC Option 2
Measurement Setup
a) Connect the EEC to Paragon-X to EEC as shown in the diagram at the beginning of this section b) Set up the Paragon-X GUI as per Section 3, including setting up ESMC with QL=PRC if using ESMC to monitor which clock the EEC is locked to.
Measurement Process
a) select the Start Capture button to start measurement.
Page 12 of 58
c) In the Frequency Offset window, add the Frequency Offset required and click the Apply Offset button
d) To remove the Frequency Offset and return it back to 0ppm click the Remove Offset button e) To stop the measurement after a pre-defined period for running the test, select Stop Capture button.
f) The Paragon-X TIE graph indicates if the EEC is in or out of lock as shown in the screenshot below
Page 13 of 58
Page 14 of 58
Input Stimulus
EEC Option 1
(Constant Temp)
Pass/Fail Criteria
MTIE & TDEV Pass/Fail masks shown in G.8262 Section 8.1.1 MTIE Pass/Fail masks shown in G.8262 Section 8.1.2. MTIE & TDEV Pass/Fail masks shown in G.8262 Section 8.1.2
EEC Option 1
(Temp effects)
EEC Option 2
(Constant Temp)
G.8262 Section 8.2 also mentions measurements in Non-locked mode and refers to section 11.2 Long-term phase transient response (Holdover) and will not be covered in this section.
Measurement Setup
a) Connect the EEC to Paragon-X to EEC as shown in the diagram at the beginning of this section b) Set up the Paragon-X as described in section 3, including setting up ESMC with QL=PRC if using ESMC to monitor which clock the EEC is locked to.
Measurement Process
a) Select the Start Capture button to start measurement.
Page 15 of 58
b) To stop the measurement, select Stop Capture. should run for approx 3,000 seconds
e) A separate window will open showing the TIE in graphical form at the top, with an MTIE/TDEV graph at the bottom.
Page 16 of 58
f)
The MTIE/ TDEV analysis can be carried out against the G.8262 masks for Wander Generation which can be selected from the measurement mask pull down selection in the left hand window.
g) Pass/Fail indication against the masks is shown also in this left hand window and will have a green background for Mask Pass and a red background for Mask Fail
Page 17 of 58
Page 18 of 58
Input Stimulus
Pass/Fail Criteria
ii. G.8262 treats 1G electrical as for further study. Calnex recommends to use the same 0.5UIpp limit until further defined by ITU.
a) Operating mode must be set to Sync-E, and Sync-E jitter selected as a measurement.
Page 19 of 58
Interface rate choice will then automatically be restricted to 1Gbit/s (Electrical or optical) or 10Gbit/s (100Mbit/s excluded). For 10Gbit/s optical, device type is automatically restricted to SFP+ (XFP excluded). Normally, the operating configuration is set to Tx+Rx mode.
b) Configure capture contains a jitter measurement threshold setting. If the checkbox is ticked, the capture (result) screen will indicate pass/fail of the peak-peak jitter result currently selected, relative to this limit. Note that the threshold result is compared to the instantaneous graph result and so its behaviour is different between long-term and short-term peak-to-peak graph selections.
c) G.8262 specifies a pp jitter measurement time of 60s. The global default capture configuration in the Paragon-X is manual (indefinite). Selecting Sync-E jitter measurement under operating mode forces the configure capture timing control to timed mode, 60s. Select other periods or manual if required, eg for experimentation and observing long -term device behaviour. However, 60s must be selected for making a G.8262 compliance measurement.
Page 20 of 58
d) Initiate jitter measurement using the start capture key. This automatically displays the statistics window, jitter measurement tab. This tab shows long-term (ie over the total elapsed time) pp and RMS jitter results. It also shows the measurement bandwidth, which is fixed depending on interface type selected. Limit checking is not applied to the jitter measurement tab. The tab can be closed, and re-opened any time via the Statistics and Results key.
e) The main capture screen shows table entries updated once per second for long-term pp, longterm RMS and short-term (one second) pp jitter results. On the graph, use Graph Display Mode to select a result for graphing. The measurement bandwidth is also displayed alongside the Y-axis legend on the graph. Note that sample periods (1-second intervals) are numbered based on the start time of each onesecond period (ie, starting from 0), so for example a 60-second total measuring period will show as containing sample periods 0-59. Compare the 60s long-term peak-to-peak value to the G.8262 0.5UI limit.
Page 21 of 58
Page 22 of 58
Input Stimulus
EEC Option 1 MTIE Wander Table 7/Figure 5 TDEV Wander Table 8/Figure 6 Sinusoidal Wander Table 9/Figure 7
Pass/Fail Criteria
The EEC is; i. Maintaining the clock within performance limits. ii. Not causing any alarms. iii. Not causing the clock to switch reference. iv. Not causing the clock to go into holdover. The EEC is;.
Notes
To Check whether the EEC is switching references or going into holdover, the Paragon can measure the wander and/or ESMC QL of the EEC output. To Check whether the EEC is switching references or going into holdover, the Paragon can measure the wander and/or ESMC QL of the EEC output.
EEC Option 2
i. Maintaining the clock within performance limits ii. Not causing any alarms. iii. Not causing the clock to switch reference. iv. Not causing the clock to go into holdover.
Measurement Setup
a) Connect the EEC to Paragon-X to EEC as shown in the diagram at the beginning of this section b) Set up the Paragon-X as described in section 3, including setting up ESMC with QL=PRC if using ESMC to monitor if the EEC is switching clock references or going into holdover.
Page 23 of 58
Measurement Process
a) Select the Wander and Jitter Generation button on the Paragon-X GUI.
c) Select the Wander Tolerance Tab and the following window will open
Page 24 of 58
d) There are three methods of generating wander into the EEC: i. ii. iii. e) MTIE/TDEV Wander Fastest and most effective way to evaluate EEC Table Sinusoidal Wander Can be used for finding Maximum Tolerable Wander Single Sinusoidal Wander Can be used for troubleshooting MTIE/TDEV Wander The Paragon-X generates MTIE and TDEV wander as defined in G.8262.
i. Select the wander mask required from the drop down list. The mask and the maximum running time will be shown on the right hand side of the window. G.8262 Option 1 MTIE - Running Time is 1000s G.8262 Option 1 TDEV - Running Time is 12000s G.8262 Option 2 TDEV - Running Time is 12000s ii. Click Generate Wander to start the test. The elapsed time will be displayed on the bottom right hand side of the window. iii. The test will stop after the max running time has elapsed, the test can also be stopped manually by clicking Stop Wander. f) Table (Sinusoidal) Can be used to test Maximum Tolerable Wander
Page 25 of 58
The user can enter up to 10 different wander parameter sets in the table. The Paragon -X will then automatically add each of the specified wander sets in turn giving full indication of progress in the Status column. Switching between the different sets is always done at a zero crossing to prevent phase steps. i. Enter the frequency, amplitude, and dwell time (number of cycles the frequency/amplitude pair will be run) for each wander test point. The same frequency with different amplitudes can be entered to find maximum tolerable wander The Restore Defaults button when checked will restore the values to that defined in table 9 in G.8262. Only rows that have the enable check box ticked will be executed in the test. To skip over a selection, un-tick the enable box for that selection. ii. iii. iv. Click Generate Wander to start the test Click Stop Wander to stop the test. The test will terminate at the next zero crossing A pop up box stating how long to the next zero crossing is displayed, click the Stop Immediate button if it is desired to stop the test instantly.
g) Single Sinusoidal Wander Can be used for troubleshooting issues at a specific frequency
Page 26 of 58
i. Enter the frequency and amplitude of the desired wander ii. Click Generate Wander to start the test iii. Click Stop Wander to stop the test. The test will terminate at the next zero crossing of the wander frequency A pop up box stating how long to the next zero crossing is displayed. h) G.8262 states that with the wander applied that the EEC should i. Maintain the clock within the prescribed performance limits (the exact performance limits are for further study) ii. Not cause any alarms iii. Not cause the clock to switch references iv. Not cause the clock to go into holdover For further insight, use the Paragon-X ESMC generation/capture capability and TIE graph to check whether the EEC is switching clock reference or going into holdover. i) Ensure the Paragon-X has been set up to Generate ESMC messages with QL=PRC as described in section 3 of this document. j) k) Select the Start Capture button to start the measurement. and then Show 2
nd
To show both the TIE and ESMC graph at the same time click Graph and then select ON
l)
To change what a graph is displaying, i. click one of the graphs (the selected graph will be non-Grey ii. Click and then Graph Display Mode, finally select the graph to display
Page 27 of 58
Time Interval Error (TIE) vs Nominal ESMC Rx Quality vs Time: (Port 1 Rx)
The screenshot above shows an EEC that has switched references or gone into holdover (top graph is ESMC, bottom graph shows the difference in Frequency between the reference and the recovered clock from the EEC) m) To stop the measurement, select Stop Capture.
Page 28 of 58
Input Stimulus
EEC Option 1 and option 2 Test packet stream from Paragon-X or thru-mode traffic Superimpose jitter to G.8262 9.2 fig 9
Pass/Fail Criteria
The EEC does not drop packets for all jitter values presented
Notes
Either as pass/fail with presented jitter values at the mask, or a margin test with jitter values above mask. Allow settling time at each measurement point
Page 29 of 58
Measurement Setup
a) Open Operating Mode dialog, change Operating Mode to Sync-E
Page 30 of 58
b) Open Setup Interface dialog c) d) Select Tx+Rx Mode Select 1GbE or 10GbE Line Rate
Page 31 of 58
Page 32 of 58
f)
From the TestPackets tab, the user configures the test packet they wish to generate.
g) From the Jitter Tolerance Table tab, the user enters the Frequency, Amplitude, and Dwell Time for each row. h) From the Jitter Tolerance Table tab, the user clicks the Start Test button. Test Packet generation will be automatically controlled as each row in the Jitter table is executed. i) The Jitter and Test Packet generation will automatically stop after all the enabled table rows have been executed. j) A PASS or FAIL result for each executed row will appear on the Jitter Tolerance Table tab. The result will indicate if any Dropped packets were detected during the test.
Page 33 of 58
k) In the Max tolerable jitter mode, operation is similar except a percentage above or below
mask can be set for the generated jitter amplitudes.
Page 34 of 58
Input Stimulus
Pass/Fail Criteria
Notes
There is no definition of the input stimulus to be used in G.8262. Without further guidance from the Standards, it is suggested that the amplitude and frequency values associated with mask points labelled f1, f2 & f3 on G.8262 Section 9.1.1, Table 9 & Figure 7 are used.
EEC Option 1
Not defined
The phase gain of the EEC should be smaller than 0.2 dB (2.3%).
EEC Option 2
Measure EEC output against TDEV Pass/Fail masks shown in G.8262 Section 10..2 Table 13/Figure 11
Measurement Setup
a) Connect the EEC to Paragon-X to EEC as shown in the diagram at the beginning of this section
Page 35 of 58
b) Set up the Paragon-X as described in section 3, including setting up ESMC with QL=PRC if using ESMC to monitor which clock the EEC is locked to.
Measurement Process
c) Select the Wander Transfer Tab and the following window will open
Page 36 of 58
Page 37 of 58
b) Enter the frequency, amplitude, and dwell time (number of cycles the frequency/amplitude pair will be run) for each wander test point. i. The Restore Defaults button when checked will restore the values to that defined in table 9 in G.8262. ii. Only rows that have the enable check box ticked will be executed in the test. To skip over a selection, un-tick the enable box for that selection. iii. Be careful to choose a dwell time of 300 cycles for any wander frequency of 1Hz or higher. Additionally, choose the shortest sample period (configure capture button) available for calibration and definitive measurement runs.
c) To calibrate the Paragon-X, connect a short Ethernet cable between Port 1 and Port 2 and click the button.
Page 38 of 58
e) Remove the short Ethernet cable between Port 1 and Port 2 and connect the EEC under test as shown in the setup diagram at the start of this section.
f)
g) The Paragon-X will show the status of the test, the measured Gain (dB) value and also a Pass/Fail indication on the right hand side of the window.
Page 39 of 58
G.8262 Option 1 Wander Transfer Test: Paragon-X single frequency test This capability can be used for fault finding issues at specific frequencies
a) Select the Single tab.
c) To calibrate the Paragon-X, connect a short Ethernet cable between Port 1 and Port 2 and click the button.
Page 40 of 58
e) Remove the short Ethernet cable between Port 1 and Port 2 and replace with the EEC under test.
f)
g) The GUI will show the estimated completion time at the bottom of the screen. The measured Gain (dB) value and also a Pass/Fail indication (Green/Red background) will be displayed.
Page 41 of 58
c) The amount of the time until completion of the test is shown at the bottom right hand side of the screen.
d) At any time during the test, it is possible to view an updated output TDEV graph by clicking the TDEV Results button. The TDEV graph will then be displayed.
Page 42 of 58
e) At the end of the test the Paragon-X will automatically show the final MTIE graph and show Pass/Fail against the MTIE mask.
Page 43 of 58
11. Phase Transient Response G.8262 Section 11 11.1 G.8262 Section 11.1 Short-term phase transient response
Input Stimulus
EEC input reference is lost for 15 seconds and a 2nd reference input signal, traceable to the same reference clock, is available simultaneously EEC input reference is lost for 15 seconds and a 2nd reference input signal, traceable to the same reference clock, is available simultaneously
Pass/Fail Criteria
Notes
To emulate the loss of the link either Change ESMC QL=DNU Remove the cable between port 2 and EEC
EEC Option 1
Maximum phase transient at the output due to reference switching to meet mask in G.8262 Figure 12
To emulate the loss of the link either: EEC output should meet MTIE mask defined by table 15/ Figure 14 in section 11.4.2 of G.8262 Change ESMC QL=DNU Remove the cable between port 2 and EEC
EEC Option 2
Measurement Setup
a) Connect the EEC to Paragon-X to EEC as shown in the diagram at the beginning of this section
Page 44 of 58
b) Set up the Paragon-X as described in section 3 (including ESMC generation if using the ESMC method of switching clock references, ensuring SSM code is set to PRC) c) Select Configure Capture , select 10mSecs sample period and then close.
Measurement Process
a) There are two methods for determining Phase Transient Response with the Paragon-X; i. Use ESMC Generation (if supported by the DUT). ii. Remove the link between Port 2 on the paragon-X and the EEC input port.
c) If using the Ethernet cable removal method please disconnect cable between port 2 and the EEC. d) If using ESMC method i. Select button. The following window will open
Page 45 of 58
ii. Ensure SSM code is set to QL PRC iii. In the SSM Code drop down menu select QL DNU/DUS and then click .
Measuring Results
The switching transient will be seen on the Paragon-X TIE graph. The Paragon-X graphs can show the TIE and also the ESMC QL for each port In the screenshot bellow the ESMC graph shows the time and QL change of the ESMC transition on Port 2 Tx of the Paragon-X. This can then be compared in time to the TIE graph showing the Transient.
Method of evaluating the capture for phase transient response is dependent upon whether EEC Option 1 or Option 2 is being evaluated.
EEC Option 1 a) To view the output TIE for EEC Option 1, closely monitor the output TIE graph on Paragon-X for the duration (15s) of the test. While the slave clock is acquiring the new reference, the output phase transient should be within the limits of figure 12 of G.8262 (provided below)
Page 46 of 58
EEC Option 2 a) To view the output MTIE mask for EEC Option 2 ONLY, perform the following: b) To display the MTIE and TDEV graphs click on the bottom right hand side of the TIE
c) A separate window will open showing the TIE in graphical form at the top, with an MTIE/TDEV graph at the bottom.
Page 47 of 58
d) The MTIE analysis can be carried out against the G.8262 masks for Wander Transient which can be selected from the measurement mask pull down selection in the left hand window.
e) Pass/Fail indication against the masks is shown also in this left hand window and will have a green background for Mask Pass and a red background for Mask Fail
Page 48 of 58
Page 49 of 58
10.
Input Stimulus
Pass/Fail Criteria
Notes
To emulate the loss of the link either
EEC Option 1
Maximum phase transient and excursion at the output due to reference switching to meet mask in G.8262 Figure 13
Change ESMC QL=DNU Remove the cable between port 2 and EEC
Measure for 10,000 seconds (approx 3hrs) Maximum phase transient and excursion at the output due to reference switching to meet mask in G.8262 11.2.2 and table 14, st nd including 1 and 2 derivatives of phase vs time To emulate the loss of the link either: Change ESMC QL=DNU Remove the cable between port 2 and EEC
EEC Option 2
Page 50 of 58
After allowing the DUT to settle, leave for at least 900s (15 minutes), measure the wander on Port 1. The resulting graph should be flat and the Offset measurement at the foot of the Wander TIE graph should indicate 0.000 ppm. The ESMC received from the DUT on Port 1 should also indicate PRC.
Restart the wander measurement and then set the ESMC on Port 2 to DNU (Do Not Use). This should cause the the DUT to search for an alternative master clock signal to lock to. If there are no other external signals then the DUT will use its internal clock. The ESMC messages on port 1 should reflect this change with the appropriate Quality Level e.g. EEC1. The TIE graph should show the SyncE signal from the DUT start to drift off. Leave the capture running for a further 10,000 seconds (approx 3 hrs) minimum. Stop the measurement. The limits of operation are given by this mask reproduced from G.8262 fig 13:
Check also that the TIE measurement shows a calculated frequency offset (shown at bottom of TIE graph) within +/-4.6ppm. For EEC-2, proceed as above, except the limit for TIE is as described in G.8262 11.2.2 and table 14, illustrated here as a mask:
Page 51 of 58
1.00E+06
1.00E+05
1.00E+02
The limits for EEC-2 1 derivative TIE (fractional frequency offset) and 2 frequency drift) are also specified in table 14.
st
nd
Currently it is not possible to compare results against the masks directly in Paragon-X, but two methods of comparing results to G.8262 limits are available: Method 1: direct observation of results from the Paragon-X (for EEC-1 and EEC-2 TIE), using marker functions. Method 2: export of Paragon-X TIE results as a .csv file, with subsequent processing in Excel. This can be used for all cases, and is the only practical way to compute the 1 and 2 the case of EEC-2. On Paragon-X, after running the TIE measurement, use FILE, EXPORT to create a .csv file, then open with Excel and perform the analysis. Paragon-X exported TIE files show the TIE values as a single series of integers based on 1ns units. To build a table of TIE versus time (S), compute the S values from the position in the series times the sampling interval selected (default is 0.033 333 320 seconds, corresponding to 30 samples/s) and the TIE values from the cumulative value of each series value minus 33333320. Compute derivatives using the central difference approximation dTIE/dS = (TIE(n+1) - TIE(n-1))/0.066666640. The constants
st nd
differentials of TIE in
Page 52 of 58
used in the calculation need to be changed to match the original capture sample rate, if it is varied from its default 30/s. See Calnex document CX5001e (on our website) as a practical example in Excel of the use of a Paragon-x exported TIE file, with central difference differentiation examples.
Page 53 of 58
Pass/Fail Criteria
Notes
Recommend to test for an hour, longer if close to limits Recommend to test for an hour, longer if close to limits
Pass/Fail Criteria
EEC starts unlocked with large offset applied EEC locks before offset reaches +/- 4.6ppm
Notes
Lock can also be monitored by using ESMC (if supported)
Pass/Fail Criteria
Not Applicable EEC should remained locked at an offset at +/-4.6ppm
Notes
Page 54 of 58
Input Stimulus
EEC Option 1 EEC is locked to the clock from the Paragon-X. The Frequency is then offset until the EEC unlocks.
Pass/Fail Criteria
EEC should remain locked at an offset at +/-4.6ppm but lock should extend beyond this. Not Applicable
Notes
G.8262 states this is for further study
EEC Option 2
Input Stimulus
EEC Option 1
(Constant Temp)
Pass/Fail Criteria
MTIE & TDEV Pass/Fail masks shown in G.8262 Section 8.1.1 MTIE Pass/Fail masks shown in G.8262 Section 8.1.2. MTIE & TDEV Pass/Fail masks shown in G.8262 Section 8.1.2
Notes
MTIE Table 1/ Figure 1 TDEV Table 3/ Figure 2 MTIE Table 1/ Figure 1 TDEV G.8262 states for further study MTIE Table 4/ Figure 3 TDEV Table 5/ Figure 4
EEC Option 1
(Temp effects)
EEC Option 2
(Constant Temp)
Page 55 of 58
Input Stimulus
EEC Option 1 MTIE Wander Table 7/Figure 5 TDEV Wander Table 8/Figure 6 Sinusoidal Wander Table 9/Figure 7
Pass/Fail Criteria
The EEC is; v. Maintaining the clock within performance limits. vi. Not causing any alarms. vii. Not causing the clock to switch reference. viii. Not causing the clock to go into holdover. The EEC is;.
Notes
To Check whether the EEC is switching references or going into holdover, the Paragon can measure the wander and/or ESMC QL of the EEC output. To Check whether the EEC is switching references or going into holdover, the Paragon can measure the wander and/or ESMC QL of the EEC output.
EEC Option 2
v. Maintaining the clock within performance limits vi. Not causing any alarms. vii. Not causing the clock to switch reference. viii. Not causing the clock to go into holdover.
Wander Transfer (G.8262 Section 10) Input Stimulus Pass/Fail Criteria Notes
There is no definition of the input stimulus to be used in G.8262. Without further guidance from the Standards, it is suggested that the amplitude and frequency values associated with mask points labelled f1, f2 & f3 on G.8262 Section 9.1.1, Table 9 & Figure 7 are used.
EEC Option 1
Not defined
The phase gain of the EEC should be smaller than 0.2 dB (2.3%).
EEC Option 2
Measure EEC output against TDEV Pass/Fail masks shown in G.8262 Section 10..2 Table 13/Figure 11
Page 56 of 58
Pass/Fail Criteria
Notes
To emulate the loss of the link either Change ESMC QL=DNU Remove the cable between port 2 and EEC
EEC Option 1
Maximum phase transient at the output due to reference switching to meet mask in G.8262 Figure 12
To emulate the loss of the link either: EEC output should meet MTIE mask defined by table 15/ Figure 14 in section 11.4.2 of G.8262 Change ESMC QL=DNU Remove the cable between port 2 and EEC
EEC Option 2
Page 57 of 58
For more information on the Calnex Paragon-X and to take advantage of Calnexs extensive experience in sync and packet testing technologies, please contact Calnex Solutions on +44 (0) 1506 671 416 or email: [email protected]
Calnex Solutions Ltd Herkimer House Mill Road Enterprise Park Linlithgow West Lothian EH49 7SF United Kingdom Tel: +44 (0) 1506 671 416 Email: [email protected]
www.calnexsol.com
CX5001 v11 Jan 2013 This information is subject to change without notice