100% found this document useful (1 vote)
222 views2 pages

SG900 Specific Gravity

The Sarasota SG900 specific gravity analyser is designed for high accuracy measurement of gases with features including accuracy to 0.2% reading, fast response, and intrinsically safe design. It has applications in energy determination, blending control, and fuel gas monitoring. The document provides dimensional drawings, specifications, ordering information, and contact details.
Copyright
© Attribution Non-Commercial (BY-NC)
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
100% found this document useful (1 vote)
222 views2 pages

SG900 Specific Gravity

The Sarasota SG900 specific gravity analyser is designed for high accuracy measurement of gases with features including accuracy to 0.2% reading, fast response, and intrinsically safe design. It has applications in energy determination, blending control, and fuel gas monitoring. The document provides dimensional drawings, specifications, ordering information, and contact details.
Copyright
© Attribution Non-Commercial (BY-NC)
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

Sarasota Specific Gravity Analyser SG900

The Sarasota SG900 specific gravity analyser is designed for high accuracy measurement of gases.

Features
accuracy to 0.2% reading integral sample conditioning fast response absolute density, temperature and pressure measurements options to suit all applications intrinsically safe design corrosion resistant construction option for heated enclosure no on-site calibration required

Applications
energy determination blending control standard volume flow calculation fuel gas monitoring

Dimensional drawing
600mm 300mm

800mm

Detailed specification and ordering information: see overleaf


Thermo Measurement reserves the right to alter specifications without notice.

ISSUE: SG900/MAY2001/UK

800mm

Specification
System: Range: Accuracy: Repeatability: Temperature coefficient: Operating temperature: Sample inlet pressure: Weight: Unlimited 0.2% reading 0.02% span 0.01%/C -20 to +55C or as limited by gas dew point 20 bar maximum for basic systems 200 bar maximum for dry and wet gas systems Maximum 4 Bar A Spool - Ni-Span-C or FV520B Body - AISI 316L stainless steel Enclosure - GRP High accuracy 13 DIN Thermowell 4 wire PT100 IP65 Intrinsically safe with approved barriers Installation: Flow range: Outputs: Frequency related to density on 2 wire current modulated loop 6mA to 18mA 4 wire PT100 4-20mA pressure Electrical: Power supply: 16-28 VDC 10mA average (peak 18mA) (density transmitter) 9-30 VDC 20mA (pressure transmitter) Screw terminals Nett 30kg max Shipping 65kg max 14 tubing compression fitting Ideally 4-20 litres/min

Exhaust pressure: Materials of construction:

Temperature measurement: General: Environmental protection: Hazardous area certification:

Connections:

Ordering Information
Density meter model code format SG900 Sarasota Specific Gravity Analyser Signal output F Frequency output System type B D W Basic system Dry gas Wet gas Internal heater O No internal heating E S Electrical heater Steam heater Spool material Z Ni-Span-C Y FV520B Certification S Non hazardous / safe area application I Intrinsically safe EEx ia IIC T4/T6 Options M N T Wetted parts material certificates (DIN 50049 BS EN 10204 3.1b) NACE conformance Traceable calibration certificate for density meter

Example: SG900-F-D-O-Z-I
SG900 specific gravity analyser with frequency output, suitable for dry gas, no internal heating, with Ni-Span-C spool for intrinsically safe application.

Thermo Electron Corporation Kings Worthy Winchester, Hampshire SO23 7QA, United Kingdom Tel: +44 (0) 1962 625000 Fax: +44 (0) 1962 885530 email: [email protected] www.thermo.com

Thermo Electron Corporation 9303 W. Sam Houston Parkway South Houston Texas 77099 5298 USA Tel: +1 (713) 272 0404 Fax: +1 (713) 272 2273

HART is a registered trademark of the Hart Foundation