0% found this document useful (0 votes)
244 views

Application Notebook

application note for sample preparation in food, forensic, medicine ...

Uploaded by

mikicacica
Copyright
© Attribution Non-Commercial (BY-NC)
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
244 views

Application Notebook

application note for sample preparation in food, forensic, medicine ...

Uploaded by

mikicacica
Copyright
© Attribution Non-Commercial (BY-NC)
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 48

Applications Notebook

Issue 1, June 2009

Thermo Scientic Hypercarb Columns

A unique solution for difcult separations

Part of Thermo Fisher Scientic

Table of Contents
Introduction

Thermo Scientific Hypercarb Columns Introduction ......................................................................................................... 4

Table of Contents
2

Application Notes

Biochemical Porous Graphitic Carbon for the Sample Preparation of Hydrophilic Biomolecules.................................................................................. 6 Porous Graphitic Carbon for the LC/MS Analysis of Hydrophilic Biomolecules.................................................................................. 7 Analysis of Glycopeptides Using Porous Graphite Chromatography and LTQ Orbitrap XL ETD Hybrid MS....................................................................... 10 The Use of Porous Graphitic Carbon LC-MS for the Analysis of Underivatised Carbohydrates from Wheat Stems.................................................... 17 Food Safety Quantitation of Acrylamide in Food Samples on the TSQ Quantum Discovery by LC/APCI-MS/MS .............................................................................................. 20 Environmental Fast LC Separation of Triazine Herbicides at Elevated Temperature .............................. 23 Analysis of Polar Metabolites of Atrazine in Ground Waters Using Hypercarb as SPE Sorbent Coupled On-Line with Hypercarb LC Column................... 27 Fast and Versatile Analysis of Desphenyl-Chloridazone and Methyl-Desphenyl-Chloridazone in Surface, Ground and Drinking Water Using LC-MS/MS and EQuan.......................................................... 30 Clinical Determination of Leucine and its Isomers by LC-MS/MS Using a Porous Graphitic Carbon Column ........................................................................... 33 Determination of Occupational Exposure to Toluene, Xylene and Styrene Through Metabolite Monitoring Using Isocratic HPLC ............................................. 35 Food and Beverage Porous Graphitic Carbon for Inorganic Ion Analysis in Drinking Water.......................... 37

Legal Notices
2009 Thermo Fisher Scientific Inc. All rights reserved. All trademarks not specifically referenced are the property of Thermo Fisher Scientific Inc. and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific Inc. products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Not all products are available in all countries. Please consult your local sales representative for details.

Applications Review

Applications Reference Guide (Application by Solute Type).......................................... 40 References............................................................................................................... 42

Table of Contents

Application Chromatograms

Biochemical ............................................................................................................. 44 Allantoin........................................................................................................................................44 Ceramides ....................................................................................................................................44 Cyclic Monophosphates ............................................................................................................44 L-Carnitine ....................................................................................................................................44 Purines and Pyrimidines (UHT-LC) ...........................................................................................44 RNB-Glycopeptides ....................................................................................................................44 Food Safety .............................................................................................................. 45 Aatoxins......................................................................................................................................45 Methylamines in Fish .................................................................................................................45 Environmental........................................................................................................... 45 Nonylphenol Isomers ..................................................................................................................45 Quaternary Ammonium Salts ....................................................................................................45 Water Pollutants .........................................................................................................................45 Clinical ................................................................................................................ 45-46 Arginine and Methylated Arginines ........................................................................................45 Creatine in Serum .......................................................................................................................46 Pharmaceutical ........................................................................................................ 46 Acyclovir.......................................................................................................................................46 Fosfomycin ...................................................................................................................................46 Glucosamine Sulfate ..................................................................................................................46 Tuberculostatics..........................................................................................................................46 Uracil and Metabolite ................................................................................................................46

Ordering Information

Hypercarb Columns, Hypercarb Drop-in Guard Cartridges, Hypercarb KAPPA Capillary Columns, Hypercarb Nanobore Columns, Hypercarb Specialized Column Hardware for High Throughput, Hypercarb Preparative Columns, Hypercarb High Temperature Columns ....................................................................... 47

Thermo Scientific Hybercarb Columns


Introduction
Porous graphitic carbon (PGC, Hypercarb) has unique properties as a stationary phase in high performance liquid chromatography (HPLC). Its chemical surface properties distinguish PGC from more conventional LC packings such as bonded-silica gels and polymers. PGC behaves as a strongly retentive alkyl-bonded silica gel for non-polar analytes, however its retention and selectivity behaviour towards polar and structurally related compounds is very different. PGC provides unique retention and separation of very polar compounds. Its surface is stereo-selective with the capability to separate geometric isomers, diastereomers and other closely related compounds. Hypercarb is stable throughout the entire pH range 0-14, and is not affected by aggressive mobile phases. Its compatibility with all solvent systems enables separation of a wide range of polarities within a single chromatographic run. The selectivity of the Hypercarb packing is different from the selectivity of silica and polymeric phases. Its retention mechanism is different from conventional C18 columns. Reference 1 gives an in-depth review of its HPLC behaviour and application areas.
Property
Particle shape Specific surface area Median pore diameter Pore volume Mean particle diameters Spherical, fully porous 120 m2/g 250 0.7 m3/g 3, 5 m 7 m 30 m 75% 100% > 400 bar

Physical and Chemical Properties of PGC


PGC particles are spherical and fully porous with a porosity of approximately 75%. The surface of PGC is crystalline and highly reproducible and does not contain micropores. At the molecular level, PGC is made up of sheets of hexagonally arranged carbon atoms linked by the same conjugated 1.5-order bonds which are present in any large polynuclear aromatic hydrocarbon.2 In principle, there are no functional groups on the surface since the aromatic carbon atoms have fully satisfied valencies within the graphitic sheets. Table 1 lists the more important physical properties of PGC. The requirements placed on its physical properties are similar to other HPLC supports where factors such as narrow particle size distribution are essential to the ultimate performance of the phase, if good bed uniformity and low operating pressures are to be achieved. PGC also has a tight pore size distribution with a mean value around 250 , allowing for good mass transfer of a wide range of analyte shapes and sizes. Surface homogeneity and absence of highly adsorptive sites are essential for good peak symmetry. PGC meets all the conventional operating criteria of a chromatographic support.

Introduction
4

To meet requirement of
No micropores Retention linearity and loading capacity Mass transfer for wide range of analytes shapes and sizes Analytical HPLC columns Preparative HPLC columns SPE applications Mass transfer within particles Chemical stability Operational particle stability; pressure gradients in packing process

Porosity %C Mechanical strength Table 1: Physical properties of PGC

Retention Mechanisms on PGC


On a molecular scale, the surface of the graphite is flat and highly crystalline unlike that of alkyl-bonded silicas, which possess a brush type surface with the bonded phase and residual silanols. Consequently, the PGC mechanism of interaction is very different. The retention by graphite from aqueous/organic eluents is determined by the balance of two factors: (1) hydrophobicity, which is primarily a solution effect that tends to drive analytes out of solution and (2) the interaction of polarisable or polarised groups in the analyte with the graphite (these are additional to the normal dispersive interactions). The strength of interaction depends on both the molecular area of an analyte in contact with the graphite surface and upon the nature and type of functional groups at the point of interaction with the flat graphite surface. The more planar the analyte, the closer its alignment is to the graphite surface, and so the greater the number of points of interaction possible hence, maximum retention. Retention is reduced for highly structured, 3-dimensional and rigid molecules that can contact the surface with only a small part of their surface, compared with planar molecules with the same molecular mass. This is illustrated in Figure 1.

Retention of Polar Compounds on PGC


In a traditional reversed phase (RP) system, analyte retention increases as its hydrophobicity increases. This is due to the increased dispersive interactions that take place between the stationary phase and the analyte. Conversely, as the polarity of the analyte increases, analyte-solvent interactions begin to dominate and retention is reduced. This simple observation holds true for all reversed-phase systems with the exception of PGC. For Hypercarb columns, it has been observed that in some cases retention increases as the polarity of the analyte increases. This effect has been called the polar retention effect on graphite or PREG. The effect of PREG makes Hypercarb columns particularly useful for the separation of highly polar compounds, such as carbohydrates, and compounds with several hydroxyl, carboxyl and amino groups, which are difficult to retain on conventional alkyl-silica phases. PREG defines the ability of molecules having lone-pair or aromatic-ring electrons to apparently interact through an electron transfer mechanism to the electronic cloud of the graphite. PREG is particularly pronounced when the polar groups are attached to a benzene ring and other larger aromatic systems. Knox et al.2 have attributed this to some type of orbital overlap between the conductivity electrons in graphite and lone pair and/or electrons in analytes. The polarizable properties of the graphite hold the key to understanding the mechanism by which polar molecules are retained at the surface (Figure 2).

Introduction

Figure 1: Effect of the solute shape on the strength of the interaction with the graphite surface: (a) Good alignment of planar molecule to the flat graphite surface; (b) Poor alignment of non-planar molecule to the flat graphite surface.

Figure 2: Schematic representation of polar analyte retention in which (a) positive charge and (b) negative charges approach the graphite surface, resulting in a charge-induced dipole at the graphite surface.

References
1. L. Pereira, J. Liq. Chrom. & Rel Technol., 2008, 31, 16871731 2. J. Knox, P.Ross, Advances in Chromatography, 1997, 37, 73-119

Porous Graphitic Carbon for the Sample Preparation of Hydrophilic Biomolecules


Luisa Pereira, Thermo Fisher Scientific, Runcorn, UK

BIOCHEMICAL
Application Notes
6

Introduction
Micro-scale solid phase extraction (SPE) can be used as a sample purication process to remove contaminants; this technique has the advantage of effectively handling limited sample volumes (low microlitre) to maximise sensitivity. When the analytes are very hydrophilic it is necessary to select a sorbent that provides good retention and minimises sample loss through breakthrough in the aqueous matrix. Desalting of hydrophilic peptides can be accomplished by using microscale tips packed with porous graphitic carbon. This enables the off-line preparation and clean-up of biological samples for further analysis and identication by mass spectrometry (MS). In this approach it is important that the sorbent in the tip is capable of retaining the hydrophilic analytes with no breakthrough in the aqueous matrix, and that good recovery of the retained analytes from the tip is achieved.

Figure 1. Recovery from HyperSep Hypercarb Tips for 2 peptides, RGES and DSDPR.

Results and Discussion


Following the procedure detailed in the experimental section, the recoveries were measured by comparison of the ESI-MS signal for the tip eluate with the ESI-MS signal for the solution of the same concentration. Recoveries are between 72 and 101%, as shown in Figure 1.

Goal
To demonstrate the capability of porous graphitic carbon (PGC) in micro-scale SPE of hydrophilic peptides.

Experimental
Peptides Arg-Gly-Glu-Ser (RGES) and Asp-Ser-Asp-Pro-Arg (DSDPR). Tips Thermo Scientic HyperSep Hypercarb Tips 10-200 L volume (part number 60109-212). Micro-scale SPE Protocol: Solvents: A H2O + 0.1%formic acid; B H2O/ACN (30:70) + 0.1% formic acid.
Tip conditioning: Aspirate and expel 5 times 20 L of solvent B. Aspirate and expel 5 times 20 L of solvent A. Sample loading (binding): Aspirate and expel 20 times 20 L of sample. Sample washing: Aspirate and expel 5 times 20 L of solvent A, discarding the expelled solvent each time. Sample elution: Aspirate and expel 20 times 20 L of solvent B, collecting the expelled solution in a clean microcentrifuge tube. Transfer solution to micro-vial for injection. A ow-through fraction from a proteolytic digest was simulated by diluting a solution containing RGES and DSDPR in Tris buffer (100 mM, pH 8.0) to concentrations of 0.1 and 0.5 ng/L respectively.

Conclusion
Micro-scale SPE with porous graphitic carbon packed tips gives good recovery of small hydrophilic peptides from buffer matrices.

Porous Graphitic Carbon for the LC/MS Analysis of Hydrophilic Biomolecules


Luisa Pereira, Thermo Fisher Scientific, Runcorn, UK

BIOCHEMICAL

Introduction
The sensitivity of the analysis of small hydrophilic peptides by mass spectrometric detection is often compromised by the presence of salts and non-volatile buffers. These peptides are not retained and, therefore, are often found in the flow-through fraction from a C18 LC column, the type of stationary phase most commonly used for the separation of proteolytic digests of proteins. The analysis of the flow-through fraction requires either a stationary phase that can retain the peptides away from the solvent front, where the biological salts and buffers elute, or a sample clean-up step to remove the salts.1 Porous Graphitic Carbon (PGC) is a material that provides strong retention of very polar compounds; the retention mechanism involves a charge-induced interaction of the polar analyte with the polarizable surface of the graphite.2 PGC is ideal to retain and resolve very polar, hydrophilic molecules, which are normally not retained under reversed-phase LC using typical MS compatible mobile phases. The work presented in this application note demonstrates the advantages of using PGC in the LC/MS analysis of a phosphopeptide and di-, tri- and penta-peptides containing polar and basic terminal amino acid residues. Analytical parameters investigated are chromatographic retention and resolution, and spectral cleanliness.

Experimental
Columns Thermo Scientic Hypercarb 5 m, 50 x 2.1 mm; (part number 35005-052130) Thermo Scientic Hypersil GOLD 5 m, 100 x 2.1 mm. (part number 25005-102130) Instrumentation Thermo Scientic Surveyor and Thermo Scientic LCQ Deca XP. Peptides Arg-Gly-Glu-Ser (RGES), Asp-Ser-Asp-Pro-Arg (DSDPR), Gly-Tyr (GY), Phe-Gln-pSer-Glu-Glu-Gln-Gln-Gln-Thr-GluAsp-Glu-Leu-Gln-Asp-Lys. LC/MS Conditions Mobile phase: A H2O + 0.1% formic acid; B ACN + 0.1% formic acid
Gradient: 5 to 100% B in 10 min Flow rate: 0.2 mL/min Temperature: 30 C Detection: +ESI

Application Notes

Results and Discussion Goal


To demonstrate the advantages of using porous graphitic carbon (PGC) in the LC/MS analysis of polar molecules of biological interest such as small hydrophilic peptides and phosphopeptides.

Phosphopeptide retention on PGC When analyzed under identical conditions the capacity factor for the monophosphopeptide on a porous graphitic carbon column is 3 times greater than when analyzed using an alkyl-silica stationary phase (Figure 1). On the PGC column

Figure 1: Comparison of the capacity factor of a monophosphorylated peptide on an alkyl-silica phase and PGC.

BIOCHEMICAL
Application Notes
8

Figure 2: Comparison of the retention of 3 hydrophilic peptides on alkyl-silica and porous graphitic carbon. PGC provides higher retention and different selectivity. Analytes: 1. Arg-Gly-Glu-Ser (RGES); 2. Asp-Ser-Asp-Pro-Arg (DSDPR); 3. Gly-Tyr (GY)

Tris

Tris

Figure 3: Comparison of the spectra obtained for 3 hydrophilic peptides in Tris buffer on alkyl-silica and porous graphitic carbon stationary phases

the retention mechanism is a combination of hydrophobic or dispersive interaction with the hydrophobic amino acid chain, and a polar interaction between the phosphate group and the polarizable surface of the graphite.

Di- to penta- hydrophilic Peptides In Figure 2 the retention of a di-, a tetra- and a penta-peptide is compared on the alkyl-silica phase and on PGC. On the alkyl-silica phase, typically used in the separation of proteolytic digests, RGES elutes at the solvent front, closely followed by DSDPR. The basic (Arg) and alcohol (Ser) terminal residues make these short peptides hydrophilic and difcult to retain under conventional reversed-phase LC/MS conditions. On the PGC column these short peptides are well retained away from the solvent front. A ow-through fraction from a proteolytic digest was simulated by diluting a solution containing three hydrophilic peptides in Tris buffer (100 mM, pH 8.0) to concentrations in the range of 15 to 30 pmol/L. This fraction was injected and separated on an alkyl-silica phase and on PGC (Figure 3). On the alkyl-silica phase the two more polar peptides (RGES and DSDPR) elute at the solvent front, co-eluting with the chromatographic peak for Tris. The spectrum for these peptides is dominated by the Tris peaks at m/z 122, 243, 265 and 327, showing very weak signals for the [M+H]+ at m/z 448 and 589. On the other hand, the PGC stationary

phase effectively separates these peptides away from the Tris chromatographic peak, producing clean and intense spectra, that allow for good identication. The chromatographic retention and resolution of the hydrophilic peptides on the PGC column enables quantitative data to be obtained. On Figure 4 the linearity of the response is demonstrated for concentrations in the range of 5 pg/L to 1 ng/L of each peptide.

BIOCHEMICAL

Conclusion
Porous graphitic carbon columns show increased capacity factors over alkyl-silica columns for phosphorylated peptides. In contrast to alkyl-silica stationary phases, porous graphitic carbon retains small hydrophilic peptides away from the solvent front under typical reversed-phase LC/MS conditions; as a result the spectra are free from biological buffer and salts allowing for good peptide identication even at low levels. Quantitative analysis of small hydrophilic peptides on porous graphitic carbon columns can be achieved with excellent linearity.

Application Notes

References
1. E. T. Chin, D. I. Papac, Anal. Biochem., 273 (1999) 179-185. 2. P. Ross, LCGC Europe, May 2000

Figure 4: Linearity data for concentrations of 5 pg/L to 1 ng/L of each peptide injected on the Hypercarb column

Analysis of Glycopeptides Using Porous Graphite Chromatography and LTQ Orbitrap XL ETD Hybrid MS
Terry Zhang, Rosa Viner, Zhiqi Hao, Vlad Zabrouskov, Thermo Fisher Scientific, San Jose, CA, USA

BIOCHEMICAL
Application Notes
10

Introduction
Glycosylation of Asn, Ser or Thr is arguably the most common known post-translational modication (PTM), resulting in a multitude of highly heterogeneous protein isoforms.1, 2 While the physicochemical differences among the glycosylated protein molecules are often minute, their characterization remains a great analytical challenge. To date, although there are reports of glycoproteome analyses based on electrophoresis techniques, an LC-MS/MS based approach offers advantages in speed, sensitivity, and automation3. It remains the most powerful and versatile technique for elucidation of glycopeptide structure. However, in addition to the difculties in capturing minor glycopeptides efciently from samples containing large amounts of non-glycosylated peptides produced by proteolytic digestion of complex protein mixtures, commonly used collision-induced dissociation (CID) has limitations for determining the modication site due to the labile nature of the glycan attachment to the peptide ion.3 CID MS/MS predominantly generates fragment ions from cleavages of glycosidic bonds without breaking amide bonds.1, 2 In contrast to CID, electron transfer dissociation (ETD) preserves labile PTMs while cleaving peptide bonds, making the identication of modication sites possible.4, 5 Since glycosylated proteins and the resulting peptides are generally very heterogeneous, their mass spectra are highly complex, consequently highquality liquid chromatography, high mass resolution, and accurate mass measurements of ETD precursors and fragments are essential for glycopeptide analysis. The high-resolution, high-mass-accuracy measurements of the Thermo Scientic LTQ Orbitrap XL ETD hybrid mass spectrometer coupled with additional capabilities such as parallel acquisition, in-source CID, and alternating CID/ETD dissociation can enable a thorough characterization of glycopeptides in a single analysis. In this study, several glycoproteins: bovine 1-acid glycoprotein, fetuin and human 1-acid glycoprotein, were analyzed using nano LC-MS/MS. The performance of C8, C18, and porous graphite columns were systematically evaluated and optimized for glycopeptide separation prior to mass spectrometry analysis by an LTQ Orbitrap XL ETD.

Experimental Conditions
Sample Preparation Glycoproteins purchased from Sigma were denatured in 0.1% SDS 50 mM Tris HCI buffer (pH 8.5), reduced with 5 mM DTT for 1 hr at 60 C and alkylated with 25 mM iodoacetamide for 2 hr in the dark at room temperature. Then reduced and alkylated proteins were precipitated with acetone, digested, and analyzed by nano-LC-MS2. The experiments were conducted on LTQ XL ETD and LTQ Orbitrap XL ETD mass spectrometers using the following conditions:
LC Separation
HPLC System: Columns: Thermo Scientic Surveyor MS Pump with a ow splitter C8 column (75 m x 5 cm); C18 column (150 m x 10 cm) Thermo Scientic Hypercarb porous graphite column, 75 m x 5 cm (part number 35005-050065) A: 0.1% Formic acid; B: 0.1% Formic acid in acetonitrile For Hypercarb column 550% B in 30 minutes For RP columns 535% B in 30 minutes

Mobile Phases: Gradient:

MS Analysis
Mass Spectrometer: LTQ XL linear ion trap mass spectrometer with ETD and nano-ESI source Spray Voltage: 2 kV Capillary Temp: 160 C Capillary Voltage: 35 V Tube Lens: 125 V MSn Target: 1e4 Mass Range: 502000 m/z or 1004000 m/z Anion Reagent: Fluoranthene Anion Reagent Isolation: On Anion Target: 2e5 Max Anion Injection Time: 50 ms ETD Reaction Time: 75150 ms Mass Spectrometer: LTQ Orbitrap XL ETD hybrid MS Mass Range: 4002000 m/z, resolution 60,000100,000 @ m/z 400 FT MS AGC Target: 5e5 FT MS/MS AGC Target: 1e5, 3 amu isolation width MS/MS Resolution: 7,500 FWHM at m/z 400, 3 microscans Monoisotopic On Precursor Selection: Exclusion Mass Tolerance: 10 ppm Max Ion Time FT MS: 500 ms Max Ion Time FT MS/MS: 500 ms Full MS Range: 4002000 m/z MS/MS Mass Range: 1002000 m/z Survey Scan: Source CID at 65 V for m/z 204

Goal
To demonstrate the advantages of porous graphite chromatography and electron transfer dissociation for analysis of N-glycopeptides on an LTQ Orbitrap XL ETD hybrid mass spectrometer.

Data Processing Thermo Scientic Xtract software was used for deconvolution of multiply charged precursors and MS/MS spectra. The GlycoMod tool from the Swiss-Prot website was used to assign possible oligosaccharide structures and compositions.

BIOCHEMICAL

Results and Discussion


LTQ Orbitrap XL ETD Instrument The LTQ Orbitrap XL ETD mass spectrometer is a highresolution, accurate-mass hybrid mass spectrometer equipped with an ETD source.6 Figure 1 is the schematic of the LTQ Orbitrap XL ETD MS. The ETD anion reagent ions travel from the ETD source through the HCD collision cell and C-trap into the linear ion trap. There they are isolated and reacted with the precursor peptides ions in the ion trap. The resulting fragments can be measured in either the ion trap or Orbitrap mass analyzers. There are three additional fragmentation modes available on this instrument: higherenergy collisional dissociation (HCD), CID, and pulsed-Q dissociation (PQD). Since CID and ETD can provide complementary glycopeptide structural information, coupling high-resolution, high-mass-accuracy full MS with alternating CID and ETD MS/MS makes comprehensive glycopeptide analysis achievable (Figure 2).4, 5
Choosing a Stationary Phase for Glycopeptide Analysis

Application Notes

Figure 2: Glycopeptide analysis ow chart using an LTQ Orbitrap XL ETD hybrid mass spectrometer

We evaluated the performance of three stationary phases for LC analysis of glycosylated peptides. Figure 3 shows analysis of bovine 1-acid glycoprotein on C18 and porous graphite columns. One pmol of protein digest was injected into C18 column versus 500 fmol on porous graphite column. The proles are the extracted ion chromatograms of m/z 1706.3, 1138.5 and 867.3 belonging to 2+, 3+ and 4+ molecular ions of the bi-antennary glycopeptide 91CVYNCSFIK99. The intensities of the 2+ and 3+ glycopeptide precursor ions from analysis with the porous graphite column were similar to the intensities from analysis with the C18 column despite 50% lower load on-column. This was likely due to the

higher afnity of porous graphite for hydrophilic peptides. In addition, chromatography using a porous graphite stationary phase promoted formation of abundant highercharge-state metal-adducted precursor ions, which improved detection limits for all observed charge states of glycopeptides. Figure 4 shows a high-resolution deconvoluted spectrum of these glycopeptides. At least one potassium adduct was observed for each glycoform. As demonstrated in Figure 3, formation of a metal adduct is likely responsible for producing higher-charge-state ions: (M+K+3H)4+ precursor at m/z 867.3 was the dominant peak for glycopeptide 5 (Figure 4) while (M+3H)3+ precursor at m/z 1138.6 was the dominant signal for glycopeptide 2 (Figure 4). Formation of higher-charge precursors can be explained by partial neutralization of sialic acid negative charges by metal cations.7, 8 Bovine and human 1-acid glycoproteins contains ve N-glycosylation sites with hybrid-type glycan structures.9, 10 Four out of ve glycopeptides were detected and identied using a porous graphite column compared to two out of ve peptides on C18 and three out of ve peptides on C8 columns

Figure 1: Schematic diagram of an LTQ Orbitrap XL ETD hybrid mass spectrometer

11

BIOCHEMICAL
Application Notes
Figure 3: Comparison of different charge state extracted ion chromatograms of bovine 1-acid glycoprotein bi-antennary peptide 91CVYNCSFIK99 using C18 and porous graphite columns Figure 4: Deconvoluted full MS spectrum of bovine 1-acid glycoprotein bi-antennary peptide 91CVYNCSFIK99
12

Type of LC Column Peptides


103QNGTLSK109 53NPEYNK58 91CVYNCSFIK99 128TFMLAASWNGTK139 19QSPECANLMTVAPITNATMDLLSGK43

MS /MS Analysis of Glycopeptides

BIOCHEMICAL

Graphite

C18

C8

Table 1: Bovine 1-acid glycoprotein glycopeptides detected by nano LC-MS/MS

(Table 1) without any prior enrichment. Only the largest and most hydrophobic peptides were not detected using any of the phases. Similar results were obtained for bovine fetuin digest. All bovine fetuin glycopeptides could be detected using C182 or C8 chromatography. However, they were mostly present as lower-charge species, resulting in poorer MS/MS ETD spectra without enough information for their unambiguous identication. Previously, Alley and co-workers reported that fetuin glycopeptides could not be observed after separation on a graphite HPLC Glycan Chip.11 Our studies indicated that larger hydrophobic peptides elute much later in the gradient. Nevertheless, three out of four fetuin peptides were identied after separation on the porous graphite column. Only the largest O-glycosylated 246-306 peptide was not detected. Figure 5 shows an example of an ETD spectrum of 5+ precursor at m/z 1307 of bovine fetuin tri-antennary peptide 72RPTGEVYDIEIDTLETTCHVLDPTPLANCSVR103, which was successfully identied after separation on porous graphite column. After optimization, all nano LC-MS/MS analyses of glycopeptides were performed using a porous graphite column and 1 pmol of glycoprotein digest.

To identify glycopeptides, we utilized a strategy described by Peterman and Mulholland2, where in-source CID generated characteristic oxonium ions at m/z 204 and/or 366 and were further fragmented by a dedicated MS3 event. This allowed for a highly sensitive and selective detection of the eluting glycosylated peptide ion(s) at a given retention time. Highresolution, accurate-mass full-scan MS was measured in the Orbitrap mass analyzer, as shown in Figure 6, followed by data-dependent MS/MS alternating CID and ETD scans with fragments analyzed in either the ion trap or Orbitrap mass analyzers. The resulting CID spectra of the 3+ precursors (Figure 6c) were used for glycan structure elucidation and the ETD spectra of the 4+ precursors (Figure 6b) were used for glycopeptide identication as demonstrated in Figures 7 and 8 for the bovine 1-acid glycoprotein peptide 103QNGTLSK109. Figure 7, inset, shows the CID spectrum of the precursor [M+3H]3+ at m/z 990.395 measured in the Orbitrap mass analyzer. The oxonium fragment ion at m/z 366 conrmed this precursor as a glycopeptide. The spectrum was deconvoluted and glycan composition was assigned as shown in Figure 7. This glycoform contains one each of N-acetylneuraminic (Neu5Ac) and N-glycolyeuraminic (Neu5Gc) acids at the glycan branch end as determined by the presence of 657/673 fragment pair. The product ion at m/z 950.4805 corresponds to [M-GlcNAc+H]+. All measured masses were within 5 ppm of their theoretical values which helped unambiguously assign the glycan portion as a bi-sialated glycopeptide structure.

Application Notes

Figure 5: Ion trap MS/MS ETD spectrum of bovine fetuin tri-antennary peptide 72RPTGEVYDIEIDTLETTCHVLDPTPLANCSVR103 (5+, m/z 1307). Insert: Magnied region for scan range 100-2000 m/z.

13

BIOCHEMICAL
Application Notes

Figure 6: a) MS full scan; b, c) zoomed in charge states 4+, 3+ showing highly heterogeneous glycoforms; d) deconvoluted full MS spectrum of bovine 1-acid glycoprotein bi-antennary peptide 103QNGTLSK109.
14

Higher-charged 4+ metal-adducted precursor ions were further selected for ETD analysis. From the ETD MS/MS spectrum of the precursor [M+K+O+3H]4+ as shown in Figure 8, the glycosylation site was clearly identied at

Asn 104 based on an almost complete series of c/z ions. No signicant loss of carbohydrate was detected and, as expected, several of the observed glycan-containing fragments retained potassium ions. The high-resolution, high-mass-

BIOCHEMICAL
Application Notes

Figure 7: Deconvoluted Orbitrap CID MS/MS spectrum of bovine 1-acid bi-antennary glycopeptide 103QNGTLSK109. Insert shows original Orbitrap CID spectrum of 3+ parent at m/z 990.395 (Figure 6d, structure 1).

Figure 8: Deconvoluted Orbitrap ETD MS/MS spectrum of bovine 1-acid bi-antennary glycopeptide 103QNGTLSK109. Insert shows original Orbitrap ETD spectrum of [M+K+O+3H] 4+ at m/z 756.278 ion (Figure 6d, structure 5). 15

accuracy and low chemical noise of the Orbitrap mass analyzer signicantly beneted ETD analysis of glycopeptides with charge states 4+ and above, allowing straightforward deconvolution and interpretation of spectra (Figure 8).

References
1. Morelle, W., Canis, K., Chirat, F., Faid, V. and Michalski, J-C. (2006) The use of mass spectrometry for the proteomic analysis of glycosylation. Proteomics, 6, 3993-15. 2. Peterman, S.M. and Mulholland, J.J.(2006) A novel approach for identication and characterization of glycoproteins using a hybrid linear ion trap/FT-ICR mass spectrometer. J. Am. Soc. Mass Spectrom., 17(2), 168-79. 3. Kaji,H, Isobe, Toshiaki, (2008), Liquid Chromatography/Mass Spectrometry (LC-MS)-Based Glycoproteomics Technologies for Cancer Biomarker Discovery. Clinical Proteom, 4:14-24 4. Syka J.E., Coon J.J., Schroder M.J., Shabanowitz J, Hunt D.F. (2004) Peptide and protein sequence analysis by electron transfer dissociation mass spectrometry. Proc. Natl. Acad. Sci. USA; 101, 9528-33. 5. Alley Jr., W.R., Merchref, Y. and Novotny, M.V. (2009) Characterization of glycopeptides by combining collision-induced dissociation and electron-transfer dissociation mass spectrometry data. Rapid Commun. Mass Spectrom., 23, 161-170. 6. McAlister, G.C., Phanstiel, D., Good D.M., Berggren, W.T. and Coon, J.J. (2007) Implementation of electron-transfer dissociation on a hybrid linear ion trap-orbitrap mass spectrometer. Molecular & Cellular Proteomics, 2007, 6, 1942-1951. 7. Newton, K.A., Amunugama, R. and McLuckey, S.A. (2005) Gas-phase ion/ion reactions of multiply protonated polypeptides with metal containing anions. J. Phys. Chem. A., 109(16), 3608-16. 8. Medzihradszky, K.F., Guan, S., Maltby, D.A. and Burlingame, A.L. (2007) Sulfopeptide fragmentation in electron-capture and electron-transfer dissociation. J. Am. Soc. Mass. Spectrom., 18(9), 1617-24. 9. Snovida, S.I., Chen, V.C., Krokhin, O. and Perreault, H. (2006) Isolation and identication of sialylated glycopeptides from bovine 1-acid glycoprotein by off-line capillary electrophoresis MALDI-TOF mass spectrometry. Anal. Chem., 78, 6556-635 10. Treuheit, M.J., Costello, C.E. and Halsall, H.B (1992) Analysis of the ve glycosylation sites of human 1-acid glycoprotein. Biochem. J., 283, 105-12. 11. Alley Jr., W.R., Merchref, Y. and Novotny, M.V. (2007) Using graphitized carbon for glycopeptides separations prior to mass spectral detection. Proceedings of the 55th ASMS conference, Indianapolis.

BIOCHEMICAL
Application Notes
16

Conclusions
The Hypercarb porous graphite column demonstrated excellent separation for glycopeptide analysis, especially for short, hydrophilic peptides containing bi- or tri-antennary glycan chains. It allowed for their sensitive detection without any prior enrichment. Formation of metal adducts promoted evolution of highercharge species, aiding ETD fragmentation of glycopeptides. ETD preserved labile glycans, facilitating the identication of both the peptide of interest and the site of modication. Combining peptide structural information obtained by ETD and the glycan composition information obtained by CID enabled condent identication and characterization of glycopeptides within a single LC-MS analysis using an LTQ Orbitrap XL ETD mass spectrometer.

The Use of Porous Graphitic Carbon LC-MS for the Analysis of Underivatised Carbohydrates from Wheat Stems
Sarah Robinson, Thermo Fisher Scientific, Hemel Hempstead, UK

BIOCHEMICAL

Introduction
The stems of wheat (Triticum aestivum) contain pith, which varies in quantity between different cultivars and at different stages of the growth cycle. It is known that stored in the pith is a pool of endogenous water soluble carbohydrate metabolites, and that these carbohydrates are translocated into the plants grains.1 It has further been suggested that some wheat genotypes, which contain high concentrations of soluble carbohydrates in their stems, may be able to deposit more carbohydrates into the grains of the plant and signicantly increase the grain yield.2 The separation and detection of such a pool of oligosaccharides is not easy owing to the fact that these analytes are very polar, potentially highly branched, isomeric structures which contain no chromophore and which are poorly or completely unretained on reversed phase HPLC columns. Many workers tackle detection issues via derivatisation of the reducing terminal of their carbohydrate analytes, however, this presupposes that all carbohydrates are reducing.3 This additional step also demands extra time and sample handling, and involves inevitable losses, as well as changing the original structure of the analyte. Here we present a rapid, robust and efficient chromatographic separation method that utilises the highly selective stationary phase of porous graphitic carbon in an on-line coupling of liquid chromatography with electrospray mass spectrometry (PGC-LC-MS). Additionally, we describe the development, characterization and use of a nanoscale LC column coupled with a nanospray ionisation source which enabled us to miniaturize the system and obtain much improved sensitivity.

HPLC and nano-HPLC Methodology Thermo Scientific Surveyor LC system


Mobile phase A: HPLC grade water Mobile phase B: acetonitrile Mobile phase C: 2-propanol HPLC column: Thermo Scientific Hypercarb 100 x 4.6 mm, 5 m (part number 35005-104630) Flow rate: 600 L/min Nano-HPLC column: Thermo Scientific Hypercarb 100 x 0.1 mm, 5 m (part number 35005-100165) Flow rate: Flow split from 400 L/min to 0.15 L/min The eluent gradient is presented in Table 1.

Application Notes

Mass Spectrometry Thermo Scientic LCQ Deca XP ion trap mass spectrometer
Start
00.00 10.00 20.00 30.00

Flow
600 600 600 600

%A
94 92 85 82

%B
3 4 6 8

%C
3 4 9 10

%D
-

Table 1: Eluent gradient program for the HPLC system. Flow rate is reported as L/min.

Ion source, polarity: ESI, positive ion mode Spray voltage: 6 kV Sheath gas: 70 units Auxillary gas: 60 units Capillary temperature: 350 C m/z 200 2000 at 5500 amu/s

Goal
To demonstrate the separation of a complex pool of branched, isomeric and underivatised oligosaccharides utilizing porous graphitized carbon liquid chromatography coupled with mass spectrometric detection.

Nanoscale Experiments Nano-ion source & polarity: nano-ESI (front coated tip), positive ion mode Nano-spray Voltage: 3.1 kV Sheath and Aux gas: 0 units Capillary temperature: 200 C

Experimental
Sample Preparation Post-harvesting, frozen wheat stems were cut into ~ 2 cm pieces and boiled for 30 min in ethanol. The ethanol extract was dried and the stem pieces further boiled in water for 30 min. The ethanol and water extracts were mixed and washed with hexane until all colouration had been removed. Each extract was ltered through an inert 0.22 m syringe lter.

Results and Discussion


A chromatographic approach that utilized a ternary gradient afforded better control of the separation of the oligosaccharide mixture. In the early part of the chromatographic separation mono, di, tri and tetrasaccharides were removed in a relatively steep aqueous/acetonitrile gradient. After these analytes were eluted the gradient was made gentler and the 2-propanol eluent was gradually increased in order to remove the larger

17

BIOCHEMICAL
Application Notes
18
0 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34

Figure 2: PGC-LC-MS positive ion mass spectra of underivatized oligosaccharides extracted from the stems of a wheat cultivar. Retention times are marked in brackets, m/z value of sodiated or disodiated molecules follows.
Reprinted with permission from Anal. Chem. 2007, 79, 2437-2445. Copyright 2009 American Chemical Society.

Figure 1: PGC-LC-MS selected ion chromatograms of underivatized oligosaccharides extracted from the stems of a wheat cultivar.
Reprinted with permission from Anal. Chem. 2007, 79, 2437-2445. Copyright 2009 American Chemical Society.

and very strongly retained oligosaccharides from the column. This method development ultimately allowed the successful separation of the complex mixtures of oligosaccharides that had been extracted from wheat stems and which ranged in size from dp 2 to dp 20, in under 30 minutes. A series of selected ion chromatograms of the oligosaccharides extracted from the stems of one cultivar type are presented in Figure 1. Although ions consistent with Hex18, Hex19 and sometimes Hex20 species were observed in the full scan mass spectra, their ion intensities were not signicantly different from background, so that their ions were not distinguishable in the SIC. In the mass spectra we observed sodiated singly charged, doubly charged and dimeric (two analyte molecules with one charge-bearing cation) species of the oligosaccharide analytes in the extracts from wheat stems (Figure 2).

It was noted that the elution order of the oligosaccharides was not necessarily in the order of increasing oligosaccharide size; sometimes larger oligosaccharides co-eluted with a smaller oligosaccharide. For example in Figure 1, a Hex6 chromatographic peak (tR: 11 min) can be observed to co-elute with a Hex4 chromatographic peak (tR: 11 min), while further Hex6 peaks elute later in the chromatographic run. Since planar molecules are more strongly retained on PGC than less planar molecules (as they do not induce as strong dipole interactions with the PGC surface), we propose that the elution order of differently sized oligosaccharides is further evidence that these analytes are isomeric structures. There are many reported benets of using miniaturized systems, for example the use of nanoscale electrospray can overcome sensitivity problems found when analysing neutral oligosaccharides in conventional microscale electrospray.4 In light of these observations, and more importantly the recent availability of nanoscale PGC columns, we decided to compare the limit of detection (LOD) of the on-line PGC-LC-MS separation with the LOD obtainable using an on-line nanoscale system. Based on the column dimensions it could be predicted that the linear velocity of the mobile phase would be 1.9 times higher in the 4.6 mm column. After optimisation we could reproduce a similar chromatographic separation of oligosaccharides extracted from wheat stems in 50 min (compared with 30 min using the 4.6 mm column) (Figure 3). Our LOD comparative study utilized a dilution series of a standard fructan trisaccharide compound, 1-kestose. The solutions were analysed using the HPLC and the nanoHPLC-MS method and samples were analysed in the order of decreasing concentration. From the SIC of each sample run in triplicate, the average peak height was calculated. An approximation of the noise level in the system was made by measuring the noise level in the baseline close to the analyte chromatographic peak. A signal three times the noise was used as an approximation to the detection limit.

The lowest concentration from the dilution series injected that could be detected in the PGC-LC-MS and PGCnanoLC-MS analysis of 1-kestose was 5 M and 50 pM respectively. From the peak height and the noise measured in the single ion chromatogram of these analyses the concentration limit of detection for this analyte in this system was calculated as 14 nm and 14 pM respectively. Thus, it may be concluded that there is a 1000 fold improvement in the concentration LODs when using the 0.1 mm column compared with the 4.6 mm column. In addition to calculating the concentration LOD of each system, the amount LOD for the two systems was calculated by multiplying the concentration LODs by the injection volumes of 20 nL and 10 L used for the 0.1 mm and 4.6 mm columns respectively. It was concluded that there is a 500,000 fold increase in absolute amount sensitivity when using the 0.1 mm column compared with the 4.6 mm column.

BIOCHEMICAL
Application Notes

Conclusion
We have developed and optimised an on-line LC-MS method, using a PGC stationary phase, for the separation of a complex mixture of non-reducing oligosaccharides extracted from the stems of wheat in 30 minutes. Our system displays high chromatographic resolution and selectivity for a broad size range of isomeric oligosaccharides in a relatively short analysis time. In addition, the system is robust and automation has allowed very high throughput of samples. We have shown that signicantly smaller quantities of analyte are required for detection when using a nanobore PGC LC column and nanoESI source compared with those required for use with the microbore PGC LC column and microESI source. The amount LODs calculated from the nanoscale experiments also suggest that analytes are detected at the zeptomol level. Nanoscale analysers are also greener as they create signicantly less organic solvent waste. Our method has been applied to the analysis of wheat cultivar crosses to study further the link between stem carbohydrate and grain yield, but also this system could be applied to the analysis of other highly polar analytes that have previously been difcult to analyse by other separation methods.

References and Acknowledgements


The full text article that further describes the data presented in this application note may be found with the following reference:
Robinson, S., Bergstrom, E., Seymour, M., Thomas-Oates, J.E. Anal. Chem. 2007, 79, 2437-2445. 1. Wardlaw, I. F., Porter, H. K. Aust. J.Biol. Sci. 1967, 20, 309-18.

Figure 3: PGC-nanoLC-MS selected ion chromatograms of underivatized oligosaccharides extracted from the stems of a wheat cultivar.
Reprinted with permission from Anal. Chem. 2007, 79, 2437-2445. Copyright 2009 American Chemical Society.

2. Ford, M. A., Blackwell, R. D., Parker, M. L., Austin, R. B. Ann. Bot. 1979, 44, 731-38. 3. Mechref, Y.; Novotny, M. V. Chem. Rev. 2002, 102, 321-69. 4. Bahr, U., Pfenninger, A., Karas, M. Anal. Chem. 1997, 69, 4530-35.

19

Quantitation of Acrylamide in Food Samples on the TSQ Quantum Discovery by LC/APCI-MS/MS


Kevin J. McHale, Witold Winnik, Gary Paul, Thermo Fisher Scientific, Somerset, NJ, USA

FOOD SAFETY
Application Notes
20

Introduction
Acrylamide has been identied as a potential human carcinogen. This is important not only because acrylamide is a common industrial chemical, but acrylamide has been shown to be present at signicant levels in food samples,1 particularly cooked foods high in carbohydrates. This has led many government health agencies around the world to assess the risk of short- and long-term exposure to acrylamide in humans. This has led to the development of LC-MS/MS methodology for the quantitative analysis of acrylamide in foodstuffs.1-5 While a GC/MS protocol for the analysis of acrylamide exists, this method requires extensive sample cleanup and chemical derivatization.6 The advantage of LC-MS/MS is that chemical derivatization is not necessary prior to acrylamide analysis. To date, most LC-MS/MS methods for the assay of acrylamide have utilized an electrospray ionization (ESI) source for the production of acrylamide ions.1-4 Yet it is well-known that ESI-MS is problematic when highly aqueous solutions, such as those required for the reversedphase LC separation of acrylamide, are used.7 On the other hand, water does not pose a problem for the formation of a stable corona discharge used in APCI. One published report has demonstrated that APCI is a viable ion source for the production of acrylamide ions for LC-MS/MS detection.5 Furthermore, a study comparing ESI and APCI ion sources for the LC-MS/MS analysis of acrylamide showed that under the same chromatographic conditions, APCI-MS/MS yielded an improved detection limit.8 This report presents data acquired on the Thermo Scientic TSQ Quantum Discovery for the analysis of acrylamide. A simple LC-MS/MS method using the APCI source is used to measure acrylamide, via selective reaction monitoring (SRM), over a wide concentration range. A small selection of food samples was analyzed for acrylamide content following extraction with water. To preclude the need for a time-consuming solid-phase extraction procedure, a column-switching method was employed to selectively fractionate acrylamide from polar matrix interferences prior to LC-MS/MS detection.

Experimental
Chemicals and Reagents: Acrylamide (> 99.0%) was purchased from Fluka (Buchs SG, Switzerland). 2,3,3-d3acrylamide (98%) was obtained from Cambridge Isotope Laboratories (Andover, MA, USA). HPLC grade water was acquired from Burdick and Jackson (Muskegon, MI, USA). All chemicals were used as received without further purication. Sample Preparation: Standards were prepared by dilution of a stock solution of 1.0 mg/mL acrylamide or 1.0 mg/mL d3-acrylamide in water. The stock solutions were stored at 4 C for a period of no longer than two weeks. Two brands of potato chips and two brands of breakfast cereals were purchased and stored at room temperature until processed. After homogenizing approximately 50 grams of a food sample, two grams were weighed into a 35 mL polypropylene centrifuge tube. Aqueous extraction of acrylamide was initiated by the addition of 20 mL water containing 1000 ng d3-acrylamide as the internal standard (nal concentration = 50 ng/mL). The sample was vortexed for 30 s then subsequently centrifuged at 18,000 g for 15 minutes. Ten milliliters of the supernatant was decanted into a clean 35 mL centrifuge tube and centrifuged at 18,000 g for 10 minutes. Prior to analysis, 0.49 mL of the aqueous extract was ltered through a 0.45 m centrifuge lter (Millipore Corp., Bedford, MA, USA) at 9,000 g for 5 minutes. Sample Analysis: LC experiments were conducted with the Thermo Scientic Surveyor HPLC system. A Thermo Scientic Hypercarb 2.1 50 mm (part number 35005-052130) column was utilized as the analytical LC column. Separations of acrylamide were achieved under isocratic conditions using 100% water as the mobile phase at a ow rate of 0.4 mL/min. The injection volume for all LC experiments was 10 L. To eliminate the need for solid phase extraction (SPE) purication prior to the analysis of the food sample extracts, a column-switching LC method was employed. Briey, the sample extract was loaded onto a 2.1 50 mm Thermo Scientic Aquasil C18 column (part number 77505-052130), which was positioned before a 6-port switching valve. The eluent from the C18 column was diverted to waste except for the period when acrylamide eluted from the C18 column, whereby the valve was switched to the Hypercarb column for MS/MS detection. This column-switching method required a second Thermo Scientic Surveyor MS pump, which also delivered 100% water at 0.4 mL/min. Both Surveyor MS pumps and the 6-port switching valve were controlled using Thermo Scientic Xcalibur version 1.3 software.

Goals
1. Development A sensitive and rugged LC/APCI-MS/MS assay for the analysis of acrylamide 2. Application An on-line column-switching technique to aqueous food extracts as an alternative to solid-phase extraction (SPE) cleanup 3. Measurement Acrylamide content in selected food samples

The experimental conditions for the TSQ Quantum Discovery were as follows: Source: APCI Ion polarity: Positive Vaporizer Temperature: 375 C Discharge Current: 5 A Ion Transfer Capillary Temperature: 250 C Source CID Offset: 6 V Scan Mode: Selective Reaction Monitoring Q2 Pressure: 1.0 mTorr argon SRM Transitions: m/z 72 55 for acrylamide; m/z 75 58 for d3-acrylamide Collision Energy: 13 eV Scan Width: 1.0 u Scan Time: 0.3 s (each SRM transition) Q1, Q3 Resolution: Unit (0.7 u FWHM)

FOOD SAFETY
Application Notes

Results and Discussion


Prior to the acquisition of acrylamide standards, it was important to determine if there was any detectable native acrylamide contribution originating from the deuterated internal standard. As shown in Figure 1, there is no acrylamide signal observed for the m/z 72 55 SRM transition at the same retention time as the 50 ng/mL d3-acrylamide standard. The limit of quantitation (LOQ) for acrylamide on the TSQ Quantum Discovery was 0.25 ng/mL acrylamide or 2.5 pg on column (Figure 2). This compares favorably to LOQs previously reported by other research groups, including an 8-fold improvement over the mass LOQ by LC/ESI-MS/MS (20 pg)1 and a 40-fold improvement over the concentration LOQ on the TSQ 7000 (10 ng/mL),5 which used an LC/APCI-MS/MS method. The calibration curve for acrylamide from 0.25 ng/mL to 2500 ng/mL is displayed in Figure 3. This calibration curve was generated using the column-switching LC method just prior to the acquisition of the food extracts data. A linear regression t to these data using 1/x weighting yielded the following equation: y = 5.5997 104 + 0.0206125x. The correlation coefcient for this curve was r2 = 0.9999, indicating excellent linearity across the four orders of magnitude dynamic range. Table 1 summarizes the statistical results for the acrylamide calibration curve. At the LOQ, the accuracy, as a percent relative error, is 1.1% and the precision, as a percent coefcient of variance (%CV), is 12.1% for ve replicate injections. Above the LOQ, the relative error varied from -2.9 to +2.4% and the %CV ranged from 0.5 to 6.7%. Results obtained from the aqueous extract of Potato Chip 2 are presented in Figure 4. By utilizing a C18 column positioned before a switching valve to selectively elute acrylamide onto the Hypercarb column, background interferences are reduced. Unlike most of the other acrylamide reports where SPE cleanup was used following extraction of the sample with water,1-4 the column-switching LC method employed here provides an on-line means of acrylamide fractionation. This has the advantage of minimizing sample losses during SPE and greatly reduces sample preparation time.

Figure 1: SRM chromatograms for 50 ng/mL d3-acrylamide

Figure 2: SRM chromatograms for 0.25 ng/mL acrylamide (LOQ) with 50 ng/mL d3-acrylamide

Figure 3: Calibration curve for acrylamide using column-switching LC method with APCI-MS/MS detection

21

Nominal (ng/mL) 0.250 0.500 1.00 5.00 10.0 100 500 1000 2500

Mean Conc. (ng/mL) 0.253 0.485 1.00(4) 4.86 10.2 101 512 1006 2481

% Rel. Error 1.1 -2.9 0.4 -2.7 2.1 0.7 2.4 0.6 -0.8

% CV 12.1 6.7 4.6 0.9 0.7 0.5 0.8 0.6 0.6

Number of Replicates 5 5 5 5 5 5 3 3 3

Table 1: Statistical data for the calibration curve of acrylamide

To monitor the consistency and reproducibility of the column-switching LC-MS/MS method, a 1 ng/mL acrylamide standard was analyzed immediately following each food sample. An example of this quality control standard analyzed after the Potato Chip 2 sample is shown in Figure 5. Although the baseline for the m/z 72 55 SRM transition is somewhat elevated near the retention time for acrylamide, the calculated concentration for this standard is 0.99 ng/mL, equating to a relative error of -1.0%. Table 2 reports the results for four different food samples that were assayed for acrylamide using the column-switching LC method and MS/MS detection. The acrylamide concentrations in each food sample were calculated by multiplying the measured solution concentration from duplicate injections by the extraction volume and dividing by the food sample mass that was extracted. The determined acrylamide concentrations correlated well to those reported elsewhere for these classes of food.1-5
Cereal 1 Injection 1 Injection 2 Mean Extraction Vol. Mass Sample 17.17 ng/mL 17.00 ng/mL 17.09 ng/mL 20.0 mL 2.003 g Cereal 2 55.93 ng/mL 56.18 ng/mL 56.06 ng/mL 20.0 mL 2.007 g 559 ng/g Potato Chip 1 57.11 ng/mL 56.52 ng/mL 56.82 ng/mL 20.0 mL 2.021 562 ng/g Potato Chip 2 29.18 ng/mL 29.14 ng/mL 29.16 ng/mL 20.0 mL 1.995 292 ng/g

FOOD SAFETY
Application Notes
22

Acrylamide Conc. 171 ng/g

Table 2: Results of acrylamide assay from food samples

Conclusions
Figure 4: SRM chromatograms of the Potato Chip 2 sample aqueous extract

An LC-MS/MS method has been developed for the measurement of acrylamide on the TSQ Quantum Discovery. Using APCI for the analysis of acrylamide from 100% water, an LOQ of 0.25 ng/mL acrylamide or 2.5 pg on column was achieved. Incorporation of a column-switching LC method prior to MS/MS detection of acrylamide eliminated the need to purify food sample extracts by SPE. The method was successfully demonstrated for the analysis of four brands of food samples using TSQ Quantum Discovery in conjunction with a column-switching LC method.

References
1. Tareke, E.; Rydberg, P.; Karlsson, P.; Eriksson, S.; Tornqvist, M. J. Agric. Food Chem., 2002, 50, 4998-5006. 2. Rosen, J.; Hellenas, K. Analyst, 2002, 127, 880-882. 3. Musser, S. M. http://www.cfsan.fda.gov/~dms/acrylami.html 4. Becalski, A.; Lau, B.P.; Lewis, D., Seaman, S.W. J. Agric. Food Chem., 2003, 51, 802-808. 5. Brandl, F.; Demiani, S.; Ewender, J.; Franz, R.; Gmeiner, M.; Gruber, L.; Gruner, A.; Schlummer, M.; Smolic, S.; Stormer, A.; Wolz, G. Electron. J. Environ. Agric. Food Chem., 2002, 1(3), 1-8.

Figure 5: 1 ng/mL acrylamide standard analyzed directly after duplicate injections of the aqueous extract of the Potato Chip 2 sample

6. Castle, L.; Campos, M.J.; Gilbert, J. J. Sci. Food Agric. 1993, 54, 549-555. 7. Ikonomou, M.G.; Blades, A.T.; Kebarle, P. J. Am. Soc. Mass Spectrom., 1991, 2, 497-505. 8. McHale, K.J.; Winnik, W.; Paul, G. Proceedings of the 51st ASMS Conference on Mass Spectrometry and Allied Topics, Montreal, 2003.

ENVIRONMENTAL

Fast LC Separation of Triazine Herbicides at Elevated Temperature


Dave Thomas, Thermo Fisher Scientific, San Jose, CA USA

Introduction
Temperature is a key variable in high performance liquid chromatography (HPLC), influencing solute diffusion rates, mobile phase viscosity, and solubility. For example, as column temperature increases, analyte diffusion increases. Increased analyte diffusion generally leads to an increase in the optimum linear velocity of the separation, so that equivalent chromatographic efficiency and resolution can be achieved at a higher flow rate. Furthermore, elevating the temperature reduces the operating backpressure. The net result is that separations can be performed faster without exceeding the pressure limitations of the instrument. This application uses a porous graphitic carbon stationary phase thermostatted in a high temperature column oven to separate triazine herbicides 5 to 10 times faster than is typical with conventional HPLC. The triazines and degradation products are separated on the Thermo Scientific Accela High Speed Liquid Chromatograph in 2 minutes on a Hypercarb 3 m, 1 x 100 mm column operated at 160 C. This application note also documents the performance of the high temperature liquid chromatographic method, including precision of retention time and peak area, resolution, and spike recovery from several environmental water matrices.

Instrumentation
Thermo Scientific Accela HPLC system with PDA Detector Thermo Scientific ChromQuest 5.0 Chromatography Data System (CDS) Polaratherm Series 9000 Total Temperature Controller (Selerity Technologies)

Chemicals
Water, LC/MS-grade Acetonitrile, LC/MS-grade Methanol, LC/MS-grade Atrazine Ametryn Cyanazine Deisopropylatrazine, 1000 mg/L Desethylatrazine, 1000 mg/L Propanil, 1000 mg/L Propazine Prometryn Simazine Simetryn Fisher Scientific W6 Fisher Scientific A998 Fisher Scientific A456 Supelco 49085 ULTRA PST-024 ULTRA PST-1360 SPEX CertiPrep S-1135 SPEX CertiPrep S-1145 SPEX CertiPrep S-3155 ULTRA PST-850 ULTRA PST-840 ULTRA PST-1130 Chem Service PS-381

Application Notes

Consumables
Autosampler vials, 1.8 mL glass, yellow septa Backpressure assembly Ferrules, high temperature Mixer, 50 L in-line static Mobile Phase Preheater, 0.005" x 70 cm Syringe filters, 0.45 m Nylon Sample Loop, 5 L Thermo Scientific A4954-010 Upchurch P-788 Selerity Technologies BM0054 Thermo Scientific 109-99-032 Selerity Technologies AD104 Thermo Scientific A5307-010 Thermo Scientific 109-99-007

Goal
Increase throughput of the HPLC method for triazine herbicides by employing ultra high-speed liquid chromatography at elevated temperature on a heat stable Hypercarb column.

Experimental
Chromatographic Conditions
Column: Mobile phase: Gradient: Thermo Scientific Hypercarb 3.0 m, 1 x 100 mm (part number 35003-101046) A: water B: acetonitrile Time %A %B 0.00 75 25 1.00 70 30 2.20 10 90 2.30 75 25 4.00 75 25 500 L/min PDA, 238 nm, 10-mm flow cell, 11 nm bw, 20 Hz, 0s rise time 160 C (housed in Selerity temperature controller) 5 L sample loop, 2 L partial loop injection Syringe Speed: 4 L/sec Flush Speed: 100 L/sec Flush Volume: 400 L Wash Volume: 200 L Flush/Wash source: Bottle with 90:10 methanol:water

Mobile Phase Proportioned mobile phase: Filled Solvent Reservoir Bottle A of the Accela pump with fresh HPLC-grade water and purged the solvent line with at least 30 mL of the water. Connected a fresh bottle of HPLC-grade acetonitrile to Reservoir B and purged as above. Calibration Standards Individual Stock Solutions, 1000 mg/L: Accurately weighed 10 mg (0.010 g) of each neat compound into a 10-mL volumetric flask, added 5 mL acetonitrile, and sonicated to dissolve. Brought to volume with acetonitrile and mix. Used desethyl atrazine, deisopropyl atrazine and propanil, purchased as solutions of 1000 mg/L in methanol, as received.
Combined Intermediate Standard 100 mg/L: Used a calibrated pipette to deliver 1000 L of each individual stock solution to a 10-mL volumetric flask. Brought to volume with acetonitrile and mix. Calibration Standards: Used a calibrated pipette to dilute the intermediate standard with mobile phase in volumetric glassware to 30, 10, 3, 1, 0.3, 0.1, and 0.03 mg/L.
23

Flow rate: Detector: Column temp.: Injection:

ENVIRONMENTAL
Application Notes
24

CAS#

4-deethyl atrazine 6190-65-4

6-deisopropyl atrazine 1007-28-9

Cyanazine 21725-46-2

Propazine 139-40-2

Prometryn 7287-19-6

Formula MW (g/mol) pKa Log Po-w Water solubility, mg/L MeOH solubility, g/L CAS#

C6H10ClN5 187.633 1.51 3200 Atrazine 1912-24-9

C5H8ClN5 173.606 1.15 670 Ametryn 834-12-8

C9H13ClN6 240.697 0.87 2.22 170 45 (ethanol) Simazine 122-34-9

C9H16ClN5 229.713 1.7 2.93 8.6 6.2 (toluene) Simetryn 1014-70-6

C10H19N5S 241.361 4.05 3.51 33 160 Propanil 709-98-8

Formula MW (g/mol) pKa Log Po-w Water solubility, mg/L MeOH solubility, g/L
a

C8H14ClN5 215.687 1.7 2.61 34.7 18

C9H17N5S 227.334 4.1 2.98 209 510

C7H12ClN5 201.66 1.62 2.18 6.2 400

C8H15N5S 213.307 4 2.8 450

C9H9Cl2NO 218.082 2.29 3.07 152 540

http://toxnet.nlm.nih.gov

Table 1: Useful properties of some triazine herbicides and degradation productsa

Analyte deisopropylatrazine desethylatrazine cyanazine propazine atrazine simazine prometryn ametryn simetryn propanil
a b

ka 2.4 2.8 5.6 6.1 7.7 9.2 10.4 11.9 13.1 15.1

Ra 1.1 1.2 7.2 1.5 3.9 3.9 3.0 4.3 3.7 6.3

Linear range, mg/L 0.03 10 0.03 30 0.03 30 0.03 30 0.03 30 0.03 30 0.03 30 0.03 100 0.03 100 0.03 30

r2 0.9999 0.9999 0.9995 0.9996 0.9995 0.9994 0.9999 0.9995 0.9999 1.0000

MDLb g/L 6 16 40 23 14 30 32 8 16 25

Precision, Retention Time % RSDc 0.25 0.22 0.16 0.13 0.10 0.09 0.08 0.04 0.04 0.02

Precision, Peak Area % RSDc 0.39 0.89 0.47 0.82 0.87 0.92 0.97 0.64 0.57 0.42

Capacity factor (k) and Resolution (R) calculated according to Reference 1. Detection limit MDL = ts,99 where t,99 = 3.14 for n = 7 replicates of the standard. c for n = 30 replicates. Table 2: Performance of high temperature method for triazines performed on Hypercarb 3 m, 1 x 100 mm column at 160 C.

ENVIRONMENTAL

Samples Samples of surface water (Salinas River, Monterey County, CA), ground water (domestic well, Santa Cruz county, CA), and drinking water (San Jose, CA tap water) were collected in accordance with established procedures, stored at 4-8 C, and were filtered through a 0.45 m nylon syringe filter into a glass autosampler vial before analysis. System Preparation To ensure good performance of this application, prepare the system as directed in Appendix A.

Results
Separation of seven triazine herbicides, two triazine degradation products often found in environmental samples, and propanil is shown in Figure 1. To optimize this separation, we adjusted the mobile phase composition to elute the first analyte with a capacity factor k > 2, thereby improving

resolution of the target analytes from sample matrix junk. Analytes spanning a wide range of polarity are well resolved by the combination of high temperature, solvent gradient, and the selectivity of the Hypercarb stationary phase. Note that because of the reduced viscosity of the mobile phase at 160 C, this separation occurs at a linear flow rate of 15 mm/s equivalent to a flow rate of over 10 mL/min on a 4.6 mm i.d. column. The system backpressure under these conditions is less than 4000 psi (272 bar). Method performance is characterized by peak resolution, linear calibration range, limits of detection, and precision of retention time and peak area, as summarized in Table 2. MDLs for each analyte were determined by performing seven replicate injections of LC/MS-grade water fortified at a concentration of three to five times the estimated instrument detection limits, calculating the standard deviation of the measured concentration, and using the equation given in the figure caption. Note the good precision of retention time and peak area, as this reflects the temperature stability maintained by the high temperature oven. We analyzed several environmental water samples to demonstrate the efficacy of this method with real matrices. Samples of surface, ground and municipal drinking water were analyzed before and after fortification with a known amount of each target analyte. Spike recovery was calculated as the amount of each analyte found in the spiked sample divided by the amount expected (i.e., the amount determined in the blank plus the amount added in the spike). Even with the dirtiest matrix, Salinas River water, the target analytes are well separated from the early eluting matrix peaks (Figure 2) and recovery of the spiked analytes exceeds 80% (Table 3).

Application Notes

Column: Temperature: Flow Rate: Detector: Injection: Solvents: Gradient:

Samples: Peaks:

Hypercarb 3 m, 100 x 1 mm 160 C 500 L/min Accela PDA at 215 nm, 20 Hz, 0s rise time 2 L partial loop from 5 L loop A: Water B: Acetonitrile Time (min) A% B% 0.00 75 25 1.00 70 30 2.20 15 85 2.10 75 25 4.00 75 25 triazines and propanil in 20% acetonitrile 1. Melamine 7. Prometryn 2. Unknown 8. Atrazine 3. Deisopropylatrazine 9. Ametryn 4. Desethylatrazine 10. Simazine 5. Cyanazine 11. Simetryn 6. Propazine 12. Propanil

Figure 1: Separation of triazine herbicides, degradation products, and propanil on the Accela High Speed LC by reversed-phase chromatography with UV absorbance detection at 215 nm. Peaks: see Figure. Sample: overlay of 30 injections of triazines in HPLC-grade water with 20% acetonitrile.

Figure 2: Chromatograms of four environmental water samples spiked with 200 g/L each of triazines and propanil. Chromatograms obtained on the Accela High Speed LC by reversed phase chromatography with UV absorbance detection at 215 nm. Peaks: see Figure 1. Samples: top trace A, surface water (Salinas River); trace B, ground water (Simoes well); trace C, drinking water (San Jose tap); bottom trace D, HPLC-grade water. Conditions: see text for details.

25

ENVIRONMENTAL
Application Notes
26

Analyte deisopropylatrazine desethylatrazine cyanazine propazine atrazine simazine prometryn ametryn simetryn propanil

HPLC water 94.0 93.0 92.5 97.0 94.5 98.0 104 97.0 99.0 101

Drinking water 104 110 102 106 104 102 117 104 98.5 103

Ground water 104 98.5 94.0 104 101 100 116 101 104 104

Surface water 93.0 85.0 97.0 101 98.5 103 110 98.5 100 80.0

Table 3. Percent recovery of analytes spiked into selected environmental water matrices. n = 3 replicates.

Conclusion
A separation performed at 160 C on the Accela high speed chromatography system equipped with a heat stable Hypercarb column and high temperature column oven resolves 11 triazine herbicides in about two minutes with retention time and peak area precision better than 1% RSD for thirty replicates.

References
1. United States Pharmacopeia 30-National Formulary 25, United States Pharmacopeia, Rockville, Maryland 20852-1790, USA.

AS: Open the Instrument Configuration and verify that the Accela AS Configuration entry for Dead volume is correct (the calibrated volume in L written on the transfer tubing between the injection port and injection valve). Verify that the entry for Loop size is correct for the currently installed sample loop. Fill the Flush reservoir with 90:10 (v/v) methanol:water and flush the syringe with solvent to purge any air bubbles from the syringe and tubing. Use the Wash/Flush conditions specified under Conditions to ensure low carryover between injections. Consult the Accela Getting Connected manual (Document 60057-97001 Revision A) for details. Polaratherm Column Oven: Install the Total Temperature Controller according to the Polaratherm Series 9000 Installation and Operation Manual. Install the Hypercarb, 3 m 1 x 100 mm column, by using a 70-cm length of precut and polished 0.005 i.d. high-pressure tubing with mobile phase preheater. It is important to use a heat stable Hypercarb column as this does not contain any PEEK components that will degrade at the temperatures used in this method. Use the high temperature graphite/Vespel ferrules and fittings described in the Series 9000 manual. Ensure that the tubing is fully pushed into the column inlet when you tighten the high-pressure fitting. Detector: Use a 10 mm light-pipe flow cell. Install a 250 psi backpressure regulator after the flow cell outlet to suppress bubble formation in the flow cell. Verify that the deuterium lamp has been used for less than 2000 hours. Use Direct Control or a downloaded method to equilibrate the Accela system under the conditions shown above. Create a method based on these operating conditions and then create a sequence to make several injections of HPLC grade water. The system is ready to run standards and samples when the peak-to-peak baseline oscillation is between 50 200 AU/min (average of 10 1-min segments) and no significant peaks elute in the retention time window of the analytes.

Suppliers
Chem Service, West Chester, PA, USA (http://www.chemservice.com) Selerity Technologies, Inc., Salt Lake City, UT, USA Sigma-Aldrich, St. Lois, MO, USA (http://www.sigmaaldrich.com) Supelco, Bellefonte, PA, USA (http://www.sigmaaldrich.com) Thermo Fisher Scientific, Waltham, MA, USA (http://www.thermofisher.com) ULTRA Scientific, No. Kingstown, RI, USA (http://www.ultrasci.com)

Appendix A
System Preparation
Pump: Always plumb the Accela system with precut and polished 0.005 i.d. high-pressure tubing and high pressure fittings as shown in Figure 15 of the Accela Pump Hardware Manual (Document 60157-97000 Revision B). For all tubing connections that you make, ensure that the tubing end is square-cut and burr-free. Firmly push the tubing into the injection valve port as you tighten the high-pressure fitting in order to maximize peak efficiency. Prime the pulse dampener and purge the solvent lines as instructed in Chapter 4 of the Accela Pump manual. Verify that the pump is performing well by monitoring the pressure pulsation and by testing the pump proportioning accuracy as described in Chapter 5 of the pump manual. If your Accela pump does not include an inline 35 L dynamic mixer, then install a 50 L static mixer between the inline high pressure filter and the Accela AS mobile phase preheater.

ENVIRONMENTAL

Analysis of Polar Metabolites of Atrazine in Ground Waters Using Hypercarb as SPE Sorbent Coupled On-Line with Hypercarb LC Column
Valrie Pichon, Christophe Flinois, Marie-Claire Hennion, Laboratoire Environnement et Chimie Analytique (UMR PECSA 7195) Ecole Suprieure de Physique et Chimie Industrielles (ESPCI ParisTech), Paris, France

Introduction
Chlorotriazine herbicides have been extensively used as preand post-emergence weed control agents on crops, mainly corn and soybean. The potential for contamination of water and sediments by the widely used herbicide atrazine is high. This is due to its relatively high solubility, its weak adsorptivity (as measured by the partition coefficient between soil organic carbon and water) and its relatively long hydrolysis half-life in some soils. These characteristics easily explain that it is the most frequently reported pesticide in agricultural areas. Its use has then been limited in several countries but without solving the problem of its degradation products found in ground and surface water. Its degradation after spreading depends on several factors such as hydrolysis, photolysis and microbial activity. The main degradation products in ground and surface waters and in soils are dealkylated metabolites and therefore deethyl- (DEA) and deisopropyl-atrazine (DIA). The European Union has set pesticide standards for drinking waters at a maximum permissible concentration for a particular pesticide at 0.1 g/L and the sum of all pesticides at 0.5 g/L. The new regulation establishes not only a maximum concentration of pesticides in drinking water but also includes their degradation products after drinking water treatment. In order to estimate the efficiency of a degradation process that will generate very polar degradation products, and to control the contamination level of ground water that can be used for the production of drinking water, it is necessary to developed analytical methods suitable for the trace analysis of these very polar organic contaminants. The objective of this study was to provide a sensitive method for the detection and quantification of several polar dealkylated degradation products of atrazine, i.e. DEA, DIA, didealkylatrazine (DDA), ammeline (2,4-diamino6-hydroxy-1,3,5-triazine) and ammelide (2-amino-4,6dihydroxy-triazine) at the trace level in water samples. Their structures are presented in Figure 1. The analysis of such organic contaminants in complex matrices at low levels of concentration requires a procedure of pretreatment in order to extract and concentrate them before their LC/UV or LC/MS analyses. Solid-phase extraction has proven to be a very efficient technique and its application to the analysis of triazines and their metabolites has been already reviewed.1 On-line coupling of solid-phase extraction with liquid chromatography was shown to be a technique of choice for the trace-level determination of organic compounds because of its speed, easy automation and reliability.2 With this approach, the monitoring of organic pollutants in drinking

Application Notes

Figure 1: Structure of the studied compounds

water at 0.1 g/L is commonly achieved using a sample volume of 50-150 mL depending on the sensitivity of the detection device. Several studies reported the use of C18 silica or apolar copolymers of styrene divinylbenzene (PS-DVB) as extraction sorbent packed in a precolumn for the on-line extraction of triazines from water samples.3 However, the high polarity of the metabolites cited below causes the breakthrough of most of them during the extraction process on these sorbents. The interest in using carbonaceous sorbents for very water-soluble compounds has been reviewed.4 The most widely used carbon-based SPE sorbent is graphitized carbon black (GCB). In spite of its low surface area (120 m2/g), its potential for trapping polar compounds with a higher efficiency than C18 silica sorbent has been largely demonstrated.4-7 Porous Graphitic Carbon (PGC, Hypercarb, Thermo Scientific) characterized by a crystalline structure made of large graphitic sheets held together by weak Van der Waals forces has also been largely used for the extraction of polar pollutants from water.4 Most applications using carbonaceous sorbents have been carried out following an off-line procedure.8-10 Their on-line coupling with classical reversed-phase C18 silica analytical columns is difficult because polar analytes may be too strongly retained on PGC and the high water content of the mobile phase required for the separation of these polar compounds on C18 silica is unable to elute the analytes from PGC.4 A solution is to couple this sorbent with a Hypercarb column because in this case, a higher amount of organic solvent can be used in the mobile phase for the separation of polar compounds. With this system, the direct determination of the polar compounds at 0.1 g/L can be achieved. The PGC/PGC coupling was automated and used for the monitoring of polar metabolites DEA and DIA in ground
27

ENVIRONMENTAL
Application Notes
28

waters with a 100 mL sample.11, 12 The application of PGC for the most polar metabolites, i.e. ammeline and ammelide has only been applied following an off-line procedure.10

Goal
The development of a miniaturized on-line coupling of a Hypercarb precolumn for the extraction of atrazine and their polar dealkylated metabolites from water, with a LC separation using a microbore Hypercarb column for increasing the sensitivity in UV detection and for facilitating the coupling with electrospray mass spectrometry detection.

Experimental
Chemicals and Reagents Atrazine, simazine and their metabolites were obtained from C.I.L. (Saint-Foy-la-Grande, France). Stock standard solutions of 100 mg/L were prepared by weighing the solutes and dissolving them in methanol or in a water-methanol (50:50) mixture for some degradation products of triazines. The stock solutions were stored at 4 C. A standard solution of 5 mg/L was obtained by dilution in methanol from the stock solution. HPLC-grade acetonitrile and methanol were purchased from Mallinckrodt Baker (Deventer, Netherlands). High purity water was obtained from a Milli-Q purification system (Millipore, Saint-Quentin en Yvelines, France). Online SPE/LC/UV Set-up LC analysis were performed with a Varian LC System Workstation including a Varian Star 9012 solvent-delivery system and a Model 9065 Polychrom diode-array detector (190-366 nm). A Thermo Scientic Hypercarb analytical column (100 x 3 mm, 5 m, part number 35005-103030) was connected to a Valco valve (VICI, Houston, TX, USA). The elution gradient was based on acetonitrile (ACN) and water. The gradient was 10% to 35% ACN from 0 to 35 min, 100% ACN at 45 min. The ow rate was set at 0.45 mL/min. The UV detection was set to 220 nm. Trace enrichment was performed on small-size precolumn using an automated programmable sample preparation unit (Prospekt, Spark Holland, Emmen, Netherlands) allowing direct elution of compounds trapped on the Hypercarb precolumn (10 x 2 mm, 5 m) to the LC column. Online SPE/LC/UV/MS Set-up The instrumentation for microLC consisted of 1100 Series pump (Agilent Technologies, Waldbronn, Germany) connected to an Acurate 1/10 microow splitter (LC Packings, Amsterdam, The Netherlands). UV detection was performed with a SPDM10A photodiode array detector (Shimadzu) equipped with a microcell with an internal volume of 35 nL (LC Packings). Analytical micro LC separations were performed on a Hypercarb column (100 x 1 mm, 5 m, part number 35005101030). The elution gradient is based on acetonitrile (ACN) and water containing 0.03% triuoroacetic acid (TFA). The gradient was 0% to 5% ACN from 0 to 5 min, 30% ACN at 15 min and 80% ACN at 30 min. The ow rate was set at 50 L/min after the microow splitter and was introduced without any split in the ESI/MS. The UV detection was set to 220 nm.

Micro-LC/ESI/MS-MS analyses were carried out on a VG Quattro (Fisons Intruments, VG Biotech, Altrincham, United Kingdom) triple quadrupole equipped with an electrospray ion source. Data analysis was controlled by MassLynx 3.3 V. Full scan spectra were acquired in the positive ion peak centroid mode over the mass range of m/z 50-1200 in 4.5 s. The energy of collision applied was set to 50 eV. The transitions were monitored with 0.3 or 0.5 sec, EI: 3.9 V. Trace enrichment was performed on Hypercarb precolumn (10 x 2.1 mm) introduced in a precolumn holder. The precolumn was connected to a preconcentration pump (Shimadzu LC 5A ) and to the analytical column via a six-port switching valve (Rheodyne, Berkeley, CA, USA). The conditioning solvents and the samples were selected using a low pressure valve (Rheodyne) that was connected to the preconcentration pump.

SPE Procedure Using the Prospekt system connected to LC/UV, the Hypercarb precolumns were conditioned by passing 5 mL of acetonitrile, 5 mL of methanol and 5 mL of pure water. Water samples were percolated and the sorbents were washed with 1 mL of pure water. The flow rate was set at 2 mL/min. After switching the valve, the desorption was carried by the mobile phase by backflush elution at 0.45 mL/min for 10 min. Using the Hypercarb precolumn connected to LC/UV/MS, the carbonaceous SPE sorbent was conditioned with 5 mL of acetonitrile and 5 mL of pure water. Samples of water were then percolated through the precolumn. The ow rate was set at 0.5 mL/min. After switching the valve, the desorption was carried by the mobile phase by backush elution at 50 L/min for 10 min.

Results and Discussion


The studied compounds belong to a broad range of polarity with log Kow values between -1.2 for ammeline and 2.7 for atrazine. It has been already mentioned that the separation of the most polar metabolites, i.e. ammeline (AME) and ammelide (ADE), necessitates the use of a mobile phase containing only acidied pure water in order to decrease as much as possible the elution strength for their retention on C18 silica.9 A Hypercarb column has then, been selected for increasing the retention of these polar compounds. The use of a Hypercarb analytical column for the separation of some metabolites of atrazine has already been described but not for ADE and AME. Moreover, this mobile phase has to be compatible with the ESI-MS detection system. The separation on Hypercarb coupled to UV detection was achieved with a water/acetonitrile gradient. For LC/MS analysis, 0.03% triuoroacetic acid was added to the water/acetonitrile mobile phase. Initially, on-line coupling based on a conventional 3 mm i.d. column connected to UV detection was developed. To reach a high enrichment factor, 50 mL water sample was percolated through the Hypercarb precolumn. With these conditions, recoveries higher than 61% were obtained for DAA, DEA, DIA, atrazine and simazine. Because of ammelide and ammeline high polarity breakthrough occurs during the percolation step and recoveries lower than 10% were obtained. Recoveries are reported in Table 1.

ENVIRONMENTAL

Compounds
Ammelide Ammeline DAA DEA DIA Atrazine Simazine

Log Kow
-0.7 -1.2 0 1.6 1.2 2.7 2.3

Recovery (%)
< 10 < 10 71.1 77.8 80.3 72.0 71.3

S.D
5.8 4.0 4.6 9.9 5.7

The sensitivity of this set-up allowed the reduction of the sample volume to 5 mL. This reduced percolated volume of sample facilitates recoveries closed to 100% for ammeline and ammelide. The use of a more sensitive detection system, i.e. MS detection, allows the analysis of all the metabolites in spring water spiked at 0.1 g/L and the monitoring of these polar metabolites with a totally automated system. The performance of the method is illustrated in Figure 3.

Table 1: Extraction recoveries obtained for the percolation of 50 mL of pure water spiked at 0.5 g/L with each compound, with the online PGC precolumn (10 x 2 mm) / PGC/UV system. S.D standard deviation (n = 5).

Conclusion
A fully automated system consisting of a Hypercarb pre-column connected to LC/UV/MS, using a Hypercarb analytical column, was set-up for the analysis of the polar dealkylated degradation products of atrazine, namely DEA, DIA, DDA, ammeline and ammelide. This methodology was applied to the detection of the metabolites present at trace levels (0.1 g/L) in spring water.

Figure 2 shows the application of this on-line coupling to the analysis of a water sample spiked with 0.2 g/L of each target analytes. In order to increase the sensitivity of the method the column diameter was reduced to 1 mm and MS detection was carried out in series with UV detection.

Application Notes

References
1. H. Sabik, R. Jeannot, B. Rondeau, J. Chromatogr. A, 885 (2000) 217. 2. V. Pichon, J. Chromatogr. A, 88 5 (2000) 195. 3. V. Pichon, M.-C. Hennion, J. Chromatogr. A, 665 (1994) 269. 4. M.-C. Hennion, C. Cau-Dit-Coumes, V. Pichon, J. Chromatogr. A, 823 (1998) 147. 5. M.-C. Hennion, J. Chromatogr. A, 885 (2000) 73. 6. M. Berg, S.R. Mller, R.P. Schwarzenbach, Anal. Chem., 67 (1995) 1860. 7. S.Y. Panshin, D.S. Carter, E.R. Bayless, Environ. Sci. Technol., 34 (2000) 2131. 8. R. Jeannot, H. Sabik, E. Sauvard, E. Genin, J. Chromatogr. A, 879 (2000) 51. 9. V. Pichon, L. Chen, S. Guenu, M .-C. Hennion, J. Chromatogr. A, 711 (1995) 257. 10. G. Machtalre, V. Pichon, M.-C. Hennion, J. High Resol. Chromatogr., 23 (2003) 437.

Figure 2: On-line analysis of water sample spiked with 0.2 g/L of each analyte using the PGC/PGC/UV set up. UV detection at 220 nm

11. S. Guenu, M.-C. Hennion, Anal. Method and Instrument., 2 (1995) 247. 12. S. Dupas, S. Guenu, V. Pichon, A. Montiel, B. Welt, M.-C. Hennion, Intern. J. Environ. Anal. Chem., 65 (1996) 53.

Figure 3: Analysis of 5 mL of spring water spiked at 0.1 g/L with each analyte by the on-line PGC/PGC/UV/MS system. Insert: MS signal (SIM mode)

29

ENVIRONMENTAL
Application Notes
30

Fast and Versatile Analysis of Desphenyl-Chloridazone and Methyl-Desphenyl-Chloridazone in Surface, Ground and Drinking Water Using LC-MS/MS and EQuan
N. Eer, B. Preu, F. Brille, Bergisches Wasser- und Umweltlabor der BTV GmbH, Wuppertal, Germany O. Scheibner, Thermo Fisher Scientic, Dreieich, Germany

Introduction
Growing concerns about contaminants in water and food has resulted in governmental regulations with lower tolerance levels nearly every year. The number of samples has increased nearly at the same speed during the last few years. Thus the demand for shorter and less laborious techniques for residue analysis, with increased sensitivity levels, is higher than ever. The analysis of pesticides and other pollutant compounds such as pharmacological substances in water samples can be highly time-consuming due to the high volumes which need to be extracted. Direct injection of water samples is limited to a volume in the range of 100 L, if one wishes to avoid affecting the peak shape. Automation of sample workup and use of fast HPLC methods for further method acceleration are the demands in modern residue analysis. The Thermo Scientific EQuan system provides a solution for these demands: injection of up to 5000 L crude water sample onto an enrichment column with subsequent HPLC analysis results in full automation and high throughput. Methods for various pesticides, antibiotics and veterinary residues have been

described already, but still the automated analysis of highly polar compounds remains a major challenge since the reliable and quantitative trapping and re-elution during pre-concentration has not yet been satisfactorily resolved. We describe a method for a fully automated analysis of metabolites of the well known herbicide Chloridazone, Desphenyl-Chloridazone and Methyl-Desphenyl-Chloridazone. Due to their polarity, quantitative trapping and sensitive HPLC analysis had not been achieved so far. Use of a Thermo Scientific Hypercarb trapping column (20 x 2.1 mm, 7 m) in combination with a Hypercarb analytical column (100 x 2.1 mm, 3 m) has resulted in a highly reliable and sensitive method for the direct analysis of surface, ground and drinking water.

Goal
The aim of our work was to establish a fast and versatile LC-MS/MS method using EQuan for both metabolites of Chloridazone providing a LOQ of less than 10 ng/L in all water samples.

Figure 1: Schematic drawing of the EQuan system

ENVIRONMENTAL

Experimental
Sample Preparation For the dilution series 10 g ultrapure water was passed through a syringe tip lter and spiked with the given levels of Desphenyl-Chloridazone and Methyl-Desphenyl-Chloridazone. Six levels with the following concentrations were prepared: 5, 6, 7, 8, 15 and 20 ng/L.
Samples were only passed through a syringe tip filter. No additional treatment of samples was necessary prior to measurement.

MS Conditions Ionisation: ESI in positive mode Spray voltage: 3500 V Sheath gas pressure (N2): 40 units Auxiliary gas (N2): 5 units Ion transfer tube temperature: 350 C Collision gas pressure (Ar): 1.5 mTorr Q1/Q3 peak resolution: 0.7 Da Scan width: 0.01 u Scan time: 0.6 s
The SRM transitions used are shown in Table 1. The product masses marked with * were used for quantication.
Analyte
DesphenylChloridazone Methyl-DesphenylChloridazone Table 1: SRM conditions

Instrumentation Thermo Scientific Surveyor LC pump for sample concentration Thermo Scientific Surveyor MS pump for sample analysis CTC PAL autosampler for sample application Thermo Scientific TSQ Quantum Access triple stage mass spectrometer. Sample Pre-concentration Hypercarb column 20 x 2.1 mm, 7 m (part number 35007-022130)
A sample volume of 1 mL was transferred to the trap column with 20% methanol in water as solvent. Sample transfer and concentration was carried out with a ow rate of 500 L/min. The trap column was switched by means of the 6-port valve (Figure 1) into the ow of the analytical pump and followed by the introduction of an analytical gradient to both columns.

Parent Ion (m/z )


145.9 159.9

Fragment Ions (m/z )


101.1 117.1* 88.2* 117.0

Collision Energy (V)


28 23 31 23

Application Notes

Results and Discussion


The chromatogram of a real water sample spiked with 10 ng/L each of Despenyl-Chloridazone and Desphenyl-MethylChloridazone is shown in Figure 2. Thermo Scientific LCQuan 2.5 software was used to process the quantitative data. Figure 4 shows the calibration curves of DesphenylChloridazone and Methyl-Desphenyl-Chloridazone and Table 2 summarizes the correlation coefficients, LODs and LOQs for both analytes.

Chromatographic Conditions Column: Hypercarb 100x2.1 mm, 3 m (part number 35003-102130) Mobile phase: A Water +0.5% methanol; B Methanol Gradient: Time (min) %A %B 0 30 70 2 30 70 11 0 100 14 0 100 17 30 70 20 30 70 Flow rate: 200 L/min
The time for sample concentration was 1.9 min. The total cycle time for sample concentration, analysis and reconditioning was 20 minutes.

Figure 2: Chromatogram of a water sample spiked with a concentration of 10 ng/L each of Desphenyl-Chloridazone and Methyl-Desphenyl-Chloridazone

31

ENVIRONMENTAL
Application Notes
32

Figure 3: Calibration curves of Desphenyl-Chloridazone and Methyl-Desphenyl-Chloridazone

A concentration range from 5 to 20 ng/L was covered with pure standards, with focusing on the lower end of the range for quantitation in residue analysis from aqueous samples as specified in current legislation (Figure 3). A real sample containing 10 ng/L each of the two analytes shows peaks well suitable for quantitation which can be considered as limit of quantitation under such conditions. Less concentrated samples showed a limit of detection at less than 5 ng/L.
Correlation Coefficient r2
0.9988 0.9992

Conclusion
The online enrichment and analysis of highly polar compounds, especially polar metabolites of pesticides, is considered challenging due to insufcient trapping capabilities of available column materials. This application demonstrates that it is possible to build a system for online enrichment and automated LC-MS/MS analysis of polar metabolites of an important pesticide widely used in Europe. A limit of quantitation of 10 ng/L, after pre-concentration from a sample volume of 1 mL, was demonstrated. Normally, sample volumes used for manual pre-concentration are much higher. The EQuan system provides capability to enlarge the sample volume up to 20 mL, thus opening the way to quantitation levels of 1 ng/L and below.

Analyte
Chloridazonedesphenyl Chloridazonemethyl-desphenyl

Limit of Detection (ng/L)


3.1 3.5

Limit of Quantitation (ng/L)


10 10

Table 2: LODs and LOQs for the two pesticide metabolites.

Determination of Leucine and its Isomers by LC-MS/MS Using a Porous Graphitic Carbon Column
Agnes Le Corre, Thermo Fisher Scientific, Paris, France Luisa Pereira, Thermo Fisher Scientific, Runcorn, UK

CLINICAL

Introduction
Leucine and valine are used as biomarkers of maple syrup urine disease in neonatal blood spot screening. Maple syrup urine disease (MSUD) is caused by a gene defect. Infants with this condition cannot break down the branched-chain amino acids leucine, isoleucine, and valine. This leads to a build-up of these chemicals in the blood. The condition gets its name from the distinctive sweet odor of affected infants urine. Beginning in early infancy, this condition is characterized by poor feeding, vomiting, lethargy, and developmental delay. MSUD affects an estimated 1 in 185,000 infants worldwide. Traditionally, the monitoring of the concentration of leucine and its isomers (isoleucine and allo-isoleucine) in blood is performed with an amino acid analyser, which can take up to 2 hours to complete the analysis. LC-MS/MS is becoming an important tool in the semi-quantitative analysis of amino acids in neonatal screening. When applied to the screening and monitoring of MSUD, this methodology requires the chromatographic resolution of leucine, leucine isomers and hydroxyproline as these are all isobaric and produce the same m/z at 188 ([M+H]). Porous graphitic carbon (PGC, Hypercarb) can be used to retain and separate polar underivatised amino acids.1 Furthermore, the flat and highly adsorptive surface of the graphite shows increased selectivity towards structurally related compounds such as structural isomers. This application note describes a LC-MS/MS method for the analysis of leucine, its isomers isoleucine and allo-isoleucine, valine and hydroxyproline, using a Hypercab column to achieve chromatographic separation of the 5 compounds.

Experimental
Chromatographic Conditions LC system: Thermo Scientific Surveyor Autosampler: HTS Pal Column: Hypercarb 100 x 4.6 mm, 5 m (part number 35005-104630) Mobile phase: water + 20 mM nonafluoropentanoic acid/Acetonitrile (75:25) Flow rate: 1500 L/min (split 1/10) Injection volume: 10 L MS Conditions MS system: Thermo Scientific TSQ Quantum Ultra Ionisation Mode: +ESI Selected reaction monitoring (SRM) transitions:
Compound
Leucine Isoleucine Allo-Isoleucine Hydroxy-Proline Valine Isoleucine Allo-Isoleucine (Conrmation ion)

Application Notes

Precursor Ion
188.09

Product Ion
86.1

Collision Energy (eV)


15

174.15 188.09

72.15 69.1

20 32

Table 1: SRM transitions used in the analysis

Sample Preparation The calibration standards were prepared by successive dilutions with water/acetonitrile (75:25) mixture. Solutions were stored at +4 C.

Goal
To develop a method for the LC/MS/MS analysis of leucine and isomers, with full chromatographic separation.

Results and discussion


The stereoselectivity of Hypercarb allows chromatographic resolution of the isobaric species in the analysis: leucine, isoleucine, allo-isoleucine and hydroxyproline. This is achieved using an isocratic mobile phase of water and acetonitrile with the addition of 0.1% nonafluoropentanoic acid, and illustrated in Figure 1. In addition to retention time, the differentiation of leucine and its isomers can be done by monitoring alternative SRM transitions. At 15 eV, leucine, isoleucine and allo-isoleucine all produce a major product ion at m/z 86 (as shown in Figure 1B). However, when collision energy is increased to 32 eV, another product ion at m/z 69 is present in the spectra of isoleucine and allo-isoleucine but absent from the spectra for leucine. Therefore monitoring of SRM transition 188.09 69.1 will produce a signal for isoleucine and allo-isoleucine but not for leucine, as demonstrated in Figure 2.

33

CLINICAL
Application Notes
Figure 1: LC-MS/MS traces showing chromatographic separation of (1) Hydroxy-proline; (2) Valine; (3) Leucine; (4) Allo-isoleucine; (5) Isoleucine. (A) Total ion chromatogram; (B) SRM chromatogram for isobaric compounds of m/z 188; (C) SRM chromatogram for valine. Figure 2: Differenciation of isoleucine and allo-isoleucine from isobaric leucine: SRM chromatogram for transition 188.09 69.10 at 32 eV collision energy. No peak for leucine is present.

Conclusion
An eight minute method is demonstrated for the analysis of leucine and isomers. Chromatographic resolution of these underivatised amino acids, valine and hydroxyproline is achieved on a porous graphitic carbon column. The sensitivity of the method could be improved by reducing the column internal diameter and therefore reduce both the split ratio into the MS and sample dilution in the column. Future work should also include the investigation of the effect of changing the electronic modifier (nonafluoropentanoic acid) and investigate its effect on chromatographic resolution and method sensitivity.

References
1. P. Chaimbault, K. Petritis, C. Elfakir, M. Dreux, J. Chromatogr. A, 2000, 896, 253-263

34

Determination of Occupational Exposure to Toluene, Xylene and Styrene Through Metabolite Monitoring Using Isocratic HPLC
Luisa Pereira, Brid Brosnan, Charlotte Blythe, Thermo Fisher Scientific, Runcorn, UK

CLINICAL

Introduction
Toluene, xylene and styrene are widely used as an industrial feedstock and in the modern scientific workplace. Styrene is the precursor compound to polystyrene. Toluene is commonly used in the paint, plastics and printing industries. Xylene is consumed in the leather, printing and rubber industry. Exposure to styrene, toluene or xylene has a detrimental effect upon the central nervous system. At low concentrations, symptoms of dizziness, headache, nausea and vomiting are experienced together with respiratory tract irritation. Exposure can cause in-coordination and mental confusion and, as the central nervous system becomes further depressed, can result in unconsciousness and death. The monitoring and evaluation of the biological effects upon health of workers exposed to such chemicals is, therefore, vitally important. Traditionally, exposure to these chemicals has been evaluated though measurement of their concentration in ambient air where workers are more likely to be exposed to, however, it has been found in recent years that biological monitoring (in blood or urine) affords a far more accurate estimate of exposure. For example, in the presence of ethanol, the adsorption of toluene into the blood is double that which occurs with exposure to toluene alone, thus illustrating that measurement of air concentration can result in misleading exposure estimates.1 Upon inhalation/absorption into the body, all three chemicals undergo biological transformation. The metabolic reaction rates are fast, consequently, the products are rapidly excreted from the body. As such, levels of compounds in blood are low, thus the concentration of metabolites in urine is measured. The main metabolites are hippuric acid, o-, m-, p-methylhippuric acid, mandelic acid and phenylglyoxylic acid. These metabolites are polar, and isomeric in nature. Consequently, Thermo Scientific Hypercarb (porous graphitic carbon, PGC) presents itself as the ideal phase for the separation of such analytes. Indeed Schlatter et al highlighted the advantageous positional isomer separation performance of Hypercarb with respect to the analysis of cresol isomers in the urine of workers exposed to toluene.2 As hippuric acid is present in the urine of individuals who have not been subject to toluene exposure, some authors believe that measurement of cresols provides a valid alternative to the monitoring of hippuric acid. This application note highlights the use of Hypercarb for the effective separation of the associated compounds of styrene, toluene and xylene metabolism. Using a simple two component isocratic mobile phase all major metabolites (and an internal standard) are separated in under 11 minutes.

Goal
To detail the effective monitoring of excreted toluene, xylene and styrene metabolites using a PGC stationary phase in isocratic mode.

Application Notes

Experimental
LC Method Column: Hypercarb 3 m, 100 x 4.6 mm (part number 35003-104630) Mobile Phase: H2O / MeOH (50:50) +0.1% TFA Flow Rate: 1 mL/min Temperature: 25 C Detection: UV at 240 nm Injection Volume: 10 L

Results and Discussion


The separation achieved for the biological metabolites using the conditions described above is illustrated in Figure 1. The analysis is achieved in less than eleven minutes with all analytes possessing good peak shape (greatest asymmetry at 10% value measured for mandelic acid at 1.18).

Figure 1: Metabolite separation: 1. Mandelic acid; 2. o-Methylhippuric acid; 3. Hippuric acid; 4. Phenyl-gyloxylic acid; 5. p-Hydroxybenzoic acid (internal standard); 6. m-Methylhippuric acid; 7. p-Methylhippuric acid.

35

CLINICAL
Application Notes
Figure 2: Linearity in the concentration range of 1 to 100 g/mL.

The suitability of the method for quantitative analysis was investigated through linearity and reproducibility assessment. The calibration curve displayed in Figure 2 shows that all analytes generate a true linear response over the concentration range of 1 100 g/mL with the lowest R2 coefficient value being recorded at 0.9978 for hippuric acid. With respect to reproducibility of area, ratio of analyte area to internal standard (p-hydroxybenzoic acid) and repeatability of retention times (Table 1), the percentage relative standard deviation values recorded over twenty injections (equating to 240 column volumes) were low. The highest value observed was for the ratio of analyte area to internal standard area for 2-methylhippuric acid which was calculated at 1.15%.
Compound
Mandelic acid o-Methylhippuric acid Hippuric acid Phenyl-gyloxylic acid p-Hydroxybenzoic acid m-Methylhippuric acid p-Methylhippuric acid

Conclusion
The use of Hypercarb for the effective separation of the associated compounds of styrene, toluene and xylene metabolism is demonstrated Using a simple two component isocratic mobile phase all major metabolites (and an internal standard) are separated in under 11 minutes. The method is reproducible and linear in the concentration range of 1 to 100 g/mL.

References
1. Walln M., Nslund P.H., Byflt Nordqvisat M., Toxicol Appl Pharmacol, 1984: 76:414-419. 2. Schlatter J., Astier A., Biomed. Chromatogr. 1995, 9, 302-304. 3. Lauweys R.R., Hoet P., Industrial Chemical Exposure Guidelines for Biological Monitoring, 3rd Edition, Lewis Publishers. 4. International Agency for Research on Cancer (IARC). IARC Monographs on the Evaluation of Carcinogenic Risk to Humans, Some Industrial Chemicals. Lyon: IARC, 1994, Vol. 60, 233-246.

%RSD Retention times


0.28 0.47 0.63 0.87 0.76 0.51 0.52

%RSD Area Ratio


1.09 1.15 1.05 0.86 1.05 1.03

Table 1: Method reproducibility (n = 20 runs).

36

FOOD

Porous Graphitic Carbon for Inorganic Ion Analysis in Drinking Water


A. Thompson, A. H. Marks Ltd, Wyke, Bradford, UK L. Pereira, Thermo Fisher Scientific, Runcorn, UK

AND

BEVERAGE

Introduction
The analysis of inorganic ions is routinely performed on a range of support materials, which include functionalised styrene-divinylbenzene copolymers, coated silica gels and even traditional ODS (C18) phases used in conjunction with ion pair reagents. Hypercarb has been previously used for the analysis of inorganic ions. Elfakir and co-workers demonstrated the separation of hydrogenophosphate, sulphate, nitrate, and perclorate with a volatile electronic interaction additive (formic, acetic, or perfluoro-carboxylic acid) in the aqueous mobile phase.1, 2 Takeuchi et al. used sodium sulphate in the mobile phase to retain and separate iodate, bromide, nitrite, bromate, nitrate and iodide on a PGC column.3 Ion interaction chromatography with tetrabutylammonium hydroxide was has also been used to separate inorganic ions on PGC.4 The same authors dynamically coated the surface of graphite with cetyltrimethylammonium (CTA) ions for the separation of seven common anions in water: F-, Cl-, NO2-, Br-, NO3-, HPO42-, SO42-.5 Using CTA-bromide as the coating agent, a permanently coated ion-exchange column was obtained which allowed separations of the anions without any coating agent in the mobile phase. This coating process was used in the work described in this application note. In this application note it is demonstrated that by dynamically coating the Hypercarb surface with an ion exchanger a stable surface suitable for the ion chromatography of common inorganic anions is produced. We investigate the effect of coating coverage on the retention and resolution of anions whilst demonstrating good reproducibility, linearity, and excellent column coating lifetime. Two brands of bottled drinking water are analysed for inorganic anions using the methodology developed.

Hypercarb Coating Process Step 1. Column flushed with H2O at 1 mL/min for 60 minutes. Step 2. Column flushed with CTA 0.5 mM in H2O/MeCN (75:25), at 1 mL/min for 120 minutes. Step 3. Column flushed with H2O at 1 mL/min for 60 minutes. Ion Chromatograph Dionex DX 300 Ion Chromatograph Anion micromembrane suppressor, Dionex 4400 integrator Rheodyne 7125 injector (20 L loop) Separation Conditions Mobile Phase: 2 mM Na2CO3/1 mM NaHCO3 + 2.5% MeCN Flow Rate: 1.2 mL/min Detection: Suppressed conductivity Injection Volume: 20 L

Application Notes

Results and Discussion


The separation of six inorganic anions commonly found in water was obtained in less that 15 minutes and is illustrated in Figure 1. Fluoride is retained away from the solvent front, allowing for easy quantification.

Goal
To study the suitability of porous graphitic carbon dynamically modified with cetyltrimethylammonium bromide for the selective ion chromatography of inorganic anions in drinking water.

Figure 1: Separation of six anion mixture: 1. Fluoride; 2. Chloride; 3. Bromide; 4. Nitrate; 5. Phosphate; 6. Sulphate.

Experimental
Column Hypercarb 5 m, 100 x 4.6 mm (part number 35005-104630) Ion exchanger Cetyltrimethylammonium bromide (CTA) is adsorbed onto the surface of Hypercarb during coating process.

The degree of coating of the surface can be manipulated to offer changes in retention times and therefore the resolution of common anions, which provides the flexibility of tailoring the column performance to suit particular application needs. The concentration of CTA was increased whilst the remaining conditions for coating remained constant, giving increased adsorption of the ion exchanger onto the Hypercarb surface. The increased ion exchange capacity leads to increased retention times and therefore greater resolution of anions, as demonstrated in Figure 2.

37

FOOD
AND

BEVERAGE
Application Notes
38

Figure 2: Effect of coating concentration on retention behavior.

The suitability of the method for quantitative analysis was investigated through a linearity assessment. Calibration curves (Figure 3) were generated by plotting the peak response area against the concentration of the standards injected. For fluoride and chloride 5 concentrations between 1 mg/L and 10 mg/L were used. For bromide, nitrate, phosphate and sulphate concentrations between 10 mg/L and 100 mg/L were used. All analytes generate a true linear response over the concentration ranges described, with the lowest R2 coefficient value being recorded at 0.9978 for phosphate.

To evaluate the reproducibility of area and repeatability of retention times, the percentage relative standard deviation (RSD) values were recorded over ten and twenty injections respectively (Table 1). The highest value observed for peak area RSD was 0.5% for fluoride and the highest RSD value for retention time was 1.1%. The lifetime of the coated column was investigated by doing repeated injections of the anion mixture. Retention times remain constant for 445 runs, which equates to 16000 column volumes of mobile phase through the column (Figure 4). As the coating process involves the use of 25% organic, it ensures that using 5% organic during analysis will not be detrimental to column coating lifetime. The dynamic coating of the PGC surface is reversible. Flushing the column with 100% organic will remove all column coating, and the column can then be re-coated using original coating conditions or used as a conventional Hypercarb column.
Compound
Fluoride Chloride Bromide Nitrate Phosphate Sulphate

%RSD Peak Area


0.51 0.47 0.26 0.20 0.57 0.27

%RSD Retention Time


0.21 0.30 0.57 0.75 0.47 1.09

Table 1: Reproducibility: 10 repeated measurements for peak areas and 20 repeated measurements for retention times.

The methodology developed was applied to the measurement of the level of inorganic anions present in two brands on bottled water, as demonstrated in Figure 5.

Figure 3: Linearity data in the concentration range of 10 to 100 mg/L for sulphate, nitrate, bromide and phosphate and 1 to 10 mg/L for fluoride and chloride.

FOOD
AND

BEVERAGE
Application Notes

Figure 4: Column coating lifetime. Reproducible retention times for 16000 column volumes of mobile phase through the column or 445 injections. Column: Hypercarb 5 m, 100 x 4.6 mm (CTA 8 mM); Mobile phase: 2 mM Na2CO3/1 mM NaHCO3 + 5% MeCN; other conditions as described in the experimental section.

Conclusion
Dynamically CTA coated Hypercarb columns possess good stability, reproducibility, linearity and lifetime performance working with organic compositions up to 25%. This approach offers an alternative to the conventional methodology for the determination on inorganic anions in drinking water.

References
1. C. Elfakir, P. Chaimbault, M. Dreux, J. Chromatogr. A, 829 (1998) 193 2. K. Petritis, P. Chaimbault, C. Elfakir, M. Dreux, J. Chromatogr. A, 1999, 833, 147 3. T. Takeuchi, T. Kojima, T. Miwa, J. High, Resol. Chromatogr., 2000, 23, 590 4. T. Okamoto, A. Isozaki, H. Nagashima, J. Chromatogr. A, 1998, 800, 239 5. H. Nagashima, T. Okamoto, J. Chromatogr. A, 1999, 855, 261-266

Figure 5: Analysis of bottled drinking water.

39

Applications Review
The table below provides an overview of the applications published between mid -1995 and 2008; these are organised in the following subsections: isomers (geometrical, positional and diastereoisomers), polar compounds (nucleotides/nucleosides/ nucleobases, amino acids and peptides, carbohydrates and sugars, and other polar molecules), charged solutes and miscellaneous. The miscellaneous and other polar molecules subsections are further categorised into application areas such as food safety, environmental, pharmaceutical, etc. Applications areas for PGC which are outside the scope of this review and therefore not covered include high temperature applications, chiral separations, PGC as a sorbent in solid phase extraction, electrically modulated chromatography and electrochromatography.

Applications Reference Guide (Modified from Reference 1.)


Applications Review
40

Application by Solute Type Application Area


Isomers

Analyte Group
Ionizable substitued benzenes Cresol Substituted aromatics p-Nonylphenol Sulfobutyl ether- -cyclodextrin Schumannificine Tropane alkaloids Estrogens F2-Isoprostanes Branched oligosaccharides Leucine, isoleucine, allo-isoleucine Xylose derivative Oligomers of methylidene malonate Metabolites II, III, and VII, VIII Nucleobases, nucleosides, nucleotides Purine bases Nucleosides and their mono-, di- and triphosphates Uracil, dihydrouracil Tegafur, 5-fluorouracil, 5-fluorodihydrouracil Underivatized amino acids Taurine derivatives Di-, tri-, tetra-peptides Small peptides in wine Phosphopeptides Glycopeptides Mono-, disaccharides Cyclodextrins Oligosaccharide alditols N-linked oligosaccharides Oligosaccharides Sugar phosphates Sulphated disaccharides Glycosaminoglycans Carrageenans Catecholamines/neurotransmiters Glutathione and conjugates Oligonucleotides Creatinine/creatine

Reference Number
2, 3 4 5 6 7 8 9 10 11, 12 13 14 15 16 17 16 19 20 21 22 23 25 24, 27 26 27, 28 29 29, 33, 42 29, 46, 47, 7 29, 35, 36 29, 34, 35, 38, 41 29, 36, 37, 39, 40, 42 30 31 32 43, 44, 45 48, 49, 50, 51 52, 53 54 55

Diastereoisomers

Nucleotides/Nucleosides

Amino Acids and Peptides

Carbohydrates and Sugars

Other Polar Species Biochemical

Application Area
Pharmaceutical

Analyte Group
Clopidogel and metabolites Acarbose and metabolites Metabolites in Escherichia coli K12 Arabinoside-CMP, cytarabine Cimetidine Levetiracetam Oxaliplatin EDTA impurities Methylamines Polar Phenolic Acrylamide Acromelic acid A Aniline Glyphosate and Ampa Cyanuric acid Alkylglycoside detergents Polyethoxylated alcohols Oligoglycerols Oligomers of nonylphenyl ethylene Polyglycerol fatty esters and fatty ethers Cyanoglycosides Glucosinolates Phosphonic acids Guanidino compounds Diphosphine-bridged complexes Pertechnetate and perrhenate ions Organometallic-charged complexes Inorganic anions Copper (II), copper (III) Diquat, paraquat and difenzoquat Wax esthers Ceramides Fatty acids methyl esters Glycosphingolipids Triacylglycerols Digalactosyldiacylglycerol Triterpenic acids Taxol Non-flavonoid polyphenols Ferrichrome and ferricrocin Morphine and metabolites Pharmaceuticals and related substances Steroids Cyclosporins A and U Tetracycline antibiotics Phthalate metabolites Boron-containing compounds Flame retardants hydrolysis products Benzo[ ]pyrene PCBs Halogenated contaminants and PAHs Betamethasone and dexamethasone Glucocorticoids and corticosteroids Fenbutin oxide Nitroaromatic and organic explosives Nitrate ester, nitramine and nitroaromatic explosives

Reference Number
56 57 58 59, 60 61 64 63 62 65 66 67, 68 69 70 71 72 73 74 75 76 77 78 79 80 81, 82 83 84 85, 86 87-89, 91, 92 90 93 94 94 -97 94, 98-100 101 107 102 103 104, 105 106 107 108 109-113 114 115 116 117 118 119 120 121-123 124 125, 126 127-129 130 131-135 136

Food Safety

Environmental

Surfactants

Applications Review

Natural Products Chemical Warfare Other


Charged Solutes

Non-polar Solutes Lipids

Natural Products

Pharmaceutical

Clinical Environmental

Food Safety

Explosives

41

References
1. L. Pereira, J. of Liq. Chromatogr. & Related Technol., 2008, 31:11, 1687-1731 2. Q. Wan, P. Shaw, M. Davies, D. Barret, J. Chromatogr. A, 1995, 697, 219-227 3. Q. Wan, P. Shaw, M. Davies, D. Barret, Anal. Chem., 1996, 68, 437-446 4. A. Schlatter, Biomed. Chromatogr., 1995, 9, 302-304 5. C. West, E. Lesellier, J. Chromatogr. A, 2005, 1099, 175-184 6. J. Gundersen, J. Chromatogr. A, 2001, 914, 161-166 7. R. Jacquet, R. Pennanec, C. Elfakir, M. Lafosse, J. Sep. Sci., 2004, 27, 1221-1228 8. P. Houghton, T. Woldermariam, Phytochemical Anal., 1995, 6, 85-88 9. S. Bieri, E. Varesio, O. Muoz, J.-L. Veuthey, P. Christen, J. Pharma. Biomed. Anal., 2006, 40, 545-551 10. J. Reepmeyer, J. Brower, H. Ye, J. Chromatogr. A, 2005, 1083, 42-51 11. K. Bohnstedt, B. Karlberg, L.-O. Wahlund, M. Jonhagen, H. Basun, S. Schmidt, J. Chromatogr. B, 2003, 796, 11-19 12. K. Bohnstedt, B. Karlberg, H. Basun, S. Schmidt, J. Chromatogr. B, 2005, 827, 39-43 13. Y. Okada, M. Semma, Y. Ito, Carbohydrate Res., 2001, 336, 203-211 14. D. Cooper, M. Morris, S. Kim, Waters Application Brief (WAB22), 2002 15. M. Chiron, N. Bonaventure, L. Boisaubert, V. Manzin, Poster presented at HPLC 2005 conference, 2005, Stockolhm 16. A. Salvadora, B. Herbreteaua, P. Bretonb, D. Royc, P. Brigandc, N. Brub, M. Dreux, Anal. Chim. Acta, 1998, 359, 57-64 17. Y.Q. Xia, M. Jemal, N. Zheng, X. Shen, Rapid Commun. Mass Spectrom., 2006, 20, 18311837 18. D. Gasparutto, J.-L. Ravanat, O. Gerot, J. Cadet, J. Am. Chem. Soc., 1998, 120, 10283-10286 19. L. Monser, Chromatographia, 2004, 59, 455-459 20. J. Xing, A. Apedo, A. Tymiak, N. Zhao, Rapid Commun. Mass Spectrom., 2004, 18, 15991606 21. G. Remaudb, M. Boisdron-Celle, C. Hameline, A. Morel, E. Gamelin, J. Chromatogr. B, 2005, 823, 98-107 22. G. Remaudb, M. Boisdron-Celle, A. Morel, E. Gamelin, J. Chromatogr. B, 2005, 824, 153-160 23. P. Chaimbault, K. Petritis, C. Elfakir, M. Dreux, J. Chromatogr. A, 2000, 896, 253-263 24. A. Adoubel, S. Guenu, C. Elfakir, M. Dreux, J. Liq. Chrom. & Rel Technol., 2000, 23, 2433-2446 25. P. Chaimbault, P. Alberic, C. Elfakir, M. Lafosse, Anal. Biochem., 2004, 332, 215-225 26. C. Desportes, M. Charpentier, B. Duteurtre, A. Maujeanc, F. Duchiron, J. Chromatogr. A, 2000, 893, 281-291 27. E. Chin, D. Papac, Anal. Biochem., 1999, 273, 179-185 28. P. Vacratsis, B. Phinney, D. Gage, K. Gallo, Biochemistry, 2002, 41, 5613-5624 29. K. Koizumi, J. Chromatogr. A, 1996, 720, 119-126 30. C. Antonio, T. Larson, A. Gilday, I. Graham, E. Bergstrm, J. Thomas-Oates, J. Chromatogr. A, 2007, 1172, 170-178 31. B. Barroso, M. Didraga, R. Bischoff, J. Chromatogr. A, 2005, 1080, 43-48 32. N. Karlsson, B. Schulz, N. Packer, J. Whitelock, J. Chromatogr. B, 2005, 824, 139-147 33. N. Nielsen, K. Granby, R. Hedegaard, L. Skibste, Anal. Chim. Acta, 2006, 557, 211-220 34. N. Packer, M. Lawson, D. Jardine, J. Redmond, Glycoconjugate J., 1998, 15, 737-747 35. S. Itoh, N. Kawasaki, M. Ohta, M. Hyuga, S. Hyuga and T. Hayakawa, J. Chromatogr. A, 2002, 968, 89-100 36. J. Yuan, N. Hashii, N. Kawasaki, S. Itoh, T. Kawanishi, T. Hayakawa, J. Chromatogr. A, 2005, 1067, 145-152 37. B. Barroso, R. Dijkstra, M. Geerts, F. Lagerwerf, P. van Veelen, A. de Ru, Rapid Commun. Mass Spectrom., 2002, 16, 1320-1329 38. B. Schulz, N. Packer, N. Karlsson, Anal. Chem., 2002, 74, 6088-6097 39. S. Robinson, E. Bergstrm, M. Seymour, J. Thomas-Oates, Anal. Chem., 2007, 79, 2437-2445 40. N. Kawasaki, M. Ohta, S. Hyuga, O. Hashimoto, T. Hayakawa, Anal. Biochem., 1999, 269, 297-303 41. N. Kawasaki, M. Ohta, S. Hyuga, M. Hyuga and T. Hayakawa, Anal. Biochem., 2000, 285, 82-91 42. M. Davies, E. Hounsell, J. Chromatogr., 1996, 720, 227-233 43. A. Antonopoulos, B. Herbreteau, M. Lafosse, W. Helbert, J. Chromatogr. A, 2004, 1023, 231-238 44. A. Antonopoulos, P. Favetta, M. Lafosse, W. Helbert, J. Chromatogr. A, 2004, 1059, 83-87 45. A. Antonopoulos, P. Favetta, W. Helbert, M. Lafosse, J. Chromatogr. A, 2007, 1147, 37-41 46. I. Clarot, D. Cldat, L. Boulkanz, E. Assidjo, T. Chiana, P. Cardot, J. Sep. Sci., 2000, 38, 38-45 47. Kwaterczak, Bielejewska, Anal. Chim. Acta, 2005, 537, 41-46 48. A. Trnkvist, P. Sjoberg, K. Markides, J. Bergquist, J. Chromatogr. B, 2004, 801, 323-329 49. P. Koivisto, A. Trnkvist, E. Heldin, K. Markides, Chromatographia, 2002, 55, 39-42 50. S. Rinne, A. Holm, E. Lundanes, T. Greibrokk, J. Chromatogr. A, 2006, 1119, 285-293 51. D. Thibaut, J. Vial, M. Michel, M.-C. Hennion, T. Greibrokk, J. Chromatogr. A, 2006, 1122, 97-104 52. A. Sakhi, K. Russnes, S. Smeland, R. Blomhoff, T. Gundersen, J. Chromatogr. A, 2006, 1104, 179189 53. T. Lindemann, H. Hintelmann, Anal. Chem., 2002, 74, 4602-4610 54. D. Gasparutto, J.-L. Ravanat, O. Grot, J. Cadet, J. Am. Chem. Soc., 1998, 120, 10283-10286 55. T. Smith-Palmer, J. Chromatogr. B, 2002, 781, 93-106 56. A. Mitakos, I. Panderi, Anal.Chim. Acta, 2004, 505, 107-114 57. Y. Daali, S. Cherkaoui, X. Cahours, E. Varesio, J.-L. Veuthey, J. Sep. Science, 2002, 25, 280-284 58. A. Buchholz, R. Takors, C. Wandrey, Anal. Biochem., 2001, 295, 129137 59. M.-H. Gouya, H. Fabre, M. Blanchin, S. Peyrottes, C. Prigaud, I. Lefebvre, Anal. Chim. Acta, 2006, 566, 178184 60. Y. Hsieh, C. Duncan, J.-M. Brisson, Rapid Commun. Mass Spectrom., 2007, 21, 629-634 61. N. Helali, L. Monser, Chromatographia, 2006, 63, 425-430 62. H. Altorfer, Pharmaeuropa, 2002, 14, 637-639 63. H.Ehrsson, I. Wallin, J. Chromatogr. B, 2003, 795, 291-294 64. J. Martens-Lobenhoffer, S. Bode-Boger, J. Chromatogr. B, 2005, 819, 197-200 65. L. Monser, G. Greenway, Anal. Chim. Acta, 1996, 322, 63-68 66. J. Vial, M.-C. Hennion, A. Fernandez-Alba, A. Aguera, J. Chromatogr. A, 2001, 937, 21-29 67. H. Rosen, Analyst, 2002, 127, 880-882 68. A. Claus, G. Weisz, D. Kammerer, R. Carle, A. Schieber, Mol. Nutr. Food Res., 2005, 49, 918 925

Applications Review
42

69. J. Bessarda, P. Saviuc, Y. Chane-Yenea, S. Monneta, G. Bessarda, J. Chromatogr. A, 2004, 1055, 99-107 70. R. Delpe, P. Chaimbault, J.-P. Antignac, M. Lafosse, Rapid Commun. Mass Spectrom., 2004, 18, 1548-1552 71. R. Phillips, LC.GC EUR, 2003, Suppl. S, 7-8 72. R. Cant, O. Evans, F. Kawahara, L. Wymer, A. Dufo, Anal. Chem., 2001, 73, 3358-3364 73. C. Elfakir, M. Lafosse, J. Chromatogr. A, 1997, 782, 191-198 74. P. Chaimbault, C. Elfakir, M. Lafosse, J. Chromatogr. A, 1998, 797, 83-91 75. P. Chaimbault, S. Cassel, S. Claude, C. Debaig, T. Benvegnu, D. Plusquellec, P. Rollin, M. Lafosse, Chromatographia, 1999, 50, 239-242 76. T. Cserhti, E. Forgcs, A. Szilgyi, J. Pharma. Biomed. Anal., 1997, 15, 1303-1307 77. S. Cassel, P. Chaimbault, C. Debaig, T. Benvegnu, S. Claude, D. Plusquellec, P. Rollin, M. Lafosse, J. Chromatogr. A, 2001, 919, 95-106 78. V. Berenguer-Navarro, R.Giner-Galvn, N. Gran-Teruel, G. ArrazolaPaternina, J. Agr. Food Chem., 2002, 50, 6960-6963 79. C. Elfakir, M.Dreux, J. Chromatogr. A, 1996, 727, 71-81 80. J.-P. Mercier, P. Morina, M. Dreux, A. Tambute, J. Chromatogr. A, 1999, 849, 197-207 81. Y. Inamoto, S. Inamoto, T. Hanai, M. Tokuda, O. Hatase, K. Yoshii, N. Sugiyama, T. Kinoshita, J. Chromatogr. B, 1998, 707, 111-120 82. T. Hanai, Y. Inamaoto, S. Inamoto, J. Chromatogr. B, 2000, 747, 123-138 83. C.-H. Li, P. Low, H. Lee, T. Hor, Chromatographia, 1997, 44, 381-385 84. C-K. Lim, Biomed. Chromatogr., 1989, 3, 92-93 85. M. Emery, C-K. Lim, J. Chromatogr., 1989, 479, 212-215 86. G. Gu, C-K. Lim, J. Chromatogr., 1990, 515, 183-... 87. C. Elfakir, P. Chaimbault, M. Dreux, J. Chromatogr. A, 1998, 826, 193-199 88. K. Petritis, P. Chaimbault, C. Elfakir, M. Dreux, J. Chromatogr. A, 1999, 833, 14789. T. Takeuchi, T. Kojima, T. Miwa, J. High, Resol. Chromatogr., 2000, 23, 590-594 90. C. Merly, B. Lynch, P. Ross, J. Glennon, J. Chromatogr. A, 1998, 804, 187- 91. T. Okamoto, A. Isozaki, H. Nagashima, J. Chromatogr. A, 1998, 800, 239-... 92. H. Nagashima, T. Okamoto, J. Chromatogr. A, 1999, 855, 261-266 93. M. Ibez, Y. Pic, J. Maes, Chromatographia, 1997, 45, 402-407 94. F. Deschamps, K.Gaudin , E. Lesellier, A. Tchapla, D. Ferrier, A. Baille, P. Chaminade, Chromatographia, 2001, 54, 607-611 95. K. Gaudin, P. Chaminade, A. Baillet, J. Chromatogr. A, 2002, 973, 69-83 96. C. West, G. Cilpa, K. Gaudin, P. Chaminade, E. Lesellier, J. Chromatogr. A, 2005, 1087, 77-85 97. L. Quinton, K. Gaudin, A. Baillet, P. Chaminade, J. Sep. Sci., 2006, 29, 390-398 98. C. Viron, P. Andr, M. Dreux, and M. Lafosse, Chromatographia, 1999, 49, 137-141 99. C. Viron, A. Saunois, P. Andre, B. Perly, M. Lafosse, Anal. Chim. Acta, 1999, 20316, 1-11 100. K. Gaudin, T. Hanai, P. Chaminade, A. Baillet, J. Chromatogr. A, 2007, 1157, 56-64 101. S. Roy, A. Delobel, K. Gaudin, D. Touboul, D. Germain, A. Baillet, P. Prognon, O. Laprvote, P. Chaminade, J. Chromatogr. A, 2001, 1117, 154-162 102. F. Deschamps, K. Gaudin, A. Baillet, P. Chaminade, J. Sep. Sci., 2004, 27, 1313-1322 103. C. Berangere, N. Caussarieu, P. Morin, L. Morin-Allory, M. Lafosse, J. Sep. Sci., 2004, 27, 964-970

104. V. Nmeth-Kiss, E. Forgcs, T. Cserhti, G. Schmidt, J. Chromatogr. A, 1996, 750, 253-256 105. V. Nmeth-Kiss, E. Forgcs, T. Cserhti, G. Schmidt, J. Pharma. Biomed. Anal., 1996, 14, 997-1001 106. E. Leira, A. Botana and R. Cela, J. Chromatogr., 1996, 724, 67-78 107. M. Moberg, S. Holmstrom, U. Lundstrom, K. Markides, J. Chromatogr. A, 2003, 1020, 91-98 108. D. Barrett, M. Pawula, R. Knaggs, P. Shaw, Chromatographia, 1998, 47, 667-672 109. A. Karlsson, M. Berglin, C. Charron, J. Chromatogr. A, 1998, 797, 75-82 110. L. Monser, F. Darghouth, J. Pharma. Biomed. Anal., 2002, 27, 851-860 111. L. Monser, F. Darghouth, J. Pharmaceut. Biomed., 2003, 32, 1087-1092 112. L. Monser, H. Trabelsi, J. Liq. Chromatogr. Rel. Techn., 2003, 26, 261-271 113. J. Xu, A.-F. Aubry, Chromatographia, 2003, 57, 67-71

Applications Review

114. E. Forgcs, T. Cserhti, J. Pharma. Biomed. Anal., 1998, 18, 15-20 115. T. Nazir, L. Gould, C. Marriott, G. Martin, M. Brown, Chromatographia, 1997, 46, 628-636 116. L. Monser, F. Darghouth, J. Pharmaceut. Biomed., 2000, 23, 353-362 117. A. Holm, K. Solbu, P. Molander, E. Lundanes, T. Greibrokk, Anal. Bioanal. Chem., 2004, 378, 1762-1768 118. E. Svantesson, J. Capala, K. Markides, J. Pettersson, Anal. Chem., 2002, 74, 5358-5363 119. K. Moller, C. Crescenzi, U. Nilsson, Anal. Bioanal. Chem., 2004, 378, 197-204 120. A. Al-Haddad, Talanta, 2003, 59, 845-848 121. K. Echolsa, R. Gale, K. Feltz, J. O'Laughlin, D. Tillitt, T. Schwartz, J. Chromatogr., 1998, 811, 135-144 122. M. Pietrogrande, A. Benvenuti, S. Previato, F. Dondi, Chromatographia, 2000, 52, 425-432 123. S. Chu, C.-S. Hong, B. Rattner, P. McGowan, Anal. Chem., 2003, 75, 1058-1066 124. J. de Boer, R. Law, J. Chromatogr. A, 2003, 1000, 223-251 125. Y. Luo, C. Uboh, L. Soma, F. Guan, J. Rudy, D. Tsang, Rapid Commun. Mass Spectrom., 2005, 19, 825-832 126. O. Van den Hauwe, J. Perez, J. Claereboudt, C. Van Peteghem, Rapid Commun. Mass Spectrom., 2001, 15, 857-861 127. O. Van den Hauwe, F. Dumoulin, J. Antignac, M. Bouche, C. Elliott, C. Van Peteghem, Anal. Chim. Acta, 2002, 473, 127-134 128. O. Van den Hauwe, F. Dumoulin, C. Elliott, C. Van Peteghem, J. Chromatogr. B, 2002, 817, 215-223 129. C. Baiocchi, M. Brussino, M. Pazzi, C. Medana, C. Marini, E. Genta, Chromatographia, 2003, 58, 11-14 130. K. Barnes, R. Fussell, J. Startin, H. Mobbs, R. James, S. Reynolds, Rapid Commun. Mass Spectrom., 1997, 11, 159-164 131. R. Batlle, H. Carlsson, E. Holmgren, A. Colmsj, C. Crescenzi, J. Chromatogr. A, 2002, 963, 73-82 132. C. Snchez, H. Carlsson, A. Colmsj, C. Crescenzi, R. Batlle, Anal. Chem., 2003, 75, 4639-4645 133. R. Batlle, C. Nern, C. Crescenzi, H. Carlsson, Anal. Chem., 2005, 77, 44241-4247 134. E. Holmgren, H. Carlsson, P. Goede, C. Crescenzi, J. Chromatogr. A, 2005, 1099, 127-135 135. C. Crescenzi, J. Albiana, H. Carlsson, E. Holmgren and R. Batlle, J. Chromatogr. A, 2007, 1153, 186-193 136. R. Tachon, V. Pichon, M. Le Borgne, J.-J. Minet, J. Chromatogr. A, 2007, 1154, 174-181

43

Application Chromatograms
Allantoin
H350-1136
100 95 90 85 80 75 70 65 60 Relative Abundance 55 50 45 40 35 30 25 20 15 10 5 0.0 0.5 1.0 1.5 Time (min) 2.0 2.5

Cyclic Monophosphates
H350-1133
mAU 120

L-Carnitine
H350-1132
100

2 4

95 90 85

100

80 75 70

80

Relative Abundance
0 2.5 5 7.5 min

65 60 55 50 45 40 35

60

40

20

30 25 20 15 10 0.0 0.2 0.4 0.6 0.8 1.0 Time (min) 1.2 1.4

Application Chromatograms
44

Column: Part Number: Mobile Phase:

Column: Part Number: Mobile Phase:

Hypercarb, 3 m, 50 x 2.1 mm 35003-052130 A H2O+ 0.1 % Formic acid; B MeCN+ 0.1 % Formic acid. Isocratic: 95% A + 5% B, run for 5 min. Flow Rate: 0.3 mL/min. Detection: UV @ 205 nm; MS: ESI ve, Temperature: 30 C. Injection Volume: 5 L

Hypercarb, 5 m, 100 x 4.6 mm 35005-104630 A = 20 mM Ammonium Acetate (no pH modification) B = MeCN Gradient: 20 45% B in 10 minutes Flow Rate: 1.0 mL/min Detection: UV @ 254 nm Temperature: 25 C Injection Volume: 5 L
1. Cytidine 3', 5' cyclic monophosphate 2. Uridine 3', 5' cyclic monophosphate 3. Guanosine 3', 5' cyclic monophosphate 4. Adenosine 3', 5' cyclic monophosphate

Column:

Hypercarb DASH HTS, 5 m, 20 x 2.1 mm Part Number: 35005-022150 Mobile Phase: H2O + 0.1% TFA Flow Rate: 0.5 mL/min Detection: MS, +ve ESI Temperature: 30 C Injection Volume: 0.5 L
1. L-Carnitnine

RNB-Glycopeptides
07100301

Purines and Pyrimidines (UHT-LC)


H350-1102

Ceramides
H350-1080

Column: Part Number: Mobile Phase: Flow Rate: Detection: Temperature:


1. Cytosine 2. Uracil

Hypercarb, 5 m, 100 x 4.6 mm 35005-104646 H2O 2 mL/min UV at 254 nm 150 to 200 C at 15 C/min; hold at 200 C
3. Thymine 4. Hypoxanthine 5. Guanine 6. Xanthine

Column: Hypercarb, 5 m, 100 x 2.1 mm Part Number: 35005-102130 Mobile Phase: A: MeOH B: CHCl3 Gradient: 45 to 80% B in 15 min Flow Rate: 0.4 mL/min Detection: ELSD Temperature: 50 C Source: K. Gaudin, Laboratoire de Chimie Analytique, Universit Paris Sud, France
Ceramides: 1. d18:1c16:0 2. d18:0c16:0 3. d18:1c18:0 4. d18:0c18:0 5. d18:1c22:1 6. d18:1c20:0 7. d18:1c23:1 8. d18:1c24:1 9. d18:1c22:0 10. d18:1c25:1 11. d18:1c23:0 12. d18:1c26:1 13. d18:1c24:0 14. d18:1c25:0 15. d18:1c26:0

Column: Hypercarb, 5 m, 100 x 4.6 mm Part Number: 35005-104630 Mobile Phase: A: H2O B: ACN Gradient: 10 to 50% B in 50 min Flow Rate: 1 mL/min Detection: UV at 210 nm Temperature: 40 C Source: J. Fan and A. Kondo, Anal. Biochem. 219, 224 (1994). Reproduced with permission
1. R-I (Man6GlcNAc2Asn) 2. R-II (Man5GlcNAc2Asn) 3. R-III (Man6GlcNAc2AsnLeu) 4. R-IV (Man5GlcNAc2AsnLeu)

Aatoxins
aa1
4 3

Nonylphenol Isomers
12020401

Quaternary Ammonium Salts


28070401

1 2

Column: Hypercarb, 5 m, 100 x 3.0 mm Part Number: 35005-103030 Mobile Phase: A: Dioxane B: CHCl3 Isocratic: 10:90 Flow Rate: 0.8 mL/min Detection: Fluorescence (exc 365 nm, em 418 nm) Source: Rhemrev-Boom, M.M, Amro Emmen
1. Aatoxin B1 2. Aatoxin B2 3. Aatoxin G1 4. Aatoxin G2

Column: Hypercarb, 3 m, 100 x 0.32 mm Part Number: 35003-100365 Mobile Phase: A: 0.1% Formic acid B: ACN + 0.1% Formic acid Gradient: 50 to 70% B in 30 min Flow Rate: 6 L/min Detection: UV at 204 nm
Sample: p-Nonylphenol (some of the possible isomer structures represented below)

Methylamines in Fish
H350-1093

Water Pollutants
H350-1131
100 90 80 70 60 50 40 30 20 10 100 90 80 70 60 50 40 30 20 10 100 90 80 70 60 50 40 30 20 10 100 90 80 70 60 50 40 30 20 10 0

Column: Hypercarb, 5 m, 50 x 4.0 mm Part Number: 35005-054030 Mobile Phase: A: H2O + 0.05% TFA B: ACN + 0.05% TFA Gradient: 5 to 35% B in 10 min Flow Rate: 0.8 mL/min Detection: UV at 295 nm to 3 min, 245 nm from 3 to 10 min Temperature: 25 C
1. Diquat

Application Chromatograms

2. Paraquat

1 2

Relative Abundance

Arginine and Methylated Arginines


01030403

5 6 4
1 2 3 4 5 6 Time (min) 7 8 9 10

Column: Part Number: Mobile Phase: Column: Hypercarb, 7 m, 100 x 4.6 mm Part Number: 35007-104630 Mobile Phase: 5 mM Heptanesulfonic acid + 5 mM KH2PO4 at pH 9 Flow Rate: 1 mL/min Source: Lofti I. Monser, Analytica Chemica Acta, 322 (1996) 63-68.
1. Trimethylamine (TMA) 2. Dimethylamine (DMA) 3. Methylamine (MA)

Hypercarb, 5 m, 100 x 2.1 mm 35005-102130 A H2O + 0.1% Formic acid B ACN + 0.1% Formic acid Gradient: 10 to 100% B in 10 mins Flow Rate: 0.2 mL/min Detection: MS, +/- ESI Temperature: 68 C Injection Volume: 10 L
1. Ammeline 2. Ammelide 3. Cyanuric Acid 4. Atrazin desethyl desisopropyl 5. Atrazin desethyl 6. Atrazin desisopropyl

Column: Hypercarb, 3 m, 100 x 2.1 mm Part Number: 35003-102130 Mobile Phase: A: 10 mM NH4COOH at pH 3.5 B: ACN Gradient: 10 to 50% B in 10 min Flow Rate: 150 L/min Detection: + ESI Temperature: 40 C
1. L-arginine 2. Methyl-L-arginine 3. Asymmetrical dimethyl arginine

45

Creatine in Serum
H350-1054

Acyclovir
12020402

Fosfomycin
01030402

Column: Hypercarb, 5 m, 100 x 4.6 mm Part Number: 35005-104630 Mobile Phase: A: ACN B: TFA C: H2O Isocratic: 3:0.1:96.9 Flow Rate: 1 mL/min Detection: UV at 210 nm Source: C. Lim, IRC, Centre for Mechanism of Human Toxicity, Leicester, UK

1. Oxalic acid 2. Creatine 3. Creatinine

Glucosamine Sulfate
01030401

Column: Hypercarb, 3 m, 100 x 2.1 mm Part Number: 35003-102130 Mobile Phase: A: H2O + 0.1% Formic acid B: ACN + 0.1% Formic acid Gradient: 30 to 100% B in 10 min Flow Rate: 0.2 mL/min Detection: + ESI Temperature: 40 C
Acyclovir

Column: Hypercarb, 3 m, 100 x 2.1 mm Part Number: 35003-102130 Mobile Phase: A: 0.1% NH3 (aq) B: ACN Isocratic: 90:10 Flow Rate: 0.15 mL/min Detection: - ESI Temperature: 30 C
1. Fosfomycin (phosphomycin)

Application Chromatograms
46

Tuberculostatics
06010403

Uracil and Metabolite


H350-1087

Column: Hypercarb, 3 m, 100 x 2.1 mm Part Number: 35003-102130 Mobile Phase: A: 0.1% NH3 (aq) B: ACN Isocratic: 50:50 Flow Rate: 0.2 mL/min Detection: - ESI Temperature: 60 C
1. Glucosamine sulfate

Column: Hypercarb, 3 m, 100 x 2.1 mm Part Number: 35003-102130 Mobile Phase: A: 0.05% TFA B: ACN + 0.05% TFA Isocratic: 70:30 Flow Rate: 0.2 mL/min Detection: + ESI Temperature: 40 C
1. Isoniazid 2. Pyrazinamide

Column: Hypercarb, 5 m, 100 x 4.6 mm Part Number: 35005-104630 Mobile Phase: A: H2O B: ACN Gradient: 5 to 50% B in 10 min Flow Rate: 1 mL/min Detection: UV at 210 nm Temperature: 40 C
1. 5,6-Dihydrouracil 2. Uracil 3. 5-Fluorouracil

Ordering Information
Hypercarb Columns
Particle Size Length (mm) 4.6 mm ID 3.0 mm ID 2.1 mm ID 1.0 mm ID

3 m

5 m

7 m

30 50 100 150 30 50 100 150 50 100

35003-034630 35003-054630 35003-104630 35003-154630 35005-034630 35005-054630 35005-104630 35005-154630 35007-054630 35007-104630

35003-033030 35003-053030 35003-103030 35003-153030 35005-033030 35005-053030 35005-103030 35005-153030 35007-053030 35007-103030

35003-032130 35003-052130 35003-102130 35003-152130 35005-032130 35005-052130 35005-102130 35005-152130 inquire inquire

35003-031030 35003-051030 35003-101030 35005-031030 35005-051030 35005-101030 35005-151030

Other column dimensions are also available. Please call Customer Service for more information. For high temperature applications please refer to columns listed below.

Hypercarb Drop-in Guard Cartridges (pk/2)


Particle Size Length (mm) 4.6 mm ID 3.0 mm ID 2.1 mm ID 1.0 mm ID

3 m 10 5 m 10 7 m 10 UNIGUARD Direct-Connect Drop-in Guard Cartridge Holder

35003-014001 35005-014001 35007-014001 850-00

35003-013001 35005-013001 35007-013001 852-00

35003-012101 35005-012101 35007-012101 852-00

35003-011001 35005-011001 35007-011001 851-00

Hypercarb KAPPA Capillary Columns


Particle Size Length (mm) 500 m ID 320 m ID 180 m ID 100 m ID 75 m ID

5 m

50 100

35005-050565 35005-100565

35005-050365 35005-100365

35005-050265 35005-100265

35005-050165 35005-100165

35005-050065 35005-100065

Ordering Information

Hypercarb Nanobore Columns


Particle Size Length (mm) IntegraFrit 75 m ID IntegraFrit Multipack 75 m ID IntegraFrit 150 m ID IntegraFrit Multipack 150 m ID

5 m

10 50

35005-017563 35005-057563
PicoFrit 75 m ID x 15 m Tip

35005-017564 (4/pk) 35005-057564 (3/pk)


PicoFrit Multipack 75 m ID x 15 m Tip

35005-011563 35005-051563

35005-011564 (4/pk) 35005-051564 (3/pk)

5 m

10 50

35005-017581 35005-057581

35005-017583 (4/pk) 35005-057582 (3/pk)

Hypercarb Specialized Column Hardware for High Throughput


Particle Size Quantity DASH HTS 20 x 2.1 mm Javelin HTS 20 x 4.0 mm Javelin HTS 20 x 2.1 mm Javelin HTS 20 x 1.0 mm

5 m

35005-022151

35005-024035

35005-022135

35005-021035

Hypercarb Preparative Columns


Particle Size Length (mm) 10 mm ID 21.2 mm ID 30 mm ID 50 mm ID

5 m

7 m

50 100 150 50 100 150

35005-059070 35005-109070 35005-159070 35007-059070 35007-109070 35007-159070

35005-059270 35005-109270 35005-159270 35007-059270 35007-109270 35007-159270

35005-059370 35005-109370 inquire 35007-059370 35007-109370 35007-159370

35005-059570 35005-109570 inquire 35007-059570 35007-109570 35007-159570

Hypercarb High Temperature Columns


Particle Size Length (mm) 4.6 mm ID 3.0 mm ID 2.1 mm ID 1.0 mm ID

3 m

5 m

30 50 100 30 50 100

35003-034646 35003-054646 35003-104646 35005-034646 35005-054646 35005-104646

35003-033046 35003-053046 35003-103046 35005-033046 35005-053046 35005-103046

35003-032146 35003-052146 35003-102146 35005-032146 35005-052146 35005-102146

35003-031046 35003-051046 35003-101046 35005-031046 35005-051046 35005-101046

Please note that these columns are for use with elevated temperatures. For other dimensions, please inquire.

To download the full text of this application note, visit www.thermo.com/msnotebook

47

Chromatography Resources
Every piece of the solution at your fingertips, precisely when you need it.

In addition to these offices, Thermo Fisher Scientific maintains a network of representative organizations throughout the world.
USA and Canada

Download a copy of our Method Development Guide for Hypercarb Columns.


www.thermo.com/hypercarb

+1 800 532 4752 +1 561 688 8731 fax [email protected]


United Kingdom

+44 (0) 845 702 3964 +44 (0) 1707 391 311 fax [email protected]
France

+33 (0) 1 60 92 48 00 +33 (0) 1 60 92 49 00 fax [email protected]


Germany

+49 (0) 6103 408 1140 +49 (0) 6103 408 1111 fax [email protected]

Our bi-monthly Separated by Experience newsletter keeps you up-to-date on the latest technical and product information of interest to chromatographers. Subscribe today!
www.thermo.com/columns

India

+1 800 22 8374 toll free +91 22 67162200 +91 22 67162244 fax [email protected]
Japan

+81 45 453 9120 +81 45 453 9122 fax [email protected]


Switzerland

+41 56 618 41 11 +41 56 618 41 41 fax [email protected]


All Other Enquiries

+44 (0) 1928 581 000 +44 (0) 1928 581 078 fax [email protected]

Our Chromatography Columns and Consumables catalogs helps you find all your chromatography needs in a single resource. Order your copy now!
www.thermo.com/columns

www.thermo.com/columns

Our Chromatography Resource Centre provides an an extensive, fully searchable library of chromatography applications which currently includes over 6000 applications and references encompassing liquid chromatography, gas chromatography and solid-phase extraction.
www.thermo.com/columns

.
ANGSCHYPERCARB0609

You might also like