Teslameter CYHT201-User's Manual
Teslameter CYHT201-User's Manual
1) Characteristics
Digital Gauss/Tesla Meter - Wide measuring range and high resolution
- A low-cost measuring device, which is easy to operate, portable
CYHT201 -
and convenient to handle and store.
Ideal for quick quality checks and comparative measurements,
with built-in polarity display.
User’s Manual
2) Technical Data
VH = k H I H B
6) Measuring Method
The Gauss/Tesla meter works with a Hall Effect sensor/probe. There are The sensing plan of the Hall
two kinds of Hall probes: transverse and axial probes, which are used for sensor/probe should be
different magnetization directions. positioned perpendicular to the
magnetic field
Electric
Charge
Vertical
Put the concave side, i.e. the side with a small circle mark of the probe on
the surface of the measured magnet or at the measuring point of a
Hall Effect and its Electrical Parameters
3 4
magnetic field carefully. Front Panel
Convex
Measurement of
surface flux density
of a magnet
DC Magnetic field
AC Magnetic field
Measurement of AC
magnetic field
generated by AC
current
CYHT201
CYHT201 Digital
7) Part and Functions Gauss/Tesla Meter
5 6
Back Panel
Hall sensor/probes
Protective tube
Concave
Transverse probe
Power
Socket
Protective tube
Concave
Axial probe
Battery
Compartment
1) Install the +9V battery to the battery compartment or connect the +9V
power supply to the measuring instrument
2) Connect the Hall sensor/probe to the measuring instrument
3) Switch on the power of measuring instrument by pressing button (5), Unit: mT Unit: Gs
LCD display shows 0.0mT
7) Reset Gaussmeter. Keep the Hall probe faraway from magnetic
field, LCD display shows zero after pressing button (7) :
0-2000mT 0-200mT
5) Select AC/DC Mode by using the button (4), the display shows the
following DC or AC measuring mode:
9 10
9) Warranty
9) Take off the protective tube of the Hall sensor/probe, position the 10) Service
measuring circle mark (10) on the surface of the measuring object
(such as permanent magnet), and read the display value (measuring Please contact us for technical questions, repairing and
value and pole display “N” or “S“).
replacement etc:
Note: the measuring circle mark (10) is measuring point of the Hall
sensor. The pole is N pole if the display is “N” when this mark orients
to the measuring object. ChenYang Technologies GmbH & Co. KG
10) Reset peak hold mode. Under the peak holding mode the display Markt Schwabener Str. 8
changes only when the actual measuring value is larger than the last 85464 Finsing, Germany
peak value (displayed). Therefore you must reset the peak holding Tel. +49-(0)8121-2574100
mode by using button (7) if you need to measure smaller magnetic
Fax: +49-(0)8121-2574101
field than the peak value. The device works still in peak holding mode
after resetting.
Email: [email protected]
http://www.chenyang-gmbh.com
Sample: last peak value: 1500mT, actual peak value 800mT after
reset the peak holding mode
11) After the measurement please put the protective tube on the Hall
probe in order to protect the Hall element and switch off the power
12) The Hall probes delivered are compatible. The measuring accuracy
under using other changed Hall probes is ±2%.
11 12