Mini MP 5200 Gas Analyzer Operator Manual
Mini MP 5200 Gas Analyzer Operator Manual
Revision: 8
Language: UK English
Operator Manual
SERVOFLEX
MiniMP (5200)
Gas Analyser
PROCESS ANALYSERS
This page intentionally blank.
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8
i
1 DESCRIPTION AND DEFINITIONS. . . . . . . . . . . . . . . . . . . . . . . . 1
1.1 Scope of this manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 Safety information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.3 Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.4 Sample measurement configurations. . . . . . . . . . . . . . . . . . . . . . . 2
1.5 Other product options. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2 SPECIFICATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.1 General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.2 Sample gas. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.3 Calibration gases . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.4 Environmental limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.5 Performance: standard oxygen sensor. . . . . . . . . . . . . . . . . . . . . . 9
2.6 Performance: industrial oxygen sensor
(standard Multi Purpose only). . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.7 Performance: high accuracy oxygen sensor. . . . . . . . . . . . . . . . . . 11
2.8 Performance: IR (infrared) sensors . . . . . . . . . . . . . . . . . . . . . . . . 12
2.9 Rechargeable battery (optional feature). . . . . . . . . . . . . . . . . . . . . 13
2.10 milliAmp outputs (optional feature). . . . . . . . . . . . . . . . . . . . . . . . . 13
3 UNPACK THE MULTI PURPOSE . . . . . . . . . . . . . . . . . . . . . . . . . 14
4 MULTI PURPOSE USER INTERFACE . . . . . . . . . . . . . . . . . . . . . 15
4.1 Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.2 Start-up and measurement screens . . . . . . . . . . . . . . . . . . . . . . . . 15
4.3 Soft key legends. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.4 Status icon bar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
4.5 Scroll bars. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
4.6 Menu options/screens and password protection . . . . . . . . . . . . . . 19
4.7 The Menu screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
4.8 The Settings screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
4.9 The Information screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
4.10 Editing on-screen data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
5 INSTALLATION AND SET-UP . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
5.1 Installation and switch-on. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
5.2 Charging/recharging the battery (Multi Purpose with
optional rechargeable battery) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
5.2.1 Charging. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
5.2.2 Recharging. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
5.3 Multi Purpose set-up. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
5.3.1 Selecting the security level and changing the password(s) . . . . . . 27
5.3.2 Setting the clock. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
5.3.3 Changing regional settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
5.3.4 Selecting power save mode (Multi Purpose with
optional rechargeable battery) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
LIST OF CONTENTS
Section Page
I
p
s
i
t
e
c
h
S
M
/
0
3
3
-
2
0
0
8
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8 ii
5.3.5 Selecting pump operation (Multi Purpose HF or standard
Multi Purpose with optional internal sample pump) . . . . . . . . . . . . 33
6 GENERAL OPERATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
6.1 Calibrating the Multi Purpose. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
6.2 Taking sample readings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
6.3 Correcting oxygen measurement for different background gases. 39
6.3.1 Overview of measurement errors . . . . . . . . . . . . . . . . . . . . . . . . . 39
6.3.2 Entering a cross-interference compensation. . . . . . . . . . . . . . . . . 39
6.4 Selecting display units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
6.5 Configuring the measurement alarms . . . . . . . . . . . . . . . . . . . . . . 43
6.5.1 Alarm modes and levels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
6.5.2 Latching/non-latching alarms. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
6.5.3 Hysteresis levels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
6.5.4 Setting the measurement alarm levels and modes . . . . . . . . . . . . 45
6.5.5 Enabling/disabling the audible measurement alarm . . . . . . . . . . . 46
6.5.6 Silencing (muting) the audible measurement alarm . . . . . . . . . . . 46
6.5.7 Unlatching measurement alarms. . . . . . . . . . . . . . . . . . . . . . . . . . 47
6.5.8 Viewing the measurement alarm status. . . . . . . . . . . . . . . . . . . . . 47
6.6 Configuring and using the milliAmp outputs (optional feature) . . . 48
6.6.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
6.6.2 Introduction to the milliAmp output parameters . . . . . . . . . . . . . . . 49
6.6.3 Set up the milliAmp output parameters . . . . . . . . . . . . . . . . . . . . . 50
6.6.4 Select the Range associated with a measurement . . . . . . . . . . . . 51
6.6.5 Calibrate a milliAmp output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
6.6.6 Check a milliAmp output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
6.7 Data logging, serial outputs and printed outputs . . . . . . . . . . . . . . 54
6.7.1 Selecting data logging/serial output/printed outputs . . . . . . . . . . . 54
6.7.2 Configuring the serial output parameters. . . . . . . . . . . . . . . . . . . . 55
6.7.3 Introduction to data logging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
6.7.4 Entering measurement data into the data log . . . . . . . . . . . . . . . . 56
6.7.5 Starting a new data log batch . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
6.7.6 Outputting the data log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
6.7.7 Viewing the data log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
6.7.8 Clearing the data log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
6.7.9 Printing a sample measurement report . . . . . . . . . . . . . . . . . . . . . 59
6.8 Adjusting the display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
6.8.1 Adjusting the backlight timer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
6.8.2 Adjusting the contrast . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
6.9 Switching off the Multi Purpose after use . . . . . . . . . . . . . . . . . . . 60
7 ROUTINE MAINTENANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
7.1 Cleaning the Multi Purpose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
7.2 Inspecting the inlet filter element. . . . . . . . . . . . . . . . . . . . . . . . . . 61
7.3 Regenerate/replace the drying agent . . . . . . . . . . . . . . . . . . . . . . 62
7.4 Use of the Multi Purpose for carbon monoxide measurements. . . 63
7.5 Preventative maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
CONTENTS (CONTINUED)
Section Page
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8
iii
8 FAULT FINDING. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
8.1 Introduction to faults and fault messages. . . . . . . . . . . . . . . . . . . . 65
8.2 Viewing fault messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
8.3 General fault finding . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
9 STORAGE AND DISPOSAL. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
9.1 Storage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
9.2 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
10 SPARES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
APPENDICES
A1 DATA LOG OUTPUT FORMATS . . . . . . . . . . . . . . . . . . . . . . . . . . 76
A2 SERIAL OUTPUT FORMATS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
A3 PRINTER OUTPUT FORMATS . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
A4 RS232 CONNECTION DETAILS . . . . . . . . . . . . . . . . . . . . . . . . . . 81
A4.1 OvervieW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
A4.2 Connecting the Multi Purpose to a PC . . . . . . . . . . . . . . . . . . . . . . 81
A4.3 Capturing data using Windowsand Hyper Terminal . . . . . . . . 82
A5 DISPLAY UNIT CONVERSION FACTORS . . . . . . . . . . . . . . . . . . 83
A6 OPTIONAL FLOWMETERS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
A6.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
A6.2 Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
A6.3 Preparing the Multi Purpose with a flowmeter . . . . . . . . . . . . . . . . 84
A6.4 Using the Multi Purpose with a flowmeter . . . . . . . . . . . . . . . . . . . 86
A6.5 Multi Purpose HF and standard Multi Purpose
with a sample pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
A6.6 Standard Multi Purpose without a sample pump . . . . . . . . . . . . . . 86
A7 OPTIONAL SAMPLE CONDITIONING PANEL . . . . . . . . . . . . . . . 87
A7.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
A7.2 Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
A7.3 Using the Multi Purpose with the sample conditioning panel . . . . . 88
A7.4 Calibrating . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
A7.5 Sampling gases . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
A7.6 Additional maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
A7.7 Draining the catchpot . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
A7.8 Replacing/regenerating the drying agent . . . . . . . . . . . . . . . . . . . . 90
A7.9 Additional spares . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
CONTENTS (CONTINUED)
Section Page
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8 iv
A8 OPTIONAL GAS PROBE ACCESSORY . . . . . . . . . . . . . . . . . . . 92
A8.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
A8.2 Preparing for use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
A8.3 Using the probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
A8.4 Additional maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
A8.5 Additional spares . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
A9 MATERIALS IN CONTACT WITH SAMPLE AND
CALIBRATION GASES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
A10 DISPOSAL IN ACCORDANCE WITH THE WASTE
ELECTRICAL AND ELECTRONIC EQUIPMENT
(WEEE) DIRECTIVE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
A11 COMPLIANCE AND STANDARDS INFORMATION . . . . . . . . . . 97
Drierite is a registered trademark of W A Hammond Co.
Krytox and Viton are registered trademarks of Dupont.
Kynar is a registered trademark of Elf Atochem North America, Inc.
Windows and Excel are registered trademarks of Microsoft Corporation.
HyperTerminal is a trademark of Hilgraeve Inc.
This manual is copyright, and no part of it may be reproduced without Servomexs written approval.
CONTENTS (CONTINUED)
Section Page
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8
1
1 DESCRIPTION AND DEFINITIONS
1.1 Scope of this manual
This manual provides installation, operation and routine maintenance instructions for
the Servomex 5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers,
abbreviated to "Multi Purpose" and "Multi Purpose HF" where applicable in the
remainder of this manual. Note that, unless otherwise stated, "Multi Purpose" is used
to mean both versions of analyser in the remainder of this manual.
1.2 Safety information
Read this manual and ensure that you fully understand its content before you attempt
to install, use or maintain the Multi Purpose. Important safety information is highlighted
in this manual as WARNINGs and CAUTIONs, which are used as follows:
This manual also incorporates Be aware of information, which is used as follows:
1.3 Description
The Multi Purpose is a lightweight gas analyser, suitable for the needs of field and
laboratory analysis, and light industrial users who require fast, accurate and reliable
gas analysis.
The Multi Purpose uses paramagnetic transducers to determine the oxygen content of
gas samples in concentrations up to 100%, and uses infrared transducers to
determine the carbon dioxide (CO
2
) or carbon monoxide (CO) content of gas samples
in concentrations of up to 100% CO
2
or 5% CO.
WARNING
Warnings highlight specific hazards which, if not taken into account,
may result in personal injury or death.
CAUTION
Cautions highlight hazards which, if not taken into account, can result in
damage to the Multi Purpose or to other equipment or property.
(not applicable)
2-measurement
Multi Purpose
Oxygen * IR
IR
IR
to 5000
meters
(-1640
to 16400
feet)
Ingress protection IP40
* Storage below 21
o
C (70
o
F) is recommended to ensure optimum battery life.
Below sea level.
Above sea level.
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8
9
2.5 Performance: standard oxygen sensor
The display indication given below is the default indication. You can
configure the Multi Purpose to provide other display indications (see
Section 6.4).
Display indication Measured volume % oxygen
Full Scale Range 0 to 100% oxygen
Resolution 0.1% oxygen
Linearity 0.1% oxygen
Intrinsic error (accuracy) 0.1% oxygen *, or 0.2% oxygen
Zero drift per week 0.4% oxygen
Output fluctuation 0.1% oxygen
Response time: standard Multi Purpose
#
Without drying tube
With drying tube fitted
Less than 10 seconds
20 seconds
Response time: Multi Purpose HF
#
Without drying tube
With drying tube fitted
8 seconds
10 seconds
Flow effect
0.4% maximum
* High calibration with certified oxygen supply (see Section 2.3).
High calibration with instrument air or other oxygen supplies (see Section 2.3).
# T
90
at 68.9 kPa gauge (10 psig, 0.69 bar gauge) supply pressure. Response
time will increase if the Multi Purpose has a sample conditioning panel: refer
to Section A7.2.
Within sample gas supply pressure range specified in Section 2.2.
A 1% change in ambient barometric pressure will result in a 1% change
in sample reading.
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8 10
2.6 Performance: industrial oxygen sensor (standard Multi Purpose only)
The display indication given below is the default indication. You can
configure the Multi Purpose to provide other display indications (see
Section 6.4).
Display indication Measured volume % oxygen
Full Scale Range 0 to 100% oxygen
Resolution 0.1% oxygen
Linearity 0.1% oxygen
Intrinsic error (accuracy) 0.1% oxygen
Zero drift per week 0.2% oxygen
Output fluctuation 0.1% oxygen
Response time
*
Without drying tube
With drying tube fitted
15 seconds
25 seconds
Flow effect
0.1% oxygen
Zero temperature coefficient 0.2% oxygen per 10
o
C (18
o
F)
Span temperature coefficient 0.3% oxygen per 10
o
C (18
o
F)
Tilt effects 0.15% oxygen per 15
o
of tilt
Pressure effects Directly proportional to ambient barometric
pressure
#
* T
90
at 68.9 kPa gauge (10 psig, 0.69 bar gauge) supply pressure. Response
time will increase if the Multi Purpose has a sample conditioning panel: refer
to Section A7.2.
Within sample gas supply pressure range specified in Section 2.2.
# A 1% change in ambient barometric pressure will result in a 1% change
in sample reading.
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8
11
2.7 Performance: high accuracy oxygen sensor
The display indication given below is the default indication. You can
configure the Multi Purpose to provide other display indications (see
Section 6.4).
Display indication Measured volume % oxygen
Full Scale Range 0 to 100% oxygen
Resolution 0.01% oxygen
Linearity 0.01% oxygen
Intrinsic error (accuracy) 0.02% oxygen *, or 0.05% oxygen
Zero drift per week 0.2% oxygen
Output fluctuation 0.01% oxygen
Response time: standard Multi Purpose
#
Without drying tube
With drying tube fitted
15 seconds
25 seconds
Response time: Multi Purpose HF
#
Without drying tube
With drying tube fitted
Less than 10 seconds
Less than 12 seconds
Flow effect
0.1% oxygen
Zero temperature coefficient 0.2% oxygen per 10
o
C (18
o
F)
Span temperature coefficient 0.3% oxygen per 10
o
C (18
o
F)
Tilt effects 0.15% oxygen per 15
o
of tilt
Pressure effects Directly proportional to ambient barometric
pressure
The display indications given below are the default indications. You can
configure the Multi Purpose to provide other indications (see Section 6.4).
Allow 1 hour warm up to reach stated accuracy.
Display indication
Carbon monoxide sensors Measured volume % carbon monoxide
Carbon dioxide sensors Measured volume % or ppm carbon dioxide
Full Scale Range *
Carbon monoxide sensors 0 to 5%
Carbon dioxide sensors 0 to 10%, 0 to 25%, 0 to 50% or 0 to 100%
Resolution Less than 0.1% Full Scale Range *
Linearity 1% Full Scale Range *
Intrinsic error (accuracy) 2% Full Scale Range *
Zero drift per week 4% Full Scale Range *
Output fluctuation 0.05% Full Scale Range *
Response time: standard Multi Purpose
Less than 4 mA
Cable requirements
Type Multi-strand twisted pair with overall screen
Maximum size 1.5 mm
2
16 AWG
* User selectable: see Sections 6.6.2 and 6.6.3.
Only available when the 4 to 20 mA output range is selected:
see Sections 6.6.2 and 6.6.3.
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8 14
3 UNPACK THE MULTI PURPOSE
1. Remove the Multi Purpose and any other equipment from its packaging.
2. Remove the protective plastic cover from the sample gas inlet on the front of the
Multi Purpose (see Figure 1).
3. Remove the protective plastic cover from the sample gas outlet (on the front or
the rear of the Multi Purpose: see Figures 1 and 2).
4. Remove the protective plastic cover from the bypass gas outlet (if fitted) on the
rear of the Multi Purpose (see Figure 2).
5. Inspect the Multi Purpose and the other items supplied, and check that they are
not damaged. If any item is damaged, immediately contact Servomex or your
local Servomex agent.
6. If you do not intend to use the Multi Purpose immediately:
Refit the protective plastic covers to the gas inlet, the sample gas outlet and
the bypass gas outlet (if fitted).
Place the Multi Purpose and any other equipment supplied back in its
protective packaging.
Store the Multi Purpose as described in Section 9.1.
Otherwise, read Section 4 (User Interface), then continue at Sections 5 onwards
to install, set up, and use the Multi Purpose.
Retain the shipping documentation and packaging for future use (for
example, return of the Multi Purpose to Servomex for servicing or repair).
CAUTION
You must remove the protective plastic covers as specified in Steps 3 and 4
above before you use the Multi Purpose. If you do not, you may damage the
Multi Purpose when you try to pass calibration or sample gases through it.
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8
15
4 MULTI PURPOSE USER INTERFACE
4.1 Introduction
The Multi Purpose user interface comprises the following (shown on Figure 1):
The Multi Purpose also has an audible alarm which will go on (emit a tone):
On initial switch-on: see Section 5.1.
When a measurement alarm condition is detected (if the audible measurement
alarm is enabled): see Section 6.5.5.
When a fault condition is detected: see Section 8.
4.2 Start-up and measurement screens
When you first switch on the Multi Purpose, a start-up screen is displayed while the
Multi Purpose carries out a self-test.
The start-up screen shows the Servomex name, a self-test time elapsed/remaining
indicator, and messages identifying the tasks being carried out as part of the self-test:
The screen will initially display the message "System Check".
If your Multi Purpose is configured for use with one or more IR sensors, the
following messages will be displayed: "Infrared Initialising" and "Infrared Warming".
When the battery less than 10% full icon starts to flash, this indicates that
the rechargeable battery is virtually empty. The Multi Purpose will
automatically shut down approximately 15 seconds after the icon starts to
flash.
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8
19
4.5 Scroll bars
On some screens (for example, see Figure 5), there may be more options available
than can be shown on the screen, and you have to scroll down the screen to view all
of the options: this is identified by a scroll bar at the right-hand side of the screen.
The height of the wide part of the scroll bar gives an indication of what proportion (of
all the options) are currently shown on the screen. As you scroll up or down the options
(using the and soft keys), the wide part of the scroll bar will also move on
the screen, indicating approximately where the currently displayed options are, within
the complete list of options. For example, compare the scroll bars in Figures 5 and 13.
4.6 Menu options/screens and password protection
The menu structure of the Multi Purpose is shown in Figure 4, which shows that some
of the options/screens are password protected.
When an option/screen is password protected, this means that the correct
corresponding password has to be entered before the option/screen can be accessed.
Password protection operates as follows:
The first time you try to access a password-protected option/screen, you will be
prompted for the corresponding password. You must then enter the correct
password (using the editing method described in Section 4.10) before the option/
screen can be displayed.
If you have already entered the corresponding password, you will gain access to all
options/screens protected by that password immediately (you do not need to enter
the password again).
Once you have entered a password, it remains active until 10 minutes after
the last soft key is pressed. After this, the password becomes inactive; you
must re-enter the password to access password-protected options/screens
again.
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8 20
Figure 4 - The Multi Purpose menu structure
Data log
View log
Output log
Clear log
Settings
Serial output
Password
Clock
Regional
Backlight
Contrast
Power save
Security
Information
Data log
Set up
Calibrate
Alarm
Settings
Service
Faults
Alarm
Mute
Unlatch
View
Set up
Audible alarm
Calibrate the Multi Purpose [6.1]
Silence the audible measurement alarm [6.5.6]
Unlatch a latched measurement alarm [6.5.7]
View the measurement alarm status [6.5.8]
Set up the measurement alarms [6.5.4]
Enable/disable the audible measurement
alarm [6.5.5]
Add measurement to data log [6.7.4]
Start new data log batch [6.7.5]
Produce printed report [6.7.9] *
* If printed outputs selected [6.7.1]
Set serial output parameters [6.7.2]
Change the password(s) [5.3.1]
Set the date/time [5.3.2]
Change regional settings [5.3.3]
Adjust the backlight timer [6.8.1]
Adjust the display contrast [6.8.2]
Select/deselect power save [5.3.4]
Select the security level [5.3.1]
View system information [4.9]
View the data log [6.7.7]
Output the data log [6.7.6]
Clear the data log [6.7.8]
Set milliAmp parameters [6.6]
Select serial output/data logging/
printed outputs [6.7.1]
Select display units [6.4]
Enter a cross-compensation error [6.3.2]
Select pump operation mode [5.3.5]
View current faults [8.2]
(Start-up screen)
[4.2]
(Measurement screen) [4.2]
Specifies that the option/screen can be
protected by the supervisor password.
Specifies the corresponding section in
this manual.
[X.Y.Z]
NOTES:
S
S
A
S
Set up
mA output
Serial type
Unit selection
X-Interference
Pump
S
S
Check milliAmp outputs [6.6.6]
Calibrate milliAmp outputs [6.6.5]
S
S
S
Specifies that the option/screen can be
protected by the operator password.
S
A
O
O
Specifies that the option/screen is always
protected by the supervisor password.
A
O
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8
21
4.7 The Menu screen
The Menu screen (see Figure 5) provides access to other screens in the menu
structure, and is displayed by pressing the soft key when the Measurement
screen is displayed.
Figure 5 - The Menu screen
Use the and soft keys to highlight the required screen option, then press
the soft key to display the selected screen:
Alternatively, press the soft key to display the Measurement screen again.
Some of the menu screens referenced below may not be available: refer to
the note at the start of Section 4.
Screen Use Section
Data Log Select this screen to view, output or clear the data log. 6.7.6 to
6.7.8
Set up Select this screen to configure the (optional) milliAmp
outputs, select serial output, data logging or printed
outputs, select the display units, or introduce a cross-
interference correction.
6.6.3,
6.7.1,
6.4,
6.3.2
Calibrate Select this screen to calibrate the Multi Purpose. 6.1
Alarm Select this screen to set up the measurement alarms, or
to silence (mute) the audible measurement alarm.
6.5.4
Settings Select this screen to change Multi Purpose settings
(password, display language and so on).
4.8
Service Select this screen to calibrate or check the (optional)
milliAmp outputs.
6.6.5,
6.6.6
Faults Select this screen to view current faults. 8.2
Scroll bar: see
Section 4.5
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8 22
4.8 The Settings screen
Alternatively, press the soft key to display the Menu screen again.
4.9 The Information screen
Note that the information shown on the screen will vary, depending on the Multi
Purpose model.
After viewing (and if necessary recording) the information shown on the screen, press
the soft key to display the Settings screen again, or press and hold the soft
key to show the Measurement screen again.
The Settings screen is shown in
Figure 6.
Use the and soft keys to
highlight the required screen option,
then press the soft key to display
the selected screen, as shown below:
Figure 6 - The Settings screen
Screen Use Section
Serial output Configuring the serial output parameters 6.7.2
Password Changing the password. 5.3.2
Clock Setting the clock time and/or date. 5.3.2
Regional Changing regional settings (language and so on). 5.3.3
Backlight Adjusting the backlight timer duration. 6.8.1
Contrast Adjusting the contrast of the screen. 6.8.2
Power save * Selecting/deselecting power save operation. 5.3.4
Security Selecting the security level. 5.3.1
Information Viewing Multi Purpose system information. 4.9
* Only available on a Multi Purpose with the optional rechargeable battery fitted.
A typical Information screen is shown
in Figure 7.
This screen shows information (such
as the Multi Purpose serial number
and the version of the operating
software embedded in the Multi
Purpose) which is useful to the
Servomex support team.
Figure 7 - Typical information screen
You may be asked to provide the information from this screen to the
Servomex support team; for example, as an aid to fault diagnosis.
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8
23
4.10 Editing on-screen data
A common method is used for editing data shown on all of the different screens.
When you press the soft key to edit an item of data, the screen changes to show
the corresponding edit screen, with the first digit highlighted; a typical edit screen is
shown in Figure 8:
Figure 8 - A typical edit screen
When the first digit is highlighted, press the soft key to exit the menu without
changing the data.
Alternatively, use the soft keys to edit the data as follows:
Note that the figures above and below the highlighted digit show the digits above and
below the currently highlighted value.
When the last digit is highlighted, press the soft key to enter the new data.
Soft key Function
Increases the highlighted digit by 1.
Decreases the highlighted digit by 1.
Moves the cursor left to the previous digit.
Moves the cursor right to the next digit.
When editing numerical values, the decimal point appears between digits
"9" and "0".
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8 24
5 INSTALLATION AND SET-UP
5.1 Installation and switch-on
1. Place the Multi Purpose in a suitable operating location, within easy reach of a
suitable electrical supply outlet.
2. If necessary (if sample or calibration gases are toxic or asphyxiant) or if
required:
Use quick-connect fittings to connect a suitable sized tube to the sample gas
outlet (on the rear or the front of the Multi Purpose: see Figures 1 and 2).
Use quick-connect fittings to connect a suitable sized tube to the bypass gas
outlet (if fitted, on the rear of the Multi Purpose: see Figure 2).
WARNING
Ensure that the cables and tubes that you connect to the Multi Purpose
are routed so that they do not present a trip hazard to people.
WARNING
Ensure that the electrical installation of the Multi Purpose and the power
supply unit conforms with all applicable local and national electrical
safety requirements.
WARNING
Sample and calibration gases may be toxic or asphyxiant. Ensure that
the external connections are leak free at full operating pressure before
you use sample or calibration gases.
WARNING
Sample and calibration gases may be toxic or asphyxiant. Ensure that
the Multi Purpose sample, bypass gas and calibration T piece (if used)
outlets are vented to an area where they will not be a hazard to people.
WARNING
Sample and calibration gases may be toxic or asphyxiant. To prevent the
build-up of such gases, ensure that the Multi Purpose is only used in a
well-ventilated environment.
CAUTION
Do not use the Multi Purpose in an area subject to high levels of vibration
or sudden jolts. If you do, sample measurements may not be accurate,
or the Multi Purpose may be damaged.
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8
25
3. If you have fitted tubes to the sample gas outlet and/or bypass gas outlet, route
the ends of the tubes so that they can freely vent to atmosphere.
4. If you have ordered and received a printer, connect the printer to the serial
connector on the rear of the Multi Purpose (see Figure 2). Alternatively, if
required, connect a PC (personal computer) or other device to the serial
connector: refer to Appendix A4.
5. If your Multi Purpose is configured to provide optional milliAmp outputs, for each
output:
Connect the wires in your cable to the screw terminals on the milliAmp
interface connector supplied: refer to Section 2.10 for the cable requirements,
and refer to Figure 9 below for the connection requirements.
Fit the interface connector to the corresponding milliAmp output connector on
the rear of the Multi Purpose (see Figure 2), and secure with the two captive
screws on the interface connector.
Figure 9 - milliAmp interface connector
6. Fit the power outlet on the power supply unit to the power socket on the rear of
the Multi Purpose.
7. Fit the power supply unit plug to a suitable electrical supply outlet.
8. Press and hold the Power On/Off key on the front of the Multi Purpose for at least
2 seconds to switch the Multi Purpose on.
The two outlets can be left to vent to local atmosphere. However if you
do fit a tube to one or both of the outlets, the tube(s) must be suitably
sized so that the gases can vent from the Multi Purpose without over-
pressurisation of the Multi Purpose or the tubes.
If you have two milliAmp outputs, ensure that you fit the correct
interface connector to the corresponding milliAmp connector on the
rear of the Multi Purpose.
Pin Use
1 +ve
2 -ve
3 screen
When the Multi Purpose is switched on, the Alarm LED, the Fault LED
and the audible alarm will all go on for 1 second to demonstrate that
they are functioning correctly, and will then go off again.
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8 26
5.2 Charging/recharging the battery (Multi Purpose with optional rechargeable
battery)
5.2.1 Charging
The first time you use a Multi Purpose with the optional rechargeable battery, you
should leave the Multi Purpose connected to the electrical supply for at least 4 hours,
to fully charge the battery.
When the battery is fully charged, you can leave the Multi Purpose connected to the
electrical supply, or you can disconnect the electrical supply and continue to use the
Multi Purpose powered by the battery.
5.2.2 Recharging
During normal use, the battery icon on the status icon bar of the display will identify
the level of charge within the battery (see Section 4.4).
You can recharge the battery as and when required during normal use. To recharge
the battery, simply connect the Multi Purpose to an external electrical supply outlet.
We recommend that you recharge the battery as soon as possible after the
battery less than 10% full icon is displayed.
During recharging, the status icon bar will continually show the battery less
than 10% full, battery 10 to 32% full, battery 33 to 65% full and battery 66
to 100% full icons in sequence.
You can recharge the battery with the Multi Purpose switched on or off.
However, recharging will take longer when the Multi Purpose is switched on.
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8
27
5.3 Multi Purpose set-up
When you switch on the Multi Purpose, a start-up screen is first displayed (see
Section 4.2), then the Measurement screen (Figure 3) is displayed.
When the Measurement screen is displayed, you can set up the Multi Purpose as
described below.
5.3.1 Selecting the security level and changing the password(s)
Introduction to security levels/passwords
You can configure the Multi Purpose to provide any of three levels of security:
Figure 4 shows the options/screens which can be password-protected within the menu
structure.
Security level Function
Low None of the options/screens are password protected *.
Standard Some of the options/screens are protected by a
supervisor password.
High Some of the options/screens are protected by a
supervisor password and some of the options/screens
are protected by an operator password
.
* Except for the change the password(s) and select the security level
options/screens: see notes below.
The supervisor password can also be used to access options/screens
protected by the operator password: see notes below.
The change the password(s) and select the security level screens/options
are always protected by the supervisor password, regardless of the security
level selected. This is to ensure that unauthorised personnel cannot change
the security level and password(s) and so lock out the Multi Purpose from
other users.
As supplied, the security level is set to high, the supervisor password is set
to "2000" and the operator password is set to "1000". We recommend that
you select your required security level and change the password(s) as
described below to provide additional protection.
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8
29
Changing passwords
Use the following procedure to change the supervisor and operator passwords:
1. With the Measurement screen displayed, press the soft key to display the
Menu screen, use the and soft keys to highlight the "Settings" menu
option, then press the soft key. The Settings screen will then be displayed
(see Figure 6).
2. Use the and soft keys to highlight the "Password" menu option, then
press the soft key. The Edit supervisor password screen will then be
displayed with the supervisor password shown, as shown in Figure 12.
Figure 12 - The Edit supervisor password screen
3. To change the supervisor password, press the soft key, then enter the new
password: use the editing method described in Section 4.10.
4. When you enter the last digit, the soft key changes to the soft key.
Press the soft key to enter the new supervisor password value.
5. To change the operator password, press the soft key to display the edit
operator password screen, press the soft key, then enter the new
password: use the editing method described in Section 4.10.
6. When you enter the last digit, the soft key changes to the soft key.
Press the soft key to enter the new operator password value.
7. Press the soft key to display the Settings screen again.
If you change a password, ensure that you record the new password
somewhere safe. Otherwise, if you cannot recall the new password, you will
have to contact Servomex or your local Servomex agent for assistance.
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8 30
5.3.2 Setting the clock
Use the following procedure to set the date and time:
1. Press the soft key to display the Menu screen, use the and soft
keys to highlight the "Settings" menu option, then press the soft key. The
Settings screen will then be displayed.
2. Use the and soft keys to highlight the "Clock" menu option, then
press the soft key. The Clock (time) screen will then be displayed, as
shown in Figure 13.
3. Press the soft key, then edit the displayed time as described in
Section 4.10. When you change the last digit, the soft key changes to the
soft key. Press the soft key to show the Clock (time) screen again.
4. Press the soft key to show the Clock (date) screen, as shown in Figure 14.
5. To change the date, press the soft key, then edit the displayed date as
described in Section 4.10. When you change the last digit, the soft key
changes to the soft key. Press the soft key to show the Clock (date)
screen again.
6. Press the soft key twice to display the Menus screen.
The pressure of your calibration gas supply must be the same as the
pressure of the gases to be sampled. If the pressures are different, sample
gas measurements may not be accurate.
For a Multi Purpose HF, the calibration gas flow rate should be as close as
possible to the flow rate of sample gases.
If you do not allow calibration gas to pass through the Multi Purpose for 3 to
5 minutes before you start the calibration procedure, the measurement
system in the Multi Purpose may not be fully purged of other residual gases,
and the calibration may not be accurate.
Do not knock or move the Multi Purpose during calibration. If you do, the
calibration measurements may be affected.
For a standard Multi Purpose with a sample pump, the following calibration
procedure assumes that you have selected manual pump operation. If you
have selected timed pump operation, you must ensure that the pump
operation time is set correctly to allow calibration gas to pass through the
Multi Purpose for sufficient time: refer to Section 5.3.5 for more information.
On a 2-measurement
Multi Purpose, this
screen will show one
Lo and one Hi
calibration for each of
the two gases.
Figure 20 - The Calibrate screen
9999d specifying days 9999h specifying hours
9999m specifying minutes Any combination of these.
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8
37
5. Press the soft key. The Calibrate target value screen will then be shown
(see Figure 21), identifying the target value and the current reading.
6. If the target value is not that for the calibration gas which you are using, change
the target value to the required value: use the edit method in Section 4.10.
7. When the current reading is stable, press the soft key. The Multi Purpose
will then carry out the specified calibration.
8. If you have a standard Multi Purpose with a sample pump, switch the pump off
(if necessary: see Section 5.3.5).
9. Disconnect the calibration gas supply from the sample gas inlet or the calibration
T piece.
10. Repeat Steps 1 to 9 of this section for the second calibration for the specific
sample gas.
11. If your Multi Purpose is configured to provide two sample gas measurements,
repeat Steps 1 to 10 of this section to carry out the Lo and Hi calibrations for
the second sample gas.
12. On a Multi Purpose with the optional sample pump: disconnect your vent
pipeline (if fitted) from the calibration T piece, then disconnect the T piece from
the sample gas inlet.
13. Press the soft key to display the Measurement screen again.
Figure 21 - The Calibrate target
value screen
Refer to Sections 2.2 and 2.3 for the required pressures, flow rates (if
applicable) and concentrations of the calibration gases.
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8 38
6.2 Taking sample readings
1. If necessary, calibrate the Multi Purpose: see Section 6.1.
2. Ensure that the Measurement screen is displayed: see Section 4.
3. Use the quick-connect fitting supplied to connect the sample gas supply to the
sample gas inlet on the front of the Multi Purpose (see Figure 1).
4. If your Multi Purpose has a sample pump, start the sample pump: see
Section 5.3.5.
5. Wait until the measurement shown on the screen has stabilised, then take note
of the reading.
6. If your Multi Purpose has a sample pump and you have selected manual pump
operation, stop the sample pump: see Section 5.3.5.
7. Turn off the sample gas supply, or disconnect it from the sample gas inlet on the
front of the Multi Purpose.
Repeat Steps 3 to 7 as necessary, for different gas samples to be measured.
Unless your sample gases are known to be dry, you must connect the drying
tube (supplied with the Multi Purpose) to the sample gas inlet on the Multi
Purpose, and then connect your sample gas supply to the drying tube.
When you select display units other than the measurement default units, you
must also enter the units conversion factor: refer to Appendix A5 to
determine the units conversion factor for your specific application.
If you select the "off" option on the units selection screen, the display units
revert to the measurement default units as supplied.
Ensure that the measurement alarm and hysteresis levels are not too close
to the expected sample measurements. (If they are, minor - and acceptable -
variations in your sample gas concentrations will result in spurious alarms.)
If you configure one measurement alarm as low and configure the other
alarm as high, ensure that the high alarm and hysteresis levels are higher
than the low alarm and hysteresis levels. (If you do not, the Multi Purpose
can be permanently in an alarm condition, until you correct the levels.)
1. On the Measurement screen,
press the soft key. The
Alarm option screen will then be
displayed, as shown in
Figure 27.
2. Highlight the "Set up" menu
option, then press the soft
key. The Alarm set up screen will
then be displayed, as shown in
Figure 28.
3. Use the and soft keys
to highlight the required alarm,
then press the soft key.
The Alarm mode screen will then
be displayed, as shown in
Figure 29.
4. If the alarm mode is not the
required mode, press the
soft key, use the and
soft keys to select the required
mode (none, low or high), then
press the soft key.
Figure 27 - The Alarm option screen
Figure 28 - The Alarm set up screen
Figure 29 - The Alarm mode screen
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8 46
6.5.5 Enabling/disabling the audible measurement alarm
1. With the Alarms option screen displayed (see Section 6.5.4), use the and
soft keys to highlight the "Audible alarm" option, then press the soft
key.
2. If the displayed alarm setting is not the required setting, press the soft key.
The Audible alarm option screen will then be displayed: see Figure 30.
Figure 30 - The Audible alarm option screen
3. Use the and soft keys to select the required option ("Yes" or "No"),
then press the soft key.
6.5.6 Silencing (muting) the audible measurement alarm
When the audible alarm is on because of a measurement alarm condition, you can
temporarily silence (mute) the audible alarm, as follows:
1. On the Measurement screen, press the soft key; the Alarm option screen
(Figure 27) will then be displayed.
2. With the "Mute" option highlighted, press the soft key. The audible alarm
will then go off and the Measurement screen will be displayed again.
The audible measurement alarm options are "Yes" (for enable) and "No" (for
disable).
(infrared sensor)
Underrange Any value below 4 mA
#
* See Section 6.6.1.
See Section 2.8.
# Only available if the 4-20 mA output range is selected; this sets the lowest out-
put current during normal operation, and allows negative gas concentrations
to be monitored through the milliAmp outputs. For example, with an under-
range setting of 3.8 mA, the milliAmp outputs can be less than 4 mA (which
indicates a zero gas concentration), down to a minimum of 3.8 mA, where an
output between 3.8 mA and 4 mA indicates a negative gas concentration.
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8 50
6.6.3 Set up the milliAmp output parameters
Use the following procedure to set up the milliAmp output parameters:
1. Press the soft key to display the Menu screen, use the and soft
keys to highlight the "Set up" menu option, then press the soft key. The Set
up screen will then be displayed (see Figure 17).
2. Use the and soft keys to highlight the "mA output" menu option, then
press the soft key: the mA configuration screen is then shown (see
Figure 32).
Figure 32 - The mA configuration screen
3. Use the and soft keys to highlight the required "Set up" option, then
press the soft key: the corresponding mA output high level screen will then
be shown: see Figure 33.
Figure 33 - The mA output high level screen
4. If necessary, change the displayed parameter using the edit method described
in Section 4.10.
5. For each of the other milliAmp parameters (see Section 6.6.2):
Use the and soft keys to select the corresponding parameter
screen.
Change the parameter as necessary: press the soft key then use the
and soft keys to select the required option, or edit the data as
described in Section 4.10.
You cannot enter a high level (span) value higher than the maximum
measurement that the corresponding gas sensor can determine (refer
to Sections 2.5 to 2.8).
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8
51
6.6.4 Select the Range associated with a measurement
Use the following procedure at any time during sampling to change the Range (see
Section 6.6.1) associated with a gas measurement:
1. Press the soft key to display the Menu screen, use the and soft
keys to highlight the "Set up" menu option, then press the soft key. The Set
up screen will then be displayed (see Figure 17).
2. Use the and soft keys to highlight the "mA output" menu option, then
press the soft key: the mA configuration screen is then shown (see
Figure 32).
3. Use the and soft keys to highlight the required gas Range option,
then press the soft key: the mA range screen will then be displayed, as
shown in Figure 34.
Figure 34 - The mA range screen
4. To change the selected Range:
Press the soft key, then use the and soft keys to highlight the
alternative Range option (Range 1 or Range 2).
Press the soft key: the mA Range screen will then be displayed again, with the
new Range shown.
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8 52
6.6.5 Calibrate a milliAmp output
Use the following procedure to calibrate a milliAmp output:
1. Press the soft key to display the Menu screen, use the and soft
keys to highlight the "Service" menu option, then press the soft key. The
mA output service screen will then be displayed: see Figure 36.
2. Select the required Calibrate option, then press the soft key. The mA
output calibrate screen will then be displayed, as shown in Figure 35.
Figure 35 - The mA output calibrate screen
3. As soon as the mA output calibrate screen is shown, the nominal milliAmp output
value is set to 20 mA:
Use your equipment (connected to the corresponding milliAmp connector on
the rear of the Multi Purpose) to monitor the actual output value.
Use the and soft keys to increase or decrease the actual output
value until your equipment indicates 20 mA output.
4. When the milliAmp output has been correctly calibrated, press the soft key:
the mA output service screen (Figure 36) will then be displayed again.
The actual milliAmp output value is controlled from the mA output calibrate
screen as long as the screen is displayed. As soon as the mA service screen
is displayed, the milliAmp output value will be updated to reflect the
corresponding gas measurement.
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8
53
6.6.6 Check a milliAmp output
If required, use the following procedure at any time to perform a check on a milliAmp
output:
1. Press the soft key to display the Menu screen, use the and soft
keys to highlight the "Service" menu option, then press the soft key. The
mA output service screen will then be displayed: see Figure 36.
Figure 36 - The mA output service screen
2. Select the required Override option, then press the soft key. The mA
output override screen will then be displayed, as shown in Figure 37.
Figure 37 - The mA output override screen
3. Edit the displayed override value as described in Section 4.10.
4. Press the soft key: an acceptance screen showing "No" will then be
displayed.
5. To apply the override, press the soft key: an Override screen will then be
shown. Use the and soft keys to highlight the "Yes" option, then press
the soft key.
The milliAmp output will now be set to the override value you have selected. Use
your equipment (connected to the corresponding milliAmp connector on the rear
of the Multi Purpose) to check that the output is correct.
6. If required, repeat Steps 3 to 5 above to check for other milliAmp output values,
or press the soft key to display the mA output service screen again.
The milliAmp output freezes at the override value as long as the Override
screen is displayed. As soon as another screen is displayed, the milliAmp
output value will be updated to reflect the corresponding gas measurement.
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8 54
6.7 Data logging, serial outputs and printed outputs
6.7.1 Selecting data logging/serial output/printed outputs
You can configure the Multi Purpose to provide data logging, to provide serial outputs,
or to provide printed outputs to an optional Servomex supplied printer:
When serial output is selected, data is continuously output to your PC (or other
device) connected to the serial output port on the rear of the Multi Purpose.
When data logging is selected, measurement information is stored in Multi Purpose
memory (see Section 6.7.3). You must manually initiate output of the stored data
log to a connected device, as described in Section 6.7.6.
When printer (printed output) is selected, you must manually initiate the printing of
a sample measurement report (to the optional Servomex supplied printer), as
described in Section 6.7.9.
Use the following procedure to select the required option:
1. With the Menus screen displayed, use the and soft keys to highlight
the "Set up" menu option, then press the soft key. The Serial type screen
will then be displayed, as shown in Figure 38.
Figure 38 - The Serial type screen
2. If the option shown on the Serial type screen is not the required option, press the
soft key; the Serial type select screen will then be displayed (see
Figure 39).
Figure 39 - The Serial type select screen
3. Use the and soft keys to select the required option, then press the
soft key.
4. Press the soft key to display the Menu screen.
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8
55
6.7.2 Configuring the serial output parameters
If you have selected data logging or serial outputs, you must configure the Multi
Purpose serial output parameters to suit the requirements of the PC (or other device)
which you have connected to the serial output port on the rear of the Multi Purpose.
The output parameters which you can configure are listed below:
To configure the parameters:
1. From the Settings screen, use the and soft keys to highlight the
"Serial output" menu option, then press the soft key. The first parameter
screen will then be displayed: if you have selected serial outputs, the RS232
period screen will be shown (see Figure 40); if you have selected data logging,
the RS232 baud rate screen will be shown (see Figure 41).
2. If necessary, change the displayed parameter using the edit method described
in Section 4.10.
3. For each of the other configurable parameters (see above):
Use the and soft keys to select the corresponding parameter
screen.
Change the parameter as necessary: use the and soft keys to
highlight the required option, or edit the data as described in Section 4.10.
The serial output parameters apply to both the data logging and serial output
options (see Section 6.7.1).
Parameter Valid settings
Period * 0 to 999 seconds.
Baud rate 2400, 4800, 9600, 19200 and 38400 are all supported.
Stop bits 1 or 2.
Data bits 7 or 8.
Parity None, odd parity, or even parity.
* Time between updates: not applicable to data logging.
Figure 40 - The RS232 period
screen
Figure 41 - The RS232 baud
rate screen
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8 56
6.7.3 Introduction to data logging
The Multi Purpose allows you to maintain a data log of sample measurements made,
within the Multi Purpose memory. The data log consists of one or more batches of
measurements, and a total of 200 measurements can be stored.
Each sample measurement entered in the log has:
an associated batch number
a sequence number of the sample measurement within the batch
the date and time that the measurement was made
an alarm indication (if the measurement caused a measurement alarm to be raised)
a fault indication (if a fault existed at the time of the measurement).
You must manually transfer measurements into the data log, specify when a new
batch is to start within the data log, and clear the data log when necessary.
At any time, you can view the currently stored data log on the Multi Purpose display,
or you can output it to a device (such as a PC) connected to the Multi Purpose serial
output port.
6.7.4 Entering measurement data into the data log
When you want to enter the sample measurement(s) from the Measurement screen
into the data log:
1. Press the soft key. When you press the soft key, the current sample
measurement at the time of the key press is captured (stored in internal
memory), and the Data logging screen will then be displayed (see Figure 42).
Figure 42 - The Data logging screen
No new measurement data is entered into the data log when you press
the soft key. To enter data into the new batch, you must press the
soft key again, or use the procedure given in Section 6.7.4.
1. With the Menu screen displayed,
use the and soft keys to
highlight the "Data log" option, then
press the soft key. The Data
log options screen will then be
displayed, as shown in Figure 44.
Figure 44 - The Data log options screen
The data log will only be output if you have correctly configured the serial
output parameters as described in Section 6.7.2.
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8 58
6.7.7 Viewing the data log
1. With the Menu screen displayed, use the and soft keys to highlight
the "Data log" option, then press the soft key. The Data log options screen
will then be displayed, as shown in Figure 44.
2. Use the and soft keys to highlight the "View log" option, then press
the soft key. The View log (batch) screen is then displayed, alternating with
the View log (date) screen: see Figures 45 and 46.
3. Use the and soft keys to scroll through all of the measurements
stored in the data log.
4. When you have finished viewing the data log, press the soft key or the
soft key to display the Data log options screen again.
6.7.8 Clearing the data log
1. With the Menu screen displayed, use the and soft keys to highlight
the "Data log" option, then press the soft key. The Data log options screen
will then be displayed, as shown in Figure 44.
2. Use the and soft keys to highlight the "Clear log" option, then press
the soft key. The screen will then show an "Are you sure?" message:
Press the soft key to clear the data log. The Data log options screen is
then displayed again.
Press the soft key to display the Data log options screen again without
clearing the data log.
Figure 45 - The View log (batch)
screen
Figure 46 - The View log (date)
screen
Ensure that you have viewed the data log (see Section 6.7.7) or have output
the data log if required (see Section 6.7.6) before you clear the log.
You cannot clear batches or individual measurements from the data log.
Batch number
Sequential number
of measurement in
the batch
Date of sample
measurement
Time of sample
measurement
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8
59
6.7.9 Printing a sample measurement report
If you have selected printed outputs (see Section 6.7.1), you must manually initiate the
printing of each sample measurement report.
Press the soft key when you want to print a report. Refer to Appendix A3 for the
format of the printed sample measurement report.
6.8 Adjusting the display
At any time, you can adjust the screen display to suit the ambient light conditions, as
described in Sections 6.8.1 and 6.8.2.
6.8.1 Adjusting the backlight timer
When the Multi Purpose is first switched on, the backlight goes on to illuminate the
screen. If no soft key is pressed, the backlight will remain on for the preset backlight
time, and will then go off. The timer associated with the backlight time is restarted
whenever a soft key is pressed (that is, the backlight remains on for the backlight time
after the last soft key press). To adjust the backlight time:
1. On the Settings screen, highlight the "Backlight" option, then press the soft
key. The Backlight timer screen will then be displayed, as shown in Figure 47.
2. Change the backlight time (Duration) setting as required, then press the
soft key.
Figure 47 - The Backlight timer screen
The backlight time (Duration) can be set between 0 and 999 seconds. Set
the backlight time (Duration) to 0 seconds to leave the backlight permanently
switched on.
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8 60
6.8.2 Adjusting the contrast
1. On the Settings screen, highlight the "Contrast" option, then press the soft
key. The Contrast screen will then be displayed, as shown in Figure 48.
Figure 48 - The Contrast screen
2. Use the and soft keys to increase or decrease the contrast to the
required level, then press the soft key.
6.9 Switching off the Multi Purpose after use
When you have finished using the Multi Purpose:
1. Switch off the Multi Purpose: press and hold the Power On/Off key for
approximately 2 seconds, then release the key when the audible alarm goes on.
2. If you will not use the Multi Purpose for several days:
Disconnect any tubes from the sample gas outlet (on the front or the rear of
the Multi Purpose: see Figures 1 and 2) and from the bypass gas outlet (if
fitted) on the rear of the Multi Purpose (see Figure 2).
Fit protective plastic caps to the sample gas outlet (on the front or the rear of
the Multi Purpose: see Figures 1 and 2), to the bypass gas outlet (if fitted) on
the rear of the Multi Purpose (see Figure 2), and to the sample gas inlet on
the front of the Multi Purpose (see Figure 1).
3. If required, disconnect the Multi Purpose from the electrical supply.
If your Multi Purpose has the optional rechargeable battery, you can leave
the Multi Purpose connected to the electrical supply. This will not affect the
service life of the battery.
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8
61
7 ROUTINE MAINTENANCE
7.1 Cleaning the Multi Purpose
Regularly use a damp (but not wet) cloth to wipe clean the outer surfaces of the Multi
Purpose, to prevent the entry of dust or other particulates into the gas sample inlet or
the interior of the Multi Purpose.
7.2 Inspecting the inlet filter element
If you only use the Multi Purpose on applications which use clean, dry cylinder gases,
you will only need to inspect the inlet filter element every 3 months. On other
applications, we recommend that you inspect the inlet filter element more frequently.
1. Ensure that the Multi Purpose is switched off.
2. Unscrew and remove the filter retainer cap (see Figure 1).
WARNING
Sample and calibration gases may be toxic or asphyxiant. Never inspect
the inlet filter, remove the side panels from the Multi Purpose, or service
or repair the Multi Purpose while such gases are still connected to it.
WARNING
Sample and calibration gases may be toxic or asphyxiant and may build
up within the Multi Purpose during use. Always inspect the inlet filter,
service the Multi Purpose, or remove the side panels from the Multi
Purpose in a well-ventilated environment, so that such gases (if present)
do not present a hazard.
WARNING
Sample and calibration gases may be toxic or asphyxiant. If you suspect
there is a gross leak or internal pipework damage in the Multi Purpose,
leave the Multi Purpose in a well-ventilated environment for at least
8 hours before servicing or repair.
The filter element is made of borosilicate glass, and the filter retainer cap
O ring is made from Viton
(fluoroelastomer).
New filter elements and filter retainer cap O rings are available as spares:
see Section 10.
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8 62
3. Inspect the condition of the white filter element (fitted to the spigot on the rear of
the filter retainer cap). If the filter element is wet or dirty:
Remove the used filter element from the filter retainer cap, then dispose of
the element.
Push a new filter element onto the spigot on the inner side of the filter retainer
cap.
4. Inspect the O ring on the inner side of the filter retainer cap. If the O ring is
twisted or damaged:
Remove the O ring, then dispose of it.
Fit a new O ring to the inner side the filter retainer cap.
5. Ensure that the O ring is correctly located in the recess in the inner side of the
filter retainer cap, then refit and tighten the filter retainer cap.
7.3 Regenerate/replace the drying agent
A Multi Purpose without a Sample Conditioning Panel is supplied with a drying tube
which can be either of two types:
A flexible drying tube, containing an indicating silica gel drying agent.
A screw-top drying tube, containing non-indicating Drierite
drying agent.
If your Multi Purpose is supplied with a flexible drying tube, you must regenerate or
replace the drying agent when it has changed colour from orange to green: if required,
use the procedure given in Section A7.4.2 to regenerate the drying agent. After
regeneration or replacement of the drying agent, take care when you refit the barbed
connections on the drying tube, to ensure that they are securely fitted.
If your Multi Purpose is supplied with a screw-top drying tube, we recommend that you
regenerate or replace the drying agent after approximately 2 hours of use sampling
wet gases. If required, use the procedure given in Section A7.4.2 to regenerate the
drying agent.
When you want to replace the drying agent, note that both types of drying agent are
available as spares (see Section A7.5) or from various suppliers. The recommended
granule sizes for the drying agents are:
0.5-1.0 mm granule size for the indicating silica gel
8 mesh (or smaller) granule size for the Drierite
The Multi Purpose is not suitable for disposal in municipal waste streams
(such as landfill sites, domestic recycling centres and so on). Refer to
Appendix A10 for disposal requirements in accordance with the WEEE
Directive within the EC.
If you send the Multi Purpose to Servomex or your local Servomex agent for
disposal, the Multi Purpose must be accompanied by a correctly completed
decontamination certificate.
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8
75
10 SPARES
The standard spares available for the Multi Purpose are shown below. You can order
these spares from Servomex or your agent.
Spare Part Number
Inlet filter element (pack of 5) 00570982
Filter retainer cap O ring 2323-7029
Inlet/outlet quick-connect fitting 202517
Power supply unit 202578
Thermal printer paper (pack of 5 rolls) 203022
Internal sample pump (standard Multi Purpose) 05200930
Internal sample pump (Multi Purpose HF) 05200948
Rechargeable battery pack 05200931
Alarms and fault information are not provided in the serial outputs.
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8
79
APPENDIX
A3 PRINTER OUTPUT FORMATS
If you have selected printed outputs (see Section 6.7.1), each time you press the
soft key, a single sample measurement report is printed (on the Servomex supplied
printer). The format of this report is as shown below:
where:
Date: <date>
Time: <time>
I.D.: <serial number>
Batch:
Product:
Measurement:
1 <gas> XXXX <units> <alarm>
2 <gas> XXXX <units> <alarm>
Signature
<date> Is the date on which the report was printed, in the currently
selected format (see Section 5.3.2).
<time> Is the time at which the report was printed, in the format
hh:mm:ss (hours:minutes:seconds).
<serial number> Is the serial number of the Multi Purpose.
1 <gas> Specifies transducer 1 and the sample gas measured.
2 <gas> Specifies transducer 2 and the sample gas measured.
Refer to your company procedures for the correct usage of the Batch,
Product and Signature fields of printed sample measurement reports.
Refer to the printer manual for additional information (for example, on how to
replace the printer roll).
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8
81
APPENDIX
A4 RS232 CONNECTION DETAILS
A4.1 Overview
The serial port on the rear of the Multi Purpose (see Figure 2) is an RS232 5.5 V
9-way D type connector.
The pins on the connector are used as shown in Figure A1 below:
Figure A1 - RS232 connection pin details
A4.2 Connecting the Multi Purpose to a PC
The Multi Purpose can be directly connected to the 9-way D type serial port (usually
designated "COM1" or "COM2") on your PC. To comply with EMC immunity
standards, we recommend that you use the serial output cable supplied. If you do not
use the cable supplied, use a compatible 9-way D Null Modem cable (with a
recommended maximum length of 3 m), with female-to-female connectors.
If your PC only has USB serial ports, use a commercially available 9-way D type serial
to USB converter to connect the Multi Purpose to one of the serial ports.
WARNING
Ensure that the electrical installation of any equipment connected to the
Multi Purpose conforms with all applicable local and national electrical
safety requirements.
Pin(s) Use
1 Not used
2 Rx (to Multi Purpose)
3 Tx (from Multi Purpose)
4 Not used
5 0 V
6, 7, 8, 9 Not used
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8 82
A4.3 Capturing data using Windows
You can drag the icon onto your desktop for quick access to
HyperTerminal.
Parameter Recommended
setting
Baud rate 38400
Data bits 8
Parity None
Stop bits 1
Flow control None *
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8
83
APPENDIX
A5 DISPLAY UNIT CONVERSION FACTORS
When you select display units as described in Section 6.4, you must ensure that you
also enter the correct units conversion factor, as shown in the table below:
To convert
from *
to
To return to the measurement default units, select the "off" units selection
option and set the units conversion factor to "1": see Section 6.4.
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8 84
APPENDIX
A6 OPTIONAL FLOWMETERS
A6.1 Overview
Two optional flowmeters are available for the Multi Purpose, as follows:
The unvalved flowmeter, for use with a standard Multi Purpose without an internal
sample pump. This flowmeter shows the flow rate of sample gas or calibration gas
into the Multi Purpose.
The flowmeter with a flow control valve, for use with a Multi Purpose HF or a
standard Multi Purpose with an internal sample pump. This flowmeter shows (and
allows you to control) the flow rate of sample gas or calibration gas into the Multi
Purpose.
A6.2 Specification
A6.3 Preparing the Multi Purpose with a flowmeter
If you have ordered the flowmeter option, the flowmeter will be supplied fitted to the
side of the Multi Purpose, as shown in Figure A2.
Before you use the Multi Purpose, ensure that the inlet pipe (2) is correctly fitted
between the flowmeter outlet and the sample gas inlet on the front of the Multi
Purpose. If the pipe is not already fitted:
1. Remove the protective plastic cover from the flowmeter outlet.
2. Remove the protective plastic cover from the sample gas inlet on the front of the
Multi Purpose (Figure 1, item 8).
3. Fit the inlet pipe supplied between the flowmeter outlet and the sample gas inlet
on the front of the Multi Purpose.
Flowmeter indicated flow rate range
Standard Multi Purpose without sample pump 1 to 10 l min
-1
*
Standard Multi Purpose with sample pump 100 to 1000 cm
3
min
-1
Multi Purpose HF 200 to 2500 cm
3
min
-1
#
* Equivalent to a flow rate range of 0.035 to 0.35 ft
3
min
-1
.
Equivalent to a flow rate range of 100 to 1000 ml min
-1
(0.0035 to 0.035 ft
3
min
-1
).
#Equivalent to a flow rate range of 0.2 to 2.5 l min
-1
(0.007 to 0.087 ft
3
min
-1
).
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8
85
* Fitted between the flowmeter outlet and the Multi Purpose inlet.
Only fitted to a valved flowmeter for use with a Multi Purpose HF or a standard Multi
Purpose with an internal sample pump.
Figure A2 - Multi Purpose with a flowmeter
Key Description Key Description
1.
2.
3.
Multi Purpose
Inlet pipe *
Sample gas inlet
4.
5.
6.
Flow control valve
Flow indicator
Flowmeter assembly
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8 86
A6.4 Using the Multi Purpose with a flowmeter
Operate the Multi Purpose with the flowmeter as described in Sections 4 to 6 of this
manual, but with the exceptions noted below.
A6.5 Multi Purpose HF and standard Multi Purpose with a sample pump
Calibrate the Multi Purpose as described in Section 6.1, but:
Connect the calibration gas supply to the sample gas inlet on the flowmeter
(Figure A2, item 3), not to the Multi Purpose.
Use the flow control valve (Figure A2, item 4) to adjust the flow of calibration gas
into the Multi Purpose to the required flow rate.
Take sample readings as described in Section 6.2, but:
Connect the sample gas supply to the sample gas inlet on the flowmeter (Figure A2,
item 3), not to the Multi Purpose.
Use the flow control valve (Figure A2, item 4) to adjust the flow of sample gas into
the Multi Purpose to the required flow rate.
A6.6 Standard Multi Purpose without a sample pump
Calibrate the Multi Purpose as described in Section 6.1, but connect the branch on the
calibration T piece to the sample gas inlet on the flowmeter (Figure A2, item 3), not
to the Multi Purpose.
Take sample readings as described in Section 6.2, but connect the sample gas supply
to the sample gas inlet on the flowmeter (Figure A2, item 3), not to the Multi Purpose.
The flow rate of your calibration gas supply must be the same as the flow
rate of the gases to be sampled. If the flow rates are different, sample gas
measurements may not be accurate.
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8
87
APPENDIX
A7 OPTIONAL SAMPLE CONDITIONING PANEL
A7.1 Overview
The optional sample conditioning panel allows for use of the Multi Purpose to measure
the oxygen content of wet gas samples, for example on a spot-check basis (when
close control of boiler efficiency is not required), or while carrying out boiler
performance surveys.
Refer to Figure A3. If you order this option, the sample conditioning panel (2) will be
supplied fitted to the side of the Multi Purpose (1).
During sampling, sample gas passes through the catchpot (6), and then the drier (3,
which contains a drying agent), before passing into the Multi Purpose.
You must drain the catchpot when necessary (see Section A7.4.1) and you must
regularly regenerate or replace the drying agent (see Section A7.4.2).
A7.2 Specification
A drying tube is not supplied if your Multi Purpose has the sample
conditioning panel fitted, as the panel has an integral drier.
Response time *
Standard oxygen sensor 60 seconds
Industrial or high accuracy
oxygen sensor
65 seconds
IR (infrared) sensor 130 seconds
Sample gas temperature
Ambient temperature 10
o
C/18
o
F
Sample gas inlet
#
Suitable for 4 mm inside diameter tube
Drying agent Indicating silica gel
* The response times are approximate and are those necessary to reach 90% of
final measurement, with indicating silica gel as the drying agent. If required, you
can use calcium sulphate as the drying agent (see Section A7.4.2), to reduce
response time to approximately 50 seconds.
For maximum operating life and efficiency.
# On the sample conditioning panel (Figure A3, item 7).
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8 88
A7.3 Using the Multi Purpose with the sample conditioning panel
A7.4 Calibrating
You must calibrate the Multi Purpose with the sample conditioning panel as described
in Section 6.1, but you must connect the calibration gas supply to the sample gas inlet
on the sample conditioning panel (Figure A3, item 7).
A7.5 Sampling gases
To obtain the best results from flue gas sampling, we recommend that you locate the
end of the sample pipe/probe as close as possible to the centre of the flue duct. If the
end of the sample pipe/probe is too close to the wall of the flue duct, the gases
sampled may not be representative of the boiler combustion products because of the
stratification of the gases, or because of leaks in the duct.
Refer to Figure A3. When you want to sample gases:
1. Ensure that the inlet pipe (4) from the sample conditioning panel is connected to
the sample gas inlet on the front of the Multi Purpose (Figure 1, item 8).
2. Take sample readings as described in Section 6.2, but note that:
You must connect a gas sample pipe/probe to the sample gas inlet (7) on the
sample conditioning panel.
Route the end of the sample pipe/probe to a suitable sample point (see
above).
Allow sufficient time for the sample measurement to stabilise (see
Section A7.2).
If the sample gases are very dirty (that is, contain soot or particulates), we
recommend that you use the catchpot as a bubbler when you carry out
sample measurements; that is: remove the top of the catchpot and fill the
catchpot with clean water to a level just above the bottom of the gas inlet
pipe in the catchpot.
If you are sampling flue gases, ensure that you position the end of the
sample pipe/probe at a location in the flue where the gases are
representative of the boiler combustion products.
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8
89
Figure A3 - Multi Purpose with the sample conditioning panel
Key Description Key Description
1.
2.
3.
4.
Multi Purpose
Sample conditioning panel
Drier
Inlet pipe
5.
6.
7.
Catchpot drain plug
Catchpot
Sample gas inlet
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8 90
A7.6 Additional maintenance
A7.7 Draining the catchpot
Refer to Figure A3. During use, liquid entrained in the sampled gases will be
condensed in the catchpot (6).
When necessary (that is, when liquid is visible in the transparent body of the catchpot):
1. Place a suitable container under the catchpot drain plug (5).
2. Remove the drain plug and allow the liquid to drain from the catchpot.
3. Refit the drain plug and safely dispose of the drained liquid: refer to Section 9.2.
A7.8 Replacing/regenerating the drying agent
The colour of the silica gel drying agent in the drier is orange as supplied, and the
column of drying agent will turn progressively green as moisture is absorbed. When
the whole column of drying agent has turned green, the drying agent needs to be
regenerated or replaced.
You can regenerate the drying agent by passing dry air or dry nitrogen through the
drier, or:
Removing it from the drier and laying it out on a tray, approximately 1 granule thick.
Heating the silica gel on the tray in an oven at a temperature of 110 to 120
o
C (230
to 248
o
F) until the granules have turned orange again, or heating the calcium
sulphate (Drierite
Do not allow the level of liquid in the catchpot to rise above the top of the
transparent body of the catchpot.
WARNING
If you replace the drying agent, ensure that you comply with any safe
handling and usage requirements specified on or with the package
containing the new drying agent.
If required, replace the silica gel drying agent supplied with calcium sulphate
(for example, use Drierite
If your probe has a protective tip, remove the tip if the temperature of the
location where you will sample gases exceeds 180
o
C/356
o
F.
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8
93
A8.5 Additional spares
The following additional spares are available for the gas probe accessory:
Order these spares from Servomex or your Servomex agent.
* Removable: only supplied fitted to the 25 cm Probe.
Optional.
# Only supplied if you also have the sample conditioning panel option.
Figure A4 - Optional gas probe assembly
Spare Part Number
Filter assembly 203148
Filter element (pack of 5) 00570982
Probe: 25 cm (approx 10 inch), with protective tip S5000942
Probe: 100 cm (approx 39 inch) S5000404B
Key Description Key Description
1.
2.
3.
4.
Protective tip *
Probe
Handle
Filter
5.
6.
7.
Coiled sample pipe
Outlet connector
Catchpot inlet adaptor
#
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8 94
APPENDIX
A9 MATERIALS IN CONTACT WITH SAMPLE AND CALIBRATION
GASES
The materials of the parts of the Multi Purpose in contact with the sample and
calibration gases are listed below. These materials have a wide range of chemical
compatibility and corrosion resistance.
Common gas path in the Multi Purpose:
Nickel
Viton
Borosilicate glass
Kynar
Krytox
GPL205 grease
* Standard Multi Purpose without internal sample pump only.
Standard Multi Purpose with internal sample pump only.
# Multi Purpose HF only.
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8
95
Industrial or high accuracy oxygen sensor:
316 stainless steel
Borosilicate glass
Platinum
Platinum/iridium alloy
Electroless nickel
Viton
IR (infrared) sensor:
316 stainless steel
Gold
Nickel
Sapphire
Epoxy adhesive (EPO-TEK H72)
Viton
Optional Flowmeters:
Acrylic
Nickel
Polyurethane
Viton
Nickel
Optional Gas Probe:
316 stainless steel
Nickel
Viton
Polyurethane
Nylon
Borosilicate glass
PVC (polyvinylchloride)
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8 96
APPENDIX
A10 DISPOSAL IN ACCORDANCE WITH THE WASTE ELECTRICAL
AND ELECTRONIC EQUIPMENT (WEEE) DIRECTIVE
The label shown in Figure A5 is fitted to the analyser.
Figure A5 - The WEEE label
This label identifies that:
The analyser is considered to be within the scope of the Waste Electrical and
Electronic Equipment (WEEE) Directive 2002/96/EC.
The analyser is not intended for disposal in a municipal waste stream, but shall be
submitted for material recovery and recycling in accordance with the local
regulations which implement the WEEE Directive.
For additional information and advice on the disposal of the analyser in accordance
with the requirements of the WEEE Directive, contact Servomex or your local
Servomex agent.
If you send the analyser to Servomex or your local Servomex agent for
disposal, the analyser must be accompanied by a correctly completed
decontamination certificate.
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8
97
APPENDIX
A11 COMPLIANCE AND STANDARDS INFORMATION
The 5200 Multi Purpose Gas Analyser complies with the European Community
"Electromagnetic Compatibility Directive", 89/336/EEC (as amended by 92/31/EEC
and 93/68/EEC).
The 5200 Multi Purpose is excluded from the scope of the European Community
"Low Voltage Directive", 73/23/EEC.
The 5200 Multi Purpose has been assessed to IEC 61010-1:2001 (+Corr 1: 2002 +
Corr 2:2003) for electrical safety, rated in accordance to IEC 60664-3 Category II,
Pollution Degree 2.
The 5200 Multi Purpose has been validated and fully complies with the
requirements of The Food and Drug Act - specifically, 21 CFR 211.165 (e) and
211.194 (a)(2) - for verification of the strength, identity and purity of: MEDICAL
OXYGEN (USP) and OXYGEN IN NITROGEN NF 19.
The power supply unit supplied with the 5200 Multi Purpose complies with the
European Community "Low Voltage Directive", 73/23/EEC.
Servomex Group Ltd is a BS EN ISO 9001:2000 and EN ISO 14001:1996 certified
organisation.
5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 8 98
This page intentionally blank.