A Technical Guide For Static Headspace Analysis Using GC
A Technical Guide For Static Headspace Analysis Using GC
Connectors
Analytical Reference Materials
GC Accessories
tatic headspace gas chromatography (GC) is a technique used for the concentra-
tion and analysis of volatile organic compounds. This technique is relatively S
Time and money are two of the many reasons why an analyst
would use static headspace analysis. Other reasons may include
ease of operation and the ability to assay a variety of sample
matrices.
simple and can provide sensitivity similar to dynamic purge and trap analysis. The
popularity of this technique has grown and has gained worldwide acceptance for
analyses of alcohols in blood and residual solvents in pharmaceutical products.
Other common applications include industrial analyses of monomers in polymers
and plastic, flavor compounds in beverages and food products, and fragrances in
perfumes and cosmetics.
Sample matrices like blood, plastic, and cosmetics contain high molecular weight,
non-volatile material that can remain in the GC system and result in poor analytical
performance. Many laboratory analysts use extensive sample preparation techniques
to extract and concentrate the compounds of interest from this unwanted non-
volatile material. These extraction and concentration techniques can become time
consuming and costly. Static headspace analysis avoids this time and cost by directly
sampling the volatile headspace from the container in which the sample is placed.
Because of the diversities in the industry and related products, this guide attempts to
cover only the basic principles of static headspace and demonstrate how to apply
them to achieve optimum chromatographic results. With an understanding of these
principles, various instrumentation will then be reviewed to help build upon this
knowledge and identify the benefits and potential problems associated with each
mode of sample transfer. Information from the Basic Principles and Instrumentation
sections of this guide can then be brought together and applied to the conditions and
methodology of common analyses. Like most applications, a variety of problems
may arise in which the System Optimization section will help to identify these
problems and offer techniques to help resolve them.
3
Figure 1
Phases of the headspace vial.
Equation 1
Partition Coefficient (K) = C
s
/C
g
Equation 2
Phase Ratio () = V
g
/V
s
C
s
=concentration of analyte in sample phase
C
g
=concentration of analyte in gas phase
V
s
=volume of sample phase
V
g
=volume of gas phase
Figure 2
K and are important variables in
headspace analysis.
Solvent K Value
cyclohexane 0.077
n-hexane 0.14
tetrachloroethylene 1.48
1,1,1-trichloromethane 1.65
o-xylene 2.44
toluene 2.82
benzene 2.90
dichloromethane 5.65
n-butyl acetate 31.4
ethyl acetate 62.4
methyl ethyl ketone 139.5
n-butanol 647
isopropanol 825
ethanol 1355
dioxane 1618
Table I
K Values of Common Solvents in Air-
Water Systems at 40C
Basic Principles of Headspace Analysis
Most consumer products and biological samples are composed of a wide variety of
compounds that differ in molecular weight, polarity, and volatility. For complex
samples like these, headspace sampling is the fastest and cleanest method for
analyzing volatile organic compounds. A headspace sample is normally prepared in
a vial containing the sample, the dilution solvent, a matrix modifier, and the
headspace (see Figure 1). Volatile components from complex sample mixtures can
be extracted from non-volatile sample components and isolated in the headspace or
vapor portion of a sample vial. An aliquot of the vapor in the headspace is delivered
to a GC system for separation of all of the volatile components.
In order to achieve the best performance when using headspace/GC, careful atten-
tion should be used in sample preparation and instrument setup. Key issues to
address when setting up headspace/GC systems include minimizing system dead
volume, maintaining inert sample flow paths, and achieving efficient sample
transfer. These issues, as well as other instrument setup-related topics, are addressed
later in the System Optimization section of this guide.
Samples must be prepared to maximize the concentration of the volatile components
in the headspace, and minimize unwanted contamination from other compounds in
the sample matrix. To help determine the concentration of an analyte in the
headspace, you will need to calculate the partition coefficient (K), which is defined
as the equilibrium distribution of an analyte between the sample phase and the gas
phase (Figure 2).
Partition Coefficient
Compounds that have low K values will tend to partition more readily into the gas
phase, and have relatively high responses and low limits of detection (Figure 3). An
example of this would be hexane in water: at 40C, hexane has a K value of 0.14 in
an air-water system. Compounds that have high K values will tend to partition less
readily into the gas phase and have relatively low response and high limits of
detection. An example of this would be ethanol in water: at 40C, ethanol has a K
value of 1355 in an air-water system. Partition coefficient values for other common
compounds are shown in Table I.
volatile
analytes
sample, dilution
solvent, and matrix
modifier
}
}
G
S
G=the gas phase (headspace).
The gas phase is commonly referred to as the headspace
and lies above the condensed sample phase.
S=the sample phase.
The sample phase contains the compound(s) of interest
and is usually in the form of a liquid or solid in combina-
tion with a dilution solvent or a matrix modifier.
Once the sample phase is introduced into the vial and the
vial is sealed, volatile components diffuse into the gas
phase until the headspace has reached a state of equilib-
rium as depicted by the arrows. The sample is then taken
from the headspace.
4
www.restek.com
(Ideal)
K
C
g
(Ideal)
Figure 3
Sensitivity is increased when K
is minimized.
K and
Figure 5
Lower K and result in higher C
g
and better sensitivity.
ammonium chloride
ammonium sulfate
sodium chloride
sodium citrate
sodium sulfate
potassium carbonate
Table II
Common salts used to decrease
matrix effects.
(Ideal)
C
g
Figure 4
Sensitivity is increased when
is minimized.
C
g
K can be lowered by changing the
temperature at which the vial is equili-
brated or by changing the composition of
the sample matrix. In the case of ethanol,
K can be lowered from 1355 to 328 by
raising the temperature of the vial from
40C to 80C. It can be lowered even
further by introducing inorganic salt into
the aqueous sample matrix. High salt
concentrations in aqueous samples
decrease the solubility of polar organic
volatiles in the sample matrix and promote their transfer into the headspace,
resulting in lower K values. However, the magnitude of the salting-out effect on K is
not the same for all compounds. Compounds with K values that are already rela-
tively low will experience very little change in partition coefficient after adding a
salt to an aqueous sample matrix. Generally, volatile polar compounds in polar
matrices (aqueous samples) will experience the largest shifts in K and have higher
responses after the addition of salt to the sample matrix. Table II lists most of the
common salts used for salting-out procedures.
Phase Ratio
The phase ratio () is defined as the relative volume of the headspace compared to
volume of the sample in the sample vial (Figure 2). Lower values for (i.e., larger
sample size) will yield higher responses for volatile compounds (Figure 4). How-
ever, decreasing the value will not always yield the increase in response needed to
improve sensitivity. When is decreased by increasing the sample size, compounds
with high K values partition less into the headspace compared to compounds with
low K values, and yield correspondingly smaller changes in C
g
. Samples that contain
compounds with high K values need to be optimized to provide the lowest K value
before changes are made in the phase ratio.
Combining K and
Partition coefficients and phase ratios
work together to determine the final
concentration of volatile compounds in
the headspace of sample vials. The
concentration of volatile compounds in
the gas phase can be expressed as
C
g
=C
o
/(K+) (where C
g
is the concen-
tration of volatile analytes in the gas
phase and C
o
is the original concentra-
tion of volatile analytes in the sample).
Striving for the lowest values for both
K and will result in higher concentra-
tions of volatile analytes in the gas
phase and, therefore, better sensitivity
(Figure 5).
5
www.restek.com
Compound of Interest Derivatizing Reagent Resulting Derivative
fatty acids
methanol
esterification
with boron trifluoride
glycerol
acetic anhydride
acetylation
with sodium carbonate
For more information on derivatization, please refer to the Handbook of Analytical
Derivatization Reactions by Daniel R. Knapp or to the text at right.
Table III
Common reagents used to derivatize compounds of interest.
Common derivatization techniques used in reaction headspace/GC are esterification,
acetylation, silylation, and alkylation. Any of these derivatization techniques can be
performed using the sample vial as the reaction vessel (see Table III for a list of
commonly used reagents). Although derivatization may improve chromatographic
performance and volatility for some compounds, derivatization reactions may
introduce other problems into the analytical scheme. Derivatization reagents as well
as the by-products from derivatization reaction may be volatile and can partition into
the headspace along with derivatized compounds. These extra volatile compounds
may pose problems by eluting with similar retention times as the compounds of
interest, causing either partial or complete coelutions.
Derivatization reactions also are typically run at elevated temperatures. Pressures
inside the sample vial may exceed the pressure handling capabilities of the vial or
the septa. Specially designed septa are available that allow excess pressure to be
vented during derivatization reactions.
Headspace Sample Size
In addition to working with K, , and derivatization reactions, sensitivity also can be
improved by simply increasing the size of the headspace sample that is withdrawn
from the sample vial and transferred to the GC. Increasing the sample size also
means that the amount of time it takes to transfer the sample to the column will
increase in proportion to the column volumetric flow rate. Sample size can be
increased only to the point that increases in peak width, as a result of longer sample
transfer times, will not affect chromatographic separations. Larger sample sizes and
longer transfer times can be offset by using cryogenic cooling and sample refocus-
ing at the head of the column.
Derivatization/Reaction Headspace
Derivatization is another technique that can be used to increase sensitivity and
chromatographic performance for specific compounds. Compounds such as acids,
alcohols, and amines are difficult to analyze because of the presence of reactive
hydrogens. When attempting to analyze these types of compounds, they can react
with the surface of the injection port or the analytical column and result in tailing
peaks and low response. In addition, they may be highly soluble in the sample
phase, causing very poor partitioning into the headspace and low response.
Derivatization can improve their volatility, as well as reduce the potential for surface
adsorption once they enter the GC system.
For more information on headspace
anaysis, check out the textbook,
Static Headspace-Gas
Chromatography, Theory and Practice
by Bruno Kolb and Leslie S. Ettre.
Instrumentation
Gas-Tight Syringe Injection
Use of a gas-tight syringe autosampling system is one of three common techniques
(gas-tight syringe, balanced pressure, and pressure loop) used to transfer a
headspace sample. Most of the autosampling units can retrofit to a standard GC with
a split/splitless injection port, making them relatively simple to use and understand.
These systems do not require the use of special configurations or special instrumen-
tation other than the autosampler itself. The gas-tight syringe autosampler is
beneficial for use with diverse samples because of the variety of sampler configura-
tions and method options available.
The gas-tight syringe technique operates by initially thermostatting the sample in an
incubation oven at a given temperature and for a given time until it has reached a
state of equilibrium (Figure 6, Step 1). Once the sample has reached an equilibrium,
an aliquot is taken from the headspace using the gas-tight syringe (Figure 6, Step
2), and the aliquot is injected into the GC as if it were a liquid sample injection
(Figure 6, Step 3).
Several concerns exist regarding this technique. Because the sample is being trans-
ferred from a heated oven, the syringe also must be heated to ensure that the sample will
not recondense in the syringe. Many manufacturers have taken this into consideration
and their samplers now come with a heated syringe assembly. There also are reproduc-
ibility issues because of possible sample loss. As the sample is transferred from the vial
to the injection port, some of it may be lost because of the pressure differences between
the vial and atmospheric conditions. Beyond these concerns, the gas-tight syringe
technique is simple to use, can retrofit into a variety of GC systems, and is best suited
for diverse samples. Examples of manufacturers and models of the gas-tight syringe
units are: the ThermoQuest TRACE
parts can
be used to minimize sample adsorption and peak tailing. Transfer line internal
diameter should be as narrow as possible to help maintain narrow sample band
widths and symmetrical peak shapes (see the following optimization of transfer lines
for more information). Analysts also should ensure that balanced-pressure instru-
ments are leak-free and operate with the least amount of dead volume in the sample
flow path. This will help obtain optimal peak shape and sensitivity.
When using pressure-loop sampling instruments, the same concerns apply as with
gas-tight syringe and balanced-pressure systems. Inert sample pathways and low
dead volume systems will yield the best chromatographic performance. In pressure-
loop systems, a gas sampling valve with a sample loop is used to transfer the sample
from the headspace unit to the GC. Adequate purging of the sample valve and loop
will guard against sample carryover. If low response or broad peaks are observed, it
may be necessary to increase the sample vial pressure to ensure that the sample loop
is being completely filled with headspace sample. If there are extraneous peaks
present due to carryover of matrix contaminants, increase the sample valve tempera-
ture to prevent sample carryover, condensation, and contamination.
Shaking or vibrating the
vial during heating can
assist in achieving
equilibrium.
10
www.restek.com
Transfer Line
After the headspace sample is withdrawn from the vial, it ready to be transferred to
the GC. In balanced-pressure and pressure-loop systems a short piece of tubing
called a transfer line is used to transfer the sample from the autosampler to the GC.
Transfer line material must be chosen that suits the sample analytes. Many different
materials can be used as transfer line tubing, including stainless steel, nickel, fused
silica, and Silcosteel
- or Siltek
tubing.
The internal diameter of the
transfer line should be chosen
depending on the internal
diameter of the analytical
column, the column flow rate,
and the flow rate delivered
from the autosampler. To elimi-
nate tubing dead volume, use the smallest diameter tubing possible. For
example, compound residence time in a 1.0mm ID transfer line is 3.6 times greater
than in the same length of 0.53mm ID tubing. Reducing the residence time of the
headspace sample in the transfer line helps to minimize band broadening. Therefore,
the flow rate should be set as high as possible to quickly move the sample cloud
through the tubing and minimize any dead volume effects.
Transfer line temperature should be set depending on the analytes of interest and the
sample matrix. Typical transfer line temperatures range from 80C to 125
C. To
minimize matrix problems and prevent water condensation from aqueous samples,
use a higher transfer line temperature (~125C to 150
C).
Injection Port Interface
The quality of the connection of the transfer line to the analytical column greatly
affects sample bandwidth. In most cases, the transfer line has a smaller internal
diameter than the injection port liner, and the vaporized headspace sample carrying
the compounds of interest will be diluted into a larger volume of carrier gas when
the sample elutes from the transfer line into the inlet liner. This can lead to broader
peaks, tailing peaks, lower sensitivity, and loss of resolution. Because headspace
samples are already in a gaseous state (vapor cloud) when they enter the injection
port, there is no need to use a large buffer volume in the liner to allow for sample
expansion as when analyzing liquid samples. Using injection port liners that have
smaller internal diameters and lower buffer volumes will help maintain a narrow
bandwidth as samples move from the end of the transfer line to the head of the
analytical column. 1.0mm ID deactivated injection port liners are recommended for
most headspace applications to achieve the lowest injection port dead volume.
If band-broadening due to excess dead volume in the system is still a problem, peak
shape may be improved by refocusing sample analytes at the analytical column head.
Highly volatile compounds can be trapped at the column head and refocused into a
narrow bandwidth by reducing the initial oven temperature below the boiling point of
compounds of interest. After the sample is completely transferred to the column, the
oven temperature can be increased to move the compounds through the column.
Use an inert transfer
line when optimizing
pressure-loop systems.
11
www.restek.com
Rtx
tubing;
Transfer line temp.: 200C; Injection port sleeve: 2mm ID; Split flow: 20mL/min.
Conc.
w/v
1. methanol 0.1%
2. acetaldehyde 0.2%
3. ethanol 0.2%
4. isopropanol 0.1%
5. acetone 0.01%
6. n-propanol 0.1%
Quantitation Technique for Blood
Alcohol Analysis (Internal Standard)
The internal standard technique uses one or
more designated compounds at known
concentrations spiked into the sample. The
response of the compounds of interest are
then compared to the results of the internal
standard. There are several advantages to
this technique. Multiple injections of the
standard are not necessary for concentra-
tion calculations; small changes in injection
volumes or detector response over time can
be determined.
Headspace Applications
Figure 12
Achieve baseline resolution of all blood alcohol components in less than 3 minutes using the
Rtx
-BAC2 columns.
12
N
e
w
Quantitation Technique for USP <467>
(External Standard)
This technique uses a separate sample
(standard) that has the compounds of interest
at known concentrations in the same matrix.
This technique is advantageous if various
samples are being analyzed, and all
compounds of interest can be assayed using
a single set external standards.
These conditions (PE Auto SYS GC and HS 40 headspace autosampler), combined
with unique columns (Rtx
1500 on 80/100
CarboBlack
-G43 column provides the resolution and detection limits needed for USP 24
th
edition <467> revised limit
test concentrations in USP Method IV.
4
1
2
3 5
2 3 4 5 6 7 8
min.
Stock Solvent Methylene Chloride Chloroform Benzene 1,1,1-Trichlorethylene 1,4-Dioxane
1 6 . 8 1 5 7 . 6 1 2 5 . 4 1 4 6 . 0 1 9 1 . 0 1 r e t a W
Water w/ Sample 8.29 9.25 11.3 13.38 15.26
3 4 . 4 1 1 0 . 8 8 1 . 8 9 5 . 8 7 3 . 7 O S M D
DMSO w/ Sample 7.25 7.54 8.8 8.67 7.37
Table V
Percent RSD for stock solutions in water vs. DMSO
Comments in the September/October 1992 Pharmacopoeial Forum
3
propose the use
of dimethyl sulfoxide as the solvent for stock standard, but this has not been
approved as of the date of this publication.
In regards to this proposal, an investigation was conducted to determine if there
were significant changes in results if dimethyl sulfoxide was used as the diluent for
stock standard. Similar RSDs can be obtained when stock solutions are diluted in
dimethyl sulfoxide as opposed to solutions made with water (Table V).
14
www.restek.com
European Pharmacopoeia Tests
The International Conference on Harmonization (ICH) has proposed a set of guide-
lines for residue solvent testing in pharmaceutical formulation and the European
Pharmacopoeia (EP) was the first to revise their regulations.
7,8
However, these
guidelines are challenging, containing over 60 compounds of regulatory interest to
manufacturers of active substances, excipients, and medicinal products. The EP
methods also allow testing limits based on either a concentration limit in a product,
or calculated from the maximum daily dosage of the product and the permissible
daily exposure limit of the solvent. These technical challenges affect the sampling
method and capillary column needed to ensure precise and accurate results.
The recommended primary capillary column for EP residual solvent testing is the
Rtx
guard column
(cat.# 16085-126). Oven temp.: 40C (hold 20 min.) to 240C @ 10C/min. (hold 20 min.);
Inj./det. temp.: 200C/250C; FID sensitivity: 1.1 x 10
-11
AFS;
Carrier gas: H
2
@ 35cm/sec.; Split ratio: 2:1.
Figure 14
The Rtx
-G27 Integra-Guard
Column Rtx
-G43 Integra-Guard
Column
Stabilwax
Capillary Columns
Rtx
-5 Capillary Columns
ID (mm) df (m) 30-Meter
0.53 5.00 10279-126
ID (mm) df (m) 30-Meter
0.53 3.00 16085-126
ID (mm) df (m) 30-Meter
0.32 3.00 10284
0.53 5.00 10279
ID (mm) df (m) 30-Meter
0.32 1.50 16069
0.53 3.00 16085
ID (mm) df (m) 30-Meter
0.32 0.25 10624
0.53 0.50 10640
0.53 1.00 10655
s n m u l o C d r a u G s n m u l o C y r a l l i p a C
Fused Silica Guard Columns
ID (mm) OD (mm) 5-Meter
0.32 0.45
+
0.04 10044
0.53 0.69
+
0.05 10045
Integra-Guard
Guard Columns
ID (mm) OD (mm) Length Suffix
0.32 0.45
+
0.04 5m -125
0.32 0.45
+
0.04 10m -128
0.53 0.69
+
0.05 5m -126
0.53 0.69
+
0.05 10m -129
Universal Angled Press-Tight
Connectors
Ideal for connecting guard columns to
analytical columns.
Designed at an angle approximating the
radius of a capillary column.
Reduces strain on column-end
connections.
5-pk. 25-pk. 100-pk.
20446 20447 20448
Universal Y Press-Tight
Connectors
Split sample flow onto two columns.
Split a single column flow into two
different detectors.
Perform confirmational analysis with a
single injection.
each 3-pk.
20405 20406
Universal Press-Tight
Connectors
Ideal for connecting guard columns to
analytical columns.
Repair broken columns.
Connect column outlets to transfer lines.
5-pk. 25-pk. 100-pk.
20400 20401 20402
Universal Angled Y
Press-Tight
Connectors
Alleviates column-end connection strain.
Inlet and outlet ends conform to the
column radius.
Perform confirmational analysis with a
single injection.
each 3-pk.
20403 20404
Universal Angled Siltek
-Deactivated
Press-Tight
Connectors
Siltek
Connectors
16
www.restek.com
European Pharmacopoeia/ICH Reference Material Mixes
Analytical Reference Materials
USP <467> Calibration Mix #2
L m / g 0 0 1 e n e z n e b
0 5 m r o f o r o l h c
1,4-dioxane 100
methylene chloride 500
trichloroethene 100
Prepared in methanol, 1mL/ampul
Ea.: cat.# 36002 10-pk.: cat.# 36102
USP <467> Calibration Mix #3
L m / g 0 0 1 e n e z n e b
0 5 m r o f o r o l h c
1,4-dioxane 100
methylene chloride 500
trichloroethene 100
Prepared in DMSO, 1mL/ampul
Ea.: cat.# 36004 10-pk.: cat.# 36104
USP <467> Calibration Mix #4
L m / g 2 e n e z n e b
0 6 m r o f o r o l h c
1,4-dioxane 380
methylene chloride 600
trichloroethene 80
Prepared in methanol, 1mL/ampul
Ea.: cat.# 36006 10-pk.: cat.# 36106
USP <467> Calibration Mix #5
L m / g 2 e n e z n e b
0 6 m r o f o r o l h c
1,4-dioxane 380
methylene chloride 600
trichloroethene 80
Prepared in DMSO, 1mL/ampul
Ea.: cat.# 36007 10-pk.: cat.# 36107
Class 1 Mix
L m / g 2 e n e z n e b
carbon tetrachloride 4
1,2-dichloroethane 5
1,1-dichloroethene 8
1,1,1-trichloroethane 1500
Prepared in water:dimethylsulfoxide 90:10, 1mL/
ampul
Each 5-pk. 10-pk.
36228 36228-510 36328
Class 2 Mix A
chlorobenzene 360g/mL
cyclohexane 3880
cis-1,2-dichloroethene 1870
dichloromethane 600
ethylbenzene 369
0 9 2 e n a x e h
methylcyclohexane 1180
N,N-dimethylformamide 880
0 9 8 e n e u l o t
1,1,2-trichloroethene 80
m 2 0 3 1 e n e l y x -
o 5 9 1 e n e l y x -
p 4 0 3 e n e l y x -
Prepared in water:dimethylsulfoxide 90:10, 1mL/
ampul
Each 5-pk. 10-pk.
36229 36229-510 36329
Class 2 Mix B
L m / g 0 1 4 e l i r t i n o t e c a
0 6 m r o f o r o l h c
1,2-dimethyoxyethane 100
N,N-dimethylacetamide 1090
1,4-dioxane 380
1,2,3,4-tetrahydronaphthalene (tetraline) 100
0 5 e n o n a x e h - 2
0 0 0 3 l o n a h t e m
nitromethane 50
0 0 2 e n i d i r y p
Prepared in water:dimethylsulfoxide 90:10, 1mL/
ampul
Each 5-pk. 10-pk.
36230 36230-510 36330
Class 2 Mix C
2-ethoxyethanol 160g/mL
ethylene glycol 620
0 2 2 e d i m a m r o f
2-methoxyethanol 50
N-methylpyrrolidone 4840
0 6 1 e n a l o f l u s
Prepared in water, 1mL/ampul
Each 5-pk. 10-pk.
36231 36231-510 36331
USP <467> Reference Material Mixes
17
www.restek.com
GC Accessories
. k p - 0 0 0 1 . k p - 0 0 1 n o i t p i r c s e D
6mL Clear Vial 21166 21167
10mL Clear Vial, Flat Bottom 24683 24684
10mL Clear Vial, Rounded Bottom 21164 21165
20mL Clear Vial, Flat Bottom 24685 24686
20mL Clear Vial, Rounded Bottom 21162 21163
27mL Clear Vial 21160 21161
. k p - 0 0 0 1 . k p - 0 0 1 n o i t p i r c s e D
Silver Seal w/ PTFE/Gray Butyl Rubber 21761 21762
Silver Seal w/ PTFE/Silicone 21763 21764
Pressure Release, Silver Seal w/ PTFE/
Gray Butyl Rubber Septa <125 6 6 7 1 2 5 6 7 1 2 C
Pressure Release, Silver Seal w/ PTFE/
Silicone Septa >125 8 6 7 1 2 7 6 7 1 2 C
Headspace Autosampler Vials
20mm Aluminum Seals w/Septa, Assembled
The crimper is adjustable for optimized sealing performance. It also is comfortable enough
for pro- longed use. For chromatographers who need to save, transfer, or dispose of their
samples, we provide a decapper that allows the user to remove a crimp-top cap safely and
easily. If you havent used an aluminum seal decapper, order one today!
Aluminum Seal Crimper and Decapper
r e p p a c e D r e p m i r C e z i S
6 3 7 1 2 5 3 7 1 2 m m 8
1 7 1 1 2 0 7 1 1 2 m m 1 1
0 4 7 1 2 9 3 7 1 2 m m 3 1
8 3 7 1 2 7 3 7 1 2 m m 0 2
Septum Diameter 25-pk. 50-pk. 100-pk.
9.5mm (
3
/8") 20359 20360 20361
0 8 3 0 2 9 7 3 0 2 8 7 3 0 2 m m 0 1
11mm (
7
/16") 20363 20364 20365
12.5mm (
1
/2") 20367 20368 20369
6 8 3 0 2 5 8 3 0 2 4 8 3 0 2 m m 7 1
Shimadzu Plug 20372 20373 20374
Thermolite
Septa
Crimper Decapper
s
18
www.restek.com
Inlet Liners for HP/Finnigan GCs
Liner ID/OD/length ea. 5-pk. 25-pk.
2mm Splitless
2.0 x 6.5 x 78.5 20712 20713 20714
Gooseneck Splitless (2mm)
2.0 x 6.5 x 78.5 20795 20796 20797
Recessed Gooseneck (2mm)
2.0 x 6.5 x 78.5 20980 20981 20982
1mm Split
1.0 x 6.3 x 78.5 20972 20973
Liner ID/OD/length ea. 5-pk. 25-pk.
1mm Split
1.0 x 6.3 x 72 20970 20971
2mm Splitless
2.0 x 6.3 x 74 20721 20722 20723
Open 0.5mm ID
0.5 x 5.0 x 54 20992 20993
Open 0.75mm ID
0.75 x 5.0 x 54 21714 21715 21716
Inlet Liners for Varian GCs
Liner ID/OD/length ea. 5-pk. 25-pk.
17A 1mm Split
1.0 x 5.0 x 94 20976 20977 20978
Inlet Liners for Shimadzu GCs
Inlet Liners for Perkin-Elmer GCs
Liner ID/OD/length ea. 5-pk. 25-pk.
2mm Splitless
2.0 x 5.5 x 79.5 20811 20812 20813
Inlet Liners for CE Instruments/ThermoQuest GCs
Liner ID/OD/length ea. 5-pk. 25-pk.
2mm Splitless
2.0 x 5.0 x 100 20730 20731 20732
Auto SYS Splitless w/Wool (2mm)
2.0 x 6.2 x 92.1 20829 20830 20831
Auto SYS XL Split/Splitless w/ wool
2.0 x 4.0 x 81.2 21717 21718
for 5000 and 6000 GCs
1mm Split
1.0 x 8.0 x 105 20916 20917
Liner ID/OD/length ea. 5-pk. 25-pk.
for TRACE and 8000 GCs
Fits Vespel
/
Ferrule Column Graphite Graphite
ID (mm) ID (mm) 50-pk. 50-pk.
0.4 0.25 20227 20229
0.5 0.32 20228 20231
0.8 0.53 20224 20230
Capillary Ferrules
(for
1
/16" compression-type fittings)
Fits Vespel
/
Ferrule Column Graphite Graphite
ID (mm) ID (mm) 5-pk. 5-pk.
0.4 0.25 20235 20241
0.5 0.32 20235 20242
0.8 0.53 20245 20246
Two-Hole Ferrules
(for
1
/16" compression-type fittings)
Fits
Ferrule Column Graphite Graphite
ID (mm) ID (mm) 2-pk. 10-pk.
0.4 0.180.25 20280 20281
0.5 0.32 20282 20283
0.8 0.50 & 0.53 20284 20285
Graphite Ferrules for
M4 Fittings
(for QCQ Fisons 8000 & TRACE 2000)
19
Restek has specially engineered a high-precision,
1
/16-inch fitting that uses standard size, two-hole
capillary ferrules. The fitting kit comes with everything needed for dual-column confirmational analysis
using 0.25 and 0.32mm ID capillary columns (two-hole ferrules must be ordered separately).
Capillary Inlet Adaptor Fitting Kit (for 0.25/0.32mm ID columns): cat.# 20633
Replacement Inlet Seal (1.2mm hole): cat.# 20390, (2-pk.); cat.# 20391, (10-pk.)
Hewlett-Packard
1
/16-Inch Capillary Inlet Adaptor Fitting Kit
Restek has specially engineered a high-precision,
1
/8-inch fitting that uses standard
1
/8-inch, two-hole capillary ferrules. The fitting kit comes with everything needed for installation.
1
/8-inch Capillary Inlet Adaptor Fitting Kit (for 0.53mm ID columns): cat.# 20645
Replacement Inlet Seal (
1
/16-inch hole): cat.# 20392, (2-pk.); cat.# 20393, (10-pk.)
Hewlett-Packard
1
/8-Inch Capillary Inlet Adaptor Fitting Kit
Includes a
1
/16-inch nut, a
1
/16-inch ferrule, a base nut and
1
/4-inch Vespel
/graphite ferrule, a
1
/16-inch
capillary nut, a 5-pack of low-bleed plug septa, and a special low-mass septum nut. Order appropriate
capillary ferrules separately.
Low-Volume Injector for Hewlett-Packard and Varian GCs
Description Kit
LVI for HP Split/Splitless GC inlets cat.# 21692
LVI for Varian Split/Splitless GC inlets cat.# 21693
Includes a
1
/16-inch nut, a
1
/16-inch ferrule, a base nut and
1
/4-inch Vespel
/graphite ferrule, a
1
/16-inch
capillary nut, a 5-pack of low-bleed plug septa, and a special low-mass septa nut. Order appropriate
capillary ferrules separately.
Low-Volume Injector for Hewlett-Packard 5890 Septum Packed Purge Port
Description Kit
LVI for HP 5890 Septum Packed Purge Port cat.# 21698
Calculates linear velocity based on column ID.
Measures N2, He, H2, 5% Ar/Me, and Air.
Measures split flow and mass flow.
Has pulse-free operation that will not interfere with EPCs.
Reads flow accurately from 5 to 500mL/min.
Description Each
Restek Veri-Flow 500 Electronic Flowmeter (110 volts) cat.# 21643
Restek Veri-Flow 500 Electronic Flowmeter (220 volts) cat.# 21645
Restek Veri-Flow 500 Electronic Flowmeter
The Leak Detective
responds in less than 2 seconds to trace leaks of gases with thermal conductivities
different than air. Helium or hydrogen can be detected at 3 x 10-4 cc/sec.* or at an absolute concentra-
tion as low as 200ppm. Leaks are indicated by an audible alarm, as well as by an LED readout. (Batteries
and AC adaptor included.)
*Caution: not designed for determining leaks of combustible gases.
Description Each
Restek Leak Detective