Metodo FIAS Enva100
Metodo FIAS Enva100
n
x)
n
s
p
=
i=1
n
x
2
- (
i=1
n
x)
2
n
n-1
Where:
Ave. %R: Average percent recovery
s
p
: Standard deviation of percent recoveries
n: Number of samples
x: Individual percent recovery
9.2.4 Quality control check sample (QCS): When beginning the use of this method, on a quarterly
basis, after the preparation of stock or calibration standard solutions or as required to meet
data quality needs, verify the calibration standards and acceptable instrument performance
with the preparation and analysis of a QCS (Sect. 7.13). To verify the calibration the QCS
must be within + 10% of the stated value. If the calibration standard cannot be verified,
performance of the determinative step of the analysis is unacceptable. The source of the
problem must be identified and corrected before either proceeding on with the initial
determination of method detection limits or continuing with on-going analyses.
9.2.5 Method detection limits (MDL): A mercury MDL must be established using an LRB
solution fortified at a concentration of two to three times the estimated detection limit. To
determine MDL values, take seven replicate aliquots of the fortified LRB and process
through the entire analytical method. Perform all the calculations defined in the method and
report the concentration values in the appropriate units. Calculate the MDL as follows
9
:
9
MDL = (t) (s)
Where:
t : Students t value for a 99% confidence level and a standard deviation
estimate with n-1 degrees of freedom [t = 3.14 for seven replicates]
s : Standard deviation of the replicate analyses.
Note: If the relative standard deviation (RSD) from the analyses of the seven aliquots is <
10%, the concentration used to determine the mercury MDL may have been inappropriately
high for the determination. If so, this could result in the calculation of an unrealistically low
MDL. Concurrently, determination of MDL in a LRB solution represents a best-case
situation and does not reflect possible matrix effects of real world samples. However,
successful analyses of LFMs (Sect. 9.4) can give confidence to the MDL value determined in
LRB solution.
The MDL must be sufficient to detect mercury at the required level according to compliance
monitoring regulation (Sect. 1.2). The mercury MDL should be determined annually, when
a new operator begins work or whenever, in the judgment of the analyst, a change in
analytical performance caused by either a change in instrument hardware or operating
conditions would dictate they be redetermined.
9.3 Assessing Laboratory Performance (mandatory).
9.3.1 Laboratory reagent blank (LRB): The laboratory must analyze at least one LRB (7.11.2)
with each batch of 20 or fewer samples of the same matrix. LRB data are used to assess
contamination from the laboratory environment. LRB values that exceed the MDL indicate
laboratory or reagent contamination should be suspected. When an LRB value is equal to
10% or more of the analyte level determined for a sample or is 2.2 times the analyte MDL,
whichever is greater, fresh aliquots of the samples must be prepared and analyzed again for
the affected analytes after the source of contamination has been corrected and acceptable
LRB values have been obtained.
9.3.2 Laboratory fortified blank (LFB): The laboratory must analyze at least one LFB (7.11.3)
with each batch of samples. Calculate accuracy as percent recovery using the following
equation:
R = (LFB-LRB) (100)
C
Where:
R: % Recovery
LFB: Mercury concentration observed in the laboratory fortified blank
LRB: Mercury concentration observed in the laboratory reagent blank
C: Mercury concentration added to the LFB
In drinking water monitoring, the control limits for the percent recovery of mercury from a
laboratory reagent blank are 85-115%. The aforementioned control limits for the percent
recovery of mercury from a laboratory reagent blank are also suggested for use in
wastewater monitoring until the 1994 version of EPA Method 245.1 or other limits are
promulgated in wastewater regulations.
10
9.3.3 The laboratory must use LFB analysis data to assess laboratory performance against the
required control limits of 85-115% (Sect. 9.3.2). When sufficient internal performance data
becomes available (usually a minimum of twenty to thirty analyses), optional control limits
can be developed from the mean percent recovery (x) and the standard deviation (s
p
) of the
mean percent recovery. These data can be used to establish the upper and lower control
limits as follows:
Upper Control Limit = x + 3s
p
Lower Control Limit = x - 3s
p
The optional control limits must be equal to or better than the required control limits of 85-
115%. After each five to ten new recovery measurements, new control limits can be
calculated using only the most recent twenty to thirty data points. Also, the standard
deviation (s
p
) data should be used to establish an on-going precision statement for the level
of concentrations included in the LFB. These data must be kept on file and be available for
review.
9.3.4 Instrument performance check (IPC) solution: For all determinations the laboratory must
analyze the IPC solution (7.12) and a calibration blank immediately following each
calibration, after every tenth sample (or more frequently, if required) and at the end of the
sample run. Analysis of the calibration blank should always be < the MDL. Analysis of the
IPC solution immediately following calibration must verify that the instrument is within +
5% of the calibration. Subsequent analyses of the IPC solution must be within + 10% of
calibration. If the calibration cannot be verified within the specified limits, analysis must be
discontinued, the cause determined and/or in the case of drift the instrument recalibrated.
All samples following the last acceptable IPC solution must be reanalyzed. The analysis
data of the calibration blank and IPC solution must be kept on file with the sample analyses
data.
9.4 Assessing Analyte Recovery and Data Quality
9.4.1 Sample homogeneity and the chemical nature of the sample matrix can affect mercury
recovery and the quality of the data. Taking separate aliquots from the sample for replicate
and fortified analyses can in some cases assess the effect. Unless otherwise specified by the
data user, laboratory or program, the following laboratory fortified matrix (LFM) procedure
(9.4.2) is required.
9.4.2 The laboratory must add a known amount of mercury to a minimum of 10% of the samples
or one sample per sample set, whichever is greater. In each case the LFM aliquot must be a
duplicate of the aliquot used for sample analysis. Select a sample with a low mercury
background that is representative of the type of water samples being analyzed. It is
recommended that this sample be analyzed prior to fortification. The concentration of
mercury added may vary based on the nature of the samples being analyzed. When possible,
the concentration should be the same as that added to the LRB, but should not exceed the
midpoint concentration of the calibration curve. It is suggested that one QC sample should
contain an organic mercury spike to monitor the effectiveness of the sample digestion. Over
time, samples from all routine sample sources should be fortified.
9.4.3 Calculate the percent recovery, corrected for background concentration measured in the
unfortified sample aliquot, and compare these values to the control limits to the designated
LFM recovery range of 70-130% for drinking water samples. Percent recovery may be
calculated using the following equation:
R = (LFM - LSM) (100)
C
11
Where:
R: % Recovery
LFM: Mercury concentration observed in the laboratory fortified matrix sample
LSM: Mercury concentration observed in the laboratory sample matrix
C: Mercury concentration added to the LFM
The aforementioned control limits for the percent recovery of mercury from a fortified
sample matrix are also suggested for use in wastewater monitoring until the 1994 version of
EPA Method 245.1 or other limits are promulgated in wastewater regulations.
9.4.4 If mercury recovery falls outside the designated range, and the laboratory performance is
shown to be in control (9.3), the sample recovery problem encountered with the fortified
drinking water or wastewater sample is judged to be matrix related, not system related. The
result for mercury in the unfortified sample must be labeled to inform the data user that the
results are suspect due to matrix effects.
9.4.5 Compute the relative percent difference (RPD) between two duplicate spiked sample results
using the following formula:
RPD = 100(LD
2
- LD
1
)
(LD
1
+ LD
2
)/2
Where:
RPD: Relative percent difference
LD
1
: The first duplicate spiked sample result
LD
2
: The second spiked sample (duplicate) result
10.0 Calibration and Standardization
10.1 Set up FIAS (used in conjunction with an atomic absorption spectrophotometer) or FIMS stand-alone
system.
10.1.1 Assemble the Flow Injection System as shown in Figure 2 or 3. Set up the operating
conditions as specified in Table III. For more detailed information refer to instrument manual
provided with the Flow Injection system. Turn on the autosampler and atomic absorption
spectrophotometer. All instrument parameters recommended by the manufacturer should be
used.
10.1.2 Align the FIAS quartz cell in the optical path.
10.1.3 Allow the system to warm up for 30 minutes.
10.1.4 Feed all reagents through the system (see Figures 1,2, and 3). Refer to the manual supplied
with the Flow Injection for more detailed information.
10.2 Mercury calibration standards: Transfer 0, 0.5, 1.0, 5.0, 10.0, and 20.0 mL aliquots of the
intermediate stock mercury solution containing 0.1 g Hg (7.6) to a series of sample containers containing
40 mL of reagent water. Add enough reagent water to make a total volume of 100 mL and mix
thoroughly. Prepare calibration standards daily. Process as described in Sections 11.1.2 and 11.1.3.
10.3 Cool the solution and add 6 mL of sodium chloride-hydroxylamine hydrochloride solution (7.6) to
each sample container to reduce the excess permanganate. When the solution becomes decolorized
transfer to the autosampler
12
10.4 Prepare the calibration curve by plotting peak absorbance of the known standards versus concentration.
The standard curve must comply with Sect. 9.2.2. Figure 5 shows a typical FIAS calibration curve and
Figure 6 shows a typical FIMS calibration curve.
11.0 Procedure
11.1 Sample Digestion
11.1.1 Transfer 100 mL of sample [or an aliquot of the sample diluted with reagent water (7.2) to
100 mL], into a sample digestion bottle (6.5.2).
11.1.2 Add 5 mL of sulfuric acid (7.2) and 2.5 mL of nitric acid (7.3) to the container.
11.1.3 Next add 15 mL of potassium permanganate solution (7.7) to each sample. Sewage samples
and samples containing high salts may require additional permanganate. If necessary, add 15
mL portions of the potassium permanganate solution until the purple color persists for at
least 15 minutes. Be sure to mix sample after each addition. The same amount of potassium
permanganate solution should be added to the standards and blanks as that added to the
samples.
Add 8 mL of potassium persulfate (7.8) to each bottle, lay the stopper across the neck, and
heat for 2 hours in a covered water bath at 95
o
C.
11.1.4 Remove the sample containers from the water bath and cool to room temperature.
11.1.5 When the samples are at room temperature. to each container add 6 mL of sodium chloride-
hydroxylamine hydrochloride (7.9) to reduce the excess permanganate. For samples
containing high chlorides, continue to add the sodium chloride-hydroxylamine hydrochloride
(up to 25 mLs) until the potassium permanganate is completely reduced. Be certain to mix
the sample after each addition.
11.1.6 Transfer the samples to an autosampler.
11.2 Sample Analysis
11.2.1 Before beginning daily calibration, the instrument should be reconfigured to the optimized
conditions. Allow the system to stabilize.
11.2.2 Following calibration, the digested samples are analyzed in the same manner as the standard
solutions.
11.2.3 At the completion of the analysis all tubing should be flushed with deionized water and then
air. Release the pump tension on all tubing.
11.2.4 During the analysis of samples, the laboratory must comply with the required quality control
described in Sections 9.3 and 9.4.
12.0 Data Analysis and Calculations
12.1 From the prepared calibration curve (10.5) compute sample concentration by comparing response with
the standard curve. The FIAS and FIMS systems will report the sample concentration directly.
12.2 The concentration of mercury in the sample is that derived from either a calibration graph or read
directly from the instrument and must be corrected for any dilution performed before the digestion.
Thus the observed concentration is multiplied by the dilution factor as follows:
13
Dilution Factor = T/A
Where:
A: Volume of aliquot taken for dilution (mL)
T: Total volume to which the sample aliquot is diluted (mL)
12.3 Report mercury concentrations to the proper significant figures in mg/L, g/L, or ng/L as required.
13.0 Method Performance
13.1 The performance of the method in a single laboratory is shown in Table IV for drinking and wastewater
spiked with organic or inorganic mercury. Table V shows the recoveries of mercury spikes in drinking
water, groundwater, and an National Institute of Standards and Technology (NIST) reference material.
14.0 Pollution Prevention
14.1 Pollution prevention encompasses any technique that reduces or eliminates the quantity or toxicity of
waste at the point of generation. Numerous opportunities for pollution prevention exist in laboratory
operation. The EPA has established a preferred hierarchy of environmental management techniques
that places pollution prevention as the management option of first choice. Whenever feasible,
laboratory personnel should use pollution prevention techniques to address their waste generation.
When wastes cannot be feasibly reduced at the source, the Agency recommends recycling as the next
best option.
Flow injection techniques reduce the volume of waste by requiring smaller amounts of reagents during
the determinative step, because of the smaller volume of sample used. If it is possible to reduce the
volume of sample digested, the waste generated will be further reduced.
14.2 For information about pollution prevention that may be applicable to laboratories and research
institutions, consult Less is Better: Laboratory Chemical Management for Waste Reduction, available
from the American Chemical Societys Department of Government Relations and Science Policy, 115
16th Street N.W., Washington, D.C. 20036, (202) 872-4477.
15.0 Waste Management
15.1 The Environmental Protection Agency requires that laboratory waste management practices be
conducted consistent with all applicable rules and regulations. The Agency urges laboratories to protect
the air, water, and land by minimizing and controlling all releases from hood and bench operations,
complying with the letter and spirit of any sewer discharge permits and regulations, particularly the
hazardous waste identification rules and land disposal restrictions. For further information on waste
management consult The Waste Management Manual for Laboratory Personnel, available from the
American Chemical Society at the address listed in Section 14.2.
16.0 References
16.1. EPA Method 245.1, In: Methods for the Determination of Metals in Environmental Samples,
Supplement 1, May 1994, EPA/600/R-94/111, Cincinnati, OH.
16.2 Perkin-Elmer FIAS Installation and System Description, Part Number B050-8399 ,The Perkin-Elmer
Corporation, 761 Main Ave. Norwalk, CT, 06859 (1992).
16.3 Perkin-Elmer FIAS Setting Up and Performing Analyses, Part Number B050-8400, The Perkin-Elmer
Corporation, 761 Main Ave., Norwalk, CT 06859 (1992).
16.4 Perkin-Elmer FIAS Recommended Analytical Conditions and General Information, Part Number B050-
1820, The Perkin-Elmer Corporation, 761 Main Ave, Norwalk, CT 06859 (1994).
14
16.5 Perkin-Elmer FIMS Installation, Maintenance, and System Description, Part Number 0993-5202, The
Perkin-Elmer Corporation, 761 Main Ave, Norwalk, CT 06859 (1994).
16.6 Perkin-Elmer FIMS Software Guide, Part Number 0993-5215, The Perkin-Elmer Corporation, 761 Main
Ave, Norwalk, CT 06859 (1994).
16.7 Perkin-Elmer FIMS Setting Up and Performing Analyses, Part Number 0993-5203, The Perkin-Elmer
Corporation, 761 Main Ave, Norwalk, CT 06859 (1994).
16.8 Specification for Reagent Water, D1193, Annual Book of ASTM Standards, Vol. 11.01 (1990).
16.9 Code of Federal Regulations 40, Pt. 136, Appendix B (July 1, 1996).
17.0 Tables and Figures
Table I
Comparison of Method Detection Limits
Instrument Configuration Method Detection Limit (MDL)
Flow Injection Analysis System (FIAS)/Hollow
Cathode Lamp (HCL)
0.2 g/L
Flow Injection Analysis System
(FIAS)/Electrodeless Discharge Lamp (EDL)
0.06 g/L
Flow Injection Mercury System (FIMS) 0.01 g/L
Table II
Sample Volume Effects on FIMS Linearity and
Instrument Detection Limit (IDL)
Sample Loop
( L)
IDL (ppt) Linear
Range ( g/L)
500 6 20
200 12 30
100 25 50
Table III
Recommended FIAS Operating Conditions
Wavelength 253.7 nm
Slit 0.7 nm
Cell Temperature 100
o
C
Sample Volume 500 L
Notes:
Argon carrier gas flow = 60-100 mL/min
Electrodeless Discharge lamp (EDL) used
3% HCl carrier flow rate = 8 mL/min
1.1% SnCl
2
reductant flow rate = 4 mL/min
15
Table IV
Precision and Recovery for Mercury-Spiked Waters using the FIAS
Sample Hg Spike
( g/L)
Mean
( g/L)
SD % RSD % Recovery
Wastewater #1 1.00 0.96 0.04 4.3 96
Wastewater #2 1.00 1.04 0.06 5.4 104
Drinking Water #1 1.00 0.99 0.04 4.4 99
Drinking Water #2 1.00 1.08 0.05 4.4 108
Notes:
SD: Standard deviation
% RSD where N = 20
Spiking solutions:
For samples #1 = 1.00 g/L Hg (HgCl
2
)
For samples #2 = 1.00 g/L Hg (2-Ethylmercurimercaptobenzoic acid, sodium salt)
Table V
Recovery from Inorganic Mercury-Spiked Water Samples Using the FIMS
Sample Hg Spike
( g/L)
Mean ( g/L) SD % RSD % Rec
Ground water 1.4 1.41 0.003 0.2 101
Drinking water 1.4 1.38 0.02 1.4 99
NIST 1641c diluted to
0.2 g/L
-- 0.184 0.004 2.2 92
Wastewater 1.4 1.36 0.006 0.4 97
Notes:
SD: Standard Deviation
% RSD where N = 3
HCl
Ar Waste
Absorption Cell
Sample
SnCl
2
Figure 1. General schematic of flow injection.
16
Figure 2. Tubing set up for Flow Injection systems with two pumps.
Figure 3. Tubing set up for Flow Injection Systems with one pump.
17
Figure 4. Mixing Manifold and Gas/Liquid Separator
Figure 5. Typical FIAS calibration curve using 500 L of sample.
Figure 6. Typical FIMS calibration curve using 500 L of sample.
Abs.
0
0.02
0.04
0.06
0.08
0.1
0.12
0 2 4 6 8 10
Concentration ( g/L)
A
b
s
o
r
b
a
n
c
e
0.00
0.05
0.10
0.15
0.20
0 5 10 15 20
Hg Concentration ( g/L)
A
b
s
o
r
b
a
n
c
e
Appendix 1
Appendix 2
Look to us. And see more
Perkin Elmer
761 Main Avenue
Norwalk, CT 06859-0010 USA
Tel: 80O-762-4000 or
(1+) 203-762-4000
Fax: (1+) 203-762-4228
ENVA-100
KG89905
Perkin Elmer is a registered trademark of affiliates of Perkin Elmer LLC.
FIMS is a trademark of Perkin Elmer
1999 Perkin Elmer LLC Printed in U.S.A.