A Practical Approach To Validation of HPLC Methods Under Current Good Manufacturing Practices - 0
A Practical Approach To Validation of HPLC Methods Under Current Good Manufacturing Practices - 0
Introduction
Regulatory analytical procedures are of two types: comAnalytical methods validation is an important regulatory
pendial and noncompendial. The noncompendial analytical
requirement in pharmaceutical analysis. High-Performance
procedures in the USP are those legally recognized as reguLiquid Chromatography (HPLC) is commonly used as an
latory procedures under section 501(b) of the Federal Food,
analytical technique in developing and validating assay
Drug and Cosmetic Act. When using USP analytical methmethods for drug products and drug substances. Method valods, the guidance recommends that information be provided
idation provides documented evidence, and a high degree of
for the following characteristics: specificity of the method,
assurance, that an analytical method employed for a specific
stability of the analytical sample solution, and intermediate
test, is suitable for its intended use. Over recent years, reguprecision. Compendial analytical methods may not be stabillatory authorities have become increasingly aware of the neity indicating, and this concern must be addressed when decessity of ensuring that the data submitted to them in appliveloping a drug product specification, because formulation
cations for marketing authorizations have been acquired
based interference may not be considered in the monograph
using validated analytical methodology. The International
specifications. Additional analytical tests for impurities may
Conference on Harmonization (ICH) has introduced guidebe necessary to support the quality of the drug substance or
1,2
lines for analytical methods validation. The U.S. Food and
drug product. Noncompendial analytical methods must be
Drug Administration (FDA) methods validation draft guidfully validated. The most widely applied validation charac3-5
ance document, as well as United States Pharmacopoeia
teristics are accuracy, precision (repeatability and intermedi6
(USP) both refer to ICH guidelines.
ate precision), specificity, detection limit,
These draft guidances define regulatory
quantitation limit, linearity, range, and staand alternative analytical procedures and Figure 1
bility of analytical solutions.
____________________
stability-indicating assays. The FDA has
The parameters that require validation
The chemical structure of
proposed adding section CFR 211.222 on
and the approach adopted for each particethyl 4-hydroxybenzoate.
analytical methods validation to the curular case are dependent on the type and
rent Good Manufacturing Practice
applications of the method. Before under7
(cGMP) regulations. This would require
taking the task of method validation, it is
pharmaceutical manufacturers to estabnecessary that the analytical system itself
lish and document the accuracy, sensitivis adequately designed, maintained, cali8
ity, specificity, reproducibility, and any
brated, and validated. The first step in
other attribute (e.g., system suitability,
method validation is to prepare a protostability of solutions) necessary to valicol, preferably written with the instrucdate test methods.
tions in a clear step-by-step format. This
29
Ghulam A. Shabir
Figure
2
______________________________________________
Graph measured peak area versus ethyl 4-hydroxybenzoate concentration demonstrating linearity.
Experimental
Chemicals and reagents.
All chemicals and reagents were of the
highest purity. HPLC-grade acetonitrile was
obtained from Merck (Darmstadt, Germany). Water was purified with a Millipore
Milli-Q system (Watford, UK). Ethyl 4-hydroxybenzoate (Batch #1005425) was supplied by Lancaster Synthesis (Morecambe, England).
HPLC instrumentation.
The HPLC system used for the validation studies consisted of a Waters Alliance 2690 Separations Module to a
996 photodiode-array (PDA) detector. The control of the
HPLC system and data collection was by a Compaq computer equipped with Waters Millennium32 software (version 3.20). The second HPLC system used for intermediate
precision studies consisted of Perkin Elmer: model series
200 UV visible detector, series 200 LC pump, series 200 autosampler, and a series 200 peltier LC column oven were
used to chromatograph the solutions. The data was acquired
via PE TotalChrom Workstation data acquisition software,
(Version 6.2.0) using PE Nelson series 600 LINK interfaces.
Both HPLC systems including software (Food and Drug Administration (FDA), 21 Code of Federal Regulations (CFR)
Part 11) were validated prior to use for the test method validation.
All chromatographic experiments were performed in the
isocratic mode. A C18 symmetry analytical column from
Waters (located in Milford, MA, United States) 3.9 x 150
mm, 5 mm particle size was used. The mobile phase consisted of a mixture of acetonitrile, water solution (65:35,
v/v). The flow rate was set to 1.0 ml/min, and the oven temperature to 25C. The injection volume was 20 l, and the
detection wavelength was set at 254nm.
30
I n s t i t u t e o f Va l i d a t i o n Te c h n o l o g y
Ghulam A. Shabir
Figure
3
______________________________________________
31
Ghulam A. Shabir
Method validation
provides documented
evidence, and a high
degree of assurance, that
an analytical method
employed for a specific
test, is suitable for its
intended use.
Figure
5
________________________________________________________________
HPLC chromatogram of ethyl 4-hydroxybenzoate.
32
I n s t i t u t e o f Va l i d a t i o n Te c h n o l o g y
Ghulam A. Shabir
Figure
6
________________________________________________________________
HPLC chromatogram for placebo. The analyte peak was eluted at 1.58 minutes.
Precision
Precision is the measure of the degree
of repeatability of an analytical method
under normal operation, and is normally
expressed as the percent relative standard
deviation for a statistically significant
number of samples. Precision may be performed at three different levels: repeatability, intermediate precision, and reproducibility.
Figure
7
______________________________________________
Demonstration of the repeatability of the HPLC assay for
ethyl 4-hydroxybenzoate as shown by the results of 10
replicate injections of one solution at 100 percent of the
test (20 mg/ml) concentration
Injection
number
RT (min)
Peak height
(V)
Peak area
(V s)
1
2
3
4
5
6
7
8
9
10
Mean
RSD (%)
1.57
1.58
1.58
1.58
1.57
1.57
1.57
1.58
1.58
1.58
1.58
0.18
855847
858249
854532
856705
857058
854755
855098
854078
856416
849916
855265
0.25
3109735
3100787
3099540
3103544
3101464
3099731
3102575
3103159
3104217
3091891
3101665
0.14
Repeatability
Repeatability (intra-day assay precision) is the results of
the method operating over a short time interval under the
same conditions (intra-assay precision). It should be determined from a minimum of nine determinations covering the
specified range of the procedure (for example, three levels,
three repetitions each), or from a minimum of six determinations at 100% of the test or target concentration. A precision criterion for an assay method is that the instrument precision (RSD) will be 1%, and for the impurity assay, at the
limit of quantitation, the instrument precision (repeatability)
will be 5%. Documentation in support of precision studies
should include the standard deviation, relative standard devi-
33
Ghulam A. Shabir
Figure
8
_____________________________________________________________
Demonstration of the intermediate precision of the HPLC assay for ethyl
4-hydroxybenzoate results in relative percent purity area
HPLC system 1
Sample
S1
(50%)
S2
(100%)
S3
(150%)
S1
(50%)
S2
(100%)
S3
(150%)
Operator 1, day 1
Operator 1, day 2
Operator 2, day 1
Operator 2, day 2
Mean
(HPLC systems)
Mean (Operators)
RSD (criteria 2%)
HPLC systems
Operators
99.83
99.76
99.71
99.53
99.71
99.79
99.74
99.76
99.62
99.73
99.76
99.74
99.77
99.57
99.71
99.76
99.82
99.75
99.79
99.78
99.83
99.80
99.76
99.81
99.80
99.83
99.78
99.69
99.82
99.78
99.79
S1+S1
0.05
0.06
99.79
S2+S2
0.05
0.04
99.78
S3+S3
0.05
0.05
99.70
99.74
99.71
Intermediate Precision
Intermediate precision (inter-day variation) is the results
from within lab variations, due to random events, such as different days, analysts, equipment, etc. In determining intermediate precision, experimental design should be employed, so
that the effects (if any) of the individual variables can be monitored. Precision criteria for an assay method is that the intraassay precision will be 2%, and for impurity assay, at the
limit of quantitation, the instrument precision will be 5%,
and the intra-assay precision will be 10%. In this study, intermediate precision (within-laboratory variation) was
demonstrated by two operators, using two HPLC systems,
and evaluating the relative percent purity data across the two
HPLC systems at three concentration levels (50%, 100%,
150%) that cover the ethyl 4-hydroxybenzoate assay method
range (5.0-40 g/ml). The mean and RSD across the systems
and analysts were calculated from the individual relative per34
HPLC system 2
I n s t i t u t e o f Va l i d a t i o n Te c h n o l o g y
cent purity mean values at 50, 100, and 150% of the test concentration. The RSD values presented in Figure 8 were less
than 1% for both systems and operators, and illustrated the
good precision of the analytical method.
Reproducibility
Reproducibility1 is determined by testing homogeneous
samples in multiple laboratories, often as part of inter-laboratory crossover studies. An example of reproducibility criteria
for an assay method could be that the assay results obtained
in multiple laboratories will be statistically equivalent, or the
mean results will be within 2% of the value obtained by the
primary testing lab. For an impurity method, results obtained
in multiple laboratories will be statistically equivalent, or the
mean results will be within 10% (relative) of the value obtained by the primary testing lab for impurities. Reproducibility is not normally expected if intermediate precision
is performed.
Ghulam A. Shabir
Figure
9
________________________________________________________________
HPLC chromatogram for limit of detection (2 ng/ml)
Figure
10
________________________________________________________________
HPLC chromatogram for limit of quantitation (5 ng/ml).
Figure
11
________________________________________________________________________
Stability of ethyl 4-hydroxybenzoate in solution (n = 6)
Time
(hour)
RT (min)
Peak area
RSD (%)
Peak Height
RSD (%)
Percent
recovery
0
24
48
0.14
0.18
0.29
0.70
0.15
0.30
0.80
0.27
0.51
99.88
99.82
99.78
Percent of initial
99.35
35
Ghulam A. Shabir
Figure
12
_______________________________________________________
Demonstration of the system suitability of the HPLC assay for ethyl
4-hydroxybenzoate
Results
System Suitability
Parameter
Acceptance
Criteria
RSD 1%
0.15
0.11
RSD 1%
0.18
0.11
T2
1.05
1.03
N = > 2000
5276
6628
System Suitability
36
I n s t i t u t e o f Va l i d a t i o n Te c h n o l o g y
Ghulam A. Shabir
Conclusion
It is clear from the various guidelines issued by regulatory authorities that analytical methodology should be thoroughly validated under Current Good Manufacturing Practice (cGMP). HPLC assay of active ingredients in pharmaceutical products, and subsequent method validation, can be
complex and time-consuming. However, a well-defined protocol and documented validation plan simplifies and shortens the process, while also providing regulatory agencies
with evidence that the analytical system and method is suitable for its intended use. This paper is intended to provide
guidance on how to perform method validation for HPLC
that generates both useful and meaningful data that meets all
FDA, USP, and ICH validation requirements for pharmaceutical analysis.
Acknowledgements
References
I thank Abbott Laboratories and MediSense for permission to publish this article. I would also thank to Dr Nigel
Forrow for his comments on the text.
Originally published in the May, 2004 issue of the Journal of Validation Technology
37