AFM Manual
AFM Manual
Instruction Manual
Version 4.31ce
Date
Ref.
DCR
Section(s) Affected
4.31ce
27OCT97
Chapters 3, 5 and 8
168, 185,
189
4.22ce
14FEB97
139
4.22
15JUL96
Released
Approval
VERSION 4.31ce27OCT97
Chapter 2
Chapter 2 Cont...
Chapter 3
Chapter 4
Chapter 5
ii
Chapter 5 Cont...
Chapter 6
Contact AFM
Introduction to Contact AFM 6-1
Preparation Prior to Imaging 6-1
Adjust the Detector Offsets 6-1
Signal settings 6-3
Adjust tip height above sample surface 6-4
iii
Chapter 7
7-1
7-2
7-3
7-5
Chapter 8
TappingMode AFM
Introduction to TappingMode AFM 8-1
Basic Principle of TappingMode 8-1
Preparation Prior to Imaging 8-3
Switch to TappingMode and check parameters 8-3
Adjust laser and photodetector 8-4
Additional preparations 8-7
Tune the cantilever 8-8
Setting the Drive amplitude and Setpoint 8-11
iv
Chapter 8 Cont...
Chapter 9
Chapter 10
vi
Chapter 11
Force Imaging
Introduction 11-1
Plotting Force in Contact AFM 11-1
The Force Curve and Piezo Extension-Retraction Cycle 11-4
vii
Chapter 12
Chapter 13
viii
Chapter 14
14-30
Chapter 15
ix
Chapter 15 Cont...
Chapter 15 Cont...
Index
xi
xii
CAUTION: Some early-model AFM tipholders may short out power supplies when used with
MM-SPMs. If you suspect your tipholder is from an older AFM, check with Digital
Instruments or your international representative before using.
2-9
ATTENTION: Certains modles de support de bras de leviers pour AFM de contact peuvent
endommager les alimentations lectriques quand ils sont utiliss avec un MultiMode (MMSPMs). En cas de doute, contacter Digital Instruments ou son reprsentant lgal pour verification.
2-9
WARNHINWEIS: Einige ltere Ausfhrungen des AFM-Spitzenhalters knnen einen
Kurzschlu verursachen falls sie mit einem MultiMode-AFM verwendet
werden. Bitte wenden Sie sich im Zweifelsfall an Digital Instruments bzw. an die fr Sie
zustndige DI-Vertretung.
2-9
WARNING! During and prior to set up of the laser, it is especially important to avoid looking
directly at the laser beam or at the laser spot. The laser head should never be plugged into the
microscope control electronics unless the head is installed in the Z-stage mount. Care should
be taken when highly reflective samples are inserted onto the chuck. Avoid looking at all
reflected laser light. Operators should use care to avoid staring into beams that may be
reflected from sample surfaces.
5-2
TOW-1
TOW-2
TOW-3
8-29
TOW-5
CAUTION: Because TappingMode cantilevers are relatively stiff, Force Mode can
potentially damage the tip and/or surface. DI recommends using Force Plot only with
extreme caution.
11-16
ATTENTION: Le pointes Tapping tant relativement raides, lutilisation du mode
Force peut endommager la pointe ou lchantillon. Digital Instruments recommande
dutiliser le mode Force Plot avec une extrme prcaution.
11-16
WARNHINWEIS: Da TappingMode-Mespitzen relativ steif sind, kann im Force
Mode unter Umstnden die Mespitze oder die Oberflche beschdigt werden. DI
empfiehlt, den Force Plot nur unter grter Vorsicht anzuwenden.
11-16
CAUTION: If employing stiffer cantilevers than those normally used, the sample may
be damaged by the cantilever tip. To minimize this damage, make the Setpoint as low
as possible.
11-34
ATTENTION! Lutilisation dun levier plus rigide que ceux utiliss habituellement
peut entraner un endommagement de lchantillon par la pointe. Afin de rduire les
risques dendommagement, ajuster la valeur cible du rtro-contrle (Setpoint) sa
valeur la plus faible possible.
11-35
WARNHINWEIS! Falls Sie mit Cantilevern hherer Federkonstanten arbeiten, als
normal blich, kann die Probe durch die Spitze beschdigt werden.. Um eine evtl.
Beschdigung zu minimieren bzw. zu vermeiden, benutzen Sie bitte einen
Arbeitspunkt (Setpoint), der nur geringfgig ber der freien Deflektion des Cantilevers
liegt. Die Differenz zwischen Setpoint im Feedback-Men und der freien Deflektion
des Cantilevers (Photodetektorsignal A-B) sollte minimal sein.
11-35
CAUTION: Before enabling the Interleave Drive Amplitude, check that its value is not
much larger than the main Drive Amplitude value to prevent possible damage to the tip.
12-6
ATTENTION: Lors dun travail en mode intercal (Interleave Mode), vrifier que la
tension applique pour osciller le bras de levier lorsquil est en positon haute nest pas
beaucoup plus importante que la tension applique lorsque ce mme bras de levier est
TOW-6
en position basse (Main Drive Amplitude). Le non respect de cette procdure peut
entraner la destruction de la pointe.
12-6
WARNHINWEIS: Bevor Sie im Interleave-Mode die Interleave Drive-Amplitude
einschalten, vergewissern Sie sich bitte, da der dort eingetragene Wert nicht
wesentlich grer ist, als der Wert der Main Drive-Amplitude, um evtl.
Beschdigungen der Spitze vorzubeugen.
12-6
CAUTION: Before enabling the Interleave Drive Amplitude, check that its value is
not much larger than the main Drive Amplitude value to prevent possible damage to
the tip.
13-12
ATTENTION: Lors dun travail en mode intercal (Interleave Mode), vrifier que la
valeur de tension applique loscillateur pizo-lectrique est infrieure celle
applique loscillateur en mode imagerie (Main Drive Amplitude). Le non respect
de cette procdure peut entraner la destruction de la pointe.
13-12
WARNHINWEIS: Bevor Sie im Interleave-Mode die Interleave Drive-Amplitude
einschalten, vergewissern Sie sich bitte, da der dort eingetragene Wert nicht
wesentlich grer ist, als der Wert der Main Drive-Amplitude, um evtl.
Beschdigungen der Spitze vorzubeugen.
13-12
WARNING: Do not insert a conducting object (e.g., screwdriver) into the Phase
Extender box while it is engergized.
13-14
ATTENTION: Ne pas insrer d objet conducteur (par exemple: un tournevis) dans le
botier dextension de phase (Phase Box) quand celui-ci est sous tension.
13-14
WARNUNG: Stecken Sie keine leitfhigen Teile (zum Beispiel Schraubenzieher) in
die Phase Extender Box, whrend diese eingeschaltet ist.
13-14
CAUTION: When closing the System.par file after viewing it, if you are asked to
SAVE the file or to SAVE CHANGES, be sure to say No.
13-15
TOW-7
TOW-8
Chapter 1
Control monitor
Display monitor
Computer
NanoScope Controller
Mouse
Keyboard
MultiMode SPM
1-1
Part I: Introduction
1.1. Introduction
The MultiMode scanning probe microscope (MM-SPM) is one of Digital Instruments original, mainstay designs and remains a highly dependable unit; several
hundred MultiModes have been shipped to laboratories and businesses worldwide.
The unit is designed for imaging small (approx. 1.5 cm dia.) samples using a series
of interchangeable scanners and is able to provide images from the atomic scale to
175 m in size. This manual is designed to assist operators with using the MMSPM and has been rewritten from earlier editions. The reader is also referred to the
Command Reference Manual.
The MM-SPM is designed around a stationary probe. That is, samples are scanned
back and forth beneath the probe. (This is opposite to other of Digital Instruments
designse.g., Dimension Series SPMswhich mount samples stationary while
scanning the probe back and forth above them.) Typically, samples are fixed to
round 1.5 cm metal disks (pucks), then magnetically attached to the top of the
scanner tube. As the scanner moves back and forth, the sample moves with it,
allowing the probe to extract information from the sample surface much like a phonograph needle plays a vinyl record.
Because the size of features imaged with SPM is often below the visible wavelength of light, all information gathered from sample surfaces is electronically
derived and rendered. Electronic controls have evolved through the years from an
early array of switches and knobs to the present software control system, designated NanoScope, version 4.x. Digital Instruments has divided its SPM software
into a two-function architecture: Real Time and Off-line. The Real Time software
functions are dedicated to running the actual microscope, changing the size and
location of scans, controlling gains, etc. Images produced from scans may be analyzed and/or modified using the microscopes Off-line functions. These yield sectional profiles, correct for noise and artifacts, analyze for depth, roughness, grain
size and power spectral density, among many other things. One major advantage to
the NanoScope design is that both modes may be run simultaneously. That is, the
microscope can save (capture) images in Real Time mode while the operator is
busy analyzing earlier images in the Off-line mode, making the MM-SPM a maximum-productivity tool.
1-2
Part I: Introduction
a deeper understanding of how SPM works will prove invaluable to the operator.
Our advice: experiment! Consult the Command Reference Manual for more information regarding software controls.
1-3
Part I: Introduction
1-4
Part I: Introduction
Contact AFM Measures topography by sliding the probes tip across the
sample surface. Operates in both air and fluids. See Chapter 6.
1-5
Part I: Introduction
1-6
Chapter 2
2.2.
2.3.
2.4.
2.5.
2.6.
Hardware 2
2.1.1. MultiMode SPM 3
2.1.2. SPM Head 4
2.1.3. Scanners 5
2.1.4. Tipholders 9
2.1.5. Probes 9
Control Mechanisms and Feedback 12
2.2.1. A Brief History of SPM Control Mechanisms 12
Feedback Gains 15
2.3.1. Proportional and Integral GainAn Analogy 15
2.3.2. Proportional Gain 16
2.3.3. Integral Gain 18
2.3.4. LookAhead Gain 19
2.3.5. Completing the Analogy 19
2.3.6. Setpoint 20
2.3.7. The SPM Electronic Feedback Loop 20
2.3.8. More about Feedback and Images 21
2.3.9. What Data Type of Image? 24
Control Parameters and Feedback 26
2.4.1. Reexamining the Control Loop 26
2.4.2. General Description of Main Menu Items 26
2.4.3. User Example 28
2.4.4. Review of General Operating Concepts 28
Review of TappingMode AFM 32
2.5.1. General Operating Concepts 32
2.5.2. Optimizing the TappingMode AFM Signal After Engagement 34
Terms and Abbreviations 35
2-1
Part I: Introduction
2.1. Hardware
This section provides a quick tour of the MultiMode SPM and its hardware. The
MultiMode SPM consists of seven major components: SPM, controller, computer,
keyboard, mouse, display monitor and control monitor. Mouse movements automatically transfer the cursor between monitors, enabling the operator to seamlessly
switch between control and display functions.
Control monitor
Display monitor
Computer
Keyboard
Controller
Mouse
SPM
2-2
Part I: Introduction
SPM tip
Head
Tipholder
X-Y head translator
Sample
Scanner
(Shown: A)
Retaining springs
Coarse adjustment
screws
Mode selector
switch
A-B
A+B
display
Base
(A+C)-(B+D)
A+B+C+D
display
2-3
Part I: Introduction
Figure 2.3. MultiMode SPM head and major components: laser (1);
mirror (2); cantilever (3); tilt mirror (4); photodetector (5).
Photodiode array The four elements of the quad photodiode (position sensitive
detector) are combined to provide different information depending on the operating
mode. In all modes the four elements combine to form the SUM signal. The amplified differential signal between the top two elements and the two bottom elements
provides a measure of the deflection of the cantilever. This differential signal is
used directly in the contact AFM. It is fed into an RMS converter (or phase module
if attached) for Tapping mode operation. Similarly, the amplified differential signal
between the sum of two left photodiodes and the sum of the two right photodiodes
provides a measure of the torsion in the cantilever and is used in Lateral Force
Microscopy (Image data type: Friction). Figure 2.4 shows the arrangement of the
photodiode elements in the MultiMode head.
2-4
Part I: Introduction
Laser
LFM
Photodetector segments
Photodetector
Mirror
AFM
Cantilever
2.1.3. Scanners
Figure 2.5 below shows the various, interchangeable scanners. The maximum scan
size and resolution of images depend upon the choice of scanner (see chart). Longer
scanners, e.g., type J, yield larger scan sizes; shorter scanners, e.g., type A,
offer smaller images down to the atomic-scale. Smaller scanners tend to be more
noise-free at acoustic frequencies because of their compact size and rigidity. Larger
scanner offer wider scans, while requiring extra noise dampening precautions at
smaller scan sizes of high resolution.
2-5
Part I: Introduction
Model
Scan Size
Vertical Range
AS-0.5 (A)
0.4m x 0.4m
0.4m
AS-12 (E)
10m x 10m
2.5m
AS-12V (E vertical)
10m x 10m
2.5m
AS-130 (J)
125m x 125m
5.0m
AS-130V (J vertical)
125m x 125m
5.0m
AS-200
200m x 200m
8.0m
Figure 2.6 depicts the electrode configuration used on one type of scanner piezo
tube. Electrodes are oriented as shown when the MultiMode is viewed from the
front. With the Scan angle parameter in the control panel set to 0.00, the fast-scan
direction is in the direction of the X-axis.
2-6
Part I: Introduction
~
Z
~
Y
Y
~
X
Figure 2.6. Typical scanner piezo tube and X-Y-Z electrical configurations.
AC signals applied to conductive areas of the tube create
piezo movement along the three major axes.
AC voltages applied to the scanner crystals X-Y axes produce a raster-type scan
motion as represented in figure 2.7. The horizontal axis presented on the display
monitor is referred to as the fast axis (in this example, the X-axis) and scans at a
Scan rate entered by the user. The orthogonal axis is known as the slow axis (in
this example, the Y-axis).
2-7
Part I: Introduction
Figure 2.7. Voltages applied to the X- and Y-axes produce a raster scan
pattern. Either axis may be designated as the fast axis.
2-8
Part I: Introduction
2.1.4. Tipholders
The sample and mode of SPM to be performed dictate the choice of tip and
tipholder. For example, if contact AFM is to be used for imaging, a silicon nitride
cantilever mounted in a standard tipholder is the usual choice. If TappingMode is to
be used for imaging a biological specimen in fluid, a special fluid cell is employed.
STM utilizes a special tipholder, having a tiny tube holder adapted for holding wire
tips. Examples of each tipholder are shown below in figure 2.8.
Top view
Bottom view
Figure 2.8. Various tipholders utilized with the MultiMode SPM.Left-toright: contact AFM and TappingMode; EFM; force modulation;STM.
Notes
2-9
Part I: Introduction
2.1.5. Probes
Probes come in a variety of sizes, shapes and materials and are chosen according to
the type of imaging to be done.
Wire Probes
STM probes usually consist of wire, cut and/or etched to produce atomically sharp
tips at one end. Usually these are made from tungsten or platinum-iridium alloy
wires. A potential is established so that electrons flow between the tip and sample.
A similar type of wire probe is used for lithography. These often consist of ordinary
tungsten STM tips and/or a wire with a tiny diamond fixed to its end. Lithography
tips are used for mechanically deforming sample surfaces in the form of controlled
dents and scratches. Essentially, the tip serves as a scribe or punch. It may be used
to test surfaces for microhardness, etch patterns or explore material characteristics.
2-10
Part I: Introduction
Cantilevered Probes
Most SPM work is done using cantilevered probes. These consist of a flexible cantilever extending from a rigid substrate, to which a tip has been attached. In contact
AFM, the cantilevers flexibility acts as a nanometric spring, allowing the tip to
measure surface forces. In TappingMode, the probe is oscillated up and down at its
resonant frequency while its amplitude and phase are monitored.
Figure 2.10. Two types of cantilevered probes: silicon nitride (left), and
crystal silicon (right).
Cantilevered ProbesTappingMode
Digital Instruments answer to minimizing contact AFM forces is TappingMode, a
proprietary form of AFM. In this instance, a stiff crystal silicon probe is oscillated
to its resonant frequency. Because the tip describes a high-frequency (e.g., 100-plus
KHz), oscillating arc, it possesses sufficient energy to break free of surface tension
2-11
Part I: Introduction
forces. The probe is considerably stiffer than silicon nitride, making it more brittle
and less forgiving. Thus, the operator must be more cautious while setting up the tip
and sample.
Cantilevered ProbesMFM
Another variation of the TappingMode tip is the MFM probe. This is basically a
crystal silicon TappingMode probe having a magnetic coating on the tip. (Such tips
are sold by Digital Instruments under the name NanoProbe.) As the magnetized
tip oscillates through magnetic fields on the sample surface, it modulates the cantilevers phase and frequency. These are monitored, providing a measure of magnetic
field strength and providing images of magnetic domains.
Cantilevered ProbesEFM
Similar to MFM (see above), EFM is also conducted using NanoProbe tips. Tips
may be electrically connected to the microscopes circuitry to obtain surface potential maps of the surface, or oscillated while monitoring phase changes due to electrostatic forces. These techniques yield images of the samples electrical domains.
Notes
Both MFM and EFM may be conducted using the MM-SPM alone; however,
best results are obtained with an Extender Electronics Module attached. For
more information regarding this attachment, contact Digital Instruments.
Specialized Probes
As the field of SPM continues its explosive growth, new probes are constantly
introduced. Here are a few examples of specialized probes:
Chemical doped tips Tips doped with a chemical species of interest to the
investigator. The chemically doped tip measures chemical bonding forces on
sample surface, images receptor sites on biological membranes, etc.
2-12
Part I: Introduction
Tip
Tunneling electrons
Surface
2-13
Part I: Introduction
STM tip
Flexible cantilever
Sample
AFM tip
Figure 2.12. Early contact AFM which allowed imaging non-conductive
samples. In this scheme, a contact AFM tip was monitored
using the STM tip directly above it.
Preceding the first SPMs, some profilometers had relied upon optical methods to
monitor the rise and fall of a sharp stylus over sample surfaces. This approach
offered good sensitivity by reflecting a laser beam off the end of the stylus and into
a photodetector to obtain an optical lever capable of detecting even the smallest
2-14
Part I: Introduction
movements (figure 2.13). This approach was then applied to SPMs. A related
method utilized interference to detect shifts in interference fringes.
Laser
Photodetector
Scanner
2-15
Part I: Introduction
obtain the highest resolution images, the balloon must track the surface as closely
as possible without crashing into it. This poses a dilemma to the balloonists: how to
tightly control the balloons position relative to the ground.
setpoint altitude
100 meters
Because the balloon will drift slightly up and down due to the effects of wind and
temperature, the balloonists must establish some minimum altitude as a safety zone.
Let us call this the setpoint altitude, and let us assume that it is set at an altitude of
100 meters.1
When the terrain is flat, the problem is simplified. The balloonists need only ensure
a constant supply of gas is supplied to the balloons burners to keep the balloon
aloft. As the terrain becomes hilly, the task becomes more complex. If the terrain
rises, the balloonists must respond by firing the burners to lift the balloon. As the
balloon clears the hill and terrain drops away, the balloonists must turn the burners
off to reduce height and continue tracking the terrain. The type and intensity of the
balloonists responses to terrain can be modeled in terms of three types of feedback:
proportional, integral and LookAhead.
1. For the sake of simplicity, setpoint in this analogy applies to the balloons altitude;
however, setpoint in SPM is applied to tip-sample forces, not the tips height above the
surface.
2-16
Part I: Introduction
sharply, Peter uses large amounts of gas to lift the balloon; where the terrain is relatively flat, Peter supplies a small, steady amount of gas to maintain the setpoint altitude above the surface.
range finder
If altitude < 100 meters,
fire burners.
A simple feedback loop emerges in this analogy: let us say Peter uses a range finder
every 30 seconds to determine the distance between the balloon and ground. If the
balloon is below its setpoint altitude, he fires the burners. If the balloon is above its
setpoint altitude, he turns off the burners to lower the balloon. The higher the proportional gain, the more Peter reacts to changes in altitude. For example, at a proportional gain of 1, if the balloon is 25 meters too low, he opens his valve at 10
liters per second; if the balloon is 50 meters too low, he opens his valve at 20 liters
per second. The proportional gain value serves as a multiplier such that at a proportional gain of 2, the gas flow rates are doubled from a proportional gain of 1, and so
on. Although this sort of feedback gain works well for simple, linear models, it does
not function as well for nonlinear models. There remains always some residual
error which causes the system to approach, but not quite reach, the target state.
Assuming that the balloonists wants to get as close as possible without crashing, the
response will depend upon, among other things, the balloons speed over the terrain. When the balloon is being carried swiftly, it is necessary to apply feedback
earlier to compensate. (That is, more gas must be used earlier.) On the other hand, if
there is little or no wind, the balloon may achieve a closer tracking of the terrain.
There may also be sufficient knowledge of the terrain to anticipate its rises and
falls. In order to compensate for these effects, integral and LookAhead gain feedbacks may also be employed. These are discussed next.
2-17
Part I: Introduction
2-18
Part I: Introduction
2.3.6. Setpoint
In our ballooning example, setpoint referred to the target altitude to be maintained. In scanning probe microscopy, setpoint refers to how much tip-sample
force is maintained. There are two ways of defining setpoint, depending upon
whether one is referring to contact AFM or TappingMode. In contact AFM, setpoint
is determined by the amount of cantilever flexionas the setpoint increases, the
cantilever flexes more and tip-sample forces increase. In TappingMode, setpoint is
determined by the RMS amplitude of the oscillating tipas setpoint decreases,
RMS amplitude decreases, but tip-sample forces increase1.
2-19
Part I: Introduction
1. At first glance, this may seem counterintuitive. However, recall that an oscillating tip in
TappingMode attains its fullest amplitude when it is in free air and not interacting with a
sample. As the oscillating tip is brought against the sample, its RMS amplitude decreases
due to damping effects. The harder the tip is pressed into the sample, the more RMS
oscillation is reduced. Thus, requesting a Setpoint of 0.00 in TappingMode commands
the system to press the tip against the sample so hard that the cantilever cannot oscillate at
all. In TappingMode, reducing setpoint increases tip-sample forces...the opposite of contact AFM.
2-20
Part I: Introduction
2-21
Part I: Introduction
Although this scan is much larger than normally found in SPM, it illustrates how an
adjacent, lagging scan line can be used to determine local scan lines on regular surfaces. In most places (e.g., the forehead), each scan line changes little from the line
next to it. In some local areas (such as under the nose) there are small, sudden
changes; however, these are relatively isolated. In contrast, a similar trace of an
irregular, random surface would reveal that each scan line bears little resemblance
to its adjacent line.
The entire purpose of LookAhead gain is to take full advantage of regular features
by using every line to anticipate the next one. In NanoScope software, the LookAhead gain value may be adjusted between a range of 0 (off) to 16 (maximum). As
values are adjusted upward from 0, the LookAhead gain is weighted to apply more
data from the adjacent (lagging) line. Although LookAhead gain is relatively useless for random surfaces, it is a tremendous help on regular surfaces.
When LookAhead gain is switched on ( > 0), it is the first gain calculated in the
feedback process and is used to weight the integral gain as follows:
LA
G int
zx
At start-up, there is no information yet recorded from an adjacent scan line; therefore, the LookAhead gain is effectively 0 until three lines have been scanned. This
allows the system to settle down and record data.
Integral gain and average error
1. Although the LookAhead gain value can be set by the operator to values between 0 and
16, these are not the values plugged into the equation. This is a digital signal feedback
process and the actual value multiplied varies between 0 and 224. A similar rule holds also
for both Integral and Proportional gain.
2-22
Part I: Introduction
The second step in the feedback process uses integral gain to correct for error by
averaging (integrating) the total error. In the ballooning example, Irene kept a running average of the balloons altitude error, then responded by firing the burners or
turning them off to bring the balloon closer to the setpoint altitude. Similarly, the
SPMs feedback process maintains a running average of the error and responds to it.
As we have seen, enabling the LookAhead gain by setting it > 0 conditions the
Integral gain entered by the operator. If LookAhead gain is turned off (= 0), integral gain enters the feedback process unchanged. The integral gain is then used to
calculate a running average of error as follows:
new
z acc
LA
= error G int
old
+ z acc
new
old
where z acc is the new average error calculated by adding the old average error z acc
to the product of the integral gain times the error. The running average represented
new
by z acc maintains itself continually until one or more of the major scanning parameters is changed by the operator. Whenever major scan parameters are changed
(e.g., Setpoint), the error accumulator is dumped and begins a new running average. With the average error calculated, the feedback system is prepared to make its
final error correction based upon proportional gain.
Proportional gain
The third and final step in the feedback process uses proportional gain to complete
error correction. Recall that proportional gain responds to error in proportion to
how much it differs from the setpoint. Proportional gain is used to calculate the
final correction voltage sent to the Z-axis piezo according to the relation
new
z voltage = z acc
+ error G prop
As suggested in the equation, by the time proportional gain is figured in, the bulk of
error correction has already been completed. This tends to make Proportional gain
a less touchy control when compared to Integral and LookAhead gain. Nevertheless, the system can be driven into oscillation whenever gains are excessive,
including Proportional gain.
REMINDER: Gain values entered on the Real Time / Feedback Controls panel
do not directly translate to any real quantity, but are merely self-referencing; e.g., a
Proportional gain value of 2.0 is not the same as Integral or LookAhead gain
values of 2.0.
2-23
Part I: Introduction
Signal in (Inaux)
Signal in (In0)
Microscope
2-24
Part I: Introduction
contact AFM, the same line now conveys Deflection voltage. In TappingMode, the
same line conveys Amplitude data. In addition, the auxiliary channels utilize line
InA through InD.
Notes
Although there are four separate auxillary lines (InA through InD), the controller
can only access one at a time, plus the In0 line.
The controller is designed to handle two input signals simultaneously (In0 and one
auxiliary line). Because it is extracted from a different (output) portion of the feedback loop, a third channel can be used to simultaneously extract Height data without affecting the input signals. This allows a maximum of three Data types for any
one sample. For example, a sample being imaged with contact AFM can show
Height, Deflection and Friction all at the same time.
2-25
Part I: Introduction
2-26
Part I: Introduction
Scan angle Combines X-axis and Y-axis drive voltages, causing the piezo to scan
the sample at varying X-Y angles.
Scan rate The number of lines scanned per second in the fast scan (X-axis on
display monitor) direction.
Number of samples Sets the number of pixels displayed per line and the number
of lines scanned per frame.
Slow scan axis Starts and stops the slow scan (Y-axis on display monitor). This
control is used to allow the user to check for lateral mechanical drift in the microscope or assist in tuning the feedback gains. Always set to Enable unless checking
for drift or tuning gains.
Z limit Limits the amount of drive voltage available to the Z piezo circuit. The Z
control system uses a 16-bit D/A converter which drives an amplifier capable of
outputting voltages from +220V to -220V. This means that the resolution of the
control over the Z direction is approximately 6.7mV per bit (440V divided by
65536). This setting defaults to 440V automatically. Reducing the Z limit is useful,
when using a E or J scanner, if scanning samples with relatively small Z features (less than 10 nm peak-to-valley). For example, setting the Z limit to 55V
means that 55 Volts is divided by the same 16-bit digital control. This gives eight
times finer control over the Z direction of the scanner.
Integral gain and Proportional gain Controls the response time of the feedback loop. The feedback loop tries to keep the output of the SPM equal to the setpoint reference chosen. It does this by moving the piezo in Z to keep the SPM's
output on track with the setpoint reference. Piezoelectric transducers have a characteristic response time to the feedback voltage applied. The gains are simply values
that magnify the difference read at the A/D convertor. This causes the computer to
think that the SPM output is further away from the setpoint reference than it really
is. The computer essentially overcompensates for this by sending a larger voltage to
the Z piezo than is truly needed. This causes the piezo scanner to move faster in Z.
This is done to compensate for the mechanical hysteresis of the piezo element. The
effect is smoothed out due to the fact that the piezo is adjusted up to four times the
rate of the display rate.
2-27
Part I: Introduction
Display both the input and the output of the feedback loop. This means setting
the display to show both Height data and the appropriate microscope signal
(STM = Current, contact AFM = Deflection; TappingMode AFM =
Amplitude).
Engage the microscope and reduce the gains until they are close to zero. The
input display data will become larger in Z, and the height data will blur or
become smeared. Raise the gains using the arrow keys until the input voltage is
minimized.
Try increasing and decreasing the scan rate parameter. This will increase or
decrease the traveling velocity of the tip. Note that it will be necessary to
increase the gain settings at faster scan speeds and decrease the gains at slower
scan speeds.
2-28
Part I: Introduction
The AFM system maintains the tip at the end of the cantilever in contact with the
sample surface. The sample is scanned under the tip in X and Y. Features on the
sample surface deflect the cantilever, which in turn change the position of the laser
spot on the photodiodes. This position change is read by the feedback loop. The
feedback loop moves the sample in Z to restore the spot to its original position.
Refer to figure 2.14 (opposite page).
Figure 2.14.A. A flat portion of the sample surface is scanned beneath the tip leftto-right, maintaining the laser beam at the center of the photodiode array.
Figure 2.14.B. As the tip encounters a raised feature, the cantilever is pushed up,
deflecting the laser beam upward onto the "A" portion of the array. With the "A"
photodiode receiving an increased portion of the laser light, its voltage increases
while portion "Bs" decreases (A > B).
Figure 2.14.C. The Vertical Deflection (A-B) voltage differential is sensed by the
feedback electronics, causing a dropped voltage to the Z piezo crystalthe piezo
retracts. As the Z piezo retracts, the cantilever recenters the laser beam onto the
photodiode array (A = B).
Figure 2.14.D. As the tip encounters a decline in the sample topology, the tip drops.
This directs more of the beam onto the "B" portion of the photodiode array. With
the "B" photodiode receiving an increased portion of the laser light, its voltage
increases while portion "As" decreases (A < B).
Figure 2.14.E. Again, the Vertical Deflection (A-B) voltage differential is sensed
by the feedback electronics, increasing voltage to the Z piezo crystalthe piezo
extends. As the Z piezo extends, the tip is pushed down until the laser beam
recenters on the photodiode array (A = B).
2-29
Part I: Introduction
Photodiode Array
Photodiode "B"
Mirror
Laser
Photodiode "A"
Laser beam
Tip
A/D
Converter
Sample
Z piezo
Scanner
Tube
Computer
B
A
Figure A
Figure D
Figure B
Figure C
Figure E
2-30
Part I: Introduction
The AFM always first engages in the repulsive region of its operating range. In
other words, the cantilever needs to exert a positive pressure on the sample surface.
The AFM block diagram shows the relationship between the cantilever movement
and the laser spot on the photodiode array. The diagram shows that the spot moves
up (more on "A") when the cantilever is pushed up. The initial setup is to have the
Vertical Deflection (A-B) voltage about 2-3 volts more negative than the Setpoint
voltage. Digital Instruments recommends starting with the Setpoint voltage set to 0
volts and the Vertical Deflection (A-B) set to -2 volts. The reason for this is that 0
volts is the middle of the control range. The indication of a good engagement is a
distinct jump of about 1V from the Vertical Deflection (A-B) voltage to the Setpoint voltage.
The displayed image is an average of the corrections made to Z in a given display
period (number of samples menu item). The two gains are set to values to effectively tune the feedback response to the particular sample topology. This will set
response time of the system so that there is no difference between the SPM's signal
and the setpoint reference during scanning.
Proportional gain The computer multiplies this number times the value read
from the comparison circuit every time the A/D converter is read. It is the high frequency feedback control.
Integral gain This number is multiplied times an accumulated average of A/D
readings. This is the low frequency feedback control.
One of the easiest ways to set the gains properly is to view the input of the feedback
loop. This means displaying the STM current, the AFM deflection, or the TMAFM
amplitude signal. Then raise the gain values until the input of the feedback is minimized. Note that this will result in an image that shows only large transitions in Z;
this is normal. There will always be a time lag between the input and the output
(height data) of the feedback loop. Usually, the Integral gain is the most sensitive
control. Raise the gains together until the input signal (current, deflection or amplitude) is minimum. Don't set them so high as to cause oscillations in the image.
Oscillations are an indication of too much feedback correction voltage sent to Z.
This is generally known as feedback oscillation. The Proportional gain can usually be set about 20 percent higher then the Integral gain, but it is not required.
2-31
Part I: Introduction
2-32
Part I: Introduction
Photodiode Array
Photodiode "B"
Mirror
Laser
Photodiode "A"
Laser beam
Reflected Laser
Beam
Oscillating tip
Figure A
Sample
Z piezo
Scanner
Tube
Computer
Figure B
Figure C
Setpoint Voltage
Figure D
RMS Voltage
Setpoint Voltage
RMS Voltage
RMS Voltage
RMS Voltage
Setpoint Voltage
Setpoint Voltage
0 Volts RMS
0 Volts RMS
0 Volts RMS
0 Volts RMS
Figure 2.15. TappingMode AFM concepts. (See sect. 2.5.2 for explanation.)
Figure 2.15 shows the relationship between the RMS and the setpoint voltage during the engage cycle. The initial setpoint voltage is determined by the computer
rather than the user. The computer sets the setpoint equal to 95 percent of the RMS
amplitude. The tip is then lowered until the RMS matches the setpoint. The computer then tests for true engagement as follows: 1) the motor halts the tips descent;
2) the setpoint is lowered slightly; 3) the feedback control monitors movement of
the Z piezo. Depending upon the tips relationship to the sample, one of the two following conditions will result:
MultiMode SPM Instruction Manual
2-33
Part I: Introduction
1. A small Z piezo movement. This indicates that the cantilever is truely engaged
with the sample surface.
2. A large Z piezo movement. This indicates that the cantilever is being damped by
air trapped between the cantilever and sample surface (not in contact with the
actual, solid surface)this is a false engagement condition. The setpoint is readjusted and the engage cycle repeated until the computer reads a small change in Z
when the setpoint voltage is lowered further. One symptom that this condition is
occuring is when the tip travel m display stops momentarily, then starts again.
2-34
Part I: Introduction
2-35
Part I: Introduction
2-36
Part I: Introduction
2-37
Chapter 3
1. NOTE: The calibrated scanners parameter files are located in the hard disks EQUIP
directory.
3-1
Part I: Introduction
3-2
Part I: Introduction
Figure 3.2. Each monitor is shipped separately (left). The SPM is packed in
a separate box (right). Heavy hardware should be removed carefully from
each box with the help of 2-3 people.
1. Clear a table for the microscope. You will require a table of at least 76 x 152 cm
(30 x 60 inches) size. The table should be level, of sturdy construction, and located
where it will not be vibrated or jarred by foot traffic. It should be located away from
sources of heat and cold; avoid windows, air conditioning and heating ducts, blowers, etc.
2. First remove the computer and controller. The controller may be used to support
the display monitor1 as shown in the setup below.
Extender box
(if supplied)
Control monitor
Printer cable
Display monitor
Computer
NanoScope
controller
SPM
Keyboard
Mouse
3-3
Part I: Introduction
3. Remove the remaining components and locate them using the layout above as a
guide. Some users find it advantageous to set the SPM on the floor or atop a heavy
concrete cinder block. This will reduce vibrational noise which can affect scans. If
a tripod vibration isolator or similar mount will be used, locate it next to the table.
All units are supplied with a round vibration isolation pad to set the SPM on. This
can be used between the SPM and table top or floor to reduce vibrational noise.
Cabling
4. Before connecting cables, review the following precautions:
DO NOT power anything until all cables are connected and double-checked.
Power cords should be connected to their source of power last.
DO NOT route cables through pinch-points or over sharp edges. The table
should be pulled away from walls so that cables will not be pinched. The cable
between the controller and SPM carries up to 440 volts and must be protected
from chaffing.
3-4
Part I: Introduction
Printer (LPT1)
COM1
(Trackball)
Display monitor
Control monitor
NanoScope
controller
3-5
Part I: Introduction
NanoScope
controller
Control monitor
Display monitor
COM1
(Trackball)
Printer (LPT1)
COM2
(Dimension SPMs only)
Mouse
Keyboard
Power
3-6
Part I: Introduction
Notes
Motorized scanner
screw must seat
with coupling here.
Front View
Rear View
3-7
Part I: Introduction
9. Carefully set the MultiModes head atop the SPM scanner. The head should mate
kinematically with the scanners contact balls. Secure the head using both retaining
springs by stretching the springs up and hooking them over posts located on both
sides of the head. Plug the head into the receptacle on the support ring as shown
below:
(Shown: "A" scanner)
Rear View
10. Recheck all cable connections. Review the steps above one last time to ensure
that all connections are properly made. Be especially certain that the controller
cable is NOT plugged into the parallel port of the computer or damage to the electronics will result.
3-8
Part I: Introduction
is also available from Digital Instruments (model TRVI). Experience suggests that
air tables often have poor horizontal isolation. If an air table is used, it may require
additional horizontal isolation for atomic-scale images.
As a final note, the best way to reduce coupling from vibrations is to eliminate as
many sources of vibration as possible. Remember that vibrations can be transmitted
to the SPM over the cable. To reduce this phenomenon, prevent tension in the cable
and keep it away from fans and other noise sources. Also, keep the microscope
away from sources of acoustic noise. Loud noises (including conversation) can disrupt atomic images, so it is best to isolate the SPM as much as possible. Use the
noise reduction hood supplied with the system to alleviate acoustic noise.
3-9
3-10
Part I: Introduction
Chapter 4
The MultiMode microscope comes furnished with etched silicon cantilever substrates for TappingMode AFM and silicon nitride cantilevers for contact AFM
modes. In both cases, the cantilever probes should be inspected under the microscope when being used for the first time, to get a better understanding of how the
probes and substrates are connected and taken apart. The procedure for removing
individual substrates from the wafer varies depending on the wafer. It will be easier
to accomplish this task with the aid of a stereo microscope having 5070X magnification.
4-1
Cantilever Preparation
silicon wafer frame. The supporting arms connecting the substrate to the bulk of the
wafer will shatter as pressure is applied. See figure 4.1 for clarification.
It may be convenient to break several substrates from the wafer at one time. Extras
may be safely stored in a specially prepared closable container. At the bottom of the
container, place X4-grade, GEL-PAK adhesive strips. Place the substrates, tips
facing up, on the adhesive to permit easy removal of the substrates when needed.
Cover the container when not in use.
Cantilever
Press here
to break out
substrate
Substrate
Supporting
Arm
Cantilever
Substrate
4-2
17.0
Cantilever Preparation
17.0
Tip
25.0
10.0
Cantilever
4-3
Cantilever Preparation
10
11
55
80
1 m - 2 m Deep Trench
Note: Any wall angle on the left wall that is > 55 deg.
will be shown as 55 deg. in the image.
4-4
Cantilever Preparation
73 73
Scan Line Profile
1 - 2 m Deep Trench
4-5
Cantilever Preparation
20 - 30
4-6
Cantilever Preparation
10
11
55
70 - 80
1 m - 2 m Deep Trench
Note: Any wall angle on the left wall that is > 55 deg.
will be shown as 55 deg. in the image.
Any wall angle on the right wall that is >70-80 deg.
will be shown as 70 -> 80 deg. in the image.
4-7
Cantilever Preparation
4-8
Cantilever Preparation
3. Place the strip down on a white piece of paper to inspect it under the microscope.
When doing this, be careful that the cantilevers are on the top side of the strip. The
cantilevers are on the same side as the reinforcing ring, while the saw-cuts are on
the opposite side from the cantilevers. Between each one on the down side of the
strip should be a saw-cut almost through the Pyrex.
4. Handling the strip by its ends, place it on a glass slide (taped down to the edge of
a table) with the substrate or spacer piece to be removed, hanging over the edge, as
shown in Figure 4.8. The saw cut should be approximately on the edge of the slide.
While holding down the next substrate on the strip with the wooden end of a cotton
swab, grip the overhanging piece with a pair of wide tweezers and rotate downward. It should pop right off. Repeat this process until as many cantilever substrates
as required have been removed.
Cantilever
Cantilever
Substrate Substrate
Saw Cuts
Saw-Cuts
Press Here to
Break Off
4-9
Cantilever Preparation
micron cantilever with the wider legs. For atomic scale images, the 200-micron triangular cantilever with the wider legs can yield good results.
Figure 4.9. Substrate shown with the 100 and 200 m long, wider
cantilevers on top and the 100 and 200 m long, slimer cantilevers on the
bottom.
4-10
Cantilever Preparation
45
35
4 m
C
D
Lever Type
100 m Wide
200 m Wide
100 m Narrow
200 m Narrow
A
115
193
115
193
B
122
205
122
205
C
21
36
15
20
D
60
113
69
150
Spring constant
(N/m)
0.58**
0.12**
0.38**
0.06**
Measured Fr
(kHz)
40
12.3-22.1
-
4-11
Cantilever Preparation
45.0
65.0
11
10.0
Scanning Profile
4-12
Chapter 5
See Chapter:
10
11
Interleave scanning
12
13
14
5-1
5-2
Sample Preparation
If it is your first time operating the microscope, we recommend that you image the
calibration sample provided with the instrument (usually a 10m-pitch grid of
180nm step height).
The calibration sample or other small sample should be placed on one of the 15 mm
diameter metal disks used for sample mounting. The MultiMode SPM is provided
with several steel sample disks that can be attached to the magnetic sample holder,
located atop the scanner tube.
Provided with the instrument are red and white colored sticky tabs which are 2
sided adhesive patches. Peel off a sticky tab from the provided sheet, and place it
on the steel small sample puck, then peel off the red-and-white paper. This leaves a
patch of the two-sided adhesive on the steel sample disk, which will hold the sample chip to the disk. Using tweezers, place the small sample to be imaged firmly on
the sticky tab adhesive (figure 5.1). Alternatively, a small sample can be glued
down to the sample puck using cyanoacrylate glue (superglue). Place the small
sample disk atop the scanner.
Figure 5.1. Gently press the sample onto the sticky tab until secured.
5-3
Retaining springs
5-4
Reattach retaining
springs (2)
5-5
Tip down
Tip up
Figure 5.4. The tip is quickly raised and lowered using the Up / Down
toggle switch on the MultiMode SPMs base.
TappingMode
For contact AFM, install a silicon nitride probe tip in the AFM tipholder. Figure 5.5
shows the AFM tipholder. Detailed procedures for silicon nitride cantilever substrate preparation are given in Chapter 4, including a description of the procedure to
break out substrates from the wafer. Ensure the gold-plated side of the substrate is
placed down towards the substrate mount, with the nitride film side attached to the
cantilever oriented up away from the substrate mount.
The procedure for installation of TappingMode, single crystal silicon probe tips is
essentially the same as for contact AFM (see above paragraph). The substrate
should be face-up, with the probes cantilever pointing away from the AFM
tipholder. This ensures that the cantilever and tip are facing toward the sample once
the tipholder is mounted in the head. Refer to Chapter 4 for a description of the procedure to break out each substrate from the wafer.
5-6
Spring clip
5-7
Probe
Wire clip
5-8
(Rear view)
Clamping
screw
5-9
5-10
Figure 5.8. Two ways of positioning a magnifier when aligning the laser.
Through the heads front window (left), and overhead (right).
2. Identify the key features.
There are many laser light reflections, particularly from the tipholder, the sample,
and various metal surfaces. It is helpful to identify the key features before starting
the alignment. Make sure that the adjustment screws are positioned such that the
cantilever is well above the sample, so reflections off the sample do not make it
hard to locate the cantilever. Compare the view through the magnifier to the earlier
view of the cantilever.
3. Position the laser spot on the substrate.
The laser spot can be moved around with the two positioning knobs on the top of
the head. With the right-positioning screw, move the spot so it is on the cantilever
substrate. This is easy to see with the silicon nitride cantilevers because the glass
substrate will glow and there will be a lot of diffused red light scattered around the
internal area of the head.
5-11
Y-axis adj.
While observing the substrate through the magnifier, slowly move the laser spot
lengthwise along the substrate with the X-axis positioning screw until the spot is
barely off the edge of the substrate (1). The laser spot may be visible on the sample
just off the substrate edge. With the rear-positioning screw, move the spot along the
edge of the substrate (2). Very small movements of the positioner are enough to
pass over the cantilever, so move slowly with small motions. Since the substrate is
mounted by hand, there is usually a small amount of angle between the substrate
and the axes of the positioners, so the best way to look for the cantilever is to make
small motions along the edge of the substrate and to move the spot back and forth a
little bit lengthwise for each small increment of edgewise motion. This way, it is
possible to stay adjacent to the edge without moving away from the substrate or
back onto it. In this fashion, move along the edge of the substrate. If a corner of the
substrate is reached without finding the cantilever, move back the other way.
1
Laser spot
X-axis adj.
5-12
5-13
Vertical adj.
1
Laser spot
Horizontal adj.
5-14
Image on paper
Figure 5.12. Laser spot off probe and resultant reflection pattern.
Intensity may vary according to reflectivity of sample.
2. Position the laser beam in X (figure 5.13).
Position the laser beam so that a small amount of light is reflected from the edge of
the cantilever substrate. The purpose of this step is to position the beam such that
most of the laser light is falling off the end of the substrate; therefore, it is important
that the amount of reflected light be small.
Spot on end
of substrate
Image on paper
Figure 5.13. Locate the laser spot on the edge of the cantilever.
5-15
Spot between
cantilever legs
near tip
Image on paper
Notes
The procedure is similar for the silicon cantilevers but because the cantilever is a
single beam, there will only be one laser spot reflected onto the paper.
4. Align the laser beam with the end of the cantilever (figure 5.15).
For silicon nitride cantilevers: move the laser beam out onto the end of the cantilever by turning the X adjustment counterclockwise while looking at the reflected
spot on the slip of paper. The spots reflected from the legs should merge into a single, well defined, bar-shaped spot. Verify that the laser beam is aligned with the end
of the cantilever by moving the beam back and forth in Y. If more than one reflected
spot is observed, then again position the beam between the two legs and move out
in X. Repeat until only a single, bright, well-defined spot is observed when the
position of the laser beam is varied in Y. If the reflected spot is dim or less well
defined, the laser beam may be improperly aligned. Attempting engagement under
these conditions may result in damage to the cantilever or sample.
Spot on tip
of cantilever
Image on paper
Good images can usually be obtained any time there is a good bar-shaped pattern
reflected off the back of the cantilever; however, the following suggestions for finetuning the alignment of the laser spot position may enhance the image quality in
some cases. The position of the laser spot on the cantilever affects the sensitivity of
the cantilever deflection signal. The closer the spot is to the tip, the higher the sensitivity. However, the strength of the deflection signal decreases and optical interference problems increase as the laser spot is moved to the end of the cantilever,
because the cantilever is triangular in shape and the laser beam begins to spill off
the cantilever as the beam is moved towards the tip. This results in a trade-off
between signal strength and sensitivity.
Increased sensitivity is desirable when imaging atoms. So, when imaging atoms,
turn the laser diode X adjustment slightly counterclockwise to move the laser beam
closer to the end of the cantilever. Stop when the reflected bar-shaped pattern on the
piece of paper begins to shorten as the beam spills off the cantilever. It is important
to keep the entire laser beam on the cantilever when imaging samples with tall features. For these and most other samples, a good, solid, bar-shaped reflection on the
piece of paper provides the best results.)
For silicon cantilevers, the paper method will be slightly different because the silicon cantilever is a single beam rather than an open triangle (refer to figure 5.16).
The general method is as follows:
1. Locate the laser spot on the substrate.
2. Move the spot along the Y-axis to locate the center of the cantilever.
3. Move the spot off the substrate and onto the cantilever.
4. Locate the spot on the center of the cantilever.
5. Move the laser spot to the end of the cantilever.
Throughout the procedure, reflections from the cantilever can be interpreted on a
slip of paper. Use figure 5.16 below as a guide.
5-17
5
~ 1/3 turn
Figure 5.16. Procedure for aligning silicon cantilevers using a slip of paper:
1) locate beam on substrate; 2) find center of substrate;
3) move toward cantilever; 4) locate cantilever;
5) find center of cantilever; 6) move to end of cantilever.
6. Align the photodiode with the reflected beam.
Contact AFM
Once the laser beam is aligned with the cantilever, it is necessary to position the
photodiode relative to the reflected spot. This is done by first adjusting the mirror
lever on the rear of the head, then adjusting the photodiode adjustment knob such
that the SUM signal displayed on the elliptical bar graph is maximized. This is a
much less sensitive adjustment than the laser beam alignment, so the maximization
need only be approximate. If the laser beam has been properly aligned, the maximized SUM signal should be over 4 volts with gold-coated silicon nitride cantilevers. If this is not the case, check the alignment of the laser beam.
TappingMode
For TappingMode: with the microscope toggled to TM AFM mode (the LED on the
face of the MultiMode base should be green), use the mirror lever on the rear of the
head and the photodiode adjustment knob to maximize the SUM signal displayed
on the elliptical bar graph, and zero the top/bottom differential signal which is
shown on the lower LCD display. If the range of movement of the photodiode stage
is inadequate to accomplish this, adjust the mirror angle with the lever in the back
5-18
of the head. If the laser beam has been properly aligned, the maximized SUM signal should be 13 volts with a silicon cantilever. If this is not the case, check the
alignment of the laser-beam.
Projecting Diffraction Patterns with Head Removed (Newer Models Only)
Yet another "paper" method of laser alignment consists of removing the head and
interpreting diffraction patterns as light spills around the cantilever (figure 5.17).
Notes
As a safety measure, all SPM heads are equipped with a cutout switch that
automatically turns off the laser when the head is moved (to prevent eye
damage). Early model heads feature a magnetic switch which turns off the laser
when the head is removed from the scanner. The method described here will
NOT work with these early model heads. Later models feature a mercury tilt
switch which allows the head to be removed from the scanner without turning
off the laser. Always handle the head with caution. DO NOT DROP!
5-19
The projection method is very similar to the method described above using a slip of
paper. The patterns are also similar, but less distinct.
5-20
(Front-Top View)
Photodetector mirror adj.
C-D signal
(Rear View)
Figure 5.19. Signals produced by the photodetector are optimized
using the heads various adjustments.
This adjustment is much less sensitive than the laser position adjustments. The
maximized value should be approximately 5.09.0 Volts for silicon nitride cantilevers. The value of this signal varies with many factors. It is important to note that it
is possible to see a large response on the elliptical bar graph without having the
laser beam on the cantilever, so it is important to visually verify that the laser beam
is on the cantilever and not rely on the elliptical bar graph alone. Attempting to
engage with the laser beam improperly aligned will usually destroy the cantilever.
5-21
Ok
STM
Edit
MultiMode
Extended MultiMode
EC AFM
EC STM
New
Delete
Cancel
Click on Ok using the mouse to close the Microscope Select dialog box.
Set the scanner configuration by using the mouse to select a scanner in the Microscope / Scanner menu.
Select Scanner
afmj
afmk
Ok
Cancel
TappingMode
For TappingMode, set the Microscope Mode to Tapping. This is done by clicking
on Microscope mode in the Other Controls panel. After this is completed, return
5-22
to the Scan Controls panel to access the most important control settings used for
imaging.
For contact AFM, set the Microscope Mode parameter on the Other Controls
panel to Contact. After this is completed, click on Scan Controls to again bring up
the most important control settings used for imaging.
B
A
Lateral deflection
Lateral deflection
C D
Vertical deflection
5-23
Contact AFM
The MultiMode SPMs bottom, elliptical (SUM) meter 1 indicates the total voltage
generated by the photodetector. That is, the combined voltage of photodetector segments. This is displayed during all modes (except STM when all meters are off).
TAPPING MODE
or CONTACT AFM
OUTPUT
SIGNAL (V)
SUM
1.2
12.0
(V)
10.8
DIFF.
3.6
VERTICAL
or HORIZONTAL
DIFFERENCE
8.4
4.8
6.0
7.2
SUM
The bottom digital meter 2 reads differences in voltage between various segments of
the photodetector. With the mode switch toggled to AFM & LFM, it indicates the
voltage difference (C - D), that is, the left segments minus the right segments. With
the mode switch toggled to TM AFM (TappingMode), it indicates the voltage difference (A - B), that is, the bottom segments minus the top segments.
The topmost digital meter 3 indicates the output signal of the SPM. Depending
upon the mode selected, the topmost meter reads either the (A - B) voltage difference (mode switch toggled to AFM & LFM), or the RMS voltage (mode switch
toggled to TM AFM).
5-24
Chapter 6
Contact AFM
Contact AFM
6-1
Contact AFM
shown on the computer screen detector graphic and top-to-bottom (short axis) on
the dark red laser spot screen mounted into the SPM head.
Photodetector vertical adjustment knob
(A - B)
Notes
Use of the laser aligning screws to adjust the laser sum signal is NOT advised;
users should adjust the laser sum signal from the photodetector adjustments
only.
6-2
Contact AFM
Vertical adj.
For initial adjustment, center the spot on the cantilever as described in Chapter 5.
Horizontal adj.
Laser spot
Notes
Large offsets are not recommended between engage and disengage (2 volts)
with etched silicon cantilevers in contact mode (450m long only) because
breakage is likely. With etched silicon contact probes, it may be necessary to
reduce the offset between engage and disengage to a value of 1.01.5 volts.
The large difference is recommended for first-time use of silicon nitride
cantilevers.
MultiModeSPM Instruction Manual
6-3
Contact AFM
The difference between the vertical deflection before engage and the setpoint is
related to the force. A larger, more positive setpoint voltage results in a larger contact force.
6-4
Contact AFM
Next, click on Panels / Feedback. Set both Integral and Proportional gain to 2.0
each and the Setpoint to 0.0 Volts. The Feedback Controls screen should appear
as shown below.
Feedback Controls
Integral gain:
2.00
Proportional gain:
2.00
LookAhead gain:
0.00
0.00 V
Deflection Setpoint:
Figure 6.3. Feedback Controls settings for initial setup (contact AFM).
Contact
2.000 m
Metric
2
Figure 6.4. Other Controls settings for initial setup (contact AFM).
6-5
Contact AFM
Height
Data scale:
300 nm
Line direction:
Trace
Scan line:
Main
Realtime Planefit:
Line
Offline Planefit:
Full
Highpass filter:
Off
Lowpass filter:
Off
6-6
Contact AFM
Panel
Z Scan Controls
Channel 1
Parameter
Setting
Scan Size
500 nm 1 ma
X offset
0.00 nm
Y offset
0.00 nm
Scan angle
0.00 deg
Scan rate
9.00 Hz
Number of samples
512
Average count
Display mode
Both
Trigger mode
Off
Start mode
Calibrate
Data type
Deflection
Z range
(Set to maximum)
What is happening?
The piezo is being retracted from the Z scan start position to the Z scan startplus-Scan size position. The Z scan start point is at the leftmost portion of the plot.
The Z scan start-plus-Scan size point is the rightmost portion of the plot. This corresponds to the Z Center Position that was being used while scanning just before
starting Force Calibration. This means that the Z scan start is recalculated at every
cycle to Force Calibration. The Scan size will be automatically changed if its value
is bigger than the difference between the Z Center Position and the fully retracted
position.
6-7
Contact AFM
Procedure
When you first enter forcecal the graph will probably look like this:
The tip is being moved up and down above the surface without touching it. If the
line is railed at the top or bottom of the image click on the Setpoint 0 button in the
Feedback Controls panel. Then you should see the above graph. To get it to touch
the surface you need to increase the Scan start. Doing this with an arrow key works
well. After a few key presses, you should see a graph similar to the one shown here:
If the line slants to the bottom of the graph and stays there, the tip is sticking to the
surface and not popping off the surface when the tip is fully retracted. To get the tip
clear of the surface, increase the Scan size until you can see it come off.
6-8
Contact AFM
Now adjust the Setpoint for imaging. Having the setpoint anywhere above the line
where the tip is off the surface will work. The farther it is above, the more force is
placed onto the sample.
To adjust tip-sample force to the minimum amount, you can run in the area where
the tip is actually pulling up but the liquid layer is holding the tip on the surface.
However, this is not a stable way to image: if the tip pops free of the surface, the
Setpoint must be increased to reattach the tip.
To minimize tip-sample forces, lower the Setpoint until the tip pulls free of the surface and record the Setpoint value: this is the pulloff value. Increase the Setpoint
again to regain the surface, then lower the Setpoint to a point just above the pulloff
value. At this point, tip-sample forces should be at their minimum.
If Setpoint is the only parameter requiring adjustment, skip ahead to section 6.5
below; otherwise, go to the next section to adjust the SPMs Sensitivity parameter.
6-9
Contact AFM
If you want to see more graph range you can change the Input attenuation to 8x.
The graph will not change in appearance until you make the graph range larger. It
can now be made eight times larger. With this you can see the bottom of those big
pulldowns. Be sure to change the Input attenuation back to 1x before imaging.
6-10
Contact AFM
If there are periodic waves in the plot of the tip in free air, these are caused by optical interference. Some of this is acceptable but it can be bad enough to be a problem. The only way to fix this without changing tips is to adjust the aim of the laser.
This can be done while in Force Cal.
6-11
Contact AFM
Silicon nitride cantilevers for this system are available in two process variations:
standard and sharpened. Usually, the standard silicon nitride cantilevers and resonant mode etched silicon probes are shipped with the MultiMode. Sharpened silicon nitride cantilevers have an almost identical appearance, but have slightly
sharper tips. Note that this system does not require the stand-alone type silicon
nitride probes which are used in some older, interferometric microscope heads;
however, they may still be used.
Each silicon nitride cantilever substrate includes four cantilever probes with different sizes and spring-constants. Two of the cantilevers on each substrate measure
115 m from the substrate to the apex of the triangular cantilever (these are referred
to as 100-m cantilevers) while the other two cantilevers measure 193 m from the
substrate to the apex of the triangular cantilever (these are referred to as 200-m
cantilevers). Both cantilever lengths are available with wide legs and narrow legs;
however, thickness of both cantilevers is the same. The calculated spring constants
for common cantilever configurations are listed below and in Appendix A of the
Command Reference Manual. These values are approximate; some variability will
occur. The tabulated values should be used to approximate the contact force unless
more accurate values are measured by the user.
Cantilever Type
k (N/m)
(narrow legs)
k (N/m)
(wide legs)
.38
.58
.06
.12
6-12
Contact AFM
6-13
Contact AFM
The LookAhead gain includes information from the previous scan line to determine the current gain setting. It should only be used on samples with step-like features which are oriented perpendicular to the fast scan direction. Otherwise, it
should be left at 0.
Notes
LookAhead gain controls are not implemented on some software releases. For
more information, contact Digital Instruments software department.
6.6.4. Setpoint
The Setpoint parameter defines the desired voltage (and, therefore, the desired
deflection of the cantilever) for the feedback loop. The setpoint voltage is constantly compared to the present photodiode cantilever deflection voltage to calculate the desired change in the piezo position. When the gain values are high, as they
should be when the Data type is set to Height, the Z piezo position changes to keep
the photodiode output signal close to the Setpoint; therefore, the cantilever deflection remains nearly constant. When the gain values are low as they should be when
the Data type is set to Deflection, the piezo height does not change, and the photodiode signal varies around the Setpoint value.
In contact AFM, increased Setpoint yields higher tip-sample forces. The Setpoint
can be adjusted to increase or decrease the cantilever deflection and, therefore, the
contact force of the tip on the sample. The Force Calibration command in the Real
Time / View menu allows the setpoint to be adjusted while viewing a graph of the
tip position versus the deflection voltage. Using this procedure, which is described
in detail in the next portion of this section, the contact force of the tip on the sample
can be minimized. This is especially important on soft materials such as biological
samples.
6-14
Contact AFM
6-15
Contact AFM
6-16
Chapter 7
7-1
Essentially, the procedure for observing samples under fluid is the same as that for
contact AFM in air; however, special hardware is utilized to contain the fluid. In
addition, minor adjustments must be made to correct for refractive effects as the
laser beam transits air-fluid boundaries. One feature of imaging samples under fluid
is that attractive forces due to surface tension effects are eliminated. This enables
the sample surface to be imaged with a minimum of cantilever tip forcea decided
advantage when imaging biological specimens and delicate materials.
Images of submerged samples may be obtained in TappingMode; however, the
drive frequency and amplitude of the cantilever must be reduced due to viscous
damping of the fluid. For more information, see Chapter 8. Chapter 8 includes a
procedure (section 8.8.3 ) that should be studied for all types of fluid imaging.
This section describes contact AFM operation of the MultiMode SPM in fluid,
including loading the cantilever into the tipholder, mounting the tipholder into the
head, aligning the laser onto the cantilever and then engaging the tip with the sample. There is also a discussion of methods of preparing the sample for fluid operation. This section assumes familiarity with contact AFM operation of the
MultiMode in air. If you are not familiar with air operation of the MultiMode,
please first follow the procedures outlined in Chapter 6 before attempting to operate
the AFM under fluid.
7-2
Fluid imaging requires a special tipholder (Figure 7.1). The fluid tipholder interchanges with the standard tipholder in a matter of a few minutes.
7-3
phobic, it should hold a fairly substantial droplet without difficulty. If the sample
surface is hydrophilic and/or porous, it may tend to absorb fluids and wet the scanner. In this case, it will have to be encapsulated within a fluid cell.
Fluid tipholder
Sample
Sample puck
Scanner
7-4
auf die oder nahe der Probenhalterung, der Elektronikbox oder anderen Komponenten, die elektronische Bauteile enthalten, verspritzt werden. Vermeiden Sie bitte,
korrosive Flssigkeiten auf freiliegende Oberflchen zu verspritzen; andernfalls
wren Beschdigungen die Folge! Falls Sie Flssigkeit verspritzt haben, subern
und trocknen Sie alle betroffenen Flchen sorgfltig.
7-5
Laser beam
Sample
Sample puck
Scanner
7-6
7.4.6. Install the Protective O-ring and Flood the Fluid Chamber
The o-ring is used to protect the SPM scanner tube from spilled liquids. It should be
inserted into the gland (recessed groove) which is ground into the underside of the
tipholder. The o-ring is designed to slide up into the gland as the tipholder is lowered until the tip accesses the sample surface.
Fill a syringe with the fluid to be used for imaging, then chase out all air bubbles.
(Injected air bubbles can become trapped inside the fluid chamber and disrupt
imaging.) Connect the syringe to the inlet tubing on the tipholder. The outlet tube
should be connected to a container where fluid can drain. Push enough fluid
through the fluid chamber to flood it completely, but do not clamp the outlet tube.
7-7
Typically, samples are softer in liquid than in air, so be careful that there is not too
large a difference between the setpoint and the vertical deflection signal before
engaging. Also, it may be desirable to reduce the setpoint once engaged to obtain
minimum tracking force.
7-8
7-9
7-10
Chapter 8
TappingMode AFM
TappingMode
Note:
TappingMode is disabled for users of NanoScope E configurations.
8-1
TappingMode AFM
nal") reveals information about the vertical height of the sample surface and some
characteristics of the sample material itself. These material characteristics may
include elasticity ("hardness"), magnetic and/or electric forces present.
Laser beam
Return signal
Cantilever
8-2
TappingMode AFM
Laser beam
Return signal
(deflected)
Sample surface
8-3
TappingMode AFM
Check that the Real Time control panel parameters are set within reasonable limits
for TappingMode operation. If you are uncertain what parameter settings to start
with, try the values below:
Panel
Scan Controls
Feedback Controls
Parameter
Setting
Scan Size
25 m
X offset
0.00 nm
Y offset
0.00 nm
Scan angle
0.00 deg
Scan rate
1.00 Hz
Number of samples
256
Enabled
Z limit
440 V
Integral gain
0.300
(V)
10.8
3.6
Proportional gain
0.300
LookAhead gain
0.00
Setpoint
6.0
Other Controls
Interleave Controls
8.4
7.2
Drive frequency
200400 kHzb
Drive amplitude
50800 mV
AFM mode
Tapping
Input attenuation
1x
Engage Setpoint
1.00
Interleave mode
Disabled
8-4
TappingMode AFM
RMS amplitude is an AC signal which does not have any real meaning until cantilever tuning is completed (see section 8.3.3). The laser spot will need to be approximately centered prior to entering the Cantilever Tune routine.
Photodetector adjustment knobs.
Photodetector output
signal (RMS AC volts).
This value is directly
influenced by the
Drive amplitude
parameter when tip
is at resonance.
Sum signal. This
value changes
according to how
much of the beam
falls on the
photodetector.
SUM
1.2
12.0
(V)
10.8
DIFF.
3.6
8.4
4.8
6.0
7.2
8-5
TappingMode AFM
SUM
1.2
12.0
(V)
10.8
er
Las
nal
sig
DIFF.
3.6
8.4
4.8
6.0
7.2
SUM
1.2
12.0
(V)
10.8
gna
r si
e
Las
DIFF.
3.6
8.4
4.8
6.0
7.2
Figure 8.4. Voltage meters on the MultiMode base reveal a great deal
about the amplitude and alignment of the TappingMode laser signal
on a tuned tip operating at its resonant frequency.
8-6
TappingMode AFM
Panel
Channel 1
Parameter
Setting
Data type
Height
Z range
04.50 ma
Line direction
Trace
Scan line
Main
Real-time Planefit
Line
Off-line planefit
Full
High-pass filter
Off
Low-pass filter
Off
8-7
TappingMode AFM
100.000 kHz
Target amplitude:
End frequency:
500.000 kHz
Peak offset:
2.000 V
0.00 %
Auto Tune
Sweep Controls
Graph Controls
Main Controls
Sweep width:
400.000 kHz
Amplitude setpoint:
Drive frequency:
300.000 kHz
Drive amplitude:
0.00 V
300 mV
256
Motor
Interleave Controls
8-8
TappingMode AFM
Note:
More than one type of cantilever exists. Different types have different dimensions
and therefore different resonance frequencies. Check the box used to ship the cantilever to see what its resonant frequency is, or refer to Appendix A in the Command
Reference Manual.
Tuning cantilevers manually
The parameter values, especially the drive frequency and the sweep width, given in
the following example apply to one type of cantilever. The nominal parameter values may vary depending upon the actual cantilever used.
For initial set-up, the Sweep Controls screens should be set to the values shown in
the example below.
Sweep Controls
Graph Controls
Main Controls
Sweep width:
400.000 kHz
Amplitude setpoint:
Drive frequency:
300.000 kHz
Drive amplitude:
0.00 V
300 mV
256
Motor
Interleave Controls
Sweep Controls
Graph Controls
Interleave Controls
Sweep width:
400.000 kHz
Amplitude Setpoint:
Drive frequency:
300.000 kHz
Drive amplitude:
0.00 V
300 mV
256
Motor
Main Controls
Figure 8.5. Sweep control panels for Main Controls (top) and Interleave
Controls (bottom).
8-9
TappingMode AFM
Set the Drive frequency parameter to a value near the center of the range of the
resonance frequencies specified for the cantilever. For example, if the frequency
range is specified as 240420 kHz, select a drive frequency of 330 kHz.
Start with a Drive amplitude of about 300 mV. It is possible to detach the
cantilever from the substrate by applying too large of a drive amplitude;
therefore, it is important to exercise some caution when adjusting the Drive
amplitude parameter. Note that the Drive amplitude is reset automatically by
the Auto Tune command.
Set the Sweep width to about the same value as the Drive Frequency, or at
least some value large enough to cover the frequency range specified for the
wafer.
Zero the Amplitude Setpoint. Note that the Amplitude Setpoint will be reset
automatically by the Auto Tune command.
3. If a peak in the frequency response plot does not appear, perform the following
steps:
8-10
TappingMode AFM
In reality, the system will work well in TappingMode if the drive frequency is at, or
below, the peak in the resonance plot. The Drive Frequency can be decreased to
the point where the vibrational amplitude reaches 90 percent of the maximum
value. It is often preferable to operate at a frequency lower than the resonant frequency due to shifting of the resonant frequency upon approach of the tip to the surface.
8-11
TappingMode AFM
Frequency Sweep
Cantilever
Response
0.50 V/div
Setpoint
8-12
TappingMode AFM
25.0 m
X offset:
0.00 nm
Y offset:
0.00 nm
Scan angle:
Scan rate:
Number of samples:
Slow scan axis:
Z limit:
0.00 deg
1.51 Hz
256
Enabled
440 V
8-13
TappingMode AFM
Metric
2
Tapping
1x
0.90
Disable
8-14
TappingMode AFM
8-15
TappingMode AFM
Figure 8.9 shows a typical response curve of the cantilever amplitude as a function
of time. To demonstrate the conflicting requirements, the performance of the system will be analyzed at two operating points.
Free Amplitude
x
Setpoint 1
Setpoint 2
t2
t1
Time
TappingMode AFM
Z Range
50.00 nm/div
Z Range
50.00 nm/div
8-17
TappingMode AFM
Setpoint
8-18
TappingMode AFM
8-19
TappingMode AFM
Note
8-20
TappingMode AFM
Distance fixed
by adjustment
screws
Retracted
Extended
8-21
TappingMode AFM
as a function of the voltage applied to the piezo tube) shows on the display monitor.
The control panel containing the parameters used to control the microscope in the
Force Calibration mode is visible on the control monitors top menu bar.
Force Calibration mode is used to compare the amplitude of cantilever oscillation
on the surface to the free-air amplitude. Other uses of Force Calibration mode
include characterizing the forces on the cantilever tip, diagnosing the performance,
and calibrating the RMS amplitude voltage as a function of the oscillation amplitude of the cantilever.
8-22
TappingMode AFM
the cantilever begins to increase. It continues to increase until the tip is free of the
surface, then it levels off at the free-standing amplitude.
Free Vibrational Amplitude
(tip is free of the surface)
RMS of Cantilever
Deflection Amplitude
Setpoint
+Z
(Scan start)
Z Piezo Voltage
-Z
(Scan start Scan size)
8-23
TappingMode AFM
100 nm
Other Controls
Trigger mode:
Z scan size:
500 nm
Trig threshold:
Z scan rate:
4.88 Hz
Trigger channel:
X offset:
0.00 V
Z step size:
Y offset:
0.00 V
Step threshold:
Number of samples:
Average count:
Sensitivity:
Units:
512
1
0.0379 V/nm
Off
0.00 V
Amplitude
2.40 V
7.53 V
Start mode:
Calibrate
End mode:
Extended
Indent Feedback:
Disabled
Metric
TappingMode AFM
using too high of a drive amplitude; therefore, it is safer to start with a small value
and increase the value incrementally. If the amplitude calibration plot consists of a
flat line all the way across, changing the value of this parameter should shift the
level of the curve. If it does not, the tip is probably fully extended into the surface
and the tip should be withdrawn before proceeding.
Sensitivity This item relates the vibrational amplitude of the cantilever to the Z
travel of the piezo. It is calculated by measuring the slope of the RMS amplitude
versus the Z voltage when the tip is in contact with the sample as shown in figure
8.12. The NanoScope system automatically calculates and enters the value from the
graph after the operator uses the mouse to fit a line to the graph. This parameter
must be accurately calibrated before Amplitude data obtained in Lift mode will be
accurate.
Note
8-25
TappingMode AFM
stiffness of the TappingMode cantilevers and the cantilever or tip can be shattered if
the deflection continues.
8-26
TappingMode AFM
8.8.1. Acknowledgments
Digital Instruments, Inc. wishes to express its appreciation to the following individuals for their assistance in preparing the following sections: Monika Fritz, Manfred
Radmacher, Magdalena Benzanilla, Helen G. Hansma, Jan H. Ho, Daniel Mueller,
Craig B. Prater.
8-27
TappingMode AFM
8.8.2. Overview
TappingMode in fluid combines the advantages of TappingMode with the ability to
image samples in native liquid environments (e.g., physiological buffer solutions).
Digital Instruments would appreciate receiving any further suggestions and is anxious to hear about new sample preparation techniques and experience with TappingMode in fluid.
Deionized water
Source of filtered (0.2 m), compressed air or dry nitrogen (chromatographicgrade)
While soaking in warm, soapy water, place a few drops of liquid dish soap on
the fluid cell and o-ring, then rub gently with a cotton swab or finger. Take care
not to scratch the glass surface with any abrasive material.
8-28
TappingMode AFM
O-ring
Sample or
substrate
Figure 8.17. Place the o-ring on the sample and slowly fill with fluid using a
micropipette.
8-29
TappingMode AFM
CAUTION! Do not add excessive amounts of fluid or the cell will overflow and
damage the piezo tube scanner.
ATTENTION! N'ajoutez pas trop de fluide. Sinon, la cellule dbordera et endommagera le scanner du tube piezo.
8-30
TappingMode AFM
Fluid Cell
Syringe
Liquid Drop
Figure 8.18. Flush the fluid cell before installation
to reduce bubble formation.
8-31
TappingMode AFM
When the fluid cell has been filled and bubbles have been removed, clamp off the
drain line with a pair of hemostats or similar clamp.
Mica
Teflon
12 mm
Stainless Steel
Sample Puck
To use this method of fluid imaging, align the SPM head and mounted fluid cell
(cantilever installed) with the dry sample puck first before applying the aqueous
sample. Allow 0.5 mm clearance between the tip and dry substrate surface. Once
the optics and laser are aligned, remove the SPM head and fluid cell. Take the dry
sample substrate off the scanner and apply the sample in a volume of approximately
25-40 l. The liquid should form a small dome over the mica and not spread onto
the surrounding teflon. Carefully replace the prepared sample puck. Slowly mount
the SPM head and fluid cell onto the sample. Since the aqueous sample will change
the light deflection, the laser photodetector will need to be re-adjusted. Course
adjustments are made by tilting the deflection mirror while fine adjustments are
made with the photodetector adjustment knobs on the left-top and left-rear of the
head. Adjust the photodiode position to give a deflection signal 0 V.
An alternative method to removing the SPM head after alignment is, with the dry
sample substrate in place on the scanner, slowly deliver 25-50l of sample solution
8-32
TappingMode AFM
Pipette
into the fluid cell using a standard micropipette (e.g. Eppendorf, Gilman) with a
200l capacity pipette tip. The end of the pipette tip fits into the fluid cell port holes
(refer to figure below). Inspect from the side to make sure the fluid is well confined
to the mica area only.
8-33
TappingMode AFM
Note
Over time, evaporation of the fluid may necessitate replenishing the fluid cell using
a standard micropipette. The pipette will fit conveniently into the fluid chamber
port.
The user should try to adjust and optimize these settings for each imaging condition and sample.
8.8.6.3. Select the drive frequency using the Cantilever Tune command.
This is similar to the Cantilever Tune process used for standard TappingMode in
air. Unlike operation in air, the cantilever resonance will be largely damped by liquid and the AutoTune function cannot be used.
8-34
TappingMode AFM
Enter the View / Sweep / Cantilever Tune menu to select a drive frequency. Users
of TappingMode in air will notice that there is no single, well-defined resonance
peak, but a number of broader peaks when viewing a wide cantilever sweep. The
user must manually select a peak that corresponds to the resonant frequency of the
cantilever being used. For the short, narrow Si3N4 cantilever, the resonant frequency in fluid is a broad peak centered around 10 kHz. Best results are achieved
by tuning the cantilever to a peak between 7-12 kHz. Higher and lower frequencies
have also been used depending on the type of cantilever employed. Start with a
Sweep width of 20 kHz and a Sweep setpoint of 9 kHz in the Sweep Control
menu. A typical Cantilever Tune screen is shown in Figure 8.20.
Note: Make certain the cantilever is very close to the sample surface when tuning
the cantilever under fluid.
8-35
TappingMode AFM
Note
The RMS amplitude must be adjusted when the AFM tip is near the sample
surface (< 50m).
8.8.6.6. Engage.
Use the Engage command to bring the tip into tapping range. The NanoScope software will automatically select a setpoint, then will stop the engagement when the
surface is detected.
Note
The slope of the Force Cal curve shows the sensitivity of the fluid TappingMode
measurement. In general, higher sensitivities will give better image quality. If
the sensitivity is very poor, check the mounting of the sample and fluid cell.
8-36
TappingMode AFM
8-37
TappingMode AFM
2. Lightly coating the area of the o-ring which contacts the sample surface with
white petrolatum or vacuum grease will allow the o-ring to slide across the surface,
minimizing lateral stress. This also forms a better fluid-tight seal between the o-ring
and sample. However, some solvents (i.e., nonpolar organic solvents) may dissolve
some of the lubricant into the fluid.
3. Replace the o-ring with a slice of thin-walled glass, plastic, or stainless steal tubing. The diameter and thickness of the ring of tubing should be chosen so that it
does not contact the inner or outer walls of the circular groove in the glass cantilever holder. This gives the optical head room to move laterally during engagement
and for positioning the tip over the sample surface. The height of the ring of tubing
must be chosen so that it is not too tall so that the tip cannot reach the sample surface, or too short so that the ring does not reach the bottom of the glass cantilever
holder before engagement. Glue the ring of tubing to the steel sample puck or to
the sample to prevent leaks.
Note
When using the ring of tubing in the fluid cell, fluid is blocked from exiting and
cannot be circulated through the cell.
4. When positioning the o-ring on the surface, adjust the positioning knob at the
base of the optical head to move it slightly forward. This will counter some of the
lateral stress on the o-ring which comes from the optical head moving back during
engagement.
Note
Remember that to prevent leaks when using the standard o-ring, it is best to
place the o-ring on the sample before placing the optical head on the scanner,
instead of inserting the o-ring into the groove in the glass cantilever holder
before placing the optical head on the scanner.
8-38
TappingMode AFM
8-39
TappingMode AFM
8-40
TappingMode AFM
proteins
buffer
pH<pI
+++
+++
mica
Figure 8.21. Proteins will typically bind to negatively charged mica when
the pH is reduced below the proteins isoelectric point, pI.
Protein binding procedure:
The following section gives a detailed procedure for preparing and imaging the protein lysozyme by TappingMode in fluid. The procedure was kindly provided by
Monika Fritz at the University of California, Santa Barbara and is described in the
following papers:
8-41
TappingMode AFM
Required Materials
Deionized water
Mica substrates
Lysozyme protein L-6876 from Sigma Chemical
Phosphate buffer solution, 10 mM KH2PO4, 150 mM KCl, pH 6 (buffer may be
adjusted for other proteins)
TappingMode AFM
Note
It is also possible to prepare samples inside the fluid cell by flowing the protein
solution through the fluid cell. In this case, it may be helpful to engage the tip in
contact mode with a zero Scan size to protect proteins from binding to the tip.
8-43
TappingMode AFM
DNA molecule
Buffer solution
Mica
Kasas, S., N.H. Thomson, B.L. Smith, H.G. Hasma, X. Zhu, M. Guthold, C.
Bustamante, E.T. Kool, M. Kashlev & P.K. Hasma (1997) "Escherichia coli
RNA polymerase activity observed using atomic force microscopy."
Biochemistry 36, 461.
8-44
TappingMode AFM
Required materials
Mica substrates
DNA: BlueScript II SK9(+) double stranded plasmid DNA, 2961 base pairs,
1mg/ml in 10mM Tris, 1mM ethylenediaminetetraacetic acid (EDTA) from
Stratagene, La Jolla, CA.
Buffer solution: 10 mM HEPES and 5 mM NiCl2 pH 7.6 (for loose binding and
air imaging), or NiCl2 (for tight binding and fluid imaging)
8.10.4. Acknowledgments
Digital Instruments, Inc. wishes to express its appreciation to the following
individuals for sharing their experience to assist in preparing this section: Monika
Fritz, Manfred Radmacher, Magdalena Bezanilla, Helen G. Hansma, Paul Hansma,
Jason Cleveland, Jan H. Hoh, Serge Magonov, Chris Johnson, Don Hersch, Tom
Kovaleski, Gouliang Yang, Jamie Vesenka, and Eric Henderson.
8-45
TappingMode AFM
8-46
TappingMode AFM
quickly lifted from the substrate surface, switched to TappingMode, then used to
image the sample before it becomes contaminated.
8-47
TappingMode AFM
8-48
Chapter 9
Scanning Tunneling
Microscopy (STM)
STM
9.1. Introduction
STM relies on tunneling current between the probe and the sample to sense the
topography of the sample. The STM probe, a sharp metal tip (in the best case,
atomically sharp), is positioned a few atomic diameters above a conducting sample
which is electrically biased with respect to the tip. At a distance under 1 nanometer
(.001 m), a tunneling current will flow from sample to tip. In operation, the bias
voltages typically range from 10 to 1000 mV while the tunneling currents vary
from 0.2 to 10 nA. The tunneling current changes exponentially with the tip-sample
separation, typically decreasing by a factor of two as the separation is increased 0.2
nm. The exponential relationship between the tip separation and the tunneling current makes the tunneling current an excellent parameter for sensing the tip-to-sample separation. In essence, a reproduction of the sample surface is produced by
scanning the tip over the sample surface and sensing the tunneling current. The first
STM operated in ultrahigh vacuum on cryogenically cooled samples. In the years
following its invention, many variations on the STM theme appeared. P.K. Hansma
and J. Tersoff wrote a good review article on the subject, containing experiments,
theory, and over 100 references for The Journal of Applied Physics, 61 pp. R1-23,
Jan 1987. Digital Instruments, Inc. introduced the first commercial STM, the NanoScope I, in 1986.
9-1
For samples having deeply relieved features, where AFM probes may not be
able to penetrate, and/or where feature verticality is very close to 90 degrees.
9-2
adjusts the height of the tip based on the tunneling current error signal, but if the
feedback gains are low (e.g., Integral gain < 1.0; Proportional gain < 0.5), the
piezo remains at a nearly constant height while tunneling current data is collected.
With the gains high (e.g., Integral gain > 1.0; Proportional gain > 0.5), the piezo
height changes to keep the tunneling current nearly constant, and changes in piezo
height are used to construct the image. The exponential relationship between tipsample separation and tunneling current allows the tip height to be controlled very
precisely. For example, if the tunneling current stays within 20 percent of the setpoint value (the current to be maintained by the feedback system), the variation in
the tip-sample separation is less than 0.02 nm.
9-3
Height
Z range:
30 nm
Line direction:
Scan line:
Retrace
Main
Realtime Planefit:
Line
Offline Planefit:
Full
Highpass filter:
Off
Lowpass filter:
Off
9-4
Current data is a measure of the tunneling current at each point tested on the sample. The DSP measures the voltage drop across a resistor in series with the tip and
calculates the tunneling current as the tip scans the surface of the sample. The tunneling current at each data point is recorded and displayed on the screen. It should
be noted that if different chemical species are present in a sample, the height data
may not be a direct representation of the topography of the surface of the sample.
Different atomic species within a sample may produce larger tunneling currents for
a given bias voltage.
The two scan modes (height and current) require subtle changes in the menu parameters to operate effectively. The parameter changes also affect the application of the
two modes. To operate effectively in collecting height data, the tip must closely
track the sample surface. The gains must be maximized to force the piezo to
respond quickly to the variations in the sample surface. The height mode is used for
most applications. Conversely, the gains must be very low to keep the piezo from
responding while collecting current data. After engaging, the tip scans the surface
of the sample with very little variation in the piezo height. This constant height provides a reference from which to measure and record the fluctuations in the tunneling current. The current mode is most useful for imaging atoms with relatively
small scan sizes.
9-5
9-6
Increasing the Setpoint current can also be helpful for larger scans. This has the
effect of raising the gain, but also brings the tip closer to the surface by a small
amount. High setpoint currents of 6 nanoamps or more can also be useful in
improving the signal-to-noise ratio for atomic images on some materials.
The Feedback type can be set to either Log, Boost, or Linear input transformations. Because the tip-to-sample separation is proportional to the log of the tunneling current, the transformation performed on the tunneling current prior to the
feedback calculation can have dramatic effects on the performance of the feedback
loop. Linear input is more protective of the tip, because the feedback error signal
responds exponentially to tip-sample separation. When the tip-to-sample separation
decreases, the error signal rises exponentially, quickly driving the tip away. However, the error signal is unsymmetrical. The same sample-to-tip separation change
that caused the tip to move away so quickly will generate a small error signal when
the tip is higher than it is supposed to be. This unsymmetrical response in linear
(Lin) mode will distort data. For this reason, the Boost and Log modeswith ln(I)
used in the feedback calculationare preferable for most samples, because they
respond in a more symmetrical fashion to positive and negative sample slopes. The
Boost mode is preferable for large scans with high vertical features such as compact disc stampers or integrated circuits. The proportional and integral gain can be
reduced greatly when the Boost mode is used. The log input has the advantage of
having a gain which is insensitive to the value of the Setpoint current.
Large scans cannot be taken at the same scan rate as small scans. When using the
large scanners with scans above a few microns, the Scan rate should be lowered
below 10 Hz. Best results can be obtained at scan rates of 1 Hz or less, although
image-taking is slow. At these scan rates, the 128 X 128 and 256 X 256 data formats are most useful, quadrupling and doubling the frame rate over the 512 X 512
format for a given scan rate. To ensure there is no image degradation due to a high
Scan rate, lower the rate and check for changes in the image. Check the Scope
Mode view to ensure that the scan is not slew-rate limited in Z, as evidenced by an
artificial sawtooth appearance in the scope trace.
The Highpass and Lowpass filter parameters provide the option of filtering the
tunneling current signal. A lowpass filter with a cut-off frequency of 25 KHz can be
applied to the analog tunneling signal in hardware. Typically, the filter is selected
for atomic-scale images; otherwise, no filtering should be selected.
9-7
9-8
9-9
4. Place the converter head atop the scanner, checking for clearance between the tip
and the sample as the head is lowered. If the tip appears to be too low and threatens
to crash into the sample, adjust the scanner screw(s) to obtain more clearance. The
head will sit securely atop the scanner under its own weight, assisted by the pull of
magnets mounted in the scanner.
5. Verify that the mode switch on the MultiModes base is toggled to STM. The
LED on the front of the base should glow blue. For "J" scanners, set the Scan Controls panel parameters to the values shown below. It is sometimes helpful to start in
Scope Mode (Real Time / View / Scope Mode) at a reduced Scan size. Once gains
are adjusted for the small scan, the Scan size may be increased.
Scan Controls
Scan size:
10.0 m
X offset:
0.00 nm
Y offset:
0.00 nm
Scan angle:
Scan rate:
Number of samples:
Slow scan axis:
Z limit:
0.00 deg
2.44 Hz
256
Enabled
440 V
Go to the Feedback Controls panel. To prevent crashing the tip into the sample,
verify that the Integral gain value is set at 12, and that the Proportional gain
value is set between 210. For now, turn off the LookAhead gain (0.00). Set the
Bias to 100.0 mV1.0 V, and Setpoint to 2 nA.
Notes
The Bias and Setpoint parameters must be set according to the conductivity of
the sample. Samples which are excellent conductors (e.g., gold) exhibit high
electron flow; therefore, the error signal is relatively large. Good conductors can
operate with a Bias of 150 mV. Conversely, if the sample is a semiconductor,
the Bias may have to be set higher (e.g., 100 mV) to obtain an image. Beware of
high bias values! If the Bias is set too high for a given conductor, electrical
arcing may occur, damaging the tip and/or sample.
6. At this point, the value of the head preamplifier offset may be measured using the
Offset command in the Real Time / Microscope menu. Checking the Offset is recommended as a good final check prior to engagement. The procedure is as follows:
9-10
Notes
Generally, the Date type field is set to Height on the Channel panel whenever
the Multimode is used in STM mode to image sample surfaces with large fields
of view. The Current setting is usually selected for imaging atomic features.
9. Click on the Scope Mode command in the View menu (or use the Scope Mode
icon) to see what the signal looks like for each scan. Return to Image Mode by
clicking on the View / Image Mode command or the Image Mode icon.
A poor tipeither because it has touched the sample, has contamination on the
end, or because of its manufacturingwill render a poor image. In this case, there
is little likelihood that the tip will recover and the best procedure is to replace the
9-11
tip. Pull down the Motor menu by clicking on Motor in the menu bar, then click on
Withdraw (or click on the Withdraw icon) to stop scanning, and retract the probe
from the surface. After the tip is withdrawn, the Secured message will appear in
the status bar of the control monitor. Remove the bad tip and install a new one. Verify there is sufficient clearance between tip and sample after installing the new
tipa new tip will extend a different distance out of the tipholder than the old tip.
To Improve Image Quality...
Vary the Scan size and the X offset and Y offset. Move to a new location on the
sample by changing the X offset and/or Y offset of the scan. The X and Y offset
parameters define the center of the scan. Often, a better image can be obtained
on a different portion of the sample. The X and Y offsets can be changed by
altering the values in the control panel, or by using the Offset command in the
menu bar on the display monitor.
Alternate the commands Withdraw and Engage a couple of times. This raises
and lowers the tip and may get rid of contamination on the end of the tip.
Use the Step Motor command to single-step the tip up a step or two. Be sure to
minimize the SPM step size before stepping the motor down. Watch the Z
Center Position scale on the display monitor to verify that the piezo has
sufficient room to retract so the tip will not harm the surface of the sample.
Stepping the tip down until the Z Center Position goes to the retracted end of
the scale where the indicator turns red will usually destroy the tip.
Vary the Setpoint current over the range 2 to 10 nA. As a last resort, current
may be increased up to 48 nA for a very brief period of time.
As a last resort, oscillate the tip by setting the Integral gain high (about 100)
for a second. The oscillation will show up on the display as a grainy pattern of
light and dark. Set the Integral gain back down and check if the signal is
quieter than before. Setting the Integral gain high for a brief period is also useful
for cleaning debris from the tip.
If none of the above procedures improves the quality of the signal, replace the
tip and try again. If that doesn't work, call Digital Instruments for technical
assistance.
10. After a clean image has been obtained, it can be saved onto disk with the Capture command in the Capture menu (or click on the Capture icon) . The Capture
command saves the current scan image (or the next image if any changes have been
made in the STM control panel) into the Capture directory. The Capture Filename command can be used to define a custom filename or the captured file will be
named with a date/time stamp. In either case, a three-digit suffix which is incremented after each capture operation is appended to the filename. If the date/time
stamp is used, a file captured at 3:32 p.m. on July 31 would be assigned the file
9-12
name 07311532.001. A second file captured the same minute would be named
07311532.002, etc.
11. Some simple filtering operations can improve the look of the captured image.
Enter the Off-line mode and select an image by clicking on it in the file list. There
are a number of image processing options available in the Off-line / Modify menu.
Click on Modify in the menu bar to display the menu. The simplest way to remove
high frequency noise is using a lowpass filter.
Pull-down the Modify menu and click on the Lowpass command to apply the lowpass filter to the image and redraw it on the display monitor. Successive Re-execute
clicks will continue to filter and redraw the image. The entire filtering operation can
be undone with the Undo command, the filtered image can be saved, or the Quit
command will return the system to the Modify menu. There is also the option of
changing the display with the Color table, Color contrast, or Color offset parameters.
Notes
The Z range parameter will have no effect if the Plot type parameter in the Top
View menu is set to Equal area.
9-13
STS i(v) The tunneling current as a function of the bias voltage is displayed. The
tip height is held constant while the current (I) versus voltage (V) plot is being
acquired. In addition to I versus V, it is also possible to plot the following:
di
------ versus V
dv
d ln ( i )
---------------versus V
dv
d ln ( i )
----------------versus V
d ln ( v )
STS i(s) The tunneling current as a function of the tip height is displayed. The
bias voltage is held constant while the current (I) versus tip height (S) plot is being
acquired. In addition to I versus S, it is also possible to plot ln ( I ) versus S.
9-14
2. The maximum input range of 100 nA on the NanoScope with the standard
preamps will restrict the current-versus-voltage plots.
3. The spectroscopic functions do not adjust for any loss of bias voltage due to IR
losses caused by the input impedance of the preamplifier. This has different effects
for the STS i(v) and STS i(s) functions which are discussed individually below:
STS i(v)
The spectroscopic plots do not correct for reductions in bias voltage caused by the
preamplifier, which for current-versus-voltage is typically small, anyway. For
example, a 1-volt scan on the bias voltage, producing a 50 nA response in the tunneling current would have a 5 percent error at the extremes of the scan (50 nA X 1
M = 50 mV or 5 percent of 1 volt).
Input impedance of the preamplifier leads to reduced bias voltages at increased tunneling currents. This effect can cause the I vs. S plot to be inaccurate. For example,
a bias voltage of 20 mV will restrict the upper limit of the tunneling current to 20
nA since that level of current would effectively reduce the bias voltage to zero.
Even before the tunneling current gets to 20 nA, the reduced bias voltage will make
the measured current appear lower than it should be for a given tip height.
9-15
Notes
Handle the converter head as little as possible during these operations to prevent
thermal drift.
4. Place the head on the three magnetic balls mounted on the threaded screws in the
scanner. (Make sure that the screws are extended enough so that the tip does not hit
the sample.) The head has a hole, a groove, and a flat which mate kinematically
with the three magnetic balls. A suggested method of placing the head on the scanner is to slowly lower the head while it is slightly misaligned until its underside
touches the screws, then slowly rotate the head until the hole and groove on the
underside mate with the screws (all the while watching the tip-sample separation).
If it looks like the tip will hit the sample, toggle the motor switch Up to raise the
head and/or use the coarse adjustment screws.
After the head is securely mounted on the magnetic balls of the base, insert the 15pin preamp plug into the socket on the scanner support ring. When placing the head
down onto the balls, try to prevent it from slamming onto the balls, as excessive pitting in the flat area can cause irregular engagement.
5. Looking at the tip with the 25X magnifier, lower the tip down with the coarseadjustment screws until the tip is about a tenth of a millimeter or less (about .004
9-16
inch or less) above the sample. It is usually easier to use only one coarse-adjustment screw to do this if you do not need to move the tip too far. Try and adjust the
screws such that the head is not tilted to one side or the other. If necessary, adjust
the rear screw so that there is a minimal amount of tilt in the head from front to
back. A reflection of the tip will be visible on the sample. Using the coarse-adjustment screws, lower the tip until there is only a slight gap between the end of the tip
and its reflected image do not bring the ends together. If you do, you will almost
certainly have to replace the tip. Don't worry at this point if the head is tilted.
If you haven't placed the microscope on the vibration-isolation platform, do so now.
Verify that the mode switch on the left side of the MultiModes base is toggled to
STM.
6. Be sure the power to the controller is on. If you have not already done so, start
the program by entering "Z".
7. Set the scan parameters. Pull down the microscope menu by clicking on the Real
Time / Microscope /Select command, then click on the scanner being used. Click
the Ok button to select parameters for the converter head. The "A" scanner is the
preferable option for imaging on an atomic scale, but the "E" scanner has also been
used effectively.
To look at graphite atoms, it is necessary to set the Scan size to about 520 nm.
Use the mouse to change the parameter. Select the parameter and drag the
mouse right and left with the left button pressed to change the parameter; the
new parameter value will take effect when the mouse button is released. The
right or left arrow keys can also be used to set this parameter. Note that the up
and down arrow keys move from one item to the next, while the right and left
arrow keys change the selected parameter. Numerical parameters may also be
set by simply typing over the existing values and pressing the enter key.
Set the remaining parameters so that the Feedback Controls panel is as shown
in figure 9.3 below.
8. As a final check, click on the Offset command in the Real Time / Microscope
menu. The offset should be very close to 0.0 nA. Proceed to the next step if the system reports a reasonable value.
If the offset reads 99 nA, the feedback loop is open (i.e., the microscope or controller cables are not properly installed), or the tip may be in contact with the sample surface. Check this by clicking on the Withdraw command and retesting the
offset . Remedy this problem before continuing.
9-17
Feedback Controls
Integral gain:
2.00
Proportional gain:
2.00
LookAhead gain:
0.00
Bias:
20.0 mV
Setpoint:
2.00 nA
Analog 1:
0.00 mV
Analog 2:
0.00 mV
Notes
If version 4.22 or higher software is used, the STM must have the In polarity
parameter set to Forward on the Microscope / Calibrate / Detector panel. On
older microscopes, the setting didnt matter and may have been set to Reverse.
This setting will cause false engagement with newer Z.exe programs.
10. At this point, an image should begin to form on the screen. Depending on the
condition of the tip, the image on the screen will be either clear or noisy. Let's treat
the two cases separately:
9-18
A good signal is one which clearly shows the repeating pattern of the carbon
atoms. Compare this image with the graphite image that may be stored in the
C:\DATA directory. Check to see how the image repeats on scans from bottomto-top and top-to-bottom. Usually, when the tip is first engaged, there will be a
drift in the piezo which will cause the image to repeat poorly. The most obvious
sign of drift is features which are compressed vertically in one frame and
vertically elongated in the next frame. This drift will decrease with time.
Click on a scan parameter and drag the mouse to change the parameter. Lower the
Integral and Proportional gains near zero and vary the Bias voltage to see how
the image changes. Vary the Scan size to get a feel for how the image can be magnified and vary the X and Y offsets to see how the image can be moved. Instead of
displaying an image of the tunneling current, select Height at the Data type line to
display the Z (vertical) signal. You will have to increase the gains and decrease the
scan rate to get a good Z-height image.
To use an "E" or "J" scanner on an atomic scale with the Data type set to Height,
the Z limit parameter (Scan Controls panel) should be decreased to 55 Volts.
Click on the Scope Mode command in the Real Time / View menu to see what the
signal looks like for each trace across the sample. Return to the Image Mode by
clicking on the View / Image Mode command, or icon.
A poor tip, either because it has touched the sample, has contamination on the
end, or because of its manufacturing, will give a noisy signal. In the worst case,
the image will have no pattern at all, and Z on the display will run off scale in
both directions. In this case, there is little likelihood that the tip will recover, and
the best procedure is to replace the tip. Pull down the Real Time / Motor menu,
then click on Withdraw (or click on the Withdraw icon) to stop scanning, and
retract the probe from the surface. After the tip is withdrawn, the "Secured"
message will appear in the status bar of the control monitor. Before removing
the head, it's a good idea to execute the Withdraw command several times to
protect the sample from being scratched when the head is picked up. Extend the
scanner screws a few turns, unplug the preamp, remove the head, and replace
the tip. Try to leave the new tip protruding the same amount as the old one. Be
sure there is sufficient clearance after installing the new tip, because the new tip
will extend a different distance out of the tipholder than the old tip.
The nominal Z position of the piezo provides useful diagnostic information. The
bar graph to the right of the image on the display monitor reflects the Z position
of the piezo during the scan. Under most circumstances, the "Z center position"
bar should be in the center of the graph to indicate that the Z voltage for the
piezo is fluctuating around zero volts.
9-19
If the sample is not level or if there is drift in the mechanical system after the tip has
engaged, the "Z center position" will fluctuate during the scan or drift off scale. If
the "Z center position" reaches the "Extended" end of the scale, the maximum voltage is being applied to the piezo, but the tip still cannot reach the surface. Similarly,
the "Retracted" end of the scale means that the maximum voltage of opposite polarity has been applied to the piezo, and the tip cannot be lifted from the surface. The
"Retracted" end of the scale is more dangerous, because damage to the tip or sample can result.
If the "Z center position" gets close to either limit, use the Withdraw command followed by the Engage command to reposition the "Z center position" near the middle of its range. If there is drift in the Z center voltage, use the Real Time / Motor /
Step Motor command with a small Step size to keep the tip within the Z piezo
range. This drift will eventually settle out.
EXCEPTION: The Z limit represents only a portion of the maximum Z range
(440 V). When using a Z limit of less than 440 V, version 4.xx software will
attempt to recenter the Z-axis if you double-click on the Z limit value or change it.
Sometimes the tip will show atomic structure on graphite with noise which is large
enough to reduce the definition of the structure. In these cases, it may be possible to
reduce the noise without replacing the tip. Any or all of the following procedures
may be tried1:
Vary the Scan size and the X and Y offsets (Scan Controls panel). Move to a
new location on the sample by changing the X and/or Y offset of the scan. The
X and Y offset parameters define the center of the scan. Often, a better image
can be obtained on a different portion of the sample. The X and Y offsets can be
changed by altering the values in the control panel or by using the Offset
command in the menu bar on the display monitor.
Alternate the commands Withdraw and Engage a couple of times. This raises
and lowers the tip and may get rid of contamination on the end of the tip.
Use the Real Time / Motor / Step Motor command to index the tip down a step
or two. Be sure to minimize the Step size before stepping the motor down.
Watch the "Z center position" scale on the display monitor to be ensure the
piezo has enough room to retract so the tip will not harm the surface of the
graphite. Stepping the tip down until the "Z center position" goes to the
retracted end of the scale where the indicator turns red will usually destroy even
a moderately noisy tip. Sometimes, though, bad tips can be reconditioned by
being driven several steps into the surface and then withdrawn and re-engaged.
1. These suggestions also apply to larger-scale STM imaging and are provided in section
9.2.1 of this chapter.
9-20
Vary the Setpoint current over the range 210 nA. As a last resort, you can
increase it up to 48 nA for a very brief period of time.
As a last resort, oscillate the tip by setting the Integral gain high for about one
second. The oscillation will show up on the display as a grainy pattern of light
and dark. Set the Integral gain back down and check if the signal is quieter than
before.
If none of the above procedures improves the quality of the signal, you should
cleave the sample, replace the tip, and try again. If that doesn't work, review all
parameter settings or call Digital Instruments for technical assistance.
9-21
7. If you have not already done so, start the main program by entering Z.
Integral gain:
Proportional gain:
5.00
LookAhead gain:
0.00
Bias:
500.0 mV
Setpoint:
5.00 nA
Analog 1:
0.00 mV
Analog 2:
0.00 mV
Change the Integral or Proportional gain parameters. Gain values which are
too high will cause graininess in the image, whereas low gains will cause image
smearing, particularly where the surface has sharp, abrupt features.
Vary the Scan rate. A Scan rate which is too high will cause smearing (like a
low gain setting) and lower scan rates may test your patience.
9-22
9-23
fouled with debris and binding. Remove the screw, clean threads and regrease per
instructions provided in section 15.3 in this manual.
bias shortedMeasure the bias by using a voltmeter between the head and stage
chuck. If this is not in agreement with the settings in the Bias voltage item in the
STM Parameter control panel and appears to be grounded, then check to see if anything is providing a conduction path between the Base and the Base Support or any
other ground.
bias not applied to sample surface In some cases (e.g., the calibration reference), the sample consists of a layer of electrically conductive material on top of an
insulator. In these cases, a conduction path must be provided between the sample
puck and the sample surface. Carbon glue or silver epoxy can be used to connect
the bias voltage to the sample surface. With NanoScope software versions 4.23b3 or
later, verify that the In polarity parameter on the Detector Calibration panel
(Real Time / Microscope / Calibrate / Detector) is set to Forward.
9-24
clean contacts The magnetic balls on the ends of the adjustment screws tend to
attract small metallic particles which may make the surface gritty. Clean them with
adhesive tape and then wipe with alcohol. Also, clean the kinematic mating surfaces on the underside of the head.
check for binding in the motorized screw Make certain the motorized adjustment screw rotation is smooth and not erratic, which can cause the screw to
advance in jumps. If this is the problem, clean the screws as described in section
15.3 in this manual. On newer scanners, the setscrew used to adjust tension on the
motorized screws plastic threaded bushing may need to be slightly loosened.
check sample conductivity there are two problems associated with sample conductivity. First, the bulk conductivity of the sample may make it more difficult to
image. If the resistance of the sample is greater than 1 k/cm, higher bias voltages
should be tried. If the resistance is greater than 1 M/cm, bias voltages of 100 mV
or more should be used. Samples with resistances of 1 M/cm or greater will be
difficult to image even with high bias voltages.
Measuring the bulk conductivity of the sample with probes may not tell the whole
story. Probes may easily penetrate oxide or contamination layers on the sample surface yielding resistance measurements. However, oxide and contamination layers
on the sample surface can make imaging very difficult. Higher bias voltages are
required for these types of samples.
check the polarity of the Z-axis piezo Use the Calibrate command in the Real
Time / Microscope menu to review the calibration parameters for the head in use.
The Z polarity should be set to Forward. (EXCEPTION: A few rare scanners
have serial numbers ending in "RP," reverse polarity; these should be set to Reverse
instead.)
9-25
sample surface is parallel to the sample puck before mounting it atop the scanner
cap. Also verify that the channel panel (e.g., Channel 1) has its Realtime Planefit
parameter set to Offset or None.
Increase Z limit parameter If set too low, try increasing the Z limit parameter
value (Scan Controls panel).
9.5.7. Z drift
If the Z center voltage tends to drift out of range rather quickly after the head is
engaged, it may be a result of the following:
thermal drift Allow some time for the temperature to stabilize if the microscope
and/or scanner has been stored in a cold place overnight. Drift can be minimized by
keeping the STM in a thermally stable environment. Avoid operating the microscope near windows and air ducts.
tip not tight Make sure the tip is held tightly in the tipholder. Push and pull on
the tip with tweezers to see that it is not loose. (On older tipholders, put a slight
bend in the tip to help it mount firmly in the tipholder tube if it seems loose.)
sample hold-down Verify that the sample is flat against the puck and cannot
move vertically. This may be become problematic if the sample has been affixed
using uncured, conductive epoxy, and/or if foam adhesives are used.
scanner screws-converter head contact Clean mating surfaces between the
underside of the converter head and the scanner adjustment screws. If dirt or oil are
evident, carefully remove with a clean wipe and alcohol. Some scanners utilize
plastic threaded bushings in the scanner body through which the adjustment screws
are threaded. If the bushings become fouled with grit, or if they are loose, drift may
result. If loose, gently tighten setscrews in the scanner body to increase tension on
the adjustment screw bushings. (Do not overtighten!) For more information regarding removal and replacement of adjustment screws, refer to section 15.3 in this
manual.
9-26
stiff rear screw If the rear screw is difficult to turn (due to an accumulation of
dust in the threads, etc.), the drive shaft can wind up while driving the screw and
unwind after the stepper motor has stopped. This will cause the tip to continue
towards the surface. Clean the threads and screw per section 15.3 in this manual.
9-27
GainThe settings of the Integral and Proportional gain parameters also tend to
be critical for large scans. They should be just sufficient to produce a slight fuzz on
the image, but not so high as to cause oscillations.
Scan rateThe Scan rate should be lowered for large scans, especially, if the sample surfaces are rough or contain large steps. Moving the tip quickly along the sample surface at high scan rates with large scan sizes will usually lead to a tip crash.
Essentially, the Scan rate should be inversely proportional to the Scan size, since
the tip must still be maintained roughly 1 nanometer above the surface.
Setpoint currentRaising the Setpoint current will effectively raise the gain of
the feedback loop which can be quite helpful for large scans. It will also bring the
tip closer to the surface but only by a small amount (i e-s).
9.7. Description
Operation of conventional STMs with current (Itun) in the nanoamp range is characterized by strong tip-sample force interactions. The forces applied to the sample
during STM imaging in air actually exceed the forces in atomic force microscopy
(AFM). The effects of high STM forces were found when imaging materials which
included graphite [1a], inorganic layered compounds[1b], organic conducting crystals [1c] and organic adsorbates on conducting substrates [1d].
9-28
For metallic surfaces, the tip-sample gap resistance, R gap = V bias I tun , can be used
as a qualitative measure of tip-sample separation. Where
V bias = bias voltage
I tun = setpoint current
In general, decreasing I tun will increase the tip-sample separation and decrease the
tip-sample force.
By operating at R gap in the G range, for example, V bias = 1.0 V
and I tun = 1 pA the tip-sample distance increases enough to allow less-destructive imaging of a variety of surfaces.
The ability to conduct STM measurements in a wide range of Rgap can also be useful for experiments aimed at observing the image variations as a function of Rgap.
Picoampere-level STM provides the opportunity to image insulating samples
deposited on conducting substrates. Indeed, it was shown that using STM at ultra
high Rgap allows imaging of the topmost layers of alkanethiol adsorbates on gold
surfaces and to reveal earlier unknown surface features.
Picoampere-level STM also opens the possibility of imaging biological macromolecules such as DNA deposited on a non-conducting mica substrate. Though such
imaging was explained by lateral conductivity across the water overlayer covering
surfaces in air, a true mechanism of electron transfer is not yet established.
9-29
9.9. Precautions
Due to the low-current heads highly sensitive electronic components, special precautions must be taken.
1. The input stage operational amplifier is extremely vulnerable to electrostatic discharge (ESD) and is easily destroyed. It is recommended that the user be grounded
with a wristband at all times that the low current head is handled, especially when
tips are installed.
2. The feedback resistor and capacitor in the input stage are extremely vulnerable to
contamination. At no time should these delicate components ever be touched with
the hands. If these components are accidentally touched, they must be rinsed clean
in a highly purified degreasing fluid immediately.
9-30
Offset current values vary slightly from head to head but are typically in the
range of 0.5-0.1 pA. Therefore, minimum operating currents are just slightly
below the 1 pA value.
5. Please note that cables connecting the SPM base, the Picoamp Boost Box and the
NanoScope controller should not be lengthened if possible.
IMPORTANT: Always disconnect the Picoamp Boost Box and Low-current STM
Converter head before performing other types of imaging.
9.10. Installation
The low-current STM converter is designed to be installed onto one of four Digital
Instruments devices: 1) the MultiMode SPM (standard); 2) the MultiMode SPM
(with Extender base); 3) the dedicated AFM; and 4) the dedicated LFM. Generally, installation consists of cable connections and nothing more; however, cabling
differs somewhat between the devices.
9-31
3. Connect the picoamp booster to the SPM base using the 25-pin-to-25-pin cable
connector.
4. Connect the low-current converter head to the MultiModes support ring using
the micro-D, 15-pin connector cable. Set switches on the picoamp booster to the
desired current level and bandwidth.
6. Before repowering the system, double-check all cable connections. If atomiclevel scans are intended, verify the cable between the SPM and picoamp booster
box is unstrained (a taut cable will transmit vibrations). The cable should rest
loosely. Before loading samples and running the SPM, attach the ESD wristband to
prevent sudden electrical discharges.
Controller
37-pin-to-37-pin cable
Extender Electronics
Module (if equipped)
37-pin-to-25-pin cable
Low-current
Converter Head
25-pin-to-25-pin cable
Picoamp Boost Box
MultiMode
Base
9.11. Operation
Once you check that leakage is absent and the offset current is small, the operation
of the low-current STM head is typically checked in the atomic-scale imaging of
graphite and the large-scale imaging of a gold-coated grating. The atomic-lattice of
graphite is well resolved in the images obtained with Itun in the 2-20 pA and Vbias in
the 20-100 mV range. Mechanically-sharpened Pt/Ir tips are generally used for
such measurements, and a tripod is good enough for the vibrational isolation of the
SPM base.
9-32
9.11.1. Peculiarities
The transition to low current, however, brings some limitations, which are absent in
STM measurements with standard heads. The bandwidth of the amplifier and, consequently, of feedback is substantially narrower than that used for standard measurements in the nA range (12 kHz). Therefore, the feedback mechanism can be
less effective at high scan speeds. This must be taken into account, especially during imaging of corrugated surfaces.
However, the low current STM head can be used for imaging of atomically flat surfaces such as graphite and inorganic layered crystals at line frequencies up to 50 Hz
and still achieve atomic-scale resolution.
NanoScope II Users: The low-current STM converter head may be used with the
AFM base and the NanoScope II version 5.5 software. The software should be run
in TipView mode. Readings will be seen in nanoamps; however, actual values are in
picoamps.
9-33
9-34
9-35
2. Use the 0.050 allen wrench, included in the converter kit, to loosen the two
retaining screws which secure the heads cover. It is not necessary to remove the
screws.
NOTE: Throughout steps #4-6 below, avoid touching the precision resistor and
capacitor. Any contaminant from the hands, including body oils, may cause excessive current flow, disabling the head. If these components are touched, they will
have to be recleaned immediately using highly purified degreasing compounds.
4. Remove the op-amp by grasping it and pulling straight up. It should slide out of
its socket.
5. Remove the new op-amp from its protective package. The op-amp is stored in
conductive foam and its leads have been preformed to fit the socket. Be careful not
9-36
to bend any of the leads. Orient the op-amp so that the metal tab on the edge of the
can is in the 10:00 position as shown below.
To remove, pull
straight up.
To install, push
straight down.
(Do not bend leads.)
6. Verify that each of the op-amps wire leads is properly aligned with the appropriate hole, then press the op-amp gently into its socket. Do not bend wire leads.
7. Replace the metal cover on the head. Retighten retaining screws to secure.
Distilled water
Ethyl alcohol
WD-40 (anti-oxidant)
Two 50-ml beakers
Tipholder (see below)
Platinum wire
Tungsten Wire, 0.010" Diameter
Miscellaneous Wire/Clips
9-37
9.13.1. Procedure
1. Mix a 5 percent (by weight) solution of Sodium Nitrite in water.
2. Pour 40 ml of the Sodium Nitrite solution into a beaker.
3. Pour 40 ml of WD 40 into a beaker.
4. Construct an electrode out of the platinum wire and insert it into the beaker.
5. Adjust the variac for 30V, and with it Off, connect one output to the platinum
electrode.
6. Cut 10 to 12 pieces of tungsten wire 1.25 cm long. Before etching, check to
make sure that at least one end of the wire has not split by inserting the ends into a
tip holder. If an end has split, you will not be able to insert it into the tipholder. You
can etch the end that has split, however, preserve the unspoiled end.
7. Place the tungsten tips into a holder. We like to use an IC socket (the low cost,
edge-grip square contact type, not the machine-grip round contact type), with all
the pins soldered together. The tips will then be held in place while inverting them
over the solution. We also find it helpful to solder the IC socket to the back of a
proto-board (perf-board). You can then invert the tips over a beaker with the protoboard sitting on the rim of the beaker.
8. Invert the tips over the Sodium Nitrite solution with 2 mm of the tips surface
submerged. More than 2 mm will cause excessive foaming of the solution during
etching, and less than 2 mm will result in tips that are too blunt.
9. Connect the other output of the variac to the common of all the tips.
10. Turn on the variac and etch the tips. While the tips are etching, the solution will
foam, and the tips will start to glow. As the tips etch towards the surface, the foaming will be reduced. Continue to etch the tip until it stops.
11. Re-submerge the tips 1mm into the solution for 15 seconds at 30V. Turn on
the variac and re-etch the tips. There should be only slight bubbling from the tip and
it should not glow.
Notes:
It possible to vary the tip shape at this point by lowering the voltage on the
variac and increasing or decreasing the amount of time the tips are submerged.
Longer time gives blunter tips.
9-38
12. Dip the tips into ethyl alcohol to clean them. If you plan to keep the tips around
for more than a day, dip them into the WD 40 after cleaning.
13. Examine the tips under the optical microscope. Ones that are too long, too
blunt, or split at the end will hardly ever be good tips and can be thrown out at this
time. Of course, this is a subjective process. As your experience in etching grows,
you will get better at throwing out the bad ones.
14. Repeat the etching procedure. Replace the etching solution when a fairly large
amount of residue is present. Typically, you can etch 60 to 80 tips in a 40 ml solution.
9-39
9-40
Chapter 10
Lateral Force
Mode
10.1. Introduction
The MultiMode SPM is capable of measuring frictional forces on the surfaces of
samples using a special feature known as lateral force microscopy (LFM). The
name derives from the fact that cantilevers scanning laterally (perpendicular to their
lengths) are torqued more as they transit high-friction sites; low-friction sites tend
to torque cantilevers less. The relative measure of lateral forces encountered along a
surface yields a map of high- and low-friction sites.
After obtaining a good topographical image in AFM mode, it is relatively easy to
switch to LFM mode to view and acquire lateral force data. It is important to obtain
a good image in AFM mode before switching to LFM mode. The NanoScope system will continue to run the feedback based on the AFM signal and feedback gains
in the control panel while LFM data is acquired and displayed. Complete the following steps to switch the system to LFM mode:
1. Set up and run the system in contact AFM mode as described in Chapter 6,
assigning the Channel 1 image to Data Type: Height and the Channel 2 image to
Data Type: Friction. Set the Scan angle to 90.00
2. Optimize the scan parameters in AFM mode for Channel 1.
3. For Channel 2, set Line direction to Trace. This will place high lateral forces
on the top of the color bar and low lateral forces on the bottom of the color bar. If
the Line direction parameter is changed to Retrace, the plus-minus sign of the lateral force changes and low lateral force will be displayed on the top of the color bar.
LFM data can now be collected and stored.
10-1
Scan Direction
without Scan
Rotation
scan angle 0
Scan Direction
for Friction
Measurements
scan angle
90 or 270
Cantilever
(Top View)
10-2
The contact mode setpoint voltage will slightly adjust the gain of the lateral force
signal. By increasing the setpoint, the contact force applied will increase, and so
will the frictional or torsional forces in an approximately linear fashion. If the frictional effects are far too large or too small, it will be necessary to resort to changing
the cantilever probe used, but if the value is near the dynamic range desired, adjustment of the contact force will produce modest changes in the lateral force or frictional signal.
Tip selection The analog-to-digital converter on the auxiliary input channel
which is used for LFM data has a maximum input range of 10 volts. This, and the
anticipated interaction between tip and sample define the selection of the cantilever
to be used for the measurement.
The 200-micron cantilever with wide legs provides a good starting point for frictional measurements. It is flexible enough to provide reasonable signal levels on
samples with moderate friction. If the signal exceeds + 10.00 volts with the 200-m
wide-legged cantilever, one of the stiffer 100-m cantilevers should be used. If the
signal level is too small, the narrow-legged 200-m cantilever will provide a larger
signal.
10-3
Trace
Retrace
Z Range
0.10 V/div
10-4
Reverse scan direction Reversing the scan direction in the Image Mode while
viewing the friction channel, is also useful in verifying the origin of the data.
Remember that changing the Line direction parameter in the Real-time control
panel changes the direction of the scan during which data is collected. If the data
results from friction between tip and sample, the relative signal strengths will invert
as the scan direction is reversed. For most color tables, the image produced from
the trace in figure 10.2 (i.e., with the Line direction parameter set to Trace) would
be darker in the low-friction areas than it is in the high-friction areas. Conversely,
for the same color table, the image produced from the Retrace (i.e., with the Line
direction parameter set to Retrace) would be lighter in the low-friction areas.
10-5
Trace
Retrace
Z Range
0.08 V/div
10-6
Trace
Retrace
4
Z Range
0.10 V/div
10-7
10-8
Force Imaging
11.1. Introduction
Force plots are used to measure tip-sample interactions and determine optimal setpoints. More recently, microscopists have begun to plot force measurements across
entire surfaces to reveal new information about the sample. This area of SPM promises to open new chapters in materials science, biology and other investigative
areas.
Chapter 11
Time
1
Scan period (sec) = -------------------------------------Z scan rate (Hz)
+ 220
Figure 11.1. Z-axis voltage during force plot scanning.
11-1
Force Imaging
As a result of the applied voltage, the cantilever tip moves up and down relative to
the sample as shown in Figure 11.4. The Z scan start parameter sets the offset of
the piezo travel (i.e., its starting point), while the Z scan size parameter defines the
total travel distance of the piezo. Therefore, the maximum travel distance is
obtained by setting the Z scan start to +220 volts, with the Z scan size set to 440
volts.
As the piezo moves the tip up and down, the cantilever-deflection signal from the
photodiode is monitored. The force curve, a plot of the cantilever deflection signal
as a function of voltage applied to the piezo tube, shows on the display monitor.
The control monitor displays various control panels used to control the microscope
in Force Plot.
In addition to dedicated force measurement experiments, Force Plot may be used
to enhance routine topographic imaging. Force Plot is frequently used to adjust,
calculate, and minimize contact forces between the cantilever and sample.
11-2
Force Imaging
Force Plot may also be used to diagnose SPM performance and determine sensitivity of the cantilever deflection voltage versus voltage applied to the piezo.
1
7
6
11-3
Force Imaging
Upward rebound
Figure 11.3. As the tip approaches the surface, it is frequently pulled down
by attractive forces (left). As the tip lifts off, it sticks to the sample
until pulled away, resulting in a sharp rebound (right).
The display monitor presents a linear scale of the Z piezo voltage. The scale ranges
from +220 volts at the extended end to -220 volts at the retracted end. The range of
piezo travel, as defined by the Z scan start and Z scan size parameters in the Force
Plot control panel, is represented by two white lines on the scale. The lower white
line corresponds to the Z scan start parameter in the control panel, while the spacing between the lines corresponds to the Z scan size. The average voltage applied
to the Z electrode on the piezo tube just prior to entering the Force Plot mode is
represented by the Z center line.
11-4
Force Imaging
Notes
In NanoScope software version 4.2 and later, the tip is retracted by the Scan size
above the surface when first entering any of the force modes (Calibrate, Step,
Indent Volume). Ramping of the Z-axis piezo begins only when the user initiates
operation using one of the Probe commands (Run Continuous, Run Single,
Approach Continuous and Approach Single). The mode is indicated
simultaneously on the gray status bar at the bottom of the control monitor.
11-5
Force Imaging
Retracting
Extending
Tip
Deflection
.25V/div
0.48
V/div
Setpoint
Notes
The Step Motor function is generally used only when the scanning range of the
Z piezo is exceeded and/or when it is necessary to position a force measurement
in the center of the scanners range. Because the Z-axis leadscrew has some
backlash, it may be necessary to rotate the screw several turns by clicking on the
Tip Up or Tip Down buttons before movement is obtained.
In an analogous manner, the photodiode positioner can be used as a coarse adjustment for Setpoint voltage. Changing the beams position on the photodiode by
rotating the mirror adjustment knob shifts the force curve on the graph. Moving the
photodiode down by rotating the positioner counterclockwise shifts the curve
down, just as decreasing the Setpoint parameter shifts the curve down. Conversely,
rotating the positioner clockwise moves the curve up by moving the photodiode up.
11-6
Force Imaging
Cantilever deflection
Down
Up
Lets begin with the simplest of SPM force plots: a contact AFM force plot using a
silicon nitride tip. Because of the pliant property (and lower spring constant) of silicon nitride probes, they are sensitive to attractive and repulsive forces. A force plot
in contact AFM is shown below. (See also Figure 11.2.)
3
Piezo extension
Piezo retraction
4
2
1
7
1
Piezo extends; tip descends. No contact with surface yet.
2
Tip is pulled down by attractive forces near surface.
3
As tip presses into the surface, cantilever bends upward.
4
Piezo retracts. Cantilever relaxes downward until tip forces
are in equilibrium with surface forces.
Piezo continues retraction. Cantilever bends downward as
surface attraction holds onto the tip.
6
7
11-7
Force Imaging
Here, the horizontal axis plots the probes movement relative to the sample. As the
probe descends toward the sample, the probe-sample distance decreases. A descent
is achieved by extending the Z-axis piezo crystal, which is plotted from right-to-left
in yellow on the NanoScope display monitor. As the probe ascends away from the
sample, the probe-sample distance increases. An ascent is achieved by retracting
the Z-axis piezo crystal, which is plotted from left-to-right in white on the NanoScope display monitor.
The cantilevers deflection is plotted on the vertical axis of the graph: when the cantilever is deflected downward, it is plotted on the graphs downward vertical; when
the cantilever is deflected upward, it is plotted on the graphs upward vertical.
The graph reveals at least two very important things:
Notes
Although attractive forces appear small, remember that the tip is extremely
sharp. Since only a few nanometers of the tip actually touch the sample, even
minute forces are distributed over an exceedingly small area, which add up
quickly. Many materials are easily dented by the tip under such conditions. Use
of force plots may be used to adjust a setpoint which applies minimal force to
the sample during contact AFM. (More on this topic below.)
11-8
Force Imaging
slope of the contacted portion of the force curve gives detailed information
about surface elasticity. It is sometimes possible to obtain quantitative measurements of sample elasticity. (See, for example: Radmacher, et al. 1994. Science,
Vol. 265:1577-1579.)
Two imaging techniques may be employed to measure and display elasticity at
multiple points on a sample surface: force modulation (see Section 11.5 below)
and force volume imaging (see Section 11.7 below).
The second click on the mouse causes the system to calculate the slope of the
line segment and enter the calculated value as the Sensitivity in the menu.
A click of the right mouse button will remove the line segment from the screen.
Note that Sensitivity can be expressed in terms of the photodiode voltage versus the
distance traveled by the piezo, or the photodiode voltage versus the voltage applied
11-9
Force Imaging
to the piezo, depending on the setting of the Units parameter. If Sensitivity is calibrated on a material much stiffer than the cantilever, it measures the value of the
AFMs optical lever sensitivity; i.e., how many volts of deflection signal are produced by a given deflection of the cantilever tip. The sensitivity will change for different cantilever lengths and styles (shorter cantilevers give higher sensitivities).
Sensitivity will also change with the position of the laser on the cantilever and the
quality of the laser beam reflection from the cantilever.
Notes
11-10
Force Imaging
differential signal on the upper MultiMode meter display will recapture the cantilever. (Slowly adjust to a more positive value until the tip is back on the surface.)
Adjusting the Setpoint a few tenths of a volt above the point where the cantilever
pulled off will provide a low contact force.
If a high initial contact force will adversely affect the sample, engage the cantilever
with a very small scan size. Then, minimize the force while the tip is confined to a
small area of the sample where it will experience the relatively high initial engagement force. Once the force is minimized, increase the Scan size or offset the scan to
a different area of the sample. However, keep in mind that if the force is minimized
in a smooth area of the sample, the cantilever may pull off when it is translated to a
rougher part of the sample.
Setpoint
VCSmin
Z Position (10.0 V / div)
Figure 11.7. Computing contact force.
11-11
Force Imaging
0.6N
stant of the cantilever is k = -----------m
12nm
and that Z piezo sensitivity = -------------- . The plot
V
above may be measured at the points where the retract portion of the curve intersects the setpoint to tip pull-off (rebound). The distance is then multiplied times the
Z piezo sensitivity to obtain Z . In this example:
Z = 7.6 div 10.0 V div 12 nm v = 152 nm
When the Data type is set to Height with the feedback gains set high, the tip tracks
the sample surface with nearly constant deflection of the cantilever. When the cantilever deflection is constant, the force is constant and the force calculation determines the force between the tip and the surface over the entire scan area.
Force calculations are not as straightforward on images captured with the Data
type set to Deflection. When collecting deflection data, the feedback gains are ideally set low so the sample stays at a constant position relative to the cantilever
holder. In this case, the cantilever deflection (and therefore the force applied to the
sample) varies as features on the surface are encountered. The nominal force
applied to the surface can be calculated from the force equation. The force applied
at other points on the sample can be calculated relative to the nominal force by
using deflection data and the spring-constant of the cantilever. Note that the Sensitivity parameter must be accuratethat is, previously determined and entered
before the deflection data will be accurate.
An simple alternative to calculating force is to follow these five steps:
1. Set sensitivity using the mouse.
2. Change units to metric.
3. Count vertical units from Setpoint to UCS min.
4. Multiply by Tip Deflection Scale.
5. Multiply by the spring-constant, k.
11-12
Force Imaging
Large adhesion
Small adhesion
Hard sample
Soft sample
Long-range repulsion
Long-range attraction
11-13
Force Imaging
Force Mode allows the imaging of forces between the tip and surface, including
chemical bonds, electrostatic forces, surface tension, magnetic forces, etc. In TappingMode, forces may be observed by measuring changes in the tips RMS amplitude, phase, or deflection. The user may represent force plots in one of two forms:
Force Plot and Force Volume (see Section 11.5). Both are essentially the same,
with Force Volume generating a map of many individual force plots. To produce
high-quality force plots, it is necessary to very precisely control the tips position
relative to the surface.
distance fixed
by adjustment
screws
11-14
Force Imaging
Piezo extension
Piezo retraction
1
Tip is clear of the surface
4
3
11-15
Force Imaging
to the surface. Over the same interval, the cantilevers amplitude is diminished
about 1.75 volts due to dampening effects.
Dividing the change in amplitude by the change in Z piezo position gives the
responsiveness of the tip-sample interaction, displayed as a Sensitivity value in the
Z Scan Controls menu. This value may be found directly by using the mouse to
draw a line parallel to the plots slope in the region between points 1 and 2 where
the tips amplitude is dampened. Tip dampening occurs as a result of mechanicalacoustic coupling between the tip and sample. As the tip descends closer and closer
to the sample, oscillation eventually ceases and the amplitude drops to zero.
For TappingMode in this sloping region, each nanometer decrease in the cantilever
position decreases the vibrational amplitude of the cantilever by two nanometers.
Channel 2 (bottom) in Figure 11.12 plots the average cantilevers deflection versus
piezo extension. The deflection signal is low-pass filtered to eliminate the high-frequency TappingMode oscillation. Even as the tips RMS amplitude is dampened
during its encounters with the sample surface, the average deflection is unchanged.
This condition changes, however, once the tip has been positioned so close to the
sample that all oscillation ceases. Pressing the tip still further causes the average
deflection to increase, applying a constant force to the sample.
At point 3 in Figure 11.10, the cantilever has begun to deflect. The region between
points 3 and 4 may be hazardous to the tip, since the tip is pressed tightly against
the sample surface. Most single crystal silicon TappingMode tips fail in this region,
depending upon the hardness of the sample.
Force Imaging
tude). If the amplitude is reduced to zero, the tip has almost certainly been blunted
or broken, and/or the sample has sustained damage.
To make a TappingMode force plot of a silicon calibration reference, try the following:
1. Verify the tipholder is fitted with a TappingMode tip. Set the AFM mode parameter to Tapping and obtain a TappingMode image. You are now in Image mode.
2. To switch to Force Mode, click on the Real Time / View / Force Mode / Plot
option. At least three panels should be visible: Z Scan Controls, Feedback Controls, plus one channel panel (e.g., Channel 1). Collectively, these panels allow the
user complete control over tip-sample interactions. If any panels are not visible,
pull down the Panels menu to select them.
The top menu bar offers a number of tip approach options detailed in the Command
Reference Manual, Chapter 10. These buttons are not generally used for TappingMode and may be ignored.
3. Set the Z Scan Controls, Feedback Controls and Channel panel parameters to
the settings shown below. (Note that the Sensitivity value shown here may differ
from yours.)
Z Scan Controls
Graph Controls
Z scan start:
Other Controls
Trigger mode:
*
*
Z scan size:
Z scan rate:
Trig threshold:
4.88 Hz
Trigger channel:
X offset:
0.00 nm
Z step size:
Y offset:
0.00 nm
Step threshold:
Number of samples:
256
Amplitude
10.0 nm
10.0 nm
Start mode:
Calibrate
Extended
Average count:
End mode:
Sensitivity:
Indent Feedback:
Units:
Off
-25.0 nm
Disabled
Metric
Channel 1
Data type:
Z range:
Amplitude
500 nm
Feedback Controls
Main Controls
Setpoint:
Input attenuation:
Interleave Controls
*
1x
Setpoint 0
The Channel 1 panel should have its Data type parameter set to Amplitude.
11-17
Force Imaging
4. To obtain a satisfactory force plot, it will be necessary to adjust the Z scan start
parameter more positive. This may be easily done using the right arrow key.
Tip
Deflection
0.5 V/div
Setpoint
11-18
Force Imaging
tional amplitude of the cantilever is small relative to the total Z travel of the longrange scanner.
Plotting phase-versus-frequency in TappingMode force plots
Interest has increased in plotting the tips phase-versus-frequency while encountering sample surfaces in TappingMode. This is available only for SPMs having an
Extender Electronics Module. Just as imaging magnetic-electric forces in phase
gives better resolution than amplitude, plotting force in terms of phase-versus-frequency allows improved observation of long-range attractive and repulsive forces
between the tip and sample. Precise interpretation of force plots using phase-versus-frequency remains under investigation. Contact Digital Instruments for more
information.
Regarding the selection of tips...
Almost any TappingMode tip may be used to obtain TappingMode force plots;
however, the ultimate choice should depend upon the delicacy of the sample and/or
the magnitude of the forces to be gauged. Recall that longer tips have lower spring
constants (i.e., they are more pliant) and therefore offer greater sensitivity for most
samples. Shorter tips, on the other hand, may afford better control when gauging
strong attractive forces and are less prone to entrapment by surface tension forces.
The user must ultimately experiment to determine the best tip.
11-19
Force Imaging
11.3.4. Triggering
The Other Controls panel within the Z Scan Controls allows the use of various
triggers when obtaining Force Plot and Force Volume plots. The idea of a trigger
is simple: it limits the total amount of force exerted by the tip upon the sample.
Depending upon which trigger is used and how it is set, it is possible to operate the
trigger independent of drift (Relative) or at some arbitrarily fixed point (Absolute).
Total force
11-20
Force Imaging
146 nm
Other Controls
Trigger mode:
Z scan size:
500 nm
Trig threshold:
Z scan rate:
4.88 Hz
Trigger channel:
X offset:
0.00 V
Z step size:
Y offset:
0.00 V
Step threshold:
512
Number of samples:
Average count:
Sensitivity:
1
0.0379 V/nm
Units:
Off
0.00 V
Amplitude
2.40 V
7.53 V
Start mode:
Calibrate
End mode:
Extended
Indent Feedback:
Disabled
Metric
Motor Control
Step size:
Quit
Tip Up
1.00 m
Tip Down
11-21
Force Imaging
Force Imaging
the slope of the RMS amplitude versus the Z voltage when the tip interacts with the
sample as shown in Figure 11.10. The NanoScope system automatically calculates
and enters the value from the graph after the operator uses the mouse to fit a line to
the graph. This parameter must be accurately calibrated before amplitude data
obtained in LiftMode will be accurate.
Units This item selects the units, either Metric lengths or Volts, used to define
the parameters and the horizontal axis of the graph. Changing this item in Force
Plot also changes it in the Real Time imaging mode. Metric units are obtained from
the Z sensitivity for the particular scanner. under microscope/calibrate/scanner.
Feedback Controls panel
Setpoint (as applied to contact AFM) The cantilever deflection voltage maintained by the feedback loop in Real Time mode can be adjusted by changing this
parameter. In Force Plot, this parameter defines the centerline of the vertical, Cantilever Deflection Voltage axis of the force plot curve shown on the display monitor. Changing the Setpoint shifts the force plot curve on the graph, because the
centerline value is always equal to the setpoint. For example, if the Setpoint is set
to -3.0 volts, the cantilever deflection axis of the graph will be centered around -3.0
volts. Raising the Setpoint to -2.0 volts will shift the force plot curve down by one
volt so the graph will be centered at -2.0 volts. Changing the value of this parameter
in the Force Plot mode also changes the Real Time setpoint in the Real Time control
panel.
Setpoint (as applied to TappingMode) The RMS value of the cantilever deflection voltage maintained by the feedback loop in the Real Time mode can be
adjusted by changing Setpoint. In the TappingMode force plot mode, Setpoint
defines the centerline of the vertical, Tip Amplitude axis of the amplitude calibration plot shown on the display monitor. Changing the Setpoint shifts the amplitude
curve on the graph, because the centerline value is always equal to the Setpoint.
For example, if the Setpoint is at 3.0 volts, the RMS amplitude axis of the graph
will be centered around 3.0 volts; raising the setpoint to 4.0 volts will shift the
amplitude curve down by one volt, so the graph will be centered at 4.0 volts.
Changing the value of this parameter in the Amplitude Calibration mode also
changes the Setpoint parameter in the Real Time control panel.
Input attenuationChanges the gain applied to the input signal. The input signal
can be reduced by a factor of eight to see more of the Force Plot plot's signal amplitude. This parameter is also changed in the Real Time menu and affects the resolution of the system; therefore, it should be changed to 1X before returning to
imaging.
11-23
Force Imaging
11-24
Force Imaging
11-25
Force Imaging
Soft Material
Large Indentation
Hard Material
Small Indentation
11-26
Force Imaging
Bimorph
11-27
Force Imaging
Modulation Tips
Model No.
Cantilever Length
Spring Constant
NP, DNP
100200 m
0.010.6 N/m
Oxide-sharpened
Silicon Nitride
NP-S,
DNP-S
100200 m
0.010.6 N/m
Contact AFM
Etched Silicon
ESP
450 m
0.020.1 N/m
Force Modulation
Etched Silicon
FESP
225 m
15 N/m
TappingMode
Etched Silicon
LTESP
225 m
2070 N/m
TappingMode
Etched Silicon
TESP
125 m
20100 N/m
Cantilever
HARDER
TABLE 7. Force
SOFTER
11-28
Force Imaging
11-29
Force Imaging
Notes:
0.000000 KHz
End frequency:
Auto Tune
500.000 KHz
Graph Controls
Main Controls
40.0000 KHz
Sweep width:
Sweep sample count:
Sweep graph range:
256
100 nm
Metric
Units:
Quit
Motor
Drive frequency:
Setpoint:
Drive amplitude:
Input attenuation:
20.0000 KHz
-0.200 V
867 mV
1x
Integral gain:
10.00
Proportional gain:
10.00
Interleave Controls
11-30
Force Imaging
adjusted later). A series of peaks should be observed on the Frequency Sweep plot.
A typical Frequency Sweep plot is shown in Figure 11.17.
11-31
Force Imaging
Tip Up
1.00 m
Tip Down
Set the Step size parameter at 1.00 m, then click on the Tip Down button repeatedly until the tip lightly contacts the sample surface. When the tip makes contact
with the surface, a rapid change will occur in the Frequency Sweep plot: multiple
peaks will give way to one or two larger peaks (e.g., at 8.00 and 20.00 KHz). An
example is shown in Figure 11.18.
Bimorph Peaks
11-32
Force Imaging
An alternative method is to use the Offset command on the display monitor to center the peak in the graph. Click on the Offset command with the left mouse button;
a vertical green line will appear on the plot. Move the line until it is centered on the
desired peak (usually the frequency with the highest amplitude), then click the left
button again to lock it. Click on Execute. The computer will automatically change
the Drive frequency so that the peak is centered. It may be necessary to click on
this command more than once to get the peak properly centered.
Once the desired frequency is chosen, the Frequency Sweep plot should be recentered similarly to the plot shown below.
Notes
If the bimorph resonant frequency was determined by placing the tip in contact
with the surface, do not use that tip for force modulation imaging. Replace the
cantilever with a new tip and double-check the resonant frequency without
contacting the tip to the surface. From this point forward, cantilever tune should
11-33
Force Imaging
11-34
Force Imaging
ATTENTION! Lutilisation dun levier plus rigide que ceux utiliss habituellement peut entraner un endommagement de lchantillon par la pointe. Afin de
rduire les risques dendommagement, ajuster la valeur cible du rtro-contrle (Setpoint) sa valeur la plus faible possible.
WARNHINWEIS! Falls Sie mit Cantilevern hherer Federkonstanten arbeiten, als
normal blich, kann die Probe durch die Spitze beschdigt werden.. Um eine evtl.
Beschdigung zu minimieren bzw. zu vermeiden, benutzen Sie bitte einen Arbeitspunkt (Setpoint), der nur geringfgig ber der freien Deflektion des Cantilevers
liegt. Die Differenz zwischen Setpoint im Feedback-Men und der freien Deflektion des Cantilevers (Photodetektorsignal A-B) sollte minimal sein.
Reduce the Setpoint using the cursor keys until the cantilever pulls off the surface.
Note the value of the Setpoint when the cantilever pulled off the surface (pull-off
value). Increase the Setpoint until the cantilever begins to touch the surface again
and an image appears. Finally, readjust the setpoint so that it is slightly more positive than the pull-off value.
Be aware that there is a trade-off here! If the Setpoint is adjusted very close to the
pull-off value, imaging will be performed with the smallest and least damaging
force. However, imaging may be more unstable, as the cantilever may pull off the
surface unexpectedly.
Also note that the setpoint value can affect the force modulation contrast as discussed below. It may be desirable to further adjust the setpoint after completing
step 11.4.19.
13. Examine the contrast of the force modulation image.
The amplitude signal shows how much the cantilever can indent into the surface.
Softer features will allow the cantilever to indent further into the sample, giving a
small deflection amplitude, while hard features will allow little indentation and create large deflection amplitudes. The force modulation software is configured so that
softer areas will appear lighter (higher) in the amplitude image. The relative elasticity of the sample is seen as contrast between dark and light areas in the amplitude
image. Therefore, the amplitude image is really the inverse. This allows soft samples mounted on firm substrates to show up as positive shapes relative to the background.
14. Optimize Drive Amplitude for Force Modulation imaging.
Next, optimize the force modulation signal, which is the amplitude data. The
amount of contrast and the quality of both Height and Amplitude images will
depend on the Drive amplitude. In general, increasing the Drive amplitude will
give greater contrast in the force modulation image.
11-35
Force Imaging
It is possible, however, to set the Drive amplitude too high. In this case, as the
Drive amplitude is increased, contrast will remain unchanged in the Amplitude
image. Instead, the overall contrast in the force modulation image will remain
roughly constant, but more artifacts will be observed in the image.
For example, if the drive amplitude is too high, the force modulation image will
become contaminated by edge effects or friction effects.
The best procedure is to operate at the lowest drive amplitude that gives sufficient
contrast to examine the sample. Low drive amplitudes will also help extend the life
of the cantilever tip and reduce sample damage.
Here is a suggested procedure for optimizing the drive amplitude:
Reduce the drive amplitude to a very small value (say 100 mV).
Increase the Drive amplitude with the arrow keys or by typing in new values.
The contrast of the Amplitude (force modulation) image should increase.
Continue increasing the Drive amplitude until sufficient contrast appears in the
Amplitude image.
If the Drive amplitude is increased to the point where contrast is no longer
improving, reduce the Drive amplitude slightly from this value.
As mentioned above, the force modulation contrast will also depend on the Setpoint. In general, if using a larger Setpoint (larger tracking force) it will be necessary to use a smaller Drive Amplitude to obtain good force modulation images
without artifacts.
If it is difficult to obtain a clear force modulation image that is free of artifacts, try
reducing (or sometimes increasing) the Setpoint and then optimizing the Drive
Amplitude again.
15. Readjust gains (if necessary).
The value of the Drive amplitude may also affect the contact mode image, making
the system more or less likely to go into unwanted oscillations. If the Drive amplitude is changed by a large amount it may also be necessary to readjust the Integral
gain and Proportional gain. The gains should be set as high as possible to track the
sample topography, but not so high that they show the oscillation due to the
bimorph oscillation. If the gains are set too high, parallel diagonal lines will be
observed on the Height image. If an oscillation is seen, the feedback system is trying to cancel the cantilever oscillation by moving the Z-axis piezo in the opposite
direction. If there is evidence of an oscillation in the Height data, turn the gains
down until the oscillation stops.
11-36
Force Imaging
Notes
Oscillations not visible in height Image Mode data may be seen more easily
using Scope Mode.
Sometimes an oscillation that appears in the data will be due to aliasing as
described in the next section. If gains cannot be adjusted to eliminate unwanted
oscillations without compromising the height images quality, see the next section
below.
16. Changing the Drive Frequency to eliminate aliasing.
Under some conditions, unwanted oscillations will appear in the data due to aliasing of the Drive Frequency with the image pixel rate. This problem can be eliminated by changing the Drive frequency by small increments. Use the arrow keys to
change the Drive frequency up or down very slightly until the oscillation disappears. On some materials, shifting the Drive frequency slightly (1-3 Hz) below or
above a resonant peak value may improve image contrast. Operators are encouraged to experiment.
Notes
It is possible to see artifacts in force modulation images that are not due to differences in elasticity. Some artifacts to watch for are outlined below:
Edge effects
Sometimes force modulation images will show changes in amplitude at the edge of
topographic features like steps or bumps. These artifacts may look like the derivative of the sample topography.
To see if a feature is an edge effect, try reversing the Scan direction from Trace to
Retrace, for example. If the contrast reverses or the amplitude change now appears
on the other side of the topographic feature, then the amplitude change is likely due
to the topographic edge, not differences in elasticity. To minimize edge effects,
reduce Drive amplitude, Setpoint or Scan speed. Also set the Integral gain and
Proportional gain as high as possible without causing unwanted oscillations.
Frictional effects
Because the cantilever is held at an angle to the sample surface, the cantilever tip
will slide laterally (skate) as the tip is pushed into the sample.
11-37
Force Imaging
1516
tip moves
down and left
11-38
Force Imaging
Notes
Force modulation contrast is very sensitive to the spring constant of the tip,
which varies according to the length and thickness of the cantilever.
TappingMode may be easily performed with either force modulation or
TappingMode cantilevers, but may prove difficult with contact AFM silicon and
silicon nitride cantilevers. (See table in Section 11.5.2 for guidance.)
11-39
Force Imaging
11-40
Force Imaging
3. Using the left arrow key, optimize the negative LiftMode image by slowly
decreasing the Lift scan height parameter in the Interleave Controls panel from
zero until the surface is reached. In most cases, it should not be necessary to go
below -60.0 nm. (This is sometimes best performed while in Real Time / View /
Scope Mode.)
Notes
Use of a negative Lift scan height value here gives rise to the term negative
LiftMode. When the interleaved scan is enabled, the tips TappingMode height
above the sample will be reduced by the Lift scan height amount. This places
the oscillating tip in contact with the sample as surface features are profiled
(from TappingMode data). The contrast between light and dark reveals areas of
high and low elasticity, with the dark area indicating harder material and the
lighter areas indicating softer material.
4. Adjust the interleaved Drive amplitude and Lift scan height until the force
modulation image is optimized. This may require some experimentation.
Notes
If you are seeing a lot of contrast in the amplitude image before reaching the
surface, try reducing the Integral and Proportional gains. It may be necessary
to adjust gains slightly lower than when performing normal TappingMode
imagery. A good way to check this is to enter a Lift scan height of 100 nm, then
adjust gains (and possibly Drive amplitude) until you see the minimum amount
of contrast in the amplitude image. (This may be easily monitored in Scope
Mode.) Once contrast is minimized, enter a Lift scan height of 0.0 nm and
begin approaching the surface.
11-41
Force Imaging
Signal Type
Contact AFM
Amplitude
Deflection
Frictiona
Phaseb
Frequencyb
Potentialb
4
4
TappingMode
4
4
4
4
4
a. The usefulness of force imaging with friction has not yet been determined.
b. Phase, frequency and potential measurements are available only on SPMs equipped
with the Extender Electronics Module.
The type of force image captured from a surface will depend upon how the SPM is
set up. For example:
If an MFM image is being captured, force volume imaging (phase) allows the
detection of long-range magnetic forces otherwise difficult to detect.
11-42
Force Imaging
For ordinary contact AFM, the use of force volume imaging (deflection) allows
the user to see otherwise invisible electrostatic forces.
Notes
The tip is lowered and pressed against the sample surface until the cantilever is
deflected to the Trig[ger] threshold value. If the surface is not contacted within
one Z scan size distance, the tip is extended one additional Z scan size (for a
total distance of two times the Z scan size), then retracted one Z scan size. This
means that the tip is incrementally lowered (or racheted) one Z scan size for
each extension-retraction cycle until the surface is contacted, or the Z piezo
reaches its total maximum of 220 volts.
Once the tip contacts the surface and deflection attains the Trig[ger] threshold
value, sample height is recorded for that Z-axis threshold. A force curve is
recorded for the X-Y coordinate after the piezo retracts. The tip is lifted clear of
the surface and translated in X-Y to the next coordinate. The entire process is
then repeated.
The general sequence for obtaining a force volume image consists of the following:
1. Obtain a height, deflection, amplitude or phase image of the sample in Image
Mode. On the Channel panel, select the same Data type intended for force plotting and for triggering.
2. Obtain a Force Plot of the sample surface (any portion of the surface will suffice). Using the force plot, determine the samples general force characteristics and
set the following parameters: Setpoint, Z range, Z scan size, Z scan start, and
Trigger mode.
11-43
Force Imaging
Notes
11-44
Force Imaging
the Sensitivity parameter, use the mouse to draw a line on the force plots parallel to
their slope. Click the left button to drop each end of the line; click on the right button to enter the Sensitivity value and erase the line.
The Start mode parameter should be set to Calibrate. The Trigger mode may be
set to Relative.
Click here
Se
ity
iv
it
ns
pe
slo
Click here
11-45
Force Imaging
Image Scan Controls Real Time parameters for the height image shown in
the upper-left corner of the display monitor, such as Scan size, X- and Y offset,
Scan angle, and Scan rate.
11-46
Force Imaging
Image Channel More Real Time parameters affecting the height image.
Z Scan Controls Parameters affecting the tips vertical movements,
including Z scan start, -size and -rate.
Notes
The total amount of data for any given force volume image file is limited to a
maximum 1 Mb by the relation
2
The following parameters are carried over from Force Plot settings: Z scan start,
Z scan size, and Z scan rate. Although these parameters may be altered during
force volume imaging, they should obtain their initial settings from the Force Plot
window.
Enter the Number of samples parameter to set the resolution (or number of
shells) of the force volume image and number of points plotted on the force curve
at the bottom of the display screen. The value of this parameter will not affect the
speed of the scan, but will affect the file size. A good default setting is 32.
Next, enter the Force per line value. This parameter determines the number of
force curves taken along each line of the sample and significantly affects capturing
speed. Enter a number which allows sufficient resolving of force-defined features,
but minimizes capturing time. A good default value is 64.
Finally, enter the Samples per line value. This parameter sets the number of samples to be displayed in the height image only and does not affect the force volume
image in any way. Increasing this parameter increases the capture time. A setting of
128 provides sufficient detail to resolve many small features while helping to
reduce capture time.
11-47
Force Imaging
The Display mode parameter may be set to show the Extend portion of the plot,
the Retract portion, or Both.
For most imaging, the Sample period parameter should be held at 30 s. (Sample
periods less than 30 s may be too fast and cause errors.) This parameter determines the sampling rate of data at the SPMs digital signal processor, which some
experimenters may wish to increase. This parameter is coupled with two other
parameters: the Z scan rate and FV scan rate; that is, changes in the Sample
period will change both Z scan rate and FV scan rate.
Keeping the Sample period minimized at 30 s will yield optimal data averaging
between Z-axis sampling points. As this parameter is increased, data averaging is
lessened and data becomes discrete.
6. Adjust Feedback Controls parameters.
Feedback Controls panel parameters carry over from the Force Plot settings; however, changes may be required to protect the tip and optimize the force volume
image.
For silicon nitride tips, the Setpoint may be adjusted to within plus or minus several volts of the microscopes vertical deflection signal value. For crystal silicon
TappingMode tips, the Setpoint should normally be increased no more than 1-3
volts below the RMS amplitude voltage; this will help protect the tip.
In Force Volume imaging, triggers are used to set the turnaround point of the Zaxis piezo. The Trigger mode determines the type of trigger to be employed. Two
types are offered: Relative and Absolute, or the trigger may be turned Off.
Notes
11-48
Force Imaging
Notes
Notes
This example assumes the Sensitivity parameter has been properly set (see step
#2 above) and that the detectors range has not been exceeded.
7. Adjust Force Channel parameters.
The Force Channel panel features parameters for the force plot at the bottom of the
display monitor. In most ways it is exactly like a Force Plot graph, except that it
superimposes up to eight multiple force plots (depending on Samples per line)
simultaneously for each scan line. When the next line is scanned, previously shown
plots are cleared.
The available settings for the Data type parameter will depend upon the type of
imaging being done. For example, Phase is available only on microscopes in TappingMode which have an Extender Electronics Module. Also available are
Amplitude, Deflection, Phase, Potential, and Thermal. Set the Data type accordingly.
Begin with a large Z range to locate the force plots. (At full size, the force plots
will probably resemble a thin line.) Slowly decrease the Z range value until the
force plots fill most of the graph.
11-49
Force Imaging
The Offset parameter determines where force plots are to be graphed relative to the
current Setpoint. When the Offset is Off, the graphs center horizontal line is positioned at the tips Setpoint value. When Offset is Enabled, the center horizontal
line is positioned at the tips free-air voltage (i.e., voltage when the tip is clear of the
surface).
8. Adjust FV [Force Volume] Channel parameters.
Parameters on the FV (Force Volume) Channel panel control the type and range of
forces viewed in the force volume image on the display monitor. In addition, the
Volume scale parameter also affects the viewable range of data captured during
force volume imaging.
The Z direction parameter determines which portion of the Real Time force curve
cycle (Extend or Retract) is shown on the display monitors force volume image.
For example, if the force of interest is material elasticity, the Extend portion of the
curve is usually selected. If the force of interest is electrostatic attraction, the
Retract portion of the curve is generally used.
The Volume scale parameter sets the range of values represented by the force volume image. Because force volume data is displayed line-by-line at periodic intervals, optimizing this parameter may require several adjustments. The results of
changing this parameter are not seen until the next line of force volume data is displayed.
The Z display and Volume offset parameters work in unison to define which portion of the superimposed force curves is to be plotted in the display monitors force
volume image. These parameters are also simultaneously represented as a green
cursor on the force plot graph. Both Z display and Volume offset may be changed
by positioning the cursor cross with the mouse, or vice versa. (The cursor is repositioned automatically whenever these parameters are changed.) Each parameters
relationship to the cursor is shown in Figure 11.23.
11-50
Z display
Volume offset
Force Imaging
Setpoint
Free-air
deflection
11-51
Force Imaging
11-52
Chapter 12
12-1
This chapter provides a general discussion of the Interleave action and commands,
with emphasis on LiftMode. Chapter 13 provides step-by-step instructions for
using Interleave scanning and LiftMode for obtaining magnetic and electric force
data. For purposes of learning to use Interleave scanning, this section may be useful
even to those users whose end application is not magnetic force microscopy.
INTERLEAVE
ENABLED
four Main and four Interleave
scan lines shown
slow scan
direction
INTERLEAVE
trace and retrace
fast scan
direction
MAIN trace
and retrace
12-2
With Interleave mode enabled, the system first performs a standard trace and
retrace with the Main Feedback Controls in effect. The tip moves at half the normal
rate in the slow scan direction. As shown on the right of figure 12.1, an additional
trace and retrace are then performed with the Interleave Feedback Controls enacted.
The frame rate halves because twice as many scan lines are performed for the same
scan rate.
Two modes are possible for Interleave scan: Normal and Lift. With Normal
selected, the feedback remains on during the Interleave pass with the values under
Interleave Feedback Controls (Setpoint, Gains, etc.) in effect. In Lift mode, the
feedback is instead turned off, and the tip lifted off the surface and scanned at a
user-selected height; see below.
Main Trace
(Height Data)
Lift
Start
Height
Lift
Scan
Height
Lift
Scan
Height
12-3
12-4
5) Display the interleave data by switching Scan line (in the Channel panels) to
Interleave.
Notes
Tip Shape
As shown in figure 12.2, the tip separation in the LiftMode is defined in terms of
the Z direction only. The Lift scan height is added to the height values taken
from previous scan lines point-by-point. However, the tip may be closer to the
sample than the Z separation indicates. On features with steep edges, the tip
may get very close to the sample even though the Z separation is constant; see
figure 12.2.
12-5
LiftMode. If a small setpoint value forces the oscillation amplitude low while the
feedback is running, the amplitude can grow considerably when the tip is lifted free
of the sample surface. The change can also be large if the main Drive amplitude
was increased or the main Drive frequency altered after the tip was engaged. (The
vibration amplitude remains at the setpoint during the main scan even if these
parameters are changed.) This could make the tip hit the surface in the lift scan for
small Lift scan Heights.
Notes
The cantilever drive circuit features a filter capacitor to limit the rate at which
the drive amplitude is changed between Main and Interleave scanning. If using
scan rates above a few hertz, it may be advantageous to remove or disable the
filter. For more information on how to disable the filter, contact Digital
Instruments technical support.
12-6
12-7
12-8
Chapter 13
13-1
4
3
5
2
1&2
3
4&5
All standard Dimension-series SPMs and MultiModes are capable of MFM imaging using amplitude detection techniques. By adding an Extender Electronics
module (see Figure 13-2), either system may also be used for frequency modulation
or phase detection, giving improved results. Amplitude detection has largely been
superseded by frequency modulation and phase detection. A more extensive discussion of force gradient detection and MFM imaging is given in the reprint Magnetic
Force Microscopy: Recent Advances and Applications. Contact Digital Instruments
to obtain a copy.
Notes
In the instructions below, steps specific to phase and amplitude imaging are
described independently. Use the icons in the margin to locate steps specific to
either frequency modulation and phase detection, or amplitude detection.
.
13-2
13-3
quency modulation detection are superior methods for magnetic force imaging,
offering greater ease of use, better signal-to-noise ratios, and reduced artifact content as compared to amplitude detection. If extensive MFM imaging is planned, the
Extender Electronics module is strongly recommended.
13.2.1. Procedure
1. Mount a NanoProbe magnetic probe on the scanner or tip holder. The tip should
be magnetized with a strong permanent magnet before installing the tip holder on
the AFM head. Tips are usually magnetized with the field aligned along the tip axis
(perpendicular to the sample surface). The MFM then senses force gradients due to
the perpendicular component of the sampless stray field. Tip magnetizers are provided with MFM probes purchased from Digital Instruments.
2. Set up the AFM as usual for TappingMode operation. In all Channel panels, the
Highpass and Lowpass filters should be Off. Set the Rounding parameter in the
Microscope / Calibrate / Scanner window to zero (0.00).
3. The procedure to tune the cantilever drive frequency and amplitude depends on
whether you are using phase detection or amplitude detection. Both cases rely on
automatic Cantilever tune just as when preparing for TappingMode; see Chapter 8
of the product instruction manual. MFM cantilevers have resonant frequencies
between 50 and 100 kHz. If using the AutoTune feature, these values can be used
as bounds for the frequency sweep. With the Extender option, two curves appear in
the Cantilever Tune box: the amplitude curve in white, and the phase curve in yellow (Fig. 229-3). Microscopes without the Extender Electronics Module display
only the amplitude curve.
Phase
Detection
MFM / EFM
Figure 13.3. Cantilever Tune for phase detection and frequency modulation.
13-4
To correctly track the cantilever phase, the Phase offset parameter must be
adjusted. This is automatically done in AutoTune; alternatively, Zero Phase can be
selected from the menu bar above the Cantilever Tune frequency sweep window.
The phase curve should appear as in Figure 13.3, decreasing with increasing frequency, and crossing the center line (corresponding to a 90 phase lag) at the peak
frequency. The phase curve then measures the phase lag between the drive voltage
and the cantilever response. Again, vertical gradients in the magnetic force cause a
shift f0 in the resonance frequency. In this case, resonance shifts give rise to phase
shifts which then give an image of the magnetic force gradients; see figure 13.4.
Notes
The Extender electronics give a measure of the phase lag of the cantilever oscillation relative to the piezo drive. This measurement is monotonic versus frequency as is the true phase lag in degrees. The Extender measurement; however,
has slightly different nonlinear characteristics vs. frequency. The measurement
technique allows optimal signal-to-noise ratios; however, absolute values of
phase data should be taken as approximate. Users requiring quantitative measures of force gradient are advised to use frequency modulation (See
Section 13.2.2). Proceed to Step #4 below.
.
180
Phase (deg)
F0
90
Drive Frequency
MFM / EFM
13-5
maximum sensitivity, set the Drive frequency to the steepest part of the resonance curve. As the tip oscillates above the sample, a gradient in the magnetic
force will shift the resonance frequency f0; (see Figure 13.6). Tracking the variations in oscillation amplitude while in LiftMode yields an image of the magnetic force gradients.
Proceed to Step #4 below.
Amplitude
F0
Drive Frequency
Figure 13.6. Shift in amplitude at fixed drive frequency.
4. Adjust the Drive Amplitude so that the RMS voltage response of the photodetector is approximately 2 volts. (Somewhat larger values may be beneficial if using
amplitude detection.) This can be done with Auto Tune by selecting an appropriate
13-6
Target Amplitude (in this case, 2V) before tuning, or by exiting Cantilever Tune
and manually adjusting the Drive Amplitude parameter under Feedback Controls.
Quit Cantilever Tune and return to Image Mode. Under Interleave Controls set
the Lift start height to 0 nm, and Lift scan height to 100 nm. (The lift height can
be optimized later.) Set the remaining Interleave parameters (Setpoint, Drive
Amplitude, Drive frequency, and gains) to the Main Controls values. This can be
done by setting the flags left of the Interleave Control column to off (grayed
bullets).
5. Under Scan Controls, set the Scan size to 5 m and Scan rate to 12 Hz. For
the left image, set the Z range to 75 nm and the Line direction to Retrace. Engage
the AFM and make the necessary adjustments to obtain a good topographical image
while displaying height data. Use the maximum possible Setpoint to ensure that the
tip is contacting the surface only lightly. The image should be similar to the topographic image shown on the left of Figure 13.7. The surface is fairly flat with lubrication nodules of various sizes. A good image of the nodules provides a check that
the tip is sharp.
Figure 13.7. Topographic (left) and magnetic force gradient image (right) of
metal evaporated tape at 100 nm Lift scan height.
6. The MFM data can be shown on Channel 2; however, the parameter settings are
different depending on whether Phase Detection or Amplitude Detection is being
used
13-7
Phase
Detection
Phase Detection
Set the Channel 2 image Data type to Phase, Z range to 3 degrees, and Line
direction to Retrace.
MFM / EFM
Notes
It is important that the Scan direction be set to Retrace for both the main and
interleave scans. If instead it is set instead to Trace, a band may appear along
the left side of the images due to the time taken for the tip to move between the
surface and the lift scan height.
Amplitude
Detection
MFM / EFM
Amplitude Detection
Set the Channel 2 image Data type to Amplitude, Z range to 1 nm, and Line
direction to Retrace.
7. Change Interleave mode to Enable to invoke LiftMode. Set the Channel 2 Scan
line to Interleave to display the interleaved data. (This can only be done after
Interleave mode is Enabled.) A magnetic force gradient image similar to that
shown on the right of Figure 13.7 should appear as the Channel 2 image. The alternating dark and light stripes represent the recorded magnetic information, and signify a varying resonant frequency and hence magnetic force gradient on the tip. As
usual, keep the Setpoint as large as possible while consistent with a good image.
Wider scans (> 25 m) will reveal separate tracks in which the magnetic stripes are
at different angles.
13-8
Phase
Detection
MFM / EFM
Amplitude
Detection
When using Amplitude Detection, optical interference may sometimes appear in
the Lift (magnetic force gradient) image when imaging highly reflective samples.
Optical interference appears as evenly spaced, sometimes wavy lines with ~12 m
MFM / EFM
spacing superimposed on the lift image. This occurs when ambient laser light (i.e.,
light passing around or through the cantilever, then reflecting off the sample) interferes with laser light reflecting from the cantilever. Interference can be alleviated by
moving the beam spot up the cantilever away from the tip; about one-third of the
cantilever length works well. The adjustment can be refined by carefully moving
the beam spot laterally a small distance on the cantilever while scanning until interference fringes are minimized. Be careful not to move the beam off the cantilever or
feedback may be lost.
Notes
13-9
13-10
Increasing the Drive Amplitude can improve the signal-to-noise ratio when
using phase detection or frequency modulation. This is because intrinsic, lowlevel noise interferes less when measuring the phase of a larger cantilever
oscillation amplitude and hence stronger photodetector output. As an
13-11
illustration, try setting the Interleave Drive Amplitude to 0; the resulting phase
image will be pure noise because measuring the phase of a non-oscillating
cantilever is meaningless.
In LiftMode, the Interleave Drive Amplitude can often be set to a value larger
than in the main scan, thus giving optimal signal-to-noise. In some cases this is
beneficial as long as the Drive Amplitude is not increased to the extent that the
tip strikes the surface on the low point of its swing. The signatures of tip-sample
contact are white and black spots in the image, or, in extreme cases, noisy, highcontrast streaks across the whole image. It is usually safe to increase the Drive
Amplitude until the first signs of tip strike are noticed, then reduce the amplitude slightly.
CAUTION: Before enabling the Interleave Drive Amplitude, check that its value
is not much larger than the main Drive Amplitude value to prevent possible damage to the tip.
ATTENTION: Lors dun travail en mode intercal (Interleave Mode), vrifier que
la valeur de tension applique loscillateur pizo-lectrique est infrieure celle
applique loscillateur en mode imagerie (Main Drive Amplitude). Le non
respect de cette procdure peut entraner la destruction de la pointe.
WARNHINWEIS: Bevor Sie im Interleave-Mode die Interleave Drive-Amplitude einschalten, vergewissern Sie sich bitte, da der dort eingetragene Wert nicht
wesentlich grer ist, als der Wert der Main Drive-Amplitude, um evtl. Beschdigungen der Spitze vorzubeugen.
Amplitude
Detection
MFM / EFM
Notes
TappingMode, the average tip-sample distance hT is equal to the oscillation amplitude, which is determined by the setpoint. Large variations in setpoint can thus
change the total tip-sample distance in Liftmode, sometimes with visible results in
the magnetic image. For this reason, reproducible results are most easily obtained
by using consistent setpoints. Note that a lift scan height of 0 nm still gives a mean
tip-sample distance of hT in LiftMode.
Lift pass
TappingMode pass
htot =
hT + hlift
hlift
hT
13-13
Dimension
MultiMode
For MultiMode SPMs, it should always be set to MultiMode. Use a pencil to
access the switch through the hole.
WARNING: Do not insert a conducting object (e.g., screwdriver) into the Phase
Extender box while it is engergized.
ATTENTION: Ne pas insrer d objet conducteur (par exemple: un tournevis)
dans le botier dextension de phase (Phase Extender Box) quand celui-ci est sous
tension.
WARNUNG: Stecken Sie keine leitfhigen Teile (zum Beispiel Schraubenzieher)
in die Phase Extender Box, whrend diese eingeschaltet ist.
13-14
Notes
Due to the many Digital Instruments microscope systems available, the variety
of scanners, when the system was purchased, and the systems history of
software updates, not all systems will have PAR files with the same name or
naming scheme. You can determine the exact name of the PAR file to be
modified by using the DOS EDIT file editor to list the contents of the file /
SPM/SYSTEM.PAR. The name of the most recent PAR file in use will be
on the line that starts with the words \Microscope file:.
CAUTION: When closing the System.par file after viewing it, if you are
asked to SAVE the file or to SAVE CHANGES, be sure to say No.
ATTENTION: A la fermeture du fichier System.par, lorsque lordinateur
demande SAUVEGARDER (SAVE) ou SAUVEGARDER LES CHANGEMENTS (SAVE CHANGES), rpondre toujours NON.
WARNHINWEIS: Falls Sie sich mit einem Editor die Datei system.par
anschauen und beim Verlassen des Editors gefragt werden, ob Sie die Datei speichern (SAVE) bzw. die nderungen speichern wollen (SAVE CHANGES), antworten Sie bitte mit Nein (No).
To modify the PAR files on a MultiMode or Dimension system, locate the proper
PAR file in the /SPM/EQUIP directory on your computer. Use a text editor to
modify the file. The text must be saved as plain ASCII (standard DOS text). DI recommends using the DOS EDIT file editor. (Microsoft Word or other similar desktop-editors normally save files with embedded formatting commands causing the
par files to be unreadable by the NanoScope software.) Using the EDIT program,
modify the file to include a new line at the bottom of the file. For example:
13-15
If the line
\Is FM: No
is present in the file, then change the line to read:
\Is FM: No
is not present, add the line:
13-16
2. Turn off the power to the NanoScope controller whenever connecting or disconnecting the Extender.
3. In LiftMode, the best performance is obtained if the RMS amplitude is kept
below 7.0 volts, the limit of the RMS outputs linear operation.
13-17
13-18
Chapter 14
14.4
14.5
14.6
14.7
14-1
14-2
3
Force Gradient Scope Data
(Interleave scan)
1
1521
Electric Fields
1
2
3
14-3
Figure 14.3. EFM probe tip holder, top and bottom view (left to right).
14-4
Amplitude
Electric field gradient imaging is analogous to standard MFM, except that gradients
being sensed are due to electrostatic forces. In this method, the cantilever is
vibrated by a small piezoelectric element near its resonant frequency. The cantilevers resonant frequency changes in response to any additional force gradient.
Attractive forces make the cantilever effectively softer, reducing the cantilever
resonant frequency. Conversely, repulsive forces make the cantilever effectively
stiffer, increasing the resonant frequency. A comparison of these force additives is
shown in Figure 14.3.
F0
Amplitude
Frequency
Attractive gradient equivalent to additional spring in tension attached
to tip, reducing the cantilever resonance frequency.
F0
Frequency
Repulsive gradient equivalent to additional spring in compression attached
to tip, increasing the cantilever resonance frequency.
Figure 14.4. Comparison of attractive and repulsive forces
to action of a taut spring attached to the tip.
Changes in cantilever resonant frequency can be detected in one of the following
ways:
14-5
material differences or regions at substantially different potentials (voltage differences of 1.0 or more). For other samples having rough surface topography or small
voltage variations, this technique may be undesirable because topographic features
will appear in the LiftMode image.
RMS Detector
Phase Detector
Laser
Beam
Amplitude Signal
Phase Signal
Signals to
NanoScope
Servo Controller
Reference
Signal
(Feedback loop
adjusts oscillation
frequency until
phase lag is zero)
Frequency
Frequency Signal
Control lines
Photodetector
High Resolution
Oscillator
Tapping
Piezo
Oscillator Signal
14-6
Notes
In most cases, it is necessary to apply a voltage across the tip or sample to
achieve a high-quality image. Various methods for applying voltages to the tip
and sample are included in the sections that follow. Samples with permanent
electric fields may not require the application of voltage.
14-7
EFM applications and surface potential operation, jumpers are usually left in their
original positions or returned to their original positions.
Jumpers, inside baseplate
window
To
er
ntroll
pe Co tronics
o
c
S
o
c
Nan
r Ele
tende
or Ex Module
Figure 14.7. Normal jumper configuration for systems without the Extender
Electronics Module as shipped from factory.
14-9
Notes
In all configurations which apply voltage to the tip, an E-field cantilever holder
is required. Contact Digital Instruments for more information.
Enabling the Analog 2 Voltage Line:
The Analog 2 voltage line is normally used by the NanoScope to control the
attenuation (1x or 8x) of the main feedback signal. This application of EFM imaging uses the Analog 2 signal for EFM data. Therefore, input attenuation must be
disabled for the duration of the EFM experiments. To do this, click on the Microscope / Calibrate / Detector option to display the Detectors Parameters window.
Switch the Allow in attenuation field to Disallow. Return to the main Feedback
Controls panel; the Analog 2 field should now be enabled. This signifies that voltage is now being supplied via the Analog 2 pin located on the baseplate header.
Remember to restore (Allow in attenuation) upon completion of EFM imaging.
Ground
Tip
Piezo Cap
Analog 2
Analog 2
Sample
Gain Select
Analog 2
To AFM Tip
Auxiliary D (to NanoScope III controller)
STM
Indicates jumpers
14-10
Tip
Piezo Cap
Analog 2
Sample
Analog 2
Gain Select
Analog 2
To AFM Tip
Auxiliary D (to NanoScope III controller)
STM
Indicates jumpers
14-11
Notes
In all configurations which apply voltage to the tip, an E-field cantilever holder
is required. Contact Digital Instruments for more information.
Ground
Tip
Piezo Cap
Analog 2
Sample
Gain Select
Unused
To AFM Tip
Auxiliary D (to NanoScope III controller)
STM
>10 M
(+)
(-)
External Voltage
Source
Indicates jumpers
14-12
Ground
External Voltage
Source
Piezo Cap
Analog 2
Gain Select
(+)
>10 M
Tip
Unused
To AFM Tip
Sample
Auxiliary D
+
>10 M
STM
Indicates jumpers
14-13
14-14
Notes
In all configurations which apply voltage to the tip, an E-field cantilever holder
is required. Contact Digital Instruments for more information.
Enabling the Analog 2 Voltage Line
The Analog 2 voltage line is normally used by the NanoScope to control the
attenuation (1x or 8x) of the main feedback signal. This application of EFM imaging uses the Analog 2 signal for EFM data. Therefore, input attenuation must be
disabled for the duration of the EFM experiments. To do this, click on the Microscope / Calibrate / Detector option to display the Detectors Parameters window.
Switch the Allow in attenuation field to Disallow. Return to the main Feedback
Controls panel; the Analog 2 field should now be enabled. This signifies that voltage is now being supplied via the Analog 2 pin located on the MultiMode baseplate header.
Remember to restore (Allow in attenuation) upon completion of EFM imaging.
Ground
Tip
Piezo Cap
Analog 2
Analog 2
Sample
Unused
Analog 2
To AFM Tip
Auxiliary D (to NanoScope III controller)
STM
Indicates jumpers
14-15
Tip
Analog 2
Unused
Sample
Analog 2
Analog 2
To AFM Tip
Auxiliary D (to NanoScope III controller)
STM
Indicates jumpers
14-16
Notes
In all configurations which apply voltage to the tip, an E-field cantilever holder
is required. Contact Digital Instruments for more information.
Ground
Tip
Piezo Cap
Analog 2
Sample
Unused
Unused
To AFM Tip
Auxiliary D (to NanoScope III controller)
STM
>10 M
(+)
(-)
External Voltage
Source
Indicates jumpers
14-17
Ground
External Voltage
Source
(+)
Piezo Cap
>10 M
Analog 2
Unused
Tip
Unused
To AFM Tip
Auxiliary D
Sample
+
>10 M
STM
Indicates jumpers
14-18
Notes
Amplitude detection is the only procedure described here that can be done
without the Extender Electronics Module; however, this method is no longer
recommended (see Section 14.4.2).
1. Locate the two toggle switches on the backside of the Extender Electronics box
(Figure 14.16), then verify that they are toggled as shown in Table 14.1.
Mode
Tip or Sample
Voltage
FM/Phase
Gnd/Surface Potential
Surface Potential
To
Mi
cro
s
To
Analog 2
cope
S
Nano
co
pe
Mode
FM/Phase
TappingMode
Contact AFM
MFM
Apply voltage to
tip or sample (Use
for electric field
gradient imaging;
tunneling AFM)
Analog 2
Surface Potential
GND/Surface
Potential
14-19
Notes
MFM / EFM
Phase Detection
Phase Detection is only available when the Extender Electronics Module has been
correctly configured into the system.
In the Cantilever Tune window, set Start frequency and End frequency to
appropriate values for your cantilever (e.g., for 225 m MFM cantilevers, set
Start frequency to 40 kHz and End frequency to 100 kHz). Select Autotune.
14-20
Two curves appear on the Cantilever Tune graph: the amplitude curve in white,
and the phase curve in yellow. (In Figure 14. 17, the phase curve is the dashed
line and the amplitude curve is the solid line.).
Phase (deg)
F0
90
Drive Frequency
Figure 14.19. Shift in phase at fixed Drive frequency.
14-21
Under Interleave Controls set the Lift start height to 0 nm, and Lift scan
height to 100 nm. (The lift height can later be optimized.) Set the remaining
Interleave parameters (Setpoint, Drive amplitude, Drive frequency, and
gains) to the main Feedback Controls values. This can be done by setting the
flags (left of the Interleave Control column) to off (grayed bullets).
Select the Interleave Controls panel. Set the Setpoint, Drive amplitude,
Drive frequency to Main (off grayed bullets). Verify that Interleave scan is
set to Lift.
Choose a Lift scan height of 100 nm. This will be optimized later. Set the
Channel 1 image Data type to Phase and choose Retrace for the scan Line
direction on both Channel 1 and 2 images.
Switch Interleave mode to Enable to start LiftMode. Set the Channel 1 Scan
line to Interleave to display interleave data. This screen should now display the
cantilever phase change due to electrical force gradients from the sample in the
left image and topography in the right image.
Optimize the Lift height. For high-resolution, make the Lift scan height as
small as possible without crashing the tip into the surface. If the tip crashes into
the surface it creates bright or dark streaks across the image. Also, if the Lift
scan height is set extremely low, the tip may continuously tap on the surface
during the LiftMode scan. Check this by toggling between the Interleave and
main scan lines for the phase image. The two images will look very similar if
the tip is continuously tapping on the surface during the LiftMode scan. In this
case, increase the Lift scan height until the Interleave scan image abruptly
changes, indicating that it is now oscillating above the surface and not
continuously tapping.
Adjust the sample or tip voltage to confirm that contrast is due to electrical force
gradients. On very rough samples, contrast in LiftMode images may be from
air damping between the tip and surface. It is often useful to look at the phase
data in Scope Mode while adjusting the tip or sample voltage up and down.
Contrast due to electrical force gradients should increase or decrease as the tipsample voltage is changed.
For more quantitative results, switch the to the frequency Data Type for
Channel 1. This technique provides a direct measure of the change in resonant
frequency felt by the cantilever. It may be necessary to optimize the FM
(frequency modulation) gain to properly track the shifts in resonant frequency.
This is described in detail in Chapter 13 of the appropriate product instruction
manual.
14-22
Amplitude
Detection
Notes
This imaging method, although described here, is not recommended without the
Extender Electronics module due to the presence of artifacts.
Amplitude Detection, unlike Phase Detection, is available with or without the
optional Extender Electronics Module. This section describes the differences in
software set up and imaging for EFM systems without the Extender module. When
EFM imaging without the Extender module, changes in the cantilever amplitude
provide an indirect measure of shifts in the cantilever resonance frequency as
shown in Figure 14.19.
Amplitude
F0
Drive Frequency
Figure 14.20. Shift in amplitude at fixed Drive Frequency
(Extender Electronics Module not installed).
14-23
Set the Drive frequency to the left side of the cantilever resonance curve, as
shown in Figure 14.20 below.
For maximum sensitivity, set the Drive frequency to the steepest part of the
resonance curve. As the tip oscillates above the sample, a gradient in the
magnetic force will shift the resonance frequency F0; (see Figure 14.19).
Tracking the variations in oscillation amplitude while in LiftMode yields an
image of the electric force gradients. Either side of the resonance may be used,
though we have obtained slightly better results on the low side, as shown in
Figure 14.19.
Notes
14-24
Notes
Surface potential detection EFM is only possible using the Extender Electronics
Module. This section does not apply to microscopes which are not equipped
with the Extender Electronics Module.
The Extender Electronics Module allows measurement of local sample surface
potential. This is similar to techniques called Scanning Maxwell Stress Microscopy
and Kelvin Probe Microscopy. Surface potential detection is a two-pass system
where the surface topography is obtained in the first pass and the surface potential
is measured on the second pass. The two measurements are interleaved, that is, they
are each measured one line at a time with both images displayed on the screen
simultaneously.
A block diagram of the surface potential measurement system is shown in
Figure 14.21. On the first pass, the sample topography is measured by standard
TappingMode. In TappingMode the cantilever is physically vibrated near its resonant frequency by a small piezoelectric element. On the second pass, the piezo that
normally vibrates the cantilever is turned off. Instead, to measure the surface potential, an oscillating voltage V ac cos t is applied directly to the cantilever tip. This
creates an oscillating electrostatic force at the frequency on the cantilever. The
oscillating force has the following amplitude:
dC
F = ------- V dc V ac
dz
dC
where ------- is the vertical derivative of the tip/sample capacitance.
dz
V dc = V tip V sample , the dc voltage difference between the tip and the sample,
14-25
and V ac is the amplitude of the oscillating voltage applied to the cantilever tip.
Cantilever Deflection Signal
Photodiode Signal
RMS Detector
Amplitude Signal
Lock-in
Amplifier
Reference
Signal
Laser
Beam
Photodetector
Tapping
Piezo
To
Tip
Servo Controller
DC Voltage
Sum
AC
Signals to
NanoScope
(Feedback loop
adjusts DC tip
voltage to zero lockin signal)
Potential Signal
Sample
Oscillator Signal
GND
High Resolution
Oscillator
14-26
Notes
14-27
3. Depending upon whether voltage is to be applied to the sample directly or indirectly, reconfigure jumpers if indicated.
Sample
Electrical Insulator
Sample Chuck
14-28
Ground
External Voltage
Source
(+)
Piezo Cap
>10 M
GND/OSC + DC
Unused
GND/OSC + DC
To AFM Tip
Auxiliary D
STM
Indicates jumpers
14-29
Mode
Tip or Sample
Voltage
FM/Phase
Gnd/Surface Potential
Surface Potential
To
cro
sc
To
e
Scop
Nano
Analog 2
Mi
op
Mode
FM/Phase
TappingMode
Contact AFM
MFM
GND/Surface
Potential
Analog 2
Surface Potential
Apply voltage to
tip or sample (Use
for electric field
gradient imaging;
tunneling AFM)
14-30
Notes
14-31
force on the cantilever and this makes for more sensitive potential measurements.
Conversely, the maximum total voltage (ac + dc) that may be applied to the tip is
10 V. So a large Drive amplitude reduces the range of the DC voltage that can be
applied to the cantilever. If the sample surface potentials to be measured are very
large, it is necessary to choose a small Drive amplitude, while small surface potentials can be imaged more successfully with large Drive amplitudes. To start choose
a Drive amplitude of 5 V.
9. Set the Channel 2 image Data type to Potential. Set the scan Line direction for
the main and interleave scans to Retrace. Remember to choose the Retrace direction because the lift step occurs on the trace scan and data collection occurs on the
retrace.
10. Choose a Lift start height of 0 nm and a Lift scan height of 100 nm. The Lift
scan height can be readjusted later.
11. Switch Interleave mode to Enable to start LiftMode. Now, when the microscope completes a topographic scan line (trace and retrace) the system turns off the
TappingMode piezo and switches the oscillator signal to the cantilever. The cantilever is driven electrostatically according to the interleave Drive amplitude that has
been selected. Also, when Potential is selected as the Data type for the Channel 2
image, a feedback circuit is enabled in the Extender box which adjusts the dc voltage on the tip to maintain the cantilever oscillation amplitude at zero. To do this, the
feedback circuit uses the lock-in signal of the cantilever oscillation and tries to keep
this value at zero volts. As detailed in the Section 14.5 above, when the cantilever
oscillation amplitude has returned to zero, the dc voltage on the tip and sample are
the same. The NanoScope records the dc voltage applied to the tip and this signal is
displayed in the Potential data type.
12. Adjust the FM gains. The feedback loop that is used by the Extender Electronics
for surface potential measurements is the same as the one used in Frequency Modulation (FM) for magnetic and electric force gradient detection, as described previously. The feedback loop should be tuned in a similar manner to the FM setup.
Select Other Controls and adjust the FM gains. Setting both FM Integral gain and
FM Proportional gain to 15.0 is a good starting point. As with the topography
gains, the scan can be optimized by increasing the gains to maximize feedback
response, but not so high that oscillation sets in. The gains often need to be much
lower for potential measurements than for standard FM measurements. More information on tuning the feedback loop is given in the Troubleshooting section below.
13. Optimize the Lift heights. Set the Lift scan height at the smallest value possible
that does not make the Potential feedback loop unstable or cause the tip to crash
into the sample surface. When the tip crashes into the surface during the Potential
measurement, dark or light streaks appear in the Potential image. In this case,
increase the Lift scan height until these streaks are minimized.
14-32
14. For large sample voltages or qualitative work, select Data type = Phase instead
of Potential. When the Extender box has been configured for surface potential
measurements, the phase signal is actually the cantilever amplitude signal, as
measured by a lock-in amplifier. If the feedback loop is not enabled by selecting the
Data type = Potential, the lock-in cantilever amplitude depends on the voltage difference between the tip and sample in a roughly linear fashion. (The lock-in amplifier produces a voltage that is proportional to the cantilever amplitude.) Qualitative
surface potential images can be collected using this lock-in signal. Also, if the sample has a surface potential that exceeds 10 V (greater than the range of the Potential signal), it is possible to use the lock-in signal to provide qualitative images that
reflect the sample surface potential. To view the lock-in signal with the reconfigured Extender box, select the Data type = Phase.
Go into Scope Mode and look at the Potential signal. If oscillation noise is
evident in the signal, reduce the FM gains. If oscillations persist even at very
low FM gains, try increasing the Lift scan height and/or reducing the Drive
amplitude until oscillation stops. If the tip crashes into the surface the Lock-in
signal becomes unstable and can cause the feedback loop to malfunction.
Increasing the Lift height and reducing the Drive amplitude can prevent this
problem. Once oscillation stops, the FM gains may be increased for improved
performance.
In Scope Mode, if the Potential signal is perfectly flat and shows no noise even
with a small Z-range, the feedback loop is probably stuck at 10 V. (You can verify
this by changing the value of Real-time planefit to None in the Channel 1 panel.)
Reduce the Scan rate and watch the display monitor which indicates the cantilever
amplitude. On the topographic trace, the voltage displayed should be the setpoint
selected for the Main scan. On the Potential trace, this voltage drops close to zero if
the cantilever oscillation is being successfully reduced. If the value on the display
monitor instead goes to a large nonzero value, the feedback loop is probably not
working properly. In this case, try reducing the Drive amplitude and increasing the
Lift scan height. It may also be helpful to momentarily turn the Interleave mode
to Disabled, then back to Enabled. Also try reducing any external voltage that is
being applied to the sample to stabilize the feedback loop, then turn the voltage
back up.
14-33
14-34
This chapter provides detailed instructions for the fine calibration of Digital Instruments MultiMode SPMs. Additionally, the latter part of the chapter focuses on
problems commonly encountered during operation of the microscope and then concludes with maintenance procedures for the MultiMode SPM adjustment screws.
Operators should establish a three-month service schedule for mainentance and calibration. Check the SPMs measuring accuracy periodically to ensure that images
are dimensionally represented within acceptable limits of error. If measuring accuracy is critical, or if environmental factors (e.g., humidity, temperature) impact the
SPM significantly, this may require a quick check at the start of each imaging session.
Adjust calibration parameters until the SPMs dimensions accord with the true
dimensions of the reference.
15-1
Part V: Utilities
Voltage
440 V
110 V
0
0
This curve typifies scanner sensitivity across the full range of movement. The vertical axis denotes voltage applied to the scanner. The horizontal axis denotes scanner
movement. At higher voltages, the scanners sensitivity increases (i.e., more movement per voltage applied). At zero volts, the scanner is motionless. Plotting each
point along the curve describes a second-order, exponential relationship which provides a rough approximation of scanner sensitivity.
Unfortunately, things are more complicated. Because piezo materials exhibit hysteresis, their response to increasing voltage is not the same as their response to
decreasing voltage. That is, piezo materials exhibit memory, which causes the
scanner to behave differently as voltages recede toward zero. The graph below represents this relationship.
15-2
Part V: Utilities
Scanner
Voltage
Scanner
Movement
Time
To produce the sharp, linear movements (triangular waveform) required for accurate back-and-forth scanning, it is necessary to shape the applied voltage as shown
on the top graph above. Moreover, the applied voltage must compensate for scan
rate and scan size. As scan rate slows, the applied voltage must compensate for
increased memory effects in the piezo material. As scan size is decreased, the piezo
exhibits more linearity. These effects are further complicated by X-Y-Z coupling
effects (the tendency for one axis to affect movement in other axes).
Through rigorous quality control of its scanner piezos, Digital Instruments has
achieved excellent modeling of scanner characteristics. Two calibration points are
typically used for fine-tuning: at 150 and 440 volts. (A third point is assumed at 0
nm/volts.) These three points yield a second-order sensitivity curve to ensure accurate measurements throughout a broad range of scanner movements.
Because scanner sensitivities vary according to how much voltage is applied to
them, the reference must be thoroughly scanned at a variety of sizes and angles.
The user then tells the software the distance between known features on the references surface, and a parameter is recorded to compensate the scanners movements. The X, Y and Z axes may be calibrated in any sequential order; however, the
linearization adjustments (Section 15.4) must be performed before any calibrations
are attempted (otherwise, they will be undone by the linearity adjustments).
15-3
Part V: Utilities
15-4
Part V: Utilities
The diagram below represents scanner crystal voltage versus photodiode voltage. In
this instance, detector sensitivity is given as volt per volt, a parameter provided in
the Force Calibration screen.
Photodiode voltage
Laser
Photodiode
array
Cantilever
-220 VDC
0 VDC
+220 VDC
Scanner
Photodiode Voltage
+3.0
Detector Sensitivity
-3.0
-220
0
Scanner Voltage
+220
15-5
Part V: Utilities
180 nm deep
10 m
10 m
15-6
Part V: Utilities
Scan angle = 0
15-7
Part V: Utilities
Panel
Parameter
Setting
Scan Controls
Scan Size
440 V
X offset
0.00 nm
Y offset
0.00 nm
Scan angle
0.00 deg
Scan rate
2.44 Hz
Other Controls
Channel 1
Number of samples
256
Enabled
Z limit
440 V
Units
Volts
Data type
Height
Z range
~ 20 Va
15-8
Part V: Utilities
below). The angle should be measured with the vertex near the center of the image
and the vertices in the upper-right or lower-left quadrant. The angle will be displayed on the white status bar at the bottom of the display monitor. Use the mouse
to drag the cursor until it is oriented correctly,
Figure 15.2. Left image: non-orthogonal image. Notice how pits are aligned
with the vertical (slow) axis but skewed with the horizontal (fast) axis.
Right image: a corrected, orthogonal scan.
then read the angle off the status bar. If the angle differs by more than a half degree,
a correction will be required; otherwise, skip ahead to the step below.
15-9
Part V: Utilities
image from the previous scan. Also, be careful to compare only the part of the scan
drawn since the last parameter change.
Because the display monitor screen itself is not linear, the best way to check mag0
and arg is with the Real-time Zoom box.
A. Adjusting Fast mag0
After engaging, click on Microscope / Calibrate / Scanner. This will invoke the
Scanner Calibration window. As parameters values are changed, the effects will
be seen on the display monitor. Now move the mouse cursor to the display monitor
and select Zoom Out. This will produce a box whose size and position can be
changed by alternate clicks on the left mouse button. Adjust the box until it is about
one-third the size of the scan. Click once on the right mouse button to park the box
and free the cursor, which will jump to Execute. (Clicking twice will execute a
Zoom, which we don't want to do.)
Select (highlight) the scanner parameter to be modified. The first one to do is Fast
mag0. Now move the zoom box to the start of the fast scan (on the left if the Line
direction is set to Trace). Move and size the zoom box until the beginning third of
the scan's features are exactly aligned with the zoom box. The beginning third of the
scan will be the standard for judging almost all of the linearity values. It is a good
technique to ignore the set of features near the edges of the scan since these may be
distorted slightly. Now move the zoom box to the ending third of the scan. Align
one side of the zoom box with some features and observe how well the other side
aligns with the features under it.
Compared to the beginning third of the scan, if the features in the ending third of
the scan are too big to fit the same number of them inside the zoom box, type a
smaller number for Fast mag0. If the features are too small, increase Fast mag0.
Change mag0s by about 0.1 to 0.3 units at a time. Remember, every time a parameter is changed, the entire scan axis is affected, so after every change it is necessary
to resize the zoom box at the beginning third of the scan and re-compare the ending.
15-10
Part V: Utilities
Center Compressed
Center Expanded
Notes
After an adjustment in Fast arg is made, Fast mag0 may need to be readjusted.
Continue steps A and B until the rulings are evenly spaced across the Fast-axis.
(If using a one-dimensional reference, it may be desirable to do step E
{Adjusting Fast mag1} before proceeding; because adjusting the slow linearity
will require the one-dimensional reference to be physically rotated 90 degrees.
After finishing step E, return to here.)
After Fast mag0 and Fast arg have been set, insert these values for Slow mag0 and
Slow arg. These will serve as close starting points before adjusting the slow linearities.
C. Adjusting Slow mag0
Follow the same instructions as for Fast mag0. Setting the slow linearities requires
a lot of time because after making a parameter change it is necessary to wait until a
fresh third of a scan has begun with the new parameters before the zoom box can be
resized. Move the resized box to the ending of the scan and be prepared to measure
the ending quickly. If the user is quick enough, a new parameter value can be typed
15-11
Part V: Utilities
in before the scan starts again. Be careful not to confuse scan top and bottom with
beginning and ending, since the scan direction alternates. Adjust the zoom box to fit
the beginning third of the scan and check the ending third. If the features at the ending third are too big for the box, reduce the parameter. If the features are too small,
increase. Be careful to compare only parts of the current scan, not the previous one.
Center Compressed
Center Expanded
15-12
Part V: Utilities
Fine Adjustment
Initial adjustment is usually adequate; however, if more precision is desired, adjust
Fast mag1 using the same procedure as in step A (see Adjusting Fast mag0). As
before, set the Zoom box for the beginning of the scan and then check the ending.
(Because the scan is small, it may be necessary to use a Zoom box up to one-half as
large as the scan). If the ending of the scan is bigger than the beginning, reduce
Fast mag1. If the ending is too small, increase the value.
F. Adjusting Slow mag1
With the Scan size still set to 150 volts, select Microscope / Calibrate / Scanner.
Select Slow mag1 and input the value from Fast mag1. For medium-sized scanners
(C to F), it is probably worth stopping here, because the Slow mag1 value is usually
100-120 percent of the Fast mag1 value. However, if more precision is desired,
adjust Slow mag1 using the same procedure as in step C (see Adjusting Slow
mag0). As before, set the Zoom box for the beginning of the scan and then check
the ending. (Because the scan is small, it may be necessary to use a Zoom box up to
one-half as large as the scan). If the ending of the scan is bigger than the beginning,
reduce Slow mag1. If the ending is too small, increase the value.
Wait one complete frame with the new value before readjusting the Slow mag1
value.
This completes the linearity adjustments. Check the final result by capturing an
image and checking it with the Off-line / Modify / Zoom window. The scanner is
now ready for calibration of the X and Y parameters.
15-13
Part V: Utilities
Quit
The Capture Calibration dialog box lists twelve parameters used in the calibration
procedure. If this is a first-time calibration, or if the microscopes calibration has
not been checked within the last three months, verify that all parameters are
selected. Click on the Capture button to initiate the automatic calibration routine.
Notes
The microscope will begin an automatic series of scans on the reference which
require about one hour total to complete. During each scan, the scanner moves
the piezo using carefully calculated movements. Many of these movements are
unusual, giving rise to a variety of images which do not look like the normal
reference. For example, pits may resemble trenches and features may be
presented at various angles.
3. As each routine is executed, the user may need to adjust the scan slightly to optimize the calibration image. This is accomplished using the Capture Control dialog
box, which is displayed on the control monitor throughout the calibration routines.
Capture Control
Capturing at 311V, 45
Setpoint:
0.00 V
Adjust Y offset
15-14
Up
Down
Skip
Abort
Part V: Utilities
Note the status bar at the bottom of the control monitor. The capture status will
start at skip 2meaning that the program will skip the current scan and one more
before capturing an image for later calibration. (This allows hysteresis and drift to
settle out when the scan changes direction and size between images.) If the scan has
not settled out by the time the capture status changes to On, click on the Skip button to increment the capture to skip 1, skip 2, or skip 3. Do not click on Abort
unless you want to stop the entire Capture Calibration program.
Features should extend across as much of the displayed image as possible. If portions of features are missing, or if the image is blank, click repeatedly on the
Adjust Y offset / Up or Down button to adjust the scan until more of the features
are imaged. When the image looks optimized, allow the software to capture the
image without disturbing it. The software will then automatically index to the next
image.
28.37 m
15-15
Part V: Utilities
Notes
All images unsuitable for calibration must be recaptured. Record the file name
extensions of any unusable files (e.g., .cxy, .dyy), then delete them. Reengage
the reference surface and select Capture Calibration again. Verify that the file
name prefix is identical to that of the usable files, then deselect all usable file
name extensions from the last capture (click on their box to remove the X).
Finally, click on the Capture button to recapture the selected files.
Quit
Verify that the file name prefix assigned to the captured files from the Capture Calibration routine described in Section 15.5 above is correct.
2. For normal calibration, verify that all parameters are selected in the dialog box,
then click on Calibrate to execute the routine.
15-16
Part V: Utilities
The software will sequentially present various calibration images on the display
monitor while prompting the user to draw either a vertical line or a horizontal line.
The control monitor will simultaneously display various dialog boxesone for
each imagerequesting that a distance be entered.
3. Use the mouse to draw a line on the image. The line should be drawn to span as
many features as possible, preferably connecting like edges. For example, consider
the following:
Autocalibration
35.95
Ok
The distance displayed in the box (in this example, 35.95 m) is the softwares estimate (based on current calibration values) of the length of the line drawn on the
image. If the line length is different from the value shown, it will have to be
changed. (In this example, the user would enter a value of 40.)
4. Enter the distance covered by the white line drawn on the image. If a 10-micron
reference is being employed, like portions of features are spaced 10 microns apart
(e.g., between bottom edges, left sides, etc.).
15-17
Part V: Utilities
Please Note: Features must be measured without regard to how they appear in calibration images. Remember, features may be represented as having stretched, distorted, or angled appearances due to the unusual movements employed during
Capture Calibration scanning. Regardless, features are separated by the same
(e.g., 10-micron) spacings.
5. Continue drawing lines and entering measured distances until all Capture Calibration images have been measured. When the software is finished, it will prompt
the user that it is done.
If Capture Calibration and Autocalibration routines have been completed correctly, the SPM should be calibrated within 1-2 percent accuracy over most of the
scanners measuring range. Should it be necessary to obtain still better accuracy, the
SPM can be fine-tuned to obtain maximum measuring accuracy. This is accomplished through the use of calibration parameters discussed in the next section.
Notes
15-18
Part V: Utilities
That is, divide the known distance by the distance displayed next to the line drawn
in step #2 above. Record this quotient.
15-19
Part V: Utilities
4. Click on the Real time / Microscope / Calibrate / Scanner option. The Scanner
Calibration dialog box will be displayed:
Scanner Calibration
X fast sens:
X fast derate:
X slow sens:
230 nm/V
0.230
nm/V2
267 nm/V
Y fast derate:
Y slow sens:
228 nm/V
0.218 nm/V2
257 nm/V
Y slow derate:
0.270 nm/V2
XsXf coupling:
0.548
nm/V2
YsYf coupling:
0.569 nm/V2
0.027 pm/V3
0.222 pm/V3
XsYf coupling:
0.078 nm V2
YsYf coupling:
0.134 nm V2
-0.04 pm/V3
0.0668 pm/V3
X slow derate:
0.286 nm/V2
Y fast sens:
Fast mag0:
1.40
Slow mag0:
1.47
Fast mag1:
1.05
Slow mag1:
0.860
3.00
Slow arg:
Fast arg:
Fast cal freq:
2.44 Hz
Piezo cal:
440 V
Rounding:
Allow rotation:
Allow
Orthogonality:
Ok
3.00
4.77 Hz
0.00
0.00 deg
Cancel
Multiply the quotient obtained in step #3 above by the X fast sens value shown on
the Scanner Calibration panel, then enter the new value. The new value will adjust
the scanners fast axis to more closely match calculated distances with actual feature distances. The new sensitivity setting will take effect as soon as it is entered. To
save it to the computers hard disk, click on the Ok button.This will also close the
Scanner Calibration panel.
15-20
Part V: Utilities
tuning will be required; go to step #3 below. If the displayed distance agrees with
the known distance, skip to Section 15.7.4 below.
3. Based upon the results in step #2 above, perform the following calculation:
That is, divide the known distance by the distance displayed next to the line drawn
in step #2 above. Write this value down.
4. Click on the Real time / Microscope / Calibrate / Scanner function to display
the Scanner Calibration dialog box. Select the Y slow sens parameter.
5. Write down the Y slow sens value shown on the Scanner Calibration panel.
Multiply the quotient obtained in step #3 above by the Y slow sens value shown on
the Scanner Calibration panel, then enter the new value. This should adjust the
scanners slow axis to more closely match calculated distances with actual feature
distances. To save the new parameter value, click on the Ok button.
15-21
Part V: Utilities
-------------------------------------------------------------440 v
Where s is the Sens value; a is the actual distance; d is the derating value; m is the
measured distance; and, v is the Scan size in volts.
4. Click on the Real time / Microscope / Calibrate / Scanner function to display
the Scanner Calibration dialog box. Select the X fast derate parameter.
5. Write down the X fast derate value shown on the Scanner Calibration panel.
Multiply the value obtained in step #3 above by the X fast derate value shown on
the Scanner Calibration panel, then reenter the new X fast derate value. This
should adjust the scanners fast axis to more closely match calculated distances
with actual feature distances. To set the new parameter value, click on the Ok button.
15-22
Part V: Utilities
4. Write down the Y slow derate value shown on the Scanner Calibration panel.
Multiply the value obtained in step #2 above by the Y slow derate value shown on
the Scanner Calibration panel, then enter the new value. This should adjust the
scanners slow axis to more closely match calculated distances with actual feature
distances. To set the new parameter value, click on the Ok button.
15.8. Calibrating Z
With respect to obtaining accurate Z-axis measurements, users should be aware that
while it is generally not difficult to obtain accurate X-Y calibration references, it is
much more difficult to obtain accurate Z-axis results. Users should be aware that Zaxis calibration is very sample-dependent. It is not easy to control Z piezo dynamics because the Z-axis does not move at a constant rate, as the X- and Y-axes do during scans. Furthermore, offsets can effect the piezo over a period of minutes. The
platinum-coated, silicon calibration references distributed by Digital Instruments
have 180 nm vertical features which are accurate to within 3 percent and are used
throughout the examples provided within this section. If greater accuracy is desired,
an appropriate calibration standard must be selected, and/or a metrology head
employed using a Digital Instruments Dimension Series microscope.
15-23
Part V: Utilities
Scope Trace
Z range
35nm/div
Scan
Figure 15.6. Scope Trace screen will display a step when the tip encounters
a pit in the reference sample.
After finding the step in the Scope mode screen, click back to Image mode. The
image should now encompass a portion of one pit.
Scanners larger than A can image an entire pit. It is not necessary to change the
Realtime and offline planefits for these scanners.
15-24
Part V: Utilities
If the image encompasses one pit, or a portion of one pit (if you are using an A
scanner), cover as much of the pit as possible with a stopband while leaving the
remainder of the top surface exposed. Click on Execute to complete the flattening
procedure, then Quit the dialog box.
15-25
Part V: Utilities
Cursor
Depth Scale
Hist Cursor
Stopband
Depth Ref
Zoom
Execute
Clear
Depth [nm]
0.50
124.12 m2
82.266 nm
78.049 m2
62.891
138.24 nm
4.361 m3
4.008 m2
3.229
182.69 nm
200
Box area
Center line av
Bearing area
Bearing ratio
Bearing depth
Bearing volume
Hist area
Hist%
Hist rel depth
Depth [nm]
100
0.25
Bearing Analysis
Cursor: Box
Depth: Auto
2.50 5.00
Hist%
Ref: Peak
7.50
50
Bearing area%
100
Scale: Full
Use the mouse to slide the top (red) line cursor on the Hist% graph to the centerline of the top data spike as shown in the example above. Select Hist Cursor / On
to generate another (green) line cursor; position this one on the centerline of the
bottom data spike. The distance between the two cursors is given at the Hist rel
depth field inside the data box. (Depending upon whether the red cursor is positioned over the top or bottom data spike, the Hist rel depth value may be either
positive or negative. It doesnt matter; take the absolute value.)
If a 10-micron, silicon calibration reference from Digital Instruments has been
used, the Hist rel depth value should be very close to 180 nm. Its variance from
this value gives a good indication of the microscopes measuring error in the Z-axis.
If the value is more than 2% from 180 nm, it will be necessary to make adjustments
(see Section 11.8.5 below); write down the Hist rel depth value now.
To exit the Bearing dialog box, click on Quit.
15-26
Part V: Utilities
That is, divide the actual depth of features (180 nm for the 10-micron calibration
reference) by the measured depth (indicated in Bearing analysis by the Hist rel
depth value). Multiply this quotient by the Z sensitivity value in the Z Calibration
dialog box, then replace the value with the product. Click on the Ok button to enter
the new Z sensitivity value.
Note
The numerator value above (180 nm) is for Digital Instruments 10-micron
silicon reference. For other calibration references, set the numerator equal to the
depth of features measured by Bearing analysis. Ideally, calibration references
used should have features with heights comparable to those being imaged and
measured on samples.
Note
15-27
Part V: Utilities
Note
180nm
1
meas. depth
Part V: Utilities
Panel
Scan Controls
Feedback Controls
Parameter
Setting
Scan Size
12 nm
X offset
0.00 nm
Y offset
0.00 nm
Scan angle
0.00 deg
Scan rate
61.00 Hz
Number of samples
512
Enabled
Z limit
440 V
Integral gain
12.00
Proportional gain
4.00
LookAhead gain
0.00
Setpoint
0V
15-29
Panel
Parameter
Setting
Other Controls
AFM mode
Contact
Input attenuation
1x
Interleave Controls
Interleave mode
Disabled
Channel 1
Data type
Height
Highpass filter
34
Lowpass filter
Part V: Utilities
3. Engage the surface and adjust the Integral gain and Setpoint to obtain a good
image. The Setpoint should be kept low if possible, and the Z Center Position
should be close to 0 V. Notice that the Scan rate is set much higher (~ 61 Hz) for
atomic-scale images, this to defeat some of the noise due to thermal drift.
If difficulty is experienced obtaining an image, Withdraw and try a different site on
the surface, then Engage again. Many users find it easiest to obtain good images
measurements if the sample is rotated until atoms are oriented vertically.
4. Once a satisfactory image is obtained, Capture it. The image should appear similar to the image of graphite shown below. The captured image will be used for the
calibration procedure described in the next several steps.
Figure 15.7. Typical atomic scan of graphite. Note the highly regular lattice
of the atoms. The cursor line describes a distance of 6.66 .
15-30
Part V: Utilities
5. Go to the Off-line / View / Top View option and measure the spacings between
atoms using the mouse. Depending upon whether the sample is graphite or mica,
the spacings should measure as shown below.
A
C
A = 0.519 nm
B = .900 nm
C = 1.37 nm
A
C
A = 0.255 nm
B = .433 nm
C = .666 nm
15-31
Part V: Utilities
The derating parameters are not adjusted for atomic-scale imaging, including:
X fast derate
X slow derate
Y fast derate
Y slow derate
Retracted offset der
Extended offset der
15-32
Part V: Utilities
15-33
Part V: Utilities
the Vertical Deflection voltage more negative than the setpoint voltage before trying
the engage sequence again.
Sources of false engagement include:
Incorrect optical alignment of laser spot on cantilever. This alignment will cause
a greater amount of laser light to reflect off of the sample.
The sample has a region on it that touches the cantilever before the tip does.
Foreign material stuck on the cantilever beam that is lower than the tip.
Check the cabling between the computer and the controller, and between the
controller and the microscope. Any discontinuity in the microscope signals can
cause an immediate engage.
The Setpoint may be set more negative than the vertical deflection (A-B)
voltage (this applies only to contact AFM modes). This false engagement is
immediate and the computer will not show any motor travel or time delay after
one gives the command. To correct for this condition, select Withdraw and
check that the vertical deflection voltage reads a voltage more negative than the
setpoint voltage.
15.10.2. Contact AFM: Head appears engaged but does not track
surface features
Occasionally, the head appears to be engaged but surface features do not appear on
the display monitor. This usually results from false engagement (see Section 15.1.1
above), poor laser alignment, or something other than the tip coming in contact
with the sample surface.
An easy way to check for laser alignment problems is to check the laser beam using
a magnifier or slip of paper. Click on the Withdraw command to remove the cantilever from the surface and recheck alignment with the magnifier or paper method
(see Chapter 5). When satisfied with the alignment, readjust the top-bottom differential signal before re-attempting engagement.
Occasionally, when the optical head is not leveled properly, something other than
the cantilever tip (usually the corner of the substrate or the cantilever clip) will contact the surface during engagement. The system may engage, but the image will not
contain any features. Testing the SUM signal after withdrawing the tip several times
is a good test for this failure mode. Generally, the SUM signal will be far from its
original value if something other than the tip contacts the surface. The head should
be leveled and the laser realigned if this has occurred.
15-34
Part V: Utilities
The tip may have started too far from the sample. Lower the tips height above
the sample surface by using the coarse alignment screws, then repeat the
engagement procedure.
The laser may be misaligned and/or the cantilever may be broken. When this
occurs, replace the cantilever substrate, realign the laser, and repeat the
engagement procedure.
AFM tip picking up contamination. The tip will effectively become longer and
this will cause the feedback loop to raise the tip to keep the same tracking force.
The contamination can come off of the tip which will cause another level shift in
the image. This problem will show up as large bands in the captured image.
Friction. Some samples have a stronger frictional interaction with the tip than
others. The cantilever will bend and straighten due to the tip sticking and
slipping as it is dragged across the surface of the sample. The result is a line-by-
15-35
Part V: Utilities
line level shift in the captured image. The trace and retrace scan directions can
actually be inverted from each other if the friction is high enough. A good
practice is to use the scope image mode in dual trace display. The trace and
retrace directions should be close to each other. Trace and retrace can invert if
there is friction present between the tip and the sample surface.
Note
The gains must be set low to reduce movement of the Z piezo when acquiring
deflection data used in height measurements.
15-36
Part V: Utilities
Mechanical drift will cause the sample-to-tip distance to change slowly bringing
them apart completely, or too close together. If this is the case, the Z center will
show either +220 (extended) or -220 (retracted), respectively.
Verify that the sample, tip and stage are all stabilized. There should be no free
movement between any of these components
Recheck that the cantilever is firmly seated in its place and that there are no dirt
particles wedged beneath it. Check that everything is tight and secure. Once this
is checked, there may still be some drift. Using the Tip Up and Tip Down
command can bring the sample back into range. (This may have to be done due
to any number of factors.)
Check for thermal stability. The SPM should not be located directly in the path
of heating or air conditioning ducts; also, avoid locating the SPM near large
windows which trap solar heat. Thermally caused drift due to thermal expansion
of SPM components is the most common cause of mechanical drift. The
MultiMode tends to heat up and drift when used with the acoustic isolation hood
because it traps heat produced by electronics in the base and head.
15-37
Part V: Utilities
15-38
Part V: Utilities
15-39
Part V: Utilities
Reduce the setpoint voltage. This increases the amount of tapping force on the
surface. This is probably the thing that will be most effective. Be careful when
doing this on soft samples. The sample surface can still be disturbed even
though the forces are very small.
Reduce the scan rate. The scan rate needs to be slower in TappingMode than in
contact AFM. Typically it should be around 1-3Hz.
Increase the integral and proportional gains. This will speed up the response
time of the Z piezo transducer.
15-40
Part V: Utilities
Repeat the steps in the lines across the image troubleshooting description to
adjust the RMS voltage.
Notes
Operating with a smaller difference between the RMS voltage and the Setpoint
voltage means that less tapping force is being used. This may damage the tip
because the tip may not be following the surface, allowing it to crash into
vertical features.
Use the fluid cell without the o-ring as described in Chapter 7 of this manual.
This technique works best on hydrophobic surfaces. If you have a surface that
does not allow a liquid drop meniscus to form or stabilize (e.g., mica), a water
repellant marker may be used to outline a boundary for holding the drop. This
type of marker, the PAP PEN, can be purchased from the The Binding Site,
5889 Oberlin Drive #101, San Diego, CA 92121; 1-800-633-4484; Fax: 619453-9189.
15-41
Part V: Utilities
Notes
For applications where fluid flow is necessary, this option should not be used.
Be certain to check for leaks and wipe up any spilled fluid as soon as possible to
avoid damage to the scanner.
The new vertical engage scanners allow the tip to approach the sample without
the lateral movement caused by the stepper motor. Thus, lateral stress on the oring is absent. Vertical engage scanners are available in the 125-micron (J),
and 10-micron (E) sizes. Contact Digital Instruments for more information.
Lightly coating the area of the o-ring which contacts the sample surface with
white petrolatum or vacuum grease will allow the o-ring to slide across the
surface, minimizing lateral stress. This also forms a better fluid-tight seal
between the o-ring and sample. Be aware that some solvents (e.g., non-polar
organic solvents) may dissolve some lubricants into the fluid.
Replace the o-ring with a slice of thin-walled glass, plastic, or stainless steel
tubing. The diameter and thickness of the ring of tubing should be chosen to
prevent contacting the inner or outer walls of the curcular groove in the glass
cantilever holder. This gives the head more room to move laterally during
engagement and for positioning the tip over the sample surface. The height of
the ring of tubing must be chosen so that it is not too tall to prevent the tip from
reaching the sample surface, or so short that the ring does not reach the bottom
of the glass cantilever holder before engagement. Glue the ring of tubing to the
steel sample puck or to the sample to prevent leaks.
Notes
When using the ring of tubing in the fluid cell, fluid cannot be circulated
through the cell as when using the supplied o-ring.
When positioning the o-ring on the surface, adjust the positioning knob at the
base of the head to move it slightly forward. This will counter some of the
lateral stress on the o-ring due to the head shifting rearward during engagement.
To prevent leaks when using the supplied o-ring, it is best to place the o-ring on
the sample before placing the head on the scanner.
15-42
Part V: Utilities
15.11.1. Inspection
Small Sample SPMs utilize fine-pitch (1/4-80) adjustment screws; in some scanners, two of them are manually turned, and a third (rear) screw is turned by a motor
located in the microscope base. The Vertical Engagement JV scanner has just one
adjustment screw. The diagram below illustrates a screw and adjacent hardware
installed on later-model (post-August 1993) microscopes. Earlier models have their
brass threaded inserts installed at the top of the scanner body (rather than the bottom) and feature a Teflon bushing. Inspection and service is similar for all types.
(top)
(Scanner body)
Ball Bearing
Setscrew (applies
pressure to bushing)
Plastic bushing
Screw
Knob
15-43
Part V: Utilities
Adjustment screws are threaded into brass inserts, which are affixed to the scanner
body with epoxy. Although screws are not heavily lubricated, a light film of oil is
applied to them at the factory to prevent galling. This allows sufficient lubrication
for fine adjustment, while minimizing drift (i.e., loosening) between the screws and
scanner body due to the slow displacement of lubricant from screw threads.
Problems develop whenever screws become fouled with fine grit: screws may be
difficult to turn and/or exhibit any of the following symptoms:
Notes
If screws are frozen (i.e., cannot be rotated), DO NOT attempt to force them!
Return the scanner body and screws to Digital Instruments for repair.
The user should inspect screws at least every three months, more often if possible.
It is a good practice to check screws whenever the scanner body is removed by turning manually and feeling for resistance. This is especially true of the rear, motoractuated screw, which may be fouled without the users notice.
15-44
Part V: Utilities
15-45
Part V: Utilities
Notes
It is not necessary to remove setscrews from the scanner base to clean bushings.
(Setscrews are used only to apply pressure to the shoulder of each adjustment
screw; see diagram above.) If setscrews are removed, they should be retorqued
to approximately 4 in/oz., or until the adjustment screw can be turned snugly
without binding.
15.11.5. Lubricate.
Apply a very fine layer of lubricant to each adjustment screw. Lubricant may consist of high-vacuum grease, optical coupling grease, or equivalent. Screws should
exhibit a slight sheen and no more, indicating that they have been finely coated.
Notes
Excessive use of oil and grease on screws can cause the head to drift slightly.
This is due to a slow displacement of lubricant between the screw and threaded
insert. As lubricant is slowly squeezed out between the screw and threaded
insert, the screw settles, causing the head to lower itself (and the tip) toward the
sample. This is especially apparent during atomic-resolution imaging.
15.11.6. Reinstall.
1. Verify that all screws, threaded inserts and plastic bushings are free of grit.
2. Replace adjustment screws in their threaded inserts. Carefully turn them down
clockwise until the shoulder of the screw engages the plastic bushing, and the
screws ball bearing appears at the top of the scanner body.
Notes
If screws bind during reinstallation, stop immediately. Back the screw out again,
reclean as described above, then redo installation. Recheck that the setscrew is
properly torqued at the plastic bushing. If screw still cannot be installed without
binding, return to Digital Instruments for repair.
3. Verify that each screw turns freely in both directions, but feels snug. (Adjustment
screws should turn freely while exhibiting sufficient rigidity to stabilize the head.)
15-46
Part V: Utilities
1. Parameter file name may be labeled with scanners serial number (e.g., 1234JV.PAR).
15-47
Part V: Utilities
Figure 15.8. Stabilizing screw for securing the vertical scanner to the
support ring. MultiMode AFM screw (left), and screw for other SPMs
(right).
One screw is supplied with a longer shoulder (P/N AFM-SCN-580); this screw
should be used with all MultiMode SPM bases. Use the short-shouldered screw
(P/N AFM-SCN-579) for all other SPM models. The screw is inserted through a
small hole on the front-underside of the support ring. Turn the screw until snug; do
not overtighten.
IMPORTANT! The use of a fastening screw is necessary to ensure optimal system
resolution and noise-free operation. DO NOT attempt to operate the scanner atop
the support ring without a fastening screw!
15-48
Part V: Utilities
NOTE:
There may be more than one \EQUIP directory, especially if the SPM uses more
than one version of the NanoScope software. If so, copy the file to ALL directories
and sub-directories labeled EQUIP.
The parameter file may be copied with any name, as long as it includes a .PAR
extension. Make certain the vertical scanners parameter file name is not the same
as a preexisting file; otherwise, it will overwrite the preexisting file. The parameter
file is copied using a DOS command. For example, for a J scanner:
COPY B:\MMAFMJ.PAR C:\EQUIP\1234JV.PAR
Once the parameter file is copied, reboot the NanoScope software to resume.
NOTE:
The vertical scanner parameter files are identical to all other J and E scanner
parameter files except for the Motor sensitivity parameter. On vertical scanners
this parameter value is larger than older models. A summary of Motor Sensitivity
parameter values is provided below.
Motor Sensitivity parameter value
(Tip travel per half step)
AFM, LFM,
TipView STM
MultiMode AFM
Vertical Scanners
80
110
3-screw Scanners
19
26
15-49
Part V: Utilities
15.13. Troubleshooting
15.13.1. Scanner is not properly calibrated.
Verify that the scanners parameter file has been properly copied to each of the
computers \EQUIP directories. Check the parameter file name; it should include a
.PAR extension (e.g., JVSCANR.PAR). Try to reselect the scanner using the Real
Time / Microscope / Select panel.
If the parameter file is correctly selected but measures inaccurately, it may be necessary to calibrate the scanner. This calibration procedure is detailed in the latest
revision of Support Note 217, Calibration Procedures.
15.13.2. Sample will not move (Tip Up/Down switch doesnt work).
If the vertical scanner is adjusted to its maximum tube height, the leadscrew may
disengage from the motor coupling on the MultiMode base. (This is a built-in feature to prevent motor burn-out.) To reengage the leadscrew, remove the scanner
from the support ring, then manually back the screw out several turns. Replace the
scanner on the support ring and reattempt motorized movement.
If the leadscrew becomes fouled with grit, rotation may become difficult or impossible. To inspect, remove the scanner from the support ring and turn the leadscrew
manually while feeling for resistance. If the leadscrew resists rotation, it will have
to be backed out, cleaned with solvent and lubricated. (DO NOT attempt to clean
the leadscrew while in the scanner body; solvents will ruin the scanner circuitry!) A
leadscrew cleaning procedure is detailed in Support Note 216 available from Digital
Instruments.
15-50
Part V: Utilities
NOTE:
If screw is frozen (i.e., cannot be rotated), DO NOT attempt to force it! Return the
scanner body and screw to Digital Instruments for repair.
15-51
Part V: Utilities
15-52
Index
A
Adjustment Screws
maintenance 15-43
Aliasing 11-37
Amplitude 8-19
Amplitude Calibration 8-25
Average count 11-22
B
Bias 9-11
Bias Voltage 9-1, 9-5, 9-6
C
Calibration
standard 15-6
Calibration Procedures 15-33
cantilever substrates 4-1
Cantilever Tune 8-8, 8-158-20
Cantilever Tune
initial settings 8-9
cantilevers 4-1
Cantilevers. See Probe Tips
Capture 9-12, 11-21
Capture Calibration 15-13
Center Frequency 8-9, 8-10
Color contrast 9-13
Color offset 9-13
Color table 9-13
Component List 3-13-2
Contact AFM
detector offsets 6-1
in fluids 7-1
principles of 2-2815-34
Contact Force 6-14, 8-25, 11-18
Index-1
D
Data type 6-13, 6-14, 8-19, 8-20, 9-11, 13-8
FM 11-34
LFM 10-1
STM 9-49-5, 9-6
Date type 13-8
Deflection 6-13
Dimension 5000
hardware described 1-2
Dimension 5000 1-1
DNA
imaging with TappingMode 8-41
Drive Amplitude 8-24, 11-24
Drive amplitude 8-10, 8-11, 11-29, 11-31, 11-35, 11-36, 11-37
Drive frequency 8-10, 8-18, 8-24, 11-24, 11-33, 13-8
with MFM 13-4, 14-21
E
ECAFM 7-1
Edge Effects 11-37
EFM 12-112-7, 13-9, 14-2
Electric Force Microscopy See EFM 12-1
Electrochemical AFM 7-1
Engage 8-14, 9-11
Engagement
false 15-35
FM 11-34
STM 9-11, 9-12, 9-13, 9-19
TappingMode 8-13
Error Signal Mode 8-19
F
False Engagement 15-35
Feedback Gains
initial settings 6-13
Feedback type 9-5, 9-7
Fluid Cell
preparation 8-278-31
Fluids, imaging in 7-1
FM See Force Modulation
Force Cal
TappingMode in fluids 8-34
Force Cal 6-13, 6-14, 11-111-12
Force Cal
adjustment 11-10
Force Calibration 15-5
Force Calibration 8-218-26, 11-19
Capture 8-23, 11-24
Drive amplitude 8-24, 11-24
Drive Frequency 8-24, 11-24
Sensitivity 8-24, 11-22
Force Curve 11-4
adjustment 11-511-12
Index-2
Index-3
Index-4
S
Samples
biological 7-18-44
minimizing surface forces 11-10
preparation 5-3, 7-2
Scan angle 10-1, 10-2
Scan direction 13-8
Scan rate 2-27, 6-14, 8-17
Scan size 2-26, 6-14, 9-11, 13-7
STM 9-6, 9-12
Scan speed 11-37
Scanner
calibration 15-33
linearity 15-13
parameters 8-188-24, 11-23
piezoelectric crystals 2-7
Scanner Calibration 15-10
Scope Mode 10-5
Sensitivity 6-13, 8-19, 8-24, 11-22
adjustment 8-23
Sensitivity
adjustment 11-9
Setpoint 6-14, 8-11, 8-16, 8-20, 8-24, 11-23, 11-37
defined 6-14
FM 11-34, 11-35, 11-36
Force Modulation 11-29
STM 9-12
Silicon Cantilever Substrates 4-1
Slow scan axis 2-27
Small Sample SPM
adjustment screws 15-43
Spring Constant
specifications 6-12
STM 9-19-30
applications 9-2
hardware 9-3
principles of 9-1
Sweep graph range 8-10, 11-31
Sweep width 8-10, 11-32
T
TappingMode 8-18-26, 11-19
in fluid 8-268-43
principles of 2-322-34
scanning parameters 8-188-24, 11-23
set-up 5-22
Tip Down 8-23, 11-6, 11-21, 11-24
Tip Holder
installation fixture 5-9
preparation 5-6
STM 9-9
Tip Up 8-23, 11-21, 11-24
to 13-7
Index-5
Trace 8-17
Troubleshooting 15-41
adjustment screws 15-43
calibration 15-33
cantilever tune plot badloose cantilever 8-43
contact force too high 15-35
control loop explained 2-26
drift during fluid imaging 15-41
loss of Z center position 15-37
no engagement 15-35
poor imagelines 15-40
poor imagerepeating images 15-41
poor imagerings around features 15-41
poor imageslines 15-36
poor imagestreaks on trailing edges 15-40
poor imagevertical distortion 15-37
poor imagewhiteness 15-41
probe tips 15-36
TappingMode in fluids 8-35
U
Units 11-22, 11-23
V
Van der Waals Forces 12-7
View
Cantilever Tune 11-30
Image Mode 9-11
Scope Mode 9-6, 9-11, 10-3
W
Wet Samples
TappingMode 8-44
X
X offset 2-26, 9-11
Y
Y offset 2-26, 9-11
Z
Z limit 2-27
Z scan rate 8-22, 11-15
Z scan size 8-21, 8-22, 8-23, 8-26, 11-2, 11-18, 11-22
Z scan start 8-21, 8-22, 8-23, 8-26, 11-2, 11-18, 11-22
Zoom In 8-10, 11-32
Index-6