Introduction To Magnetic Particle Inspection
Introduction To Magnetic Particle Inspection
Magnetic particle inspection (MPI) is a nondestructive testing method used for defect
detection. MPI is fast and relatively easy to apply, and part surface preparation is not
as critical as it is for some other NDT methods. These characteristics make MPI one
of the most widely utilized nondestructive testing methods.
MPI uses magnetic fields and small magnetic particles (i.e.iron filings) to detect flaws
in components. The only requirement from an inspectability standpoint is that the
component being inspected must be made of a ferromagnetic material such as iron,
nickel, cobalt, or some of their alloys. Ferromagnetic materials are materials that can
be magnetized to a level that will allow the inspection to be effective.
The method is used to inspect a variety of product forms including castings, forgings,
and weldments. Many different industries use magnetic particle inspection for
determining a component's fitness-for-use. Some examples of industries that use
magnetic particle inspection are the structural steel, automotive, petrochemical, power
generation, and aerospace industries. Underwater inspection is another area where
magnetic particle inspection may be used to test items such as offshore structures and
underwater pipelines.
Basic Principles
In theory, magnetic particle inspection (MPI) is a relatively simple concept. It can be
considered as a combination of two
nondestructive testing methods: magnetic
flux leakage testing and visual testing.
Consider the case of a bar magnet. It has a
magnetic field in and around the magnet. Any
place that a magnetic line of force exits or
enters the magnet is called a pole. A pole
where a magnetic line of force exits the magnet is called a north pole and a pole
where a line of force enters the magnet is called a south pole.
When a bar magnet is broken in the center of its length, two complete bar magnets
with magnetic poles on each end of each piece will result. If the magnet is just
cracked but not broken completely in two, a north and south pole will form at each
edge of the crack. The magnetic field exits the north pole and reenters at the south
pole. The magnetic field spreads out when it
encounters the small air gap created by the
crack because the air cannot support as much
magnetic field per unit volume as the magnet
can. When the field spreads out, it appears to
leak out of the material and, thus is called a
flux leakage field.
If iron particles are sprinkled on a cracked magnet, the particles will be attracted to
and cluster not only at the poles at the ends of the magnet, but also at the poles at the
edges of the crack. This cluster of particles is much easier to see than the actual crack
and this is the basis for magnetic particle inspection.
The first step in a magnetic particle inspection is to magnetize the component that is
to be inspected. If any defects on or near the surface are present, the defects will
create a leakage field. After the component has been magnetized, iron particles, either
in a dry or wet suspended form, are applied to the surface of the magnetized part. The
particles will be attracted and cluster at the flux leakage fields, thus forming a visible
indication that the inspector can detect.
Magnetism
Magnets are very common items in the workplace and household. Uses of magnets
range from holding pictures on the refrigerator to causing torque in electric motors.
Most people are familiar with the general properties of magnets but are less familiar
with the source of magnetism. The traditional concept of magnetism centers around
the magnetic field and what is know as a dipole. The term "magnetic field" simply
describes a volume of space where there is a change in energy within that volume.
This change in energy can be detected and measured. The location where a magnetic
field can be detected exiting or entering a material is called a magnetic pole. Magnetic
poles have never been detected in isolation but always occur in pairs, hence the name
dipole. Therefore, a dipole is an object that has a magnetic pole on one end and a
second, equal but opposite, magnetic pole on the other.
A bar magnet can be considered a dipole with a north pole at one end and south pole
at the other. A magnetic field can be measured leaving the dipole at the north pole and
returning the magnet at the south pole. If a magnet is cut in two, two magnets or
dipoles are created out of one. This sectioning and creation of dipoles can continue to
the atomic level. Therefore, the source of magnetism lies in the basic building block
of all matter...the atom.
The Source of Magnetism
All matter is composed of atoms, and atoms are
composed of protons, neutrons and electrons. The protons
and neutrons are located in the atom's nucleus and the
electrons are in constant motion around the nucleus.
Electrons carry a negative electrical charge and produce a
magnetic field as they move through space. A magnetic
field is produced whenever an electrical charge is in
motion. The strength of this field is called the magnetic
moment.
This may be hard to visualize on a subatomic scale but
consider electric current flowing through a conductor.
When the electrons (electric current) are flowing through the conductor, a magnetic
field forms around the conductor. The magnetic field can be detected using a
compass. The magnetic field will place a force on the compass needle, which is
another example of a dipole.
Since all matter is comprised of atoms, all materials are affected in some way by a
magnetic field. However, not all materials react the same way. This will be explored
more in the next section.
Magnetic Domains
Ferromagnetic materials get their magnetic properties not only because their atoms
carry a magnetic moment but also because the material is made up of small regions
known as magnetic domains. In each domain, all of the atomic dipoles are coupled
together in a preferential direction. This alignment develops as the material develops
its crystalline structure during solidification from the molten state. Magnetic domains
can be detected using Magnetic Force Microscopy (MFM) and images of the domains
like the one shown below can be constructed.
Magnetic Force
Microscopy (MFM)
image showing the
magnetic domains in a
piece of heat treated
carbon steel.
During solidification, a trillion or more atom moments are aligned parallel so that the
magnetic force within the domain is strong in one direction. Ferromagnetic materials
are said to be characterized by "spontaneous magnetization" since they obtain
saturation magnetization in each of the domains without an external magnetic field
being applied. Even though the domains are magnetically saturated, the bulk material
may not show any signs of magnetism because the domains develop themselves and
are randomly oriented relative to each other.
Ferromagnetic materials become magnetized when the magnetic domains within the
material are aligned. This can be done by placing the material in a strong external
magnetic field or by passing electrical current through the material. Some or all of the
domains can become aligned. The more domains that are aligned, the stronger the
magnetic field in the material. When all of the domains are aligned, the material is
said to be magnetically saturated. When a material is magnetically saturated, no
additional amount of external magnetization force will cause an increase in its internal
level of magnetization.
Unmagnetized Material
Magnetized Material
It can be seen in the magnetograph that there are poles all along the length of the
magnet but that the poles are concentrated at the ends of the magnet. The area where
the exit poles are concentrated is called the magnet's north pole and the area where the
entrance poles are concentrated is called the magnet's south pole.
Magnetic Fields in and around Horseshoe and Ring Magnets
Magnets come in a variety of shapes and one of the more
common is the horseshoe (U) magnet. The horseshoe magnet
has north and south poles just like a bar magnet but the magnet
is curved so the poles lie in the same plane. The magnetic lines
of force flow from pole to pole just like in the bar magnet.
However, since the poles are located closer together and a
more direct path exists for the lines of flux to travel between
the poles, the magnetic field is concentrated between the
poles.
flow in loop within a material. (See section on circular magnetism for more
information).
Electromagnetic Fields
Magnets are not the only source of magnetic fields. In
1820, Hans Christian Oersted discovered that an electric
current flowing through a wire caused a nearby compass
to deflect. This indicated that the current in the wire was
generating a magnetic field. Oersted studied the nature of
the magnetic field around the long straight wire. He found
that the magnetic field existed in circular form around the
wire and that the intensity of the field was directly
proportional to the amount of current carried by the wire.
He also found that the strength of the field was strongest
next to the wire and diminished with distance from the
conductor until it could no longer be detected. In most
conductors, the magnetic field exists only as long as the
current is flowing (i.e. an electrical charge is in motion).
However, in ferromagnetic materials the electric current
will cause some or all of the magnetic domains to align
and a residual magnetic field will remain.
Oersted also noticed that
the direction of the
magnetic field was
dependent on the direction
of the electrical current in
the wire. A three-dimensional representation of the
magnetic field is shown below. There is a simple rule
for remembering the direction of the magnetic field
around a conductor. It is called the right-hand rule. If
a person grasps a conductor in one's right hand with
the thumb pointing in the direction of the current, the
fingers will circle the conductor in the direction of the
magnetic field.
A word of caution about the right-hand rule
For the right-hand rule to work, one important thing that must be remembered about
the direction of current flow. Standard convention has current flowing from the
positive terminal to the negative terminal. This convention is credited to Benjamin
Franklin who theorized that electric current was due to a positive charge moving from
the positive terminal to the negative terminal. However, it was later discovered that it
is the movement of the negatively charged electron that is responsible for electrical
current. Rather than changing several centuries of theory and equations, Franklin's
convention is still used today.
The strength of a coil's magnetic field increases not only with increasing current but
also with each loop that is added to the coil. A long, straight coil of wire is called a
solenoid and can be used to generate a nearly uniform magnetic field similar to that of
a bar magnet. The concentrated magnetic field inside a coil is very useful in
magnetizing ferromagnetic materials for inspection using the magnetic particle testing
method. Please be aware that the field outside the coil is weak and is not suitable for
magnetizing ferromagnetic materials.
SI Units
(Somerfield)
A/m
H
Field
Flux Density
(Magnetic Induction)
Flux
Magnetization
SI Units
(Kennelly)
A/m
CGS Units
(Gaussian)
oersteds
tesla
tesla
gauss
Weber
Weber
Maxwell
A/m
erg/Oe-cm3
The units for magnetic field strength H are ampere/meter. A magnetic field strength
of 1 ampere/meter is produced at the center of a single circular conductor with a one
meter diameter carrying a steady current of 1 ampere.
The number of magnetic lines of force cutting through a plane of a given area at a
right angle is known as the magnetic flux density, B. The flux density or magnetic
induction has the tesla as its unit. One tesla is equal to 1 Newton/(A/m). From these
units, it can be seen that the flux density is a measure of the force applied to a particle
by the magnetic field. The Gauss is the CGS unit for flux density and is commonly
used by US industry. One gauss represents one line of flux passing through one
square centimeter of air oriented 90 degrees to the flux flow.
The total number of lines of magnetic force in a material is called magnetic flux, .
The strength of the flux is determined by the number of magnetic domains that are
aligned within a material. The total flux is simply the flux density applied over an
area. Flux carries the unit of a weber, which is simply a tesla- meter2.
The magnetization is a measure of the extent to which an object is magnetized. It is a
measure of the magnetic dipole moment per unit volume of the object. Magnetization
carries the same units as a magnetic field: amperes/meter.
reached the point of magnetic saturation. When H is reduced to zero, the curve will
move from point "a" to point "b." At this point, it can be seen that some magnetic flux
remains in the material even though the magnetizing force is zero. This is referred to
as the point of retentivity on the graph and indicates the remanence or level of
residual magnetism in the material. (Some of the magnetic domains remain aligned
but some have lost their alignment.) As the magnetizing force is reversed, the curve
moves to point "c", where the flux has been reduced to zero. This is called the point of
coercivity on the curve. (The reversed magnetizing force has flipped enough of the
domains so that the net flux within the material is zero.) The force required to remove
the residual magnetism from the material is called the coercive force or coercivity of
the material.
As the magnetizing force is increased in the negative direction, the material will again
become magnetically saturated but in the opposite direction (point "d"). Reducing H
to zero brings the curve to point "e." It will have a level of residual magnetism equal
to that achieved in the other direction. Increasing H back in the positive direction will
return B to zero. Notice that the curve did not return to the origin of the graph because
some force is required to remove the residual magnetism. The curve will take a
different path from point "f" back to the saturation point where it with complete the
loop.
From the hysteresis loop, a number of primary magnetic properties of a material can
be determined.
1. Retentivity - A measure of the residual flux density corresponding to the
saturation induction of a magnetic material. In other words, it is a material's
ability to retain a certain amount of residual magnetic field when the
magnetizing force is removed after achieving saturation. (The value of B at
point b on the hysteresis curve.)
2. Residual Magnetism or Residual Flux - the magnetic flux density that
remains in a material when the magnetizing force is zero. Note that residual
magnetism and retentivity are the same when the material has been
magnetized to the saturation point. However, the level of residual magnetism
may be lower than the retentivity value when the magnetizing force did not
reach the saturation level.
3. Coercive Force - The amount of reverse magnetic field which must be applied
to a magnetic material to make the magnetic flux return to zero. (The value of
H at point c on the hysteresis curve.)
4. Permeability, - A property of a material that describes the ease with which
a magnetic flux is established in the component.
5. Reluctance - Is the opposition that a ferromagnetic material shows to the
establishment of a magnetic field. Reluctance is analogous to the resistance in
an electrical circuit.
Permeability
As previously mentioned, permeability is a material property that describes the ease
with which a magnetic flux is established in a component. It is the ratio of the flux
density to the magnetizing force and is represented by the following equation:
= /
It is clear that this equation describes
the slope of the curve at any point on
the hysteresis loop. The permeability
value given in papers and reference
materials is usually the maximum
permeability or the maximum
relative permeability. The maximum
permeability is the point where the
slope of the B/H curve for the
unmagnetized material is the
greatest. This point is often taken as
the point where a straight line from
the origin is tangent to the B/H curve.
The relative permeability is arrived at by taking the ratio of the material's permeability
to the permeability in free space (air).
The shape of the hysteresis loop tells a great deal about the material being
magnetized. The hysteresis curves of two different materials are shown in the graph.
Relative to other materials, a material with a wider
hysteresis loop has:
Lower Permeability
Higher Retentivity
Higher Coercivity
Higher Reluctance
Higher Residual Magnetism
Higher Permeability
Lower Retentivity
Lower Coercivity
Lower Reluctance
Lower Residual Magnetism.
Watch this short movie showing the effect of field direction on indication visibility.
(775 KB mov)
A second technique involves using clamps or prods, which are attached or placed in
contact with the component. Electrical current flows through the component from
contact to contact. The current sets up a circular magnetic field around the path of the
current.
Magnetizing Current
As seen in the previous pages, electric current is often used to establish the magnetic
field in components during magnetic particle inspection. Alternating current and
direct current are the two basic types of current commonly used. Current from single
phase 110 volts, to three phase 440 volts, are used when generating an electric field in
a component. Current flow is often modified to provide the appropriate field within
the part. The type of current used can have an effect on the inspection results, so the
types of currents commonly used will be briefly reviewed.
Direct Current
Direct current (DC) flows continuously in one direction at a constant voltage. A
battery is the most common source of direct current. As previously mentioned, current
is said to flow from the positive to the negative terminal. In actuality, the electrons
flow in the opposite direction. DC is very desirable when inspecting for subsurface
defects because DC generates a magnetic field that penetrates deeper into the
material. In ferromagnetic materials, the magnetic field produced by DC generally
penetrates the entire cross-section of the component. Conversely, the field produced
using alternating current is concentrated in a thin layer at the surface of the
component.
Alternating Current
Alternating current (AC) reverses in direction at a rate of 50 or 60 cycles per second.
In the United States, 60 cycle current is the commercial norm but 50 cycle current is
common in many countries. Since AC is readily available in most facilities, it is
convenient to make use of it for magnetic particle inspection. However, when AC is
used to induce a magnetic field in ferromagnetic materials, the magnetic field will be
limited to narrow region at the surface of the component. This phenomenon is known
as the "skin effect" and occurs because induction is not a spontaneous reaction and the
rapidly reversing current does not allow the domains below the surface time to align.
Therefore, it is recommended that AC be used only when the inspection is limited to
surface defects.
View a short video on the AC versus DC. (195 KB mov)
Rectified Alternating Current
Clearly, the skin effect limits the use of AC since many inspection applications call
for the detection of subsurface defects. However, the convenient access to AC, drives
its use beyond surface flaw inspections. Luckily, AC can be converted to current that
is very much like DC through the process of rectification. With the use of rectifiers,
the reversing AC can be converted to a one directional current. The three commonly
used types of rectified current are described below.
Stationary magnetic particle equipment wired with three phase AC will usually have
the ability to magnetize with AC or DC (three phase full wave rectified), providing
the inspector with the advantages of each current form.
The field strength varies from zero at the center of the component to a
maximum at the surface.
The field strength at the surface of the conductor decreases as the radius of the
conductor increases when the current strength is held constant. (However, a
larger conductor is capable of carrying more current.)
The field strength outside the conductor is directly proportional to the current
strength. Inside the conductor, the field strength is dependent on the current
strength, magnetic permeability of the material, and if magnetic, the location
on the B-H curve.
The field strength outside the conductor decreases with distance from the
conductor.
In the images below, the magnetic field strength is graphed versus distance from the
center of the conductor. It can be seen that in a nonmagnetic conductor carrying DC,
the internal field strength rises from zero at the center to a maximum value at the
surface of the conductor. The external field strength decrease with distance from the
surface of the conductor. When the conductor is a magnetic material, the field
strength within the conductor is much greater than it was in the nonmagnetic
conductor. This is due to the permeability of the magnetic material. The external field
is exactly the same for the two materials provided the current level and conductor
radius are the same.
In a hollow circular conductor there is no magnetic field in the void area. The
magnetic field is zero at the inside wall surface and rises until it reaches a maximum
at the outside wall surface. As with a solid conductor, when the conductor is a
magnetic material, the field strength within the conductor is much greater than it was
in the nonmagnetic conductor due to the permeability of the magnetic material. The
external field strength decreases with distance from the surface of the conductor. The
external field is exactly the same for the two materials provided the current level and
conductor radius are the same.
Demagnetization
After conducting a magnetic particle inspection, it is usually necessary to demagnetize
the component. Remanent magnetic fields can:
producing a measurable voltage between the two sides of the conductor. The presence
of this measurable transverse voltage is called the Hall-effect after Edwin H. Hall,
who discovered it in 1879.
used. Permanent magnets can also be made small enough to fit into tight areas where
electromagnets might not fit.
Electromagnets
Today, most of the equipment used to
create the magnetic field used in MPI
is based on electromagnetism. That
is, using an electrical current to
produce the magnetic field. An
electromagnetic yoke is a very
common piece of equipment that is
used to establish a magnetic field. It
is basically made by wrapping an
electrical coil around a piece of soft
ferromagnetic steel. A switch is
included in the electrical circuit so
that the current and, therefore, the
magnetic field can be turned on and off. They can be powered with alternating current
from a wall socket or by direct current from a battery pack. This type of magnet
generates a very strong magnetic field in a local area where the poles of the magnet
touch the part being inspected. Some yokes can lift weights in excess of 40 pounds.
prods are connected by any insulator (as shown in the image) to facilitate one hand
operation. This is referred to as a dual prod and is commonly used for weld
inspections.
If proper contact is not maintained between the prods and the component surface,
electrical arcing can occur and cause damage to the component. For this reason, the
use of prods are not allowed when inspecting aerospace and other critical
components. To help prevent arcing, the prod tips should be inspected frequently to
ensure that they are not oxidized, covered with scale or other contaminant, or
damaged.
The following applet shows two prods used to create a current through a conducting
part. The resultant magnetic field roughly depicts the patterns expected from an
magnetic particle inspection of an unflawed surface. The user is encouraged to
manipulate the prods to orient the magnetic field to "cut across" suspected defects.
Portable Coil
Conductive Cable
Watch these short movies showing a cable being used to establish magnetic fields in
parts. Cable wrapped around part (690 KB mov). Cable wrapped through the part
(425 KB mov).
Portable Power Supplies
Portable power supplies are used to provide the necessary electricity to the prods,
coils or cables. Power supplies are commercially available in a variety of sizes. Small
power supplies generally provide up to 1,500A of half-wave direct current or
alternating current when used with a 4.5 meter 0000 cable. They are small and light
enough to be carried and operate on either 120V or 240V electrical service. When
more power is necessary, mobile power supplies can be used. These units come with
wheels so that they can be rolled where needed. These units also operate on 120V or
240V electrical service and can provide up to 6,000A of AC or half-wave DC when 9
meters or less of 0000 cable is used.
The features of a portable power supply are explained in this short movie (800KB
mov).
Lights for
Magnetic Particle Inspection
Magnetic particle inspection can be
performed using particles that are
highly visible under white light
conditions or particles that are highly
visible under ultraviolet light
conditions. When an inspection is
being performed using the visible color
contrast particles, no special lighting is
required as long as the area of
inspection is well lit. A light intensity
of at least 1000 lux (100 fc) is
recommended when visible particles
are used, but a variety of light sources
can be used.
When fluorescent particles are used,
special ultraviolet light must be used. Fluorescence is defined as the property of
emitting radiation as a result of and during exposure to radiation. Particles used in
fluorescent magnetic particle inspections are coated with a material that produces
light in the visible spectrum when exposed to near-ultraviolet light. This "particle
glow" provides high contrast indications on the component anywhere particles collect.
Particles that fluoresce yellow-green are most common because this color matches the
peak sensitivity of the human eye under dark conditions. However, particles that
fluoresce red, blue, yellow, and green colors are available.
Ultraviolet Light
Ultraviolet light or "black light" is light in the 1,000 to 4,000 Angstroms (100 to
400nm) wavelength range in the electromagnetic spectrum. It is a very energetic form
of light that is invisible to the human eye. Wavelengths above 4,000A fall into the
visible light spectrum and are seen as the color violet. UV is separated according to
wavelength into three classes: A, B, and C. The shorter the wavelength, the more
energy that is carried in the light and the more dangerous it is to the human cells.
Class
UV-A
UV-B
UV-C
Wavelength Range
3,2004,000 Angstroms
2,8003,200 Angstroms
2,8001,000 Angstroms
The desired wavelength range for use in nondestructive testing is between 3,500 and
3,800A with a peak wavelength at about 3,650A. This wavelength range is used
because it is in the UV-A range, which is the safest to work with. UV-B will do an
effective job of causing substances to fluoresce, however, it should not be used
because harmful effects such as skin burns and eye damage can occur. This
wavelength of radiation is found in the arc created during the welding process. UV-C
(1,000 to 2,800A) is even more dangerous to living cells and is used to kill bacteria in
industrial and medical settings.
The desired wavelength range for use in NDT is obtained by filtering the ultraviolet
light generated by the light bulb. The output of a UV bulb spans a wide range of
wavelengths. The short wavelengths of 3,120 to 3,340A are produced in low levels. A
peak wavelength of 3650A is produced at a very high intensity. Wavelengths in the
visible violet range (4050A to 4350A), green-yellow (5460A), yellow (6220A) and
orange (6770A) are also usually produced. The filter allows only radiation in the
range of 3200 to 4000A and a little visible dark purple to pass.
Basic Ultraviolet Lights
UV bulbs come in a variety of shapes and sizes. The more common types are the low
pressure tube, high pressure spot, the high pressure flood types. The tubular black
light is similar in construction to the tubular fluorescent lights used for office or home
illumination. These lights use a low pressure mercury vapor arc. Tube lengths of 6 to
48 inches are common. The low pressure bulbs are most often used to provide general
illumination to large areas rather than for illumination of components to be inspected.
These bulbs generate a relatively large amount of white light, which is concerning
since inspection specifications require less than two foot-candles of white light at the
inspection surface.
Flood lights are also used to illuminate the
inspection area, since they provide even
illumination over a large area. Intensity levels
for flood lamps are relatively low because the
energy is spread over a large area. They
generally do not generate the required UV
light intensity at the given distance that
specifications require.
Spot lights, on the other hand, provide
concentrated energy that can be directed to
the area of inspection. A spot light will
generate a six inch diameter circle of high
intensity light when held fifteen inches from the inspection surface. One hundred watt
mercury vapor lights are most commonly used, but higher wattages are available.
In the high pressure mercury vapor spot or flood lamps, UV light is generated by a
quartz tube inside the bulb. This tube contains two electrodes that establish an arc.
The distance between electrodes is such that a starting electrode must be used. A
resister limits the current to the starting electrode that establishes the initial arc that
vaporizes the mercury in the tube. Once this low level arc is established and the
mercury is vaporized, the arc between the main electrodes is established. It takes
approximately five minutes to "warm up" and establish the arc between the main
electrodes. This is why specifications require a "warm up time" before using the high
pressure mercury vapor lights. Flood and spot black lights produce large amounts of
heat and should be handled with caution to prevent burns. This condition has been
eliminated by newer designs that include cooling fans. The arc in the bulb can be
upset when exposed to an external magnetic field, such as that generated by a coil.
Care should be taken not to bring the lamp close to strong magnetic fields, but if the
arc is upset and extinguished, it must be allowed to cool before it can be safely
restarted.
The advantages of Hall Effect devices are: they provide a quantitative measure of the
strength of magnetizing force tangential to the surface of a test piece, they can be used
for measurement of residual magnetic fields, and they can be used repetitively. Their
main disadvantages are that they must be periodically calibrated and they cannot be
used to establish the balance of fields in multidirectional applications.
Quantitative Quality Indicator (QQI)
The Quantitative Quality Indicator (QQI) or Artificial Flaw Standard is often the
preferred method of assuring proper field direction and adequate field strength. The
use of a QQI is also the only practical way of ensuring balanced field intensity and
direction in multiple-direction magnetization equipment. QQIs are often used in
conjunction with a Gauss meter to establish the inspection procedure for a particular
component. They are used with the wet method only, and like other flux sharing
devices, can only be used with continuous magnetization.
The QQI is a thin strip of either 0.002 or 0.004 inch thick AISI 1005 steel. A photo
etch process is used to inscribe a specific pattern, such as concentric circles or a plus
sign. QQIs are nominally 3/4 inch square, but miniature shims are also available.
QQIs must be in intimate contact with the part being evaluated. This is accomplished
by placing the shim on a part etched side down, and taping or gluing it to the surface.
The component is then magnetized and particles applied. When the field strength is
adequate, the particles will adhere over the engraved pattern and provide information
about the field direction. When a multidirectional technique is used, a balance of the
fields is noted when all areas of the QQI produce indications.
Some of the advantages of QQIs are: they can be quantified and related to other
parameters, they can accommodate virtually any configuration with suitable selection,
and they can be reused with careful application and removal practices. Some of the
disadvantages are: the application process is somewhat slow, the parts must be clean
and dry, shims cannot be used as a residual magnetism indicator as they are a flux
sharing device, they can be easily damaged with improper handling, and they will
corrode if not cleaned and properly stored.
Above left is a photo of a typical QQI shim. The photo on the right shows the
indication produced by the QQI when it is applied to the surface a part and a magnetic
field is established that runs across the shim from right to left.
Pie Gage
The pie gage is a disk of highly permeable material divided into four, six, or eight
sections by nonferromagnetic material. The divisions serve as artificial defects that
radiate out in different directions from the center. The diameter of the gage is 3/4 to 1
inch. The divisions between the low carbon steel pie sections are to be no greater than
1/32 inch. The sections are furnace brazed and copper plated. The gage is placed on
the test piece copper side up and the test piece is magnetized. After particles are
applied and the excess removed, the indications provide the inspector the orientation
of the magnetic field.
The principal application is on flat surfaces such as weldments or steel castings where
dry powder is used with a yoke or prods. The pie gage is not recommended for
precision parts with complex shapes, for wet-method applications, or for proving field
magnitude. The gage should be demagnetized between readings.
Several of the main advantages of the pie gage are that it is easy to use and it can be
used indefinitely without deterioration. The pie gage has several disadvantages, which
include: it retains some residual magnetism so indications will prevail after removal
of the source of magnetization, it can only be used in relatively flat areas, and it
cannot be reliably used for determination of balanced fields in multidirectional
magnetization.
Watch this short movie to see a Pie field gage in action (600KB mov).
Slotted Strips
Slotted strips, also known as Burmah-Castrol Strips, are pieces of highly permeable
ferromagnetic material with slots of different widths. They are placed on the test
object as it is inspected. The indications produced on the strips give the inspector a
general idea of the field strength in a particular area.
Advantages of these strips are: they are relatively easily applied to the component,
they can be used successfully with either the wet or dry method when using the
continuous magnetization, they are repeatable as long as orientation to the magnetic
field is maintained, and they can be used repetitively. Some of the disadvantages are
that they cannot be bent to complex configuration and they are not suitable for
multidirectional field applications since they indicate defects in only one direction.
Magnetic Particles
As mentioned previously, the particles that are used for magnetic particle inspection
are a key ingredient as they form the indications that alert the inspector to defects.
Particles start out as tiny milled (a machining process) pieces of iron or iron oxide. A
pigment (somewhat like paint) is bonded to their surfaces to give the particles color.
The metal used for the particles has high magnetic permeability and low retentivity.
High magnetic permeability is important because it makes the particles attract easily
to small magnetic leakage fields from discontinuities, such as flaws. Low retentivity is
important because the particles themselves never become strongly magnetized so they
do not stick to each other or the surface of the part. Particles are available in a dry mix
or a wet solution.
Dry Magnetic Particles
Dry magnetic particles can typically be
purchased in red, black, gray, yellow and
several other colors so that a high level of
contrast between the particles and the part
being inspected can be achieved. The size of
the magnetic particles is also very important.
Dry magnetic particle products are produced to
include a range of particle sizes. The fine
particles are around 50 m (0.002 inch) in size,
and are about three times smaller in diameter
and more than 20 times lighter than the coarse
particles (150 m or 0.006 inch). This make them more sensitive to the leakage fields
from very small discontinuities. However, dry testing particles cannot be made
exclusively of the fine particles. Coarser particles are needed to bridge large
discontinuities and to reduce the powder's dusty nature. Additionally, small particles
easily adhere to surface contamination, such as remnant dirt or moisture, and get
trapped in surface roughness features. It should also be recognized that finer particles
will be more easily blown away by the wind; therefore, windy conditions can reduce
the sensitivity of an inspection. Also, reclaiming the dry particles is not recommended
because the small particles are less likely to be recaptured and the "once used" mix
will result in less sensitive inspections.
The particle shape is also important. Long,
slender particles tend align themselves along
the lines of magnetic force. However, research
has shown that if dry powder consists only of
long, slender particles, the application process
would be less than desirable. Elongated
particles come from the dispenser in clumps
and lack the ability to flow freely and form the
desired "cloud" of particles floating on the
component. Therefore, globular particles are
added that are shorter. The mix of globular and
elongated particles result in a dry powder that flows well and maintains good
sensitivity. Most dry particle mixes have particles with L/D ratios between one and
two.
The particles used with the wet method are smaller in size than those used in the dry
method for the reasons mentioned above. The particles are typically 10 m (0.0004
inch) and smaller and the synthetic iron oxides have particle diameters around 0.1 m
(0.000004 inch). This very small size is a result of the process used to form the
particles and is not particularly desirable, as the particles are almost too fine to settle
out of suspension. However, due to their slight residual magnetism, the oxide particles
are present mostly in clusters that settle out of suspension much faster than the
individual particles. This makes it possible to see and measure the concentration of
the particles for process control purposes. Wet particles are also a mix of long slender
and globular particles.
The carrier solutions can be water or oil-based. Water-based carriers form quicker
indications, are generally less expensive, present little or no fire hazard, give off no
petrochemical fumes, and are easier to clean from the part. Water-based solutions are
usually formulated with a corrosion inhibitor to offer some corrosion protection.
However, oil-based carrier solutions offer superior corrosion and hydrogen
embrittlement protection to those materials that are prone to attack by these
mechanisms.
Suspension Liquids
Suspension liquids used in the wet
magnetic particle inspection method can
be either a well refined light petroleum
distillate or water containing additives.
Petroleum-based liquids are the most
desirable carriers because they provided
good wetting of the surface of metallic
parts. However, water-based carriers are
used more because of low cost, low fire
hazard, and the ability to form indications
quicker than solvent-based carriers.
Water-based carriers must contain wetting
agents to disrupt surface films of oil that
may exist on the part and to aid in the dispersion of magnetic particles in the carrier.
The wetting agents create foaming as the solution is moved about, so anti-foaming
agents must be added. Also, since water promotes corrosion in ferrous materials,
corrosion inhibitors are usually added as well.
Petroleum based carriers are primarily used in systems where maintaining the proper
particle concentration is a concern. The petroleum based carriers require less
maintenance because they evaporate at a slower rate than the water-based carriers.
Therefore, petroleum based carriers might be a better choice for a system that gets
only occasional use or when regularly adjusting the carrier volume is undesirable.
Modern solvent carriers are specifically designed with properties that have flash
points above 200oF and keep nocuous vapors low. Petroleum carriers are required to
meet certain specifications such as AMS 2641.
dry particles is commonly used when inspecting large castings for hot tears and
cracks.
Steps in performing an inspection using dry particles
Prepare the part surface - the surface should be relatively clean but this is not as
critical as it is with liquid penetrant inspection. The surface must be free of grease, oil
or other moisture that could keep particles from moving freely. A thin layer of paint,
rust or scale will reduce test sensitivity but can sometimes be left in place with
adequate results. Specifications often allow up to 0.003 inch (0.076 mm) of a
nonconductive coating (such as paint) and 0.001 inch max (0.025 mm) of a
ferromagnetic coating (such as nickel) to be left on the surface. Any loose dirt, paint,
rust or scale must be removed.
Apply the magnetizing force - Use permanent magnets, an electromagnetic yoke,
prods, a coil or other means to establish the necessary magnetic flux.
Dust on the dry magnetic particles - Dust on a light layer of magnetic particles.
Gently blow off the excess powder - With the magnetizing force still applied,
remove the excess powder from the surface with a few gentle puffs of dry air. The
force of the air needs to be strong enough to remove the excess particles but not
strong enough to dislodge particles held by a magnetic flux leakage field.
Terminate the magnetizing force - If the magnetic flux is being generated with an
electromagnet or an electromagnetic field, the magnetizing force should be
terminated. If permanent magnets are being used, they can be left in place.
Inspect for indications - Look for areas where the magnetic particles are clustered.
Field Strength
The applied magnetic field must have sufficient strength to produce a satisfactory
indication, but not so strong that it produces nonrelevant indications or limits particle
mobility. If the magnetizing current is excessively high when performing a wet
fluorescent particle inspection, particles can be attracted to the surface of the part and
not be allowed to migrate to the flux leakage fields of defects. When performing a dry
particle inspection, an excessive longitudinal magnetic field will cause furring.
Furring is when magnetic particles build up at the magnetic poles of a part. When the
field strength is excessive, the magnetic field is forced out of the part before reaching
the end of the component and the poles along its length attract particles and cause
high background levels. Adequate field strength may be determined by:
Formulas for calculating current levels should only be used to estimate current
requirements. The magnetic field strength resulting from calculations should be
assessed for adequacy using one of the two method discussed above. Likewise,
published current level information should also be used only as a guide unless the
values have been established for the specific component and target defects of the
inspection at hand.
Using a Pie Gage
A pie gage is placed copper side up and held
in contact with the component as the
magnetic field and particles are applied.
Indications of the leakage fields provide a
visual representation of defect direction
within the component. Pie gages work well
on flat surfaces, but if the surface is concave
or convex, inaccurate readings may occur.
The pie gage is a flux sharing device and
requires good contact to provide accurate
readings.
Using Quantitative Quality Indicator (QQI) Shims
Quantitative Quality Indicator (QQI) flaw shims are used to establish proper field
direction and to ensure adequate field strength during technique development. The
QQI flaw shim is the most efficient means of determining balance and effectiveness
of fields. The QQI's are also flux sharing devices and must be properly attach so as
not to allow particles to become trapped under the artificial flaw. Application using
Super glue is the preferred way of attaching the artificial flaw, but does not allow for
reuse of the shims. Shims can also be attached with tape applied to just the edge of the
shim. It is recommended that the tape be impervious to oil, not be fluorescent, and be
1/4 to 1/2 inch in width.
The QQI must be applied to locations on the component where the flux density may
vary. One example would be the center area of a yoke or Y shaped component.
Oftentimes, the flux density will be near zero in this area. If two legs of a Y are in
contact with the pad in circular magnetization, it must be determined if current is
flowing evenly through each leg. A QQI on each leg would be appropriate under such
conditions.
QQI's can be used to establish system threshold values for a defect of a given size. By
attaching a QQI shim with three circles (40%, 30% and 20% of shim thickness) to the
component, threshold values for a specific area of the component, can be established.
Begin by applying current at a low amperage and slowly increasing it until the largest
flaw is obtained. The flux density should be verified and recorded using a Hall effects
probe. The current is then increased until the second circle is identified and the flux
density is again recorded. As the current is further increased, the third ring is
identified and the current values are recorded.
Hall Effects Gauss Meter
There are several types of Hall effects probes
that can be used to measure the magnetic
field strength. Transverse probes are the type
most commonly used to evaluate the field
strength in magnetic particle testing.
Transverse probes have the Hall effect
element mounted in a thin, flat stem and they
are used to make measurements between two
magnetic poles. Axial probes have the
sensing element mounted such that the
magnetic flux in the direction of the long
axis of the probe is measured.
To make a measurement with a transverse probe, the probe is positioned such that the
flat surface of the Hall Effect element is transverse to the magnetic lines of flux. The
Hall Effect voltage is a function of the angle at which the magnetic lines of flux pass
through the sensing element. The greatest Hall Effect voltage occurs when the lines of
flux pass perpendicularly through the sensing element. If not perpendicular, the output
voltage is related to the cosine of the difference between 90 degrees and the actual
angle. The peak field strength should be measured when the magnetizing force is
applied. The field strength should be measured in all areas of the component to be
inspected.
Suspension Contamination
The suspension solution should also be examined for
evidence of contamination. Contamination primarily comes
from inspected components. Oils, greases, sand, and dirt will
be introduced to the system through components. If the area is
unusually dusty, the system will pickup dust or other
contaminates from the environment.
This examination is performed on the carrier and particles
collected for concentration testing. The graduated portion of
the tube is viewed under ultraviolet and white light when
fluorescent particles are being used, and under white light
when visible particles are being used. The magnetic particles should be examined for
foreign particles, such as dirt, paint chips and other solids. Differences in color,
layering or banding within the settled particles would indicate contamination. Some
contamination is to be expected but if the foreign matter exceeds 30 percent of the
settled solids, the solution should be replaced.
The liquid carrier portion of the solution should also be inspected for contamination.
Oil in a water bath and water in a solvent bath are the primary concerns. If the
solution fluoresces brightly when fluorescent particles are being used, this can be an
indication that dye is being dislodged from the particles by the mixing pump. While
not technically contamination, this condition should be further evaluated by allowing
the collected sample bath to set for 10 to 12 hours and viewed under ultraviolet light.
If a band that fluoresces brighter than the bulk of particles is evident on top of the
settled solids, the bath contains excessive unattached fluorescent pigments and should
be discarded.
Water Break Test
A daily water break check is required to evaluate the surface wetting performance of
water-based carriers. The water break check simply involves flooding a clean surface
similar to those being inspected and observing the surface film. If a continuous film
forms over the entire surface, sufficient wetting agent is present. If the film of
suspension breaks (water break) exposing the surface of the component, insufficient
wetting agent is present and the solution should be adjusted or replaced.
Lighting
Magnetic particle inspection predominately relies on visual inspection to detect any
indications that form. Therefore, lighting is a very important element of the inspection
process. Obviously, the lighting requirements are different for an inspection
conducted using visible particles than they are for an inspection conducted using
fluorescent particles. The lighting requirements for each of these techniques, as well
as how light measurements are made, is discussed below.
Light Requirements When Using Visible Particles
Magnetic particle inspections that use visible particles can be conducted using natural
or artificial lighting. When using natural lighting, it is important to keep in mind that
daylight varies from hour to hour. Inspector must constantly stay aware of the lighting
conditions and make adjustments when needed. To improve the uniformity of lighting
from one inspection to the next, the use of artificial lighting is recommended.
Artificial lighting should be white whenever possible and white flood or halogen
lamps are most commonly used. The light intensity is required to be 100 foot-candles
at the surface being inspected. It is advisable to choose a white light wattage that will
provide sufficient light, but avoid excessive reflected light that could distract from the
inspection.
Light Requirements When Using Fluorescent Particles
Ultraviolet Lighting
When performing a magnetic particle inspection using fluorescent particles, the
condition of the ultraviolet light and the ambient white light must be monitored.
Standards and procedures require verification of lens condition and light intensity.
Black lights should never be used with a cracked filter as the output of white light and
harmful black light will be increased. The cleanliness of the filter should also be
checked regularly since a coating of solvent carrier, oil, or other foreign material can
reduce the intensity or light by as much as 50%. The filter should be checked visually
and cleaned as necessary before warming-up the light.
For UV lights used in component evaluations, the normally accepted intensity is
1000mwatts/cm2 when measured at 15 inches from the filter face (requirements can
vary from 800 to 1200mwatts/cm2). The required check should be performed when a
new bulb is installed, at startup of the inspection cycle, if a change in intensity is
noticed, or every eight hours if in continuous use. Regularly checking the intensity of
UV lights is very important because bulbs lose intensity over time. In fact, a bulb that
is near the end of its operating life will often have an intensity of only 25% of its
original output. Black light intensity will also be affected by voltage variations, so it is
important to provide constant voltage to the light. A bulb that produces acceptable
intensity at 120 volts will produce significantly less at 110 volts.
Ambient White Lighting
When performing a fluorescent magnetic particle inspection, it is important to keep
white light to a minimum as it will significantly reduce the inspectors ability to detect
fluorescent indications. Light levels of less than 2fc are required by most procedures
and some procedures require it to be less than 0.5fc at the inspection surface. Some
Eye Consideration
Eye Adaptation
Just as lighting is an important consideration
in the inspection process, so is the eye's
response to light. Scientists have recently
discovered that a special, tiny group of cells
at the back of the eye help tell the brain how
much light there is, causing the pupil to get
bigger or smaller. The change in pupil
diameter is not instantaneous, therefore,
eyes must be given time to adapt to
changing lighting conditions. When performing a fluorescent magnetic particle
inspection, the eye must be given time to adapt to the darkness of the inspection booth
before beginning to look for indications. Dark adaptation time of at least one minute
is required by most procedures. Some studies recommend adaptation time of five
minutes if entering an inspection area from direct sunlight. Inspectors should carefully
adhere to the required adaptation time as it is quite easy to overlook an indication
when an inspection is started before the eyes have adjusted to the darkened
conditions.
Eyeball Fluorescence
When ultraviolet light enters the human eye, the fluid that fills the eye fluoresces.
This condition is called eyeball fluorescence, and while it is considered harmless, it is
annoying and interferes with vision while it exists. When working around ultraviolet
lights, one should be careful not to look directly into lights and to hold spot lights to
avoid reflection. UV light will be reflected from surfaces just as white light will, so it
is advisable to consider placement of lights to avoid this condition. Special filtered
glasses may be worn by the inspector to remove all UV light from reaching the eyes
but allowing yellow-green light from fluorescent indications to pass. Technicians
should never wear darkened or photochromatic glasses as these glasses also filter or
block light from fluorescent indications.
Ultrasonic:
The study and use of pressure waves that have a frequency in excess of 20,000
Hz and therefore inaudible to the human ear.
One of the advantages that a magnetic particle inspection has over some of the other
nondestructive evaluation methods is that flaw indications generally resemble the
actual flaw. This is not the case with NDT methods such as ultrasonic and eddy
current inspection, where an electronic signal must be interpreted. When magnetic
particle inspection is used, cracks on the surface of the part appear as sharp lines that
follow the path of the crack. Flaws that exist below the surface of the part are less
defined and more difficult to detect. Below are some examples of magnetic particle
indications produced using dry particles.
~ First name ~
~ Last name ~
10
11
12
13
14
15
16
17
18
Dipole
Inductor
Capacitor
A ferromagnetic material
19
20
North pole
Dipole
Flux density
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
10
11
12
13
14
15
16
17
18
19
20
21
22
23
Permeability
24
25
26
27
28
29
30
31
32
33
poles
All of the above
34
35
36
37
38
39
40
41
42
43
44
45
46
47
Permanent Magnets
48
49
50