FEI Tecnai f20 Operations Manual
FEI Tecnai f20 Operations Manual
Nanofabrication
This manual is to be used as a guide to the basic operation of the Tecnai F-20 microscope. It
should compliment the certification course offered by the Center for Electron Microscopy and
Nanofabrication (CEMN). Additional resources should be consulted for more advanced operation
procedures.
Aaron
I.
Besser
and
Zhiqiang
(Tony)
Chen
[email protected]
Portland
State
University
Center
for
Electron
Microscopy
and
Nanofabrication
Contents
1
Microscope Operation
1.1 FEI Tecnai F-20 TEM Anatomy
1.1.1 External Controls
1.1.2 Tecnai Optics
1.1.3 Microscope Features
1.2 Initial Conditions
1.2.1 Facility Online Management (FOM)
1.2.2 Software
1.2.3 Log Sheet & Microscope Status
1.2.4 Cold finger
1.3 Sample Preparation and Insertion
1.3.1 Single Tilt
1.3.2 Double Tilt
1.3.3 Holder Insertion
1.3.4 Microscope operation
1.4 Shutdown Procedure
1.4.1 Holder Extraction
1.4.2 Cryo Cycle
1.4.3 Software Shutdown
TEM Alignment
2.1 Condenser Apertures
2.2 Condenser Astigmatism
2.3 Gun Tilt
2.4 Gun Shift
2.5 Beam Pivot Points
2.6 Beam Shift
2.7 Eucentric Height
2.8 Rotation Center
2.9 Objective Astigmatism
2.10 Alignments from a File
TEM Analysis
3.1 Objective Focus
3.2 Image Acquisition
3.3 Selected Area Electron Diffraction (SAED)
3.3.1 Diffraction Astigmatism
3.4 Zone Axis
3.5 Bright-Field (BF)
3.6 Dark-Field (DF)
3.7 High Resolution Imaging (HRTEM)
STEM Alignments
4.1 Rotation Center
4.2 Pivot Points
4.3 Condenser Aperture & Condenser Astigmatism
4.4 Detector Alignment
4.4.1 Bright-Field
4.4.2 Dark-Field
STEM Analysis
5.1 Image Acquisition
5.2 Energy Dispersive Spectroscopy
5.2.1 Spectrum Acquisition
5.2.2 Line & Map Scans
5.2.3 EDS Quantification
Gatan Image Filter
6.1 Introduction
6.2 Alignment
6.2.1 Tune
6.2.2 Spectrum Alignment
6.3 Electron Energy Loss Spectrum
6.3.1 Imaging Mode
6.3.2 Diffraction Mode
6.4
6.5
7
Appendix
7.1
7.2
7.3
7.4
7.5
7.6
7.7
Chapter 1
Microscope Operation
1.1 FEI Tecnai F-20 Anatomy
1.1.1 External Controls
Filter Control.
Gatan DigitalMicrograph.
Tecnai imaging and analysis.
Microscope control.
Procedure:
1. Wait until the column vacuum
recovers after the insertion of
the holder. The column should
be less than 15 log before
beginning operation.
2. Click the Col. Valves Closed
button under the Vacuum panel
in microscope control. This will
expose the specimen to the
beam.
a. The FEG must operate in ultra high vacuum conditions, opening the column valve when the column
pressure is too high will result in damage.
3. The electron beam may not appear on the screen for any of the following reasons:
a. Reduce the magnification.
b. Move the specimen.
c. Remove the objective and SAD apertures.
Chapter 2
TEM Alignment
2.1 Condenser Aperture
Purpose: This procedure will center the condenser aperture down the optical axis. The proper alignment of this
aperture has two important aspects. Convenience, changing strength of condenser lens does not change beam
position. Reproducible illuminating conditions, misalignment results in beam tilt and thus changes rotation center.
Center C2 Procedure:
1. Find a vacuum region.
2. Set magnification to 13500X.
3. Insert condenser aperture #4 and remove all other apertures.
a. Starting at the largest condenser aperture makes the gun tilt alignment easier.
b. To use a different condenser aperture finish the gun tilt alignment then repeat this procedure with the
desired aperture.
4. Focus beam around center pip on the screen using intensity knob.
5. Center beam using beam shift track ball.
6. Expand the beam (clockwise) just around larger pip on the screen.
7. Center the expanded beam with the knobs on the condenser aperture housing.
8. Repeat steps 4 through 7 until the beam expands evenly when adjusting the intensity knob.
C1 Procedure (Only for aperture numbers 1, 2 and 3):
The C1 aperture is used for EDS, EELS, and other advanced data collection. For most cases the C1 aperture will not
be used and it should be set to position #4.
1. Reduce the magnification.
2. Spread the beam to a maximum with the intensity knob.
3. Center the aperture in the screen with the aperture housing knobs.
Notes:
It is very important to align this aperture in a reproducible manner so that the rotation center does not have to
be changed every time the aperture is changed.
The alignment of the condenser aperture will change between STEM (nanoprobe) and normal (microprobe)
imaging modes. When entering a new mode the aperture should be realigned.
Notes:
Each multifunction knob controls a perpendicular direction in the stigmator coils.
It often helps to adjust the condenser lens strength (intensity knob) through a range of values to correct the
astigmatism.
The condenser astigmatism will be different for each spot size and condenser aperture.
10
11
Notes:
There are two pivot point adjustments which can be made for each direction. The compensator will have to be
adjusted more (multifunction x) than the corrector (multifunction y).
Set Obj current value to ~91.8% with focus knob. Or press the eucentric focus button.
Remove objective aperture.
Over an amorphous region (like lacy carbon) condense the beam to a point using the intensity knob.
Minimize airy rings by adjusting the specimen height (Z-axis) buttons.
12
Notes:
The first technique is more difficult since it allows for a more subjective interpretation of the image. It may also
be difficult to locate a planar hole in the specimen depending on the type of grid used. For these reasons it is
often preferable to use the CCD camera.
It can also help to draw a circle on the screen to help align this pattern, under the Standard tools window
select the circle and hold down the shift key to draw a perfect circle around the FFT.
13
The rings will be hard to see if the specimen is completely in focus, try moving to slightly underfocus.
Figure 1 - Fourier transforms of amorphous carbon with various degrees of objective astigmatism.
14
Chapter 3
TEM Analysis
3.1 Objective Focus
Focus is a measurement of how far the image plane is from the specimen plane. When the image plane is above the
specimen the resulting image is said to be over-focused (a more strongly excited objective lens). When the image
plane is below the specimen then the resulting image is under-focused (A weakly excited objective lens). When the
image plane is in the same position as the specimen plane then the image is in-focus or at Gaussian defocus
(Objective lens = 91.8%). Defocus is measured in units of length and is usually on the order of nanometers and is
controlled by the focus knob on the right control pad.
In some cases it may be difficult to determine the current focus condition. The current focus condition manifests in the
final image as changes in the fresnel fringe around the edges of the specimen and in the frequency domain as
changes in the radius of the airy rings, see figure 2. Over-focused images will have a dark fresnel fringe and airy rings
of finite radius in the frequency domain. Under-focused images will have a bright fresnel fringe and airy rings of finite
radius in the frequency domain. In-focus images will have no fresnel fringe and airy rings of infinite radius in the
frequency domain.
There are a few important focus values for certain types of analysis. For normal imaging it is best to operate the
image slightly under-focused, Obj. Lens <91.8%. For high resolution imaging the spatial frequencies present in the
image are dependant upon the specific focus condition. At Scherzer defocus the number of spatial frequencies will be
maximized. This occurs at approximately -60nm.
Figure 2 - Series of images showing different focus conditions and how they manifest.
15
16
6. Adjust the astigmatism with the multifunction knobs until the pattern forms a three pointed star.
Notes:
The diffracted beams can be extremely bright making it difficult to capture them on the CCD camera without
causing damage. This actually makes it impossible to capture some types of patterns on the camera. The use
of the beam stop is important along with increasing the strength of C2 to dim the beam. If the pattern is still
too bright try decreasing the size of the SA aperture or the condenser aperture.
The camera introduces a magnification so before capturing the image decrease the camera length. The
camera is not exactly down the optical axis so it might be necessary to shift the diffraction pattern slightly to
the left.
Assuming that the specimen is at eucentric height it can be tilted with the and to tweak which zone axis it
is on.
Notes:
Method 1 is generally the easiest but requires thick specimens.
Method 2 can be used to achieve a rough alignment first before continuing on to method 3.
17
Notes:
The objective aperture is used to select which beams will contribute to the final image. In bright field the
diffracted beams are excluded which results in enhanced contrast, for this reason the objective aperture is
sometimes referred to as the contrast aperture.
It may be necessary to correct for objective astigmatism again upon the insertion of an objective aperture.
18
f = 1.2 C s
Chapter 4
STEM Alignments
Scanning Transmission Electron Microscopy is an operating mode available in the Tecnai F-20 instrument. In this
mode images are formed by rastering a focused beam across the specimen then collecting the electrons that emerge
to form an image. In this way the function of a STEM is very similar to a SEM. Due to the change in operating
conditions when entering STEM mode the microscope will have to be realigned in nanoprobe mode.
19
Remove the Gatan Ultrascan CCD camera by clicking insert under the camera panel.
Insert the STEM bright field detector by depressing the button on the control box.
Set the camera length above 100mm.
Using the diffraction shift move the beam to the 9 oclock position around the inner pip on the phosphorous
screen.
5. Raise the screen and click the search button under the STEM imaging panel.
6. Reduce the magnification and maximize the brightness of the image using the diffraction shift.
4.4.2 Annular dark field detector
This detector generates an image from the diffracted electrons exiting the specimen. In addition to diffraction contrast
Z contrast is also possible.
Procedure:
20
Chapter 5
STEM Analysis
5.1 Image Acquisition
Procedure:
1. Find the area of interest while using the continuous search.
2. Focus the image using the focus knob.
3. Set the desired contrast and brightness levels either
manually under the STEM detector panel or automatically by
clicking the Auto C/B button.
4. To acquire an image click acquire under the STEM imaging
panel.
a. Dwell time and resolution can be adjusted under the
STEM imaging panel.
21
Notes:
Alternatively the Automap function can be used which will identify elements and generate maps automatically.
5.2.3 EDS Quantification
Procedure:
1. Collect spectrum with a dead time of ~20-30%.
a. Optimal dead times are extremely important for accurate quantification.
2. Go to the TIA window by clicking the blue icon in the bottom right of microscope control.
3. Under the Quant Setup tab click Spectrum Conditions.
4. Set the thickness and mass density of the specimen.
5. Under the EDX Quant tab click Quantify.
a. The quantification information will appear in the analysis window near the bottom.
Notes:
Precise quantification is very difficult and requires a detailed understanding of the sample and machine
parameters.
22
Chapter 6
Gatan Image Filter
6.1 Introduction
When the electron beam passes through the specimen it losses energy through a variety of different processes.
These inelastically scattered electrons contain additional information about the specimen. The GIF system is used to
sort out and collect this information. The energy distribution of this signal contains a large amount of information about
the elemental composition and valence states of the sample. This technique is considered complimentary to EDS
since it has improved detection of lighter elements and a higher energy resolution.
The GIF system consists of several different components located on the base of the column, see figure ###. Electrons
are passed through a magnetic prism which sorts them out based upon their energy. A series of magnetic lenses
project the resulting signal through an energy slit and then onto the GIF CCD camera. Images are captured using the
Gatan DigitalMicrograph and the Filter Control software.
The GIF system is extremely sensitive to external interference. Small changes in the surrounding electromagnetic
fields will cause large changes in the alignment of the GIF system. For this reason it is recommended to use to the
wooden chairs (which lack movable magnetic components). There is a field cancellation system around the room
which should eliminate some interference.
6.2 Alignment
The GIF must be aligned precisely in order for it to function. Any change in the conditions within the microscope will
require it to be realigned. A smaller condenser aperture will improve the spherical and chromatic aberrations but at the
cost of brightness.
6.2.1 Tune
This procedure must be preformed at the beginning of the session and every time the magnification or other optical
conditions are changed. The GIF system comes with a set of scripts which will perform most of the alignments
automatically.
Procedure:
1. Align the microscope in TEM mode.
2. Set the desired magnification.
a. The prism introduces an additional 20X magnification.
3. Go to a vacuum region.
4. Set the GIF to imaging mode, the imaging button will be lit.
5. Turn on the TV monitor.
6. Click TV image in the AutoFilter panel.
7. Set the GIF to imaging mode in the AutoFilter panel, the imaging button will
be lit.
8. Center the beam at the 9 oclock position of the center pip.
9. Expand the beam to the edge of the screen with the intensity knob,
clockwise.
10. Raise the screen and view the image on the TV monitor.
11. Make sure the image is not too bright.
a. Reduce intensity if it is.
12. Extract the TV camera by clicking the TV image button in the AutoFilter
panel.
13. Click Align ZLP. This process will take 1-2 minutes.
a. Follow the directions given by the software. It may ask you to adjust
the intensity, do so with the intensity knob.
b. Repeat this step if the energy shift is more than 2eV.
14. Click the Tune GIF button in the AutoFilter panel. This process will take 3-5
minutes.
a. Follow the directions given by the software. It may ask you to adjust
intensity.
b. Repeat this step.
23
8.
9.
10.
11.
Lower phosphorous screen and center the feature at the 9 oclock position at the center pip.
Decrease the intensity of the beam with the intensity knob.
Click spectroscopy in AutoFilter.
Click enter to confirm the reduction in intensity.
Select the desired entrance aperture.
a. 0.6mm or 2mm apertures are often the best.
b. A smaller entrance aperture will improve the energy resolution at the cost of intensity.
Reduce the search time to a minimum (0.01s).
Select the region of interest in the spectra by either clicking the Zero Loss, Plasmon, Pre-C, or Custom
button.
a. If a custom energy shift is desired enter the appropriate value in the energy shift section then click the
enter button.
Raise the phosphorus screen.
Click search.
a. Adjust the intensity such that the signal is visible. As the energy loss increases the signal decreases
so it may be necessary to increase the intensity.
b. Note the search time and highest intensity of the signal on the graph.
Adjust the acquisition time such that the GIF CCD is not over exposed.
Click Acquire
Notes:
24
When the microscope is in imaging mode the GIF system is called diffraction-coupled.
If the spectrum ever becomes yellow or red IMMEDIATELY lower the screen.
Notes:
When the microscope is in imaging mode the GIF system is called image-coupled.
The Tune GIF and Align ZLP routines cannot be preformed while the microscope is in diffraction mode.
If the spectrum ever becomes yellow or red IMMEDIATELY lower the screen.
6.3.3 STEM Mode
The microscope can be configured to simultaneously obtain an EELS spectra while generating an ADF STEM image.
This allows the user to generate EELS data from small volumes while correlating it to an image.
Procedure:
1. Align the GIF system while in nanoprobe imaging mode.
a. See procedures from section 6.2
2. Set the desired energy shift such that the edges of interest are visible.
a. Be careful when projecting the ZLP onto the GIF while in STEM mode as it may be to bright
3. Insert the TV camera by clicking the TV Image button in the AutoFilter panel.
a. This will block the beam from hitting the GIF CCD.
4. Enter STEM mode.
5. Perform the STEM alignments described in chapter 4.
6. Align the beam onto the ADF detector. See procedure 4.4.2
7. Acquire a STEM image.
8. Dwell the beam at a position of interest using the beam marker tool.
9. Look at the spectra on the TV camera and make sure that it is not too intense.
10. Set the search time to a minimum (0.01s).
11. Extract the TV camera.
12. Click search.
a. The spectrum should be visible in DigitalMicrograph.
b. At this point the beam position can be moved by dragging the red point around on the acquired STEM
image.
13. Note the intensity of the spectrum on the search and adjust the acquire time accordingly.
14. Acquire the spectra.
Notes:
If the spectrum ever becomes yellow or red IMMEDIATELY lower the screen.
Current must be controlled precisely for this technique to work. There is a delicate balance between intensity
of the spectra and STEM spatial resolution. It is always better to start with less current (higher spot size,
smaller C2) and work up to higher currents.
25
26
7. Align the resulting images by dragging the images with the mouse while holding down the Ctrl key until the
image movement is minimized.
Notes:
Longer exposure times may be required. Be sure not the damage the GIF CCD.
Remember to use the wooden chair.
To change the parameter in the Elemental Mapping Setup page click other items.
27
Chapter 7
Appendix
7.1 Alignment Quick reference
Initial conditions
Software: GIF, DigitalMicrograph, TIA, Microscope Control.
Fill out log sheet and billing information.
Full LN2 Dewar, Closed Col Valve, Stage at Home.
Condenser Aperture
Magnification: 13500X
Condense beam to center pip, center with beam shift trackball; Expand beam to larger circle, adjust aperture
knobs.
Condenser Astigmatism
Magnification: 125kX, Multifunction: Condenser stigmators.
Condense beam, adjust multifunction to make beam circular.
Gun Tilt
Magnification: 125kX; C2 aperture: #4; Multifunction: Gun Tilt, reduce sensitivity.
Condense beam, center central bright spot with multifunction.
Gun Shift
Magnification: 125kX; Crossover beam; Direct Alignments: Gun Shift
Spot: 3; center beam with Multifunction knobs
Spot: 9; center beam with beam shift trackball
Repeat
Pivot Points
Magnification: 125kX; Crossover beam
Multifunction: ppx, converge spots
Multifunction: ppy, converge spots
Beam Shift
Magnification: 125kX; Crossover beam; Direct Alignments: Beam shift
Move beam to center of screen using beam shift trackball
Eucentric Height
Magnification: 125kX; Crossover beam onto carbon.
Adjust Z-axis until no airy rings.
Rotation Center
Magnification: 125kX; Expand beam past screen; Multifunction: Rotation Center; Specimen in focus
Minimize image movement
Objective Astigmatism
Magnification: 125kX; Expand beam; insert CCD camera; Multifunction: Obj. Stig.
Raise screen and search for image.
Process Live FFT
Adjust multifunction until rings are circular.
28
TEM BF (microprobe)
2mm, #4
Any
10m, #5
1-5
1
Slightly under-focused, Obj. Lens < 91.8%
TEM BF (microprobe)
2mm, #4
50-100 m, #1-3
Depends, used to select specific range of diffracted beams.
3-5
1
Scherzer, Defocus = -60nm.
STEM (nanoprobe)
30m, #1
150m, #4
None, #7
5-8 (depending on current required)
6-9 (depending on current required)
In-focus, Obj. Lens = 92.5%
TEM BF (microprobe)
2mm, #4
Any
None, #7
3-5
Any
In-focus, Obj. Lens = 91.8%
High Magnification
Image focus
Diffraction focus
Contrast forming
Area selection
Image stigmation
Diffraction stigmation
Low Magnification
Diffraction (LAD) focus
Image focus
Area selection
Contrast forming
Diffraction stigmation
Image stigmation
29
H
He
Li
Be
B
C
N
O
F
Ne
Na
Mg
Al
Si
P
S
Cl
Ar
K
Ca
Sc
Ti
V
Cr
Mn
Fe
Co
Ni
Cu
Zn
Ga
Ge
As
Se
Br
Kr
Rb
Sr
Y
Zr
Nb
Mo
Tc
Ru
Rh
Pd
Ag
Cd
In
Sn
Sb
Te
I
Xe
Cs
Ba
La
K
13
22
55
111
188
284
401
532
685
867
1,072
1,305
1,560
1,839
2,146
2,472
2,822
3,203
3,607
4,038
4,493
4,966
-
30
O
15
14
58
59
60
61
62
63
64
65
66
67
68
69
70
71
72
73
74
75
76
77
78
79
80
81
82
83
90
92
Ce
Pr
Nd
Pm
Sm
Eu
Gd
Tb
Dy
Ho
Er
Tm
Yb
Lu
Hf
Ta
W
Re
Os
Ir
Pt
Au
Hg
Tl
Pb
Bi
Th
U
31
883
931
978
1,027
1,080
1,131
1,185
1,241
1,295
1,351
1,409
1,468
1,528
1,588
1,662
1,735
1,809
1,883
1,960
2,040
2,122
2,206
2,295
2,389
2,484
2,580
3,332
3,552
110
113
118
120
129
133
140
147
154
161
168
180
185
195
214
229
245
260
46
60
71
83
98
118
138
157
335
381
20
22
21
21
22
20
25
26
20
29
32
24
28
31
36
36
35
45
50
52
54
58
75
19
24
88
96
32
M
0.833
0.883
0.929
60
61
62
63
64
65
66
67
68
69
70
71
72
73
74
75
76
77
78
79
80
81
82
83
84
85
86
87
88
89
90
91
92
Nd
Pm
Sm
Eu
Gd
Tb
Dy
Ho
Er
Tm
Yb
Lu
Hf
Ta
W
Re
Os
Ir
Pt
Au
Hg
Tl
Pb
Bi
Po
At
Rn
Fr
Ra
Ac
Th
Pa
U
37.185
38.535
39.914
41.323
42.761
44.229
45.728
47.257
48.818
50.410
52.035
53.693
55.362
57.106
58.864
60.655
62.482
64.346
66.246
68.185
70.106
72.176
74.228
76.321
78.460
80.636
82.855
85.124
87.437
89.790
92.190
94.643
97.143
5.230
5.431
5.636
5.846
6.059
6.275
6.495
6.720
6.948
7.181
7.414
7.654
7.898
8.145
8.396
8.651
8.910
9.173
9.441
9.711
9.987
10.266
10.549
10.836
11.128
11.424
11.724
12.029
12.338
12.650
12.966
13.291
13.613
0.978
1.032
1.081
1.131
1.185
1.240
1.293
1.348
1.406
1.462
1.521
1.581
1.645
1.710 (Si, Rb)
1.775 (Si, Sr)
1.843
1.910
1.980
2.051
2.123
2.195
2.271
2.346 (S, Mo)
2.423
2.501
2.581
2.662
2.744
2.825
2.911
2.996
3.082
3.171
7.6 Glossary
7.6.1 List of Acronyms
ADF
BF
CBED
CCD
DF
DP
EDS
EELS
EFTEM
FFT
GIF
HRTEM
IGP
ROI
SAED
SEM
STEM
TEM
TIA
ZLP
7.6.3 Terminology
Accelerating Voltage
The
electrical
potential electrons experience after exiting the
gun which causes them to accelerate. This
controls the energy of each electron.
Aperture
A hole or opening which the
electrons travel through. There position is
controlled through the aperture housing knobs on
the side of the column.
Astigmatism Astigmatism refers to an uneven
level of focus in the lens. There are three different
types of astigmatism that the user can correct for:
Condenser, Objective, and Diffraction.
Backfocal plane
The plane just below the
specimen plane which contains the diffraction
pattern.
Beam Blanker
A deflection coil which
shifts the beam completely off the specimen.
Beam Stop Located on the right of the column
this device is used to block out the zero-order or
forward scattered spot so that it does not damage
the CCD chip.
Binning
The combination of several pixels
into one. Increasing the binning reduces image
size and increases intensity.
Binoculars These are used for a finer
inspection of the image projected onto the lift up
screen.
Brightness The current density per unit solid
angle of the source with SI units of AM-2sr-1.
C1 Aperture The top aperture on the column.
For most cases this aperture will be left out, in
position #4. This aperture is located just above the
crossover below the C1 lens.
C2 Aperture The second aperture from the top
of the column. For most cases this aperture will be
used as the beam defining aperture and is located
in the second condenser lens.
Center pip The small central ring on the
phosphorus screen.
Cold Finger Improves the vacuum conditions
within the column and prevents contamination.
Column
The main section of the
microscope that houses the optics and detectors.
Column Valve
Also referred to as the V7,
this valve isolates the gun from the rest of the
column.
Compustage This devices controls the position
of the specimen within the column.
Condenser Lens
This lens controls the size
of the probe on the specimen by adjusting the
C = I
Io
35
Extractor
The anode plate below the FEG tip
which draws electrons out of the tungsten
filament.
FEG emission
A measure of the amount of
current from the gun to the extraction anode. This
should be on the order or ~60m.
Field-cancellation system The Tecnai room is
surrounded with a field cancellation system that
limits the amount of electromagnetic interference.
The control box is located on the right side of the
work bench.
Field Emission Gun The component of the
microscope responsible for generating the electron
beam. The Tecnai is equipped with schotkey field
emission gun.
Forward scattered spot
Also referred to as
the zeroth-order spot (000). This beam in the
diffraction pattern is formed by electrons which
have been not been scattered by the specimen.
Gatan Image Filter This system has several
different components that allow the user to form
images using electrons of particular energies or
spectra showing the distribution of electron
energies.
GIF CCD
The CCD which forms both EEL
spectra and EFTEM images. This camera is
extremely sensitive and expensive.
Goniometer The device on the column which
provides an tilt.
Guassian Defocus Refers to the focus
condition in which the image plane is at the
specimen plane, objective lens ~91.8%. This is
the same as in-focus.
Gun Lens
An electrostatic lens just below the
gun which controls how spread of the electron
beam entering the condenser system.
High-tension The voltage applied to the anodes
below the gun which accelerate the electrons to
high energies.
Holder
Refers to the device which holds
the TEM specimen.
Image Plane The plane which is being projected
onto the screen or CCD. The position of this plane
is controlled by the objective lens.
In-focus
Refers to the focus condition in
which the image plane is at the specimen plane,
objective lens ~91.8%. This is the same as
Guassian defocus.
Intensity
The amplitude of signal being
generated by the detector, related to the current
Zone axis
A specific specimen orientation in
which
two
crystallographic
planes
are
intersecting.
7.7 References
1.
2.
3.
4.
5.
6.
7.
8.
9.
37