QA For IGRT
QA For IGRT
IGRT QA
Agenda
IGRT QA Tests
QA Tools
Examples of QA Protocols
Summary
On Board Imager
Radiographic mode
Fluoroscopic mode
Acuity
Radiographic Mode
Setup patient
2.
Extend arms
3.
4.
5.
Automated
Manual
6.
7.
8.
9.
10.
Treat
Sample images
Kilovoltage Radiographs
Cone-Beam CT
CBCT Mode
1.
Setup patient
2.
Extend arms
3.
4.
5.
Manual
Automated
6.
7.
8.
9.
10.
Treat
VARIAN Medical Systems
CT
CBCT
CBCT
CBCT
CT Simulator
Intrafraction motion
Setup patient
Extend arms
Treat
IGRT QA
Yoo et all., A quality assurance program for the onboard imager, Medical Physics, Vol. 33, No. 11,
November 2006
IGRT QA
Primarily a machine QA
Parameters to be tested
Safety
X-Ray beam
Imaging
Functionality
Anti-collision system
MV detector arm
Detector
Front
Side
kV source arm
kV detector arm
Arm
Arm
Source
Detector
Front
Front
Side
Side
Back
IGRT QA Safety
Daily checks
Motion Enable
Detector and arm motion is only possible when motion enable bars on the
hand pendant or motion enable button on the OBI console are pressed
VARIAN Medical Systems
10
Acquire an image
11
Tolerance
displacement < 2 mm
Procedure
The Cube Isocenter Phantom was used for this daily QA procedure
Two sets of 1-mm lead shots (bbs) were used to define 2 separate
spatial locations (points)
In each set, one bb is on the anterior side (AP bb) and the other (Lat
bb) is on the lateral side of the phantom
The first set defines the isocenter and the second set defines a point
1.5 cm anterior, inferior, and lateral from the isocenter
12
The OBI daily QA procedure starts with setting up the Cube phantom
center (central bb set) at the mechanical isocenter of the linac by using
the cross hairs projections anteriroly and laterally
VARIAN Medical Systems
Lat DRR
Lat KV Image
Using the plan created from Eclipse, AP MV and Lat KV images were acquired
(shown only Lat image)
VARIAN Medical Systems
13
Lat DRR
KV Lat Image
AP DRR
MV AP Image
The alignment of the central bbs with the digital graticule (representing the imager
isocenter) will be checked from both AP and Lat images
VARIAN Medical Systems
Long: __ mm OK 1.5 mm
Long: __ mm OK 1.5 mm
Lat:
Vert: __ mm OK 1.5 mm
__ mm OK 1.5 mm
14
Results
Daily mismatch
distance (mm)
MV Isocenter Check
1.6
1.4
1.2
1
0.8
0.6
0.4
0.2
0
S/I
R/L
3 5 7 9 11 13 15 17 19 21 23 25 27 29 31
Daily m easurem ent (days)
Results
Daily mismatch
distance (mm)
KV Isocenter Check
1.4
1.2
1
0.8
0.6
0.4
0.2
0
S/I
A/P
1 3
5 7 9 11 13 15 17 19 21 23 25 27 29 31
Daily Measurem ent (days)
15
Safety
Blades calibration
Couch motion
Linear
Rotation
IGRT QA Safety
16
17
18
Blades calibration
MV Image Quality
kV Image quality
Shift calculation
Measurement tools
19
Isocenter Accuracy
Tolerance
Preparation
Isocenter Accuracy
Procedure
Blade Calibration
Plate
Marker phantom
20
Preparation
Isocenter accuracy
(May 2005 Apr 2006)
Mechanical accuracy
Isocenter over gantry rotation (Dec, 05 July, 06)
S
R
G0
P
S
A
G90
R
S
L
G180
A
S
P
0.2
Displacement (cm)
G270
0.1
KVS
0.0
-0.1
-0.2
KVD
KVS
Ver
KVD
Long
KVS KVD
Lat
21
MV Image Quality
Set up detector
Weekly checks
Monthly checks
Image geometry
Distance
Angle
22
Fluoroscopic
Radiographic
Crosshair #1 reference
Crosshair #2 test
23
Shift Calculation 2D / 2D
kV images (AP/Lat)
Preparation
Phantom
2D2D match
1.
2.
Before match
After match
Results
2D2D match and couch shift
(May 2005 Apr 2006)
2
1
1
Lat
0
Ver
-1
Long
Difference (mm)
Tolerance
Disagreement (mm)
1
0
Ver
Long
3
Lat
-1
-2
24
Measurement tools
Measurement of
Distance
Area
Angle
Annual checks
HVL
25
Daily checks
Geometry
Annually (Monthly)
Uniformity
Contrast
Resolution
HU calibration
Acquire CBCT
Verify geometry
Verify HU
26
Acquire CBCT
Results HU Uniformity
ROI
Full fan
Half fan
Top
- 4.8 8.7
- 14.6 10.4
Right
- 3.4 8.9
- 14.6 11.6
Bottom
- 7.3 8.4
- 13.4 10.1
Left
- 8.5 7.8
- 15.9 9.6
Center
18.0 9.0
15.6 11.3
Full-fan
Half-fan
27
Contrast
Performance
0.5 %
Resolution
Performance
10 lp/cm
28
Half fan
Low contrast
resolution
Disk 4
Disk 5
Disk 4
Spatial
resolution
Group 6
Group 8
Group 6
Slice thickness
0.1 mm
2.39 2.60
2.45 2.51
Spatial linearity
0.5 mm
49.8 50.2
49.8 50.0
Half-fan
Spatial
resolution
Full fan
Low contrast
resolution
Full-fan
Tolerance
HU Calibration
1. Air =-1000
2. LDPE = -100
3. Acrylic = 120
4. Teflon = 990
29
Results HU linearity
Full fan
Half fan
Teflon
989.0 10.5
1007.8 22.8
Delin
363.1 10.7
354.0 15.1
Acrylic
96.0 9.0
116.1 12.3
Poly-
- 42.8 8.8
- 48.5 10.5
LDPE
- 101.2 8.8
- 106.9 16.1
PMP
- 196.6 8.6
- 206.5 12.7
Air
- 998.3 2.4
- 995.7 13.3
Full-fan
Half-fan
On Board Imager
Isocenter cube
RPM
30
Isocenter Cube:
alignment of the kV to the
MV beam
RPM-Gating phantom:
gated radiographs
and pretreatment fluoroscopic
verification of gating
Cone Beam CT
CATPHAN
31
Norm calibration
Norm Phantom
CATPHAN
Geometry
2 Norm phantoms
Resolution
HU
Contrast
Head
Body
Examples of QA Protocols
32
Daily OBI
Weekly OBI
Monthly OBI
(continues)
VARIAN Medical Systems
33
Monthly CBCT
(continues)
Weihua Mao, Louis Lee, and Lei Xing Development of a QA phantom and automated
analysis tool for geometric quality assurance of onboard MV and kV x-ray imaging systems,
submitted to Med. Phys.
34
Summary
QA should cover
Safety
Imaging
Functional tests
35