Turbidity Monitors Comparison

The document compares specifications for various on-line turbidimeters. It lists the method, consumables, measuring range, accuracy, sample temperature range, calibration, sensor cleaning, response time, water usage, flow monitoring, flow control, sample inlet pressure, analog outputs, and relay outputs for each turbidimeter model. Over 20 turbidimeter models are compared side-by-side based on their technical specifications.

Uploaded by

maserras
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
210 views

Turbidity Monitors Comparison

The document compares specifications for various on-line turbidimeters. It lists the method, consumables, measuring range, accuracy, sample temperature range, calibration, sensor cleaning, response time, water usage, flow monitoring, flow control, sample inlet pressure, analog outputs, and relay outputs for each turbidimeter model. Over 20 turbidimeter models are compared side-by-side based on their technical specifications.

Uploaded by

maserras
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 1

ON-LINE TURBIDIMETERS COMPARISON SPREADSHEET

Method

Consumables

SWAN AMI Turbitrace*

SWAN AMI Turbitrack*

SWAN AMI Turbiwell*

HF Scientific Micro TOL*

ATI A15/76*

ABB 4670 / 4675

Yokogawa TB750G

Yokogawa TB450G

Rosemount T 1056 Clarity II

Rosemount T 1055 Clarity II

Royce TU 7685

Hach 1720E with sc100

Hach FilterTrak 660 sc with sc100

GLI (Hach) Accu4

Sigrist AquaScat WTM/HT

Sigrist (Siemens) DualScat

Mettler Toledo Trb 8300/8400/8500

Bran + Luebbe Turbidity Sensor

AquaSensors Turbidity with AV38

E & H CUE21 & CUE22

Wallace & Tiernen** TMS 561

Severn Trent MicroScatter 90

High Precision Nephelometer (LED)

High Precision Nephelometer (LED)

Laser-Based Optical, Non-Contact

Nephelometer (White Light or LED)

Nephelometer (White Light or LED)

Absorption or Nephelometric

Nephelometric (White Light)

Tungsten Lamp Surface Scatter

Nephelometric White Light or LED

Nephelometric White Light or LED

Nepholometric with LED

Nephelometer (White Light)

Nephelometric with Laser

Nepholometric with LED

90 LED in Free Falling Water

90 / 25 Scattered Light

Forward (8400) or Forward with 90

nephelometric, with LED

White Light (tungsten lamp)

Buyout of HF Scientific MicroTOL

Buyout of HF Scientific MicroTOL

Buyout of Rosemount T 1055 Clarity II

650 nm

scattered light (8500), Both LED

Sub-Micron Filter used for Automatic

Humidity Absorbing Cartridge

None

Cuvet Bottles, Desiccant Pouches

None

Not Stated

Zeroing, Humidity Absorbing Cartridge


Measuring Range

0.000-100.0 NTU

Filter Element (if using zero water filter),

0.000-100.0 NTU

0.000-200.0 NTU

0-100 or 0-1000 NTU

0-4 NTU up to 0-4000 NTU

0.00-1.00 NTU up to 0-1000 NTU

0-0.2 NTU up to 0-100 NTU

depending on sensor unit


Accuracy

Sample Temperature

0.001 NTU or 1% of Reading

1 to 40C

Lamp (2 per year), Filters for Zero Water

Not Stated

Not Stated

Not Stated

Lamp Once per Year Recommended

Not Stated

None

Not Stated

None

Not Stated

None

None

See HF Scientific MicroTOL

See HF Scientific MicroTOL

See Rosemount T 1055 Clarity II

0-0.1 to 0-100 NTU Low Range

0.001-200 NTU

0.001-200 NTU

0-4.000 NTU up to 0-4000 NTU,

0.000-100.0 NTU

0.000-5000 mNTU

0.000-100.0 NTU

From 0.0-0.1 up to 0-4000 FNU

0-2000 NTU

0.00-400 NTU

2-1000 FNU

0.001-100 NTU

See HF Scientific MicroTOL

See HF Scientific MicroTOL

See Rosemount T 1055 Clarity II

2% of reading

Not Stated

Not Stated

</= 1%

3% of measured value

2% of reading or 0.015 NTU

See HF Scientific MicroTOL

See HF Scientific MicroTOL

See Rosemount T 1055 Clarity II

Desiccant, Lamp Assembly?

0.001 NTU or 1% of Reading

1 to 40C

0.003 NTU or 1% of Reading

Up to 40C

0-10 to 0-2000 NTU High Range

2% of reading or 0.02 Below 40 NTU

5% of reading or 0.02 NTU,

2 % of reading (limited by uncertainty in

0-1 NTU: 2% of reading or 0.015 NTU,

0-1 NTU: 2% of reading or 0.015 NTU,

5% of reading for 0/400 NTU

2% of reading or 0.015 NTU

From 0 to 1000 mNTU: 3% of reading or

5% of reading above 40 NTU

whichever is greater, on 40/400 scale

Formazine standards) from 0-100 NTU

whichever is greater.

whichever is greater.

10% of reading for 400/4000 NTU

(whichever is greater) from 0 to 40 NTU;

5 mNTU, whichever is greater

whichever is greater

10% of reading or 2 NTU, whichever

5 % from 100-250 NTU

0 - 20 NTU: 2% of reading

0 - 20 NTU: 2% of reading

5% of reading from 40 to 100 NTU

From 1000 to 5000 mNTU: 2% of reading

5% of reading above 40 NTU.

is greater, on 400/4000 scale

up to 10 % FSD with high range models

0 to 50C

0 to 50C

5 - 35C

5 - 35C

0 to 50C

0 to 50C

0 to 50C

Not Stated

0 to 50C

Not Stated

manually selectable

0 to 50C

0 to 50C

0-60C

0 to 40C

-10C to 80C short-time

0 to 140C

-4C to 45C

-5C to 50C

See HF Scientific MicroTOL

See HF Scientific MicroTOL

See Rosemount T 1055 Clarity II

Factory defined (from sensor), single-

Not Stated

Manual zero and span

See HF Scientific MicroTOL

See HF Scientific MicroTOL

See Rosemount T 1055 Clarity II

Not Stated

Function available on relays for

See HF Scientific MicroTOL

See HF Scientific MicroTOL

See Rosemount T 1055 Clarity II

120C up to 195C with cooling option


Calibration

Automatic or manual zero with built-in

Low & high level span verification with

Factory calibrated with Formazine

Manual with HF Scientific reuseable

sub-micron filter, span calibration not

2 test units

No field calibration - verification only

standards or with Formazine standards

required however it can be checked

Manual zero and span

Manual with Formazine or dry

Manual Zero and Span (Formazine) or

Automatic zero with zero water (2 hrs.)

Manual zero and span with DI water

Manual zero and span with DI water

secondary standards

grab sample

or with light source off

and Formazine or commercial standard,

and Formazine or commercial standard,

Span verification check block

Manual span with formazine (10+ litres)

or grab sample

or grab sample, or lamp calibration

2 or 3 point calibration depending on range

manually with test unit


Sensor Cleaning

Response Time

Manual with Formazine

Manual with Formazine or Hach

Single point at 800 mNTU 50 mNTU.

Manual with Formazine or lab sample

Optional automatic with internal reference

standards

with Hach standard

Periodic verification with Cal-Cube

standard

Calibration unit with a solid reference

point with sampling, 2-5 point with


standards, pre-defined from factory,

or with calibration disk

Manually (auto-zero prolongs cleaning

Manually

Not required on optics

Ultrasonic autoclean available for

intervals)

Automatic purging of optical chamber,

Manual cleaning of sample chamber,

most ranges

tubing, and pressure valve

optional cleaning module available

Typically T90 < 15 seconds with

Typically T90 < 15 seconds with

T90 typically 3 minutes

sample set at 10 lph

sample set at 10 lph

Adjustable (5 to 500 seconds)

Optional with LED version sensor only

Wiper system on some models

Optional ultrasonic transducer

Automatic (2 hr. cycle)

or Cali-Cap offline tool


Manual

Manual

With submersible probe using air

Manual

Manual

Manual

Not Stated

Manual

blast initiated by controller or manual

95% in 10 seconds

Varies with flow rate, typically 90 % step

Not Stated

Not Stated

change in 2 minutes at 1 lpm or in 45 sec.

90% of final value after a step change

90% of final value after a step change

in 4.5 minutes

in 4.5 minutes

10 seconds for 98% of input

Relay provided for external cleaning


system

Initial response in 1 minute 15 seconds

external wash

Not Stated

9.5 seconds at 1 GPM (residence time)

Not Stated

Not Stated

Not Stated

< 5 seconds

90% in 120 seconds

See HF Scientific MicroTOL

See HF Scientific MicroTOL

See Rosemount T 1055 Clarity II

for a full-scale step change

with high range unit


Water Usage

5 to 20 lph

5 to 20 lph

20 to 60 lph

100 ml/min to 1 lpm

Not Stated

0.5 to 1.5 lpm

0.05 to 20 lpm

1.5 to 2 lpm

250 - 750 ml/min

250 - 750 ml/min

0.2 to 25 lpm (if using flow cell)

200 to 750 ml/min

100 to 750 mL/min

0.19 to 26.5 lpm

2.5 to 7 lpm

In-Pipe

In-line

Not Stated

250 to 500 ml/min

See HF Scientific MicroTOL

See HF Scientific MicroTOL

See Rosemount T 1055 Clarity II

Flow Monitoring

Integral paddlewheel flow sensor with

Integral paddlewheel flow sensor with

No

Flow alarm optional

None

Not Stated

Not Stated

None

None

None

None

None

None

None

None

None

None

None

None

See HF Scientific MicroTOL

See HF Scientific MicroTOL

See Rosemount T 1055 Clarity II

monitoring

monitoring

Built-In manual flow control valve

Built-In manual flow control valve

Constant-head flow assembly

Built-In regulator and backpressure

None

By User

By User

By user with numerous available

By User

Optional rotameter

Yes, on flow cell

By User

By User

By User

By User

By User

By User

By User

Optional flowmeter and regulator

See HF Scientific MicroTOL

See HF Scientific MicroTOL

See Rosemount T 1055 Clarity II

Optional head hanks

parts from Yokogawa - see manual

72.5 psi max.

14 to 72.5 psi

Not Stated

See HF Scientific MicroTOL

See HF Scientific MicroTOL

See Rosemount T 1055 Clarity II

Sensor only, control unit may be

1 x 4-20mA standard, 1 optional, also

See HF Scientific MicroTOL

See HF Scientific MicroTOL

See Rosemount T 1055 Clarity II

available separately

available with none

See HF Scientific MicroTOL

See HF Scientific MicroTOL

See Rosemount T 1055 Clarity II

Flow Control

valve
Sample Inlet Pressure

15 to 145 psi

15 to 145 psi

Not Stated

1 to 200 psi (built-in regulator set at

0 to 100 psig

Up to 43.5 psi

available
30 psig max.

30 psig max.

87 psi max. at 20C

Not Stated

Not Stated

15 psi)
Analog Outputs

2 x standard 0/4-20mA

2 x standard 0/4-20mA

2 x standard 0/4-20mA

1 x optional 0/4-20mA

1 x optional 0/4-20mA

1 x optional 0/4-20mA

4-20mA (supercedes RS-485)

0-50 psig at 20C

Pressureless

290 psi max.

0-150 psig at 20C (high pressure design)


1 x 4-20mA

1 x 0-10 mA, 0-20 mA or 4-20 mA

1 x 4-20 mA, 1 x 0/4-20mA

1 x 4-20mA or 0-5VDC Output

Optional 2nd current output supercedes

2 x 0/4-20mA, optional superimposed

1 or 2 x 0/4-20mA, depending on number

1 x 0/4-20mA

HART on Output 1

of sensor inputs

1 x optional 0/4-20mA

4 x Relays

3 x Optional Relay Outputs

2 x 5A 220V

2 x 0/4-20mA

2 x 0/4-20mA

2 x 0/4-20mA

Up to 150 psi with NPT fitting (1/2" or 1")

Not Stated

Up to 362 psi with other fittings


2 x 0/4-20mA

2 x 0/4-20mA (one for each measuring

4 x 0/4-20mA

angle)

the RS485 output

Relay Outputs

Inputs

3 x Relays 1A/250 VAC

3 x Relays 1A/250 VAC

3 x Relays 1A/250 VAC

1 x Potential-free contact, for hold or

1 x Potential-free contact, for hold or

1 x Potential-free contact, for hold or

remote off

remote off

remote off

2 x Hi/Low 2A/240VAC

None

3 x Relays 5A/220 VAC

None

2 x Relays 3A/250 VAC/VDC

None

3 x Relays 2A/250VAC, 3A/30VDC

1 x 250VAC or 220VDC, 2A for range

1 Digital Output

3 x Contacts for alarms

2 Contacts

1 Digital input for range switching

3 x Relays 5A/230 VAC

3 x Relays 5A/230 VAC

4 x Relays 5A 115/230VAC,

2 x Relays 250VAC, 4 A

2 x Relays 250VAC, 4 A

5A @ 30VDC resistive
None

None

None

None

None

None

None

None

4 x Relays (2 limit, 1 alarm, 1 wash)

Sensor only, control unit may be

1 standard, 1 optional, also available

250 VAC 5A, 30VDC 5A

available separately

with none

4 buffered digital inputs (0 to 5V)

Sensor only, control unit may be

None

See HF Scientific MicroTOL

See HF Scientific MicroTOL

See Rosemount T 1055 Clarity II

1 Digital Hold Input

available separately

Optional RS-232, RS-485, Modbus,

See HF Scientific MicroTOL

See HF Scientific MicroTOL

See Rosemount T 1055 Clarity II

3 digital inputs for stored parameter sets


Digital Communications

Control Function

Optional Profibus DP Interface

Optional Profibus DP Interface

Optional Profibus DP Interface

RS-485 (supercedes 4-20mA) for

Optional Modbus RS232 or RS485

Optional Modbus RS232 or RS485

HART Protocol

Optional Modbus Interface

Optional Modbus Interface

Optional Modbus Interface

Modbus, HF Simplebus, HF Online

Optional LonWorks

Optional LonWorks

RS-232 Interface

Optional HyperTerminal Interface (RS232)

Optional HyperTerminal Interface (RS232)

Optional HyperTerminal Interface (RS232)

Interface

Wireless IR Port

Wireless IR Port

None

None

None

Programmable P, PI, PID, or PD for

Programmable P, PI, PID, or PD for

Programmable P, PI, PID, or PD for

2 x relays or 2 x 0/4-20mA

2 x relays or 2 x 0/4-20mA

2 x relays or 2 x 0/4-20mA

None

On/Off

Optional RS485 Modbus

None

Serial RS422 or RS232C

None

RS-232C

None

HART Optional

None

None

None

Optional RS232 Output

Not Stated

With relays: high/low phasing,

Optional Profibus DP

Optional Profibus DP, Profibus FMS,

RS-232 Interface

CAN-Bus

Interbus, or CAN OPEN

DeviceNet

None

None

Not Stated

Not Stated

PID on Optional 4-20mA Output

See HF Scientific MicroTOL

See HF Scientific MicroTOL

See Rosemount T 1055 Clarity II

setpoint, deadband, overfeed timer, off


delay, and on delay

Power

Display

85-265 VAC, 47-63 Hz

85-265 VAC, 47-63 Hz

85-265 VAC, 47-63 Hz

or 24 VDC (optional)

or 24 VDC (optional)

or 24 VDC (optional)

Multi-function backlit LCD

Multi-function backlit LCD

Multi-function backlit LCD

100-240 VAC, 47-63 Hz

110/220 VAC 10%, 50/60 Hz

100-130 VAC, 200-260 VAC, 50/60 Hz

100-240 VAC -15%/+10%, 50/60 Hz

100/110/200/220 VAC, 50/60 Hz

20-30 VDC, 85-265 VAC, 47.5 to 65.0 Hz

85-265 VAC, 47.5-65.0 Hz

110/220 VAC +/-10% 50/60 Hz

100-230 VAC, 50/60 Hz

100-230 VAC, 50/60 Hz

90-130 VAC or 180-260 VAC, 50/60 Hz

85-264 VAC 47-63 Hz, or 24 VDC

85-264 VAC 47-440 Hz, or 24 VDC

100-240 VAC, 47-63 Hz or 20-32 VDC

24 VDC

24 VDC

See HF Scientific MicroTOL

See HF Scientific MicroTOL

See Rosemount T 1055 Clarity II

Multiline Backlit LCD

16 character alphanumeric backlit LCD

Backlit 2-line LCD with alarm LED's

Multi-function LCD

4-digit LED with LED indicators

Graphic backlit LCD

2-Line 16-character backlit LCD

Single line alphanumeric backlit

Graphical LCD

Graphical LCD

Graphical backlit LCD

Plain-text LCD

Plain-text LCD

4-Line, 20-character backlit

None on sensor

2 lines, 16 character LCD, backlit

See HF Scientific MicroTOL

See HF Scientific MicroTOL

See Rosemount T 1055 Clarity II

alphanumeric
Datalogging

Yes

Yes

Yes

None

None

None

None

None

None

None

None

Yes

Yes

None

None

Not Stated

None

Not Stated

Not Stated

See HF Scientific MicroTOL

See HF Scientific MicroTOL

See Rosemount T 1055 Clarity II

Protection

IP66

IP66

IP66

NEMA 4X, IP66

NEMA 4X Wall-Mount or Panel-Mount

IP66 / NEMA 4X (front only on panel)

IP65

Not Stated

NEMA 4X / CSA 4 (IP65)

NEMA 4X / CSA 4 (IP65)

Not Stated

NEMA 4X / IP66

NEMA 4X / IP66

NEMA 4X

IP54

IP65

IP65 sealed face

IP65

NEMA 4X

See HF Scientific MicroTOL

See HF Scientific MicroTOL

See Rosemount T 1055 Clarity II

Mounting

Wall (all components on panel)

Wall (all components on panel)

Wall (all components on panel)

Wall

Panel, Wall (optional)

Wall, Panel

Pipe, Panel, Rack, Wall

Pipe or Rack

Panel, Pipe, Rack

Panel, Pipe, Wall

Panel

Wall, Floor Stand (optional)

Wall, Floor Stand (optional)

Panel, Wall, Pipe

Wall

Wall for Controller, In-Pipe (1" to 6") for

Panel, Wall

In-Line

Wall (all components on panel),

See HF Scientific MicroTOL

See HF Scientific MicroTOL

See Rosemount T 1055 Clarity II

Sensor
Password Protection

Yes

Yes

Yes

Yes

Keypad Lock

Yes

Yes

No

Yes

Yes

Not Stated

Not Stated

Not Stated

Yes

Not Stated

Not Stated

Yes

Not Stated

Not Stated

See HF Scientific MicroTOL

See HF Scientific MicroTOL

See Rosemount T 1055 Clarity II

Country of Origin

Switzerland

Switzerland

Switzerland

U.S.A.

U.S.A

U.K.

Japan

Japan

U.S.A.

U.S.A.

Australia?

U.S.A.

U.S.A.

U.S.A.

Switzerland

Switzerland

Switzerland

Germany

U.S.A.

See HF Scientific MicroTOL

See HF Scientific MicroTOL

See Rosemount T 1055 Clarity II

Features

- Automatic zero function with

- Sample measured at line pressure to

- Non-contact turbidimeter

- Dessicant alarm

- Sample can be measured at line

- Non-contact sensor

- CIP/SIP-compatible

- Self-diagnostics

- Standalone system?

- Bubble removal chamber

See HF Scientific MicroTOL

See HF Scientific MicroTOL

See Rosemount T 1055 Clarity II

- Overvoltage protection

- Optional ultrasonic cleaning

- Manual or automatic sensor-check

- Data storage in EEPROM

sub-micron filter
- Sample measured at line pressure to
eliminate bubble interference
- Error diagnostics
- Electronic record of major process
events and calibration data

eliminate bubble interference

pressure to eliminate bubble interference

- Error diagnostics

- Non-volatile data storage

- Cuvette monitoring

- Sensor monitoring

- Electronic record of major process

- Case temperarature monitoring

- Cuvette eliminates direct contact with

- Optional sensor cleaning with IR version

events and calibration data


- Automatic purging of optical chamber in
programmable intervals

- Overvoltage protection

- Overvoltage protection

- Non-volatile data storage

- Non-volatile data storage

- Case temperarature monitoring

- Case temperarature monitoring

- Built-in 5-way valve

- Real-time clock with calendar

- Real-time clock with calendar

- All components pre-mounted on PVC


panel for wall-mounting
- Error diagnostics

optics

- Bubble rejection function

- Bubble trap system

- Self-diagnostics

- 1 or 2 Sensor Inputs

- 1 or 2 sensor inputs

- Optional submersible probe

- Bubble rejection system

- Bubble rejection system

- Bubble trap

- 1m or 2m submersible version available

- Self-diagnostics

- Automatic water jet cleaning

- Debubbler

- CSA Class I Div. 2 Gropus A,B,C,D

- Autoclean relay for air pressure blast

- Controller accepts up to 2 turbidity

- Controller accepts up to 2 turbidity

- System diagnostics

- Autocleaning on some models

- Non-volatile data storage

- Signal smoothing of air bubbles

- Self-diagnostics

- Debubbler

- Flow control to avoid air bubbles

- Automatic zero with zero water

- Bubble-rejection

- Self-diagnostics

- Dirty lens monitoring

- Multiple sensor mounting options

- Bubble rejection

sensors
- Electronic record of major process
events and calibration data

sensors

function

events and calibration data

- Bubble rejection system

- Compatible with Hach networks

- Compatible with Hach networks

- Reusable Calibration Kit

- Error diagnostics

- Error diagnostics

- Automatic purging of optical chamber in

- User settings stored in EEPROM

- Electronic record of major process

- Can operate without the local display

- 3 configurable parameter sets

unit

- Cleaning funtion

- Plug & play communications adapters

- Colour influence compensation

- Integral temperature sensor

- Logs 12 calibrations and verifications

programmable intervals
- Heated optics, windows, and sample
compartment to avoid condensation
- Laser beam automatically switches
off when turbidimeter is opened
- Real-time clock with calendar

Options

- 24 VDC power

- 24 VDC power

- 24 VDC power

- Flow alarm

- Digital communications

- Digital communications

- Digital communications

- Flow controller

- RS-232 output

- RS-232 output

- RS-232 output

- Remote panel meter

- Sample degasser

- Software for data collection

- Submersion sensor for wastewater


or raw water
- 1.5" In-line tee assembly

- Dry secondary calibration standard

- In-Line analysis

- Dip System

- Two head tanks available (presurized

- De-Bubbler

- Head tank

- DC power supply

- Backplate mounted system

- 2nd sensor Input

- 2nd sensor input

- Flow, inline, submersible sensors

or simple)

- LonWorks card

- LonWorks card

- HART Interface

- 24VDC Power

- 24VDC power

- DC power supply

- Add-on sensor

- Add-on sensor

- Software for data collection from

- High pressure flow cell (P Model) for

- EExd II

- Multiple sensor fitting types

- sc1000 controller

- sc1000 controller

- High temperature options

- ATEX or FM (pending) approvals for

- Epoxy paint

multiple units

pressure to 87psi, 0.2-2 lpm flow rate

- Low and high pressure pipe mount kits

Not Stated

- Turbidity sample chamber

See HF Scientific MicroTOL

See HF Scientific MicroTOL

See Rosemount T 1055 Clarity II

classified areas

- Cleaning module

- Air-purge for optical system

- Verification kits

Other Parameters

Flow, Temperature, FTU

Flow, Temperature, FTU

Available

Restrictions

ppm (Oil-In-Water), Flow, Temperature,

None

mg/l or ppm (suspended solids)

FTU

Not Stated

Not Stated

Sample temperature max. 5C

FTU, mg/l (up to 0-2000 mg/l range) on

None

None

FTU, FNU, mg/l, ppm, or no units

FTU, FNU, mg/l, ppm, or no units

mg/l, ppm

mg/l, TE/F, FTU, Degree

mFTU, mg/l, Degree

FTU, FNU, TEF (German unit)

Not Stated

EBC (default)

FTU, EBC, ppm, g/l

Not Stated

Temperature

See HF Scientific MicroTOL

See HF Scientific MicroTOL

See Rosemount T 1055 Clarity II

Not Stated

Not Stated

Not Stated

Not Stated

Not Stated

Not Stated

Not Stated

Installation of restrictor valve on the

Not Stated

Not Stated

Not Stated

No AC Power

Not Stated

See HF Scientific MicroTOL

See HF Scientific MicroTOL

See Rosemount T 1055 Clarity II

some models

Not Stated

over ambient temperature

Not Stated

Not Stated

sensor outlet is recommended for flows


above 1.8 lpm with air in sample

Website Information

http://www.swan.ch/products/potablewater/turb http://www.swan.ch/products/potablewater/turb http://www.swan.ch/products/potablewater/turb http://www.hfscientific.com/micro_tol_specs.htmhttp://analyticaltechnology.com/cms/Default.asphttp://www.abb.com/product/seitp330/0a8791e http://www.yokogawa.com/an/tubidity/an-tb750 http://www.yokogawa.com/an/tubidity/an-tb750 http://www.emersonprocess.com/raihome/liquidhttp://www.emersonprocess.com/raihome/Liqui http://www.roycewater.com.au/index.php?page http://www.hach.com/hc/search.product.details http://www.hach.com/hc/search.product.details http://www.hach.com/hc/search.product.details http://www.photometer.com/en/products/detailshttp://www.automation.siemens.co.uk/main/ext http://ca.mt.com/mt/filters/products-applicationshttp://www.branluebbe.com/sites/branluebbe/gl http://www.aquasensors.com/e/aquacatalog/sh http://container.endress.com/ProductContainer http://www.usfwt.com/prod/prodspec.cfm?id=96http://www.severntrentservices.com/Online_Qua

Photo

Notes:

* WJF Instrumentation is a distributor for SWAN Analytical Instruments, HF Scientific, and Analytical Technology Inc.
**Wallace & Tiernen (W&T) may also affiliated with Siemens and U.S. Filter.
The specifications and information have been gathered from the public domain i.e. websites, product brochures, manuals, etc.
"Not Stated" means no information could be found relating to that topic.
The information in this comparison has been compiled by Sheldon Wrubleski of WJF Instrumentation Ltd.
In no way does WJF Instrumentation Ltd. warrant this information to be 100% accurate. If in doubt, please contact the manufacturer of the equipment directly.
If you wish to contact WJF Instrumentation, we can be reached at (403)291-5570 or at [email protected].
You can find additional free product comparisons at www.wjf.ca/articles . Your feedback of this comparison is always appreciated.