HPLC Troubleshooting Guide
HPLC Troubleshooting Guide
Troubleshooting
Guide
n Application
n Outstanding
TABLE OF CONTENTS
I.
Introduction....................................................................................... 1
II.
Abnormal Pressure........................................................................... 2
III.
Leaks................................................................................................ 4
IV.
V.
VI.
VII.
I. Introduction
LOCATING AND CORRECTING THE PROBLEM
A systematic approach to identifying the problem is the best approach to troubleshooting your HPLC system. This guide is organized by five major categories
of symptoms to help you quickly identify the source of the problem(s) you are
encountering:
pressure abnormalities
leaks
problems with the chromatogram
injector problems
other problems detected by the senses of smell, sight, and sound
When you have corrected the problem, record the incident in the system recordbook to help with future problems.
PREVENTION
Many LC problems can be prevented with routine preventive maintenance. For
example, replacing pump seals at regular intervals should eliminate pump-seal
failure and its associated problems. Section VII lists the most common problem
areas for each LC module, and preventive maintenance practices that will reduce their frequency. These suggestions should be modified to fit your particular
model of LC, and then made a regular part of your laboratory routine.
WHERE TO GET ADDITIONAL HELP
The operators and service manuals for the instrument should be consulted.
These contain exploded diagrams, troubleshooting procedures for specific
models, and part numbers to help you order replacement parts.
Other people in the lab may have had experience solving a problem which is
giving you trouble; they can be a helpful resource.
The manufacturer of your instrument can help you. Most LC manufacturers
offer free technical support to their customers.
Phenomenex has experienced technical consultants who can assist you with
almost any problem. We welcome your phone calls, faxes or emails.
Phenomenex offers seminars, as well as a complete line of reference books
on HPLC.
There are a number of reference sources that can give you guidance
in problem solving:
J.W. Dolan and L.R. Snyder, Troubleshooting LC Systems, Humana Press, NJ
(1989). Phenomenex Order No.: AA0-1717
L.R. Snyder and J.J. Kirkland, Introduction to Modern Liquid Chromatography,
2nd ed., Wiley, NY (1979). Phenomenex Order No.: AA0-1700
D.J. Runser, Maintaining and Troubleshooting HPLC Systems - A Users
Guide, Wiley, NY (1981).
J.W. Dolan, Troubleshooting, LC/GC Magazine. This is a monthly column.
SOLUTION
1. Power off
1. Turn on power
2. Fuse blown
2. Replace fuse
4. Replace piston
8. Major leak
SOLUTION
1. Faulty meter
1. Replace meter
2. Replace transducer
SOLUTION
1. Adjust setting
5. Injector blockage
6. Raise temperature
7. Controller malfunction
* Check manufacturers column warranty first. Removal of end-fittings may void column warranty.
SOLUTION
2. Leak in system
3. Improper column
4. Lower temperature
5. Controller malfunction
E. Pressure climbing
POSSIBLE CAUSE
SOLUTION
1. See section C
1. See section C
SOLUTION
SOLUTION
1. See section D
1. See section D
H. Pressure cycling
POSSIBLE CAUSE
SOLUTION
1. Air in pump
1. a. Degas solvent
b. Bleed air from pump
2. Faulty check valve(s)
4. Insufficient degassing
4 a. Degas solvent
b. Change degassing methods
(use Degassex on-line degasser)
5. Leak in system
III. Leaks
Leaks are usually stopped by tightening or replacing a fitting. Be aware, however, that overtightened metal compression fittings can leak and plastic fingertights
can wear out. If a fitting leak does not stop when the fitting is tightened a little,
take the fitting apart and inspect for damage (e.g. distorted ferrule, or particles
on the sealing surface); damaged fittings should be discarded.
A. Leaky fittings
possible cause solution
1. Loose fitting
1. Tighten
2. Stripped fitting
2. Replace
3. Overtightened* fitting
3. a. Loosen and retighten
b. Replace
4. Dirty fitting
4. a. Disassemble and clean
b. Replace
5. Mismatched parts
B. Leaks at pump
possible cause solution
4. Repair or replace
5. Repair or replace
* Use fingertight end-fittings to avoid sealing problems and the need for wrenches
2. Blocked loop
2. Replace loop
3. Adjust
D. Column leaks
possible cause solution
1. Loose endfitting
1. Tighten endfitting
E. Detector leaks
possible cause solution
2. Replace window(s)
3. Leaky fittings
3. Tighten or replace
5. Rebuild or replace
0.5 m
3 - 4 m
2 m
5 - 20 m
A. Peak tailing
possible cause solution
1. Blocked frit
1. a. Reverse fIush column (if allowed)
b. Replace inlet frit*
c. Replace column
2. Column void
2. Fill void
3. Interfering peak
3. a. Use longer column
b. Change mobile phase
and/or column/selectivity
4. Wrong mobile phase pH.
4. Adjust pH. For basic compounds,
lower pH usually provides more
symmetric peaks
5. Sample reacting with active sites. 5. a. Add ion pair reagent or volatile
basic modifier
b. Change column
Normal
Problem
PEAK TAILING
B. Peak fronting
possible cause solution
1. Low temperature
4. Bad column
C. Split peaks
possible cause solution
1. Contamination on guard or
1. Remove guard column and
analytical column inlet. attempt analysis. Replace guard
continued on next page
* Check manufacturers column warranty first. Removal of end-fittings may void column warranty.
if necessary. If analytical column
is obstructed, reverse and flush. If
problem persists, column may
Normal
Problem
be fouled with strongly retained
contaminants. Use appropriate
restoration procedure. If problem
persists, inlet is probably plugged.
Change frit or replace column
SPLIT PEAKS
2. Sample solvent incompatible with 2. Change solvent. Whenever possible,
mobile phase. inject samples in mobile phase
1. Sample overload
1. Extra-column effects
1. a. Replumb system
(shorter, narrower tubing)
b. Use smaller volume detector cell
1. Secondaryretention effects,
1.
reversed-phase mode
2. Secondaryretention effects,
2. a. Add triethylamine
normal-phase mode (basic compounds)
b. Add acetic acid
(acidic compounds)
2. Secondaryretention effects,
2. c. Add water
normal-phase mode (poly-functional compounds).
Only for normal-phase methods
which use water-miscible solvents.
d. Try a different LC method
3. Secondaryretention effects, ion-pair 3. Add triethylamine (basic samples)
1. Inadequate buffering
1. a. Use 50-100 mM buffer
concentration
b. Use buffer with pKa equal
to pH of mobile phase
c. See page 19
I. Extra peaks
POSSIBLE CAUSE solution
1. Normal
4. Filter sample
1. Thermostat column
1. Reset flowrate
1.
2.
1. Leak
1.
Normal
Problem
BASELINE NOISE
10
8. Replace lamp
9. Replace column
O. Broad peaks
POSSIBLE CAUSE solution
4. Adjust settings
5. Extra-column effects:
5. a. Inject smaller volume (e.g.,
a. Column overloaded 10 L vs. 100 L) or 1:10 and
1:100 dilutions of sample
b. Detector response time or cell b. Reduce response time or use
volume too large smaller cell
c. Tubing between column and c. Use as short a piece of 0.007-
detector too long or ID too large 0.010 in. ID tubing as practical
d. Recorder response time too high d. Reduce response time
Normal
Problem
BROAD PEAKS
11
6. Increase concentration
7. Guard column
contaminated/worn out
P. Loss of resolution
POSSIBLE CAUSE solution
Problem
LOSS OF RESOLUTION
12
1. Reduce attenuation
13
1. Valve misaligned
1. Adjust alignment.
2. Blocked loop
2. Replace loop
3. Dirty syringe
4. Blocked lines
2. Adjust
3. Valve misaligned
3. Adjust alignment
1. Blockage
2. Jammed mechanism
3. Faulty controller
14
A. Solvent smell
POSSIBLE CAUSE solution
1. Leak
2. Spill
2. a. Check for overflowing waste
container
b. Locate spill and clean up
B. Hot smell
POSSIBLE CAUSE solution
1. Overheating module
1. a. Check for proper ventilation,
adjust
b. Check temperature setting,
adjust
c. Shut module off, see service
manual
1. Pressure abnormality
1. See section II
3. Replace lamp
D. Warning lamps
POSSIBLE CAUSE solution
15
1. Bearing failure
2. Poor lubrication
2. Lubricate as necessary
3. Mechanical wear
16
Reservoir
Problem preventive maintenance
Pump
Problem preventive maintenance
1. Air bubbles
2. Replace (3 mo.)
Injector
Problem preventive maintenance
Column
Problem preventive maintenance
1. Blocked frit
1. a. Filter mobile phase
b. Filter samples
c. Use in-line filter and/or
guard column
2. Void at head of column
2. a. Avoid mobile phase pH > 8.
(Most silica-based columns)
b. Use guard column
c. Use precolumn
(saturator column)
17
General
Problem preventive maintenance
1. Corrosive/abrasive damage
1. Flush buffer from LC and clean
when not in use
18
How it works*:
From injector
contaminants
compounds of interest
Holder
HPLC column
Universal fit*:
With the patented design, SecurityGuard
can adjust to fit virtually any manufacturers
female/inverted endfitting.
19
SecurityGuard is a trademark of
Phenomenex, Inc.
Phenex Offers:
Low absorption
Maximum chemical
compatibility
Certified quality
Minimum extractables
Excellent flow rate
High total throughput
Membrane Types
RC
(Regenerated Cellulose)
NY
(Nylon)
PTFE
(Polytetrafluoroethylene)
CA
(Cellulose Acetate)
PES
(Polyethersulfone)
GF
(Glass Fiber)
This publication is distributed free of charge. Additional copies are available from:
mail:
tel.:
fax:
email:
mail:
tel.:
fax:
email:
mail:
tel.:
fax:
email:
Australia
PO Box 4084
Lane Cove, NSW 2066
Australia
Austria
Zeppelinstr. 5
63741 Aschaffenburg
Germany
Belgium
Maarssenbroeksedijk 13D
3542 DL Utrecht
Netherlands
Canada
411 Madrid Ave.
Torrance, CA
90501-1430
USA
Denmark
Gydevang 39-41
3450 Allerd
Denmark
02-9428-6444
02-9428-6445
auinfo@
phenomenex.com
01-319-1301
01-319-1300
anfrage@
phenomenex.com
+31 (0)30-2418700
+31 (0)30-2383749
beinfo@
phenomenex.com
(800) 543-3681
(310) 328-7768
info@
phenomenex.com
4824 8048
4810 6265
dkinfo@
phenomenex.com
France
Parc des Grillons, Bat.3
60 route de Sartrouville
78232 Le Pecq Cedex
France
Germany
Zeppelinstr. 5
63741 Aschaffenburg
Germany
Ireland
Queens Avenue,
Hurdsfield Ind. Est.,
Macclesfield, Cheshire
SK10 2BN, UK
Italy
Via M. Serenari, 15/D
40013 Castel Maggiore (BO)
Italy
Luxembourg
Maarssenbroeksedijk 13D
3542 DL Utrecht
Netherlands
01 30 09 21 10
01 30 09 21 11
franceinfo@
phenomenex.com
06021-58830-0
06021-58830-11
anfrage@
phenomenex.com
01 247 5405
+44 1625-501796
eireinfo@
phenomenex.com
051 6327511
051 6327555
italiainfo@
phenomenex.com
+31 (0)30-2418700
+31 (0)30-2383749
nlinfo@
phenomenex.com
Netherlands
Maarssenbroeksedijk 13D
3542 DL Utrecht
Netherlands
New Zealand
P O Box 31-601
Milford 0741
North Shore City
New Zealand
Puerto Rico
273 Sierra Morena,
Suite #104
San Juan,
Puerto Rico 00926
United Kingdom
Queens Avenue,
Hurdsfield Ind. Est.,
Macclesfield, Cheshire
SK10 2BN, UK
USA
411 Madrid Ave.
Torrance, CA
90501-1430
USA
030-2418700
030-2383749
nlinfo@
phenomenex.com
09-4780951
09-4780952
nzinfo@
phenomenex.com
(800) 541-HPLC
(310) 328-7768
info@
phenomenex.com
01625-501367
01625-501796
ukinfo@
phenomenex.com
(310) 212-0555
(310) 328-7768
info@
phenomenex.com
6116_L
www.phenomenex.com
Phenomenex products are available worldwide. For the distributor in your country,
contact Phenomenex USA, International Department by telephone, fax or email: [email protected].