ACECT System Specification: 1. Working Environment
ACECT System Specification: 1. Working Environment
1. Working environment
General
Temperature
Humidity
Power
Laboratory conditions
0 ~ 50C ambient
Without condensation
110-240 V~
2. System hardware
Hardware units
Capacitance measurement board
DDS signal
generator board
Capacitance resolution
Signal-to-noise ratio (SNR)
Power supply
Board format:
Connector to electrode
Connector to backplane board
Number of generators
Signal type
Signal frequency
Signal amplitude
Multiplexer board
Backplane board
Data acquisition
board
Power supply
2
Sinusoidal excitation and phase-sensitive
demodulation
Programmable up to 500 kHz (Default frequency:
200 kHz)
0 ~ 2 pF (Different measurement range with
different electrode pair/combination)
Up to 0.1 fF
40 dB
+5 V, 15 V
Eurocard**
SMB
DIN41612 a/c-64 way plug
2
Sinusoidal
Programmable up to 500 kHz (Default frequency:
200 kHz)
Programmable up to 20 Vp-p (Default amplitude: 16
Vp-p)
0 ~ 360 programmable
+5 V, 15 V
Eurocard
DIN41612 a/c-64 way plug
16
1, 2, 4, 8, 16
+5 V, 15 V
Eurocard
DIN41612 a/c-64 way plug
Compatible with Eurocase
DIN41612 a/c-64 way socket
16
12 bits
2
12 bits
24
PCI
>100 frames per second for a 12-electrode sensor
85-264 V~
47 ~ 440 Hz
+5 V (3 A), +15 V (1.5 A) DC, -15 V DC (0.5 A)
25 W
Yes
Ribbon cable
Eurocase
Number of wires
Type of connector to PC
50
50 pin D shape
50 pin ribbon
Type
Approximate size
3. System software
Software package
System test program
Image data acquisition program
Image reconstruction programs
Data conversion program
Software details
Overall
System test
program
Data acquisition
program
Working environment
Windows XP
Computer language for system software
Visual C++
Test offset
Test DC gain
Test data acquisition of a single set of capacitance data
System calibration
Combined excitation electrode
1, 2 and 4***
Acquisition of defined sets of image data
Calculation of normalised capacitance
Save binary raw measurement data
Generic sensitivity maps provided for 8- and 12-electrode sensors
Linear back-projection algorithm
Image display
Volume concentration estimation from normalised capacitance and image
Relaxation factor
Programmable (Default value: 1.2)
Number of iterations
Programmable
Plus all other functions in LBP algorithm
Read binary raw measurement data
Save text measurement data
Number of electrode
12
Cable
RG174 A/U
Connector
SMB
Diameter
1.5
Length
21
Weight
1 kg
USB data
Working with laptop
Instead of PCI board
acquisition board
Number of channels of analogue input
16
Resolution of ADC
12 bits
Number of analogue output
2
Resolution of DAC
12 bits
Number of DI/DO
8
Computer interface
PCMCIA
Data acquisition rate
500 k sample/s****
Accessories
Cable
RG174 A/U
Connector
SMB
SMB plug cramp tool
Self-adhesive copper foil to construct sensor electrodes
**** The overall data acquisition rate with PCMCIA interface is lower than PCI interface.
Declaration: The above specification is accurate when this document is written. Some specification may change at the
discretion of ECT Instruments Ltd.
If you are interested in this product, please contact us by
Tel. +44 161 962 7923, Email: [email protected] or [email protected]
Copyright ECT Instruments Ltd 2004 - 2006. All rights reserved.