Chapter 10 Acceptance Testing and Commissioning
Chapter 10 Acceptance Testing and Commissioning
Commissioning Measurements
Set of 189 slides based on the chapter authored by
J. L. Horton
of the IAEA publication (ISBN 92-0-107304-6):
Review of Radiation Oncology Physics:
A Handbook for Teachers and Students
Objective:
To familiarize the student with the series of tasks and
measurements required to place a radiation therapy machine into
clinical operation.
Slide set prepared in 2006
by G.H. Hartmann (Heidelberg, DKFZ)
Comments to S. Vatnitsky:
[email protected]
Version 2012
IAEA
International Atomic Energy Agency
CHAPTER 10.
10.1
10.2
10.3
10.4
10.5
TABLE OF CONTENTS
Introduction
Measurement Equipment
Acceptance Tests
Commissioning
Time Requirements
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.
10.1 INTRODUCTION
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.1 Slide 1
10.1 INTRODUCTION
Subsequent QA actions
Purpose
process
Acceptance testing
(including calibration)
Periodic QA tests
program
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.1 Slide 2
10.1 INTRODUCTION
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.1 Slide 3
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.2 Slide 1
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.2.1 Slide 1
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.2.1 Slide 2
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.2.1 Slide 3
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.2.2 Slide 1
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.2.2 Slide 2
Adequate type of
ionization chamber
Thimble ionization chambers with
volumes on the order of
0.1 - 0.2 cm3
Output factors
Measurements in rapidly
changing gradients
Calibration measurements
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.2.2 Slide 3
Example:
Congruence of radiation and
light field (as marked by pinholes)
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.2.3 Slide 1
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.2.3 Slide 2
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.2.3 Slide 3
Important:
Since the absorption peaks occur at wavelengths different
from conventional radiographic film, the adequacy of the
densitometer must be checked before use.
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.2.3 Slide 4
Diode
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.2.4 Slide 1
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.2.5 Slide 1
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.2.5 Slide 2
After the gantry has been leveled with the beam directed
vertically downward, leveling of the traversing mechanism can
be accomplished by scanning the radiation detector along the
central axis of the radiation beam indicated by the image of the
cross-hair.
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.2.5 Slide 3
Set up of RFA
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.2.5 Slide 4
Plastic phantoms
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.2.5 Slide 5
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.2.5 Slide 6
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.2.5 Slide 7
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.2.5 Slide 8
Note:
In spite of the popularity of plastic phantoms, for
calibration measurements (except for low-energy xrays) their use of is strongly discouraged, as in general
they are responsible for the largest discrepancies in the
determination of absorbed dose for most beam types.
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.2.5 Slide 9
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.2.5 Slide 10
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.2.5 Slide 11
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3 Slide 1
Characteristics (continued)
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3 Slide 2
1. Safety checks.
2. Mechanical checks.
3. Dosimetry measurements.
A number of national and international protocols exist to
guide the physicist in the performance of acceptance tests.
Example: Comprehensive QA for Radiation Oncology, AAPM Task Group 40
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3 Slide 3
Interlocks.
Warning lights.
Patient monitoring equipment.
Radiation survey.
Collimator and head leakage.
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.1 Slide 1
Interlocks
Initial safety checks should verify that all interlocks are
functioning properly and reliable.
Door interlocks.
Radiation beam-off interlocks.
Motion disable interlocks.
Emergency off interlocks.
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.1 Slide 2
1. Door interlocks:
Door interlock prevents irradiation
from occurring when the door to the
treatment room is open.
1. Radiation beam-off
interlocks:
Radiation beam-off interlocks halt
irradiation but they do not halt the
motion of the treatment unit or
patient treatment couch.
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.1 Slide 3
3. Motion-disable interlocks:
Motion-disable interlocks halt motion of the treatment unit and patient
treatment couch but they do not stop machine irradiation.
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.1 Slide 4
4. Emergency-off interlocks:
Emergency-off interlocks typically
disable power to the motors that drive
treatment unit and treatment couch
motions and power to some of the
radiation producing elements of the
treatment unit. The idea is to prevent
both collisions between the treatment
unit and personnel, patients or other
equipment and to halt undesirable
irradiation.
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.1 Slide 5
Warning lights
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.1 Slide 6
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.1 Slide 7
Radiation survey
In all areas outside the treatment room a radiation survey
must be performed.
Typical floor plan for an
isocentric high-energy
linac bunker.
X
Green means:
All areas outside the
treatment room must be
"free" of radiation
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.1 Slide 8
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.1 Slide 9
After location of these hot-spots the ionization chambertype survey meter may be used to quantify the leakage
values.
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.1 Slide 10
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.1 Slide 11
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.1 Slide 12
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.1 Slide 13
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.2 Slide 1
1.
Aim of test.
2.
Method used.
3.
Practical suggestions.
4.
Expected results.
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.2 Slide 2
Aim
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.2 Slide 3
Method
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.2 Slide 4
Practical suggestions
With the rod rigidly mounted, the collimator is rotated through its
range of motion. The point of the rod will trace out an arc as the
collimator is rotated.
Point of the rod is adjusted to be near the center of this arc. This point
should be the collimator axis of rotation.
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.2 Slide 5
Expected result
In most cases this arc will reduce to a point but should not
exceed 1 mm in radius in any event.
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.2 Slide 6
Aim
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.2 Slide 7
Method
Indicator is attached to a
point on the collimator
housing that remains
stationary during rotation
of the collimator jaws.
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.2 Slide 8
Practical suggestions
Feeler of the indicator is brought into contact with one set of jaws and
the reading is recorded.
Collimator jaw symmetry about the rotation axis is one half of the
difference in the two readings. This value projected to the isocenter
should be less than 1 mm.
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.2 Slide 9
Expected result
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.2 Slide 10
Aim
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.2 Slide 11
Method
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.2 Slide 12
Method
With the jaws in the position of the jaw motion test the
collimator angle indicators are verified. These indicators should
be reading a cardinal angle at this point, either 0, 90, 180, or
270 depending on the collimator position. This test is repeated
with the spirit level at all cardinal angles by rotating the
collimator to verify the collimator angle indicators.
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.2 Slide 13
Aim
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.2 Slide 14
Method: Adjustment
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.2 Slide 15
Method (continued)
The edges of the light field are marked with radio-opaque objects
or by pricking holes with a pin through the ready pack film in the
corners of the light field.
Plastic slabs are placed on top of the film such, that the film is
positioned near zmax
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.2 Slide 16
Expected result
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.2 Slide 17
Aim
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.2 Slide 18
Method
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.2 Slide 19
Practical suggestions
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.2 Slide 20
Practical suggestions
If not, the treatment couch height and length of the front pointer
are adjusted until this condition is achieved as closely as
possible.
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.2 Slide 21
Expected result
The tip of the rod affixed to the treatment table indicates the
position of the gantry axis of rotation.
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.2 Slide 22
Aim
Note:
Whereas the collimator
and gantry rotation axis
can hardly be changed
by a user, the position
of the couch rotation axis
can indeed be adjusted.
gantry
axis
collimator
axis
treatment couch
axis
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.2 Slide 23
Method
Axis of rotation of the patient treatment couch can be found
by observing and noting the movement of the cross-hair
image on a graph paper while the gantry with the collimator
axis of rotation is pointing vertically downward.
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.2 Slide 24
Expected result
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.2 Slide 25
Aim
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.2 Slide 26
Method
Location and dimension of
the radiation isocentre
sphere can be determined
by a film using the "starshot" method.
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.2 Slide 27
Practical suggestions
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.2 Slide 28
Practical suggestions
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.2 Slide 29
Expected result
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.2 Slide 30
Method
Convenient method to
verify the accuracy of the
optical distance indicator
over the range of its readout consists of projecting
the indicator on top of a
plastic phantom with
different heights.
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.2 Slide 31
Practical suggestions
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.2 Slide 32
Expected results
Deviation of the actual height from that indicted by the optical
distance indicator must comply with the specification.
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.2 Slide 33
Method
Accuracy of the gantry
angle indicators can be
determined by using a
spirit level.
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.2 Slide 34
Practical suggestions
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.2 Slide 35
Expected results
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.2 Slide 36
Method
Collimator field size indicators can be checked by comparing
the indicated field sizes to values measured on a piece of
graph paper.
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.2 Slide 37
Practical suggestions
Graph paper is fixed to the treatment couch with the top of the
couch raised to isocentre height.
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.2 Slide 38
Expected results
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.2 Slide 39
Aim
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.2 Slide 40
Method
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.2 Slide 41
Practical suggestions
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.2 Slide 42
Expected results
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.2 Slide 43
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.3 Slide 1
Photon energy.
Photon beam uniformity.
Photon penumbra.
Electron energy.
Electron beam bremsstrahlung contamination.
Electron beam uniformity.
Electron penumbra.
Monitor characteristics.
Arc therapy.
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.3 Slide 2
1.
2.
Method used.
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.3 Slide 3
Specification
80
PDD
60
40
20
0
0
10
15
20
25
depth / cm
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.3 Slide 4
Method
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.3 Slide 5
Specification
Uniformity index.
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.3 Slide 6
Beam flatness F
obtained from the
profile in 10 cm depth:
1.1
1.0
Dmin
Dmax
0.9
Dmax Dmin
F 100
Dmax Dmin
relative dose
0.8
central area
=80%
0.7
0.6
0.5
Col 1 vs Col 2
Col 1 vs Col 2
0.4
0.3
field size
0.2
0.1
0.0
-100
-50
50
100
mm
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.3 Slide 7
Beam symmetry S
110
arealeft +arearight
90
relative dose in %
S 100
arealeft arearight
100
80
70
60
50
40
30
area left
20
area right
10
0
-100
-50
50
100
mm
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.3 Slide 8
area 90%
area 50%
area90%
UI
area50%
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.3 Slide 9
Specification
Profile at 10 cm depth
1.1
Photon penumbra is
typically defined as the
distance between the
80 % and 20 % dose
points on a transverse
beam profile measured
in water phantom at
depth of 10 cm.
1.0
0.9
0.8
1.1
0.7
1.0
0.6
0.9
0.8
0.5
0.7
0.4
0.6
0.3
0.5
0.2
0.4
0.3
0.1
0.2
0.0
-150
-100
-50
0.1
50
100
150
0.0
mm
-60
-50
-40
mm
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.3 Slide 10
Method
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.3 Slide 11
Specification
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.3 Slide 12
Method
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.3 Slide 13
Specification
Bremsstrahlung
contamination of the
electron beam is the
radiation measured
beyond the practical range
of the electrons in percent
of the maximum dose.
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.3 Slide 14
Method
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.3 Slide 15
Specification
Uniformity index
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.3 Slide 16
Note
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.3 Slide 17
Specifications
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.3 Slide 18
Methods: Linearity
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.3 Slide 19
Methods: Linearity
Integrated ionization
chamber current
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.3 Slide 20
Negative x-intercept:
more radiation is delivered than
indicated by the monitor unit setting.
Positive x-intercept:
less radiation is delivered than indicated
by the monitor unit setting.
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.3 Slide 21
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.3 Slide 22
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.3 Slide 23
Specification
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.3 Slide 24
Method
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.3.3 Slide 25
10.4 COMMISSIONING
Characteristics
Clinical use can only begin when the physicist responsible for
commissioning is satisfied that all aspects have been
completed and that the equipment and any necessary data,
etc., are safe to use on patients.
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4 Slide 1
10.4 COMMISSIONING
Commissioning of an external beam therapy device includes a
series of tasks:
5.
6.
7.
system.
Developing all dosimetry, treatment planning, and treatment
procedures.
Verifying the accuracy of these procedures.
Establishing quality control tests and procedures.
Training all personnel.
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4 Slide 2
10.4 COMMISSIONING
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4 Slide 3
10.4 COMMISSIONING
10.4.1 Photon Beam Measurements
1.
2.
3.
4.
5.
6.
7.
8.
9.
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.1 Slide 1
10.4 COMMISSIONING
10.4.1 Photon Beam Measurements: PDD
Method
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.1 Slide 2
10.4 COMMISSIONING
10.4.1 Photon Beam Measurements: PDD
Note:
Ionization chambers always provide depth-ionization curves.
Since stopping-power ratios and perturbation effects for photon beams are
almost independent of depth, relative ionization distributions can be
used in a very good approximation as relative distributions of absorbed
dose.
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.1 Slide 3
10.4 COMMISSIONING
10.4.1 Photon Beam Measurements: PDD
Method (continued)
If a cylindrical ionization
chamber is used instead,
then the effective point
of measurement ( ) of
the chamber must be
taken into account.
real
depth
0.6 rcyl
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.1 Slide 4
10.4 COMMISSIONING
10.4.1 Photon Beam Measurements: PDD
Practical suggestions
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.1 Slide 5
10.4 COMMISSIONING
10.4.1 Photon Beam Measurements: PDD
Note:
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.1 Slide 6
10.4 COMMISSIONING
10.4.1 Photon Beam Measurements: Output factors
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.1 Slide 7
10.4 COMMISSIONING
10.4.1 Photon Beam Measurements: Output factors
OF is defined as:
DP ( zmax , A,SSD, E )
OF RDF
DP ( zmax ,10,SSD, E )
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.1 Slide 8
10.4 COMMISSIONING
10.4.1 Photon Beam Measurements: Output factors
Method
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.1 Slide 9
10.4 COMMISSIONING
10.4.1 Photon Beam Measurements: Output factors
Square fields
OF
1.0
0.9
0.8
0.7
0.6
0.5
0
10
15
20
25
30
35
40
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.1 Slide 10
10.4 COMMISSIONING
10.4.1 Photon Beam Measurements: Output factors
Rectangular fields
2ab
aeq
ab
This assumption must be verified by measuring the output for a
number of rectangular fields with high and low aspect ratios.
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.1 Slide 11
10.4 COMMISSIONING
10.4.1 Photon Beam Measurements: Output factors
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.1 Slide 12
10.4 COMMISSIONING
10.4.1 Photon Beam Measurements: Output factors
Asymmetric fields
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.1 Slide 13
10.4 COMMISSIONING
10.4.1 Photon Beam Measurements: Output factors
Asymmetric fields
=
=
=
OAR(zmax,y) =
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.1 Slide 14
10.4 COMMISSIONING
10.4.1 Photon Beam Measurements: Output factors
OF
CF
SF
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.1 Slide 15
10.4 COMMISSIONING
10.4.1 Photon Beam Measurements: Output factors
Collimator scatter factor is measured in air with a buildup cap large enough to provide electronic equilibrium.
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.1 Slide 16
10.4 COMMISSIONING
10.4.1 Photon Beam Measurements: Output factors
OF CFSF
phantom scatter factor SF may be simply found by
dividing the output factor by the collimator scatter factor.
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.1 Slide 17
10.4 COMMISSIONING
10.4.1 Photon Beam Measurements: Blocking tray factors
Purpose
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.1 Slide 18
10.4 COMMISSIONING
10.4.1 Photon Beam Measurements: Blocking tray factors
Method
Ratio of the ionization chamber signal with the tray in the beam
to the signal without the tray is the blocking tray transmission
factor.
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.1 Slide 19
10.4 COMMISSIONING
10.4.1 Photon Beam Measurements: Multileaf collimators
Purpose
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.1 Slide 20
10.4 COMMISSIONING
10.4.1 Photon Beam Measurements: Multileaf collimators
Penumbra
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.1 Slide 21
10.4 COMMISSIONING
10.4.1 Photon Beam Measurements: Multileaf collimators
Output factor
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.1 Slide 22
10.4 COMMISSIONING
10.4.1 Photon Beam Measurements: Multileaf collimators
Leakage
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.1 Slide 23
10.4 COMMISSIONING
10.4.1 Photon Beam Measurements: Multileaf collimators
Leakage
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.1 Slide 24
10.4 COMMISSIONING
10.4.1 Photon Beam Measurements: Wedge transmission factors
Specification
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.1 Slide 25
10.4 COMMISSIONING
10.4.1 Photon Beam Measurements: Wedge transmission factors
relative dose
250
d=1.5 cm y-axis
200
d= 5cm x-axis
d=5 cm y-axis
150
d=10 cm x-axis
d=10 cm y-axis
100
d=20 cm x-axis
50
d=20 cm y-axis
0
-200
-100
100
200
off-axis distance, mm
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.1 Slide 26
10.4 COMMISSIONING
10.4.1 Photon Beam Measurements: Wedge transmission factors
Method
Rotation of the wedge itself which reveals whether or not the side
rails are symmetrically positioned about the collimator axis of
rotation.
Rotation of the collimator which verifies that the ionization
chamber is positioned on the collimator axis of rotation.
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.1 Slide 27
10.4 COMMISSIONING
10.4.1 Photon Beam Measurements: Wedge transmission factors
Otherwise:
IAEA
WF
WF0 WF180
2
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.1 Slide 28
10.4 COMMISSIONING
10.4.1 Photon Beam Measurements: Dynamic wedges
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.1 Slide 29
10.4 COMMISSIONING
10.4.1 Photon Beam Measurements: Dynamic wedges
Method
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.1 Slide 30
10.4 COMMISSIONING
10.4.1 Photon Beam Measurements: Dynamic wedges
Note:
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.1 Slide 31
10.4 COMMISSIONING
10.4.1 Photon Beam Measurements: Transverse beam profiles
Purpose
For calculation of 2-D and 3-D dose distributions,
off-axis dose profiles are required in conjunction with central
axis data.
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.1 Slide 32
10.4 COMMISSIONING
10.4.1 Photon Beam Measurements: Transverse beam profiles
Method
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.1 Slide 33
10.4 COMMISSIONING
10.4.1 Photon Beam Measurements: Transverse beam profiles
10MV, TPS *******, OPEN, 10x10 cm2, d = 200 mm, error bars 1%, 2mm
100
100
80
80
Relative dose, %
Relative dose, %
10MV, TPS ******, OPEN, 10x10 cm2, d = 50 mm, error bars 1%, 2mm
60
40
60
40
20
20
0
-100
-80
-60
-40
-20
20
Off axis, mm
40
60
80
100
-100
-80
-60
-40
-20
TPS
TPS
IAEA
20
40
60
80
Off axis, mm
Phantom
Phantom
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.1 Slide 34
100
10.4 COMMISSIONING
10.4.1 Photon Beam Measurements: Entrance/interface dose
Purpose
Examples:
Entrance dose between the patient surface and zmax,
Interfaces at small air cavities such as the nasopharynx,
At the exit surface of the patient,
At bonetissue interfaces
Interfaces between a metallic prosthesis and tissue.
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.1 Slide 35
10.4 COMMISSIONING
10.4.1 Photon Beam Measurements: Entrance/interface dose
Method
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.1 Slide 36
10.4 COMMISSIONING
10.4.1 Photon Beam Measurements: Entrance/interface dose
Method (continued)
constant SSD
1. measurement
2. measurement
3. measurement
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.1 Slide 37
10.4 COMMISSIONING
10.4.1 Photon Beam Measurements: Entrance/interface dose
Method (continued)
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.1 Slide 38
10.4 COMMISSIONING
10.4.1 Photon Beam Measurements: Virtual source position
Purpose
IAEA
Da
Db
Db Da fa / fb
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.1 Slide 39
10.4 COMMISSIONING
10.4.1 Photon Beam Measurements: Virtual source position
Method
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.1 Slide 40
10.4 COMMISSIONING
10.4.1 Photon Beam Measurements
Method (continued)
If the straight line passes through the origin the virtual and
nominal source positions are the same.
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.1 Slide 41
10.4 COMMISSIONING
10.4.2 Electron Beam Measurements
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.2 Slide 1
10.4 COMMISSIONING
10.4.2 Electron Beam Measurements: PDD
Method
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.2 Slide 2
10.4 COMMISSIONING
10.4.2 Electron Beam Measurements: PDD
Note:
Depth-ionization curve
of electrons differs from
the depth-dose curve by
water-to-air stopping
power ratio.
100
18 MeV
the
depthdose
curve
PDD
80
60
depthionization
curve
40
20
0
0
10
12
14
16
depth / cm
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.2 Slide 3
10.4 COMMISSIONING
10.4.2 Electron Beam Measurements: PDD
Note (cont.)
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.2 Slide 4
10.4 COMMISSIONING
10.4.2 Electron Beam Measurements: PDD
Measurement of R50
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.2 Slide 5
10.4 COMMISSIONING
10.4.2 Electron Beam Measurements: PDD
Practical suggestions
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.2 Slide 6
10.4 COMMISSIONING
10.4.2 Electron Beam Measurements: PDD
When using an ionization chamber, the measured quantity is the halfvalue of the depth-ionization distribution in water, R50,ion. This is the
depth in water (in g/cm2) at which the ionization current is 50 % of its
maximum value.
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.2 Slide 7
10.4 COMMISSIONING
10.4.2 Electron Beam Measurements: PDD
For electron beam qualities with R50 4 g/cm2 (i.e., for electron
energies larger than 10 MeV) a cylindrical chamber may be used.
Nominal
depth
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.2 Slide 8
10.4 COMMISSIONING
10.4.2 Electron Beam Measurements: PDD
For beam qualities R50 < 4 g/cm2 (i.e., for electron energies
smaller than 10 MeV) a plastic phantom may be used.
cpl
0.949
PMMA
0.941
White polystyrene
0.922
(Table is from the IAEA TRS 398 dosimetry protocol)
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.2 Slide 9
10.4 COMMISSIONING
10.4.2 Electron Beam Measurements: PDD
For depths beyond zref,pl it is acceptable to use the value for hpl at zref,pl
derived from the Table below.
hpl
1.008
PMMA
1.009
White polystyrene
1.019
Table from the IAEA TRS 398 dosimetry protocol .
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.2 Slide 10
10.4 COMMISSIONING
10.4.2 Electron Beam Measurements: PDD
Practical suggestion
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.2 Slide 11
10.4 COMMISSIONING
10.4.2 Electron Beam Measurements: PDD
Practical suggestion
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.2 Slide 12
10.4 COMMISSIONING
10.4.2 Electron Beam Measurements: Output factors
9 MeV electrons
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.2 Slide 13
10.4 COMMISSIONING
10.4.2 Electron Beam Measurements: Output factors
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.2 Slide 14
10.4 COMMISSIONING
10.4.2 Electron Beam Measurements: Output factors
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.2 Slide 15
10.4 COMMISSIONING
10.4.2 Electron Beam Measurements: Output factors
Cones or electron
collimators are
available in a limited
number of square
fields typically
55 cm2 to
2525 cm2 in 5 cm
increments.
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.2 Slide 16
10.4 COMMISSIONING
10.4.2 Electron Beam Measurements: Output factors
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.2 Slide 17
10.4 COMMISSIONING
10.4.2 Electron Beam Measurements: Output factors
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.2 Slide 18
10.4 COMMISSIONING
10.4.2 Electron Beam Measurements: Output factors
Irregularly shaped
electron fields are
formed by placing
metal cutouts of
lead or low melting
point alloy in the
end of the cone
nearest the patient.
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.2 Slide 19
10.4 COMMISSIONING
10.4.2 Electron Beam Measurements: Output factors
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.2 Slide 20
10.4 COMMISSIONING
10.4.2 Electron Beam Measurements: Output factors
Output factor is the ratio of dose at zmax for the small field
to dose at zmax for the 1010 cm2 field.
Since zmax shifts toward the surface for electron fields with
dimensions smaller than the range of the electrons, it must
be determined for each small field size when measuring
output factors.
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.2 Slide 21
10.4 COMMISSIONING
10.4.2 Electron Beam Measurements: Output factors
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.2 Slide 22
10.4 COMMISSIONING
10.4.2 Electron Beam Measurements: Output factors
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.2 Slide 23
10.4 COMMISSIONING
10.4.2 Electron Beam Measurements: Output factors
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.2 Slide 24
10.4 COMMISSIONING
10.4.2 Electron Beam Measurements: Transverse beam profiles
Method using a water phantom
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.2 Slide 25
10.4 COMMISSIONING
10.4.2 Electron Beam Measurements: Transverse beam profiles
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.2 Slide 26
10.4 COMMISSIONING
10.4.2 Electron Beam Measurements: Extended SSD applications
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.2 Slide 27
10.4 COMMISSIONING
10.4.2 Electron Beam Measurements: Extended SSD applications
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.2 Slide 28
10.4 COMMISSIONING
10.4.2 Electron Beam Measurements: Extended SSD applications
PDD changes
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.2 Slide 29
10.4 COMMISSIONING
10.4.2 Electron Beam Measurements: Extended SSD applications
Penumbra changes
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.2 Slide 30
10.4 COMMISSIONING
10.4.2 Electron Beam Measurements: Extended SSD applications
Penumbra changes
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.4.2 Slide 31
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.5 Slide 1
IAEA
Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 10.5 Slide 2