Total Chrom Tutorial
Total Chrom Tutorial
Part Number
N515-6021
Release
B
Publication Date
November 2004
Any comments about the documentation for this product should be addressed to:
User Assistance
PerkinElmer
710 Bridgeport Avenue
Shelton, CT 06484-4794
Or emailed to: [email protected]
Notices
The information contained in this document is subject to change without notice.
Except as specifically set forth in its terms and conditions of sale, PerkinElmer makes
no warranty of any kind with regard to this document, including, but not limited to,
the implied warranties of merchantability and fitness for a particular purpose.
PerkinElmer shall not be liable for errors contained herein for incidental consequential
damages in connection with furnishing, performance or use of this material.
Table of Contents
Welcome
Welcome to PerkinElmer's Line of Chromatography Software Products................................... 1
Where Do I Begin?
Register with Software Support.................................................................................................. 2
Users New to Computers ............................................................................................................ 2
Users New to Microsoft Windows.............................................................................................. 3
About Installation
How to Prepare for Installation................................................................................................... 5
What If I Install TotalChrom Myself and Need Help ................................................................. 6
Table of Contents
Tutorial Introduction
About the TotalChrom Tutorial.................................................................................................21
Before You Begin......................................................................................................................22
Checklist.............................................................................................................................22
Other Helpful Hints............................................................................................................23
Tutorial Part 1
The TotalChrom Navigator .......................................................................................................24
Collecting Your First Data File .................................................................................................26
Building a Quick Method ...................................................................................................30
Building a Vial List ............................................................................................................35
Injecting your Standard and Acquiring Data ......................................................................38
Viewing the Real-Time Plot...............................................................................................39
Zooming In on the Real-Time Plot..............................................................................40
Rescaling a Plot...........................................................................................................41
The Default Report and Plot........................................................................................42
Tutorial Part 2
Optimizing a Method Using Graphic Editing............................................................................43
Opening the Graphic Method Editor ..................................................................................44
Changing the Plot Display..................................................................................................47
Fine Tuning Peak Detection ...............................................................................................50
Creating a Component List.................................................................................................55
Using the Method Editor ...........................................................................................................59
Setting Calibration Parameters ...........................................................................................59
Performing a Manual Calibration.......................................................................................63
Saving a Method.................................................................................................................68
Building a Report Format ..........................................................................................................69
Building a Sequence ..................................................................................................................72
Using the Sequence Template ............................................................................................72
Using the Sequence Spreadsheet ........................................................................................75
Batch Reprocessing a Sequence File .........................................................................................80
Viewing Results using Reprocess Results.................................................................................82
Congratulations! You have successfully completed the TotalChrom Tutorial...................90
ii
Welcome
Where Do I Begin?
About Installation
Where Do I Begin?
TotalChrom Workstation customers: You must fill out your permanent License
PAK application and send it to PerkinElmer by following the directions on the
form. If you do not currently have your permanent License PAKs from the
TotalChrom PAK Administrator, you are probably not registered with us.
TotalChrom Turnkey customers: Turnkey systems are shipped with the software
and permanent PAK installed on the hard drive.
Note: Turnkey customers should only follow the installation instruction card
provided with their system and skip all sections in this tutorial that are related
to installation.
Where Do I Begin?
are available at local community colleges, technical schools, and even local software
stores.
If you did not purchase a computer from PerkinElmer, make sure the
computer used for TotalChrom is configured according to the minimum
TotalChrom requirements (refer to the product release notes). We
cannot support our products on any system configuration that does not meet
these requirements.
2.
3.
About Installation
Note: Turnkey systems are shipped with the software installed on the hard drive. Turnkey
customers should only follow the installation instruction card provided with their system
and skip all sections in this tutorial that are related to installation.
Before your computer can collect data, you need to do the following:
1.
2.
3.
LINK instrument control users may have additional installation steps to prepare third
party analytical instrumentation for LINK instrument control. In some cases involving
non-PerkinElmer instruments, you may have to order and install additional boards.
Make certain you have consulted with your PerkinElmer sales/service representative
regarding connections to third party vendor instruments so these pre-installation
requirements can be met.
If you would like to schedule a service representative to perform this installation,
contact your local PerkinElmer Service Center for instructions. In the U.S., contact
the PerkinElmer Customer Support Center at 1-800-762-4000.
About Installation
Unpack all equipment, both hardware and software. Compare the items that
arrived against your sales order to make sure you received all items.
Cables that need to be routed through walls, ceilings, etc. should be in place by
the time of installation.
Provide the required bench space for the data system. Make certain there is
adequate electrical power and that the space is clean, clear, and safe.
Place the A/D interfaces and/or LINK modules in the desired locations.
The detector leads and remote-start leads for your PerkinElmer instruments are
connected by the PerkinElmer service engineer. For non-PerkinElmer
instruments, please have these signals connected with twisted pair cabling for
connection to the 900 series interface. We cannot open up other vendors
equipment.
Have available a copy of your computer and/or operating system manuals. Also,
have the manuals and disks for third party software, detectors, computer,
expansion cards, and peripheral equipment.
Make sure you have the required cables. These include analog lead lines, printer
cables, IEEE/serial cables, remote lead lines, etc.
About Installation
If you purchased LINK instrument control for analytical instruments other than
PerkinElmer, you may need to make changes to this equipment in advance of
your installation visit. Your service engineer or sales representative will have
details on these changes.
Method Creation
Know how to create a basic method and how to use the naming convention
for files created by your system. Make sure you understand how to adjust
the parameters in each method, and what the consequence of that adjustment
is, without trying to apply those to your application. All parameters are
defined in the users guide, and we provide example method files with the
software.
Sequences
Make sure you understand the concept behind sequences for TotalChrom.
Practice by making and using several sequences with the supplied data files
to analyze the effect of changing parameters. The sequence is the main
blueprint for TotalChrom. It tells the system what to do next, with what
method, where the method is located, and what adjustments, if any, are
necessary to arrive at the final result. The sequence does exactly what you
tell it to do, whether it is what you intended or not. If you are not sure of an
entry, check the help files or users guide rather then making a guess.
Graphics
TotalChrom graphics routines provide countless ways to view, edit, and
manipulate data. The ability to view raw data down to a single sampled data
point, or to select a small portion of the entire chromatogram for expansion,
is probably foreign to former users of chart recorders or simple integrators.
Use the sample data files with this tutorial to practice the variety of graphics
options your package allows. These files can be restored from your
TotalChrom CD so you can experiment with them to your hearts content. It
is best to try all those features now and become proficient with the test data,
so when the real data comes along, you will be prepared to work with it
quickly. It is a good idea to copy these over after each practice session so
the next user has fresh files when they start learning.
10
Does your system at least meet the MINIMUM requirements as stated in the
TotalChrom Release Notes?
If it does not, you WILL have problems. These truly are the basic
requirements for the version of software.
11
Note: LINK customers should be aware that any instrument control errors displayed
in TotalChrom are actually returned from the analytical instrument. They are usually
explained in the manual that comes with your instrument (GC or LC) and not in the
TotalChrom or LINK documentation.
12
b) Can the problem be duplicated with the test data files that came with the
system? If not, this may point to a problem with the new files you created.
c)
Are you on a network? See if the problem still exists when you are
disconnected from the network to rule out a network problem.
Remember that system operation is dependent on everything communicating
in harmony. In the case of data acquisition software, the timing and number
of these communications are incredibly complex and demanding on the
system, involving both internal (computer, disk drive, printer) and external
(instrument control, IEEE and serial communications) devices.
Occasionally, something interferes with this communication and simply
cannot be traced. Perhaps it is the result of electrical disturbances in your
lines from voltage brownouts or surges; a strong magnetic field created by
equipment in the next lab; a loose connection; a failing piece of hardware; or
just the night cleaning crew plugging their vacuum cleaners into the same
circuit as the system analyzing your overnight run. If the problem seems to
have disappeared, make a note of it just in case, but you can generally
proceed without concern.
It is not impossible that a software file gets corrupted somehow, but before
you reinstall the software, check out the obvious culprits first. These can
often be diagnosed quickly by rebooting your system with a blank system
disk containing basic autoexec and config files. If the problem disappears
with the use of this system disk, look to the startup files for a potential
conflict.
Hardware on the other hand does break. Even worse, it can break with an
insidious style that drives users and support specialists crazy. It can work
fine one day and create havoc the next. Often we cannot catch this until it
finally breaks for good. Every system should have a good diagnostic
software tool that puts its hardware through the paces so hardware faults can
13
be determined or ruled out. Do not forget the obvious here, or maybe not so
obvious, too.
Did someone wrap that extra length of cable around the base of an old
centrifuge to clean up the lab counter top?
More than half of all the calls to Software Support have nothing to do with
the actual TotalChrom software, but are the result of problems with the
operating system (DOS), Windows, network configuration, hardware system
configuration, or device drivers for printers and other devices.
14
Have your Serial Number, Software Licenses and/or your original software
media available.
You will need these to access most centralized phone support programs and
services.
Know the version and service packs of all relevant software packages.
Have available the make and model numbers of your computer and
associated hardware.
Note: DO NOT make calls on behalf of the end user having the problem unless you
fully understand both the problem and the system. These calls are seldom resolved
quickly because of the unfamiliarity of the caller with the situation.
Contacting PerkinElmer
Customer Support
Outside the US
15
16
What Is an Upgrade?
There are two types of upgrades Product Upgrades and Software Upgrades.
Product Upgrades exist so customers have an option to buy into the PerkinElmer line
of products with minimal features, then migrate up to more feature-packed products at
a reduced price as conditions permit. Upgrading is usually associated with the
strategic needs of a growing lab as determined by the lab manager in consultation
with a PerkinElmer sales engineer.
For example, you may have bought a Product Upgrade to TotalChrom from our
Turbochrom Professional chromatography software. We only mention product
upgrades here to help distinguish them from software upgrades.
Software Upgrades are major releases of an established product that incorporate
significant enhancements to the software, or take advantage of a major new platform,
operating system, or user interface. TotalChrom v6.2 is a Software Upgrade from the
Turbochrom Professional v4 revision.
17
Summary
We hope this introduction has been helpful. All of us in Software Support want your
experiences with our products and the entire PerkinElmer team to be successful for
many years. Please let us know how we can be of more help, whether it is a feature
you would like to see in the software, or a support program you would like to see us
implement in the future.
19
20
Tutorial Introduction
Creating a report format file for printing the results of the analysis
A report format file contains the specifications for the analysis report and
some of the reports formatting characteristics.
This tutorial does not cover every feature in TotalChrom. It follows a specific path,
while TotalChroms flexibility and the variety of instruments it supports offer many
paths.
For a complete description of all TotalChrom functions and operations, see the
TotalChrom Users Guide.
21
Tutorial Introduction
For electronic information, TotalChrom has online help that you can access at any
time. For information about online help, see The Help System in Chapter 1 of the
TotalChrom Users Guide.
Checklist
Complete the following preparations before you start the tutorial:
22
Make sure that at least one interface (Model 900, Model 600 LINK, SoftLINK or
NCI) is connected to a chromatograph.
Ensure that the above directory has RWXD privileges for both the TCProcess
account as well as the users logon account.
Tutorial Introduction
Follow the directory and file name instructions carefully. You will save everything
in the C:\PenExe\...\tutorial directory then delete all the files after you have
completed the tutorial.
When you collect data in Part 1 of the tutorial, you must provide values that are
appropriate for the analysis that you are performing and for the equipment you
are using.
The defaults on your system may be different from what the tutorial describes or
illustrates. The tutorial may direct you to enter or select something that is already
present or selected, or where your instruments do not provide a choice.
When there are multiple ways to do something, the choices are listed and
separated by OR.
23
Tutorial Part 1
Click the Start button and point to Programs, then click TotalChrom
Workstation Ver6.3 to display the Links window. Double-click on the
TCNav icon in this window.
Note: You must logon to a user account that has manager permissions enabled. If you
have any questions, consult your system manager.
24
Tutorial Part 1
2.
Enter your Logon User Name and your Password. Then choose OK.
The Navigator window appears:
Title Bar
Menu Bar
Instrument
Selection
Panel
Status Bar
Note: This screen shows that the TCPublisher reporting option was selected. To
change to Review and Approve, select TotalChrom Review & Approve in the
Reporting Options dialog. This is displayed by selecting Reporting from the
Build>Configuration>User> menu.
We will use the instrument selection panel and several function buttons in the tutorial.
25
Tutorial Part 1
Acquire data
Adjust the real-time plot of the data points as they are acquired
26
Select the instrument you want to use for data collection from the instrument
selection panel in the Navigator (you will use this instrument throughout the
entire tutorial).
Tutorial Part 1
2.
In the Setup Instrument dialog box, select Quick start as the Setup Type.
4.
Click the folder icon next to the Raw File Path text box as shown above.
27
Tutorial Part 1
5.
6.
Perform the same procedure for the Result File Path text box.
7.
To enter a Base File Name, click on the button to the right of the text box.
Tutorial Part 1
8.
In the Default Base Names dialog box, select NONE for the Separator and
type TEST into the Base File Name field.
This dialog discusses Tokenized filenames (a new feature to version 6.2). Please
consult your Users Guide for a more detailed description about this option.
All files related to this quick method that we are creating raw, result,
method, and sequence files will use this base file name.
TotalChrom will append a three-digit run or cycle number to this name
(based on the number of digits specified for the suffix under the Default
Base Names dialog) when it creates the raw and result file names.
9.
10. Leave the Suppress processing and Suppress reports/plots check boxes
unchecked as shown in the above screen.
We want to process the data file immediately after it is collected, and we
want a printed report and plot at the end of the run.
11. Do not click the OK button yet.
You have completed the entries in the Setup dialog box. The next
step is to build a quick method.
For additional information on the Setup dialog box, see Acquiring and Viewing
Data in Chapter 12 of the TotalChrom Users Guide.
29
Tutorial Part 1
In the Setup Instrument dialog box, click on the button to the left of Quick
method.
The Quick Method Editor window opens, with the Documentation dialog box.
This is the first in the series of dialog boxes that will be displayed
automatically.
2.
You may enter any descriptive information about the method in this window
or click on NEXT to leave it blank. Do Not Select Audit Trail.
The next window to appear is the Instrument Notes dialog box.
30
Tutorial Part 1
3.
You may enter any instrument specific information into this dialog or choose
from the template drop-down menu for one of the pre-designed instrument
templates. You may also choose to leave it blank, if so, click NEXT to display
the Data Acquisition dialog box.
The information displayed on the Data Channels tab varies depending on the
instrument you selected at the beginning of this procedure. This is shown in
the following two figures:
31
Tutorial Part 1
4.
Under Data Channel, select the channel (A, B, or Dual) that matches the way
your detector output is connected to the interface.
5.
6.
If you know how wide your narrowest peak is at its baseline, select By peak
width at Base and enter the number of seconds in the text box. TotalChrom
will use this to calculate the optimum sampling rate for that peak.
OR
Select By sampling rate and select or enter a rate (depending on the
instrument).
Suggested sampling rates are:
Capillary GC
=
HPLC or IC
=
Packed column GC =
32
20 pts/sec
5 pts/sec
2 pts/sec
Tutorial Part 1
7.
We want to save the complete run rather than using delay and run times to
store only a part of it.
For a LINK-controlled instrument, select Store all data from run.
OR
For a 900/NCI Series Interface, leave Delay time set to 0, and enter a
Run time that corresponds to the length of your run.
8.
Choose NEXT to display the first in a series of dialog boxes for the
instrument you selected.
9.
Complete each series of dialog boxes using the appropriate values for your
sample and equipment. (Since the people following this tutorial will be using
a variety of equipment, we cannot anticipate the specific entries.)
Note: Press F1 when a dialog box is open to display the online Help for that box.
The dialog boxes displayed in the rest of this series depend on the instrument
you selected and on the instrument modules you configured. We will not
enter any processing parameters for this quick method. It is often easier to
set these parameters after the fact using the Graphic Method Editor. We will
use that function in Part 2 of the tutorial.
When you reach the Components dialog box, which is the last screen in the
series for all instrument types, click on Finish.
33
Tutorial Part 1
When you complete Quick Method dialog boxes, the Method Summary,
Component List, or a combination of the two screens will be displayed
showing the initial parameters you have set.
2.
Choose Exit from the File menu to return to the Setup Instrument dialog box.
For more information about quick methods, see Using Quick Method
in Chapter 12 of the TotalChrom Users Guide.
Although it is not required for all instrument configurations, the next task we
will perform in this tutorial is building a vial list to specify the samples to be
injected.
34
Tutorial Part 1
In the Setup Instrument dialog box, click on the Vial list button.
This opens the Vial List window with the Sequence Information window:
35
Tutorial Part 1
2.
Choose Global Parameters from the Change menu to display the Global
Parameters dialog box.
b.
Select the type of tray installed in your autosampler from the Tray dropdown list.
c.
Click on OK.
3.
In the Sequence Information window, select the first cell in the Vial column
by clicking in it.
4.
5.
36
Do not enter any more vials. Instead, delete the remaining rows, 2 through 5.
Tutorial Part 1
6.
7.
Choose Delete from the Edit menu. Row 2 is deleted and rows 3-5 move up
in the spreadsheet.
8.
You have completed the vial list and are ready to save it and start the run.
2.
Choose Exit from the File menu to close the vial list window and return to
the Setup Instrument dialog box.
3.
37
Tutorial Part 1
OR
Tutorial Part 1
In the Navigator, click on the Real-Time Plot button to open the Real-Time
Plot window.
The parameters for the initial display come from the Real-Time Plot Scale
dialog box that you completed in the quick method by accepting its defaults.
There are a number of things that you can do from the Real-Time Plot
window. We will experiment with changing the display from the Options
menu. Display changes are temporary, affecting the current display only;
they do not change the settings in the method and they do not affect data
acquisition.
39
Tutorial Part 1
2.
Choose Zoom In (for the channel you want to view) to enlarge the plot.
3.
Choose Zoom In again until the view is as large as you want it to be.
4.
Choose Zoom Out until the plot view is restored to its original size.
Note: You can zoom in and out without displaying the Options menu by using the
keyboard shortcuts F5 and F6 while the real-time plot is displayed.
Zoom In and Zoom Out act on the full-scale value for the voltage axis,
halving and doubling it respectively. Our next step is to change the voltage
axis value explicitly from the Rescale dialog box.
40
Tutorial Part 1
Rescaling a Plot
The Rescale command in the Options menu lets you change the scaling parameters of
the plot while data are being collected.
In the Real-Time Plot window, choose Rescale from the Options menu to
open the Rescale dialog box.
Under Voltage Axis, the value for Full Scale is the value that is currently
being used for the display. That value may be (a) from the method, (b) as
altered by zooming, or (c) as you specifically changed it in this dialog box.
2.
3.
4.
To close the Real-Time Plot window, choose Exit from the Instrument menu
to return to the Navigator.
When the run is completed (or stopped), the default report and annotated
replot will be printed.
41
Tutorial Part 1
Once you have acquired a data file, you can use the stored data file to set up better
processing parameters. You do this by using the Graphic Method Editor. We will use
this function in Part 2 of the tutorial.
First, lets look at the printed output from our data analysis.
The Default Report and Plot
If your printer was attached and turned on, a report and plot were printed after data
analysis. Lets take a quick look at these printouts. Both were formatted using default
settings.
The top part of the report contains information from the method, such as data channel
and run time. These items, and the information on the report itself, are configurable.
We will customize a report in Part 2 of the tutorial.
The plot is also headed by identifying information. The way the plot is printed is also
configurable in the method (it has no relationship to the scale of the real-time plot.)
Note: We will not reuse any of the files created in this part of the tutorial. In Part 2
we will use sample files that are shipped with TotalChrom.
Note: Every step in the rest of the tutorial is built on, and is a continuation of, the
previous step. Therefore, you need to complete it as a whole.
42
Tutorial Part 2
A method contains the parameters that determine how TotalChrom analyzes raw data.
With the Graphic Method Editor, you can adjust those method parameters and see the
effects of the change immediately.
Viewing the data in graphic form helps you determine if peaks are being detected
correctly, if baselines are being placed properly, and if expected retention times and
search windows for components are identifying peaks accurately.
To edit a method in the Graphic Method Editor, you specify a raw data file that
represents a chromatogram of the analysis that you want to optimize. TotalChrom
processes the raw data with the method that is referenced in the raw file header and
then displays the chromatogram, showing the resulting integration and peak
identification.
For more information about the Graphic Method Editor, see Chapter 10 in the
TotalChrom Users Guide.
43
Tutorial Part 2
The Graphic Method Editor window opens as well as a file selection dialog
box that displays a list of raw data files.
2.
44
Select the Examples directory from the TotalChrom Quick Paths dropdown list or from the list of directories.
Tutorial Part 2
3.
4.
Choose Open.
The default display shows a working chromatogram in the main work area of
the window and a reference chromatogram above it. The name of the raw
data file appears in the status bar and the name of the method used to process
the data appears in the title bar.
45
Tutorial Part 2
In the following steps, we will explore some of the ways that you can
optimize a method to achieve this kind of result.
Note: For this tutorial, we will create an unsatisfactory method so that we have
something to optimize. The step below is for that purpose; it will most likely not be
your normal procedure.
Before we change any of the method parameters, we will experiment with changing
the way the chromatographic information is displayed.
46
Tutorial Part 2
2.
3.
Move the cursor (arrow) into the highlighted area in the reference
chromatogram.
47
Tutorial Part 2
The mouse pointer changes to a thick crosshair cursor. This indicates that
you may move the selection rectangle.
4.
Hold down the left mouse button and drag the selection rectangle to another
part of the reference chromatogram.
5.
6.
48
Position the cursor anywhere in the working chromatogram, then click the
right button to restore the chromatogram to its original scale.
Tutorial Part 2
If you moved the rectangle more than once, click the right mouse button
until the chromatogram returns to its original scale. A message will be
displayed when there are no more expansions.
Next we will change the display by directly altering the values that
define it. We want to see only the area before the first peak to
have a detailed view of the baseline noise so we can begin to
optimize the method.
With the chromatogram back to its original scale showing the entire run,
choose Rescale Plot from the Display menu.
2.
3.
a.
b.
Change Offset to 0.
c.
49
Tutorial Part 2
For additional information on changing the display, see Changing Display Options
in Chapter 10 of the TotalChrom Users Guide.
50
Tutorial Part 2
2.
Select the entire baseline by holding down the left mouse button and
dragging the mouse to outline the area.
3.
Release the mouse button to display the Noise/Area Threshold dialog box.
Click on OK to reprocess the data and redraw the plot using the suggested
thresholds.
51
Tutorial Part 2
As shown in the example below, the new noise and area threshold values
have eliminated the spurious peaks.
6.
Click the right mouse button to return the chromatogram to the previous
level of expansion (full size chromatogram).
The number of data points across a peak influences peak detection. If you have too
many points across the peak, you can either decrease the sampling rate for future runs
(you cannot change the sampling rate for data that is already collected) or increase the
bunching factor to filter out unneeded data from existing files.
52
Tutorial Part 2
2.
Outline the width of the first peak (do not select the tailing portion) by
holding down the left mouse button and dragging the mouse.
3.
With the Bunching Factor option selected, choose OK to reprocess the data
and redraw the chromatogram using the new bunching factor.
53
Tutorial Part 2
You have now optimized the method for peak detection and
integration. The next step is to have TotalChrom assign names to
some of the peaks and calculate the amount (or concentration) of
the peaks. To do this you must create a component list.
54
Tutorial Part 2
The plot is narrowed to make room for the Edit Components dialog box:
2.
3.
Select the Ref. check box to have this component act as a reference peak for
other components.
The reference peak is used to adjust the expected retention times during
component identification.
4.
Select the ISTD check box to have this component act as an internal standard
for other components.
55
Tutorial Part 2
The internal standard adjusts the responses for any variation in the injection
volume.
5.
6.
Change the entries for both Absolute Window and Relative Window to 5.
The total search window is the sum of the absolute and relative windows,
applied to either side of the expected retention time of the component.
7.
8.
Choose Prev to move the selection rectangle to peak 4 (the peak preceding
peak 5).
Note: To save component information, you MUST choose Next, Prev, or Return. If
you move to another peak by clicking on it instead, your entries are not saved. (This,
however, is a way to discard entries that you do not want to save.)
56
Tutorial Part 2
Your screen should now look like the example below, with the component
name below peak 5, and with peak 4 selected but not yet defined.
9.
10. From both the Reference and ISTD drop-down lists, select Chlorotoluene.
11. Change the Absolute Window entry to 5 (leave Relative Window set to 3).
12. Select the Find Tallest Peak in Window check box.
Your dialog box should look like the following example:
13. Choose Prev to move the selection rectangle to 3 (the peak preceding peak 4).
57
Tutorial Part 2
Your chromatogram is now labeled with the three components you just
created.
58
Tutorial Part 2
You use the Method Editor to develop a complete method. It includes the parameters
that are not available through the Graphic Method Editor or when you develop a
quick method.
Choose Method from the Other menu to switch to the Method Editor and
display the summary information for each section of the method.
59
Tutorial Part 2
2.
The names of the components that we want to identify in the analysis are on
the left side of the window. Information for the selected component is on the
right side. We will edit this existing information to add a calibration level for
each component.
60
Tutorial Part 2
3.
Note: The entries below are TotalChroms initial defaults. The default values may be
different on your system.
4.
5.
6.
7.
8.
61
Tutorial Part 2
9.
Click in the Level text box and enter Level 1 for the level name.
It is recommended to enter an alphanumeric level name in this box.
62
Tutorial Part 2
13. Make the same entries for 1,2-Dichloroethene that you made for Chloroform:
Calibration Type = Use Curve
Response = Area
Curve Fit Type = Point to Point
Scaling/Weighting = None
Origin = Include
Calibration level = Level 1
Calibration amount = 30
Delete the default calibration level entries, if any
63
Tutorial Part 2
2.
Then click Open. This will be the file that we define as a calibration standard.
64
Tutorial Part 2
3.
Select Level 1, from the drop-down menu, in the Calibration Parameters Level box.
This is the calibration level of the standard file.
4.
5.
Choose Add to add the calibration entry to the Filename/Level/Cal Type list
on the bottom of the dialog box.
If you had more levels to calibrate at this time, you would select the next
file, level, and type combination and add it to the list.
6.
Select the Identify peaks before calibrating check box so that the peaks in
the result file will be identified before calibration.
Your screen should look like the example below.
7.
65
Tutorial Part 2
8.
To view the updated calibration, return to the Calibration tab by selecting Edit
Component from the Components menu, then click on the Calibration tab.
Notice that the area has been calculated and is displayed at the bottom right
of the window.
9.
10. To view the calibration curve, you can Choose Component List from the
Window menu to switch from the Method Summary screen.
66
Tutorial Part 2
You can also switch between Method Summary and Component List by
minimizing and maximizing the windows in the Method Editor, by clicking
on the appropriate icons in the upper right corner of the window.
11. To return to the Method Summary window, choose Window and select
Method Summary.
Although there are many ways that you can further customize a
method, this completes what we will do in the tutorial. Our next
step is to save what we have done.
67
Tutorial Part 2
Saving a Method
2.
When the Description dialog box opens, enter any information that you want
to store with the method file. Do not select Audit trail.
3.
Choose OK.
3.
You have finished optimizing this method and have saved it for
future use.
The next step is to create a format for the report that will be
printed after an analysis that uses this method.
68
Tutorial Part 2
With the Method Editor window still displayed, choose Report Format
Editor from the Other menu.
Since there is no report format with this name, TotalChrom uses the name of
the method that you have open with default values to set up a new report
format.
3.
In the Title dialog box, change the default title to TUTORIAL REPORT.
4.
69
Tutorial Part 2
You can display most of the edit dialogs by clicking directly on the desired
area of change (for example, click on TUTORIAL REPORT to display the
Title dialog box).
Next well add a column to the report.
5.
To see a list of all the options for column headings, choose Report to display
the menu.
6.
7.
8.
Choose Insert/Add.
A column for Component Name is inserted before the Area column. The font
becomes smaller so that the report fits on the screen, and a compressed font
would be used when this report is printed.
70
Tutorial Part 2
9.
10. In the Adjusted Amount dialog box, change the column number from 10 to 7.
11. Choose Replace.
Adjusted Amount is added to the report as column 7, in place of Normalized
Area.
The last change we will make before saving the file is to delete a
column from the report.
71
Tutorial Part 2
Notice that with fewer column headings the font returns to its original size.
Building a Sequence
Sequences are used to control acquisition and analysis of chromatographic data
produced by a series of injections.
For the purpose of this tutorial, we will build a sequence that we will use to reprocess
stored data. To do this, we will use the sequence template and the sequence spreadsheet.
Note: In most cases, an .idx file would be used for reprocessing data. We will discuss
.idx files, and their use, at the end of this section.
72
With the Method Editor window still displayed, choose Sequence Editor
from the Other menu.
Tutorial Part 2
When the Sequence Editor window opens, the sequence spreadsheet may
also be displayed. We will come back to it later.
2.
In the Sequence Editor window, choose New from the File menu.
This opens the Global Parameters dialog box:
3.
In the Build Parameters tab, select your instrument name from the
Instrument pull down menu.
4.
5.
6.
7.
Tutorial Part 2
8.
Select Same name for Raw data and Result files option.
9.
This dialog box is where you enter all the information that you need to create
a basic sequence.
10. In the Study text box, type the name TUTORIAL to identify the analysis
being performed.
11. In the Method text box, browse to the examples directory, by clicking on
the yellow folder icon to the right of the method box, and select
C:\penexe\\practice.mth.
We want to use our newly optimized method to reprocess existing raw data files.
12. Change the Base file name to HALO for Channel A.
Note: HALO is the name of a group of raw files that are supplied with TotalChrom as
examples. We cannot use the raw file that you acquired in Part 1 of the tutorial since
every user has a different one.
This will be the base name of the raw and result files that will be listed in the
sequence spreadsheet.
13. In the Calibration box, deselect Include calibration standards by clicking
on the check box.
We already calibrated the method manually in the Method Editor.
74
Tutorial Part 2
14. In the Samples box, enter 3 for the Number of samples to be analyzed in
the sequence.
15. In the Samples box, enter 1 in the # Injections per sample.
16. Choose OK to close the Sequence Template dialog box.
When you complete the sequence template, TotalChrom builds the sequence. You
can then view and edit the sequence data in spreadsheet form.
Your sequence spreadsheet should have 3 rows, the number of samples you specified
in the template. Each row contains information about the sample and the values that
will be used when the raw data is reprocessed. To see the following columns and
information you will have to scroll the sequence spreadsheet window to the right:
The Report column is the name of the report format file that will be used to
generate the report.
When you build a sequence to acquire a series of injections, this entry will
be the base file name of the raw files that are created.
The parameters that you did not change contain either the default value or no
value.
Note: The columns that are displayed may vary depending on which options were
selected in the Global Parameters, Configuration tab or user defined hidden
columns.
75
Tutorial Part 2
The Edit menu is the same menu we saw when we developed the vial list in the Part 1
of the tutorial. It offers several choices for manipulating rows in the spreadsheet.
We will use the Change menu to change the paths for the data files so TotalChrom
will know to look for them in the Examples directory rather than the Tutorial
directory.
76
Choose Path from the Change menu to open the Change Path dialog box.
2.
3.
4.
In the New path text box, browse to the Examples directory by clicking on
the yellow file folder to the right of the text box.
Tutorial Part 2
5.
6.
Choose OK.
You will notice that the full path is now displayed in the Data column and
that it now points to the Examples directory.
Scroll the spreadsheet window horizontally until the printer and plotter
columns are visible. They are at the far right of the spreadsheet.
2.
Select the cell for row 1 of that column by clicking inside it.
77
Tutorial Part 2
3.
Click on the Printer pull down menu and select the entry that represents the
printer you want to use.
4.
5.
Choose Fill Down from the Change menu to copy the entry from row 1 to
every cell in the column.
Right clicking on the cell and selecting Fill Down can also accomplish this.
78
2.
When the Description dialog box opens, enter any information that you want
to store with the sequence file. Do not select Audit trail.
Tutorial Part 2
3.
Choose OK.
4.
5.
Choose Save to close the dialog box and save the new sequence file.
For more information on sequences, see Chapter 11 in the TotalChrom Users Guide.
Now we are ready to use this sequence with the Batch Reprocessing
function.
79
Tutorial Part 2
2.
Choose your Actual data path by clicking on the yellow folder and
browsing to the Examples directory.
The Data Path is the directory where your Raw and Result files are stored. If
you used the full path for your data files in the sequence spreadsheet, which
we did in the previous exercise, you can leave this field blank.
80
Tutorial Part 2
3.
Click OK.
4.
5.
Choose Local (if you have Client/Server) or Interactive (if you have
Workstation) for the Batch execution.
6.
7.
81
Tutorial Part 2
The Batch Reprocessing window opens, listing the raw files that are
scheduled for reprocessing. They are the three files from our sequence.
8.
Choose Start from the Reprocess menu to begin reprocessing the data. You
can also start the Batch process by clicking on the green arrow on the toolbar.
The entries on the list are deleted as they are processed. The report and replot
for each file will be printed once data analysis is complete.
9.
To close the Batch Reprocessing window and return to the Navigator, choose
Exit from the File menu.
82
Tutorial Part 2
2.
3.
Click Open.
You will notice that the Halo003.rst file now has the method parameters
applied to it that you specified in the practice.mth that you created in the
Graphic Method Editor.
4.
By either using Print Preview, or looking at the printed report, you will
notice that the processing method that was used for this data file is the
practice.mth method that you created.
83
Tutorial Part 2
84
Tutorial Part 2
The display of any window may be toggled on or off using menu commands or
toolbar buttons. Each window may also be closed using the close button on the
docked window itself. Each window may be toggled between normal size and
maximized in the usual way for docking windows (i.e. the arrow button incorporated
in the frame).
The docking scheme will be implemented such that the open docked windows
completely fill the parent window. While it will be possible to undock a window there
is little point in doing so. Docking will be used here to allow a flexible layout with
easy relative sizing of windows, rather than any other purpose.
When a single file is selected in the file list, the plot window displays the
chromatogram from that file and the report window displays results from that run (if
available). The chromatogram can be annotated with sample information, peak names
and/or other peak information, baselines and labels added by the user. The result data
85
Tutorial Part 2
can be displayed in the form of a simple peak table and/or in a specific TotalChrom
report format as defined by a .rpt file.
When more than one file is selected in the file list, the plot window displays all the
chromatograms, in either a stacked or overlay display format and the report window
displays a table containing a summary of results from all the files. There is no
arbitrary limit to the number of chromatograms that can be displayed the stacked
mode is quite valid for any number of plots.
The following example is a scenario of how the Review and Approve process could
be used. This example has been developed to illustrate some key aspects of the
Review and Approve specification; it is neither the simplest nor the most complex
path through the application. While the scenario aims to be accurate in its depiction of
review and approve, it does not attempt to be completely realistic in all matters.
Set Up Conditions
The configuration of the TotalChrom system for this example contains the following
key items:
Review and Approve signature levels:
Review, Final Review, Final Approve and Hold
Review and Approve options:
The same user cannot sign two levels for a report = True. You must enter a comment
for each Hold report = True
Job Types:
Manager, Analyst, Technician
Users:
Susanna (Manager), Michael (Senior Analyst), Vicky (Analyst), and James (Lab
Technician)
Review and Approve permissions:
Susanna Final Approve, Hold
Michael Final Review, Hold
Vicky Review
James no permissions
The sample data set for this example consists of three samples (CK46123, CK46124
and CK46125) each of which has two associated reports (Actives and Impurities).
Processing of these samples yielded three TCR files, each containing two report
definitions.
Vicky has reviewed the results in the Review and Approve environment and signed
each report, such that each now has the status of Reviewed.
86
Tutorial Part 2
In this Example
1.
Michael logs on to TotalChrom on one of the lab computers and opens the
Review and Approve environment by clicking the
button in the
TotalChrom Navigator. He prefers to be away from his office phone when
reviewing reports.
2.
He selects the three TCR files representing the CK46123, CK46124 and
CK46125 sample reports in the File Open dialog and clicks OK.
The Signature Status column in the file list shows Reviewed for all three TCR
files, as Michael expects (knowing Vicky came in on Sunday to review the
samples after James had finished the chromatographic runs). The chromatogram
and Actives report for the first sample are displayed in the Plot and Report
windows.
3.
4.
Now he wants to review the reports more closely prior to signing them so he
chooses the All Reports command from the Sign menu.
The Report Review window appears with the Actives report for sample CK46123
displayed. Michael notes that the Review, Final Review and Hold buttons are
enabled on the Sign toolbar, while the Approve and Final Approve buttons are
disabled.
He will not use the Review button since the company SOP only requires two
levels of review. He also hopes he wont be using the Hold button because
rejecting batches of product leads to a lot of paperwork and unwelcome attention
from upper management. Once he has applied a Final Review signature to these
reports they will be ready for Susanna to give Final Approval.
5.
Michael clicks the Zoom In button so that the report text is at a size that is
comfortable for him to read. He then checks the header information carefully to
make sure the correct methods were used. Clicking the scroll bar, he examines
the peak table to ensure the reported amount of each active component is within
limits and that the suitability parameters are within tolerance.
6.
When he is satisfied that all is as it should be Michael clicks the Final Review
button.
87
Tutorial Part 2
The Impurities report for CK46123 is then displayed. Michael reviews the header
information again and checks the amounts for each impurity, as well as the figure
for total impurities. All these values are well within limits and so Michael clicks
Final Review once again.
7.
Michael repeats his examination on the reports from CK46124. Both of these are
also satisfactory, so he marks both using the Final Review button.
8.
The Actives report for CK46125 also appears acceptable, so he clicks Final
Review once more. This causes the Impurities report for CK46125 to be
displayed. Michael notes that the individual impurities are just within limits but
the total is too high.
9.
10. He clicks OK and the notes that the report name in the status bar of the Review
Reports window is now followed by (Hold Unsigned). Since that was the final
report in the set, the Close button has been replaced by Finish, indicating he has
reviewed everything.
11. Michael clicks the Finish button and the Sign reports dialog is displayed. This
shows the three TotalChrom Report files he originally selected as top level nodes
on the tree control. Each top level node has two secondlevel nodes under it,
representing the individual reports. The Status to be Signed column shows three
Reviewed Final and one Hold:
C:\data\september\09\CK46123.tcr
C:\data\september\09\actives_r2.rpt
C:\data\september\09\impurities_r3.rpt
Reviewed Final
Reviewed Final
C:\data\september\09\CK46124.tcr
C:\data\september\09\actives_r2.rpt
C:\data\september\09\impurities_r3.rpt
Reviewed Final
Reviewed Final
C:\data\september\09\CK46124.tcr
C:\data\september\09\actives_r2.rpt
C:\data\september\09\impurities_r3.rpt
Reviewed Final
Hold
12. Michael clicks on the Sign button and the Electronic Signature dialog appears.
13. Michael enters his user name (logon ID) in the first field and his password in the
second field.
88
Tutorial Part 2
14. When he clicks the OK button the display returns to the main Review and
Approve screen. The file list still contains the three TCR files for samples
CK46123, CK46124 and CK46125. The Status column now displays Reviewed
Final for the first five reports and Hold for the final report.
At that moment Susanna walks into the lab. Debbie said that the analyses that
had to be run again over the weekend. Well, I can tell you that Debbie isn't
going to be happy because one sample is no good. CK46125 has too high a total
level of impurities. We cant let it go. How about the others, asks Susanna.
CK46123 was well within limits and CK46124 was between the two. Well I
suppose we should be grateful that two out of three are OK. Do you want to
look these over, since you are here right now? asks Michael, offering his chair
to Susanna. Sure, she replies, and sits down at the computer.
15. Susanna chooses the All Reports command from the Sign menu and examines
each report in turn in the Review Reports window. The first five reports are all
marked with the Reviewed Final status that Michael signed off.
16. Susanna clicks the Final Approve button for each of these. After the fifth click,
the final report is displayed and Susanna notes the Hold designation in the status
bar.
17. She clicks the Comment button, which is the only button in the Sign toolbar
enabled, since the current unapproved report has a comment attached. A dialog
is displayed showing Michael's reason for holding the report.
You misspelled phenylethyllonamide, Michael, Susanna teases. Thats not
like you.
18. Susanna closes the comment dialog and clicks the Finish button, reviews the
names of the five reports she is signing in the Sign Reports dialog, clicks the Sign
button and finally enters her logon name and password in the Electronic
Signature dialog.
When she accepts that dialog the file list in the main Review and Approve
window is updated to show the following Signature Statuses for the three sample
files:
C:\data\september\09\CK46123.tcr
Reviewed Final
C:\data\september\09\CK46124.tcr
Reviewed Final
C:\data\september\09\CK46124.tcr
Hold
At that moment James rushes into the lab. Good Morning, James says Michael.
Hi Mike, I was hoping Susanna would be with you. Debbie is waiting for you in
your office, Susanna. She asked me to get you right away. OK, I'll go and
break the bad news to her. Call on me if you need support, Michael calls after
89
Tutorial Part 2
Susanna as she leaves the lab. That's what the PerkinElmer salesperson said,
replies Susanna, but that TotalChrom software is so easy to use I have never had
to..
Congratulations!
You have successfully completed the TotalChrom Tutorial.
This tutorial was designed to give new users a brief overview of how to setup and run
a chromatograph with the TotalChrom software. There are many different features
and options available in TotalChrom that were not discussed in this tutorial. We
strongly recommend that all users read the TotalChrom Users Guide carefully in
order to fully understand this application and how it works.
It is also advised that new users consider attending a TotalChrom training class to
supplement this tutorial. To find out more information about a training class in your
area, please contact 1-800-762-4000.
The global TotalChrom Support Team is readily willing to help you with any
questions or difficulties you might experience while using the TotalChrom software.
To reach your TotalChrom support representatives, call 1-800-762-4000 (U.S. and
Canada). For our international users, please contact your local PerkinElmer office.