DigitalMultimeterCalibration
Purpose
Thepurposeofthisprocedureistoprovideinstructionforthecalibrationofdigital
multimeters(DMM).
DownloadCalibrationControl
Email:*
Name:*
Scope
Company:*
Allmultimeters(multirangeinstrumentsusedtomeasurevoltage,current,and
DayPhone:
resistance)aretobeverifiedincalibrationnolessthanonceperyear.
SignIn&DownloadSoftware
Procedure
PreliminaryInstructionsandNotes
1. Readthisentireprocedurebeforebeginningthecalibration.
2. CalibrationshallbeperformedinanenvironmentthatconformstoManufacturer
Specifications.
http://www.apesoftware.com/calibrationprocedures/digitalmultimeter.aspx
1/12
23/01/2016
FreeDigitalMultimeterCalibrationProcedure
3. ThedigitalmultimeterwillhereafterbereferredtoastheInstrumentUnderTest
(IUT).
4. VerifythattheIUTisclean.
5. VisuallyexaminetheIUTforanyconditionthatcouldcauseerrorsinthecalibration.
6. Ifanyoftherequirementscannotbemet,refertotheapplicablemanufacturer
manual.
7. Ifamalfunctionoccursoradefectisobservedwhilecalibrationisinprogress,the
calibrationshallbediscontinuedandnecessarycorrectiveactiontakenifcorrective
actionaffectsameasurementfunctionpreviouslycalibrated,thefunctionshallbe
recalibratedbeforetheremainderoftheprocedureimplemented.
ReferenceMaterial
ApplicableManufacturersManualorBrochures.
Specifications
ThespecificationsoftheIUTaredeterminedbytheapplicablemanufacturer's
documentation.Ifthemanufacturer'sdocumentationisnotavailable,thenthe
specificationsidentifiedinthisprocedureareused.
EquipmentRequired
TheStandardslistedbelowshouldbeselectedonthebasisoftheirhigheraccuracylevel
whencomparedtotheunitundertest.EquivalentStandardsmustbeequaltoorbetter
thantheMinimumUseSpecification.
http://www.apesoftware.com/calibrationprocedures/digitalmultimeter.aspx
2/12
23/01/2016
FreeDigitalMultimeterCalibrationProcedure
MinimumUseSpecificationsforStandardslistedare1/4theaccuracyrequiredbythe
IUT.
RotekModel2500Calibratororequivalent
DiodeTestFixture
TestLeads
ContinuityTestFixture
CapacitanceTestFixture
Thermometer
Hydrometer
SetUp
Turnallpoweron(calibratorandtheunittobecalibrated).Allowtheinstrumentsto
stabilizeforapproximately5minutes.Conductthetestsinanambienttemperatureof25
5Candarelativehumidityoflessthan80%.
Note:IftheIUThasanautomaticrangingfeature,thenthefunctionbeingchecked(i.e.
voltage,current,orresistance)mustbeverifiedinthevariablerangemodeaswell.
BatteryReplacement
ReplacethebatteryintheIUTandmakesurethebatterycontactsarenotdirty.
DetailedProcedure
DISPLAYANDSWITCHTEST
http://www.apesoftware.com/calibrationprocedures/digitalmultimeter.aspx
3/12
23/01/2016
FreeDigitalMultimeterCalibrationProcedure
TurntheIUTonandverifythatallLCDsegmentsareworkingandnotdim.Verifythatthe
selectorswitch(s)is/areworkingproperly.MakesuretheLowBatteryindicatorisnot
showing.
RESISTANCECHECK
ConnectthecalibratorbetweentheV/ohmandcommoninputterminalsoftheIUT.Use
thefollowingcharttoverifytheresistancevalues.
Step
Range
Input
Display
200ohm
Short
00.0to00.5
2kohm
Short
0.000to0.001
2kohm
1kohm
.998to1.002
20kohm
10kohm
9,98to10.02
200kohm
100kohm
99.8to100.2
2000kohm
1Mohm
998to1002
2000Mohm
Open
0.10to00.0
CONTINUITYTEST
1. Selectthecontinuitycheckfunctiononthecalibrator.
2. ConnectthetestleadstotheV/ohmandcommonterminalsoftheIUT.
3. Momentarilyshortthetestleadstogetherandobservethatthetonesounds.
http://www.apesoftware.com/calibrationprocedures/digitalmultimeter.aspx
4/12
23/01/2016
FreeDigitalMultimeterCalibrationProcedure
4. Connectthetestleadstothecontinuitytestfixturebetweenthe100ohmtest
points.Notoneshouldbeheardindicatingnonterminatingconnection.
5. Connectthetestleadstothecontinuitytestfixturebetweenthe50ohmsetup.The
testtoneshouldindicatecontinuitybetweenthetestpoints.
DCVOLTAGETEST
1. UsethecalibratortosupplythecorrectDCvoltagelistedbelow.
2. Setthecalibratorforazerovoltinput.
3. ConnectthecalibratoroutputtotheV/ohmandcommoninputterminalsoftheIUT.
Note:Connecttheground/common/lowsideoftheRotek2500tothecommonon
theIUT.
4. Withreferencetothetablebelow,selecttheIUTvoltagerangeandsetthe
calibratoroutputtothecorrespondingIUTinputvoltage.Testperthefollowing
chartandverifythatthedisplayfallswithinthelimitsgiven.
Step
Range
Input
Display
200mV
190mV
189.7to190.3
2V
0.0V
0.001to0.001
2V
1.9V
1.897to1.903
20V
19V
18.97to19.03
http://www.apesoftware.com/calibrationprocedures/digitalmultimeter.aspx
5/12
23/01/2016
FreeDigitalMultimeterCalibrationProcedure
200V
100V
99.84to100.16
1000V
400V
399.3to400.6
ACVOLTAGETEST
UsethecalibratortosupplythecorrectACvoltageinthefollowingtest:
1. Setthecalibratorto60Hz.
2. ConnectthecalibratoroutputtotheV/ohmandcommoninputterminalsoftheIUT.
Connecttheground/common/lowsideofthecalibratortocommonontheIUT.
3. Withreferencetothetablebelow,selecttheIUTvoltagerangegiveninstep1and
setthecalibratoroutputtothecorrespondingIUTinputvoltage.Verifythatthe
displayreadingiswithinthelimitsshown.
Step
Range
Input
200mV 100mV
Display
99.0to101.0
2V
1V
0.990to1.010
20V
10V
9.90
200V
100V
99.0to100.1
750V
400V
294to406
DCCURRENTTEST
UsethecalibratortosupplythecorrectDCcurrentlistedinthetablebelow:
http://www.apesoftware.com/calibrationprocedures/digitalmultimeter.aspx
6/12
23/01/2016
FreeDigitalMultimeterCalibrationProcedure
1. SettheoutputofthecalibratortozeromA.
2. Connecttheoutputofthecalibratortothelowestcurrentterminal(typically200
mA)andcommoninputterminalontheIUT.
Note:Donotexceedthemaximumcurrentratingofthemeter.Onlytestfor
valuesthemeteriscapableofreading.
3. Withreferencetothetablebelow,selecttheIUTcurrentrangeandsetthe
calibrationoutputtoprovidethecorrespondingIUTinputcurrent.Verifythatthe
displayreadingiswithinthelimitsshown.
Step
Range
Input
Display
200?A
190?A
189.7to192.3
2mA
1.9mA
1.897to1.923
20mA
19mA
18.97to19.23
4. DisconnectthecalibratorandchangethecurrentterminalontheIUTtothehighest
setting(typically10A)leavingthecommoninputterminalconnected.
5. Usethefollowingtabletoverifythatthemeterisfunctioningwithinthespecified
parameters.
Step Range Input
Display
10A
1A
0.98to1.02
10A
5A
4.90to5.10
http://www.apesoftware.com/calibrationprocedures/digitalmultimeter.aspx
7/12
23/01/2016
FreeDigitalMultimeterCalibrationProcedure
10A
9A
8.82to9.18
ACCURRENTTEST
UsethecalibratortosupplythecorrectACcurrentlistedinthetablebelow:
1. Usethefollowingtabletoverifythatthemeterisfunctioningwithinthespecified
parameters.
2. SettheoutputofthecalibratortozeromA.
3. Connecttheoutputofthecalibratortothelowcurrentinputterminal(typically200
mA)andthecommonterminaloftheunitundertest.
Note:Donotexceedthemaximumcurrentratingofthemeter.Onlytestfor
valuesthemeteriscapableofreading.
4. Withreferencetothetablebelow,selecttheIUTcurrentrangeandsetthe
calibratoroutputtoprovidethecorrespondingIUTinputcurrent.Verifythatthe
displayreadingiswithinthelimitsshown.
Step
Range
Input
Display
200?A
190?A
18.27to19.28
2mA
1.9mA
1.872to1.928
20mA
19mA
18.72to19.28
5. DisconnectthecalibratorandchangethecurrentterminalontheIUTtothehighest
http://www.apesoftware.com/calibrationprocedures/digitalmultimeter.aspx
8/12
23/01/2016
FreeDigitalMultimeterCalibrationProcedure
setting(typically10A)leavingthecommoninputterminalconnected.
6. Usethefollowingtabletoverifythatthemeterisfunctioningwithinthespecified
parameters.
Step Range Input
Display
10A
1A
0.97to1.03
10A
5A
4.87to5.13
10A
9A
8.77to9.23
LOGICTEST
Usethisproceduretotestweatherthelogicfunctionisindicatingthecorrectstate.
1. SettheIUTtothelogictestmodeandinsertthetestleadsintotheV/ohmand
commonterminals.
2. Shorttheleadstogetherandverifythatthemeterindicatesalowlogicstate.
3. Setthecalibratorto0.400VDCandconnectthemetertotheoutputterminals.
Verifythatthemeterindicatesalowlogicstate.
4. Setthecalibratorto3.500VDCandverifythattheIUTindicatesahighlogicstate.
FREQUENCYTEST
1. ConnectanoscilloscopetoafunctiongeneratortomeasureACsinewaves.
2. Setthefunctiongeneratortooutput10VACandusetheoscilloscopetosetthe
frequenciesfromthefollowingtable.
http://www.apesoftware.com/calibrationprocedures/digitalmultimeter.aspx
9/12
23/01/2016
FreeDigitalMultimeterCalibrationProcedure
3. ConnecttheIUTleadstotheV/ohmandcommonterminalsontheIUTandtothe
groundandoutputofthefunctiongenerator.
4. Verifythemeterisdisplayingproperresultsfromthetable.
Step Range
Input
Display
Hz
60Hz@10V
59.4to60.6
Hz
1kHz@10V
.990to1.010k
Hz
1MHz@10V .990to1.010M
CAPACITANCETEST
1. SettheIUTtomeasurecapacitance,insertthetestleadsintotheV/ohmand
commontestterminals.
2. Connectthemetertothecapacitancetestfixtureandusethefollowingtableto
determinetheabilityoftheIUT.
Step
Range
Input
Display
Capacitance
10microF
9.85to10.15
Capacitance
1microF
.985to1.015
DIODETEST
1. SettheIUTtothediodecheckfunctionandconnectthetestleadstotheV/ohm
http://www.apesoftware.com/calibrationprocedures/digitalmultimeter.aspx
10/12
23/01/2016
FreeDigitalMultimeterCalibrationProcedure
andcommonterminals.
2. Connectthetestleadstothediodetestfixtureandfollowthestepsinthe
proceedingtabletoverifydiodetestoperation.
Step
Range
Setup
Polarity
Display
Tone
DiodeCheck
Short
N/A
.000to.001
Yes
DiodeCheck
Si
Forward .800to.600
Yes
DiodeCheck
Si
Reverse
No
DiodeCheck
Ge
Forward .300to.500
Yes
DiodeCheck
Ge
Reverse
No
1.
1.
Note:Somemetersarenotdesignedtoemitatone.
ReasonforReissue
RevisionAFirstRelease
Software
Software
GetHelp
Forum
About
ContactUs
CalibrationControl
Printers
HelpPages&Videos
ForumHome
AboutUs
ContactUs
Download
Scanner
RemoteSupport
CalibrationControl
OurCustomers
EmailSales
VersionHistory
MinimumRequirements
OnlineMeeting
Feedback
ImportantLinks
EmailSupport
FeatureList
PDFBrochure
OnSiteServices
SiteMap
DesktopSupport
http://www.apesoftware.com/calibrationprocedures/digitalmultimeter.aspx
11/12
23/01/2016
FreeDigitalMultimeterCalibrationProcedure