DV3T Manual
DV3T Manual
Viscometer
Operating Instructions
Manual No. M13-2100-A0415
SPECIALISTS IN THE
MEASUREMENT AND
CONTROL OF VISCOSITY
with offices in:
Table of Contents
I. INTRODUCTION................................................................................................... 5
I.1 Components................................................................................................................................6
I.2 Utilities.......................................................................................................................................8
I.3 Specifications..............................................................................................................................8
I.4 Installation..................................................................................................................................9
I.5 Safety Symbols and Precautions...............................................................................................11
I.6 Key Functions...........................................................................................................................11
I.7 Cleaning....................................................................................................................................12
V. SETTINGS........................................................................................................... 57
I. INTRODUCTION
The Brookfield DV-III Rheometer series has been the leading industrial rheometer since it was first
introduced in 1989. Brookfield has continued to develop and improve the DV-III to provide the best
value in the market for both quality control and research customers. The Brookfield DV3T Rheometer
continues in this tradition of innovation, quality and value. The incorporation of a full color graphical
touch screen display has allowed for a new and improved user interface which preserves the single
speed data collection methods while making all of the advanced features of the DV3T readily available.
The Brookfield DV3T Rheometer measures fluid viscosity at given shear rates. Viscosity is a measure
of a fluids resistance to flow. You will find a detailed description of the science of viscosity in the
Brookfield publication More Solutions to Sticky Problems, a copy of which was included with your
DV3T.
This manual covers the DV3T Rheometer and the DV3T Extra Rheometer. The DV3T Extra is a special
package that includes several optional features to maximize the utility of the DV3T. The DV3T Extra
includes: Ball Bearing Suspension (not included in LV versions), EZ-Lock spindle connection, Model
Q Lab Stand, and RheocalcT Software. These differences are detailed in the Components section of
the manual. All references to the rheometer in this manual will be made as DV3T. All of the optional
features of the DV3T Extra may be added individually to the DV3T Rheometer.
The DV3T offers exceptional versatility in modes of control, allowing choice of traditional standalone
operation and automatic operation through programs downloaded from the PC or with complete control
by PC using Brookfield RheocalcT Software.
The DV3T can be used as a traditional Brookfield rheometer for collection of single speed
viscosity data through the easy to use touch screen; just select the spindle and speed and read
the value from the display. [See Section II: Getting Started]
T he DV3T can collect multiple data points through programs created directly on the
touchscreen display.
The DV3T can make a direct measurement of yield stress when using optional vane spindles.
The Brookfield PG Flash Software can be used to program the DV3T to control all aspects of
the test and data collection without the need for the operator to monitor the instrument; just
start the program and return to the printed test data (printer is optional). [See Section V: PG
Flash Software]
The Brookfield RheocalcT Software will perform all control and data collection functions
of the DV3T from the PC while also providing a platform for advanced data collection and
analysis.
In any of these modes of control, the DV3T will provide the best in viscosity measurement and control.
The principal of operation of the DV3T is to drive a spindle (which is immersed in the test fluid)
through a calibrated spring. The viscous drag of the fluid against the spindle is measured by the spring
deflection. Spring deflection is measured with a rotary transducer. The measurement range of a DV3T
(in centipoise or milliPascalseconds) is determined by the rotational speed of the spindle, the size and
shape of the spindle, the container the spindle is rotating in, and the full scale torque of the calibrated
spring.
There are four basic spring torque series offered by Brookfield:
Spring Torque
Model
dynecm
milliNewtonm
DV3TLV
673.7 0.0673
DV3TRV
7,187.0
0.7187
DV3THA
14,374.0
1.4374
DV3THB
57,496.0
5.7496
The higher the torque calibration, the higher the measurement range. The measurement range for each
torque calibration may be found in Appendix B.
All units of measurement are displayed according to either the CGS system or the SI system.
Brookfield Engineering Labs., Inc.
Page 5
1. Viscosity appears in units of centipoise (cP), Poise (P), milliPascal-seconds (mPas) or Pascalseconds (Pas) on the DV3T Rheometer display or centistokes (cSt) or millimeter squared
per second (mm2/sec).
2. Shear Stress appears in units of dynes/square centimeter (D/cm2) or Newtons/square
meter (N/m2)/or Pascals (Pa).
3. Shear Rate appears in units of reciprocal seconds (1/sec).
4. Torque appears in units of dyne-centimeters or Newton-meters (shown as percent % in
both cases) on the DV3T Rheometer display.
5. Density appears in units of grams/cubic centimeter (g/cm3) or kilograms/cubic meter
(kg/m3).
Note: To change CGS to SI units on the display (see Section IV.3: Global Settings).
The equivalent units of measurement in the SI system are calculated using the following conversions:
SI
CGS
Viscosity:
1 mPas
= 1 cP
Shear Stress:
1 Newton/m2 = 10 dyne/cm2
Torque:
1 Newtonm = 107 dynecm
References to viscosity throughout this manual are done in CGS units. The DV3T Rheometer provides
equivalent information in SI units.
I.1 Components
Please check to be sure that you have received all components, and that there is no damage. If
you are missing any parts, please notify Brookfield Engineering or your local Brookfield agent
immediately. Any shipping damage must be reported to the carrier.
Components
Part Number
DV3T Rheometer
Model G Laboratory Stand
Model Q Laboratory Stand
Quantity
DV3T
DV3T Extra
varies
MODEL G
MODEL Q
varies
DVP-94Y
B-20Y or B-20KY
B-21Y or B-21KY
DVE-7Y
CD ProgA
GV-3003
GV-1044
GV-1045
GV-1043
M13-167
N/A
GV-1020
1
1
1
1
1
1
1
1
1
1
1
or
1
1
1
1
1
1
1
1
1
1
1
1
Page 6
COMPONENT DIAGRAM
DV3T
Rheometer
Model G
Laboratory Stand
Bubble
Level
Spindle Set
Shipping
Cap
Guard Leg
LV Spindle Set
shown above
Leveling Screw
Cone/Plate Option
Temperature Probe
Wrench
Tension Bar
Cone Spindle
Sample Cup
Temperature Probe
Figure I-1
Page 7
I.2 Utilities
Input Voltage:
Input Frequency:
Power Consumption:
Power Cord Color Code:
Hot (live)
Neutral
Ground (earth)
Black
White
Green
Brown
Blue
Green/Yellow
Main supply voltage fluctuations are not to exceed 10% of the nominal supply
voltage.
I.3 Specifications
Speeds:
Weight:
Gross Weight
23 lbs.
10.5 kg.
Net Weight
20 lbs.
9 kg.
Carton Volume
1.65 cu. ft.
0.05 m3
Carton Dimensions 22 in. (56 cm) W x 11 in. (28 cm) L x 22 in. (56 cm) H
Temperature Sensing Range:
-100C to 300C (-148F to 572F)
USB A Port for use with an attached PC.
USB B Port (x3) for use with a USB Flash Drive and Dymo Printer.
Viscosity Accuracy:
Viscosity Repeatability:
Temperature Accuracy:
Operating Environment:
If you ordered the ball bearing suspension system with your new instrument, please note
the following:
1) The ball bearing suspension in your Brookfield instrument is noted on the serial tag on
the back of the head by the letter B in the part number (the ninth digit; for example:
XDV3THBTB00U00.
2) When attaching and detaching the spindle, it is not necessary to lift the coupling where
the spindle connects to the instrument.
3) The Oscillation Check, explained in Section IV.1: Device Setup, does not pertain to
this instrument.
Page 8
Electrical Certifications:
Conforms to CE Standards:
BSEN 61326:
Electrical equipment for measurement, control and laboratory use - EMC
requirements.
BSEN 61010-1: Safety requirements for electrical equipment, for measurement, control
and laboratory use.
Airborne Noise Emissions - levels do not exceed 70 dB(A).
Notice to customers:
This symbol indicates that this product is to be recycled at an appropriate collection center.
Users within the European Union:
Please contact your dealer or the local authorities in charge of waste management on how to dispose
of this product properly. All Brookfield offices and our network of representatives and dealers can
be found on our website: www.brookfieldengineering.com.
Users outside of the European Union:
Please dispose of this product according to your local laws.
I.4 Installation
Note: IQ, OQ, PQ, an abbreviated guideline document for installation, operation and
performance validation for your DV3T digital rheometer, can be downloaded from
our website www.brookfieldengineering.com. A more detailed IQ, OQ, PQ procedure
is available for purchase from Brookfield or your authorized dealer.
1. Assemble the Model G Laboratory Stand (refer to assembly instructions in Appendix I).
5. Remove the shipping cap which secures the coupling nut on the rheometer to the pivot
cup. For Cone/Plate Models, hold the Sample Cup and swing the tension bar away
from the bottom of the cup. Lower the cup and remove the foam insert. (Save for future
shipments.)
6. Optional: Install the screen protector per the instructions on the package (and also
shown in Appendix K). Failure to properly install the screen protector may result
in touch screen malfunction. Additional installation help can also be found on our
YouTube channel: www.youtube.com/user/BrookfieldEng
The AC input voltage and frequency must be within the appropriate range as shown
on the nameplate of the viscometer (see Section I.2)
Brookfield Engineering Labs., Inc.
Page 9
7. Make sure that the AC power switch at the rear of the DV3T is in the OFF position.
Connect the power cord to the socket on the back panel of the instrument and plug it into
the appropriate AC line. For Cone/Plate Models, be sure that the toggle switch, used to
activate the electronic gap, is to the left position. (Left when facing the rheometer.)
The AC input voltage and frequency must be within the appropriate range as
shown on the nameplate of the rheometer. (See Section I.2.)
Note: The DV3T must be earth grounded to ensure against electronic failure!!
8. Turn the power switch to the ON position and allow the rheometer to warm up for 10
minutes before performing autozero.
9. For Cone/Plate models, refer to Appendix A.
10. If appropriate, connect USB cable (DVP-202) to USB port for connection of DV3T
to PC or printer.
11. Review Read Me file. The Read Me file contains notes on the DV2T Firmware. This file
can be found on the CD that contains the PG Flash Software.
Temperature
Probe
Temperature
Controller
USB B
USB A
Ethernet
Power Input
(Future Use)
On/Off
Switch
Figure I-2
Brookfield Engineering Labs., Inc.
Page 10
Refer to the manual for specific warning or caution information to avoid personal
injury or damage to the instrument.
Precautions
If this instrument is used in a manner not specified by the manufacturer, the protection
provided by the instrument may be impaired.
In case of emergency, turn off the instrument and then disconnect the electrical cord
from the wall outlet.
The user should ensure that the substances placed under test do not release poisonous,
toxic or flammable gases at the temperatures which they are subjected to during the
testing.
Data Fields
More/Less Bar
Command Keys
Figure I-3
Brookfield Engineering Labs., Inc.
Page 11
Status Bar:
The Status Bar provides information relating to the date and time (as configured
by the user) and various connections to the DV3T Rheometer.
Title Bar:
The Title Bar identifies the activity to be conducted in the current view and
includes any navigation icons that are relevant.
Test Name:
The Test Name identifies the name of the currently loaded test.
Data Fields:
More/Less Bar: The More/Less Bar informs the user when more Data Fields are available and
when the number of Data Fields can be reduced.
Command Keys: The Command Keys indicate action that can be taken. These keys will vary
from view to view depending on what actions are relevant.
I.7 Cleaning
Make sure that the instrument is in a clean, dry working environment (dust-free, moderate
temperature, low humidity, etc.).
Make sure the instrument is on a level surface.
Hands/fingers must be clean and free of sample residue. Not doing so may result in deposit
build up on the upper part of the shaft and cause interference between the shaft and the pivot
cup.
Be sure to remove the spindle from the instrument prior to cleaning. Note left-handed thread.
Severe instrument damage may result if the spindle is cleaned in place.
Instrument and Display:
Immersed Components (spindles): Spindles are made of stainless steel. Clean with a nonabrasive cloth and solvent appropriate for sample material.
When cleaning, do not apply excessive force, which may result in bending spindles.
Page 12
II.1 Power Up
The DV3T will go through a Power Up sequence when the power is turned on. The Rheometer
will issue a beep, present a blue screen, and finally show the DV3T About screen for 5 seconds.
The About screen is shown below and includes several critical parameters about the rheometer
including; rheometer torque (LV, RV, HA, HB, or other), firmware version number, model number
(DV3TLV for example) and the serial number.
Figure II-1
The About screen can also be accessed through the Settings Menu
The DV3T Rheometer will automatically transition from the About screen to the AutoZero screen.
TIP: When contacting Brookfield or your authorized Brookfield dealer for technical support or
repair services, please record the information on the About screen and include this detail
in your email.
II.2 AutoZero
The DV3T Rheometer must perform an AutoZero prior to making viscosity measurements. This
process sets the zero reading for the measurement system. The AutoZero will be performed every
time the instrument is turned on. Additionally, you may perform an AutoZero at any time through
(see Section V.1).
the Settings Menu
The AutoZero screen will be presented automatically, after the About screen, during a power up.
Page 13
Figure II-2
The operator must ensure that the rheometer is level (see Section I.4) and remove any attached
spindle or coupling. When the Next button is pressed, the rheometer will operate for approximately
13 seconds. After the AutoZero is complete and the operator presses the Next button, the rheometer
will transition to the Configure Viscosity Test screen. If the AutoZero was performed from the
Settings Menu, then the rheometer will return to the Settings Menu.
TIP: Do not touch the rheometer during the AutoZero process to ensure the best zero value.
This status bar will indicate: time of day, date, and connection status for a variety of connection
devices. The status icons are defined as:
USB Icon 1,2,3: The DV3T Rheometer can store data and tests to USB storage
device (USB B) such as a USB Flash Drive. There are three
USB ports. These ports are represented as 1, 2 and 3 based on
the order of connection.
Printer Icon:
II.4 Navigation
The DV3T Rheometer uses a touch screen display. Navigation of the instrument features is done
using a variety of Data Fields, Arrows, Command Keys and Navigation Icons. The operating
system has been designed for intuitive operation and employs color to assist the user in identifying
options.
Data Fields require that the user touch the screen
to initiate the data entry / selection process. These
fields are normally outlined in black. They may
also include a blue arrow.
Blue Arrows indicate that options exist for a Data
Field. The user may be required to press anywhere
in the Data Field or they may have to press the Blue
Arrow specifically.
Command Keys are buttons which direct the DV3T
to perform a specific action such as SAVE a data set
or STOP a program. Command Keys are presented
in a variety of colors. These keys are normally found
at the bottom of the screen.
Navigation Icons are normally found in the Title Bar
to the left and right. These buttons will take you to
specific areas of the operating system.
Navigation Icons are shown below.
Perform data entry from a scroll
list
Home Menu
Log In (will only be available if
the administrator has activated user
log in)
Settings Menu
Service Menu (only available for
authorized Brookfield service
centers)
Page 15
Figure II-3
Configure Viscosity Test: Create and Run viscosity tests.
Configure Yield Test: Create and Run yield tests.
Load Test: Load a test that has previously been saved or created with PG Flash software. Tests
may be loaded from internal memory or a USB Flash Drive.
View Results: Load Results (saved test data) that have previously been saved. Results may be
loaded from internal memory or a USB Flash Drive.
Manage Files: Manage the file system in the internal memory or on a USB Flash Drive for test
programs and saved data. Create new folder structures, delete files, rename files and move files.
External Mode: Direct the DV3T to communicate with Brookfield RheocalcT Software for
complete rheometer control.
Each of the Main Menu items are detailed in the following sections of the manual.
II.5.1
Viscosity measurements are made on the DV3T Rheometer through the Configure Viscosity Test
function. The user is presented with Configure Viscosity Test at the conclusion of the AutoZero
.
function or by selection on the Home Menu
All elements related to the measurement of viscosity are selected within Configure Viscosity
Test. Tests that are created can be saved to the internal memory of the DV3T Rheometer or onto
a connected USB Flash Drive. Tests can be loaded from memory by selecting Load Test from the
.
Home Screen
TIP: Many aspects of Configure Viscosity Test can be limited when User ID and Log In functions
are implemented (see Section V.4.2).
Page 16
The basic Configure Viscosity Test view is shown in Figure II-4. This view includes the Status Bar
(Section II.3), Title Bar (which includes the Home and Settings icons), test name, test parameters,
the More/Less bar, and Command Keys (see Section II.4).
Status Bar
Title Bar
Test Name
Test Parameters
More/Less Bar
Command Keys
Figure II-4
The user can see the name of any test that has been loaded through the Load Test function. In Figure
II-4, the file name is listed as Unsaved Test, indicating that the current test has not been saved.
The More/Less bar is seen just below the test parameters. In Figure II-4, this bar includes a
down arrow, which indicates that more information is available. Figure II-5 shows the additional
information that can be accessed. The More/Less bar , in this view, now has an up arrow indicating
that the additional information can be hidden.
Status Bar
Title Bar
Data Fields
More/Less Bar
Command Keys
Figure II-5
Page 17
Clear all data that has been entered into the test parameters
and restore the values to the factory default.
Save:
Run:
The Test Parameter area includes many elements of the viscosity test as well as live measurements
of Torque % and Temperature. Temperature data will only be displayed if a Brookfield temperature
probe is connected to the DV3T Rheometer.
Torque:
Spindle:
Speed:
II.5.2
Temperature:
End Condition:
Data Collection:
Instructions:
Create a message that the user will see when the test
begins.
Reports:
QC Limits:
Density:
Yield measurements are made on the DV3T Rheometer through the Configure Yield Test function.
.
The user can select Configure Yield Test from the Home Menu
Brookfield Engineering Labs., Inc.
Page 18
All elements related to the measurement of yield are selected within Configure Yield Test . Tests that
are created can be saved to the internal memory of the DV3T Rheometer or onto a connected USB
.
Flash Drive. Tests can be loaded from memory by selecting Load Test from the Home Screen
TIP: Many aspects of configure Yield Test can be limited when User ID and Log In functions are
implemented (see Section V.4.2).
The basic Configure Yield Test view is shown in Figure II-6. This view includes the Status Bar
(Section II.3), Title Bar (which includes the Home and Settings icons), test name, test parameters,
the More/Less bar, and Command Keys (see Section II.4).
Status Bar
Title Bar
Test Name
Test Parameters
More/Less Bar
Command Keys
Figure II-6
The user can see the name of any test that has been loaded thorough the Load Test function. In
Figure II-6, the file name is listed as Unsaved Test, indicating that the current test has not been saved.
The More/Less bar is seen just below the test parameters. In Figure II-6, the bar includes a
down arrow, which indicates that more information is available. Figure II-7 shows the additional
information that can be accessed. The More/Less bar in this view now has an up arrow indicating
that the additional information can be hidden.
Status Bar
Title Bar
Test Name
Data Fields
More/Less Bar
Command Keys
Figure II-7
Brookfield Engineering Labs., Inc.
Page 19
Clear all data that has been entered into the test parameters
and restore the values to the factory default.
Save:
Run:
The test parameter area includes many elements of the yield test as well as live measurement of
temperature. Temperature data will only be displayed if a Brookfield temperature probe is connected
to the DV3T Rheometer.
Spindle:
Immersion Mark:
Pre-Shear Speed:
Pre-Shear Time:
Zero Speed:
Wait Time:
Specify the duration of a rest period prior to starting the yield test.
Run Speed:
Torque Reduction: Specify the end point of the test. Normally set to 100%.
Temperature:
Yield Limit:
Instructions:
Create a message that the user will see when the test begins.
Save Path:
Page 20
II.5.3
Load Test
Test programs that are created (Configure Viscosity Test/Configure Yield Test) can be saved to the
internal memory of the DV3T or to a USB Flash Drive. These files can be reloaded into the DV3T
for immediate use through the Load Test function. A file that is placed onto a USB Flash Drive can
be loaded onto any DV3T Rheometer.
Within the Load Test function, the user can access the internal memory of the rheometer or any
. The rheometer will point to the USB Flash
USB Flash Drive that is connected to a USB port
Drive according to the order in which it is connected. The first USB Flash Drive that is connected
will be referred to as #1 on both the Load Test screens and the Status Bar (see Section II.3). You
can have as many as three USB Flash Drives connected to the DV3T at any time.
Test files that are displayed on the screen can be sorted by date of creation or by alpha/numeric.
This sorting is selected by pressing the Navigation Icon
or
.
TIP: You can use the Manage Files function
USB Flash Drive.
II.5.4
View Results
Test results (data files) can be saved to the internal memory of the DV3T or to a USB Flash Drive.
Theses files can be reloaded into the DV3T for review, analysis, or printing through the View Results
function. A file of Test Results that is saved onto a USB Flash Drive can be viewed on any DV3T
Rheometer.
Within the View Results function, the user can access the internal memory of the DV3T Rheometer
. The rheometer will point to the
or any USB Flash Drive that is connected to a USB port
USB Flash Drive according to the order in which it is connected. The first USB Flash Drive that
is connected will be referred to as #1 on both the View Results screen and the Status Bar. You can
have as many as three USB Flash Drives connected to the DV3T at any time.
Results files that are displayed on the screen can be sorted by date of creation or by alpha/numeric.
This sorting is selected by pressing the Navigation Icon
or
.
TIP: You can use the Manage Files function
a USB Flash Drive.
II.5.5
Manage Files
Result Files and Test Files can be managed in the internal memory or on USB Flash Drives from the
Manage Files function. Folder structures can be added or changed to assist with data management.
Files may be copied, moved, renamed or deleted. Access to this function can be limited when User
ID and Log In functions are implemented (see Section V.4.2).
Files that are displayed on the screen can be sorted by date of creation or by alpha/numeric. This
or
.
sorting is selected by pressing the Navigation Icon
Page 21
II.5.6
External Mode
The DV3T Rheometer can be controlled from a computer through the use of optional Brookfield
Software RheocalcT. The rheometer must be placed into external control mode from the Main
. The rheometer must be connected to the computer with a USB A cable (DVP-202).
Menu
TIP: The Status Bar will indicate a proper connection to the computer by displaying the Computer
.
Icon
The DV3T will display External Mode when configured for operation with the computer. This
display includes a Return button that will reset the rheometer to stand alone operation.
II.6 Range
The DV3T Rheometer will calculate the measurement range for a specific spindle and speed
combination. This information is displayed on the screen while selecting the spindle number as
shown in Figure II-6. The Range is also shown in the Running Viscosity Test view during the
and is
measurement. Viscosity will be displayed in the unit of measure specified in Settings
set to centipoise (cP) from the factory.
Range
Figure II-6
TIP: The Range value is the same as the AutoRange available on earlier Brookfield rheometer
models.
Page 22
Figure II-7
Measurement data should not be collected when the %Torque reading is out of range. The out of
range condition can be resolved by either changing the speed (reduce speed when reading is out of
range: high) or changing the spindle (increase the spindle size when the reading is out of range: low).
TIP: When comparing data, the test method is critical. Be sure that you know the proper spindle
and speed required for the test method. If readings are out of range, this condition should
be reported as the test result.
Page 23
II.8 Printing
The DV3T Rheometer can communicate to a Dymo Label Writer 450 Turbo label printer. This
printer can be purchased from Brookfield (Part No. GV-1046). The communication to the printer is
by USB (cable provided with the printer). When the printer is connected to the DV3T, the printer
icon will be visible in the status bar.
There is no provision at this time to have other drivers for other printers work with the DV3T.
The DV3T Rheometer can configure the print out for several formats of paper/labels. These
various paper/label stocks are available from Brookfield.
1.
2.
3.
Figure II-8
Printer formats are detailed below:
Data Label (Small)
Page 24
TIP: When printing to a label, if the data set includes more than one point, only the last point
will be printed.
Data Continuous
Page 25
A) RHEOMETER: The DV3T should be turned on, leveled and autozeroed. The level is adjusted
using the two feet on the bottom of the base and confirmed using the bubble level on the front
of the head. Adjust the feet until the bubble is inside the center target. Set the level prior to
autozero
and check the level prior to each measurement.
The proper level is essential for correct operation of the DV3T.
B) SAMPLE: The fluid to be measured (sample) must be in a container. The standard spindles
supplied with the DV3T [LV (1-4), RV (2-7), or HA/HB (2-7)] are designed to be used with
a 600 mL low form Griffin beaker (or equivalent container with a diameter of 8.25 cm). The
same applies to the optional RV1, HA/HB1, and Vane spindles. Many other spindle systems
are supplied from Brookfield with specific sample chambers such as the Small Sample Adapter,
UL Adapter and Thermosel.
Brookfield recommends that you use the appropriate container for the selected spindle. You
may choose to use an alternate container for convenience, however, this may have an effect on
the measured viscosity. The DV3T is calibrated considering the specified container. Alternate
containers will provide results that are repeatable but may not be true.
The LV (1-4) and RV (1-7) are designed to be used with the guardleg attached. Measurements
made without the guardleg will provide repeatable results but may not provide true results.
When comparing data with others, be sure to specify the sample container and presence/
absence of the guardleg.
Brookfield Engineering Labs., Inc.
Page 26
Many samples must be controlled to a specific temperature for viscosity measurement. When
conditioning a sample for temperature, be sure to temperature control the container and spindle
as well as the sample.
Please see our publication, More Solutions to Sticky Problems, for more detail relating to
sample preparation.
III.3 Programming
The DV3T Rheometer provides a powerful programming capability for data collection. The interface
in Configure Viscosity Test allows for control of all instrument parameters including Spindle, Speed,
Temperature (optional), End Condition, Data Collection, Instructions, Reports, Density, and QC Limits
(described in the next sections of this chapter). Any collection of these parameters is considered a
Test Step. The DV3T allows you to create multiple steps to better evaluate your sample material.
Several parameters are only active in the first step including Spindle and Instructions.
Multiple steps in a Test can be useful to evaluate the rheology of a sample or to simply aid in data
collection. A multi step Test that has a change in speed in each step can show Newtonian or non
Newtonian flow behavior. A first step with No Data can be useful when pre shearing is required. A
step with No Data can be useful when a temperature change is effected and some time is required
for thermal equilibrium.
The Configure Viscosity Test screen (see Figure III-1) includes several tools for creating Tests with
multiple steps.
Figure III-1
Navigation / Add:
Use the
to add a new step at the end of the
Test. Use the to navigate to previous steps.
Delete:
Insert:
Two multi step Tests are shown below. The first test shows a three step test where each step has a
different speed with a single data point collection. The results will include three viscosity points,
one for each speed. The second test shows a two step test where the first step is a pre shear step.
The pre shear is conducted at 200 rpm for one minute and no data is collected. The second step
generates a single data point.
Three Step Test:
Page 28
The process of selecting a spindle and speed for an unknown fluid is normally trial and error. An
appropriate selection will result in measurements made between 10-100 on the instrument %
torque scale. There are two general rules that will help in the trial and error process:
In other words: to measure high viscosity, choose a small spindle and/or a slow speed. If the chosen
spindle/speed results in a reading above 100%, then reduce the speed or choose a smaller spindle.
Experimentation may reveal that several spindle/speed combinations
will produce satisfactory results between 10-100%. When this
circumstance occurs, any of the spindles may be selected.
Non-Newtonian fluid behavior can result in the measured viscosity and
yield stress changing if the spindle and/or speed is changed. See our
publication, More Solutions to Sticky Problems, for more details.
When viscosity data must be compared, be sure to use the same
test methodology: namely the same instrument, spindle, speed,
container, temperature and test time.
Figure III-4
Page 29
Figure III-5
Manual No. M13-167-A0415
Figure III-6
Select Temperature Controller and you will be presented with the Default Temperature menu (see
Figure III-7). This menu allows the user to 1) initiate communication with the temperature controller
and 2) set a default temperature.
Brookfield Engineering Labs., Inc.
Page 30
Figure III-7
Initiate temperature control by pressing the Test Connection. The connection status will be confirmed
with a message box and the appearance of the bath icon in the status bar (see Section II.3).
The Default Temperature is the temperature value that the control device will return to at the
conclusion of a Test. This value is set only if the check box is checked. When the check box is
unchecked, the temperature controller will hold at the last set point used in the Test.
TIP: When a test involves several temperature set points, set the Default Temperature to the first
set point used in the Test. This will reduce the transition time from one Test run to the next
Test run.
Note that temperature controllers can also be controlled through the Rheocalc T software when
they are connected to the computer using the proper cable. The Thermosel requires the HT-106
cable (RS-232), available from Brookfield. The water bath requires the 225-173 cable (RS-232),
which comes with the bath.
Page 31
allow for a more complete characterization of a fluid and may help in formulation and production
of a product. The DV3T is capable of collecting multiple data points for comprehensive analysis
of flow behavior.
More information on flow behavior, shear rate and rheograms is available in our publication, More
Solutions to Sticky Problems.
Figure III-8
Figure III-9
Single Point: Collect only a single data point when the End Condition is met.
Single Point Averaging: Specify an amount of time over which to average measured data.
Collect a single data point when the End Condition is met. This
data point is an averaged value. If the time for averaging is shorter
than the total time for the step, then the average will be performed
for the specified time at the end of the test.
Example 1: The End Condition is Time with a value of 1 minute and 30 seconds, the Single
Point Averaging Duration is 30 seconds, the single data point collected from this
step will be an average of the data measured from 1 minute to 1 minute 30 seconds.
Multi Point: Collect multiple data points based on time. The Data Interval is specified in
Hours:Mins:Secs. If the End Condition is set to Time, then the total number of
Brookfield Engineering Labs., Inc.
Page 32
points will be calculated and displayed in the Data Collection screen. If the End
Condition is not based on time then it is possible that the step will conclude prior
to a data point being collected. If you want a data point at the conclusion of the
step regardless of the time interval, you can check the check box in the Data
Interval screen.
Example 2: End Condition is Time = 2 minutes, Multi Point Data Interval is 10 seconds. Total
points collected will be 12 with the last data point taken in the last second of the
step.
Example 3: End Condition is set to Viscosity = 200 cP, Multi Point Data Interval is 10 seconds.
During the test the total time required to reach 200 cP is 65 seconds. Total points
collected will be 6 with the last data point taken at 60 seconds, 5 seconds before
the test is finished.
Example 4: End Condition is set to Viscosity = 200 cP, Multi Point Data Interval is 10 seconds.
Check the check box to Also Collect Single Point at Step End. During the test, the
total time required to reach 200 cP is 65 seconds. Total points collected will be 7
with the last data point taken at 65 seconds, 5 seconds after point #6 taken at 60
seconds.
Multi Point Averaging: Specify an amount of time over which to average measured data.
Collect multiple data points based on time until the End Condition is
met. Each data point is an averaged value. If the Averaging Duration
is shorter than the Data Interval for the step, then the average will be
performed for the specified time at the end of the Data Interval. The
total number of points to be collected will be displayed if the End
Condition is set to Time. If the End Condition is not based on time,
then it is possible that the step will conclude prior to a data point
being collected. If you want a data point at the conclusion of the step
regardless of the time interval, you can check the check box in the
Data Interval screen.
Example 5: End Condition is Time = 2 minutes, Multi Point Data Interval is 10 seconds.
Averaging Duration is 5 seconds. Total points collected will be 12 with the last
data point taken in the last second of the step. Each data point will be an average
of the data measured in the last 5 seconds of each Data Interval.
Example 6: End Condition is set to Viscosity = 200 cP, Multi Point Data Interval is 10 seconds.
Averaging Duration is 5 seconds. During the test, the total time required to reach
200 cP is 63 seconds. Total points collected will be 6 with the last data point taken
at 60 seconds, 3 seconds before the test is finished. Each data point will be an
average of the data measured in the last 5 seconds of each Data Interval.
Example 7: End Condition is set to Viscosity = 200 cP, Multi Point Data Interval is 10 seconds.
Averaging Duration is 5 seconds. Check the check box to Also Collect Single Point
at Step End. During the test, the total time required to reach 200 cP is 65 seconds.
Total points collected will be 7 with the last data point taken at 65 seconds, an
average of the 5 seconds after point #6.
Example 8: End Condition is set to Viscosity = 200 cP, Multi Point Data Interval is 1 minute.
Averaging Duration is 20 seconds. Check the check box to Also Collect Single
Point at Step End. During the test, the total time required to reach 200 cP is 10
minutes 40 seconds. Total points collected will be 11 with the last data point taken
at 10 minutes 40 seconds, an average of the last 20 seconds of the step.
Page 33
Figure III-10
Time: The test will complete when the specified amount of time has elapsed. Time
is entered in Hours, Minutes, and Seconds.
Hours: 0-99
Minutes: 0-59
Seconds: 0-59
A time of zero hours, minutes and seconds may be selected. With this End
Condition, the DV3T will operate at the selected speed until the operator
selects Stop Test. Data will be collected according to the Data Collection
setting (see Section III.7).
TIP: An End Condition of zero Time can be useful when measuring a new
material. During the test, the speed can be changed without ending and
then rerunning the test (see Section III.10). This method can allow you
to quickly evaluate the spindle selection to determine the best speeds
for testing.
# of Points: The test will complete when the specified number of data points has been
collected. Data is collected according to the Data Collection setting (see
Section III.6). The range of data points is: 1 9,999.
Brookfield Engineering Labs., Inc.
Page 34
# of Revolutions: The test will complete when the specified number of revolutions of the spindle
has occurred. Data is collected according to the Data Collection setting (see
Section III.6). The range for number of revolutions is: 1 9,999.
Torque: The test will complete when the specified Torque value is measured. Data
is collected according to the Data Collection setting (see Section III.6). The
range of values for the measured Torque is: -10.0 100.0.
Viscosity: The test will complete when the specified Viscosity value is measured. Data
is collected according to the Data Collection setting (see Section III.6). The
range of measured Viscosity is 0 10,000,000,000 cP.
Note: The Viscosity End Condition should be selected in consideration of the
Range provided by the spindle and speed selected. Check the Range
value by selecting Spindle in Configure Viscosity Test. If you choose
a Viscosity End Condition greater than the Range of the spindle and
speed settings, then the value shown in the End Condition field will
be EEEE. In this case, the end condition will never be achieved.
Temperature: The test will complete when the specified Temperature value is measured
to within the indicated Tolerance. The Tolerance specifies how close the
measured temperature should be to the specified temperature value to consider
the target reached. Temperature is measured through the use of a connected
Brookfield temperature probe (DVP-94Y, CPA-44ZPY cup, SC4-XXRPY
chamber). Data is collected according to the Data Collection setting (see
Section III.6).
Temperature: -100 300 C
Tolerance: 0.0 9.9 C
TIP: A small tolerance value will require a much longer time to reach the
End Condition.
TIP: Consider using a larger Tolerance value for the End Condition and
then use a second step for thermal equilibrium (End Condition Time).
This two step method can reduce the variability in test time caused by
variations in temperature sensors.
III.9 Additional Test Parameters
The Configure Viscosity Test screen includes a More/Less Bar. When pressed, the More Bar will
present several additional Test Parameters including: QC Limits, Instructions, Reports, and Density.
These parameters are always available and active regardless of the position of the More / Less Bar.
QC Limits: Select an acceptable range for measurement results. The range may be defined
by Viscosity, Torque, Time, Temperature, or Shear Stress. The possible range for
Viscosity and Shear Stress will be defined by the spindle and speed selected. QC
Limits are a visual and audible signal to the operator during the test. The data set
does not include an indication of QC Limits violation.
A violation of the QC Limits during the test will be indicated by a flashing yellow box around the
display for the specified parameter. Note that Time does not have a dedicated display. A violation
will also result in an audible beep and a one time warning message (see Figure III-11). Data will
continue to be collected while the warning message is displayed.
Brookfield Engineering Labs., Inc.
Page 35
Figure III-11
When Viscosity or Torque are the selected QC Limits parameter, QC Limits will be shown during
the test on the Real Time Graph if that parameter is selected for display (see Figure III-12). The
QC Limits are represented by the dashed lines.
TIP: Audible alarms may be turned off in the User Settings
TIP: An indication of a QC Limit violation is not part of the data set. The user can record a
violation by using the Notes available when viewing test Results (see Section III.11).
Trend Bar
Figure III-12
Instructions: Record specific instructions to the operator. This information will be presented
immediately when the program is Run (see Figure III-13). The operator is required
to acknowledge the message before the program will continue.
Page 36
Figure III-13
TIP: If the operator selects Do not show this message again within the
Instructions message box, then Instructions will no longer be displayed for
any test. This condition can be reset to allow Instructions to be displayed
.
in the Display section of User Settings
TIP: Instructions can be viewed during the test by selecting instructions (more/
less bar) in the View Test screen.
Reports: Define how data will be presented at the conclusion of the test and specify the path for
saving data files. Upon the conclusion of a test, the user can be presented with either: a
Graph, Table, Math Model, or an Average of collected data points. Once in the Results
can be used to change the view.
section, the down arrow
Post Test Averaging can be specified as either Test Averaging or Step Averaging. Test
Averaging allows you to select which step from the program (created with PG Flash) is
to be averaged and displayed. Step Averaging is used when there is a single test step.
See Section III.10 for more detail relating to Data Averaging.
The data path for saving data can be preselected as part of the test report.
TIP: Selecting a data path can help to organize data storage in the internal
memory of DV3T. By preselecting the data path, operators are directed
to put data from specific tests into specific areas.
TIP: If you select a data path that includes a USB Flash Drive location and the
USB Flash Drive has been removed, an error message will be displayed.
Density: Input a density value for the sample to be tested. This value will be used to calculate
viscosity when kinematic viscosity units have been chosen for display (see Section
V.4.2).
Page 37
Measurement Data
Command Keys
Figure III-14
The Running Viscosity Test screen provides information on the current measurement including:
Torque, Viscosity, Shear Stress, Shear Rate, Temperature and Speed.
Torque is the deflection of the Rheometer torque sensor. It is described as a percent (%) and has
a range of 0 100%. The DV3T will provide measurement results within the stated accuracy
provided the Torque reading is between 10 and 100%. If the Torque reading falls below 10% the
labels in the data fields will flash to indicate an error condition. Brookfield does not recommend
that data be collected below 10% Torque; however, data collection is not restricted.
TIP: Torque on the DV3T is equivalent to the dial reading from the Brookfield Dial Reading
Viscometer or the % reading from Brookfield Digital Viscometers (DV-E, DV-I,
DV-II, DV-III).
Viscosity is calculated from the measured Torque based on the selected spindle and speed of
rotation. The units of viscosity are defined in the Global Settings section of the Settings Menu
. If the Torque reading falls below 10%, the labels in the data fields will flash to indicate
an error condition.
Shear Stress is calculated from the measured Torque based on the selected spindle. The units of
. If the Torque
shear stress are defined in the Global Settings section of the Settings Menu
reading falls below 10%, the labels in the data fields will flash to indicate an error condition.
TIP: Shear Stress will be displayed as zero for spindles that do not have SRC values.
Shear Rate is calculated from the selected speed based on the selected spindle.
TIP: Shear Rate will be displayed as zero for spindles that do not have SRC values.
Brookfield Engineering Labs., Inc.
Page 38
Temperature is the input value from a connected Brookfield temperature probe. The DV3T is
provided with a DVP-94Y probe that can be inserted into the test sample or a water bath. The
Cone/Plate version of the DV3T can be utilized with a sample cup that includes an embedded
temperature probe. Some Brookfield accessories include temperature probes (Thermosel) or
optional chambers with embedded temperature probes (Small Sample Adapter). The units of
.
temperature are defined in the Global Settings section of the Settings Menu
The temperature display will be ---- when no temperature probe is connected.
Speed is the selected speed from the Configure Viscosity Test screen.
TIP: Speed can be changed by switching from the Running Viscosity Test screen to the
View Test screen using the command key at the bottom of the screen.
TIP: When Speed is changed in the View Test screen, the Test Name will be changed to
unsaved.
The status of the test configuration is indicated in the grey bar above the data fields. A configuration
that has been saved will be indicated by a display of the test file name. A configuration that is not
saved will be indicated by a display of Unsaved Test.
The Real Time Graph is displayed at the bottom of the Running Viscosity Test screen. This graph
shows all measurements during the life of the test. The Y-axis can represent any data field. The
data field represented on the graph is shown by the blue dot; for example, in Figure III-14, Torque
is shown on the graph. Any data field can be selected by touching the screen.
The Real Time Graph will indicate data points collected as part of the test with a blue dot on the
data line. At the conclusion of the test, the Real Time Graph is displayed for review. Once the OK
button is pressed, only the data points collected as part of the test will be available for review/save.
The Real Time Graph cannot be seen again once the Results page is in view.
QC Limits will be indicated on the Real Time Graph when the selection for Y-Axis is the same as
the selection for QC Limits (see Section III.9).
Two Command Keys are available on the Running Viscosity Test screen: Stop Test and View Test.
Stop Test: Immediately stops the current test. The user will be presented with a confirmation
box. The test will continue to run until Yes is selected in the Confirmation box.
If any data had been collected during the test, the user will be presented with the
Results screen (see Section III.11). If data was not collected, the user will be
returned to the Configure Viscosity Test screen.
TIP: If the operator selects Do not show this message again within the
Confirmation box, then the Confirmation box will no longer be displayed when
Stop Test is pressed. This condition can be reset in the Display section of User
.
Settings
View Test: Change the view from measurement to test parameters including: Spindle, Speed,
Temperature, End Condition, Data Collection, QC Limits, Instructions, and Density.
The View Test screen includes a live display of Torque and a Time parameter that
shows the total elapsed time for the test (see Figure III-15).
Page 39
Figure III-15
The More/Less Bar can be used to reduce the number of parameters shown in the display.
The Speed parameter is active in this view. The operator can change the speed of test without
returning to the Configure Viscosity Test screen.
TIP: If the speed is changed during the execution of a saved test, then the test status will be
changed to Unsaved Test. This will also be reflected if the collected data is saved.
View Test includes the Stop Test command key and a Back Arrow navigation key. The Back Arrow
is used to return the display to the Running Viscosity Test view.
III.11 Results
Measurement data is viewed in the Results screen. This screen is presented at the conclusion of a
test or when data is loaded through the View Results selection from the Home Menu.
The DV3T utilizes a comprehensive data format. Data files include the complete set of measurement
results and calculated values along with the test protocol. All elements of the test can be viewed
in the Results screen.
The DV3T Rheometer allows for 5000 total data points per file. When viewing large data files,
additional time is required when moving from the various Results options listed below. There may
be some delay on the screen while the DV3T prepares the data.
Page 40
The Results screen includes several Navigation Icons and Command Keys.
Home:
Down Arrow:
Blue Arrow:
Print:
Save:
Save data.
Scroll Bar:
The Results screen offers several options for viewing test data. These options are accessed via the
Down Arrow in the Navigation Bar at the top of the screen (see Figure III-16).
Figure III-16
Brookfield Engineering Labs., Inc.
Page 41
Table: Display all data points. Data will be shown in a scroll list where each page can hold a
maximum of 50 points. If the data set has more than 50 points, then additional pages
will be indicated at the bottom of the screen. Additional pages of data can be accessed
through the use of the Blue Arrows
Each data point includes: Viscosity, Torque, Speed, Temperature, Time, Shear Stress
(SS), Shear Rate (SR), Density, and Accuracy.
Post Test Averaging : Average and Standard Deviation are calculated for measured and calculated
parameters including: Viscosity, Torque, Shear Stress, and Temperature.
Post Test Averaging is calculated regardless of the Data Collection setting.
If Data Collection was set to Multipoint Averaging, then the Post Test
Averaging will calculate an average of the averaged data.
When using multiple step programs, the test step utilized for Post Test
Averaging is specified in the Report setting (see Section III.9).
Test Used: Display the test elements used to generate the data set. In this view, the Configure
Test button is available. Selecting Configure Test will program the DV3T to run
the same test utilized to collect the data set being viewed and present the Configure
Viscosity Test screen.
Notes: Document any relevant information about the test or data. This information will be
stored with the data set once saved.
Compare Results: Load two data sets and view side by side.
Figure III-17
Device Info: Display basic information about the data file and the specific DV3T Rheometer
used to collect the data including: date and time the test started, completed and
was saved; if user accounts are active, which user saved the data; rheometer serial
number, firmware version, rheometer Torque range (LV, RV, HA, HB, other).
Page 42
Graph: Display a graph of the collected data. The axis will default to the values selected in the
Report section of the Test. The value displayed on each axis can be changed by selecting
from the blue drop down arrow in each field. Individual data points can be selected
and the X / Y coordinate values will be displayed. The selected point is represented
by a blue dot (other points are shown as grey dots).
TIP: Some spindle selections do not allow for the calculation of shear stress and
shear rate. Shear Stress and Shear rate will be represented as zero for all data
points for this type of spindle.
TIP: If the temperature probe is not connected to the DV3T, then the temperature
field will not contain any data. The Graph will not display temperature under
this condition.
The Graph can be printed if there is an attached printer. The buttons at the bottom of
the graph are for zooming in, zooming out, and resetting.
Figure III-18
Math Model: Analyze collected data through several defined equations. The selected math
model will default to the selection from the Report section of the Test. You can
change the selected model by pressing the blue drop down arrow.
Page 43
Figure III-19 shows the Power Law equation for a typical set of data. A brief
definition of the equation will be shown if the i (information) button is pressed.
Figure III-19
Detailed information about Math Models is presented in Section VII.
Page 44
Multi Point Averaging requires a definition of 1) Data Interval, the frequency of data collection
and 2) Averaging Duration, the amount of time for which readings will be averaged. These two
parameters will work in conjunction to generate multiple data points each of which represent
the average of all data measured during the average duration within the specified Data Interval.
Post Test Averaging of data occurs after the test is complete, through the Results screen. Averages
and standard deviation values can be generated for data collected in a single step or across several
steps. If the test used to collect data utilized Live Averaging (described above), then the Post Test
Averaging will produce an average value of averaged data points.
Post Test Averaging can be specified in the Report section of Configure Viscosity Test (see Section
III.9). When configured as part of the test, the average values will be displayed immediately
upon test conclusion. If not specified in Report, Post Test Averaging can be selected from the
Down Arrow in the Results screen (see Section III.11).
Post Test Averaging offers two options: Step Averaging and Test Averaging (see Figure III-20).
Figure III-20
Page 45
Step Averaging: Calculate Average and Standard Deviation for all data collected within a single
step test. Step Averages will be displayed as shown below in Figure III-21.
Figure III-21
Test Averaging: Calculate Average and Standard Deviation for all data collected within step
specified. Test Averages will be displayed as shown below in Figure III-22.
Figure III-22
Page 46
A) RHEOMETER: The DV3T should be turned on, leveled and autozeroed. The level is adjusted
using the two feet on the bottom of the base and confirmed using the bubble level on the front
of the head. Adjust the feet until the bubble is inside the center target. Set the level prior to
autozero and check the level prior to each measurement.
The proper level is essential for correct operation of the DV3T.
B) Sample: The measured yield value is very dependent on how the sample is handled prior to the
measurement. The sample should be disturbed as little as possible prior to inserting the vane
spindle. Brookfield recommends testing the sample in the container in which it is provided when
possible. If it is necessary to transfer the sample to a new container, then it is recommended to
include a pre-shear and/or wait time in the Yield Test. This pre-shear and wait-time will provide
a common history to each sample to aid in sample to sample comparison.
Brookfield recommends the use of vane spindles for measuring yield value. The vane spindle
provides a unique geometry that minimizes disturbance of the sample during spindle insertion.
Many samples must be controlled to a specific temperature for yield measurement. When
conditioning a sample for temperature, be sure to temperature control the container and spindle
as well as the sample.
Page 47
In other words: to measure high yield, choose a small spindle and/or a slow speed. If the chosen
spindle/speed results in a reading above 100%, then reduce the speed or choose a smaller spindle.
Experimentation may reveal that several spindle/speed combinations will produce satisfactory
results between 10-100%. When this circumstance occurs, any of the spindles may be selected.
When yield data must be compared, be sure to use the same test methodology: namely the
same instrument, spindle, speed, container, temperature and test time.
Vane spindles are optional equipment and are not part of the standard DV3T package. DV3TLV
Rheometers are provided with a set of four spindles and a narrow guardleg; DV3TRV Rheometers
come with a set of six spindles and a wider guardleg; DV3THA and DV3THB Rheometers come
with a set of six spindles and no guardleg. (See Appendix G for more information on the guardleg.)
Do not use the guard leg with vane spindles.
The spindles are attached to the rheometer by screwing them onto the coupling nut on the lower
shaft (see Figure IV-1). Note that the spindles have a left-hand thread. The lower shaft should
be secured and slightly lifted with one hand while screwing the spindle to the left. The face of
the spindle nut and the matching surface on the lower shaft should be smooth and clean to prevent
eccentric rotation of the spindle. Spindles can be identified by the number on the side of the spindle
coupling nut.
The motor should
be OFF whenever
spindles are being
removed or attached.
If your instrument has the EZ-Lock
system, the spindles are attached as
follows:
With one hand hold the spindle, while
gently raising the spring-loaded outer
sleeve to its highest position with
the other hand, as shown in Figure
IV-2. Insert the EZ-Lock Spindle
Coupling so that the bottom of the
coupling is flush with the bottom of
the shaft, and lower the sleeve. The
sleeve should easily slide back down
to hold the spindle/coupling assembly
Figure IV-1
Page 48
Figure IV-2
Manual No. M13-167-A0415
in place for use. [Spindles can be identified by entry code; look for the number on the side of the
EZ-Lock spindle coupling.]
Note:
Keep the EZ-Lock Spindle Coupling and outer sleeve as clean as possible and free from
debris that could become lodged inside the adapter.
Figure IV-3
Figure IV-4
TIP: Embedded temperature probes used with the Cone/Plate and Small Sample Adapter can be
used for set point control of a water bath. Making this selection can improve the temperature
control of the sample by off setting the temperature bath to account for any heat loss in
the tubing or water jacket. To use the embedded temperature probe select Probe in the
Temperature menu.
Temperature control must be initiated each time the DV3T is powered on. Be sure that the
temperature control device is powered on prior to initiating communication. Choose Device Setup
Brookfield Engineering Labs., Inc.
Page 49
in the Settings menu. Choose Temperature and you will be presented with the Temperature menu
(see Figure IV-3). Temperature Offset is described in Section IV.1.
Select Temperature Controller and you will be presented with the Default Temperature menu (see
Figure IV-4). This menu allows the user to 1) initiate communication with the temperature controller
and 2) set a default temperature.
Initiate temperature control by pressing the Test Connection. The connection status will be confirmed
with a message box.
The Default Temperature is the temperature value that the control device will return to at the
conclusion of the Test. This value is set only if the check box is checked. When the check box is
unchecked the temperature controller will hold at the last set point used in the Test.
TIP: When a test involves several temperature set points, set the Default Temperature to the first
set point used in the Test. This will reduce the transition time from one Test run to the next
Test run.
Primary
Immersion
Mark
Secondary
Immersion
Mark
Figure IV-5
Brookfield Engineering Labs., Inc.
Page 50
Page 51
Torque Reduction
Torque Reduction is percentage change in torque value between successive data points. That is, the
material yields or begins to break down and, as a result, the measured incremental torque begins to
decrease. A value of 100% for this parameter causes the test to stop as soon as there are no torque
increases during a base time increment. Some users may wish to see a drop in torque after the
yield point. Setting this parameter to values greater than 100% allows data to be collected after the
yield point by the EZ-YieldTM software so the decrease in torque may be more easily visualized.
However, this will also affect the calculated yield stress value.
TIP: Brookfield recommends a torque reduction value of 100%.
Yield Limits
These stress values can be used as a Quality Control tool. If the resulting yield stress from a test
falls outside these limits, an appropriate message is displayed and printed with the results. Entering
zero (0) for both these parameters disables this QC feature.
Instructions
Record specific instructions to the operator. This information will be presented immediately
when the program is Run (see Figure IV-6). The operator is required to acknowledge the message
before the program will continue.
Figure IV-6
TIP: If the operator selects Do not show this message again within the Instructions message
box, then Instructions will no longer be displayed for any test. This condition can be
.
reset to allow Instructions to be displayed in the Display section of User Settings
TIP: Instructions can be viewed during the test by selecting instructions (more/less bar) in
the View Test screen.
Save Path
Define the memory location (internal or USB Flash Drive) where the Test Results will be saved.
TIP: Selecting a data path can help to organize data storage in the internal memory of the
DV3T. By preselecting the data path, operators are directed to put the data from specific
tests into specific areas.
TIP: If you select a data path that includes a USB Flash Drive location and the Flash Drive
has been removed, an error message will be displayed.
Brookfield Engineering Labs., Inc.
Page 52
Measurement Data
Command Keys
Figure IV-7
The Running Yield Test screen provides information on the current measurement including: Delta
Torque, Torque, Stress, Strain, Temperature, and Speed.
Delta Torque is the difference between the current %Torque value and the previous %Torque
value. As the sample approaches its yield point, the Delta Torque percentage should decrease.
If the Torque Reduction for the test is 100%, the yield point occurs when the % decrease in
Torque = 100.
Torque is the deflection of the Rheometer torque sensor. It is described as percent (%) and has
a range of 0 100%. The DV3T will provide measurement results within the stated accuracy
provided the Torque reading is between 10 and 100%. If the torque reading falls below 10% the
labels in the data fields will flash to indicate an error condition. Brookfield does not recommend
that data be collected below 10% Torque however data collection is not restricted.
TIP: Torque on the DV3T is equivalent to the % reading from the DV-III Ultra and the YR-1
Rheometer.
Stress is the calculated stress based on the Speed, Torque, and Spindle.
Strain is the Apparent Strain placed on the sample. The Apparent Strain is the angular distance
that the spindle rotates in the sample.
Temperature is the input value from a connected Brookfield temperature probe. The DV3T is
provided with a DVP-94Y probe that can be inserted into the test sample or a water bath. The
Cone/Plate version of the DV3T can be utilized with a sample cup that includes an embedded
temperature probe. Some Brookfield accessories include temperature probes (Thermosel) or
Brookfield Engineering Labs., Inc.
Page 53
optional chambers with embedded temperature probes (Small Sample Adapter). The units of
temperature are defined in the Global Settings section of the Settings Menu
.
The temperature display will be ---- when no temperature probe is connected.
Speed is the selected speed from the Configure Yield Test screen.
The status of the test configuration is indicated in the grey bar above the data fields. A configuration
that has been saved will be indicated by a display of the test file name. A configuration that is not
saved will be indicated by a display of Unsaved Test.
The Real Time Graph is displayed at the bottom of the Running Yield Test. This graph shows all
measurements during the life of the test. The Y-Axis can represent any data field. The data field
represented on the graph is shown by the blue dot; for example, in Figure IV-7, Torque is shown
on the graph. Any data field can be selected by touching the screen.
The Real Time Graph will indicate data points collected as part of the Test with a blue dot indicating
the final point. At the conclusion of the test, the Real Time Graph is displayed for review. Once
the OK button is pressed, the screen will transition to Results.
QC Limits will be indicated on the Real Time Graph when the selection for Y-Axis is the same as
the selection for QC Limits (see Section III.9).
Two Command Keys are available on the Running Yield Test Screen: Stop Test and View Test.
Stop Test: Immediately stop the current test. The user will be presented with a confirmation
box. The test will continue to run until Yes is selected in the Confirmation box. The user will
be presented with the Results screen with any data that was collected.
View Test: Change the view from measurement to the test parameters. The View Test screen
includes a live display of temperature. View Test includes the Stop Test command key and a
Back Arrow in the Navigation Bar. The Back Arrow is used to return the display to the Running
Yield Test view.
IV.7 Results
Measurement data is viewed in the Results screen. This screen is presented at the conclusion of a
.
test or when data is loaded through the View Results selection from the Home Menu
The DV3T utilizes a comprehensive data format. Data files include the complete set of measurement
results and calculated values along with the test protocol. All elements of the test can be viewed
in the Results screen.
The DV3T Rheometer allows for 5000 total data points per file. When viewing large data files,
additional time is required when moving from the various Results options listed below. There may
be some delay on the screen while the DV3T prepares the data.
Page 54
The results screen includes several Navigation Icons and Command Keys.
Home:
Down Arrow:
Blue Arrow:
Print:
Save:
Save data.
Scroll Bar:
The Results screen offers several options for viewing test data. These options are accessed via the
Down Arrow in the Navigation Bar at the top of the screen (see Figure IV-8).
Figure IV-8
Brookfield Engineering Labs., Inc.
Page 55
Table: Display all data points. Data will be shown in a scroll list where each page can hold a
maximum of 50 points. If the data set has more than 50 points, then additional pages
will be indicated at the bottom of the screen. Additional pages of data can be accessed
through the use of the Blue Arrows
Each data point includes: Viscosity, Torque, Speed, Temperature, Time, Shear Stress
(SS), Shear Rate (SR), Density, and Accuracy.
Test Used: Display the test elements used to generate the data set. In this view, the Configure
Test button is available. Selecting Configure Test will program the DV3T to run
the same test utilized to collect the data set being viewed and present the Configure
Viscosity Test screen.
Notes: Document any relevant information about the test or data. This information will be
stored with the data set once saved.
Compare Results: Load two data sets and view side by side (see Figure IV-9).
Device Info: Display basic information about the data file and the specific DV3T Rheometer
used to collect the data including: date and time the test started, completed and
was saved; if user accounts are active, which user saved the data; rheometer serial
number, firmware version, rheometer Torque range (LV, RV, HA, HB, other).
Graph: Display the final data point and graph (see Figure IV-10). The graph axis (X and Y)
can be changed by selecting the blue drop down arrow in the appropriate data field.
Zoom and reset controls are indicated under the X-Axis.
Figure IV-9
Figure IV-10
Page 56
V. SETTINGS
The Settings menu provides access to the many controls and features of the DV3T Rheometer. This
which is often present in the Title Bar.
menu can be accessed through the Settings Navigation Icon
Figure V-1 shows the Settings Menu which is divided into: Device Setup, User Settings, Global Settings,
and Admin Functions. Global Settings include items that affect the complete range of features within
the DV3T. Admin Functions include items related to administrator level controls.
Figure V-1
Page 57
The temperature menu provides access to three functions: Temperature Offset, Temperature Controller
Access, and Temperature Display/Settings.
Figure V-2
Temperature: The Temperature Offset setting allows the user to create up to 10 temperature
offset values for connected temperature probes. Brookfield offers several types of
temperature probes for use with the DV3T including: DVP-94Y immersion probe
(supplied with the DV3T), CPA-44PYZ embedded probe (optional with Cone/Plate
DV3T Rheometers), and SC4-13RP embedded probe (optional with Small Sample
Adapter). Any of these probes can be calibrated locally against a standard reference
thermometer to determine an offset (how far from the actual temperature does the
probe read). This offset can be entered into the DV3T Rheometer and identified
with a name defined by the user.
When Temperature is selected in the Device Setup menu, the Temperature Offset
menu is presented (see Figure V-3). From this menu you can create new offset values
by pressing the Add Probe Offset command key at the bottom of the screen and
you can select which offset to utilize with the DV3T by pressing the circle beside
the name. In Figure IV-3, the offset DVP-94Y 3 has been selected as indicated by
the blue circle.
Page 58
Figure V-3
The creation of a Temperature Offset requires input of the offset value and a name. The
offset value must be in the range of -9.9 to 9.9C. The Name can be up to 14 characters
long. To delete an existing Temperature Offset, first select the offset from the list, then
press the Delete command key at the bottom of the screen.
The use of a Temperature Offset will be indicated in the Temperature field displayed in
the Configure Viscosity Test screen with a (o) beside the Live indication.
The Temperature menu also allows the user to initiate control over a connected
temperature control device such as the Brookfield Thermosel (HT-106 controller) or
Water Bath (model TC-XX0SD or TC-XX0AP). See Section II.4 for further details.
Establish control over the external temperature controller by selecting Temperature
Controller in the Temperature menu (see Figure V-4).
Figure V-4
Brookfield Engineering Labs., Inc.
Page 59
Communication is established when the Test Connection command button is pressed. You must
press this button every time the DV3T Rheometer is powered on. Be sure that the temperature
controller is turned on prior to pressing the Test Connection button.
The Default Temperature screen also allows the user to set a Default Temperature. The Default
Temperature is the set point value that will be used by the DV3T when a test (Viscosity Test or
Yield Test) is not in use. To set the Default Temperature first, enter a temperature value and second
check the Default Between Tests check box.
TIP: Use a Default Temperature when the test program includes a change in temperature and
set the value equal to the first temperature control point in the test. At the conclusion
of the test, the control device will be reset to the first temperature set point which will
minimize the time required between tests.
The Temperature screen also allows you to choose which temperature probe will be used for data
collection and temperature control. When the DV3T is controlling temperature through an external
device, there may be two temperature probes available for data collection: 1) the temperature probe
connected to the DV3T (Probe) and 2) the temperature probe associated with the temperature
control device (Controller). Press either Probe or Controller in the Temperature menu, figure the
first one above.
Printer Setup: The DV3T is configured for use with the Dymo 450 printer. No other printers can
be utilized with the rheometer. Printer Setup allows the user to choose from three
different configurations for print out.
Label Small: A small label (1.13inch x 3.5inch) that allows for printing of limited
data for a single data point. The label stock is available from
Brookfield; part number GV-1048.
Label Large: A large label (2.31inch x 4.0inch) that allows for printing of a
complete set of parameters for a single data point. The label stock
is available from Brookfield; part number GV-1049.
Receipt: A continuous roll of paper that allows for printing of multiple data
points. The paper stock is available from Brookfield; part number
GV-1047.
The Printer Setup is selected by touching the desired configuration, which will
result in a blue circle to appear to the left of the item. The OK command key must
be selected to confirm the setting.
AutoZero: AutoZero is an operation performed by the DV3T Rheometer automatically during
the power up sequence. This operation sets the instrument transducer to a correct zero
value. The zero value should not shift over time, however, if the user determines that
the zero is not correct, you can force an AutoZero to be performed without powering
down the rheometer. When the AutoZero function is selected, the DV3T is immediately
sets to AutoZero mode and the user is presented with the AutoZero screen (Figure II-2).
Upon completion, the user will be returned to the Device Setup menu.
Oscillation Check: The Oscillation Check is a check of the performance of the lower bearing of the
DV3T Rheometer. This lower bearing is either Point and Jewel or Ball Bearing.
Correct performance of the lower bearing is essential for proper operation of the
rheometer. A failed lower bearing will normally cause viscosity measurements
to be lower than expected.
Page 60
Figure V-5
A successful Oscillation Check will show a very smooth decrease of %Torque
with a final value within +/-0.2 of 0.0. Any value above 0.2 % or below -0.2%
indicates that a calibration check should be performed (see Appendix F).
TIP: Higher Torque ranges (HA and HB) will quickly return to the zero
value without displaying much oscillation. This is normal behavior.
The Oscillation Check is not a guarantee of proper calibration. It is only an
indication of the performance of the lower bearing of the rheometer. Calibration
can only be verified through the use of a calibration standard such as Brookfield
Silicone and Mineral Oil standards.
Technical Support Info: The information contained in the Technical Support Info menu is designed
to support Brookfield for problem resolution. When contacting Brookfield
for support, you may be asked to provide information contained in one of
the menu choices within Technical Support Info. There are no settings
within this section and the performance of the DV3T Rheometer can
not be affected by accessing this section.
About: The About screen is shown during the start up sequence of the DV3T Rheometer. Selecting
About in the Device Setup menu will show the same information (see Figure V-6). Information
Brookfield Engineering Labs., Inc.
Page 61
in the About screen includes: Torque Range (LV, RV, HA, HB), Model Number, Firmware
Version, Rheometer Serial Number, TK version (torque table), Spindle version (spindle
table).
Figure V-6
TIP: The information on the About screen will be needed when contacting Brookfield
for support.
To return from the About screen, press the Left Arrow navigation button in the Title Bar.
TIP: Each User can set their own preferences for Sound and Display.
Sound: The Sound menu provides adjustment of the volume for the sounds utilized by the DV3T.
Additionally, specific sounds can be toggled on/off including: Button Click, Test End Alarm,
QC Limits Alarm, Global Alarm, and Pop-Up.
TIP: Sound settings can be returned to Factory Setting through Settings Reset found in
the Admin Function Settings menu.
Page 62
Display: The Display menu provides adjustment of Screen Brightness, Language Selection, and
Pop-Up Message management. The Screen Brightness adjustment applies to all screen
views within the DV3T.
Note: The screen brightness is dimmed automatically after 5 minutes of no touch screen
activity. The brightness will be dimmed to the lowest value. A single touch of the touch
screen will return the display to the set brightness value.
The Language Setting is selected by pressing the Language field and choosing the appropriate
language from the list. The selected language will be implemented immediately.
Pop-Up Messages can be toggled on/off through the check box (see Figure V-7). This check
box applies to all Pop-Up Messages. Pop-Up Messages can be toggled off individually
through the check box that appears within the message box. Once the Pop-Up is turned
off, it will not be shown again. Pop-Up Messages can be restored through the use of the
Restore Hidden Pop-Ups command key at the bottom of the Display Settings screen.
This command key will restore all Pop-Ups.
Figure V-7
Change Password: A specific user can change their own password at any time. The user must
first enter the current password. Then the user will be prompted to enter and
confirm the new password.
TIP: If the administrator password is lost, it can be reset. Please contact
Brookfield or your Brookfield representative. Remember to have the
information found in the About screen available (see Section V.1).
Page 63
Measurement Units: The Measurement Units menu displays the current selections for measurement
units. Each unit can be changed by pressing the blue down arrow and
selecting the desired value. Measurement Units are part of the data file for
saved results. Available units include:
Viscosity
cP
P
mPas
Pas
cSt
mm2/s
centipoise
Poise
milliPascal seconds
Pascal seconds
centistokes
millimeters squared per second
RPM
1/s
C
F
Centigrade
Fahrenheit
Density
g/cm3
kg/m3
9,999,999.00
9.999.999,00
9 999 999,00
Time Format
13:59:59
1:59:59 PM
24 Hour Format
12 Hour Format
Day Separator
/
.
Forward Slash
Dash
Period
Date Format
Year/Month/Day
Month/Day/Year
Day/Month/Year
ISO Standard
USA Standard
International Standard
Global Alarm: The Global Alarm establishes a single measurement parameter with a range of
acceptance. Whenever the measured value exceeds this range of acceptance,
an audible and visual alarm will be issued. The Global Alarm will apply to any
test that is run. The QC Limit also establishes a measurement parameter with a
range of acceptance; however, the QC Limit is applied only to the test in which
it is defined.
The Global Alarm can be set for any of the four measurement parameters including;
Shear Stress, Temperature, Torque and Viscosity. Once a parameter is selected,
both low and high limits must be entered. These limits are independent of the
spindle and speed selected. When the global alarm is triggered, the DV3T will
provide an audible alarm and a Warning message on the screen (see Figure V-8).
The error message will provide some detail about the alarm condition. The test
Brookfield Engineering Labs., Inc.
Page 65
and data collection will continue while the Warning message is displayed. The
audible alarm will continue to sound until the operator touches the OK button
within the Warning box.
Figure V-8
The Running Viscosity Test display will show a colored box around the data
parameter associated with the Global Alarm. When the alarm condition is first
met, the colored box will be yellow. If the measurement parameter returns to the
acceptable range, then the colored box will be removed. When the alarm condition
is met a second time, the colored box will be red (see Figure V-9). The audible
alarm will only sound once.
Figure V-9
TIP: The Global Alarm sound can be turned off in the Sounds menu of User
Settings.
The Global Alarm is deactivated by choosing None in the drop down list of
available measurement parameters. You must press OK to confirm your Global
Alarm choice.
Page 66
Spindle List: The selection of Spindle in Configure Viscosity Test can be done by using either a
number pad or a scroll list. (This choice is selected through the navigation button
in the Set Spindle screen.) The contents of the scroll list are defined in the Spindle
List menu. Each spindle can be individually selected or deselected through the use
of a check box. Spindles, shown with both their name and their number, are listed
numerically according to their spindle entry number.
The restricted spindle will not appear on the scroll list of available spindles when
choosing Spindle in Configure Viscosity Test. If the user selects to enter the spindle
by number, then the choice of a restricted spindle will present an Invalid Input
message and prevent the selection of the spindle (see Figure V-10).
List Button
Figure V-10
TIP: To ensure that the correct spindle code is always entered, create a test that
is saved to the DV3T memory.
TIP: The selection of spindle can be restricted by User Log In. Use this restriction
combined with a saved test configuration to ensure that the correct spindle
is always used for critical tests. See Section V.4.2 for details of Users and
Access.
A Special Spindle can be created when the user has developed a new spindle geometry
or when using a standard Brookfield spindle in a non standard measurement container.
The Special Spindle, once created, will be available to users through the Spindle
selection in Configure Viscosity Test.
A Special Spindle is created by selecting the Special Spindle command button in the
Edit Spindle List menu. A Special Spindle requires a code, a name, and an SMC
value. Optional values include SRC and YMC.
Page 67
Code
The code is the numerical value used to select the spindle. This value
must be unique and can not be the same as a standard Brookfield
spindle. The acceptable range for Code is 100-199.
Name
The Name is a unique value to describe the spindle. The Name will
be displayed in the Spindle field. The Name can be comprised of
alpha and numeric characters.
SMC
SRC
The SRC (Shear Rate Constant) is used to convert the set RPM to
shear rate and calculate Shear Stress. The acceptable range of SRC
is 0.001 500.
The SRC value can be determined if the shear rate characteristics of
the spindle geometry are known. Consult the Brookfield publication,
More Solutions to Sticky Problems, for a discussion about shear
rate and some recommended equations.
Shear Rate (1/s) = SRC * RPM
YMC
Speed List: The selection of Speed in Configure Viscosity Test can be done by using either a
number pad or a scroll list. (This choice is selected through the navigation button in
the Set Spindle screen.) When scroll list is selected, the speed entry is restricted to
the classic Brookfield speed sets for the LV and RV/HA/HB rheometers. The contents
of the scroll list are defined in the Speed List menu. Each speed can be individually
selected or deselected through the use of a check box. Select or deselect all speeds
by using the Select All check box. Select or deselect all LV (or RV/HA/HB) speeds
by using the Select All LV check box. The classic Brookfield speed sets are defined
below.
LV:
RV/HA/HB:
Additionally, a new speed can be added to the list by selecting the New Speed command
key at the bottom of the display. The New Speed must be selected within the range
of available speeds on the DV3T: 0.0 250 RPM.
Speed entry through the number pad is never restricted.
TIP: To ensure that the correct speed is always entered, create a test that is saved
to the DV3T memory.
TIP: The selection of speed can be restricted by User Log In. Use this restriction
combined with a saved test configuration to ensure that the correct speed
is always used for critical tests. See Section V.4.2 for details of Users and
Access.
Page 68
Default Path
Settings Reset
Device Reset
Calibration Reminder
Save Audit Trail
Figure V-11
The Log In check box controls the Log In requirement. When Log In is checked, the DV3T will
require a User ID and Password prior to allowing any activity. The Log In screen is presented on
is selected from the Home Menu. The user must select their
power up or when the User Icon
User ID from the drop down list and then enter their password. The Log In requirement is removed
when the check box is unchecked.
Brookfield Engineering Labs., Inc.
Page 69
Page 70
General Settings
Administrator
Power User
User
N/A
Set
N/A
N/A
Allow
QC Alarm
Global Alarm
Pop-Up Warnings
Customize
Load A Test
Save a Test
Load Data
Save Data
Other Options
Spindle List
Speed List
Regional Settings
Measurement Units
Language
Instrument
USB
Instrument
USB
Instrument
USB
Instrument
USB
Print Data
Spindle
Speed
Temperature Setpoint
QC Limits
Density
Graph Axes
Math Model
Instructions
Save Path
Page 71
Set
Spindle
Administrator
Power User
Immersion Mark
Pre-Shear Speed
Pre-Shear Time
Wait Time
Run Speed
Zero Speed
Torque Reduction
Yield Limits
Temperature Setpoint
Graph Axes
Instructions
Save Path
User
Page 72
Add Language: Language files can be added to the DV3T as they become available from Brookfield.
Insert a USB Flash Drive into the DV3T that contains the language file. Select Add Language from
the Backup and Import menu and select the appropriate path. Press the Import command button to
bring the new language file into the DV3T. Select the language in the Display section of the User
Settings menu.
Update Software: The internal operating software of the Brookfield DV3T can be updated from
a USB Flash Drive as new versions become available from Brookfield. Insert a USB Flash Drive
into the DV3T that contains the operating software file. Select Update Software from the Backup
and Import menu and select the appropriate path. Press the Update command button to bring the
new software file into the DV3T.
Figure V-12
TIP: The Default Path can be overwritten at the time that data is saved.
Note that there is a limitation of 99 Results files which can exist in a single Path location. If this
limit is exceeded, specify a different data Path to save new files to, or remove files from the Path
location to decrease the number below 99. This limitation applies to all data Paths created on either
the USB Flash Drive or Internal Memory.
Brookfield Engineering Labs., Inc.
Page 73
Device Reset will return the DV3T to the factory default condition. All data, tests and Audit Trails
will be deleted from internal memory. All User accounts will be deleted.
TIP: Move data files and test files to a USB Flash Drive prior to selecting Device Reset.
Figure V-13
The Calibration Reminder will be presented during the first power up on or after the specified day.
The message box (see Figure V-14) requires that the user press OK before the AutoZero will be
performed.
Page 74
Figure IV-14
The calibration reminder will be reset by Brookfield or your authorized Brookfield dealer when the
unit is returned for calibration.
Page 75
General
Parameters
Program
Grid
Figure VI-1
The Instrument Bar includes the setting for DV3T or DV2T. PG Flash can be used to create
programs for both the DV3T and DV2T. The DV3T has additional control parameters which are
not available on DV2T. PG Flash should be set to DV3T when creating programs for the DV3T.
When set to DV3T, the yield test tab becomes available. The Instrument Bar also includes the
Load Test key and Reset Key. Load Test allows the user to load tests that have previously been
saved to the PC or a Flash USB Flash Drive. Reset All clears all data within the PG Flash interface
including: General Parameters, Test Parameters, and the Program Grid.
Page 76
The General Parameters section includes all Viscosity Test program controls that will apply to
every step of the program.
Spindle:
Report Screen:
Instructions:
The Test Parameters section includes all Viscosity Test program controls that may vary if there
is more than one test step.
Speed:
Data Collection:
Time Interval:
Density:
End Condition:
The Program Grid shows all Test Parameters for each step in a line view.
All measurement units can be set through the Units Setup button (see Figure VI-2). This menu
can be pinned to the screen. If the menu is not pinned, then it will minimize when the mouse is
moved outside of the window.
Page 77
Units Setup
Figure VI-2
Sample Viscosity Test programs can be seen in the following figures. Figure VI-3 demonstrates
a typical single step test. Figure VI-4 demonstrates a multiple step test.
Figure VI-3
Figure VI-4
Brookfield Engineering Labs., Inc.
Page 78
Yield Test parameters are shown in a single area within PG Flash without a Program Grid (only
a single step for Yield Tests). All yield parameters are available including Spindle Settings, Test
Settings and Test Instructions & Path Settings.
Spindle: The currently selected spindle. All Yield and Stress calculations will be made
based on the Spindle selection and the Immersion Mark. Spindle choice is
limited to vane spindles or user defined custom spindles.
Immersion Mark: Specify the immersion location of the Brookfield vane spindle. Primary is at
the immersion mark shown on the spindle shaft. Secondary is at the mid-point
of the vane surface.
Pre-Shear Speed: Specify a sped of rotation for pre-shear of the sample
Pre-Shear Time: Specify the duration of the Pre-Shear step.
Zero Speed: Specify the speed of rotation used to return the %Torque reading to zero before
starting the yield measurement.
Wait Time: Specify the duration of a rest period prior to starting the yield test.
Test Run Speed: Specify the speed of the yield test.
Torque Reduction: Specify the end point of the test. Normally set to 100%.
Temperature Set Point: A live signal from the rheometer when a temperature probe is attached (Brookfield
Part No. DVP-94Y). Also shown is the set point used to control the attached
temperature controller (Brookfield Thermosel or TC-XX0SD / TC-XX0AP
Temperature Baths).
Yield Low Limit: Parameter based on the chosen QC Limits.
Yield High Limit: Parameter based on the chosen QC Limits.
Save Results Path: Define the memory location where the data will be saved.
Instructions: Create a message that the user will see when the test begins.
Figure VI-5
Page 79
Figure VII-1
When n < 1 the product is shear-thinning or Pseudoplastic. This means the apparent viscosity
decreases as shear rate increases. The closer n is to 0, the more shear thinning the material is.
When n > 1 the product is shear-thickening or Dilatant. Their apparent viscosity increases as
shear rate increases.
When should you use it?
This model should be used with non-Newtonian, time-independent fluids that do not have a yield
stress. These fluids will begin to flow under any amount of shear stress. Graphs of such material
generally intersect the y-axis at 0.
Brookfield Engineering Labs., Inc.
Page 80
Shampoo
Flow Index (n) = 0.08
Consistency Index (k) = 91071 cP
With the new ingredient the shampoo has a flow index (n) of 0.08. This indicates that the
shampoo is shear-thinning enough to flow properly during processing and that it will flow
properly for the end-user. The consistency index, k, indicates how the shampoo behaves
when it experiences low shear rates. The power law values show that the shampoo becomes
quite thin at process shear rates and therefore it can be easily pumped into filling equipment,
hold tanks, etc. The consistency index of 91,071 cP shows that the shampoo is very viscous
at low shear rates, and as a result, it will appear to customers to be rich and creamy while
still being easy to apply.
Page 81
VII-2
Page 82
yielding, the product creates a flow curve and behaves as a Power Law fluid so that n indicates
where there is a shear-thinning or shear-thickening tendency. (In this case, if n =1, the material
is behaving as a Bingham fluid, which is discussed next.)
An Example of the Herschel-Bulkley Model at Work
A company uses a gel-like substance as part
Gel-Like Substance
of their production process. Upon arrival
they test the material and apply the Herscheln = 0.66
Bulkley model to ensure it will perform
o = 51.0 dynes/cm2
correctly during process. The results in Figure
k = 8550 cP
VII-2 show that the consistency index is 8,550
cP, the flow index is 0.66, and the yield stress
is 51.0 dynes/cm2. These results indicate that
this batch of gel does not quite meet specification. While the consistency index is within
spec, the yield value is higher than normal so the fluid will not begin to flow as easily. With
a flow index of 0.66, this batch is also less shear thinning than normal. Pump and mixer
speeds must be adjusted before using this material.
Page 83
Figure VII-3
Page 84
Drilling Fluid
Plastic Viscosity (h) = 6621 cP
Yield Stress (o) = 166.4 dynes/cm2
Page 85
Figure VII-4
Page 86
Page 87
= shear rate)
The NCA/CMA Casson model is designed by the National Confectioners Association and the
Chocolate Manufacturers Association as the standard rheological model for the industry. This
model determines yield and flow properties under specified conditions and closely approximates
the plastic behavior of chocolate before final processing.
Figure VII-5
When chocolate is used for enrobing, it must have a yield stress high enough to stay in place once
it enrobes the filling. In the case of decorating chocolate, the yield stress must be high enough so it
can keep its shape once it has been squeezed into place through a nozzle. For molding chocolate,
the plastic viscosity must be low enough to completely fill the mold.
(The NCA/CMA lists Brookfields HA-spring range viscometer with a Small Sample Adapter,
SC4-27 spindle and SC4-13R sample chamber as the approved apparatus.)
The IPC Paste Model was developed for solder pastes. It calculates the viscosity of solder pastes
at 10rpm. The IPC Paste Model requires that the product be tested with a Brookfield Spiral
Adapter at multiple speeds. More details can be found in the IPC-TM-650 Test Methods Manual
(methods 2.4.34.2 and 2.4.3).
Brookfield Engineering Labs., Inc.
Page 88
This model is a variation of the Power Law Model. Unlike the Power Law Model, which relates
apparent viscosity to shear rate, the IPC Paste Model relates apparent viscosity to the testing
speed (rpm).
Figure VII-6
Figure VII-7
Page 89
Programmable DV3T Cone/Plate Rheometers, S/N 50969 and higher, have an Electronic Gap Setting
feature. This feature enables the user to easily find the 0.0005 inch gap setting that was established at
Brookfield prior to shipment.
Brookfield recommends that the maximum particle size in the sample material for measurement with
cone/plate geometry be less than 5 times the gap settings. A more conservative approach is to limit the
maximum particle size to less than 10 times the gap setting.
The following information explains how to set the Electronic Gap and verify calibration of the DV3T
Rheometer.
TOGGLE SWITCH
Pilot Light
(red)
Toggle Swtich
Contact Light
(yellow)
Sliding Reference
Marker
Micrometer
Adjustment Ring
Figure A-1
Page 90
A.2 Setup
1. Be sure that the Rheometer is securely mounted to
the Laboratory Stand, leveled and zeroed with no
cone or cup attached and 0% torque is displayed.
2. Figure A-2 shows a typical water bath setup. Connect the sample cup inlet/outlet ports to the water
bath inlet and outlet and set the bath to the desired
test temperature. Allow sufficient time for the bath
to reach the test temperature.
3. The Rheometer has been supplied with a special
cone spindle(s) which contains the Electronic
Gap Setting feature. The CPE or CPA part
number designation on the cone spindle verifies
the Electronic Gap Setting feature.
Note:
Bath
Outlet
Cup
Inlet
Bath/Circulator
Sample
Cup
Figure A-2
Micrometer
Adjustment
Ring
4. With the motor off, thread the cone spindle by
using the spindle wrench to secure the rheometer
coupling nut (see Figure A-3); gently push up on
the coupling nut and hold this securely with the
wrench. Thread the cone spindle by hand.
Cup
Outlet
Bath
Inlet
Coupling Nut
Spindle
Wrench
These surfaces
must be clean!
Cone Spindle
Figure A-3
Figure A-4
Page 91
Figure A-5
Sliding
Reference
Marker
Full Scale
Division Marks
Figure A-6
1. The cup may be removed and replaced without resetting the gap if the micrometer
adjustment ring has not been moved.
2. Remove the spindle from the rheometer when cleaning.
3. Re-establish the hit point every time the spindle is attached/detached.
Page 92
Brookfield uses mineral oil viscosity standard fluids to calibrate Wells Brookfield Cone/Plate
Rheometers at the factory. Brookfield recommends that customers use mineral oil viscosity
standard fluids when you perform a calibration check.
If you decide to use a silicone viscosity standard fluid, do not use a fluid with a viscosity value
greater than 5000 cP with a Cone/Plate. Brookfield offers a complete range of mineral oil viscosity
standards suitable for use with Cone/Plates for viscosities above 5,000 cP or shear rates above
500 sec-1; see Table E-2 in Appendix F for a list of available fluids.
It is best to use a viscosity standard fluid that will be close to the maximum viscosity for a given
cone spindle/speed combination.
Example:
At 60 RPM, the full scale viscosity range is 10.0 cP. Thus, the Rheometer reading should be
97% torque and 9.7 cP viscosity 0.197 cP. The allowable error (0.197 cP) is a combination
of Rheometer accuracy and fluid tolerance (refer to Interpretation of Calibration Test Results
in Appendix F).
3. With the motor off, remove the sample cup and place the viscosity standard fluid into the cup.
Table A-1
Cone Part No.
CPA-40Z,
CPA-41Z,
CPA-42Z,
CPA-51Z,
CPA-52Z,
CPE-40,
CPE-41,
CPE-42,
CPE-51,
CPE-52,
Sample Volume
CP-40
CP-41
CP-42
CP-51
CP-52
0.5 mL
2.0 mL
1.0 mL
0.5 mL
0.5 mL
4. Attach the sample cup to the Rheometer and allow sufficient time for the sample, cup and cone
to reach temperature equilibrium.
5. Turn the motor on. Set the desired speed(s). Measure the viscosity and record the reading in
both % torque and centipoise (cP).
6. Verify that the viscosity reading is within the allowable 1% deviation, as explained earlier, for
the specific viscosity standard fluid(s) that you are using.
The CPE or CPA designation on the cone spindle indicates use with Electronic Gap Setting Cone/
Plate Rheometers/Rheometers only.
Page 93
Minimum
Maximum
DV3TLV
15
6,000,000
DV3TRV
100
40,000,000
DV3THA
200
80,000,000
DV3THB
800
320,000,000
Viscosity (cP)
DV3TLV
DV3TRV
DV3THA
DV3THB
Shear
Rate
sec-1
S SC4-14
.40N
S SC4-15
23.4 -
.48N
S SC4-16
.29N
SC4-18
1.5 -
30,000 16 - 320,000 32 -
1.32N
SC4-21
2.4 -
.93N
SC4-25
.22N
SC4-27
11.7 -
.34N
SC4-28
23.4 -
.28N
SC4-29
46.9 -
.25N
SC4-31
.34N
SC4-34
.28N
T HT-DIN-81
3.4 -
10,000 36.5 -
10,000
73 -
10,000 292 -
10,000
1.29N
S SC4-DIN-82 3.4 -
10,000 36.5 -
10,000
73 -
10,000 292 -
10,000
1.29N
S SC4-DIN-83 11.3 -
37,898 121.3 -
50,000 242.6 -
50,000 970.4 -
50,000
1.29N
T
S
Page 94
UL Adapter
Viscosity (cP)
UL Spindle
YULA-15 or 15Z
Shear Rate
sec-1
DV3TLV
DV3TRV
DV3THA
DV3THB
1 - 2,000
3.2 - 2,000
6.4 - 2,000
25.6 - 2,000
1.22N
DV3THB
Shear Rate
sec-1
DAA
Spindle
DV3TLV
5,000
DV3TRV
85
0.6 -
6.1 -
5,000
86
1.8 - 10,000
18.2 - 10,000
87
5.7 - 50,000
61 - 50,000
DV3THA
12.2 -
5,000
48.8 -
5,000
1.29N
36.5 - 10,000
146 - 10,000
1.29N
121 - 50,000
485 - 50,000
1.29N
Spiral Adapter
Viscosity (cP)
Spiral
Spindle
SA-70
DV3TLV
DV3TRV
DV3THA
DV3THB
2,100 - 2
,100,000
8,400 - 8,400,000
Shear
Rate
sec-1
0.667N
Cone/Plate Rheometer
Viscosity (cP)
13.1 - 261,000
Shear
Rate
sec-1
7.5N
49.1 - 982,400
2.0N
CPA-42Z
25.6 - 512,000
3.84N
205 - 4
,096,000
3.84N
393 - 7
,864,000
2.0N
Cone
Spindle
DV3TLV
CPA-40Z
.15 -
.3 -
DV3TRV
DV3THA
DV3THB
Viscosity (cP)
DV3TLV
DV3TRV
DV3THA
DV3THB
T-A
2,000,000 4,000 -
T-B
4,000,000 8,000 -
T-C
T-D
T-E
T-F
Page 95
Vane Spindles
Spindle
Torque
Range
V-71
Shear Stress
Range (Pa)
Viscosity Range
cP (mPas)
V-72
LV
.188-1.88
104.04-1.04K
V-73
LV
.938-9.38
502-5.02K
V-74
LV
9.38-93.8
5.09K-50.9K
V-75
LV
3.75-37.5
1.996K-19.96K
V-71
RV
.5-5
262-2.62K
V-72
RV
2-20
1.11K-11.1K
V-73
RV
10-100
5.35K-53.5K
V-74
RV
100-1K
54.3K-543K
V-75
RV
40-400
21.3K-213K
V-71
HA
1-10
524-5.24K
V-72
HA
4-40
2.22K-22.2K
V-73
HA
20-200
10.7K-107K
V-74
HA
200-2K
108.6K-1.086M
V-75
HA
80-800
42.6K-426K
V-71
HB
4-40
2.096K-20.96K
V-72
HB
16-160
8.88K-88.8K
V-73
HB
80-800
42.8K-428K
V-74
HB
800-8K
434.4K-4.344M
V-75
HB
320-3.2K
170.4K-1.704M
V-71
5XHB
20-200
10.48K-104.8K
V-72
5xHB
80-800
44.4K-444K
V-73
5XHB
400-4000
214K-2.14M
V-74
5xHB
4K-40K
2.172M-21.72M
V-75
5xHB
1.6K-16K
852K-8.52M
Note: 1. 1 Pa = 10 dyne/cm2
2. Viscosity Range is given at rotational speed of 10 RPM
3. 5xHB is the highest torque range available
4. Not for use with DV-E Viscometers
M = 1 million
K = 1 thousand
Pa = Pascal
cP = Centipoise
mPas = Millipascalseconds
Page 96
Special Considerations
In taking viscosity measurements with the DV3T Rheometer, there are two considerations which
pertain to the low viscosity limit of effective measurement.
1) Viscosity measurements should be accepted within the equivalent % Torque Range from
10% to 100% for any combination of spindle/speed rotation.
2) Viscosity measurements should be taken under laminar flow conditions, not under turbulent
flow conditions.
The first consideration has to do with the precision of the instrument. All DV3T Rheometers have
an accuracy of +/- 1% of the range in use for any standard spindle or cone/plate spindle. (Note that
accuracy values may be higher than 1% when using accessory devices with the DV3T). We discourage
taking readings below 10% of range because the potential viscosity error of +/- 1% is a relatively
high number compared to the instrument reading.
The second consideration involves the mechanics of fluid flow. All rheological measurements of
fluid flow properties should be made under laminar flow conditions. Laminar flow is flow wherein
all particle movement is in layers directed by the shearing force. For rotational systems, this means
all fluid movement must be circumferential. When the inertial forces on the fluid become too great,
the fluid can break into turbulent flow wherein the movement of fluid particles becomes random
and the flow can not be analyzed with standard math models. This turbulence creates a falsely high
rheometer reading with the degree of non-linear increase in reading being directly related to the degree
of turbulence in the fluid.
For the following geometries, we have found that an approximate transition point to turbulent flow
occurs:
1) No. 1 LV Spindle:
15 cP at 60 RPM
2) No. 2 LV Spindle:
100 cP at 200 RPM
3) No. 1 RV Spindle:
100 cP at 50 RPM
4) UL Adapter:
0.85 cP at 60 RPM
5) SC4-18/13R:
1.25 cP at 240 RPM
Turbulent conditions will exist in these situations whenever the RPM/cP ratio exceeds the values
listed above.
Page 97
- These fluids have the same viscosity at different Shear Rates (different RPMs)
and are called Newtonian over the Shear Rate range they are measured.
Non-Newtonian
Time Independent
Pseudoplastic
Time Dependent
Thixotropic
Brookfield publication, More Solutions to Sticky Problems, includes a more detailed discussion of
rheological properties and non-Newtonian behavior.
Page 98
1)
2)
3)
4)
5)
6)
7)
8)
9)
Test temperature
Sample container size (or spindle/chamber geometry)
Sample volume
Rheometer model
Spindle used
Whether or not to attach the guard leg
Test speed or speeds (or the shear rate)
Length of time or number of spindle revolutions to record viscosity
How the sample was prepared and/or loaded into the container
Page 99
SPINDLE
ENTRY CODE
SMC
SRC
RV1
01
RV2
02
RV3
03
10
RV4
04
20
RV5
05
40
RV6
06
100
RV7
07
400
HA1
01
HA2
02
HA3
03
10
HA4
04
20
HA5
05
40
HA6
06
100
HA7
07
400
HB1
01
HB2
02
HB3
03
10
HB4
04
20
HB5
05
40
HB6
06
100
HB7
07
400
LV1
61
6.4
LV2
62
32
LV3
63
128
LV4 or 4B2
64
640
LV5
65
1280
LV-2C
66
32
0.212
LV-3C
67
128
0.210
SA-70
70
105
0.677
Page 100
SPINDLE
ENTRY CODE
SMC
SRC
T-A
91
20
T-B
92
40
T-C
93
100
T-D
94
200
T-E
95
500
T-F
96
1000
ULA
00
0.64
1.223
DIN-81
81
3.7
1.29
DIN-82
82
3.75
1.29
DIN-83
83
12.09
1.29
DIN-85
85
1.22
1.29
DIN-86
86
3.65
1.29
DIN-87
87
12.13
1.29
SC4-14
14
125
0.4
SC4-15
15
50
0.48
SC4-16
16
128
0.29
SC4-18
18
3.2
1.32
SC4-21
21
0.93
SC4-25
25
512
0.22
SC4-27
27
25
0.34
SC4-28
28
50
0.28
SC4-29
29
100
0.25
SC4-31
31
32
0.34
SC4-34
34
64
0.28
40
0.327
7.5
41
1.228
42
0.64
3.84
51
5.178
3.84
52
9.922
SPINDLE
ENTRY CODE
SMC
SRC
YMC
V-71
71
2.62
0.5
V-72
72
11.1
V-73
73
53.5
10
V-74
74
543
100
V-75
75
213
40
Table D-1
Page 101
Table D-2 lists the model codes and spring torque constants for each Rheometer model.
Table D-2
MODEL
TK
DV3TLV
0.09375
LV
DV3TL3
0.234375
2.5LV
DV3TL5
0.46875
5LV
DV3TRQ
0.25
1/4 RV
DV3TRH
0.5
1/2 RV
DV3TRV
RV
DV3THA
HA
DV3TA2
2HA
DV3TA3
2.5HA
DV3THB
HB
DV3TB2
16
2HB
DV3TB3
20
2.5HB
DV3TB5
40
5HB
The full scale viscosity range for any DV3T model and spindle may be calculated using the equation:
Page 102
Entry Code
LV
RV
HA
RV1
RV2
HB
01
937
10,000
20,000
80,000
02
3,750
40,000
80,000
320,000
RV3
03
9,375
100,000
200,000
800,000
RV4
04
18,750
200,000
400,000
1,600,000
RV5
05
37,500
400,000
800,000
3,200,000
RV6
06
93,750
1,000,000
2,000,000
8,000,000
8,000,000
32,000,000
RV7
07
375,000
4,000,000
HA1
01
937
10,000
20,000
80,000
HA2
02
3,750
40,000
80,000
320,000
HA3
03
9,375
100,000
200,000
800,000
HA4
04
18,750
200,000
400,000
1,600,000
HA5
05
3,7500
400,000
800,000
3,200,000
HA6
06
93,750
1,000,000
2,000,000
8,000,000
HA7
07
375,000
4,000,000
8,000,000
32,000,000
HB1
01
937
10,000
20,000
80,000
HB2
02
3,750
40,000
80,000
320,000
HB3
03
9,375
100,000
200,000
800,000
400,000
1,600,000
HB4
04
18,750
200,000
HB5
05
37,500
400,000
800,000
3,200,000
HB6
06
93,750
1,000,000
2,000,000
8,000,000
HB7
07
375,000
4,000,000
8,000,000
32,000,000
LV1
61
6,000
64,000
128,000
512,000
LV2
62
30,000
320,000
640,000
2,560,000
LV3
63
120,000
1,280,000
2,560,000
10,240,000
LV4 or 4B2
64
600,000
6,400,000
12,800,000
51,200,000
LV5
65
1,200,000
12,800,000
25,600,000
102,400,000
LV-2C
66
30,000
320,000
640,000
2,560,000
LV-3C
67
120,000
1,280,000
2,560,000
10,240,000
Page 103
Range Coefficient
Spindle
Entry Code
LV
RV
HA
HB
T-A
91
18,750
200,000
400,000
1,600,000
T-B
92
37,440
400,000
800,000
3,200,000
T-C
93
9,3600
1,000,000
2,000,000
8,000,000
T-D
94
187,200
2,000,000
4,000,000
16,000,000
10,000,000
40,000,000
T-E
95
468,000
5,000,000
T-F
96
936,000
10,000,000
20,000,000
80,000,000
Spiral
70
98,400
1,050,000
2,100,000
8,400,000
ULA
00
600
6,400
12,800
51,200
HT-DIN-81
81
3,420
36,500
73,000
292,000
SC4-DIN-82
82
3,420
36,500
73,000
292,000
SC4-DIN-83
83
11,340
121,300
242,600
970,400
ULA-DIN-85
85
1,144
12,200
24,400
97,600
ULA-DIN-86
86
3,420
36,500
73,000
292,000
ULA-DIN-87
87
11,340
121,300
242,600
970,400
SC4-14/6R
14
117,200
1,250,000
2,500,000
10,000,000
SC4-15/7R
15
46,880
500,000
1,000,000
4,000,000
SC4-16/8R
16
120,000
1,280,000
2,560,000
10,240,000
SC4-18/13R
18
3,000
32,000
64,000
256,000
SC4-21/13R
21
4,688
50,000
100,000
400,000
SC4-25/13R
25
480,000
5,120,000
10,240,000
40,960,000
SC4-27/13R
27
23,440
250,000
500,000
2,000,000
SC4-28/13R
28
46,880
500,000
1,000,000
4,000,000
SC4-29/13R
29
93,750
1,000,000
2,000,000
8,000,000
SC4-31/13R
31
30,000
320,000
640,000
2,560,000
SC4-34/13R
34
60,000
640,000
1,280,000
5,120,000
40
307
3,270
6,540
26,160
41
1,151
12,280
24,560
98,240
42
600
6,400
12,800
51,200
51
4,854
51,780
103,560
414,240
52
9,300
99,220
198,440
793,760
V-71
71
2,456
26,200
52,400
209,600
V-72
72
10,404
111,000
222,000
888,000
V-73
73
50,146
535,000
1,070,000
4,280,000
10,860,000
43,440,000
4,260,000
8,520,000
V-74
74
508,954
5,430,000
V-75
75
199,645
2,130,000
Table E-1
Page 104
Conditions:
5,000
HT-30,000
10
12,500
HT-60,000
50
30,000
HT-100,000
100
60,000
500
100,000
1,000
Table F-1
29
B200
200
B600
600
B1060
1,060
B2000
2,000
B10200
10,200
B21000
21,000
B73000
73,000
B200000
200,000
B360000
360,000
Table F-2
Brookfield Engineering Labs., Inc.
Page 105
7) The viscosity reading should equal the cP value on the fluid standard to within the combined
accuracies of the Rheometer and the viscosity standard (as discussed in the section, at the
end of this Appendix, entitled Interpretation of Calibration Test Results) which appears
later in this section.
Page 106
Page 107
When a UL or DIN UL Adapter is used, the water bath is stabilized at the proper temperature:
1) Put the proper amount of viscosity standard fluid into the UL Tube (refer to the UL
Adapter instruction manual).
2) Attach the spindle (with extension link and coupling nut) onto the DV3T.
3) Attach the tube to the mounting channel.
4) Lower the tube into the water bath reservoir, or if using the ULA-40Y water jacket,
connect the inlet/outlets to the bath external circulating pump.
5) Allow 30 minutes for the viscosity standard, sample chamber and spindle to reach test
temperature.
6) Measure the viscosity and record the Rheometer reading. Note: The spindle must rotate
at least five (5) times before readings are taken.
7) The viscosity reading should equal the cP value on the viscosity fluid standard to within
the combined accuracies of the rheometer and the standard (as discussed in the section
entitled, Interpretation of Calibration Test Results). However, instrument accuracy is
2% of the maximum viscosity range and not the standard 1%.
Page 108
It is best to use a viscosity standard fluid that will be close to the maximum viscosity for a
given cone spindle/speed combination.
Example:
At 60 RPM, the full scale viscosity range is 10.0 cP. Thus, the Rheometer reading should be
97% torque and 9.7 cP viscosity 0.197 cP (0.1 cP for the rheometer plus 0.097 cP for the
fluid). The accuracy is a combination of Rheometer and fluid tolerance (refer to Interpretation
of Calibration Test Results).
4) With the rheometer stopped, remove the sample cup and place the viscosity standard fluid into
the cup, waiting 10 minutes for temperature equilibrium.
5) Connect the sample cup to the Rheometer. Allow sufficient time for temperature to reach
equilibrium. Typically 15 minutes is the maximum time that you must wait. Less time is
required if spindle and cup are already at test temperature.
6) Measure the viscosity and record the Rheometer reading in both % torque and centipoise (cP).
Notes: 1) The spindle must rotate at least five (5) times before a viscosity reading is
taken.
2) The use of Brookfield Viscosity Standard fluids in the range of 5 cP to 5000 cP is
recommended for cone/plate instruments. Please contact Brookfield Engineering
Laboratories or an authorized dealer if your calibration procedure requires more
viscous standards.
3) Select a viscosity standard fluid that will give viscosity readings between 10%
and 100% of full scale range. Refer to Appendix B for viscosity ranges of
cone spindles. Do not use a silicone viscosity standard fluid with a viscosity
value greater than 5000 cP with a Cone/Plate Rheometer. Brookfield offers a
complete range of mineral oil viscosity standards suitable for use with Cone/Plate
Rheometers as shown in Table E-2. Consult with Brookfield or an authorized
dealer to determine which fluid is appropriate.
Interpretation of Calibration Test Results:
When verifying the calibration of the DV3T, the instrument and viscosity standard fluid error
must be combined to calculate the total allowable error.
The DV3T is accurate to (+/-) 1% of the range in use when using spindles LV #1-3, RV #2-6,
HA #2-6 and HB #2-6. When using an accessory device with the DV3T such as Small Sample
Adapter, UL Adapter, Thermosel, Spiral Adapter, and DIN Adapter the accuracy value may be
increased. In general the increase in accuracy will be minimal, however, it could be as much as
1% for a total accuracy of +/- 2% of the range in use.
Brookfield Engineering Labs., Inc.
Page 109
Brookfield Viscosity Standards Fluids are accurate to (+/-) 1% of their stated value.
Example: Calculate the acceptable range of viscosity using DV3TRV with RV-3 Spindle at
2 RPM; Brookfield Standard Fluid 12,500 with a viscosity of 12,257 cP at 25C:
1) Calculate full scale viscosity range using the equation:
2) The viscosity standard fluid is 12,257 cP. Its accuracy is (+/-)1% of 12,257 or
(+/-)122.57 cP.
3) Total allowable error is (122.57 + 500) cP = (+/-) 622.57 cP.
4) Therefore, any viscosity reading between 11,634.4 and 12,879.6 cP indicates that the Rheometer
is operating correctly. Any reading outside these limits may indicate a Rheometer problem.
Contact the Brookfield technical sales department or your local Brookfield dealer/distributor
with test results to determine the nature of the problem.
Example:
1) Calculate the full scale viscosity range either by using the Spindle Range Coefficient in
Appendix B of More Solutions to Sticky Problems or by using the Auto Range button on your
rheometer.
The Spindle Range Coefficient for the 21 spindle on an LV Torque instrument is 4,688.
At 6 RPM, the Full Scale Range (FSR) viscosity is 781 cP. Allow +/- 1% for the rheometer
and +/- 1% for the Small Sample Adapter. Total allowable accuracy is:
A similar calculation at 12 RPM gives FSR = 391 cP: +/- 2% x 391 cP = +/- 7.8 cP
A similar calculation at 30 RPM gives FSR = 156 cP: +/- 2% x 156 cP = +/- 3.1 cP
+/- 1% x 101.5 cP = +/- 1.015 cP or roughly +/- 1.0 cP for further calculations.
Page 110
4) Therefore, at each speed, the acceptable windows within which the measured viscosity value
must lie is calculated relative to the viscosity value of the standard:
If your measured values fall outside of these windows, contact Brookfield or your authorized
dealer to discuss your results and determine whether your instrument is out of calibration.
Page 111
Page 112
should be known that this type of torque reading will not convert into a correct centipoise value when
using a Brookfield factor if the boundary conditions are not those specified by Brookfield.
The guard leg is a part of the calibration check of the Brookfield LV and RV series Viscometer/
Rheometer. Our customers should be aware of its existence, its purpose and the effect that it may have
on data. With this knowledge, the rheometer user may make modifications to the recommended method
of operation to suit their needs.
B-20Y
Guard Leg
For LV Torque
B-21Y
Guard Leg
For RV Torque
5 9/32
4 11/16
3 3/16
1 7/16
B-20KY
Guard Leg
For LV Torque
B-21KY
Guard Leg
For RV Torque
5 7/8
5 5/32
3 3/16
1 7/16
Page 113
The DV3T and earlier Brookfield DV-III Series Rheometers offer additional speeds to enhance
measurement capabilities. The DV3T offers speeds from 0.01 - 250 RPM with two options for
speed selection: numeric key pad for direct entry; Scroll List. The user can choose from these two
entry options by using the Navigation Bar (see Section IV.3: Global Settings). The Scroll List is
the format used with previous versions of the DV-III rheometer. You may choose the Scroll List
to provide the rheometer user with a familiar method for speed selection. The Scroll List can
be customized to limit the available speeds and to add new speeds to the list of standard speeds.
Page 114
2
3
4
Item
1
2
3
4
5
Part No.
VS-CRA-14S
VS-CRA-18S
GV-1201
GV-1203
502028071S33B
50S311832S01B
Description
Upright Rod and Clamp Assembly
Upright Rod and Clamp Assembly
Base, includes 2 GV-1203 leveling screws
Leveling Screws available separately or in assembly above
Flat Washer 5/16 X 7/8 X .071
Screw, 5/16 - 18 X 1 Hex Head
Qty
1
Optional
1
2
1
1
Page 115
Model QB is an optional laboratory stand which can be ordered for use with the DV3T Rheometer.
The advantage is the rapid speed of movement for lowering and raising the rheometer head.
4
5
Item
1
2
3
4
5
Part No.
VSQA-100Y
GV-1201
502028071S33B
50S311832S01B
GV-1203
Description
Upright Rod and Clamp Assembly
Base, includes 2 GV-1203 leveling screws
Flat Washer 5/16 X 7/8 X .071
Screw, 5/16 - 18 X 1 Hex Head
Leveling Screws available separately or in assembly above
Qty
1
1
1
1
2
Page 116
Unpacking
Check carefully to see that all the components are received with no concealed damage.
1 Base, GV-1201, with 2 Leveling Screws, GV-1203, packed in a cardboard carton
1 Upright Rod with attached Clamp Assembly in the instrument case
Assembly (Refer to Figures I-1 or I-2)
1. Remove the base assembly from the carton.
2. Remove the screw and washer from the upright rod. Place the rod and clamp assembly into
the hole in the top of the base.
Note: The Front designation on the clamp assembly should face toward you.
3. Rotate the rod/clamp assembly slightly until the slot on the bottom of the rod intersects the
pin located in the base.
4. While holding the rod and base together, insert the slotted screw and washer as shown and
tighten securely.
5. Adjust the tension screw so that the clamp assembly is not loose on the upright rod.
Rheometer Mounting
Insert the Rheometer mounting rod into the hole (with the cut-away slot) in the clamp assembly.
Adjust the instrument level until the bubble is centered from right to left and tighten the clamp
knob (clockwise). Use the leveling screws to fine adjust the rheometer level.
Note: If the Digital Rheometer cannot be leveled, check to insure that the rod is installed
with the gear rack facing forward.
Caution: Do not tighten the clamp knob unless the rheometer mounting rod
is inserted in the clamp assembly.
Caution: Do not use the DV3T Rheometer with any laboratory stand that
does not utilize the GV-1201 base. This large base is necessary for
stability of the DV3T Rheometer during use. Earlier versions of the
Brookfield Laboratory Stand including the Model A and Model S
should not be used with the DV3T.
Operation
Rotate the UP/DOWN knob to raise or lower the rheometer. Adjust the tension screw if the UP/
DOWN movement of the rheometer head is not acceptable, i.e. too easy or too difficult.
Page 117
Figure J-1
Figure J-2 shows the Probe Clip (with RTD temperature probe installed) mounted on the LV
guard leg.
Figure J-3 shows the Probe Clip mounted in a 600 mL low form Griffin beaker. This mounting
may be used with LV, RV, HA and HB series instruments.
Caution:
Note:
DVE50
Probe Clip
Probe Clip
DVE-50
R TD Temperatur e
Probe
Figure J-3
Figure J-2
Brookfield Engineering Labs., Inc.
Page 118
Retain the soft card and screen cloth with any unused screen protector.
Install the Screen Protector
Installation instructions for the screen protector are available in three methods:
Figure K-1
Step 2:
Alignment of the screen protector to the bezel (frame) of the DV3T
is critical. Be sure to avoid mis-alignment which may result in the
screen protector contacting or locating under the bezel (frame).
A foam alignment tool is attached to the screen protectors outer
layer and should be used to locate the screen protector to the lower
right hand corner of the touch screen (see Figure K-2).
Figure K-2
Page 119
Step 3:
Completely remove Tab 1 (see Figure K-3).
Lightly press the screen protector to the touch screen.
Inspect the alignment of the screen protector. If the screen protector
is touching the bezel (frame), them remove and reapply.
Step 4:
Figure K-3
Peel back Tab 2 all the way to remove the backing, which also
removes the alignment tool (see Figure K-4). The screen protector
is now in place.
Step 5:
Use the soft card provided to push trapped air bubbles from the
center of the screen to the edge (see Figure K-5). Small bubbles
should go away within 48 hours.
DO NOT use the soft card to push the screen protector under the
bezel (frame). If the screen protector is not properly aligned,
remove and reapply. Two screen protectors are provided in this
package in case the first attempt is unsuccessful.
Test the viscometer (power on) to ensure that the screen protector
is properly attached. The screen protector is properly aligned if
the DV3T Rheometer responds normally to a touch on the touch
screen. If the DV3T Rheometer is unresponsive to touch then
remove the screen protector and reapply.
Figure K-4
Figure K-5
Page 120
Inspect threads for wear; if the threads are worn, the unit needs service (see Appendix O). Check
to see if spindles rotate eccentrically or wobble. There is an allowable runout of 1/32-inch in
each direction (1/16-inch total) when measured from the bottom of the spindle rotating in air.
Check to see if the rheometer coupling appears bent; if so, the unit is in need of service (see
Appendix O: Warranty Repair and Service).
Inaccurate Readings
If % readings are under-range (less than 10%), the display will flash; change spindle and/or
speed.
EEEE on the digital display means the unit is over-range (greater than 100%); reduce speed
and/or change spindle.
Page 121
* This procedure does not apply to instruments with ball bearing suspension (see Section I.3).
Display Reading Will Not Stabilize
Page 122
Page 123
The Brookfield website is a good resource for additional and self-help whenever you need it. Our
website offers a selection of how-to videos, application notes, conversion tables, instructional manuals,
material safety data sheets, calibration templates and other technical resources.
http://www.youtube.com/user/BrookfieldEng
Brookfield has its own YouTube channel. Videos posted to our website can be found here as well as
other home-made videos made by our own technical sales group.
ViscosityJournal.com
Brookfield is involved with a satellite website that should be your first stop in viscosity research. This
site serves as a library of interviews with experts in the viscosity field as well as Brookfield technical
articles and conversion charts. Registration is required, so that you can be notified of upcoming
interviews and events, however, this information will not be shared with other vendors, institutions, etc..
Article Reprints
Learn more about viscosity and rheology with our most popular publication. This informative booklet will
provide you with measurement techniques, advice and much more. Its a must-have for any Brookfield
Viscometer or Rheometer operator. More Solutions is available in print and also as a downloadable
PDF on the Brookfield website by following this path: http://www.brookfieldengineering.com/support/
documentation.
Training/Courses
Whether it is instrument-specific courses, training to help you better prepare for auditing concerns, or
just a better understanding of your methods, who better to learn from than the worldwide leaders of
viscosity measuring equipment? Visit our Services section on our website to learn more about training.
** Downloads will require you to register your name, company and email address. We respect your
privacy and will not share this information outside of Brookfield.
Brookfield Engineering Labs., Inc.
Page 124
Brookfield Viscometers/Rheometers are guaranteed for one year from date of purchase against defects
in materials and workmanship. They are certified against primary viscosity standards traceable to
the National Institute of Standards and Technology (N.I.S.T.). The Rheometer must be returned to
Brookfield Engineering Laboratories, Inc. or to the Brookfield dealer from whom it was purchased for
a warranty evaluation. Transportation is at the purchasers expense. The Rheometer should be shipped
in its carrying case together with all spindles originally provided with the instrument. If returning to
Brookfield, please contact us for a return authorization number prior to shipping.
All Brookfield DV3T Rheometers are supplied from the factory with a Calibration Seal (located on
the back of the viscometer). The warranty stated above will be voided if the Calibration Seal has
been damaged. Only Brookfield or our authorized servicing dealer may break the Calibration Seal for
purposes of instrument repair or recalibration.
For a copy of the Repair Return Form, go to the Brookfield website,
www.brookfieldengineering.com
For repair or service in the United States return to:
Brookfield Engineering Laboratories, Inc.
11 Commerce Boulevard
Middleboro, MA 02346 U.S.A.
Telephone: (508) 946-6200 FAX: (508) 946-6262
www.brookfieldengineering.com
For repair or service outside the United States, consult Brookfield Engineering Laboratories, Inc. or
the dealer from whom you purchased the instrument.
For repair or service in the United Kingdom return to:
Brookfield Viscometers Limited
Brookfield Technical Centre
Stadium Way
Harlow, Essex CM19 5GX, England
Telephone: (44) 1279/451774 FAX: (44) 1279/451775
www.brookfield.co.uk
For repair or service in Germany return to:
Brookfield Engineering Laboratories Vertriebs GmbH
Hauptstrasse 18
D-73547 Lorch, Germany
Telephone: (49) 7172/927100 FAX: (49) 7172/927105
www.brookfield-gmbh.de
For repair or service in China return to:
Guangzhou Brookfield Viscometers and Texture Instruments Service Company Ltd.
Suite 905, South Tower, Xindacheng Plaza
193 Guangzhou Da Dao Bei, Yuexiu District
Guangzhou, 510075 P. R. China
Telephone: (86) 20/3760-0548 FAX: (86) 20/3760-0548
www.brookfield.com.cn
On-site service at your facility is also available from Brookfield. Please contact our Service
Department in the United States, United Kingdom, Germany or China for details.
Brookfield Engineering Labs., Inc.
Page 125
MODEL
SPINDLE
RPM
DIAL READING
% TORQUE
BY:
DATE:
FACTOR
VISCOSITY SHEAR
cP
RATE
TEMP C
FOR:
TIME
NOTES
BROOKFIELD ENGINEERING LABORATORIES, INC. 11 Commerce Blvd. Middleboro, MA 02346 TEL: 508-946-6200 or 800-628-8139 FAX: 508-946-6262 www.brookfieldengineering.com VTR1207
CONCLUSIONS:
SAMPLE
TEST INFORMATION:
This tear-off sheet is a typical example of recorded test data. Please photocopy and retain this
template so that additional copies may be made as needed.