User Manual For Ultrascope
User Manual For Ultrascope
RIGOL
Product Description
UltraScope Software introduces a family of Windows 95, 98, 2000, NT, and Windows XP
applications that connect your RIGOL DS1000 series oscilloscopes to your PC desktop by
USB or RS-232 .
The UltraScope software provides the following control and analysis features:
z
Use DSO Controller to control the oscilloscope locally or over the network.
System Requirements
To install and run UltraScope software, you must have an IBM-compatible PC with the
following installed:
z
10 MB of disk space
You must have a DS1000 series oscilloscope and the firmware version must be higher or
equal to 01.02.22 . To connect to PC, the oscilloscope must install USB dirver or have an
available RS-232 COM port.
Contents
This manual is the user guide for the Ultrascope software, and it contains four chapters.
List
PRODUCT DESCRIPTION .....................................................................................................3
SYSTEM REQUIREMENTS ...................................................................................................4
CONTENTS..............................................................................................................................5
CHAPTER 1: INSTALLATION AND UNINSTALLATION...................................................7
CHAPTER 2: OPERATING THE ULTRASCOPE ..................................................................8
Description of the user interface............................................................................................9
Description of the software menu........................................................................................10
Description of the Toolbar...................................................................................................12
How to use DSO Controller.................................................................................................14
Shortcut Menu.....................................................................................................................15
Instant Running Control......................................................................................................16
Use DSO Controller to set up CH1& CH2 .........................................................................17
Use DSO Controller to set up Time Base............................................................................20
Use DSO Controller to set up Trigger System ....................................................................22
Use DSO Controller to set up Acquisition System .............................................................34
Use DSO Controller to set up Display System ...................................................................35
How to use the Virtual Keyboard........................................................................................36
How to use Data Browser ....................................................................................................37
How to display a waveform ................................................................................................39
How to Measure with Data Browser...................................................................................42
How to get waveform data with Data Browser ...................................................................45
General Options ..................................................................................................................52
CHAPTER 3: PROMPTING MESSAGES AND TROUBLESHOOTING ...........................55
Prompting Messages ...........................................................................................................55
Troubleshooting ..................................................................................................................55
CONTACT RIGOL.................................................................................................................56
Uninstallation
To uninstall Ultrascope for DS1000 Series Oscilloscopes, you can click by the follow steps:
Start-> Program -> Ultrascope for DS1000 Series->Uninstall. Or use the Add/Remove
program in the Microsoft Windows Control Panel .
Both uninstall methods activate the Uninstall wizard, which takes you through the uninstall
process for Ultrascope.
1. File
New:
Open:
Save Data Sheet:
Save Controller:
Save All:
Page Setup:
Exit:
2. View
Toolbar:
Status Bar:
Show Controller:
Show Data Browser:
3. Tools
Connect to Oscilloscope:
Disconnect:
Options:
License Setup:
4. Window
Cascade:
Tile Horizontal:
Tile Vertical:
Arrange Icons:
10
5. Help
Contents:
About:
11
1( New ): To create a new document of Data Sheet, Controller, Waveform, Measure or Wave
Data.
2( Open ): To open a document of Data Sheet or Controller
3( Save All ): To save all files
4( Connect to Oscilloscope ): Connect PC to the oscilloscope
5( Disconnect ): Disconnect the link between PC and the oscilloscope
6( DSO Controller ): Show the DSO Controller window
7( Data Browser ): Open or close the Data Browser window
8( Add New Waveform ): Add new waveform window
9( Add New Measure ): Add new measure window
10( Add New Data ): Add new data window
11( Add New Record ): Add new record window
12( Add New Memory Waveform ): Add new memory waveform window
13( Options ): I/O settings and general settings
14( Connection Status Lamp ): The lamp will turn blue in connecting status, and it will turn
red after disconnect the link between PC and Oscilloscope.
Copyright 2006 Rigol Technologies, Inc.
UltraScope For DS1000
12
Note: After connecting the PC to the oscilloscope, the keys and knobs on the front panel will
be locked. To activate the front panel control, you must disconnect the link in Ultrascope
software or pressing the FORCE key on the front panel.
13
14
Shortcut Menu
The Shortcut menu contains commands and settings available in DSO Controller. You could
access the menu by right-click on an icon.
Opens a dialog box that lets you open a DSO Controller file.
Saves the DSO Controller to its current directory and file name.
Save Control File As: Opens a dialog box that lets you save a DSO Controller file.
Connect To DSO:
Disconnect:
Upload:
Download:
Expand All:
15
Figure 4
Double-click Run|Stop to pop up a dialogue box, you could select RUN or STOP to let
the instrument run or stop.
Double-click the other three action keys to execute the Autoset, Set Trig to 50% or
Force Trigger actions instantly.
16
Figure 5
Double-click the lowest subtrees to pop up a list, you could select a value and set the
instrument to use that value.
17
Table 1:
Subtree
Display
Volt/Div
Offset
Coupling
BW
Limit
Probe
Settings
Comments
ON
OFF
2.00mV~
5.00V
-40.00V~+40.00V
(Volt/Div> 100mV)
-2.00V ~ +2.00V
(Volt/Div<200mV)
AC
DC
ON
OFF
1X
10X
100X
1000X
18
Figure 6
Double-click the lowest subtrees or the second-subtrees to pop up a list, you could select a
value and set the instrument to use that value.
Table 2:
Subtree
Second-subtree
Settings
Comments
Dipslay
D15-D0
ON
OFF
Position
D15-D0
0-15
TTL
CMOS
ECL
User
Threshold
Reset waveform
D15-D0.
Reset
of
the
channel
19
Figure 7
Table 3:
Subtree
Settings
Time/Div
5ns50s
Delay
Format
Comments
Select the timebase of CH1 from 5ns to 50s
Display the delay value
Y-T
X-Y
Roll
20
When the trigger mode is Alternative, you can set up the value as below figure and table.
Figure 8
Table 4:
Subtree
CH1
Second-subtree
Time/Div
Settings
Comments
5ns50s
Delay
CH2
Time/Div
Delay
21
22
Method
Source
Slope
Mode
Coupling
Settings
EDGE
PULSE
VIDEO
SLOPE
ALTERNATIVE
PATTERN
DURATION
CH1
CH2
EXT
EXT5
AC
D15-D0
RISING
FALLING
BOTH
AUTO
NORMAL
SINGLE
DC
AC
HF
LF
Comments
Select trigger mode as EDGE
Select trigger mode as PULSE
Select trigger mode as VIDEO
Select trigger mode as SLOPE
Select trigger mode as ALTERNATIVE
Select trigger mode as PATTERN
Select trigger mode as DURATION
Select the input source as the trigger signal
EXT and EXT5 use the signal applied to the EXT TRIG
connector as the source.
Select any digital channel in D15-D0 as trigger source
(only in mixed-signal oscilloscope).
Select to trigger on the rising, falling or both edge of the
signal
Select the type of triggering
Selects the components of
the trigger circuitry
Level
-12.00div
to +12.00div
Sensitivit
y
0.10div
to 1.00div
Holdoff
100ns ~ 1.5s
23
24
Settings
Comments
Source
CH1
CH2
EXT
EXT5
Polarity
POSITIVE
NEGATIVE
All line
Line num
Odd Field
Even Field
Line number
1 ~ 525/625
Video
Standard
NTSC
PAL/SECAM
Select the video standard for sync and the number count
Level
-12.00div
to +12.00div
Sensitivity
0.10div
to 1.00div
Holdoff
100ns ~ 1.5s
Sync
25
26
Source
Settings
Comments
CH1
CH2
EXT
EXT5
D15-D0
Pulse mode
+Less Than
+Greater Than
+Equal
-Less Than
-Greater Than
-Equal
Pulse width
20ns ~ 10s
Mode
Coupling
Level
Sensitivity
Holdoff
AUTO
NORMAL
SINGLE
DC
AC
HF
LF
-12.00div
to +12.00div
0.10div
to 1.00div
100ns ~ 1.5s
Note: The change of trigger methods will make the trigger subtrees different.
27
28
Slope mode
Slope time
Mode
Coupling
LevelA
LevelB
Sensitivity
Holdoff
Settings
Comments
CH1
CH2
EXT
EXT5
+Less Than
+Greater Than
+Equal
-Less Than
-Greater Than
-Equal
20ns ~ 10s
AUTO
NORMAL
SINGLE
DC
AC
HF
LF
-12.00div
to +12.00div
0.10div
to 1.00div
100ns ~ 1.5s
29
30
Second-subtree
Trig type
CH1
Settings
EDGE
PULSE
VIDEO
SLOPE
Different
Second-subtrees
Trig type
CH2
Comments
Select trigger mode as EDGE
Select trigger mode as PULSE
Select trigger mode as VIDEO
Select trigger mode as SLOPE
The change of trigger type will make
the trigger second-subtrees different
EDGE
PULSE
VIDEO
SLOPE
Different
Second-subtrees
Note: The change of trig type will make the alternative trigger second-subtrees different.
31
Code
Second-subtree
Settings
H
L
X
Rising
Falling
AUTO
NORMAL
SINGLE
D15-D0
Mode
100ns
~ 1.5s
Holdoff
Comments
Set the pattern on the digital
channels
32
Second-subtree
Settings
Comments
H
L
X
Qualifier
Less Than
Greater Than
Equal
Duration
time
20ns ~ 10s
Mode
AUTO
NORMAL
SINGLE
Select the
triggering
Holdoff
100ns ~ 1.5s
Code
D15-D0
type
of
33
Figure 16:
Table 12:
Subtree
Acquisition
Settings
NORMAL
AVERAGE
PEAK
DETECT
Comments
Set to Normal Acquisition mode
Set to Average Acquisition mode
Set to Peak Detect Acquisition mode
Sampling
Mode
REAL-TIME
EQU-TIME
Averages
2 to 256, step
2
34
Figure 17
Table 13:
Subtree
Settings
Comments
VECTORS
Type
DOT
ON
Persist
OFF
Clear
Grid
35
36
Shortcut Menu
The Shortcut menu contains commands and settings available in Data Browser. You could
access the menu by right-click on an icon.
New Data Sheet:
Opens a dialog box that lets you open a Data Sheet file.
Saves the Data Sheet to its current directory and file name.
37
Opens a dialog box that lets you save a Data Sheet file.
Add New:
Rename:
Remove:
Sorted:
38
The waveform panel provides waveform display, zooming, storage, and printing for analysis.
The following description illuminates how to operate these functions.
39
Refresh
After you click the Refresh button on this panel or press the F5 key on your PC keyboard,
the signal currently on the oscilloscopes screen will be displayed on the waveform window.
You could select CH1, CH2, Math, LA(Mixed-Signal Oscilloscope) or all of them to display
the signal connected to channel 1, channel 2, Math, LA(Mixed-Signal Oscilloscope) or all
channels. You must open the channel that you want to display in the DSO.
Controller. Otherwise, the selection item in conjunction with this channel will be disabled as
gray color.
Display
The panel toolbar contains controls for zooming the waveform, selecting into the Best Fit
mode, setting the display in Color or B&W mode, setting the Grid ON or OFF.
The percent zoom field displays the current data sheet magnification setting, where 100% is
the system-default zoom value. Clicking on the drop-down menu arrow displays a list of
magnification values from which you can select. You can also enter numeric values into this
field by moving the cursor into the field, and typing in a new value after the Best Fit mode
disabled.
The Best Fit mode fits the graticule to the current window size, allowing you to view the
entire graticule.
Output
You could output the waveform by exporting currently displayed window into BMP format
file to its directory and file name.
Previewing and Printing
To preview and set up printing page of waveform, click the Preview button. To print the
page directly, just click the Print button.
First Page Go to Page Last Page
Previous Page
Next Page
Zoom Level
Close Preview
Select Page
40
41
This panel provides 18 parameters for measuring, including Vpp, Vmax, Vmin, Vavg, Vamp,
Vtop, Vbase, Vrms, Vover, Vpre, Frequency, Rise Time, Fall Time, Period, +Pulse Width,
-Pulse Width,+Duty or -Duty.
42
The description of the functional buttons on the measurement panel is shown as follows:
Refresh
You could get the auto measurement results from the oscilloscope by clicking the Refresh
button on this panel or pressing the F5 key on your PC keyboard. The panel will display a list
of measurement values for currently selected channel.
CH1, CH2
You could select CH1 or CH2 for measuring the signal connected to channel 1 or channel 2.
The oscilloscope must display the channel that you want to measure; you could change this
setting in the DSO Controller. If the channel is close, the selection item associated with this
channel will be disabled as gray color.
Export
You could save the measurement results as TXT or EXCEL file in your computer.
Info Bar
This bar displays channel, vertical scale, horizontal scale, sampling rate and date info in
conjunction with the waveform for measuring.
43
The Ultrascope software enables users to set pass/fail test parameters. These parameters
include Vpp, Vmax, Vmin, Vavg, Vamp, Vtop, Vbase, Vrms, Vover, Vpre, Frequency, Rise
Time, Fall Time, Period, +Pulse Width, -Pulse Width,+Duty or -Duty.
You could set pass/fail factors range. If the measurement value is out of range, the
measurement panel will display test result as fail. If the measurement value is under limit,
it will show pass on the Pass&Fail list in the measurement panel.
To set the range of these pass/fail factors according to the following steps:
1. Select parameters for pass/fail factors from the list, and click to fill in the square
block to the right of the list.
2. After enable the parameter, input number and select unit for this parameter to set pass
/fail range.
44
46
The record panel provides waveform display, storage, and printing for analysis. The
following description illuminates how to operate these functions.
47
48
Replay
The record waveform will be played back in continues.
Pause
Pause the waveform to play back.
Stop
Stop the waveform to play back.
Go to start frame
Press to go to the start frame.
Go to end frame
Press to go to the end frame.
The last frame
Press to go to the last frame.
The next frame
Press to go to the next frame.
Circulation and single replaying
Choosing to circulation or single replaying the record waveform.
Default time
Press to set the frame interval time is 50.00ms.
The frame interval time
This message display the frame interval time.
Current and total frame
This message display the current and total frame.
Sub time
Press to set the frame interval time sub.
Add time
Press to set the frame interval time add.
49
The memory waveform panel provides the saved waveform display, zooming, storage, and
printing for analysis. The following description illuminates how to operate these functions.
50
Open
You can press this button to select the waveform file which you have saved.
Display
The panel toolbar contains controls for zooming the waveform, setting the display in Color
or B&W mode, setting the Grid ON or OFF.
The percent zoom field displays the current data sheet magnification setting, where 100% is
the system-default zoom value. Clicking on the drop-down menu arrow displays a list of
magnification values from which you can select. You can also enter numeric values into this
field by moving the cursor into the field.
Output
You could output the waveform by exporting currently displayed window into BMP format
file or the Memory data into CSV format file to its directory and file name.
Previewing and Printing
To preview and set up printing page of waveform, click the Preview button. To print the
page directly, just click the Print button.
Volt/Div
You can change the volt/div value to zoom the waveform in vertical direction.
Time/Div
You can change the time/div value to zoom the waveform in horizontal direction.
51
General Options
I/O Selection
You could select USB or RS-232.
Startup Option
In this section, you could set three operations while the Ultrascope software is starting up.
1. Startup Hardware Check: It will check the connection between the oscilloscope and
the computer. If the communication is available, it will connect the oscilloscope with PC
automatically while the software starting up.
52
2. Recent Data Sheet File Load: It will open the data sheet file that you use last time after
the software starts up.
3. Recent Control File Load: It will open the control file that you use last time after the
software starts up.
Refresh
You could select to refresh the waveform or measurement data in single or auto mode.
Time Setting
You could select the refreshing time interval from a list of 1s, 2s, 5s, 10s and 20s. The default
value is 1s.
53
54
Prompting Messages
Device not found: Prompting that the software does not detect the connection between the
oscilloscope and the computer.
File not found: Prompting that Ultrascope could not find the control file or data sheet file
when starting up.
This Item Query Only: Prompting that this item is only for display the setting or status of
the oscilloscope. You could not double-click this item to setup the oscilloscope.
Save the changes to the Control File: Prompting users to save the changes of the control
file before quit Ultrascope software.
Save the changes to the Data Sheet File: Prompting users to save the changes of the data
sheet file before quit Ultrascope software.
Troubleshooting
1. After running the Ultrascope, the software could not detect the connection between
the instrument and the computer. Please inspect the settings of the software and the
oscilloscope according to the following steps:
(1) Make sure that the oscilloscope is powered on.
(2) Check the cable connection between the oscilloscope and PC.
(3) Check if the settings for the baud rate of the oscilloscope are same as those of the
Ultrascope software. (The default baud rate is 9600.)
(4) After the inspections above, restart the Ultrascope software.
(5) If the problem still remains, please contact RIGOL for help.
55
Contact RIGOL
Product Support: For questions about using RIGOL measurement products, call in China:
Tel: 86-10-82899325
9:00 a. m.5: 00 p. m. Beijing time
Or contact us by e- mail:
[email protected]
For product support outside of China, contact your local RIGOL
distributor or sales office.
For a listing of worldwide service centers, visit our web site.
Web sitewww.rigol.com
56