Summary of Redesign of Indonesian-Made Osteosynthesis Plates To Enhance Their Mechanical Behavior
Summary of Redesign of Indonesian-Made Osteosynthesis Plates To Enhance Their Mechanical Behavior
20
20
20
20
20
20
Narrow
Narrow
Narrow
Narrow
Narrow
Narrow
DCP
DCP
DCP
DCP
DCP
DCP
from
from
from
from
from
from
Indonesian Manufacturer A
Indonesian Manufacturer A, SMAT treated
Indonesian Manufacturer B
Indonesian Manufacturer B, SMAT treated
Indonesian Manufacturer B, shot peened
Mathys/ Synthes as the standard plate (S)
Each group of plates were tested to study the strength (proportional load limit,
bending stress) and the fatigue limit (Load, bending stress limit).
The results were shown that, during the single cycle bending test, shot peened and
SMAT-treated plates showed improvements in mean failure loads and mean bending
stresses, compared to those of the original A and B plates. SMAT-treated plate from
Manufacturer A exceeded the mean failure load of the standard European plate (S).
None of the modified treated plates from Manufacturer A nor B could exceed the
calculated bending stress of the standard plate (S). The fatigue test of 3 million
cycles were also concluded. SMAT-treated plates from Manufacturer A showed
considerable improvements, resulting in exceeding the fatigue load limit of the
standard plate.
Effects of surface mechanical attrition treatment (SMAT) on a rough
surface of AISI 316L stainless steel
In this study, Surface mechanical attrition treatment (SMAT) were used to smoothen
a rough-surfaced AISI 316L stainless steel material. SMAT modifies the morphology
and the roughness of the specimen. Previous studies have been done on using SMAT
to roughen a smooth surface, but this time is used to smoothen a rough surface. A
variation of size of milling balls were used during this experiment, ranging from
3.18mm to 6.35mm. The modification of subsurface micro hardness, surface
morphology, roughness, and mass reduction were taken into account. The specimen
were formed into steel strips or plates with a dimension of 100mm x 50mm x 4mm.
A clean-up procedure were taken into action using 70% ethanol solution to remove
impurities at the surface. The specimen was placed in a tubular chamber with a
dimension of 150mm in length and 80mm in diameter containing 250 milling balls
of various size. The SMAT was carried out by shaking the chamber for 20 minutes
with a rotational speed of 1400rpm.
In the study it was found that the fatigue life of the knee joint would go down from
1.2 million cycles to 9531 cycles as the force increased from 10N to 1010N.
An experimental approach
polyethylene tibial inserts
to
determining
fatigue
crack
size
in
Each year, 700 thousand knee surgeries are performed in the US to treat
degenerative joint diseases by replacing the damaged natural knee with a
prosthetic knee joint. The performance of the prosthetic knee joint have been
documented to decline dramatically after 15-20 years of usage. The primary reason
of these failures are caused by fatigue wear. Existing methods to locate, and inspect
fatigue crack damage have several shortcomings such as limited resolution,
destructive testing approach, and high cost. In this study, the author propose an
alternative fatigue crack damage visualization and measurement methods that
overcome these shortcomings of the existing methods available. This new method
of inspection is based on trans-illumination. It is a non-destructive method of
inspection and has low-cost set-up.
In this experiment, two tibial inserts are being measured for fatigue crack damage
using the proposed method. Tibial insert 1 is a 7mm thick posterior that was used in
a knee wear simulation, while tibial insert 2 is a 14mm tibial insert that was
retrieved from 9 years of use in vivo. Tibial 1 is tested using a knee simulator, in
accordance with ISO standard 14243-1 with the following applied loading axial load