DIVA-D

0123

Digital Imaging Video Acquisition


with full DICOM compatibility

Users manual
Release 4 January 2006 (Rev. 0)

USERS MANUAL
Contents

Contents
1

INTRODUCTION
1.1

Icon appearing in the manual ................................................................. 1

SAFETY ASPECT

2.1

Warning.................................................................................................. 3

2.2

Environmental risks and displacement ................................................... 5

2.3

Symbols used ......................................................................................... 6

DESCRIPTION

3.1

Identification labels ................................................................................ 7

3.2

Description ............................................................................................. 8
3.2.1 DIVA-D composition ................................................................... 9
3.2.2 Monitors..................................................................................... 9
3.2.3 Accessories................................................................................. 9

TECHNICAL DATA

10

4.1

DIVA-D AS 25....................................................................................... 10

4.2

CRT Monitor ......................................................................................... 15

4.3

LCD Monitor ......................................................................................... 17

4.4

Environmental conditions..................................................................... 18

4.5

Standards............................................................................................. 18

KEYBOARD AND FUNCTIONS


5.1

19

Functions description ........................................................................... 20

UNIT SWITCHING ON

22

6.1

CCD camera not connected .................................................................. 24

6.2

System fault ......................................................................................... 25

PATIENT DATA

26

FLUOROSCOPY

28

8.1

Fluoro special function ......................................................................... 30

MODALITY SELECTION

31

10

GRAPHY

33

(Rev. 0)

DIVA-D

USERS MANUAL
Contents

11

POST PROCESSING

35

11.1 Special function .................................................................................... 39


11.2 Multi image........................................................................................... 41
11.3 Image print ........................................................................................... 42

12

DICOM FUNCTIONS

43

12.1 Waiting list ........................................................................................... 44


12.2 Queue view and flush ........................................................................... 45
12.3 Print ..................................................................................................... 47
12.4 Store ..................................................................................................... 49
12.5 CD - ROM ............................................................................................. 51

13

MAINTENANCE AND CLEANING

54

13.1 Frequency of maintenance and cleaning actions ................................... 54

This publication can only be reproduced, transmitted, transcribed, or translated


into any human or computer language with the written consent of VILLA SISTEMI
MEDICALI S.p.a.
This manual in English is the original version.
DIVA-D

ii

(Rev. 0)

USERS MANUAL
Introduction

INTRODUCTION

NOTE:
The present manual is updated for the product it is sold with in order to
grant an adequate reference to use properly and safely the product.
The manual may not reflect changes to the product not impacting
operating modes or safety.

DIVA-D is an acquisition, storing and processing digital system studied


to process digitalized images.
The purpose of this manual is to give the User instructions for a safe and
efficient use of the DIVA-D.
The device must be used in compliance with the procedures described in
the user manual and never used for purposes different than the ones
described in the user manual.
Qualified personnel only, having necessary knowledge on X-ray
protection can use the device.

1.1

Icon appearing in the manual

This icon indicates a NOTE: please read thoroughly the items marked
by this picture.

This icon indicates a WARNING message; the items marked by this icon
refer to the safety aspects of the patient and/or of the operator.

(Rev. 0)

DIVA-D

USERS MANUAL
Safety aspect

SAFETY ASPECT

WARNING: Please read this chapter thoroughly.

VILLA SISTEMI MEDICALI devices comply with the related safety


requirements; furthermore it supplies all information necessary for a
correct use and the warnings related to danger associated with X-rays
generating units.
VILLA SISTEMI MEDICALI, has not to be held responsible for:

use of DIVA-D different from the intended one as described in this


manual;

damages to the unit, to the operator, to the patient, caused either by


wrong installation or maintenance procedures, different from those
described in the Servcie manual and in the user manual supplied
with the unit, and by wrong operations;

mechanical and/or electrical modifications performed during and


after the installation, different from those described in the official
documentation of the product.

Any technical intervention must only be performed by VILLA


SISTEMI MEDICALI qualified technicians.
Only the authorized personnel can remove the covers and/or have
access to the components under voltage.

DIVA-D

(Rev. 0)

USERS MANUAL
Safety aspect

2.1

Warning
This device has not been designed to be used in environments where
vapors, anaestethic mixes flammable with air, or oxygen and nitrous
oxide can be detected.
Avoid pouring water, even accidentally, or other liquids into the device,
as this could cause short-circuits.
Before cleaning the device, please disconnect it from the line voltage.
Wherever necessary, use the fit accessories, such as the leaded aprons,
to protect the patient from radiation exposure.
While performing the radiography, no one, apart from the operator and
the patient, must remain in the room.
Though this unit has been designed with a quite acceptable protection
level from electromagnetic interference, it is advisable to install it at a
certain distance from electrical energy transformation rooms, from static
continuity units, from portable receiving-transmitting units for
amateurial use and for cellular use. Cellular telephones are only
admitted at a distance of more than 1.5 m from any component of the
device.
Other medical instruments and devices that must be used in the same
installation area of the unit must comply with the Electromagnetic
Compatibility rules in force. Non-complying instruments, of which the
poor immunity from electromagnetic fields is well known, must be
installed at least 3 m away from the DIVA-D and supplied by a different
electrical line.
DIVA-D cannot be installed in operating theatres.
DIVA-D must be off while using devices such as electrical scapels or
similar instruments.
The equipment automatically performs some self-test functions and
therefore is able to detect some anomalous conditions. However,
whenever there are doubts on the performance of the equipment (e.g.
images too clear, too dark or with artifacts), the activity must be
suspended and a Villa Sistemi Medicali qualified and authorized engineer
must be contacted.

WARNING:
If during the self-test phase the equipment shows a failure message
caused by the electronic circuits or the Ram disks, the images of a given
patient retrieved for post processing may not belong to the requested
patient.
In this case, immediately contact an authorized Villa Sistemi Medicali
engineer and do not make any diagnosis or evaluation on the stored
images.
In addition, as the failure may have occurred during the last working
session, diagnosis carried out after the last exam cannot be
considered valid.
(Rev. 0)

DIVA-D

USERS MANUAL
Safety aspect

During the use of the equipment, the operator must comply with the
local work safety regulation, in particular with the ones referring to the
use of video terminals.
The cathode X-ray tube (CRT) of the monitors, when present, is
manufactured in order to minimize the risk of implosion. However, if
heavily hit it may break thus projecting out pieces of glass, powder and
toxic gas. Avoid hitting the CRT with any object and make sure that it
does not hit any object.
These risks are not present when LCD monitors are used.

DIVA-D

(Rev. 0)

USERS MANUAL
Safety aspect

2.2

Environmental risks and displacement


The DIVA-D contains in some of its parts, materials that at the end of the
units life, must be recycled at an approved location.
Particularly the device contains the following materials and/or
components:

Central unit and control console: iron, copper, aluminum, nonbiodegradable plastic material, glass-resin for printed circuit board
(PCB)

CRT Monitor, TV camera: glass vacuum tubes, iron, copper,


aluminum, glass-resin for printed board, non-biodegradable plastic
material.

WARNING
Glass vacuum tube may implode with projection of glass pieces, powder
and toxic gas.

(Rev. 0)

DIVA-D

USERS MANUAL
Safety aspect

2.3

Symbols used
Symbol

Description
Warning, read the relevant documentation
The equipment contains in some of its parts,
materials that at the end of the units life, must be
recycled at an approved location

A.C.

Complies with Medical Equipment Directive 93/42


0123

Device with type B applied parts

Dangerous voltage

DIVA-D

(Rev. 0)

USERS MANUAL
Description

DESCRIPTION

3.1

Identification labels
1
4

2
3

1
Main label (rack)

2
Control keyboard label

3
CRT Monitor label

4
LCD Monitor label

(Rev. 0)

DIVA-D

USERS MANUAL
Description

3.2

Description
DIVA-D is a digital acquisition, storing and processing system for
radiological images. The large storing capacity allows performing the
following functions:

Image acquisition and storing in FLUOROSCOPY

Image acquisition and storing in EXPOSURE with the following


working programs: SPOT, BONE, ANGIO, PERIF, STEP, TOMO

Post-processing of the acquired images, thanks to a dedicated


program

Printing of the acquired images through a laser printer.


Sending via Dicom network acquired images to Pacs system.

The images coming from the image intensifier are acquired through a
CCD camera, then fed to the digitalization stage (A to D Converter) after
that they can be stored in the permanent memory (Ram disk).
From this moment on the images can be processed thanks to the
dedicated software programs.
The system has a large accessible database, which allows for the entry
and retrieval of patient information along with exam description, remark
and notes.
DIVA-D is available in 2 versions:
DIVA-D RF 12
Ram Disk capacity

36 Gbytes

Operator monitor

1 or 2 on request

"In room" monitor

0, 1 or 2 on request

Imaging acquisition (max speed and


matrix)

12 imag/sec matrix 1024x1024


12 imag/sec matrix 1024x1024

Functions

standard acquisition:
imaging subtraction:
road mapping:
fluoro store:

YES
NO
NO
YES

DIVA-D AS 25
Ram Disk capacity

72 Gbytes

Operator monitor

1 or 2 on request

"In room" monitor

0, 1 or 2 on request

Imaging acquisition (max speed and


matrix)

25 imag/sec matrix 1024x1024

Functions

DIVA-D

25 imag/sec matrix 1024x1024

standard acquisition:
imaging subtraction:
road mapping:
fluoro store:

YES
YES
YES
YES
(Rev. 0)

USERS MANUAL
Description

3.2.1

DIVA-D composition
DIVA-D contains the following components:

3.2.2

Rack containing all electronics (computer, PCB's, Ram disk, power


supply, etc).

Control keyboard.
CCD TV camera.

POD (Printed Ouput/input Device) pcb interface for table and


generator.

Laser interface for standard digital protocol (P831) when Dicom


module is not present only.

Monitors
All DIVA-D models are available with CRT or LCD monitors for operators
and "in room".
The size of these monitors depends on customer request.

3.2.3

Accessories
Depending on the request, DIVA-D can be supplied with accessories:

(Rev. 0)

DICOM module equipped with SW package for:


print
store
worklist
CD write

Remote control function.


Ceiling suspension for "in room" monitor.

DIVA-D

USERS MANUAL
Technical data

TECHNICAL DATA

4.1

NOTE:
Technical data, performance and instructions provided in this manual
are referred to the DIVA-D AS 25 model. Therefore, some of these
characteristics are not applicable for DIVA-D RF 12 systems.

DIVA-D AS 25

Identification
Type

DIVA-D AS 25

Model

NDR 632 E00 + NDR transf 02

Components
Central unit

Rack composed by:


Power supply; Computer; Pcb's; Ram
disk

Control system

Flat control keyboard

Acquisition by

CCD Camera 1024 x 1024

Interface system

Generator and Remote Control Table


interface PCB

Connection cables

Prepared interconnection cables

Monitor

Operator: 1 or 2 on request
"In room": 1 or 2 on request

Dimension and weight Central Unit (Rack)


Weight

15 Kg

Height

254 mm

Depth

345 mm

Width

440 mm

Dimension and weight Control keyboard


Weight

6 Kg

Height

50 mm

Depth

235 mm

Width

460 mm

DIVA-D

10

(Rev. 0)

USERS MANUAL
Technical data

Dimension and weight CCD Camera


Weight

2 Kg

Height

95 mm

Diameter

140 mm

Dimension and weight Interface Card (POD)


Weight

about 0,5 Kg

Height

20 mm

Depth

98 mm

Width

200 mm

Dimension and weight Laser Printer Interface Box


Weight

about 1 Kg

Height

38 mm

Depth

150 mm

Width

235 mm

Technical characteristics / Acquisition and memorization


Acquisition, visualization and
memorization

With digital processor and an A/D


converter 12 bits

Memorization of the images

12 bit

Visualization of the images

12 bit

A/D frequency conversion

66 MHz

Processor management

AMD

Clock frequency of the processor

1,2 GHz

CCD camera

With medical grade 2/3" CCD sensor


offering high resolution with progressive
scanning.
1024x1024 matrix.
Integrated optical system with focus and
diaphragm.

CCD camera signal-noise ratio

During exposure acquisition <65 db

Control keyboard

Alphanumerical with special function.


Qualified functions indicated by LED
display.
Flat and waterproof surface.

Generator and remote control table


interface card

Remote table: with dedicated input \


output system.
Generator: Digital mode; analogic mode;
serial mode (RS 232)

TM

Laser printer interface box (when Dicom With parallel output for digital
module not present).
connections.
Communication protocol P831, P859.
Maximum connection distance 500m.
(Rev. 0)

11

DIVA-D

USERS MANUAL
Technical data

DICOM module (optional) Technical characteristics


Processor management

Intel

Working frequency

1 GHz

Ram memory

512 Mbytes

Transit memory

Ram disk 40 Gbytes

Dicom images backing

Ram disk 40 Gbytes

Network access speed

100 Mbit

SW package available

-Print (1024x1024 16bit)


-Store (1024x1024 16bit @ 3 Mbyte)
-CD write (1024x1024 8bit)
(512x512 8 bit)

-Worklist
Imaging acquisition (speed and matrix)
Mode

12 bits progressive scanning acquisition


and visualization

Fluoroscopy

Matrix 1024x1024
Acquisition max rate 25 fr/sec

Pulsed fluoroscopy

Matrix 1024x1024
Acquisition selectable rate:
1-3-6-12-25 fr/sec

Exposure and angiography

Matrix 1024x1024
Acquisition max rate 25 fr/sec

Imaging memorization
Storage

Automatic and permanent.

Maximum access time to disk for image


write and read (with matrix 1024x1024
16 bit)

30 msec.

Memory capacity

72 Gigabytes

Maximum storage capacity of images


with matrix 1024x1024

36000 images

Electrical characteristics
Standard power supply (transf 02)

230 V ac 15%

Standard power supply (DIVA-D rack)

24 V ac 15%

Frequency

50-60 Hz

Current absorbed

650 mA @ 230 V ac

Power absorbed

150 W @ 230 V ac

DIVA-D

12

(Rev. 0)

USERS MANUAL
Technical data

Working configuration
Remote R\ F control table

- Apollo
- Mercury 332

R\ F tilting table

- Vision

Integration with generator

Genius 65/80 HF-A with serial mode


(RS 232) interface

Clinical application
Fluoroscopy

Continuous and pulsed HCF

Standard exams and with contrast agent Gastroenterological, bone exams, ERCP,
Urology, Mielography, etc.
Tomography

In low dose modality

Vascular exams

Thoracic angiography, carotid,


peripheral, phlebography.
Exams with on-line subtraction and
post processing.

Peripheral steps Angiography

Optional. Possible application with


Apollo remote control table only.

Working characteristics
User interface

Windows representation

Fluoroscopy mode
Storage

Manual start on Ram disk

Regulation of the parameter

Auto video level regulation.


Auto radiological parameter regulation.
Manual selection of six field video levels

Edge enhancement

Two selectable levels

Positive/Negative function (grey scale


inversion)

In line/off line

Noise reducer

Two selectable level

Road mapping

Direct masking

Imaging comparison

Automatic or manual transfer of the


image

Last image hold (LIH)

With integrated image

Manual regulation of the video level

Three selectable levels

Motion detection filter

Automatic exclusion of the recursive


filter during the movement of the image
(no blurring effect)

Dynamic recursive filter DRF

Automatic noise/blurring optimization


effect offers zero persistence

(Rev. 0)

13

DIVA-D

USERS MANUAL
Technical data

Radiography mode
Storage

On Ram disk

Five standard acquisition modality with


different functions.
One optional acquisition modality

Spot Bone Angio - Peripheral angio


Tomo.
Stepping angio (optional)

Anatomic program for each acquisition


modality

20

Low dose acquisition

For pediatric exams

Maximum opacification function


(maxop)

Track due to the storing of the contrast


liquid flow with subtraction acquisition
mode

Acquisition time selection

Auto-self in according with the selected


program.
Manual in real time.

Positive/Negative (grey scale inversion)

In line/off line

Last image hold (LIH)

Each exposure

Post processing mode


Electronic collimator

With scrolling

Edge enhancement

Nine regulation levels

Windowing

Auto/manual

Road map contrast

Nine regulation levels

Positive/negative (grey scale inversion)

Default or manual

Image visualization

Full screen or multi format size

Angle measurement

Displayed data

Length measurement

Displayed data and ratio

Editing on the image

Available in full screen

Zoom

Zoom x 4 with scrolling

Exams review

With acquisition speed; image stand;


single reply

Visualization with subtraction

Possible with standard acquisition


exams

Exams with subtraction review

With Remasking and Maxop functions

Optimization of the subtraction

Pixel shift function

Image printing

Full screen

DIVA-D

14

(Rev. 0)

USERS MANUAL
Technical data

4.2

CRT Monitor

NOTE:
The characteristics of the monitors are the same for both DIVA-D's
model.

Technical data
Type

Monitor CRT B&W

Manufacturer

Nical

Model

Refer to optional

Dimensions and weight

Monitor 17

Monitor 21

Weight

17,5 Kg

21,8 Kg

Height

335 (359) mm

394 (418) mm

Depth

350 mm

380 mm

Width

405 mm

489 mm

17" (414 mm)

21" (511 mm)

Screen dimensions
Electrical characteristics
Standard power supply

230 V ac 15%

Frequency

50-60 Hz

Current absorbed

350 mA

Power absorbed

80 W

Scanning
Type

Interlaced

Line number

1049

"Refresh" frequency

60 Hz

Standards

EIA

Video input signal


Type

Video input connected ac

Connection type

BNC

Impedance

75

Input video signal

0,5 to 1,5 V pp (negative


synchronism)

(Rev. 0)

15

DIVA-D

USERS MANUAL
Technical data

Display characteristics
Deflection angle

90

Aspect ratio

4:3

Band width

40 MHz 3 dB

Raster

0.25 mm

Phosphor

White P45

Geometric error

<1.5 %

Vertical/horizontal linear error

<1.5 %

Maximum brightness

850 cd/m2

Light transmission

46%

Dynamic focus

Yes

Stabilized EHT

Yes

Optional
Contrast and brightness control with digital
push-button on front panel and rear remote
control connection

Available on request on all models

Self luminosity sensor on front panel


Contrast and brightness control only with
rear remote control connection

DIVA-D

16

Available on request on all models

(Rev. 0)

USERS MANUAL
Technical data

4.3

LCD Monitor

NOTE:
The characteristics of the monitors are the same for both DIVA-D's
model.

Technical data
Type

Monitor LCD B&W

Manufacturer

Nical

Model

Refer to optionals

Dimensions and weight

Monitor 18

Monitor 20

Weight

5,5 Kg

6,0 Kg

Height

340 mm

355 mm

Depth

60 mm

60 mm

Width

430 mm

457 mm

18"

20"

Screen dimensions
Electrical characteristics
Standard power supply

230 V ac 15%

Frequency

50-60 Hz

Current absorbed

250 mA

Power absorbed

50 W

Display characteristic
Visibility angle

180

Contrast

600:1

Resolution

1280x1024

Gray scale

768 levels

Max luminance

700 Cd/m2

Calibration

automatic

Optional
Contrast and brightness control with digital
push-button on front panel and rear remote
control connection

Available on request on all models

Contrast and brightness control only with


rear remote control connection

Available on request on all models

(Rev. 0)

17

DIVA-D

USERS MANUAL
Technical data

4.4

Environmental conditions
Operative

Temperature: from -10 to +45 Celsius


Humidity:
up to 95 % not
condensing
Temperature: from -35 to +65 Celsius
Humidity:
up to 95 % not
condensing

Transport and storage

4.5

Standards
DIVA-D system with accessories and optional complying with the relevant
standards:
EN 60601-1

Electro medical equipment. General safety requirement


(included amendment A1+A2+A11+A12+A13)

EN 60601-1-2

Electro medical equipment. General safety


requirement.
2 Collateral standards: Electromagnetic
compatibility.

EN 60950

Information device safety

0123

The symbol CE grants that DIVA-D complies with


European directive 93/42

Further information regarding standards complying can be requested to:


VILLA SISTEMI MEDICALI S.p.A.
Via delle Azalee, 3
20090 - Buccinasco - (MI) - Italia
Tel: +39-02-48.859.1
Fax: +39-02-48.818.44
The manufacturer of the equipment DIVA-D is (according to the
93/42/CEE European directive):
NICAL S.p.A.
Via Soffredini, 43
20126 - Milano - Italia
Tel.: +39-02-25.711.10
Fax: +39-02-25.722.07

DIVA-D

18

(Rev. 0)

USERS MANUAL
Keyboard and functions

KEYBOARD AND FUNCTIONS


All functions of DIVA-D are activated through buttons located on a
dedicated keyboard.
The following picture and relevant buttons functions descriptions permits
to locate easily all reference of the present manual.

(Rev. 0)

19

DIVA-D

USERS MANUAL
Keyboard and functions

5.1

Functions description

To adjust the kV value when automatic mode is


active.
It is possible to increase and decrease the image
level with XX steps.
R.O.I. selection. 6 different areas can be selected for
the auto kV correction in function of dimension and
position of the image area to be optimized.
To store fluoroscopy images.
To enable the road mapping function. It is used as
support during catheter insertion.
To select the fluoro modality between continuous
and pulsed.
To select the frame rate exposure acquisition.

When a new patient must be recorded the key


permits to access the patient data window.
Mode is used to enter into modality and anatomic
exam selection.
To enter into patient data base to perform post
processing mode.
In post processing mode, when a patient is recalled
from data base, to select run into the patient and
the frame into run.
To play a run that contains more then one frame
and to stop the sequence visualization.

DIVA-D

20

(Rev. 0)

USERS MANUAL
Keyboard and functions

These keys allows to enter into different post


processing mode:
symmetric and asymmetric image limitation
negative-positive
zoom x4 with scroll
cine loop review
multi image runs selection.
These keys allows to enter into different post
processing angiographic mode:
mask shift with 1/8 pixel step
mask present or removed
manual mask selection
remasking.
To

modify images quality in post processing:


manual windowing
edge enhancement
auto windowing
reset of the modifications.

To access into fluoroscopy or post processing


special function.
To enter into DICOM (optional) modality, to save
data into DICOM module and to print or store the
captured images.
This group of keys are used as auxiliary functions
during the different modality.

Alphanumeric keys used to write characters in


different working modalities.

This manual is divided in different chapters, each of one is dedicated to a


working modality which DIVA-D supports.
In any chapter it is explained how to use the buttons located on the
keyboard to optimize the DIVA-D use.

(Rev. 0)

21

DIVA-D

USERS MANUAL
Unit switching ON

UNIT SWITCHING ON
Any time that DIVA-D is switched on, it performs an auto test to verify if
the CCD camera is connected and correctly working and if the electronics
of the main rack, including RAM disks are normally operating.
At start up, self test starts and the window is displayed:

where software and hardware information is displayed.


In the first area the software versions present in the unit are indicated.
This information allows the user and service engineer to identify the unit
configuration.
In the second area the hardware characteristics and the license level
are displayed. This field permits to identify which functions are available
on the DIVA-D.
At the end of start up, three conditions are possible:
-

No error code
The Unit automatically goes to full operation and the display shows.

SYSTEM READY
Device is ready to operate.

DIVA-D

22

(Rev. 0)

USERS MANUAL
Unit switching ON

If a DIVA-D is equipped with DICOM module after about 2 minutes


from this condition, the monitors shows:

SYSTEM READY
DICOM READY
This delay is determined by the DICOM module initialization time
that is longer than the DIVA-D one.

(Rev. 0)

Camera not connected (see paragraph 6.1)


It is necessary to press ENTER key to escape from this condition.
Unit goes to off line operation only.

System fault (see paragraph 6.2)


Unit stops.
A service engineer is required.

23

DIVA-D

USERS MANUAL
Unit switching ON

6.1

CCD camera not connected


During auto test, the unit could recognize a fault in the CCD camera
connection.
In this condition the window shows:

It is necessary to press ENTER key to escape from this message.


In this condition the device goes into post processing mode in which it
will be possible to perform review only of exams already stored.
No acquisition of new images is possible.

DIVA-D

24

(Rev. 0)

USERS MANUAL
Unit switching ON

6.2

System fault
If auto test recognizes an analogic, digital or disks fault the window
shows:

The unit is out of work and a service engineer must be contacted.

(Rev. 0)

25

DIVA-D

USERS MANUAL
Patient data

PATIENT DATA
Before starting with a new exam is necessary to insert into DIVA-D the
data of the new patient.
If this is not done, as soon the exposure images acquisition is required,
the message:

is displayed.
It is necessary to press NEW PATNT key to open the patient data
entering window.

NOTE:
If fluoroscopy mode is activated, without storing any LIH or any fluoro
sequence, the images visualization is present also if no patient data is
inserted. This because a patient could perform a fluoroscopic
investigation only. In all other cases, because all radiographic images are
automatically stored on RAM disks, the patient data must be entered into
the data base.

When the key NEW PATNT is activated the new patient window is
displayed:

DIVA-D

26

(Rev. 0)

USERS MANUAL
Patient data

If DICOM module is present the worklist selection is allowed, otherwise


no.
For this function, please refer to DICOM functions chapter.
By using UP or DOWN arrow keys activate the selection required between
"DATA ENTRY" or "LAST PATIENT".
Press ENTER key to confirm.
If data entry is selected, the window appears:

By using UP or DOWN arrow keys select the required fields.


The surname field is mandatory, the others are optional.

NOTE:
If the birth date is not entered, the patient age in the data base will be
recorded as 0.
If sex is not inserted, O (other) will be present in the data base).

Confirming by ENTER, the data are stored into the data base and the
modality window is displayed to permit the acquisition mode selection.
If "last patient" is selected, the PATIENT DATA window appears with the
data relevant to the last patient stored.
This selection is used when during exam on a patient, the unit has been
switched off.
This feature is very useful to avoid 2 files for the same patient.

(Rev. 0)

27

DIVA-D

USERS MANUAL
Fluoroscopy

FLUOROSCOPY
If DIVA-D is equipped with 2 monitors in the control desk area, one of
them is reserved to fluoroscopy images only, otherwise the fluoro window
is present any time the fluoro command is activated.

The following functions and information are available in this modality:

Status
The visualization could be X-ray during acquisition or L.I.H. (Last
Image Hold) in off line.

Reg. of int
The visualization is modified according to selection.
ROI (Region Of Interest) must be modified in function of position and
dimension of the organ. 6 different ROI are available trough
correspondent keys. Maintaining pressed the key relevant to the ROI
selected for 3 sec, the ROI dimension and position is displayed on the
image.

Modality
This field could be STD or PULSE xx fr/sec.
the fluoro mode can be selected.
By means of the key
When led is off, STD is displayed in the window, meaning that the
fluoro mode selected is continuous.
When led is on PULSE with the frame rate is displayed in the
window. This rate can be modified using keys FR/s + and FR/s -.

DIVA-D

28

(Rev. 0)

USERS MANUAL
Fluoroscopy

Store
In normal use the condition is OFF. To activate the fluoro storing
modality it is necessary to press STORE. The relevant led turns on
and the images number stored are indicated at the bottom of the
display.
When fluoro acquisition is interrupted also the storage is stopped.
At new fluoro acquisition it will be necessary to press the STORE key
again to perform a new storage.
In LIH condition, pressing the STORE key, the LIH will be stored into
the hard disk.

Edge
By default the value is OFF. It means the edge level used is the
default one. Using E key, the visualization is ON and the edge level
selected is the one configured at the installation.

Noise
By default the value is OFF. It means the noise suppression level
used is the default one. Using N key, the visualization is ON and the
noise level selected is the one configured at the installation.

Road map
Road mapping function is available on DIVA-D AS25 only.
The function is activated by using the ROAD MAP key. Pressing the
key, the relevant led turns on and the fluoro images will be in
subtractions mode.
Releasing fluoro, on the display the LIH will be in sub mode and
following fluoro images will have the mask overlapped.
To reset the road mapping mode and the mask, press the key with
fluoro off.
The Led will be turned off and next fluoro will be performed in
standard mode.

Kv mA Dose
In these fields the actual kV and mA are displayed.
Dose field is not yet activated.
Run Frame
In these fields the fluoro run number stored for the present patient
and the images contained in the run are displayed.

The dose level can be modified by using the keys kV+ and kV-.
The effect of this action is the image contrast level becoming higher or
lower depending by the key activated. 3 decreasing steps and 3
increasing steps are available starting from the standard level.
The radiological parameters are automatically adjusted by the system,
during fluoro acquisition to optimize the image video level.
This adjustment is calculated based on ROI selection.

(Rev. 0)

29

DIVA-D

USERS MANUAL
Fluoroscopy

8.1

Fluoro special function


In fluoro off line mode, by pressing the SPEC FUNCT key, the relevant
window is displayed:

By using UP or DOWN arrow keys select the field required.


Press RESET key to activate the relevant selection.

DIVA-D

Store XX Fr/s
The storage rate in fluoro can be selected to preserve disk space.
This selection modifies the storage speed only without modifying the
visualization rate.

Standard / Pulsed
This selection can also be done in on line mode by using the pulsed
key as described above.

Ref image
This selection appears when DIVA-D is equipped with 2 monitors in
control desk area only.
When this selection is active the graphy monitor is used to display
the LIH while on fluoro monitor the live images are displayed.

Pattern 1 / 2
Allows to display 2 different patterns in order to adjust correctly the
monitor brightness and contrast.

30

(Rev. 0)

USERS MANUAL
Modality selection

MODALITY SELECTION
When patient data are inserted, the modality and relevant anatomic
programs must be chosen.
The access to this function is automatic when exiting from patient data
window with ENTER key or any time the MODE key is activated.
This because when a new patient is inserted it is necessary to select the
modality for the first acquisition, and to change this modality if a patient
requires different exams type.
The acquisition modality can be:
SPOT
BONE
ANGIO
PERIF
STEP
TOMO.
SPOT, BONE and TOMO are allowed for both DIVA-D models RF and AS,
ANGIO and PERIF on AS only, STEP could be present on a AS model only
as supplementary option.
The step modality is allowed only if the system is based on APOLLO table
only.
The modality window looks as follows:

It is divided in 3 areas:
modality selection
anatomic selection
parameters modification.
By default the left area is active when modality is selected. To inform
about the area selected, the background is light gray.

(Rev. 0)

31

DIVA-D

USERS MANUAL
Modality selection

By using UP and DOWN arrow keys, select the required modality, by


using RIGHT arrow key select the anatomic area, this background
becomes light grey and select now the desired anatomic program.

NOTE:
The not available modality (depending by the DIVA-D license level) will be
displayed in gray and not reachable by selection.

On the right area the parameters associated to the selected anatomic


program are displayed.
To modify permanently one or more parameters, press RIGHT arrow key,
the background will be light gray (as in picture) and by using relevant
keys, the associated parameter can be changed.
Using alphanumeric keys the anatomic name will be modified and the
message ENTER to save will appear on window bottom side.
By ENTER key the new anatomic configuration will be stored.
If a temporary modification, on an existing anatomic program, need to be
done it will be sufficient to modify the relevant parameter by using the
associated key.
Confirming by ENTER key, the acquisition mode will be selected with the
modified parameters.
In modality function, the parameters associated will be different as
summarized below:
SPOT

BONE

ANGIO

PERIF

STEP

TOMO

ROI

ROI

ROI

ROI

ROI

ROI

Frame rate

Edge

Frame rate

Frame rate

Frame rate

Tomo frames

Edge

Low dose

Edge

Edge

Edge

Edge

Low dose

Neg Pos

Low dose

Low dose

Low dose

Low dose

Neg Pos

Neg Pos

Neg Pos

Neg Pos

Auto mask

Auto mask

Auto mask

Pulsed fluoro

Sub cont

Sub cont

Neg -Pos

Sub cont
Kv reduction

DIVA-D

32

(Rev. 0)

USERS MANUAL
Graphy

10

GRAPHY
When the modality selection is done and confirmed, DIVA-D is ready for
graphy acquisition.
The window displayed is:

on which the patient data, modality and anatomic parameters are


displayed.
Now the acquisition can start or anatomic parameters modification can
be done by using the associated keys.
If a key for modification is pressed (i.e. ROI, Frame/sec ecc) the modality
window will be displayed.
Before to start with graphy acquisition, a fluoro acquisition must be
done.
This action allows DIVA-D to calculate the correct exposure parameters
(technique factors).
As matter of fact at the exposure request, DIVA-D informs the generator
about the technique factors to be used for the graphy acquisition.
The area Run is reserved to the number of exams stored for the present
patient.
During acquisition the areas Run, Frame, mSec and kV are
updated.

(Rev. 0)

33

DIVA-D

USERS MANUAL
Graphy

At the end of acquisition the following window will be displayed:

DIVA-D

34

(Rev. 0)

USERS MANUAL
Post processing

11

POST PROCESSING
To review the exams contained for the current patient, 2 options are
possible:
by MENU key
by STOP key.
Using STOP the Post Processing window is immediately displayed,
selecting MENU the patient selection window will appears.

Select the required patient using UP and DOWN keys.


Press ENTER to confirm.

(Rev. 0)

35

DIVA-D

USERS MANUAL
Post processing

The window reporting the data relevant to the different runs of the
patient is displayed.
Select the desired run using UP and DOWN keys.
Press ENTER to confirm.
The post processing window appears:

On this window, in addition to the data relevant to patient, acquisition


modality, anatomic program etc, the following information is present:
ROI used to acquire the present run
Frame rate used to acquire the present run
I.I. field used to acquire the present run
mSec exposure time.
In

DIVA-D

addition to these information the following parameters are present:


total runs number and run showed
total frame number present on the active run and the frame showed
in angio or perif mode only, the mask number selected
in angio or perif mode only, the sub contrast value
the windowing level
the edge level.

36

(Rev. 0)

USERS MANUAL
Post processing

All of them and other settings can be modified using the dedicated keys:

To select run in a patient and frame in a run.


The frame keys allow to review a run frame by frame
independently by frame rate acquisition.
To play a run containing more then one frame and
to stop the visualization sequence.
The frame rate in play mode is the same used
during acquisition and cannot be modified.
To introduce electronic diaphragm.
When this function is activated (led ON) by using
UP, DOWN, LEFT and RIGHT keys, upper and right
shutters will be closed and opened.
Using the four arrows contemporarily with SHIFT
key the lower and left shutters will be activated.
To reset the function press the key.
To reverse the gray scale of image from negative to
positive and reverse.
To zoom the present image 4 times.
When this function is activated (led ON) by using
UP, DOWN, LEFT and RIGHT keys, it is possible to
scroll the enlarged image.
To reset the function press the key.
When this function is activated the frames present
in the run are displayed with the frame rate used in
acquisition. The number of sequence depends by
the technical setting.
At the end of the last sequence the led turns OFF.
Any time the cine loop is required, press the key to
activate the function.
This function is available for exams acquired in
angio mode only. When the subtracted image is
displayed the pixel shift function can be activated
by relevant key. The led become ON and the mask
position can be modified by 1/8 pixel step. This
movement is obtained by UP, DOWN, LEFT and
RIGHT keys.
This function is available for exams acquired in
angio mode only. When the subtracted image is
displayed the mask can be removed or replaced by
means of the relevant key.
This function is available for exams acquired in
angio mode only. When a run is scrolling the
contained frames, using this key a new mask can be
selected. The mask will be captured at the exact
moment when the key is pressed.
This function is available for exams acquired in
angio mode only. When subtracted image is
displayed it is possible to increase or decrease the
image mask number.

(Rev. 0)

37

DIVA-D

USERS MANUAL
Post processing

To modify images characteristics by means of the


relevant keys it is possible to change:
windowing (changing the black and white levels)
edge enhancement.
With AUTO WNDW key the windowing will be
automatically optimized at default level. Pressing
reset key any modification described above is reset.
When subtracted image is displayed it is possible to
increase or decrease the image background level by
means of the keys on the left side.

In post processing special functions are available in


addition to the functions described. This functions
are described in the next paragraph.

DIVA-D

38

(Rev. 0)

USERS MANUAL
Post processing

11.1

Special function
When the post processing modality is activated by means of the SPECIAL
FUNCTION key the lower part of the window:

is represented as in the above figure.


By means of the UP and DOWN key it is selectable one of the four special
functions available:
overwrite
angle measure
length measure
test pattern display.

(Rev. 0)

Overwrite
Allows to report comments on the image. To position the cursor on
the required place use UP, DOWN, LEFT and RIGHT arrow keys.
Press ENTER key to confirm when the text insertion is completed.
Changing the image or using STOP key, the text inserted is cancelled.

Angle measure
Allows to take a measure of the angle of a part of image.
Using UP and DOWN keys select the approx level from the table top
on which the part to be measured is located between the 3 levels
proposed 0, 10 or 20 cm. Press ENTER to confirm.
To position the cursor on the required place use UP, DOWN, LEFT
and RIGHT arrow keys.
Press ENTER key to confirm the starting measuring point. Move the
cursor on the angle vertex and confirm whit ENTER. Move the cursor

39

DIVA-D

USERS MANUAL
Post processing

at the end of second segment of the angle, the angle value is


displayed.
Changing the image or using STOP key, the text inserted is cancelled.

WARNING:
Because of different reasons ( II deformation, error in the level organ
position and others) the angle value obtained are only indicative. Do not
use this value for diagnostic purpose.

Length measure
Allows to take a measure of a part of the image.
Using UP and DOWN keys select the approx level from the table top
on which the part to be measured is located between the 3 levels
proposed 0, 10 or 20 cm. Press ENTER to confirm.
To position the cursor on the required place use UP, DOWN, LEFT
and RIGHT arrow keys.
Press ENTER key to confirm the starting measuring point. Move the
cursor on the final measuring point and confirm with ENTER, the
length value is displayed.
Changing the image or using STOP key, the text inserted is cancelled.
More than one measure can be done on the same image. To do that
repeat the above described steps.
Ratio between 2 measure can be done selecting the relevant window.

WARNING:
Because of different reasons (II deformation, error in the level organ
position and others) the measures obtained are only indicative. Do not
use this value for diagnostic purpose.

DIVA-D

Test pattern
Allow to display a test pattern to adjust correctly the monitor
brightness and contrast.
Starting with any modality the test pattern is reset from monitor.

40

(Rev. 0)

USERS MANUAL
Post processing

11.2

Multi image
When the patient menu window is displayed, multi image function can be
used.
It is a short and easy way how to select the run to be review.
Using the
key, on monitor the first image of any run present into
the selected patient is displayed.

By means of the UP, DOWN, LEFT and RIGHT arrow keys, select the
desired run and press ENTER to confirm.
The run selected will be displayed in full screen mode.

(Rev. 0)

41

DIVA-D

USERS MANUAL
Post processing

11.3

Image print
If DIVA-D is not equipped with DICOM module, it is possible to send the
image to a laser printer directly.
The image present on display is also present on the printer output
channel.
To capture this image it is necessary to confirm from the printer
keyboard supplied by the printer manufacturer.

DIVA-D

NOTE:
Do not change the image displayed until the laser printer acquisition is
completed.

42

(Rev. 0)

USERS MANUAL
DICOM functions

12

DICOM FUNCTIONS
If the DIVA-D is equipped with DICOM module, the DICOM key is active.
By means of this key the following window is displayed:

Depending from the license present the following functions can be


activated:
print (to send images to a DICOM laser printer)
store (to send images to a Pacs via network)
CD-ROM (to burn CD-ROM with images).
In addition to these functions another software package can be present:
waiting list.

(Rev. 0)

43

DIVA-D

USERS MANUAL
DICOM functions

12.1

Waiting list
This function is allowed if the network is equipped with HIS/RIS.
When new patient window is displayed (chapter 7) the waiting list
selection is allowed.
Confirm by ENTER, the window

is displayed.
On this window the patients present on the worklist are displayed.
By means of the UP and DOWN arrow key select the desired one, press
ENTER to confirm.
The patient data window will be displayed (chapter 7) reporting all data
inserted in HIS/RIS.
By this function any manual patient data entering is avoided.

DIVA-D

44

(Rev. 0)

USERS MANUAL
DICOM functions

12.2

Queue view and flush


To enter into DICOM functions the dedicated key must be used.
The window

will be displayed.
If no queue error or messages are present, the last used function will be
proposed, anyway the queue view will be reachable by means of ENTER
key directly.
In any case UP and DOWN arrow keys can be used to reach the function
desired or queue view.
If errors or messages are present in the queue, the relevant field is with
white background; pressing ENTER key the window

will appears.
(Rev. 0)

45

DIVA-D

USERS MANUAL
DICOM functions

The queue list will be showed and erasable.


To cancel the list, using UP and DOWN arrow keys, activate FLUSH
function and confirm with ENTER key.

DIVA-D

46

(Rev. 0)

USERS MANUAL
DICOM functions

12.3

Print
When DICOM window is selected, the last function used in DICOM mode
is proposed.
By means of UP and DOWN arrow keys select the required function and
confirm with ENTER key.
If print function is selected the printer window selection is showed.
With UP and DOWN arrow keys, select the required printer and confirm.
With LEFT and RIGHT arrow key select the film subdivision and confirm.
The DICOM print window

is displayed.
Select the image required and modify, if necessary the characteristic as
in standard post processing mode.
In the lower section of the window, the frame number and position are
displayed.

(Rev. 0)

47

DIVA-D

USERS MANUAL
DICOM functions

To add the image, the


time the window

key must be pressed 2 times. After the first

is displayed.
Pressing once more the addition is confirmed and the image stored into
DICOM print temporary archive.
The previous window will be showed allowing a new image addition.
When the above window is present, 3 other functions are available (use
UP and DOWN arrow keys to select and press ENTER to confirm):

Delete & exit


To reset the temporary DICOM print archive and exit from DICOM
functions.

Study print
To print the image stored.
To print the study it is also possible to press 2 times the
starting from the main DICOM print window.

DIVA-D

key

Change frame
To store the present image instead of an image already present into
the DICOM print archive.
To do that use LEFT and RIGHT arrow keys, the white portion of the
film will be moved on the film representation showing the place on
which the present image will be located.
Press ENTER to confirm, the previous image will be erased and the
new one stored.

48

(Rev. 0)

USERS MANUAL
DICOM functions

12.4

Store
By means of UP and DOWN arrow keys select the store function and
confirm with ENTER key.
When store function is selected the window is showed:

On the lower part of the window the following information are listed:

(Rev. 0)

Total Mbytes
Mbytes used by the images stored in the present DICOM store study.
This is referred to the temporary DICOM archive.

Total Patient
This value can be 0 or 1.
As matter of fact a DICOM study to be stored cannot contain more
than 1 patient.

Total runs
Is the number of runs stored to be transferred in the network.

Total frames
Is the number of single image to be transferred in the network.

Total sub-runs
Is the number of sub-runs stored to be transferred in the network.

49

DIVA-D

USERS MANUAL
DICOM functions

To add the image, the


first time the window

key must be pressed two times. After the

is displayed. Pressing once more adds the selection to the archive.


By means of UP and DOWN arrow keys select if the memory addition
will be of the showed frame only, of the run or of the total patient.
Press ENTER to confirm.
The previous window will be showed allowing a new addition.
When the above window is present, 2 other functions are available
(use UP and DOWN arrow keys to select and press ENTER to
confirm):

Delete & exit


To reset the temporary DICOM store archive and exit from DICOM
functions.

Study send
To send the study stored.
To send the study it is also possible to press 2 times the
starting from the main DICOM store window.

DIVA-D

50

key

(Rev. 0)

USERS MANUAL
DICOM functions

12.5

CD - ROM
By means of UP and DOWN arrow keys select the CD-ROM function and
confirm with ENTER key.
When CD-ROM function is selected the media type window is showed:

allowing to select the CD-ROM format between High resolution


1024x1024 or High speed 512x512.
High resolution saves images with 1024x1024 matrix but for static
images only.
High speed has matrix 512x512 but is capable to store images up to
25 images/sec.
By means of UP and DOWN arrow keys select the media type required
and confirm with ENTER.

(Rev. 0)

51

DIVA-D

USERS MANUAL
DICOM functions

The following window will be displayed:

On the lower part of the window the following information are listed:

DIVA-D

CD free space
The disk percentage space still available.

Total Patient
This value can be 0 or 1.
As matter of fact a DICOM study to be stored cannot contain more
than 1 patient.

Total runs
Is the number of runs stored to be transferred in the network.

Total frames
Is the number of single image to be transferred in the network.

Total sub-runs
Is the number of sub-runs stored to be transferred in the network.

52

(Rev. 0)

USERS MANUAL
DICOM functions

To store into temporary DICOM archive, select the patient, the run or the
image required and press ENTER. The window:

is displayed.
By means of UP and DOWN arrow keys select if the memory addition will
be of the showed frame only, of the run or of the total patient.
Press ENTER to confirm.
The previous window will be showed allowing a new addition.
When the above window is present, 2 other functions are available (use
UP and DOWN arrow keys to select and press ENTER to confirm):

Delete & exit


To reset the temporary DICOM CD-ROM archive and exit from
DICOM functions.

Burn disk
To send the study stored to CD-ROM.

Before to select "burn disk", put a new CD-ROM into driver and press
key.
The dicomization procedure will start and at the end of this operation the
CD-ROM burning start.
This sequence is displayed on the window with a waiting notice.

(Rev. 0)

NOTE:
If no CD-ROM or not valid one is present into the driver, an error will be
present in the queue.
DIVA-D will try again to access the CD after 30 sec.
Insert a valid CD-ROM into the driver before a new writing attempt.

53

DIVA-D

USERS MANUAL
Maintenance and cleaning

13

MAINTENANCE AND CLEANING


Like all electrical device, this equipment requires periodical checks.
This will insure an efficient and correct functioning for years to come.
This equipment is made up of electrical and electronic parts that need to
be checked out in order to guarantee high quality radiological results as
described in previous chapter.
The maintenance and cleaning procedures described in this chapter
must be done by the user.
In addition to the above, the authorized service technician will have to
perform the maintenance actions described in the Service Manual.

WARNING:
Before proceeding with the cleaning of the equipment, disconnect it
from the mains.

13.1

DIVA-D

Make sure that water and any other liquids are not spilled inside the
equipment as this may cause corrosion and short circuits.

Clean the painted surfaces by using a moist cloth and neutral


detergent, taking care to wipe them by using a dry cloth. Do not use
solvents (alcohol, petrol, and trichloroethilene) or corrosive or
abrasive products.

Frequency of maintenance and cleaning actions


Frequency

Description

1 month

Check the presence of labels and their condition

1 months

Check of the image quality.

6 months

Clean the monitors and the external parts.

6 year

Visual checking of the status of the interconnection cables


and relevant connection.

54

(Rev. 0)

Cod. 6995001003_Rev.0

VILLA SISTEMI MEDICALI


Via Delle Azalee, 3
20090 Buccinasco (MI) - ITALY
Tel. (+39) 02 48859.1
Fax (+39) 02 4881844

0123