DIVA-D (User's Manual)
DIVA-D (User's Manual)
0123
Users manual
Release 4 January 2006 (Rev. 0)
USERS MANUAL
Contents
Contents
1
INTRODUCTION
1.1
SAFETY ASPECT
2.1
Warning.................................................................................................. 3
2.2
2.3
DESCRIPTION
3.1
3.2
Description ............................................................................................. 8
3.2.1 DIVA-D composition ................................................................... 9
3.2.2 Monitors..................................................................................... 9
3.2.3 Accessories................................................................................. 9
TECHNICAL DATA
10
4.1
DIVA-D AS 25....................................................................................... 10
4.2
4.3
4.4
Environmental conditions..................................................................... 18
4.5
Standards............................................................................................. 18
19
UNIT SWITCHING ON
22
6.1
6.2
PATIENT DATA
26
FLUOROSCOPY
28
8.1
MODALITY SELECTION
31
10
GRAPHY
33
(Rev. 0)
DIVA-D
USERS MANUAL
Contents
11
POST PROCESSING
35
12
DICOM FUNCTIONS
43
13
54
ii
(Rev. 0)
USERS MANUAL
Introduction
INTRODUCTION
NOTE:
The present manual is updated for the product it is sold with in order to
grant an adequate reference to use properly and safely the product.
The manual may not reflect changes to the product not impacting
operating modes or safety.
1.1
This icon indicates a NOTE: please read thoroughly the items marked
by this picture.
This icon indicates a WARNING message; the items marked by this icon
refer to the safety aspects of the patient and/or of the operator.
(Rev. 0)
DIVA-D
USERS MANUAL
Safety aspect
SAFETY ASPECT
DIVA-D
(Rev. 0)
USERS MANUAL
Safety aspect
2.1
Warning
This device has not been designed to be used in environments where
vapors, anaestethic mixes flammable with air, or oxygen and nitrous
oxide can be detected.
Avoid pouring water, even accidentally, or other liquids into the device,
as this could cause short-circuits.
Before cleaning the device, please disconnect it from the line voltage.
Wherever necessary, use the fit accessories, such as the leaded aprons,
to protect the patient from radiation exposure.
While performing the radiography, no one, apart from the operator and
the patient, must remain in the room.
Though this unit has been designed with a quite acceptable protection
level from electromagnetic interference, it is advisable to install it at a
certain distance from electrical energy transformation rooms, from static
continuity units, from portable receiving-transmitting units for
amateurial use and for cellular use. Cellular telephones are only
admitted at a distance of more than 1.5 m from any component of the
device.
Other medical instruments and devices that must be used in the same
installation area of the unit must comply with the Electromagnetic
Compatibility rules in force. Non-complying instruments, of which the
poor immunity from electromagnetic fields is well known, must be
installed at least 3 m away from the DIVA-D and supplied by a different
electrical line.
DIVA-D cannot be installed in operating theatres.
DIVA-D must be off while using devices such as electrical scapels or
similar instruments.
The equipment automatically performs some self-test functions and
therefore is able to detect some anomalous conditions. However,
whenever there are doubts on the performance of the equipment (e.g.
images too clear, too dark or with artifacts), the activity must be
suspended and a Villa Sistemi Medicali qualified and authorized engineer
must be contacted.
WARNING:
If during the self-test phase the equipment shows a failure message
caused by the electronic circuits or the Ram disks, the images of a given
patient retrieved for post processing may not belong to the requested
patient.
In this case, immediately contact an authorized Villa Sistemi Medicali
engineer and do not make any diagnosis or evaluation on the stored
images.
In addition, as the failure may have occurred during the last working
session, diagnosis carried out after the last exam cannot be
considered valid.
(Rev. 0)
DIVA-D
USERS MANUAL
Safety aspect
During the use of the equipment, the operator must comply with the
local work safety regulation, in particular with the ones referring to the
use of video terminals.
The cathode X-ray tube (CRT) of the monitors, when present, is
manufactured in order to minimize the risk of implosion. However, if
heavily hit it may break thus projecting out pieces of glass, powder and
toxic gas. Avoid hitting the CRT with any object and make sure that it
does not hit any object.
These risks are not present when LCD monitors are used.
DIVA-D
(Rev. 0)
USERS MANUAL
Safety aspect
2.2
Central unit and control console: iron, copper, aluminum, nonbiodegradable plastic material, glass-resin for printed circuit board
(PCB)
WARNING
Glass vacuum tube may implode with projection of glass pieces, powder
and toxic gas.
(Rev. 0)
DIVA-D
USERS MANUAL
Safety aspect
2.3
Symbols used
Symbol
Description
Warning, read the relevant documentation
The equipment contains in some of its parts,
materials that at the end of the units life, must be
recycled at an approved location
A.C.
Dangerous voltage
DIVA-D
(Rev. 0)
USERS MANUAL
Description
DESCRIPTION
3.1
Identification labels
1
4
2
3
1
Main label (rack)
2
Control keyboard label
3
CRT Monitor label
4
LCD Monitor label
(Rev. 0)
DIVA-D
USERS MANUAL
Description
3.2
Description
DIVA-D is a digital acquisition, storing and processing system for
radiological images. The large storing capacity allows performing the
following functions:
The images coming from the image intensifier are acquired through a
CCD camera, then fed to the digitalization stage (A to D Converter) after
that they can be stored in the permanent memory (Ram disk).
From this moment on the images can be processed thanks to the
dedicated software programs.
The system has a large accessible database, which allows for the entry
and retrieval of patient information along with exam description, remark
and notes.
DIVA-D is available in 2 versions:
DIVA-D RF 12
Ram Disk capacity
36 Gbytes
Operator monitor
1 or 2 on request
0, 1 or 2 on request
Functions
standard acquisition:
imaging subtraction:
road mapping:
fluoro store:
YES
NO
NO
YES
DIVA-D AS 25
Ram Disk capacity
72 Gbytes
Operator monitor
1 or 2 on request
0, 1 or 2 on request
Functions
DIVA-D
standard acquisition:
imaging subtraction:
road mapping:
fluoro store:
YES
YES
YES
YES
(Rev. 0)
USERS MANUAL
Description
3.2.1
DIVA-D composition
DIVA-D contains the following components:
3.2.2
Control keyboard.
CCD TV camera.
Monitors
All DIVA-D models are available with CRT or LCD monitors for operators
and "in room".
The size of these monitors depends on customer request.
3.2.3
Accessories
Depending on the request, DIVA-D can be supplied with accessories:
(Rev. 0)
DIVA-D
USERS MANUAL
Technical data
TECHNICAL DATA
4.1
NOTE:
Technical data, performance and instructions provided in this manual
are referred to the DIVA-D AS 25 model. Therefore, some of these
characteristics are not applicable for DIVA-D RF 12 systems.
DIVA-D AS 25
Identification
Type
DIVA-D AS 25
Model
Components
Central unit
Control system
Acquisition by
Interface system
Connection cables
Monitor
Operator: 1 or 2 on request
"In room": 1 or 2 on request
15 Kg
Height
254 mm
Depth
345 mm
Width
440 mm
6 Kg
Height
50 mm
Depth
235 mm
Width
460 mm
DIVA-D
10
(Rev. 0)
USERS MANUAL
Technical data
2 Kg
Height
95 mm
Diameter
140 mm
about 0,5 Kg
Height
20 mm
Depth
98 mm
Width
200 mm
about 1 Kg
Height
38 mm
Depth
150 mm
Width
235 mm
12 bit
12 bit
66 MHz
Processor management
AMD
1,2 GHz
CCD camera
Control keyboard
TM
Laser printer interface box (when Dicom With parallel output for digital
module not present).
connections.
Communication protocol P831, P859.
Maximum connection distance 500m.
(Rev. 0)
11
DIVA-D
USERS MANUAL
Technical data
Intel
Working frequency
1 GHz
Ram memory
512 Mbytes
Transit memory
100 Mbit
SW package available
-Worklist
Imaging acquisition (speed and matrix)
Mode
Fluoroscopy
Matrix 1024x1024
Acquisition max rate 25 fr/sec
Pulsed fluoroscopy
Matrix 1024x1024
Acquisition selectable rate:
1-3-6-12-25 fr/sec
Matrix 1024x1024
Acquisition max rate 25 fr/sec
Imaging memorization
Storage
30 msec.
Memory capacity
72 Gigabytes
36000 images
Electrical characteristics
Standard power supply (transf 02)
230 V ac 15%
24 V ac 15%
Frequency
50-60 Hz
Current absorbed
650 mA @ 230 V ac
Power absorbed
150 W @ 230 V ac
DIVA-D
12
(Rev. 0)
USERS MANUAL
Technical data
Working configuration
Remote R\ F control table
- Apollo
- Mercury 332
R\ F tilting table
- Vision
Clinical application
Fluoroscopy
Standard exams and with contrast agent Gastroenterological, bone exams, ERCP,
Urology, Mielography, etc.
Tomography
Vascular exams
Working characteristics
User interface
Windows representation
Fluoroscopy mode
Storage
Edge enhancement
In line/off line
Noise reducer
Road mapping
Direct masking
Imaging comparison
(Rev. 0)
13
DIVA-D
USERS MANUAL
Technical data
Radiography mode
Storage
On Ram disk
20
In line/off line
Each exposure
With scrolling
Edge enhancement
Windowing
Auto/manual
Default or manual
Image visualization
Angle measurement
Displayed data
Length measurement
Zoom
Exams review
Image printing
Full screen
DIVA-D
14
(Rev. 0)
USERS MANUAL
Technical data
4.2
CRT Monitor
NOTE:
The characteristics of the monitors are the same for both DIVA-D's
model.
Technical data
Type
Manufacturer
Nical
Model
Refer to optional
Monitor 17
Monitor 21
Weight
17,5 Kg
21,8 Kg
Height
335 (359) mm
394 (418) mm
Depth
350 mm
380 mm
Width
405 mm
489 mm
Screen dimensions
Electrical characteristics
Standard power supply
230 V ac 15%
Frequency
50-60 Hz
Current absorbed
350 mA
Power absorbed
80 W
Scanning
Type
Interlaced
Line number
1049
"Refresh" frequency
60 Hz
Standards
EIA
Connection type
BNC
Impedance
75
(Rev. 0)
15
DIVA-D
USERS MANUAL
Technical data
Display characteristics
Deflection angle
90
Aspect ratio
4:3
Band width
40 MHz 3 dB
Raster
0.25 mm
Phosphor
White P45
Geometric error
<1.5 %
<1.5 %
Maximum brightness
850 cd/m2
Light transmission
46%
Dynamic focus
Yes
Stabilized EHT
Yes
Optional
Contrast and brightness control with digital
push-button on front panel and rear remote
control connection
DIVA-D
16
(Rev. 0)
USERS MANUAL
Technical data
4.3
LCD Monitor
NOTE:
The characteristics of the monitors are the same for both DIVA-D's
model.
Technical data
Type
Manufacturer
Nical
Model
Refer to optionals
Monitor 18
Monitor 20
Weight
5,5 Kg
6,0 Kg
Height
340 mm
355 mm
Depth
60 mm
60 mm
Width
430 mm
457 mm
18"
20"
Screen dimensions
Electrical characteristics
Standard power supply
230 V ac 15%
Frequency
50-60 Hz
Current absorbed
250 mA
Power absorbed
50 W
Display characteristic
Visibility angle
180
Contrast
600:1
Resolution
1280x1024
Gray scale
768 levels
Max luminance
700 Cd/m2
Calibration
automatic
Optional
Contrast and brightness control with digital
push-button on front panel and rear remote
control connection
(Rev. 0)
17
DIVA-D
USERS MANUAL
Technical data
4.4
Environmental conditions
Operative
4.5
Standards
DIVA-D system with accessories and optional complying with the relevant
standards:
EN 60601-1
EN 60601-1-2
EN 60950
0123
DIVA-D
18
(Rev. 0)
USERS MANUAL
Keyboard and functions
(Rev. 0)
19
DIVA-D
USERS MANUAL
Keyboard and functions
5.1
Functions description
DIVA-D
20
(Rev. 0)
USERS MANUAL
Keyboard and functions
(Rev. 0)
21
DIVA-D
USERS MANUAL
Unit switching ON
UNIT SWITCHING ON
Any time that DIVA-D is switched on, it performs an auto test to verify if
the CCD camera is connected and correctly working and if the electronics
of the main rack, including RAM disks are normally operating.
At start up, self test starts and the window is displayed:
No error code
The Unit automatically goes to full operation and the display shows.
SYSTEM READY
Device is ready to operate.
DIVA-D
22
(Rev. 0)
USERS MANUAL
Unit switching ON
SYSTEM READY
DICOM READY
This delay is determined by the DICOM module initialization time
that is longer than the DIVA-D one.
(Rev. 0)
23
DIVA-D
USERS MANUAL
Unit switching ON
6.1
DIVA-D
24
(Rev. 0)
USERS MANUAL
Unit switching ON
6.2
System fault
If auto test recognizes an analogic, digital or disks fault the window
shows:
(Rev. 0)
25
DIVA-D
USERS MANUAL
Patient data
PATIENT DATA
Before starting with a new exam is necessary to insert into DIVA-D the
data of the new patient.
If this is not done, as soon the exposure images acquisition is required,
the message:
is displayed.
It is necessary to press NEW PATNT key to open the patient data
entering window.
NOTE:
If fluoroscopy mode is activated, without storing any LIH or any fluoro
sequence, the images visualization is present also if no patient data is
inserted. This because a patient could perform a fluoroscopic
investigation only. In all other cases, because all radiographic images are
automatically stored on RAM disks, the patient data must be entered into
the data base.
When the key NEW PATNT is activated the new patient window is
displayed:
DIVA-D
26
(Rev. 0)
USERS MANUAL
Patient data
NOTE:
If the birth date is not entered, the patient age in the data base will be
recorded as 0.
If sex is not inserted, O (other) will be present in the data base).
Confirming by ENTER, the data are stored into the data base and the
modality window is displayed to permit the acquisition mode selection.
If "last patient" is selected, the PATIENT DATA window appears with the
data relevant to the last patient stored.
This selection is used when during exam on a patient, the unit has been
switched off.
This feature is very useful to avoid 2 files for the same patient.
(Rev. 0)
27
DIVA-D
USERS MANUAL
Fluoroscopy
FLUOROSCOPY
If DIVA-D is equipped with 2 monitors in the control desk area, one of
them is reserved to fluoroscopy images only, otherwise the fluoro window
is present any time the fluoro command is activated.
Status
The visualization could be X-ray during acquisition or L.I.H. (Last
Image Hold) in off line.
Reg. of int
The visualization is modified according to selection.
ROI (Region Of Interest) must be modified in function of position and
dimension of the organ. 6 different ROI are available trough
correspondent keys. Maintaining pressed the key relevant to the ROI
selected for 3 sec, the ROI dimension and position is displayed on the
image.
Modality
This field could be STD or PULSE xx fr/sec.
the fluoro mode can be selected.
By means of the key
When led is off, STD is displayed in the window, meaning that the
fluoro mode selected is continuous.
When led is on PULSE with the frame rate is displayed in the
window. This rate can be modified using keys FR/s + and FR/s -.
DIVA-D
28
(Rev. 0)
USERS MANUAL
Fluoroscopy
Store
In normal use the condition is OFF. To activate the fluoro storing
modality it is necessary to press STORE. The relevant led turns on
and the images number stored are indicated at the bottom of the
display.
When fluoro acquisition is interrupted also the storage is stopped.
At new fluoro acquisition it will be necessary to press the STORE key
again to perform a new storage.
In LIH condition, pressing the STORE key, the LIH will be stored into
the hard disk.
Edge
By default the value is OFF. It means the edge level used is the
default one. Using E key, the visualization is ON and the edge level
selected is the one configured at the installation.
Noise
By default the value is OFF. It means the noise suppression level
used is the default one. Using N key, the visualization is ON and the
noise level selected is the one configured at the installation.
Road map
Road mapping function is available on DIVA-D AS25 only.
The function is activated by using the ROAD MAP key. Pressing the
key, the relevant led turns on and the fluoro images will be in
subtractions mode.
Releasing fluoro, on the display the LIH will be in sub mode and
following fluoro images will have the mask overlapped.
To reset the road mapping mode and the mask, press the key with
fluoro off.
The Led will be turned off and next fluoro will be performed in
standard mode.
Kv mA Dose
In these fields the actual kV and mA are displayed.
Dose field is not yet activated.
Run Frame
In these fields the fluoro run number stored for the present patient
and the images contained in the run are displayed.
The dose level can be modified by using the keys kV+ and kV-.
The effect of this action is the image contrast level becoming higher or
lower depending by the key activated. 3 decreasing steps and 3
increasing steps are available starting from the standard level.
The radiological parameters are automatically adjusted by the system,
during fluoro acquisition to optimize the image video level.
This adjustment is calculated based on ROI selection.
(Rev. 0)
29
DIVA-D
USERS MANUAL
Fluoroscopy
8.1
DIVA-D
Store XX Fr/s
The storage rate in fluoro can be selected to preserve disk space.
This selection modifies the storage speed only without modifying the
visualization rate.
Standard / Pulsed
This selection can also be done in on line mode by using the pulsed
key as described above.
Ref image
This selection appears when DIVA-D is equipped with 2 monitors in
control desk area only.
When this selection is active the graphy monitor is used to display
the LIH while on fluoro monitor the live images are displayed.
Pattern 1 / 2
Allows to display 2 different patterns in order to adjust correctly the
monitor brightness and contrast.
30
(Rev. 0)
USERS MANUAL
Modality selection
MODALITY SELECTION
When patient data are inserted, the modality and relevant anatomic
programs must be chosen.
The access to this function is automatic when exiting from patient data
window with ENTER key or any time the MODE key is activated.
This because when a new patient is inserted it is necessary to select the
modality for the first acquisition, and to change this modality if a patient
requires different exams type.
The acquisition modality can be:
SPOT
BONE
ANGIO
PERIF
STEP
TOMO.
SPOT, BONE and TOMO are allowed for both DIVA-D models RF and AS,
ANGIO and PERIF on AS only, STEP could be present on a AS model only
as supplementary option.
The step modality is allowed only if the system is based on APOLLO table
only.
The modality window looks as follows:
It is divided in 3 areas:
modality selection
anatomic selection
parameters modification.
By default the left area is active when modality is selected. To inform
about the area selected, the background is light gray.
(Rev. 0)
31
DIVA-D
USERS MANUAL
Modality selection
NOTE:
The not available modality (depending by the DIVA-D license level) will be
displayed in gray and not reachable by selection.
BONE
ANGIO
PERIF
STEP
TOMO
ROI
ROI
ROI
ROI
ROI
ROI
Frame rate
Edge
Frame rate
Frame rate
Frame rate
Tomo frames
Edge
Low dose
Edge
Edge
Edge
Edge
Low dose
Neg Pos
Low dose
Low dose
Low dose
Low dose
Neg Pos
Neg Pos
Neg Pos
Neg Pos
Auto mask
Auto mask
Auto mask
Pulsed fluoro
Sub cont
Sub cont
Neg -Pos
Sub cont
Kv reduction
DIVA-D
32
(Rev. 0)
USERS MANUAL
Graphy
10
GRAPHY
When the modality selection is done and confirmed, DIVA-D is ready for
graphy acquisition.
The window displayed is:
(Rev. 0)
33
DIVA-D
USERS MANUAL
Graphy
DIVA-D
34
(Rev. 0)
USERS MANUAL
Post processing
11
POST PROCESSING
To review the exams contained for the current patient, 2 options are
possible:
by MENU key
by STOP key.
Using STOP the Post Processing window is immediately displayed,
selecting MENU the patient selection window will appears.
(Rev. 0)
35
DIVA-D
USERS MANUAL
Post processing
The window reporting the data relevant to the different runs of the
patient is displayed.
Select the desired run using UP and DOWN keys.
Press ENTER to confirm.
The post processing window appears:
DIVA-D
36
(Rev. 0)
USERS MANUAL
Post processing
All of them and other settings can be modified using the dedicated keys:
(Rev. 0)
37
DIVA-D
USERS MANUAL
Post processing
DIVA-D
38
(Rev. 0)
USERS MANUAL
Post processing
11.1
Special function
When the post processing modality is activated by means of the SPECIAL
FUNCTION key the lower part of the window:
(Rev. 0)
Overwrite
Allows to report comments on the image. To position the cursor on
the required place use UP, DOWN, LEFT and RIGHT arrow keys.
Press ENTER key to confirm when the text insertion is completed.
Changing the image or using STOP key, the text inserted is cancelled.
Angle measure
Allows to take a measure of the angle of a part of image.
Using UP and DOWN keys select the approx level from the table top
on which the part to be measured is located between the 3 levels
proposed 0, 10 or 20 cm. Press ENTER to confirm.
To position the cursor on the required place use UP, DOWN, LEFT
and RIGHT arrow keys.
Press ENTER key to confirm the starting measuring point. Move the
cursor on the angle vertex and confirm whit ENTER. Move the cursor
39
DIVA-D
USERS MANUAL
Post processing
WARNING:
Because of different reasons ( II deformation, error in the level organ
position and others) the angle value obtained are only indicative. Do not
use this value for diagnostic purpose.
Length measure
Allows to take a measure of a part of the image.
Using UP and DOWN keys select the approx level from the table top
on which the part to be measured is located between the 3 levels
proposed 0, 10 or 20 cm. Press ENTER to confirm.
To position the cursor on the required place use UP, DOWN, LEFT
and RIGHT arrow keys.
Press ENTER key to confirm the starting measuring point. Move the
cursor on the final measuring point and confirm with ENTER, the
length value is displayed.
Changing the image or using STOP key, the text inserted is cancelled.
More than one measure can be done on the same image. To do that
repeat the above described steps.
Ratio between 2 measure can be done selecting the relevant window.
WARNING:
Because of different reasons (II deformation, error in the level organ
position and others) the measures obtained are only indicative. Do not
use this value for diagnostic purpose.
DIVA-D
Test pattern
Allow to display a test pattern to adjust correctly the monitor
brightness and contrast.
Starting with any modality the test pattern is reset from monitor.
40
(Rev. 0)
USERS MANUAL
Post processing
11.2
Multi image
When the patient menu window is displayed, multi image function can be
used.
It is a short and easy way how to select the run to be review.
Using the
key, on monitor the first image of any run present into
the selected patient is displayed.
By means of the UP, DOWN, LEFT and RIGHT arrow keys, select the
desired run and press ENTER to confirm.
The run selected will be displayed in full screen mode.
(Rev. 0)
41
DIVA-D
USERS MANUAL
Post processing
11.3
Image print
If DIVA-D is not equipped with DICOM module, it is possible to send the
image to a laser printer directly.
The image present on display is also present on the printer output
channel.
To capture this image it is necessary to confirm from the printer
keyboard supplied by the printer manufacturer.
DIVA-D
NOTE:
Do not change the image displayed until the laser printer acquisition is
completed.
42
(Rev. 0)
USERS MANUAL
DICOM functions
12
DICOM FUNCTIONS
If the DIVA-D is equipped with DICOM module, the DICOM key is active.
By means of this key the following window is displayed:
(Rev. 0)
43
DIVA-D
USERS MANUAL
DICOM functions
12.1
Waiting list
This function is allowed if the network is equipped with HIS/RIS.
When new patient window is displayed (chapter 7) the waiting list
selection is allowed.
Confirm by ENTER, the window
is displayed.
On this window the patients present on the worklist are displayed.
By means of the UP and DOWN arrow key select the desired one, press
ENTER to confirm.
The patient data window will be displayed (chapter 7) reporting all data
inserted in HIS/RIS.
By this function any manual patient data entering is avoided.
DIVA-D
44
(Rev. 0)
USERS MANUAL
DICOM functions
12.2
will be displayed.
If no queue error or messages are present, the last used function will be
proposed, anyway the queue view will be reachable by means of ENTER
key directly.
In any case UP and DOWN arrow keys can be used to reach the function
desired or queue view.
If errors or messages are present in the queue, the relevant field is with
white background; pressing ENTER key the window
will appears.
(Rev. 0)
45
DIVA-D
USERS MANUAL
DICOM functions
DIVA-D
46
(Rev. 0)
USERS MANUAL
DICOM functions
12.3
Print
When DICOM window is selected, the last function used in DICOM mode
is proposed.
By means of UP and DOWN arrow keys select the required function and
confirm with ENTER key.
If print function is selected the printer window selection is showed.
With UP and DOWN arrow keys, select the required printer and confirm.
With LEFT and RIGHT arrow key select the film subdivision and confirm.
The DICOM print window
is displayed.
Select the image required and modify, if necessary the characteristic as
in standard post processing mode.
In the lower section of the window, the frame number and position are
displayed.
(Rev. 0)
47
DIVA-D
USERS MANUAL
DICOM functions
is displayed.
Pressing once more the addition is confirmed and the image stored into
DICOM print temporary archive.
The previous window will be showed allowing a new image addition.
When the above window is present, 3 other functions are available (use
UP and DOWN arrow keys to select and press ENTER to confirm):
Study print
To print the image stored.
To print the study it is also possible to press 2 times the
starting from the main DICOM print window.
DIVA-D
key
Change frame
To store the present image instead of an image already present into
the DICOM print archive.
To do that use LEFT and RIGHT arrow keys, the white portion of the
film will be moved on the film representation showing the place on
which the present image will be located.
Press ENTER to confirm, the previous image will be erased and the
new one stored.
48
(Rev. 0)
USERS MANUAL
DICOM functions
12.4
Store
By means of UP and DOWN arrow keys select the store function and
confirm with ENTER key.
When store function is selected the window is showed:
On the lower part of the window the following information are listed:
(Rev. 0)
Total Mbytes
Mbytes used by the images stored in the present DICOM store study.
This is referred to the temporary DICOM archive.
Total Patient
This value can be 0 or 1.
As matter of fact a DICOM study to be stored cannot contain more
than 1 patient.
Total runs
Is the number of runs stored to be transferred in the network.
Total frames
Is the number of single image to be transferred in the network.
Total sub-runs
Is the number of sub-runs stored to be transferred in the network.
49
DIVA-D
USERS MANUAL
DICOM functions
Study send
To send the study stored.
To send the study it is also possible to press 2 times the
starting from the main DICOM store window.
DIVA-D
50
key
(Rev. 0)
USERS MANUAL
DICOM functions
12.5
CD - ROM
By means of UP and DOWN arrow keys select the CD-ROM function and
confirm with ENTER key.
When CD-ROM function is selected the media type window is showed:
(Rev. 0)
51
DIVA-D
USERS MANUAL
DICOM functions
On the lower part of the window the following information are listed:
DIVA-D
CD free space
The disk percentage space still available.
Total Patient
This value can be 0 or 1.
As matter of fact a DICOM study to be stored cannot contain more
than 1 patient.
Total runs
Is the number of runs stored to be transferred in the network.
Total frames
Is the number of single image to be transferred in the network.
Total sub-runs
Is the number of sub-runs stored to be transferred in the network.
52
(Rev. 0)
USERS MANUAL
DICOM functions
To store into temporary DICOM archive, select the patient, the run or the
image required and press ENTER. The window:
is displayed.
By means of UP and DOWN arrow keys select if the memory addition will
be of the showed frame only, of the run or of the total patient.
Press ENTER to confirm.
The previous window will be showed allowing a new addition.
When the above window is present, 2 other functions are available (use
UP and DOWN arrow keys to select and press ENTER to confirm):
Burn disk
To send the study stored to CD-ROM.
Before to select "burn disk", put a new CD-ROM into driver and press
key.
The dicomization procedure will start and at the end of this operation the
CD-ROM burning start.
This sequence is displayed on the window with a waiting notice.
(Rev. 0)
NOTE:
If no CD-ROM or not valid one is present into the driver, an error will be
present in the queue.
DIVA-D will try again to access the CD after 30 sec.
Insert a valid CD-ROM into the driver before a new writing attempt.
53
DIVA-D
USERS MANUAL
Maintenance and cleaning
13
WARNING:
Before proceeding with the cleaning of the equipment, disconnect it
from the mains.
13.1
DIVA-D
Make sure that water and any other liquids are not spilled inside the
equipment as this may cause corrosion and short circuits.
Description
1 month
1 months
6 months
6 year
54
(Rev. 0)
Cod. 6995001003_Rev.0
0123