TLC Scanner4
TLC Scanner4
TLC SCANNER 4
Quantitative evaluation
supports the following calibration modes: single level calibration, multi level calibration with linear or nonlinear regression using internal or external standards. Statistics are
available as relative standard deviation (cv) or confidence
interval (ci).
Sub-component evaluation
(included in the option Quantitative Evaluation) can be used
to quantify unknown peaks by relating them to the main component as is prescribed by European or US pharmacopoeias
(Related Compounds).
Dual-wavelength scan
The chromatogram is scanned at two individually selectable wavelengths. During integration the signal from the second wavelength
is subtracted from that of the first wavelength to eliminate matrix
effects. Dual-wavelength scanning is also useful for the quantitation of incompletely resolved peaks.
Multi-wavelength scan
The chromatogram can be scanned automatically at up to
36 different wavelengths. This permits multiple measurements between 190 and 900 nm in order to achieve optimum selectivity.
Track optimization
Each track of a chromatogram is scanned several times with small
lateral offset. From this data the optimum track following the peak
maxima is calculated and used for quantitation. Thereby distorted
chromatograms can be correctly evaluated.
Spectrum library
Enables the user to create his own library files, which provides
the only way the user can save and compare spectra recorded on
different plates.
Note
The TLC Scanner 4 with winCATS is compliant with the requirements
of GMP/GLP and can be IQ/OQ qualified.
If you want to use the instrument in a 21 CFR Part 11 environment,
the option 21 CFR Part 11 compliance ready is required for each
winCATS workstation.
2|3
Planar chromatogram
Analog curve
Calibration function
User-friendly operation
All functions of the scanner are controlled by winCATS software.
The electronic amplification is set automatically. In fluorescence mode this is performed by a
quick scan, which measures the complete plate, e.g. 18 tracks in 20 seconds, and determines
the optimized settings from the result. In absorbance mode only one track is quick scanned;
either track 1 (automatic mode) or a track of choice (manual mode).
The 16 bit A/D converter ensures optimum resolution of the signal. This is particularly useful
for the measurement of both low absorbance and weak fluorescence signals.
Scanner operation with winCATS software is convenient and easy to learn.
4|5
Lamp selector
Entrance lens
system
Monochromator
entry slit
Monochromator grating
Any of the three light sources, high pressure mercury lamp, deuterium lamp, or halogen-tungsten lamp can be positioned in the light
path by a motor drive.
The signal of the measuring photomultiplier is continuously offset
against the signal of the reference photomultiplier.
This compensates for lamp aging and short-time fluctuations. It also
reduces the warm-up time required to reach lamp stabilization.
All components of the optical system, lamps, monochromator, scanning stage, and photomultiplier are mounted on one sturdy metal
support. This ensures high precision of the detector signal.
Mirror
Light sources
The light beam strikes the object at right angle. The photomultiplier
is aligned at an angle of 30.
Optical system
Monochromator
Mirror
Reference
photomultiplier
Beam splitter
Measuring
photomultiplier
Scanning object
Secondary filter
Mains voltage
115 V and 230 V selectable; 50/60 Hz; maximum energy use 180 W
(tungsten and mercury lamp ignited).
A/D converter
Scanning slit
Revolving disk with 20 fixed apertures; length of slit images selectable between 0.2 and 12 mm, width between 0.1 and 1.2 mm in
42 combinations.
Connections/interfaces
Detector
Two matched broad band photo multipliers, multi alkali type, spectral
sensitivity 185 900 nm.
Dimensions
Width = 590 mm, depth = 650 mm, height = 367 mm; net weight 39 kg.
Stage drive
Independent in both directions by stepper motors, micro step driven for smooth movement; reproducibility of positioning better than
50 m in Y-direction, better than 100 m in X-direction; maximum
scanning speed 100 mm/s, positioning at 150 mm/s.
6|7
Manual Input
Stationary Phase
Sample
Preparation
Software Controlled
Nanomat
Horizontal
Developing Chamber
Immersion Device
Developing Chamber
Spray
Definition
Sample
Application
Chromatogram
Development
Samples
Automatic TLC
Sampler ATS 4
Automated Multiple
Development AMD 2
Standards
Linomat 5
Automatic Developing
Chamber ADC 2
Derivatization
winCATS SOFTWARE
winCATS for all steps of instrumental
Thin-Layer Chromatography
winCATS features
Image Import
Image Annotation
Detection
Evaluation
TLC Scanner
Quantitative
Evaluation
Scanner Selftest
Dual-Wavelength Scan
Documentation
Report
TLC Visualizer
21 CFR Part 11
(FDA Regulation)
Image
Enhancement
Spectrum
Library
Image
Comparison
Multi-Wavelength Scan
Track Optimization
Modular design
Workflow
Using the explorer type view the operator at all times remains
in control over all active steps of the workflow. These are:
Sample application
Chromatogram development
Detection
Spectra recording
Quantitative evaluation
Documentation
page 12
page 14
page 14
page 15
Multi-wavelength scan
Scanner selftest
21 CFR Part 11 compliance ready
page 15
page 16
page 16
8|9
Measurement
For routine analysis winCATS starts off with a method file containing all relevant data for the current task such as instruments,
parameters, etc. This method can be validated for GxP work.
The TLC Scanner 4 scans the plate and transfers raw data back
to winCATS where baseline correction and peak recognition is
performed automatically in the background.
A display of integrated peaks with substance names. This display mode enables manual baseline correction and peak integration by mouse-click.
Spectra list
Evaluation
Recording of spectra
Spectra can be measured from 190 to 900 nm. If the emission range of
the deuterium lamp is exceeded, the scanner automatically switches
to the halogen-tungsten lamp. Both lamps remain ignited.
10 | 11
QUANTITATIVE EVALUATION
Quantitative evaluation is based on the comparison of peak
heights or peak areas of the unknowns with those of calibration standards chromatographed on the same plate.
According to the European and the US Pharmacopoeia, related compounds can be quantified by comparison with small amounts of the
main component in case their identification is not required. This main-/
subcomponent evaluation is part of the standard quantitative evaluation program.
Calibration modes
Single-level calibration is suitable for analyses where the concentration of the unknowns shall be checked within narrow limits. Singlelevel calibration requires reduced calibration efforts.
Multi-level calibration is used when the target values are expected in a comparatively wide range. winCATS offers a choice of four
calibration modes , the selection is made based on which mode
gave the best result during method validation.
winCATS offers a wide choice for data input: from entering amount
per fraction to entering amounts from balance readings or from
stock solutions.
Ordering information
027.6315 Quantitative chromatogram evaluation
Application example
Analysis of 25 water-soluble food dyes in food
HPTLC is the most effective, rapid and cost-efficient solution for
the analysis of water-soluble food dyes in food. Up to 20 runs
can be chromatographed under identical conditions (by use of
the Automatic Developing Chamber ADC2) within 12 minutes
(see CBS 105 or www.camag.com).
Calibration function of the red food dye E122 (x) and determination in
fruit drink, energy drink and bakery ink formulation (+).
Calibration function of the blue food dye E131 (x) and determination
in the yoghurt sample (+).
12 | 13
SPECTRUM LIBRARY
TRACK OPTIMIZATION
In order to compare spectra of substances chromatographed on different plates or to compare spectra with those of a spectra collection,
the software option spectrum library is required.
The spectrum library can be used during substance assignment for
validation of assigned substances as well as for identification of
unknown fractions.
During the search process for identification the spectrum library shows
a hit list of the closest substances including their spectra.
Ordering information
027.6342 Spectrum library
Each track is scanned several times with small lateral offset. The number of scans and the distance in between is selected according to the
requirements of the particular chromatogram. When all scans of one
track are completed, the software selects the maximum signal for all
peaks. Only those data are then used for the calculation of the result
from the corrected chromatogram.
Note:
When used with chromatograms with proper track alignment, the results
obtained with or without track optimization are about the same
Ordering information
027.6344 Track optimization
Ordering information
027.6346 Dual-wavelength scan
Ordering information
027.6348 Multi-wavelength scan
14 | 15
21 CFR PART 11
COMPLIANCE READY
SCANNER SELFTEST
With this option an automatic selftest of the TLC Scanner 4 and
a report can be generated.
Features :
Stage positioning
Documentation of all activities in History Log/Audit Trial for secure tracking of electronic signatures according to 21 CFR Part 11
requirements
Ordering information
027.6340 Scanner selftest
Ordering information
027.6380 21 CFR Part 11 compliance ready
INSTRUMENT QUALIFICATION
INSTRUMENT SERVICE
For all CAMAG instruments for which we offer IQ/OQ qualification we also offer service contracts.
16 | 17
TECHNICAL DATA
Light sources
Scanning slit
Revolving disk with 20 fixed apertures; length of slit images selectable between 0.2 and 12 mm, width between 0.1 and 1.2 mm in
42 combinations.
Detector
Two matched broad band photo multipliers, multi alkali type, spectral
sensitivity 185 900 nm.
The slit is automatically illuminated with visible light when the compartment illumination is switched on. The scanning compartment is
illuminated with a 4 watt fluorescent tube UV 254 nm which the user
can replace by a UV 366 nm or a white light tube.
Optical system
Stage drive
Independent in both directions by stepper motors, micro step driven
for smooth movement; reproducibility of positioning better than 50
m in Y-direction, better than 100 m in X-direction; maximum scanning speed 100 mm/s, positioning at 150 mm/s.
115 V and 230 V selectable; 50/60 Hz; maximum energy use 180 W
(tungsten and mercury lamp ignited).
Monochromator
A/D converter
Secondary filter
Motor-driven filter wheel with three automatically selected filters for
the elimination of second order wavelengths; 400 nm cut-off filter for
fluorescence measurements; three positions for user selected filters.
Mains voltage
Connections/interfaces
Serial interface RS232 for communication to the computer, EquiLink
for connection to winCATS software.
Dimensions
Width = 590 mm, depth = 650 mm, height = 367 mm; net weight 39 kg.
ORDERING INFORMATION
027.6200 CAMAG TLC SCANNER 4
for scanning by absorbance and fluorescence, equipped for objects up to 200 200 mm, wavelength range
190 900 nm, complete with deuterium lamp, tungstenhalogen lamp, and mercury vapor lamp, including Equilink
(027.6312) to winCATS, but without winCATS license.
18 | 19
CAMAG (Germany)
Bismarckstrasse 2729
D-12169 Berlin
Phone +49 30 516 555 0
Fax
+49 30 795 70 73
E-Mail [email protected]
www.camag.com
CAMAG (Switzerland)
Sonnenmattstrasse 11
CH-4132 Muttenz 1
Phone +41 61 467 34 34
Fax
+41 61 461 07 02
E-Mail [email protected]