PDC Bits
PDC Bits
a. A minimum offset
b. high reset
c. low sensitivity
d. narrow proportional band
11. A controller with settings PB = 110%, Reset = 1.5 minutes is the same controller
with setting of:
12. A controller with settings PB = 180%, Reset = 2.3 min is subjected to a 20% step
change at time t = 0, after 2 min the controller output change in percent will be?
(consider open loop only)
a. 20.8%
b. 1.45%
c. 11.6%
d. 67.3%
13. An error signal that can be represented by the function e = At . psi/min where A
= 1.4, n = 1.89 and t = time in mins., is transmitted to a direct acting controller having
a PB = 24%, and an integral setting of 1.6 repeats/min. If the system is in an open
loop situation, what is the output from the controller after 1.81 min when the initial
value of the output was 7.3 psi
a. -28.54 psi
b. 52.64 psi
c. 43.14 psi
d. 77.05 psi
14. An error signal that can be represented by the fuction e = At psi/min where A =
1.6, n = 1.62 and t = time in mins, is transmitted to a reverse acting controller having a
PB = 350%, and an integral setting of 1.2 min/repeat. If the system is in an open loop
situation, what is the output from the controller after 1.36 min when the initial value
of the output was 10 psi
a. 8.395 psi
b. 8.922 psi
c. 8.490 psi
d. 11.28 psi
n
15. An error signal that can be represented by the function e = At psi/min where A
= 1.88 and t = time in mins., is transmitted to a reverse acting controller having a PB =
440%, and an integral setting of 1.2 min/repeat. If the system is in an loop situation,
what is the output from the controller after 0.729 min when the initial value of the
output was 8.0 psi
a. 3.695 psi
b. 7.742 psi
c. 9.354 psi
d. insufficient data
n
16. An error signal that can be represented by the function e = At psi/min where A
= 1.8, and t = time in mins., is transmitted to a reverse acting controller having a PB =
410%, and an integral setting of 0.74 min/repeat. If the system is in as open loop
situation, what is the output from the controller after 0.798 min when the initial value
of the output was 8.0 psi
a. 7.272 psi
b. 8.462 psi
c. 7.443 psi
d. 7.451 psi
17. The main purpose for reset action in a control loop is to:
a. increase controller gain
b. eliminate offset
21. A proportional controller will have an offset difference between set point and
control point:
a. that depends on process load
b. at all times
22. If it were possible for a proportional controller to have true 0 % proportional band,
the controller gain would have to be:
a. 0
b. unity
c. 100
d. infinite
23. If the proportional band of the controller is adjusted to minimum possible value,
the control action is likely to be:
a. on/off
b. with maximum off/set
c. excelent
d. inoperative
24. All process control systems may be assumed to be:
a. linear
b. nonlinear
c. linear for the purpose of initial consideration but with the full knowledge that
this may not be the case.
d. nonlinear for the purpose of analysis with the exception that the system may
prove to be linear
29. If the closed loop control system has too much gain it will cycle. The only loop
component that has a conveniently adjustable gain is the:
a. valve operator
b. measuring transmitter
c. process
d. controller
30. A P & I controller with an error of 9%, a gain of 2.3 and a reset of 0.94 min/repeat
in open loop would produce an output signal that would ramp at the rate of :
a. 42.7%/min
b. 19.5%/min
c. 20.7%/min
d. 22.0%/min
d. highly fluctuating
33. A stable system is
a. for which the output response is bounded for all bounded input
b. which exhibits an unbounded reponse to a bounded input
c. which satisfies the conditions of a dynamic system
d. which can be easily controlled
a. 2
b. 3
c. 0
d. infinte
36. The lapse time between the moment of the step change is made and the moment
the recorder first begins to reflect that change is called :
a. process rate
b. integral time
c. dead time
d. error
37. During the start-up of the hypochlorate plant you have the opportunity to tune the
three mode level controller installed on the inlet separator. Using the Ultimate
Sensitivity Method you establish that the system oscillates with aperiod of 1.26 min.
The proportional setting that creates the oscillations is 190 %. The optimum settings
for good control ( One quarter amplitude decay QAD ) using the Ziegler and Nichols
method will be :
f.
g.
h.
i.
j.
k.
l.
m.
n.
2. L-1
(a) sin t
is
(b) cos t
(c) sin t
(d) cos t
o.
p.
3.L-1 A is
q.
r.
(a) A
u(t) (e) A-1 u(t)
s.
t. 4. L 1 F (s-a) is
u. a) e at f (t) u(t)
at
b) e
-at
f(t) u(t),
c) e
(d) A t
at
f (t-a) u(t),
d)
f (t a)
v.
w.
5.
is
x.
y.
z.
6. ex is
aa.
ab.
ac.
ad.
ae.
7. sin x is
af.
ag.
ah.
ai.
aj.
8.
ak.
al.
am.
an.
ao. 9. Tick
ap. System
of these
aq.
coefficient
ar. OPTION
d
Characteristics
Both
Time
Characteristics
a
None
a and b
b
constant
Damping
c
as.
i) Zero Order
au. ii) First Order
av. iii) Second Order
aw. iv) Third Order
at.
ax.
ay.
az.
ba.
bb.
bc.
bd.
be.
bf.
bg.
bh.
bi.
bj.
bk.
bl.
bm.
bn.
bo.
bp.
bq.
br.
bs.
a) exponentially
increasing,
b) exponentially
decreasing,
c) sinusoidal,
d) an
impulse
function,
none of these,
13. Amplitude ratio for a 1st or 2nd order system is
a) 1
b) > 1,
c) < 1
d) none of
these
bt.
bu.
bv.
bw.
bx.
by.
bz.
ca.
cb.
cc.
cd.
ce.
cf.
cg.
ch.
ci.
cj.
ck.
cl.
cm.
cn.
co.
cp.
ds
18. e
is the transfer function due to
a) Process lag
b) transfer lag
c) transportation lag
d) exponential lag
e) quadratic lag
cq.
cr. 19. The nature of the solution for the roots of the
characteristic equation lying on the left-hand half of the
complex plane would be approximated as
cs.
e at ( C1 Cos t + C2 Sin t )
cu. e at ( C1 Cos t + C2 Sin t )
cv. C1 Cos t + C2 Sin t
cw. d) none of these
ct.
cx.
cy.
cz.
da.
db.
dc.
dd.
de.
df.
dg.
dh.
dn.
do.
dp.
dq.
dr.
a) Proportional
b) Proportional derivative
c) Proportional integral
d) Proportional
integral derivative
ds.
dy.
dz.
ea.
eb.
ec.
ed.
ee.
ef.
eg.
eh.
ei.
ej.
ek.
eo.
ep.
eq.
er.
es.
et.
eu.
d) none of these
ex.
ey.
ez.
fa.
fb.
fc.
fd.
fe.
ff.
fg.
fh.
fi.
fj.
fk.
fl.
fm.
fn.
fo.
fp.
fq.
fr.
fs.
ft.
fu.
fv.
fw.
fx.
fy.
fz.
b) load to be constant
d) neither a nor b
ga.
gb.
gc.
gd.
ge.
gf.
gg.
gh.
gi.
gj.
gk.
a) large
b) small
c) variable from small to large values d) none of these
gl.
gm. 41.
gn.
go.
gp.
gq.
gr.
gs.
gt.
gu.
gv.
gw.
gx.
gy.
gz.
ha.
hb.
hc.
hd.
proportional
he.
hf.
hg.
hh.
hi.
a) 1/6 Cfm/psi
c) 24 Cfm
b) 6 psi/Cfm
d) none of these,
hj.
hk.
hl.
hm. a)
hn.
ho.
hp.
b)
c)
d)
hq.
hr.
hs.
ht.
hu.
hv.
hw.
hx.
hy.
hz.
a) deviation
c) error ratio
b) derivative time
d) differential gap
ia.
ib.
ic.
id.
ie.
a) Error ratio
c) proportional sensitivity,
b) derivative time
d) gain
if.
ig.
ih.
ii.
ij.
a) off set
c) deviation
b) error ratio
d) static error
ik.
il.
im.
in.
io.
a) time constant
c) rise time
b) settling time
d) derivative time
ip.
b) integral action
c) derivative action
d) none of these
b) derivative action
d) none of these
b) reset rate
d) off set
b) slower,
c) same
b) 20 loge AR
d) either of a or b
b) reciprocal of amplitude
c) radian frequency
d) cyclical frequency
b) - 1800
+ 900
d)
- 900
b) resonant frequency
c) cross-over frequency
d) natural frequency
b) an unstable system
is
a) n
b) n2
c) n (n+1)
d) n (n-1)
e) none of these
f) all of (a) to (d)
70. The time constants for a fractionating column in a process
industry are of the order of
a) Seconds
b) fraction of a second
c) minutes
d) hours
e) days,
f) none of these
b) minutes
c) hours
d) none of these
b) clock
d) energy motor
e) telephone
g) L I C
h) tachometer
mercury-in-glass thermometer.
c) grocers scale
f) T R C
i) rotameter
k)Bourdon gauge
l) Stetheoscope
73. A temperature sensing device of accuracy + 0.5% measures
temperature in the range of 500 10000C. The span of the
instrument and the true value of temperature corresponding to a
reading of 8000C are respectively
a) 5000C and 797.5 to 802.50C b) 10000C and 800 + 0.50C
c) 5000C and 796 to 8040C
1.50C
e) none of these
b) threshold value
c) sensitivity
d) time delay,
b) sensitivity
c)
e) fidelity
Lag
(a)
Speed
of
(b) Fidelity
(c)Reproducibility
Sensitivity
(d)
b) no
b) internal friction
c) external friction
d) backlash
80. Definitions of
appropriate one for
some
errors
are
given
below. Identify
b) measuring elements
signals
attached
to
the
b) Circular recorder
c) Oscilloscope
d) Teleprinter
e) Line printer
f) Paper type
g) Punch card
h) Cassette type
Floppy disc.
85. Dead zone is :
a) Same as time constant,
b) Same as transportation log
c) The maximum change in the measured variable that does
not change the reading of the instrument.
None of these
86. Degree to which an instrument indicates the changes in
measured variable without dynamic error is called
a) Speed of response
b) Fidelity
c)
Reproducibility
characteristics
87. Instrumentation in a process plant offers :
d)
Static
b) Better quality of
d) All a, b and c.
d) None of these
90. Which of the following
measurement accuracy ?
a) Static and dynamic error
factors
does
not
influence
b) Reproducibility
c) Dead zibe
d) None of these
e) All of these
91. Pick out the symbol for Locally mounted instrument in
instrumentation diagram:
93. Representation
diagram is :
of
capillary
tube
in
instrumentation
following
laws
enables
calibration
of
b) Peltier effect
c) Thomson effect
d) Joule-Thomson effect
pyrometer
b) Thermocouple
d) Bimetallic
b) Radiation pyrometer
c) Mercury thermometer
d) Resistance thermometer
and
oxidising
action
on
b)
bimetallic
d) none of these
b) MV
c) V
d) KV
b) resistance thermometer
d)
bimetallic
b) 4000C
c) 12000C
d) 16000C
Platinum +
Nickel +
Copper +
Thermistor
Silicon semiconductor .
109. Which
of
the
following
is
suitable
for measuring the temperature of a red hot moving object ?
a) Thermocouple
c) Radiation pyrometer
b) Thermistor
d) none of these
110. Thermocouples
a) have very slow speed of response
b) cannot be connected to the measuring
remotely located
need cold junction compensation
instrument
200 to 30
5 to 200
114. Full compensation in filled-system thermometers means
(a) Compensation for changes of temperature of the
capillary tube and that within the instrument case
(b) Compensation in temperature for the full length of the
capillary tube, if used
(c) None of these
115. Self-heating in resistance thermometer is minimised when
a) The resistance is immersed in a high thermal conductivity
fluid
b) The current is kept low
c) Several sensore are provided in series
d) None of these
e) all of a, b and c
116. Piezo electric effect is
a) Conversion of a changing pressure into electrical voltage
signal
b) Creation of a voltage by the movement of a coil within a
magnetic field
c) Generation of an emf or current through the use of photoelectric effect.
117. What is the standard of reference in the range of 0
6000C,
a) gas thermometer
b) platinum resitance thermometer
c) platinum/Rhedium 10 Platinum thermocouple
118. Which thermocouple would you choose for use in an
oxidising environment ?
a) copper constantan
b) iron-constantan
c) charomel-alumel,
119. The brightness of a hot object is the basis of temperature
measurement in certain instruments Guess,
a) optical pyrometer
b) radiation pyrometer
c) pyrometric cones
d)
resistance
pyrometers
120. The low expansion and high expansion materials of a
bimetallic strip are respectively
a) Chromel and alumel
c) brass and invar
iii) + 2000C:
a) gas filled thermometer
b) mercury filled thermometer
c) bimetalic thermometer
d) resistance thermometer
iv) + 4500C:
a) liquid filled thermometer
b) gas filled thermometer
c) bimetallic thermometer
d) copper constantan thermocouple
e) iron-constantan thermocouple
f) platinum
Rhodium
10
thermocouple
g) platinum resistane thermometer
h) optical pyrometer
i) chromel alumel thermocouple
platinum
v) + 5500C: a) - i) as above
vi) + 8000C: a) Chromel-alumel thermocouple
b) Chromel-constantan thermocouple
c) Platinum & Rhodium 10 +
thermocouple
d) Optical pyrometer
e) Radiation pyrometer
f) Platinum resistance thermometer
Platinum
vii) + 14000C:
a)
Platinum
/
Rhodium
10
/
Platinum
thermocouple
b) Optical pyrometer
c) Radiation pyrometer
d) Tridium / Rhodium-Iridium thermocouple
viii) + 20000C:
a) Tungston / Rhenium Thermocouple
b) Platinum / Rhodium 10 thermocouple
c) Optical pyrometer
d) Radiation pyrometer
122. Gas analysis is commonly done using
Platinum
b) X-ray diffraction
d)
emission
b) polarograph
d) refractometer
b) surface tension
c) refractive index
d) optical activity
129. Stroboscope is used for the measurement of
a) rpm of a flywheel
b) frequency of light
c) depression of freezing point
d) liquid level under pressure
130. Small depression of freezing point of a liquid solvent on
addition of a solid solute can be accurately measured by
a) Beckman thermometer
b) dilatometer
c) mercury thermometer
d) thermocouple
131. Pitot tube measures
a) average velocity
b) maximum velocity
c) point velocity
d) static pressure
132. Hotwire anemometer is used for the measurement of
a) flow rates of fluids
b) flow rates of granular solids
c) very high temperature
d) thermal conductivity of gases
133. Flow rate through an orifice is
a) proportional to the pressure differential
b) inversely proportional to the pressure differential
c) proportional to the square root of pressure differential
d) 1 / Q
b) concentric orifice
d) rotameter
b) pitot tube
d) segmental orifice meter
b) pitot tube
c) venturi meter
d) orifice meter
b) pitot tube
d) rotameter
b) Venturi meter
c) Magnetic flowmeter
b)
oscillating
piston
type
d) nozzle
pipe dia.
ratio : (0.3D 0.8D).
1D
8D
b) 1
c) 10
ii) Flange tap :
a) Varying with
size of orifice
b)
1
c)
1D
Varying with
size of pipe
1
1D
8D
b) 2
4D
D
8D
b) U-tube manometer
d) Inclined tube manometer
b) 30 psi
c) 40 psi
d) 50 psi
tester
b)
precision
d) diaphrgm gauges
b)
differential
pressure
d) air-trap system
b) liquid purge
d) U tube
e) Ekstrom gauge
169. Determination of the level of a liquid by weighing the tank
with its contents or that of a solid by weighing solids in bins or
on conveyor belts is a method recommended for the materials to
be
a) too corrosives
c) injurious to health
b) too viscous,
d) chemically inert
b) no