An Introduction To Time Waveform Analysis
An Introduction To Time Waveform Analysis
AN INTRODUCTION TO
TIME WAVEFORM ANALYSIS
PRESENTED BY
J. Alex Bradley
with
This spectrum was created from the time waveform on the previous page. Since the time waveform is very sinusoidal the
amplitude of the spectral peak at 1 times the run speed of the shaft is nearly the same as the peak to peak (PK-PK)
amplitude of the time waveform. The spectrum cursor shows a peak of 5.709 mils at 46 hertz or 2760 rpm. The peak marked
on the time waveform on the previous page is 2.9 mils, or about half of the peak-to-peak value shown on the spectrum. A
peak to peak (PK-PK) value of the waveform could be estimated to be about 5.8 mils. The frequency of the waveform was
shown on the previous page to be 45.92 hertz.
Very simple time waveforms have good amplitude correlation with their associated spectra. This is because the Fourier
transform (FFT) can very accurately determine the amplitude and frequency of the components of the waveform as it
converts them to the frequency domain, that is, into a spectrum. Time waveforms with random energy do not convert so
cleanly to the frequency domain.
The fast Fourier transform (FFT) could be considered to be able to rotate each frequency in the time waveform 90 degrees.
The time length of a waveform depends upon the maximum frequency of the spectrum, Fmax, that is to be examined. In
order to filter out unwanted high frequency spectral information, the data acquisition rate MUST be over 2 times the maximum
frequency of the spectrum, Fmax. This is required to perform proper anti-aliasing. Virtually all modern digital spectrum
analyzers sample raw vibration data at a rate equal to 2.56 times the maximum spectral frequency, Fmax, being examined.
This means that the analyzer stores "samples" of the waveform data coming from the transducer at this "RATE". This is
called the "SAMPLING RATE".
The analyzer must "SAMPLE" enough points to perform any necessary Fourier calculations from these points. More
importantly, after the data has been sampled, the maximum spectral frequency, Fmax, CANNOT be increased°
To calculate a 400 line resolution spectrum, 1024 "SAMPLES" of incoming data must be collected. By knowing the
"SAMPLING RATE" and number of "SAMPLES" that must be collected, the time length, in seconds, of the waveform can be
calculated.
This chart clearly shows that low maximum frequencies require long time periods of data collection.
This order based plot shows horizontal, vertical, and axial measurements on both the outboard and inboard turbine bearings.
The major peaks in the radial directions are all found at 1 order of run speed, i.e. 1 X run speed.
A clear, repeatable, waveform can be seen occurring once per shaft revolution, 1 X RPM. The waveform is showing the
acceleration created on the bearing housing by the shaft imbalance. High frequency energy can be seen riding on top of the
dominant 1 X RPM sine wave. The high frequency energy is created by the gearbox that this turbine is driving.
The repeatability of the waveform both in time (with respect to shaft turning speed) and amplitude mean that the vibration
force is tied to the shaft run speed. This is particularly true of waveforms taken in acceleration, as they can be influenced
greatly by high frequency energy bursts, such as impacts. These impacts make the time waveform non-repeatable. The
absence of impacting helps to conclude that the problem is imbalance.
This multi-spectral plot shows inboard and outboard exciter points. At first glance the problem with the exciter might appear to
be looseness. Many harmonics of run speed are present in the radial directions, which often indicates a looseness problem.
This is a good time to examine the time waveform.
These peaks are very repetitive both in amplitude and spacing. It is this repeatable, repetitive pattern that confirms that this
machine is misaligned, and does not have a primary looseness problem.
This is a diagram of an overhung pump that has a worn out bushing in the coupling. The motor turns at about 1750 RPM°
The worn bushing makes the pump-motor system very loose torsionally. As is typical of many looseness problems this
problem has grown progressively worse over time.
MULTI-SPECTRAL PLOT OF LOOSENESS
A multi spectral plot of all 5 motor points is shown above. Inboard and outboard radial measurements as well as an outboard
axial measurement were made. Many harmonics of run speed are visible on all measurement positions. Baseline, or floor,
energy is also very visible.
Patterns that are non-periodic and random do not convert well in the FFT process. It is very difficult to assign specific
frequencies and amplitudes to patterns to waveforms like this one. This difficulty leads to the broad band energy humps in
the spectrum. Broader humps indicate more random energy. Higher humps indicate more impacting in the waveform.
The noise levels remained high even after replacing the motor.
This probably means that the motor is slightly misaligned to the pump. Since the vibration levels are so low in amplitude it is
probably not necessary to immediately correct the misalignment.
All the measurements made on this pump show broad banded humps of energy, along with some low frequency peaks. The
humps of energy are in the 200 to 400 hertz (12000 RPM to 24000 rpm) range. This is about 10 to 20 orders of run speed.
The low frequency peaks should be examined more closely. The peaks might be harmonics of run speed, or they could be
harmonics of some bearing frequency. The full scale range of all the plots is low at .1 inches/second, but quite a bit of energy
is being produced by the broad humps of energy.
This pump spectrum is a good example of the reason a single overall number cannot be used to determine the condition of a
machine.
The top spectrum is from the pump with the bad bearing. It has an overall value of just under .3 inches/second. No particular
peaks are very high. Broad banded energy is dominant.
The bottom spectrum comes from a steam turbine with an imbalance problem. It also has an amplitude of about .3
inches/second. Virtually all of the energy of the spectrum comes from the single peak at 1X RPM. This may be seen by
comparing the amplitude of the single peak in the spectrum to the overall value of the spectrum.
TIME WAVEFORM OVERALL VALUE COMPARISON
The time waveforms from the spectra on the previous page should now be examined to determine which machine has the
worst problem.
The bad pump bearing time waveform is at the top of the page. It shows high levels of impacts. 0neimpact reached almost 5
g's. Most impacts are in the ± 2 g range. The RMS overall value of the waveform is about .80 g's.
The bottom waveform comes from the imbalanced turbine. It shows virtually no impacting. No peaks exceed ± 1 g's. A
sinusoidal shaped pattern is also seen. This is the type of signal pattern that can be well transformed into a spectral value.
The RMS value of the waveform is about .23 g's. Even though the pump waveform RMS value is more than 3.3 times larger
than the turbine waveform, the ease of transform of the turbine waveform ends up giving the turbine a higher spectral value.
TIME WAVEFORM COMPARISONS
Another reason that the pump waveforms might be examined could be to determine which of the pump bearings had
degraded the most° Higher impacts would be indicative of a bearing with more and larger defects.
The three time waveforms on this page are from the inboard pump bearing housing. The RMS values of the three waveforms
are found at the upper right of the plot.
Once again it is seen that the axial point is the most energetic point.
The RMS levels for the outboard points are consistently lower than the levels for the inboard points. More importantly the
level of the spikes or impacts is much higher on the inboard points. This can be seen by examining the scale of the plots.
In this manner it can be determined that the inboard bearing is likely to be the one with the most severe defects. If only one
bearing were to be changed it should be the inboard one.
All contents copyright © 1998 - 2006, Computational Systems, Inc.
All Rights Reserved.