SSPC PA 2 Measurement DFT
SSPC PA 2 Measurement DFT
--`,,,,,,`,```,,,`,,````,`````,-`-`,,`,,`,`,,`---
DISCLAIMER
This standard has been developed
in accordance with voluntary consensus proceduresby SSPC Advisory Committeesand
is believed to represent good current practice.
All SSPC specifications, guides and recommendations are monitored
and revised
as practices improve,
and suggestions for revision are welcome.
Other methods, materials, and specifications may
be equally
effective or superior. SSPC is not responsible for the application,
interpretation, or administrationof its specifications, guides and
recommendations. Moreover,SSPC does not issue interpretations of its specifications, guides or recommendations; and no
person is authorized to issuean interpretation ofanSSPC
specification,guideorrecommendationonbehalf
of SSPC.
SSPC specifically disclaims
responsibilityforthe useor misuse of
its specifications, guides and recommendations. The supplying of
details aboutpatented formulations, treatments, or processesis
not to be regarded as conveying any right permitting
or
the user
of this guideto use or sell any patented invention. When
it is
known
that the subject matter of the text is covered by patent, such
patents are reflected in the text.
--`,,,,,,`,```,,,`,,````,`````,-`-`,,`,,`,`,,`---
Copyright 1997 by
SSPC
The Society for Protective Coatings
(fomlerly Steel Structures Painting Council)
40 24th St., 6th Floor
Pittsburgh, PA 15222
Printed in U.S.A.
SSPC-PA 2
June 1,1996
SSPC
--`,,,,,,`,```,,,`,,````,`````,-`-`,,`,,`,`,,`---
3. ReferenceStandards
3.1 The documents andstandards referenced in this standard are listed in Section 3.4 and form a part of this standard.
3.2 The latest issue, revision or amendment of the referenced documents in effect on the date of invitation to bid shall
govern unless otherwise specified.
2.2.2 Type l-Pull-Off Gages: In pull-off gages, a permanent magnet is brought into direct contact with the coated
surface and a calibrated scale measures the force necessaryto
pull the magnet from the surface. Less force is required to
remove the magnet from a thick coating. The scale is nonlinear.
4.1 GENERAL
4.1.1 Access to Bare Substrate: All gages are affected to
some degree by
substrate conditions such as roughness,shape,
thickness and composition. To correct for this effect, access to
the uncoated substrate is recommended. Another option is to
use separate uncoated reference panels with similar roughness, shape, thickness and composition. (See Sections 8.3 to
8.9.) These would be used as the bare substrate in the procedures of Sections 4.2 and 4.3. Reference panels shall be of
sufficient size to preclude edge effects. (See Section 8.7.)
Measurements on the bare substrate can be taken before
June 1, 1996
SSPC-PA 2
June 1.1996
5.1.4 If the dry film thickness for any 1O m2 (1O0 ft2) area
(see Sections 5.1.2 and 5.1.3) is not in compliance with the
requirements of Sections 5.2.1 and 5.2.2, then additional measurements must be made to isolate the non-conforming area.
8. Notes
Notes are not requirements of this standard.
8.1 PRINCIPLES OF THE MAGNETIC GAGE: Each of
these gages can senseand indicate only the distance between
the magnetic surface of the steel and the small rounded tip of the
magnet or probe that restson the top
surface of the coating. This
measured distance, from the top surface of the coating, must be
corrected for the thickness of any extraneous films or other
interfering conditions on the surface
of the steel. Such correction
is described in Section 4.2 for Type 1 gages and manufacturers
instructions for Type 2 gages.
8.1.1 Type 1 (pull-off)gages useacalibratedspring mechanism to measure the force needed to pull a small permanent
magnetfrom the surface of thecoated steel. The magneticforce
holding the magnet to the surface varies inversely as a nonlinear function of the distance between magnet and steel, .e.,
the thickness of the dry coating (plus any other films present).
The Type 1A banana gages use a helical spring to pull a
small permanent magnet from the surface. Internal balancing
mechanisms in most banana gages compensate for horizontal,
vertical and overhead positions so that there is no need to
recalibrate when changing orientation.
In a Type 16 pencil gage, a calibrated coil spring measures the force necessaryto pull the permanent magnet from the
surface. Because of gravitational effects, these gages must be
recalibrated when the orientation of the surface changes; e.g.,
a gage calibrated on a horizontal surface will not be accurate
when measuring a vertical surface. Some gages have three
separate indicators which compensate for horizontal, vertical
and overhead positions. Type 1B gages are generally not as
precise as Type 1A gages.
Normally, Type 1 gages are not adjusted or reset for each
new series of measurements.
Shims of sheet plastic or ofnon-magnetic metals, which are
permissible for calibrating Type 2 (constant pressure probe)
gages, should not be used for calibration of Type 1 gages. Such
shims are usually fairly rigid and curved and do not lie perfectly
flat, even on a smooth steel test surface. Near the pull-off point
of the calibration measurementswith any Type 1 gage, the shim
frequently springs back from the steel surface, raising the
magnet too soon and causing erroneous calibration readings.
6. Accuracy
6.1 To qualify under this standard, a gage must have an
accuracy at least within +IO%. Forthicknesses less than 25 pm
(1 mil), the gage must have an accuracy at least within e . 5 pm
(0.1 mil).
7. Disclaimer
7.1 While every precaution is taken to ensure that all
information furnished in SSPC standards and specifications is
as accurate, complete and useful as possible, SSPC cannot
assume responsibility nor incur any obligation resulting from the
use of any materials, coatings or methods specified therein, or
of the specification or standard itself.
--`,,,,,,`,```,,,`,,````,`````,-`-`,,`,,`,`,,`---
SSPC-PA 2
June 1,1996
S T D - S S P C PA 2-ENGL l 7 7 b W 8 b 2 7 9 4 0 0 0 0 4 b b b 8 5 2
8.12 EXTREMES OFTEMPERATURE: Most of the magnetic gages operate satisfactorily at 4C and 49C (40F and
120F). Somegagesfunction well at much highertemperatures.
However, if such temperature extremes are met in the field, the
gage might well be checked with at least one reference standard
after both the standard and the gage are brought to the same
ambient temperature. Most electronic gages compensate for
temperature differences among the gage, probe and surface.
--`,,,,,,`,```,,,`,,````,`````,-`-`,,`,,`,`,,`---
8.15 Polished metal calibration standards are manufactured by the National Institute of Standards and Technology
(NIST). The chrome plated panels are flat smooth steel 2.86 x
2.86 cm (1.125 x 1.125 in) in size. Examples of some NIST
standards are:
8.7 PROXIMITY TO EDGES: Magnetic gages are sensitive to geometrical discontinuities of the steel, such as holes,
corners or edges. Thesensitivityto
edgeeffectsand
discontinuities varies from gage to gage. Measurementscloser
than 2.5 cm (1 in) from the discontinuity may not be valid unless
the gage is calibrated specifically for that location.
COPYRIGHT 2003; The Society for Protective Coatings
SSPC-PA 2
June 1, 1996
(100 ft2).
80,225,lo00 pm (3,9,40
mil)
48,140,505,800 pm (2,5.5,20,31mil)
40,80, 140, 205 pm (1.6, 3, 5.5,8 mil)
Single standards from 3 pm (0.1 mil)
to 62 pm (2.4 mil)
FIGURE A l
Part "A" of Structure (Area Approx.10 m2 [lo0 W)
GAGE READINGS
/"
spot 2
spot 1
2.0
0
O
Avg 2.5
3.6
t-2.6
2.7
"
Avg 3.0
H
1.5 in
9ft
Avg 2.1
spot 4
-/3.2
~
O
spot 5
2.6
a-
Avg 3.0
2.8
2.6
4H
Avg
2.3
Avg 2.3
11 ft
5
COPYRIGHT 2003; The Society for Protective Coatings
--`,,,,,,`,```,,,`,,````,`````,-`-`,,`,,`,`,,`---
June 1,1996
to within
The thicknessof a plastic shim is typically accurate