Operation Manual

This document provides operation manuals for two analytical balance models, the VE-204 and VE-303. It includes sections on technical specifications, installation, operation instructions, calibration procedures, and maintenance. The technical specifications section lists the key specifications of each model including capacity, readability, repeatability, stabilization time, and power requirements. The operation section provides instructions on preparation, startup, unit selection, integration time adjustment, sensitivity settings, and calibration. It describes the internal calibration process for the VE-204 and external calibration using a standard weight for the VE-303. The document also covers parts counting, output settings, weighing procedures, and reading deviations.

Uploaded by

Guzman Susan
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
100% found this document useful (1 vote)
799 views

Operation Manual

This document provides operation manuals for two analytical balance models, the VE-204 and VE-303. It includes sections on technical specifications, installation, operation instructions, calibration procedures, and maintenance. The technical specifications section lists the key specifications of each model including capacity, readability, repeatability, stabilization time, and power requirements. The operation section provides instructions on preparation, startup, unit selection, integration time adjustment, sensitivity settings, and calibration. It describes the internal calibration process for the VE-204 and external calibration using a standard weight for the VE-303. The document also covers parts counting, output settings, weighing procedures, and reading deviations.

Uploaded by

Guzman Susan
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
You are on page 1/ 18

Operation Manual

Analytical
Balances

Models: VE-204 and VE-303

Content
1)
2)
3)
4)

Introduction
Technical specifications
Installation diagram
Operation of equipment
a) Preparation
b) Operation
b.1) VE-04
b.2) VE-03
c)
d)
e)
f)

5)
6)
7)
8)

change in unit of measure


Adjustment of integration time
Sensitivity settings RSD
Calibration of the balance
a) VE-204
b) VE-303
g) Output settings PRT
h) Parts count.
i) Weighing, Tare, add objects, reading of deviations
Maintenance and troubleshooting solution
interface
RS232C.
Accessories Extra Dreessy

1.

Introduction

This manual contains instructions for installation, operation and maintenance of


analytical balances VELAB , VE- 204 and VE- 303.

Please read fully the manual before using the balance. This balance has electronics
multifunctions using a system MCS- S4 patented by our company has heavy, ct,
ounces, grams, etc.. Including automatic calibration, adjustment of stability this
balance has RS232C port to connect a computer and printing.
These balance are used in laboratory areas.

Proprietary Medicinal Products, Industry, etc. Offering fast, efficient and accurate
weighing. Dressy

12

2. Technical specifications
Model

VE-204

VE-303

II

0-220

300

0.1

0-220

0-300

Repeatability (g) ( standard


deviation )
Linearity

+ 0.0002

+ 0.001

Stabilization time (seconds)

-6

Integration time (adjustable)

2.5 /5/10

2/4/8

90

80

Dimensions (mm)

360 x 217 x 345

360 x 217 x 345

Equipment weight

68

68

90 -12.0R1000
MA
15 W

9.0 -12.0 A =1000


MA
15 W

Degree of accuracy
Range of weighed (Capacity)
Accuracy of reading (Legibility)
(Mg)
Range of tare

0.0005

Platform size

Power source
Power consumption
Calibration

Internal

Electrical Requirements

Adapter AC
(included)

0.002

External with
standard weight
Adapter AC
(included)

Note: This balance light when the power button (ON) is pressed. The panel
only shows activation in the display. The connection of electric power must be
disconnected if the balance is not used in a certain time (5 days approx.) If you
use it daily there isnt need to disconnect, you have to press the (ON) button
located in the display. jorge

12

12

3. Installation diagram Erick josheep


1) Display
2) Weighing platform
3) Platform
2

4) Interface
5) Electric connection

CAL
UNI
T
COU

On
OF
IN
T

ASD
TA
Sensitivity R

PR
T

Screw and cover


Levelling feet

12

4. Operation of equipment
a) Preparation

Unwrapping perfectly the equipment, extracts all the accessories and free
pieces of the packing of polystyrene foams
Place And properly assemble the equipment as indicated in the instructions
Place the balance in the table proper working, away from vibrations, light of
the Sun, air, wind and unfavourable conditions that alter the perfect work of
the balance
Environment -Temperature 20 C +- 5 C
Relative -humidity 50% -75%
Suitable -Voltage 110V + - 11 V, 50 HZ Erick Joshep

b) b)Operation

Once placed the scale in the desk (there is recommended a fixed,


completely stable table and designed for balance made of granite, etc.)
To check that the bubble of leveling is in the middle of the circle or
otherwise to jousts the legs of the balance to achieve the above
The tactile buttons of the scale, allow you to control efficiently, with its
multiple functions
Connect the equipment and the display shows "OFF"
For the balance VE-204 press "ON" and is show in the display of form
descendant 30 minutes, wait the time displayed and after finished the time
can press the button "ON"
At Turn its balance displayed on the display "0.0000g" means that your
balance was successfully calibrated and can work with the team. Note:
When in the heating process by pressing the "tare" you can open the
equipment without calibrate the balance
"OFF" display indicates shutdown. Adan

Press gently "OFF" (Switched off) and on the display will show "OFF"
The scale in protection way of screen "Screen Protection", later presses "TAR".
Place the container in the plate of the tray with the weight.
Before operating the Balance check that the bubble is positioned in the middle
of the circle, and this is achieved adjusted their legs thereof.

12

Press the "TAR" tare button, the display will show zeroes. The tare is completed

When removing the container, a negative value is shown on the display.

Press "TAR" tara again and the display will change showing zeroes Tara is ready
Salvador

For the Scale VE-3Q3


Connect and turn "ON", for recommendation one asks to leave her an hour
ignited before using it (in case that the display does not show anything).
Function control panel "ON" display on.
Gently press the "ON" button and the display on Czech "the
function display, after 2 seconds, the balance model appear and
then light up and then its heavy function.
"OFF" Display Off fidencio

Press the "OFF" button and the display will disappear. If the balance is not
used a certain time, please disconnect the unit from the power supply.
"TAR". Place the container on the pan of the balance and the weight keeps
the following.

Press "TAR" tare, the display changes and showing zeros on the display. The
tare is ready.

When you remove the weight, a negative value is shown on the display

12

Press "TAR" again, the display will

mark zeros.

C. Change In Unit Of Measure


If the accuracy of the reading needs to be 0.1mg, check at the end of the
display and it indicates the unit of measure.

The balance VELAB can be formed to measure several units. Sub-menu of


units is in use for enable or disable a specific unit "UNITSW"

Press the button "ERY" until the display shows the following figure

And every value is explained example; "g" means grams, "ct" means value of
heavy so called Carat, "ozt" ounce troy, "GN" grain , "oz" ounce, "N" Newton,
"Ib" pound, etc. The unit will appear by means of RNG.

D. Adjustment of integration time matilde


4 modes integration time can be selected as shown in Figure.

The integration time is divided as follows. INT-0 fast, INT-1 very short, INT-2
Short, INT-3 long. The selection of the integration time is the same as RNG.

E. Sensitivity settings RSD


Like the adjustment knob of time INT, there are 4 other ways to provide this to
adjust sensitivity.

ASD-0 higher. ASD-1 High, ASD-2 Medium, Low ASD-3. ASD-0 is used in test
equipment should not be used in user.
12

The selection of the sensitivity is the same as RNG.

Here is a list of how the ASD INT is used as reference only.


The fastest way to Heavy; INT-1 ASD-3
Normal, INT-3 ASD-2
With unfavorable environment. INT-3 ASD-3

F. Calibration of the balance


Normally, your balance should be calibrated after long unused or stored, with a
sudden movement, or a different environment that you work, is when
calibration for accurate measurements of heavy is recommended.
There are 2 types of balance Orlando
A) With Internal Calibration (VE-204)
Remove all objects that are on the weighing platform, press "CAL" is displayed
until the "CAL int" display, then automatically unppshed and the balance will go
to calibration mode automatically. When the display shows "0.0000g" means
that the calibration was successful. The balance will go to normal working
mode.
Note 1: "CAL int" means internal calibration
Note 2: After heating process 30 minutes, the balance in your display shows
nothing but internal calibration will automatically after 10 minutes without
performing any weighing.
B) With External calibration (VE-303)
Note the external weight ready for use
Remove all objects on the weighing platform, press "CAL" until is displayed on
the display "CRL-E", the balance will go the way of external calibration, it is
displayed flashing in the "CAL-200" display . Place the external weight (200 g)
in the center of the plate, then the balance will go to calibration status. After a
few seconds, it will be displayed on the display "200.000g". Remove the
calibration weight and the display indicates "0.000g". Balance Calibrated and
ready for use. Note: It is recommended 2 times this procedure to ensure
perfect calibration of equipment. The calibration sequence as shown below.

12

g. Output settings PRT Marco /mauricio


PRT and 4 hold modes will be shown on the display PRT-0: The indefinite mode
output setting time. PRT press while the weight shown in the output interface.
Attention: Press the button quickly or possibly the following display mode
PRT-1 Output every 30 seconds
PRT-2: Output every minute
PRT-3 Output house 2 minutes

h. Parts count
The balance has a counting function. The average is: 5. I0, 25 and 50 with a
total of four averages. COU press and hold, the display circulated as follows:

The percentage of value, (5, 10, 25 and 50) objects respectively.

If functions is Required Heavy press when the display shows COU-00 and
indicate a wait state "_" at least "0.000g" is displayed on the display. If the
percentage of the required value is five press when the display shows COU-5,
and place 5 items on the plate and so on depending 5,10, 25 and 50.

12

Weighing
parts count
weight Cumulative

Put the container on the plate, the weight is shown on the display.
Press TAR and shows zero, that's tara. Despite putting objects in the
container or tray weighing weight value shown is the net weight of the
object only.
Place objects to weigh on the weighing plate one by one with the TARE
mode and cleaned one by one. Remove all objects that have been
weighed, the absolute value of the total weight is to be displayed.

deviations Reading of

After selecting the mode (all modes can be used for weighing, after
pressing OFF, depending on the memory function of the balance), press
"TAR" and zero is shown on the display. Place the weighing on the
plate. When the number is set to "0" on the right side of the display will

Tare

i. Weighing, Tare, add objects, reading of deviations


Susan /narciso

Put the weight or sample on the pan and press TARE, then remove the
reference and displayed on the display a negative value. Place the
object to be weighed on the plate. The sum shown deviations show
more or less, compared to the weight of the object, is what you are
looking for.

Apply INT-0 mode and put the tray or container on the plate, then press
the TARE button. Add the objects to be weighed (either liquid or solid)
the container one by one continuously reading weighing obtaining
quickly. When added objects reach "0" on the right side of the display
of balance and the value required by the user. When you add all
objects so with seal, the net weight of each object can be measured by

12

5. Maintenance and troubleshooting solution pepe


The scales themselves are equipment to be used in delicate, have always clean
the platform. Wipe with a soft cloth, never use agents or solutions that can
damage the equipment.

Nmb
er

Problem

The display does not


turn

Medium shows
Superior Display
Power

Medium shows
On lower display

The weighing is not


stable, is in constant
motion

The result is not


correct

Case
-It is not on
-Switch is not lit
-Interference
-Overloaded equipment
-Calibration memory is
damaged
-The dish was placed
incorrectly
-No plate Installed
-Dish improperly installed
-Air flow
-The worktable is not
stable
-Integration time is very
short
-No zero before weighing
frame
-The scale is not red, or the
calibration weight is not
correct
The line voltage is
incorrect

12

Solution
-Check and turn
-Press' "ON"
-Check Switch and
connection
-Immediately
reduce weight
-Recalibration with
proper procedure
-Remove the plate
and reinstall
-Properly position
the weighing plate
-Check the glass
components are
closed
-Change Long Time
Integration
-Press button tare
-Recalibrate
-Change electrical
connection

Nmb
er
Problem
6
Display
It remains in a certain
digit or Indicates a
meaningless code
Display is in standby
7

Case
-Interference
-Erroneous line of
Voltage

Solution
-Come back to
connect the balance
and check connection
-change The line
voltage
-The position of the
-Put Less Sensitivity
balance is not correct -Improve the work
(Vibration and draught) environment
- very high sensibility
- There is an object on -Remove The
the platform that does object
-Clean And
not allow calibration
Calibrating
-Press CTRL

El Display shows "CRL


Err"

Display shows Err-1. Err- -Instan Interference


2
- A critical error in the
balance

10

- It is not calibrated
The measure of
-The Number of
weighed in the right
internal
calibration
side of the display is not
memory
has been
the sample
deleted

11

Display shows "CouErr"

- No constant
-On Estimated
weight
-under Estimated
weight

Octavio

12

--Turn Off and turn


on the balance
-Bring a Technical
Service
-Balance calibrate

- Programme the
percentage of count
of parts adequately

6. Interface Ruben
The information that adopts the Interface Refers to a program (9-core RS-232C)
,comes with 2 Interface RS232C universal connections which can be connected
to a microprocessor or possibly any printer Pin and corresponding signals are:
PIN
1
2
3

SIGNAL
SI
GND
SO

GND

SIGNAL
Internal Signal
GND
External Signal (Range
1200)
GND

7. RS232C
Computer(9 PINS)

Balance (9 PINS)
2
3

2
3
5

1. Range Band port is 1200.


2. The data format is 10 digits, the first digit (0), 8 digits are the position data.
(ASCII Code, small digits shown). One digit from Paro.
3. Data is Mandaran to the output port without any special reading commands,
a form of sailda detailed below:
12

8. Accessories extras Dressy/jorge


1)
2)
3)
4)
5)

External weighs 200 g and 100 g, VE-303 model only 1 piece


Connection cable 1 piece
Handkerchief 1 piece
Operation Manual 1 piece
Certificate 1 piece

12

LF

15 16

CR

14

Unt 2

13

Unt 1

12

Data

11

Data

IO

Data

Data

Data

Data

Data

Data

*/Space
+/-

Space

Type

12

www.velab.net

12

You might also like