Principles of Radiotherapy
Principles of Radiotherapy
Principles of Radiotherapy
DAVID YOO x CHRISTOPHER WILLETT
KEY POINTS
The clinical application of radiation oncology balances
the dual goals of improved tumor control with reduced
treatment-related morbidity.
Over time, technological and biological advances have
offered many opportunities to further reduce normal
tissue complications and improve tumor control with
radiotherapy.
Therapeutic Ratio
Introduction
The discipline of radiation oncology involves the therapeutic
application of ionizing radiation in the management of benign
and malignant diseases. In fact, x-rays were used to treat a
patient with ulcerated breast cancer less than 2 months after
their discovery in November 1895.1 Therapeutic radiology, like
its diagnostic counterpart, has undergone considerable evolution since then. Still, the practice of radiation oncology has at
its core several fundamental concepts that have guided, and will
continue to guide, management decisions.
Tumors and their neighboring organs will remain in proximity to one another. These tumors and their normal tissue
counterparts will have varying responsiveness to ionizing radiation. Therefore, an ideal radiation treatment will balance optimal tumor control with acceptable acute side effects and late
toxicities. Over time, new technological or biological advances
will promise improved differentiation of these two populations,
with better clinical outcomes. Some of these advances will actually work, shifting paradigms and reinforcing the ongoing need
for rigorous testing before widespread adoption and changes to
accepted standards of care.
With each new patient encounter, the radiation oncologist
considers these concepts and determines the most appropriate treatment approach. A simplified algorithm is shown in
Figure 4-1, showing the general thought processes and decision points involved. The global decisions regarding treatment and goals of therapy are usually reached within a
multidisciplinary context with other medical providers and
informed consent of the patients and their support networks.
Once the decision has been made to incorporate radiation
therapy into the treatment plan, the more specific questions
regarding timing, dose, treatment volume, and technology
are considered. The solid and dashed arrows between the various boxes represent the interrelated nature of these questions
40
TREAT?
Contraindications
Alternatives
Literature support
GOAL?
Cure
Palliation
Performance status
Patient preferences
Target Volumes
TIMING?
Neoadjuvant
Definitive
Adjuvant
Emergent
Urgent
Delayed
VOLUME?
Primary site
Nodal basins
Distant sites
Normal tissues
Critical organs
Margins
DOSE?
Fraction size
Total dose
Treatment time
Chemotherapy
Biologics
Hyperthermia
TECHNOLOGY?
Target delineation
Immobilization
Inverse planning
Image guidance
Adaptive therapy
100
100
75
75
50
50
25
25
When radiation oncologists discuss treatment targets and volumes (Figure 4-3), they utilize definitions and acronyms developed by the International Commission of Radiation Units and
Measurements (ICRU). The ICRU 50 report published in 1993
introduced the concepts of gross tumor volume (GTV), clinical
target volume (CTV), and planning target volume (PTV).9 The
GTV represents grossly visible disease, either clinically or radiographically. The CTV represents areas of potential microscopic
extension of disease from the GTV. These microscopic deposits
may be in proximity to the primary tumor and/or dispersed
throughout the relevant draining regional lymph node basins.
The PTV includes sufficient margin around the CTV to ensure
adequate coverage and accounts for various uncertainties such as
patient/organ motion and daily setup errors.
Technologic advances such as 4DCT, which captures organ
motion, and immobilization devices/onboard imaging modalities that can reduce setup errors, have helped characterize and
eliminate some of these uncertainties. A modern linear accelerator treatment machine with onboard imaging and respiratory motion tracking capabilities is shown in Figure 4-4. The
subsequent ICRU 62 report considered these advances with the
addition of internal target volume (ITV) and setup margin
(SM).11 Now, PTV 5 CTV 1 ITV 1 SM. Further advances in
diagnostic radiology with functional imaging modalities are
providing potential biologic target volumes (BTVs) to further
improve the therapeutic ratio.12
These target volume definitions help to determine how much
technology and complexity is required to safely and effectively
deliver the radiation treatments to optimize the therapeutic
ratio. Depending on the clinical scenario, treatment plans may
PTV
ITV
Increasing dose
Figure 4-2 Therapeutic ratio. The blue solid curve represents tumor
control probability (TCP), whereas the red solid curve shows normal tissue
complication probability (NTCP). As radiation dose increases (dashed
vertical line to solid vertical line), both tumor control and toxicity are likely
to increase. The final dose chosen strikes the optimal balance between
the two outcomes. Various strategies can be employed to separate the
two curves. Factors that sensitize tumors to radiation may shift the TCP
curve to the left (dashed blue curve), whereas agents that protect normal
tissues may mitigate treatment-related toxicity, shifting the NTCP curve
to the right (dashed red curve).
4 Principles of Radiotherapy
Motion detector
Gantry head
kV image
source
kV image detector
Treatment
couch
MV image detector
Figure 4-4 Linear accelerator treatment machine with onboard imaging and respiratory motion tracking devices. The radiation is delivered
via the gantry head that can rotate 360 degrees. The treatment couch
also rotates, allowing for various beam angle approaches. The kilovoltage imaging source and detector allow for acquisition of static portal
images as well as cone-beam CT images to confirm patient setup.
Megavoltage portal films may also be captured to ensure proper targeting. An infrared tracking device is seen in the ceiling, allowing for
detection of respiratory motion.
41
Figure 4-5 Three-dimensional treatment fields for rectal cancer. The beam apertures are custom-designed with multileaf collimators located in the
head of the treatment machine to encompass the target volumes with adequate margin while sparing surrounding normal tissue structures. Normal
tissues and tumor volumes are contoured and projected onto digital radiographic reconstructionleft femoral head (blue), right femoral head (tan),
small bowel (magenta), bladder (purple), rectum (yellow), iliac blood vessels (turquoise), and gross tumor volume (red). (A) Posterioranterior field.
(B) Left lateral field.
42
Breath-hold (BH)
Free-breathing (FB)
Heart
(BH)
the heart has physically moved away from the chest wall, allowing the same tangent field to cover the breast while minimizing
heart irradiation.
Because the late effects of radiation-induced changes to the
heart may require 1020 years to manifest, longer follow-up
and experience with this breath-hold technique are required
before the potential benefits of this change may be seen. Still,
the hope would be that the benefits of radiation already seen in
terms of locoregional control and survival would be further
enhanced if more patients were no longer dying from subsequent cardiac complications. Other strategies being employed
to minimize heart dose include treating patients in the prone
position or the use of IMRT to minimize heart dose in treatment planning. A much more radical change currently under
debate within the field is a redefinition of the appropriate CTV
itselffrom treating the whole breast to just the lumpectomy
cavity alone.19
Despite the continuing refinement and focus on target volumes, the tolerance of the surrounding normal tissues often
drives decisions regarding treatment. Acute side effects happen
during therapy and are often transient, requiring supportive care,
whereas late effects, as discussed, can develop months to years
following radiation therapy and tend to be irreversible. The issue
is further complicated by the fact that organs often have different
expressions of toxicity with different time courses. For example,
the skin may respond acutely to radiation with mild erythema
and dry desquamation. With increasing doses, patients may develop moist desquamation and even ulceration. Late effects may
include tanning changes, skin fibrosis and contracture, atrophy,
telangiectasias, or permanent hair loss.
Heart
(FB)
C
Figure 4-6 Utilization of breath-hold for treatment of left-sided breast cancer. (A) A merge of axial slices is seen from both free-breathing and deep
inspiration breath-hold CT simulation scans. The outer body contour shows increased chest wall expansion with deep inspiration. (B) Free-breathing
CT scan, with the thick yellow line representing the medial border of the tangential radiation fields used to treat the whole breast. In this scenario,
a small portion of the left ventricle is within the radiation treatment fields. (C) The use of deep inspiration breath-hold allows the same radiation
tangential fields to cover the left breast without irradiating the left ventricle.
4 Principles of Radiotherapy
Not all normal tissues are created equal. Some are organized
in a parallel fashion, with multiple independent functional subunits.20 For example, the kidney is composed of approximately
1 million separate nephrons, capable of performing their duties
independently from their neighboring subunits. Therefore,
organs arranged in parallel, such as the kidney, lung, and liver,
may have significant portions surgically removed or irradiated
beyond tolerance, with the remaining undamaged subunits able
to maintain normal global organ function. Other tissues, such
as the small bowel and spinal cord, are arranged in series, where
excessive damage along any point may compromise the entire
organs function.
These differences have significant implications in treatment
planning. When the target lies in proximity to a parallel structure, the dose parameters to avoid excessive toxicity are typically
expressed in mean dose or volume percentages. For example,
when evaluating lung cancer plans, a primary concern is the
potential development of radiation pneumonitis. One wellstudied dosimetric parameter is the volume of normal lung
receiving at least 20 Gy (V20), with higher percentages associated with increased risk of pneumonitis.21 For serial organs, the
goal is to avoid any dose beyond the expected tolerance of that
structure. With either the small bowel or spinal cord, the maximum dose is generally in the 4550-Gy range for conventionally
fractionated regimens.
Not all toxicities are created equal as well. Radiation-induced
cataracts may develop after doses beyond 10 Gy to the lens. These
can be easily removed surgically. Spinal cord myelopathy resulting in paralysis or blindness due to optic chiasm damage are
irreversible catastrophic complications. Figure 4-7 shows a patient
with nasopharyngeal carcinoma with intracranial extension
treated with concurrent radiation and chemotherapy.22 Panels A
and B demonstrate the utility of magnetic resonance imaging
(MRI)23 in the delineation of both tumor and normal tissues
compared to CT. The light-blue contour represents the 4400-cGy
isodose line. This dose level with chemotherapy is felt to be sufficient for microscopic disease control in the cervical neck nodes.
The yellow contour shows the 7000-cGy isodose line used for
gross disease control.
This is an example of an IMRT plan, where the intensity of
the beams is modulated during therapy to allow for more conformal dose distributions. The isodose lines show the radiation
dose being shaped away from the bilateral parotid glands and
the larynx in order to minimize long-term xerostomia and
edema/dysphagia. More critically, panel C shows the proximity
of the optic chiasm to the target volumes. Visual loss from radiation-induced optic neuropathy has been carefully studied and
reviewed. The risk is very low when the maximum dose to the
chiasm/optic nerves is less than 55 Gy at 1.82.0 Gy fraction
sizes. The incidence rises to 3%7% with maximum doses of
5560 Gy and rises quickly beyond 60 Gy.24 In this example, the
chiasm has been contoured and expanded by 1 mm to allow for
daily setup uncertainties (chiasm 1 1 mm). During IMRT
planning, priority was placed on covering the target volumes
while keeping the dose to the chiasm 1 1 mm structure less
than 54 Gy.
Plan evaluation is performed by careful examination of
the isodose distributions in the various axial, sagittal, and coronal planes and by review of a dose volume histogram (DVH)
that shows the various structures and their associated dosimetric characteristics (panel D). Based on the plan, the expanded
chiasm 1 1 mm structure is receiving a maximum dose of
43
5334 cGy. In essence, this 5334-cGy dose represents the worstcase scenario if the patients setup position is off by 1 mm every
day toward the high-dose areas. The chiasm itself is receiving
4731 cGy. Other organs, such as the parotids, are arranged in a
more parallel fashion. For these, the maximum dose has not
been found to be as critical as the mean dose received. Generally, mean doses less than 26 Gy are associated with improved
salivary function.25 Given the need to balance cure with toxicity,
and that patients are unlikely to die from xerostomia, target
volume coverage is usually given priority over parotid gland
sparing.
Over time, the typical tolerances of normal tissues to conventionally fractionated radiation have been meticulously gathered
and reexamined.26,27 The rapidly growing use of stereotactic techniques, which deliver much higher doses per fraction, has
spawned significant discussion and research in the establishment
of guidelines and tolerances of normal tissues to these larger
doses. For example, the recommended optic chiasm tolerance
for single fraction radiosurgery procedures is in the 1012-Gy
range.24
Dose Fractionation
In the earliest days of radiation therapy, treatments often involved
a single large fraction of radiation. Any clinical responses were
unfortunately accompanied by significant early and late toxicities.1 Subsequent strategies employed more protracted courses of
therapy, splitting the radiation treatment into multiple fractions
with smaller doses. Classical experiments from the 1920s with
rams testes showed the ability of small daily radiation fractions to
sterilize the animals without the resulting skin necrosis associated
with single large doses. The radiobiology behind fractionation
was best described by Withers and what he labeled the four Rs
repair of sublethal injury, cell repopulation, redistribution into
more radiosensitive phases of the cell cycle, and reoxygenation.28
Nonmalignant cells are more adept at the repair of radiationinduced DNA damage than their malignant counterparts. Therefore, the use of multiple radiation fractions allows for killing of
tumor cells, while giving the normal surrounding tissues time to
repair sublethal damage. One potential drawback of protracted
treatment courses is the opportunity for repopulation of both
normal and cancerous cells. For this reason, more fractionated
regimens typically require higher total doses to achieve similar
degrees of tumor control. Still, with each subsequent fraction, the
potential exists for more cancer cells to redistribute into the more
radiosensitive G2/mitosis phase of the cycle, allowing for more
effective radiation-induced cell kill. One final advantage of fractionation is the potential for reoxygenation of tumors during
treatment progression, as hypoxic cells are much more radioresistant compared to normoxic cells.
Different fractionation regimens have been used in a number of clinical scenarios. For example, common prescriptions
for the palliation of bone metastases include a single fraction of
8 Gy, 5 fractions of 4 Gy to a total dose of 20 Gy, 10 fractions of
3 Gy to 30 Gy, and 14 fractions of 2.5 Gy to 35 Gy.29 To make
comparisons between various schedules, radiation oncologists
often speak in terms of the biologically effective dose (BED) of
a particular treatment regimen. This equation takes into account a number of parameters, including total dose, fraction
size, overall treatment time, and the relative sensitivity of the
tissue to changes in fraction size.30 This final factor is known as
the alpha/beta ratio. Most tumors and normal tissues with
44
PTV44
PTV70
chiasm
cord 5 mm
R parotid
chiasm 1 mm
L parotid
D
Figure 4-7 Intensity-modulated radiation therapy (IMRT) for nasopharyngeal carcinoma. Isodose distributions from the IMRT plan are superimposed
on coronal images from the CT (A) and MRI (B) simulation planning. The isodose lines show target volume coverage with sparing of the bilateral parotid
glands and larynx. The light-blue 4400-cGy isodose line covers all areas of potential microscopic tumor spread, whereas the yellow 7000-cGy isodose
line encompasses areas of gross disease. (C) The use of IMRT allows more precise coverage of target volumes while simultaneously sparing the optic
chiasm. To account for potential setup errors that could result in exceeding organ tolerance, the chiasm is expanded by an additional millimeter
(chiasm 1 1 mm) for treatment planning purposes. (D) The dose volume histogram (DVH) displays the dose delivered to various normal tissues and
target volumes.
4 Principles of Radiotherapy
45
46
C
Figure 4-8 Stereotactic radiosurgery for brain metastasis. (A) The target volume is shown in purple. The optic structures and brainstem
are also contoured. The fanlike structures show the paths of the four dynamic arcs used to deliver the radiation dose. (B) This reconstruction
shows the beams and field apertures. As the gantry arm moves along the arc path, the micro multileaf collimators continuously change
the field aperture shape to conform to the target volume. (C) The resulting isodose lines show the tight conformality of dose around the
metastasis.
4 Principles of Radiotherapy
47
E
Figure 4-9 Stereotactic body radiation therapy (SBRT) for early-stage lung cancer. (A) Sagittal images from 10-phase four-dimensional CT scan
(4DCT) demonstrate tumor motion during normal respiration. (B) Images from the4CT are used to generate a maximum-intensity projection (MIP) of
the tumor mass from all phases of the respiratory cycle. The isodose lines from the SBRT plan are superimposed, showing target volume coverage.
(C) The nine-field beam arrangement used for the treatment plan. (D) Cone-beam CT obtained on treatment table just prior to treatment to confirm
accurate target localization. (E) Cine-megavoltage images acquired during treatment delivery from one of the nine fields to confirm tumor coverage
throughout therapy.
Summary
Technological innovations continue to change the field of
radiation oncology. The dual goals of improved treatment efficacy and reduced toxicity remain at the core of the discipline.