Basic Principles of Instrument Calibration
Basic Principles of Instrument Calibration
Every instrument has at least one input and one output. For a pressure
sensor, the input would be some fluid pressure and the output would (most
likely) be an electronic signal. For a variable-speed motor drive, the input
would be an electronic signal and the output would be electric power to the
motor. To calibrate an instrument means to check and adjust (if necessary)
its response so the output accurately corresponds to its input throughout a
specified range
Calibration is one exercise that is often taken for granted within an
industrial plant. Even the most important industrial equipment will become
useless if it is not calibrated. Through the process of calibration, adjustments
are made to a piece of equipment or device to ensure that it performs as
expected to deliver predictable, accurate and reliable results that meet
quality standards. Adjustments made during calibration must fall within
certain tolerances. Such tolerances represent very small, acceptable
deviations from the equipments specified accuracy
Definition of Calibration.
Instrument calibration can be defined in several ways. Put simply,
calibration is the process of adjusting an instrument or equipment to meet
the manufacturers specifications.
Calibration can also be defined as the process of issuing data including
a report or certificate of calibration that assures an end user of a products
conformance with its specifications.
To the instrument engineer or technician, calibration is the process of
determining the relationship between the values of the quantity being
measured and that indicated on a measuring instrument. The calibration of
an instrument can be carried out by comparing the readings on the
instrument with those given by a reference instrument or calibrator. From
time to time, the manufacturers reference instruments are sent to a
calibration center to be calibrated against national standards.
When an instrument is purchased, the manufacturers calibration data
is generally supplied. Most instrument manufacturers have sets of reference
instruments against which all instruments they produce are calibrated
Why Calibrate an Instrument?
Virtually all equipment degrades in some fashion over time, and
electronic equipment, a mainstay of todays manufacturing process, is not an
exception. As components age, they lose stability and drift from their
Instrument Range
The instrument range refers to the capability of the instrument. It is
often the nameplate rating of the instrument. For example an instrument
nameplate rating may read:
Instrument range:
0 -800psig
Output:
4 to 20mA.
Never confuse the instrument range with the calibration range. They
are two different things. Although our instrument range is 0 800psig, we
may decide to calibrate it to a range 0 400psig or even 0 800psig for an
application with high input pressure in which case the instrument range
becomes the calibration range of the device.
Ranging an Instrument
To range an instrument means to set the lower and upper range values
so it responds with the desired sensitivity to changes in input. Suppose we
want to use a pressure transmitter to measure pressure in the range 0 -100
bar to give an output of 4 20mA. To range this transmitter, we simply set:
0 bar= 4mA
100 bar = 20mA
Closely related to ranging is re-ranging which simply means resetting
the lower and upper range values to a different measurement range. For
example, suppose we want to re-range the above transmitter to now
measure pressure in the range 50 150 bar we simply reset as follows:
50 bar = 4mA
150 bar = 20mA
Zero and Span Adjustments
Zero and Span Adjustments are commonly done on analog and smart
instruments. By adjusting both zero and span, we may set the instrument for
any range of measurement within the manufacturers limits. For most analog
instruments, zero and span adjustments are interactive. That is, adjusting
one has an effect on the other. Specifically, changes made to the span
adjustment almost always alter the instruments zero point. An instrument
with interactive zero and span adjustments requires much more effort to
accurately calibrate, as one must switch back and forth between the lowerand upper-range points repeatedly to adjust for accuracy
For smart instruments however, there is no interaction between the zero and
span adjustments.
The
The
The
The
The
as found data
current calibration date
final calibration or as left data
name or initials of the technician who did the calibration
date the instrument is due for the next calibration
As Found Data
The as found data of an instrument to be calibrated is the response
(reading) from the device at the points of calibration (0%, 25%, 50%, 75%
and 100%) before the actual calibration exercise begins.
As Left Data
The as left data of an instrument is the response (reading) from the
device at the points of calibration (0%, 25%, 50%, 75% and 100%) after the
instrument has been calibrated.
Traceability
All calibrations should be performed traceable to a nationally or
internationally recognized standard. Traceability is defined as the property of
a result of a measurement whereby it can be related to appropriate
standards, generally national or international standards, through an
unbroken chain of comparisons. This means that the calibrations performed