Calibration of Differential Pressure Transmitter
Calibration of Differential Pressure Transmitter
For your application, this could be modified slightly. For example if the
pressure source is a hand pump, you can easily control the pressure applied
to the DP cell. However, if you are doing field calibration that requires the
use of the actual process pressure, you will need a pressure regulator in
conjunction with a pneumatic calibrator to help you control the pressure
applied to the DP cell.
Step 4:
Most transmitter calibration done is a five point calibration. That is for
0%, 25%, 50%, 75% and 100% of input span or range (in this case pressure
input into the DP transmitter). This should correspond exactly to 0%, 25%,
50%, 75% and 100% of the transmitter output span (4- 20mA). The graph
below illustrates the correlation between input and output values.
Readings are taken for both increasing and decreasing input values
and the corresponding transmitter output values are recorded
Step 5:
The low port of the DP transmitter cell is vented to the atmosphere (as
shown in the connection diagram above) and the high port of the DP
transmitter connected to a pressure source e.g a hand pump or any other
suitable pressure source in a bench calibration or the actual process pressure
through a pressure regulator and a pneumatic calibrator in a field calibration.
So once your equipment is well setup, power it up and pressurize the high
port of your DP transmitter. Record the current reading in (m A) which will be
your first data point. Continue pressurizing the transmitter and recording
your readings for the five points (0%, 25%, 50%, 75% and 100% of input
pressure). All the reading obtained will be the as found readings. If you
calibrate the DP transmitter before first testing and recording the as found
data, the history of the device performance data will be lost.
Step 6
Start to calibrate the DP transmitter. Note that during the calibration
process, the transmitters zero percent, (LRV), is to be calibrated to the,
Adjust the SPAN screw while observing the meters current indication
to cause the meter to indicate 20 mA, which is the 100% (URV) output value
signal
for
the
DP
transmitter.
Step 12:
100% input to the transmitter (pressure) exactly equals the
transmitters 0 % through 100 % output (4 - 20 mA current). A correctly
calibrated DP transmitter can be described as one where the % input equals
the % output for all values between 0 and 100 percent.
Once you are satisfied with the level of accuracy of the calibration result, you
are done with the calibration of the DP transmitter otherwise you will have to
continue fine tuning the calibration process until a reasonable accuracy is
achieved.
Effective Tips in Calibration of DP Pressure Transmitters
Calibration of a DP pressure transmitter involves a process by which
the output of the transmitter is adjusted to properly represent a known
pressure input. Calibration is one of the most frequently performed
maintenance operations on pressure transmitters. If well performed, the
transmitters performance improves otherwise its performance could
deteriorate with grave consequences. A pressure input is used to provide
zero and span adjustments to the transmitter in the calibration process.
The following tips are general guides that you should have at the back
of your mind when calibrating a DP pressure transmitter:
Tip 1: - Read and understand the calibration procedure in the
manufacturers instruction manual. The calibration procedures in the manual
should be followed carefully to ensure a proper calibration.
Tip 2: - The use of proper calibration equipment is crucial. The pressure
source and any readout device in use must be of greater accuracy than the
instrument being calibrated. Some experts in calibration have posited that as
a general rule, the pressure source and readout device should be at the
minimum four times more accurate than the device being calibrated. High
accuracy measurements cannot be obtained when the calibration is done
with low-accuracy equipment. It should be a regular practice to check the
accuracy of calibration equipment against a higher standard on a regular
basis to maintain the accuracy of the calibration equipment.
Tip 3: - When doing calibration, leaks are a potential source of error.
Eliminate all leaks in the calibration system. Use TEFLON tape on all pressure
connections.
Transmitter Re-ranging
Configuration function
that
changes
transmitter
MWP
MWP means the Maximum Working Pressure of the DP transmitter. MWP refers to the
amount of gauge pressure common to each port (High and Low), not the differential pressure
between ports, which the DP transmitter can safely handle without being damaged.
Transmitter Damping
Output function that increases the response time of a transmitter to smooth the
output when there are rapid input variations
Zero Trim
A zero-based, one point adjustment used in different pressure applications to
compensate for mounting position effects or zero shifts caused by static pressure.