Accurate Optical Biometry Using IOL Master
Accurate Optical Biometry Using IOL Master
Lyn Millbank
4
ational Journal of Ophthalmic Practice. Downloaded from magonlinelibrary.com by 194.074.145.068 on July 20, 2016. For personal use only. No other uses without permission. . All rights res
Practical
all the different versions of software for each
scenario, therefore, the reader is advised to
consult the manual provided to work within
the limitations of the model used in clinic.
Learning how to
perform biometry using
the IOLMaster
Read instructions, attend
courses, ask colleagues
Start by reading the manual carefully (available in several languages) and also attend
Calibration
Do not forget to always check the calibration of the equipment every day see the
manual section for Test Eye.
6
Normal value
mean (range)
Beware
Notes
Adjust AL
settings if
measuring
other than
phakic eye
Corneal curvature
Distance between
4244D
(Ks)
reflected infrared
(Expect steeper in
images projected onto
hyperopes and flatter
the cornea
with myopes)
Refractive power
of cornea approx
43D
Corneal thickness
approx 0.5 mm
Anterior chamber
Distance between the
3.24 mm
Do not measure ACD on
depth (ACD)
anterior vertex of the
(Varies, may be
IOLMaster in pseudophakia
cornea and the anterior
shallower if dense
or aphakia
vertex of the natural lens
cataract, high hyperopia
Ks are required before ACD
or glaucoma)
measurement can be taken
Average lens
thickness approx
4.63 mm
Refractive power
of lens approx
20D
ACD required for
Haigis and Haigis
L formula
White to White
Horizontal diameter of
12 mm
(WTW)
the iris
(11.512.5)
Provides distance
of visual axis
from centre of
iris
ational Journal of Ophthalmic Practice. Downloaded from magonlinelibrary.com by 194.074.145.068 on July 20, 2016. For personal use only. No other uses without permission. . All rights res
Practical
case of myopic shift induced by the cataract.
Access the patients ophthalmic history to
account for any out of range measurements.
Prior refractive surgery will alter the Ks and
if this information is ignored the chances
are that the outcome will lead to a refractive
surprise. Patients who have had refractive
surgery have high expectations. However,
they do not always volunteer the information and may not consider they have had
surgery.
Do not measure patients with known
retinal detachments or macula oedema
because optical biometry measures the
AL along the visual axis from the tear film
to the retinal pigment epithelium (RPE)
using partial coherence interferometry. If,
for example, macular pathology is present,
a double peak can occur in the AL and the
value may be variable and imprecise (Kojima
et al, 2010).
Ideally biometry should be performed
before dilating drops are instilled and before
tonometry or pachymetry is performed
(Royal College of Ophthalmologists, 2010),
because the shape of the cornea will be
changed and the Ks may be inaccurate. If
the patient wears contact lenses consult
departmental protocol or refer to RCO guidelines (Royal College of Ophthalmologists,
2010) and inform the patient to remove
contact lenses accordingly in the appointment letter. Contact lenses will also affect K
measurements.
For axial lengths from 22.50 mm to
26.00 mm, and central corneal powers
ranging from 41.00 D to 46.00 D, almost
any modern IOL power calculation formula
will give good outcomes. Based on certain
indications (e.g. axial lengths and/or
corneal powers outside the normal range,
previous surgery or laser treatment) the
IOLMaster uses different formulae to make
the correct calculations. Table 2 shows the
formulae and the parameters needed for
every formula, together with the constant
that is used. Normal values for the different
parameters are given in Table 1.
Conclusion
Today, with sophisticated IOL power calculations, about 90% of cases fall within 1.0
D and 99.9% within 2.0 D of their targets,
assuming optimized conditions. Prediction is more accurate in long eyes and less
Requirements
A constant
Haigis
AL K1 K2 ACD
a0 a1 a2
Hoffer Q
AL K1 K2
Indications
All lengths of eyes if
each of the three A
constants optimized
pACD (personalized
Short eyes <22 mm
ACD)
Holladay 1
AL K1 K2
sf (surgeon factor)
Axial length 24
26mm
SRK/T
AL K1 K2
A constant
Average to longer
eyes
Haigis L
AL K1 K2
aO a1 a2
Phakic IOL
K1 K2 ACD WTW pre op
refraction back vertex
distance target refraction
A constants preloaded
on IOLMaster
according to lens
selected by surgeon
Prior refractive
surgery clinical
history method
According to the
formula to which the
recalculated Ks
are applied
AL pre op average Ks
pre-op refraction Stable
post-op refraction
Corneal vertex distance
ational Journal of Ophthalmic Practice. Downloaded from magonlinelibrary.com by 194.074.145.068 on July 20, 2016. For personal use only. No other uses without permission. . All rights res
Practical
Measurement procedure
8
ational Journal of Ophthalmic Practice. Downloaded from magonlinelibrary.com by 194.074.145.068 on July 20, 2016. For personal use only. No other uses without permission. . All rights res
Practical
most difficult measurement on the IOLMaster, although software
version 5 and the IOLMaster 500 has automated mode for ACD
with traffic light aid.
Note that the Haigis, Haigis L and Holladay 2 formulae all require
accurate measurements of ACD.
Do not measure ACD using the IOLMaster in patients with pseudophakia or aphakia because the alignment is based on reflection
from the lens structure.
White to white (WTW) and pupil diameter (PD) (keyboard
shortcut W)
The WTW is determined from the image of the iris. Ensure that the
patient keeps the eye wide open and maintains fixation on the target
light.
WTW provides useful information when an anterior chamber IOL is
required. This enables calculation of the size, not the power, of the
IOL to be inserted into the anterior chamber. The WTW provides
the distance of the visual axis from the centre of the iris (Figure 3).
References
Kojima T, Tamaoki A,
Yoshida N, Kaga T, Suto
C, Ichikawa K (2010)
Evaluation of axial length
measurement of the eye
using partial coherence
interferometry and ultrasound in cases of macular
disease. Ophthalmology
117(9): 17504. Epub 2010
May 14
Santodomingo-Rubido J,
Mallen EA, Gilmartin B,
Wolffsohn JS (2002) A new
non-contact optical device
for ocular biometry. Br J
Ophthalmol 86: 45862
Key points
Selection of the correct IOL power to meet individual patient expectations is crucial.
It is important to have knowledge of the structures measured in biometry and to be aware of average measurements of the eye.
Ensure you have relevant clinical and refractive information about the patient prior to performing biometry.
It is important to use optimized A constants for each individual IOL being applied to the biometry measurements.
It is recommended to use third or optimally fourth generation formulae to calculate IOL power. However, this decision is made by the operating surgeon.
Optical biometry Axial length measurement Keratometry Anterior chamber depth measurement White to
white measurement Optimized A constants Modern IOL calculation formulae
ational Journal of Ophthalmic Practice. Downloaded from magonlinelibrary.com by 194.074.145.068 on July 20, 2016. For personal use only. No other uses without permission. . All rights res