Using The Akta Prime
Using The Akta Prime
Loading your sample into the Sample/Super Loop and Injecting onto the Column:
1. Start a manual run and set the following parameters
Method base: min
Flow rate: 1.0 mL/minute
Pressure limit: column pressure limit
2. Load your sample into the sample/super loop by first loading it into a luer lock syringe and then
attaching the syringe to the connector on the face of the injection valve. Slowly depress the
plunger to push the sample from the syringe into the loop.
Make sure that there are NO bubbles in the syringe before you attach the syringe to the
connector.
The largest syringe volume you can use is 25 mL. If you are loading a larger volume into
a super loop, you can iteratively fill the syringe and load into the super loop.
6|Jenni fer Arr uda
To save the file make note of the manual run number. It will be displayed at the top of the screen in
Primeview.