Usermanual Nt3 Veterinary Monitor
Usermanual Nt3 Veterinary Monitor
Operation Manual
Newtech.Inc.
Information contained in this document is copyrighted by Newtech and may not be duplicated in
full or part by any person without prior writte approval of Newtech. Its purpose is to provide the
user with adequately detailed documentation to efficiently install, operate, maintain and order
spare parts for the divice supplied. Every effort has been made to keep the information contained
in this document current and accurate as of the date of publication or revision. However, no
guarantee is given or implied that the document is error free or that it is accurate regarding any
specification.
Content
1. Overview.................................................................................................................................................................1
4. Basic Operation....................................................................................................................................................10
I
Content
4.19 Recorder Setup (for details see on Printing Function................................................................................19
5. Alarm ....................................................................................................................................................................20
10. Printing...............................................................................................................................................................37
II
Content
11.1.3 Screen showing other leads channels ....................................................................................................43
11.2 Parameter display ..........................................................................................................................................43
11.3 Trouble shooting............................................................................................................................................44
12.1 Overview.......................................................................................................................................................46
12.2 Cuff Selection and Placement .......................................................................................................................46
12.3 Cuff sites on differents animals.....................................................................................................................46
12.4 Measurement limitations ...............................................................................................................................47
12.5 Display ..........................................................................................................................................................48
12.6 Functions.......................................................................................................................................................48
12.6.1 Blood pressure measurement.................................................................................................................49
12.6.2 Venipuncture..........................................................................................................................................49
12.7 Maintenance and cleaning.............................................................................................................................50
12.8 Trouble shooting ...........................................................................................................................................50
16.1 Overview.......................................................................................................................................................57
16.2 SpO2 Pleth priciple .......................................................................................................................................57
16.3 Precautions in SpO2 monitoring ...................................................................................................................57
16.4 SpO2 monitoring procedures ........................................................................................................................58
16.5 Application Guide .........................................................................................................................................58
16.6 Measurement limitations ...............................................................................................................................59
16.7 SpO2 setting menu ........................................................................................................................................60
16.8 Maintenance and Cleaning ..............................................................................................................................60
III
Content
18. Monitors Accessories and Purchase Information ..........................................................................................63
IV
Overview
1. Overview
Welcome to use NT3 Series Veterinary monitorThe main purpose of this Manual is to provide operating guide,
instrument repair and maintenance information to users. The detailed performance indexes, routine installation,
operation and maintenance method as well as safety information are illustrated in this Manual. Before use, the
user shall carefully read this Manual in order to properly and correctly operate this monitor to ensure its safety
standards and performance functions.
The NT3 Series Veterinary monitor is used to monitor physiological signals: ECG, NIBP, SpO2, RESP,
and TEMP of patients in hospitals. This product has multi-parameter functions which can be selected,
combined and configured by the user according to his/her requirements. Therefore the NT3 Series
Veterinary monitor which you are using probably only has the parameters and corresponding accessories
selected by you at the time of purchase. This product is composed of main unit and corresponding
functional accessories such as ECG cable, blood pressure cuff, SpO2 sensor, body temperature sensor, and
BP hose. This product has four input/output interfaces for printer, network communication, external CRT
and external telephone line.
Human physiological signals (ECG, blood pressure etc.) are enlarged by each sensor connected to both patients
and monitor, transmitted through extension cable to various parameter modules, and communicated with the
main control board. The main control board displays measurement results on the monitor screen by waveform
and numeric form. The results can be printed if required. Each parameter result may be stored in a certain
timeframe.
1.3 Warranty
The manufacturer offers, since the date of the purchase, a one year warranty for the NT3 Series Veterinary
monitor, a three-month warranty for the SpO2 sensor, blood pressure cuff, ECG cable and body temperature
sensor. The consumables are excluded from this warranty.
The following situations are not included in the warranty:
1. The serial number of the monitor was ripped off or unreadable.
2. The monitor was damaged due to improper connection with other equipment.
3. The monitor was damaged due to accident.
4. The user modified the monitor without manufacturers written authorization.
The monitor is equipped with an internal lithium battery which supplies power during transportation or in a
situation where no AC power is available. A fully charged battery enables the monitor to function continously
for 2 hours (under the condition of blood presure measuring for every 10 minutes).
The internal battery may discharge during long-term storage or in transportation. If this monitor is stored for
2 months, it is necessary to charge the battery for about 30 minutes before use.
Regardless normal use or power off of the monitor, connect the monitor with an AC power for 6 hours, the
battery will be fully charged.
Attention:
Only authorized service person can replace the battery. The battery needs to be replaced every
two years.
For protect environment, Callback or deal with the exhaust batteries in correct ways are
necessary.
1
Overview
1.5 Safety information
The electrical safety classification of this product is class I. The symbol (electrocardiogram) indicates
CF defibrillation prevention type. The symbol indicates BF defibrillation prevention type. The marked
symbol on the Manual means: attention, refer to document attached with the machine; or
caution/warning, pay more attention to guarantee the safety of the patient, operator or equipment.
Prohibition
Do not use the NT3 Series Veterinary monitor for asphyxiation monitoring.
Do not use the NT3 Series Veterinary monitor during magnetic resonance imaging (MRI) or CT
inspection process.
Do not use the NT3 Series Veterinary monitor in an environment filled with inflammable gas and
anaesthesia gas.
Caution
The monitor should be operated by qualified person! The operator should have special medical
application skills, and operates according to the instruction for use.
NT3 Patient monitor just can be used on one patient one time
In order to make the monitor with safe grounding, the hospital must provide power sockets with
proper wiring, zero line and protective grounding in accordance with national standard. Or the
hospital bears all consequences.
Do not open the case of the monitor at will for possible electric shock. The maintenance and
upgrade of the monitor shall be carried out by qualified and authorized service personnel.
Keep the monitor dry from water and humidity, and avoid strong collision.
Because parts of the equipment are not provided with protections against burning, the monitor
can not be used with electricity surgical equipment, in case of burn to endanger patients lives.
When using this monitor, the F-type application section can not be connected to other conductivity
or ground.
In order to avoid time loss for diagnosis or treatment, configure adequate alarm settings according
to different conditions of each patient (for detailed setting methods, refer to the Alarm System in
this Manual).
When various equipment are connected to the same patient, pay attention to the danger of any
leakage.
Do not put the monitor in an environment such as high temperature, high pressure, gas
fumigation or liquid immersion. Unplug the monitor before cleaning or sterilizing.
2
Overview
Use the accessories recommended by the manufacturer to avoid possible mal-function of the
monitor.
No requirement for warm-up time before the monitors and sensor be used.
1.6 Functions
The NT3 Series Veterinary monitor can be used to monitor major parameters such as electrocardiogram
(ECG), body temperature (TEMP), respiration (RESP), saturation oxygen (SpO2), and noninvasive blood
pressure (NIBP). It integrates parameter measurement, display function and output function to form a
compact and portable monitor.
3
The equipment installation
z Ground
The installation place should be suitable for Monitors size in a hospital.
Monitor should be placed stable with appropriate height so it will not hurt people in case it falls.
There should be power with ground line within the length of Monitors power cable.
z Working Environment
Suitable for hospital use.
Not suitable for home use.
z Work condition safety
Environment temperature: 5 ~ 40
Relative humidity range: 80 %
Atmosphere pressure: 86 ~ 106kPa
Power Voltage: AC 100V ~ 240V
Power Frequency: 50/60Hz
z Influence to other ambient equipment
The monitor can produce, use and emit radio frequency. If installed or used not according to the
instruction of the manual, it may cause interference to the nearby equipement. If unstable readings,
interruption of operation without any obvious reasons, or malfunction of the monitor occur, there may be
some interference. If any of the above happens, check the environment and locate the sources of the
inference. Then take some measures to remove the interference sources.
Conduct a switch test on nearby equipment to determine interference sources.
Change the direction of other equipment or change its installation place.
Build/increase barrier between the interfering equipment and this monitor.
2.2 Inspection
Open the package according to the marks on the box. Take out the monitor and its accessories carefully.
z Count the accessories according to the packing list
z Check the monitor and accessories for any mechanical damage.
The packaging material should be saved for future transportation and storage.
z Confirm the AC power source is in conformity with the requirements of this equipment: 100~240V
ac, 50/60Hz.
z Connect the monitor to the AC power.
z The connection with the external equipment should be in conformity with the requirements pecified
in this Manual.
z Electric balancing terminal (refer to the rear panel introduction) should be connected to the
grounding end of the public power grid or properly connected to the other protective grounding ends.
Caution:
The monitor power cable must be connected to the special socket for hospital use.
If internal batteries are installed, they must be recharged after the monitor has been transported
or stored. By turning the monitor on without using AC power, it will not work properly because of
4
The equipment installation
weak battery power. By connecting to AC power, the batteries get recharged even if the monitor is
not turned on.
There are paralle printer interface (for external printer) and network interfact (for future software upgrade) on
the rear panel of the monitor.
To connect external equipement through this machines interface, refer to EN60601-1-1 standards.
Caution:
Do not place the monitor close to the wall. Do not block the vents on the back and the sides of the
monitor.
Turn on the power switch. After about 3 seconds, the system will trigger a dee sound. At the same time, LED
will twinkle one time among red, yellow, green.It means system enters main monitoring mode with a
successful self-test. The User now can start operating the monitor.
Attention:
Check all usable monitoring funtions to make sure they function normally.
The internal batteries must be recharged every time after use to assure full power.
If you have found and proved that the functions of the monitor have been damaged or it has shown
error messages, the monitor can not be used for monitoring patients. You shall contact Newtechs
service department.
Connect the needed sensors to the monitor and the patient. Refer to the appropriate s for instructions and
requirements.
Check if the recorder is installed on the monitors left side. Refer to the printing for instructions.
5
Description of Monitor
3. Description of Monitor
3.1 Description of external appearance
The front panel of the NT3 Series Veterinary monitor is shown as Figure3.1:
Figure 3.1
(1) DISPLAY: Waveforms, menu, alarm and measuring parameters are displayed here.
(2) CHARGE INDICATOR: The built-in battery is being charged when the indicator is on.
(3) POWER INDICATOR: When lit, it indicates that the machine is turned on.
(4) SILENCE: Silence setting button. Press this button to start or stop the monitor alarm sound.
(5) ALARM: Alarm setting button. Press this button to enter alarm setting.
(6) FREEZE: Freeze button. When the waveforms are updated, press this button to freeze the displayed
waveforms.
(7) START/STOP: Start/stop NIBP measuring button. In NIBP manual mode, press this button to start
NIBP measurement. In NIBP measuring mode, press this button to stop current NIBP measurement.
(8) PRINT: Print button. When the device is equipped with a recording/printer device, press this button to
activate printout of the monitoring results. When the printer is operating, press this button to stop
printing.
(9) MENU: Main menu button. When no menu is shown on display, press this button to show the system's
main menu. Whatever status the menu appears in, exit the menu by pressing this button.
(10) ROTATING KNOB: Rotate the knob clockwise or anti-clockwise to select the menu item. Press the
knob to operate the menu item.
(11) VOLUME ADJUSTING KNOB: Rotate the knob to adjust the monitors sound volume. Rotate the
knob clockwise to increase in volume. Rotate the knob anti-clockwise to decrease in volume.
(12) BRIGHTNESS ADJUSTING KNOB: Rotate the knob clockwise to increase the brightness. Rotate
the knob anti-clockwise to decrease the brightness.
Reminder:
Only NT3A, NT3C have brightness adjusting knob. NT3B, NT3E, and NT3F do not have it.
The side panel of the NT3 Series Veterinary monitor is shown as Figure 3.2
6
Description of Monitor
Figure 3.2
1 Temperature sensor 1
2 Temperature sensor 2
3 NIBP cuff connector
4 SpO2 probe connector
5 ECG lead connector
Figure 3.3
Vents
Network interface - connect to the central monitor
Printer interface
CRT interface
AC socket
Voltage balance interface with symbol. When the monitor is used in conjunction with
other equipment, use this to balance electrical voltage.
Telephone interface
3.2 Display
The display of this monitor is divided into four major areas: channel display area, parameter display area,
status bar and menu bar. The status bar is displayed on the top of the screen. The channel display area is under
the status bar on the left of the screen. The menu bar is always displayed at the bottom of the screen. The
parameters are always displayed on the right of the screen. The menu bar is a pop-up one, when poped up, it
will block part of the channel display area and parameter display area. The basic layout of the display is as
7
Description of Monitor
following:
The display content at the channel display area can be set by the operator. The parameter display area can be
automatically adjusted according to the channel configuration format, so that each parameter can be equally
displayed at the right part or lower right part of the screen along with corresponding waveform channels.
Channel display area The channel display area is divided into several channels. Total seven channels
are provided. The content of the channel can be selected to display ECG
waveform, RESP waveform, blood Pleth waveform, trend map, and trend table.
The name of each channel is usually displayed on the left corner of the channel.
Select one channel to enter its menu. Displayed on the same row of the channel
is the status information for this channel.
Parameter display area The heart rate (HR), oxygen saturation (SpO2), respiration rate (RESP), body
temperature (T1/T2) and noninvasive blood pressure (NIBP) are displayed in this
area. Every parameter has a corresponding label. The important parameters are
configured with alarm prohibition signs. The prohibition signs are displayed on
the left side of the label. The operational menu of each parameter can be
accessed by selecting parameter label.
Menu bar When no menu is displayed on the screen, the MENU, the ALARM menu and the FREEZE
menu can be activated and displayed in singe line, the relevant menus can also be activated by
selecting the labels under the display with the rotating knob.
Status bar
Information bar: Display the Monitor status information, including patients and alarm information.
Patient type: HORSE, CAT, and DOG.
Date: Its display format can be modified through menu.
Time: Its display format can be modified through menu.
The alarm status icon: Alarm is allowed, Alarm is prohibited.
The alarm silence status icon: Audio alarm is allowed, Audio alarm is prohibited.
Audio alarm is paused 30s, 60s, 120s respectively. Pause or resume alarm sound is
carried out by silence key, turn off alarm sound is carried out by menu<alarm sound/off>.
Central monitoring network function status icon: The red icon means the monitor is not connected to
the central monitoring system. The blue icon means the monitor is connected to the central
monitoring system.
Alarm information: Parameter name > high limit, or parameter name < low limit. The font color display: The
high, middle and low priority alarm display as red, yellow, yellow characters separately.
Memory icon: Blinking means the system is memorizing.
Caution:
When the symbol shows, the system can not produce alarm sound, so the operator shall be
8
Description of Monitor
especially cautious to use this function.
When the symbol appears, the system will prohibit alarm. It is recommended
that the operator does not use alarm forbitten function unless it is necessary.
9
Basic Operation
4. Basic Operation
4.1 How to use rotating knob
At the main display, there is a framed label which indicates your current position. You can select different
parameter names with the rotating knob, or you can click at the current parameter name to prompt the
corresponding functional menu.
By turning the volume knob you can adjust the volume. Turn anti-clockwise, the volume decreases; turn
clockwise, the volume increases.
3. Press can switch the character group in the text frame , there are four
groups switchable, capital letters group(A-Z), small letters group(a-z), Arabic numerals
group(0-9), punctuations group.
4. Select or , can page the characters in the text frame .
5. Select can insert character at the cursor in password frame.
6. Select , can delete the characters before the cursor in password frame.
7. Move cursor to , rotate the mouse and select the password characters you
want to input one by one, password you input will appears in password frame, press
Caution:
Change date or time will influence the memory of the trend and events. After
modification of time or date, you must restart the Monitor!
Select Menu -> System Setup -> Set Time
10
Basic Operation
4.6 Set language
The monitors can supply four languages: Chinese, English, Spanish, Portuguese
When you can not clear about display present language, Please press MENU -> Press the fifth menu frame->
Press the third menu frame -> Press the fourth menu frame, the optional window is as above,Select the
language you need, press [OK].
If it is not the language you need, please consult Newtech or the local Newtech representatives.
Select Menu -> System Setup -> Region Settings -> Select Units
Decrease the waveform speed, the waveform will be compressed to show longer time quantum. Increase the
wave speed, the waveform will be expanded to display more detailed view. The user can change ECG
waveform speed and RESP waveform speed.
The common monitoring formats include Standard 7 ECGs, large numeric display, and five formats defined by
the user. The detailed information for the format required, refer to the format setting.
11
Basic Operation
4.11 Remote maintenance
Select Menu -> System Setup -> Maintenance -> Remote maintenance
Enter Remote Maintenance mode, the Central Management System can remotely upgrade the Monitor.
The Monitor provided with the second cofirm information, when you would like to accessing the Remote
Maintenan.
Start menu:
Press corresponding menu keys (MENU, FREEZE, ALARM), the corresponding menu will pop up.
Select the concerned waveform parameter label on the screen using the rotating knob. Press
down the knob, the corresponding menu pops out.
Menu browse:
In the menu bar, the field with black background and white character is the field selected by the cursor. Using
the rotating knob to select the required field, then press it down.
Menu status:
In the menu bar, there are three different menu colors corresponding to three different menu status:
1 The menu field with black background and white character means the current operating menu.
2 The menu field with green background and black character means the menu to be operated.
3 The menu field with gray background and black character is an inactivated field. It means that current
field is forbidden to operate or has not been configured.
Return to the previous menu:
Select exit at the far right end of the menu bar to return to the previous menu.
Press <Menu> key to exit all menu.
If no key is pressed for more than one minute, the Monitor will automatically exit all menus.
<Fixed Format> set the display channel as 1ECG, 2ECGS, 3ECGS, 7ECGS, or large numerics.
12
Basic Operation
z 2ECGS
z 7 ECGS
z Large numerics
<User Format> sets the display channel as the channel format defined by the user.
It supports five user-defined channel formats.
<Format Setup> define the display channel.
13
Basic Operation
Attention:
Select the ECG channel, the system will automatically configure the lead of the ECG channel.
The configured channel can be saved as a user-defined format, and can be selected later at any
time from <User Format>.
<Waveform Speed> adjust waveform speed of the ECG / PLETH and RESP.
14
Basic Operation
<Other Settings> set other display formats.
z Menu font can be switched between large font and small font.
z Screen Font can be switched between large font and small font.
<Trend> used to review trend data. Refer to sand sections on Trend Setting.
<Wave Review> used to replay waveform. Refer to s and sections on Trend Setting.
<Alarm Review> used to review alarm memory records. See alarm review.
15
Basic Operation
4.17 Patient menu
<Set Sound> sets up volume of current alarm sound and pulse sound, switch on/off pulse sound.
Click <Device Tone> and enter to set up Pulse Tone and Alarm Tone:
z Select Units
16
Basic Operation
z Date Format
z Power Frequency
Attention:
Set appropriate power frequency, otherwise the signal noise of ECG can be loud.
z Language
<Network Setting>
z IP Address
The user can set the IP address of the Monitor. Refer to the s and sections of the network settings.
17
Basic Operation
The Menu maintenance is as follows:
z System Status
Attention:
18
Basic Operation
The user should pay more attention to the following operation functions:
z Remote maintenance: used to upgrade the Monitors software;
z Setup password: password protection;
z Reserved 1, Reserved 2;
z Demo Switch: Turns on the demonstration display.
19
Alarm
5. Alarm
5.1 General alarm
This monitor includes physiological alarm, technical alarm, and normal alarm.
z Physiological alarm happens when patients physiological parameters exceed limits (e.g. heart rate and
blood pressure exceed limits).
z Technical alarm is triggered by system abort, monitor mal-function or inaccurate measurement.
Technical alarm is system error message also, e.g. plug off, sensor off, monitor abort.
z Normal alarm is within the normal range. It may not threaten the patients health. This is
different from the first two alarm types but needs attention.
z High level alarm: When a patients life is threatened or the monitor appears to have serious technical
problems. This is the most serious alarm. For example: Leadwires off, sensor off, the pulse search
failed, the patient suffocates, or the alarm exceeding limit of related parameters is setted as upper
alarm.
z Mid level alarm: This is a major alarm. For example: low battery, the alarm exceeding limit of
related parameters is setted as median alarm.
z Minor alarm: This is a general alarm. For example: the alarm exceeding limit of related parameters
is setted as lower alarm.
Attention:
Only doctors can set physiological parameters upper limits for alarm level according to patient's
conditions and his/her own experiences.
NT3 series have both audible and visual alarm modes. Visual alarm is displayed on the monitor screen with
appropriate blinking numerics. Audible alarm is through the monitors speaker.
1. High priority: Two groups of five Dee with an interval of 10 seconds. Numerics blinking frequency:
1.5Hz.
2. Medium priority: Continued three Dee with an interval of 18 seconds. Numerics blinking frequency:
0.625Hz.
3. Low priority: one Dee with an interval of 20 seconds. No blinking.
Attention:
When different priority alarms occur simultaneously, only the highest priority alarm is displayed.
Monitor can set overall alarm functions, as well as separate parameters alarm function. Each individual
parameter has alarm function and alarm limits can be set. The user may select the parameter field by the
rotating knob and set alarm limits, press it down and follow the menu.
Overall alarm setting: Presses down ALARM button, choose alarm menu to carry on the setting.
20
Alarm
Partial parameter alarm setting: Select some parameters, press down the knob, choose [alarm switch], [alarm
setting] menu to carry on the operation.
Attention:
When overall alarm is switched off, if physiological parameters exceed limits, there will be no
alarm sound and no blinking numerics. Yet the words relating to technical alarm will blink, but no
audible sound.
Warning:
User should pay serious attention to the overall alarm switching. It is recommended not to turn off
this switch.
The below chart shows alarm prohibition icon in a red circle.
When the overall alarm is switched on, there is no need for confirmation. When the monitor is turned on for the
first time, the default alarm switch is ON. Restarting this monitor, alarm setting remains the same as when the
monitor is turned off.
21
Alarm
Warning:
User need to pay serious attention to use switch alarm sound function. It is recommended not to
switch off overall alarm sound.
Select ALARM key -> Alarm Sound/off
When switching off alarm sound, a confirmation dialog bar will appear as a reminder.
On the right of the status bar, different icons indicate current status of alarm sound.
These represent five alarm sound status: On, Off, Interval 30s, interval 60s, Interval 120s. To switch on or off
alarm sound, go to menu [Alarm Sound/off]. To pause alarm sound, use Silence key.
When alarms occur, they are displayed on the left side of the status bar.
z Alarm format: parameter > high limit, or parameter < low limit.
z Alarm level colors: High = red, Medium = yellow, Low = yellow words.
z Alarm location: in the status bar. When an alarm occurs, parameter name >high limit or parameter
name <low limit replaces parameter name. When an alarm disappears, the parameter name returns.
z If multi alarm occurs, alarm parameters will be displayed alternately on the left side of the status bar.
From alarm menu, enter [Alarm Settings] to configure all parameter alarm high and low limits. On
the current parameter, press knob, enter [Alarm Settings] to configure current parameter alarm
high and low limits.
z Set all parameter alarm
Select ALARM key -> Alarm Settings
22
Alarm
See attachment "Product Technical Specifications" for the range of various parameters alarm limits and factory
default value.
Attention:
Alarm high limit cannot be lower than low limit. If high limit is lower than low limit, the monitor
will not respond.
When the monitor is turned on for the first time, it has adults alarm configuration as default.
If the monitor restarts after 30 minutes, its alarm limits are for the current patient. Otherwise its
alarm limits remain the same as when the monitor is turned off.
If [patient information] has been reset, the user should reset alarm limits based on new patient
type.
Alarm control is used to shield and set priority sequence when all parameter alarms occur.
Select ALARM key -> Alarm Control
z Alarm switch: If the single parameter alarm switch is set as Off, this parameters alarm sound and the
blinking is shielded. Once the overall alarm switch is set as Off, any single parameter alarm switch will not
be active. Alarm printing is also off.
z Print switch (with printer option) can be set as On or Off. Once overall print switch is set as Off, any single
parameter printer switch will not be active.
z Priority sequence based on patient conditions can be set as high, medium, or low. But HR and SpO2 cannot
be set as minor alarm.
Alarm priority chart:
Alarm content Type Priority
HR Over limit above medium
SpO2 Over limit above medium
ST Over limit Set different priority
RR Over limit based on patient
NIBP Over limit conditions
IBP Over limit
Temperature Over limit
23
Alarm
Adjust alarm tone by MENU> System Setup > Set Sound > Device Tone
Adjustment range for alarm tone: 400 Hz ~ 2950 Hz. Factory default: 900 Hz.
24
Alarm
Returning to default setting applies to all parameters, alarm limits, alarm switch, alarm sound and
all other related values. There are three sets of default values for alarm limits: adult, pediatric,
and neonate. See Product Specifications for detailed alarm default values for all parameters.
Alarm records can be arranged by alarm time sequence. Press Line up, Line Dn, Page up, or Page Dn to review
more records.
Parameters' name and definition
Name Definition
HR Heart Rate
ST_I ST segment
ST_II
ST_M
SBP(NIBP) Shrink Blood Pressure
MAP(NIBP) Mean Average Blood Pressure
DBP(NIBP) Diastolic Blood Pressure
T1 Temperature 1
T2 Temperature 2
SPO2 SpO2
RR Respiration Rate
ET et CO2
INS ins CO2
Warning:
When an alarm happens, check the patient's conditions first.
25
Alarm
When there is an alarm sound from the speaker, the user can not determine which physiological
parameter or which type of alarm is occurring. The user must look at the display and the blinking
numerics to judge the alarm type and reason. The following are some actions to take:
1. Check the patients condition
2. Identify which parameter is alarming or which type of alarm is occurring
3. Identify the alarms reason
4. Silence alarm when necessary
5. Check if alarm is off after the alarm status has been adjusted.
26
Trend Analysis
6. Trend Analysis
Trend data is a patients data collected during a period of time. Trend analysis is patient condition chart
displayed based on the trend data. The trend map or trend table can be prepared according to the parameters
selected by the user. Except noninvasive blood pressure, all trends are consecutive curves. The trend map (table)
can provide time zoom and range zoom for detailed analysis of the patients conditions.
2. Parameter selection menu: select any parameter, click OK. The corresponding trend map will be displayed
in the trend map display area;
3. Time frame of the trend map: click [Page Up], [Page Dn];
4. Interval: as shown below: the image will be compressed or expanded according to selected intervals.
27
Trend Analysis
5. Range: as shown below, it is used to control the Y axel value range of the trend map. The adjusted range
will be stored in the Monitor. It will be applied when the monitor restarts. The range has three adjustment
modes: alarm limit range, maximum range, and manual adjustment. Once the user sets the range, the trend
data takes the upper limit/lower limit according to this range. Any value over the limit is not valid.
2. Adjust the time frame of data: use [Line up], [Line Dn], [Page Up], and [Page Dn];
3. Adjust the intervals: as shown below: the data in the trend table is compressed or expanded according to
the desired intervals. The new data in the trend table will be added accordingly.
Enter the history database from the system menu, click the trend display. The trend window can show analysis
of one parameter or several parameters via the trend map or trend table. The user can expand the time frame
and adjust the value range manually.
1. Select the parameter: The user can select up to three parameters combination in the menu. The selected
parameter is highlighted in the parameter table. Click the selected parameter again will reverse the
selection. Click [OK], the corresponding trend map will be displayed in the trend map display area.
2. The map time frame: Using Page Up, Page Dn, the user can browse the trend map back and forth until the
required time frame is shown on the screen.
3. Cursor: Click this knob and rotate it to move the cursor. The time value and events where the cursor is will
be displayed in the parameter table.
4. Intervals: Select any time interval as shown below. The trend map will be compressed or expanded
according to the selection and displayed in the trend map display area.
5. Range: As shown below, this is used to control the Y axel on the trend map. The adjusted range will be
saved in the Monitor and applied in the next start up. There are three areas: alarm limit range, maximum
range, and manual adjustment for the use to choose. The trend data will take this range as the upper limit
or lower limit value. Any value above or below the range will be invalid.
Alarm Limit: The alarm limit range set by the user will be used for the value limit range.
Max Range: The default range of this Monitor serves as the value range.
Manual Adjust: Used to adjust upper/lower limit range. Its value range is limited by the maximum range.
6. Trend table
The information in the trend table is displayed as a list. The parameter number and category are the same as the
trend map. In the trend table, the median time is the cursor time value in the trend map. The value of each
parameter is date and time (except NIBP). The NIBP value is the first value in the current interval.
Menu description:
29
Trend Analysis
Adjust data display time frame: Use [Line Up], [Line Down], [Page Up], [Page Dn] to set the time frame.
Adjust time intervals: Select one interval for the time frame of each data in the trend table.
30
Freeze Waveform and Replay
Save waveform: Press [Save] to save the frozen waveform section to replay it later. If the waveform is
successfully saved, the following appears:
Press [Wave Review], the previous frozen waveform is displayed on the screen.
Browse waveform: Use [Page Up] and [Page On] to review different waveforms.
31
Freeze Waveform and Replay
8. Record Events
Events can be recorded for future reference. Use [Tools], click [Event], the following appears:
32
Drug Calculator
9. Drug Calculator
The information of the drugs given through intravenous infusion includes dosage, infusion speed, volume, and
concentration. The drug calculator function can help calculate unknown value by the known formula and
therefore help control drug infusion.
Caution:
Before applying any drug, correct calculation unit and patient weight should be considered. If
there are doubts regarding these, consult the hospital pharmacy.
It is always a doctors decision to determine the type of drug and its dosage to be applied to a patient. The drug
calculator will calculate the dosage based on the entered value. It can not verify the validity of the calculated
data.
Drug unit
When the drug name is any drug, you can select: g, mg, mcg, unit, k unit, m unit, mEq.
Once the drug name is identified, the unit is set by the drug calculator automatically. The user cannot
modify it.
After the drug measurement unit is set, the weight unit in the drug calculator changes accordingly as well
as the pace of the infusion process.
Time unit
Min (minute) and hr (hour) are used for time unit. After setting the time unit, the unit in the drug calculator
changes accordingly as well as the pace of the infusion process.
9.3 Glossary
z Total drug volumn: The total weight of the drug used by a patient within a certain time period.
z Liquid volume: The volume of the solution including the drug.
z Dosage/min or dosage/hr: The drug quantity injected into a patient within unit time (per
minute/hour).
z Dosage/(kg*min) or dosage/(kg*hr): The amount of drug per 1kg of a patients weight injected
in the patient within the unit time period (per minute or hour)
dosage/(kg*min) * patient weight dosage/min
dosage/(kg*hr) * patient weight dosage/hr
z Infusion speed: The volume of the drug solution injected into a patients body within unit time
(per minute, hour). The unit is ml/min, ml/hr.
z Infusion speed/kg: The volume of the drug solution per 1kg of a patients weight injected into
the patient within unit time (per minute, hour) , the unit is ml/min, ml/hr.
infusion speed/kg * patient weight transfusion speed
33
Drug Calculator
z Time: The consumed time for drug infusion (the unit is min, hr).
z Concentration: The concentration of the drug in the solution.
concentration total drug / liquid volume
z guttae speed/ min, guttae speed/ hr: The infusion within unit time
z guttae/ml: The volume of each guttae dropped form the infusion device, the unit is GTT.
a) Entered numbers
z To enter numbers, the cursor selects the first digit from the left. Turning the rotating knob the
cursor will select each digit consecutively from left to right. When it reaches the last, it will
jump back to the first digit repeatedly. Press the rotating knob and enter the digit selection
status.
z The selection range for each digit is 0 - 9.
z When the digit is larger than the display value, it displays ---.--. When it is smaller than the
display value, it displays 0.00.
z All digits after the decimal point will be rounded up.
b) Calculation rules
z The equation for drug volume:
total drug=dosage/min* duration
dosage/min=dosage/(kg*min)*patient weight
z The equation for solution volume
liquid volume=infusion speed*duration
Infusion speed=infusion speed/kg*patient weight
z concentration=total drug/liquid solution volume
c) Known value and calculation results
z Total drug, dosage/min, dosage/(kg*min), volume, infusion speed, infusion speed/kg, time ,
concentration can be entered as known items or be put out as a calculation result.
z The user can enter at least three known items according to the calculation requirements. The
drug calculator will automatically calculate using the default equations.
z The known items entered by the user are expressed with blue background. The calculation result
is expressed with gray background.
z All fields are initially displayed as 0.00 and can be filled with value. After the user inputs the
known item, the drug calculator start calculating and the result will be displayed in real time.
The calculation result is then locked. The user can only modify the entered value. To put in 0.00
means to cancel this entry.
z Pressing calculate knob returns to the initial status of the entry.
34
Drug Calculator
e) Conduct calculation for certain drug
z The drug calculator has been preset to do calculation for more than ten types of general drugs,
e.g. Aminophylline and Amrinone Lactate (see Drug Dosage Range Limitation Table). Once a
drug name is selected, the drugs relevant value such as concentration, total drug amount,
dosage/min will have corresponding limitation ranges. When any information entered or any
calculation result is out of range, the drug calculator will show it in red color.
z The drug dosage over range prompted from the drug calculator can only warn the user to pay
attention to apply appropriate dosage of the current drug. The final infusion dosage and infusion
process should be determined by the physitian in charge.
The Titration Process Table allows the user a quick glance of how much liquid has been infused into the
patients body, and how much time remains.
z The titration process table displays the remaining amount of the drug and liquid at each time
frame after the user inputs the data into the drug calculator.
z Select titration process table to enter this function.
z The left side of the screen shows the data (and calculation result) entered by the user. The right
side is equally divided into 15 segments, according to the total infusion process, listing the total
drug amount and liquid volume for each segment.
z All the fields in this window can not be modified.
z The unit for the total drug amount is the same as the unit entered at the drug calculator main
window. The liquid volume unit is ml.
z Press OK to exit the table and return to the drug calculator main widow.
35
Drug Calculator
At a glance the user can see the drug dosage accepted by the patient at different infusion speeds. The higher the
infusion speed, the larger the gaps between the segments of the table.
z The titration table shows the relationship between the dosage/min and the infusion speed under
the condition of the same liquid concentration.
z Select the titration table to enter this function.
z The left side of the screen shows the data (and calculation result) entered by the user. The right
side is equally divided into 30 segments showing the relationship between the dosage/min and
infusion speed.
z There are two fields in the list: dosage/min (in the case where the unit selected is hr, then this
field will be dosage/hr) and infusion speed.
z The range used in calculating the Dosage/min (dosage/hr) is one to two times of the value
entered by the user in the drug calculator at the beginning.
z Press OK to return to the drug calculator screen.
9.7 Recalculation
The drug calculator result remains in the monitor when the user exits this function or turns off the Monitor.
When the user enters the drug calculator function again, the latest data appears. If the user wants to start a new
calculation, press Recalculation in the drug calculator screen which clears all existing data and a new
calculation starts.
36
Printing
10. Printing
The printing function is optional with NT3 series monitors. The user needs to determine whether to include this
function which prints one real time waveform of the main lead ECG (see other s and sections for the main lead
definition) or two real time waveforms of the main lead ECG and blood Pleth volume. This function has two
printing speeds: 25mm/s and 12.5 mm/s. Timer printing and alarm printing are also available.
The printer used for this Monitor is a thermal printer which is installed on the side of the Monitor. Do not
install any other types of printers which may cause damage to the monitor. Contact the manufacturer
immediately for any questions.
If the printer is properly installed, the printer indicator will be green when the monitor is turned on.
If it is red, it means that the printer is without paper. Do not print without paper, which may cause
damage to the printer.
The printing paper is single sided thermal printing paper. Press printer knob to open the cover.
Insert the paper with the smoother side up and close the printer door. Tthe red error alarm indicator
should be off.
Check to see if the printer is properly connected. If not, the printer can not be normally detected
will appear.
Curve mode is used to select the number of curve channels and content to be printed. Mode 1 prints one
main lead ECG, while Mode 2 prints two waveforms: the main lead ECG + blood Pleth waveform.
The user can select printing speed: 25mm/s, or 12.5mm/s.
Total length is the total printing length after print is pressed. The unit is second. The minimum length is 5
seconds. The maximum is 30 seconds.
Alarm printing is to print all parameters when alarms occur, including heart rate, SpO2, blood pressure,
respiration rate and body temperature.
Timer printing prints only parameters. The default setting for both alarm printing and timer printing is
off.
Caution:
When it is time to perform a timer printing but the printer is printing, the Monitor will dismiss
this timer printing.
37
Printing
10.3 Printing
Press print to start printing. It will stop when it finishes. While the printer is printing, press print again to
stop the current printing job.
Caution:
When the thermal recorder is printing, the user should not open the printers cover. Only after it
stops printing, the user can retrieve the print-out. The printing paper needs to be installed before
restarting the recorder. Never start printing without printing paper.
38
Monitoring ECG
Warning:
Do not use the ECG-cables, which do not belong to our company, or it will lead the safety problem
in defibrillating
Using five-electrode ECG cable can obtain up to 12 selectable ECG leads. The Monitor displays a patients
ECG waveform from those pre-selected, effective leads.
The ECG cable includes two parts: the main cable connecting the Monitor and the electrode leads connecting
the patient.
1) Selection and use of electrodes
The electrodes use for monitoring ECG are disposable electrodes made from Ag-AgCl (silver - silver chloride)
with the disposable electrode mark: . Before use, make sure the electrode has valid expiration date.
Normal electrodes should enable the monitor to display and maintain data in 5 seconds after the defibrillation
electric discharge. Defected electrodes can lead to inaccurate data.
2) Skin preparation
Because the skin is not a good conductor, it needs to be treated before affixing the electrodes.
1 Completely clean the skin with soap and water, but not with aether or pure alcohol which will increase
skin resistance.
2 If necessary, shave the body hair where the electrode is to be affixed.
3 Rub the skin to accelerate blood flow and get rid of the dead cells and greases on the skin.
4 Install the electrode on the patient. Apply some conducting gel first if the electrode comes without it.
5 Connect the electrode lead with the patient cable.
6 Make sure the Monitor is power on.
7 Make sure the electrode, wire leads, and the trunk cable are not connected with any other conductor
parts including the ground.
(1) For the installation of the fivelead ECG electrodes, refer to the ECG electrode installation as shown in
Figure 11.1
Installing electrode for five-lead device
39
Monitoring ECG
Attention:
If there are cardiac pacemaker inside the patients body, Montior will count the packmaker pulse
as the normal ECG wave, it caused the ECG signal of patient can not be measured.
The label name and color of the electrode is American standard. The square brackets [-] indicate
European standard.
Caution:
40
Monitoring ECG
Select a place with stable ECG signals or little activities to place the electrodes.
The electrosurgical unit (ESU) and non-grounded equipment near the patient may disturb
waveform.
Never put the electrode near the grounding of a surgical electrocauterize unit, which will cause
huge interference to the ECG signals. The electrodes should have equal distance to the
electrocauterize blade and its ground.
The electrode should be affixed tight to the skin to ensure proper conduction.
Caution:
For the five-lead configuration, place chest (V) electrode in the above mentioned position.
Check daily to see if the ECG electrode patch irritates the skin. If irritation occurs, replace the
electrode or change the electrode position every 24 hours.
Always make sure all leads are nomal before start monitoring. If the ECG cable is not connected,
the screen will show lead off.
Attention:
To protect the environment, all used electrodes must be recycled or disposed properly.
The ECG waveform is green including two ECG channels. The top ECG channel is the main channel as shown
below:
Waveform speed scale scale range lead off sign ECG waveform
Select position with the cursor, press the rotating knob and the ECG menu appears.
lead selection: To switch lead type, press the knob. There are seven lead types: I, II, III, aVR, aVL, aVF, V
for selection. The main lead is the most important lead. To get accurate heart rate and gain
control performance, the doctor should select the lead with large range and little noise as the
main lead.
Caution:
To avoid duplicate leads, when switching the lead type in the current channel, the lead
41
Monitoring ECG
types in other channels will be changed automatically.
gain mode: To switch between automatic gain and manual gain, press the knob once.
Gain means to magnify the ECG signals. This monitor has five magnifying stages: x1/4, x1/2, x1, x2, and x4.
x1 is one time magnification. Under this magnification, the range of the scale at the left side of the ECG
waveform is 1mV. Under the standard waveform display format, the length of the scale is 10mm. Then
under the condition of x2 magnification, the waveform of the 10mm displayed on the standard screen will be
for the ECG signal of 0.5 mV. Below is a detailed table:
X1 adopted applicable 1 mV
Caution:
This Monitor provides two methods to regulate the electrocardiogram waveform range:
Automatic mode: In automatic mode, the monitor will automatically regulate the gain to amplify the ECG
waveform from the main lead without possible distortion. The automatic mode can be slow.
Using this mode, the manual mode forgain regulationis not activated.
Manual mode: In manual mode, the monitor will not automatically regulate the ECG gain value. The ECG
gain is regulated via thegain regulationmenu. This mode generates fast response. The
waveform will change immediately after the gain is regulated, and can be amplified with
some distortion.
gain regulation: To manually regulate ECG gains, press down the rotating knob. Rotate anti-clockwise, the
gain and waveform range becomes smaller. Rotate clockwise, the gain and waveform range becomes larger.
After the regulation is done, press down the rotating knob again.
measurement mede: The measurement mode refers to the filter bandwidth of the ECG channel. Under
diagnostic mode, the filter bandwidth is set to the widest, which can get the detailed ECG signals. This
detailed information helps to accurately interpret the status of the ECG signals. Yet the detailed information
may be accompanied with some ambient noise, e.g. noise of the HF electrotome. The noise is mixed with the
real ECG signals; therefor the true ECG signals can be difficult to distinguish. To accommodate noisy
environment, the monitor provides two measurement modes: the monitoring mode and the theater operation
mode. In these two modes, narrower bandwidth is used to obtain smoother signals. The user can choose for
appropriate situations. See below noise comparison table:
42
Monitoring ECG
Caution:
1mV standard signal: Used for the square wave signal 1Hz 1mV.
Waveform speed: Used to regulate the waveform speed for parameters relating to the heart, including all
ECG channels and blood Pleth waveform channel. There are four speeds: 6.25 mm/s,
12.5 mm/s, 25 mm/s, 50 mm/s.
ECG parameters include the heart rate and ST section. Except in large numeric mode, the heart rate and ST
sections are displayed on the right corner of the screen. See below.
ECG waveform heart rate and main lead ST alarm limit heart rate alarm off sign heart rate label
heart rate source heart rate main lead ST section other lead STsection ST section label
Move the cursor to the heart rate label, and press down the rotating knob to use heart rate setting menu, as
shown below:
43
Monitoring ECG
AVRG.: Used to set up calculation average periods of the heart rate and ST section. Press down the knob
and the following menu appears:
Label The color of the heart rate value Heart rate source mark Source
heart rate green I, II, III, aVR, aVL, aVF, V Form ECG signal
Attention:
Pulse rate and heart rate share the same set of alarm limits and alarm on/off.
Warning:
lead off ECG electrode fell off from the patients Ensure the electrode lead and cable
body or the ECG cable fell off from the are properly connected.
monitor.
ECG signal noise too The measurement mode is not correctly set Change the electrocardiogram
loud. according to the environment. measurement mode.
The power frequency is not set in Set right power frequency. Refer to
accordance with the local power frequency. Three.
44
Monitoring ECG
The monitor is not grounded properly Check the grounding system of the
monitor.
The electrode is not connected properly to
the patient Keep the patient stable to obtain
reliable connections of the electrodes.
No heat rate display The ECG signal is very weak, <0.25mV. Check the connection between the
electrode and patient to ensure
No heat beat sound The electrode is not connected to the patient
proper contact of the electrode and
properly.
patients skin.
The patient type is not set correctly.
Correctly set the patient type.
The gain is not enough.
If it is in manual gain regulation, tune
the gain to the right magnification.
45
Monitoring NIBP
For a CAT
For a conscious patients, measurements from the coccygeal artery can be taken by wrapping the cuff around
the base of the tail. For anesthetized patients, measurements from the median artery on the foreleg can be used
by wrapping the cuff around the forelimb, between the elbow and carpus. For cats less than five pounds when
measurements are difficult to obtain, place the cuff around the leg above the elbow to obtain measurements
from the brachial artery. Hair need not be clipped except when heavily matted.
Concerning measurements in dogs, its preferable to use the right lateral, stemal or dorsal recumbent position.
If the dog is in a sitting position, place the front paw on the operators knee and take the measurements from
the metacarpus.
The metacarpus, metatarsus and anterior tibial are recommended for the cuff placement. For
anesthetized patients, most surgeries are done on the posterior part of the body so that the
metacarpal area of the forelimb is most convenient In situations where this is not possible, place
the cuff around the metatarsus just proximal to the tarsal pad or around the hind leg next to the
hock. For conscious patients, measurements from the coccygeal artery can be used to over tail
site.
46
Monitoring NIBP
Its preferable for a large animal, such as a horse and cow, to be in a stock, standing still.
Measurements from the coccygeal artery on the ventral surface may be used by placing the cuff
around the base of tail.
Caution:
Extremity and cuff motion should be minimized during blood pressure determinations.
Proper blood pressure cuff size and placement are essential to the accuracy of the blood pressure
determination.
Any blood preesure recording can be affected by the position of the patient, his or her physiologic
condition and other factors.
Use only the accessories approved by the manufacturer to avoid any malfunction of
the monitor.
The width of the cuff should be either 40% of the limb circumference or 2/3 of the
upper limb length. The inflatable part of the cuff should be long enough to circle
50-80% of the limb. The wrong size cuff can cause erroneous reading. If the cuff size is
in questions, use a larger cuff.
Dont apply the cuff to a limb that has an intravenous infusion or catheter in place.
This could cause tissue damage around the catheter when infusion is slowed or
blocked during cuff inflation.
Make sure that hoses are not kinked, compressed or restricted.
The limb chosen for taking the measurement should be placed at the same level as the
patients heart.
If the animals hair over the artery site is too thick or matted for good contact, it
should be clipped.
The tube between the blood pressure cuff and the monitor should be unobstructed
and cannot be tangled.
The blood pressure measurement of this monitor uses oscillometric method. This method looks for a regular
artery pressure pulsation. If a patients conditions are not applicable to use this method, the measurement may
become unreliable and the time needed for measurement will be increased. The following situations may have
some interference to this measurement method, which can make measurement become unreliable or a longer
measurement time. If this happens, measurement can become impossible.
47
Monitoring NIBP
Patients movements: If the patient is moving, shivering or jerking, it will make the measurement result
unreliable or even impossible. These conditions may interfere the artery pressure
pulsation measurement, and prolong the measurement time needed.
Arrhythmia: If the patient has irregular heart beat due to cardiac arrhythmia, the measurement result will be
unreliable or impossible. The measurement time will be prolonged as well.
Heart-lung machine: if the patient is connected to a heart-lung machine, the measurement can not be carried
out.
Unstable blood pressure: if the patients blood pressure changes rapidly during certain measurement
process, the measurement will be unreliable or can not be conducted.
Severe coma: If the patient is in coma or with very low body temperature, the measurement will be unreliable.
The insufficient blood flow to the periphery can lead to the decrease of the artery pulsation.
Extreme heart rate: when the heart rate is below 30bpm (heart beat /minute) or over 300bpm, the
measurement cannot be carried out.
Obese patient: The excess body fat level can reduce the accuracy of the measurement, because the fat can
block the artery pulsation to reach the cuff.
12.5 Display
Alarm On/Off: To switch simultaneously alarm on/off for systolic blood pressure, diastolic blood
pressure and mean blood pressure.
Alarm Setting To check and set the alarm for systolic blood pressure, diastolic blood pressure and
mean blood pressure.
Auto/Manual To switch between automatic measurement mode and manual measurement mode.
Test Interval To set automatic measurement intervals in automatic measurement mode.
Venous Puncture To enter and exit the venipuncture status.
Start TestTo start the continuous measurement.
12.6 Functions
The NIBP has two functions: blood pressure measurement and the venipuncture function.
48
Monitoring NIBP
12.6.1 Blood pressure measurement
When starting blood pressure measurement, the blood pressure cuff will be inflated. The systolic blood
pressure, diastolic blood pressure and mean blood pressure are measured through sensors. A normal
measurement process takes about 40 seconds.
There are three methods for blood pressure measurement: manual, automatic, and consecutive.
Manual measurement: Each blood pressure measurement process shall start manually. Press <Start> on
the panel of the monitor, the blood pressure cuff air pressure is set manually.
To enter manual measurement mode, press automatic/manual. When the manual displays on the
position , it indicates that it is in manual measurement mode. Position does not display any
information.
Automatic measurement: The monitor will periodically start the blood pressure measurement
according to the selected time intervals. Press automatic/manualto switch to this function. When
position displays automatic, it indicates that it is in automatic measurement mode. The position
now will show automatic measurement interval time.
Automatic measurement mode has two statuses: pause with the clock at position stopped. It means
that although it has entered automatic mode, the automatic measurement cycle has not been started yet.
With the clock at position running, it means the automatic measurement cycle has been started. The
clock at position indicates how much time still remains before next measurement. During pause status,
conducting one manual blood pressure measurement will restart the automatic measurement operational
status and the clock at position begins to run. The user can switch between automatic measurement
and manual mode. Press <start> during the blood pressure measurement can stop the measurement
immediately; this can also switch the automatic measurement operational status to pause status.
Attention:
No matter in automatic mode or manual mode, pressing <start> will begin the blood
pressure measurement.
During blood pressure measurement process, press <start>, the monitor stops the
blood pressure measurement immediately and deflates the cuff at the same time.
Consecutive blood pressure measurement
The consecutive blood pressure measurement mode is used during operations or emergencies. This mode
measures blood pressure consecutively for five minutes with a simpler method. After five minutes, the
monitor returns to the previous status (automatic or manual).
12.6.2 Venipuncture
In venipuncture mode, the monitor inflates the blood pressure cuff to the pre-assigned level and stabilizes it for
the pre-assigned time period before deflating the cuff.
49
Monitoring NIBP
The blood pressure measurement type should be appropriate with each patient type. Set correct
patient type to ensure safety. Refer to the relevant for information on the monitor setting.
Warning:
Do not allow water or detergent solution to get inside the NIBP connector on the monitor panel.
The instrument may be damaged.
When the reusable blood pressure cuff is not connected to the monitor or being washed, do not
allow any liquid to enter the tube to avoid accidental leakage into the monitor.
Do not clean the interior of the monitor. Only clean the exterior of the connector sockets.
To protect the environment, disposable blood pressure cuffs should be disposed properly.
Failure to start Monitor hardware error Stop the NIBP measurement function, inform the
supplier to dispatch qualified maintenance
personnel to repair.
Blood pressure cuff not Blood pressure cuff is not Tie up the blood pressure cuff.
connected properly tied up or no cuff
pressure cuff
50
Monitoring NIBP
Blood pressure cuff air Blood pressure cuff, bladder or Check and replace the leakage part. If necessary,
leakage connector is damaged inform the supplier to dispatch qualified
maintenance personnel to repair.
Weak signal Blood pressure cuff is too Use other method to measure the blood pressure.
loose or the patients pulse is
too weak.
Over pressure The pressure is above the If failure continues, stop the NIBP measurement
protection specified safety upper limit. function and inform the supplier to dispatch
qualified maintenance personnel to repair.
51
Monitoring Impedance Respiration
The impedance plethysmography is to monitor a patients respiration by the impedance variation between two
electrocardiogram electrodes. Respiration monitoring is achieved via electrocardiogram I, II or III lead and
their corresponding electrodes. No additional electrodes are required and the installation of the electrodes is
very important. With some patients, it is better to place the two respiration electrodes at the central line from
the right armpit or the most active area on the left side of the thoracic cage to obtain optimum respiration
waveform. See Figure 2.8 for recommended placements of five lead electrodes in respiration monitoring.
Warning:
The respiration monitoring is not recommended on patients with activity. It may lead to faulse
alarm.
Install white and red electrodes in opposite positions to obtain optimum respiration waveform.
Avoid placing the cable above liver and ventricle to reduce false readings produced by the heart
coverage or pulsation blood flow.
Figure14.1
52
Monitoring Impedance Respiration
Press Resp mode to enter respiration measurement mode and then choose between impedance respiration
and nasal tube respiration.
Press Lead to select lead types.
Press Gain Mode for automatic or manual.
Press Adjust Gain for four levels: x1/2, x1, x2, and x4.
Status prompting bar
Press Waveform Speed for four levels: 6.25mm/s, 12.5mm/s, 25mm/s, and 50mm/s. This also adjusts the
speed for impedance respiration and nasal tube respiration.
Respiration waveform.
Scale.
The range unit is Ohm. The range and scale vary with different gains.
The respiration rate is displayed on the right side of the monitor, as shown below:
[Auto]: the monitor determines respiration rate source based on its detection of the connection of the
impedance respiration lead. If the connection is good, the respiration rate comes from the impedance
respiration. If the impedance respiration lead is off, the respiration rate source comes form the nasal tube
respiration or CO2.
[ECG]: selects the impedance respiration as the respiration rate source.
[T- Resp]: selects nasal tube respiration as the respiration rate source.
13.4 Messages
Lead off: it means that currently the leads are not properly connected. When the lead cable is not inserted into
the monitor or connected to the patient, this message appears.
Connection off: it means that the impedance respiration function is cut off internally. Go to settings in the
<menu>[System Setup][Calibration Function][ECG On/Off][Impedance
Respiration]. Refer to ECG On/Off operation s and sections.
Ops normal: except the above mentioned two conditions, the system will prompt Ops normal.
53
Monitoring Nasal Tube Respiration
Press T-Resp to enter respiration measurement mode and then choose between impedance respiration and
nasal tube respiration.
Press Gain Mode for automatic or manual.
Press Adjust Gain for four levels: x1/2, x1, x2, and x4.
Status prompting bar
Press Waveform Speed for four levels: 6.25mm/s, 12.5mm/s, 25mm/s, and 50mm/s. This also adjusts the
speed of the impedance respiration and nasal tube respiration.
14.2 Messages
54
Monitoring Temperature
The NT3 Series Veterinary monitor uses two types of temperature sensors: internal and external. Here external
mentions forehead and oxter, internal mentions anus. Place temperature sensors nicely on a patients body to
obtain two body temperatures and their difference.
Caution:
Before monitoring starts, check if the sensor cable is properly connected. Pull the temperature
sensor cable from the channel 1 jack, an error message T1 sensor is off is displayed on the screen
and the alarm sound can be heard. The same happens with the other channel.
Be cautious when handling temperature sensors and cables. When it is not in use, the sensor and
cable ought to be in a loose loop. Tightly folded cable can cause mechanical damage.
Temperature calibration should be done once a year (or according to the schedule specified in the
hospital procedure). If calibration is needed, contact the manufacturer.
During monitoring process, the temperature measurement instrument will conduct a self-check
every hour lasting for two seconds. This will not affect normal operation of the monitor.
Stable temperature values will be get mesureemnt for about 3 min. The temperature measurement result is
displayed on the screen as shown below:
55
Monitoring Temperature
1. Conncect the sensor to monitor, inspect it carefully to make sure it doesnt damaged.
2. Sterilize the sensor by dipping it in rubbing alcohol.
3. Allow it to dry, then apply petroleum jelly or a similar lubricant to the tip.
4. Allow your pet to stand, or lay her on her side, and gently hold her down.
5. Slowly but firmly raise the tail up and to the side, and insert the sensor. Push it in gently until only about half
an inch is sticking out. Then, clip the alligator clamp on the horse's tail hair.
6. Leave the sensor in for about three minutes
7. Wipe or rinse the sensor after use, and then sterilize it by dipping it in rubbing alcohol.
Application site
Caution:
When cleaning the sensor connected to the monitor, always power off and unplug the AC.
Reusable temperature senor
The heat for the temperature senor can not be over 100 (212). It can only endure for a short time period
80 (176) 100 (212).
1 The sensor can not be sterilized with high pressure vapor.
2 After use, surface-clean sensor and cable with a soft gauze pad by saturating it with a solution such as 70%
isopropyl alcohol, if low-level disinfection is required, use a 1:10 bleach solution.
3 When cleaning the sensor, hold the front end with one hand and use the other hand to wipe the sensor
towards the connector with a wet cloth(scrub away for two times.
Warning:
56
Monitoring SpO2
SpO2 measures functional blood oxygen saturation. It measures the percentage of oxyhemoglobin. It does not
measure carboxyhemoglobin or methemoglobin. For example, if 97% of red blood cells in the artery are
oxygenated, then blood has 97% blood oxygen saturation. The monitor SpO2 value reading would be 97. SpO2
measurement is a non-invasive, continuous measurement through an SpO2 sensor attached to a patients finger.
The sensor is connected directly to the SpO2 module. There are three types of display for SpO2: percentage
(%), pulse rate, and SpO2 waveform.
z SpO2 measurement is a non-invasive, continuous method. It uses a light emitter with red and
infrared LEDs that shine through a reasonably translucent site with good blood flow. Typical sites
are tongue, toe, paw.
z The red light used is 660nm wavelength light band. Infrared light is 940nm wavelength light band.
The biggest LED output power is 4mW.
z The quantity of light passing through depends on many factors, most of which are permenant. One
of those factors is artery current changing with time. Through measurement of the light absorbed
during pulsation, oxygen saturation of artery is obtained. This meausre can generate Pleth
waveform and PR signal.
z SpO2 value and Pleth waveform can be shown on the main screen.
Warning:
1. Do not use the same limb for both SpO2 sensor and NIBP cuff. During NIBP measuring, blood flow
abstruction can influence SpO2 reading.
2. Make sure the finger nail blocks the light window.
3. Sensor cable should be on the back of the hand.
4. PR value displays only when the HR source is set fom SpO2.
Caution:
Before monitoring starts, check if the sensor cable is in good condition. If the SpO2 sensor cable is
unpluged, sensor off will appear on the screen. And touch off audio alarm at same time.
If the package or the sensor seems to be damaged, do not use this SpO2 sensor and return it to the
manufaturer.
Warning:
Using a damaged sensor may cause inaccurate readings, possibly resulting in patient injury or
death, inspect each sensor. If a sensor appears damaged, dont use it. Use another sensor or contact
your local distributor.
Prolonged use or the patients condition may require changing the sensor site periodically. Change
sensor site and check skin integrity, circulatory status, and correct alignment at least every 4 hours.
If any of the integrity checks fail, dont attempt to monitor the patient. Use another sensor or
contact your local distributor for help if necessary.
1. Choose the sensor and clip that is appropriate for the patient, inspect it carefully to make sure it
doesnt damaged.
2. Connect the sensor to the monitor.
3. Clean or disinfect the sensor.
4. Attach the sensor to the proper site of the patient.
5. Power on the monitor.
The preferred sensor site for feline, canine and equine animal is on the tongue, with the optical
components of the sensor positioned to the tongue. Alternatively, the sensor and clip may be
placed to the toe, lip, ear, vulva or prepuce of the animal.
Select the appropriate size, either for a large animal, a small animal or a medium animal. Position
the lingual clip on the base of the tongue, and placement is dependent on the thickness of the
tongue. Start at the tip and work your way towards the base. Always direct the light downward
(towards the floor) regardless of the animals position to reduce the effects of ambient light. Keep
the tongue moist during longer procedures and monitor for significant temperature loss. Ensure
that there is a pulse strength bars displayed on the pulse oximeter.
Veterinary sensor with large clip Veterinary sensor with small clip
(Applicate site: tongue, paws and well-vascularized areas)
If necessary, the lingual clip may also be positioned on toe, lip, cheeks, ear, prepuce or vulva of
the animals. Moisten the hock area with isopropyl alcohol, water, and clip hair if needed. Toget
better reading on the samller tongues, bringing the sides of the tongue up and pass the light
through both layers. Dont fold the tip of the tongue back because you will restrict blood flow to
the tongue.
The small clip proves effective on the small breeds and especially on smaller cats. The clip will
work on a cats tongue, ear and toe-webbing. The small clip also works well on larger animals.
58
Monitoring SpO2
Small clip
In order to protect the animal from injure by clip, and confirm the the Y-type sensor easily.
Solaris has the two sizes of cushions for its clips accordingly. See followed photo.
Make sure the sensor cable is positioned along the side of the animals face and body to avoid
entanglement with animal.
Caution:
The maximum time duration for one sensor site in use should not be over 4 hours. The sensor
surface temperature should not be higher than 41C, or it may cause burn.
59
Monitoring SpO2
During continous monitoring, sensor site should be cleaned at least every 12 hours. Otherwise it
may result in inaccurate measurements.
SpO2 display:
SpO2 alarm high/low limits
Symbol for SpO2 alarm off
SpO2 parameter label. Press it, SpO2 menu will appear.
SpO2 value
In SpO2 menu, the user can set alarm switch, alarm high/low limits, and alarm priority.
Average periods: calculate average time of SpO2 value. Select [AVRG.], there are 3 options. Select 4 Beats,
the result is more sensitive. Select 16 Beats, the result is more accurate.
Warning:
Must power off and disconnect AC power before cleaning monitors or sensors.
2. Clean and disinfect cables with 3% hydrogen peroxideor, 70% isopropyl alcohol.
60
Maintenance and Cleaning
Warning:
If the hospital/institution does not implement a satisfactory maintenance plan, it will result in
malfunction of the equipment, which in turn may threaten human lives.
When equipment is in accidental humidity, please notice do not power-on, ask qualified service
personal open the monitor and airing it immediately
Dispose or callback to the dispose spare parts (including batteries) should keep to local
garvenments rules about the related dispose or callback to the spareparts.
1. Switch off power and unplug AC before cleaning the monitor or sensor.
2. Monitor should be dust free.
3. Clean the surface with soap and water.
4. Do not use strong solvents such as acetone.
2. Use diluted cleaning agents according to the manufacturer's instructions.
3. Do not scrub the monitor with metal brush.
4. Do not allow any liquid into the monitor or immerse any part in liquid.
5. Do not leave any cleaner residue on the surface of the monitor.
17.3 Cleaner
Sodium hypochlorite with consentration range about 500ppm (Diluted bleaching powder for
family use 1:100) to 5000 ppm (Diluted bleaching powder for family use 1:10) is very effective. The
ppm can be determined by the amount of stain (blood, mucilage glue of animal or plant) on the
surface of the monitor.
61
Maintenance and Cleaning
Clean the surface of the monitor and sensors using medical grade alcohol. Let it air-dry or wipe it
dry with cloth.
Disinfect the monitor when necessary. Clean the monitor before disinfecting it.
Disinfect the monitor when necessary. especially equipement are polluted by animalcules during
the storage, transportation and operationClean the monitor before disinfecting it.
Suggested cleaning materials are: Ethanol and Acetaldehyde. See relevant s for sterilization materials for ECG
leads and NIBP cuffs.
Attention:
Do not leave any cleaner stain on the surface of the monitor. Wipe with wet cloth immediately.
17.4 Disinfecting
Only when it is necessary as deemed by the hospital/institution, should the monitor be disinfected. Always
clean the monitor before disinfecting it.
Warning:
Do not use ECG leadwires, SpO2 sensor, NIBP cuff, temperature probe between infections and
non-infections without disinfection.
62
Monitors Accessories and Purchase Information
The following accessories are recommended by manufacturer. Use other models may cause damage
to the monitor.
Note:
The life for 5 pin ECG cable, SpO2 sensor, temperature, NIBP cuff is one year.
Name Model PN
5 leads ECG cable and lead HP-12PIN 30101112R00042
ECG clip 31078135KENDALL 30106310781352
Name PN(Model)
Y-type digital SpO2 Sensor Y400N-160087
Y-type digital SpO2 Sensor Y400A-160087
Y-type digital SpO2 Sensor Y400N-300087
Y-type digital SpO2 Sensor Y400A-300087
Name Model PN
Rectal temperature sensor 30104000000022
Name Model PN
Nose tube JEIII-A
63
Appendix A
64
Appendix A
<5 % UT <5 % UT
(>95 % dip in (>95 % dip in
UT) UT)
for 0,5 cycle for 0,5 cycle
40 % UT 40 % UT
Voltage dips, short (60 % dip in (60 % dip in Mains power quality should be that of a typical
interruptions and UT) UT) commercial or hospital environment. If the user
voltage variations for 5 cycles for 5 cycles of the NT3requires continued operation during
on power supply 70 % UT 70 % UT power mains interruptions, it is recommended
input lines (30 % dip in (30 % dip in that the NT3 be powered from an
IEC 61000-4-11 UT) UT) uninterruptible power supply or a battery.
for 25 cycles for 25 cycles
<5 % UT <5 % UT
(>95 % dip in (>95 % dip in
UT) UT)
for 5 sec for 5 sec
7
d= P 800 MHz to 2.5 GHz
E1
65
Appendix A
These limits are designed to provide reasonable protection against harmful interference in a typical
medical installation. However, because of the proliferation of radio-frequency transmitting
equipment and other sources of electrical noise in the health-care and home environments (for
66
Appendix A
example, cellular phones, mobile two-way radios, electrical appliances), it is possible that high levels
of such interference due to close proximity or strength of a source, may result in disruption of
performance of this device.
NT3 generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance
with these instructions, may cause harmful interference with other devices in the vicinity. Disruption may
be evidenced by erratic readings, cessation of operation, or other incorrect functioning. If this occurs, the
site of use should be surveyed to determine the source of this disruption. Take below actions to eliminate
the source:
Turn the equipment in the vicinity off and on to isolate the offending equipment.
Reorient or relocate the other receiving devices.
Increase the separation between the interfering equipment and NT3.
67
Appendix B
1. Electric shock type: type I equipment with inside power supply and internal power supply.
2. Electric shock degree: all application parts are BF type, except ECG and TEMP which are CF type.
3. Harmful liquid material proof degree:: ordinary equipment (no seal agains liquid).
4. Disinfection: follow manufacturers recommended methods.
5. Safety on flammable gas: not suitable to use where flammable gas is present.
6. Input power: no bigger than 80VA.
7. The monitor is a continual working system.
B.5 ECG
HRheart rate:
B.6 NIBP
Measurement range
Horse: 20~250mmHg
Dog: 20~250mmHg
Cat: 20~250mmHg
Measurement accuracy: 0.4 kPa (3 mmHg) or 2%take the bigger one.
Display unit: kPa or mmHg.
B.7 RR
B.8 TEMP
B.9 SpO2
B.10 PR
69
Appendix B
Measurement accuracy: 1 bpm or 2%. Take the bigger one.
B.11 Alarm
Default value of alarm high limit Default value of alarm low limit
Alarm parameter (unit)
Horse Dog Cat Horse Dog Cat
SpO2% 99 99 97 85 85 85
RR 30 40 50 8 8 8
et CO2mmHg 50 50 100 20 20 40
ins CO2mmHg 20 20 20 0 0 0
mV 23 22 20 23 22 20
mV 20 20 20 20 20 20
mV 20 20 20 20 20 20
aVRmV 23 22 20 23 22 20
aVLmV 20 20 20 20 20 20
aVFmV 20 20 20 20 20 20
VmV 20 20 20 20 20 20
70
Appendix B
B.13 Settings range and allowable tolerance of alarm high/low limits
Alarm parameter Setting range of alarm high limit Setting range of alarm low limit
HR 60 bpm ~300 bpm 30 bpm ~200bpm
ST -2.55 mV ~2.55 mV -2.55 mV ~2.55 mV
NIBP 0 mmHg ~300 mmHg 0 mmHg ~300 mmHg
IBP -59 mmHg300 mmHg -59 mmHg300 mmHg
SPO2 0%~100% 0%~100%
PR 25 bpm ~400 bpm 25 bpm ~400bpm
TEMP 35.5~43.5 35.5~43.5
RR 0 bpm -120 bpm 0 bpm -120 bpm
Int CO2 0mmHg76mmHg 0mmHg76mmHg
Et CO2 0mmHg76mmHg 0mmHg76mmHg
Internal power supply: 2 hoursthe fully charged battery under condition of blood presure measuring for
every 10 minutes
Packaging
71
Appendix B
Place the monitor in a plastic bag. Place it in a corrugated carton filled with the foam or other fillers. Seal the
carton.
Transportation
The monitor can be transported by airplane, train, or automobile. Prevent fierce collision during transportation.
Do not keep it with perishables. The transportation environment should be:
a)
b) Environment temperature range-20~70
c) Relative humidity range95%
d) Air pressure range: 500kPa~1060kPa
Storage
The monitor should be stored indoors with a temperature range -10~+40, relative humidity 80%, no
corrosive gas, and with good ventilation.
72
Appendix B
13 EN 1060-1:1995/A1:2002 General Rrequirements
14 EN 1060-3:1997/A1:2005 Supplemental Requirements for Mechanical-Electronic Blood
Pressure Measurement Systems
15 EN 60601-2-30:2000 Medical Electrical Equipment Part II:
Safety and Performance Requirements for Autocycle NIBP
Patient Monitoring Equipment
16 EN 12470-4:2000 Performance of Continuous Electronic Temperature Monitoring
17 EN 60601-2-27:1994 Special Safety Requirements for ECG Monitoring Equipment
18 EN 60601-2-49:2001 Special Safety Requirements for Multi-Parameter Monitoring
Equipment
19 IEC 60601-2-34:2000 Special performance and safety requirements for IBP
monitoring equipment
<Multi-parameter monitor> production specification
73