Scheme Tutorial 1
Scheme Tutorial 1
2 2014/15)
Tutorial 1
Question 1
Measurement is the process of determining the amount, degree, or capacity by comparison (direct or
indirect) with the accepted standards of the system units being used.
Instrument is a device or mechanism used to determine the present value of the quantity under
measurement
(b). A Digital Clamp Meter is used to measure the current flow into a machine at different
interval. The results are shown in Table 1. Calculate:
Arithmetic mean;
Deviation for each reading;
Average deviation;
Standard deviation;
Precision for most frequent reading.
Table 1
No. of measurement Frequency of reading Current, I (A)
1 1 2.002
2 3 1.997
3 1 2.001
4 4 1.996
5 1 1.998
Sol:
x
n I 1.9976
n
Dav = 0.00164
1.996 1.9976
(vi) Precision for the most frequent reading= P 1 0.9992
1.9976
Question 2
(b) The expected value of the current through a resistor is 22 mA. However the
measurement yields a current value of 19 mA. Calculate
Yn 22mA, X n 19mA
Absolute error , e Yn X n (22 19) mA 3mA
e 3
%E 100 100 13.64%
Yn 22
e
Relative accuracy, A 1 1 0.1364 0.9864
Yn
% of accuracy, a A 100 98.64%
(c) The output voltage of an amplifier was measured at eight different intervals using the
same digital voltmeter with the following results: 20.00, 19.80, 19.85, 20.05, 20.10,
19.90, 20.25, 19.95. Which is the most precise measurement?
Xn Xn
Pr ecision 1
Xn
n 1 2 3 4 5 6 7 8 n=8
voltage 20.00 19.80 19.85 20.05 20.10 19.90 20.25 19.95 =159.9
deviation 0.0125 -0.1875 -0.1375 0.0625 0.1125 -0.0875 0.2625 -0.0375 Xn bar=19.9875
Precision 0.9994 0.9906 0.9931 0.9969 0.9944 0.9956 0.9869 0.9981
The nearest precision value to 1 is the most precise value, or The lowest deviation is the most precise value..
Hence the 1st measurement is the most precise measurement.
EKT212/4 Principles of Measurement and Instrumentation (sem.2 2014/15)
Question 3
(a) Discuss the difference between random error and systematic error of a
measurement.
Systematic Error is the deviation of the true value from the desired value
Random Error occurs when different results in magnitude or sign obtained on repeated measurement of
one or the same quantity.
(b) In a calibration test, 10 measurements using a digital voltmeter have been made of the
battery voltage that is known to have a true voltage of 6.11 V. The results are shown
in Table 2. Calculate:
Table 2
Number of measurement Reading (V)
1 5.98
2 6.05
3 6.10
4 6.06
5 5.99
6 5.96
7 6.02
8 6.09
9 6.03
10 5.99
Solution:
EKT212/4 Principles of Measurement and Instrumentation (sem.2 2014/15)
(iii)Deviatio
Number of measurement Reading (V)
n
1 5.98 -0.047
2 6.05 0.023
3 6.10 0.073
4 6.06 0.033
5 5.99 -0.037
6 5.96 -0.067
7 6.02 -0.007
8 6.09 0.063
9 6.03 0.003
10 5.99 -0.037
N =10 =60.27
Question 4
(b) The expected value of the voltage across a resistor is 100 V. However, the measurement
gives a value of 98 V. Calculate
Yn 100V , X n 98V
Question 5
(a) A voltmeter reading 185 V on its 200 V range and an ammeter reading 350 mA on its
500 mA range are used to determine the power dissipated in a resistor. Both these
instruments are guaranteed to be accurate within 1.5% at full scale deflection.
(b) With a suitable block diagram, briefly explain each of the main components in
instrumentation system.
Transducers
Device that converts a change in physical quantity into a change of electrical
signal magnitude.
Power Supply
Provide energy to drive the transducers.
Signal Conditioning Circuits
Electronic circuits that manipulate, convert the output from transducers into
more usable electrical signal.
EKT212/4 Principles of Measurement and Instrumentation (sem.2 2014/15)
Amplifiers
Amplify low voltage signal from transducers or signal conditional circuit.
Recorders
Used to display the measurement for easy reading and interpretation.
Data Processors
Can be a microprocessor or microcontroller.
Process Controllers
Used to monitor and adjust any quantity of the specified level or value.
Command Generator
Provide control voltage that represents the difference of the parameter in a
given process.