Lecture Notes On Impedance Spectros PDF
Lecture Notes On Impedance Spectros PDF
Editor
Olfa Kanoun
Chair for Measurement and Sensor Technology,
Chemnitz University of Technology, Chemnitz, Germany
All rights reserved. No part of this publication or the information contained herein may be
reproduced, stored in a retrieval system, or transmitted in any form or by any means, elec-
tronic, mechanical, by photocopying, recording or otherwise, without written prior permis-
sion from the publisher.
Although all care is taken to ensure integrity and the quality of this publication and the
information herein, no responsibility is assumed by the publishers nor the author for any
damage to the property or persons as a result of operation or use of this publication and/or
the information contained herein.
Table of contents
Preface vii
Corrosion
The impact of decreasing chloride concentration on the corrosion
state of steel in alkaline solution 91
T. Eichler, A. Faulhaber, K. Weidauer & B. Isecke
vi
Preface
vii
ABSTRACT: Many real devices such as batteries are nonlinear and their impedance values
depend on either current or voltage. However, impedance analysis is based on the assump-
tion of a linear system. If certain conditions such as sufficiently small excitation amplitude
to ensure quasi-linearity and measurements at several operating points are fulfilled, also the
impedance of a nonlinear system can be interpreted. Equations to calculate the large sig-
nal impedance from small signal impedance measurements of current, voltage and mixed
dependent devices are derived and analyzed.
Keywords: component, nonlinear, large signal impedance, small signal impedance, negative
resistance, negative capacitance, inductive semicircle, inductive behaviour
1 INTRODUCTION
Impedance spectroscopy is a powerful and widely used tool for the parameterization of simu-
lation models. From a single impedance spectrum of a linear device, both the structure of its
equivalent circuit and the parameter values can be extracted. For nonlinear devices, the same
information can be obtained, but measurements at several operating points are necessary to
obtain the complete behavior. Besides the dependency on the operating point, which means
the bias current or voltage, the impedance of a nonlinear device also depends on the excita-
tion amplitude. In order to guarantee quasi-linear conditions and thus usable spectra, the
amplitude has to be kept small enough (Barsoukov and Macdonald 2005). Furthermore it
has to be considered thatin contrast to the linear casethe measured (small signal) imped-
ance of a nonlinear device is not equal to the large signal impedance which is needed for a
time domain simulation model.
There are several publications dealing with nonlinear circuits; already in the 1950s, 1960s
and 1970s, researchers focused on the mathematical description of polarized electrodes
(Macdonald and Brachman 1954; Macdonald 1954; Macdonald 1955), nonlinear capacitors
(Macdonald and Brachmann 1955) and nonlinear networks (Popov and Paltov 1963; Chua
1969; Chua and Lin 1975). Those researchers had to face the problem that they could not
employ their theory on a grand scale because of the limitations in computing power. The
focus of these works was to calculate current and voltage behaviour of a circuit with given
nonlinearities, while the focus in this work is to determine the nonlinearities from current
and voltage measurements. More recent publications deal with this problem by the usage of
impedance spectroscopy, but they typically only determine the small signal impedance in the
frequency domain, e.g. (Darowicki 1994; Darowicki 1997; Darowicki 1998).
In a previous paper by the authors, a procedure was developed to determine the large
signal impedance of a voltage dependent device from its small signal impedance (Kowal and
Sauer 2009). Starting from the differential equation of the current-voltage relationship, the
equation is linearized after an AC and DC perturbation. The resulting AC part is trans-
formed into the frequency domain giving differential equations for the small signal imped-
ance elements. In this paper, the method is applied to current and mixed dependency.
Yssi (U DC )
Clsi (U DC ) = Im
DC = Cssi (U DC ) (1)
U DC U DC
Rlsi (U DC)=U
DC DC
=U
DC
(2)
{Yssi (u )} u + c Gssi (u )du + c
0 0
where Yssi is the measured small signal admittance, = 2 f is the frequency and UDC is the
bias voltage of the operating point. The small signal capacitance Cssi is used as an abbrevia-
tion for Im{Yssi} and the small signal conductance Gssi is used as an abbreviation for Re{Yssi}.
Using the same procedure for a current dependent RC circuit gives similar equations:
Cssi ( I DC ) Rssi ( I DC )
C)=
Clsi ( I D
DC
(3)
Rlsi ( I DC)
DC
I DC
Rlsi ( I D
)= 0
Rssi (i )di c
(4)
DC
C
I DC
Figure 1. RC circuit.
Fig. 2 shows the given voltage dependencies of the large signal parameters for the resistor
and the capacitance and the resulting simulated impedance spectra of the RC circuit. It can
be seen that for higher voltages the real part of the small signal impedance becomes negative
although the large signal parameters are always positive in the considered voltage range. This
behaviour can be explained by considering the equation to calculate the small signal conduct-
ance from the large signal conductance Glsi = 1/Rlsi, which is the inverse of equation 2:
dGlsi
lsi ( u )
Gssi (U DC ) = U DC Glsi (U DC
DC ) (5)
du u U DC
It can be seen that the small signal conductance depends on the derivative of the large sig-
nal conductance and if the first term of equation 5 becomes sufficiently negative, the small
signal impedance becomes negative. This relationship is illustrated in Fig. 3: For voltages
larger than 5V, the slope of the large signal conductance is negative and above about 7V,
also the small signal conductance is negative. Fig. 4 shows the simulated impedance spectra
(left hand picture) for a given current dependency of R and C (right hand picture). Again
both large signal parameters are always positive in the considered current range, but for high
Figure 2. Simulated small signal impedance of an RC element according to Fig. 1 (left) with voltage
dependent large signal impedance characteristics for R and C (right).
Figure 3. Large signal (upper figure) and small signal (lower figure) conductance corresponding to 2.
currents, both the real and the imaginary part become negative. Similarly to voltage depend-
ency, this behaviour can be explained using the relationships between small and large signal
impedance (inverse of equations 3 and 4):
Clsi ( I DC ) Rlsi ( I DC )
Cssi ( I DC ) = DC (6)
Rssi ( I DC )
dRlsi
lsi ( u )
Rssi (U DC ) = I DC Rlsi ( I DC)
DC (7)
di i I DC
Equation 7 shows that also the current dependent small signal resistance depends on the
derivative of the large signal resistance and if it is sufficiently negative, the small signal resist-
ance becomes negative. Since also the small signal capacitance (equation 6) depends on the
small signal resistance, it becomes negative for the same current range. This is illustrated
in Fig. 5.
6
Also the equations of mixed dependencies, e.g. R(u) with C(i) and R(i) with C(u), can be
derived applying the same procedure. As in the cases of pure current or voltage dependency,
the large signal resistance can be calculated independently from the capacitance according
to equation 2 for voltage dependent resistance and equation 4 for current dependent resist-
ance. However, the calculation of the large signal capacitance can depend on the large and
small signal resistance. The equations for the large signal capacitance for all four cases of
current or voltage dependencies are given in Table 1. It can be seen that the capacitance
equation depends on the dependency of the resistance. If the resistance is voltage dependent,
the large signal capacitance is equal to the small signal capacitance and if the resistance is
current dependent, the large signal capacitance depends on the large and small signal resist-
ance as well. The equations were verified by simulated impedance spectroscopy of the four
dependencies of RC circuits: A sinusoidal current with a fixed frequency is applied to the
parallel connection of resistance and capacitance and the impedance is calculated from the
FFT of current and voltage. This is repeated for a set of frequencies for each direct current
offset. Fig. 6 shows the Simulink model used for simulation. For each of the four combina-
tions, the large signal resistance is calculated according to equation 2 for voltage depend-
R(u) R(i)
Rssi (i )
C(u) Clsi(u) = Cssi(u) Clsi (u ) = Cssi (u )
Rlsi (i )
Rssi (i )
C(i) Clsi(i) = Cssi(i) Clsi (i ) = Cssi (i )
Rlsi (i )
Figure 6. Comparison of possible equations to calculate the large signal capacitance from simulations
of mixed dependent RC circuits. Left hand figure: R(u) with C(i), right hand figure: R(i) with C(u).
Figure 7. Simulink model for simulated impedance spectroscopy of parallel connection of current and
voltage dependent resistance and capacitance.
5 CONCLUSION
Based on the procedure introduced in (Kowal and Sauer 2009) to derive equations for the
large signal impedance as a function of the small signal impedance, the equations for an RC
parallel circuit with pure voltage, pure current or mixed voltage and current dependency of
the elements have been derived and verified by simulations. It turned out that the equation to
calculate the large signal capacitance depends on whether the resistance is current or voltage
dependentindependent from the kind of dependency of the capacitance. These equations
apply to any current or voltage dependent system. Most systems can be modelled both as cur-
rent and voltage dependent, but typically one dependency is more convenient or makes more
sense because of physical or chemical interpretation. According to the decision to model
either current or voltage or mixed dependency, the corresponding equations have to be used
for parameter extraction from measurement.
Simulations of RC circuits current and voltage dependent elements have shown that for
suitable resistance characteristics, the resulting impedance spectra (small signal impedance)
show apparently negative resistance semicircles for voltage dependency and apparently
negative resistance and capacitance semicircles for current dependency, although both the
large signal resistance and capacitance are positive at all times in the performed simulations.
This behaviour can be explained mathematically from the corresponding equations for the
small signal impedance. The inductive behaviour in the case of current dependent resistance
can also be modelled with positive small signal resistance and inductance. However, if such
behaviour is observed in measured spectra, it does not necessarily mean that the large signal
impedance is also inductive, but a detailed analysis is needed. Before searching for a physical
or chemical explanation for inductive behaviour, the large signal impedance should be calcu-
lated to find out if it is inductive as well. In the case of voltage dependent resistance, a nega-
tive small signal resistance can be observed. Negative large signal resistances are of course
impossible for passive circuits, so if such behaviour is measured, it is most probably caused
by the nonlinearities of the circuit or otherwise, active elements are present.
REFERENCES
Barsoukov, E. and J.R. Macdonald (2005). Impedance spectroscopy, theory, experiment and applications
(2nd ed.). Number ISBN: 0-471-64749-7. John Wiley & Sons, Inc.
Chua, L.O. (1969). Introduction to nonlinear circuit theory. McGraw Hill.
Chua, L.O. and P.-M. Lin (1975). Computer aided analysis of electronic circuits: algorithms and compu-
tational techniques. Prentice Hall.
Darowicki, K. (1994). Fundamental-harmonic impedance of first-order electrode reactions. Electrochim.-
Acta. 39, 27572762.
Darowicki, K. (1997). Linearization in impedance measurements. Electrochimica Acta 42, 17811788.
Darowicki, K. (1998). Differential analysis of impedance data. Electrochimica Acta 43, 22812285.
Kowal, J., Hente, D. and D.U. Sauer (2009). Model parameterization of nonlinear devices using imped-
ance spectroscopy. IEEE Transactions on Instrumentation and Measurement 58, 23432350.
Macdonald, J.R. (1954). Static space-charge effects in the diffuse double layer. J. Chem. Phys. 22(8),
13171322.
Macdonald, J.R. (1955). Note on theories of time-varying space-charge polarization. J. Chem. Phys.
23(12), 23082309.
Macdonald, J.R. and M.K. Brachman (1954). Exact solution of the debye-hckel equations for a polar-
ized electrode. J. Chem. Phys. 22, 13141316.
Macdonald, J.R. and M.K. Brachmann (1955). The charging and discharging of nonlinear capacitors.
In Proceedings of the IRE, Volume 43, pp. 7178.
Popov, E.P. and I.P. Paltov (1963). Nherungsmethoden zur Untersuchung nichtlinearer Regelungssys-
teme. Akadem. Verlags-Gesellschaft. Geest & Portig K.-G.
1 INTRODUCTION
Time domain simulations of electrochemical systems like batteries or super capacitors are
important for example in electric vehicles or stationary uninteruptable power supplies.
The energy management systems can use the simulation results to decide about further charg-
ing or discharging the battery or the capacitor by not exceeding the end of charge or end of
discharge voltage or other limiting factors. Charge or discharge power can be regulated to
avoid aging and to extend lifetime. For time domain simulations adequate simulation models
are required that sufficiently precise describe the underlying physical mechanisms inside the
energy storage. In the frequency domain a lot of well matching impedance models have been
developed to describe underlying mechanisms. Many of these impedance models models
contain constant phase elements for the description of double layer capacities and diffusion
processes (Trltzsch and Kanoun 2006). Describing the behavior of constant phase elements
in the frequency domain can easily be done by an analytical mathematical function. Describ-
ing a constant phase element in the time domain is more complex because the inverse fourier
or laplace transform will not lead to an ordinary differential equation but to a fractional dif-
ferential equation. A direct solution of an fractional differential equation in the time domain
is not suitable for online simulations in technical systems because it needs a lot of calculation
power and memory, especially for systems containing more than one constant phase element.
Therefore approximations of constant phase elements are required to use established meth-
ods for the solution of ordinary differential equations. These methods requiring only little
computational power and memory and are suitable for online simulation. Digital filters, a
widely used and well understood technique in communications engineering, are one method
9
The temporal and spatial concentration of charge carriers in electrochemical systems is influ-
enced by many mechanisms. One important mechanism is diffusion. Ficks second law is the
describing equation for diffusion phenomena and allows the calculation of the charge car-
rier density c in electrochemical systems. The diffusion constant D is the proportional factor
between spatial and time behavior (Hamann and Vielstich 1997).
2c (1)
sc = D
x 2
With Ficks first law the relation between charge carrier density c and current density jdiff is
given. Hereby F is the faraday constant and n is the amount of unit charges of the considered
charge carrier species.
c (2)
jdiff nDF
x
For solving Ficks second law for semiinfinite diffusion the following boundary conditions are
valid. At the surface of the electrode a constant current density jdiff, 0 is assumed. At an infinite
distance from the electrode the concentration of the species is assumed to be constant at c0.
1 : jdiff ( 0) = jdiff ,0 , 2 : c ( ) = c0 (3)
For calculating the impedance arising from diffusion the concentration at the electrode
surface at x = 0 is required.
jdiff ,0
c (x = ) = c0 + (4)
nF Ds
From the concentration at the electrode surface the corresponding diffusion overvoltage
diff can be calculated using the nernst equation. Assuming small voltage changes leads to the
linearization of the nernst equation. This linearization is valid if the overvoltages are less
than 25mV.
RT c RT
diff ,0 = ln jddiff ,0 (5)
nF c0 c0 n2 F 2 Ds
The impedance resulting from diffusion is the ratio of diffusion overvoltage diff and
diffusion current A jdiff at the electrode surface. Hereby A is the area of the electrode.
10
2W 2
2 1 j
ZW ( s ) = = W = W (6)
s s
1 RT
W = 2 2 s
A c0 n F 2 D
Equation 6 represents the warburg impedance in the frequency domain with the frequency
= 2f and also in the laplace domain with the complex frequency s = j. The laplace rep-
resentation is important for the constant phase element representation ind the later sanc-
tions. In practice the impedance of diffusion is not observed according to equation 6. Very
often the impedance of the diffusion has not a fixed exponent of = 0.5 but an exponent of
0 < < 1. So the warburg impedance will become a constant phase element.
2 2
ZW ( s ) = W (7)
s
The corresponding impulse response of a system having the transfer function of the con-
stant phase element according to equation 7 can be calculated by inverse laplace transform.
The result is the following equation. Hereby is the gamma function.
t 1
zW (t ) = 2W
2 (8)
( )
If the impedance is considered as transfer function in the frequency domain there exists a
corresponding differential equation in the time domain. Exciting the system using a certain
current i results in a corresponding voltage u. In the frequency or laplace domain the voltage
spectra is calculated by multiplying the current spectra with the transfer function.
U ( s ) = ZW ( s ) I ( s ) (9)
This multiplication in the laplace domain corresponds to a convolution in the time domain.
2
2W t
u (t ) = (t ) 1i ( ) d (10)
( ) 0
For time domain simulation of the voltage u of a diffusion element the solution of this
integral is required. For short time simulations the solution of this integral is reasonable, for
example by approximation the integral by a sum. For long time- and online simulations the
computation time and memory usage will rise with simulation time. The solution of this inte-
gral is also not reasonable for systems containing more than one diffusion element because
they will interact. The interaction leads to more than one convolution integral. To overcome
these problems several methods for constant phase element approximation were developed.
These methods allow the determination of transfer functions using poles and zeros in the
laplace domain in terms of control theory. These transfer functions can easily be simulated
using methods for ordinary differential equations or digital filters. The following sections
discuss one method to overcome the problem of high computational effort by approximation
the constant phase element and to simulate the approximation using digital filters.
Several methods for approximating the transfer function of a constant phase element by rational
fractions are known (Podlubny, Petras, Vinagre, OLeary, and Dorcak 2002). They mainly base
11
The impulse response of this system can be calculated by the inverse laplace transform.
The result is the following equation.
1
d (t ) = t n 1 (12)
u ( n)
n
The first approximation of the constant phase element is done by the following equation.
n n
s b 1 + s / b
1 + s/ (13)
u u h
The frequency u is defined as center frequency between lower edge frequency b and the
upper edge frequency h of the approximation.
u b h (14)
So the transfer function of the constant phase element is replaced by its approximation in
a certain frequency band, what is sufficient for many applications.
n
1 + s / b (15)
DA ( s ) = b
u 1 + s/ h
The approximation is synthesized from a polynomial fraction containing poles and zeros.
If the number of poles and zeros goes to infinity the approximation according to equation 15
will be the same as the infinite product.
DA ( s ) = lim DN ( s ) (16)
N
n
N
1 + s/ k
DN ( s ) = b
u k N 1 + s / k (17)
The transfer function according to equation 17 can be used for time domain simulation
solving its corresponding ordinary differential equation. The zeros and poles are defined by
the corresponding frequencies k and k that can be calculated according to the following
formulas (Oustaloup, Levron, Mathieu, and Nanot 2000).
( k + N + . n// ) / ( N + )
k = b h (18)
b
( k N + . + n/ ) / ( N + )
k b h (19)
b
12
N = 2
n = 0.594
b = 0.0001 1/s
h = 1 1/s
u (calculated) = 0.01 1/s
105 0
D
DN
1
|D|
100
105
Im(D)
106 104 102 100 102 3
0
4
20
(D)/
5
40
6
60
106 104 102 100 102 0 2 4 6
Re(D)
Figure 1. Comparison of the approximation DN (s) and the constant phase element D (s).
The approximation of the constant phase element now is visualized for some example val-
ues. Table 1 shows the values used for simulation. Figure 1 shows the transfer function of the
approximation DN (s) according to equation 17 and the transfer function of the constant phase
element D (s) according to equation 11. Within the the approximated frequency range between
b and h the absolute value of both transfer function is matching well. Also the angle of the
transfer function is in accordance. Outside this frequency range there is a big deviationas
expected. Figure 2 shows the deviation between the two approximations DA (s) according to
equation 15 and DN (s) according to equation 17. For the number of 2N + 1 = 5 zeros and poles
there is an maximum relative deviation of 3.5%. The phase angle maximally differs about 1.4.
So the function DN (s) according to equation 17 is a very good approximation for the function
DA (s) according to equation 15 also for a small number of poles and zeros.
Digital filters are the preferred method for time domain simulations especially for online
simulations of technical systems because of their easy implementation using microproces-
sors. Digital filters require the definition of filter coefficients. These filter coefficients can
be determined from poles and zeros of transfer functions in the laplace domain by several
transforms. Widely used transforms are the direct transform, the bilinear transform and the
13
(|DN| |DA|)/|DA|
0.02
0.02
0.04
106 105 104 103 102 101 100 101 102
1.5
1
(DA) (DN)/
0.5
0.5
1
106 105 104 103 102 101 100 101 102
Figure 2. Relative deviation between the approximation DA (s) and the approximation DN (s).
impulse invariant transform (van den Enden and Verhoeckx 1990). Because of their very
simple realization the direct transform will be used in this article. The poles and zeros of a
digital filter in the domain of the z transform are calculated from the poles and zeros in the
s domain of the laplace transform. Using the direct transform poles and zeros zz/p in the z
domain are calculated using the following equation from poles and zeros sz/p in the s domain.
Hereby ts is the sampling time of the digital system.
sz pts (20)
zz p e
The transfer function DN (s) is the transfer function that should now be used for time
domain simulations. For applying the direct transform for digital filter design the transfer
function must be rewritten in terms of poles and zeros. So equation 17 will become the fol-
lowing equation.
n
N
k N
k + s
DN ( s ) = b (21)
u k N k k N k + s
From this equation the poles and zeros in the laplace domain can easily be obtained.
sz ,k = k (22)
s p k = k (23)
From the poles and zeros in the laplace domain the poles and zeros in the z domain are
calculated using the direct transform according to equation 20.
zz , k e k ts (24)
zp k e k ts (25)
Knowing the zeros and poles in the z domain, the transfer function in the z domain is the
following.
14
0.07
d
0.06
dN
0.05 dZ
0.04
0.03
0.02
0.01
0 100 200 300 400 500 600 700 800 900 1000
102
|dZ d|/d
100 |dZ dN|/dN
102
104
106
0 100 200 300 400 500 600 700 800 900 1000
t/s
Figure 3. Impulse responses d of the constant phase element, dN of the system in the laplace domain
and dz of the digital filter and corresponding relative deviations.
15
2
a
aN
1
aZ
0
0 100 200 300 400 500 600 700 800 900 1000
102
|aZ a|/a
100 |aZ aN|/aN
102
104
106
0 100 200 300 400 500 600 700 800 900 1000
t/s
Figure 4. Step responses a of the constant phase element, aN of the system in the laplace domain and
az of the digital filter and corresponding relative deviations.
approximation at low frequencies. The relative deviation between the impulse response of
the approximated system in the laplace domain according to equation 17 and the digital
filter dz is very small and arises from the direct transform of zeros and poles from laplace
domain to z domain. Figure 4 shows the step responses of the different system consid-
ered. The step response of the constant phase element a is further rising after 370s. The
step response aN of the system in the laplace domain according to equation 17 and the
step response aZ of the digital filter tend to reach a stable value. This is required due to
system stability.
The application of the described method will now be shown by a simulation of the voltage
response of a battery. As an example a 18650 Lithium Ion cell with 1.2Ah at 80% state of
charge was selected.
16
0 1000 Hz
Zmeas
101 Hz Zmodel
0.01 10.5 Hz 0.1 Hz Zmeas,PP
Zmodel,PP
0.02
1 Hz 0.01 Hz
0.03 Pulse Pattern:
Im (Z)/
0.06
0.07
0.08 0.1 0.12 0.14 0.16 0.18 0.2 0.22
Re (Z)/
0.18
0.16
||/
0.14
0.12
0.1
2
4
6
()/
8
10
12
14
102 101 100 101 102 103
f/Hz
Figures 6 and 7 show the Nyquist and Bode diagram of the measured impedance spec-
trum and the fitted model. The model is matching very well at low frequencies. There are
big deviations at high frequencies. Because diffusion occurs at low frequencies and the time
domain simulation aims on long simulation times, deviations in the high frequency range are
neglected.
For verification reasons also the impedance ZPP extracted from the pulse pattern data
measured in the time domain is shown. The impedance ZPP was extracted from time domain
17
U/V
ubattery
3.9
uN
3.8 uZ
0.5
I/A
0.5
1
0 200 400 600 800 1000 1200
t/s
pulse pattern data applying the fourier transform to the voltage and current data measured
at the battery that are shown in figure 8. The model parameters of the model according to
equation 28 extracted from pulse pattern data are very similar to the parameters extracted
from an impedance spectrum. Also the impedance data show a good agreement.
Comparing equation 7 and equation 29 gives the relation between the parameters.
n = (30)
2 2
k = W (31)
u
From the known model parameter the corresponding parameter n of the constant phase
element approximation according to equation 17 can be calculated. The frequency range
within the approximation is valid is determined by the edge frequencies b and h, which must
be selected. The center frequency u is calculated using equation 14. From the known model
parameter W and the center frequency u the remaining parameter k can be determined.
uN (t ) = U RE i (t ) + k (i (t ) d N (t )) (32)
18
N = 2
n = 0.594
b = 0.0001 1/s
h = 1 1/s
u (calculated) = 0.01 1/s
RE = 0.14773 1
W = 0.0092473 2 / s
k (calculated) = 0.089214
U0 = 3.9776 V
102
U/V
103
|uN ubattery|/ubattery
104
0 200 400 600 800 1000 1200
50
Q/As
100
The voltage response of the system in the laplace domain can be calculated using
equation 32. Hereby * is the convolution operator, dN (t) is the impulse response of the
approximated system in the laplace domain according to equation 17 and i (t) is the measured
time signal of the current.
Z (t ) = U 0 RE i (t ) + k filter (i (t ) , DZ ( z )) (33)
The voltage response of the system in the z domain can be calculated using equation 33.
The filter operation for filtering the current i (t) using a digital filter having the transfer func-
tion DZ (z) according to equation 26 is done by the filter () function.
Figure 8 shows the simulation results. No significant deviation between the simulation
results for uN and uZ was observed, the relative deviation was less than 105. The relative
deviation between the measured battery voltage ubattery and the output of the digital filter
uZ is shown in figure 9. Also the amount of charge Q charged or discharged is shown.
The relative deviation of the simulated voltage uZ has its largest values for large changes in
charge and for large currents and large current changes. The large deviations at large cur-
rent changes arise from model properties. The model is not able to simulate fast transients
better because in the high frequency range there is a big deviation between model and
battery impedance. The deviations at big changes in charge arises from the change of the
state of the battery due to the change in state of charge. This change is not considered by
changing model parameters. The model parameters are kept constant for the whole simula-
tion. At the end of the simulation the state of charge is the same as in the beginning. Also
19
6 CONCLUSION
In this article a method for time domain simulations of battery voltage was presented. The
method bases on a constant phase element model for diffusion that is approximated by
an ordinary differential equation and simulated using a digital filter. The used model has
only 4 model parameters and allows the simulation of battery voltage for at least 1000s at a
maximum relative voltage deviation of 1%. The computational effort is very small because
of the use of digital filters instead of convolution integrals. Further improvements of the
presented method can be made by adjusting model parameters with changing battery state
for example state of charge, temperature or age. The presented method is well suitable for
online prediction of battery voltage in a battery management system. This is for example very
important in electrical vehicles for predicting the available power of a battery.
REFERENCES
Hamann, C.H. and W. Vielstich (1997). Elektrochemie (3. Auflage ed.). Weinheim: Wiley-VCH.
Kanoun, O., U. Trltzsch, and H.-R. Traenkler (2006). Benefits of evolutionary strategy in modeling of
impedance spectra. Electrochimica Acta 51(89), 14531461.
Muralidharan, V.S. (1997). Warburg impedance-basics revised. Anti-Corrosion Methods and Materials
44(1), 2629.
Oustaloup, A., F. Levron, B. Mathieu, and F. Nanot (2000). Frequency-band complex noninteger differ-
entiator: Characterization and synthesis. IEEE Transactions on Circuits and Systems I: Fundamental
Theory and Applications 47(1), 2539.
Podlubny, I., I. Petras, B. Vinagre, P. OLeary, and L. Dorcak (2002). Analogue realizations of fraction-
alorder controllers. Nonlinear Dynamics 29(14), 281296.
Trltzsch, U., H.-R. Kanoun, and O. Traenkler (2006). Characterizing aging effects of lithium ion
batteries by impedance spectroscopy. Electrochimica Acta 51(89), 16641672.
van den Enden, A. and N. Verhoeckx (1990). Digitale Signalverarbeitung (1. Auflage ed.). Braunschweig:
Friedr. Vieweg & Sohn Verlagsgesellschaft mbH.
Warburg, E. (1899). Ueber das Verhalten sogenannter unpolarisierbarer Elektroden gegen Wechselstrom.
Annalen der Physik und Chemie 67(3), 493499.
20
A. Mahmoud
Universite de Toulouse, Mines Albi, CNRS, Ecole des Mines dAlbi Carmaux, Centre RAPSODEE,
81013 ALBI CT Cedex 09, France
ABSTRACT: The work describes preliminary studies for the application of impedance
spectroscopy for the determination of the packed bed conductivity. Electrical conductivity
of a packed bed varies with the mobility and affinity of ions with which solid particles are in
contact. For very low specific conductivity of the liquid phase, the specific conductivity of
the packed bed is enhanced by the presence of solid particles. In contrast, when liquid phase
conductivity is high, their contribution to electrical transport becomes more significant.
The paper concerns the investigation of the electrical behaviour of a packed bed, with
particular emphasis on bed conductivity determination. For this, different packed-beds have
been constituted using potassium chloride solutions, as liquid phase, and synthetic mineral
and organic suspensions materials (talc, kaolin, cellulose 101, 105, ), with different particle
size distribution, as solid phase. On the other hand, the case of copper sulphate solutions
using Dowex resins of different cross-linking degrees is treated also here.
First, the electrical conductivities of the packed beds were determined, using a model for
the conductivity of two-phase media, by measurement of the bed impedance in a lab cell with
two facing Pt sheets. Secondly, the conductivity of the bed was analyzed using a porous-
plug model to understand the current path. Choosing the magnitudes of the components
in accordance with the geometry of the dedicated model leads to quantitative agreement
between the model theory and the results.
1 INTRODUCTION
A renewed international interest appears during the last decade for electrical techniques
based on impedance/or electrical measurement. This type of technique has four main modal-
ities: electrical capacitance technique (ECT), electromagnetic technique (EMT), electrical
impedance technique (EIT) and electrical resistance technique (ERT), as shown in Table 1.
Each of these techniques has its advantages, disadvantages and limitations. The choice of
a particular technique is usually dictated by many, very often contradictory, factors. These
include: physical properties of the constituents of multiphase flow, cost of the equipment,
its physical dimensions, human resources needed to operate it, and potential hazards to the
personnel involved (e.g. radiation). In electrical capacitance technique or ECT (Beck, Byars,
Dyakowski, Waterfall, He, Wang, and Yang 1997; Gamio 1997; Plaskowski, Beck, Thorn,
and Dyakowski 1995), normally used with mixtures where the continuous phase is non-
conducting, the sensor employed is made of a circular array of electrodes distributed around
the cross-section to be examined, and the capacitance between all the different electrode-pair
combinations is measured. It shows the variation of the dielectric constant (or relative per-
mittivity (r)) inside the sensor area, thus providing an indication of the physical distribution
of the various components of the mixture.
23
plates
EIT (ERT) Impedance Permittivity Water/Saline
(Resistance) r = 100 102 biological tissue,
Z/R Conductivity ion-exchange resins
0.1 107 s/m rock/geological
materials,
semi conductor
Electrode Array
EMT Self/mutual Permeability Metals, some
Inductance r = 100 104 minerals magnetic
L/M Conductivity materials and
102 107 s/m ionized water
Coil array
The electrical current flow through packed bed is dependent on the conductivity of the solid
and the liquid phases. One of these phases will be the dispersed phase and the other the
continuous phases. Several theories have been proposed for the transport of electrical cur-
rent in mixtures (Helfferich 1962). However, most of the theories are based on the assump-
tions of either a regular lattice-type solid phase arrangement or a completely random
24
b
s = (1)
2 +
b
where is the conductivity of the continuous phase, b is the porous medium conductivity, s
is the volume fraction of the dispersed phase of conductivity , and is given by
1
= = with = (2)
+ 2 +1
The electrical conductivity of porous medium is complex because of the presence of two
conducting phases, namely the dispersed, and the continuous phases, .
According to Maxwell, the validity of Equation 1 is limited to small volume fractions.
However, various researchers have found that Equation 1 produced good agreement with
experimental data over a wide range of void fractions (Neal and Nader 1974; Turner 1976).
If the dispersed phase is assumed to be a non-conducting material, then Equation 9 can
be simplified as follows:
1 K
s = (3)
1 + 0.5K
K = (1 s)x (4)
where the value of the exponent x is 1.5, as originally determined by (Bruggemann 1935).
Reference (Rayleigh 1892) attempted to calculate the conductivity of regular arrays of
spheres when their interactions could not be neglected. The result of Rayleighs treatment,
after correcting a numerical error in the original paper, is
(5)
3 s
K = 1+
1 s 0.525 ( ) 110 / 3
s
which is strictly only applicable to when the spheres are arranged in cubic order, and is fur-
ther limited to moderate values of s.
Another theoretical solution that takes into account the effect of the presence of the
dispersed phase on the electrical flux field was proposed by (Meredith and Tobias 1960;
Meredith and Tobias 1961) for oil-in-water emulsions:
K=
( 21+ ) 2 s + 0 409 ( 64+33 ) s7 / 3 2.133( 34+23 ) s10 / 3 (6)
( 21+ ) s + 0 409 ( 64+33 ) s7 / 3 0.906 ( 34+23 ) s10 / 3
25
8(2 s ) (1 s )
K = (7)
( 4 + s )( 4 s )
However this equation has little advantage over Equation 4 which has an empirically deter-
mined value for the exponent x. The relative conductivity with respect to the volume fraction
of the dispersed phase has been plotted in Fig. 1, for non-conducting particles, together
with theoretical solutions by Maxwell, Rayleigh, Bruggeman and Meredith, respectively
(Helfferich 1962; Maxwell 1873; Neal and Nader 1974; Turner 1976; Bruggemann 1935;
Rayleigh 1892; Meredith and Tobias 1960; Meredith and Tobias 1961). In contrast to the
case of a porous medium of solid inert particles, the geometrical, electrical and chemical
properties of particles are sensitive to the chemical composition of the liquid (continuous
phase) as adsorption equilibrium prevails. First attempts in modeling often involved regular
lattice arrangements, as that suggested by Baron (Helfferich 1962) who considered a statisti-
cal cage model. However, most approaches were not valid for solid particles in contact with
one another. A more realistic model was developed even earlier (Wyllie and Spiegler 1955;
Mahmoud, Muhr, Grvillot, Valentin, and Lapicque 2006). It relies upon the porous-plug
model described below. The electrical current is considered to pass through three different
paths within the bed (Fig. 2(A)): (i) through alternating layers of particles and interstitial
Figure 2. The porous-plug model. A: schematic representation of the three paths which the current
can take. B: the simplified model consisting of three conductance elements in parallel. C: Extension of
the same model.
26
a
b = 1 + 2 + 3 = + b c (8)
d e
where the parameters a, b and c are the fractional cross-section equivalents of the three
elements:
a+b+c=1 (9)
The ( ) term in Equation 7 expresses the contribution of the liquid, with weight
d, and of the solid, weight e in the first conduction process. Parameters d and e obey the
relation:
d+e=1 (10)
For a packed bed, the parameters of the model must be estimated to predict the electro-
chemical properties of the dispersed phase.
The different materials analyzed in the experiments can be classified into synthetic suspen-
sion (mineral and organic) and copper sulphate solutions using Dowex resins. These materi-
als are presented as follows:
27
28
where is the pulsation, j the complex operator, Z the resistive component and Z the reac-
tive component. The resistance value of the medium investigated can be deduced from the
intercept of the curve (Nyquist diagram) or with the real axis. Therefore the intercept of the
impedance spectrum recorded at rest potential with the real axis led to the ohmic resistance,
R, and the conductivity of the medium, , was deduced using the relation:
l
= (12)
RS
where l is the electrode gap (2 cm), and S the electrode area (2 2 cm2). Nevertheless after
calibration, the l/S ratio was taken at 0.52 1/cm. Besides, former experiments conducted with
packed beds of inert materials with conducting solutions allowed estimation of parameter c.
Numerous measurements conducted with the softest model materials led to c = 0.33 (0.28
for ion-exchange resins). This value agrees well with the value obtained by (Raghavan and
Martin 1995) from modelling of heat transfer in packed beds, with a liquid void fraction of
40%. Then, for these model materials, only two independent parameters had to be deter-
mined amongst the series a-e.
Figure 5. Variation of the bed conductivity with the solution conductivity. Dotted lines are for
theoretical results.
exchange resins are more conductive than the solution to be treated, as expected. These res-
ins will significantly reduce the device resistance and increase the available surface area for
ion exchange, which is beneficial when performing continuous electro-deionisation operation
with dilute solutions. Although the porous-plug model illustrates the pattern of the ionic
flux, the conductivity of the backed bed is an important operating variable for the sake of
maximum ion flux.
30
H+ 26.5 26.6
Dowex 2% 50 100
Cu2+ 2.9 3.1
H+ 26.5 26.6
Dowex 4% 50 100
Cu2+ 2.9 3.1
H+ 26.5 26.6
Dowex HCR-S 8% 20 50
Cu2+ 2.9 3.1
Cellulose 105 20 0.72103 0.77103
Cellulose 101 60 5.65103 6.05103
Microchrystallin
cellulose (MMC) 40 0.62103 0.68103
Kaolin 6 2.10103 2.17103
Talc 11 0.47103 0.45103
Figure 6. Representation of the ratio between the solid and solution phases; grey blocks are solid
phases and white blocks are solution phases. Current flows vertically in the representation.
In case of the synthetic materials, the current is mainly transported through (i) the alter-
nating layers of particles and solution, and (iii) in the channel of the liquid phase. Therefore,
the synthetic materials used in this study are less conductive than the solution to be treated,
as expected. The conductivity parameters were used to predict the conductivity of synthetic
material beds impregnated with a 1 M potassium chloride solutions. Conductivity of the
31
e
fe = (13)
b
Where e and e are the specific conductance of the element and the total plug respectively.
Hence, the fraction of current carried by element (i) equals:
1 d ae
f1 = = (14)
b b
Figure 7. Representation of the fraction of current carried by elements (i), (ii) and (iii).
32
2 b 3 c
f2 = = ; f3 = = (15)
b b b b
As shown f1, f2 and f3 are given in terms of geometrical parameters determined from con-
ductance measurements. Fig. 7 represents the fraction of the current passing through the
elements of the backed bed.
As can be seen from Fig. 7, it is clear that more than 20% flows through the resin phase
and more than 70% of the current flows through both the solution and the resins phase
alternately based on the porous plug model. In case of synthetic materials, more current is
expected to flow through the first and the third paths.
5 CONCLUSION
REFERENCES
Al-Zeibak, S. and N.H. Saunders (1993). A feasibility study of in vivo electromagnetic imaging. Phys.
Med. Biol. 38, 151160.
Beck, M.S., M. Byars, T. Dyakowski, R. Waterfall, R. He, S.M. Wang, and W.Q. Yang (1997). Princi-
ples and industrial applications of electrical capacitance tomography. Measurement and Control 30,
197200.
Beck, M.S., A. Williams, and R.A. Williams (1995). Process Tomography: Principles, Techniques and
Applications. Butterworth-Heinemann.
Bond, J., J.C. Cullivanu, N. Climpson, I. Faulkes, X. Jia, J.A. Knstuch, D. Paylon, M. Wang, S.J. Wang,
R.M. West, and R. Williams (1999). Industrial monitoring of hydrocyclone operation using electrical
resistance tomography. Mineral Engineering 12, 12451252.
Brown, B. (2001). Medical impedance tomography and process impedance tomography: A brief review.
Measurement Science and Technology 12, 991996.
Bruggemann (1935). Berechnung verschiedener physikalischer konstanten von heterogenen substanzen.
i. dielectrizittkonstanten und leitfhigkeiten der mischkrper aus isotropen substanzen. Ann. Phys.
24, 636679.
Gamio, J.C. (1997). A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System.
Ph.D. thesis, Institute of Science and Technology, University of Manchester.
Helfferich, F.G. (1962). Ion Exchange. McGraw-Hill.
Mahmoud, A., L. Muhr, G. Grvillot, G. Valentin, and F. Lapicque (2006). Ohmic drops in the ion-
exchange bed of cationic electrodeionisation cells. Journal of Applied Electrochemistry 3, 277285.
Maxwell, J.C. (1873). A Treatise on Electricity and Magnetism. Clarendon Press, Oxford.
Meredith, R.E. and C.W. Tobias (1960). Resistance to potential flow through a cubical array of spheres.
Journal of Applied Physics 31, 12701273.
Meredith, R.E. and C.W. Tobias (1961). Conductivities in emulsions. Journal Electrochemical Society
108, 286290.
33
ABSTRACT: Various metals and their alloys can be characterized, among other meas-
ures, also by their electrical conductivity value, for DC and AC. Precise measurements on
different frequencies can be used to validate metals, their alloys and their mechanical and
physical structure. Specially developed instrumentation (hardware and software) for meas-
uring precisely (with error below 1%) the conductivity of specimens with dimensions of
80 mm*80 mm*3 mm is described. The measurement tolerances are estimated by simulations
(in COMSOL Multiphysics environment). Also practical measurements with different mate-
rials have been carried out in the conductivity range of about 2.5 ... 60 MS/m and frequency
range from DC to 100 kHz AC.
1 INTRODUCTION
One important parameter of a metal (or alloy) material is its electrical conductivity (Rossiter
1991). Validation of materials is one important application of electrical conductivity meas-
urements. By measuring on various frequencies, physical and structural non-homogeneity
of the material under test can be studied. In current study, one objective has been to design
a measurement probe and instrumentation for conductivity measurements of high con-
ductivity metals with a very high accuracy (less than 1% error). The size of the samples is
80 mm*80 mm and the thickness can vary from sample to sample (210 mm).
35
Figure 2. All 4 electrodes are conical tip needle electrodes with spring and 30 deg tip angle. A cable is
attached to the tip area of the probe. Typical wobble of needle tip is shown on the right.
Figure 4. The results of temperature simulation showing temperature in the electrode tip cross-section
at 1 ms, 4 ms, 7 ms and 10 ms after the start of 1 A current feed.
We answer the second question with a current density calculation combined with heat-
transfer calculation in the conical contact tip. We model the sample in this case as a much
smaller objectnamely coin-sizeand only run calculations on one half (half coin, 1 elec-
trode feeds current, current distribution is assumed symmetrical in the other coin-half).
1 A current is entered through a small cone that has 22 m contact diameter and led into the
sample. The modeling domain (half-coin and conical electrode tip) is discretized into 26000
elements and the coupled electricand induction currents with heat transfer calculation is
performed with 290000 DOF. The element size is also very variable here, being extremely
small in electrode contact and largest in edges of the metal sample (see Figure 3). The element
volume varies 2e10 (20 billion) times in this case.
The results of this simulation show a current density of 8 GA/m2 (8e9 A/m2) in the edge of
the contact area. With a measurement that lasts 10 ms, the temperature in the conical electrode
tip increases by only 0.64 degrees. Due to very high thermal conductivity of the materials, the
energy generated in the very edge of the contact area is immediately dissipated into the contact-
probe and the metal sample. Based on the simulation, it is safe to say, that if the current were
10 A and contact area half that of the modeled diameter 22 mthe thermal conductivity
would still dominate over the energy density in the contact region. The current density is high-
est in the edges of the contact (the 22 m contact diameter). With higher frequency the energy
would be even more concentrated to the edges of the contact, but thermal conductivity would
still dominate and would be able to dissipate the heat into the sample and up into the electrode.
Therefore we do not expect to find any problems arising from the heating in the contact.
37
Figure 6. The frequency characteristic of an actual measurement with the developed hardware, just
like it is shown real-time with LabView software. The metal sample used in this preliminary test had a
high copper-scale conductivity.
A special test fixture for holding metal samples for 4-electrode probe has been developed
together with hardware and software for high-accuracy conductivity measurements.
The measurement principle used is synchronous detection and is realized in LabView
software environment on PC. In order to get the best signal to noise ratio, the excitation
waveform is single sinusoid. USB-6259 (multichannel 16 bit, up to 1 Ms/s analog-to-digital
and digital-to-analog converters) interface box from National Instruments acts as a bridge
between the LabView software and custom electronics.
The developed electronics part consists of a discreet power amplifier, with up to 5 A of
output current capability and bandwidth from DC up to 1 MHz and low-noise voltage ampli-
fier, with programmable gain up to 20000. Configuration of the developed interface box is
done via USB. In addition, since the impedance of metals is temperature dependant, the
temperature of the metal sample is measured using Pt100 temperature sensor. The need for
high excitation current and high sensitivity of the voltage pick-up section poses several diffi-
culties in the realization of the measurement system. Sensitive sections can be easily shielded
from various capacitive interferences by using Faraday cage. Unfortunately this method is
not efficient against strong electromagnetic field, which emanates from the excitation section
38
consisting of the power amplifier, distribution wires, probes and the device under test itself.
This tends to interfere into the sensitive pick-up amplifier and cause errors that cannot be
rejected by synchronous detection as the interfering signal has the same frequency as is used
for measurement. To reduce this effect, the area of the frame created by the current loop is
kept as small as possible and the fixture with probes and pickup amplifier is placed further
from the power amplifier. The current is fed to the sample via a frame and not by cables since
this helps to keep the interference more constant making it feasible to be compensated.
The voltage pick-up probes are connected to the high-gain measurement amplifier using
custom made twisted pair cable to ensure that the interferences picked up by both leads are
equal and rejected by PSRR of the input amplifier.
The high-accuracy measurement system for measuring electrical conductivity of metal plates
has been developedfirst the important aspects of measurement have been simulated and
then a practical experimental solution has been realized using a 4-electrode measurement
probe. Both simulations and experiments show that the accuracy of 1% can be achieved
for DC and for AC low frequencies (at 10100 Hz). At higher frequencies significant chal-
lenges come up for simulation as well as the instrumentation (similar, as described in (Hall,
Henderson, Ashcroft, Harmon, Warnecke, Schumacher, and Rietveld 2004)). Induced para-
sitic signals are observed at the test fixture and electronics as well as skin-effect in the sample.
We are currently working on the next and considerably improved version of the system,
where we expect to overcome or compensate some of the shortcomings in the high frequency
capability of the system. It seems necessary to simulate the electro-magnetic fields and inter-
ferences that they cause during the development of the next generation measurement device.
The software environment chosen for that task is CST STUDIO SUITE.
As an alternative, we are also working on eddy current methods to measure the proper-
ties of metals and alloys in wider frequency range (up to 500 kHz). One of our goals is to
39
ACKNOWLEDGEMENT
We would like to thank O Baitaim for the excellent work on physical realization of the
measurement test fixture. Current work in this field (development and investigation of sig-
nal processing solutions for AC and other measurements) has been supported by Enterprise
Estonia (support of Competence Center ELIKO), Estonian IT Foundation (establishment
of the sensor-signal processing chair at Tallinn University of Technology), grant ETF7243
and target financing SF0142737s06 and excellence center CEBE at Tallinn University of
Technology.
REFERENCES
Bowler, N. and Y. Huang (2005a). Electrical conductivity measurement of metal plates using broadband
eddy-current and four-point methods. Meas. Sci. Technol. 16, 21932200.
Bowler, N. and Y. Huang (2005b). Model-based characterization of homogeneous metal plates by
four-point alternating current potential drop measurements. IEEE Transactions on Magnetics 41,
21022110.
Hall, M., L. Henderson, G. Ashcroft, S. Harmon, P. Warnecke, B. Schumacher, and G. Rietveld (2004).
Discrepancies between the dc and ac measurement of low frequency electrical conductivity. In 2004
Conference on Precision Electromagnetic Measurements Digest.
Koon, D.W., A.A. Bahl, and E.O. Duncan (1989). Measurement of contact placement errors in the van
der Pauw technique. Review of Scientific Instruments 60, 275276.
Rietveld, G., L.C.A. Koijmans, Ch. V. and. Henderson, M.J. Hall, S. Harmon, P. Warnecke, and
B. Schumacher (2003). Dc conductivity measurements in the van der pauw geometry. IEEE Transac-
tions on Instrumentation and Measurement 52, 449453.
Rossiter, P.L. (1991). Electrical Resistivity of Metals and Alloys. Cambridge University Press.
Schroder, D.K. (2006). Material and Device Characterization. Wiley-IEEE Press.
40
A. Pokatilov
AS Metrosert, Tallinn, Estonia
A. Kolyshkins
Riga Technical University, Latvia
1 INTRODUCTION
43
Specimen- material pieces under test, has been used of plate shape, with size 80 80 mm2 area
and 3 mm thick. Industrial LCR meters has been used for measurement of the coil impedance.
Different coils have been tested the best results were obtained with the planar (PCB)-coil with
N = 80 turns, diameter D = 25 mm (Fig. 1). A software developed by Dr. Theodoros P. Theo-
doulidis and his colleagues (described also in [7]) has been converted into C/C++ software -by
using freeware C/C++ compilers and free libraries for complex calculations, Bessel, Struve
and other required functions. Furthermore, an iterative reverse task solver has been added
into the C/C++ software, making it possible to estimate both actual lift-off and conductivity
of the material under test from measured complex impedance of the coil.
44
Experimental investigation shows, that for described setup and software models, accuracy of
conductivity measurement better than 10% is achieved, in the frequency range up to 500 kHz.
One example set of measurement results is given in the Tables 1 and 2 below.
Combining theoretical and experimental studies and development and experimental works
show, that with single planar coil measurement probe, absolute measurements can be carried
out, with accuracies, much better than 10%, for non-magnetic materials with conductivities
from 2.5 to 15 MS/m, in the frequency ranges up to 500 kHz. Measurement coil, setup (instru-
mentation) and software for conductivity measurements (and simultaneously lift-off estimation
and compensation) and measurement results has been given in the paper. Further improvement
of accuracy is related to more precise using of the additional correction factors- e.g. from finite
size of the specimen, parasitic capacitance and other non-idealities of the measurement coil.
ACKNOWLEDGEMENT
Current work in this field (development and investigation of signal processing solutions
for AC and other measurements) has been supported by OLAF (the European Anti-Fraud
Office), Mint of Finland, Enterprise Estonia (support of Competence Center ELIKO), Esto-
nian IT Foundation (establishment of the sensor-signal processing chair at Tallinn University
of Technology), grant ETF7243 and target financing SF0142737s06 and excellence center
CEBE at Tallinn University of Technology.
Special thanks to Dr. Theodoros P. Theodoulidis and his colleagues for providing of very
valuable software for eddy current calculations (simulations) and Mr. Douglas Reid and
Dr. Toomas Kbarsepp for practical suggestions.
REFERENCES
Bowler, N. and Y. Huang (2005). Electrical conductivity measurement of metal plates using broadband
eddy-current and four-point methods. Meas. Sci. Technol. 16, 21932200.
Dodd, C.V. and W.E. Deeds (1968). Analytical solutions to eddy-current probe-coil problems. Journal
of Applied Physics 39, 28292838.
Dodd, C.V. and W.E. Deeds (1975). Calculation of magnetic fields from time-varying currents in the
presence of conductors. Technical report, Oak Ridge National Laboratory.
Hall, M., L. Henderson, G. Ashcroft, S. Harmon, P. Warnecke, B. Schumacher, and G. Rietveld (2004).
Discrepancies between the dc and ac measurement of low frequency electrical conductivity. In 2004
Conference on Precision Electromagnetic Measurements Digest.
Rossiter, P.L. (1991). Electrical Resistivity of Metals and Alloys. Cambridge University Press.
Snyder, P.J. (1995). Method and apparatus for reducing errors in eddy-current conductivity measure-
ments due to lift-off by interpolating between a plurality of reference conductivity measurements.
Theodoulidis, T.P. and M.K. Kotouzas (2000). Eddy current testing simulation on a personal computer.
In 15th World Conference on Nondestructive Testing Roma (Italy).
45
O. Kanoun
Fakultt fr Elektro- und Informationstechnik, Technische Universitt Chemnitz, Germany
J. Himmel
Laboratory for Sensor Technology and Measurement Engineering, University of Applied Sciences,
Koblenz, Germany
ABSTRACT: For the rod shape measurement of hot rolled round steel bars (rods) the high
frequency eddy current method is especially well suited as it requires no contact point and
is not limited to bellow the Curie Temperature. Defects of the rods shape can be detected
by measuring the impedance spectrum of the RLC-oscillator. In the first laboratory setup
(Weidenmller, Knopf, Sehestedt, and Himmel 2008) an Agilent Impedance Analyser was
used for initial tests. Nevertheless, this setup cannot be applied in a steel plant due to the
difficult environmental conditions and low data acquisition rate. Hence, a Vector Network
Analyser for passive impedance measurement that is applicable in these surroundings was
developed, (Weidenmller, Sehestedt, Kanoun, and Himmel 2009).
1 INTRODUCTION
In measurement and sensor technology the eddy current principle is one of the most promis-
ing methods for noncontact detection of material properties, cracks, distances and pressures
(Weidenmller, Knopf, Sehestedt, and Himmel 2008). Currently, this highly industrial-suited
principle is not in use for shape measuring for electroconductive semifinished products, like
rods. Three typical defects of ovalities, which occur most frequently during the rod produc-
tion process, are shown in Figure 1. In order to avoid rejections these typical defects should
be detected in the production process as the rolls and the rods velocity in the production line
can be readjusted immediatly. This requires a dynamic measurement system for rods veloci-
ties up to 70 m/s. Currently, in rod shape testing applications measurement systems based
on optical sensors are used. Integrating such optical systems into places with harsh environ-
mental conditions, especially close to redhot rods, is difficult. Whereas measurement systems
based on high frequency eddy current are much more robust. Also the earlier designed mate-
rial tracking sensor based on the same high frequency eddy current method confirmed the
good industrially applicability of this measurement principle. However, the evaluation elec-
tronics implemented in these sensors is not transferable for the rod shape testing application.
In order to reconstruct the rod shape, information about the complex impedance spectrum
around the resonant frequency of the tuned parallel RLC-oscillator are required. Hence,
a robust impedance measurement system was developed as the Agilent Impedance Analyser
4294A used in the laboratory setup is not applicable for tests in a steel plant, Wei08.
47
2 PHYSICAL BACKGROUND
The measurement principle can be most easy described using the transformer model (Vyroubal
2004). In this model a primary winding, the detection coil, carries the time-variable current. The
time-variable current causes
an alternating rotational field, which can be described in terms of the
magnetic vector potential A. The magnetic vector potential produces a circular electric field.
E j A (1)
Hence, the induced eddy current on the rod surface is circular as well and has no radial
component.
i (t )eddy = EdA (2)
A
As a result, the secondary part of the transformer model, the rod, is characterized with
many single turn windings which are fed by respective circular eddy currents. The primary
inductance Lc and the secondary inductance Ln are coupled through a shared magnetic field.
Consequently, the measured inductance L can be described in terms of the primary induct-
ance Lc and the mutual inductance Mc,n caused by the rod.
(3)
L Lc (Mc , Mc ,n + Mc ,n ) Lc + Mc ,n
1 2Rc2C p Rc2C p
res = +1 (4)
L C p L L
To gain a better understanding Figure 2 displays the resonance curves for different rod
positions inside a detection coil, starting at centre position moving outwards in steps of 1 mm.
In this setup the rod diameter is 20 mm and the inner coil diameter is 121 mm. As expected,
due to eddy current losses resulting in rising mutual inductance Mc,n, the effective inductance
Ldrops for declining gaps between the coil and the rod (Eq. 3). Due to the interrelationship
given by equation 4 the detected resonant frequency increases, (Simonyi 1980; Philippow
2000; Zinke and H. Brunswig 2000).
3 PRELIMINARY CONSIDERATIONS
In the material tracking application, which was successfully tested at the DEW steel plant
(Fig. 3), the resonant frequency was detected using a frequency counting system integrated
48
into a microcontroller (Himmel, Arend, Otto, and Sehestedt 2006). This method is suitable
for material tracking but initial setups indicate that a more extensive data set is required for
the rod shape testing application. Therefore a measurement system which also provides infor-
mation about the complex impedance spectrum near the resonant frequency was developed.
For the first experimental setups an Agilent Impedance Analyser 4294A was used (Fig. 2).
However, this device is not applicable for such harsh environmental conditions as occurring
in steel plants and also not capable of a measurement speed that is sufficient to support proc-
ess control. Additionally, the advantages such as the wide frequency and impedance range
and further analysis options are not required. Consequentially, a new customized HF Vector
Network Analyzer (HF VNA) will be implemented into the sensor. The main idea in develop-
ing a new HF VNA is to separate the measurement part, including source, bridge and detec-
tor, from the control part, that drives the signal source and provides the data collection and
processing part. This makes the HF VNA cost efficient and suitable for applications close
to hot rolled steel bars as no PC can be located in the immediate vicinity of the production
process, (Kiciak 2008a; Kiciak 2008b).
49
The HF VNA used is based on the measurement of the reflection, the Agilent 4294A on the
other hand side utilizes an RF I-V method. The RF I-V method is suitable for a wide range in
impedance as well as in frequency but the reflection measurement has comparable sensitivity
in a narrow range around the null balance point that the bridge is matched for. In this chapter
the key elements of the designed VNA will be introduced, which are:
A software controlled synthetized signal source.
A bridge to separate incident from reflected signals.
A detector capable of amplitude and phase detection.
Controls, data collection, and processing.
The Device Under Test (DUT), the RLC-oscillator.
OSL calibration standards (Open, Short, Load).
50
measurement starts using the OSL-calibration standards. During the production process it
is not possible to connect the OSL calibration standard to the bridge manually, hence a soft-
ware controlled switching system using a microcontroller will be implemented. In this way
the VNA can be calibrated via software before the process starts.
4.3 Detector
Basically, the VNA detector is realized using a linearized Gilbert cell mixer (MC1496), analog
filtering and buffering, a 24 bit analog to digital converter (LTC2440CGN), and a precise and
temperature stable 2.5 V reference for the ADC (Fig. 4). Consequentially, the theoretical res-
olution of the ADC is approximately 0.149 V. The conversation speed of the LTC2440CGN
is adjustable with the fastest performance being 3.5 kHz. As both DDSs are programmed
with the same frequency, the voltage at the detector output is DC and contains the desired
carrier information. Please consider, this is only obtained by using a synchronised LO DDS.
Otherwise if the LO DDS is asynchronous with its RF DDS the information is contained in
the detected sidebands. As the phase of the LO DDS can be programmed to values of 0 and
90, the detector output takes on two DC values that together represent the vector compo-
nents of the applied RF signal, as shown in equation 5 and 6.
VDC ,0 GDE
DETT VRF
RF ( RF DET ) Vof
ff ,0
off (5)
b1 Vreflection
S11 = = = (7)
a1 Vtowards
In this case the measurement system consists of just one port, therefore the S11 parameter
equals the reflection coefficient . Hence the complex impedance can be determined with
equation 8.
+1 (8)
Zx Z0
1
As mentioned above, the reflection measurement has high impedance measurement sensi-
tivity around the null balance point. Thus a 50 bridge is applied, as the expected impedance
lies within the high sensitivity range between 10 and 200 (Fig. 7). If the impedance antici-
pated is out of that range the bridge needs to be matched to the range of interest.
52
5 RESULTS
Figure 8 displays the resonance curve of the RLC-oscillator measured with different VNAs.
The inductance of the detection coil in use is 240 nH and the desired resonant frequency of
about 29 MHz is tuned with a variable parallel capacitor. In order to compare both meas-
urement systems this setup is first tested with the Agilent 4294A. After a warm-up time of
about 15 minutes of the new designed VNAs, which is recommended in order to gain stable
temperature conditions, the four VNAs are tested in the frequency spectrum between 20 and
40 MHz. Whereas the ADC sample rate in one version is set to 7 Hz, the so called slow ver-
sion, the other three ADCs are set to the fastest performance of 3.5 kHz. As mentioned in
a previous chapter, this high speed performance is required in rod shape testing. As a result,
the new designed VNAs yield good precision but a static offset of about 3 ohm compared to
the Agilent 4294A. Possible reasons therefore are:
Temperature drift
The use of different calibration standards
Unshielded experimental setup
Those possible reasons will be investigated in the next step of development.
6 CONCLUSION
The new high frequency eddy current rod shape testing approach has several advantages
compared to currently in use sensor systems. The promising test results using the labora-
tory setup (Weidenmller Knopf, Sehestedt, and Himmel 2008) can be verified in a steel
plant as the designed HF VNA is applicable into places with harsh environmental condi-
tions. Due to the compact and robust front-end electronic the HF VNA can be located
close to the detection unit which is necessary in order to gain stable measurement condi-
tions. Also the data acquisition speed is sufficient to support process control. In the next
step of development the frond-end electronic will be placed into a temperature controlled
and shielded box. Additionally, the OSL calibration standards will be automatically con-
nected to thereflection bridge via a software controlled switching system. After comparing
53
ACKNOWLEDGEMENTS
REFERENCES
54
ABSTRACT: Elderly people and geriatric patients often suffer from dehydration which can
be diagnosed by bioimpedance spectroscopy (BIS). In this work, the capability of segmen-
tal BIS to reflect the total body composition is analyzed. Therefore, measurements of body
segments have been simulated using the finite integration technique (FIT) and the results
have been verified by measurements on subjects. We show that results obtained from seg-
mental BIS measurements correlate with wholebody bioimpedance spectroscopy measure-
ments. Furthermore, simulations of segmental BIS have been successfully accomplished and
the simulation results show similar characteristics compared to measurements performed for
reference.
1 INTRODUCTION
Since the beginning of the 20th century, a demographic change is observable in Europe which
leads to a steadily aging society. Looking at Germany as an example, the average life expect-
ancy of women is supposed to rise from 74.2 to 87.1 years between 1913 and 2035. Important
factors are the continuous progress in medical technology and the decreasing birth rate of
actually around 1.4 children per couple (Statistisches Bundesamt 2007). Hence, society will
have to face more and more geriatric patients, which leads to additional costs and burdens
on medical personnel. Thus, it is reasonable to establish methods to treat elderly people more
cost-effectively and more easily by improving diagnostics for certain diseases that prevalently
occur among elderly people.
One of these diseases is dehydration, which describes a water deficiency in the body. This
deficiency is primarily caused by a lower thirst sensation and a disturbed hormone balance
(Lavizzo-Mourey 1987). Also athletes can suffer from dehydration during physical exercise,
which can lead to lifethreatening situations. The project NutriWear, funded by the German
Federal Ministry of Education and Research (BMBF), aims at developing a wearable system
to monitor the nutrition and hydration status based on intelligent textiles. What is more, the
patient shall be monitored 24 hours per day (Bundesministerium fr Bildung und Forschung
2007). Such a system can be used for preventive monitoring of dehydration.
Dehydration can be easily and cost-effectively diagnosed by bioimpedance spectroscopy
(BIS). Currently, BIS is not commonly used as diagnostic method because it is not considered
to be valid, even though there are a lot of studies implying the opposite (Wabel 2009). One
possible reason is that many processes in the human body during BIS are widely unknown.
One way to analyze where the current paths run, and which tissue contributes significantly
to the measurement result, is to use computer simulations employing FIT (Clemens 2001).
In the following, correlations of segmental BIS and wholebody BIS measurements shall be
analyzed and compared with simulations of segmental BIS measurements.
57
The basic idea of BIS is that each cell consists of a capacitive cell membrane, separating extra-
cellular and intracellular space. When injecting an alternating current into a certain tissue, these
membranes lead to a current flow between the cells through the extracellular water (ECW) at low
frequencies. At high frequencies, the membranes are no barrier for the current so that it flows
through all cells now as a function of ECW as well as intracellular water (ICW) (cf. Fig. 1).
There are three major dispersion regions: (mHzkHz), (0.1100 MHz) and
(0.1100 GHz). The frequency range between 1 kHz and 10 MHz within the dispersion
region is generally the most interesting one for diagnosis because physiological and pathophysi-
ological processes lead to changes in body impedances with high dynamics in this range. The
reason for this is polar proteins and organelles which behave like dipoles in the alternating field
(Grimnes and Martinsen 2000). In addition, in this frequency region safety regulations permit
higher alternating currents than in the frequency range below 1 kHz. Thus, the degree of meas-
urement accuracy is kept at an optimum. BIS is a method for measuring bioimpedances using
a frequency spectrum between 5 kHz and 1 MHz with a current between 500 A and 10 mA.
Usually, whole-body BIS measurements are accomplished using one electrode pair to
inject the current and one electrode pair to measure the voltage at wrist and ankle so that four
electrodes have to be used for one measurement. The resulting impedance curve represents a
complex cole-cole curve (cf. Fig. 2, left).
The model which fits very well to reality is represented by the equivalent circuit for this
plot shown on the right side of Fig. 2, whose impedance can be described by the following
formula. Here Cm is the membrane capacity, Re the extracellular, Ri the intracellular resistance
and R = Re||Ri (Grimnes and Martinsen 2000).
Re R
Z ( ) R + (1)
1 + ( j Cm (Ri Re ))
Figure 2. Whole-body BIS measurement, typical impedance progression and equivalent circuit diagram.
58
Here KB
is a dimension less shape factor calculated from an atomic measurements,
e and tbw are the specific resistivities of the extracellular and overall fluid, W is the body
weight and H the body height (Matthie 2005).
Approximating the human body as one cylinder is of course an assumption which leads to
modeling errors. These errors can be reduced by sub-dividing the body into several segments
(Zhu 2006). Segmental and whole-body measurements have been compared to reference
methods for detecting the body composition, such as Dual Energy X-Ray Absorpiometry
(DEXA), 40K or D2O solution method.
DEXA technology uses two different x-ray sources for emitting x-rays through the human
body. A detector receives the non-absorbed radiation which depends on the density of the
irradiated tissue. Thus, bone density and body fat can be calculated. Comparing segmental
and whole-body measurements with fat free mass (FFM) assessments by DEXA, both bio-
impedance methods generate correct values. In addition, BIS measurement results of single
segments show high correlations with results of DEXA analysis of the same segments (Salmi
2003; Bracco 1996).
The second reference method introduced here is the 40K method. 40K is a radioactive potas-
sium isotope which is administered orally. It preferably accumulates in body fat and there
emits x-rays when decaying. ECW and ICW measured by segmental BIS and the 40K method
show good correlations (Carter 2007).
The other substance taken as reference method is deuterium oxide (D2O), also called heavy
water, which is taken orally and is distributed in all body tissues except fat. By taking a blood
sample and measuring the D2O content in plasma, the total body water can be calculated.
Segmental measurements are superior to whole-body measurements when detecting the
hydration status compared to D2O measurements (Thomas 2003).
It has been also shown without reference measurements, that thoracic segmental BIS
measurements are better than wholebody measurements concerning the hyperhydration
detection for dialysis patients (Nescolarde 2007). To sum it up, the reason why segmental
measurements and simulations have been accomplished is that several studies showed that
segmental BIS produces reliable and in some cases even better results than conventional
whole-body BIS.
3 METHODS
Classically, the body is divided into five segments: arms, legs and thorax. However, within
this work the body was divided into nine segments: thighs, knees, lower legs, arms and thorax
59
300000 km/s
c= = = 60 m (4)
f 5 MHz
This leads to a problem concerning the imaginary part of the complex current because the
boundary conditions for electroquasistatic and magnetoquasistatic cases, respectively, are:
D B
=0 a d =0 (5)
t t
60
The material of the borders of the cubic simulation domain have been set to PEC. In addi-
tion, the interspace between simulation object and boundaries has been filled with vacuum
(permittivity r = 1, conductivity = 0). After having set the mesh density to create 2 million
tetrahedrons, the simulation consumed 5 GB RAM and took 25 hours per sweep for the lower
extremities. A whole body simultation would have taken 72 hours per frequency. Apart from
knee, thigh and lower leg, knee-to-knee and foot-to-foot measurements have been simulated
additionally as further scenarios. The impedances obtained by the simulations have been ana-
lyzed in Matlab to compute extra- and intracellular resistances. The calculation frequency
ranged from 5 kHz to 5 MHz, covering the measuring range of common BIS-analyzers.
For validation purposes, each segment has been measured at five male subjects with similar
age (2628 years) and Body Mass Index (2125.5). The measurement device was a Xitron
Hydra 4200 from Xitron Technologies, San Diego, USA. Standard adhesive electrodes from
Fresenius Medical Care, Bad Homburg, Germany with a size of 19 mm 80 mm have been
used. All subjects were not allowed to drink and eat two hours before the measurements
began. The subjects changed from an upright to a supine position remaining supine for
30 minutes so that a fluid shift from the lower extremities to the rest of the body occurred.
To gather information about this shift, whole-body bioimpedance has been obtained after
each segmental BIS measurement (Medrano 2009).
4 RESULTS
Before the human model has been built and used for simulation, the ability of the simulation
program to simulate BIS has been tested by creating a capacitor with a muscle dielectric. This
capacitor was intended to reproduce the measurements accomplished by Gabriel et al. who
measured permittivity and conductivity of human tissues between 10 Hz and 1 GHz using
network analysers. The results showed that the CST software could correctly reproduce the
measurements by Gabriel et al. with, , and dispersion regions.
The model
used for the simulation results is shown in Fig. 4.
The D-Field results of a foot-to-foot measurement simulation are presented here by arrows
floating through the body from the left foot to the right foot. Thickness and direction of the
arrows symbolize the displacement current density and its direction.
Simulation results were analyzed using frequency response locus plots which resemble
cole-cole curves (cf. Fig. 2, left), although the curve progression deviates slightly. Further-
more, the calculated values of the simulation results yield higher impedances in relation to
61
Re,sim Re,literature
Segment () ()
62
5 DISCUSSION
The task of this work was to derive correlations of segmental BIS and whole-body BIS meas-
urements and to compare them with simulations of segmental BIS measurements. Promising
results have been obtained comparing measurements and simulations.
First, it hast been shown by segmental BIS measurements that intra- and extracellular resist-
ances of segments correlate with whole-body resistances. The fluid shift from the lower extremi-
ties to the rest of the body during these measurements is explainable due to the position change
of the subjects (Rush 2006). Second, results of the measurements reveal on the one hand that
extremities have the highest relative share on the wholebody resistances which is explainable
because they contain joints which are composed of low conducting bone tissue. On the other
hand, the thighs contains much more good conducting material and thus they have the lowest
influence on the whole-body resistance. Third, comparing measurements and simulations, this
work also showed that simulations offer a reliable method for imitating processes in the body
during BIS because they deliver the same dependencies of segments and the whole body.
However, some future work has to be done. Reasons for the simulation model to produce
too high absolute impedances should be analyzed further to consider it in future simulations.
The main reason could be the coarse model resolution of 8 8 8 mm that leads to inaccura-
cies. During the conversion process small structures can be lost. Combined with some volumes
such as small blood vessels (e.g. arterioles), which are too small to be considered in the model,
63
ACKNOWLEDGEMENTS
This work was funded by the German Federal Ministry of Education and Research within
the program Micro Systems.
REFERENCES
Bracco, D. (1996). Segmental body composition assessed by bioelectrical impedance analysis and dexa
in humans. Journal of Applied Physiology 81, 25802587.
Bundesministerium fr Bildung und Forschung (2007). Gesundheitsforschung hautnah erleben. Web-
site. http://www.bmbf.de/press/2171.php.
Carter, M. (2007). Estimation of Arm Muscle Mass in Hemodialysis Patients by Segmental Bioimped-
ance (SBIS) MRI and 40K Techniques. In IFBME Proceedings.
Clemens, M. (2001). Discrete electromagnetism with the finite integration technique. Progress In Elec-
tromagnetics Research 32, 6587.
Grimnes, S. and O.G. Martinsen (2000). Bioimpedance and Bioelectricity Basics. Academic Press.
Lavizzo-Mourey, R. (1987). Dehydration in the elderly: A short review. Journal of the National Medical
Association 79, 10331038.
Lorenzo, A.D. (1997). Predicting body cell mass with bioimpedance by using theoretical methods.
Journal of Applied Physiology 82, 15421558.
Matthie, J. (2005). Second generation mixture theorie equation for estimating intracellular water using
bioimpedance spectroscopy. Journal of Applied Physiology 99, 780781.
Medrano, G. (2009). Model-based correction of the influence of body position on continous segmental and
hand-to-foot bioimpedance measurements. Submitted to: Medical & Biological Engineering & Computing.
National Library of Medicine (2000). The visible human project. Website. http://www.nlm.nih.gov/
research/visible/visible_human
Nescolarde, L. (2007). Whole-body and thoracic bioimpedance measurement: Hypertension and hyper-hydration
in hemodialyisis patients. In Proceedings of the 29th Annual International Conference of the IEEE EMBS.
Rush, E.C. (2006). Validity of hand-to-foot measurement of bioimpedance: Standing compared with
lying position. Obesity, a research journal 14, 252257.
Salmi, J. (2003). Body composition assessment with segmental multifrequency bioimpedance method.
Journal of Sports Science & Medicine 2, 129.
Statistisches Bundesamt (2007). Entwicklung der Lebenserwartung Fernere Lebenserwartung im Alter
von 60 Jahren, 1901 bis 2050. Website. http://www.bpb.de/files/XH3MK2.pdf.
Thomas, B.J. (2003). A comparison of the whole-body and segmental methodologies of bioimpedance
analysis. Acta Diabetologica 40, 236237.
Wabel, P. (2009). Importance of whole-body bioimpedance spectroscopy for the management of fluid
balance. Blood Purification, 27, 7580.
Zhu, F. (2006). Segment-specific resistivity improves body fluid volume estimates from bioimpedance
spectroscopy in hemodialysis patients. Journal of Applied Physiology 100, 717724.
64
Qinghai Shi
Chair for Measurement and Sensor Technology, Chemnitz University of Technology, Chemnitz, Germany
Andreas Heinig
Lifetronics, Fraunhofer Institute for Photonic Microsystems, Dresden, Germany
Olfa Kanoun
Chair for Measurement and Sensor Technology, Chemnitz University of Technology, Chemnitz, Germany
1 INTRODUCTION
Evaluation of the hemodynamic parameters of a patient has always been a subject of interest
to clinicians. It has been difficult to capture hemodynamic parameters and invasive methods
have been used to measure them. These invasive techniques are expensive, time-consuming,
demand complicated equipment and trained staff and are not always possible to use because
of the condition of the patient, which maybe either too serve or else too good to run the
risks associated with invasive techniques (Sodolski and Kutarski 2007).Therefore they are
not suitable for long time or repeated measurements because of their invasive nature. Several
non-invasive techniques capable of monitoring cardiac activity were in past two decades
developed. They are ultrasound Doppler, magnetic resonance imaging, and impedance cardi-
ograph (ICG). However, all these techniques except ICG are not suitable for long-term con-
tinuous monitoring of cardiac activity. Impedance cardiograph is a simple, inexpensive, and
non-invasive method for hemodynamic parameters measurements. The method is based on
65
66
2 METHODS
The hardware module in this work is based on the tradition method where a constant current
I(t) (approximately 1 mA at 100 kHz) fed into current electrodes [Fig. 2] and the resulting
voltage U (t) is measured through inner voltage electrodes, thereby allowing the calculation
of the electrical bio-impedance Z (t):
U (t )
Z (t ) = (1)
I (t )
To realize this simple method, the following issues have to be kept in mind: the current
fed into the patient needs to be a well defined constant signal with fixed magnitude, time
and frequency. A direct digital synthesis (DDS) IC is used to generate the sine wave signal.
This sin wave voltage can be fed into the patient through the ECG Electrodes. The resulting
patient voltage is detected by an instrumentation amplifier and the current is measured by
the other instrumentation amplifier at the same time. Then the gain phase detector can be
used to detect the magnitude and phase of the bio-impedance of the selected segments. The
magnitude of bio-impedance consists of two important signals: the base thoracic impedance
and thoracic impedance changes. There are some noises in this signal: the impedance changes
because of breath and motion which should be eliminated. Then a differentiator is used
to derivate the thoracic impedance changes. All of the analog signals are sent to analog to
digital convert of the microprocessor MSP430f1611 and over there digital signal processing
for calculation of hemodynamic parameters can be programmed. MSP430f1611 is the micro-
controller configurations with two build-in 16-bit timers, the fast 12-bit analog to digital con-
verters, dual 12-bit digital to analog converters, universal serial synchronous/asynchronous
communication interfaces (USART), I2C, DMA and 48 I/O pins.
67
In this capital the electrode configurations by Sramek, Penney and Qu are measured and
compared with this ICG System. As illustrated in [Fig. 3] the base thoracic impedance Z0 is
28 (2.8V, G = 100). The thoracic impedance changes is minimal 0.5 and maximal 0.8
which is ca. 3% of the base thoracic impedance (0.6V to 0.96V, G = 1200). The four current
spot electrodes are symmetrically placed on both side of the neck and chest. Therefore that
generates homogeneous electrical field and the impedance of electrode can be neglected so
68
that these can improve the accuracy of measurement. The reproducibility of measurement
of this electrode configuration is best than the other plot electrode configurations. The base
thoracic impedance Z0 is 18 (1.8V, G = 100) [Fig. 4]. The thoracic impedance changes is
minimal 0.5 and maximal 0.8 (0.6 to 0.96V, G = 12000) and is ca. 4% of the base thoracic
impedance. The base thoracic impedance by this configuration is smaller than by Sramek. As
showed in Fig. 5 the base thoracic impedance Z0 is 18 as same as the electrode configura-
tion by Penney (1.8V, G = 100). The thoracic impedance changes is minimal 0.6 and maxi-
mal 0.9 (0.72 to 1.1V, G = 1200) and ca. 5% of the base thoracic impedance which is greater
than another electrode configuration. In this work the influence of breath for the thoracic
changes by these three spot electrode configurations is evaluated. As illustrated in Fig. 68
we can make a conclusion, that Qu spot electrode configuration has the best signal noise ratio
than another two configurations. Because from Fig. 8 we can see that the thoracic impedance
changes has the constant basis line.
69
Figure 7. Results of Penney electrode configuration by a) normal breath and b) depth breath.
Figure 8. Results of Penney electrode configuration by a) normal breath and b) depth breath.
70
4 CONCLUSIONS
The electrodes configurations and signal processing play an important role in signal repro-
ducibility. In this study the results by different electrodes configuration (Sramek, Penney and
Qu) are measured and compared with the same circuit. The results of electrodes configura-
tion by Sramek are similar as band electrodes configuration. But this electrodes configura-
tion is sensitive to breath and motion. The results of electrodes configuration by Qu have best
accuracy and the breath and motion have no effect on the results of this electrodes configura-
tion. In this work a basis line system is used to eliminate the effect of breath and motion on
the results. With this system the section of heart-action caused changes of impedance and
of breath caused changes of impedance can be separated. That means, the noise of breath
caused changes of impedance not only can be eliminated from the basic thorax impedance
and heart-action caused changes of impedance but this signal can be also measured in this
system. In the future, more advanced techniques such as adaptive filter, specific model and
digital signal processing for calculation of hemodynamic parameter can be implemented for
detection to increase the accuracy and reliability. The SD card can be used in the future to
storage the large data. A fewer components will be hoped to use and therefore as few as pos-
sible power can be used that a long time monitoring system can be achieved in the future.
71
Kubicek, W.G. and Patterson, R.D. (1970). Impedance cardiography as a non-invasive method of moni-
toring cardiac function and other parameters of the cardiovascular system. Ann. N. Y. Acad. Sci.,
724732.
Osypka, M. and D. Bernstein (1999). Electrophysiological principles and theory of stroke volume deter-
mination by thoracic electrical bio-impedance. AACN Clin Iss. Vol. 10, 385395.
Penney, B. (1986). Theory and cardiac applications of electric impedance measurements. Critical
Reviews in Biomedical Engineering Vol. 3, 227281.
Penney, B.C., N.A.P. and H.B. Wheeler (1985). Simplified electrode array for impedance cardiography.
Medical and Biological Engineering and Computing Vol. 23, 17.
Qu, M. and Zhang, Y.W. J.T.W. (1986). Motion artifact from spot and band electrodes during imped-
ance cardiography. IEEE Transactions on Biomedical Engineering BME 33, 10291036.
Sodolski, T. and A. Kutarski (2007). Impedance cardiography: A valuable method of evaluating
hemodynamic parameters. Journal Vol. 14. No. 2, 115126.
Sramek, B.B. (1981). Non-invasive technique for measurement of cardiac output by means of electrical
impedance. Tokyo. Proc. of the 5th ICEBI, 3942.
Strobeck and M. Silver (2004). Beyond the four quadrants: the critical and emerging role of impedance
cardiography in heart failure. Congest Heart Fail Vol. 10, 16.
72
S. Wegner
Sciospec Scientific Instruments, Schmlen, Germany
1 INTRODUCTION
2 CONCEPT OVERVIEW
In order to adapt one concept to basically any possible application of BIS it is necessary to
enable both high resolution, range and precision, as well as cost and time effective multichan-
nel realizations. For all steps in the instrumentation chain easily interchangeable modules have
been developed, e.g. either analog lock-in or direct sampling techniques are used depending
on the applications SNR, frontends can be easily trimmed for the exact bandwidth required,
adaptive gain settings and signal compression enable high dynamic ranges and for cost sensi-
tive high channel count applications a fully integrated frontend solution is available. The heart
of the electronic system is a FPGA based controlling platform. Depending on the specific
demands this main unit either functions plainly as communication junction between the sys-
tem and the PC based controlling software or incorporates process control, data processing
and analysis, error correction as well as data management and storage, making the system
self contained. Combinations of the frontend choices result in an application specific channel
73
3 POSSIBLE APPLICATIONS
A fair number of studies based on electrical impedance spectroscopy have been published
ranging from characterization of cell types, over bioreactor process control to cell parameter
extraction and studies on the reaction of cell properties to changes in environment (e.g. phar-
macological screening). With suitable means of implementing BIS in ready to use cost effec-
tive applications it is now possible to benefit from all this research and ultimately translate
the efforts already made into process control, monitoring or diagnostic solutions.
Examples of applications include electrochemical characterization, cell toxicity, blood char-
acterization (e.g. hematocrit, blood oxygen and glucose levels or cell differentiation), patch
clamp based single cell analysis or general process monitoring in bioreactors (cell growth and
adherence) as well as environmental process and parameter monitoring. For most of these
applications laboratory based setups and corresponding results have already been described
in according publications. As for the interface between cell (single cell, cell-suspension, mixed
volumes, tissue etc.) and the measurement electronics many possibilities are available, includ-
ing multi electrode arrays, multi well plates, patch clamp (planar or pipette) configurations,
needle electrodes, micro fluidic channels with electrodes etc. The diversity of electrode con-
figurations ensures a suitable fit for any BIS application that comes to mind.
The trend towards system biologythe systematic approach to understanding and model-
ling biologic systemsdemands for efficient ways to gain insight into these systems. Using
methods with sparse information output might facilitate the mathematical effort for modelling
the system, but in order to create more accurate models and better understand the system,
more data has to be collected in a way, that can be used for the modelling task at hand. The
combined measurement of high resolution, precise impedance spectra with other electrophysi-
ological or optical techniques opens up entirely new possibilities of modelling biological cells.
In addition to the above mentioned combination of different electrochemical analyses, the
application specific implementation of BIS offers the possibility to extend the impedance
spectrum analysis in several ways. Fast acquisition times enable good time resolution for
process control (spectrogram analysis) and the variation of input signals allows for nonlinear
spectroscopy and thus nonlinear modelling.
One major disadvantage of many analytic applications is the fact that the cells have to be
disposed after the test and cannot be used for further analysis. This also means that in-process
measurements are very limited. The main reason for this is that up to now almost exclusively
high throughput applications have been realized, making multi well or multi electrode array
assemblies necessary. With cost efficient ways to create application specific solutions it is now
possible to create small volume applications, targeting flow-through, in-system setups for
non-invasive, online measurements.
5 CONCLUSION
Bio impedance spectroscopy has proven to be a powerful analytical tool in the laboratory for
decades and even though many studies based on BIS have been done, it still shows largely
74
REFERENCES
Ayliffe, H.E., A.B. Frazier, and R.D. Rabbitt (1999). Electric impedance spectroscopy using microchan-
nels with integrated metal electrodes. IEEE Journal of Microelectromechanical Systems 8, 5056.
Bao, J.-Z., C.C. Davis, and R. Schmuklert (1992). Frequency domain impedance measurements of
erythrocytes constant phase angle and impedance characteristics and a phase transition. Biophysical
Journal 61, 14271434.
Coverdale, R., H.M. Jennings, and E.J. Garboczi (1995). An improved model for simulating impedance
spectroscopy. Computational Materials Science 3, 465474.
Gawad, S., M. Wthrich, L. Schild, O. Dubochet, and P. Renaud (2001). On-chip impedance-spectroscopy
for flow-cytometry using a differential electrode sensor. In Transducers 01: Eurosensors XV, Digest of
Technical Papers Vol. 1 and 2.
Orazem, M.E. and B. Tribollet (2008). Electrochemical Impedance Spectroscopy. Number ISBN
978-0-470-04140-6. Wiley.
Pftzner, A., A. Caduff, M. Larbig, T. Schrepfer, and T. Forst (2004). Impact of posture and fixa-
tion technique on impedance spectroscopy used for continuous and noninvasive glucose monitoring.
Diabetes Technology & Therapeutics 6, 435441.
Schanne, O.F. and E. Ruiz P.-Ceretti (1978). Impedance Measurements in Biological Cells. Wiley Interscience.
75
ABSTRACT: Wideband excitation enables to perform the fast spectroscopy, that is, to
cover a wide frequency range during a short time interval. It extends the measurement system
to provide spectral information matched with the dynamics of the subject under study. Power
spectrum of the chirp pulse excitation covers nearly the same frequency range in despite of
different duration of pulses in time. The paper describes the results of the study of spectral
properties of chirp signals depending on their duration.
1 INTRODUCTION
The simplest wideband excitation is a short rectangular pulse (Pliquet, Gersing, and Pliquett
2000), either unipolar or bipolar one (10 and 20 s pulses in Figs. 1a and b). The problem
is that only about 65% of generated energy falls into the useful bandwidth BW = 44 kHz
(at 3dB or 0.707 RMS level corresponding to 6dB or 0.5 power level), whereby the RMS
spectral density reduces to zero at 100 kHz (Figs. 2a and b). The spectrum reduces 20 dB/decade
outside the BW, also the crest factor (ratio of the peak and RMS values) of the signal is high
and affects the matter under test. The 10 s chirp pulse in Fig. 3 a covers the excitation band-
width BW = 100 kHz and its RMS spectral density is almost flat within the BW (Fig. 3 b).
About 80% of generated energy lies in the useful BW.
The chirp pulse in Fig. 3a is comparable to the rectangular pulse in Fig. 1a. This chirp
pulse (Fig. 3a) can be described as
(1)
ch(t ) sin [ / T ) t /tt ]
( B /T
where 0 t Texc and duration Texc = T/2 of the chirp pulse (1) is equal to half-cycle of sine
wave. The chirp rate TBexc , Hz/s, corresponds to the excitation bandwidth BW = 100 kHz
(Fig. 3 b) covered by the chirp pulse spectrum during one half-cycle T/2 = 10 s of sine func-
tion (1). A bipolar chirp pulse with duration of one full wave cycle T = 20 s in Fig. 4a cor-
responds to the rectangular excitation in Fig. 1b. Its RMS spectrum is shown in Fig. 4b.
79
Figure 2. Density of RMS spectra of the unipolar (a) and the bipolar (b) rectangular pulses shown
in Fig. 1a and b.
Figure 3. Half-cycle sine wave pulse (a) and its RMS spectral density (b).
80
Figure 6. RMS spectral density of the chirp pulse consisting of 10 cycles, Texc = 200s.
Texc = 10T = 200s, the RMS spectrum density of this signal is depicted in Fig. 6. In Fig. 7 is
given the RMS spectrum of the chirp with duration Texc = 105 T = 2s.
Excitation energy depends proportionally on duration of the excitation pulse Texc. There-
fore, it is reasonable to use longer excitation pulses for obtaining a better signal-to-noise
ratio. Besides other requirements, the main limiting factor is a speed of impedance variations
(dynamics). Thanks to specific properties of the chirp function, it possible to match the needs
for bandwidth, time, signal-to-noise ratio and dynamic requirements when implementing the
81
impedance spectroscopy in lab-on-chip type analyzers (Min, Pliquett, Nacke, Barthel, and
Annus 2008; Sun, Gawad, Bernabini, Green, and H. 2007) and in medical devices (Nahvi and
Hoyle 2008).
ACKNOWLEDGEMENTS
The authors thank their colleagues from the EU FP6 Marie Curie ToK project no. 29857
InFluEMP workgroup at the Institut fr Bioprozessund Analysenmesstechnik in Heilbad
Heiligenstadt, Germany, for the design of microfluidic system and making it available for
experiments. Special thanks belong to Dr. Uwe Pliquett for his guiding discussions on the
principles of impedance spectroscopy.
REFERENCES
Min, M., U. Pliquett, T. Nacke, A. Barthel and P. Annus (2008). Broadband excitation for short-time
impedance spectroscopy. Physiological Measurement 26, 185192.
Nahvi, M. and B.S. Hoyle (2008). Wideband electrical impedancetomography. Measurement Science and
Technology 19, 19.
Pliquet, U., F. Gersing and F. Pliquett (2000). Evaluation of Fast Time-domain Based Impedance Meas-
urements on Biological TissueBeurteilung schneller Impedanzmessungen im Zeitbereich an biolo-
gischen Geweben. Biomedizinische Technik/Biomedical Engineering 45, 613.
Sun, T., S. Gawad, C. Bernabini, N.G. Green and M.H. (2007). Broadband single cell impedance
spectroscopy using maximum length sequences: theoretical analysis and practical considerations.
Measurement Science and Technology 18, 28592868.
82
1 INTRODUCTION
2 MEASUREMENT CONCEPT
The system should be capable of operating across a wide frequency range and flexible with
respect to the interface and the software to cover as much practical applications as possible. This
leads to the idea of the modular concept with basic units and special modules (see Fig. 1).
While the basic units are necessary for all applications, the special modules are interchange-
able and they can be adapted to the material under test (MUT). The general purpose unit
(GPU) controls data acquisition; it is used for any advanced data post-processing required
for a particular application. This unit can be used to control a number of additional sensors
and capture data in several user programmable channels supporting the high versatility of the
entire system. In typical applications, it is convenient to use a Laptop-PC as GPU. The data
processing unit (DPU) is the heart of the whole measurement system. The DPU is able to
generate stimulus signal for low-frequency AU and to capture the data in two parallel meas-
urement channels. Moreover, it can be used for critical data analysis (for some application
this makes GPU redundant which simplifies the system and makes it cheaper).
83
The application unit (AU) is available as: a) a low frequency application unit (LF-AU) up
to 10 MHz and b) an ultra wideband application unit (UWB-AU) up to 5 GHz (see Fig. 2 and
Fig. 3). In LF-AU electrical current and voltage drop are measured by means of electrodes
in order to estimate complex impedance. Differential signals are used to minimize the influ-
ence of transmission lines. The measurement concept allows a stimulus signal of arbitrary
shape for this unit and thus, a multisine signal with minimum crest factor is usually chosen.
In short, the choice of such stimulus allows us to take advantage from the same source as
laboratory impedance analyzers, since similar approach is used in those devices.
The low crest factor of the sine waves promotes the handling of signals rich in energy
resulting in high Signal to Noise and Distortion (SINAD)-values. Moreover, signal sources
provide for stable operational conditions so that effective methods can be applied to remove
systematic errors.
In UWB-AU, however, completely different approach is used in order to meet the system
requirements. A digital shift-register of order m with appropriate feedbacks provides Maxi-
mal Length Sequence (an M-Sequence) having a bandwidth up to several GHz (depending
on clock-rate fc). M-sequence is a periodical signal with all possible combinations of binary
numbers of order m (excluding all zeros) inside one period. This signal stimulates the MUT
(by means of UWB antennas, coaxial probe, etc). In contrast to the impulse technique (and
similar to frequency sweep approach), M-Sequence distributes its energy over the complete
measurement period. That is, for maximal appropriate value of signal amplitude determined
by electronic components resulting power and Signal to Noise Ratio (SNR) is higher than
that in the impulse technique and it is comparable to those of frequency-sweep devices (see
(Sachs, Peyerl, and Zetik 2003) for thorough comparison of M-sequence devices with other
methods). Signals with low amplitudes are easy to handle from hardware point of view and
they allow a monolithic circuit integration resulting in an improved RF behavior (realization
of critical system components on a circuit is shown in Fig. 3). The envelope of the power
84
spectrum of the M-Sequence has sinc-squared shape with zeros at a multiple of the clock
rate fc. It can be seen in Fig. 4 that roughly 80% of the signal energy is concentrated to the
frequency band below fc/2. Therefore, cutting the frequency band of interest at fc/2 does not
lead to dramatic loss in performance. Respecting the Nyquist theorem leads again to an
(equivalent) sampling rate fs which is equal to the clock rate fs = fc i.e. one sample per chip of
the M-Sequence. The period of an M-Sequence always includes N = 2n 1 chips in which n is
the length of the shift register. Since there is no number N of such kind that could be divided
by 2, a binary divider of arbitrary length can be used for generating sampling events.
The steep flanks of the divider as well as the accuracy of the temporal sample spacing con-
tribute positively to the stability of entire measurement system. A sampling event is gener-
ated only after the binary divider has run completely through all its internal states. Moreover,
85
this sampling approach creates the opportunity to adapt the speed of data gathering to the
actual needs of the particular measurement task. In highly dynamic applications, short binary
dividers are used so that more than one data sample per period is captured. In the end, one
limiting factor concerning the measurement speed is the digital data processing which has to
follow the data stream. In more stationary measurement scenarios, longer binary dividers are
used resulting in reduced requirements concerning the processing speed.
1+ R f )
Z f ) = Z0 (1)
1 R f )
1
( f ) = (2)
C0 j f Z ( f )
In (1) and (2) Z0 is the impedance of coaxial probe (Z0 = 50 ) and C0 is its capacitance
(C0 = 2 1014 F).
Real parts of the dielectric constant versus frequency for a number of liquids calculated
from measured data as described are shown in Fig. 6. Evidently, results of IMPSPEC do not
86
Figure 6. Dielectrical permittivitz of liquids, real part, measured with vector network analzyer (red)
and UWB AU (black).
deviate from the results measured with the laboratory network analyzer more then a few per-
cent and both devices give results, similar to those described in literature (Kaatze 2007).
87
The simplest way is to use for this purpose empirical dependency between measured signal
amplitudes and actual nanotube content (Fig. 7).
4 CONCLUSION
The modular concept of broadband impedance spectroscopy analyzers is presented. Con-
figuration of hardware and software can be relatively easily adjusted to meet a broad range
of particular requirements for the specific measurement task. Therefore, the system can be
applied in process monitoring to solve a number of problems. The system has high measure-
ment speed, good long-term stability and is relatively easy to use. This opens up new applica-
tions (such as in moisture and liquid sensing, non-destructive testing of polymers, etc.) which
are difficult to implement with state of the art measurement techniques.
REFERENCES
Kaatze, U. (2007). Reference liquids for the calibration of dielectric sensors and measurement instru-
ments. Measurement Science and Technology 18, 967976.
Nacke, T., A. Barthel, J. Friedrich, M. Helbig, J. Sachs, P. Peyerl, and U. Pliquett (2007). A new hard
and software concept for impedance spectroscopy analyzers for broadband process measurements.
In ICEBI 2007, IFMBE Proceedings 17.
Sachs, j., P. Peyerl, and R. Zetik (2003). Stimulation of UWB-Sensors: Pulse or Maximum Sequence?
In International Workshop on UWB System.
88
1 INTRODUCTION
Usually, the reinforcement of reinforced concrete structures is well protected against corro-
sion by a dense and pore free oxide layer, the so called passive layer that forms on the steel
surface in the highly alkaline environment, which is provided by the surrounding concrete.
This layer has only a thickness of a few nanometres but it almost completely suppresses the
anodic dissolution of iron (1).
When chlorides from external sources like de-icing salts or seawater etc. take ingress
into the concrete, the passive layer may become instable if a critical chloride concen-
tration is exceeded at the steel surface. The result of this process is local depassivation
followed by pitting corrosion. The depassivated reinforcement acts as anode, where the
dissolution of iron (1) takes place, while the remaining passive steel surfaces form cath-
odes (2).
91
In following reaction-steps corrosion products arise at the anode (4) and (5).
Fe2 + + 2Cl
C + H2O Fe
2 Cl F (OH)Cl + HCl (3)
4 Fe(OH)Cl + 2 H2O + O2 FeO OH) + 4 HCl
F O( OH
Fe2 + + 2OH
2 OH Fe(OH)2 (4)
4Fee(OH)2 O2 FeO(OH) + 2 H2O
The resulting potential difference between the anode and cathode provides the driving
force for macro cell corrosion. The macro cell current is the dominating part of the total cor-
rosion current (Raupach 1992) which can be expressed by (5).
Ec Ea
I corr = + I self (5)
R
In order to control corrosion in technical systems, as for example in the case of chlo-
ride induced pitting corrosion of the reinforcement in reinforced concrete structures and
to reduce remaining corrosion rates to negligible values, since more than 30 years cathodic
protection is a well established and approved rehabilitation measure. The reinforcement of
the building protected by CP is subjected to a cathodic current that is introduced by an
external retrofitted anode. The previously existing potential differences between anodic
and cathodic surface areas are reduced by the cathodic polarisation which is caused by the
cathodic current flow from external anode to the reinforcing steel. Considering the character-
istic polarization curves for active and passive iron in aqueous solutions, schematically shown
in Fig. 1, it becomes obvious that a comparable small potential shift to cathodic direction
can reduce corrosion current densities of a corroding system to values, which are in the same
order of magnitude as corrosion current densities of passive systems, which is, according to
(Boukamp 1995) below 0.1 A/cm2.
In addition to the cathodic polarisation the so called secondary protection mechanisms
occur, such as the increase of pH at the steel/concrete interface due to forcing the cathodic
reaction (2) as well as the migration of chloride ions, caused by the necessarily applied elec-
tric field. A derived and modified Ficks law can be used to describe the flux of ions, when
assuming the flux of ions to be the sum of diffusion and migration and in the case of one-
dimensional penetration (6).
Figure 1. Schematic representation of polarisation curves for active and passive iron in aqueous
solution.
92
Table 1. Survey of the test solutions used and size of the test series.
126000 3 12.6 0
126050 8 12.6 50 1.26
126100 8 12.6 100 2.51
126150 8 12.6 150 3.77
1261250 8 12.6 1250 31.4
93
2.2 Results
Figure 2 shows exemplarily the records of the electrode potential development of the test
series 1261250 (samples 15 at gradually decreasing chloride concentrations and samples
68 at constant chloride concentration). The first vertical red line on the left side marks the
beginning of the experiment; each remaining line represents one dilution step. Initially, the
samples show active behavior at a potential of about 500 mV. Considering to the entire
potential record a transition phase between active and passive behaviour can be observed
between the fourth and fifth dilution step. These thresholds, 500 mV for active behavior and
500 mV for passive behavior and the development of electrode potentials were also found in
the remaining test series.
Figure 3 shows exemplarily the graphical evaluation of Fig. 2, the development of the
mean polarisation potentials over time together with the chloride concentration of the test
solution within the shown time intervals for test series 1261250.
Figure 2. Development of electrode potential over the time against chloride concentration of the test
series 1261250.
Figure 3. Development of electrode potential (mean value) over the time against chloride concentra-
tion of the test series 1261250.
94
Figure 4. Used equivalent fitting models for the test series 1260001261250.
95
Figure 5. Bode plots of the three samples from the test series 126000.
96
Figure 8. Bode plots, exemplarily samples of a passive (series 126000), re-passivated and active
reinforcement steel samples from series 126100.
97
Figure 10. Bode plots, exemplarily samples of a passive (series 12600), re-passivated and active
reinforcement steel samples from series 1261250.
98
Time series
including sample Rs,mean Y0,mean mean ct, mean EOC,mean
number () (S sa/cm2) amean ( cm s )
2
(kcm2) (mV)
The anodic polarisation tests have shown that the severity of pitting corrosion on reinforcing
steel can be lowered when reducing the chloride concentration of the test solution. For all
tested conditions the c(Cl)crit for E/t = max is in the same order of magnitude and in the
range of 7 and 25 mmol/1.
On the basis of the impedance spectra and the equivalent circuit modelling it could be
figured out that although no repassivation in a commonly used context could be reached by
decreasing the chloride concentration, a sustainable reduction of the remaining corrosion
rate of the system could be obtained.
Future research work will have to show, if the results found by anodic polarisation tests are
also valid for cathodic olarisation tests. The impact of increasing the pH at the steel/medium
interface is subject of another test series in the framework of the presented research project.
The cathodic partial process with respect to the ratio of reduction of oxides and oxygen has to
be investigated in order to enable the implementation of such data in the mathematical model.
The transferability to practical systems and to steel in concrete respectively will have to be
demonstrated by adequate test series.
ACKNOWLEDGEMENTS
This work is a first part of the joint research project Numerical Simulation of Cathodic
Protection of the Rear Reinforcement of Reinforced Concrete which is funded by the DFG
(German Research Foundation). The Authors are grateful that the DFG is supporting this
research work.
REFERENCES
Andrade, C., M.A. Sanjun, A. Recuero, and O. Rio (1994). Calculation of chloride diffusivity in con-
crete from migration experiments, in non steady-state conditions. Cement and Concrete Research 24,
12141228.
Bertolini, L., F. Bolzoni, A. Cigada, T. Pastore, and P. Pedeferri (1993). Cathodic protection of new and
old reinforced concrete structures. Corrosion Science 35, 16331639.
Bertolini, L., p. Pedeferri, E. Redaelli, and T. Pastore (2003). Repassivation of steel in carbonated con-
crete induced by cathodic protection. Materials and Corrosion/Werkstoffe und Korrosion 54, 163175.
Bhni, H. (1974). Die Bestimmung von Lochfrapotentialen unter besonderer Bercksichtigung der gal-
vanokinetischen Memethodik. Materials and Corrosion/Werkstoffe und Korrosion 25, 97104.
99
100
Author index
101