Qualification of Equipment Core Document PDF
Qualification of Equipment Core Document PDF
PA/PH/OMCL (08) 73 2R
QUALIFICATION OF EQUIPMENT
CORE DOCUMENT
QUALIFICATION OF EQUIPMENT
CORE DOCUMENT
Introduction
The standard ISO/IEC 17025 requires an appropriate choice and qualification of equipment to
be used for testing purposes. Particularly, checks and calibrations before and during use and,
if needed, intermediate tests (see ISO/IEC 17025 chapter 5.5.10) are necessary.
In order to guarantee a harmonized interpretation and application within the OMCL Network,
the guideline Qualification of Equipment has been elaborated.
From experience, the terms DQ, IQ, OQ and PQ (not explicitly mentioned by ISO/IEC 17025)
have been used in a non-harmonized way amongst the different OMCLs. Therefore their
mention has been avoided in this document. This does not exclude their use in OMCLs
quality systems where already approved and in application, or a reference to literature using
this nomenclature.
In order to simplify the management of the guideline, the present document contains only the
general introduction and the first two levels of qualification, which are common to all type of
equipment. The third and fourth levels of qualification can be found in separate instrument-
related annexes. When considered appropriate, additional requirements and/or examples
related to Level I and/or Level II have also been included in the annexes, which are to be used
in combination with the general recommendations given in the core document.
The list of annexes, included in this document, will be updated as soon as new annexes are
issued.
This document should be considered as a guide to OMCLs for planning, performing and
documenting the equipment qualification process. It should not be taken as an exhaustive list
of compulsory tests. It is left to the professional judgement and background experience of
each OMCL to decide on the most relevant tests and the most appropriate tolerance limits for
each of the parameters, in order to give evidence that the instrument is working properly and
is appropriate for its intended use.
For the more technique-related aspects of equipment qualification checks, practical examples
of possible approaches are also presented in the annexes.
The following four levels of Equipment Qualification should be considered by the OMCLs:
The selection and purchase of new instruments shall follow a conscious decision process,
based on the needs related to the intended use of the instrument.
An example for setting and documenting such specifications and decisions taken is given in
Table I.
When receiving an instrument, the OMCL should check that it is received in good conditions,
as ordered, and should monitor and document the installation process of the instrument in the
selected environment. This includes the start up checks done by the supplier, followed by a
full periodic check as described in Level III.
The release for use shall be documented and authorised by the person responsible for the
instrument.
An example for documenting the instrument installation and release for use and decisions
taken is given in Table II.
When instruments are installed or moved into a new environment a series of checks have to
be carried out to verify the key performance parameters of the instrument avoiding additional
contributory effects from the analytical method. Depending on the frequency of use and the
experienced stability of the instrument this shall be repeated periodically.
The same verifications (or a relevant part of them) shall be carried out following events like
significant repair or maintenance operations.
Examples of parameters to be checked on instruments and their typical acceptance limits can
be found in the Table III of the corresponding instrument-related annex.
The specifications from the manufacturer of the instrument should be taken into account when
setting the tolerance limits.
Some examples on how these checks may be performed on each type of instrument are also
provided in the corresponding Annexes.
During the day-to-day use of the instruments, checks are necessary to demonstrate continued
evidence of satisfactory performance by the instrument itself and compliance with the system
suitability criteria as defined in the applied analytical procedure for each product or group of
products tested at this occasion.
Examples of parameters to be checked on instruments and their typical acceptance limits can
be found in the Table IV of the corresponding instrument-related annex.
In the case of OMCLs performing routine testing (Type A: batch release of vaccines and
blood products), the use of control charts provides supplementary information on equipment
performance, which can also be used in this context.
The qualification levels dealt with in each annex are indicated in brackets.
TABLE I
Manufacturer:
Provider/Distributor:
Attribute Assessment
Benefits
(This list may be adapted if Specifications
(Instrument/supplier)
necessary) Pass Fails
Technique
TABLE I (cont.)
Comments / Decisions
Date / Signature:
TABLE II
Identification code:
Check of damages
Pass Fails (description of deficiencies)
1
if available
Comments
Date / Signature:
Disclaimer:
The present Core Document of the OMCL Guideline Qualification of Equipment, as well as
all its Annexes, have been drafted by ad-hoc working groups of technical experts, mainly
coming from Official Medicines Control Laboratories (OMCLs) and only occasionally from
other public institutions. These working groups do not include any representative from any
commercial organisation.
This Core Document and its Annexes may contain trade names of laboratory instruments,
materials and/or reagents. These are exclusively given as example in order to make these
guidelines easier to understand and implement, and were found to be suitable when the
guideline was being developed. These references do not imply in any way that the mentioned
instruments, materials or reagents or their suppliers are especially endorsed, recommended or
certified by the EDQM, the OMCL Network or the Council of Europe, in preference to others
of a similar nature which are not mentioned. It is therefore acceptable to use instruments,
materials and reagents from another source, provided that they fulfil the necessary criteria laid
down in these documents and appropriately satisfy the needs of the concerned laboratories in
the frame of their specific activities.