DNA User Manual en PDF
DNA User Manual en PDF
User Manual
Version 2.0
English
2
Digital Level Product ID
Congratulations on your purchase of the new The type and the serial-no. of your instrument are
Leica Geosystems digital level. written on the label located on the underside of the
This manual contains important instrument. Enter the type and serial number in your
safety directions as well as manual and always refer to this information when
instructions for setting up the you need to contact your agency or Leica Geosys-
product and operating it. Refer to "Safety directions" tems authorized service workshop.
for further information. Type:_____________ Serial-no.:______________
Please read these instructions carefully before put-
ting the instrument into operation.
Trademarks
Windows and Windows CE are a registered
trademark of Microsoft Corporation
CompactFlash and CF are trademarks of
SanDisk Corporation
All other trademarks are the property of their respec-
tive owners.
Symbols used
DANGER
Immediate hazards of use that lead to
personal injuries or even death.
WARNING
Hazards of use or inappropriate applica-
tion that lead to personal injuries or even
death.
CAUTION
Hazards of use or inappropriate applica-
tion that lead to minor personal injuries
but appreciatble material, financial or
enviromental damage.
3
4
Contents
Introduction ......................................... 8 Centering........................................................ 30
Principal of measurement .......................... 9 Measuring ..........................................31
Validity........................................................ 9 General notices ........................................31
Special features................................. 10 Height reading ..........................................31
Most important elements.......................... 11 Distance measurement.............................32
Measurement values................................ 14 Angle measurement .................................33
Applications.............................................. 15 Operating the instrument..................34
Line levelling .................................................. 15 Keyboard and display ...............................35
Area levelling.................................................. 17 Fixed keys ...................................................... 36
Computer software package Key combinations........................................... 36
Leica Geo Office (LGO) ........................... 18 Navigation keys.............................................. 37
PCMCIA or CF card ................................. 20 Entry keys ...................................................... 38
Equipment ................................................ 21 Display keys .............................................39
Unpack ........................................................... 21 Navigating the menus .................................... 41
Batteries ......................................................... 22 Illumination menu ........................................... 41
Batteries ......................................................... 23 User input .................................................42
Memory card .................................................. 24 Entering numeric values................................. 42
External power supply.................................... 25 Entering alphanumeric values........................ 43
Measurement preparations .............. 26 Inserting letters and numbers......................... 43
Deleting letters and numbers ......................... 44
Levelling up .................................................... 27
Focussing the telescope ................................ 29 Set of charaters.............................................. 44
Find point ................................................. 45 Start programs...................................72
Wildcard-search ............................................. 48 Set job ............................................................ 73
Technical hints for measurements .. 49 Set line ........................................................... 74
Special measuring situations ................... 49 Set tolerances ................................................ 76
Select method ................................................ 78
Important instrument settings......................... 50
Measurement modes (MODE) ....................... 51 Check list........................................................ 78
Start programs error messages ..................... 79
Measuring progress ....................................... 53
Repeating a measured sight .......................... 54 Measurement programs....................80
Managing point IDs ........................................ 55 Introduction...............................................80
Data and memory management............... 56 Line levelling.............................................81
Measure & Record............................. 57 Typical measurement display of
Starting display (1st backsight) .................... 59 Line points (B/ F)............................................ 82
Foresight display ............................................ 60 Last measurement backsight ......................... 83
Backsight display ........................................... 61 Last measurement foresight........................... 84
Switching to intermediate sight or set Intermediate sight and set out........................ 84
out sight.......................................................... 61 Station results ................................................ 85
Surveying intemediate points ......................... 62 Exceeding tolerances..................................... 86
Set out............................................................ 64 Line cut .....................................................87
Line Adjustment........................................89
Functions (FNC) ................................ 67
Data-Management ......................................... 92
Test measurement ......................................... 68
View measurement ........................................ 68
Check & Adjust .........................................93
"A x Bx" method ........................................... 95
Code............................................................... 69
"A x x B" method ............................................ 97
Point ID and increment................................... 70
Manual entry of measured values .................. 70 Measuring procedure ..................................... 99
5
6
Coding .............................................. 101 Memory information................................121
Entering a code ...................................... 102 Data export .............................................122
Quick Code ............................................ 103 Data import ............................................124
Menu settings .................................. 105 Data storage.....................................126
All settings.............................................. 107 Start programs ............................................. 126
System ......................................................... 107 Measurement program................................. 127
Measuring .................................................... 108 Measurement mode and corrective
Communication ............................................ 109 parameter..................................................... 128
Unit selection................................................ 110 Coding.......................................................... 129
Date and time............................................... 111 Fix points coordinates .................................. 129
System information ................................ 111 RS232 interface ........................................... 130
Check with collimator ................................... 112 Safety Directions .............................131
Data manager................................... 114 Intended use of instrument .....................131
Card functions ........................................ 115 Adverse uses ............................................... 132
View /edit data........................................ 116 Limits of use ...........................................132
Measurements ............................................. 116 Responsibilities.......................................133
Fix points...................................................... 118 Hazards of use ............................................. 134
Code list ....................................................... 119 Electromagnetic Compatibility (EMC).....137
Delete memory............................................. 120 FCC Statement, applicable in U.S. .............. 139
Product labelling........................................... 140
Care and storage .............................141
Transport...................................................... 141
Check and adjust............................. 145
Stand............................................................ 145
Circular level ................................................ 145
Reticle .......................................................... 146
Technical data ................................. 147
Corrections/ formulas ..................... 150
Accessories ..................................... 151
Sensor-error messages .................. 152
Index ................................................. 153
7
8
Introduction
With the acquisition of this Leica digital level you The unique concept of format files enables the out-
have chosen a product of excellence, with state of put of store data in almost any format. The format
the art ergonomy and outstanding measuring accu- files can be created individually and modified asy
racy. Both types of the instrument feature electronic desired. The logfile e.g., can be completed in the
reading of the measuring staff height. The bull's eye field and transferred to an internal memory card.
bubble only has to be roughly set for every sta-
tioning. The fine adjustment of the target beam is
done automatically by the high-precision compen-
sator. A key press triggers the electronic measure-
ment. Should it not be possoble for once, to make
an electronic measurement, then the height can be
optically read from a conventional, metered staff
and manually enterred with the keypad.
Leica digital levels come with a wide range of soft-
ware functions. Single height measurements are
easy to make and also just as easy is measuring all
the elements in a line levelling job. With the pro-
gram "Line Adjustment", measured heights can be
compared directly with the height of fixed points and
adjusted to them if desired. Staking out absolute
heights or height differences or point-to-point mea-
surements are all easily possible.
Principal of measurement
The bar code of the staff is stored in the instrument
as a reference signal. When measuring, the visible
section of the staff within the field of view is captu-
red by the line decoder as a measuring signal. Then
the measuring signal is compared to the reference
signal. This results in the height reading and hori-
zontal distance. The staff must be perpendicular
during measurement, as for optical levelling. With
artificial lighting of the staff, mesurements can be
made in the dark (The sensitivity of the sensor ran-
ges from the highest frequiencies of visible light
down to the frequency of infra red light).
Validity
This manual is valid for both instruments of the DNA
series. Sections only valid for the DNA03 are mar-
ked accordingly.
9
10
Special features
Large display, alphanumeric keyboard
Bi-directional horizontal drive
Camcorder batteries
Magnetically damped compensator
Onboard programs
Data storage to internal memory
Data backup to PCMCIA-card or to CF-card
with adapter
DNA03_01
1 On/ off button
Most important elements 2 Base plate
3 Foot screws
8 11 12 13 14 4 Horizontal circle
5 Lever to unlatch battery
6 Battery compartment
7 Button to unlatch card compartment cover
7 8 Card compartment cover
6 19 9 Display
5 10 Circular level
4 11 Hand grip with aiming sight
3
2 12 Ocular
1 9 10 13 Keyboard
14 Objective
15 Battery GEB111 (optional)
16 PCMCIA or CF-card with adapter (optional)
17 Battery GEB121 (optional)
18 Battery adapter GAD39; 6 single cells (optional)
19 Light duct for circular level
20 Plug stopper for crosshair adjustment knob
15 16 17 18 20 21 22 23 24 21 RS232 serial interface with external power
DNA03_02 supply
22 Measuring button
23 Focussing drive
24 Endless horizontal drive (bi-directional)
11
Vertical axis tilt Compensator 12
1
2 2
DNA03_05 DNA03_06
After centering the circular level the instrument is The compensator corrects for the vertical axis tilt on
almost horizonal. A minimal instrument tilt remains, the line-of-sight so that the instruments aim is exa-
this is the vertical axis tilt. ctly horizontal.
1 Vertical axis
2 Plumb line
3 Line-of-sight
Collimation error
DNA03_07
13
F Staff height foresight. for double 14
Measurement values observations: F1, F2
int Staff height intermediate sight / set out sight
DB Backsight-distance
C1
2 C2
DF Foresight-distance
1
DB S DF Dint Dint DInt Intermediate sight distance/ set out sight
distance
F Z int
B Ho Starting point height, e.g. as height above
dh
HCol dH sea level
H H
H Height of foresight point/ intermediate point
Ho
H=0.0000m dH Height difference between backsight and
foresight/ intermediate sight/ set out sight
DNA_Messgrssen dh Height difference between two
S Station measurements taken in sequence
1 Staff 1 (backsight staff) (intermediate sight/ set out sights/ foresight)
2 Staff 2 (foresight staff) HCol Instrument horizon (= height of line-of-sight)
C Staff C (intermediate point: intermediate Additional calculations with these dimensions can
sight while surveying, set out sight while be found in section, Corrections/ formulas.
setting out)
B Staff height backsight. for double
observations: B1, B2
Line levelling
Applications
Dependent on the required accuracy the same rules
DNA10 for levellling and regulations within the given coun-
Mainly for technical levelling tasks. tries apply as for optical levelling.
DNA03 Adhere to the following general rules:
Technical levelling and precision levelling. Maintain same target distance for back and
Staff selection foresight.
Measuring accuracy depends on the staff used in Measure a fore and back run and check it by
combination with the instrument. Use standard the error after the closing point.
staffs for low to medium range accuracy and invar Exclusively for precision levelling:
levelling staffs for the highest precision. Limit target distance, < 30m.
Range of application Minimum ground clearance of >0.5m required
Simple measurements using staff and distance to minimize refractionary influences of ground
readings proximity.
Line levelling Double observance (BFFB, aBFFB) to increase
Sureying and set out intermediate points. the reliability of the measurement and to reduce
Online operations with a computer. possible errors caused by the staff sinking.
Applying alternating observations procedures
(aBFFB = BFFB FBBF) to eliminate horizontal
tilt (residual error of the automatic
compensator).
Use an umbrella in strong sunlight.
15
Precision Mode: "Precise" is activated in the increase measuring accuracy. Activating 16
tolerance-settings for line levelling, the Precision Mode for line levelling with typical
instrument monitors the distance of the height accuracy is possible hut not necessary.
reading (Ziellinie) to both ends of staff, top and
bottom. The reduced number of staff code
elements may slightly lower the measuring
accuracy of measurements taken on the edge
of the staff. If the distance is less than 50cm a
warning is displayed. When this mode is
activated, the top and bottom limits of the staff
are automatically converted to a 3m Invar staff.
In order to use different staff sizes, the limit
values can be manuallly adjusted. The
precision mode also monitors measuring
distances critical to the staff. These distances
depend on the physical properties of the
instrument and of the staff. The measuring
accuracy of height measurements within these
distance ranges may also be slightly lower. A
warning is displayed if the measuring distance
is within the following ranges: 13.250m -
13.500m and 26.650m - 26.900m. The
Precision Mode is meant as a helpful tool to
Area levelling
Unlike line levelling, the individual target distances
in area levelling may be very different. Depending
on the required accuracy, a possible line-of-sight
error or the influence of the earths curvature may
have be taken into consideration.
17
Format Manager 18
Computer software package Creates and edits user defined output formats.
Leica Geo Office (LGO) Configuration Manager
Creates and edits user defined instrument
The Leica Geo Office (LGO) software package
settings.
includes a number of programs and tools that sup-
port working with the instrument. LGO-Tools is a DNA GSI Converter
part of the complete LGO package and can be Converts DNA03/ DNA10-data in the new GSI-
installed from the supplied CD. format into data in the old GSI-format of the
NA3003/ NA2002.
After the tools have been successfully installed, the
following program modules are available: For more information on the Leica Geo
Data Exchange Manager Office refer to the detailed Online-help.
Data exchange of fixed points, measurements, LEICA Geo Office (LGO) is available as a separate
code lists and data output formats between the software package and includes the basic module as
instrument (internal memory) and the computer. well as the LGO Tools package. The base module
Data exchange between PCMCIA-card (in the and the corresponding options offered supportvi-
instrument) and the computer. sualization, calculation, quality testing and recor-
Coordinate Editor ding of measurement data of various Leica
Import/ export/ create/ edit coordinate data. instruments.
Codelist Manager
Creates and edits code lists.
Software Upload
Uploads system software and measuring
programs.
The following options are available for the evalua-
tion of levelling data:
Display, editing and evaluation of single line
levelling
Creating and adjusting1D height grids.
Further information on the LGO will gladly be given
by your local Leica representative.
Data flow:
We recommend using the XML-format to transfer
measurement data from the DNA to the LGO and its We advise against transferring measured data in
modules. The required Format file (DnaXml.frt) is GSI format from the level to the LGO software pak-
located on the supplied CD and can be installed on kage. As the GSI format does not include all infor-
the instrument with the "Data Exchange Manager". mation, errors may occur during height calculations
The transfer of job data from the instrument and lead to wrong results.
to the computer is then also done through Transferring measurement data from the level test
the "Data Exchange Manager". from the DNA to LGO is also not recommrnded in
As LGO / LevelPak-Pro read in *.lev files by default, XML format.
we recommend giving the file to be transferred to
the computer a *.lev name. The following input
mask is displayed in the LGO Data Exchange
Manager when transferring data to the computer. As
an example, the format file DNAXml5 was selected.
19
20
PCMCIA or CF card
Measured data is stored in the internal memory of
the DNA03/ DNA10 and remains there. In addition,
data can be backed up to a PCMCIA or CF card
from the internal memory.
The system supports the PCMCIA standard for
ATA-Flash, SRAM or CF memory cards. Data
exchange with the computer is done via an internal
PCMCIA- drive or with the external OMNI-drive offe-
red by Leica Geosystems.
Files can also be exchanged between the memory
card in the instrument and the computer via the RS
232 serial interface using the Leica Survey Office
software.
5
1
6 7
8
9
DNA03_03
21
The battery charge is not displayed correctly when 22
Batteries
using single cells. Use the single cells with the bat-
tery adapter as emergency power supply. The
advantage of the single cells is in a lower rate of
discharge even over long periods.
Use the Leica Geosystems batteries, char-
gers and accessories or accessories
1 2 3 recommended by Leica Geosystems to ensure the
correct functionality of the instrument.
DNA_GEB
1 GEB121
2 GEB111
3 Single cells in the battery adapter GAD39
Your Leica Geosystems instrument is powered by
rechargeable plug-in batteries. For DNA instru-
ments, we recommend the basic battery (GEB111)
or the Pro battery (GEB121). Optionally six single
cells can be used with the GAD39 battery adapter.
Six single cell batteries (1.5 V each) supply 9 Volts.
The voltmeter on the instrument is designed for a
voltage of 6 Volts (GEB111/ GEB121).
Batteries Removing the batteries
DNA_BTTR_1 DNA_BTTR_2
First insert battery towards the objective (contact Place one of your hands below the open battery
in a). Then pull the lever towards the display, press compartment to catch the battery and with the other
battery upwards until it locks in place. hand, pull the lever towards the display. The battery
will fall out into your waiting hand.
23
Removing the card 24
Memory card
Card compartment cover
Open: Press latch.
Close: Press cover downwards until it locks in
place.
DNA03_PCMCIA_1
Ferrit_01
25
26
Measurement preparations
DNA03_Stativ2 DNA03_Stativ3
DNA03_Stativ1
DNA03_Stativ4 NA03_Stativ5
DNA03_Horiz_1
27
Centering the circular level 28
A B
NA03_Horiz_3
NDNA03_Monok_fok_l
29
30
Centering
Zentrier
31
32
Distance measurement
27
Example of an optical measurement:
27
26
26
25
25
DDNA_03_LatteF-Kreuz
5 Fine aim with horizontal drive. Perform points 1 to 6 according to height readings.
6 Check if bubble is centered . Reading
7 Read height H at the middle line of the reticle.
Distance line above: 2.670 m
Example shown: H = 2.586 m
Perform electronic measurement according to steps Distance line below: 2.502 m
1...6 and trigger measurement. Difference L: 0.168 m
Distance d: 16.8 m
Result:
Distance d = 100 x L
Angle measurement
DNA_Winkel-Mess
33
34
Operating the instrument
The shown displays are examples and local
software versions may deviate from the
base version.
DNA03_03
3 6 9 10
4 5
DNA03_04
35
36
Fixed keys Key combinations
Switching to point-to-point measurement. SET OUT
Set measurement mode.
Key with any function from the FNC-menu. Switch to setout.
Measurement programs, main menu. INV
Data manager.
Quit measurment program, function or edit Invert staff (0-mark at the top) for measurement.
mode step-by-step, retrieving the old The "T" symbol appears as long as INV is active.
parameters. Cancel /stop measurement. Revert with a renewed press on INV.
Measured values with inverted staffs are negative.
Switching to the 2nd key level (SET OUT,
INV, FNC, MENU, Lighting, PgUp, PgDn, FNC
<<Back, INS) and toggling numeric/
alphanumeric.
Functions that support the measurements.
Delete character/ field, Cancel/ stop
MENU
measurement.
Confirm entry, continue to next field.
Instrument settings, system-Information, line-of-
sight check with collimator (DNA03 only).
"Page Up"= scroll up to previous "page" if display The navigation keys take on different functions,
contains multiple pages.
depending on the mode in which they are used:
PgDn Focus control
Cursor control
"Page Down" = scroll down to next "page" if display Navigate through the selection
contains multiple pages. Select resp. confirming parameters
<<Back
37
38
Entry keys
... Enter numbers, letters and special
characters.
Enter decimal point and special characters.
Toggle positive/ negative signs; enter
special characters.
In alphanumeric mode:
Quick consecutive presses call up the next sign
(letter/ special character, number).
After a pause of approx. 0,5 seconds, the
presently displayed sign is accepted and the
cursor jumps to the next position.
39
Symbols 40
N Numeric digits activated.
Symbols with the following meaning are displayed
Alphanumeric characters activated.
on the right side of the screen:
1/3 Page number of the total pages or for
QC QuickCode. QuickCode is activated when a
search results the count of the counter codelist is loaded into the instrument and
from the total sum. Navigate with [PgUp] when the cursor is over a button.
resp.[PgDn]. Enter the corresponding two digit Quick-
Signal the selection from a list. Code to trigger the measurement and to
Navigate the list. save it together with the assigned code.
QuickCode is not activated when the cursor
Quit. is over an input field or when no codelist is
Quit. loaded into the instrument. In this case
entering a two digit code produces an error
The battery symbol shows the level of bat- message.
tery capacity remaining. (Example: 50%
full).
EC Earth curvature correction switched on.
Electronically measured or manually
entered staff heights are automatically
corrected for earth's curvature.
T Inverse staff switched on. It is only possible
to take measurement with the staff inverted.
FUNCTIONS ILLUMINATION
Sequence, layout and texts of the menu Display and circular level illuminations
may vary depending on local configuration. remain on permanently.
Switches on the circular level illumination.
43
44
Deleting letters and numbers Set of charaters
Deleting single characters: Numeric digits Alphanumeric characters
Deleting single characters in edit mode. Key Numeric Alpha 1 Alpha 2 Alpha 3 Alpha 4
Example:
0 / $ % 0
1 AB C 3 2 A B32 . # @ & .
8 D E F 8
9 G H I 9
DNA-Dde 4
45
Direct search: Search results: 46
It is possible to search for a specific point number
(e.g. "P13"). The search result contains all the
points with the relevant point number.
Example:
Entry: P13
Two fixed points and three measurements are dis-
played.
Scroll through the found selection.
DNA-Dde 5
Explanations
2/5
The displayed point P13 is the second of 5 in the
relevant job.
Type
The displayed point is a fixed point.
<PtSearch+>
Calling up the extended point search. Enter new
search criteria.
If no points are found a message is
displayed with that information.
The Search function always finds fixed point (in The entered point number is searched for in the
Fixed Points-jobs) before measured point (in Meas. memeory after confirming it with ENTER. The
Jobs) that fulfill the search criterion. If several points search result is displayed in the dislog "Point
are found that fulfill the search criterion, they are Search - Search Results" and includes all points
listed chronologically. The instrument lists the oldest with corresponding point numbers.
fixed point first Example:
Extended point search: Enter : P13 --> 2 fixed points and 2 measurements
The search for a specific point number (e.g. "P13") are displayed.
is possible with <PtSearch+>. The dialog permits <-- --> Scroll through the selection found
searching the point a random job or in all jobs.
47
Examples: 48
Wildcard-search
Wildcard-searches are characterized by the "*". The * finds all points of any given length.
"*" serves as a wild-card for any given sequence of A finds all points with the exact point ID "A".
characters. A* finds all points of any given length that start
Wildcards are used when the exact point ID is not with "A" (e.g.: A, A9, A15, ABCD).
known or when a series of points are to be found. *1 finds all points of any given length that have
a "1" as a second digit (e.g. A1, B12, A1C).
EXTENDED POINT SEARCH A*1 finds all points of any given length with an
"A" as the first character and a "1" as the
third digit (e.g.: AB1, AA100, AS15).
JOB : HEERBRUGG
PtID. : 13*
<QUIT>
DNA-Dde 6
51
reached by eliminatig outliers (measurement At "n min" = "n max" no measurements are 52
value with the greatest improvement). discarded by the wild shot test.
Entries: Repeat single
n min Minimum amount of measurements "Repeat single measurement". The instrument
(2..99) continuously makes single measurements
n max Maximum amount of measurements (maximum 99), until the observer stops the
(2.99) process as follows:
sDev/20m Standard deviation of the average The last valid single measurement is stored
value at 20m immediately.
For the measurement, this value is converted to
If any other key is pressed [not the DATA key]:
the specific distance measured and compared
The last valid single measurement is displayed.
to the current standard deviation of the average
(sDevM). Repeated measurements (of median)
Example: increase the integrity and quality of the
Measured distance= 60m measured data, especially during heat shimmer or
sDevM/20m= 0,0007 m ground vibrations due to road traffic.
0, 0007m 60
S = sDevM/60m = -----------------------------------
- =0.0021m
20
The maximum allowable standard deviation at
60m is 0,0021m.
Measuring progress Staff
Different screens are displayed while the measure- Current staff height according to mode (mean,
median or single measurement) after "n" measure-
remnts are in progress, depending on the measure-
ments.
ment mode selected.
sDev
Single measurement
Current standard deviation of a single measurement
Measurement time is very short. An hour glass is
after "n" measurements.
displayed to indicate that measurement is in pro-
gress. sDevM
Current standard deviation of the mean after "n"
Multiple measurements
measurements.
Spread
Measuring... Spread of the single measurements after "n" mea-
Mode : Median surements.
Count: 5
Spread =
Staff: 2.8005 m EC largest measured value - smallest measured value
sDev : 0.0003 m
sDevM: 0.0001 m After the last measurement the display
Spread: 0.0007 m remains static for approx. 3 seconds.
or Shortens the display time.
DNA-Dde 8
All the important information required to evaluate Manual cancellation of multiple measurements
multiple measurements is placed in a single display. The last valid measured value is accepted
and stored.
Count
Amount of measurements made (n).
53
If any other key is pressed [not the DATA key]: 54
Repeating a measured sight
the last valid single measurement is displayed: A sight just carried out can be repeated by pressing
the <<Back key. In line levelling, several sights, but
Measuring... the entire station at most (B and F, resp. B1, F1, F2,
B2), can be repeated. When a sight is repeated the
Mode : Median
calculations are updated. The original measure-
Count: 7
ment is deleted from the internal memory.
Staff: 2.8004 m EC
sDev : 0.0003 m Example: repeating the foresight measurement with
sDevM: 0.0001 m point ID = 2.
Spread: 0.0009 m <<Back opens the display:
<CANCEL> <CONT> <OK>
DNA-Dde 9
<OK>
Accept measured value and continue.
<CANCEL>
Reject measured value and cancel measurement.
<CONT>
Continue measurement.
DNA-Dde 10
Managing point IDs
Point IDs are handled differently if they are for line
points (foresight), intermediate or setout sights.
Line points (foresight)
The instrument will suggest a running automati-
cally incremented number as a foresight point ID.
The starting point ID and the increment are defined
in the function [FNC]/ "PtID & Increment". Swit-
ching the instrument on, set PtID value to A1.
Manually entered foresight point IDs are taken as
individual numbers and are only valid for single
measurements. The next foresight number will
automatically be a running point ID.
Intermediate points (intermediate/ setout sight)
A special range of numbers are reserved for inter-
mediate and setout sights. When switching on the
instrument they begin at point ID 1001. A manually
entered point ID is always a running number and
is automatically incremented. Set increment in
[FNC].
55
56
Data and memory management Job: Measure and Job 1 Job 2 Job 3
Record
The data is stored in jobs, comparable to direc-
tories.They can be copied, edited or deleted indivi- Line 1
Line_01
dually. Line 2
Line_02
Within a job data is stored in two memory areas:
Line 3
1 Measurement memory:
Line_99
Measurements and codes Line n
2 Fixed point memory:
Fixed and setout points. The job "Measure & Record" and its line designa-
Internal memory is divided into 16 sectors of equal tions are fixed and preset in the system.
sizes that are individually assigned to store measu- In a job, only the most recent (last) line can be
rements or fixed points. selected as the current line. It can also be supple-
When starting a job, sectors are reserved for mea- mented.
surements or fixed points. When a memory sector is Line
full, another free sector is used. A maximum of 16
jobs can be stored to internal memory. Each sector
stores approximately 350 measurements or 700
fixed points (PtID, X, Y, Z).
The hierarchy of jobs and lines
The measured data of a measuring program selec-
ted in [PROG] is stored in a line within a job.
Measure & Record
After start up, either the basic measuring program The procedure of MEAS & REC corresponds mostly
"MEAS & REC" is called up or message is dis- to line levelling with the BF method . The first back-
played. The message is displayed if the instrument sight is the starting point of the line. The height of
had been shut down during a line levelling job. The the starting point is either searched for in the mer-
last measured line can be continued from by confir- mory or it can be entered manually. All basic
ming the message. functions of levelling can then be easily carried out.
Single point measurements, staff readings
and distances to different points:
If height difference are not required and only
Continue saving staff readings and distances are desired,
Level Line? then directly in the first dialog of Meas & Rec
Line: X any amount of single points can be measured.
57
The point numbers are not automatically If the data back up settings in the menu is set to 58
incremented in this dialog. If required, the "RS232" , then the data are transferred to the imter-
numbers must be incremented manually. face in GSI-format and not saved internally. Backing
BF line levelling : up to the RS232 interface triggers a warning to the
Press<REC> to save the backsight and to user.
switch to foresight. Now measure the foresight In the back and foresight displays a single
and save it with <REC>. This sequence point can be measured as often as desired
corresponds tot the BE line levelling procedure. (PtID is not incremented). The observer has to
Point-to-point and staking out of heights, switch to the next sight, like e.g. between back and
height differences and distances: foresights.
Before measuring a foresight, it is possible to
measure point-to-point or stakeout points. Before measuring the next sight make sure
you have moved to the next empty data
The measurements are saved in the internal field.
memory in current job. If not job was created, then a
"DEFAULT" job is automatically created.
In the Start dialog of "Meas & Rec" set whether
every measurement (every measurement triggered
with the red measuring button) should be saved or
only the last measurement made before pressing
<REC>.
H0
Starting display (1st backsight)
Height of starting point (standard value = 0.00000).
First enter all required values and then trigger the If the point is recorded in the list of fixed points of
measurement with the measurement key. the job "Measure & Record", its height is automati-
cally entered.
MEAS & REC BF After the measurement:
PtBS: A1 HCol, Staf, and Dist are displayed with their
Rem : ------ respective values. The measurement can be repea-
H0 : 251.6670 m ted as often as desired. PtID is not incremented in
HCol: 253.5223 m EC the display for the same sight.
Staf: 1.8553 m
<JOB/LINE>
Dist: 9.65 m
Call up the Start dialog to enter names of the job
<JOB/LINE> <REC>QC
and line.
DNA-Dde 11
59
<REC> 60
Foresight display
First enter all required values and then trigger the Save measurement and continue to backsight.
measurement. <LAST>
Data and measured values of the last backsight are
displayed.
MEAS & REC BF
PtFS: 1 <END>
Rem : ------ Quits the "Meas & Rec" program. As long as no
H FS: 251.0257 m new line is started, the current line can always be
dH : -0.6413 m EC continued.
Staf: 2.4966 m
Dist: 12.67 m
<END> <LAST> <REC>QC
DNA-Dde 12
Entries:
PtFS
Retain running point IDs or replace with individual
PtIDs.
Rem
Remarks to the measurements.
After the measurement:
H FS (Height Foresight), dH, Staf and Dist are
displayed with corresponding vales.
Backsight display H BS (height backsight), dH, Staf and Dist are
First enter all required values and then trigger the displayed with corresponding vales.
measurement. <REC>
Save measurement and continue to foresight.
<LAST>
MEAS & REC BF Data and measured values of the last foresight are
PtBS: 1 displayed.
Rem : ------
H BS: 251.0257 m <END>
HCol: 254.1417 m EC Quits the "Meas & Rec" program. As long as no
Staf: 3.1160 m new line is started, the current line can always be
Dist: 16.56 m continued.
<END> <LAST> <REC>QC
Switching to intermediate sight or set out
DNA-Dde 13 sight
Entries: Switching is only possible from the foresight to cal-
PtBS culate the intermediate and setout sights. Requires
Point number for the starting point. a valid backsight for the station.
The standard value is "1". Note that this is a significant difference to
line levelling-programs: Point-to-point and
Rem
stakeout points can only be called up after comple-
Remarks to the measurements.
ting all measurements of a station i.e after measu-
After the measurement: ring the foresight.
61
Entries: 62
Surveying intemediate points
There are two displays for the intermediate sight: Next:
a) Height difference as related to the backsight (Pt Enter the point number to be measured next. Point
numbers are sequential and are incremented with
to BS)
every measurement.
b) Height difference as related to the previous
After the measurement
intermediate point (point to point).
Pt2:
Point ID is incremented after each The point number of the current measured point
measurement.
Staff:
Starts measurement display for intermediate The staff reading of the current measured point
points.
dH:
First enter all required values and then trigger the The height difference from point-to-point to back-
measurement. sight point
Pt2H
INTERMEDIATE (BS to Pt) The height of the current measured intermediate
Next: 1001 point.
Pt2 : ------ <Pt to Pt>
Staf: ----.---- m EC Switches to display "point to point"
Dist: ---.-- m
dH : ----.---- m <QUIT>
Pt2H: ----.---- m Exit intermediate sight, return to foresight.
<QUIT> <Pt to Pt>QC
DNA-Dde 14
Point-to-point Pt2:
First enter all required values and then trigger the The point number of the current measured point
measurement with the trigger key.
dH2:
The height difference from the current measured
INTERMEDIATE (BS to Pt) intermediate point to the last measured point.
Next: 1003 Pt2H
RemN: ------ The height of the current measured intermediate
Last: A1 point.
Pt2 : 1002 EC
<Pt to BS>
dH2 : -1.0000 m
Returns to the "intermediate point to backsight
Pt2H: 110.0000 m
point" screen
<QUIT> <BS to Pt>QC
DNA-Dde 15 <QUIT>
Exit from intermediate sight display and return to
Entries:
foresight display.
Next
Enter the point number to be measured next. Point
numbers are sequential and are incremented with
every measurement.
RemN:
Remarks on measurements
After the measurement
Last:
The point number of the last measured point
63
Entries: 64
Set out
Normally, height values are staked out. These stake StkP:
out heights can be loaded as fixed points into the Enter the running point number. The height of the
corresponding job, in order for the heights to be cal- entered point is searched for in the current job as
led up for the stake out, simply by the point number. soon as the entry is confirmed with ENTER. Should
These set out values have to be entered manually. a suitable point number be found, the dialog "Find
Of the three possible set out parameters only one Point - Result" is displayed. From here it is also pos-
can be used. sible to search in other jobs or to search with the
wildcard "*" for random point numbers.
[SET OUT] Starts set out point display:
Rem
Remarks to the measurements.
STAKE
SO H
If any set out point heights are stored in the fixed
StkP: 1004
point memory they are displayed, else new heights
Rem : ------
have to be entered.
SO H: 414.0000 m
SOdH: ---.---- m SOdH
SO D: ---.-- m Set out height differences with respect to backsight.
<QUIT> <CONT> SO D
DNA-Dde 16 Set out distance.
<CONT>
Continue to set out display.
<QUIT>
Exit set out, return to foresight.
Set out display Set out according to distance:
Trigger measurement. The calculated values and
cut are displayed.
STAKE DISTANCE 1/2
Set out according to height or height difference: PtBS: A1
PtID: 1001
Rem : -----
STAKE HEIGHT 1/2
SO D: 25.00 m EC
PtBS: A1
Dist: 24.85 m
StkP: 1002
Out : 0.15 m
Rem : ------
<EXIT> <REC> <NEXT> QC
SO H: 414.0000 m EC
H : 412.3750 m DNA-Dde 18
Fill: 1.6250 m Dist
<QUIT> <REC> <NEXT> QC Measured distance:
DNA-Dde 17
in/ out
H/ dH Amount to displace:
Measured height/ measured heigth difference. out (+) = move staff further away
fill/ cut in (-) = move staff closer
Amount to displace: Page 2
fill (+) = raise staff Measured values (staff height and distance).
cut (-) = lower staff
Page 2
Measured values (staff height and distance).
65
Procedure within the set out display 66
Move staff and repeat measurement until the diffe-
rence (fill/ cut, in/ out) meets the cut. Then select
one of the three functions:
<REC>
Store measurements and results, with the possibility
of subsequent measurements.
<NEXT>
Set out next point.
<QUIT>
Exit from set out, return to foresight.
Functions (FNC)
[FNC] opens the main menu for functions that sup- Example:
port measurements: Place the function " View measurement " in the
[USER]-key if you want to frequently inspect the
FUNCTIONS results of measurements.
1 Test Measurement
2 VIEW MEASUREMENT
3 CODE
4 PtID & Incremen
5 Manual Input
<QUIT>
DNA-Dde 19
67
68
Test measurement View measurement
"Test measurement" provides a measurement dis- This function displays the result of the last measure-
play, in which as many measurements as desired ment made again.
can be made without storing the data. This mode is
Calling up the "View measurement" display
intended for test measurements or to optimize tar-
get distances. Test measurements are always sin- Example for the mean measuring mode:
gle measurements disregarding the currently set
measuring mode. VIEW MEASUREMENT
Calling up "Test measurement": Mode : Mean s
Staff : 1.68859 m
n : 5
TEST MEASUREMENT sDev : 0.00036 m EC
sDevM : 0.00016 m
Spread: 0.00075 m
Staff: ----.---- m EC <QUIT>
Dist.: ---.-- m DNA-Dde 21
<QUIT>
DNA-Dde 20
Code Example:
This function permits a code to be entered Code entered manually (no code list available):
Two entry modes are available:
1 Select a code from the code list. A code would CODE & ATTR ENTRY 1/2
have to be stored in the instrument. If no code
list is stored then the second mode is Code : ------
automatically suggested. Info1: ------
Info2: ------
2 Enter a code manually. Info3: ------
Info4: ------
<QUIT> <REC>
Call up the "Code" function. A check is made to
see if a code list is stored in the instrument. DNA-Dde 22
Entries on page 1:
Code and info 1-4
Entries on page 2:
Info 5-8
69
70
Point ID and increment Manual entry of measured values
Enter starting value for the running point ID and Electronic measurements to staff less than 1.8 m
enter the increment. away are not possible. These measurements must
be done optically and the data entered into the
Call up the function
instrument manually.
PtID & Increment:
The following applies:
<SET>
The PtID is set ready for the current or next fore-
sight.
Calling up the "Manual input" function:
MANUAL INPUT
Staff: 0.00000 m
Dist : 0.00 m
<QUIT> <CONT>
DNA-Dde 24
71
72
Start programs
The following start programs are available as
measuring programs in [PROG]: LINE LEVELLING
73
74
Set line
In the selected job, a new line name is automatically NEW LINE
created and displayed. Should a different line name
be desired, then it has to be changed before starting Name: LINE00003
with the measurement. If the job is empty, then a Meth: BF
PtID: P13
new line is switched to. HO : 0.0000 m
Example: Stf1: ------
Stf2: ------
Display of Actual line line levelling: <QUIT> <PtSearch+> <SET>
DNA-Dde 29
ACTUAL LINE Entries:
Name: LINE00001
Name
Meth: BF
One time line name (no same line names possible
PtID: A1
in same job).
HO : 426.00000 m
Stf1: INVAR1 Meth
Stf2: ------ Observation method: BF/ aBF/ BFFB/ aBFFB.
<QUIT> <NEW> <SET> PtID
DNA-Dde 28
Start PtID.
<SET>
Stf1/ Stf2
Accept current line.
Designation for staffs 1 and 2.(optional).
<NEW>
Branches off to enable the entry of a new line.
Display New line line levelling:
After entering the start PtID the job checks if it is <PtSearch+>
already stored as a fixed point, measured point or Extented point search, including other jobs.
as a previous start PtID (manual input/ standard If the point is not found in the memory, even with the
value). If it is stored, it is selected from the list. extended search, then the manual entry box opens
automatically:
FIND POINT - RESULT
( 1/3) NEW POINT
Job : HEERBRUGG
PtID: P13
Job : 123
H : 412.2259 m
PtID: P50045
Type: Fixpoint HO : 0:00000 m
<QUIT> <PtSearch+> <OK>
DNA-Dde 30 <QUIT> <SEARCH> <OK>
H DNA-Dde 31
Point height. Entries:
Type PtID
Type of point: fixed point/ measured point/ manually Start PtID.
input point/ standard value (0.000).
H0
Height of start PtID. (standard value: 0.0000).
75
tion) to both ends of staff, top and bottom. The redu- 76
Set tolerances
Set tolerances must be adhered to during line level- ced number of staff code elements may slightly
lower the measuring accuracy of measurements
ling, depending on the application. Tolerance
taken on the edge of the staff. If the distance is less
checks are activated and deactivated here. With
tolerance check activated a message is displayed than 50cm a warning is displayed. When this mode
is activated, the top and bottom limits of the staff are
as soon as the set tolerance is exceeded. This per-
automatically converted to a 3m Invar staff. In order
mits corrective measures to be taken immediately.
to use different staff sizes, the limit values can be
BF, aBF method: manuallly adjusted.
The precisioin mode also monitors measuring
SET TOLERANCES distances critical to the staff. These distances
Precise : Off depend on the physical properties of the instrument
TDistBal: Off and of the staff. The measuring accuracy of height
MaxDist : Off measurements within these distance ranges may
StafEnds: Off also be slightly lower. A warning is displayed if the
measuring distance is within the following ranges:
13.250m - 13.500m and 26.650m - 26.900m.
<QUIT> <VALUE> <SET> The Precision Mode is meant as a helpful tool to
increase measuring accuracy. Activating Precision
DNA-Dde 32
Mode for line levelling with typical accuracy is possi-
Activate or deactivate relevant tolerance check: ble nut not necessary.
Precise:
Precision Mode: "Precise" is activated in the tole-
DistBal
rance-settings for line levelling, the instrument
"Distance-balance" = Distance balance between
monitors the distance of the height reading (collima-
foresight and backsight distances.
MaxDist Change tolerances
Maximum target distance. To check or change tolerance values open the dis-
StafEnds: play as follows:
Enables lowest or highest approach to the ends of <VALUE>
the staff. Branches to the entry of tolerance values:
BFFB, aBFFB method:
ENTER TOLERANCES
SET TOLERANCES TDistBal: 3.00 m
Precise : Off MaxDist : 50.00 m
DistBal : Off StafHigh: 2.5000 m
MaxDist : Off StafLow : 0.5000 m
StafEnds: Off StatDiff: 0.0003 m
StatDif : Off B-B/F-F : 0.0002 m
B-B/F-F : Off <PREV> <DEFLT> <SET>
<QUIT> <VALUE> <SET> DNA-Dde 34
DNA-Dde 33
77
78
Select method Check list
Select level test procedure. With the measuring programs Meas & Rec and
Line Levelling, after starting a new line, the first
thing displayed before starting the measurement is
SELECT METHOD a check list with important settings. To change the
settings, call up the corresponding function.
Method: A x Bx
Stf1: ------ CHECK LIST
Stf2: ------ Meas-Mode: Single
n Meas : 6
<QUIT> <SET> sDevM/20m: -.----- m
DNA-Dde 35 USER-Key : TestMeas EC
Entries: Pt-Incr. : 1
Methode : BF
Method <OK>
"A x x B" or "A x Bx". DNA-Dde 36
Stf1/ Stf2 Meas-Mode, n Meas, sDevM/20m
Designation for staff 1. and 2. (optional). Change values in [MODE].
USER-Key
Change assigned keys in [MENU]/ Quick settings.
PtID, Incr
Change values in [FNC]/ PtID & increment (PtID =
PtID of foresight).
Start programs error messages
The error messages are self explanatory. The countermeasures can usually be derived.
79
80
Measurement programs
The contents displayed, especially the rows, may
differ in local software versions. The functions Introduction
however, are always the same. As opposed to the simple measuring program Meas
PROG, MENU and DATA & Rec, the display switched automatically to the
These functions can be called up from the basic next screen in the measuring programs Line-
measurement program "Meas & Rec" and also from Levelling and Level Test. This simplifies and acce-
other parts of the program. Therefore the data lerates measurement procedures.
saved in the instrument can be displayed with the Calling up the program-Menu. The program-
[DATA] key at almost any time. Menu is the highest level menu in the hierar-
Online-mode chy of the Level-operating interface. From here, all
measuring programs can be called up.
In all measuring programs the instrument receives
and processes commmands sent via a serial inter-
face from a computer. Transferring measured data
via the RS232 interface by triggering a measure-
ment is only possible in GSI-format and only from
the basic measurement program "Meas & Rec".
Line levelling
PROGRAMS
The line levelling measuring program supports the
1 MEAS & REC BF, aBF, BFFB and aBFFB method that are select-
2 LINE LEVELLING able in the "Set line" start program.
3 LINE ADJUSTMENT Meaning of the methods:
4 CHECK & ADJUST
DNA-Dde 37
BF BF BF
With the start of a measuring program, the start
screen is displayed with Job, Line and other set- aBF BF FB
(alternating BF)
tings of the corresponding measuring program
(refer to section Start Up programm). BFFB BFFB BFFB
81
PtID 82
Typical measurement display of
Line points (B/ F) PtID of next aiming (can only be edited in foresight).
Rem
BF BF Remarks to the next measurement (optional).
LINE LEV
ST.4 dH T
PtID: 3 Total height difference between currrent backsight
Rem : ------ and starting point.
dH T: 1.0179 m EC H
H : 427.1299 m Height of the current backsight point.
TBal: 4.20 m
TBal
<END> <CL> <LAST> QC
Total current distance compensation between all
DNA-Dde 38
back and foresights.
Heading
Displays the method (here BF) with even and
uneven station.
2. row
The arrow marks the station (here the even station)
and within the station the next measurement that
will be made (here B).
St.4
Displays the current station number, starting with 1.
With the help of the current station number Last measurement backsight
and the indicator arrow, it can be
determined if you are on an even or uneven <LAST>
Example for BF method:
numbered station. This information is helpful, if you
have to close line levelling with a even station Displaying measured values with the instruments
(levelling with two staffs). height of collimation:
<END>
Quits the Line Levelling program. The line can be View Last BACK
continued at any time as long as no new line is ope- PtBS : 3
ned, no data from other programs were saved in the Rem : ------
current job and the current job was not changed. HCol : 428.7973 m
<CUT> H : 427.1299 m
Complete the line at a known point. Display all infor- Staff: 1.6674 m
mation on the line. Dist : 16.80 m
Refer to section "Line Cut". TBal : 21.00 m OK
DNA-Dde 39
<LAST>
Last measurement with calculated values.
83
84
Last measurement foresight Intermediate sight and set out
<LAST> Calculations of the intermediate sight and set out
always reference the last backsight. With the BF
Example for BF method:
method switching between intermediate and set out
Measured values displayed with height difference sight is permitted in any sight (B/F). With most
and height of the foresight point:
methods, the station must be completely measured
before intermediate or stakeout points are possible.
View Last FORE The displays and the procedures are analogous to
PtFS : "Meas & Rec".
Rem : ------
Opens the display to record intermediate
dH : -0.7200 m
sights.
H : 427.1299 m
Staff: 2.6000 m [SETOUT] Opens the display for set outs.
Dist : 9.60 m After a station change, always complete the
TBal : 4.20 m OK measurement of the backsight or the entire
DNA-Dde 40 station, before measuring intermediate or stake out
sights. The results would otherwide relate to the
backsight of the previous station and would
therefore be wrong. The display switches
automatically to the next station when all
measurements of a station are completed. Even
though the display already shows the next station to
be measured, the intermediate and stakeout sights Station results
for the current station must be measured at this time 'When using the double sight procedure (BFFB,
before the first measurement of the next station is
aBFFB) at the end, after completing the 4th measu-
made.
rement each the station results are displayed.
Example with the BFFB method:
The 4th measurement to an uneven station is
completed.
85
<STATION> Page 2: 86
Switches to station results, page 1.
STATION RESULTS 2/2
STATION RESULTS 1/2
Stat.No : 3
Stat.No : 3 StatDist: 30.55 m
StatDif : -0.0200 m B1-B2 : -0.0200 m
StatDif: 0.0080 m F1-F2 : 0.0000 m
dH : -0.7900 m
H 282.5023 m <CONT>
<PREV> DNA-Dde 43
DNA-Dde 42
Stat.No
Stat.No Station number.
Station number (running numbers,
B1-B2
begins with 1).
Difference of the two backsight measurements.
StatDif
F1-F2
Station difference.
Difference of the two foresight measurements.
StatDif Exceeding tolerances
Accumulated station differences.
If the tolerence is exceeded during measurement
dH
with the activated tolerence check (see Section Set
Height difference (B-F).
tolerences), a message is displayed showing the
H current parameters.
Height of foresight point.
Example:
Distance balance tolerance was exceeded:
Line cut
With all methods of the program line-levelling, there
is the possibility of comparing the height of the last
DISTANCE BALANCE CHECK point of a line with a known fixed point and to calcu-
late the misclosure.
TBal : 6.75 m After completing the measurement of a station, the
Limit: 3.00 m <CUT> key is displayed. It starts the calculation of
Distance Bal. too big! the line misclosure.
87
88
LINE INFO QUICK CLOSURE
DNA-Dde 46
DNA-Dde 48
91
H: 92
Data-Management
new: the adjusted height The original measurements are kept in the job. With
ori: the original measured height every adjustment, a new data-block with all new
Resid: point-heights is recorded in the current job.
Residual or difference between H new and H ori. The calculated heights can be viewed using the on-
board data manager.
Using a suitable format mask, the adjustment re- to view the adjusted heights.
sults and the heights of all adjusted points can be
read out and stored in a file on the PC-card or trans-
ferred to the PC.
The sequence of the staff to be measured (A1, B1,
Check & Adjust B2, A2) must be adhered to in all four procedures:
With the Leica DNA03/10 the optical as well as an Station 1 A1, B1
electronic collimation error can occur. Station 2 B2, A2
Electronic staff readings are automatically corrected The shorter of two aimings must always be
by the collimation error stored in the instrument. done first (not valid for measurements from
Optical collimation errors have to be eliminated by the center).
adjusting the reticle.
Where possible the distance ranges are checked.
Procedure and general notices A wrong instrument position is reported at once with
The instrument offers two integrated field procedu- the correction factor.
res for electronic measurements.
After the fourth measurement the electronic line-of-
"A x Bx" and "A x x B" (A and B are the staffs posi- sight error is compared with the stored line-of-sight
tions, x is the instrument position). Both procedures error and displayed. The new line-of-sight error is
cover two procedures each. now set as corrective factor. The value for the opti-
A x Bx cal reading for the reticle adjustment is displayed at
Procedure "from the center" (classic) and by the same time.
Kukkamki.
AxxB
Procedure by Frstner and Nbauer.
The line-of-sight error is displayed in arc seconds.
With the following approximation arc seconds are
converted to circular measure.
1" = 0.1mm/ 20m resp. 2" = 0.001ft/ 100ft
93
Data storage: 94
Level test measurements are stored in a special line
of the selected job. The line name is predetermined
and is called "Check & Adjust" .
min.
A 1/2 1/2 B 2.5 m
1 2
a b
DNA03_Nivellierprobe_1
1 1st Position
2 2nd Position
A Staff A
B Staff B
Conditions on the distances:
1. Station The center has to be within an accuracy of +/-1 m
2. Station b 2.5 m
95
Kukkamki procedure: 96
Instrument positioned in the center of the staffs and on the outside of staff B at a distance b (b = a).
a = approx. 20 m.
Distance conditions as before.
.
A B
1/2 1/2
2
1
a b
DNA03_Nivellierprobe_2
1 1st Position
2 2nd Position
A Staff A
B Staff B
"A x x B" method
Features of this method: The distances have a 1:2 ratio in every position.
Frstner procedure:
Instrument positioned at each third between the staffs. Staff distance D approx. 45 m - 60 m.
Distance conditions:
1. Station: 0.2 x D < Dist_A1 < 0.4 x D
2. Station: 0.2 x D < Dist_B2 < 0.4 x D
D = Dist_A1 + Dist_B1
A 1/3 2/3 B A 2/3 B
1/3
Dist_A1 Dist_B1 Dist_B2
1 2
DNA03_Nivellierprobe_3/4
1 1st Position
2 2nd Position
A Staff A
B Staff B
97
Nbauer procedure: 98
Instrument position outside of the staffs. a = approx. 15 m - 20 m.
Distance condition: same as above.
a A a B a
2
1
DNA03_Nivellierprobe_5
1 1st Position
2 2nd Position
A Staff A
B Staff B
Measuring procedure A1, Dist
The program shows which station is to be placed Displays the measurement A1
where using self explanatory displays: B1, Dist
Procedure: Displays the measurement B1 (in this display not
taken yet)
Trigger measurement. Repeated
measurements can be made. <CONT>
<CONT> switches to next aiming. Continue to the next Set Up and Measurement B2
Example of a measurement display: and A2 proceed in the same manner.
Display of the final result:
DNA-Dde 49
Heading DNA-Dde 50
101
Coding without code list 102
Coding without code list is the same as the form of Entering a code
coding in GSI-format known from the old generation Codes are entered in FNC (see section Code).
of instruments. Should no code list exist then the prompt for a
The elements are: manual input appears automatically.
(see section Code).
Calling up CODE with an existing code lists:
Designation Input
Procedure:
Find
Enter a search criteria ("*" = all values ).
Code
Make selection from the list of codes found.
Desc
The description of the found code is displayed. Quick Code
<REC> Quick Code enables entering two 2 digit to trigger a
Store code as data block. measurement und to directly save the assigned
code. Quick-Code is available in most measure-
<ATTR>
ment displays and is usually shown in the lower
Attribute display; attribute values can be changed
right with QC. The cursor can be placed on the soft
as desired.
key QC. Press ENTER to determine if a codelist
Except with fixed assigned status in the code with Quick-Code numbers is loaded in the insru-
list manager: ment or if the codelist is empty. The corresponding
Status message is then displayed for a brief period.
"fix Value is write protected. The entry of Quick-Code digits in a measurement
"mandatory" Input or confirmation required. display activates QC, as soon as the cursor is pla-
"normal Value is editable as desired. ced on a soft key of the current display. After the
<MAN> two desired Quick-Code digits are entered, a mea-
Manual code entry, as if no codelist was loaded. surement is triggerred, saved and the assigned
code is saved. The code is saved before or after the
measurement, depending on the settings.
2 Measuring
Code setting
Decimal position
GSI-format
DNA-Dde 52
Earth curvature correction
MENU over view:
1 QUICK SETTINGS 3 Communication
Contrast Baudrate
Earth curvature correction Data bit
USER-key Parity
Decimal position
105
End mark Example: 106
Stop bit [MENU]/ Quick settings:
4 Units
Distance
Temperature
5 Date and time
3 SYSTEM INFO
Free jobs
USER-key
Battery
Instrument temperatue DNA-Dde 53
Display heater
Select appropriate setting from the list.
Collimation error
Software version <SET>
Accept and store new settings.
4 CHECK WITH COLLIMATOR
Measuring program (DNA03) <QUIT>
Quit display, possible setting changes are ignored.
Standby (Sleep)
All settings The instrument switches into energy saving
mode about 15 minutes after the last key press.
System At the next key press it reverts to normal mode.
Contains the system settings.
Contrast
Beep Sets contrasts in steps of 10%.
Acoustic signal trigger key:
USER-key
Off, loud, normal.
Has a function assigned in FNC.
Data output TestMeas
RS232 Performs a test measurement without storing.
Data output is through the (RS232) serial ViewMeas
interface. Only possible in the "Measure & Displays the last measurements with standard
Record" program. deviation and spread for repeated
Internal measurements (measurement progress
Data is stored to internal memory. display).
Auto-OFF Code
Inactive Code-input and code-selection.
Does not switch off. The instrument is PtID&Inc
permanently on. Enter running PtID and increment.
Active ManInput
The instrument switches off about 15 minutes Manual input of staff reading and distance.
after the last keypress.
107
Display heater 108
Measuring
When the instrument is switched on the display Important settings for measuring.
heater is set to Off.
Set code
On
Refers to Quick code.
Display heater starts when the inner temperature of
the instrument falls below -5C. Before
Stores code before making a measurement.
Off
Display heater is off. After
Stores code after making a measurement.
Collimation error
Displays the current collimation error. The value can Decimals
be edited to a new collimation value if e.g. it was Number of decimals displayed and for manual input
determined by other means than with the integrated (three steps).
level test.
GSI-format Communication
Data via interface and data export in GSI-format. Communication parameters of the RS232 interface
GSI-8 for data transfer between computer and instrument.
8-data characters output format
Leica standard settings
(83..00+12345678).
19200 baud, 8 data bits, no parity, CR/LF, 1 stop bit.
These data words can hold alpha as well as
numeric data. Baudrate
GSI-16 Data transfer rates: 1200, 2400, 4800, 9600, 19200
16-data characters output format (bits/second).
(*83..00+1234567890123456 ). Data bits
These data words can hold alpha as well as 7
numeric data. 7 data bits are set automatically for data transfer
Earth curvature correction when parity is set to "Even" or "Odd".
For electronically measured or manually entered 8
staff heights. 8 data bits are automatically set when for parity
Yes "No" is set.
With correction. Parity
No Even
Without correction. Even parity.
Odd
Odd parity.
No
No parity (when data bits = 8).
109
Endmark 110
Unit selection
CR/ LF
Distance
New line and line feed. Meter Meter
CR US foot US foot
New line.
INT foot International foot
Stop bit US foot/ inchUS-foot-inch (DNA03 only)
Set to 1.
Temperature
C Degree Centigrade
F Degree Fahrenheit
Date and time
System information
Display and setting of system date/ time. After date
and time have been entered, they are stored in the Display important and useful information.
system. Free jobs
Date Amount of free jobs (maximum 16).
Format:dd/mm/yy (day,month, year) USER-key
Time Current assignment of the USER-key.
Format:hh:mm:ss (hour, min., sec.) Battery
remaining battery capacity.
Instr.-Temp
Inner temperature of the instrument.
Display heater
Settings (on/ off).
Collimation error
Currently set collimation error.
111
Procedure: 112
Check with collimator
The measuring program, only available with the [MENU]/ Testing with the collimator:
DNA03, contains electronic collimation error deter-
mination using a collimator. Reticle adjustment
(optical collimation error) is not supported.
The special collimator contains a graticule with a
reticle with the scale of the digital level and is used
to make the adjustments. The collimator is not listed
in the scope of delivery of the measuring instru-
ment. For more information please contact Leica
Geosystem's technical services.
DNA-Dde 54
Entries:
Co-Staff
Collimator calibration value for the staff height.
Co-Dist
Collimator calibration value for the distance.
<CONT> Distance
Confirm and continue to measurement display. Distance.
Coll.err.old
Old collimation error.
Coll.err.new
New collimation error.
<SET>
Store new collimation error.
DNA-Dde 55
Procedure:
[MODE]
All the measuring modes that are also vailable with
the staff.
Aim at and focus on the bar code of the staff.
Trigger measurement with the measurement key or
with the external command (GET/M/WI32/WI330).
Staff
Staff height.
113
114
Data manager
Use the data manager to enter, edit, view or copy Points 1-3 refer to the internal memory. The corre-
data to other data mediums. It manages the internal sponding functions referencing the storage card can
memory, where all measured data is stored and the be found under <CARD>.
PCMCIA-card. VIEW/ EDIT DATA
Call up the data manager- selection display: Change, create, view and delete job data,
measurements, fixed points and code lists.
INITIALIZE JOB
DATA MANAGER Delete the entire memory, single jobs or single
storage areas (fix points, measurements).
1 VIEW / EDIT DATA
2 DELETE JOBS MEMORY INFORMATION
Information on jobs and of memory status.
3 MEMORY INFORMATION
4 DATA EXPORT DATA EXPORT
5 DATA IMPORT Export measured data/ fix points of a job from
<QUIT> <CARD> internal memory to a card or via the interface.
DNA-Dde 56 DATA IMPORT
Load fix points/ code lists from the card to the
internal memory.
1. Page:
Card functions Dir.
Insert or delete data on the Directory of the card.
PCMCIA-card or format the card: Size
File size in KB.
/ <CARD> calls up the card display: Free
Free memory space on the card in KB.
2. Page:
File
File name.
Dir.
Directory of the card.
Date
Date the file was created.
File, first line:
DNA-Dde 57 Displays the first 23 characters (as identification
Procedure: aid).
<FORMAT>
Search Format card.
Select file extenstion from the list as search criteria. <DELETE>
File Delete selected files from the card.
Select file from the list.
115
116
Measurements
View /edit data
Measured data in memory can be searched, dis-
/ Call up "View / Edit data": played and partially deleted. First, enter search cri-
teria of the job, line and point.
VIEW / EDIT DATA
1 MEASUREMENTS
2 IMPORT FIXPOINTS
3 JOBS
4 CODELIST
<QUIT>
DNA-Dde 58
DNA-Dde 59
Procedure:
Job
Select job from the list.
Line
Select line from the list.
PtID
Set criteria for point search.
<VIEW> <DEL>
Point search starts and result is displayed: Deletes record (necessary records for line calcula-
tion can not be deleted).
VIEW 1/3 4 <SEARCH>
Class:Meas Meas&Rec Returns to data search display ready for the entry of
PtID : A1 new search criteria.
Rem : ------ More information on the structure of data block is in
Back : 1.6500 m section Data storage.
Dist : 16.30 m
Type : measured
<QUIT> <DEL> <SEARCH>
DNA-Dde 60
117
Display: 118
Fix points
A fix point must contain either all coordinates (E, N, PtID
H) or only height (H). Lists the found points.
E/N/H
Point coordinates. With height points only enter the
height.
<DEL>
Delete single points.
<NEW>
Enter new point (PtID and height with or without
position coordinates).
DNA-Dde 61 Jobs
Procedure: Contents:
Job Displays jobs with relevant supplemantary
Select job from the list. information.
<DEL> Delete single jobs.
Find
Enter search criteria for the PtID ("*" = all). <NEW> Create new job.
Code list <NEW>
Find, display and supplement data in the code list. It Enter a new code:
is possible to scroll through the code list from the
main display. The code has a maximum of eight INPUT CODELIST 1/2
attributes. The information 1-8 on a code extend
over two pages. Code : ------
Rem : ------
VIEW CODELIST 1/2 QCNo : --
Find : * Info1: ------
Code : Info2: ------
Rem : ------ <QUIT> <PREV> <SAVE>
QCNo : -- DNA-Dde 63
Job : DEFAULT
Data: Measurements
Procedure:
Job
Select relevant job.
Data
Select data area (fix point/ measurements/ fix points
& measurements).
<DEL>
Deletes selected data area.
FixPoints
Memory information Amount of strored fix points in the job.
Provides information on contents (sectors for mea- Free Jobs
surements or fix points) of individual jobs and the Displays the amount of free jobs available if only
number of free jobs available: measurements were stored. Jobs containing mea-
surements and fix points display twice the amount
of space for free jobs than is actually available.
DNA-Dde 65
Job
Select job.
Lines
Amount of stored lines in the job.
MeasRecds
Amount of stored data blocks in the job (measure-
ments, codes, etc.).
121
122
Data export
The data export function backs up the data in inter-
nal memory to the storage card or sends it to the
serial interface. Data transfer via the serial interface
is done without protocol. As standard output format
the GSI-format is available in both options, GSI-8
and GSI-16. The description is on the CD-ROM as
a PDF-file in the directory GSI_Online. To collect
data in your own format the instrument can load four DNA-Dde 66
additional user formats. Create and load the user
Target
formats with the Leica Survey Office program.
Select memory target (card or serial interface).
Target directory on card:
Job
GSI-Formats: \GSI
Select job.
User-Formats: \DATA
Data
Select data type (measurements or fix points).
Format
Select output format (GSI-8, GSI-16 or user for-
mat).
<EXPORT>
Start data export.
123
Dir. 124
Data import Card directory.
Loading fix points or code list from the memory card Data
into the internal memory of the instrument. Existing Data file was created.
data is completely replaced by the new data. Fixed
<OK>
points and code lists must be in GSI-format.
Confirm and continue.
Data are searched and viewed in all
directories of the card.
For fix points a job has to be selected as the target.
DNA-Dde 67
Type
Select data type (fix points or code list).
Find DNA-Dde 68
Select data extension (GSI or all).
Job
File Select target job.
Select file
File
Displays the file for checking.
<IMPORT>
Start data import.
<NEWJOB>
Create new job.
125
126
Data storage
The internal memory stores data by jobs. However, Start programs
fix points and measurements are stored seperately.
Data blocks of different sizes are created, depen- Job
ding on the measuring program. The data is stored Job = Job name
right after the completion of an action. Oper = Operator name
Cmt1 = Comment 1
Example:
Cmt2 = Comment 2
The "Line" type data block is stored immediately Date = Date
after the line definition in the start program "Set line" Time = Time
is done. The data blocks of the measurements are
Line
displayed in the data manager in the sequence they
Name = Line name
were measured and stored.
Method = Method
The following list shows the data blocks in the
Staff1 = Designation of 1st staff
sequence of their creation with the most important
variables as displayed in the data manager. Staff2 = Designation of 2nd staff
Measurement program Measurement
PtID = Point number
Start PtID
Rem = Remarks
PtID =Point number
Back/... = Staff height (Back/B1/B2/Fore/F1/F2/
E =Coordinate (Eastings) Intm/ SOut/ Meas)
N =Coordinate (Northings)
Dist = Horizontal distance
H =Height
Type = Measurement/ manual input
Type =Type of point (fix point/ entered dH = Height difference refering to backsight
measurement/ standard).
dH_seq = Height difference of two successive
Rem = Remarks
measured points
Date = Date H = Height
Time = Time
Date = Date
Time = Time
n = Number of measurements made
sDev = Standard deviation (single
measurement)
sDevM = Standard deviation (mean)
Spread = Spread (max. - min.)
127
Target point 128
Measurement mode and corrective
PtID = Point number parameter
E = Coordinate (Eastings)
These data blocks are stored at the beginning of a
N = Coordinate (Northings)
new line, even when a setting is changed during a
H = Height
measurement.
Rem = Remarks
Date = Date Measurement mode
Time = Time Mode = Measurement mode
n = Preset amount of measurements (2-99)
Station
for mean and median modes
No = Running station number
n min = Minimum preset amount of
dH = Height difference
measuremenrs for "Mean s"
H = Foresight point height
mode.
DBal = Distance balance
n max = Maximum preset amount of
DTot = Total distance
measuremenrs for "Mean s"
DSta = Station distance
mode.
Date = Date
sDevM/20m=Preset standard deviation of "Mean s"
Time = Time
mode.
StatDif = Station difference
StatDif = Accumulated station difference Corrective parameter
B1 - B2 = Difference double sighting (backsight) Earth curv= Earth curvature correction
F1 - F2 = Difference double sighting (foresight) yes/ no
Collim.Err= Collimation error
Set out results
Diff (dH, H, D) = Set out result
Coding Fix points coordinates
Coding with code list PtID = Point number
Code = Code name E = Coordinate (Eastings)
Desc = Description N = Coordinate (Northings)
Attr1 *) = Attribute name 1 H = Height
...
Attr8 *) = Attribute name 8
*) Instead of attr1... attr8 a user defined attribute
name can be used.
Coding without code list
Code = Code name
Info1 = Information 1
...
Info8 = Information 8
129
130
RS232 interface
Measuring via serial interface RS232 is only possi-
ble in the basic measuring program "Meas & Rec"
and only in the GSI-format. Set data output to
RS232 ([MENU]/ All settings/ System) and select
GSI-8 or GSI-16 format ([MENU]/ All settings/
Measuring).
Safety Directions
The following directions should enable the person
responsible, and the person who actually uses the Intended use of instrument
instrument, to anticipate and avoid operational Permitted uses
hazards. The person responsible for the instrument Electronic and optical height and distance
must ensure that all users understand these direc- measurements to a staff.
tions and adhere to them. Angle measurement with the horizontal circle.
Recording of measurement data.
Calculations using measuring programs.
131
WARNING 132
Adverse uses
Use of the level without previous instruction Adverse use can lead to injury, malfunction,
Use outside of the intended limits and material damage.
Disabling safety systems. It is the task of the person responsible for the pro-
Removal of hazard notices. duct to inform the user about hazards and how to
Opening the product using tools (screwdriver, counteract them. The product is not to be operated
etc.), unless this is specifically permitted for until the user has been instructed on how to work
certain functions. with it.
Modification or conversion of the instrument
Use after misappropriation Limits of use
Use of products with obviously recognizable Environment
damages or defects
Suitable for use in an atmosphere appropriate for
Use with accessories from other manufacturers permanent human habitation. Not suitable for use in
without the prior express approval of Leica aggressive or explosive environments. Use in rain is
Geosystems permissible for limited periods (spash-water-proof).
Aiming directly into the sun
Inadequate safeguards at the surveying site
(e.g. when measuring on roads, etc.)
Person in charge of the product
Responsibilities The person in charge of the product has the follo-
Manufacturer of the product wing duties:
Leica Geosystems AG, CH-9435 Heerbrugg, her- To understand the safety instructions on the
einafter referred to as Leica Geosystems, is respon- product and the instructions in the user manual.
sible for supplying the product, including the user To be familiar with local regulations relating to
manual and original accessories, in a completely safety and accident prevention.
safe condition. To inform Leica Geosystems immediately if the
product and the application becomes unsafe.
133
CAUTION 134
Hazards of use
Watch out for erroneous measurement
WARNING
results if the product has been dropped or has been
The absence of instruction, or the inadequate misused, modified, stored for long periods or trans-
imparting of instruction, can lead to incorrect or ported.
adverse use, and can give rise to accidents with far-
Precautions:
reaching human, material, financial and environ-
mental consequences. Periodically carry out test measurements and per-
form the field adjustments indicated in the User
Precautions: Manual particularly after the instrument has been
All users must follow the safety directions given by subjected to abnormal use and before and after
the manufacturer and the directions of the person important measurements.
responsible for the instrument.
CAUTION
WARNING
Strong magnetic fields in the immediate vici-
Using a battery charger not recommended by nity (e.g. transformers, melting furnaces...) may
Leica Geosystems can destroy the batteries. This influence the compensator and lead to measuring
can cause fire or explosions. errors.
Precautions: Counter measures:
Only use chargers recommended by Leica Geosy- When measuring near strong magnetic fields, check
stems to charge the batteries. results for plausibility.
DANGER CAUTION
Because of the risk of electrocution, it is very Be careful when pointing the product towards
dangerous to use poles and extensions in the vicin- the sun, because the telescope functions as a mag-
ity of electrical installations such as power cables or nifying glass and can injure your eyes and/or cause
electrical railways damage inside the product..
Precautions: Precautions:
Keep at a safe distance from electrical installations. Do not point the product directly at the sun.
If it is essential to work in this environment, first con-
tact the safety authorities responsible for the electri- WARNING
cal installations and follow their instructions. Inadequate securing of the surveying site can
lead to dangerous situations, for example in traffic,
on building sites and at industrial installations.
Precautions:
Always ensure that the surveying site is adequately
WARNING secured. Adhere to the local regulations governing
By surveying during a thunderstorm you are accident prevention and road traffic.
at risk from lightning. CAUTION
Precautions: If the accessories used with the product are
Do not carry out field surveys during thunderstorms. not properly secured and the product is subjected to
mechanical shock, for example blows or falling, the
product may be damaged or people may sustain
injury.
135
Precautions: WARNING 136
When setting-up the product, make sure that the If computers intended for use indoors are
accessories, for example tripod, tribrach, connecting used in the field there is a danger of electric shock.
cables, are correctly adapted, fitted, secured, and
locked in position. Precautions:
Adhere to the instructions given by the computer
Avoid subjecting the product to mechanical stress.
manufacturer with regard to field use in conjunction
CAUTION with Leica Geosystems instruments.
When using a vertical staff supported by one CAUTION
brace there is always the danger of falling (e.g. by
wind gusts) and therefore danger of damage to During the transport, shipping or disposal of
batteries it is possible for inappropriate mechanical
equipment and danger of personal injuries.
influences to constitute a fire hazard.
Precautions:
Precautions:
Never leave a vertical staff supported by a brace
unsupervised (person at the staff). Before shipping the product or disposing of it,
discharge the batteries by running the product until
they are flat.
When transporting or shipping batteries, the person
in charge of the product must ensure that the appli-
cable national and international rules and regula-
tions are observed. Before transportation or
shipping contact your local passenger or freight
transport company.
WARNING Product specific treatment and waste management
If the product is improperly disposed of, the information can be downloaded from the Leica Geo-
following can happen: systems home page at http://www.leica-geosys-
tems.com/treatment or received from your Leica
If polymer parts are burnt, poisonous gases are
produced which may impair health. Geosystems dealer.
If batteries are damaged or are heated strongly, CAUTION
they can explode and cause poisoning, burning, Only Leica Geosystems authorized work-
corrosion or environmental contamination. shops are entitled to repair these products.
By disposing of the product irresponsibly you
may enable unauthorized persons to use it in
contravention of the regulations, exposing them-
Electromagnetic Compatibility (EMC)
selves and third parties to the risk of severe Description
injury and rendering the environment liable to The term "electromagnetic compatibility" is taken to
contamination. mean the capability of the instrument to function
Improper disposal of silicone oil may cause correctly in an environment where electromagnetic
environmental contamination. radiation and electrostatic discharges are present,
Precautions: and without causing electromagnetic disturbances
The product must not be disposed with in other equipment.
household waste. Dispose of the product WARNING
appropriately in accordance with the
national regulations in force in your Electromagnetic radiation can cause distur-
country. Always prevent access to the product by bances in other equipment.
unauthorized personnel.
137
Although this product meet the strict regulations Although the product meets the strict regulations 138
and standards which are in force in this respect, and standards which are in force in this respect,
Leica Geosystems cannot completely exclude the Leica Geosystems cannot completely exclude the
possibility that other equipment may be disturbed. possibility that the product may be disturbed by very
intense electromagnetic radiation, for example,
CAUTION near radio transmitters, two-way radios or diesel
There is a risk that disturbances may be generators.
caused in other equipment if the product is used in
Precautions:
conjunction with accessories from other manufac-
Check the plausibility of results obtained under
turers, for example field computers, personal com-
these conditions.
puters, two-way radios, non-standard cables or
external batteries. WARNING
Precautions: If the product is operated with connecting
Use only the equipment and accessories recom- cables attached at only one of their two ends, for
mended by Leica Geosystems. When combined example external supply cables, interface cables,
with this product, they meet the strict requirements the permitted level of electromagnetic radiation may
stipulated by the guidelines and standards. When be exceeded and the correct functioning of other
using computers and radios, pay attention to the products may be impaired.
information about electromagnetic compatibility pro- Precautions:
vided by the manufacturer. While the product is in use, connecting cables, for
CAUTION example product to external battery, product to
computer, must be connected at both ends.
Disturbances caused by electromagnetic
radiation can result in erroneous measurements.
FCC Statement, applicable in U.S. Increase the separation between the equipment
and receiver.
WARNING Connect the equipment into an outlet on a
This equipment has been tested and found to circuit different from that to which the receiver is
comply with the limits for a Class B digital device, connected.
pursuant to part 15 of the FCC Rules. Consult the dealer or an experienced radio/TV
These limits are designed to provide reasonable technician for help
protection against harmful interference in a residen-
WARNING
tial installation.
Changes or modifications not expressly
This equipment generates, uses and can radiate approved by Leica Geosystems for compliance
frequency energy and, if not installed and used in
could void the user's authority to operate the equip-
accordance with the instructions, may cause harm-
ment.
ful interference to radio communications.
However, there is no guarantee that interference
will not occur in a particular installation.
139
140
Product labelling
ImFeld_2
Im Feld_1
141
In a vehicle Shipping 142
ImAuto PerFracht
Never transport the instrument loose in a vehicle. For shippping by train, plane or ship always use
Shocks and vibrations may damage the instrument. the Leica Geosystems original packaging
Always transport the instrument secured in the (transport case and shipping container) or other sui-
casing. table packaging. The packaging protects the instru-
ment from shock and vibrations.
Storage
F C
Lagerung Koffer_gedreht
Keep an eye on temperature limits when If the instrument gets wet remove it from
storing your euipment, especially in the case, clean and dry it. Do the same with
summer inside a vehicle. the conainer, foam insert and accessories (at most
(-40C to +70C; -40F to +158F) at +40C/ +104F). Only place the instrument back
into the case after everything has dried.
Keep the transport case closed as often as possi-
ble.
143
Cleaning 144
Do not use aggressive liquids, as they
may damage the plastics parts.
Casing, keypad and display:
Only clean them with a clean, soft cloth and
if necessary dampened with a small amount of
suds.
Objective and Ocular:
Blow dust off the lense.
Do not touch the glass with your fingers.
Use only a clean, soft and lint-free cloth for
cleaning. If necessary, moisten the cloth
with water or pure alcohol.
1 2
2
1
Stativ_just Dosenlibelle_1
145
146
Reticle
/2
Dosenlibelle_2
Repeat steps 1 to 4 until the circular level bubble is If the collimation error exceeds 3mm per 30m, the
centered at any random telescope pointing. collimation needs to be adjusted.
Marked screw may not be used to adjust 1 Turn Allen key until design value is reached.
the bubble. 2 Check collimation.
147
Heating on/off switch, sets in below -5C Data backup PCMCIA-card (flash, 148
Dimensions SRAM),
up to 32MB capacity
Instrument
Height (incl. hand grip) 168mm +/-5mm Temperature range
Storage: -40C - +70C
Width
Operation -20C - +50C
on the side drives 240mm
instrument body 206mm Environmental conditions
Water and dust-proof IP53 (acc. IEC60529)
Length 210mm
Humidity up to 95% humidity no condensation
Container 468 x 254 x 355mm (L x B x H)
Magnetic field sensitivity
Weight
Line-of-sight difference in horizontal constant
incl. battery GEB111 2.85kg
magnetic field at a field strength of 0T up to
Measured values corrections 400T [4 Gauss]. 1"
Collimation error correction automatically
Battery powered
Earth curvature correction on/off switch; level Batteries (NiMh) GEB111 GEB121
probe with correction
Voltage 6V 6V
Record
Capacity 1800mAh 3600mAh
Internal storage approx. 6000
measurements or about Operating life DNA 12h 24h
1650 stations (BF) Battery adapter GAD39 only for Alkaline batteries,
Serial interface RS232 from "Measure & Record" 6 x LR6/AA/AM3, 1.5V
in GSI-8/16-format
Powered via serial interface
Voltage range using
an external cable 11.5V - 14V (DC)
Current consumption at 12V
maximum 500mA
Instrument on and
without lighting, typically: 70mA
149
150
Corrections/ formulas
Height differences d1, d2, d3, d4 = distances of the respective
dH = height difference of all sights in relation to staff heights.
back sight.
Distance balances
dh = height difference of two successive measured D Bal = DB- DF
points from the first measurement after the back
sight. DB =back sight distance
DF = fore sight distance
Example for a measure set of a station BF with inter-
mediate sightings Int1 and Int2: B - Int1 - Int2 - F. Total distance
dh1 = B - Int1 D Tot = DB + DF
dh2 = Int1 - Int2 Station distances
dh3 = Int2 - F D Stat = DB + DF
Earth curvature correction
Station differences
E = x2 / (2R) StatDiff = (B1 - F1) - (B2 - F2)
x = measured distance B1, F1, B2, F2, = staff heights
R = 6'378'000m (earth radius)
Line-of-sight error
= arctan [(A1 - B1 + B2 - A2) / (d1 - d2 + d3 - d4)]
A1, B1, B2, A2 = staff height
Accessories
Tripod
Staffs
ground plate of staff
struts
Power supply
batteries
charger
Recording
PCMCIA-card
Computer cable - serial interface
Computer software
Leica Geo Office
LevelPak-Pro
Documentation
User Manual
Field Manual
GSI Online
151
152
Sensor-error messages
List of the "Measurement not possible" messages:
Staff inverted or wrong INV setting. Check staff position and INV-setting.
S U
Safety Directions .............................................. 131 Unit ................................................................... 106
Unit selection .................................................... 110
Sensor .............................................................. 152
USER-key ................................................ 107, 111
Set ...................................................................... 64
Set code ........................................................... 108 V
Set of charaters .................................................. 44 Vibrations ........................................................... 49
Set out .......................................................... 64, 65 View /edit data .................................................. 116
Set out results .................................................. 128 W
Set tolerances .................................................... 76
Wildcard-search ................................................. 48
Single measurement .......................................... 53
Special characters .............................................. 45
Start programs .................................................... 72
Station results ..................................................... 85
Surveying intemediate points ............................. 62
157
Printed in Switzerland - Copyright Leica Geosystems AG, Heerbrugg,
726203-2.0.0en Translation from original text 726204-2.0.0de
Switzerland 2006
Total Quality Management: Our commitment to total customer satisfaction.
Ask your local Leica dealer for more information about our TQM program.
Leica Geosystems AG
Heinrich-Wild-Strasse
CH-9435 Heerbrugg
Switzerland
Phone +41 71 727 31 31
www.leica-geosystems.com