RM-Method Validation Protocol
RM-Method Validation Protocol
TABLE OF CONTENTS
Serial No. Contents Page No.
1 Purpose 02
2 Objective 02
3 Scope 02
4 Experimental Plan 02
5.2 Regents 03
5.3 Column 03
8.1 Accuracy 06
8.2 Precision 07
8.4 Speclficity L0
10 Deviation 11
L2 Revision History 12
1" purpose: To detail the standard procedures for validation of analytical method of Montelukast Sodium"
for analysis of
objective: The objective of validation of the analytical method is to demonstrate that the method is suitable
2.
Montelukast Sodium with accuracy, precision and reliability within defined range.
3. Scope: To evaluate the acceptability of analytical method for its intended purpose. lt will define the
procedure,
Montelukast Sodium'
documentation, references and acceptance criteria to be used in the validation of assay method of
4. Experimental Plan:
Repeatability Prepare six samples atLOO% concentration label and RSD of the active contents of six
(Method Precision) determine active content of each sample and also samples must be less than 2%.
calculate their RSD value.
lntermediate Prepare six samples at L00% concentration label RSD of the active contents of six
Precision from the same batch at next day by another analyst samples must be less than 2Yo.as per
and determine active content of each sample and Repeatability day 1.
also calculate their RSD value.
Linearity & Range Linearity to be performed in the range of about 60- The calibration curve should be linear (
140% with LOO% as working level concentration visually) and the value of R2 must be >
0.99
Specificity lnject standard, sample and blank solution according will be no interference at the
There-legion
to method. peak of Montelukast due to
blank.
System Suitability Prepare a standard solution al !00% concentration RSD of the area, retention time &
label and run six injections. height of peaks is within 2%, Avg' tailing
factor not more than 2.
QA Manaqer
Desiqnation
Signature
PD Officer
ta.v haro-
Asst. Manaoer, QC
'trtJ"-(
1At 0<'/
&;{
9t.,efb'lV
Date 1-<lh, I 4 11
ANALYTICAL METHOD VALIDATION PROTOCOL
Material Name Montelukast Sodium INN
5,2 REAGENT
Manufacture Batch No. Shelf Life
Sr. No. Reagent Name
01 Ammonium Acetate
02 Glacial Acetic Acid
03 Methanol
5.3 Column
ID
Sr. No. Description
01 Kromacil l-00-Crs HD, 1-50 x4.6 mm, 5pr or Equivalent
Principle:
of High
are injected consecutively into a suitable column
A prepared standard solution and sample solution
sodium present in sample is calculated
I
I
performance Liquid chromatography (Hplc). The content of Monterukast
I
I
bycomparingthepeakareaofactiveMontelukastSodiumpresentinsamplepreparationwiththatofstandard
preparation.
Procedure:
detector
Apparatus PerkinElmer Flexar HPLC integrated with PDA
I
Temperature 30'c
1.5 ml/rnin.
Flow rate
Load
L0 pl
Sample PreParation:
about
in to a crean and dry 100 mr vorumetric flask' Add
weigh accuratery and transfer about r.0.0 mg of sampre mark with
and sonicate for 5 minutes. Vorume up to the
60 mr of diruting sorution in to it, shake for 5 minutes
0,2 micron disk firter. cau this sorution B.
same sorvent and shake to mix weil.. Firter this through
Procedure:
A and sample preparation B into the tray of the auto sampler of the
Place vials containing standard preparation
chromatogram. Calculate the quantity of Montelukast in sample using
HPLC. Run the instrument and record the
the following equation.
Calculation:
S.METHoDSVAIIDATIoNANALYTICALPERFoRMANCECHARACTERISTICSToBEEVALUATED
8.1 AccuracY
accepted as either
between the varue found and the varue that is
a
Accuracy expresses the croseness of agreement the assay of known'
value. lt may often be expressed as the recovery by
conventional true value or an accepted reference g over a minimum of
using preferably a minimum of determinations
added amounts of anaryte. Accuracy shourd be assessed procedure')
(e.g. 3 concentrations /3 repricates each of totar analytical
3 concentration revers covering the specified range
Experimental Determination of Accuracy
less than 2%
Average recovery at each level within 98% to 102% & RSD of nine determinations must be
Acceptance criteria:
Result: n Complies tr Not ComPlies
Data Attached
8.2 Precision
The precision of an analytical procedure expresses the closeness
of agreement (degree of scatter) between a series of
measurements obtained from multiple sampling of the
same homogeneous sample under the prescribed conditions'
precision may be considered at three levels: repeatability, intermediate precision and reproducibiiity'
Result
AcCeptancg critgria: RSD of six samples should not be more than 2'0%
Analyst 1
Analyst 2
Done By
Location
lnstrument
Used
Date of
Analysis
Mean
Mean Label spl
Label spl Assay% Assay STDEV RSD%
Assay% STDEV RSD%
Assay% Claim% No.
Claim% No. %
L
7
2
2
100 3
100 3
,4 4
5
5
6
6
Acceptance criteria: RSD of six determinations must be less than 2/o as per repeatability day1"
Result:
Peak area of
Montelukast
Montelukast
of must be > 0'99
The calibration curve should be linear (visually) and the value
R2
Acceptance criteria:
Remarks: I Complies E Not Complies
Data Attached
8.4 SpecificitY
8.4.1 Blank effect
presence of components that may be expected to be
specificity is the ability to assess unequivocally the analyte in the
must be inarguable data for a method to be specific'
present such as impurities, degradation products, and excipients. There
from other species which might be present;
specificity measures only the desired component without interference
separation is not necessarily required'
Experimental Determination of Specificity:
*Preparestandardsolutionaccordingtomethod&measurePeakArea.
* Prepare Blank & run according to propose method
i. Determine percentage of interference due to blank"
%lnterference bY Blank
Result:
Avg. (n=6)
%RSD
10" Deviation
justified.
lf any deviation observed it must be mentioned and
11. REPORT
blank spaces in calculation and result will filled and will
be
Upon completion of the method validation experiments, the
piesented to QAM for, evaluation, review, and comment, after which a final
report will be issued'
& Recommendation as a result of Method Validation study shall be prepared and
summary & Discussion and conclusion
documented in the following format:
$ritas
ProtocolNo.: Method No. TP11.1.73010/01 Test parameter Assay by HPLC
vPr/MV-01s Revision No. 00 Date Revised New
Seritas
Department : Product DevetL
Material Name:
Summary & Discussion :
The data collected so far suggests that method is suitable and there is no significant change with respect to the established
limits.
Conclusion & Recommendation :
Based on this it is concluded that this method can be assigned as Analytical method for the assty of
C"rt''f'*t" .f Vrl'td.tl"*
Prepared by:
Signature:
Date :
Checked by:
Signature
Date
Approved by:
Signature:
Date :
Sl. No. Versi6n & lssue details Revision/Change in brief Reason for change