Manual UT Thickness Specific Exam From SAEP-1146 Name: Badge#: Date
Manual UT Thickness Specific Exam From SAEP-1146 Name: Badge#: Date
4a. UT equipment must be calibrated (checked using a standard) for all of these EXCEPT:
A) Calibration checks are required at the beginning and the end of each shift
B) Calibration checks are required every four (4) hours during the examination
C) Calibration checks are required if personnel or equipment are changed
D) Calibration checks are required at any time if equipment malfunctioning is suspected
E) Calibration checks are required after the crew returns from lunch
5. Equipment shall be calibrated using a two-point calibration range that overlaps the
expected measurement range. You must select a calibration block or step wedge block
thickness value greater than the maximum expected measurement value and a calibration
block or step wedge value less than the expected measurement value.
A) True B) False
5a. Per SAEP-1146, Equipment shall be calibrated using how many known standard values?
A) Single known point B) Two points C) Three points D) None of these are correct
5b. Equipment calibration ranges must:
A) Equal the known thickness range B) Overlap expected measurement ranges
C) Exceed the known thickness range D) None of these are correct
Manual UT Thickness Specific Exam from SAEP-1146
7. Test areas shall be free of what items that might interfere with the examination?
A) Corrosion including rust B) Insulation C) Weld Spatter
D) Surface irregularities E) Foreign matter F) All of the above
7a. Test areas to be examined shall be free of all of the following items EXCEPT:
A) Corrosion deposits including rust B) Surface Irregularities & insulation
C) Weld Spatter & other foreign matter D) Oil residues
10. For UT Thickness examinations on 19mm pipe, examiners must consider the effects of
the surface temperature. If the surface temperature is 121C, what must be done?
A) Surface temperature shall be measured and recorded
B) The calibration block must be heated up to allow correct readings to be taken
C) Temperature correction must be calculated and applied per SAEP-1146
D) All of the above are correct
10a. For UT Thickness examinations on 38mm plate, examiners must consider the effects of
the surface temperature. If the surface temperature is 150C, what must be done?
A) Surface temp is measured & recorded on UT report & no temp correction is required
B) Only the temperature correction need be calculated & applied per SAEP-1146 rules
C) Surface temp is measured & recorded on the report & temp correction is now required
D) None of the above is the correct way per SAEP-1146
Manual UT Thickness Specific Exam from SAEP-1146
14. UT thickness readings (on high temp surfaces and other surfaces) must remain on the
instrument display in what manner & for how long before they are considered valid?
A) Readings hold steady value within +/- 0.002 in. (0.05 mm) for at least 3 seconds
B) Readings hold steady value within +/- 0.002 in. (0.05 mm) for at least 4 seconds
C) Answer A is correct for high temp work & answer B is correct for all other work
D) All the above are correct! SAEP-1146 needs to be revised to show a 3 second rule
14a. UT thickness readings must remain on the instrument display in what manner & for how
long before they are considered valid?
A) Readings hold steady value within +/- 0.002 in. (0.05 mm) for at least 3 seconds
B) Readings hold steady value within +/- 0.002 in. (0.05 mm) for at least 4 seconds
C) Both A & B are correct & anything longer than 1 second is considered a valid reading
D) None of the above are correct, because a valid reading is held for several seconds
14b. UT thickness readings for high temp services must remain on the instrument display in
what manner & for how long before they are to be accepted and recorded on reports?
A) Readings are steady & hold for no longer than 1 second to avoid transducer damage
B) Readings are steady & hold for no longer than 2 seconds to avoid transducer damage
C) Readings are steady & hold for no longer than 3 seconds to avoid transducer damage
D) None of the above is correct, because a few seconds are required by SAEP-1146
15. UT Thickness measurements must be recorded using how many decimal points?
A) 2 decimal places for inches & mm B) 2 decimal places for inches & 3 for mm
C) 3 decimal places for inches & mm D) 3 decimal places for inches & 2 for mm
Manual UT Thickness Specific Exam from SAEP-1146
22. Water based couplants must be completely removed after the post-exam cleaning. How?
A) Surfaces must be wiped with a dry absorbent cloth or paper towel
B) Surfaces must be fresh water washed or otherwise cleaned in a professional manner
C) The immediate area is visually inspected for slipping hazards due to leftover couplant
D) Either A or B technique is acceptable provided that ALL couplant is removed, and
answer C is common sense & Good Safety Practice and should be practiced on the job
24. SAEP-1146 procedure for UTT is limited to contact testing & which wave techniques?
A) Shear or transverse waves
B) Plate or Lamb waves
C) Longitudinal or compression waves
D) Any of the above techniques may be used with a Level III approved procedure
25. What is the nominal wall thickness range of base materials per SAEP-1146?
A) 0.050 6.000 in B) 1.2 150 inches C) 1.2 mm 150 mm D) Both A & C
25a. Which of these below are not covered & may not be examined per SAEP-1146?
A) Plate used in the manufacture of any equipment (Tank, vessel, pipe, etc)
B) Tubing used in the construction of Heat Exchangers & other components
C) Piping including seamless wrought piping, welded piping, and others
D) Castings for valve bodies/fittings & forgings for LWN nozzles & flanges
E) All of the above may be UTT examined, so this is the correct answer
26. What API Standard is followed when taking thickness measurements on Plant piping?
A) API 510 B) API 653 C) API 1104 D) API 570
27. What API Standard is followed when taking thickness measurements on Pipelines?
A) API 510 B) API 653 C) API 1104 D) API 570