CLB21103 Process Instrumentation Experiment 3: Density Measurement
CLB21103 Process Instrumentation Experiment 3: Density Measurement
The purpose of the experiment were to determined the measurement of density varied by
different pressure or level transmitter and different temperature and also to check the density
calibration system using hand pump, DP calibrator, pressure module, multimeter and hand held
terminal.There were two different type of experiment being conducted that were with heater on
and heater off with both open and close tank.Two sets of reading also were taken for all run of
the experiment and the average was calculated so that the reading of experiment was
accurated.Throughout the experiment, it showed that the calculated specific gravity in both tank
is same which 0.9959 when heater off while when heater on was 0.9911.From open tank with
heater on and heater off,it can be seen that the higher the temperature,the lower the specific
gravity.It also can be observed that higher density is achieved when heater is turned
on.Meanwhile, from close tank with heater on and heater off,it can be said that, the higher
temperature, the lower specific gravity and the higher density is also achieved when heater is
on.To improved this experiment, several steps must be done such as proper handling must be
used when using the hydrometer and also eyes must be perpendicular with the scale to avoid any
parallax error.
OBJECTIVES
2. To check the density calibration systems using hand pump, DP calibrator, pressure
REMARKS:
B: Read from the Digital Density Indicator (DI). Take average value for further calculation.
w at opearting temperature TG
SG
w at reference temperature 4C
F: Calculate from equation deviation = (E-hydrometer)
REMARKS:
B: Read from the Digital Density Indicator (DI). Take average value for further calculation.
w at opearting temperature TG
SG
w at reference temperature 4C
DISCUSSION
The experiment is to evaluate and study the measurement of density. In this experiment
that was conducted, several parameters were taken into consideration where one experiment was
done with heater off and heater on. Both experiments were done with open tank closed tank,
where the pressure is considered as atmospheric pressure, Patm =1 atm. Closed tank on the other
hand is fully closed and the content is not exposed to air. The heater off tank is the tank with
water bath temperature at 29.0 C while the heater on tank is the tank with hot water temperature
The density is calculated using hydrometer by first obtained specific gravity, digital
density indicator (DI) and read from Table 2-28, Perry VII at temperature TG. Based on these
three method to calculate density, the method of using digital density indicator (DI) to obtain
From the comparison of first parameter which is pressure in open tank and closed tank
which is results in Table 1 (Run 1-OPEN TANK/HEATER OFF) and Table 1 (Run 2-CLOSED
TANK/HEATER OFF), it can be seen that the calculated specific gravity in open tank which is
0.9959 is same with closed tank. It should be the specific gravity in open tank higher than in
closed tank because of open tank have higher pressure than in closed tank. The actual density is
also higher in open tank 965.5 kg/m3 as compared to closed tank 955.5 kg/m3 because open tank
has higher pressure compared to closed tank. Thus, the deviation from hydrometer is same in
From the comparison of second parameter is the condition of temperature of open tank
with heater off and heater on which is results in Table 1 (Run 1-OPEN TANK/HEATER OFF)
and Table 2 (Run 1-OPEN TANK/HEATER ON), it can be seen that the higher temperature, the
lower specific gravity achieved after running experiment where the calculated specific gravity in
Table 1 (Run 1-OPEN TANK/HEATER OFF) is 0.9959 which is higher than Table 2 (Run 1-
OPEN TANK/HEATER ON) which is 0.9911. The actual density is also higher in
Table 1 (Run 1-OPEN TANK/HEATER OFF) is 965.5 kg/m3 compared to Table 2 (Run 1-
OPEN TANK/HEATER ON) which is 960.0 kg/m3 because the higher temperature has lower
specific gravity and actual gravity. Thus, the deviation from hydrometer in Table 1 (Run 1-
OPEN TANK/HEATER OFF) is 0.0388 higher than in Table 2 (Run 1-OPEN TANK/HEATER
ON) which is 0.0340. It can be observed that a higher density is achieved when the heater is
turned on.
closed tank with heater off and heater on which is results in Table 1 (Run 2-CLOSED
TANK/HEATER OFF) and Table 2 (Run 2-CLOSED TANK/HEATER ON), it can be seen that
the higher temperature, the lower specific gravity achieved after running experiment where the
calculated specific gravity in Table 1 (Run 2-CLOSED TANK/HEATER OFF) is 0.9959 which
is higher than Table 2 (Run 2-CLOSED TANK/HEATER ON) which is 0.9911. The actual
density is lower in Table 1 (Run 2-CLOSED TANK/HEATER OFF) is 955.5 kg/m3 compared
to Table 2 (Run 2-CLOSED TANK/HEATER ON) which is 958.5 kg/m3. It should be the higher
temperature has lower specific gravity and actual gravity, but it pressure in closed tank may be
affect the results. Thus, the deviation from hydrometer in Table 1 (Run 2-CLOSED
TANK/HEATER OFF) is 0.0388 higher than in Table 2 (Run 2-CLOSED TANK/HEATER ON)
which is 0.0340. It can be observed that a higher density is achieved when the heater is turned
on.