Mcclary Terry G 1963
Mcclary Terry G 1963
Abstract Approved
Major Professor)
reactor at several different flow rates. Since all flow rates were
in laminar flow, provision was made to develop a laminar velocity
profile in the reactant fluid before the reactants entered the reactor.
All four experimental reactors had the same volume. Reactors
of 1-, 1 -1 /4 -, 1 -1 /2 -, and 2-inch diameters were used. Six different
flow rates were used with each reactor. The amount of reaction at
each flow rate was determined by means of a titration method.
The 1 -, 1-1/4-, and 1-1/2-inch diameter reactors gave data
very similar to each other. At high flow rates all four reactors gave
similar data; however, at the low flow rates the 2 -inch diameter
reactor gave a marked decrease in amount of reactant conversion.
Comparison of the experimental data with the data calculated
for a theoretical laminar flow reactor showed that the 1-, 1-1/4-,
only at the very low flow rates. At high flow rates all four experi-
by
A THESIS
submitted to
OREGON STATE UNIVERSITY
in partial fulfillment of
the requirements for the
degree of
MASTER OF SCIENCE
June 1963
APPROVED:
Page
INTRODUCTION 1
THEORETICAL CONSIDERATION 4
EXPERIMENTAL APPARAT US 14
EXPERIMENTAL PROCEDURE 22
CALCULATIONS 28
DISCUSSION OF RESULTS 30
SUMMARY 37
NOME NC LA T URE 42
BIBLIOGRAPHY 44
APPENDIX 46
LIST OF FIGURES AND TABLES
Figure Page
1. Schematic Diagram of Reactor Test Apparatus 15
Table
1. Experimental Results 31
Reactor 49
8. Reactor Measurements 52
INTRODUCTION
reactor to obtain the desired residence time with the high turbu-
lent flow- rates. In order to achieve a sufficient residence time
while using a reasonable reactor length, it is often necessary to
flow rates.
The use of low flow rates, within tubular reactors, tends
to lower the percent conversion of reactants for a given average
residence time. It has been shown (3, 5, 9), for the same
2
turbulent flow. The laminar velocity profile is such that the longer
times some molecules spend in the reactor do not compensate for the
the reactants was studied at six different flow rates. The reaction
THEORETICAL CONSIDERATION
dx k (a-x)(b-x)
dt - 2 (1)
V = reactor volume
F = volumetric flow rate
a
then x = F +1 (3)
V ak
inar velocity profile was assumed and the effect of molecular dif-
fusion was neglected. Bosworth (1, p. 58) has shown that the effect
of molecular diffusion both axial and longitudinal, can be neglected
in most liquid flow reactors.
where,
= the pressure gradient acting down a cylindrical reaction
vessel of length L.
R = the radius of the reactor.
L 4r}L 1
(6)
v ci) (R2 - r2)
8
Upon differentiating
dt = 8rLr 1 2 (7)
dr cio
( R2 r2)
= 4(R2 - r2)r dr
(8)
R4
The fraction ftdt of the molecules which pass through the re-
action zone with transit time lying between t and t + dt is given by
1
ftdt = frdr = fr dt
dt
dr
= 4(R2-r2)3r dt
R4 8r1Lr
= 32n2L2 1 dt (9)
.1)2R4 t3
9
Since the minimum time of passage through the reactor, to, exists
to =
L 41,11 (10)
vo cia2
ftdt = 2t o
2
-
dt
t3
t= t ft dt
to (12)
= 2to
00
S.: f dt
t
to
Thus the average time of passage is just twice the time of passage
dx k2 (a - x)2 (1)
dt
This equation is the same equation as given for plug flow. If each
10
Then,
oo
X = 1 xftdt
co
o
= Jx ft dt (13)
Go to
C ft dt
to
and ak to
x = 2a2k2to[ l+toak21n( 2 )]. (15)
ak2to+1
a 22 k ak
1+ ak2V ln
V
x_ 2V [
( 2 )] (17)
F 2 F ak2V + 2F
product in the overflow leaving the tank will reach a constant value
which is equal to the concentration within the tank itself. Greenhalgh
the particles lose their identity. The contents of the reactor are
or
x2-x( F +2a)+a2= 0 (19)
k2V
F
k2V + 2a] ± (k + 2a)2- 4a210. 5
X =
[ C
(k2v V
2a
or
(a - x) =
F
+
aF 10. 5
- F (20)
L
( 2k 2 V )2 kk2V 2k2V
investigators (8, 12, 13, 15, 16) to be bimolecular and second order.
These investigators have shown that this reaction is essentially an
irreversible reaction.
The rate constants as a function of temperature are in excellent
neutralize the sodium hydroxide in the reactants and thus stop fur-
ther reaction. Since the hydrolysis of ethyl acetate by acid is very
much slower than with base (12), this method provides an accurate
way of determining the amount of reactant conversion for the ethyl
EXPERIMENTAL APPARATUS
and an overflow line at the top of the tank. The inlet, outlet and
J
I
I
The relay in turn was wired to a 110 volt power source and to a knife
were rated at 250 watts. Each constant head tank contained one
knife blade heater and one temperature controller.
The reactant outlet line from each constant head tank was con-
pyrex floats for the low flow rates and stainless steel floats for the
high flow rates. Flow rates to the reactor were controlled by the
two 3/8-inch needle valves.
ter holes. This was done at the end of the tubing opposite from the end
connected to the middle of the IT The holes in the tubing were spaced
in six rows along the length of the tubing. The reactants flowed into both
ends of the brass "T " and then through the seven inch long copper tubing
and out the holes into the calming section. The perforated tubing was
screwed into a 2 -inch cast iron flange with the aid of a 2 to 1 -inch
brass bushing. The 1-inch copper pipe was screwed through the
brass bushing until the copper pipe was flush with the face of the
flange. The space between the 1 -inch copper pipe and the flange
body was filled with lead and then filed flush with the flange face.
so that there was a five degree slope upward from the end where the
reactants entered. The slight upward slope prevented any air being
trapped in the reactor.
Both ends of each reactor were threaded and screwed into cast
of the reactor was flanged to the calming section of pipe using a rub-
ber gasket between the flange faces to prevent leakage. The other
end of the reactor was flanged to a 1 -inch diameter copper pipe 2-
1/2 inches long. This short piece of copper pipe was fitted to a
flange exactly like the calming section.
Inserted in the open end of the seven inch long piece of copper
pipe was a rubber stopper. Protruding through the stopper was a
0 to 51° C thermometer, conductivity electrodes and supports for a
copper disc. The disc had a diameter slightly smaller than the
diameter of the 1-inch copper pipe. Thus liquid could flow between
the wall of the pipe and the edge of the disc. In this way the velocity
19
3/8 -inch copper tubing fittedto the 1 -inch copper pipe. The exit
pieces of sheet platinum mounted in glass and spaced 5/8 -inch apart.
Before the electrodes were used they were coated with a film of
platinum black. The conductivity electrodes were wired to a stand-
ard Wheatstone bridge circuit. The Wheatstone bridge consisted of
two 0- 1,000 ohms, number 4775, and one 0- 100 ohms, number
type 1212 -A unit null detector was used to ascertain the point of
the extent of reaction but was used to determine when the reaction
Water
The distilled water used for all experimental work was freed of
Sodium Hydroxide
Ethyl Acetate
Kodak No. 37 benzoic acid. The benzoic acid was heated at 115°C
for two hours to remove any water. All standard solutions of sodi-
um hydroxide and hydrochloric acid were made with carbon dioxide
EXPERIMENTAL PROCEDURE
and then letting the head tank overflow back to the storage tank.
After mixing for over 1/2 -hour, a sample of solution was withdrawn
and the normality of the sodium hydroxide solution was determined
This was the amount required to make a 0.1 normal ethyl acetate
solution. The hole in the tank cover was then plugged with a rubber
stopper and the transfer pumps were started on both the sodium hy-
droxide and ethyl acetate systems. At the same time, both temper-
ature controllers were turned on. After a period of one and one-
half hours, two 50 milliliters samples of ethyl acetate solution were
withdrawn and added to a glass- stoppered flask containing 50 milli-
liters of 0.2484 normal sodium hydroxide. The samples were set
aside for latter analysis.
Flow into the reactor was begun by opening the needle valves on
24
the constant head outlet lines. The flow of each reactant was con-
trolled by adjusting the needle valve until the rotameter indicated
that the desired amount of reactant was flowing into the reactor.
The progress of the reaction was followed with the conductivity
apparatus. After the conductivity of the reactor product stream be-
came constant, samples of the product were taken and analyzed.
Four different size reactors were used, with six different flow
rates through each size reactor. Three different flow rates were
run using pyrex floats in the rotameters and three flow rates were
obtained using stainless steel floats in the rotameters. Due to the
quantity of reactants required for each run, additional reactant solu-
tions were prepared after three experimental runs.
After the reactor had reached steady state, as indicated by the
conductivity apparatus, and the inlet and outlet temperatures were
within 29.4 °C ± 0.2 °C, samples of the reactor product were taken.
Two samples were taken at each flow rate. Each sample was col-
lected in a 250 milliliter volumetric flask, to which an amount of
acid had been previously added. The amounts of acid added to the
flasks for each flow rate are shown in Tables 2 -7 in the appendix.
The acid in the flask had been calculated to be slightly in excess of
thus stop the reaction between ethyl acetate and sodium hydroxide.
25
order to completely mix the acid with the sample. The sample was
collected until the level of the liquid in the flask was just below the
neck of the flask. The volumetric flask was then filled with distilled
water, so that the total volume of liquid contained in the flask was
exactly 250 milliliters. The amount of distilled water added was
recorded. The inlet and outlet reactor temperatures were recorded
after the sampling.
After shaking the flask to thoroughly mix the distilled water
with the sample, a 100 milliliter sample from the flask was pipeted
into a 250 milliliter conical flask. The 100 milliliter sample was
recorded.
Before another sample was taken at the same flow rate, the
rotameters, thermometers and conductivity apparatus were checked
to make sure the operating variables were at the desired values. A
second sample was then taken, after which the needle valves were
used to change the flow rates to the next desired value. After re-
cording the inlet and outlet temperatures, the second sample was
treated exactly as the first one
This procedure was repeated for each desired flow rate until
26
three successive flow rates were run. The pumps and temperature
controllers were then shut off. Two 50 milliliter samples of the
ethyl acetate feed solution were withdrawn and treated as the first
two ethyl acetate samples. All of the ethyl acetate solution was
sucked out of both the storage and constant head tanks. Carbon di-
sodium hydroxide and ethyl acetate was exactly the same as the
procedure described above. The higher flow rate runs were made
using exactly the same procedure as previously described.
The flasks containing 50 milliliter samples of the ethyl acetate
known excess of standard acid was added to each flask. The sam-
ples were then back -titrated, using 0. 02050 normal sodium hydrox-
ide with phenolphthalein as the indicator. This procedure, used for
analyzing ethyl acetate, is similar to the one described by
Siggia (11, p. 24).
27
taken before and two after completion of a run. This was done in
order to verify the concentration of the ethyl acetate solution. Due
to the high vapor pressure of the ethyl acetate, the storage tank con-
taining the ethyl acetate solution was carefully sealed with poly-
ethylene tape. However, it was soon discovered that a leakproof
system was essentially impossible; accordingly, the double sam-
pling of the ethyl acetate feed was believed necessary to establish
its average concentration level.
After all the data were complete for a given size reactor,
another size was inserted into the system and the complete pro-
cedure was repeated again. This was done for a total of four
reactors. In order to check the consistency of the data two of the
CALCULATIONS
therefore
a
0. 09937(H) m 0. 05125(S) (24)
N
250-HmW
then
E (0.05-N)(100) (25)
0. 05
or
E _ 100 m
9. 937(H) 5.125(S)
+ (26)
0.05(250aH-W)
30
DISCUSSION OF RESULTS
reactor diameter was increased, but there was not enough differ-
ence between the reactors to establish a trend. However, the 2-
inch reactor at low flow rates was checked by rerunning the 2 -inch reac-
tor at the flow rates. Both sets of data agreed well with each other. The
difference between data obtained with the 2-inch reactor and with the oth-
er three reactor sizes might be caused either by molecular diffusion
or by "short circuiting" of the fluid in the 2 -inch reactor. However,
Bosworth (1, p. 858) was able to show mathematically that axial
molecular diffusion could be neglected when the length, L, is
greater than 255 h,FDT and radial diffusion could be neglected when
w
Figure 2 32
70
60
Reactants
50
b
á
of
40
Percent Conversion
>
0
U
q0 30
,REACTANT CONVERSION IN TUBU-
w
LAR REACTORS AS A FUNCTION OF
a RESIDENCE TIME
10
g." o diameter reactor.
Calculated values for a plug flow
reactor
Calculated values for a laminar
flow reactor
0
I ® I
Calculated values for a stirred
tank reactor
i Í
reactants and t is the average time the reactants spend in the reactor.
According to Bosworth °s criteria, the effect of molecular diffusion
can be neglected in all four sizes of reactors which were used in this
experiment. If diffusion can be eliminated as a factor in the varia-
tion of the data from the 2 -inch reactor with the other reactor sizes,
the variation probably was caused by the short circuiting of the re-
these two areas the main flow of liquid could short circuit and not
flow into these volumes. A dead space in the reactor would essen-
tially lower the effective reactor volume, thus lowering the percent
conversion of reactants in the overall reactor. At higher flow rates
in the 2 -inch reactor the fluid velocity was evidently high enough to
eliminate the dead space in the reactor. The 1-1/4-inch and 1-1/2-
inch diameter reactors probably had a high enough fluid velocity at
all flow rates to eliminate any dead spaces within the reactors.
Comparison of the experimental data with values calculated
for a laminar flow reactor shows that at low flow rates the 2 -inch
reactor gives less conversion than the calculated data. The three
other size reactors give more percent conversion at all flow rates
34
than the theoretical laminar flow reactor. At the lowest flow rate
the three smallest diameter reactors seem to be approaching the
flow rates, the experimental data were slightly lower than the cal-
culated values. The data obtained by Hovorka did not agree with the
values obtained in the present investigation, except possibly at the
very low flow rates. It is possible that the different method used to
introduce reactants had a great effect on the velocity profile through-
out the reactor and in turn on the conversion level. Also in the
present investigation the flanged joints used at both ends of the re-
actor section would have affected the velocity profile in the reactor.
Any radial mixing would increase the amount of conversion of react-
ants and axial mixing would decrease the conversion.
35
rates the experimental curves lie slightly above the conversion theo-
retically obtained for a plug flow reactor. It is, of course, theoretic-
ally impossible for the experimental reactors to be more efficient
than a plug flow reactor. This discrepancy was probably due to the
finite amount of time required to stop the reaction when obtaining a
sample from the reactor. The extra reaction time would have the
same effect as increasing the reactor volume and, hence, would
make the amount of reactant conversion high by only about 0.7 per-
cent. The additional residence time would not change the relative
amounts of conversion for the different size experimental reactors;
however, it would make a slight difference in a comparison with cal-
culated reactor conversions such as those for plug flow and stirred
tank reactors.
SUMMARY
ants were first introduced into a short calming section of pipe and
then allowed to flow into the tubular reactor. Four different diame-
ter reactors were used. Six different flow rates were used with
each of the experimental reactors. All of the reactors had the same
internal volume. Samples were collected from the exit of each re-
actor and added to a measured volume of acid. The acid was in
sufficient quantity to allow the sample to be back titrated with a
standard base. The percent completion of the reaction could then be
determined from the volumes of acid and base used for each sample.
The experimental data obtained for the three smallest di-
conversion was only apparent at the lower flow rates. At the higher
flow rates all diameter reactors gave approximately the same
38
mine what effect these variations would have on the percent conver-
sion of the reactants.
Some of the above proposed investigations could be performed
device would be needed to make the reactor fluid uniform across the
cross section of the reactor where the conductivity is to be
measured.
42
NOMENCLATURE
BIBLIOGRAPHY
At Start At Finish
Ethyl Acetate Reactant Concentration 0. 1000 N. O. 0993 N.
Sodium Hydroxide Reactant Concentration 0.1004 N. ----
1 336.7 32 5.70
53
Table 10. Calibration of the Ethyl Acetate Rotameter Using a Stainless Steel Float
Table 11. Calibration of the Sodium Hydroxide Rotameter Using a Pyrex Float
1 300.0 57 11.40
55
Table 12. Calibration of the Sodium Hydroxide Rotameter Using a Stainless Steel Float
Table 13. Calculated Data for the Equimolar Reaction of 0. 1 Normal Ethyl Acetate with 0. 1
Normal Sodium Hydroxide at 29. 4°C
0. 25 9.48 8.83 8. 80
57