Service Manual Ventilator SystemPb840
Service Manual Ventilator SystemPb840
Puritan Bennett
TM
Phone: +800.255.6774
Preface
Definitions
This manual uses three special indicators to convey information of a specific nature.
They include:
Warning
Indicates a condition that can endanger the patient or the ventilator operator.
Caution
Indicates a condition that can damage the equipment.
NOTE:
Indicates points of particular interest that make operation of the ventilator more
efficient or convenient.
Year of manufacture
The year of manufacture for ventilators whose serial numbers begin with 4200 is indicated on
the rear panel of the BDU as shown here. The two digit number following the year indicates
the month of manufacture.
2003 – 10
For ventilators whose serial numbers begin with 3510, the year of manufacture is indicated by
the serial number’s 5th and 6th digits. In the example below, a ventilator with serial number
3510021223 was the 1223rd unit built in 2002.
3510 02 1223
Sequential number
Year of manufacture
Electromagnetic susceptibility
The Puritan Bennett™ 840 Ventilator System complies with the requirements of IEC 60601-1-2
(EMC Collateral Standard), including the E-field susceptibility requirements at a level of
10 volts per meter, at frequencies from 26 MHz to 1 GHz, and the ESD requirements of this
standard. However, even at this level of device immunity, certain transmitting devices
(cellular phones, walkie-talkies, cordless phones, paging transmitters, etc.) emit radio
frequencies that could interrupt ventilator operation if located in a range too close to the
ventilator. It is difficult to determine when the field strength of these devices becomes
excessive. Practitioners should be aware radio frequency emissions are additive, and that the
ventilator must be located a sufficient distance from transmitting devices to avoid
interruption. Do not operate the ventilator in a magnetic resonance imaging (MRI)
environment. Chapter 7 describes possible ventilator alarms and what to do if they occur.
Consult with your institution’s biomedical engineering department in case of interrupted
ventilator operation and before relocating any life support equipment.
Warning
Accessory equipment connected to the analog and digital interfaces must be certified
according to IEC 60601-1. Furthermore, all configurations shall comply with the system
standard IEC 60601-1-1. Any person who connects additional equipment to the signal
input part or signal output part of the Puritan Bennett 840 ventilator configures a
medical system, and is therefore responsible for ensuring the system complies with
the requirements of the system standard IEC 60601-1-1. If in doubt, consult Covidien
Technical Support at 1.800.255.6774 or your local representative.
Customer assistance
If you require further assistance, contact Covidien at 1.800.255.6774 or your local Covidien
representative.
Additional information
For online technical support, visit the SolvITSM Center Knowledge Base
at http://www.puritanbennett.com. The SolvIT Center provides
answers to frequently asked questions about the Puritan Bennett
840 Ventilator System and other Covidien products 24 hours a day,
7 days a week.
2 Theory of Operation
2.1 Major ventilator subassemblies ................................................................................. 2-1
2.1.1 Breath delivery unit (BDU) ........................................................................... 2-2
2.1.2 Graphic user interface (GUI)......................................................................... 2-3
2.1.3 806 Compressor Unit..................................................................................... 2-4
2.1.4 802 or 803 Backup Power Source (BPS) ....................................................... 2-4
2.1.5 Cart ................................................................................................................ 2-5
2.1.6 Patient system ............................................................................................... 2-6
2.2 Operational overview ................................................................................................ 2-7
2.3 Pneumatic system .................................................................................................... 2-11
2.3.1 Inspiratory module...................................................................................... 2-16
2.3.1.1 Gas supply conditioning subsystem ........................................... 2-17
2.3.1.2 Flow control subsystem .............................................................. 2-22
2.3.1.3 Safety valve and inspiration monitoring subsystem ................. 2-25
2.3.1.4 Inspiratory module operation .................................................... 2-31
2.3.2 Patient System............................................................................................. 2-35
2.3.2.1 Patient system components........................................................ 2-36
2.3.2.2 Patient system operation............................................................ 2-36
2.3.3 Exhalation module...................................................................................... 2-37
2.3.3.1 Exhalation module components ................................................ 2-38
2.3.3.2 Exhalation module operation .................................................... 2-39
2.3.4 806 Compressor Unit................................................................................... 2-42
2.3.4.1 806 Compressor unit components ............................................. 2-43
2.3.4.2 806 Compressor unit operation ................................................. 2-46
2.4 Electrical system ....................................................................................................... 2-47
2.4.1 Summary of electrical components............................................................ 2-47
2.4.2 Overview of electrical system operation ................................................... 2-52
2.4.3 AC distribution components ..................................................................... 2-61
3 Self Tests
3.1 Introduction................................................................................................................ 3-1
3.2 How to enter Service Mode ....................................................................................... 3-1
3.3 Self tests and background checks.............................................................................. 3-1
3.3.1 POST (Power On Self Test)............................................................................ 3-1
3.3.2 SST (Short Self Test) ...................................................................................... 3-1
3.3.3 EST (Extended Self Test) ............................................................................... 3-2
3.3.4 Background checks ....................................................................................... 3-2
3.3.5 When self tests are run................................................................................. 3-2
3.4 Power on self test (POST)........................................................................................... 3-4
3.4.1 Safety ............................................................................................................ 3-7
3.4.2 POST characteristics ...................................................................................... 3-7
3.4.3 POST following power interruptions........................................................... 3-8
3.4.4 POST user interface....................................................................................... 3-8
3.4.5 Structure of POST ......................................................................................... 3-9
3.5 SST (short self test) ................................................................................................... 3-12
3.5.1 When to run SST ......................................................................................... 3-13
3.5.2 Hardware requirements ............................................................................ 3-14
3.5.3 Running SST ................................................................................................ 3-15
3.6 EST (extended self test)............................................................................................ 3-19
3.6.1 Description of EST ....................................................................................... 3-19
3.6.2 When to run EST ......................................................................................... 3-19
3.6.3 Hardware requirements ............................................................................ 3-20
3.6.4 How to run EST ........................................................................................... 3-20
3.6.5 How to run Single Test EST ........................................................................ 3-28
4 Service Mode
4.1 Accessing Service Mode ............................................................................................. 4-1
4.2 Service Mode functions ............................................................................................. 4-3
4.2.1 SST RESULT: Displaying SST results .............................................................. 4-5
4.2.2 DIAG LOG: Displaying error and status logs................................................ 4-5
4.2.3 Alarm log: Displaying the alarm history ..................................................... 4-7
4.2.4 VENT CONFIG: Displaying software revisions and serial numbers ............. 4-7
4.2.5 OPERATION TIME: Displaying ventilator and compressor elapsed hours.. 4-7
4.2.6 TEST SUMMARY: Displaying results from EST and SST ............................... 4-7
4.2.7 EST: Extended self test.................................................................................. 4-7
4.2.8 DATE/TIME: Adjusting date and time .......................................................... 4-7
4.2.9 EXIT: Exiting Service Mode .......................................................................... 4-8
4.2.10 Other screens............................................................................................... 4-8
4.2.10.1 Service Mode Setup .................................................................... 4-8
4.2.10.2 External Test Control: Performing remote ventilator testing .. 4-9
4.2.10.3 Exp Valve Calibration.................................................................. 4-9
4.2.10.4 Vent Inop Test ........................................................................... 4-12
4.2.10.5 Flow Sensor Calibration............................................................ 4-13
4.2.10.6 Atmospheric Pressure Transducer calibration ......................... 4-15
4.2.10.7 Data key Update ....................................................................... 4-16
5 Performance Verification
5.1 Tools, test equipment, and service materials ........................................................... 5-1
5.2 When to run ............................................................................................................... 5-1
5.3 Preliminary ventilator cleaning and inspection ....................................................... 5-4
5.4 Preliminary ventilator setup ...................................................................................... 5-4
5.5 Preliminary calibrations and tests ............................................................................. 5-5
5.6 Performance verification guidelines ......................................................................... 5-7
5.7 Performance verification tests................................................................................... 5-7
5.7.1 Electrical safety test ..................................................................................... 5-8
5.7.2 Ground isolation check................................................................................. 5-9
5.7.3 Extended self test (EST) ................................................................................ 5-9
5.7.4 Regulator setting verification .................................................................... 5-10
5.7.5 Serial loopback test (10.4-inch GUI only)................................................... 5-10
5.7.6 Performance verification using PTS 2000 Performance Test System and
BreathLab 840 VTS software .............................................................................. 5-10
5.7.6.1 How to view the test results log ................................................ 5-11
5.7.7 Manual ventilator check using equipment other than PTS 2000 Performance
Test System.......................................................................................................... 5-14
5.8 Regulator calibration hose repair instructions ....................................................... 5-15
6 Diagnostic Codes
6.1 Introduction................................................................................................................ 6-1
6.2 Reference Tables ........................................................................................................ 6-1
6.3 Troubleshooting......................................................................................................... 6-2
6.4 POST fault handling .................................................................................................. 6-2
6.5 Diagnostic CPU LED arrays ......................................................................................... 6-3
6.6 Diagnostic codes......................................................................................................... 6-4
6.6.1 How to interpret diagnostic codes .............................................................. 6-4
6.7 Organization of diagnostic codes table.................................................................... 6-5
6.8 System Diagnostic Log and BDU POST analog devices test ................................... 6-75
6.9 Diagnostic codes for POST faults............................................................................. 6-77
6.9.1 POST interrupt errors and test failures...................................................... 6-85
6.10 SST and EST test sequences and diagnostic codes................................................ 6-87
6.11 How to troubleshoot VGA LCD controller PCB faults (UT0002) (9.4-inch GUI only) 6-124
6.12 How to troubleshoot GUI LCD screen messages ............................................... 6-125
6.12.1 Example 1: Missing data key information ............................................. 6-125
6.12.2 Example 2: No communication between the GUI and BDU ................. 6-127
6.12.3 Example 3: POST or Background check errors....................................... 6-128
7 Alarm Handling
7.1 Alarm classifications ................................................................................................... 7-1
8.15.11.9 Exhalation heater (EXH HTR) and check valve (CV5) .......... 8-101
8.15.12 BDU housing ......................................................................................... 8-102
8.15.12.1 Removing BDU housing........................................................ 8-102
8.15.12.2 Installing BDU housing ......................................................... 8-104
8.15.12.3 Motherboard PCB ................................................................. 8-104
8.15.12.4 BDU alarm assembly ............................................................. 8-105
8.15.13 Power indicator..................................................................................... 8-106
8.15.13.1 Removing power indicator................................................... 8-106
8.15.13.2 Installing power indicator .................................................... 8-107
8.15.13.3 Alarm blindmate cable ......................................................... 8-107
8.15.13.4 Inspiratory blindmate cable ................................................. 8-108
8.15.13.5 DC power supply blindmate cable ....................................... 8-109
8.15.13.6 AC power supply blindmate harness ................................... 8-110
8.15.14 Release handle ...................................................................................... 8-112
8.15.14.1 Removing release handle ..................................................... 8-112
8.15.14.2 Installing release handle....................................................... 8-112
8.16 806 compressor unit............................................................................................. 8-112
8.17 Servicing the 806 compressor .............................................................................. 8-112
8.17.1 Compressor inlet filter ............................................................................ 8-113
8.17.2 Removing and installing the compressor inlet filter ............................ 8-113
8.17.3 Removing compressor from cart ............................................................ 8-113
8.17.3.1 Disconnecting compressor from BDU .................................... 8-113
8.17.3.2 Removing compressor from cart ............................................ 8-113
8.17.4 Removing top cover................................................................................ 8-114
8.17.5 Installing top cover ................................................................................. 8-115
8.17.6 Cooling fans ............................................................................................ 8-115
8.17.6.1 Removing a cooling fan.......................................................... 8-115
8.17.6.2 Installing a fan ........................................................................ 8-116
8.17.7 Removing the back panel....................................................................... 8-116
8.17.8 Reinstalling the back panel .................................................................... 8-116
8.17.9 Replacing an accumulator fitting and o-ring........................................ 8-117
8.17.10 Pneumatic hoses ................................................................................... 8-117
8.17.11 Plenum assembly................................................................................... 8-118
8.17.11.1 Removing the plenum assembly .......................................... 8-119
8.17.12 Heat exchanger (HE) ............................................................................. 8-120
8.17.12.1 Replacing the heat exchanger ............................................. 8-120
8.17.13 Removing and replacing the AC power cord ...................................... 8-120
8.17.14 Removing and replacing the data cable.............................................. 8-120
8.17.15 Compressor PCB .................................................................................... 8-121
8.17.15.1 Removing and replacing the compressor PCB..................... 8-121
8.17.16 Air dryer (dryer) and solenoid valve assembly .................................... 8-121
8.17.16.1 Removing and reinstalling the air dryer, solenoid valve assembly,
and replacing filters8-122
8.17.17 Compressor assembly............................................................................ 8-123
8.17.17.1 Removing the compressor assembly .................................... 8-123
8.17.17.2 Replacing the coalescing filter element .............................. 8-124
8.17.18 Replacing the compressor panels ........................................................ 8-125
8.17.19 Reinstalling the plenum assembly ....................................................... 8-126
8.17.20 Reconnecting electrical cables ............................................................ 8-126
9 Parts List
9.1 How to use this parts list ........................................................................................... 9-1
9.2 Puritan Bennett™ 840 Ventilator System patient system and accessories ............ 9-3
9.2.1 Puritan Bennett™ 840 Ventilator System NeoMode patient system and accesso-
ries ........................................................................................................................ 9-6
9.3 Flex arm assembly, oxygen and air hose assemblies, power cords ........................ 9-8
9.3.1 Flex arm assembly ........................................................................................ 9-8
9.3.2 Oxygen hose assemblies ............................................................................ 9-10
9.3.3 Air hose assemblies .................................................................................... 9-12
9.3.4 Power cords ................................................................................................ 9-14
9.4 Ventilator major assemblies ................................................................................... 9-16
9.4.1 Major ventilator assemblies (Puritan Bennett™ 800 Series Ventilator Compres-
sor Mount Cart) .................................................................................................. 9-18
9.4.2 Major ventilator assemblies (shown on Puritan Bennett™ 800 Series Ventilator
Pole Cart)............................................................................................................. 9-20
9.4.3 10.4-inch GUI (graphic user interface) assembly with Trending-enabled (TE) GUI
CPU PCB .............................................................................................................. 9-30
9.4.4 10.4-inch GUI (graphic user interface) assembly ...................................... 9-34
9.4.5 10.4-inch GUI (graphic user interface) handle ......................................... 9-36
9.4.6 10.4-inch graphic user interface (GUI) rotor assembly ............................ 9-38
9.4.7 10.4-inch GUI keyboards and speaker assembly ...................................... 9-40
9.4.8 9.4-inch graphic user interface (GUI) ........................................................ 9-43
9.4.8.1 9.4-inch GUI front housing assembly ......................................... 9-46
9.4.8.2 9.4-inch GUI rear housing assembly........................................... 9-48
9.4.9 Breath delivery unit (BDU) ........................................................................ 9-50
9.4.9.1 Exhalation module...................................................................... 9-54
9.4.9.2 Inspiratory module...................................................................... 9-58
9.4.9.3 BDU cover .................................................................................... 9-68
9.4.9.4 BDU chassis assembly .................................................................. 9-70
9.4.9.5 BDU chassis kit............................................................................. 9-72
9.4.10 806 compressor unit ................................................................................ 9-74
9.4.10.1 806 compressor base assembly................................................. 9-76
9.4.10.2 806 compressor unit enclosure assembly................................. 9-78
9.4.10.3 806 compressor unit plenum assembly.................................... 9-80
9.4.11 Backup power source (802 BPS) .............................................................. 9-82
9.4.12 803 Backup Power Source ....................................................................... 9-84
9.4.13 840 RTA (ready-to-assemble) Cart assembly ........................................... 9-86
9.4.14 Puritan Bennett™ 800 Series Ventilator Compressor Mount Cart assembly ........ 9-88
9.4.15 Puritan Bennett™ 800 Series Ventilator Pole Cart assembly ................. 9-91
Figures
Figure 1-1. Remote alarm (nurse’s call) port pinout (view from back of GUI) . . . . . . . . . . . . . . . . . 1-7
Figure 1-2. Puritan Bennett™ 840 Ventilator System RS-232 serial port pinout. . . . . . . . . . . . . . . . 1-7
Figure 1-3. Monochrome GUI front view (showing all keys) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-38
Figure 1-4. 10.4-inch GUI rear view . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-46
Figure 1-5. BDU front view . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-47
Figure 1-6. BDU I/O panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-50
Figure 1-7. BDU right-side panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-52
Figure 1-8. BDU rear view. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-53
Figure 1-9. GUI rear view . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-54
Figure 1-10. 802 BPS controls and indicators. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-55
Figure 1-11. 803 BPS controls and indicators. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-55
Figure 1-12. Puritan Bennett™ 800 Series Ventilator Compressor Mount Cart charging status and battery
indicator labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-56
Figure 1-13. Puritan Bennett™ 800 Series Ventilator Pole Cart charging status and battery indicator labels
1-57
Figure 2-1. L-R Puritan Bennett™ 840 Ventilator shown on RTA cart,
Puritan Bennett 800 Series Ventilator Compressor Mount Cart,
and Puritan Bennett 800 Series Ventilator Pole Cart . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Figure 2-2. BDU . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Figure 2-3. 10.4-inch GUI . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Figure 2-4. Compressor unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Figure 2-5. 802 BPS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Figure 2-6. 803 BPS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Figure 2-7. RTA cart, Puritan Bennett™ 800 Series Ventilator Compressor Mount Cart, and Puritan
Bennett 800 Series Ventilator Pole Cart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Figure 2-8. Patient system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Figure 2-9. NeoMode patient system. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
Figure 2-10. Puritan Bennett™ 840 Ventilator System block diagram . . . . . . . . . . . . . . . . . . . . . . . . 2-8
Figure 2-11. Pneumatic system block diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
Figure 2-12. Pneumatic system diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12
Figure 2-13. Inspiratory module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-16
Figure 2-14. Inspiratory module in ventilator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-16
Figure 2-16. Gas supply conditioning subsystem. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-17
Figure 2-17. Gas supply conditioning subsystem gas flow diagram . . . . . . . . . . . . . . . . . . . . . . . . . 2-18
Figure 2-18. Gas supply conditioning subsystem components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-20
Figure 2-19. Flow control subsystem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-22
Figure 2-20. Flow control subsystem gas flow diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-23
Figure 2-21. Hot film . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-23
Figure 2-22. Flow control subsystem components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-24
Figure 2-23. Safety valve and inspiration monitoring subsystem . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-26
Figure 2-24. Inspiration monitoring subsystem gas flow diagram. . . . . . . . . . . . . . . . . . . . . . . . . . . 2-26
Figure 2-25. Safety valve and inspiration monitoring subsystem components. . . . . . . . . . . . . . . . . 2-28
Figure 2-26. Safety valve open gas flow diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-30
Figure 2-27. Air flow diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-31
Figure 2-28. Oxygen flow diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-33
Figure 2-29. Patient system (minus exhalation collector vial and expiratory filter) . . . . . . . . . . . . . 2-35
Figure 2-30. Patient system flow diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-36
General Information
Cart can also be used with a one-hour battery or an optional four-hour battery, located in the
cart base. The Puritan Bennett™ 800 Series Ventilator Pole Cart cannot be used with a
compressor.
Warning
Install only ventilator BDUs with serial numbers starting with 3512 onto the newer
Puritan Bennett 800 Series Ventilator Compressor Mount Cart and Puritan Bennett 800
Series Ventilator Pole Cart. Other ventilator serial numbers are not compatible with
the newer carts.
The Puritan Bennett™ 800 Series Ventilator Compressor Mount Cart and the Puritan Bennett
800 Series Ventilator Pole Cart may not be available in all regions. Please contact your local
Covidien representative for more information.
The BDU is the core of the ventilator. It contains the pneumatics that deliver gas and
electronic and electrical systems that control pneumatics, monitor alarms, and distribute
power. The ventilator includes two independent central processing units (CPUs): one for the
BDU that controls ventilation, and one for the GUI that monitors ventilator and patient data.
The GUI CPU verifies the BD CPU is functioning properly and prevents a single fault from
causing a simultaneous failure of controlling and monitoring operations.
The GUI provides a communications path between the ventilator and the operator. By using
the touch screen, keys, and knob on the GUI the practitioner gives initial instructions and
data to the ventilator. The GUI CPU processes this information and stores it in the ventilator’s
memory. The BD CPU uses this stored information to control and monitor the flow of gas to
and from the patient.
The Puritan Bennett 840 Ventilator System supplies mandatory or spontaneous breaths with
a preset oxygen concentration. A mandatory breath can be pressure- or volume-controlled. A
spontaneous breath allows the patient peak inspiratory flows up to 200 L/min with or
without pressure support.
The 802 BPS (Backup Power Supply) or 803 BPS (extended BPS) provides DC power to the BDU
in the event AC power is lost. A new, fully charged 802 BPS allows the ventilator to function
(without compressor or humidifier) for at least 60 minutes (30 minutes on ventilators built
prior to July 2007); thus, the BPS can power the ventilator for transport purposes within the
respiratory care facility. The 803 BPS can power the ventilator for at least four hours under
the same conditions.The BPS should always be connected to the ventilator during operation.
The optional compressor unit provides compressed air to the BDU, and can be used in place of
wall or bottled air for normal operation. The compressor unit is powered by and
communicates with the BDU.
NOTE:
An external air source is required when performing service mode calibrations and
performance verification testing.
1.4 Accessories
The following accessories are either required or can be used with the ventilator.
Patient circuit: A variety of reusable Covidien patient circuits, adult and pediatric, with and
without water traps, and with and without heated wire, is available. Consult the “Part
numbers” appendix of the Puritan Bennett™ 800 Series Ventilator System Operator’s and
Technical Reference Manual for patient circuit ordering information.
Humidification device: The Puritan Bennett 840 Ventilator System supports the use of an
optional humidification device, including a heated humidifier, heat and moisture exchanger
(HME), or heated wire. A mounting bracket is available for the Fisher & Paykel™* Humidifier.
An AC socket for the humidifier is available on 100 and 120 V ventilators.
Remote nurse call unit: An analog output connector permits connection to a remote nurse
call unit. (Covidien does not supply nurse’s call units or cables.) Refer to Communication
Capabilities in Table 1-1 for remote alarm port (nurse call) specifications.
Caution
• Do not connect or disconnect the ventilator’s graphic user interface (GUI), backup
power source (BPS), or compressor while the power switch is on or the ventilator is
connected to AC power.
• All components must be securely mounted and connected by qualified service
personnel according to the appropriate Covidien installation instructions.
• Do not obstruct the breath delivery unit (BDU), GUI, or compressor cooling vents or
fan vents.
• To avoid possible damage to ventilator components, do not use the horizontal
surfaces of the ventilator to place or stack objects.
1.5 Specifications
Physical characteristics
Dimensions BDU: 330 mm high x 457 mm wide x 254 mm deep (13 in. high x 18 in. wide x 10 in. deep)
GUI: 460 mm high x 394 mm wide x 170 mm deep (18.1 in. high x 15.5 in. wide x 6.7 in. deep)
802 BPS: 83 mm high x 244 mm wide x 254 mm deep (3.25 in. high x 9.6 in. wide x 10 in. deep)
803 BPS (extended BPS for use with RTA cart): 95 mm high x 438 mm wide x 260 mm deep
includes housing and bracket (3.75 in. high x 17.25 in. wide x 10.25 in. deep)
RTA Cart: 998 mm high x 582 mm wide x 602 mm deep (39.3 in. high x 22.9 in. wide x
23.7 in. deep)
Puritan Bennett 800 Series Ventilator Compressor Mount Cart: 1041 mm high x 686 mm wide
x 839 mm deep (41 in. high x 27 in. wide x 33 in. deep with wheels in outermost position)
Puritan Bennett 800 Series Ventilator Pole Cart: 1041mm high x 686 mm wide x 839 mm deep
(41 in. high x 27 in. wide x 33 in. deep with wheels in outermost position)
804 Compressor: 417 mm high x 458 mm wide x 362 mm deep
(16.4 in. high x 18 in. wide x 14.25 in. deep) (no longer available)
806 Compressor: 425 mm high x 458 mm wide x 362 mm deep
(17 in. high x 18 in. wide x 14.25 in. deep)
Environmental requirements
Pneumatic specifications
Oxygen and air inlet Pressure: 241 to 690 kPa (35 to 100 psi)
supplies
Warning
Due to excessive restriction of the Air Liquide™*, SIS, and Dräger™* hose assemblies,
reduced ventilator performance levels may result when oxygen or air supply pressures
< 50 psi (345 kPa) are employed.
Oxygen sensor life The oxygen sensor should be replaced one year after date of manufacture, or as often as
necessary. Actual sensor life depends on operating environment; operation at higher
temperature or O2% levels will shorten sensor life.
Gas mixing system Range of flow from the mixing system: Can be set to 150 L/min standard temperature and
pressure, dry (STPD). Additional flow is available (up to 80 L/min for pediatric patients whose
IBW 24 kg, and up to 200 L/min for adults whose IBW > 24 kg) for compliance
compensation.
Leakage from one gas system to another: Meets standard EN 60601-2-12
Operating pressure range: 35 to 100 psi (241 to 690 kPa)
Air/oxygen regulator bleed: Up to 3 L/min
Electrical specifications
NOTE:
Above values obtained using the following ventilator settings at 22 C ambient temperature: mode, A/C;
mandatory type, PC; IBW, 85 kg; fTOT, 20/min; PSUPP, 30 cmH2O; TI, 1 s; Rise Time Percent (was Flow
Acceleration), 50%; O2%, 50%; PMEAN, 50 cmH2O; PSENS, 3 cmH2O. Input power specifications are for
ventilators with Fisher & Paykel™* MR730 humidifiers. (Humidifier connection only available on 100 – 120 V
ventilators.)
Warning
In the event of a defective earth conductor, connecting equipment to the auxiliary
mains socket outlet(s) (that is, the humidifier or compressor connections) may increase
patient leakage current to values that exceed the allowable limits.
NOTE:
BPS battery life specifications are approximate. To ensure maximum battery life,
maintain full charge and minimize the number of complete discharges.
Communications Remote alarm (nurse’s call) port (Figure 1-1). Allows medium- and high-urgency alarm
capabilities conditions to be annunciated at locations away from the ventilator (for example, when the
ventilator is in an isolation room). The ventilator signals an alarm using a normally open or a
normally closed signal. The ventilator asserts a remote alarm when there is an active
medium- or high-urgency alarm condition, unless the alarm silence function is active, and
when the ventilator power switch is turned off. The remote alarm port is a 4-pin female
connector. Allowable current is 500 mA at 30 V DC (maximum).
2 3
1 4
Pin Signal
2 Relay common
4 Not connected
Figure 1-1. Remote alarm (nurse’s call) port pinout (view from back of GUI)
Communication RS-232 (serial) port (Figure 1-2). A 9-pin male connector configured as data terminal
capabilities (cont) equipment (DTE). Allowable current is 0.2 A at 10 V DC (maximum).
1 2 3 4 5
6 7 8 9
Figure 1-2. Puritan Bennett™ 840 Ventilator System RS-232 serial port pinout
Pin Signal
1 Not connected
5 Ground (GND)
6 Not connected
9 Not connected
NOTE:
The connection of accessories or equipment to the ventilator’s RS-232 and remote
alarm ports requires electrical separation of conductive earth (ground) connections
between different items of equipment of a system. The system cable assembly
shielding can interconnect the ventilator and other equipment, increasing the risk of
excessive enclosure leakage current from external equipment. For safe connection
and prevent a double earth ground, the shield conductor must have separation of
4 mm between the equipment and ventilator conductors, as shown below:
To Puritan Bennett™
840 Ventilator System Minimum shield
RS-232 and remote alarm Cable conductive separation
input connectors shielding 4 mm
North America
Authorized to bear the CSA certification mark with 120 V, 60 Hz Canadian Standards
NRTL/C indicator, signifying the product has been 220 – 240 V, 50 Hz Association (CSA)
evaluated to the applicable ANSI/Underwriters 220 – 240 V, 60 Hz
Laboratories Inc. (UL) and CSA standards for use in the US
and Canada.
CSA Std. No. 601-1-M90
CSA 601-1 Supplement 1:1994
CSA Std. No. 60601-2.12-1994
UL No. 60601-1 (1st Edition)
IEC 60601-1:1988
IEC 60601-1 Amendment 1:1991
IEC60601-1 Amendment 2:1995
IEC 60601-2-12:2001
Manufacturer self-
IEC 60601-1-2:2007
certification
International
European
Approved to the type test requirements of Annex III of 220 – 240 V, 50 Hz TÜV Product Service
the Medical Device Directive. 220 – 240 V, 60 Hz
EN 60601-1:1990
EN 60601-1 Amendment 1:1993
EN 60601-1 Amendment 11:1993
EN 60601-1 Amendment 12:1993
EN 60601-1 Amendment 2:1995
EN 60601-1 Amendment 13:1996
IEC 60601-2-12:2001
0123
EN 60601-1-2:2001+A1:2006 Manufacturer self-
certification
Warning
Portable and mobile RF communications equipment can affect the performance of the
Puritan Bennett 840 Ventilator System. Install and use this device according to the information
contained in this manual and the Puritan Bennett™ 800 Series Ventilator System Operator’s and
Technical Reference Manual.
Warning
• The Puritan Bennett 840 Ventilator System should not be used adjacent to or stacked with
other equipment, except as specified in this manual and the Puritan Bennett™ 800 Series
Ventilator System Operator’s and Technical Reference Manual. If adjacent or stacked use is
necessary, the Puritan Bennett 840 Ventilator System should be observed to verify normal
operation in the configurations in which it will be used.
NOTE:
This is a class A product and is intended to be used in a hospital environment only. If used outside
of the hospital environment, this equipment may not offer adequate protection to radio-
frequency communication services. The user may be required to take mitigation measures, such
as relocating or re-orienting the equipment.
The Puritan Bennett™ 840 Ventilator System is intended for use in the electromagnetic environment specified below.
The customer or the user of the Puritan Bennett 840 Ventilator System should ensure it is used in such an
environment.
Radiated RF emissions Group 1 The Puritan Bennett 840 Ventilator System uses RF
CISPR 11 Class A energy only for its internal functions. Therefore, its RF
emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
Conducted RF emissions Group 1 The Puritan Bennett 840 Ventilator System is suitable
CISPR 11 Class A for use in all establishments including domestic
establishments and those directly connected to the
Harmonic emissions Class A public low-voltage power supply network that supplies
IEC 61000-3-2 buildings used for domestic purposes.
The Puritan Bennett™ 840 Ventilator System is intended for use in the electromagnetic environment specified below.
The customer or the user of the Puritan Bennett 840 Ventilator System should ensure it is used in such an environment.
Electrical fast ± 2 kV for power ± 2 kV for power Mains power quality should be that of a
transient/burst supply lines supply lines typical commercial or hospital
IEC 61000-4-4 environment.
± 1 kV for input/ ± 1 kV for input/
output lines output lines
Voltage dips, short < 5% UT < 5% UT Mains power quality should be that of a
interruptions and (> 95% dip in UT (> 95% dip in UT typical commercial or hospital
voltage variations on for 0.5 cycle) for 0.5 cycle) environment. If the user of the
power supply input Puritan Bennett 840 Ventilator System
lines requires continued operation during
40% UT 40% UT
IEC 61000-4-11 power mains interruptions, it is
(60% dip in UT (60% dip in UT recommended the Puritan Bennett
for 5 cycles) for 5 cycles) 840 Ventilator System be powered from
an uninterruptible power supply or a
70% UT 70% UT battery.
(30% dip in UT (30% dip in UT
for 25 cycles) for 25 cycles)
< 5% UT < 5% UT
(> 95% dip in UT for (> 95% dip in UT
5 s) for 5 s)
NOTE:
UT is the AC mains voltage prior to application of the test level.
The Puritan Bennett™ 840 Ventilator System is intended for use in the electromagnetic environment specified below. The
customer or the user of the Puritan Bennett 840 Ventilator System should ensure it is used in such an environment.
NOTE:
• At 80 MHz and 800 MHz, the higher frequency range applies.
• These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects,
and people.
a The ISM (industrial, scientific, and medical) bands between 150 kHz and 80 MHz are 6.765 MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to
the likelihood mobile/portable communications equipment could cause interference if it is inadvertently brought into patient areas. For this reason, an
additional factor of 10/3 is used in calculating the recommended separation distance for transmitters in these frequency ranges.
c Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters,
an electromagnetic site survey should be considered. If the measured field strength in the location in which the Puritan Bennett™ 840 Ventilator System is
used exceeds the applicable RF compliance level above, the Puritan Bennett 840 Ventilator System should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the Puritan Bennett 840 Ventilator System.
d Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
Table 1-6: Recommended separation distances between portable and mobile RF communications equipment
and the Puritan Bennett™ 840 Ventilator System
The Puritan Bennett 840 Ventilator System is intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled. The customer or the user of the Puritan Bennett 840 Ventilator System can help
prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Puritan Bennett 840 Ventilator System as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum
output power of Separation distance according to frequency of transmitter (m)
transmitter (W)
150 kHz to 80 MHz 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
outside ISM bands in ISM bands
100 3.5 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE:
• At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
• The ISM (industrial, scientific, and medical) bands between 150 kHz and 80 MHz are 6.765 MHz to 6.795
MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz.
• An additional factor of 10/3 is used in calculating the recommended separation distance for transmitters in
the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to 2.5 GHz to
decrease the likelihood mobile/portable communications equipment could cause interference if it is
inadvertently brought into patient areas.
• These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects, and people.
Covidien does not supply remote alarm (nurse call) or serial port cables. In order to maintain compliance to
International Electromagnetic Compatibility (EMC) standards, Covidien recommends using shielded cables for these
applications.
Warning
The use of accessories and cables other than those specified, with the exception of parts sold by Covidien as
replacements for internal components, may result in increased emissions or decreased immunity of the
Puritan Bennett™ 840 Ventilator System.
4-078107-00, 4-078107-SP 10 ft (3 m)
Power cord, latching, North America
4-078108-00, 4-078108-SP 10 ft (3 m)
Power cord, latching, Europe
4-078109-00, 4-078109-SP 10 ft (3 m)
Power cord, latching, Japan
4-078110-00, 4-078110-SP 10 ft (3 m)
Power cord, latching, Australia
4-071421-00 10 ft (3 m)
Power cord, Denmark
4-071422-00 10 ft (3 m)
Power cord, India/S. Africa
4-071423-00 10 ft (3 m)
Power cord, Israel
4-078144-00 10 ft (3 m)
Power cord, UK
4-031323-00 10 ft (3 m)
Power cord, Italy
4-031325-00 10 ft (3 m)
Power cord, Switzerland
4-071441-00 10 ft (3 m)
Cable assembly, GUI to BDU
NOTE:
When pressure units are set to hPa, pressure delivery and spirometry are subject to an
additional 2% error.
Maximum working pressure 100 cmH2O (98.1 hPa), ensured by high pressure limit
90 cmH2O (pressure-based ventilation)
Volume measurements:
Type: Hot film anemometer.
Sensing position: Exhalation compartment
Measurements:
Exhaled tidal volume
Range: 0 to 6,000 mL
Total minute volume
Range: 0 to 99.9 L
Results of ventilator patient circuit testing (using circuits Inspiratory pressure drop from inlet of open safety valve
identified for use with Puritan Bennett™ 840 Ventilator to outlet port without inspiratory filter:
Systems) At 5 standard liters per minute (SL/min): 0.06 cmH2O
At 30 SL/min: 0.28 cmH2O
At 60 SL/min: 0.95 cmH2O
Inspiratory pressure drop across inspiratory filter:
At 5 SL/min: 0.17 cmH2O
At 30 SL/min: 0.56 cmH2O
At 60 SL/min: 1.37 cmH2O
Results of ventilator patient circuit testing (using circuits Pressure drop across Fisher & Paykel™* humidifier and
identified for use with Puritan Bennett™ 840 Ventilator lead-in tube:
Systems) (continued) Neonatal patient circuit at 5 SL/min: 0.14 cmH2O
Pediatric patient circuit at 30 SL/min: 0.28 cmH2O
Adult patient circuit at 60 SL/min: 0.93 cmH2O
Internal volume:
Inspiratory pneumatics: 50 mL 5 mL
Expiratory pneumatics: 1000 mL 25 mL (including
expiratory filter and collector vial)
The Puritan Bennett 840 Ventilator System automatically
adjusts for volume losses due to gas compressibility (that
is, automatic compliance compensation), subject to a
maximum delivered volume of 2500 mL.
NOTE:
• Patient circuit testing specifications are with the ventilator powered off, and are based on the
recommended configurations shown in the Puritan Bennett™ 800 Series Ventilator System Operator’s and
Technical Reference Manual.
• To ensure that compliance compensation functions correctly, the user must run SST with the circuit
configured as intended for use on the patient.
Bacteria filter efficiency 99.97% for nominal particle size of 0.3 μm (micron) at
100 L/min
Ventilator settings
Apnea ventilation
Apnea tidal volume (VT) Range: 3-315 mL* with NEONATAL patient circuit
25 mL with PEDIATRIC or ADULT patient circuit
2500 mL with ADULT patient circuit
IBW based range: 1.16 x IBW minimum, 45.7 x IBW maximum
Resolution: 0.1 for 3 to 5* mL
1.0 mL for 5 to 100 mL
5 mL for 100 to 400 mL
10 mL for 400 to 2,500 mL (full scale)
Accuracy (INVASIVE vent type only):
For TI < 600ms, (10+10% x 600ms/TIms of setting), mL
otherwise (10+10% of setting), mL with PEDIATRIC or ADULT patient
circuit;
± (4+10% of setting), mL with NEONATAL patient circuit
Applicable during steady state when 3 consecutive breaths are within
1% of each other, and only when the flow limit and compliance
compensation limits are not reached.
New patient: Maximum of 3* mL or (7.25 x IBW); with NEONATAL patient
circuit
(7.25 x IBW); with PEDIATRIC or ADULT patient circuit
*Assumes NeoMode 2.0 software option is installed
Constant during rate change Range: Inspiratory time, I:E ratio, or expiratory time; TH, TL, TH:TL in BILEVEL
Resolution: Not applicable
Accuracy: Not applicable
New patient: Inspiratory time
Expiratory sensitivity (ESENS) Range: 1 to 80% (1 to 10 L/min when Spontaneous Type is PA)
Resolution: 1%
Accuracy: Not applicable
New patient: 25% (3 L/min when Spontaneous Type is PA)
Flow sensitivity (VSENS) Range: 0.1 to 10.0 L/min with NEONATAL patient circuit
0.2 to 20.0 L/min with PEDIATRIC or ADULT patient circuit
Resolution: 0.1 L/min
Accuracy: Not applicable
New patient: 1.0 L/min with NEONATAL patient circuit
2.0 L/min with PEDIATRIC patient circuit
3.0 L/min with ADULT patient circuit
Humidification type Range: HME, non-heated expiratory tube, or heated expiratory tube
Resolution: Not applicable
Accuracy: Not applicable
New patient value: previous setting
Ideal body weight (IBW) Range: 0.3 kg (0.66 lb)* to 7 kg (15.4 lb) with NEONATAL patient circuit
3.5 kg (7.7 lb) to 35 kg (77 lb) with PEDIATRIC patient circuit
7.0 kg (15.4 lb) to 150 kg (330 lb) with ADULT patient circuit
Resolution: 0.1 kg for 0.3* kg to 3.5 kg
0.5 kg for 4.0 to 9.5 kg
1 kg for 10 to 50 kg
5 kg for 50 to 100 kg
10 for 100 to 150 kg
Accuracy: Not applicable
New patient: 3.0 kg with NEONATAL patient circuit
15 kg with PEDIATRIC patient circuit
50 kg with ADULT patient circuit
*Assumes NeoMode 2.0 software option is installed
I:E ratio Range: 1:299 I:E 4.00:1; 1:299 < I:E < 149:1 (BILEVEL mode only)
Resolution: 0.01 for 1:9.99 to 4.00:1
0.1 for 1:99.9 to 1:10.0
1 for 1:299 to 1:100
Accuracy: 0.01 s of the inspiratory time determined by the I:E ratio and
respiratory rate settings
New patient: 1: TE /TI
Patient circuit type Range: Neonatal (if Neo-mode option is active), Pediatric, or Adult
Resolution: Not applicable
Accuracy: Not applicable
Peak inspiratory flow (VMAX) Range: 1.0 to 30 L/min with NEONATAL patient circuit
3.0 to 60 L/min with PEDIATRIC patient circuit
3 to 150 L/min with ADULT patient circuit
Resolution: 0.1 L/min for flows of 1 to 20 L/min; 1 L/min for flows above
20 L/min
Accuracy: ± (0.5 + 10% of setting) L/min of the flow command input to the
flow controller, at the end of each control interval, after the first
100 milliseconds of inspiration.
Safety ventilation Settings are identical to new patient values, except: mode = A/C,
mandatory type = PC, respiratory rate = 16/min, inspiratory time = 1 s,
inspiratory pressure = 10 cmH2O, PEEP = 3 cmH2O, trigger type = pressure,
Rise time % = 50%, PSUPP = 2 cmH2O, O2% = 100% or 40% in NeoMode
(21% if O2 not available), patient circuit type = last set value or ADULT if
none available, humidification type = last set value or NON-HEATED EXP
TUBE if none available, humidifier volume = last set value or 480 mL if none
available.
Alarm settings in safety ventilation: High circuit pressure (2PPEAK) =
20 cmH2O, high exhaled minute volume = OFF, high exhaled tidal
volume = OFF, high respiratory rate = OFF, low exhaled mandatory tidal
volume = OFF, low exhaled minute volume = 0.05 L, low exhaled
spontaneous tidal volume = OFF
Setting limits for volume control (VC) Tidal volume: 25 mL VT 2500 mL; 1.16 mL/kg VT 45.7 mL/kg (default
mandatory breaths 7.25 mL/kg)
Inspiratory time: 0.2 s TI 8 s
Expiratory time: 0.2 s TE 59.8 s
I:E ratio: 1:299 I:E 1:4.00
Flow (at 1/min f 100/min):
3 L/min V 60 L/min for IBW 24 kg
150 L/min Vfor IBW > 24 kg
Minute volume (using square flow pattern, I:E = 1:1, and f 30/min):
30 L/min VE for IBW < 24 kg
30 L/min <VE < 75 L/min for IBW 24 to 54 kg
MaximumVE = 75 L/min for IBW 55 to 150 kg
TI is a function of VT, flow pattern, TPL, and VMAX
TE is a function of VT, flow pattern, TPL, and f
I:E is the result of TI and TE
Any combination of settings for VT, VMAX, TPL, f, and flow pattern that
violates these boundaries is rejected. Refer to the Technical Reference
section of the Puritan Bennett™ 800 Series Ventilator System Operator’s
and Technical Reference Manual for more details.
Setting limits for pressure control (PC) Inspiratory pressure: PI = 5 to 90 cmH2O; PI + PEEP 90 cmH2O;
mandatory breaths PI + PEEP + 2 cmH2O 1PMEAN
Inspiratory time: 0.2 s TI 8 s
Expiratory time: 0.2 s TE 59.8 s
I:E ratio: 1:299 I:E 1:4.00
Respiratory rate: 1/min f 100/min
High circuit pressure limit: 7 cmH2O PMEAN 100 cmH2O
TI is a function of f (for I:E or TE constant during rate change) and TE. TE is a
function of f (for I:E or TI constant during rate change) and TI. I:E is a
function of f (for TI or TE constant during rate change), TI, and TE.
Any combination of settings for PI, PEEP, 1PMEAN, f, TI, I:E, or TE that violates
these boundaries is rejected. Refer to the Technical Reference section of the
Puritan Bennett™ 800 Series Ventilator System Operator’s and Technical
Reference Manual for more details.
Setting limits when selected Support pressure: PSUPP = 0 to 70 cmH2O; PSUPP + PEEP 90 cmH2O
spontaneous type is pressure support PEEP: PEEP = 0 to 45 cmH2O; PEEP + 7 cmH2O 1PMEAN
(PS) High circuit pressure limit: PSUPP + PEEP + 2 cmH2O 1PMEAN
Any combination of settings for PSUPP, PEEP, or 1PMEAN that violates the
above boundaries is rejected. Refer to the Technical Reference section of
part of the Puritan Bennett™ 800 Series Ventilator System Operator’s and
Technical Reference Manual for more details.
Alarm settings
High exhaled minute volume limit Range: OFF or 0.05 to 100 L/min and > low exhaled minute volume limit and
(2V E TOT) Neonatal: 10 L/min
Pediatric: 30 L/min
Adult: 100 L/min
Resolution: 0.005 L/min for 0.05 to 0.495 L/min; 0.05 L/min for 0.5 to
4.95 L/min; 0.5 L/min for 5 to 100.0 L/min
New patient: ((20 x 7.25 x IBW x 1.30/1000) + 0.05) with NEONATAL patient
circuit
((14 x 7.25 x IBW x 1.30/1000) + 0.05) with PEDIATRIC patient
circuit
((10 x 7.25 x IBW x 1.30/1000) + 0.05) with ADULT patient
circuit
Resolution:
0.5 cmH2O for PEEP < 20 cmH2O
1.0 cmH2O for PEEP 20 cmH2O
NOTE:
The O2 sensor must be Enabled for the High/Low delivered O2%
alarm to function.
Manufacturer/model or Covidien
Description Where used
part number
Puritan Bennett™ 840 Software 4-075731 or Local Supplier Software Download only
Download Cable (BNC to BNC)
Puritan Bennett™ 840 VTS 4-075359-00 and current revision Performance verification
(Ventilator Test Software) Software Download CD-ROM
Manufacturer/model or Covidien
Description Where used
part number
Digital multimeter (DMM) accurate Fluke Model 87 or equivalent Performance verification, general
to 3 decimal places, with test leads troubleshooting
Double banana jack for ground Pomona Electronics, P/N 1330-2 Performance verification
isolation test
Electrical safety analyzer capable of Fluke Model ESA 620 or equivalent Performance verification
measuring ground resistance and http://www.flukebiomedical.com
leakage current 6920 Seaway Blvd.
Everett, WA USA 98203
Telephone: (425) 446-6945
Toll Free (US): (800) 850-4608
* These parts may be purchased individually or by ordering a Puritan Bennett™ 840 VTS Accessories Kit,
P/N 4-076599-00.
Manufacturer/model or Covidien
Description Where used
part number
* These parts may be purchased individually or by ordering a Puritan Bennett™ 840 VTS Accessories Kit,
P/N 4-076599-00.
Manufacturer/model or Covidien
Description Where used
part number
Pneumatic calibration analyzer or Puritan Bennett™ PTS 2000 Performance verification, regulator
equivalent devices capable of Performance Test System adjustment. Barometer used for
measuring oxygen percent, flow, (4-76185-00) (includes accessory kit atmospheric pressure transducer
BTPS volume, pressure, and with serial cable) calibration.
barometric pressure. Oxygen
analyzer connector tee. Required
accuracies: NOTE:
• Flow: 2.75% of reading If you use a device other than
±0.05 slpm the PTS 2000 Performance
Test System, refer to the
• Volume: 2% of reading or ±1 digit “Manual ventilator check”
• Low pressure (-150 to section in Chapter 5 of this
+150 cmH2O): 0.75% of reading manual.
±0.04 cmH2O
• High pressure (0 to 150 psig):
1.0% of reading ±0.1 psi
• Oxygen percentage: ±2% oxygen
• Barometric pressure: Range: 10 to
16 psia; resolution: 0.0 to 1 psia;
operating temperature: 10 to
40 C; measurement accuracy:
0.75 of reading;
response: 100 ms.
Serial Card, Socket I/O Ruggedized or Socket Communications Corp. Performance verification
equivalent (Provides second serial Newark, CA
port) http://www.socketcom.com/
* These parts may be purchased individually or by ordering a Puritan Bennett™ 840 VTS Accessories Kit,
P/N 4-076599-00.
Manufacturer/model or Covidien
Description Where used
part number
Stoppers, no. 1, 5, and 5.5 Local supplier SST, EST, performance verification
(with center bore and luer fitting:
4-071856-00)
* These parts may be purchased individually or by ordering a Puritan Bennett™ 840 VTS Accessories Kit,
P/N 4-076599-00.
Manufacturer/model or Covidien
Description Where used
part number
* These parts may be purchased individually or by ordering a Puritan Bennett™ 840 VTS Accessories Kit,
P/N 4-076599-00.
Table 1-11 lists the periodic maintenance activities required for the Puritan Bennett
840 Ventilator System. See the Ventilator Information screen for total hours of operation for
the ventilator and compressor. For details on patient system maintenance, refer to the
Puritan Bennett™ 800 Series Ventilator System Operator’s and Technical Reference Manual.
Several times a day or as required by Patient circuit: inspiratory and Check for water build-up, empty,
your institution’s policy expiratory limbs and clean as necessary.
Collector vial, water traps, and drain Check and empty as needed.
bag
Every 250 hours (or more often, if Compressor inlet filter Either vacuum filter or wash filter in
required) a warm detergent solution, rinse,
and dry well. Replace filter when it
shows signs of wear.
Varies: Every year or 100 autoclave Reusable expiratory or expiratory Replace. Sterilize between patients
cycles. Consult product Directions for bacteria filters and circuit changes, or according to
Use. your institution’s policy. Sterilize
before nondestructive disposal.
Every year or as necessary Oxygen sensor Replace. Actual sensor life depends
on operating environment;
operation at higher temperature or
O2% levels will result in shorter
sensor life.
2 years or as necessary 802 BPS battery pack Actual BPS life depends on the
803 BPS battery pack history of use and ambient
conditions.
NOTE:
If any part found in a
preventive maintenance kit
requires replacement before
the recommended interval
elapses, consider installing
the entire kit anyway.
10,000 hours 9.4“ GUI and BDU 4-079056-00 Filter, air inlet (F2) with o-ring
Filter, oxygen (F1)
Filter, oxygen inlet (F3) with o-ring
Installation instructions
Label, preventive maintenance
Lamps, fluorescent backlight
(for 9.4“color LCD displays)
Spring
10.4” GUI and BDU 4-078179-00 Filter, air inlet (F2) with o-ring
Filter, oxygen (F1)
Filter, oxygen inlet (F3) with o-ring
Installation instructions
Label, preventive maintenance
Spring
10,000 hours BPS (6 month shelf 10000007 (for use Battery pack
(continued) life) with 802 BPS)
10030284 (for use in
803 BPS)
4-070532-SP (for use
in BPS of Compressor
Mount Cart with
one-hour battery)
10030284 (for use in
Compressor Mount
Cart BPS with four-
hour battery)
4-070532-SP (for use
in Pole Cart with one-
hour battery)
10041274 (for use in
Pole Cart with four-
hour battery)
Index
Labeling Function
(Figure 1-3)
1 Screen lock key. When the yellow light on the screen lock key is
lit, touching the screen or off-screen controls (including the knob
and ACCEPT key) has no effect until you press the screen lock key
again. New alarms (or when an alarm’s urgency level escalates)
automatically unlock the screen and controls.
The screen lock allows you to clean the touch screen and prevents
inadvertent changes to settings and displays.
or
Index
Labeling Function
(Figure 1-3)
4 Alarm volume key. Allows you to adjust the alarm volume when
you hold down this key while turning the knob. You cannot turn
off alarm volume.
or
5 Alarm silence key. Turns off alarm sound for 2 minutes. The yellow
light on the alarm silence key lights during the silence period, and
turns off if you press the alarm reset key or the 2-minute interval
times out. A new, high-urgency alarm cancels the silence. Alarms
that can be silenced (lockable) include all Patient-Data Alarms
and Circuit Disconnect (see Table 7-2 for details).
Each time you press the alarm silence key, the silence period resets
to 2 minutes. Each time you press the alarm silence key (whether
or not there is an active alarm), the keypress is recorded in the
alarm log.
or
US version only
Index
Labeling Function
(Figure 1-3)
10 Allows you to measure auto-PEEP. Hold this key down until the
maneuver begins (at the next ventilator-initiated inspiration), and
release when the measurement is stable, up to 20 s. The measured
values for intrinsic and total PEEP are displayed at the end of the
expiratory pause. Displays and freezes the most recently selected
graphics, allowing you to see when expiratory pressure stabilizes.
EXP PAUSE is not functional in SPONT. It has no effect during the
inspiratory phase of a breath. The pause is canceled and
inspiration begins if you release the EXP PAUSE key, the patient
triggers an inspiration, an alarm occurs, or the maximum 20-s
pause interval elapses. If flow triggering is active, backup pressure
sensitivity is used to detect patient effort. Only one expiratory
pause is allowed during a breath. Expiratory pause requests are
ignored in apnea ventilation, safety ventilation, occlusion status
cycling (OSC), and idle mode.
During an expiratory pause, occlusion alarm detection is
suspended. During expiratory pause, the apnea interval is
extended by the amount of time the pause is active. In SIMV, the
cycle during which the pause becomes active (and the next
scheduled VIM will occur) is extended by the amount of time the
pause is active. For purposes of I:E ratio calculation, expiratory
pause is considered part of the exhalation phase.
Index
Labeling Function
(Figure 1-3)
15 System operation
indicators
non-US version
or
US version
Index
Labeling Function
(Figure 1-3)
non-US version
or
US version
non-US version
or
US version
Index
Labeling Function
(Figure 1-3)
15 Green BPS ready indicator. The ventilator senses that the BPS is
(cont) installed, operational, and has at least 2 minutes of estimated run
time.
or
BATTERY
READY
US version only
On BPS power indicator. When yellow bar to the right of a lit BPS
ready indicator (battery symbol) is lit, ventilator is operating on
BPS, and AC power is insufficient to support ventilator operation.
During BPS operation, power to the compressor unit and the
humidifier outlet (if available) is off.
or
BATTERY
ON
US version only
or
COMPRESSOR
READY
US version only
or
COMPRESSOR
ON
US version only
Index
Labeling Function
(Figure 1-3)
2 3
Index
Labeling Function
(Figure 1-4)
2 RS-232 10.4-inch GUI only: Two serial ports with 9-pin male connector
configured as data terminal equipment (DTE).
NOTE:
Allowable current is 0.2 A at 10 V DC (maximum).
2 3
5
6
1
Index
Labeling Function
(Figure 1-5)
2 Ventilator operation
indicators
non-US version
or
US version
Index
Labeling Function
(Figure 1-5)
2 Red safety valve open (SVO) indicator. (See Table 1-13, item 15.)
(cont)
non-US version
or
US version
non-US version
US version
Index
Labeling Function
(Figure 1-5)
non-US version
or
US version
non-US version
or
US version
7 (Collector vial drain port) Collector vial drain port. Use to attach drainage bag.
2
1
Index
Labeling Function
(Figure 1-6)
1 TEST TEST (service) button. Enables service mode. When you turn on
the ventilator and press this button after the first beep from the
BDU, the ventilator is placed into service mode (for example, to
run EST).
2 PTS 2000 Puritan Bennett™ PTS 2000 Performance Test System connection.
Caution
Do not remove the data key. The data key cover can only
be removed with a screwdriver. The data key enables
software options, and stores ventilator operational hours
or and the serial numbers for the BDU and GUI. The data key
is for use by a qualified service technician only.
Data Key
US version
Index
Labeling Function
(Figure 1-6)
or
Compressor
US version
or
Display (GUI)
US version
2 3
1
5
Index
Labeling Function
(Figure 1-7)
NOTE:
A humidifier connection is only available on 100 – 120 V
ventilators.
or
Compressor & humidifier
circuit breaker
US version
or
Ventilator circuit breaker
US version
or
AC input
US version
Index
Labeling Function
(Figure 1-7)
or
Compressor outlet:
5.6 A max
US version
Index
Labeling Function
(Figure 1-8)
1 (High-pressure air fitting) DISS male, DISS female, NIST, Air Liquide™*, or SIS fitting
2 (High-pressure oxygen DISS male, DISS female, NIST, Air Liquide™*, or SIS fitting
fitting)
Index
Labeling Function
(Figure 1-9)
1 (Warning label) Warns user of hazards associated with the operation of the
Puritan Bennett™ 840 Ventilator System and GUI
2 (Serial number label) Unique assigned number. Must be the same as the GUI serial
number stored on the data key.
1
Figure 1-11. 803 BPS controls and indicators
Figure 1-12. Puritan Bennett™ 800 Series Ventilator Compressor Mount Cart
charging status and battery indicator labels
1
2
Labeling Function
Index
(Figure 1-10)
or
US version
Labeling Function
Index
(Figure 1-11)
1 803 BPS charging status. Indicates the charging status of the 803
BPS. A yellow LED next to the partially full battery icon indicates
the battery is charging. A green LED next to the full battery icon
indicates that the battery is charged.
All countries
Index
(Figure 1-12)
All countries
All countries
All countries
Labeling Function
Index
(Figure 1-13)
All countries
All countries
Theory of Operation
Figure 2-1. L-R Puritan Bennett™ 840 Ventilator shown on RTA cart,
Puritan Bennett 800 Series Ventilator Compressor Mount Cart,
and Puritan Bennett 800 Series Ventilator Pole Cart
2.1.5 Cart
The optional carts are shown in Figure 2-7. The cart mounts system components and
accessories, including the compressor unit on the RTA and Puritan Bennett™ 800 Series
Ventilator Compressor Mount Carts. They also provide mobility for the ventilator. Brakes on
the front casters prevent the cart from rolling and turning.
8-01119
Figure 2-7. RTA cart, Puritan Bennett™ 800 Series Ventilator Compressor Mount Cart, and
Puritan Bennett 800 Series Ventilator Pole Cart
NeoMode, a software option, requires the use of the neonatal patient system. In addition to
the patient tubing and inspiratory filter, the neonatal patient system includes a disposable
expiratory filter, a collector vial, and a special mounting plate, as shown in Figure 2-9.
Graphic user
interface (GUI)
Exhalation Inspiratory
module: module:
Active exhalation valve PSOLs
Pressure transducer Safety valve
Flow sensor Oxygen sensor
Pressure transducers
Flow sensors
Air Oxygen
regulator regulator
Air
supply
Expiratory
filter Oxygen
supply
(Expiratory (Inspiratory
Collector limb) limb) Inspiratory
vial filter
Patient
Humidification
circuit
device
The GUI logs an event in the diagnostic log and declares a ventilator inoperative condition if:
• Any raw signal data from three BDU transmissions within 24 hours is corrupted.
• The GUI does not receive data from the BDU within the time required.
• The GUI determines that raw data is valid, but settings or alarm limits are not being
handled properly.
The ventilator uses flow or pressure triggering to recognize patient effort. When pressure
triggering (P-TRIG) is selected, the ventilator monitors pressure in the patient circuit. As the
patient draws gas from the circuit and the airway pressure drops by at least the value selected
for pressure sensitivity (VSENS), the ventilator triggers a breath.
When flow triggering (V-TRIG) is selected, the BDU maintains a constant flow of gas through
the patient circuit (called base flow) during the later part of exhalation. The ventilator
indirectly measures patient flow (assuming minimal leaks) by monitoring the difference
between the inspiratory and expiratory flow sensor measurements. If the patient is not
inspiring, any difference between the delivered and the exhaled flow is due to sensor
inaccuracy or leaks in the patient system. To compensate for leaks in the patient system, the
operator can increase the flow sensitivity, which ideally equals desired flow sensitivity + leak
flow.
As the patient inspires from the base flow, the ventilator measures less exhaled flow while
delivered flow remains constant. As the patient continues to inspire, the difference between
the two flows, measured by the inspiratory and expiratory transducers, increases.
The ventilator declares an inspiration when the flow inspired by the patient (that is, the
difference between the measured flows) is equal to or greater than the operator-selected
value for flow sensitivity (VSENS).
As a backup method of triggering inspiration, a pressure sensitivity of 2 cmH2O is also in
effect. This setting is the most sensitive setting that is still large enough to avoid
autotriggering, yet will trigger with acceptable patient effort.
Air and oxygen from cylinders, wall supplies, or compressor (air only) enter the ventilator
through hoses and fittings (the fittings are available in several versions). Once inside the
ventilator, air and oxygen are regulated to pressures appropriate for the ventilator, then
mixed according to the operator-selected O2%.
The ventilator delivers the mixed air and oxygen through the inspiratory module, and out to
the patient. The oxygen concentration of the delivered gas is monitored here, using a
galvanic oxygen sensor. The galvanic sensor generates a voltage proportional to the oxygen
concentration. The ventilator alarms if the monitored oxygen concentration is more than
seven percent above or below the O2% setting, or below 18%. A safety valve relieves patient
pressure if necessary (for example, if the patient circuit is kinked or occluded). The operator
selects the humidification type, which the ventilator uses to correct for gas temperature and
humidity.
Ventilator inspiratory pneumatics consist of two parallel circuits: one for oxygen and one for
air. The primary elements of the inspiratory pneumatics are two proportional solenoid valves
(PSOLs), which control the flow of gas delivered to the patient. Air and oxygen flow sensors,
along with pressure signals from the inspiratory and exhalation modules, provide feedback
that is used by the BDU CPU to control the PSOLs. As a result, the ventilator supplies mixed
breathing gas to the patient according to operator-set variables. The mixed air and oxygen
passes through the patient circuit external to the ventilator.
The patient circuit includes the components external to the ventilator that route gas between
the ventilator and the patient. These components include the inspiratory filter (which
protects against contamination between the patient and ventilator), a humidification device,
the inspiratory and expiratory limbs of the patient circuit (the tubing through which the gas
travels), a collector vial (which protects the expiratory system from bulk moisture in the
exhaled gas), and an expiratory filter (which limits the bacteria in the patient’s exhaled gas
from escaping to room air or contaminating the ventilator).
The ventilator actively controls the exhalation valve, which is accurately positioned by
software throughout inspiration and exhalation, and allows the ventilator to deliver
aggressive breaths while minimizing pressure overshoots, controlling PEEP, and relieving
excess pressures. The exhalation system monitors the gas leaving the patient circuit for
spirometry.
Throughout the respiratory cycle, pressure transducers monitor inspiratory, expiratory, and
atmospheric pressures. The temperature of the exhaled gas is heated to a temperature above
its dew point to prevent condensation in the exhalation compartment.
Power to operate the ventilator comes from AC mains (wall) power or the BPS. The integral
power supply is designed to protect against excessive voltages, temperatures, or current
draws. A power cord retainer prevents accidental disconnection. A power switch cover
protects against spills and accidental AC power-off.
The ventilator includes the BPS that supplies DC power to the ventilator if AC power is lost. A
fully charged BPS operating under nominal ambient conditions can power the ventilator for
at least 60 minutes (30 minutes on ventilators built prior to July 2007). The 803 BPS can power
the ventilator for at least four hours under the same conditions. The BPS recharges during
operation from AC power. The GUI and BPS indicate when the ventilator is operating on the
BPS. The BPS does not power the compressor unit or the humidifier power outlet.
In some situations, the ventilator declares a ventilator inoperative condition. A hardware
failure or critical software error that could compromise safe ventilation triggers this
condition. When this occurs, the ventilator inoperative indicator lights and the ventilator
enters the safety valve open (SVO) state. To correct a ventilator inoperative condition, the
ventilator must be turned off, then powered on again; at power on, a qualified service
technician must run extended self test (EST). The ventilator must pass EST before normal
ventilation can resume.
In the safety valve open (SVO) state, the safety valve allows the patient to breathe room air
unassisted. The ventilator remains in the SVO state until the condition that caused the SVO
state is corrected or, if the ventilator declared a ventilator inoperative condition, the power
on self test (POST) verifies that power levels to the ventilator are acceptable and that the
major electronics systems are functioning correctly.
If the ventilator enters the SVO state, the safety valve open indicator lights and a high-
urgency alarm sounds. The ventilator enters the SVO state if a hardware or software failure
occurs that could compromise safe ventilation, both air and oxygen supplies are lost, or an
occlusion is detected. In case of a malfunction that prevents software from opening the
safety valve, there is also an analog circuit that opens the safety valve if system pressure
exceeds 100 to 120 cmH2O.
Inspiratory module
Exhalation
module
Air Oxygen
Exhaled
supply supply
gas
Electronic
modules
Compressor (hardware Backup
module and software) power
source
Reference
Component Description
designator
Inspiratory module
Fitting, inlet Connects external oxygen and air sources to ventilator via hoses.
Fittings include diameter index system standard (DISS) (male or
--
female), noninterchangeable screw thread (NIST), Air Liquide™*,
and Sleeved Index System (SIS).
Manifold, flow sensor Holds bases of flow sensors Q1 and Q2, and houses pressure valves
TP1 and TP2 and pneumatic noise filters F6 and F7.
Manifold, PSOL/SV Houses proportional solenoid valves (PSOL1 AND PSOL2) and
--
safety valve (SV).
Orifice, inspiratory pressure Bypasses inspiratory check valve to relieve pressure when an
-- relief occlusion is present in exhalation circuit while safety valve is open.
Minimizes rebreathing of exhaled gas during SVO.
Check valve, air/compressor CV2 opens to admit external compressed air and CV4 closes to
isolate compressor unit. When CV4 opens to allow compressor-
CV2
supplied compressed air, CV2 closes to prevent compressed air
CV4
(compressor source) from venting out the air (external source)
inlet fitting.
Check valve, inspiratory Opens to supply inspiratory gas and restricts exhalation flow in
CV3
reverse direction.
F5 Filter, screen, oxygen/air Filters large debris from REG1 and REG2. These filters are part of
F4 impact the regulator assemblies (one in each regulator).
Filter, pneumatic noise, Conditions gas flow by eliminating swirling of gas induced by
F7
oxygen/air elbows and restrictions. These filters are part of the flow sensor
F6
manifold (two in each manifold).
Sensor, oxygen (percentage) Measures partial pressure of oxygen in inspired gas. Range is 21 to
OS
100% O2.
PS1 Pressure switch, oxygen/air Opens when pressure is less than 20.0 psig nominal.
PS2 Closes when pressure is greater than 31.5 psig nominal.
PSOL1 Proportional solenoid valve, 0 to 200 L/min BTPS output (intermittent) or 0 to 180 L/min BTPS
PSOL2 oxygen/air output (steady state).
Reference
Component Description
designator
REG1 Regulator, oxygen/air Reduces input supply pressure (35 to 100 psig, flow up to 200 L/min
REG2 BTPS) to output pressure (9 psig minimum to 12 psig maximum).
TP1 Pressure valve, oxygen/air Allows measurement of REG1 and REG2 output.
TP2
WT1 Water trap, air Houses air inlet filter (F2) and includes a manual drain.
Patient system
Collector vial, exhalation Collects water (up to 250 mL with the full line at 200 mL) resulting
ECV from condensation in patient circuit. Neonatal collector vial
collects up to 25 mL water (with the full line at 15 mL).
Filter, inspiratory (main flow) Filters matter greater than 0.3 μm (micron) (nominal) at 100 L/min
F8
flow.
Filter, expiratory Filters matter greater than 0.3 μm (micron) (nominal) at 100 L/min
F9
flow.
WT Trap, water Collects excessive water. Present only on certain patient circuits.
Mounting plate (neonatal Allows neonatal expiratory filter to be installed in place of adult
--
only) filter.
Exhalation module
Check valve, exhalation Opens during exhalation to let exhaled gas into exhalation system.
CV5
Prevents rebreathing when safety valve is open.
Reference
Component Description
designator
Air dryer Removes water vapor from the compressed air to lower the dew
Dryer
point below ambient temperature.
F10 Filter 0.3 micron filter that filters gas going to the ventilator.
F12 Filter, intake silencer Filters and silences the compressor intake.
Housing base Collects water emptied from the water trap where it evaporates
HB
using heat from the compressor motor and air flow from the fans.
HE Heat exchanger Cools the compressed air allowing water vapor to condense.
Water trap Collects condensate as the compressed air cools in the heat
WT2 exchanger. Automatically drains collected water to the housing
base where it evaporates.
Oxygen
inlet filter
(F3)
Check valve
assembly
(CV2/CV4) Water trap (WT1)
Air inlet
filter (F2)
The air side of the gas supply conditioning subsystem includes the following components:
• The air hose and inlet fitting connect an external high-pressure air source to the
ventilator. Inlet fittings are available in DISS male, DISS female, NIST male, Air Liquide™*,
and SIS male fittings.
• The air inlet filter (F2) protects the pneumatic systems from debris by trapping particles
down to 0.3 μm (micron).
• The coalescing air water trap (WT1) removes condensation from the high-pressure air.
This condensation can be caused by pressure drops between the wall outlet and water
trap. WT1 also houses F2. It has a screw-type drain valve.
• The air pressure switch (PS2) senses a loss of supply pressure. This 12 V, normally open
switch closes when gas pressure reaches 31.5 psig nominal and opens when gas pressure
reaches 20.0 psig nominal.
• The air and compressor check valves (CV2 and CV4), housed in the check valve manifold,
are high-pressure unidirectional valves that prevent ventilator backflow into the air
system. CV2 opens and CV4 closes when wall air pressure exceeds compressor air pressure
(or no compressor is installed). CV4 opens and CV2 closes when compressor air exceeds
wall air pressure (or wall air is not connected). When wall air pressure is equal to
compressor air pressure, both check valves are open.
• The air regulator (REG2) is a spring-loaded, venting line regulator that maintains 9 to
12 psig output throughout a 35 to 100 psig supply range, from less than 1 L/min to
200 L/min. Its nominal output is 10.5 psig at 2 L/min. REG2 vents inside the BDU.
The air impact screen filter (F4), housed in REG2, removes debris entering REG2.
• The air pressure valve (TP2) checks the oxygen regulator setting (REG2).
• The air pneumatic noise filter (F6), housed in the flow sensor manifold, conditions gas
flow by eliminating swirling of gas induced by elbows and restrictions.
Oxygen
inlet filter
(F3)
Oxygen
impact
filter
(F1)
Air regulator
(REG2)
Oxygen regulator
(REG1)
CV4 Seat
CV2 Seat Compressor
check valve
(CV4)
Oxygen seat
NOTE:
Use an external (optional) water trap if there is condensation in the piping system.
On the oxygen side, two filters protect the pneumatic system from debris. The first filter, the
oxygen impact filter (F1), traps particles down to 65 μm (microns) and the subsequent filter,
the oxygen inlet filter (F3), traps particles down to 0.3 μm (micron). The oxygen transfer tube
directs oxygen to the check valve manifold. There is no check valve for oxygen.
Normally-open pressure switches, PS1 and PS2, located between their respective O2 and air
inlet filters and the check valve manifold, monitor oxygen and air supply pressures. When gas
pressure reaches 31.5 psig nominal, the respective pressure switch closes, signaling the
inspiratory electronics PCB that the supply is adequate. When the oxygen or air pressure
drops to 20.0 psig nominal, the corresponding pressure switch opens, causing a NO AIR
SUPPLY or NO O2 SUPPLY alarm. (See Chapter 7 for alarm handling.) The patient is ventilated
with the remaining gas source. If both gas sources are lost, the ventilator alarms and opens
the safety valve, and the patient breathes air from the room, unassisted by the ventilator.
On both the oxygen and air sides, filtered gas passes through screen filters F5 and F4, which
are part of regulators REG1 and REG2. The regulators maintain a stable nominal output of
10.5 psig. Pneumatic noise filters F7 and F6, part of the flow sensor manifold, condition gas
flow by eliminating swirling of gas induced by elbows and restrictions.
Pressure valves TP1 and TP2 let you check the regulator settings. This is done as part of the
performance verification (see Chapter 5).
PSOL1
PSOL2
Q1
Q2
• Q1 and Q2 use a hot film (Figure 2-21) within each sensor to measure gas velocity (flow).
The hot film constitutes one leg of a bridge circuit. The bridge is supplied by a constant
current. With no flow, all legs of the bridge circuit have equal resistance. As flow increases
(velocity increases), hot film temperature (and resistance) decreases. Current through the
hot film leg increases, and the bridge circuit becomes unbalanced. Output voltage VOUT
increases proportionally to gas flow.
• The oxygen and air proportional solenoid valves (PSOL1 and PSOL2) are software-driven
and individually controlled to deliver the prescribed air/oxygen mixture to the patient.
Each proportional solenoid valve includes a linear motor that controls a poppet valve. The
poppet valves open in proportion to the applied current.
Oxygen and air flow sensors (Q1 and Q2) with ferrites installed
Inspiratory check
Absolute pressure valve (CV3) housing
transducer (PA)
Inspiratory
pressure
transducer
(PI)
Oxygen Safety
sensor valve (SV)
(OS)
Inspiratory
pressure
transducer
autozero
solenoid
(SOL1)
Leaf Seat
Check valve
housing
Oxygen
sensor
Oxygen
sensor
port
8-0112
PA
PI
Absolute and inspiratory pressure transducers (PA and PI) on inspiratory electronics PCB
1. Inlet fitting
2. Air inlet filter (F2)
3. Air water trap (WT1)
4. Air pressure switch (PS2)
5. Transfer tube
6. Check valve manifold
7. Air check valve (CV2)
8. Compressor check valve (CV4)
9. Air impact screen filter (F4)
10. Air regulator (REG2)
11. Flow sensor manifold
12. Air pressure valve (TP2)
13. Air pneumatic noise filter (F6)
Gas in. Wall air enters the ventilator at the air inlet fitting. The ventilator requires wall air
pressure maintained between 35 and 100 psig with intermittent flow up to 200 L/min. If wall
air falls below compressor pressure, an optional compressor unit supplies air to the ventilator
at the check valve manifold.
Filtration. The air inlet filter (F2) traps particles down to 0.3 μm (micron), protecting the
pneumatic system from debris. Coalescing-type water trap WT1 collects condensation.
Pressure monitoring. Wall air is monitored by PS2 (normally open). As air pressure increases
above 31.5 psig nominal, PS2 closes, indicating sufficient wall air pressure to operate the
ventilator.
Wall or compressor air. Externally supplied air continues through a transfer tube to the check
valve manifold, which contains the air check valve (CV2) and the compressor check valve
(CV4). Depending on wall air pressure, either wall air or compressor output is supplied. If wall
air pressure falls below 26 psig, compressor air is supplied to REG2.
Regulation. From the check valve manifold, air enters REG2 via the air impact screen filter
(F4). F4 traps particles. REG2 reduces air pressure from 35 – 100 psig to 9 – 12 psig (10.5 psig
nominal). It vents 2 L/min inside the BDU.
Turbulence smoothing. Low-pressure air enters the flow sensor manifold, which houses the
air pressure valve (TP2) and the air pneumatic noise filter (F6). TP2 is used to check REG2
settings. F6 conditions gas flow by eliminating swirling of gas induced by elbows and
restrictions.
Air and oxygen mixing. Air continues on to the air flow sensor (Q2), where air flow is
measured. Software calculates air flow at Q2 and target flow. Software-controlled current is
applied to PSOL2, opening it to allow a predetermined air flow through PSOL2. The PSOL/SV
manifold houses proportional solenoid valves PSOL1 and PSOL2, and the safety valve (SV).
Mixed gas is delivered to the patient circuit through the inspiratory outlet.
Safety valve. Should the operating pressure of the patient circuit exceed 100 cmH2O
(measured at PE), the safety valve opens. The outlet manifold houses the oxygen sensor (OS),
the inspiratory check valve (CV3), and the pressure relief orifice. CV3 prevents rebreathing
from the room by ensuring that exhalation occurs only through the exhalation valve.
Inspiration monitoring. The oxygen concentration of the delivered gas is measured by the
oxygen sensor (OS). The ventilator alarms if the monitored oxygen concentration is not
within ±7% of the oxygen percentage setting. Pressure measurements are made by two
differential pressure transducers.
The inspiratory pressure transducer autozero solenoid (SOL1) is energized periodically to set
a zero reference or offset voltage for the inspiration pressure transducer.
Turbulence smoothing. Low-pressure oxygen enters the flow sensor manifold, which houses
the oxygen pressure valve (TP1) and the oxygen pneumatic noise filter (F7). TP1 checks REG1
settings. F7 conditions gas flow by eliminating swirling of gas induced by elbows and
restrictions.
Oxygen and air mixing. Oxygen continues on to the oxygen flow sensor (Q1), where oxygen
flow is measured. Software calculates oxygen flow at Q1 and target flow. Software-
controlled current is applied to PSOL1, opening it and allowing a predetermined oxygen flow
through PSOL1. The PSOL/SV manifold houses proportional solenoid valves PSOL1 and PSOL2
and safety valve SV. Mixed gas is delivered to the patient circuit through the inspiratory
outlet.
Safety valve. If the operating pressure of the patient circuit exceeds 100 cmH2O (measured at
PE), SV opens. The outlet manifold houses the galvanic oxygen sensor (OS), the inspiratory
check valve (CV3), and the pressure relief orifice. CV3 prevents rebreathing from the room by
directing exhaled gas to the exhalation valve.
Inspiration monitoring. The oxygen concentration of the delivered gas is measured by OS.
The ventilator alarms if the monitored oxygen concentration is not within 7 percentage
points of the O2 percentage setting. Pressure measurements are made by two solid-state
differential pressure transducers.
SOL1 is energized periodically to set a zero reference or offset voltage for the inspiration
pressure transducer.
Inspiratory
filter
(F8)
Humidification
device
Water trap
(2 PL)
Wye
Figure 2-29. Patient system (minus exhalation collector vial and expiratory filter)
Caution
To protect the exhalation assembly, always use a Puritan Bennett™ expiratory filter or
a recommended DAR™ expiratory filter.
inspiration gas is delivered to the patient through the wye. Exhaled gas flows through the
wye, to the ECV, and through F9.
NOTE:
Exhalation module parts are shown here for reference only. For assembly illustration,
refer to Figure 9-19 on page 9-53.
Figure 2-34. 806 compressor on Puritan Bennett 800 Series Ventilator Compressor Mount Cart
Pressure
relief valve
To heat exchanger
Silencer Motor/compressor
Cooling fans
0.3 micron
filter
Filter foam
Water
trap
Filter
element
To accumulator
To pressure
Unloading transducer (PC)
solenoid (SOL3)
To solenoid Filter (F10)
From heat
exchanger
Air dryer Silencer
Accumulator
Air dryer
assembly
Motor/
Silencer
compressor
During the start-up sequence, a solenoid valve (SOL3) opens 1 second prior to the time the
compressor motor starts, and remains open for 0.5 seconds afterwards, reducing the load on
the compressor. A transducer on the PCBA measures the accumulator pressure, and the signal
is used to control the compressor operation. If pressure in the accumulator drops below
13 psig, the compressor shuts down and the ventilator delivers 100% O2 to the patient.
In stand-by mode, the compressor shuts off when the pressure reaches 27 psig in the
accumulator, and turns on when the pressure drops to 22.5 psig. (This pressure drop can occur
because of small leaks in the compressor system or because of cooling of compressed air in
the accumulator.)
The compressor PCBA contains an hour meter that records the elapsed run time of the
compressor. This information is stored in the EEPROM on the PCBA, as is the compressor’s
serial number. If a replacement PCBA is required, the EEPROM from the existing PCBA can be
removed and installed on the new board, preserving the compressor elapsed time and serial
number.
For more information regarding the 806 compressor operation, see Section 2.4.12 and
Figure 2-84.
NOTE:
The 10.4-inch GUI features color LCD panels, updated GUI CPU, touch frame, and
backlight inverter PCBs, and three serial communications ports. There is a single serial
communications port on the older 9.4-inch monochrome GUI (no longer available)
and the 9.4-inch color GUI color LCD panels, and one serial communications port.
Both 10.4-inch and 9.4-inch GUIs (except monochrome) accept the Trending-enabled
(TE) GUI CPU PCB. This PCB contains a compact flash memory component used by the
Puritan Bennett 840 Ventilator System Trending software option.
For a discussion of electropneumatic parts, or electrical parts used in close conjunction with
pneumatic parts, see Section 2.3. For wiring details, see Figure 2-42, Figure 2-43, Figure 2-44,
Figure 2-45 Figure 2-47, Figure 2-50, Figure 2-56, Figure 2-68, Figure 2-78, Figure 2-80, and
Figure 2-86. Table 2-2 describes the electronic parts.
Reference
Component Description
Designator
CB1 Main circuit breaker Provides over-current protection to the power supply.
CB2 Auxiliary circuit breaker Provides over-current protection to the humidifier and compressor.
J3 Humidifier receptacle Located at the front of the ventilator, receives AC power from
power relay K1 and provides AC power to the humidifier.
J4 Compressor receptacle Receives AC power from power relay K1 and provides AC power to
the compressor via the compressor AC power cord.
-- Back-up power source (BPS) With new, fully charged batteries, provides at least 60 minutes
(30 minutes on ventilators built prior to July 2007) of back-up
power in the event of an AC line failure or temporary disconnect.
The 803 BPS provides at least four hours of back-up power under
the same conditions. Output signals from the BPS PCB include
charging*, charged*, BPS model, IBATT, VBATT, and E-BP.
NOTE:
When the ventilator is running on battery power, the
compressor and humidifier are non-operational.
Reference
Component Description
Designator
-- Motherboard PCB Provides the main electrical interconnect between the breath
delivery CPU PCB, analog interface PCB, breath delivery LED PCB,
exhalation module, inspiratory module, power supply, power
switch, and BD alarm.
-- Breath delivery (BD) CPU PCB Contains the electronics and software that control all breath
delivery functions in the Puritan Bennett™ 840 Ventilator System.
Communicates with the GUI CPU to respond to operator inputs
and display ventilation parameters.
-- Analog interface (AI) PCB Provides the interface for all analog signals in the ventilator. The
AI and BD CPU PCBs together provide the main intelligence and
drive for the mechanical devices and electronic sensors used in
ventilation.
-- BD LED PCB Contains visual alarm indicators and LED drivers to annunciate
safety valve open, loss of GUI, and ventilator inoperative
conditions. All indicators are powered by +12V.
S1 Power switch Switches power to the power relay (K1), and power supply,
enabling its secondary outputs used by ventilator electronics.
-- Data key Required for proper ventilator operation, the data key stores BDU
and GUI serial numbers, hours of ventilator operation, and enables
ventilator software options.
-- Exhalation transducer PCB Produces a voltage signal proportional to pressure in the patient
circuit exhalation limb using the exhalation pressure transducer
(PE). The PCB uses a +10V reference.
PE Exhalation transducer Measures the pressure difference in the patient circuit exhalation
limb relative to atmospheric air pressure.
EXH HTR Exhalation heater Heats the patient’s exhaled gas, preventing condensation in the
exhalation module. The 16 W exhalation heater uses +12V.
Q3 Exhalation flow sensor Measures exhalation flow for use in determining net flow to the
patient and for spirometry. Flow sensor drivers use +12V and +5V.
Thermistor supply voltage is +12V.
SOL2 Exhalation pressure This +12V, 3-way solenoid periodically exposes the exhalation
transducer autozero solenoid pressure transducer to atmospheric pressure for setting a zero-
reference offset voltage.
Reference
Component Description
Designator
-- Inspiratory electronics PCB Produces voltage signals proportional to the pressure in the
patient circuit inspiratory limb and atmospheric pressure using the
inspiratory pressure transducer (PI) and absolute pressure
transducer (PA), respectively. Also provides an interface for the
oxygen sensor. The PCB uses a +10V reference.
PA Absolute pressure transducer Measures atmospheric pressure for use in volume delivery
calculations.
PI Inspiratory pressure Measures pressure in the patient circuit inspiratory limb relative to
transducer atmospheric pressure.
SV Safety valve This +12V valve opens when de-energized (during POST, at
pressures > 100 cm H2O, during ventilator inoperative conditions,
or when both source gases are lost). Energized closed all other
times.
PSOL2 Air PSOL A +12V proportional solenoid valve. Regulates the flow of air in
the inspiratory circuit by adjusting flow proportionally to the
supplied current which is under feed-back control from the Q2
flow sensor signal and the inspiratory pressure transducers.
PSOL1 Oxygen PSOL A +12V proportional solenoid valve. Regulates the flow of oxygen
in the inspiratory circuit by adjusting flow proportionally to the
supplied current which is under feed-back control from the Q1
flow sensor signal and the inspiratory pressure transducers.
SOL1 Inspiratory pressure This +12V, 3-way solenoid periodically exposes the inspiratory
transducer autozero solenoid pressure transducer to atmospheric pressure for setting a zero-
reference offset voltage.
Q1 Oxygen flow sensor Measures inspiratory oxygen flow. Measurements are used to
control PSOL1. Flow sensor drivers use +12V and +5V. Thermistor
supply voltage is +12V.
Q2 Air flow sensor Measures inspiratory air flow. Measurements are used to control
PSOL2. Flow sensor drivers use +12V and +5V. Thermistor supply
voltage is +12V.
PS1 Oxygen pressure switch Senses inlet oxygen pressure and opens if pressure drops below
20.0 psig nominal. Closes when pressure is greater than 31.5 psig
nominal.
PS2 Air pressure switch Senses inlet air pressure and opens if pressure drops below 20.0
psig nominal. Closes when pressure is greater than 31.5 psig
nominal.
-- Oxygen sensor A galvanic sensor that measures the oxygen percentage of the
delivered gas at the patient wye. Delivers a voltage proportional
to the oxygen concentration. Ventilator alarms if the monitored
oxygen concentration is not within 7% of the oxygen
percentage setting.
Reference
Component Description
Designator
-- Graphical user interface (GUI) Provides microprocessor control and monitoring of the display and
CPU PCB user interface system. Communicates with the BD CPU PCB via an
Ethernet connection.
Backlight inverter PCB Converts +12V input into high-voltage AC that powers the
fluorescent lamps that backlight the LCD display panels. Pulse
width modulation from the GUI CPU board determines the
magnitude of the AC voltage which determines the brightness of
the displays. Two backlight inverter PCBs individually control the
upper and lower LCD panels on the 10.4-inch GUI. On the 9.4-inch
GUI, one PCB performs the backlight functions for both displays.
Touch frame PCB Contains the electronics and firmware to detect inputs to the GUI
touchscreen by means of locating the logical X/Y coordinate
position of an operator's touch.
LCD displays Two variable voltage, 9.4-inch (older version) or 10.4-inch (current
version) color LCD flat panels display ventilator settings and
patient information.
GUI LED PCB Contains LED drivers and light bars that illuminate visual alarm
and status indicators on the GUI LED panel.
RS-232 connectors Three (10.4-inch GUI) or one (9.4-inch GUI) RS-232 connectors
provide for serial communications with external devices such as
printers or ventilator monitors.
Remote alarm connector Enables/disables an external alarm device (such as a nurse’s call).
Relay contacts are closed during normal ventilation, and open
during an alarm condition or when unit is off.
VGA LCD controllers Plug-in modules (on some versions of the GUI CPU board) that
provide a video interface between the GUI CPU PCB and the LCD
flat panels. The VGA LCD controllers are integrated into the
current version of the GUI CPU PCB.
GUI alarm assembly The ventilator’s primary alarm, emits multiple frequency alarm
sounds under control of the GUI CPU PCB.
-- Compressor PCB Controls the functions of the compressor motor, cooling fans, and
unloading solenoid.
SOL3 Unloading solenoid A +12V, 3-way solenoid controlled by logic signals that reduces
back pressure on the compressor during start-up and periods of
low demand. Also vents excess accumulator pressure to assist in
transporting water vapor from the air dryer.
Reference
Component Description
Designator
M/C Motor compressor AC powered and logic-controlled, supplies compressed air to the
ventilator when adequate wall air pressure is unavailable.
Contains internal over-temperature protection device.
Figure 2-42. Electrical system block diagram (10.4-inch GUI—Trending-enabled CPU PCB shown)
Figure 2-43. Electrical system block diagram (10.4-inch GUI—Non-Trending-enabled CPU PCB shown)
Figure 2-44. Electrical system block diagram (9.4-inch GUI—Trending-enabled CPU PCB shown)
Figure 2-45. Electrical system block diagram (older 9.4-inch GUI shown)
2.4.3.2 AC panel
The AC panel (Figure 2-46 and Figure 2-47), a single field-replaceable unit (FRU), includes
components that apply AC to the ventilator head, compressor unit, and optional humidifier.
The AC panel includes AC receptacle J1, circuit breakers CB1 and CB2, AC filter PCB, power
relay K1, compressor receptacle J4, and potential equalization connector J2. A humidifier
receptacle (J3), which is connected to but is not a part of the AC panel, is on the front of the
ventilator.
• The AC receptacle (J1) receives facility AC power via the power cord.
Warning
Ensure that the power cord retainer bracket is properly installed and secures the power
cord to the AC receptacle (J1).
• The AC panel houses two push-to-reset type circuit breakers. The 5 A main circuit breaker
(CB1) limits current to the power supply. The 10 A (100 – 120 V) or 5 A (220 – 240 V)
auxiliary circuit breaker (CB2) limits current to the compressor and humidifier circuits.
• The AC filter PCB reduces AC line noise to and from the ventilator.
• The power relay (K1) switches power to humidifier receptacle J3 and compressor
receptacle J4 when the power switch (S1) is on. The relay is powered from a dedicated
+12 V power supply output, which is active when the power switch is on.
• The compressor receptacle (J4) receives AC power via K1 and provides the AC power
outlet for the compressor power cord.
• The humidifier receptacle (J3) receives AC power via K1 and provides an AC power outlet
for the humidifier power cord.
• The potential equalization connector (J2) interfaces the equipment and the potential
equalization bus bar.
AC receptacle
Main
(J1)
circuit
breaker
(CB1)
Auxiliary Potential equalization
(compressor/ connector (J2)
humidifier)
circuit breaker
(CB2) Compressor
receptacle (J4)
Power
relay
(K1)
Fuse
Auxiliary (compressor/
humidifier circuit breaker (CB2)
Auxiliary mains
circuit breaker
AC power
receptacle
EXTERIOR OF
AC PANEL
Potential
Compressor/humidifier equalization
circuit breaker point
Compressor
receptacle
GRN/YEL
BRN
J1
C1 BLU
2 4
L N
INTERIOR OF
G
AC PANEL BLU
CB1
E2 BRN
E1
J2
1 3
C2
GRN/YEL
K1 2 4
E4 E3 BRN
CB2
J4
L N 1 3
L G
BRN BLU
GRN/YEL
BRN BRN
BRN
BLU BLU
BLU
1 3 BLU
4 6 BLU BRN
7 BRN
9
A B
LEGEND:
CB1 Power supply circuit breaker
CB2 Auxiliary mains circuit breaker
BRN
J1 Mains inlet BRN
BLU To power supply
J2 Potential equalization point compartment
RED BLK
J4 Compressor outlet
K1 Power relay
AC PANEL MODULE
Figure 2-47. Puritan Bennett™ 840 Ventilator System interconnect diagram – AC panel
Power LED
switch Indicator
(S1)
BD alarm
MOTHERBOARD PCB
Inspiratory
blindmate DC blindmate Exhalation I/O
cable assy cable assy cable assy
Comms
outlet/hmd
cable assy
AC
indicator
POWER SUPPLY
To backup
power supply
(BPS)
Figure 2-50. Puritan Bennett™ 840 Ventilator System interconnect diagram – Power distribution
2.4.5 BPS
The 802 BPS, shown in Figure 2-51, provides a DC power source during a brief AC line failure
or temporary disconnect. A fully charged BPS supplies at least 60 minutes (30 minutes on
ventilators built prior to July 2007) of power backup to the ventilator. A new, fully charged
803 BPS (Figure 2-52) provides at least four hours of back-up power to the ventilator. The BPS
does not power the optional compressor or humidifier; these operate from AC line power.
NOTE:
It is assumed that the ventilator is powered from an outlet connected to the hospital’s
emergency generator AC backup system.
When the ventilator power switch (S1) is on, the BPS provides a +19 to +30 V supply voltage
to the power supply. The ventilator software monitors this voltage to determine whether the
battery is present and the battery status. The software triggers a medium-urgency alarm if
the battery capacity falls below 2 minutes runtime.
The BPS electronics charge the battery pack as necessary whenever AC is present and
sufficient (> 80 V), even if the power switch is off. An indicator displays the batteries’
charging status: yellow means the batteries are charging, and green means the batteries are
approximately 75% charged relative to nominal. Good batteries can be charged within 8
hours. The BPS has overcurrent (overcharging) protection.
NOTE:
Battery packs, BPS assemblies, or units with BPS’s that are unused or in storage for
over 6 months, or have been depleted by continuous usage require a minimum of 8
hours to recharge the battery pack. Because of the larger battery capacity, the 803
BPS may take up to 20 hours to recharge.
The BPS contains a battery pack (Figure 2-53) (which includes two +12 V, lead-acid batteries
and a 15 A fuse) and a PCB (Figure 2-54).
Fuse
Motherboard PCB
BD alarm Power
switch (SI)
To AC
panel
BDU LED
display PCB Inspiratory Exhalation I/O
blindmate cable assy
DC blindmate
cable assy
cable assy
J9 J10
J7 J8
J6
J4
J2
P2 P3
CPU PCB
BREATH DELIVERY
UNIT (BDU)
TEST
switch
(SW2)
P1
LED
indicator
GUI-to-BDU
cable assy
P9 P10
AI PCB
PTS 2000
J1
connection J6
P5 9-pin
Data BDU/compressor
key DC cable
Figure 2-56. Puritan Bennett™ 840 Ventilator System interconnect diagram – Card cage
AI PCB connectors
BD CPU PCB
connectors
Motherboard PCB
P1 J7
BD +5 V, BD +12 V
J8
J11
J9
PWR SW POSITIONS
FROM PWR SW
TO ALARM
J13
DATA, ADDRESS,
STATUS
+15 V, -15 V
CTRL, STATUS
TO/FROM
AI PCB
TO/FROM EXH MOD PWR & SENS
EXHALATION
MODULE
BD ALARM
+12 V EXH AZ
J12
TEST button
GUI port
• The real-time clock tracks the ventilator operational time and short-duration power losses
(up to 60 minutes). When the ventilator is operating, the real-time clock is powered by
VCC. When power is lost or removed, a built-in battery powers the clock. This battery has a
10-year life expectancy.
• The watchdog circuit monitors the BD CPU for safe operation. When a time-out occurs,
indicating a lack of bus activity, a system reset occurs. Problems with the CPU, clock, or
software can cause a watchdog time-out.
• The GUI-BDU interface provides for Ethernet communication between both BD and GUI
CPU PCBs. This controller sends an interrupt to the BD CPU to inform it of a
communication from the GUI CPU. Commands and messages to be transferred (via direct
memory access, DMA) are stored in DRAM. The GUI-BDU interface uses isolation
transformers and an isolated power supply, but is connected to standard system ground.
There is no isolation from any external device that may be attached.
• An RS-232C channel is available for developmental use only.
• A diagnostic LED array (with supporting circuitry) indicates the status of the BD CPU PCB.
During POST, they indicate the current test step. A ninth LED displays the supervisory
mode status of the CPU.
• The power fail detection circuit causes an NMI to be generated if the power supply
generates a power fail signal. The power supply generates a power fail signal if the
output of the bulk supply is less than 19.30 V.
• The digital I/O interface circuit provides these outputs:
– A signal (redundant) to declare a ventilator inoperative condition and to light an LED.
– A signal to place the BDU hardware into a safety valve open (SVO) state and to light an
LED.
– BD LED PCB control signals, including loss of user interface (LOUI) LED, safety valve
open (SVO) LED, and ventilator inoperative (VENT INOP) LED.
– Compressor control signals (representing on, standby, and off). The on command runs
the compressor continuously when the ventilator is operating on compressed air (that
is, when the wall air supply has dropped below 14 psig). Standby causes the
compressor to cycle to maintain accumulator pressure; wall air pressure drops to
between 14 and 26 psig, and the compressor supplements the wall air pressure. The
off command overrides the compressor automatic cycling control (to maintain
accumulator pressure) to ensure the compressor is off for self test purposes.
– Power mode signal suspends battery charging to allow a true reading of battery
voltage.
– Low AC power signal (AC < 80% of nominal) to compressor.
– Signals to place the ventilator into the service mode (in conjunction with a momentary
push-button switch) and to disable the two POST 10-s timers (necessary to allow test
mode).
The digital I/O interface provides these inputs:
– Compressor status signals, including: compressor present; compressor compartment
overtemperature; AC input to compressor low; compressor accumulator pressure
adequate for ventilation; compressor in standby mode, compressor off, or compressor
on.
– BPS status signals including: battery being charged, battery in use, and battery fully
charged.
– Service mode enabled
• The AI PCB interface provides for communication between the BD CPU and the AI PCB.
• The BDU LED display drivers interface the BDU LED PCB.
• The service mode switch is a push-button on the board edge used to activate this mode.
The BD CPU PCB generates an NMI under any of these conditions:
• Ethernet parity error detected
• Power fail signal
• A/D converter system error
Background alarm Status of all DEVICE ALERT alarms BD alarm statuses are stored
status in BD NOVRAM. GUI alarm
statuses are stored in GUI
NOVRAM.
Diagnostic logs This includes the System Diagnostic Log, System Information BD events are stored in BD
Log, and EST/SST Diagnostic Log (which are actually stored in NOVRAM. GUI events are
two separate locations -- one for EST, one for SST). stored in GUI NOVRAM.
Alarm history log History of alarm events for the current/previous patient GUI NOVRAM
NOTE:
The alarm history log is automatically cleared upon
successful completion of a new patient setup. (A new
patient setup is required to start normal ventilation
after exiting the service mode.)
EST and SST result Result and status information for each EST and SST test BD information is stored in
information BD NOVRAM. GUI
information is stored in GUI
NOVRAM.
POST test status Status of each of the POST/kernel tests BD statuses are stored in BD
NOVRAM. GUI statuses are
stored in GUI NOVRAM.
PTS 2000
connector
Data key
connector
Compressor
connector
AI LEDs
• The BD alarm circuit activates the BD alarm. The alarm is activated under any of these
conditions:
– A ventilator inoperative condition is declared
– The BD CPU detects a loss of communication with the GUI
– A GUI alarm fault is detected
– The power switch fails
– The +5 V sentry on the AI PCB detects a low voltage condition
Under most conditions, the +5 V powers the alarm. If the +5 V sentry circuit on the AI PCB
detects a power failure, a power fail capacitor powers the alarm for a minimum of 2 min.
• The PSOL power disconnect circuit removes power from the PSOL drivers when the
ventilator is in the safety valve open state.
• The exhalation valve circuit drives the exhalation valve motor. Pressure readings provide
feedback to the software that determines the amount of valve dampening required. The
circuit includes power disconnect circuitry, which removes power from the exhalation
valve driver when the ventilator is in the safety valve open state.
• The safety valve circuit controls and drives the safety valve. Full power to the safety valve
pulls the valve closed. After a short interval, power is reduced to keep the valve closed.
The safety valve circuit ensures full power is applied in the event of a loss of +12 V
(required to reopen the valve).
• The voltage sentry circuit creates signals that are proportional to voltages used in the
system. These signals are used in software monitoring of these voltages: +10 V reference
and +5, +12, +15, and -15 V supply voltages.
In addition, the sentry circuit generates a global reset if +5 V is out of range. A relay,
which is part of the sentry circuit, ensures that a power fail reset occurs during voltage
transients.
• The BPS model signal buffer provides a signal that yields BPS model type information in
an analog form. The software reads this signal to determine whether the BPS is currently
connected to the ventilator.
• Other buffers provide conditioning for various ventilator signals.
• The compressor interface circuit, in conjunction with the compressor PCB, controls and
monitors the compressor operation.
NOTE:
Call your Covidien representative if the data key requires replacement due to loss or
failure.
The Puritan Bennett™ 840 Ventilator System uses a data key (Figure 2-62) to record data
specific to a particular ventilator unit. The data key provides a way to retain data when PCBs
or the battery are removed from the ventilator. The data key plugs into a receptacle on the
ventilator. Data is read from or written to flash memory on the data key. This serial data is
transferred to the BD CPU through the data key interface. The data key must always be
installed in the ventilator for proper operation.
The data key stores this data:
• Serial numbers of the GUI and BDU.
• Hours of ventilator operation.
2.4.8 GUI
The GUI (Figure 2-67 through Figure 2-69), which detaches from the ventilator head for
servicing, is an enclosure that houses the GUI CPU PCB, keyboard, GUI LED PCB, backlight
inverter PCBs, knob, two 10.4-inch or 9.4-inch color LCD flat panels, a touch frame PCB, and
an alarm speaker.
compact flash
component
GUI CPU
PCB
Backlight
inverter PCB
Figure 2-65. 9.4” Non-TE GUI CPU PCB and backlight inverter PCB in place
compact flash
component
The GUI CPU PCB includes these circuits, which perform the indicated functions:
• The CPU and control circuit includes a Motorola 68040 microprocessor (which includes an
onboard math coprocessor), operating at 24 MHz, along with supporting devices to
perform ventilator control functions. In addition to executing instructions, these functions
include passing data back and forth between memory and I/O devices; generating address
signals used to access memory locations; generating read/write, timing, and other control
signals; and processing interrupts from the various system devices.
• The PCB’s memory includes 4 MB of flash memory (soldered; no sockets) that contains the
operational software for the GUI. A boot PROM contains ventilator initialization and
POST code and provides a basic communications program to permit downloading of
system software. Downloaded software is not written to the boot memory PROM; the
PROM cannot be written to in the field. 4 MB of DRAM is used for ongoing calculations
and data storage and as a message and command buffer for the Ethernet controller.
• 64 KB of NOVRAM (nonvolatile RAM) stores essential, persistent variables and
configuration information; current breath mode settings (for obtaining default
parameters at power on); POST and EST fault information; the reset stack pointer; and
others (see Table 2-3). The NOVRAM devices are soldered in place (no sockets). Although
information in these devices is lost when the GUI CPU PCB is replaced, much of this data is
updated during POST and EST. Critical data specific to a particular unit (including system
serial number) is stored on a data key device, which stays with the unit.
Each NOVRAM device consists of an internal flash memory device (which has the
advantages of being able to retain data without the application of power, and which also
can be easily written to) and a static RAM device. During power on, data stored in the
flash memory is retrieved and stored in static RAM. During normal operation, the
NOVRAM behaves like static RAM. Upon detecting low supply voltage, the NOVRAM
stores data from static RAM into internal flash memory for nonvolatile storage. This
operation takes 10 ms. The GUI CPU PCB provides an isolated, decoupled power supply
(VCAP), independent of the +5 V supply, to the NOVRAM to ensure the power-down ramp
is long enough to store the static RAM contents.
• The compact flash component, available only on the Trending-enabled GUI CPU PCB,
provides additional memory required for use with the Puritan Bennett™ 840 Ventilator
System Trending software option.
• The real-time clock tracks the ventilator’s operational time and short-duration power
losses (up to 60 minutes). When the ventilator is operating, the real-time clock is powered
by VCC. When power is lost or removed, a built-in battery powers the clock. This battery
has a 10-year life expectancy. The real-time clock can be removed from its socket for field
replacement.
• +5 and +12 V monitors signal the microprocessor (via a nonmaskable interrupt) when +5 V
is too high or +12 V is out of range. This circuit also generates a power fail reset if the +5 V
output drops below +4.8 V.
• The watchdog circuit monitors the CPU for safe operation. When a time-out occurs,
indicating a lack of bus activity, a system reset occurs. Problems with the CPU, clock, or
software can cause a watchdog time-out.
• The GUI-BDU controller provides for Ethernet communication between both BD and GUI
CPU PCBs. The controller sends an interrupt to the GUI CPU to inform it of a
communication from the BD CPU. Commands and messages to be transferred (via DMA)
are stored in DRAM. The GUI-BDU interface uses isolation transformers and an isolated
power supply, but is connected to standard system ground. There is no isolation from any
external device that may be attached to it.
• Three RS-232 C channels provide output for digital communications interface and
external communications for service mode. They are electrically isolated for safety. The
9.4-inch GUI has only one RS-232 channel, also electrically isolated.
Backlight PCB
Backlight cable Upper LCD
panel
4-078267-SP
Flex circuit,
video
Backlight PCB
Lower LCD
panel
GUI LED PCB 4-078267-SP
P1
J3
J5
J7
J11
Touch frame PCB
GUI-to-BDU cable assy J10
Keyboard
J2 cable assy
Rear of keyboard
Touch frame
cable assy
GUI MODULE
• Diagnostic LED array uses eight discrete LEDs, in conjunction with supporting circuitry, to
indicate the status of the GUI. During POST, the array indicates the current test step. A
ninth LED displays the supervisory mode status of the CPU, and a 10th LED indicates
power on.
• The VGA video control circuit, comprising two VGA LCD controller ICs with 1 MB video
RAM, interfaces the two 640 x 480 x 256K color 10.4-inch LCD panels. Older GUI versions
with 9.4-inch, 4K color or monochrome (16 greyscale levels) displays containing the older
GUI CPU board contain plug-in IP interface VGA LCD controllers that interface the LCD
panels.
• An infrared touch scanner interface, which includes its own microcontroller, monitors the
touch screen.
• The LED PCB interface lets the microprocessor control the discrete LEDs on the GUI LED
PCB.
• The LCD backlight control circuit controls the brightness of the two fluorescent lamps that
light the flat panels based on software input. It generates a pulse width modulated signal
to the backlight inverter PCB(s) (two PCBs on 10.4-inch GUIs; one PCB on 9.4-inch GUIs),
which controls the brightness of both LCD panels. There is no user-adjustable brightness
control on color displays.
• The remote alarm relay interface enables/disables an external alarm device (such as a
nurse’s call). The relay contacts are closed when the unit is in a normal ventilation state
and opened when the unit is off or in an alarm state.
• The sound generator interface produces the ventilator’s alarm sounds. Consisting of a
microcontroller and audio signal processing hardware, the interface produces ISO-
standard alarm sounds and provides keyboard entry audible feedback.
• The rotary encoder/decoder circuit determines the direction and amount of knob
movement. The optical encoder incorporates an emitter section, two codewheels, and a
detector section. Each codewheel has a pattern photographically plated on it. As the knob
shaft revolves, the codewheels rotate with respect to the emitter and photodetector
sections, causing the light beam to be interrupted by the pattern of spaces and bars on
the codewheels. The detectors are positioned such that a light period on one
photodetector corresponds to a dark period on the other photodetector. Using the
photodetector outputs, the decoder can determine the knob position.
• The keyboard interface reads the keyswitches and controls the lighting of the LEDs on the
keyboard.
The GUI CPU PCB generates an NMI under any of these conditions:
• GUI +5 V is high
• GUI +12 V is out of range
• Ethernet parity error detected
• SAAS microcontroller failure
Backlight
control
cable assy
Cable assy
Color display
GUI cable assy
alarm assy J6
GUI CPU PCB
P1
J4 J5
J3 J8
J7 Color display
cable assy
J9
Color display
J16 cable assy
Touch Frame PCB
J18Color display
To BDU J19 J20
cable assy
RS-232
Remote alarm
Keyboard
cable assy
J5
GUI MODULE
Figure 2-68. Puritan Bennett™ 840 Ventilator System interconnect diagram – GUI 9.4-inch LCD panels
(Non-TE GUI CPU PCB P/N 4-073200-SP shown)
Backlight PCB J3
GUI cable assy, long Upper LCD
J1 panel
Backlight
control
cable assy
Lower LCD
GUI LED PCB panel
Cable assy
Color display
alarm assy
cable assy
J4
GUI CPU PCB
P1
J3
J5 J6
Color display
J7 cable assy
Color display
J12
J15Color display
J19 J20
To BDU cable assy
Keyboard
J5
Touch Frame
Rear of keyboard
cable assy
GUI MODULE
Photosensitive
diodes
Light-emitting
diodes
A micro-controller located on the touch frame PCB scans the touchscreen for operator inputs
by switching on the LEDs one at a time in a pseudo random pattern, and then sampling the
output results from the logically associated IR detectors. A complete scan of all the LEDs
occurs 39 times per second. When the GUI screen is touched, some of the light beams don’t
reach their associated IR detectors, and as a result, there is no output from those detectors
and they are considered "blocked." The microcontroller checks pairs of LEDs and IR detectors
for blocked beams, uses a mathematical coordinate conversion routine to locate the touch in
the intersecting area of light beams, and sends the information to the GUI CPU board for
processing.
To improve touch frame reliability, each LED output is received by two IR detectors. The
blocked beams are detected simultaneously in different intersecting areas providing
redundancy which allows continued operation of the touch frame in the event an LED or IR
detector burns out. During normal conditions, when all LEDs and IR detectors are functioning
properly, the redundant detectors are used to provide a more accurate output by averaging
the locations of the blocked beams.
On the 9.4-inch GUI, the touch frame communicates with the GUI CPU PCB via a
CarrollTouch™* proprietary interface and operates from +5V and +12V supplies. Two sides of
the touch frame PCB (Figure 2-70) have rows of infrared (IR) LEDs, while their opposite sides
have corresponding rows of phototransistors. Whenever the ventilator is on, the IR touch
scanner interface on the GUI CPU PCB sequentially pulses the LEDs, creating an invisible grid
of IR light beams just in front of the display surface. The phototransistors sense these light
beams.
When the screen is touched, the light beams from certain LEDs are obstructed and do not
reach their corresponding phototransistors. The interface constantly monitors the presence,
or in this case the absence, of the IR light beams received by the phototransistors and can
thus determine the X- and Y-coordinates of the touch activation.
On both the 10.4-inch and 9.4-inch GUIs, the touch frame is attached to the face of the
display, where it is concealed behind an IR-transparent bezel (a type of enclosure that is
opaque to the eye, but that allows IR light to pass through).
The multiposition knob assembly, which is part of the keyboard assembly, permits ventilator
setting selections or changes. Knob encoder/decoder circuitry on the GUI CPU PCB determines
the direction and position of the shaft based on encoder outputs.
LCD panels
Tube
element
PA
PI
The PCB includes these circuits, which perform the indicated functions:
• The inspiratory pressure transducer (PI) senses the inspiratory pressure difference relative
to ambient air pressure. Pressure is sensed at the inspiratory module.
• The absolute pressure transducer (PA) senses absolute pressure in the inspiratory module.
It is used to determine atmospheric pressure for volume delivery.
• The oxygen sensor (OS) amplifier provides an interface for the oxygen concentration
sensor mounted on the PCB.
Oxygen
sensor
4-072214-00
Inspiratory
pressure transducer/
autozero solenoid
(SOL1)
4-072215-SP
Blindmate connector
P2 44 pin
Air flow
sensor (Q2)
Oxygen flow
4-072211-SP
sensor (Q1)
4-072211-SP
Oxygen flow
Air flow sensor
sensor cable assy
cable assy 4-071322-00
4-071321-00
Press switch
harness assy
4-071326-00
INSPIRATORY MODULE
Figure 2-78. Puritan Bennett™ 840 Ventilator System interconnect diagram – Inspiratory module
Expiratory
pressure transducer/
autozero solenoid (SOL2)
Exhalation
heater assembly
Exhalation
flow sensor (Q3)
To
motherboard
Exhalation
module Exhalation I/O
cable assy cable assy
P1 J2
J13
Expiratory
Exhalation pressure transducer
valve (PE)
Exhalation
transducer PCB
EXHALATION MODULE
Figure 2-80. Puritan Bennett™ 840 Ventilator System interconnect diagram – Exhalation module
BD alarm
assembly
Data
cable
Pressure
transducer Power
hose cord
Solenoid
Compressor cable
cable
Fan
cables
Ground
The PCB includes these circuits, which perform the indicated functions:
• The flash memory circuits include a flash memory device, an elapsed time counter, and
circuits to permit serial data transfer between the AI PCB and the compressor PCB. The
counter monitors (in seconds) compressor motor operation. The counter value is
periodically read by the ventilator, which then stores the total elapsed time (in hours) in
compressor flash memory. During a power-on reset, the ventilator loads the current
compressor flash memory value into the counter. The flash memory device also contains
other compressor-specific information, including the compressor identification (ID).
• The motor start-up circuit includes a relay, timers, and logic that together orchestrate the
motor start-up sequence. This sequence involves energizing SOL3, which vents compressor
output, and then energizing the relay to apply power to the compressor. (Venting
compressor output momentarily reduces the load on the compressor.)
• The compressor pressure transducer (PC) circuit monitors accumulator pressure via several
comparators in the circuit. When a comparator detects an accumulator pressure equal to
or greater than predefined trip points, it signals the compressor control logic.
• The compressor control logic controls the motor, fans, and SOL3. It also provides
compressor status signals. Section 2.4.12.3 describes how this logic controls compressor
component operation. This logic shuts off the compressor when AC is inadequate or the
thermostat on the PCB detects overtemperature.
COMPRESSOR PCB
J1
12 V COMPS
UNLOADING
SOLENOID SOLENOID
CKT (SOL3)
J2
FAN1
J6
FROM
AC PANEL J3
FAN2
J4
COMP COMPRESSOR
PRESS PRESS RELAY
XDUCER STARTING
(PC) CKTS CAPACITOR
J5
MOTOR
COMP
COMPRESSOR
CONTROL
HOUSING
CKTS
TO/FROM
AI PANEL
EEPROM
CKTS
+5 V COMP
THERMOSTAT
+12 V COMP
Figure 2-86. Puritan Bennett™ 840 Ventilator System interconnect diagram – Compressor unit
27 psig
Accumulator
pressure
22.5 psig
Compressor
motor
27 psig
Accumulator
pressure
22.5 psig
SOL3
Compressor
motor
Breath
interval
SOL3
(Dump 1 3
compressor
head
pressure)
1.0 s 0.5 s
Motor
relay
NOTE:
• Inspiratory and expiratory flows and tidal volumes in the ventilator are
compliance-compensated and corrected to body temperature and pressure,
saturated (BTPS).
• Exhaled volumes delivered/measured by the ventilator meet specified
accuracies when conventional humidification, heated wire systems, and heat-
moisture exchangers (HMEs) are used and SST is successfully completed.
2.5.1 Inspiration
During inspiration (Figure 2-89), the exhalation valve (EV) is energized closed and gas flows
to the patient.
An inspiration is triggered when any of the following happens:
• The ventilator senses patient inspiratory effort (pressure or flow triggering).
• The ventilator’s breath timing dictates.
• The operator presses the MANUAL INSP key.
• The ventilator’s software otherwise dictates (such as apnea time-out).
NOTE:
The ventilator does not autotrigger when pressure sensitivity is greater than 2 cmH2O
or when flow sensitivity is greater than 1 L/min for pediatric patients or 1.5 L/min for
adult patients.
2.5.2 Exhalation
During patient exhalation (Figure 2-90), the exhalation valve (EV) opens (maintaining PEEP)
and gas expelled from the patient is heated, filtered, and vented to the atmosphere.
The ventilator declares exhalation based on internal triggers or backup exhalation limits.
The ventilator software can trigger exhalation (internally triggered exhalation) using:
• The time-cycling method (monitoring elapsed time)
• The end-inspiratory flow method (monitoring end-inspiratory flow)
• The airway pressure method (monitoring airway pressure at the end of inspiration)
Backup limits (time, circuit pressure, and ventilator pressure) prevent inspirations of excessive
duration or pressure. If a particular breath is subject to more than one backup limit,
exhalation is triggered by whichever method goes into effect first.
Breath type Indicates the type and phase of the Type: Control, assist, or
breath being delivered. spontaneous
Phase: Inspiration or
exhalation
Delivered O2% (O2%) Percentage of oxygen in the gas Oxygen sensor (OS) 0 to 103%
delivered to the patient. The high
and low O2% alarms are set
internally and are based on the set
O2% value.
End expiratory Pressure at end of previous breath’s Based on expiratory -20.0 to 100 cmH2O
pressure (PE END) expiratory phase. Reflects level of and inspiratory
any active lung PEEP. pressure transducer (PE
and PI) readings
End inspiratory Pressure at end of current breath’s Based on expiratory -20.0 to 130 cmH2O
pressure (PI END) inspiratory phase. If plateau is active, and inspiratory
it reflects level of end-plateau pressure transducer (PE
pressure. and PI) readings
Exhaled minute Calculated total of exhaled volumes Net flow, based on 0.00 to 99.9 L
volume (V. E TOT) for mandatory and spontaneous exhalation (Q3),
breaths for previous 1-minute oxygen (Q1), and air
interval. It is compliance- and BTPS- (Q2) flow sensor
compensated. readings
Exhaled tidal volume Exhaled volume for previous Net flow, based on 0 to 6000 mL
(VTE) mandatory or spontaneous breath. It exhalation (Q3),
is compliance- and BTPS- oxygen (Q1), and air
compensated. (Q2) flow sensor
readings
NOTE:
A significant change to the O2% setting can cause the VTE (exhaled tidal volume) to
be transiently displayed as lower or higher than the actual exhaled volume. This is a
result of initial spirometry calculations and does not reflect actual volume exhaled
by the patient.
Mean airway pressure Average airway pressure over entire Based on expiratory -20.0 to 120 cmH2O
(PMEAN) breath cycle of previous breath, and inspiratory
regardless of type. pressure transducer (PE
and PI) readings
Maximum circuit Indicates the maximum pressure Based on expiratory -20.0 to 130 cmH2O
pressure during the previous breath, relative and inspiratory
(PPEAK) to the patient wye, including the pressure transducer (PE
inspiratory and expiratory phases. and PI) readings
Updated at the beginning of the
next inspiration.
Spontaneous minute Displays a calculated total of the Net flow, based on 0.00 to 99.9 L
volume (VE SPONT) volumes exhaled by the patient for exhalation (Q3),
spontaneous breaths for the oxygen (Q1), and air
previous 1-minute interval. Values (Q2) flow sensor
for mandatory breaths during this readings
period are not included. The
displayed value is compliance- and
BTPS-compensated. Updated at the
beginning of the next inspiration.
AC input < 80 V AC
level
31.50 V DC 24 V DC
Bulk
supply
24 V DC nominal 19.30 V DC
BPS or external
battery output
NMI
(generated by
power fail)
802 BPS: 60 minutes minimum (30 minutes on ventilators built prior to July 2007)
803 BPS (Four-hour battery life with new fully charged battery)
GUI CPU PCB If voltage is low, this circuit issues a PCB reset. If
voltage is high, an NMI is generated.
+10 V reference AI PCB AI PCB Converted into a scaled voltage for reading by
software during background and other checks.
NOTE:
During a severe occlusion, apnea detection, expiratory pause, manual inspirations,
and maneuvers are suspended, and the PMEAN (high airway pressure) alarm limit is
disabled. The GUI does allow you to change ventilator settings.
Self Tests
3.1 Introduction
This chapter describes the Puritan Bennett™ 840 Ventilator System built-in self
tests, including how to run them. For a listing of ventilator diagnostic codes and
messages generated during these self tests, refer to Chapter 6.
checks the exhalation filter resistance. SST is a user-initiated check primarily intended for use
by the operator. The ventilator does not begin SST if it senses that a patient is connected.
NOTE:
EST must always be performed with an ADULT exhalation filter installed. Attempting
to run EST with a NeoMode filter can cause some EST tests to fail.
SST A short sequence of tests intended to be run • Before patient is connected to ventilator
primarily by the operator. Checks pneumatics • After patient circuit or humidifier is
and electronics. SST also characterizes system changed
leaks and system/tubing compliance to • When ventilator is serviced
compensate during breath delivery. POST is • Every 15 days
run as part of SST, both at the start and end.
NOTE: NOTE:
• SST is not run as part of EST. To EST must always be performed with an
determine patient circuit adult-sized exhalation filter installed.
resistance and compliance, run Attempting to run EST with a
SST. NeoMode filter can cause some EST
• The performance verification, tests to fail.
described in Chapter 5 of this
manual, is a more thorough test
of the ventilator to verify
specifications are met.
Single Test EST One or more EST tests can be run in any order. • A singe Test EST test is run at the service
A prompt will appear on the GUI if special technician’s discretion, for troubleshooting
equipment is required. purposes. When all single Test EST tests are
completed, the technician must run EST
with passing results before a patient may
be ventilated.
Ongoing Includes these checks: memory, power supply, Automatically and continually during
background checks DC voltages, pressure transducers, analog ventilation
input validity, and more. These checks do not
disrupt normal operation.
1. POST will run only if there is no BPS installed, or if the BPS battery is discharged. Otherwise, during a temporary loss of AC power,
the vent will automatically switch to BPS power, and POST will not run.
AC monitor voltage X
BPS model X X
Patient system
Pneumatics
System leak X
Pneumatics (continued)
Electronics
ADC/DAC functionality X X
Electronics (continued)
Keyboard X partially X
Knob X
Ethernet controller X X
Safety system
NOTE:
A fault identified in POST indicates that the ventilator or an associated component is
defective. A defective ventilator or associated component should be repaired before
the ventilator is returned to service, unless it can be determined with certainty that
the defect cannot create a hazard for the patient, or add to the risks which may arise
from other hazards.
There are two versions of POST: long POST, which is run when the user cycles power, and short
POST, which is run following an unintended hardware reset.
NOTE:
The graphic user interface (GUI) indicates a POST is in progress but does not indicate
a long or short POST.
The graphic user interface (GUI) and the breath delivery unit (BDU) subsystems each has its
own POST that tests the major hardware electronics systems. POST does not check the
ventilator’s pneumatics, options, or accessories that are not directly related to ventilation.
POST is designed to detect major problems before proceeding to normal ventilation, and to
provide a confidence check before a patient is connected to the ventilator.
POST routines are ordered so that each routine requires successively more operational
hardware than the last. This sequence allows POST to systematically exclude electronic
components as causes of system malfunctions.
3.4.1 Safety
The ventilator does not provide ventilatory support to the patient during POST. The
ventilator alarms if POST lasts longer than 10 seconds or if an unexpected fault is detected.
POST is designed to minimize the delay until normal ventilation begins and to provide
immediate notification in case a fault is detected.
During POST the ventilator proportional solenoid valves (PSOLs) are closed and the
exhalation valve and safety valve are open to allow the patient to breathe room air, and the
ventilator displays a message that POST is in progress.
Once POST is complete, ventilator startup (following power-up) or normal ventilation begins,
unless service mode is requested or the ventilator detects any of the following:
• An uncorrected major system fault.
• An uncorrected major POST fault.
• An uncorrected short self test (SST) failure or non-overridden SST alert.
• An uncorrected extended self test (EST) failure or non-overridden EST alert.
• The ventilator is turned on for the first time following a software download, but has not
yet successfully completed one of the following: exhalation valve calibration, flow sensor
calibration, atmospheric pressure transducer calibration, SST, or EST.
• An uncompleted system initialization.
NOTE:
When a compressor is installed and wall air is not present, there may be a short
interval following a successful POST before the compressor achieves operational
pressures. If so, the ventilator annunciates a NO AIR SUPPLY alarm, which resets as
soon as the compressor charges the system to operational pressure.
To ensure that there is an alarm if the central processing unit (CPU) fails, audio, visual, and
remote alarms are normally on, and turn off once system initialization (that is, the process
that occurs between POST completion and the start of ventilation) is completed and
communication is established.
An alarm turns on if POST lasts more than 10 seconds or if POST restarts three times without
completion. The 10-second timer is a redundant check in case POST fails to alarm upon
detecting a fault. The check for three restarts can detect a continuous loop, and prevents
POST from running for more than 10 seconds.
NOTE:
Covidien recommends that a BPS is always installed on the ventilator. If there is a loss
of AC power with a charged BPS installed, the ventilator will switch to the DC source
and will not run POST upon restoration of AC power.
GUI CPU POST • Initializes processor • Tests GUI on-board • Tests GUI CPU • Updates NOVRAM
• Tests power, GUI electronics, GUI 5V memory • Initializes
CPU, time of day sentry, GUI 12V management hardware
clock, kernel sentry, POST and floating
DRAM/NOVRAM, NOVRAM, bus point units,
boot PROM, timer, DRAM, download OS
interrupt timer, SAAS, Ethernet boot,
watchdog timer, controller, application boot
flash memory unexpected reset
checksum, EEPROM umpire test
checksum
No communication occurs between GUI and • GUI and BDU • Ventilator begins
BDU CPU boards during Phases 1 and 2 CPUs establish flowing air at
communication 10 L/min to check
for patient
connection
BDU CPU POST • Initializes processor • Tests BDU on- • Tests BDU CPU • Initializes
• Tests power, BDU board electronics, memory hardware
CPU, time of day VH 5V, VH 12V, management
clock, kernel POST NOVRAM, and floating
DRAM/NOVRAM, bus timer, DRAM, point units,
boot PROM, vent INOP, download OS
interrupt timer, Ethernet controller, boot,
watchdog timer, service mode application boot
flash memory switch, unexpected
checksum, EEPROM reset umpire test,
checksum, rolling • AI PCB test
thunder test • Analog devices
test:
PS1/PS2
Q1/Q2/Q3
PSOL1/PSOL2
PA/PI/PE
O2 sensor
safety valve
exhalation valve
exh. valve heater
• ADC and DAC tests
• AC voltage test
Visual and • GUI LEDs: SVO LED • GUI LEDs: all LEDs • GUI LEDs: all OFF • GUI: start-up
audible indicators ON ON except green screen displays if
• BDU LEDs: SVO LED • BDU LEDs: all LEDs “normal” POST passes
ON, loss of UI LED ON indicator • BDU: all LEDs off if
ON • BDU continuous • BDU LEDs: all POST passes
• BDU continuous tone alarm OFF OFF • No audible
tone alarm ON • GUI audible alarms/LEDs
NOTE: Prior to the alarm turned on active1
beginning of Phase 1, then off (2
all of the LEDs on the beeps if OK)
Trending-enabled GUI • BDU continuous
turn on for 200 to tone alarm
turned on and
350 ms.
off twice
AI PCB LEDs Red LED ON • Red LED ON Green LED ON Green LED flashing
(checks BD CPU
PCBA)
• Red & Green LED
ON (checks AI
PCBA)
• Red & Green LED
ON (checks 32
Analog Devices)
• Green LED ON
(checks
communication
between BD CPU
PCBA and AI PCBA)
1. This is true only if both air and O2 supplies are connected and have adequate pressure, and the patient circuit is not connected, and
no failures or fault conditions are detected by ongoing diagnostics (including POST 10 second timer).
Symptoms
How to respond
What you see and hear What is happening
• POST continues to end. • Previous run of POST failed • Run EST/service ventilator.
• Ventilator inoperative and was not corrected,
alarm (audible and visible). although POST passed
• Ventilation disallowed. subsequent run.
• Ventilator inoperative
condition declared.
• POST continues to end. • Minor fault detected. • Check System Diagnostic Log
• Ventilation proceeds. for any associated error
• A DEVICE ALERT is codes.
annunciated. • Rerun POST/EST.
NOTE:
• If the ventilator has not reached operating temperature from recent usage, allow
it to warm up for at least 10 minutes in service mode before running SST, to ensure
accurate testing.
• Covidien recommends that you run SST every 15 days, between patients, after a
major service or repair (refer to Table 5-2 on page 5-6), and when you change the
patient circuit. Covidien recognizes that the protocol for running SST varies widely
among health care institutions. It is not possible for Covidien to specify or require
specific practices that will meet all needs, or to be responsible for the effectiveness
of those practices.
Table 3-6 lists the tests that comprise SST. For more details about these tests and associated
diagnostic codes, see Chapter 6 of this manual. For a theory of operation of SST, consult the
Puritan Bennett™ 800 Ventilator System Operator’s and Technical Reference Manual.
NOTE:
To ensure that compliance compensation
functions correctly, the user must run SST
with the circuit configured as intended
for use on the patient.
NOTE:
Due to the excessive restriction of Air
Liquide™*, Dräger™*, and SIS air/oxygen
hose assemblies, certain SST tests may fail
when supply pressures< 50 psi (345 kPa)
are used in conjunction with these hoses.
NOTE:
Due to the excessive restriction of Air
Liquide™*, Dräger™*, and SIS air/oxygen
hose assemblies, certain SST tests may fail
when supply pressures< 50 psi (345 kPa)
are used in conjunction with these hoses.
1. If necessary, the compressor can be used as the only gas source. However, only the gas pathway, flow sensor, and
PSOL of the single gas (air or O2) will be verified during SST.
NOTE:
To ensure accurate circuit resistance measurement, check that the circuit is not
obstructed and is properly connected to the ventilator.
2. Ensure that patient is not connected to circuit and that patient wye is unblocked.
3. Turn on power switch to start normal ventilation.
4. Verify that the ventilator is operating on full AC power before running SST. If full AC
power is not present, test failures may result.
5. At Ventilator Startup screen (lower GUI screen), touch SST, then press TEST button (on
side of ventilator) within 5 seconds. Waiting longer than 5 seconds cancels the SST
prompt.
Tubing configured
for use on patient
6. At SST Setup screen (lower GUI screen), select patient circuit and humidification types,
then press ACCEPT.
Warning
Incorrectly specifying the patient circuit type or humidifier during SST setup, or
changing either type after you have run SST can affect the accuracy of compliance
calculation and the delivered and measured exhaled tidal volumes. You must rerun SST
after changing the patient circuit type, installing a new patient circuit, or after adding
or changing accessories.
NOTE:
During ventilation, the humidification type can be changed by touching the OTHER
SCREENS button, then the More Settings button. However, the preferred operation in
this case is to rerun SST after making such a change.
Warning
• To ensure ventilation that correctly compensates for circuit resistance and
compliance, do not exit SST and begin normal ventilation until the entire SST has
been successfully completed with the circuit to be used on the patient installed.
• To ensure reliable SST results, do not repeat an individual test with a different
patient circuit if the test result is FAILURE or ALERT. If you suspect a defective
patient circuit, restart SST from the beginning with a different patient circuit.
NOTE:
• If SST is interrupted, you can restart normal ventilation as long as no errors have
occurred that would otherwise prevent it.
• Tests that result in an alert or failure and are successfully rerun are not written into
the EST/SST Diagnostic Log if they are successfully rerun during the same SST test
sequence.
10. When all of the tests in SST are complete, the SST Status screen displays all individual test
results and SST outcome. Table 3-8 summarizes overall SST outcomes and how to proceed
in each case.
11. To begin normal ventilation (if SST has not detected an ALERT or FAILURE), touch EXIT
SST, then press ACCEPT. The ventilator reruns POST, then displays the Ventilator Startup
screen.
SST Flow Sensor Test 1. Verifies that at least one gas is con- When prompted, do the following:
nected. (If only one gas is connected • Make sure inspiratory filter is installed.
then only that gas pathway (flow • Connect patient circuit with inspiratory
sensor and PSOL) is verified during filter, but without the humidifier.
SST.) • Block wye.
2. Cross-checks air and oxygen flow • When later prompted, and If
sensors against exhalation flow sen- applicable, install the humidifier with
sor at four different gas flows. jar filled with water.
3. Verifies that proportional solenoid
(PSOL) command current is in range
for each test flow. NOTE:
To ensure proper compliance
compensation when using a
humidifier, make sure the jar is full
of water.
Circuit Pressure Test 1. Verifies that system is running on FAILURE if not passed (cannot be
AC. overridden).
2. Exercises air and oxygen PSOLs.
3. Checks inspiration and exhalation
autozero solenoids.
4. Cross-checks inspiration and exhala-
tion pressure transducers at various
pressures.
Circuit leak Displays drop in circuit pressure in 10 s. Overriding an ALERT could cause
(Determines ability of circuit to hold improper compliance compensation,
pressure.) inaccurate tidal volume delivery, or
autotriggering.
FAILURE reported if test detects excessive
leak.
Expiratory filter Displays pressure drop across expiratory Connect and disconnect tubing when
filter. prompted.
Overriding an ALERT could cause
inaccurate patient pressure estimation.
FAILURE if test detects exhalation
compartment occlusion, expiratory filter
occlusion or damage, or you did not
follow prompts to detach and reattach
tubing correctly.
Circuit Resistance Displays pressure drop across inspiratory Unblock wye when prompted.
and expiratory limbs, including effect of Overriding an ALERT could cause
all devices on each limb (filters, inaccurate patient pressure estimation.
humidifier, water traps). FAILURE if test detects excessive or low
limb resistance, or if you did not follow
the prompt to unblock the wye.
Compliance calibration Displays patient circuit compliance. Block and unblock wye when prompted.
Overriding an ALERT could cause
improper compliance compensation or
inaccurate tidal volume delivery. FAILURE
if test detects out of range compliance.
ALERT Test results not ideal, but not Repeat test by touching REPEAT.
critical. SST halts. Skip to next test by touching NEXT.
Repeat SST from the beginning by touching RESTART SST.
Exit SST in order to service ventilator by touching EXIT SST.
FAILURE A critical problem has been Repeat failed test by touching REPEAT.
detected, and SST cannot Repeat SST from the beginning by touching RESTART SST.
complete until the ventilator Exit SST in order to service ventilator by touching EXIT SST.
passes the failed test.
If the SST
It means: Do this:
outcome is:
ALERT One or more faults were Repeat SST from the beginning by touching RESTART SST.
detected. If it can be determined Override ALERT, as allowed by your institution’s protocol, by
with certainty that this cannot touching OVERRIDE, followed by EXIT SST.
create a hazard for the patient,
or add to the risks which may
arise from other hazards, the
user can choose to override the
ALERT status and authorize
ventilation.
FAILURE One or more critical faults were Repeat SST from the beginning by pressing RESTART SST.
detected. A ventilator Exit SST in order to service ventilator or review error codes by
inoperative condition is declared. pressing EXIT SST.
NOTE:
• If the ventilator has not reached operating temperature from recent usage, allow
it to warm up for at least 10 minutes in service mode before running EST or single
Test EST to ensure accurate testing.
• Be aware that each time EST is performed, the patient alarm log is cleared.
Manufacturer or model or
Description
Covidien part number
Compressed air source, 241 to 690 kPa (35 to 100 Local supplier
psi) or compressor option
NOTE:
Due to excessive restriction of Air
Liquide™*, Dräger™*, and SIS air/oxygen
hose assemblies, certain EST tests may fail
when supply pressures< 50 psi (345 kPa)
are used in conjunction with these hoses.
Oxygen source, 241 to 690 kPa (35 to 100 psi) Local supplier
Adult
expiratory
filter
Gold
standard
tube
4. Touch the ALL TESTS button, then press ACCEPT to confirm.The ventilator automatically
starts the test sequence (Table 3-10). Some tests require your intervention, and will wait
indefinitely for your response. Others, such as the GUI Knob test and the GUI Keyboard
test, require your response within a 15-second timeout period. Failure to respond before
the timeout period expires will result in test failure or alert.
Otherwise, you don’t need to do anything until a test result is ALERT or FAILURE, or EST
is complete.
5. As each test is performed, the displays resemble those in Figure 3-3, with test data
displayed on the upper GUI. Overall test results are shown in the “Result” column (see
Table 3-11). Error codes beginning with an A indicate an alert and error codes beginning
with an F indicate a failure. You can review resulting error information through the
EST/SST Diagnostic Log after EST is completed (see Chapter 6).
6. You can touch EXIT EST during EST to halt testing. You can touch EXIT EST again to
resume testing, or press ACCEPT to restart the ventilator (if EST has not detected an
ALERT or FAILURE).
NOTE:
• If EST is interrupted, you can restart normal ventilation as long as no errors have
occurred that would otherwise prevent it. You must press the EXIT button on the
GUI screen then the ACCEPT key to exit EST properly.
• Tests that result in an alert or failure and are successfully rerun during the same
test sequence are not written into the EST/SST Diagnostic Log.
• It may be useful to complete EST even with errors, because information on
multiple errors can facilitate troubleshooting.
7. When all of the tests in EST are complete, the Extended Self Test screen displays all
individual test results and EST outcome. Refer to Tables Table 3-10, Table 3-11, and
Table 3-12 for additional information.
NOTE:
The screens given above are intended as examples only. Your unit may vary slightly in
appearance.
Circuit Pressure Test Checks that the system is running on AC power, • Leaks or occlusions in the
prompts user to set up the vent for EST, and exercises expiratory filter or elsewhere
the PSOLs. will cause this test to fail.
Calibrates inspiratory and exhalation transducers to • Connect AC power if ventilator
atmospheric pressure. is running on battery power and
AC is not connected.
Cross-checks inspiration and exhalation pressure
• Make sure green BPS ready
transducers at various pressures. indicator is lit on GUI LED panel.
Checks inspiration and exhalation autozero • Make sure air and oxygen are
solenoids. connected.
• Remove inspiratory filter and
connect gold standard circuit as
directed.
Flow Sensors Cross Cross-checks air and oxygen flow sensors against • Leaks or occlusions in the
Check Test exhalation flow sensor at five different gas flows. expiratory filter or elsewhere
Verifies that proportional solenoid (PSOL) command will cause this test to fail.
current is in range for each test flow. Performs air/O2 • Low source gas pressures may
PSOL lift-off current calibration. also cause this test to fail.
• Exhalation flow sensor has
separate air and O2 calibration
tables. During EST, the system
uses the calibration table(s)
appropriate for the source
gas(es) used during EST.
Gas Supply/SV Test Tests air and oxygen pressure switch gross • Leaks or occlusions will cause
functionality, tests PSOL forward leak, tests safety this test to fail.
valve cracking pressure, and tests flow sensor zero • Block and unblock To patient
offsets. port as directed.
• Disconnect and reconnect air
and oxygen as directed.
NOTE:
If you are using the
compressor as the air
source, press the CLEAR key
when prompted to connect
air.
SM Leak Test Checks for system leaks by pressurizing the system • Leaks or occlusions in the
and verifying that pressure drops by no more than expiratory filter or elsewhere
5 cmH2O in 10 s. will cause this test to fail.
GUI Keyboard Test Verifies that GUI keys are functioning by prompting Press keys within 15 s when
the operator to press a key. prompted.
GUI Knob Test Verifies GUI knob rotation direction by prompting Turn GUI knob as prompted, within
the operator to rotate the knob in one direction, 15 s.
then the other.
GUI Lamp Test Verifies that GUI LEDs are functioning by prompting Observe GUI indicators and press
the operator to acknowledge that the LEDs are ACCEPT (if on) or CLEAR (if not on).
turned on.
BD Lamp Test Verifies that BDU lights are functioning. Observe BDU indicators and press
ACCEPT (if on) or CLEAR (if not on).
NOTE:
The ventilator inoperative and safety valve
open LEDs on the GUI are hard-wired to the
BDU and are considered BDU LEDs.
GUI Audio Test Verifies that GUI audio is functioning by prompting Listen for GUI audio. Press ACCEPT
the operator to listen for the alarm sound. (if on) or CLEAR (if not on).
GUI Nurse Call Verifies that nurse’s call relay is functioning. If not testing nurse’s call, press
CLEAR. If testing nurse’s call, press
ACCEPT, then verify nurse’s call
operation as prompted.
BD Audio Test Verifies that BDU audio is functioning. Also verifies Listen for BDU audio alarm.
that power fail capacitor can operate loss-of-power
alarm for at least 2 minutes.
Safety System Test Tests safety valve and inspiratory check valve
operation by checking for occlusions when the safety
valve is open, checking safety valve loopback current,
and checking inspiratory check valve reverse flow.
Exp Valve Seal Test Verifies that exhalation valve seal and poppet friction
are acceptable.
Exp Valve Test Verifies that current versus pressure values in flash
memory correspond with actual installed exhalation
valve.
Exp Heater Test Verifies that gas flow temperature increases and
decreases in response to commanding the heater on
and off.
Compressor Test Tests compressor operation. This test checks the timer, • If wall air is connected,
compressor pressure switch, and compressor disconnect wall air as directed.
operating states (run, standby, and disabled). • If AC is not connected, connect
AC power as directed.
Compressor Leak Test Verifies that a fully charged compressor (if installed) • Disconnect and connect wall air
does not leak enough to cause the compressor to as directed.
turn on within 1 minute while in standby mode with
wall air connected.
NOTE:
If you are using the
compressor as the air
source, press the CLEAR key
when prompted to connect
wall air.
Analog Data Display Reads and displays analog data channels in O2 sensor data is displayed in DAC
engineering units. counts.
GUI Serial Port Test Verifies that GUI serial port can send and receive
data.
Battery Test With AC disconnected, tests BPS under discharging Test takes 5 minutes to complete.
and charging conditions.
If the test
It means: Do this:
result is:
ALERT Test results not ideal, but not Repeat test by touching REPEAT.
critical. EST halts. Skip to next test by touching NEXT.
Repeat EST from the beginning by touching RESTART EST.
Exit EST in order to service ventilator or review error codes by
touching EXIT EST.
In Single Test EST, touching “NEXT” allows you to choose
another test to run.
FAILURE A critical problem has been Repeat failed test by touching REPEAT.
detected, and EST cannot Skip to next test by touching NEXT.
complete until the ventilator Repeat EST from the beginning by touching RESTART EST.
passes the failed test.
Exit EST in order to service ventilator or review error codes by
touching EXIT EST.
In Single Test EST, touching “NEXT” allows you to choose
another test to run
NEVER RUN New ventilator software has Run EST by touching the ALL TESTS button.
(This message been downloaded or a Single
appears in the Test EST was run.
Ventilator Test
Summary)
If the test
It means: Do this:
result is:
Outcome: ALL Single Test EST test was initiated. Run EST by touching the ALL TESTS button.
TESTS required You must run EST with a passing
(This message result to resume ventilation.
appears in the
Diagnostic Code
log)
If the EST
It means: Do this:
outcome is:
ALERT One or more faults were Repeat EST from the beginning by touching RESTART EST.
detected. Override the alert in order to service ventilator or review error
codes by touching OVERRIDE, followed by EXIT EST.
NOTE:
Although you can override an alert in EST, normal
ventilation is not allowed until the problem is
corrected and all EST tests passed.
FAILURE One or more critical faults Repeat EST from the beginning by touching RESTART EST. Exit EST
detected. A ventilator in order to service ventilator or review error codes by touching
inoperative condition is EXIT EST.
declared.
NOTE:
If special equipment for the particular test you selected is needed, the GUI prompts
you to install the equipment.
7. If you have completed Single Test EST tests, and the test(s) passed, run the full suite of
EST tests (Section 3.6.4). If failures occur, repair the ventilator and run EST.
8. If a test fails or completes with an ALERT outcome, REPEAT to retry the test, touch
“NEXT” to choose another test, or exit Single Test EST by touching the EXIT button.
Service Mode
The Puritan Bennett™ 840 Ventilator System operates in two modes: Patient
Ventilation and Service Mode. The Service Mode is intended for use by a trained
service technician to aid in ventilator testing and troubleshooting and to perform
system calibrations.
This chapter details the Service Mode, including prompts and expected system
responses. It also includes how to calibrate the oxygen sensor, which is required
during almost all service operations, but occurs outside of Service Mode.
Warning
Do not enter the Service Mode while a patient or other person is connected
to the ventilator. To do so may cause patient injury or death. Normal
ventilator operation is not accessible while in the Service Mode.
NOTE:
During Service Mode startup, certain functions may automatically execute
to program data into BDU and/or GUI NOVRAM. Refer to Section 4.2.11
for details. While in the Service Mode, normal ventilation is not allowed.
Service Mode
Function Purpose
Upper screen
SST RESULT Displays the latest SST test results, including test name, date and time each test was
last run, individual test results, and overall test outcome.
NOTE:
When reviewing SST results, be aware that not all of the results for each SST
test may be from the same date and time run.
DIAG LOG Displays the contents of System Diagnostic Log, System Information Log, and
EST/SST Diagnostic Log.
Alarm log Displays the history of most recent ventilator alarms for the previous patient.
NOTE:
Running EST or selecting New Patient will erase this log.
VENT CONFIG Displays software revisions and serial numbers. Software revisions displayed
include: GUI and BDU boot PROM and application software as well as GUI audible
alarm subsystem (SAAS). Serial numbers displayed are for BDU, GUI, and compressor
unit (if installed); these serial numbers are read from the data key and compressor
(if installed).
OPERATION TIME Displays accumulated ventilator and compressor operational time, in hours.
TEST SUMMARY Displays the latest EST and SST test results, including date and time each was last
run and overall test outcome.
Lower screen
EST Displays the details of the last EST run. Runs EST to thoroughly test the operational
integrity of the ventilator, both electronics and pneumatics.
DATE/TIME Adjusts current date and time and allows the user to select a date format.
UPPER SCREEN
Compact
Flash Test*
Nominal External
Line Pressure
Control Unit
Voltage Baud Rate
LOWER SCREEN
NOTE:
SST lets you exit and resume patient ventilation if required. Thus, the results shown
are not necessarily all from the same SST run. If a partial SST run is performed, the
results of the first few tests might be more recent than the results of later tests.
NOTE:
System Diagnostic Information and EST/SST Diagnostic logs can only be erased by
using the Puritan Bennett™ 840 VTS Breath Labs Software. The alarm log is
automatically erased each time a new patient is setup is selected and when EST is run.
This function lets you read the contents of the ventilator’s error and status logs. The logs are
as follows:
• The System Diagnostic Log (Figure 4-3) stores data on unexpected conditions detected
during POST, background checks. This includes alerts and failures.
• The System Information Log (Figure 4-4) stores data on ventilator events, including details
on GUI/BDU communication.
• The EST/SST Diagnostic Log (Figure 4-5) stores data on overall test outcomes and events
during SST and EST.
NOTE:
The performance verification, described in Chapter 5 of this manual, is a more
thorough test of the ventilator to verify it meets specifications.
Function Purpose
Service Mode Setup Lets you adjust nominal line voltage lower limit, baud rate for RS-232 port, and lets
you choose the unit for display of pressures (cmH2O or hPa).
External Test Control Lets you test the ventilator using a computer, Breath Labs 840 VTS Software, and
PTS 2000 tester, to run performance verification. Also used to download new
application software to the Puritan Bennett™ 840 ventilator.
Exp Valve Calibration Performs calibration. For use after a new exhalation valve, AI PCB, or BD CPU PCB is
installed, or as part of the performance verification.
Vent Inop Test Exercises ventilator inoperative circuitry during five test phases.
Flow Sensor Calibration Performs calibration. For use after a flow sensor replacement or as part of the
performance verification.
Atmospheric Pressure Performs calibration. For use after an Inspiratory or Expiratory Electronics PCB
Transducer replacement, new vent installation, if the vent has been moved ± 1000 feet in
elevation, or as part of the performance verification.
Data key Update Transfers the operational hours from the ventilator to the new data key. For use
after software options have been installed.
Serial Loopback Test (10.4-inch Tests the serial communication ports on the 10.4-inch GUI.
GUI only)
Compact Flash Test Only available if Trending software option is installed. Performs internal diagnostics
and functional tests of the compact flash memory on the TE GUI CPU PCB.
Function Purpose
Nominal Line Voltage Nominal line voltage lower limit. This setting is used to detect low power alarm
conditions. It has no direct effect on power supply function. Improper line voltage
settings will cause a Device Alert, preventing normal operation. You must run EST
to reset this Device Alert.
External Control Baud Rate For GUI RS-232 port. The baud rate takes effect when you restart the ventilator in
the Service Mode.
NOTE:
• If the ventilator has not reached operating temperature from recent usage, allow
it to warm up for at least 10 minutes in Service Mode before performing this
calibration, to ensure a successful calibration.
• Ensure that the ventilator has passed the Leak Test in EST prior to performing this
calibration.
This function builds the exhalation valve calibration table. This table lists exhalation valve
DAC commands versus exhalation pressure. The exhalation valve calibration requires a gold
standard test circuit and lasts approximately 3 minutes.
If the calibration fails, try rerunning it. If it continues to fail, refer to Table 4-4 to troubleshoot
any errors. Run EST to assist in troubleshooting.
NOTE:
A system leak is a common source of failure, particularly expiratory filter leaks. Make
sure the expiratory filter collector vial is secured tightly, the drain port is capped, and
the filter is mounted securely to the ventilator. Make sure the test circuit is securely
attached.
Bad exp valve loopback Failure Exhalation valve and/or analog 1. Clean the exhalation valve if
current interface PCB may be out of inspection warrants and re-run
specification. the calibration.
2. Replace the exhalation valve.
Both insp and exp Failure Analog Interface PCB may be out of 1. Replace the AI PCB.
autozero failed specification. 2. Replace the BD CPU PCB.
Calibration failed: Failure Exhalation valve and/or analog 1. Verify that there are no leaks or
current limit exceeded interface PCB may be out of occlusions.
specification. 2. Verify expiratory module
connections.
3. Clean the exhalation valve.
4. Replace the exhalation valve.
Calibration failed: gain Failure Exhalation valve and/or analog 1. Verify that there are no leaks or
resolution interface PCB may be out of occlusions.
specification. 2. Verify expiratory module
connections.
3. Clean the exhalation valve.
4. Replace the exhalation valve.
Calibration failed: Failure Verify that gold standard circuit is 1. Verify that there are no leaks or
Pressure build time-out. still connected and expiratory filter occlusions.
is locked in place. Verify that seal is 2. Run EST circuit pressure and leak
placed over collector vial drain and tests.
vial is securely fastened to 3. Clean the exhalation valve.
expiratory filter. 4. Replace the exhalation valve.
5. Replace expiratory pressure
transducer.
6. Replace inspiratory pressure
transducer.
Exhalation valve Failure Expiratory pressure transducer DAC 1. Verify that there are no leaks or
pressure transducer count at 0 cmH2O is out of range. occlusions.
autozero failed 2. Replace expiratory pressure
transducer autozero solenoid.
3. Replace expiratory pressure
transducer.
Exhalation valve Failure Exhalation valve temperature out of 1. Verify that the ventilator has
temperature out of range. warmed up for at least 10 minutes
range in Service Mode.
2. Replace exhalation valve.
3. Replace AI PCB.
Flow sensor cross-check: Failure Flow controller is not able to 1. Verify that there are no leaks or
unable to establish flow establish a 5 L/min test flow via the occlusions.
air flow sensor. 2. Swap air flow sensor (Q2) and
oxygen flow sensor (Q1). If
calibration passes, replace Q2.
3. Swap air PSOL (PSOL2) and
oxygen PSOL (PSOL1). If
calibration passes, replace PSOL2.
4. Replace expiratory flow sensor
(Q3).
Flow sensor cross-check Failure Expiratory air flow is out of range 1. Verify that there are no leaks or
failed with respect to the 5 L/min occlusions.
inspiration-controlled test flow. 2. Exit exhalation valve calibration
Verify gold circuit is connected. and perform flow sensor
Check for expiratory filter leaks. calibration.
3. Swap air flow sensor (Q2) and
oxygen flow sensor (Q1). If
calibration passes, replace Q2.
4. Swap air PSOL (PSOL2) and
oxygen PSOL (PSOL1). If
calibration passes, replace PSOL2.
5. Replace expiratory flow sensor
(Q3).
Flow sensor info invalid Failure Flow sensor calibration table 1. Re-run flow sensor calibration.
checksum error occurred. 2. Replace expiratory flow sensor
(Q3).
3. Replace air flow sensor (Q2).
Inspiratory autozero Failure Inspiratory pressure transducer DAC 1. Verify that there are no leaks or
failed count at 0 cmH2O is out of range. occlusions.
2. Replace inspiratory autozero
solenoid (SOL1).
3. Replace inspiratory pressure
transducer.
No air connected Failure Wall and compressor air pressure 1. Verify air is connected.
switches detect no air connected. 2. Replace air pressure switch (PS2).
Pressure sensor cross- Failure Pressurization algorithm is unable to 1. Check for expiratory filter leaks.
check: unable to build pressurize system to one or more of 2. Replace inspiratory pressure
pressure five test pressures via the inspiratory transducer.
pressure transducer.
Pressure sensor cross- Alert Pressure cross-check alert occurred 1. Verify that there are no leaks or
check alert: approaching at one or more of five test pressures, occlusions.
spec limit no cross-check failures occurred. 2. Replace inspiratory pressure
transducer.
3. Replace expiratory pressure
transducer.
Pressure sensor cross- Failure Pressure transducer cross-check 1. Verify that there are no leaks or
check failed failed at one or more of these five occlusions.
test pressures (5, 25, 50, 75, and 2. Verify adequate supply pressure.
100 cmH2O). 3. Replace inspiratory pressure
transducer.
4. Replace expiratory pressure
transducer.
Unable to establish flow Failure Flow controller is not able to 1. Verify that there are no leaks or
establish a 5 L/min test flow via the occlusions.
expiratory flow sensor. 2. Exit exhalation valve calibration
and perform flow sensor
calibration.
3. Swap air PSOL (PSOL2) and
oxygen PSOL (PSOL1). If
calibration passes, replace PSOL2.
4. Swap air flow sensor (Q2) and
oxygen flow sensor (Q1). If
calibration passes, replace Q2.
5. Replace expiratory flow sensor
(Q3).
Unable to program flash Failure Unable to program calibration table 1. Check BDU +12 V supply and BD
into BDU NOVRAM. CPU PCB.
2. Replace power supply.
3. Replace BD CPU PCB.
4. Replace AI PCB.
If a test phase fails (“Vent Inop Test failed” message in lower screen), powering the ventilator
down and back up into Service Mode will invoke the normal Service Mode. Troubleshoot as
indicated in Table 4-6. You must repeat the Vent Inop Test and pass it with no failures before
ventilation is allowed.
GUI vent inop test phase BDU alarm (alone) Check BDU alarm or AI PCB.
NOTE:
The BD CPU PCB controls the
ventilator inoperative and
safety valve open LEDs on
both the BDU and GUI CPU
PCBs, sending the two LED
signals to the GUI via the
BDU-GUI cable.
BDU alarm and ventilator Check BDU-GUI cable, GUI LED PCB,
inoperative and safety valve open or GUI CPU PCB.
LEDs
Any other test phase Any Check BD LED PCB, BD alarm, BD CPU
PCB or AI PCB.
NOTE:
• If the ventilator has not reached operating temperature from recent usage, allow
it to warm up for at least 10 minutes in Service Mode before performing this
calibration, to ensure a successful calibration.
• It is very important that the ventilator passes the leak test during EST prior to
running a flow sensor calibration.
• A system leak is a common source of failure, particularly expiratory filter leaks.
Make sure the expiratory filter collector vial is secured tightly, the drain port is
capped, and the filter is mounted securely to the ventilator. Make sure the test
circuit is securely attached.
• Use both air and oxygen gas sources to run Flow Sensor calibration. Ensure gas
source is capable of supplying 200 L/min.
This function builds a table of expiratory flow sensor (Q3) offsets. The flow sensor calibration
requires a gold standard test circuit.
If the calibration fails, make sure that the test circuit is attached and that the expiratory filter
is securely connected and free of leaks, then try rerunning it. If the calibration continues to
fail, refer to Table 4-7 to troubleshoot any errors.
Air offset out of range Failure Determines maximum air flow 1. Monitor air pressure at TP2 with regu-
capability in 500 ms. Measures lator calibration hose, P/N 4-079050-00
inspiratory and expiratory flow to ensure adequate supply pressure.
offsets at maximum flow, then 2. Swap air PSOL (PSOL2) and oxygen
decreases to 1 L/min in 1 L/min PSOL (PSOL1). If calibration passes,
increments. replace PSOL2.
3. Swap air flow sensor (Q2) and oxygen
flow sensor (Q1). If calibration passes,
replace Q2.
4. Try installing a known good
inspiratory module. If calibration
passes, replace inspiratory electronics
PCB. If calibration fails, replace
inspiratory module blindmate cable.
5. Replace expiratory flow sensor (Q3).
6. Replace AI PCB.
Cannot achieve Failure System cannot sustain a 1. Monitor air pressure at TP2 with regu-
minimum air flow minimum 60 L/min flow. lator calibration hose, P/N 4-079050-00
to ensure adequate supply pressure.
2. Swap air PSOL (PSOL2) and oxygen
PSOL (PSOL1). If calibration passes,
replace PSOL2.
3. Swap air flow sensor (Q2) and oxygen
flow sensor (Q1). If calibration passes,
replace Q2.
4. Try installing a known good
inspiratory module. If calibration
passes, replace inspiratory electronics
PCB. If calibration fails, replace
inspiratory module blindmate cable.
5. Replace expiratory flow sensor (Q3).
6. Replace AI PCB.
Cannot achieve Failure System cannot sustain a 1. Monitor oxygen pressure at TP1 with
minimum O2 flow minimum 60 L/min flow. Check regulator calibration hose,
oxygen supply pressure and P/N 4-079050-00 to ensure adequate
REG1. supply pressure.
2. Swap oxygen PSOL (PSOL1) and air
PSOL (PSOL2). If calibration passes,
replace PSOL1.
3. Swap oxygen flow sensor (Q1) and air
flow sensor (Q2). If calibration passes,
replace Q1.
4. Try installing a known good
inspiratory module. If calibration
passes, replace inspiratory electronics
PCB. If calibration fails, replace
inspiratory module blindmate cable.
5. Replace expiratory flow sensor (Q3).
6. Replace AI PCB.
No air connected Failure Wall air pressure switch (PS2) Verify air is connected.
detects no air connected.
O2 offset out of range Failure 1. Replace oxygen flow sensor 1. Monitor oxygen pressure at TP1 with
(Q1). regulator calibration hose,
2. Replace expiratory flow P/N 4-079050-00 to ensure adequate
sensor (Q3). supply pressure.
3. Replace AI PCB. 2. Swap oxygen PSOL (PSOL1) and air
PSOL (PSOL2). If calibration passes,
replace PSOL1.
3. Swap oxygen flow sensor (Q1) and air
flow sensor (Q2). If calibration passes,
replace Q1.
4. Try installing a known good
inspiratory module. If calibration
passes, replace inspiratory electronics
PCB. If calibration fails, replace
inspiratory module blindmate cable.
5. Replace expiratory flow sensor (Q3).
6. Replace AI PCB.
NOTE:
When calibrating the atmospheric pressure transducer, make sure that the pressure in
the room is stable and not subject to pressure changes caused by door closures, heavy
foot traffic, or the startup of an air conditioning system. Any such environmental
changes will result in an incorrect adjustment being made to the ventilator.
This function calibrates the atmospheric pressure transducer using an external barometer.
Refer to Table 1-10 for barometer accuracy specifications.
Allow the barometric pressure reading to stabilize, then touch the button and turn the knob
to enter the pressure (in mmHg) measured by this external barometer, as prompted.
(1 mmHg = 1.333 mbars = 133.33 Pa = 0.019 psia.)
If the calibration displays the Atmospheric Pressure Transducer OOR (out of range) message,
try rerunning it, making sure you correctly enter the barometric pressure. If the calibration
continues to fail, try rerunning EST, then replacing the inspiratory pressure transducer
autozero solenoid (SOL1), the inspiratory electronics PCB, and then the AI PCB.
NOTE:
If the data key is updated before calibrations have been run, the operational hours
will be lost or become corrupted.
NOTE:
If the compact flash component test fails, the Trending option will be disabled until a
functional component is installed. The ventilator’s performance is otherwise
unaffected.
Unable to read air flow sensor Error reading air flow sensor (Q2) serial EEPROM
data. Check Q2 and AI PCB.
Unable to read O2 flow sensor Error reading oxygen flow sensor (Q1) serial
EEPROM data. Check Q1 and AI PCB.
Unable to read the expiratory flow sensor Error reading expiratory flow sensor (Q3) serial
EEPROM data. Check Q3 and AI PCB.
Unable to receive flash data GUI did not receive all calibration data from
BDU within 10 s. Check BDU/GUI cable, BD CPU
PCB, or GUI CPU PCB.
Caution
When troubleshooting with a GUI not originally matched to the BDU, do not press the
ACCEPT key when prompted to match serial numbers. If you press ACCEPT, a software
download will be required to restore the original serial numbers.
NOTE:
• You are given the option to “Proceed without setting serial number.” You may
want to use this option for troubleshooting purposes. For example, if a new BDU
or GUI CPU PCB is inserted temporarily for testing, you may find it useful to
maintain the default serial number on that PCB; otherwise, by setting the serial
number the test CPU PCB will become “committed” to the ventilator used to
initialize it. If you choose not to initialize the serial numbers, be aware that
ventilation is not allowed until serial numbers are copied from the data key.
• It is possible to have one CPU with a valid serial number in NOVRAM but the other
CPU containing the default serial number. For example, when the BDU only is
downloaded with new software, the BDU serial number in NOVRAM is set to the
default, but the GUI serial number in NOVRAM is unaffected.
• Do not use the same data key to initialize the serial numbers in more than one
ventilator.
Cannot program flash BDU and/or GUI NOVRAM cannot be successfully programmed. Check BDU and
GUI +12 V supplies, BD CPU PCB, and GUI CPU PCB.
Data key is not installed If a data key is installed but not being detected, try inserting the key into a
known good ventilator whose serial numbers have already been initialized.
Then power the ventilator up into Service Mode and verify that the ventilator
sees an invalid data key (that is, the message “Serial number(s) doesn’t match
the data key” is displayed). If the data key is still not detected, replace it;
otherwise, replace the AI PCB. Alternatively, try using a known good data key
in the suspect ventilator and verify that it is detected.
NOTE:
If no data key is inserted, you can proceed with Service Mode but
ventilation is not allowed.
Serial number(s) doesn’t match the Press ACCEPT to confirm and proceed with Service Mode.
data key
NOTE:
If an invalid data key is inserted, you can proceed with Service Mode
but ventilation is not allowed.
Performance Verification
NOTE:
Performance verification procedures do not apply to ventilator
accessories. Refer to the specific accessory operator’s or service manual for
performance verification information. Malfunctioning accessories may
affect some ventilator functions and may result in false test results.
Table 5-1: Tools, equipment, and service materials required for performance verification
Required for
Manufacturer/model or
Description PTS 2000
Covidien part number
use?
Puritan Bennett 840 VTS (Ventilator Test Software) 4-075359-00 and current X
revision Software Download
CD-ROM
Digital multimeter (DMM) accurate to 3 decimal places, with test Fluke Model 87 or equivalent
leads
Electrical safety analyzer capable of measuring ground resistance Dale Model 600, Dale
and leakage current Technology Inc., P.O. Box 196,
401 Claremont Ave.,
Thornwood, NY 10594 USA,
800.544.3253
http://www.daletech.com/
or equivalent
Filter, inspiratory, Re/Flex (for use on PTS 2000 tester port) 4-074600-00 X
Flex tube (Gold standard tube), 21.0 in. (53.4 cm) (quantity = 2) *4-018506-00 X
Hose, regulator calibration (refer to Section 5.8 for regulator *4-079050-00 (quick X
calibration hose repair instructions) disconnect) (4-079050-00
4-079051-00 (female) only)
* These parts may be purchased individually or by ordering a Puritan Bennett™ 840 VTS Accessories Kit,
P/N 4-076599-00.
Table 5-1: Tools, equipment, and service materials required for performance verification (continued)
Required for
Manufacturer/model or
Description PTS 2000
Covidien part number
use?
Pentium 166 Mhz computer, equipped, at a minimum, with the Local supplier X
following:
• 32 MB RAM (64 MB for NT systems)
• 800x600 screen display resolution with 256 colors
• Two available serial ports or USB/serial adapters
• Windows™* 95, 98, 2000, XP or NT 4.0 operating system
• Mouse or equivalent pointer device
• 2x or higher CD-ROM drive
• 1 Gigabyte available hard disk space
• One available TCP/IP network port with BNC connector
(if software download desired)
Pneumatic calibration analyzer or equivalent devices capable of Puritan Bennett™ PTS 2000 X
measuring oxygen percent, flow, BTPS volume, pressure, and Performance Test System (also requires
barometric pressure. Oxygen analyzer connector tee. Required (4-076185-00) (includes Puritan
accuracies: accessory kit with serial cable) Bennett
• Flow: 2.75% of reading ± 0.05 slpm for performance verification; 840 VTS
or equivalent device for software)
• Volume: 2% of reading or ± 1 digit
• Low pressure (-150 to +150 cmH2O): manual ventilator check
0.75% of reading ± 0.04 cmH2O
• High pressure (0 to 150 psig): 1.0% of reading ±0.1 psi
• Oxygen percentage: ± 2% oxygen
• Barometric pressure: Range: 10 to 16 psia; resolution:
0.0 to 1 psia; operating temperature: 10 to 40 °C; measurement
accuracy: 0.75 of reading; response: 100 ms
Serial Card, Socket I/O Ruggedized or equivalent (Provides second Socket Communications Corp.
serial port) Newark, CA
http://www.socketcom.com/
* These parts may be purchased individually or by ordering a Puritan Bennett™ 840 VTS Accessories Kit,
P/N 4-076599-00.
Table 5-1: Tools, equipment, and service materials required for performance verification (continued)
Required for
Manufacturer/model or
Description PTS 2000
Covidien part number
use?
Tubing, silicone, 3/16 in. ID x 5/16 in. OD, 36 in. (91.4 cm) 4-008577-00 X
Tubing, silicone, 1/8 in. ID x 1/4 in. OD, 4.75 in. (12.1 cm) *4-008578-00 X
Vacuum cleaner, ESD-safe, with 0.2 μm filter (rated for Local supplier
photocopiers and laser printers)
* These parts may be purchased individually or by ordering a Puritan Bennett™ 840 VTS Accessories Kit,
P/N 4-076599-00.
Warning
To prevent disease transmission, use personal protective equipment when handling
contaminated bacterial filters or other patient accessories. Refer to the
Puritan Bennett™ 800 Series Ventilator System Operator’s and Technical Reference
Manual for instructions on sterilizing patient system parts.
Caution
Do not soak any portion of the ventilator in solvent, alcohol, or any other cleaning
agent. Soaking ventilator components may damage the ventilator.
NOTE:
A humidification device is not necessary for performance verification, but is required
for completion of SST.
Expiratory valve
Performance
Flow sensor
calibration
calibration
calibration
Download
Software
Type of service
Ventilator installation 1 -- 2 3 4 5 6 7 8 -- 92
10,000-hour preventive 1 -- 2 3 4 5 6 7 8 -- 9
maintenance (ventilator or
compressor)
Oxygen sensor 1 -- -- -- -- 3 -- 4 2 -- --
NOTE:
To ensure that the ventilator performs within specifications, Covidien recommends
that you use the Puritan Bennett™ PTS 2000 Performance Test System in conjunction
with the BreathLab 840 Verification Test Software. If this equipment is not available,
you can use an alternative test that checks some parameters but does not guarantee
the integrity of the entire system. Refer to Table 5-3 to determine the approximate
test limits if you use equipment other than the PTS 2000 Performance Test System.
To ensure systematic performance verification and logical fault diagnosis, perform tests in the
order given. Follow these general guidelines when running the performance verification:
Warning
Follow accepted safety procedures for electrical equipment when making connections,
adjustments, or repairs.
Warning
Do not enter the service mode while a patient or other person is connected to the
ventilator. To do so may cause patient injury or death. While in the service mode,
normal ventilator operation is not accessible.
NOTE:
The procedures that follow do not verify the performance of accessories. Verify the
performance of accessories using the appropriate procedures in the applicable
operator’s or service manual.
Warning
If the ventilator fails an electrical safety test, do not proceed to the next electrical
safety test until the problem is corrected and the ventilator is retested.
Caution
To prevent possible equipment damage, do not attempt to perform HI-POT testing on
the ventilator. The ventilator design incorporates high-voltage protective devices that
may be damaged if HI-POT testing is performed on the entire system. The ventilator is
HI-POT-tested during manufacturing, using processes that do not damage the
protective devices.
NOTE:
• After servicing the ventilator, perform an electrical safety test before putting
the unit back into operation.
• Before performing the electrical safety test, make sure the compressor unit (if
installed) and all accessories are connected and operational.
pin 3
Figure 5-1. Puritan Bennett™ 840 ventilator side view — PTS 2000 port
NOTE:
• A test (gold standard) patient circuit and a no. 1 stopper are required to
perform EST.
• The nurse’s call circuit is tested during EST and may be bypassed by selecting
the CLEAR button.
• If the ventilator has not reached operating temperature from recent usage,
allow it to warm up for at least 10 minutes in service mode before running EST,
to ensure accurate testing.
NOTE:
During testing, it is sometimes necessary for the user to perform test-related tasks,
such as disconnecting a gas supply or blocking the To Patient port. For each such step,
messages are displayed in the prompt area.
3. Verify that EST completes all tests with a PASS result. Performance verification is not
complete if any EST test results in an ALERT, OVERRIDE, or FAILURE. If a test fails, refer to
Chapter 8 for repair information. You will enter the results of EST when prompted by the
Breathlab 840 VTS software (Section 5.7.6)
Caution
To prevent damage to ESD-sensitive components, always follow ESD guidelines when
performing this procedure.
NOTE:
Regulator performance must be verified each time performance verification or a
manual ventilator check is performed. Use the appropriate type of regulator
calibration hose (see Table 5-1).
NOTE:
If regulators are out of specification, refer to Chapter 8 for service and repair of the
inspiratory module.
5.7.6 Performance verification using PTS 2000 Performance Test System and
BreathLab 840 VTS software
NOTE:
If performing full performance verification testing, use the results recorded from the
previous performance verification tests to complete this step of Performance
Verification Testing.
1. Assemble test equipment indicated in Table 5-1. Refer to Figure 5-4 and Figure 5-5 for
equipment setup information. (Software installation instructions are provided with
BreathLab 840 VTS Software.)
2. Turn on computer, and bring up the VTS program.
3. Turn ventilator on, and bring up in service mode.
4. Using ventilator test program, press START TEST to begin test application.
5. Follow prompts displayed on computer screen to complete performance verification.
Access the Breathlab 840 VTS Software Help Screen for additional information.
Reports icon
2. Click the desired report button and then click to view a test results log or to
cancel. The test results log appears in a text document to save and print as desired.
4-L
Use for PCV tests only Lung
To Patient Port
Puritan Bennet™
840 BDU Adult
PTS
Patient 2000
Circuit
Insert cork into high flow
exhaust port during SST.
From Patient Port After SST, attach lungs here.
Figure 5-4. Puritan Bennett 840 Ventilator System PVT setup with PTS 2000
Puritan Bennett™
Port 1
840 BDU
Trigger
cable Serial cable
5.7.7 Manual ventilator check using equipment other than PTS 2000
Performance Test System
NOTE:
• To ensure proper ventilator performance, Covidien recommends that you use
the Puritan Bennett™ PTS 2000 Performance Test System along with the
BreathLab 840 VTS software.
• If you are using a pneumatic analyzer other than the PTS 2000 Performance
Test System, you must also have an oxygen analyzer available.
• The manual ventilator check (i.e., without PTS 2000 and Breathlab 840 VTS
software) does not satisfy the manufacturer’s performance testing
requirements. The manual ventilator check can be used as a general indicator
of ventilator condition.
NOTE:
To accurately measure the parameters listed in Table 5-3, add the tolerance of your
measurement device to the target value of the ventilator. The test results you obtain
are approximate and may not guarantee the accuracy of the ventilator.
Diagnostic Codes
6.1 Introduction
This chapter contains information to assist in the interpretation of diagnostics
codes, indicated in the Puritan Bennett™ 840 Ventilator System’s error and status
logs, as well as diagnostic codes reported by the diagnostic LED arrays on the BD
CPU PCB and the GUI CPU PCB.
The ventilator’s diagnostic and status logs also contain other information that can
be useful when troubleshooting the ventilator. The logs are:
• System Diagnostic Log: Stores data on unexpected conditions detected during
POST, background checks, and calibrations. This includes alerts and failures.
• System Information Log: Stores data on ventilator events, including details on
GUI/BD communication.
• EST/SST Diagnostic Log: Stores data on unexpected conditions detected during
SST and EST, along with overall results at completion of SST and EST.
You access these logs through the service mode, as described in Section 4.2.2.
Table Page
Table 6-1: Values of diagnostic code digits 6-4
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes 6-7
Table 6-3: Address codes for BDU POST analog devices test errors 6-75
Table 6-4: BDU and GUI— Phase 1 (kernel) POST diagnostic codes 6-77
Table 6-5: BDU only — Phase 2 POST diagnostic codes 6-79
Table 6-6: GUI only — Phase 2 POST diagnostic codes 6-82
Table 6-7: BDU or GUI—Phase 3 POST diagnostic codes 6-84
Table 6-8: Interrupt errors and test failures – POST self tests 6-85
Table 6-9: SST diagnostic codes 6-87
Table 6-10: EST test sequence 6-98
Table 6-11: UT0002 Fault Addresses for VGA LCD controller PCB errors 6-124
Diagnostic Codes
6.3 Troubleshooting
The table of diagnostic codes (Table 6-2) lists repair and replacement actions that are
sequenced to correct the most probable malfunction or to present the most efficient
corrective action first. (The proposed fixes listed, however, may not always correct the
particular problem.)
Also, keep the following in mind when troubleshooting the ventilator:
• Verify secure connections of cables and ventilator modules.
• If possible, run full EST, bypassing any failures or faults that may occur, to further diagnose
a problem.
• Replace BD or GUI CPU PCB only after all other remedies have been attempted. Contact
Covidien Technical Support for further assistance.
• Check for system leaks, particularly at the expiratory filter. Make sure the expiratory filter
collector vial is secured tightly, the drain port is capped, and the filter is mounted securely
to the ventilator. Make sure the test circuit is securely attached.
• Before running SST or EST, ensure the unit has been warmed up for at least 10 minutes in
ambient temperature. Failure to warm up the ventilator may result in false flow sensor or
pressure transducer failures.
• When troubleshooting SST or EST, always use a known good patient circuit or gold
standard circuit and filter(s).
NOTE:
The alphanumeric diagnostic code for each of the self tests is not identified in this
manual. This information is not necessary for field service repairs.
If the system detects a fault during POST (or a background check), each of the four pairs of
LEDs in the corresponding LED array will light, in sequence, creating a flashing effect. The
same LED array then displays the diagnostic code corresponding to the detected fault. This
LED pattern of display—flashing LED pairs followed by a diagnostic code—cycles continuously
and is a visual indicator of a system fault. Use the tables in this section to convert the
diagnostic code LED pattern to an alphanumeric diagnostic code.
Additionally, under a fault condition, the processor of the corresponding CPU PCB attempts
to log the fault information, with a time-stamp, in its nonvolatile random access memory
(NOVRAM). Upon completion of its POST, each processor reports its test results to the GUI
processor for display and transfer into the System Diagnostics Log.
POST failures are classified as minor or major faults:
Minor POST fault is a fault that does not affect ventilation or patient safety checks. A minor
fault does not interrupt the regular POST sequence. The GUI displays POST fault information
and logs it into NOVRAM of the corresponding CPU PCB. The System Diagnostic Log displays
an alert. Normal ventilation will begin if POST detects a minor fault.
Major POST fault is a fault that affects ventilation or patient safety checks. A major fault
interrupts the regular sequence of POST and causes a VENT INOP condition. Fault information
is sent to the GUI (if possible) and to the LED array on the corresponding CPU PCB (BD or GUI).
The safety valve and exhalation valve remain open to allow the patient to breathe room air.
The software does not permit normal ventilation while a major fault exists. To clear a major
fault, both CPU processors must complete their respective POST, with no major POST faults. If
only a minor fault exists, normal ventilation will begin.
Status LED
diagnostic code LEDs
(read from top
to bottom)
S GUI event
b B Background check
C Communications test
E EST
N Nonmaskable interrupt
S SST
• The fifth column in the table provides additional information and/or identifies possible
causes.
• The sixth column suggests how to correct the condition. These actions are sequenced to
correct the most probable malfunction or to present the most efficient corrective action
first. The proposed fixes listed, however, may not always correct the particular problem.
In addition to the actions suggested in the tables in this section, keep the following in mind
when troubleshooting the ventilator:
• Verify secure connections of cables and ventilator modules.
• If possible, run full EST, bypassing any failures or faults that may occur, to further diagnose
a problem.
• Replace BD or GUI CPU PCB only after all other remedies have been attempted. Contact
Covidien Technical Support for further assistance.
• Check for system leaks, particularly at the expiratory filter. Make sure the expiratory filter
collector vial is secured tightly, the drain port is capped, and the filter is mounted securely
to the ventilator. Make sure the test circuit is securely attached.
• Before running SST or EST, ensure the unit has been warmed up for at least 10 minutes in
ambient temperature. Failure to warm up the ventilator may result in false flow sensor or
pressure transducer failures.
When troubleshooting SST or EST, always use a known good patient circuit or gold standard
circuit and filter(s).
Message Information /
Code Test Response Corrective action
displayed possible cause
AE0110 Unable to EST Flow Alert Flow controller 1. Make sure oxygen sup-
establish O2 sensors cross unable to establish ply is connected and
flow check Test and control oxygen unrestricted.
flow at 120 L/min. 2. Verify oxygen regulator
pressure is set between
9-12 psi.
3. Switch PSOLs to see if
failure transfers to the
other gas side. If yes,
return the air PSOL to
its original position and
replace the oxygen
PSOL.
4. Run the leak test in EST
to check for leaks/
occlusions.
5. Perform a flow sensor
calibration.
6. Replace Q1.
7. Replace the Inspiratory
Electronics PCB.
8. Replace the AI PCB.
AE0111 Unable to EST Flow Alert Flow controller 1. Make sure air supply is
establish air sensors cross unable to establish connected and unre-
flow check Test and control air flow stricted.
at 120 L/min. 2. Verify air regulator
pressure is set between
9-12 psi.
3. Switch PSOLs to see if
failure transfers to the
other gas side. If yes,
return Oxygen PSOL to
its original position and
replace the Air PSOL.
4. Run the leak test in EST
to check for leaks/
occlusions.
5. Perform a flow sensor
calibration.
6. Replace Q2.
7. Replace the Inspiratory
Electronics PCB.
8. Replace the AI PCB.
AE0213 Air PSOL leak EST Gas Alert Air PSOL (PSOL2) 1. Check for leaks around
Supply/SV forward leak was the Q2 o-rings.
Test detected via 2. Make sure air supply is
excessive pressure connected and unre-
buildup stricted.
( > 50 cmH2O but 3. Reseat Air PSOL.
< 100 cmH2O) in 4. Replace Air PSOL.
blocked inspiratory 5. Replace AI PCB.
module.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
AE0214 O2 PSOL leak EST Gas Alert Oxygen PSOL 1. Check for leaks around
Supply/SV (PSOL1) forward the Q1 o-rings.
Test leak was detected 2. Make sure oxygen sup-
via excessive ply is connected and
pressure buildup unrestricted.
( > 50 cmH2O but 3. Reseat oxygen PSOL.
< 100 cmH2O) in 4. Replace oxygen PSOL.
blocked inspiratory 5. Replace AI PCB.
module.
AE0306 Test circuit not EST Leak Test Alert Pressure not 1. Make sure test circuit is
connected detected on properly connected.
expiratory side 2. Replace expiratory bac-
teria filter.
3. Check for leaks around
the Q3 flow sensor.
4. Check/replace the exha-
lation valve.
5. Replace the expiratory
pressure transducer
PCB.
AE0601 GUI High Alarm EST GUI Lamp Alert CLEAR key pressed 1. Check/replace intercon-
LED fails. Test to indicate LED not nect cable between the
on. GUI LED PCB and the
GUI CPU PCB.
2. Replace the GUI LED
PCB.
AE0602 GUI Medium EST GUI Lamp Alert CLEAR key pressed 1. Check/replace intercon-
Alarm LED fails. Test to indicate LED not nect cable between the
on. GUI LED PCB and the
GUI CPU PCB.
2. Replace the GUI LED
PCB.
AE0603 GUI Low Alarm EST GUI Lamp Alert CLEAR key pressed 1. Check/replace intercon-
LED fails. Test to indicate LED not nect cable between the
on. GUI LED PCB and the
GUI CPU PCB.
2. Replace the GUI LED
PCB.
AE0604 GUI Normal LED EST GUI Lamp Alert CLEAR key pressed 1. Check/replace intercon-
fails. Test to indicate LED not nect cable between the
on. GUI LED PCB and the
GUI CPU PCB.
2. Replace the GUI LED
PCB.
AE0605 GUI Batt Backup EST GUI Lamp Alert CLEAR key pressed 1. Check/replace intercon-
LED fails. Test to indicate LED not nect cable between the
on. GUI LED PCB and the
GUI CPU PCB
2. Replace the GUI LED
PCB.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
AE0606 GUI On Batt Pwr EST GUI Lamp Alert CLEAR key pressed 1. Check/replace intercon-
LED fails. Test to indicate LED not nect cable between the
on. GUI LED PCB and the
GUI CPU PCB.
2. Replace the GUI LED
PCB.
AE0607 GUI Compressor EST GUI Lamp Alert CLEAR key pressed 1. Check/replace intercon-
Ready LED fails. Test to indicate LED not nect cable between the
on. GUI LED PCB and the
GUI CPU PCB.
2. Replace the GUI LED
PCB.
AE0608 GUI Compressor EST GUI Lamp Alert CLEAR key pressed 1. Check/replace intercon-
Operating LED Test to indicate LED not nect cable between the
fails. on. GUI LED PCB and the
GUI CPU PCB.
2. Replace the GUI LED
PCB.
AE0609 GUI 100% O2 or EST GUI Lamp Alert CLEAR key pressed 1. Check/replace intercon-
INCREASE O2 2 Test to indicate LED not nect cable between the
min LED fails. on. GUI LED PCB and the
GUI CPU PCB.
2. Replace keyboard.
AE0610 GUI Alarm EST GUI Lamp Alert CLEAR key pressed 1. Check/replace intercon-
Silence LED fails. Test to indicate LED not nect cable between the
on. GUI LED PCB and the
GUI CPU PCB.
2. Replace the GUI LED
PCB.
AE0611 GUI Screen Lock EST GUI Lamp Alert CLEAR key pressed 1. Check/replace intercon-
LED fails. Test to indicate LED not nect cable between the
on. GUI LED PCB and the
GUI CPU PCB.
2. Replace the GUI LED
PCB.
AE0702 Bad Vent inop EST BD Lamp Alert CLEAR key pressed Replace BD LED PCB.
LED Test to indicate one or
both ventilator
inoperative LEDs not
on.
AE0703 Bad SVO LED EST BD Lamp Alert CLEAR key pressed Replace BDU LED PCB.
Test to indicate one or
both SVO LEDs not
on.
AE0704 Bad Loss of GUI EST BD Lamp Alert CLEAR key pressed Replace BDU LED PCB.
LED Test to indicate loss of
GUI LED not on.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
AE1001 Air PSOL EST PSOL Alert Air PSOL (PSOL2) 1. Verify that the air sup-
loopback Loopback loopback current ply is good.
current OOR Test out of range of drive 2. Verify air regulator
current. pressure set to
between 9 and 12 psi.
3. Switch PSOLs to see if
failure transfers to the
other gas side. If yes,
return oxygen PSOL
(PSOL1) to its original
position and replace
PSOL2.
4. Switch Q1 and Q2, run
a flow sensor calibra-
tion and rerun test. If
the problem transfers
to the other gas side,
return Q1 to its original
position and replace
Q2.
5. Replace the AI PCB.
6. Replace the Inspiratory
Electronics PCB.
AE1002 O2 PSOL EST PSOL Alert Oxygen PSOL 1. Verify that the oxygen
loopback Loopback (PSOL1) loopback supply is good.
current OOR Test current out of range 2. Verify oxygen regulator
of drive current. pressure set to
between 9 and 12 psi.
3. Switch PSOLs to see if
failure transfers to the
other gas side. If yes,
return the air PSOL to
its original position and
replace PSOL1.
4. Switch Q1 and Q2, run
a flow sensor calibra-
tion and rerun test. If
the problem transfers
to the other gas side,
return Q2 to its original
position and replace
Q1.
5. Replace the AI PCB.
6. Replace the Inspiratory
Electronics PCB.
AE1104 Insp check valve EST Safety Alert It took too little time 1. Make sure test circuit is
test failed System Test to relieve excess connected.
pressure through 2. Make sure CV3 is not
open safety valve, installed backward.
indicating 3. Replace CV3.
inspiratory check
valve (CV3) may be
damaged or
incorrectly
mounted.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
AE1201 Exp valve EST Exp Valve Alert Exhalation valve 1. Verify that the system
loopback Loopback loopback current is has no leaks or occlu-
current OOR Test out of range of drive sions by running the
current. leak test in EST.
2. Clean exhalation valve
diaphragm.
3. Calibrate the exhala-
tion valve.
4. Clean/replace the exha-
lation valve.
5. Replace the AI PCB.
AE1305 Seal test failed EST Exp Valve Alert Seal test pressure 1. Verify that the system
Seal Test is above alert level has no leaks or occlu-
but below failure sions by running the
level. leak test in EST.
2. Clean the exhalation
valve.
3. Calibrate exhalation
valve.
4. Replace the exhalation
valve.
5. Replace the AI PCB.
AE1600 Compressor Test EST Status Ventilator did not No action required.
- Not installed Compressor sense a compressor
Test attached and
skipped test.
AE1601 Wall air EST Alert Wall air pressure 1. Make sure air supply is
pressure Compressor switch (PS2) disconnected.
detected Test detected air after 2. Disconnect PS2 and
user was prompted rerun test. If test
to disconnect air. passes, replace PS2.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
AE1604 Run mode time EST Alert Compressor timer is Listen for motor. If motor is
OOR Compressor not running while on, replace compressor
Test compressor motor is PCB. Otherwise, replace
on. compressor.
AE1607 Disabled mode EST Alert Compressor timer is Replace compressor PCB.
time OOR Compressor running while
Test compressor motor is
off.
AE1609 Standby mode EST Alert Compressor motor is Replace compressor PCB.
time OOR Compressor still running. During
Test standby mode test
phase, compressor
motor should
eventually turn off.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
AE1702 Wall air EST Alert Wall air pressure 1. Make sure air supply is
pressure Compressor switch (PS2) disconnected.
detected Leak Test detected air 2. Disconnect PS2 and
presence after user rerun test. If test
was prompted to passes, replace PS2.
disconnect air.
AE1703 Wall air EST Alert Wall air pressure 1. Make sure air supply is
pressure not Compressor switch (PS2) did not connected.
detected Leak Test detect air presence 2. Run Gas Supply/SV Test.
after user was
prompted to
connect air.
AE1704 Compressor leak EST Alert Compressor leak was Troubleshoot compressor
detected Compressor detected using compartment or
Leak Test compressor timer to inspiratory module for
detect compressor leaks.
turning on
momentarily during
test.
AE1901 GUI touch: Error EST GUI Alert Touch screen error 1. Clean touch screen,
Touch Test occurred (e.g., removing any obstruc-
blocked beam or tions.
low-level error), 2. Replace touch frame
buffer cannot be PCB.
read, or report is 3. Replace GUI CPU PCB.
invalid.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
AE2001 Bad GUI serial EST GUI Serial Alert While in loopback Replace GUI CPU PCB.
port Port Test mode, failed to
verify received
message was
identical to
transmitted
message.
AE2101 Battery not EST Battery Alert BPS not fully 1. Allow BPS to fully
charged Test charged at start of charge, then repeat
test. test.
2. Replace BPS PCB.
3. Replace battery pack.
4. Replace BD CPU PCB.
AE2102 Battery not EST Battery Alert BPS not discharging 1. Verify AC power is dis-
discharging Test after AC power was connected when
disconnected. prompted.
2. Replace BPS PCB.
3. Replace battery pack.
4. Replace BD CPU PCB.
AE2103 Bad Backup EST Battery Alert While BPS was 1. Ensure the battery pack
Power Supply Test discharging, BPS is fully charged by
voltage dropped checking that the
below accepted green LED on the BPS is
level or dropped too lit prior to initiating an
quickly. EST. If the amber LED is
lit, allow the unit to
charge the batteries
prior to rerunning EST.
2. Replace the battery
pack.
3. Replace the BPS PCB.
4. Replace the AI PCB.
5. Replace the power sup-
ply.
AE2104 Battery not EST Battery Alert BPS not charging 1. Verify that AC power is
charging Test after AC power was reconnected when
reconnected. prompted and that AC
is good.
2. Replace BPS PCB.
3. Replace the battery
pack.
4. Replace the power sup-
ply.
5. Replace the BD CPU.
6. Replace the AI PCB.
AE2300 GUI Nurse Call EST Test Status User pressed CLEAR No action required.
Test - Not to indicate nurse’s
installed call device not
installed. Test was
skipped.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
AS0010 Unable to SST flow Alert Flow controller 1. Make sure oxygen sup-
establish O2 sensor test unable to establish ply is connected and
flow and control oxygen unrestricted.
flow at 120 L/min. 2. Verify that the oxygen
supply meets minimum
pressure requirements.
3. Verify that the patient
circuit system has no
leaks or occlusions. If
not sure, run the first
four tests in EST to get
to the leak test. If the
leak test passes in EST,
exit out of EST. Correct
the leak or occlusion in
the patient circuit used
in SST and rerun or try
another patient circuit.
4. Check the oxygen inlet
filter assembly.
5. Verify that the oxygen
regulator pressure is set
between 9-12 psi.
6. Run a flow sensor cali-
bration.
7. Switch Q1 and Q2.
Rerun the flow sensor
calibration and EST. If
the failure transfers to
the air side, return Q2
back to its original posi-
tion and replace Q1.
8. Switch PSOLs to see if
the failure transfers to
the air side. If yes,
return PSOL2 back to its
original position and
replace PSOL1.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
AS0011 Unable to SST flow Alert 1. Flow controller 1. Make sure air supply is
establish air sensor test unable to estab- connected and unre-
flow lish and control stricted.
air flow at 2. Verify that the air sup-
120 L/min. ply meets minimum
pressure requirements.
3. Verify that the patient
circuit system has no
leaks or occlusions. If
not sure, run the first
four tests in EST to get
to the leak test. If the
leak test passes in EST,
exit out of EST. Correct
the leak or occlusion in
the patient circuit used
in SST and rerun or try
another patient circuit.
4. Check the air inlet filter
(F2).
5. Verify that the air regu-
lator pressure is set
between 9-12 psi.
6. Run a flow sensor cali-
bration.
7. Switch Q1 and Q2.
Rerun the flow sensor
calibration and EST. If
the failure transfers to
the oxygen side, return
Q1 back to its original
position and replace
Q2.
8. Switch PSOLs to see if
the failure transfers to
the oxygen side. If yes,
return PSOL1 back to its
original position and
replace PSOL2.
AS0012 O2 pressure not SST flow Alert Oxygen pressure not Make sure oxygen supply is
detected sensor test detected via PS1. connected.
Only air available for
SST.
AS0013 Wall air SST flow Alert Air pressure not Make sure air supply is
pressure not sensor test detected via PS2 or connected.
detected PC. Only oxygen
available for SST.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
AS0202 Excessive leak SST Circuit Alert Pressure drops to 85 1. Make sure patient cir-
leak cmH2O in 10 s, then cuit is connected and is
in 10 s more by not leaking.
10 cmH2O. 2. Check exhalation valve
operation.
3. Verify connections of
bacteria filters and
humidifier are secure.
AS0311 Low expiratory SST Alert Pressure drop 1. Repeat test, following
filter P Expiratory across expiratory directions more closely.
filter filter < 0.4 cmH2O, 2. Replace filter.
but > 0.1 cmH2O.
AS0403 Occluded SST Circuit Alert Inspiratory limb Check for occluded patient
inspiratory limb Resistance pressure > 8.5 tubing.
cmH2O (adult) or
5.5 cmH2O
(pediatric), but
< 12.5 cmH2O (adult)
or 7.5 cmH2O
(pediatric),
indicating occlusion.
AS0406 Occluded SST Circuit Alert Expiratory limb Check for occluded patient
exhalation limb Resistance pressure > tubing.
8.5 cmH2O (adult) or
5.5 cmH2O
(pediatric) but > 12.5
cmH2O (adult) or
7.5 cmH2O
(pediatric),
indicating occlusion.
AS0407 Unable to reach SST Circuit Alert During Check for kinked or
min peak flow Resistance characterization of occluded patient tubing.
total circuit
resistance over a
range of flows, peak
flow < 80 L/min but
> 60 L/min.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
AS0411 Unable to reach SST Circuit Alert During Check for kinked or
min peak flow Resistance characterization of occluded patient tubing.
total circuit
resistance over a
range of flows, peak
flow was less than
alert threshold.
AS0413 Insp limb SST Circuit Alert Inspiratory limb 1. Make sure inspiratory
resistance low Resistance pressure < 0.6 filter is installed.
cmH2O (adult) or 2. Repeat test, following
5.5 cmH2O directions more closely.
(pediatric), but 3. Replace inspiratory fil-
> 0.2 cmH2O. ter.
4. Replace patient circuit.
AS0414 Exp limb SST Circuit Alert Expiratory limb Replace patient circuit.
resistance low Resistance pressure
< 0.6 cmH2O (adult)
or 0.5 cmH2O
(pediatric), but
> 0.2 cmH2O.
AS0507 Compliance low SST Alert High compliance 1. Make sure correct
Compliance falls below patient tubing type
calibration 1.56 mL/cmH2O was specified in SST.
(adult) or 2. Replace patient circuit
1.34 mL/cmH2O with a known good cir-
(pediatric) but cuit and filter set.
not lower than 3. Run EST to check the
1.05 mL/cmH2O. operation of PI and PE
during the Circuit Pres-
sure Test. Replace appli-
cable pressure
transducer.
4. Run Atmospheric Pres-
sure Transducer calibra-
tion.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
AS0511 Compliance low SST Alert Low compliance 1. Make sure correct
Compliance falls below patient tubing type
calibration 1.56 mL/cmH2O was specified in SST.
(adult) or 2. Replace patient circuit
1.34 mL/cmH2O with a known good cir-
(pediatric), but cuit and filter set.
not below 3. Run EST to check the
1.05 mL/cmH2O. operation of PI and PE
during the Circuit Pres-
sure Test. Replace appli-
cable pressure
transducer.
4. Run Atmospheric Pres-
sure Transducer calibra-
tion.
Dxxxxx Assertion Background BDU reset System generated a 1. See specific error code.
Checks (BD) reset to correct a 2. Replace the BD CPU
boundary check or PCB.
possible data
corruption of
control variables.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
DT0002 Bus error / Background BDU reset Hardware trap from Replace the BD CPU PCB.
Access fault Checks (BD) an access fault due
to a bad memory
chip, bad control
line, or a hardware
timing issue.
DT0011 Line 1111 Background BDU reset Replace the BD CPU PCB.
Emulator Checks (BD)
(Unimplemen-
ted F-line
Opcode)
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
NOTE:
For all "FE" codes (failures reported during EST), diagnose the problem and perform the repair.
Run a complete EST to reset the EST test failure in memory. If there is a failed EST test result in
memory, normal operation is prevented until the failed EST test is rerun and passes.
FE0003 Failed to reach EST Circuit Failure Unable to build 1. Make sure that test cir-
test pressure Pressure Test pressure (air) to cuit is installed and air
10 cmH2O at 5 L/min supply or compressor is
within time-out available and good.
period. 2. Verify that the air regu-
lator pressure is set
between 9-12 psi.
3. Verify that the patient
circuit system has no
leaks or occlusions by
running the first four
tests in EST to get to
the leak test.
FE0004 Cross-check EST Circuit Failure Inspiratory/ 1. Verify that the system
failed Pressure Test expiratory pressure has no leaks or occlu-
transducer readings sions. If not sure, run
at 10 cmH2O test the first four tests in
pressure are too far EST to run the leak test
apart. 2. If the leak test passes in
EST, exit out of EST.
Review the data for the
Circuit Pressure Test to
determine which pres-
sure transducer is out
of range: PI or PE.
Replace applicable
pressure transducer.
3. Replace AI PCB.
FE0005 Bad insp EST Circuit Failure Inspiratory pressure 1. Replace SOL1.
autozero Pressure Test reading (taken after 2. Replace Inspiratory
solenoid inspiratory pressure Electronics PCB.
transducer autozero 3. Replace AI PCB.
solenoid (SOL1)
actuated) out of
range (-0.60 to
0.60 cmH2O).
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
FE0006 Bad exp EST Circuit Failure Expiratory pressure 1. Replace SOL2.
autozero sol Pressure Test reading (taken after 2. Replace exhalation
expiratory pressure transducer PCB.
transducer autozero 3. Replace AI PCB.
solenoid (SOL2)
actuated) out of
range (-0.60 to
0.60 cmH2O).
FE0008 Failed to reach EST Circuit Failure Unable to build 1. Make sure air supply or
test pressure Pressure Test pressure (air) to 50 compressor is available.
or 100 cmH2O at 2. Check for system leak
5 L/min within time- at the expiratory bacte-
out period. ria filter or O2 sensor. If
not sure, run the first
four tests in EST to run
the leak test.
FE0010 AC power not EST Circuit Failure System is still 1. Plug in ventilator
connected Pressure Test running on battery power cord.
power after 2. Disconnect BPS to iso-
prompting user to late problem.
connect AC power. 3. Replace power supply.
EST can only run on
AC (facility) power.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
FE0101 O2 flow sensor EST Flow Failure Oxygen flow sensor 1. Make sure oxygen sup-
cross check sensors cross (Q1) cross-check ply is connected and
failed check Test failed. unrestricted.
2. Verify that the oxygen
supply is good.
3. Verify that the system
has no leaks or occlu-
sions. If not sure, run
the first four tests in
EST to run the leak test.
4. Run a flow sensor cali-
bration.6.Switch Q1
and Q2. Rerun the flow
sensor calibration and
EST. If the failure trans-
fers to the air side,
return Q2 back to its
original position and
replace Q1.
5. Switch PSOLs to see if
the failure transfers to
the air side. If yes,
return PSOL2 back to its
original position and
replace PSOL1.
6. Replace exhalation
flow sensor (Q3)
7. Replace AI PCB.
FE0102 O2 PSOL current EST Flow Failure Oxygen PSOL 1. Make sure oxygen sup-
out of range sensors cross (PSOL1) current is ply is connected and
check Test out of range with unrestricted.
respect to flow 2. Verify that the oxygen
sensor (Q1). supply is good.
3. Verify that the system
has no leaks or occlu-
sions. If not sure, run
the first four tests in
EST to run the leak test.
4. Switch PSOLs to see if
the failure transfers to
the air side. If yes,
return PSOL2 back to its
original position and
replace PSOL1.
5. Run a flow sensor cali-
bration.
6. Switch Q1 and Q2.
Rerun the flow sensor
calibration and EST. If
the failure transfers to
the air side, return Q2
back to its original posi-
tion and replace Q1.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
FE0103 Air flow sensor EST Flow Failure Inspiratory module 1. Make sure air supply is
cross check sensors cross air flow sensor (Q2) connected and unre-
failed check Test cross-check failed. stricted.
2. Verify that the air sup-
ply is good.
3. Verify that the system
has no leaks or occlu-
sions. If not sure, run
the first four tests in
EST to run the leak test.
4. Run a flow sensor cali-
bration.
5. Switch Q1 and Q2.
Rerun the flow sensor
calibration and EST. If
the failure transfers to
the oxygen side, return
Q1 back to its original
position and replace
Q2.
6. Switch PSOLs to see if
the failure transfers to
the air side. If yes,
return PSOL back to its
original position and
replace PSOL2
7. Replace exhalation
flow sensor (Q3).
8. Replace AI PCB.
FE0104 Air PSOL current EST Flow Failure Air PSOL (PSOL2) 1. Make sure air supply is
out of range sensors cross current is out of connected and unre-
check Test range with respect stricted.
to air flow sensor 2. Verify that the air sup-
(Q2). ply is good.
3. Verify that the system
has no leaks or occlu-
sions. If not sure, run
the first four tests in
EST to run the leak test.
4. Switch PSOLs to see if
the failure transfers to
the oxygen side. If yes,
return PSOL back to its
original position and
replace PSOL2.
5. Run a flow sensor cali-
bration.
6. Switch Q1 and Q2.
Rerun the flow sensor
calibration and EST. If
the failure transfers to
the oxygen side, return
Q1 back to its original
position and replace
Q2.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
FE0106 Unable to EST Flow Failure Flow controller 1. Make sure oxygen sup-
establish O2 sensors cross unable to establish ply is connected.
flow check Test and control oxygen 2. Check regulated oxy-
flow at 60, 5, and 1 gen pressure.
L/min. 3. Replace PSOL1 or Q1.
4. Replace AI PCB.
FE0107 Unable to EST Flow Failure Flow controller 1. Make sure air supply is
establish air sensors cross unable to establish connected.
flow check Test and control air flow 2. Check regulated air
at 60, 5, and 1 L/min. pressure.
3. Perform flow sensor
calibration.
4. Replace PSOL2 or Q2.
5. Replace AI PCB.
FE0108 O2 zero flow EST Flow Failure Inspiratory flow > 1. Verify no leaks at the
check failed sensors cross 0.153 L/min with Q1 o-rings.
check Test oxygen PSOL 2. Run EST Gas Supply Test
(PSOL1) commanded to check for PSOL1
to 0 (closed). leak. Remove and then
reseat PSOL1.
3. Replace PSOL1.
4. Perform a flow sensor
calibration.
5. Replace Q1.
FE0109 Air zero flow EST Flow Failure Inspiratory flow 1. Verify no leaks at the
check failed sensors cross > 0.153 L/min with Q2 o-rings.
check Test air PSOL (PSOL2) 2. Run EST Gas Supply Test
commanded to 0 to check for PSOL2
(closed). leak. Remove and then
reseat PSOL2.
3. Replace PSOL2.
4. Perform a flow sensor
calibration.
5. Replace Q2.
FE0204 Wall air EST Gas Failure Wall air not 1. Make sure air supply is
pressure not Supply/SV detected initially connected.
detected Test and wall air still not 2. Check PS2: Remove the
detected after user wiring connectors from
was prompted to PS2 and jumper the
connect wall air. connectors to each
other. If the system
now recognizes the air
supply, replace PS2.
FE0205 O2 pressure not EST Gas Failure Oxygen pressure not 1. Make sure oxygen sup-
detected Supply/SV detected initially ply is connected.
Test and oxygen pressure 2. Check PS1: Remove the
still not detected wiring connectors from
after user was PS1 and jumper the
prompted to connectors to each
connect oxygen. other. If the system
now recognizes the air
supply, replace PS1.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
FE0206 O2 pressure EST Gas Failure Oxygen pressure 1. Make sure oxygen sup-
detected (O2 Supply/SV switch detected ply is disconnected.
not Test presence of oxygen 2. Replace Inspiratory
disconnected) after user was Electronics PCB.
prompted to 3. Replace PS1.
disconnect oxygen.
FE0207 Air PSOL leak EST Gas Failure Air PSOL (PSOL2) 1. Check for leaks around
Supply/SV forward leak was the Q2 o-rings.
Test detected via 2. Verify that the safety
excessive pressure valve relieves pressures
buildup (> 100 above 100 cmH2O. Lis-
cmH2O) in blocked ten for the pressure
inspiratory module. relief while watching
the numeric digital dis-
play of system pressure
in the upper screen. If
the safety valve does
not crack open, replace
the safety valve.
3. Replace PSOL2.
FE0208 Wall air EST Gas Failure Wall air pressure 1. Make sure air supply is
pressure Supply/SV switch (PS2) disconnected.
detected Test detected wall air 2. Replace Inspiratory
after user was Electronics PCB.
prompted to 3. Replace PS2.
disconnect it.
FE0209 O2 PSOL leak EST Gas Failure Oxygen PSOL 1. Check for leaks around
Supply/SV (PSOL1) forward the Q1 o-rings.
Test leak was detected 2. Verify that the safety
via excessive valve relieves pressures
pressure buildup above 100 cmH2O. Lis-
(> 100 cmH2O) in ten for the pressure
blocked inspiratory relief while watching
module. the numeric digital dis-
play of system pressure
in the upper screen. If
the safety valve does
not crack open, replace
the safety valve.
3. Replace PSOL1.
FE0210 SV pressure EST Gas Failure Safety valve cracking 1. Make sure To patient
relief failed Supply/SV pressure and/or port is blocked.
Test peak steady-state 2. Check for leaks at the
pressure is out of O2 sensor, SOL1 or PI.
range. 3. Replace the safety
valve.
4. Replace PI.
FE0211 O2 pressure not EST Gas Failure PS1 did not detect 1. Make sure oxygen sup-
detected Supply/SV oxygen after user ply is connected.
Test was prompted to 2. Replace Inspiratory
connect it. Electronics PCB.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
FE0212 Compressor EST Gas Failure Compressor pressure Replace compressor PCB.
pressure Supply/SV transducer (PC)
detected Test detected that
compressor was
pressurized.
FE0215 Air zero flow EST Gas Failure During zero-flow 1. Make sure gas supplies
check failed Supply/SV check, air flow are disconnected.
Test sensor (Q2) reads 2. Make sure compressor
> 0.05 L/min. is not running.
3. Verify no leaks at the
Q2 o-rings.
4. Check for PSOL2 leak.
5. Remove and then
reseat PSOL2.
6. Perform a flow sensor
calibration.
7. Replace PSOL2.
8. Replace Q2.
FE0216 O2 zero flow EST Gas Failure During zero-flow 1. Make sure gas supplies
check failed Supply/SV check, oxygen flow are disconnected.
Test sensor (Q1) reads 2. Make sure compressor
> 0.05 L/min. is not running.
3. Verify no leaks at the
Q2 o-rings.
4. Check for PSOL2 leak.
5. Remove and then
reseat PSOL2.
6. Perform a flow sensor
calibration.
7. Replace PSOL2.
8. Replace Q2.
FE0217 Exp zero flow EST Gas Failure During zero-flow 1. Make sure gas supplies
check failed Supply/SV check, exhalation are disconnected.
Test flow sensor (Q3) 2. Make sure compressor
reads > 0.1 L/min. is not running.
3. Verify no leaks at the
Q3 o-rings.
4. Perform a flow sensor
calibration.
5. Replace Q3.
FE0301 Excessive leak EST Leak Test Failure System pressure 1. Make sure test circuit is
dropped below connected and is not
failure pressure leaking.
level. 2. Check for system leaks
or occlusions especially
at the O2 sensor and
expiratory filter.
3. Verify secure connec-
tion of expiratory filter.
4. Check exhalation valve
operation.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
FE0305 Unable to EST Leak Test Failure System cannot 1. Make sure test circuit is
establish attain leak test connected and is not
pressure starting pressure leaking.
using oxygen or air 2. Check exhalation valve
within time-out operation, and verify
period. secure connection of
expiratory filter.
FE0401 Accept key fails. EST GUI Failure Wrong key pressed 1. Repeat test.
Keyboard or key not pressed 2. Replace keyboard.
Test within 15 s.
FE0402 Clear key fails. EST GUI Failure Wrong key pressed 1. Repeat test.
Keyboard or key not pressed 2. Replace keyboard.
Test within 15 s.
FE0403 Insp. Pause key EST GUI Failure Wrong key pressed 1. Repeat test.
fails. Keyboard or key not pressed 2. Replace keyboard.
Test within 15 s.
FE0404 Exp. Pause key EST GUI Failure Wrong key pressed 1. Repeat test.
fails. Keyboard or key not pressed 2. Replace keyboard.
Test within 15 s.
FE0405 Man Insp fails. EST GUI Failure Wrong key pressed 1. Repeat test.
Keyboard or key not pressed 2. Replace keyboard.
Test within 15 s.
FE0406 100% O2/CAL EST GUI Failure Wrong key pressed 1. Repeat test.
key or INCREASE Keyboard or key not pressed 2. Replace keyboard.
O2 2 min key Test within 15 s.
fails.
FE0407 Info key fails. EST GUI Failure Wrong key pressed 1. Repeat test.
Keyboard or key not pressed 2. Replace keyboard.
Test within 15 s.
FE0408 Alarm Reset key EST GUI Failure Wrong key pressed 1. Repeat test.
fails. Keyboard or key not pressed 2. Replace keyboard.
Test within 15 s.
FE0409 Alarm Silence EST GUI Failure Wrong key pressed 1. Repeat test.
key fails. Keyboard or key not pressed 2. Replace keyboard.
Test within 15 s.
FE0410 Alarm Volume EST GUI Failure Wrong key pressed 1. Repeat test.
key fails. Keyboard or key not pressed 2. Replace keyboard.
Test within 15 s.
FE0411 Screen EST GUI Failure Wrong key pressed 1. Repeat test.
brightness fails. Keyboard or key not pressed 2. Replace keyboard.
Test within 15 s.
FE0412 Screen contrast EST GUI Failure Wrong key pressed 1. Repeat test.
key fails. Keyboard or key not pressed 2. Replace keyboard.
Test within 15 s.
FE0413 Screen lock key EST GUI Failure Wrong key pressed 1. Repeat test.
fails. Keyboard or key not pressed 2. Replace keyboard.
Test within 15 s.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
FE0501 Bad knob EST GUI Knob Failure Knob was not 1. Repeat test.
Test turned in direction 2. Replace keyboard.
as prompted within
15 s.
FE0801 SAAS EST GUI Failure CLEAR key pressed 1. Verify the GUI alarm
(Safety Audible Audio Test to indicate GUI cable connection to the
Alarm System) audio diagnostic GUI CPU PCB.
test failed failed. 2. Replace GUI alarm
assembly.
3. Replace GUI CPU PCB.
FE0901 Bad alarm cable EST BDU Failure Alarm cable voltage 1. Make sure BD alarm
Audio Test is out of range (< 3.5 cable is connected.
or > 5.05 V). 2. Replace BD alarm.
3. Replace AI PCB.
4. Replace BD alarm
cable.
FE0902 Bad power fail EST BDU Failure Power failure 1. Make sure BD alarm
cap Audio Test capacitor initial cable is connected.
voltage is out of 2. Replace BD alarm.
range (< 4.5 or 3. Replace AI PCB.
> 5.05 V). 4. Replace BD alarm
cable.
5. Replace Motherboard
PCB.
FE0903 Bad power fail EST BDU Failure Power failure 1. Make sure BD alarm
cap Audio Test capacitor final cable is connected.
voltage is out of 2. Replace BD alarm.
range or RC constant 3. Replace AI PCB.
< 60 s.
FE0904 Bad BD audio EST BDU Failure CLEAR key pressed 1. Make sure BD alarm
Audio Test to indicate user did cable is connected.
not hear alarm, 2. Replace BD alarm.
although alarm was 3. Replace AI PCB.
active. 4. Replace BD alarm
cable.
FE1101 Safety valve EST Safety Failure Excessive safety Replace safety valve.
occluded System Test valve back pressure
when safety valve is
open.
FE1102 Bad safety valve EST Safety Failure Safety valve 1. Replace safety valve.
driver or System Test loopback current is 2. Replace AI PCB.
loopback out of range during 3. Replace Inspiratory
one or more of Electronics PCB.
timed test points.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
FE1103 Insp check valve EST Safety Failure It took too long to 1. Make sure proper test
test failed System Test relieve excess circuit is used.
pressure through 2. Make sure CV3 is not
open safety valve, installed backward.
indicating 3. Replace CV3.
inspiratory check 4. Replace Safety Valve
valve (CV3) is (SV).
occluded or test
circuit is too large.
FE1105 Unable to EST Safety Failure Flow controller 1. Make sure air supply is
establish flow System Test unable to establish connected and unre-
and control gas flow stricted.
at 60 L/min. 2. Check air regulator
pressure.
3. Run flow sensor cali-
bration.
4. Switch PSOLs to see if
test passes. If it does,
return PSOL1 to its
original position and
replace PSOL2.
5. Replace Q2.
FE1301 Seal test failed EST Exp Valve Failure Seal test pressure 1. Clean exhalation valve.
Seal Test is above failure 2. Calibrate exhalation
level. valve.
3. Run flow sensor cali-
bration.
4. Replace exhalation
valve.
5. Replace AI PCB.
6. Replace PE.
FE1302 Exp valve temp EST Exp Valve Failure Exhalation valve 1. Ensure the unit has
OOR Seal Test magnet been warmed up at
temperature out of ambient temperature
range (10 to 100oC). for at least 10 minutes.
2. Calibrate the exhala-
tion valve (EV).
3. Replace exhalation
valve (EV).
4. Replace AI PCB.
FE1303 Unable to EST Exp Valve Failure Flow controller 1. Make sure proper test
establish exp Seal Test unable to establish circuit it used and that
flow and control air flow there are no leaks or
measured by occlusions.
exhalation flow 2. Make sure air supply is
sensor (Q3). still connected.
3. Run flow sensor cali-
bration.
4. Replace Q3.
5. Replace Q2.
6. Replace the exhalation
valve.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
FE1304 Exp valve not EST Exp Valve Failure Exhalation valve 1. Check for leaks or
calibrated Seal Test table checksum is occlusions. Correct and
not valid or last then run the exhalation
calibration valve calibration.
performed was not 2. Replace the exhalation
completed valve.
successfully. 3. Replace the AI PCB.
4. Replace Q3.
5. Replace Q2.
FE1401 Bad calibration EST Exp Valve Failure Measured system 1. Calibrate exhalation
Test pressure at one or valve.
more test points is 2. Replace exhalation
out of range. valve.
3. Replace AI PCB.
4. Replace the Exhalation
Pressure Transducer
PCB.
FE1402 Exp valve not EST Exp Valve Failure Exhalation valve 1. Check for leaks or
calibrated Test table checksum is occlusions. Correct and
not valid or last then run the exhalation
calibration valve calibration.
performed was not 2. Replace the exhalation
completed valve.
successfully. 3. Replace the AI PCB.
4. Replace Q3.
5. Replace Q2.
FE1403 Unable to EST Exp Valve Failure Flow controller 1. Check for leaks or
establish exp Test unable to establish occlusions. Correct and
flow and control air flow then run the exhalation
measured by valve calibration.
exhalation flow 2. Replace the exhalation
sensor (Q3). valve.
3. Replace the AI PCB.
4. Replace Q3.
5. Replace Q2.
FE1501 Unable to EST Exp Failure Flow controller 1. Make sure air supply is
establish air Heater Test unable to establish connected and unre-
flow and control 60 L/min stricted.
air flow. 2. Check air regulator
pressure.
3. Run flow sensor cali-
bration.
4. Switch PSOLs to see if
test passes. If it does,
return PSOL1 to its
original position and
replace PSOL2.
5. Replace Q2.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
FE1502 Bad exp heater EST Exp Failure Exhalation heater 1. Replace exhalation
Heater Test temperature did not heater.
rise sufficiently 2. Replace AI PCB.
when heater was 3. Replace interconnect
turned on. cable between exhala-
tion compartment and
motherboard.
FE1503 Bad exp heater EST Exp Failure Heater temperature 1. Replace exhalation
Heater Test did not drop heater.
sufficiently during 2. Replace AI PCB.
an interval after
heater was turned
off.
FE2201 Low exp P EV Velocity Failure Expiratory P too 1. Check for leaks or
Transducer low, indicating occlusions. Correct and
Test velocity transducer then run the exhalation
not responding valve calibration.
properly. 2. Run flow sensor cali-
bration.
3. Replace exhalation
valve.
4. Replace Q3.
5. Replace Q2.
FE2301 Nurse call stuck EST GUI Nurse Failure CLEAR key pressed 1. Make sure test equip-
on Call Test to indicate nurse’s ment is connected
call (remote alarm) properly to remote
relay is stuck on alarm port.
when it should be 2. Replace GUI CPU PCB.
off.
FE2302 Nurse call stuck EST GUI Nurse Failure CLEAR key pressed 1. Make sure test equip-
off Call Test to indicate nurse’s ment is connected
call (remote alarm) properly to remote
relay is stuck off alarm port.
when it should be 2. Replace GUI CPU PCB.
on.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
NOTE:
For all "FS" codes (failures reported during SST), if the corrective action indicated is to run an EST
to further diagnose the problem, perform the repair after the EST, verify the fix in EST, and then
rerun SST in order to enter normal ventilator operation. If there is a failed SST test result in
memory, normal operation until is prevented until the failed SST test is rerun and passes.
FS0001 O2 flow sensor SST Flow Failure Oxygen flow sensor 1. Correct the leak or
cross check Sensor Test (Q1) cross-check occlusion in the SST
failed failed. patient circuit and
rerun or try known
good patient circuit
and filter set.
2. Verify that the ventila-
tor system has no leaks
or occlusions by run-
ning the first four tests
in EST to run the leak
test.
a. If the leak test fails
in EST, troubleshoot
the expiratory filter
or ventilator system
for leaks or occlu-
sions.
b. If the leak test passes
in EST, troubleshoot
the SST patient cir-
cuit.
3. Run a flow sensor cali-
bration.
4. Switch Q1 and Q2.
Rerun the flow sensor
calibration and EST. If
the failure transfers to
the air side, return Q2
back to its original posi-
tion and replace Q1.
5. Replace Q3
6. Switch PSOLs to see if
the failure transfers to
the air side. If yes,
return PSOL2 back to its
original position and
replace PSOL1.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
FS0002 O2 PSOL current SST Flow Failure Oxygen PSOL 1. Make sure oxygen sup-
out of range Sensor Test (PSOL1) command ply is connected and
current is out of unrestricted.
range with respect 2. Verify that the oxygen
to oxygen flow supply is good.
sensor (Q1). 3. Verify that the system
has no leaks or occlu-
sions. If not sure, run
the first four tests in
EST to run the leak test.
4. Run a flow sensor cali-
bration.
5. Switch Q1 and Q2.
Rerun the flow sensor
calibration and EST. If
the failure transfers to
the air side, return Q2
back to its original posi-
tion and replace Q1.
6. Replace Q3.
7. Switch PSOLs to see if
the failure transfers to
the air side. If yes,
return PSOL2 back to its
original position and
replace PSOL1.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
FS0003 Air flow sensor SST Flow Failure Air flow sensor (Q2) 1. Correct the leak or
cross check Sensor Test cross-check failed. occlusion in the SST
failed patient circuit and
rerun or try known
good patient circuit
and filter set.
2. Verify that the ventila-
tor system has no leaks
or occlusions by run-
ning the first four tests
in EST to run the leak
test.
3. If the leak test fails in
EST, troubleshoot the
expiratory filter or ven-
tilator system for leaks
or occlusions.
4. If the leak test passes in
EST, troubleshoot the
SST patient circuit.
5. Run a flow sensor cali-
bration.
6. Switch Q1 and Q2.
Rerun the flow sensor
calibration and EST. If
the failure transfers to
the oxygen side, return
Q1 back to its original
position and replace
Q2.
7. Replace Q3.
8. Switch PSOLs to see if
the failure transfers to
the oxygen side. If yes,
return PSOL1 back to its
original position and
replace PSOL2.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
FS0004 Air PSOL current SST Flow Failure Air PSOL (PSOL2) 1. Make sure air supply is
out of range Sensor Test command current is connected and unre-
out of range with stricted.
respect to air flow 2. Verify that the air sup-
sensor (Q2). ply is good.
3. Verify that the system
has no leaks or occlu-
sions. If not sure, run
the first four tests in
EST to run the leak test.
4. Run a flow sensor cali-
bration.
5. Switch Q1 and Q2.
Rerun the flow sensor
calibration and EST. If
the failure transfers to
the oxygen side, return
Q1 back to its original
position and replace
Q2.
6. Replace Q3.
7. Switch PSOLs to see if
the failure transfers to
the oxygen side. If yes,
return PSOL1 back to its
original position and
replace PSOL2.
FS0005 Gas not SST Flow Failure No gas is available Connect gas supply.
connected Sensor Test (neither PS1, PS2,
nor PC detects
pressure).
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
FS0006 Unable to SST Flow Failure Flow controller 1. Make sure oxygen sup-
establish O2 Sensor Test unable to establish ply is connected and
flow and control oxygen unrestricted.
flow at 60, 5, and 1 2. Verify that the oxygen
L/min. supply is good.
3. Check regulated oxy-
gen pressure.
4. Verify that the system
has no leaks or occlu-
sions. If not sure, run
the first four tests in
EST to run the leak test.
5. Run a flow sensor cali-
bration.
6. Switch Q1 and Q2.
Rerun the flow sensor
calibration and EST. If
the failure transfers to
the air side, return Q2
back to its original posi-
tion and replace Q1.
7. Replace exhalation
flow sensor (Q3).
8. Switch PSOLs to see if
the failure transfers to
the air side. If yes,
return PSOL2 back to its
original position and
replace PSOL1.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
FS0007 Unable to SST Flow Failure Flow controller 1. Make sure air supply is
establish air Sensor Test unable to establish connected unrestricted.
flow and control air flow 2. Verify that the air sup-
at 60, 5, and 1 L/min. ply is adequate.
3. Check regulated air
pressure.
4. Verify that the system
has no leaks or occlu-
sions. If not sure, run
the first four tests in
EST to run the leak test.
5. Run a flow sensor cali-
bration.
6. Switch Q1 and Q2.
Rerun the flow sensor
calibration and EST. If
the failure transfers to
the oxygen side, return
Q1 back to its original
position and replace
Q2.
7. Replace exhalation
flow sensor (Q3).
8. Switch PSOLs to see if
the failure transfers to
the oxygen side. If yes,
return PSOL1 back to its
original position and
replace PSOL2.
FS0008 O2 zero flow SST Flow Failure Inspiratory flow 1. Verify no leaks at the
check failed Sensor Test > 0.153 L/min with Q1 o-rings.
oxygen PSOL 2. Run EST Gas Supply Test
(PSOL1) commanded to check for PSOL1
to 0 (closed). leak.
3. Remove and then
reseat PSOL1.
4. Replace PSOL1.
5. Perform a flow sensor
calibration.
6. Replace Q1.
FS0009 Air zero flow SST Flow Failure Inspiratory flow 1. Verify no leaks at the
check failed Sensor Test > 0.153 L/min with Q2 o-rings.
air PSOL (PSOL2) 2. Run EST Gas Supply Test
commanded to 0 to check for PSOL2
(closed). leak.
3. Remove and then
reseat PSOL2.
4. Replace PSOL2.
5. Perform a flow sensor
calibration.
6. Replace Q2.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
FS0103 Failed to reach SST circuit Failure Unable to build 1. Verify that the patient
test pressure pressure test pressure (air) to circuit system has no
10 cmH2O at 5 L/min leaks or occlusions or
within time-out use a known good
period. patient circuit and filter
set.
2. Verify that the air sup-
ply or compressor is
available and ade-
quate.
3. Verify that the ventila-
tor system has no leaks
or occlusions. If not
sure, run the first four
tests in EST to run the
leak test.
FS0104 Cross-check SST circuit Failure Inspiratory/ 1. Verify that the patient
failed pressure test expiratory pressure circuit system has no
transducer readings leaks or occlusions or
at 10 cmH20 test use a known good
pressure are too far patient circuit and filter
apart. set.
2. Verify that the ventila-
tor system has no leaks
or occlusions. If not
sure, run the first four
tests in EST to run the
leak test.
3. If the leak test passes in
EST, exit out of EST.
Review the data for the
Circuit Pressure Test to
determine which pres-
sure transducer is out
of range: PI or PE.
Replace applicable
pressure transducer.
4. Replace AI PCB.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
FS0105 Bad insp SST circuit Failure Inspiratory pressure 1. Replace SOL1.
autozero pressure test reading taken after 2. Replace Inspiratory
solenoid actuating Electronics PCB.
inspiratory pressure 3. Replace AI PCB.
transducer autozero
solenoid (SOL1) is
out of range
(-0.60 to 0.60
cmH2O).
FS0106 Bad exp SST circuit Failure Expiratory pressure 1. Replace SOL2.
autozero sol pressure test reading taken after 2. Replace exhalation
actuating expiratory transducer PCB.
transducer autozero 3. Replace AI PCB.
solenoid (SOL2) is
out of range (-0.60
to 0.60 cmH2O).
FS0108 Failed to reach SST circuit Failure Unable to build 1. Make sure air supply or
test pressure pressure test pressure (air) to 50 compressor is available.
or 100 cmH2O at 2. Check for system leak.
5 L/min within time-
out period.
FS0201 Excessive leak SST Circuit Failure System pressure 1. Make sure patient cir-
leak dropped below cuit is connected and is
failure pressure not leaking.
level. 2. Check exhalation valve
operation, and verify
secure connections of
bacteria filters, water
traps, and humidifier.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
FS0205 Unable to SST Circuit Failure Unable to attain 1. Make sure patient cir-
establish leak leak test starting cuit is connected and is
pressure pressure within not leaking.
time-out period 2. Check exhalation valve
using oxygen or air. operation, and verify
secure connections of
bacteria filters, water
traps, and humidifier.
FS0301 Unable to SST Failure Flow controller Make sure gas supply is still
establish flow Expiratory unable to establish connected.
filter and control gas
flow.
FS0304 Occluded SST Failure Pressure drop across Replace expiratory filter.
expiratory filter Expiratory filter > 3 cmH2O.
filter
FS0306 Patient circuit SST Failure A test flow was used Make sure patient circuit is
not Expiratory to verify that circuit disconnected at expiratory
disconnected filter was disconnected, filter.
and an expiratory
flow was detected.
FS0307 Patient circuit SST Failure A test flow was used Make sure patient circuit is
not reconnected Expiratory to verify that circuit connected at expiratory
filter was reconnected, filter.
and an expiratory
flow was not
detected.
FS0310 Low expiratory SST Failure Pressure drop across 1. Repeat test, following
filter P Expiratory expiratory filter directions more closely.
filter < 0.1 cmH2O. 2. Replace filter.
FS0401 Unable to SST Circuit Failure Flow controller Make sure gas supply is
establish flow Resistance unable to establish connected.
and control gas
flow.
FS0402 Occluded SST Circuit Failure Inspiratory limb Check for occluded patient
inspiratory limb Resistance pressure tubing.
> 12.5 cmH2O
(adult) or 7.5 cmH2O
(pediatric),
indicating occlusion.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
FS0404 Unable to reach SST Circuit Failure During Check for kinked or
min peak flow Resistance characterization of occluded patient tubing.
total circuit
resistance over a
range of flows, peak
flow was less than
failure threshold.
FS0405 Occluded SST Circuit Failure Expiratory limb Check for occluded patient
exhalation limb Resistance pressure tubing.
> 12.5 cmH2O
(adult) or 7.5 cmH2O
(pediatric),
indicating occlusion.
FS0408 Insp limb SST Circuit Failure Inspiratory limb 1. Make sure inspiratory
resistance low Resistance pressure filter is installed.
< 0.2 cmH2O. 2. Repeat test, following
directions more closely.
3. Replace inspiratory fil-
ter.
4. Replace patient circuit.
FS0409 Exp limb SST Circuit Failure Expiratory limb Replace patient circuit.
resistance low Resistance pressure
< 0.2 cmH2O.
FS0410 Unable to reach SST Circuit Failure During Check for kinked or
min peak flow Resistance characterization of occluded patient tubing.
total circuit
resistance over a
range of flows, peak
flow was less than
failure threshold.
FS0412 Wye not SST Circuit Failure Unable to pressurize 1. Block wye when
blocked Resistance system within time- prompted.
out period. 2. Check for circuit leaks.
FS0501 Unable to SST Failure System cannot Check patient circuit for
pressurize Compliance pressurize to one of leaks. Replace as necessary.
circuit calibration the test pressure
points.
FS0502 Unable to SST Failure System cannot Check patient circuit for
pressurize Compliance pressurize to one of leaks. Replace as necessary.
circuit calibration the test pressure
points.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
FS0506 Compliance low SST Failure High compliance 1. Make sure correct
Compliance falls below patient tubing type
calibration 1.05 mL/cmH2O was specified in SST.
2. Replace patient circuit.
FS0510 Compliance low SST Failure Low compliance 1. Make sure correct
Compliance falls below patient tubing type
calibration 1.05 mL/cmH2O was specified in SST.
2. Replace patient circuit.
HN0002 NMI: Ethernet Other (BDU) BDU reset Replace BD CPU PCB.
Parity Error
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
HN0129
NMI: DRAM Other (BDU) BDU reset 1. Check GUI/BD cable
Parity Error connections.
2. Replace power supply.
3. Replace BD CPU PCB.
HN0130 NMI: Ethernet Other (BDU) BDU reset 1. Check GUI/BD cable
Parity Error connections.
2. Replace power supply.
3. Replace BD CPU PCB.
KB0001 Bad safety valve Background Vent inop Measured voltage 1. Replace safety valve.
switched side check (BDU) on safety valve 2. Replace AI PCB.
switched side 3. Replace Inspiratory
indicates valve is not Electronics PCB.
in expected state. 4. Replace interconnect
cable between the
mother board and the
Inspiratory Electronics
PCB.
KB0002 Bad expiratory Background Vent inop Exhalation flow 1. If possible, run EST and
flow check (BDU) sensor (Q3) reading use the Flow Sensor
out of range Cross Check test to
determine if Q3 is fail-
ing. Replace Q3 if appli-
cable.
2. Perform flow sensor
calibration.
3. Replace AI PCB.
KB0003 Bad O2 PSOL Background Vent inop Oxygen PSOL 1. If possible, run EST to
current check (BDU) (PSOL1) current out determine if PSOL1 is
of range failing. Switch PSOLs to
see if the failure trans-
fers to the other side.
Replace PSOL1 if appli-
cable.
2. Replace AI PCB.
3. Replace Inspiratory
Electronics PCB.
4. Replace interconnect
cable between the
motherboard and the
Inspiratory Electronics
PCB.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
KB0004 Bad air PSOL Background Vent inop Air PSOL (PSOL2) 1. If possible, run EST to
current check (BDU) current out of range determine if PSOL2 is
failing. Switch PSOLs to
see if the failure trans-
fers to the other side.
Replace PSOL2 if appli-
cable.
2. Replace AI PCB.
3. Replace Inspiratory
Electronics PCB.
4. Replace interconnect
cable between the
motherboard and the
Inspiratory Electronics
PCB.
KB0005 Bad exp motor Background Vent inop Current to 1. Perform exhalation
current check (BDU) exhalation valve valve (EV) calibration.
motor out of range 2. Replace EV.
3. Replace AI PCB.
KB0007 Bad exp Background Vent inop PE reading out of 1. Replace exhalation
pressure check (BDU) range transducer PCB.
2. Replace AI PCB.
KB0008 Bad insp Background Vent inop PI reading out of 1. Replace Inspiratory
pressure check (BDU) range Electronics PCB.
2. Replace AI PCB.
KB0009 Air flow out of Background Vent inop Q2 reading too high 1. Perform flow sensor
range HIGH check (BDU) calibration.
2. Replace Q2.
3. Replace AI PCB.
KB0012 O2 flow out of Background Vent inop Q1 reading too high 1. Perform flow sensor
range HIGH check (BDU) calibration.
2. Replace Q1.
3. Replace AI PCB.
KB0016 Bad BD 10V Background Vent inop BDU +10 V supply Replace AI PCB.
supply check (BDU) out of range
KB0017 Bad BD 12 V Background Vent inop BDU +12 V supply 1. Replace AI PCB.
supply check (BDU) out of range 2. Replace power supply
or other power input
components.
KB0018 Bad BD 15 V Background Vent inop BDU +15 V supply 1. Replace AI PCB.
supply check (BDU) out of range 2. Replace power supply
or other power input
components.
KB0019 Bad BD -15 V Background Vent inop BDU -15 V supply 1. Replace AI PCB.
check (BDU) out of range 2. Replace power supply
or other power input
components.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
KB0022 Bad BD 5 V Background Vent inop BDU +5 V supply out 1. Replace AI PCB.
supply check (BDU) of range 2. Replace power supply
or other power input
components.
KB0023 O2 PSOL stuck Background Vent inop Oxygen PSOL 3. If possible, run EST to
check (BDU) (PSOL1) command determine if PSOL1 is
current out of range failing. Switch PSOLs to
see if the failure trans-
fers to the other side.
Replace PSOL1 if appli-
cable.
4. Replace AI PCB.
5. Replace Inspiratory
Electronics PCB.
6. Replace interconnect
cable between the
motherboard and the
Inspiratory Electronics
PCB.
KB0024 Air PSOL stuck Background Vent inop Air PSOL (PSOL2) 1. If possible, run EST to
check (BDU) command current determine if PSOL2 is
out of range failing. Switch PSOLs to
see if the failure trans-
fers to the other side.
Replace PSOL2 if appli-
cable.
2. Replace AI PCB.
3. Replace Inspiratory
Electronics PCB.
4. Replace interconnect
cable between the
motherboard and the
Inspiratory Electronics
PCB.
KB0025 Air PSOL stuck Background Vent inop Air PSOL (PSOL2) 1. If possible, run EST to
open check (BDU) command current determine if PSOL2 is
reading indicates failing. Switch PSOLs to
PSOL stuck open see if the failure trans-
fers to the other side.
Replace PSOL2 if appli-
cable.
2. Replace AI PCB.
3. Replace Inspiratory
Electronics PCB.
4. Replace interconnect
cable between the
motherboard and the
Inspiratory Electronics
PCB.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
KB0026 O2 PSOL stuck Background Vent inop PSOL1 command 1. If possible, run EST to
open check (BDU) current reading determine if PSOL1 is
indicates PSOL stuck failing. Switch PSOLs to
open see if the failure trans-
fers to the other side.
Replace PSOL1 if appli-
cable.
2. Replace AI PCB.
3. Replace Inspiratory
Electronics PCB.
4. Replace interconnect
cable between the
motherboard and the
Inspiratory Electronics
PCB.
KB0030 Bad safety valve Background Vent inop Current to safety 1. Replace safety valve.
current check (BDU) valve out of range 2. Replace AI PCB.
3. Replace Inspiratory
Electronics PCB.
4. Replace interconnect
cable between the
motherboard and the
Inspiratory Electronics
PCB.
KB0031 Insp pressure Background Vent inop Inspiratory pressure 1. Check or replace the
stuck check (BDU) transducer (PI) tubing between SOL 1
reading indicates and the outlet mani-
transducer is stuck fold to ensure no
restrictions to pressure
measurement.
2. Replace Inspiratory
Electronics PCB.
3. Replace AI PCB.
KB0032 Exp pressure Background Vent inop Expiratory pressure 1. Replace exhalation
stuck check (BDU) transducer (PE) transducer PCB.
reading indicates 2. Replace AI PCB.
transducer is stuck
KB0033 Insp pressure Background Vent inop Inspiratory pressure 1. Replace inspiratory
autozero offset check (BDU) transducer (PI) pressure transducer
failed reading following autozero solenoid
autozero out of (SOL1).
range 2. Replace Inspiratory
Electronics PCB.
3. Replace AI PCB.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
KB0034 Exp pressure Background Vent inop Expiratory pressure 1. Replace expiratory
autozero offset check (BDU) transducer (PE) pressure transducer
failed reading following autozero solenoid
autozero out of (SOL2).
range 2. Replace exhalation
transducer PCB.
3. Replace AI PCB.
KB0044 BD NOVRAM Background Vent inop BD NOVRAM failure 1. Replace BD CPU PCB.
checksum error check (BDU) 2. Replace BPS battery
pack.
KB0053 BD EEPROM Background Vent inop BD EEPROM failure Replace BD CPU PCB.
checksum error check (BDU)
KB0065 SV current out Background Vent inop Safety valve current 1. Replace BD CPU PCB if
of range check (BDU) out of range accompanied by error
code XB0074 or
XB0071.
2. Replace safety valve.
3. Replace Inspiratory
Electronics PCB.
4. Replace AI PCB.
5. Replace interconnect
cable between the
motherboard and the
Inspiratory Electronics
PCB.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
KB0079 BK vent inop Background Vent inop Ventilator 1. Run the VENT INOP
occurred check (BDU) inoperative signal test.
read by BDU is a. If the test passes, run
active. This could EST to reset the
mean GUI has Device Alert. Run the
declared a ventilator unit for 48 hours and
inoperative then put back into
condition. service if the code
cannot be dupli-
cated.
b. If the test fails, trou-
bleshoot per Table 6-
5, VENT INOP Test
troubleshooting.
2. Replace BD CPU PCB.
KP0002 Integer Unit POST (BDU) Vent inop Replace BD CPU PCB.
Test
KP0003 DRAM Refresh POST (BDU) Vent inop Replace BD CPU PCB.
Timer Test
KP0004 Kernel DRAM POST (BDU) Vent inop Replace BD CPU PCB.
Test
KP0005 Boot EPROM POST (BDU) Vent inop Replace BD CPU PCB.
Checksum Test
KP0006 POST Phase 2 POST (BDU) Vent inop Replace BD CPU PCB.
Initialization
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
KP0009 Rolling Thunder POST (BDU) Vent inop POST has started 1. Replace AI PCB.
Test (but not run to 2. Replace BD CPU PCB.
completion) three or 3. Replace motherboard.
more times, for
reasons unrelated to
power.
KP0012 Timer Test POST (BDU) Vent inop Timer failed 1. Replace BD real-time
comparison test clock.
with real-time clock 2. Replace BD CPU PCB.
(external timing
reference)
or
Timer failed to
generate interrupt
KP0013 Watchdog POST (BDU) Vent inop Watchdog timer did 1. Replace BD real-time
Timer Test not time out as clock.
expected 2. Replace BD CPU PCB.
KP0015 FLASH Memory POST (BDU) Vent inop Replace BD CPU PCB.
Checksum Test
KP0082 Bus Timer Test POST (BDU) Vent inop Activity has ceased Replace BD CPU PCB.
on local bus
or
Bus time-out circuit
did not generate an
interrupt as
expected
KP0083 NMI Source POST (BDU) Vent inop Replace BD CPU PCB.
Register Test
KP0084 POST DRAM POST (BDU) Vent inop Replace BD CPU PCB.
Test
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
KP0087 Unexpected POST (BDU) Vent inop POST has been 1. Check associated errors
Reset Umpire invoked three times in the System Diagnos-
Test in 24 operational tic and System Informa-
hours due to tion log to determine
unexpected resets. why POST was invoked
three times within 24
hours.
If the codes indicate
that the BD CPU
generated the resets,
replace the BD CPU
(Other codes may be
present indicating that
the GUI lost
communications with
the BD).
2. Run a complete EST to
check for analog device
failures.
KP0089 Floating Point POST (BDU) Vent inop Replace BD CPU PCB.
Unit Test
KP0090 DRAM Parity POST (BDU) Vent inop Replace BD CPU PCB.
Circuit Test
KP0114 ADC Test POST (BDU) Vent inop Analog-to-digital 1. Replace AI PCB.
converter data not 2. Replace motherboard.
available to CPU as
expected
KP0115 DAC Test POST (BDU) Vent inop Digital-to-analog 1. Replace AI PCB.
converter data not 2. Replace motherboard.
as expected
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
KP0116 Analog Devices POST (BDU) Vent inop Critical analog data 1. Refer to Section 6.8
Test out of range and Table 6-3 and use
the PC: error address to
identify which analog
device is failing.
2. Run EST to get more
information on the
analog device failure.
3. Replace the AI PCB.
4. Replace the mother-
board.
5. Replace the intercon-
nect cable between the
Inspiratory Electronics
PCB and the mother-
board.
KP0120 Safe State POST (BDU) Vent inop Places PSOLs and 1. Look up associated
System Test exhalation valve codes reported in Diag-
into safe state and nostic Log and/or the
verifies that they BD CPU LED array to
cannot be get more information
commanded. about other possible
system, software, or
hardware faults. Trou-
bleshoot those codes.
2. Run a Ground Isolation
Test to troubleshoot
shorts.
3. Try another inspiratory
module and see if the
code can be reset by
running the VENT INOP,
EST and POST tests.
a. If the unit passes
with a known good
inspiratory module,
replace the Inspira-
tory Electronics PCB.
b. If the unit fails,
replace the inspira-
tory blind mate
interconnect cable.
4. Replace the Inspiratory
Electronics PCB.
5. Replace the AI PCB.
6. Replace the BD CPU
PCB.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
KP0130 PB-MON Boot POST (BDU) Vent inop Replace BD CPU PCB.
LB0006 Bad exhalation Background Alert Exhalation valve coil 1. Replace exhalation
valve coil temp. check (BDU) temperature out of valve.
range 2. Replace AI PCB.
3. Replace interconnect
harness between the
exhalation compart-
ment connector and
the motherboard.
LB0010 Air flow out of Background Alert Air flow sensor (Q2) 1. Perform flow sensor
range LOW check (BDU) reading too low calibration.
2. Replace Q2.
3. Replace AI PCB.
LB0011 Bad air flow Background Alert Air flow sensor (Q2) 1. Replace Q2.
temperature check (BDU) thermistor reading 2. Replace AI PCB.
out of range
LB0013 O2 flow out of Background Alert Oxygen flow sensor 1. Ensure the ventilator
range LOW check (BDU) (Q1) reading too low has been warmed up
for at least 10 minutes
in service mode at
ambient temperature.
2. Run a flow sensor cali-
bration.
3. Replace Q1.4.Replace
the AI PCB.
LB0014 Bad O2 flow Background Alert Oxygen flow sensor 1. Replace Q1.
temp. check (BDU) (Q1) thermistor 2. Replace AI PCB.
reading out of range
LB0015 Bad expiratory Background Alert Expiratory flow 1. Ensure the ventilator
flow temp check (BDU) sensor (Q3) has been warmed up
thermistor reading for at least 10 minutes
out of range in service mode at
ambient temperature.
2. Replace Q3.
3. Replace AI PCB.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
LB0020 Bad GUI 12V Background Alert GUI +12 V supply out 1. Check cabling from BD
supply check (BDU) of range CPU PCB to GUI CPU
PCB.
2. Replace AI PCB.
3. Replace power supply.
4. Check whether or not
the cable from the GUI
to the BDU was recon-
nected to the unit
while the ventilator
power was on. If yes,
run EST to reset the
DEVICE ALERT.
LB0021 Bad GUI 5V Background Alert GUI +5 V supply out 1. Check cabling from BD
supply check (BDU) of range. CPU PCB to GUI CPU
PCB.
2. Replace AI PCB.
3. Replace power supply.
4. Check whether or not
the cable from the GUI
to the BDU was recon-
nected to the unit
while the ventilator
power was on. If yes,
run EST to reset the
DEVICE ALERT.
LB0028 Bad O2 sensor Background Alert Oxygen sensor 1. Calibrate oxygen sen-
OOR check (BDU) (percentage) sor.
reading out of range 2. Replace oxygen sensor.
LB0035 Bad power fail Background Alert Power fail capacitor 1. Replace AI PCB.
capacitor check (BDU) voltage out of range 2. Replace power supply.
voltage
LB0036 Alarm cable Background Alert Alarm cable voltage 1. Check cabling to main
error check (BDU) out of range alarm.
2. Replace BD alarm.
3. Replace AI PCB.
LB0043 AC switch stuck Background Alert Power switch is in 1. Replace power switch.
check (BDU) off state but power 2. Replace AI PCB.
fail NMI did not
occur
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
LB0048 Bad backup Background Alert BPS voltage out of 1. Replace BPS battery
power supply check (BDU) range pack.
voltage 2. Replace AI PCB.
3. Replace BPS PCB.
LB0049 Bad backup Background Alert BPS current out of 1. Replace BPS battery
power supply check (BDU) range pack.
current 2. Replace AI PCB.
3. Replace BPS PCB.
LB0050 Bad backup Background Alert Model information Replace BPS PCB.
power supply check (BDU) incorrect
model
LB0051 Bad exp heater Background Alert Exhalation heater 1. Replace exhalation
check (BDU) thermistor reading heater.
out of range 2. Replace AI PCB.
3. Replace interconnect
cable between the
exhalation compart-
ment and the mother-
board.
LB0057 Compressor bad Background Alert Data in compressor Replace compressor PCB.
eprom data check (BDU) EPROM not as
expected
LB0062 EST required Background Alert EST is required, but Run EST.
check (BDU) user cycled power
into normal
operation, rather
than into service
mode
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
LB0088 10000 hours Background Alert Compressor hours Replace the compressor
stored for check (BDU) corrupted. Contact PCB.
elapsed time your Covidien
representative to
reset.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
LB0092 Battery Event Background Alert A status message 1. Connect the ventilator
checks (BDU) regarding the state to a known good AC
of the battery. power source.
2. Verify that the AC LED
Message indicator near the
displayed in power switch is lit.
NOTES column 3. Check the AC circuit
of System breaker on the AC
Information log: panel and reset if nec-
essary.
ErrCode:0x1 Bad LED indicator or
4. Remove the power
open circuit
cord retainer (if pres-
ErrCode:0x2 On battery power ent) and ensure the
power cord is plugged
ErrCode:0x3 Off battery power, into the ventilator
running on AC properly; reinstall the
retainer.
ErrCode:0x4 Open circuit
5. Ensure the green LED
ErrCode:0x5 Less than 2 minutes on the front of the BPS
of battery operation is lit.
time remain a. If the green LED is
not lit, allow the BPS
to charge (indicated
by a yellow LED)
until the green LED
lights.
b. If the green LED
does not light within
6-8 hours or less:
• Verify that the BPS
charging voltage at the
power supply connector
is 34.2 V DC minimum.
To do this, disconnect
the BPS from the power
supply. Connect the
negative (-) DMM lead
to ground pin A1 and
the positive (+) DMM
lead to the connector at
pin 1.
A1
pin 1
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
LB0092
Caution
(continued)
To avoid damage to
ventilator
electronics, do not
cause a short circuit
between pins 1 and
A1 or pin 1 and
chassis ground when
measuring BPS
charging voltage.
• With battery
disconnected, check
charging voltage on
BPS PCB at battery
connector. (Refer to
Section 8.18 for BPS
disassembly
instructions.) If the
voltage does not
read 30 V DC- 37.8
V DC, replace the
BPS PCB.
6. Check/replace the bat-
tery pack fuse.
7. Check that the battery
voltage reads 26 V DC
minimum. If not,
replace the battery
pack.
8. Once the green LED is
lit, run EST to test the
BPS.
9. Replace the BPS cable.
10. Replace AI PCB.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
LP0008 Kernel POST (BD) Alert Noncritical Replace the BD CPU PCB.
NOVRAM Test NOVRAM data
failure
LP0087 Unexpected POST (BDU) Alert POST was invoked 1. Check all logs to see
Reset Umpire due to a watchdog associated errors occur-
Test time-out ring around the same
(unexpected reset) time and date as this
code.
2. Run EST to get more
information.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
LP0088 POST NOVRAM POST (BDU) Alert Noncritical 1. Check all logs to see
Test NOVRAM data associated errors occur-
failure ring around the same
time and date as this
code.
2. Run EST to get more
information.
LP0112 Ventilator INOP POST (BDU) Alert Ventilator 1. Replace the intercon-
Test inoperative signal nect cable between the
not in expected Inspiratory Electronics
state PCB and the mother-
board.
2. Run VENT INOP Test.
3. Replace BD CPU PCB.
LP0116 Analog Devices POST (BDU) Alert Noncritical analog 1. Refer to Section 6.8
Test data out of range and Table 6-3 and use
the PC: error address to
identify which analog
device is failing.
2. Replace AI PCB.
LP0121 Service Switch POST (BDU) Alert Service switch 1. Turn off ventilator.
Stuck Test appears to be Release switch.
closed. (Only 2. Ensure you are enter-
requests to enter ing SST and Service
service mode made Mode correctly. (Do not
after this test are press the switch for lon-
honored by system.) ger than one second.)
This code can be 3. Check for mechanical
generated if the obstruction to switch.
service switch is 4. Replace BD CPU PCB.
pressed for longer
than one second.
LP0122 AC Voltage Test POST (BDU) Alert AC monitor voltage 1. Make sure ventilator is
not within set plugged into AC
nominal line voltage source.
±20% 2. Make sure nominal line
voltage is set correctly.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
U01044 Assertion Background GUI reset Replace the GUI CPU PCB if
check (GUI) the failure recurs.
U04004 Assertion Background GUI reset Can occur if there is 1. Check all cable connec-
check (GUI) an intermittent tions.
connection to any 2. Perform Ground Isola-
digital I/O line. tion Test.
3. Replace GUI CPU PCB.
U05013 Assertion Background GUI reset Can occur if there is 1. Check all cable connec-
check (GUI) an intermittent tions.
connection to any 2. Perform Ground Isola-
digital I/O line. tion Test.
3. Replace GUI CPU PCB.
U08002 Assertion Background GUI reset Occurs when the GUI 1. Check for wet expira-
check (BDU) is overloaded with tory or inspiratory fil-
invalid data from ter.
the serial port. 2. Check all cable connec-
Caused by the host tions: GUI, Serial, Inter-
device transmitting nal cables.
bad data that is 3. Review the ventilator
generating over-run, logs, including patient
parity, or framing logs, observe any occur-
errors. The GUI rences of rapidly
generates a soft repeating messages.
reset or assertion to This pattern may con-
clear the condition tribute to a buffer
and resynchronize overflow situation that
with the BDU. the ventilator will
attempt to clear by run-
ning POST.
4. Perform Ground Isola-
tion Test.
5. Replace GUI CPU PCB.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
U08012 Assertion Background GUI reset The GUI generates a 1. Check all cable connec-
check (GUI) soft reset or tions.
assertion to clear the 2. Perform Ground Isola-
condition and tion Test.
resynchronize with 3. Replace GUI CPU PCB.
the BDU.
U14019 Assertion Background GUI reset The GUI generates a 1. Check all cable connec-
check (GUI) soft reset or tions.
assertion to clear the 2. Perform Ground Isola-
condition and tion Test.
resynchronize with 3. Replace GUI CPU PCB.
the BDU.
U14021 Assertion Background GUI reset If the time stamps 1. Check all cable connec-
check (GUI) between the GUI tions.
and the BD are out 2. Perform Ground Isola-
of range, the GUI tion Test.
generates a soft 3. Replace BD real-time
reset or assertion to clock.
clear the condition 4. Replace GUI real-time
and re synchronize clock.
with the BDU. 5. Replace GUI CPU PCB.
UT0003 Address Error Background GUI reset Replace GUI CPU PCB if
check (GUI) failure recurs.
UT0010 Line 1010 Background GUI reset Replace GUI CPU PCB if
Emulator check (GUI) failure recurs.
(Unimplemen-
ted A-line
Opcode)
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
UT0011 Line 1111 Background GUI reset Replace GUI CPU PCB if
Emulator check (GUI) failure recurs.
(Unimplemen-
ted F-line
Opcode)
UT0106 Unassigned Background GUI reset A reset of the GUI 1. Check all cable connec-
User-defined check (GUI) has occurred as a tions.
vector result of a 2. Perform Ground Isola-
hardware-trap tion Test.
detection. 3. Replace GUI CPU PCB if
failure recurs.
VN0004 NMI: +5V Other (GUI) Failure 1. Replace the power sup-
Overvoltage ply.
2. Replace the GUI CPU
PCB.
VN0129 NMI: DRAM Other (GUI) GUI reset 1. Check GUI/BD cable
Parity Error connections.
2. Replace power supply.
3. Replace GUI CPU PCB.
VN0130 NMI: Ethernet Other (GUI) GUI reset 1. Check GUI/BD cable
Parity Error connections.
2. Replace power supply.
3. Replace GUI CPU PCB.
VN0136 NMI: +12 V DC Other (GUI) GUI reset 1. Check GUI/BD cable
Overvoltage connections.
2. Replace power supply.
3. Replace GUI CPU PCB.
VN0144 NMI: +12 V DC Other (GUI) GUI reset 1. Check GUI/BD cable
Undervoltage connections.
2. Replace power supply.
3. Replace GUI CPU PCB.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
XP0001 Processor POST (GUI) GUI inop Replace GUI CPU PCB.
Initialization
XP0002 Integer Unit POST (GUI) GUI inop Replace GUI CPU PCB.
Test
XP0003 DRAM Refresh POST (GUI) GUI inop Replace GUI CPU PCB.
Timer Test
XP0004 Kernel DRAM POST (GUI) GUI inop Replace GUI CPU PCB.
Test
XP0005 Boot EPROM POST (GUI) GUI inop Replace GUI CPU PCB.
Checksum Test
XP0006 POST Phase 2 POST (GUI) GUI inop Replace GUI CPU PCB.
Initialization
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
XP0007 Addressing POST (GUI) GUI inop Replace GUI CPU PCB.
Mode Test
XP0008 Kernel POST (GUI) GUI inop Replace GUI CPU PCB.
NOVRAM Test
XP0010 Interrupt POST (GUI) GUI inop Replace GUI CPU PCB.
Controller Test
XP0011 Time of Day POST (GUI) GUI inop 1. Replace GUI real-time
Clock Test clock.
2. Replace GUI CPU PCB.
XP0012 Timer Test POST (GUI) GUI inop 1. Replace GUI real-time
clock.
2. Replace GUI CPU PCB.
XP0013 Watchdog POST (GUI) GUI reset Watchdog timer did 1. Make sure ventilator is
Timer Test not time out as plugged in to wall
expected power and BPS is fully
charged.
2. Replace GUI real-time
clock.
3. Replace GUI CPU PCB.
XP0015 FLASH Memory POST (GUI) GUI inop Replace GUI CPU PCB.
Checksum Test
XP0081 Memory POST (GUI) GUI inop Replace GUI CPU PCB.
Management
Unit Test
XP0082 Bus Timer Test POST (GUI) GUI inop Replace GUI CPU PCB.
XP0083 NMI Source POST (GUI) GUI inop Replace GUI CPU PCB.
Register Test
XP0084 POST DRAM POST (GUI) GUI inop Replace GUI CPU PCB.
Test
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
XP0087 Unexpected POST (GUI) GUI inop POST has been Check associated errors in
Reset Umpire invoked three times log. Call Technical Support
Test in 24 operational or your Covidien
hours due to representative.
unexpected resets.
XP0089 Floating Point POST (GUI) GUI inop Replace GUI CPU PCB.
Unit Test
XP0090 DRAM Parity POST (GUI) GUI inop Replace GUI CPU PCB.
Circuit Test
XP0097 SAAS Self-Test POST (GUI) GUI inop Replace GUI CPU PCB.
Start
XP0098 SAAS Self-Test POST (GUI) GUI inop 1. Check the GUI alarm
End connection.
2. Replace the GUI alarm.
3. Replace GUI CPU PCB.
XP0128 Download OS POST (GUI) GUI inop Replace GUI CPU PCB.
Boot
XP0129 Application OS POST (GUI) GUI inop Replace GUI CPU PCB.
Boot
XP0130 PB-MON Boot POST (GUI) GUI inop Replace GUI CPU PCB.
XP0131 Application POST (GUI) GUI inop Replace GUI CPU PCB.
Boot
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
ZB0047 GUI NOVRAM Background Alert Noncritical 1. Replace GUI CPU PCB.
checksum error checks (GUI) NOVRAM data 2. Call Technical Support
failure or your Covidien repre-
sentative.
ZB0054 GUI EEPROM Background Alert This code may be 1. Ensure the GUI is
checksum error checks (GUI) generated when installed onto the cor-
there is a serial rect BDU.
number mismatch 2. Replace GUI CPU PCB.
between the GUI
and BDU.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
ZB0086 Init Loss of BD Background Alert A status message See associated error codes.
communication checks (GUI) indicating that
communication with
BD was lost.
ZB0093 Compact FLASH Background Status Status message 1. Perform Service Mode
error checks (GUI) indicates a compact compact flash test to
flash error. Trending reset device.
option disabled, but 2. Replace GUI compact
ventilator flash component.
performance 3. Replace GUI CPU PCB.
unaffected.
Message
displayed in
NOTES column
of System
Information
Log:
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
ZC2000 DCI parity error Communi- Alert Check parity of host device.
Port 1 cations error
(GUI)
ZC2002 DCI command Communi- Alert Undefined 1. Check host software for
error Port 1 cations error command received compatibility with ven-
(GUI) tilator.
2. Check cabling between
ventilator and host.
ZC2010 DCI parity error Communi- Alert Check parity of host device.
Port 2 cations error
(GUI)
ZC2012 DCI command Communi- Alert Undefined 1. Check host software for
error Port 2 cations error command received compatibility with ven-
(GUI) tilator.
2. Check cabling between
ventilator and host.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
ZC2020 DCI parity error Communi- Alert Check parity of host device.
Port 3 cations error
(GUI)
ZC2022 DCI command Communi- Alert Undefined 1. Check host software for
error Port 3 cations error command received compatibility with ven-
(GUI) tilator.
2. Check cabling between
ventilator and host.
ZP0011 Time of day POST (GUI) GUI inop 1. Replace GUI real-time
clock test clock.
2. Replace GUI CPU PCB.
Table 6-2: Puritan Bennett™ 840 Ventilator System diagnostic codes (continued)
Message Information /
Code Test Response Corrective action
displayed possible cause
ZP0085 Ethernet Self- POST (GUI) GUI inop Replace GUI CPU PCB.
Test Start
ZP0086 Ethernet Self- POST (GUI) GUI inop Replace GUI CPU PCB.
Test End
ZP0087 Unexpected POST (GUI) Alert POST was invoked Check associated errors in
Reset Umpire due to an log. No action required
Test unexpected reset. unless other errors are
present.
ZP0088 POST NOVRAM POST (GUI) Alert Noncritical Replace GUI CPUPCB.
Test NOVRAM data
failure
6.8 System Diagnostic Log and BDU POST analog devices test
Analog device errors reported during the BDU POST analog devices test are recorded in the
System Diagnostic Log. These errors are reported as diagnostic codes KP0116 and LP0116.
Information in the NOTES column of the log identifies the analog channel on which the error
occurred.
Using an example of a BDU POST analog device error, as it might be reported in the System
Diagnostic Log:
The error address (see NOTES column above) identifies the analog channel:
PC: 0x0 002 02DD
The diagnostic codes, KP0116 (BDU) and LP0116 (BDU), are major faults (VENT INOP) that can
occur during the Analog Devices Test (Phase 2 POST).
If there are multiple KP0116 and/or LP0116 diagnostic codes, check the PC channels and
determine if they are all the same. If the channels are all the same, replace the associated
analog device or the AI PCB. If the channels are different, but originate from the same
module, check and/or replace the associated module, interconnect cable or AI PCB.
Using the fifth and sixth digits of error address reported in the System Diagnostic Log and
Table 6-3, the analog channel on which the error occurred can be identified.
Table 6-3: Address codes for BDU POST analog devices test errors
Table 6-3: Address codes for BDU POST analog devices test errors (continued)
12 O2 Sensor O2 Sensor
Table 6-4: BDU and GUI— Phase 1 (kernel) POST diagnostic codes
Table 6-4: BDU and GUI— Phase 1 (kernel) POST diagnostic codes (continued)
BDU
BDU Test BDU Corrective
BDU Test Fault
Description Diag. Code Action
Type
KP0114
Analog Devices Test Performs a range check on Major 1. Check the diagnostic code’s
each of the 32 analog or corrective actions in Table 6-2.
channels Minor 2. Refer to Section 6.8 for addi-
tional troubleshooting infor-
KP0116 or mation regarding this fault
LP0116 (specifically check the PC: error
address to determine the
affected analog device).
3. Replace affected component.
4. Replace AI PCB.
5. Call Technical Support or your
Covidien representative.
Analog Interface Tests analog interface Major 1. Check the diagnostic code’s
Test circuitry corrective actions in Table 6-2.
2. Replace BD CPU PCB.
3. Replace AI PCB.
KP0113 4. Replace mother board.
Bus Timer Test* Monitors activity on the Major 1. Check the diagnostic code’s
local bus corrective actions in Table 6-2.
2. Replace BD CPU PCB.
KP0082
KP0115
DRAM Memory Test* Tests DRAM by writing and Major 1. Check the diagnostic code’s
verifying data patterns to corrective actions in Table 6-2.
memory 2. Replace BD CPU PCB.
KP0084
*
Tested in Short POST
BDU
BDU Test BDU Corrective
BDU Test Fault
Description Diag. Code Action
Type
DRAM Parity Error Checks the ability to detect Major 1. Check the diagnostic code’s
Test* a parity error upon reading corrective actions in Table 6-2.
defective parity from 2. Replace BD CPU PCB.
memory
KP0090
LP0086
Ethernet Self-Test Internal routine verifies Major 1. Check the diagnostic code’s
Start functionality of Ethernet corrective actions in Table 6-2.
controller 2. Replace BD CPU PCB.
KP0085
NMI Register Test* Verifies NMI source register Major 1. Check the diagnostic code’s
is in reset state (no bits set) corrective actions in Table 6-2.
2. Replace BD CPU PCB.
KP0083
KP0006
Phase 2 NOVRAM Compares NOVRAM data to Minor 1. Check the diagnostic code’s
Test* valid states and ranges corrective actions in Table 6-2.
2. Call Technical Support or your
Covidien representative.
LP0088
Safe State Test Places PSOLs and exhalation Major 1. Check the diagnostic code’s
valve into safe state and corrective actions in Table 6-2.
verifies that they cannot be 2. Replace AI PCB.
commanded 3. Replace BD CPU PCB.
KP0120
Serial Device Test Writes to serial EEPROM to Major 1. Check the diagnostic code’s
verify accessibility corrective actions in Table 6-2.
2. Replace AI PCB.
3. Replace BD CPU PCB.
KP0117
BDU
BDU Test BDU Corrective
BDU Test Fault
Description Diag. Code Action
Type
Service Switch Test Checks that the service Minor 1. Check the diagnostic code’s
Start* mode switch is not stuck, corrective actions in Table 6-2.
and prevents entry into 2. Turn off ventilator. Release
Service Mode from a power switch.
fail condition LP0121 3. Check for mechanical obstruc-
tion to switch.
4. Replace BD CPU PCB.
Unexpected Reset Checks for unexpected Major 1. Check the diagnostic code’s
Umpire Test* resets of the CPU. Three or corrective actions in Table 6-2.
resets (strikes) in 24 hours Minor 2. Check associated errors in the
will cause a VENT INOP System Diagnostic and System
condition. KP0087 or Information log to determine
LP0087 why POST was invoked three
times within 24 hours.
a. If the codes indicate that
the BD CPU generated the
resets, replace the BD CPU
(Other codes may be pres-
ent indicating that the GUI
lost communications with
the BD).
b. If the codes indicated that
the GUI CPU generated the
resets, replace the GUI CPU
(Other codes may be pres-
ent indicating that the BD
lost communications with
the GUI.
3. Run a complete EST to check
for analog device failures.
VENT INOP Test Checks to see if the Major 1. Check the diagnostic code’s
ventilator is in the VENT corrective actions in Table 6-2.
INOP state 2. Check cabling from BD CPU
PCB to GUI CPU PCB.
KP0112 3. Replace BD CPU PCB.
4. Call Technical Support or your
Covidien representative.
*
Tested in Short POST
DRAM Parity Error Checks the ability to Major 1. Check the diag-
Test detect a parity error nostic code’s cor-
upon reading rective actions in
defective parity from Table 6-2.
memory XP0090 XP0090 2. Replace GUI CPU
PCB.
NMI Register Test Verifies NMI source Major 1. Check the diag-
register is in reset nostic code’s cor-
state (no bits set) rective actions in
Table 6-2.
XP0083 XP0083 2. Replace GUI CPU
PCB.
Table 6-8: Interrupt errors and test failures – POST self tests
Table 6-8: Interrupt errors and test failures – POST self tests (continued)
Code and
Test description and failure modes Corrective actions
message
Failure: No gas is available (neither PS1, PS2, nor FS0005 Connect gas supply.
PC detects pressure). Gas not
connected
Failure: Only air is available (only PS2 or PC AS0012 Make sure oxygen supply is connected.
detects pressure). O2 pressure not
detected
Failure: Only oxygen is available (only PS1 AS0013 Make sure air supply is connected.
detects pressure). Wall air
pressure not
detected
Failure: Flow controller unable to establish and AS0010 1. Make sure oxygen supply is connected and
control oxygen flow at 120 L/min. Unable to unrestricted.
establish O2 2. Verify that the oxygen supply is good.
flow 3. Verify that the patient circuit system has no
leaks or occlusions. If not sure, run the first
four tests in EST to get to the leak test. If
the leak test passes in EST, exit out of EST.
Correct the leak or occlusion in the patient
circuit used in SST and rerun or try another
patient circuit.
4. Verify that the oxygen regulator pressure is
set between 9-12 psi.
5. Switch PSOLs to see if the failure transfers
to the air side. If yes, return PSOL2 back to
its original position and replace PSOL1.
6. Run a flow sensor calibration.
7. Switch Q1 and Q2. Rerun the flow sensor
calibration and EST. If the failure transfers
to the air side, return Q2 back to its original
position and replace Q1.
Code and
Test description and failure modes Corrective actions
message
Failure: Flow controller unable to establish and FS0006 1. Make sure oxygen supply is connected and
control oxygen flow at 60, 5, and 1 L/min. Unable to unrestricted.
establish O2 2. Verify that the oxygen supply is good.
flow 3. Check regulated oxygen pressure.
4. Verify that the system has no leaks or occlu-
sions. If not sure, run the first four tests in
EST to run the leak test.
5. Switch PSOLs to see if the failure transfers
to the air side. If yes, return PSOL2 back to
its original position and replace PSOL1.
6. Run a flow sensor calibration.
7. Switch Q1 and Q2. Rerun the flow sensor
calibration and EST. If the failure transfers
to the air side, return Q2 back to its original
position and replace Q1.
8. Replace Q3.
Failure: Q1 and Q3 readings are too far apart. FS0001 1. Check for leak.
O2 flow sensor 2. Verify that safety valve is closed.
cross check 3. Perform flow sensor calibration.
failed 4. Replace Q1.
5. Replace Q3.
6. Replace AI PCB.
Failure: Oxygen flow > 0.153 L/min. FS0008 1. Verify no leaks at the Q1 o-rings.
O2 zero flow 2. Run EST Gas Supply Test to check for PSOL1
check failed leak.
3. Remove and then reseat PSOL1.
4. Replace PSOL1.
5. Perform a flow sensor calibration.
6. Replace Q1.
Code and
Test description and failure modes Corrective actions
message
Failure: Flow controller unable to establish and AS0011 1. Make sure air supply is connected and unre-
control air flow at 120 L/min. Unable to stricted.
establish air 2. Verify that the air supply is good.
flow 3. Verify that the patient circuit system has no
leaks or occlusions. If not sure, run the first
four tests in EST to get to the leak test. If
the leak test passes in EST, exit out of EST.
Correct the leak or occlusion in the patient
circuit used in SST and rerun or try another
patient circuit.
4. Verify that the air regulator pressure is set
between 9-12 psi.
5. Switch PSOLs to see if the failure transfers
to the oxygen side. If yes, return PSOL1 back
to its original position and replace PSOL2.
6. Run a flow sensor calibration.
7. Switch Q1 and Q2. Rerun the flow sensor
calibration and EST. If the failure transfers
to the oxygen side, return Q1 back to its
original position and replace Q2.
Failure: Flow controller unable to establish and FS0007 1. Make sure air supply is connected unre-
control air flow at 60, 5, and 1 L/min. Unable to stricted.
establish air 2. Verify that the air supply is adequate.
flow 3. Check regulated air pressure.
4. Verify that the system has no leaks or occlu-
sions. If not sure, run the first four tests in
EST to run the leak test.
5. Switch PSOLs to see if the failure transfers
to the oxygen side. If yes, return PSOL1 back
to its original position and replace PSOL2.
6. Run a flow sensor calibration.
7. Switch Q1 and Q2. Rerun the flow sensor
calibration and EST. If the failure transfers
to the oxygen side, return Q1 back to its
original position and replace Q2.
8. Replace Q3.
Code and
Test description and failure modes Corrective actions
message
Failure: Q2 and Q3 readings are too far apart. FS0003 1. Correct the leak or occlusion in the SST
Air flow sensor patient circuit and rerun or try known good
cross-check patient circuit and filter set.
failed 2. Verify that the ventilator system has no
leaks or occlusions by running the first four
tests in EST to run the leak test.
3. If the leak test fails in EST, troubleshoot the
expiratory filter or ventilator system for
leaks or occlusions.
4. If the leak test passes in EST, troubleshoot
the SST patient circuit.
5. Switch PSOLs to see if the failure transfers
to the oxygen side. If yes, return PSOL1 back
to its original position and replace PSOL2.
6. Run a flow sensor calibration.
7. Switch Q1 and Q2. Rerun the flow sensor
calibration and EST. If the failure transfers
to the oxygen side, return Q1 back to its
original position and replace Q2.
8. Replace Q3.
Failure: PSOL2 command current out of range. FS0004 1. Make sure air supply is connected and unre-
Air PSOL current stricted.
out of range 2. Verify that the air supply is good.
3. Verify that the system has no leaks or occlu-
sions. If not sure, run the first four tests in
EST to run the leak test.
4. Switch PSOLs to see if the failure transfers
to the oxygen side. If yes, return PSOL1 back
to its original position and replace PSOL2.
5. Run a flow sensor calibration.
6. Switch Q1 and Q2. Rerun the flow sensor
calibration and EST. If the failure transfers
to the oxygen side, return Q1 back to its
original position and replace Q2.
7. Replace Q3.
Code and
Test description and failure modes Corrective actions
message
Failure: Air flow > 0.153 L/min. FS0009 1. Verify no leaks at the Q2 o-rings.
Air zero flow 2. Run EST Gas Supply Test to check for PSOL2
check failed leak.
3. Remove and then reseat PSOL2.
4. Replace PSOL2.
5. Perform a flow sensor calibration.
6. Replace Q2.
Failure: Inspiratory pressure transducer ADC FS0101 1. Replace Inspiratory Electronics PCB.
count at 0 cmH2O is out of range. Inspiratory 2. Replace AI PCB.
autozero out of
range
Failure: Expiratory pressure transducer ADC FS0102 1. Replace exhalation transducer PCB.
count at 0 cmH2O is out of range. Expiratory 2. Replace AI PCB.
autozero out of 3. Replace SOL1.
range
Failure: Unable to build pressure to 10 cmH2O FS0103 1. Verify that the patient circuit system has no
at 5 L/min within time-out period. Failed to reach leaks or occlusions or use a known good
test pressure patient circuit and filter set.
2. Verify that the air supply or compressor is
available and adequate.
3. Verify that the ventilator system has no
leaks or occlusions. If not sure, run the first
four tests in EST to run the leak test.
Code and
Test description and failure modes Corrective actions
message
Failure: Inspiratory/expiratory pressure FS0104 1. Verify that the patient circuit system has no
transducer readings are too far apart. Cross-check leaks or occlusions or use a known good
failed patient circuit and filter set.
2. Verify that the ventilator system has no
leaks or occlusions. If not sure, run the first
four tests in EST to run the leak test.
3. If the leak test passes in EST, exit out of EST.
Review the data for the Circuit Pressure Test
to determine which pressure transducer is
out of range: PI or PE. Replace applicable
pressure transducer.
4. Replace AI PCB.
Failure: Unable to build pressure to 50 or FS0108 1. Make sure gas source is available.
100 cmH2O at 5 L/min within time-out period. Failed to reach 2. Check for system leak.
test pressure
Code and
Test description and failure modes Corrective actions
message
Circuit leak
Failure: Unable to attain leak test starting FS0205 1. Make sure test circuit is connected and is
pressure within time-out period. Unable to not leaking.
establish 2. Check exhalation valve operation, and ver-
pressure ify secure connection of expiratory filter.
Failure: System pressure dropped below FS0201 1. Make sure test circuit is connected.
85 cmH2O. Excessive leak 2. Check for system leaks.
3. Check exhalation valve operation.
4. Verify secure connection of expiratory filter.
Failure: Pressure drops to 85 cmH2O in 10 s, FS0201 1. Make sure patient circuit is connected.
then in 10 s more by ³ 30 cmH2O. Excessive leak 2. Check for system leaks.
3. Check exhalation valve operation.
4. Verify secure connections of bacteria filters
and humidifier.
Failure: Pressure drops to 85 cmH2O in 10 s, AS0202 1. Make sure patient circuit is connected.
then in 10 s more by ³ 10 cmH2O. Excessive leak 2. Check for system leaks.
3. Check exhalation valve operation.
4. Verify secure connections of bacteria filters
and humidifier.
Failure: Inspiratory/expiratory pressure AS0206 Make sure test circuit is connected properly.
transducer readings are too far apart. Test circuit not
connected
Expiratory filter
Failure: Unable to deliver stable gas flow. FS0301 Make sure gas supply is connected.
Unable to
establish flow
Code and
Test description and failure modes Corrective actions
message
Failure: Expiratory flow detected, indicating FS0306 Make sure patient circuit is disconnected at
that circuit not disconnected. Patient circuit expiratory filter.
not
disconnected
Failure: Unable to deliver stable gas flow. FS0301 Make sure gas supply is connected.
Unable to
establish flow
Failure: Expiratory flow reading indicates that FS0307 Make sure patient circuit is connected at
circuit not reconnected. Patient circuit expiratory filter.
not reconnected
Failure: Pressure drop across filter > 2 cmH2O, AS0305 Consider replacing expiratory filter.
but < 3 cmH2O. Occluded
expiratory filter
Failure: Pressure drop across filter > 3 cmH2O. FS0304 Replace expiratory filter.
Occluded
expiratory filter
Failure: Pressure drop across expiratory filter AS0311 1. Repeat test, following directions more
< 0.4 cmH2O, but > 0.1 cmH2O. Low expiratory closely.
filter P 2. Replace filter.
Failure: Pressure drop across expiratory filter FS0310 1. Repeat test, following directions more
< 0.1 cmH2O. Low expiratory closely.
filter P 2. Replace filter.
Code and
Test description and failure modes Corrective actions
message
Circuit Resistance
Failure: Unable to pressurize system within FS0412 1. Block wye when prompted.
time-out period. Wye not 2. Check for circuit leaks.
blocked
Failure: Peak flow < alert threshold but > failure AS0411 Check for kinked or occluded patient tubing.
threshold. Unable to reach
Alert threshold (adult/pediatric) = 80 L/min min peak flow
Alert threshold (neonatal) = No alert state for
neonatal; pass or fail only
Failure threshold (adult/pediatric) = 60 L/min
Failure threshold (neonatal) = 20 L/min
Failure: Peak flow < failure threshold. FS0410 Check for kinked or occluded patient tubing.
Failure threshold (adult/pediatric) = 60 L/min Unable to reach
Failure threshold (neonatal) = 20 L/min min peak flow
Failure: Flow controller unable to establish and FS0401 Make sure gas supply is connected.
control gas flow. Unable to
establish flow
Failure: Inspiratory limb pressure > 8.5 cmH2O AS0403 Check for occluded patient tubing.
(adult) or 5.5 cmH2O (pediatric), but Occluded
< 12.5 cmH2O (adult) or 7.5 cmH2O (pediatric), inspiratory limb
indicating occlusion.
No Alert state for neonatal; pass or fail only
Failure: Inspiratory limb pressure > 12.5 cmH2O FS0402 Check for occluded patient tubing.
(adult), 7.5 cmH2O (pediatric), or 3.5 cmH2O Occluded
(neonatal), indicating occlusion. inspiratory limb
Failure: Inspiratory limb pressure < 0.6 cmH2O AS0413 1. Make sure inspiratory filter is installed.
(adult) or 0.5 cmH2O (pediatric), but Insp limb 2. Repeat test, following directions more
> 0.2 cmH2O. resistance low closely.
No Alert state for neonatal; pass or fail only 3. Replace inspiratory filter.
4. Replace patient circuit
Failure: Inspiratory limb pressure < 0.2 cmH2O. FS0408 1. Make sure inspiratory filter is installed.
Insp limb 2. Repeat test, following directions more
resistance low closely.
3. Replace inspiratory filter.
4. Replace patient circuit
Code and
Test description and failure modes Corrective actions
message
Failure: Peak flow < alert threshold but > failure AS0407 Check for kinked or occluded patient tubing.
threshold. Unable to reach
Alert threshold (adult/pediatric) = 80 L/min min peak flow
Alert threshold (neonatal) = No alert state for
neonatal; pass or fail only
Failure threshold (adult/pediatric) = 60 L/min
Failure threshold (neonatal) = 20 L/min
Failure: Peak flow < failure threshold FS0404 Check for kinked or occluded patient tubing.
Failure threshold (adult/pediatric) = 60 L/min Unable to reach
Failure threshold (neonatal) = 20 L/min min peak flow
Failure: Expiratory limb pressure > 8.5 cmH2O AS0406 Check for occluded patient tubing.
(adult) or 5.5 cmH2O (pediatric) but Occluded
> 12.5 cmH2O (adult) or 7.5 cmH2O (pediatric), exhalation limb
indicating occlusion.
No Alert state for neonatal; pass or fail only
Failure: Expiratory limb pressure > 12.5 cmH2O FS0405 Check for occluded patient tubing.
(adult), 7.5 cmH2O (pediatric), or 3.5 cmH2O Occluded
(neonatal). exhalation limb
Failure: Expiratory limb pressure < 0.6 cmH2O AS0414 Replace patient circuit
(adult) or 0.5 cmH2O (pediatric), but Exp limb
> 0.2 cmH2O. resistance low
No Alert state for neonatal; pass or fail only
Failure: Expiratory limb pressure < 0.2 cmH2O. FS0409 Replace patient circuit
Exp limb
resistance low
Compliance calibration
Failure: System cannot pressurize to one of the FS0501 Check patient circuit for leaks. Replace as
test pressure points. Unable to necessary.
pressurize
circuit
Code and
Test description and failure modes Corrective actions
message
Failure: System cannot pressurize to one of the FS0502 Check patient circuit for leaks. Replace as
test pressure points. Unable to necessary.
pressurize
circuit
Failure: Out-of-range pressure ratio calculation. FS0503 1. Repeat test, making sure circuit is undis-
Compliance turbed during test.
calculation 2. Repeat SST to retest pressure transducers.
failure 3. Run EST and verify all pneumatics tests pass.
Failure: High compliance > 6 mL/cmH2O (adult) AS0505 1. Make sure correct patient tubing type was
or 4.5 mL/cmH2O (pediatric), or 2.5 mL/cmH2O Excessive specified in SST.
(neonatal) but < 12 mL/cmH2O (adult) or compliance 2. Replace patient circuit.
9 mL/cmH2O (pediatric), or 3 mL/cmH2O
(neonatal).
Failure: High compliance > 12 mL/cmH2O FS0504 1. Make sure correct patient tubing type was
(adult), 9 mL/cmH2O (pediatric), or 3 mL/cmH2O Excessive specified in SST.
(neonatal). compliance 2. Replace patient circuit.
Failure: High compliance falls below AS0507 1. Make sure correct patient tubing type was
1.56 mL/cmH2O (adult), 1.34 mL/cmH2O Compliance low specified in SST.
(pediatric), or 0.6 mL/cmH2O (neonatal) but not 2. Replace patient circuit with a known good
lower than 1.05 mL/cmH2O (adult or pediatric) circuit and filter set.
or 0.25 mL/cmH2O (neonatal). 3. Run EST to check the operation of PI and PE
during the Circuit Pressure Test. Replace
applicable pressure transducer.
4. Run Atmospheric Pressure Transducer cali-
bration.
Failure: High compliance falls below FS0506 1. Make sure correct patient tubing type was
1.05 mL/cmH2O (adult or pediatric) or Compliance low specified in SST.
0.25 mL/cmH2O (neonatal). 2. Replace patient circuit.
Failure: Low compliance > 6 mL/cmH2O (adult), AS0509 1. Make sure correct patient tubing type was
4.5 mL/cmH2O (pediatric), or 2.5 mL/cmH2O Excessive specified in SST.
(neonatal), but < 12 mL/cmH2O (adult), compliance 2. Replace patient circuit with a known good
9 mL/cmH2O (pediatric), or 3 mL/cmH2O circuit and filter set.
(neonatal). 3. Run EST to check the operation of PI and PE
during the Circuit Pressure Test. Replace
applicable pressure transducer.
4. Run Atmospheric Pressure Transducer cali-
bration.
Failure: Low compliance > 12 mL/cmH2O (adult), FS0508 1. Make sure correct patient tubing type was
9 mL/cmH2O (pediatric), or 3 mL/cmH2O Excessive specified in SST.
(neonatal). compliance 2. Replace patient circuit.
Code and
Test description and failure modes Corrective actions
message
Failure: Low compliance falls below AS0511 1. Make sure correct patient tubing type was
1.56 mL/cmH2O (adult), 1.34 mL/cmH2O Compliance low specified in SST.
(pediatric), or 0.6 mL/cmH2O (neonatal) but not 2. Replace patient circuit with a known good
below 1.05 mL/cmH2O (adult or pediatric) or circuit and filter set.
0.25 mL/cmH2O (neonatal). 3. Run EST to check the operation of PI and PE
during the Circuit Pressure Test. Replace
applicable pressure transducer.
4. Run Atmospheric Pressure Transducer cali-
bration.
Failure: Low compliance falls below FS0510 1. Make sure correct patient tubing type was
1.05 mL/cmH2O (adult or pediatric) or 0.25 Compliance low specified in SST.
(neonatal). 2. Replace patient circuit.
Code and
Description Display Corrective actions
message
Failure: System is still running on battery FE0010 1. Plug in ventilator power cord.
power. AC power not 2. Disconnect BPS to isolate
connected problem.
3. Replace power supply.
Code and
Description Display Corrective actions
message
Failure: Unable to build pressure (air) to FE0003 Inspiratory 1. Make sure that test circuit is
10 cmH2O at 5 L/min within time-out Failed to and expiratory installed and air supply or com-
period. reach test pressure (only pressor is available and good.
pressure if test fails) 2. Verify that the air regulator pres-
sure is set between 9-12 psi.
3. Verify that the patient circuit sys-
tem has no leaks or occlusions by
running the first four tests in EST
to get to the leak test.
Code and
Description Display Corrective actions
message
Failure: Unable to build pressure (air) to 50 FE0008 Inspiratory 1. Make sure air supply or compres-
or 100 cmH2O at 5 L/min within time-out Failed to and expiratory sor is available.
period. reach test pressure (only 2. Check for system leak at the expi-
pressure if test fails) ratory bacteria filter or O2 sensor.
If not sure, run the first four tests
in EST to run the leak test.
Failure: Flow controller unable to establish AE0110 1. Make sure oxygen supply is con-
and control oxygen flow at 120 L/min. Unable to nected and unrestricted.
establish O2 2. Verify oxygen regulator pressure
flow is set between 9-12 psi.
3. Switch PSOLs to see if failure
transfers to the other gas side. If
yes, return the air PSOL to its
original position and replace the
oxygen PSOL.
4. Run the leak test in EST to check
for leaks/occlusions.
5. Perform a flow sensor calibration.
6. Replace Q1.
7. Replace the Inspiratory Electron-
ics PCB.
8. Replace the AI PCB.
Code and
Description Display Corrective actions
message
Failure: Flow controller unable to establish FE0106 1. Make sure oxygen supply is con-
and control oxygen flow at 60, 5, and Unable to nected.
1 L/min. establish O2 2. Check regulated oxygen pressure.
flow 3. Replace PSOL1 or oxygen flow
sensor (Q1).
4. Replace AI PCB.
Failure: Q1 and Q3 readings are too far FE0101 1. Make sure oxygen supply is con-
apart. O2 flow nected and unrestricted.
sensor cross 2. Verify that the oxygen supply is
check failed good.
3. Verify that the system has no
leaks or occlusions. If not sure,
run the first four tests in EST to
run the leak test.
4. Switch PSOLs to see if the failure
transfers to the air side. If yes,
return PSOL2 back to its original
position and replace PSOL1.
5. Run a flow sensor calibration.6.
Switch Q1 and Q2. Rerun the flow
sensor calibration and EST. If the
failure transfers to the air side,
return Q2 back to its original
position and replace Q1.
Failure: PSOL1 command current is out of FE0102 1. Make sure oxygen supply is con-
range. O2 PSOL nected and unrestricted.
current out 2. Verify that the oxygen supply is
of range good.
3. Verify that the system has no
leaks or occlusions. If not sure,
run the first four tests in EST to
run the leak test.
4. Switch PSOLs to see if the failure
transfers to the air side. If yes,
return PSOL2 back to its original
position and replace PSOL1.
5. Run a flow sensor calibration.
6. Switch Q1 and Q2. Rerun the flow
sensor calibration and EST. If the
failure transfers to the air side,
return Q2 back to its original
position and replace Q1.
Code and
Description Display Corrective actions
message
Failure: Oxygen flow > 0.153 L/min. FE0108 1. Verify no leaks at the Q1 o-rings.
O2 zero flow 2. Run EST Gas Supply Test to check
check failed for PSOL1 leak. Remove and then
reseat PSOL1
3. Replace PSOL1.
4. Perform a flow sensor calibration.
5. Replace Q1.
Failure: Flow controller unable to establish AE0111 1. Make sure air supply is connected
and control air flow at 120 L/min. Unable to and unrestricted.
establish air 2. Verify air regulator pressure is set
flow between 9-12 psi.
3. Switch PSOLs to see if failure
transfers to the other gas side. If
yes, return Oxygen PSOL to its
original position and replace the
Air PSOL.
4. Run the leak test in EST to check
for leaks/occlusions.
5. Perform a flow sensor calibration.
6. Replace Q2.
7. Replace the Inspiratory Electron-
ics PCB.
8. Replace the AI PCB.
Failure: Flow controller unable to establish FE0107 1. Make sure air supply is con-
and control air flow at 60, 5, and 1 L/min. Unable to nected.
establish air 2. Check regulated air pressure.
flow 3. Perform flow sensor calibration.
4. Replace PSOL2 or air flow sensor
(Q2).
5. Replace AI PCB.
Code and
Description Display Corrective actions
message
Failure: Q2 and Q3 readings are too far FE0103 1. Make sure air supply is connected
apart. Air flow and unrestricted.
sensor cross 2. Verify that the air supply is good.
check failed 3. Verify that the system has no
leaks or occlusions. If not sure,
run the first four tests in EST to
run the leak test.
4. Switch PSOLs to see if the failure
transfers to the air side. If yes,
return PSOL back to its original
position and replace PSOL2
5. Run a flow sensor calibration.
6. Switch Q1 and Q2. Rerun the flow
sensor calibration and EST. If the
failure transfers to the oxygen
side, return Q1 back to its original
position and replace Q2.
Failure: PSOL2 command current out of FE0104 1. Make sure air supply is connected
range. Air PSOL and unrestricted.
current out 2. Verify that the air supply is good.
of range 3. Verify that the system has no
leaks or occlusions. If not sure,
run the first four tests in EST to
run the leak test.
4. Switch PSOLs to see if the failure
transfers to the air side. If yes,
return PSOL back to its original
position and replace PSOL2
5. Run a flow sensor calibration.
6. Switch Q1 and Q2. Rerun the flow
sensor calibration and EST. If the
failure transfers to the oxygen
side, return Q1 back to its original
position and replace Q2.
Failure: Air flow > 0.153 L/min. FE0109 1. Verify no leaks at the Q2 o-rings.
Air zero flow 2. Run EST Gas Supply Test to check
check failed for PSOL2 leak. Remove and then
reseat PSOL2
3. Replace PSOL2.
4. Perform a flow sensor calibration.
5. Replace Q2.
Code and
Description Display Corrective actions
message
Failure: ACCEPT key was pressed but PS2 FE0204 1. Make sure air supply is con-
still open. Wall air nected.
pressure not 2. Check PS2: Remove the wiring
detected connectors from PS2 and jumper
the connectors to each other. If
the system now recognizes the air
supply, replace PS2.
Code and
Description Display Corrective actions
message
Failure: ACCEPT key was pressed but PS1 FE0205 1. Make sure oxygen supply is
still open. O2 pressure connected.
not detected 2. Check PS1: Remove the wiring
connectors from PS1 and jumper
the connectors to each other. If
the system now recognizes the air
supply, replace PS1.
Failure: Stable flow could not be FE0210 1. Make sure To patient port is
established within allotted time. SV pressure blocked.
relief failed 2. Check for leaks at the O2 sensor,
SOL1 or PI.
3. Replace the safety valve.
4. Replace PI.
Failure: Cracking pressure out of range. FE0210 1. Make sure To patient port is
SV pressure blocked.
relief failed 2. Check for leaks at the O2 sensor,
SOL1 or PI.
3. Replace the safety valve.
4. Replace PI.
Failure: Neither PSOL could deliver FE0210 1. Make sure To patient port is
100 L/min within prescribed count. SV pressure blocked.
relief failed 2. Check for leaks at the O2 sensor,
SOL1 or PI.
3. Replace the safety valve.
4. Replace PI.
Failure: Peak pressure is too high. FE0210 1. Make sure To patient port is
SV pressure blocked.
relief failed 2. Check for leaks at the O2 sensor,
SOL1 or PI.
3. Replace the safety valve.
4. Replace PI.
Code and
Description Display Corrective actions
message
Failure: PS1 closed instead of open. FE0206 1. Make sure oxygen supply is dis-
O2 pressure connected.
detected 2. Replace Inspiratory Electronics
(O2 not PCB.
disconnected) 3. Replace PS1.
Failure: Air PSOL (PSOL2) forward leak was AE0213 1. Check for leaks around the Q2
detected via excessive pressure buildup Air PSOL leak o-rings.
(> 50 cmH2O but < 100 cmH2O) in blocked 2. Make sure air supply is connected
inspiratory module. and unrestricted.
3. Reseat Air PSOL.
4. Replace Air PSOL.
5. Replace AI PCB.
Failure: Air PSOL (PSOL2) forward leak was FE0207 1. Check for leaks around the Q2 o-
detected via excessive pressure buildup Air PSOL leak rings.
(> 100 cmH2O) in blocked inspiratory 2. Verify that the safety valve
module. relieves pressures above 100
cmH2O. Listen for the pressure
relief while watching the numeric
digital display of system pressure
in the upper screen. If the safety
valve does not crack open,
replace the safety valve.
3. Replace PSOL2.
Failure: PS2 closed instead of open. FE0208 1. Make sure air supply is discon-
Wall air nected.
pressure 2. Replace Inspiratory Electronics
detected PCB.
3. Replace PS2.
Code and
Description Display Corrective actions
message
Failure: During zero-flow check, Q2 reads FE0215 1. Make sure gas supplies are dis-
> 0.05 L/min. Air zero flow connected.
check failed 2. Make sure compressor is not run-
ning.
3. Verify no leaks at the Q2 o-rings.
4. Check for PSOL2 leak.
5. Remove and then reseat PSOL2.
6. Perform a flow sensor calibration.
7. Replace PSOL2.
8. Replace Q2.
Failure: During zero-flow check, Q1 reads FE0216 1. Make sure gas supplies are dis-
> 0.05 L/min. O2 zero flow connected.
check failed 2. Make sure compressor is not run-
ning.
3. Verify no leaks at the Q2 o-rings.
4. Check for PSOL2 leak.
5. Remove and then reseat PSOL2.
6. Perform a flow sensor calibration.
7. Replace PSOL2.
8. Replace Q2.
Failure: During zero-flow check, Q3 reads > FE0217 1. Make sure gas supplies are dis-
0.1 L/min. Exp zero flow connected.
check failed 2. Make sure compressor is not run-
ning.
3. Verify no leaks at the Q3 o-rings.
4. Perform a flow sensor calibration.
5. Replace Q3.
Failure: PS1 open instead of closed. FE0211 1. Make sure oxygen supply is con-
O2 pressure nected.
not detected 2. Replace Inspiratory Electronics
PCB.
Code and
Description Display Corrective actions
message
Failure: Oxygen PSOL (PSOL1) forward leak AE0214 1. Check for leaks around the Q1 o-
was detected via excessive pressure buildup O2 PSOL leak rings.
(> 50 cmH2O but < 100 cmH2O) in blocked 2. Make sure oxygen supply is con-
inspiratory module. nected and unrestricted.
3. Reseat oxygen PSOL.
4. Replace oxygen PSOL.
5. Replace AI PCB.
Failure: Oxygen PSOL (PSOL1) forward leak FE0209 1. Check for leaks around the Q1 o-
was detected via excessive pressure buildup O2 PSOL leak rings.
(> 100 cmH2O) in blocked inspiratory 2. Verify that the safety valve
module. relieves pressures above 100
cmH2O. Listen for the pressure
relief while watching the numeric
digital display of system pressure
in the upper screen. If the safety
valve does not crack open,
replace the safety valve.
3. Replace PSOL1.
Leak test
Failure: System cannot attain leak test FE0305 1. Make sure test circuit is con-
starting pressure using oxygen or air within Unable to nected.
time-out period. establish 2. Check for system leaks.
pressure 3. Check exhalation valve operation.
4. Verify secure connection of expi-
ratory filter.
Failure: System pressure dropped below FE0301 1. Make sure test circuit is con-
85 cmH2O. Excessive leak nected.
2. Check for system leaks.
3. Check exhalation valve operation.
4. Verify secure connection of expi-
ratory filter.
Code and
Description Display Corrective actions
message
Failure: Pressure not detected on expiratory AE0306 Pressure drop 1. Make sure test circuit is properly
side (inspiratory/expiratory pressure Test circuit in 10 s connected.
transducer readings are too far apart). not 2. Replace expiratory bacteria filter.
connected 3. Check for leaks around the Q3
flow sensor.
4. Check/replace the exhalation
valve.
5. Replace the expiratory pressure
transducer PCB.
Failure: Pressure drops to 85 cmH2O in 10 s, FE0301 1. Make sure test circuit is con-
then in 10 s more by 5 cmH2O. Excessive leak nected and is not leaking.
2. Check for system leaks or occlu-
sions especially at the O2 sensor
and expiratory filter.
3. Verify secure connection of expi-
ratory filter.
4. Check exhalation valve operation.
Code and
Description Display Corrective actions
message
Failure: CLEAR key pressed to indicate LED AE0601 1. Check/replace interconnect cable
not on. GUI High between the GUI LED PCB and
Alarm LED the GUI CPU PCB.
fails. 2. Replace the GUI LED PCB.
Code and
Description Display Corrective actions
message
Failure: CLEAR key pressed to indicate LED AE0602 1. Check/replace interconnect cable
not on. GUI Medium between the GUI LED PCB and
Alarm LED the GUI CPU PCB.
fails. 2. Replace the GUI LED PCB.
Failure: CLEAR key pressed to indicate LED AE0603 1. Check/replace interconnect cable
not on. GUI Low between the GUI LED PCB and
Alarm LED the GUI CPU PCB.
fails. 2. Replace the GUI LED PCB.
Failure: CLEAR key pressed to indicate LED AE0604 1. Check/replace interconnect cable
not on. GUI Normal between the GUI LED PCB and
LED fails. the GUI CPU PCB.
2. Replace the GUI LED PCB.
Failure: CLEAR key pressed to indicate LED AE0605 1. Check/replace interconnect cable
not on. GUI Batt between the GUI LED PCB and
Backup LED the GUI CPU PCB.
fails. 2. Replace the GUI LED PCB.
Failure: CLEAR key pressed to indicate LED AE0606 1. Check/replace interconnect cable
not on. GUI On Batt between the GUI LED PCB and
Pwr LED fails. the GUI CPU PCB.
2. Replace the GUI LED PCB.
Failure: CLEAR key pressed to indicate LED AE0607 1. Check/replace interconnect cable
not on. GUI between the GUI LED PCB and
Compressor the GUI CPU PCB.
Ready LED 2. Replace the GUI LED PCB.
fails.
Failure: CLEAR key pressed to indicate LED AE0608 1. Check/replace interconnect cable
not on. GUI between the GUI LED PCB and
Compressor the GUI CPU PCB.
Operating 2. Replace the GUI LED PCB.
LED fails.
Failure: CLEAR key pressed to indicate LED AE0609 1. Check/replace interconnect cable
not on. GUI 100% O2 between the GUI LED PCB and
or INCREASE the GUI CPU PCB.
O2 2 min key 2. Replace keyboard.
LED fails.
Failure: CLEAR key pressed to indicate LED AE0610 1. Check/replace interconnect cable
not on. GUI Alarm between the GUI LED PCB and
Silence LED the GUI CPU PCB.
fails. 2. Replace keyboard.
Failure: CLEAR key pressed to indicate LED AE0611 1. Check/replace interconnect cable
not on. GUI Screen between the GUI LED PCB and
Lock LED the GUI CPU PCB.
fails. 2. Replace keyboard.
Code and
Description Display Corrective actions
message
BD lamp test
Failure: CLEAR key pressed to indicate one AE0702 Replace BDU LED PCB.
or both ventilator inoperative LEDs not on. Bad Vent
inop LED
Failure: CLEAR key pressed to indicate one AE0703 Replace BDU LED PCB.
or both SVO LEDs not on. Bad SVO LED
Failure: CLEAR key pressed to indicate LED AE0704 Replace BDU LED PCB.
not on. Bad Loss of
GUI LED
Failure: CLEAR key pressed to indicate GUI FE0801 1. Verify the GUI alarm cable con-
audio diagnostic failed. SAAS nection to the GUI CPU PCB.
(Safety 2. Replace GUI alarm assembly.
Audible 3. Replace GUI CPU PCB.
Alarm
System) test
failed
Failure: CLEAR key pressed to indicate relay FE2301 1. Make sure test equipment is con-
is stuck on when it should be off. Nurse call nected properly to remote alarm
stuck on port.
2. Replace GUI CPU PCB.
Failure: CLEAR key pressed to indicate relay FE2302 1. Make sure test equipment is con-
is stuck off when it should be on. Nurse call nected properly to remote alarm
stuck off port.
2. Replace GUI CPU PCB.
Code and
Description Display Corrective actions
message
BD audio test
Failure: Voltage out of range FE0901 1. Make sure BD alarm cable is con-
(< 3.5 or > 5.05 V). Bad alarm nected.
cable 2. Replace BD alarm.
3. Replace AI PCB.
4. Replace BD alarm cable.
Failure: Voltage out of range FE0902 1. Make sure BD alarm cable is con-
(< 4.5 or > 5.05 V). Bad power nected.
fail cap 2. Replace BD alarm.
3. Replace AI PCB.
4. Replace BD alarm cable.
5. Replace Motherboard PCB.
Failure: CLEAR key pressed to indicate user FE0904 1. Make sure BD alarm cable is con-
did not hear alarm, although alarm was Bad BD audio nected.
active. 2. Replace BD alarm.
3. Replace AI PCB.
4. Replace BD alarm cable.
Failure: Final capacitor voltage higher than FE0903 1. Make sure BD alarm cable is con-
initial voltage. Bad power nected.
fail cap 2. Replace BD alarm.
3. Replace AI PCB.
Code and
Description Display Corrective actions
message
Failure: PSOL2 loopback current out of AE1001 1. Verify that the air supply is good.
range (OOR) of drive current. Air PSOL 2. Verify air regulator pressure set
loopback to between 9 and 12 psi.
current OOR 3. Switch PSOLs to see if failure
transfers to the other gas side. If
yes, return oxygen PSOL (PSOL1)
to its original position and
replace PSOL2.
4. Switch Q1 and Q2, run a flow
sensor calibration and rerun test.
If the problem transfers to the
other gas side, return Q1 to its
original position and replace Q2.
5. Replace the AI PCB.
6. Replace the Inspiratory
Electronics PCB.
Failure: PSOL1 loopback current out of AE1002 1. Verify that the oxygen supply is
range (OOR) of drive current. O2 PSOL good.
loopback 2. Verify oxygen regulator pressure
current OOR set to between 9 and 12 psi.
3. Switch PSOLs to see if failure
transfers to the other gas side. If
yes, return the air PSOL to its
original position and replace
PSOL1.
4. Switch Q1 and Q2, run a flow sen-
sor calibration and rerun test. If
the problem transfers to the
other gas side, return Q2 to its
original position and replace Q1.
5. Replace the AI PCB.
6. Replace the Inspiratory
Electronics PCB.
Code and
Description Display Corrective actions
message
Failure: Flow controller unable to establish FE1105 1. Make sure air supply is connected
and control gas flow at 60 L/min. Unable to and unrestricted.
establish flow 2. Check air regulator pressure.
3. Run flow sensor calibration.
4. Switch PSOLs to see if test passes.
If it does, return PSOL1 to its orig-
inal position and replace PSOL2.
5. Replace Q2.
Failure: Excessive safety valve back pressure FE1101 Replace safety valve.
when safety valve is open. Safety valve
occluded
Failure: Safety valve loopback current out FE1102 1. Replace safety valve.
of range. Bad safety 2. Replace AI PCB.
valve driver 3. Replace Inspiratory Electronics
or loopback PCB.
Failure: It took too long to relieve excess FE1103 1. Make sure proper test circuit is
pressure through open safety valve, Insp check used.
indicating inspiratory check valve (CV3) is valve test 2. Make sure CV3 is not installed
occluded or test circuit is too large. failed backward.
3. Replace CV3.
4. Replace Safety Valve (SV).
Failure: It took too little time to relieve AE1104 1. Make sure test circuit is con-
excess pressure through open safety valve, Insp check nected.
indicating inspiratory check valve (CV3) valve test 2. Make sure CV3 is not installed
may be damaged or incorrectly mounted. failed backward.
3. Replace CV3.
Code and
Description Display Corrective actions
message
Failure: Exhalation valve loopback current AE1201 1. Verify that the system has no
is out of range (OOR) of drive current. Exp valve leaks or occlusions by running the
loopback leak test in EST.
current OOR 2. Clean exhalation valve
diaphragm.
3. Calibrate the exhalation valve.
4. Clean/replace the exhalation
valve.
5. Replace the AI PCB.
Failure: Exhalation valve table checksum FE1304 1. Check for leaks or occlusions. Cor-
invalid or last calibration performed was Exp valve not rect and then run the exhalation
not completed successfully. calibrated valve calibration.
2. Replace the exhalation valve.
3. Replace the AI PCB.
4. Replace Q3.
5. Replace Q2.
Failure: Exhalation valve magnet FE1302 1. Ensure the unit has been warmed
temperature out of range (OOR) Exp valve up at ambient temperature for at
(10 to 100 °C). temp OOR least 10 minutes.
2. Calibrate the exhalation valve
(EV).
3. Replace exhalation valve (EV).
4. Replace AI PCB.
Failure: Flow controller unable to establish FE1303 1. Make sure proper test circuit it
and control air flow measured by Unable to used and that there are no leaks
exhalation flow sensor (Q3). establish exp or occlusions.
flow 2. Make sure air supply is still con-
nected.
3. Run flow sensor calibration.
4. Replace Q3.
5. Replace Q2.
6. Replace the exhalation valve.
Code and
Description Display Corrective actions
message
Failure: Seal test pressure is above failure FE1301 1. Clean exhalation valve.
level. Seal test 2. Calibrate exhalation valve.
failed 3. Run flow sensor calibration.
4. Replace exhalation valve.
5. Replace AI PCB.
6. Replace PE.
Failure: Seal test pressure is above alert AE1305 1. Verify that the system has no
level but below failure level. Seal test leaks or occlusions by running the
failed leak test in EST.
2. Clean the exhalation valve.
3. Calibrate exhalation valve.
4. Replace the exhalation valve.
5. Replace the AI PCB.
Failure: Exhalation valve table checksum FE1402 1. Check for leaks or occlusions. Cor-
invalid or last calibration performed was Exp valve not rect and then run the exhalation
not completed successfully. calibrated valve calibration.
2. Replace the exhalation valve.
3. Replace the AI PCB.
4. Replace Q3.
5. Replace Q2.
Failure: Flow controller unable to establish FE1403 1. Make sure air supply is con-
and control air flow measured by Unable to nected.
exhalation flow sensor (Q3). establish exp 2. Make sure proper test circuit is
flow used.
Code and
Description Display Corrective actions
message
Failure: Expiratory P too low, indicating FE2201 1. Calibrate exhalation valve and
velocity transducer not responding Low exp P retest.
properly. 2. Replace exhalation valve.
Failure: Flow controller unable to establish FE1501 Make sure air supply is connected.
and control 60 L/min flow. Unable to
establish air
flow
Failure: Heater temperature did not rise FE1502 1. Replace exhalation heater.
sufficiently during an interval after heater Bad exp 2. Replace AI PCB.
was turned on. heater
Failure: Heater temperature did not drop FE1503 1. Replace exhalation heater.
sufficiently during an interval after heater Bad exp 2. Replace AI PCB.
was turned off. heater
Code and
Description Display Corrective actions
message
Compressor test
Failure: System is still running on battery AE1602 1. Plug in ventilator power cord and
power (i.e., when polled, either BPS AC power not check the cord connection at the
charged or BPS charging line is active). connected ventilator.
Compressor can only run on AC (facility) 2. Check AC.
power. 3. Disconnect BPS to isolate prob-
lem.
4. Replace power supply.
Failure: PS2 is closed when it should be AE1601 1. Make sure air supply is discon-
open. Wall air nected.
pressure 2. Disconnect PS2 and rerun test. If
detected test passes, replace PS2.
Failure: Calculated run time not within AE1604 Listen for motor. If motor is on,
range, perhaps because compressor timer is Run mode replace compressor PCB. Otherwise,
not running while compressor motor is on. time OOR replace compressor.
Code and
Description Display Corrective actions
message
Failure: Calculated disabled time not within AE1607 Replace compressor PCB.
range, perhaps because compressor timer is Disabled
running while compressor motor is off. mode time
OOR
Failure: Calculated standby mode time not AE1609 Replace compressor PCB.
within range, perhaps because compressor Standby
motor is still running. During standby mode mode time
test phase, compressor motor should OOR
eventually turn off.
Code and
Description Display Corrective actions
message
Failure: PC reading indicates compressor AE1611 1. Run compressor leak test to check
unable to maintain minimum pressure Compressor for compressor leak.
under worst-case breath delivery load test a. Use leak detector to check for
waveform. failed leaks at the accumulator fit-
tings and other tubing connec-
tions.
b. Verify no leaks at CV2 within
the ventilator by plugging the
air inlet fitting.
2. Replace compressor.
3. Replace compressor PCB.
Failure: Compressor leak test cannot be run, AE1701 1. Troubleshoot accompanying code
because of inability to verify timer Unable to AE1604 or AE1607.
functionality. perform 2. Replace compressor PCB.
compressor
leak test
Failure: PS2 is closed, but it should be open. AE1702 1. Make sure air supply is discon-
Compressor control is restored. Wall air nected.
pressure 2. Disconnect PS2 and rerun test. If
detected test passes, replace PS2.
Failure: PS2 is open, but it should be closed. AE1703 1. Make sure air supply is con-
Compressor control is restored. Wall air nected.
pressure not 2. Run Gas Supply/SV Test.
detected
Code and
Description Display Corrective actions
message
Failure: Touchscreen error occurred (e.g., AE1901 1. Clean touchscreen, removing any
blocked beam or low-level error), buffer GUI touch: obstructions.
cannot be read, or report is invalid. Error 2. Replace touchframe PCB.
3. Replace GUI CPU PCB.
Failure: Received message not identical to AE2001 Replace GUI CPU PCB.
transmitted message. Bad GUI serial
port
Code and
Description Display Corrective actions
message
Battery test
Failure: BPS not fully charged. AE2101 1. Allow BPS to fully charge, then
Battery not repeat test.
charged 2. Replace BPS PCB.
3. Replace battery pack.
4. Replace BD CPU PCB.
Failure: BPS voltage dropped below AE2103 1. Ensure the battery pack is fully
accepted level or dropped too quickly. Bad Backup charged by checking that the
Power Supply green LED on the BPS is lit prior
to initiating an EST. If the amber
LED is lit, allow the unit to charge
the batteries prior to rerunning
EST.
2. Replace the battery pack.
3. Replace the BPS PCB.
4. Replace the AI PCB.
5. Replace the power supply.
Failure: BPS not charging after AC power AE2104 1. Verify that AC power is recon-
was reconnected. Battery not nected when prompted and that
charging AC is good.
2. Replace BPS PCB.
3. Replace the battery pack.
4. Replace the power supply.
5. Replace the BD CPU.
6. Replace the AI PCB.
The NOTES column in the log contains the UT0002 Fault Address (FA). Using the FA, you can
determine if there is a problem with a VGA LCD controller, and identify whether it is the
upper LCD PCB or the lower LCD PCB. If you cannot locate a fault code in Table 6-11 that
matches the three digits of the reported fault code then replace the GUI CPU PCB.
Table 6-11: UT0002 Fault Addresses for VGA LCD controller PCB errors
The example given indicates that the upper VGA LCD Controller PCB should be replaced.
NOTE:
Covidien offers an in-depth training course on the Puritan Bennett™ 840 Ventilator
System, that includes error troubleshooting and repair. Call Covidien Technical
Support at 1.800.255.6774 for more information.
Software Property of
Covidien
(Released/Unformatted)
Lower screen:
Ventilator Startup
Analysis:
• Ventilator Test Summary SST and EST PASSED messages indicate that there is
communication between the BDU and the GUI. The message on the lower GUI screen
indicating an incorrect serial number provides a clue that the data key is faulty, because
the BD and GUI serial numbers are stored in the data key. The appearance of the
“Software Property of Covidien” message indicates that the data key information is not
being read by the GUI.
• The System Diagnostic Log code LB0089 indicates that the data key is faulty or missing and
should be replaced (see Table 6-2).
• The System Information Log indicates that the GUI and BD have resumed communicating,
which suggests that the BD CPU PCB and AI PCB are operating properly. This is also
indicated by the BD CPU PCB LEDs strobing and AI PCB LED flashing properly (BDU side
panel must be removed to verify LEDs).
• If you enter Service Mode, the lower GUI screen displays a message, “Data key not
installed.”
Possible corrective actions based on the messages above include replacing the data key, AI
PCB, or motherboard.
Software Property of
Covidien
(Released/Unformatted)
Lower screen:
Ventilator Startup
Analysis:
• Ventilator Test Summary SST and EST NEVER RUN and calibration messages indicate that
there is no communication between the BDU and the GUI (EST, SST, and calibration data
did not get uploaded to the GUI). The “GUI timeout occurred” and “Software Property of
Covidien” messages are further evidence that the GUI and BDU are not communicating.
• More than one cause could manifest itself in this way. The following list suggests ways to
pinpoint the cause of the communication problem:
– Check the System Diagnostic and System Information logs for more information. Use
the time stamps to help establish the order in which the codes were generated.
– If you can enter Service Mode, run EST for more information.
– Look at the BD PCB LEDs for more information. For example, if no LEDs are lit, it
suggests that the BD +5V signal is not getting to the BD CPU PCB. This could be caused
by a faulty power supply, AI PCB, or BD CPU PCB.
– Check, and replace if necessary, the Ethernet cable between GUI and BDU. If the cable
has broken wires or connections at the pins, the communication error could result.
Software Property of
Covidien
(Released/Unformatted)
Lower screen:
Ventilator Startup
Analysis:
• In this example, the message “BD POST failed” may also appear as “GUI POST failed,”
“BD Background failed,” “GUI Background failed,” or “Timeout/Inop occurred.” Use these
messages as your starting point to determine the cause of the problem.
• Consult the System Diagnostic and System Information Logs to view the associated
diagnostic codes. Use the time stamps to establish the order in which the codes were
generated.
• If you can enter Service Mode, run EST for more information.
• Look at the BD and GUI CPU PCB LEDs to determine POST diagnostic codes (refer to
Table 6-1 for information on how to interpret diagnostic codes, Table 6-2, and Table 6-4
through Table 6-8 for POST diagnostic LED patterns). Replace indicated components based
upon recommended diagnostic code corrective actions.
NOTE:
If the message “BD TUV failed” appears on the upper GUI screen, replace the BD CPU
PCB.
Alarm Handling
} priority active
alarm messages are
displayed here.
3. Correct the condition, if necessary, referring to Table 7-2. You can check the alarm log
(Figure 7-2). You may want to check the contents of the System Diagnostic Log
(accessible through the service menu, Chapter 4) for related diagnostic codes. The
troubleshooting procedures listed are sequenced to correct the most probable
malfunction or to present the most efficient corrective action first. The proposed fixes
listed, however, may not always correct the particular problem.
NOTE:
• You can change alarm settings even when alarms are active.
• Completing a new patient setup or running EST clears the alarm log.
• When more than one alarm is active and their alarm messages vary in their
degree of seriousness, you should assume that the most serious message is
applicable.
4. If necessary, clear the message or reset the alarm by pressing the alarm reset key.
For specifics about alarm operation, see the Puritan Bennett™ 800 Series Ventilator System
Operator’s and Technical Reference Manual.
"Lockable" alarms include all PATIENT DATA and CIRCUIT DISCONNECT alarms and can be
silenced for two minutes by pressing the alarm silence key. A new, high-urgency alarm cancels
the silence. Each time you press the alarm silence key, the silence period resets to two
minutes. Each time you press the alarm silence key (whether or not there is an active alarm),
the keypress is recorded in the alarm log.
Analysis Remedy
Base message Description Corrective action
message message
APNEA The set apnea interval has Check patient and settings.
(lockable) elapsed without the ventilator,
patient, or operator triggering
a breath. Resets when patient
initiates 2 consecutive breaths.
Analysis Remedy
Base message Description Corrective action
message message
CIRCUIT No Check patient/ Ventilator has recovered from 1. Check patient circuit con-
DISCONNECT ventilation. ventilator unintended power loss, nections.
(lockable) status. detects circuit disconnect, and 2. Run EST.
switches to idle mode; upper
screen displays elapsed time
without ventilator support.
Resets when ventilator senses
reconnection.
Analysis Remedy
Base message Description Corrective action
message message
Last spont
breath set
1PPEAK limit
Last 3 spont
breaths set
1PPEAK limit
Last 4 or
more spont
breaths set
1PPEAK limit
3VTE SPONT Check patient Exhaled spontaneous tidal
& settings volume set limit
Analysis Remedy
Base message Description Corrective action
message message
1VTI SPONT Delivered inspiratory volume 1. For TC: Check for leaks,
(lockable) high inspiratory volume limit tube type/I.D. setting.
(breath type = TC or VS) 2. For VS: Check patient and
settings.
LOSS OF Power switch is on, but there is Turn power switch off to
POWER insufficient AC and BPS power. reset alarm. Restore AC
There may not be a visual power.
indicator for this alarm, but an
independent audio alarm
sounds for at least 120 s.
LOW AC Ventilator Power Mains (AC) power has dropped 1. Check integrity of AC
POWER currently not interrupt below 80% of nominal for 1 s. power source.
affected. possible. Ventilator continues operation 2. If ventilator software was
as close to settings as possible. recently downloaded or if
Resets when there is no low AC ventilator was moved
power signal for 1 s. between locations with
different voltages, check
Service Mode Setup con-
NOTE: figuration.
Compressor operation 3. Check connections of AC
may be disrupted power distribution com-
during low AC events. ponents.
LOW BATTERY Operational Replace or Resets when BPS has more 1. Check integrity of AC
time < 2 allow than approximately 2 min of power source.
minutes. recharge. operational time remaining. 2. Check connections of AC
power distribution com-
ponents.
Analysis Remedy
Base message Description Corrective action
message message
3O2% Measured Check patient, The O2% measured during any 1. Calibrate oxygen sensor.
O2% < set gas sources, phase of a breath cycle is 7% 2. Run EST.
O2%. O2 analyzer & (12% during the first hour of 3. Replace oxygen sensor.
ventilator. operation) or more below the 4. Inspect oxygen PSOL
O2% setting for at least 30 s, (PSOL1).
or below 18%. (These
percentages increase by 5%
for 4 minutes following an
increase in the O2% setting.)
Alarm updated at 1-s intervals.
3VTE MAND 2 of last Check for Exhaled mandatory tidal Run SST.
(lockable) 4 mand. leaks, changes volume set limit. Alarm
breaths set in patient‘s updated whenever exhaled
limit. RSTAT & CSTAT. mandatory tidal volume is
recalculated.
3 of last Check for
4 mand. leaks, changes
breaths set in patient‘s
limit. RSTAT & CSTAT.
3V E TOT VE TOT set Check patient Total minute volume set Run SST.
(lockable) limit for & settings. limit. Alarm updated
20s. whenever exhaled minute
volume is recalculated.
VE TOT set Check patient
limit for & settings.
> 20s.
Analysis Remedy
Base message Description Corrective action
message message
NO AIR SUPPLY Ventilation Check air Ventilator delivers 100% 1. Check air inlet filter.
continues as source. oxygen. Resets if air supply 2. Check pressure switch
set. Only O2 connected. connections.
available. 3. Run EST.
Compressor Check air Ventilator delivers 100% 1. Check air inlet filter.
inoperative. source. oxygen. Resets if air supply 2. Check pressure switch
Ventilation connected. connections.
continues as 3. Run EST.
set. Only O2
available.
Ventilation Check patient Ventilator delivers 100% 1. Check air inlet filter.
continues as & air source. oxygen instead of set O2%. 2. Check pressure switch
set except Resets if air supply connected. connections.
O2% = 100 3. Run EST.
Compressor Check patient Ventilator delivers 100% 1. Check air inlet filter.
inoperative. & air source. oxygen instead of set O2%. 2. Check pressure switch
Ventilation Resets if air supply connected. connections.
continues as 3. Run EST.
set, except
O2% = 100.
NO O2 SUPPLY Ventilation Check patient Resets if oxygen supply 1. Check oxygen inlet filter.
continues as and O2 connected. 2. Check pressure switch
set. Only air source. connections.
available. 3. Run EST.
Ventilation Check patient Ventilator delivers 21% 1. Check oxygen inlet filter.
continues as & O2 source oxygen instead of set O2%. 2. Check pressure switch
set, except Resets if oxygen supply connections.
O2% = 21. connected. 3. Run EST.
Analysis Remedy
Base message Description Corrective action
message message
SCREEN BLOCK Possible Remove Possible blocked beam (GUI) or 1. Remove obstruction.
blocked obstruction or touch screen fault. 2. Service ventilator.
beam or service
touch screen ventilator.
block.
VOLUME NOT Last 2 mand Check patient Ventilator cannot deliver Check patient and settings.
DELIVERED breaths, and setting target volume.
(This alarm pressure for
applies to VC+ > max 2PPEAK.
and VS allowable
breaths.) level
Last 10 or
more mand
breaths,
pressure
> max
allowable
level
• To prevent possible personal injury, never attempt to push or pull a ventilator that is
installed on a cart, while the brakes are set on the casters.
• To prevent possible personal injury and equipment damage, make sure the brakes on the
casters are set to prevent inadvertent movement of the ventilator during service.
• To prevent possible personal injury and equipment damage, have someone assist you
when lifting the ventilator or any of its major components.
• Chemicals from a broken LCD panel are toxic when ingested. Use caution when handling
a GUI with a broken display panel.
• Replacing some ventilator parts requires the use of leak detector fluid to ensure a good
gas seal. Use extreme caution when using leak detector in the vicinity of electronics.
Thoroughly dry all components following use of leak detector.
Warning
The failure to replace damaged warning, caution, and identification labels may result
in personal injury or equipment/property damage.
Performance
Flow sensor
calibration
calibration
calibration
Download
Software
cycle1
Type of service
Ventilator installation 1 -- 2 3 4 5 6 7 8 -- 92
10,000-hour preventive 1 -- 2 3 4 5 6 7 8 -- 9
maintenance (ventilator or
compressor)
Oxygen sensor 1 -- -- -- -- 3 -- 4 2 -- --
Warning
Connectors and tubes with the proprietary Bennett barbed cuff fittings are intended
for use only with like fittings. They are not interchangeable with ISO-standard cone
and socket fittings. A leak-tight connection cannot be ensured if these two fitting
types are combined. Adapters may be used to connect Bennett barbed cuff fittings to
ISO-standard cone and socket fittings.
There are two configurations of the GUI in the field: the older 9.4-inch GUI and the newer
10.4-inch GUI. Both GUI configurations have color LCD screens. Additionally, a Trending-
enabled (TE) GUI CPU PCB may be installed in either the 9.4-inch or 10.4-inch GUI. This PCB
contains an additional compact flash memory device used by the Puritan Bennett
840 Ventilator System Trending software option. The table below lists the major differences
between the versions.
Screen size 9.4 inch 9.4 inch 10.4 inch 10.4 inch
Communication Single null port Single null port Single null port and Single null port and
Ports two serial ports two serial ports
Minimum software No Yes; S/W revision V Yes; H or higher Yes; S/W revision V
level required or higher (US only or or higher (US only or
equivalent) equivalent)
Print Screen Yes; S/W revision H Yes Yes; S/W revision H Yes
function (US only or (US only or
equivalent) and equivalent) and
higher higher
NOTE:
Externally, there are no differences between the 10.4-inch GUI with or without the
TE CPU PCB, or the 9.4-inch GUI with or without the TE CPU PCB. Internally, the
GUI CPU PCBs differ in hardware components. Where applicable, the service
instructions make the distinction between the 10.4-inch GUI with Non-TE CPU PCB
and 10.4-inch GUI with TE GUI CPU PCB; or 9.4-inch GUI with Non-TE CPU PCB and the
9.4-inch GUI with TE GUI CPU PCB. In all other cases involving the GUI, the service
instructions for the respective 9.4-inch or 10.4-inch GUIs with TE CPU PCBs can be used
interchangeably with those for GUIs with Non-TE CPU PCBs.
The GUI, shown in Figure 8-1, is a detachable module that can be mounted on the optional
cart or shelf. When the ventilator is mounted on a shelf, an optional cable allows you to place
the GUI up to 10 ft from the BDU.
Touch screen
interface
Rotary
encoder
knob
Keyboard
Table 8-3 shows the GUI parts that can be replaced with the GUI mounted or detached from
the cart. Any other GUI parts must be replaced with the GUI detached and the rear housing
removed.
Keyboard assembly X
The service and repair procedures that follow are organized into procedures for the 10.4-inch
GUI and 10.4-inch GUI with TE CPU PCB (where applicable), followed by procedures for the
9.4-inch GUI and 9.4-inch GUI with TE GUI CPU PCB (where applicable).
8.13.1 Removing or installing the 10.4-inch GUI when mounted on RTA cart
1. Disconnect GUI interface cable from side of BDU (Figure 8-3).
2. If GUI is mounted to optional RTA cart, detach cable from cable guide at rear of cart.
3. Remove GUI from its mount:
a. Grasp handle at back of GUI.
b. Push GUI release knob to right to disengage rotor housing from mounting platform.
c. While holding knob to right, lift module up and away from mounting platform.
d. Release knob and it will snap back to left, locked position.
GUI interface
GUI GUI release knob
cable
interface
cable GUI
interface
connection
Warning
Lock cart in place by locking casters prior to servicing any part of the ventilator.
If GUI is mounted on optional Puritan Bennett 800 Series Ventilator Compressor Mount Cart
or Puritan Bennett 800 Series Ventilator Pole Cart, follow these steps:
1. Disconnect GUI interface cable from side of BDU (Figure 8-3)
2. Detach cable from p-clip on underside of cart handle (Figure 8-4).
p-clip
Figure 8-4. P-clip and GUI cable installed on Puritan Bennett™ 800 Series Ventilator Pole Cart
7. Facing the front of the ventilator, turn GUI clockwise as far as it will go; this pulls the GUI
cable as far as possible from its BDU connection and allows GUI the full range of motion
once connected to the BDU.
8. Remove p-clip from underside of handle (at back of ventilator) (Figure 8-4) and clamp
the cable into position, ensuring adequate strain relief by rotating GUI slightly counter-
clockwise so that the p-clip can be installed. The measurement from GUI handle to the p-
clip is approximately 15.5 inches.
NOTE:
Do not use existing shrink tubing on cable, if present, as a reference. Tubing is used
specifically for placing the GUI cable in the strain relief on the RTA cart.
Caution
Use care not to overtighten captive screws.
2. Align the tabs on the top handle into the slots in the bottom handle. Press handle into
place.
3. Insert screw and gently tighten until snug.
Caution
If you use a small flat-blade screwdriver to separate the rear and front housings, take
care not to damage the GUI front and rear panels.
8-01324
Figure 8-7. 10.4-inch GUI rear warning label and serial number tag
NOTE:
• Remove only those screws that go through the shield and not those located in
a cutout.
• On GUIs manufactured after June 2007, a grounding strap connecting the GUI
keyboard assembly to the EMI shield is present. Ensure that this strap is
fastened to the EMI shield as shown in Figure 8-8 upon reassembly.
EMI shield
GUI LED
cable
RS-232
Keyboard
grounding strap
Keyboard
PCB cable
GUI LED
PCB
Touchframe
PCB cable
Caution
When reassembling the EMI shield, do not apply excessive force to the shield in the
vicinity of the CPU that may permanently deform the thermal pad on the TE GUI CPU
PCB. Permanent deformation could reduce or eliminate the thermal pad contact area
between the CPU heatsink and the EMI shield, causing increased CPU operating
temperatures.
Caution
To maintain an isolated ground system, ensure the nylon washer is placed between the
inverter PCB and the standoff on the GUI CPU PCB when mounting the backlight
inverter PCBs on the GUI CPU PCB.
4. For each of the backlight inverter PCBs, disconnect the harness that connects the LCD to
the backlight inverter and the cable from the backlight inverter to the GUI CPU PCB
(Figure 8-9).
5. Remove each of the PCBs and place into a conductive bag.
backlight inverter/
CPU harness
LCD/backlight Backlight
driver harness driver PCB
(one of two)
Standoffs (2)
Caution
Handle the flex circuit cable with extreme care to prevent damage to the flex circuit
connectors.
NOTE:
If you are replacing a 10.4-inch GUI CPU PCB with a TE GUI CPU PCB, properly dispose
of the old RS-232 shield, four female screw-lock screws, and flex circuit; otherwise
retain for reassembly.
RS-232 shield
Screw-lock screws
Socket savers
Spacers
4. Using a Phillips screwdriver, remove the 10 screws securing the PCB to the GUI bracket.
5. Gently tilt the GUI CPU PCB to the right and disconnect the LCD panel flex cables (2) from
the board (J2, J11).
6. Remove the GUI CPU PCB and place into a conductive bag. Set the PCB aside.
NOTE:
If installing a new 10.4-inch Non-TE CPU PCB or TE GUI CPU PCB, ensure the touch
panel jumpers are set to the 10.4-inch touchframe style, “MKG TOUCH”
(B-C position), as shown in Figure 8-11. The five jumpers are located at the lower left
corner of the PCB.
B
Jumpers
C
Figure 8-11. GUI CPU PCB touch panel jumpers configured for “MKG TOUCH”
8.13.10 Replacing the 10.4-inch Non-TE GUI CPU PCB with the TE GUI CPU PCB
The 10.4-inch GUI CPU PCB is replaced by the TE GUI CPU PCB for use with the
Puritan Bennett™ 840 Ventilator System Trending software option. Hardware differences
between this PCB assembly and previous 10.4-inch GUI CPU PCBs assembly include:
• the addition of spacers and socket savers to the RS-232 connectors (J23 and J24)
• a new RS-232 shield with added EMI gaskets
• longer female screw-lock screws and lock washers
• addition of a thermally conductive pad to the CPU heatsink
• compact flash memory required for the Puritan Bennett 840 Ventilator System Trending
software option
NOTE:
You will need to re-use the backlight inverter PCBs, including nylon washers and
screws, from the old 10.4-inch GUI PCB (removed in Section 8.13.8).
To replace an older 10.4-inch GUI CPU PCB (Non-TE) with a TE GUI CPU PCB:
1. Ensure that the touch panel jumpers are configured for MKG Touch as shown in Figure 8-
11.
2. Place the TE GUI CPU PCB onto the CPU bracket, aligning holes in the PCB with standoffs
on the CPU bracket.
3. Using a Phillips screwdriver, fasten the TE GUI CPU PCB with the ten screws removed in
Section 8.13.9.
4. Place the four spacers over the standoffs that protrude through the two RS-232
connectors (J23 and J24) as shown in Figure 8-12, making sure that the counterbore-end
of the spacer is located over the standoff.
5. Push one socket saver onto each of the RS-232 connectors (J23 and J24) as shown in
Figure 8-12. Retain the four included female screw-lock fasteners and washers for
securing the RS-232 shield in 8.
Socket saver
Spacer
Standoff
RS-232 connector
Figure 8-12. Socket savers and spacers installed onto RS-232 connectors
6. Peel off the paper backing and adhere the EMI gaskets to the RS-232 shield, matching
the lengths of the gaskets with the corresponding surfaces of the shield (Figure 8-13).
3.6 in
2.5 in
3 in
7. Route the nurse call connector dust cap through the hole in the RS-232 shield while
installing the shield onto the PCB.
8. Fasten the RS-232 shield onto the PCB by installing the 4 screw-lock fasteners with lock
washers through the socket savers, spacers, and into the standoffs below. Tighten the
screw-lock fasteners evenly until snug. Do not overtighten.
9. Inspect the conductive thermal pad on the CPU (U45) for damage prior to assembly. If
the pad looks permanently deformed or damaged, replace it with a new one before
assembling. To do this, remove the damaged pad, then peel the adhesive backing from
one side of a new pad and adhere it to the CPU heatsink. Ensure that the pad fits entirely
on the heatsink with minimal overhanging edges (Figure 8-14).
10. Peel the remaining adhesive backing from the thermal pad.
Thermal pad
Caution
Do not squeeze the thermal pad. When assembled, this pad contacts the EMI shield to
conduct heat away from the CPU’s heatsink. Permanently deforming the pad may
reduce or eliminate contact with the EMI shield which may result in higher CPU
operating temperatures.
NOTE:
You must re-use the two backlight inverter PCBs with nylon screws and washers.
Follow the steps in Section 8.13.9 and steps 1-5 and 7-12 in Section 8.13.10 to remove and
replace the TE GUI CPU PCB.
8.13.12 Replacing the compact flash component (TE GUI CPU PCB only)
To replace the compact flash component:
1. Slide the component out of the socket (J28) towards the edge of the PCB (Figure 8-15).
2. Insert the new component into the socket. The socket is designed so that the component
can be inserted only one way.
Real-time clock
pin 1
2. Insert the new component, ensuring that pin 1 is properly oriented with the socket.
3. Set the ventilator time and date after replacing the real-time clock.
LCD
shield
EMI
gaskets Harness from
backlight
inverter to
LCD panel
LCD
mounting
screws
(4 ea.) CPU
shield
Caution
• The LCD panels are easily soiled or damaged. Use caution not to touch the surface of
the LCD panel or the inside of the GUI window as fingerprints will result.
• Do not allow the LCD panels to be exposed to the ambient environment any longer
than necessary as dirt and debris will be deposited on the LCD panels. Should this
occur, use an ESD-protected air source to gently blow away any dust or debris.
NOTE:
An updated LCD panel was released that requires additional drivers which are found
in US English software revision P or your country’s language equivalent. If you have
US English software revision P or later (or an equivalent version for your country’s
language), you may replace a single LCD panel, as drivers for both new and old LCD
panels are included in these versions. Contact Covidien Technical Support at
1.800.255.6774 to determine the correct software version for languages other than
US English.
1. With the LCD panels exposed, remove the four screws that secure each of the two LCD
panels in place. Set screws aside (Figure 8-17).
2. Gently lift each LCD panel while routing the cables through the panel.
Caution
The touchframe PCB should not be flexed or bent as damage will occur. Use caution
when handling and avoid touching emitters or receptors mounted on the board.
GUI LED
PCB
GUI LED
cable
GUI
Touchframe
alarm
PCB
Touchframe
cable Keyboard
cable
Figure 8-18. 10.4-inch GUI touchframe PCB and GUI LED PCB locations
Caution
When handling the GUI alarm assembly, be careful not to damage the speaker
element.
Speaker
harness
Speaker
element
GUI front
housing
Ground strap
Figure 8-21. How to attach the grounding strap to the EMI shield
3. Separate the grounding strap connectors (if not already separated) and fasten the end
of the grounding strap to the EMI shield with the screw you removed in 2. (Figure 8-21).
4. Reassemble the GUI rear housing, cable, and handle, making sure that the grounding
strap or connector is not twisted or pinched between the front and rear housings.
5. Orient the GUI assembly so that the keyboard is facing up.
6. Slide a flat tool under one edge of the keyboard and gently lift to break the adhesive
bond with the front housing.
7. Once the adhesive has been broken, disconnect the keyboard cable from the keyboard.
8. If you are replacing a newer style keyboard that has a grounding strap attached,
separate the grounding strap connectors.
9. Remove the old keyboard.
10. Clean the surface of the GUI housing with alcohol to remove any adhesive residue.
11. Slide the Clear/Accept and control key language inserts into the appropriate slots on the
new keyboard.
12. Connect the new keyboard’s grounding strap to the connector on the strap attached to
the GUI EMI shield. Position the grounding strap so that it is not twisted or pinched.
13. Connect the ribbon cable from the GUI CPU PCB to the keyboard. Verify that the ribbon
cable is in the upper row of the keyboard connector and is firmly latched into place.
14. Remove the adhesive cover strips from the keyboard.
15. Align the keyboard with the opening, sliding the PCB as far up as possible on the front
housing. Press firmly into place.
Caution
Do not touch the inside (coated side) of the GUI window. If necessary, use finger cots or
protective gloves. The window is treated for glare resistance and touching it creates
smudges, which cannot be removed without damaging the window.
Fascia
panel
GUI rear
panel
Rotor
Rotation
spring
Brake shoes
Caution
Be careful not to touch the inside (coated side) of the window. If necessary, use finger
cots or protective gloves. The window is treated for glare resistance and touching it
creates smudges which cannot be removed without damaging the window.
Bezel
Gasket
Screwdriver
placement Window
Window
gasket
LCD
panels
Caution
• Be careful not to touch the inside (coated side) of the window. If necessary, use
finger cots or protective gloves. The window is treated for glare resistance and
touching it creates smudges which cannot be removed without damaging the
window.
• Avoid cleaning the coated side of the window. If necessary, gently blow off dust or
debris using compressed air.
The window serves as the GUI touch screen; it is held in place by the bezel.
Caution
Be careful not to touch the inside (coated side) of the window. If necessary, use finger
cots or protective gloves. The window is treated for glare resistance and touching it
creates smudges which cannot be removed without damaging the window.
NOTE:
When installing the window, be sure the coated-side (with the painted black strip) is
facing to the inside of the unit.
GUI tilt
positions
NOTE:
The top of the bezel is narrower than the bottom. When replacing the bezel, position
the narrower end at the top.
Caution
Carefully remove the keyboard assembly to avoid damaging the GUI front housing.
2. Slide a flat tool under one edge of keyboard laminate and pull up to break keyboard's
adhesive bond to front housing.
3. Once the adhesive has been broken, disconnect ribbon cable from keyboard.
4. Continue to pull keyboard assembly away from front housing. Remove entire assembly
including keyboard laminate, keyboard PCB, and rotary encoder.
5. Gently remove keyboard assembly and place in conductive bag or on ESD-protected mat.
6. Using alcohol, clean any remaining adhesive from front housing, being careful not to let
debris fall into the GUI housing.
NOTE:
Do not attempt to reuse the 9.4-inch keyboard assembly after removal.
Keyboard
assembly
Keyboard
laminate
fig8-19c.tif
Caution
The keyboard assembly is held in place by a quick bonding adhesive that surrounds the
inside perimeter of the keyboard laminate. Once the adhesive is exposed, it
immediately bonds to anything it touches. Therefore, use extreme care when removing
the protective strip concealing the adhesive, and while aligning the keyboard laminate
to the GUI's front panel.
1. Align keyboard over housing and connect keyboard PCB ribbon cable (dark stripe goes
to pin 1 of GUI CPU PCB J20) (Figure 8-25). Verify that cable is positively locked into place.
2. On inside of keyboard laminate, carefully peel back protective strip that conceals
adhesive.
3. Taking care not to touch keyboard assembly to GUI front housing, align keyboard
laminate (starting with side containing rotary encoder) with keyboard inset on front
housing.
4. Carefully press keyboard in place by applying pressure to perimeter of laminate; then,
check keyboard perimeter to ensure it is securely bonded to GUI housing.
Cable
Cable
Mounting
Mounting guide
guide
bracket
bracket
GUI
interface
connection
Release knob
GUI
interface
connection
Caution
To prevent damage to ESD-sensitive components, always follow ESD guidelines when
disassembling the GUI.
Top handle
Interface cable
Bottom
handle
Captive screw
(2 PL)
Interface cable
connector
Figure 8-27. Removing 9.4-inch GUI handle assembly and interface cable
5. Loosen the two captive screws that retain RS-232 port cover; remove cover.
6. Unplug nurse’s call port cover.
7. Remove 19 screws from GUI rear housing (Figure 8-28) and set aside.
8. Gently separate GUI rear and front housing. If necessary, use a small flat-bladed
screwdriver to separate the GUI rear panel from the front.
Caution
If you do use a small flat-blade screwdriver to separate the GUI rear and front housings,
use care not to damage the GUI front or rear housing.
1-1/2-in. screw
(13 PL)
3/4-in. screw
3/8-in. screw
(5 PL)
NOTE:
The screws that secure the GUI rear housing to the front housing come in three
lengths. Refer to Figure 8-28 for correct lengths.
1. With GUI front housing face down, align rear and front housings.
2. Feed remote alarm cover through rear housing.
3. Check perimeter of housing to make sure cables are not being pinched.
4. Gently press rear housing into front housing until two enclosures fully seat.
5. Install GUI rear housing (Section 8.14.6.1).
Caution
Ensure cables are not pinched between the front and rear housings. Otherwise,
damage to the cables may result.
6. Install and hand-tighten 19 screws; then, using a screwdriver, tighten until snug. Do not
overtighten.
7. Insert the bottom handle into handle seating on rear housing (Figure 8-27). Install and
hand-tighten two screws; then, using a screwdriver, tighten until snug. Do not
overtighten.
8. Connect female end of GUI interface cable to GUI CPU PCB connector located inside
bottom handle. Tighten two interface cable captive screws to GUI CPU PCB until snug. Do
not overtighten.
9. Position the interface cable into the handle slots.
10. Snap top handle into place. Install screw and tighten until snug. Do not overtighten.
11. Install GUI (Section 8.14.5).
J3
J2
Backlight
inverter
PCB
Figure 8-29. 9.4" GUI interior (TE GUI CPU PCB shown)
Alarm
assembly
8.14.8 9.4-inch GUI backlight inverter PCB and GUI LED PCB
The backlight inverter PCB, which is part of the LCD backlight control circuit, controls the
brightness of both LCD panels.
The GUI LED PCB interface lets the GUI microprocessor control the discrete diagnostic LEDs on
the LED PCB.
Backlight inverter
PCB
Standoff
(4 PL)
GUI LED
PCB
Figure 8-31. Replacing 9.4-inch backlight inverter and GUI LED PCBs
CPU
Screw (13 PL) protective
shield
J20
J19
Keyboard
cable
Touch
frame
cable
Caution
When reassembling the EMI shield, do not apply excessive force to the shield in the
vicinity of the CPU that may permanently deform the thermal pad on the TE GUI CPU
PCB. Permanent deformation could reduce or eliminate the thermal pad contact area
between the CPU heatsink and the EMI shield, causing increased CPU operating
temperatures.
1. Align GUI EMI shield to 13 screwholes on CPU bracket (Figure 8-32). Install and hand-
tighten 13 screws; then, using a screwdriver, tighten until snug. Do not overtighten.
2. Connect these cables to bottom of GUI (Figure 8-28):
• Touchframe PCB ribbon cable to GUI CPU PCB connector J19
• Keyboard ribbon cable to GUI CPU PCB connector J20
8.14.10.1 Removing 9.4-inch VGA LCD controller PCBs (older CPU PCB)
1. Remove GUI rear housing (Section 8.14.6).
2. Remove CPU EMI shield (Section 8.14.9.1).
3. Grasp sides of desired VGA LCD controller PCB and pull up to disengage pins from GUI
CPU PCB connectors (Figure 8-33). Place in conductive bag or on ESD-protected mat.
VGA LCD
controller
PCBs
J5
J4
J15
J14
8.14.10.2 Installing 9.4-inch VGA LCD controller PCBs (older CPU PCB)
NOTE:
The two plug-in IP interface VGA LCD controller PCBs are identical and can be
mounted at either GUI CPU PCB location (J14/J15 or J4/J5). The pin sets at each end of
the VGA LCD controller PCBs are different, preventing the VGA LCD controller PCBs
from being mounted backward on the GUI CPU PCB.
1. Orient pins on VGA LCD controller PCB so they align with corresponding connectors on
GUI CPU PCB (Figure 8-33).
2. Gently press VGA LCD controller PCB until fully seated.
3. Install GUI EMI shield (Section 8.14.9.2).
4. Install GUI rear housing (Section 8.14.6.2).
4. Disconnect the backlight extender cable assemblies from the backlight inverter PCB
connectors J2 and J3.
5. Disconnect these cables from bottom of GUI (Figure 8-29):
• Touchframe PCB ribbon cable to GUI CPU PCB connector J19
• Keyboard ribbon cable to GUI CPU PCB connector J20
6. Remove the eleven (11) screws that secure GUI CPU PCB/bracket assembly to the
standoffs in front housing.
7. Lift the CPU bracket assembly off standoffs.
Caution
When removing the GUI CPU PCB, be careful not to damage surface components. Do
not flex the PCB.
Jumpers
1. Align GUI CPU PCB to eleven (11) screwholes in mounting bracket (Figure 8-29). Install
and hand-tighten 11 screws. Then use a screwdriver to tighten the screws until snug. Do
not overtighten.
2. Connect the backlight extender cable assemblies to the backlight inverter PCB
connectors J2 and J3.
3. Connect these cables to GUI CPU PCB:
4. If you are installing a TE GUI CPU PCB, inspect the conductive thermal pad on the CPU
(U45) for damage prior to assembly. If the pad looks permanently deformed or damaged,
replace it with a new one before assembling. To do this, remove the damaged pad, then
peel the adhesive backing from one side of a new pad and adhere it to the CPU heatsink.
Ensure that the pad fits entirely on the heatsink with minimal overhanging edges
(Figure 8-35).
NOTE:
If you are replacing an older GUI CPU PCB with a TE GUI CPU PCB, you will only need
the thermal pad from the bag of included hardware shipped with the GUI PCB FRU
kit.
Thermal pad
Caution
Do not squeeze the thermal pad. When assembled, this pad contacts the EMI shield to
conduct heat away from the CPU’s heatsink. Permanently deforming the pad may
reduce contact with the EMI shield which may result in higher CPU operating
temperatures.
6. Align CPU EMI shield to 13 screwholes on CPU bracket (Figure 8-32). Install and hand-
tighten 13 screws. Then, use a screwdriver to tighten the screws until snug. Do not
overtighten.
7. Connect these cables to bottom of GUI (Figure 8-29):
• Touchframe PCB ribbon cable to GUI CPU PCB connector J19
• Keyboard ribbon cable to GUI CPU PCB connector J20
8. Install GUI rear housing (Section 8.14.6.2).
8.14.12 Replacing the compact flash component (TE GUI CPU PCB only)
To replace the compact flash component:
1. Slide the component out of the socket (J28) towards the edge of the PCB (Figure 8-36).
2. Insert the new component into the socket. The socket is designed so that the component
can be inserted only one way.
Real-time clock
pin 1
Note: On older GUI
CPU PCBs, clock is
located at U4.
Figure 8-37. GUI CPU PCB real-time clock (TE GUI CPU PCB shown)
2. Insert the new component, ensuring that pin 1 is properly oriented with the socket.
3. Set the ventilator time and date after replacing the real-time clock.
Caution
• Handle the touchframe PCB with care: do not flex the assembly, damage any of the
surface components, or touch emitters or receivers.
• Take care when disconnecting the LCD harnesses; do not pull on wires. These
harnesses are delicate.
Touchframe PCB
LCD
adapter
Backlight
cable
extender
cable LCD
assemblies panels
LCD
harnesses
LCD
adapter
CPU bracket cable
assembly
LCD
harnesses
Figure 8-38. Replacing the 9.4-inch touchframe PCB and LCD panel assembly
Short
cable
Long
Foam
cable
side
Standoff
Plastic spacer
Caution
• Avoid touching the front of the LCD panels. Smudges are difficult to remove.
• Use care not to scratch the front of LCD panels.
NOTE:
LCD panels are not interchangeable due to backlight wire lengths. Do not reverse
displays. Complete the backlight replacement for one LCD panel first, then complete
the second backlight replacement.
7. Gently lift the LCD panel from the bracket and place the panel face down on an ESD-
protected mat.
8. Remove the LCD interface PCB (with LCD harnesses attached) from the rear of the LCD
panel and set aside (Figure 8-40).
9. Remove the backlight panel (Figure 8-40):
a. Using a small flat-bladed screwdriver, straighten the small metal tabs around the
perimeter of the LCD panel.
b. Carefully remove the backlight panel.
c. Discard if replacing the panel.
LCD
interface
PCB
Detail:
Bottom of LCD
Backlight
interface PCB
panel
Reattach to
backlight cable
extender assembly
(not shown)
Bend and
straighten metal
tabs and secure
backlight panel
LCD panel
Caution
To prevent damage to filter-retaining studs, do not attempt to remove the pushnuts
that hold the foam filters in place.
Caution
Ensure cables are not pinched between the front and rear housings as you reassemble
the GUI unit. Otherwise, damage to the cables may result.
Pushnut
Foam
filter
Foam
filter Pushnut
Fascia
panel
GUI rear
panel
Spring
The BDU is a removable unit that can be mounted to the optional cart or shelf (using the
shelf-mount kit).
Many BDU parts can be replaced without removing the unit from its mounting platform.
Table 8-4 indicates the level of disassembly required for all BDU field-replaceable units (FRUs).
As you conduct the following replacement and installation procedures for the BDU, you may
find it useful to refer to Table 2-2, and Figure 2-11 in Chapter 2 of this manual.
Caution
To prevent damage to ESD-sensitive components, always follow ESD guidelines when
servicing the BDU.
Modules removed
BDU
replacement part Power Inspiratory Exhalation BDU Card cage
BDU BPS
supply module module housing PCBs
AI PCB X
BD CPU PCB X
Real-time clock X
(on BD CPU PCB)
Power supply X
assembly
Power indicator X X X X X X X
Oxygen filters
(F1/F3)
Pressure switches
(PS1/PS2)
PSOL cartridges X
(PSOL1/PSOL2)
Inspiratory X
electronics PCB
Oxygen sensor X
Safety valve X
Inspiratory check X
valve (CV3)
Regulators (REG1/ X
REG2)
Inspiratory X
pressure
transducer
autozero
solenoid (SOL1)
Exhalation valve
(EV)
Modules removed
BDU
replacement part Power Inspiratory Exhalation BDU Card cage
BDU BPS
supply module module housing PCBs
Exhalation flow
sensor (Q3)
Exhalation X
transducer PCB
Filter seal
Expiratory X
pressure
transducer
autozero
solenoid (SOL2)
Exhalation heater X
Exhalation check X
valve (CV5)
Sample port X
Motherboard PCB X X X X X X X
BDU alarm X X X X X X
assembly
Alarm blindmate X X X X X X X
cable
Inspiratory X X X X X X X
blindmate cable
DC power supply X X X X X X X
blindmate cable
AC power supply X X X X X X X
blindmate cable
AC panel X
Humidifier X
receptacle
Warning
To prevent personal injury, take appropriate measures when removing the BDU from
its mounting platform as the unit weighs 18.2 kg (40.1 lb) when fully loaded.
Caution
Use extreme care when moving the BDU. Do not jar or drop the unit.
2. Pull out BDU release handle (Figure 8-46) to disengage BDU locking bracket; while
holding handle out, slide BDU off bracket while resting it on edge of mounting
platform. Release handle.
3. Carefully lift BDU off platform from rear of ventilator, and place on a firm, clean surface.
GUI cable
BDU
release
handle Compressor
power cord
Older field units can be retrofitted with this new combination, using a FRU kit version. (Refer
to Chapter 9 for part numbers.)
1. Remove the two screws securing the current bracket to AC receptacle of the BDU.
2. Install the Sure-Lock™* retainer over the receptacle and secure with the two screws
provided.
3. Insert the power cord and ensure that it clicks firmly into place.
Warning
Install only ventilator BDUs with serial numbers starting with 3512 onto the newer
Puritan Bennett 800 Series Ventilator Compressor Mount Cart and Puritan Bennett 800
Series Ventilator Pole Cart. Other ventilator serial numbers are not compatible with the
newer carts.
1. Lifting BDU from rear, slide BDU onto mounting bracket from front of cart.
2. Make sure BDU is locked in place by pushing and pulling it on slide.
3. Connect cables from GUI, BPS, compressor, and AC power, as required (Figure 8-44)
8.15.6 Analog interface (AI) PCB and breath delivery (BD) CPU PCB
The AI and the BD CPU PCBs (Figure 8-47) reside in the card cage of the BDU.
Data
key
Looking into the card cage, the AI PCB sits in the far right slot (toward the front of the
ventilator), and the BD CPU PCB sits in second slot from the right (behind the AI PCB). Refer
to Figure 8-47.
Caution
The data key is matched to each ventilator. Ventilator is inoperable without the correct
data key installed. Do not lose the data key.
Card
cage
BD CPU
PCB
2 PL
AI PCB
8 PL
I/O panel
Test button
PTS 2000
connector
Data key
GUI
connector
Compressor
Data connector
Real-time clock
pin 1
2. Insert the new component, ensuring that pin 1 is properly oriented with the socket.
3. Set the ventilator time and date after replacing the real-time clock.
Caution
Grounding of the BDU can be compromised is the I/O panel is incorrectly installed.
5. Using a 3/16-in. nutdriver, install two latching post screws that secure panel to PTS 2000
connector, and tighten until snug. Do not overtighten.
6. Insert data key, install data key cover, and connect cables to panel.
Caution
To prevent electrical shock hazard, always unplug the power cord from facility power
and disconnect the BPS cable from the power supply before servicing the power supply.
Warning
To prevent electrical shock hazard, always unplug the power cord from facility power
and disconnect the BPS cable from the power supply assembly before servicing AC
components.
Power switch
2. From front of BDU, orient power switch so “|”is at top of switch, and slide switch into
slot.
3. From inside power supply compartment, connect harness to power switch (Figure 8-52).
4. Install power supply (Section 8.15.7.2).
Caution
Route the humidifier cables so they do not obstruct the AC power supply harness and
blindmate bracket, located at the bottom rear of the power supply compartment.
NOTE:
Consult this manual or the humidifier vendor for the appropriate bracket model
when mounting a humidifier to the ventilator.
8.15.9.3 AC panel
8.15.9.3.1 Removing AC panel
1. Remove inspiratory module (Section 8.15.10.8). This is necessary to provide access to the
panel.
2. If connected, disconnect ventilator and compressor (if installed) power cords from AC
panel, and BPS cable from BDU.
3. Remove six screws that secure panel (Figure 8-55) to BDU chassis (Two screws also secure
BPS connector cover to panel).
4. Taking care not to disturb electrical connections on inside of AC panel, gently separate
panel from chassis.
5. To provide better access to component connections, cut tie wrap that secures wires to
wall of AC power distribution compartment.
6. Using needlenose pliers, disconnect the wires from the AC panel components:
• Red wire from terminal 10 of relay
• Black wire from terminal 12 of relay
• Blue wire from top left-hand terminal of AC filter PCB
• Brown wire from top right-hand terminal of AC filter PCB
• Brown wire from middle left-hand terminal of compressor AC receptacle
• Blue wire from middle right-hand terminal of compressor AC receptacle
7. Using a 5/16-in. nutdriver, disconnect ground wires from ground stud of AC panel.
8. Remove AC panel.
Compressor AC receptacle
Warning
Replacing key inspiratory module components requires the use of leak detector fluid to
ensure a good gas seal. Use extreme caution when using leak detector in the vicinity of
electronics. Thoroughly dry all components following use of leak detector.
4. Place the third core in the middle of the oxygen transducer harness, using care not to
pinch or damage the wires.
5. Visually inspect the cables and ensure that both of the latches are secure and that no
stress is being placed on any discrete wires.
6. Reinstall the fascia panel (Section 8.15.10.5.2).
Caution
When replacing damaged o-rings, carefully remove and install to prevent damage to
the manifold ports.
Check for damaged o-rings as you disassemble the inspiratory module. If o-rings are
damaged, pry them off of the ports. Lubricate replacement o-rings with Krytox™* grease,
then gently stretch o-ring over base of the port and position in place.
Allow replaced o-rings to settle in place; then verify their integrity.
NOTE:
Replace the air inlet filter (F2) and o-ring every 10,000 hours. They are part of the
10,000-hour preventive maintenance kit.
8.15.10.4.1 Removing F2
1. Make sure air source is disconnected from ventilator.
2. Drain any condensate from air inlet filter bowl by unscrewing captive drain plug at
bottom (Figure 8-57). Tighten plug.
3. Unscrew bowl.
4. Unscrew filter.
Bowl
Top
Groove
Filter
Air filter
assembly
(F2)
o-ring
Drain
plug
8.15.10.4.2 Installing F2
1. Make sure o-ring is seated in unthreaded groove atop F2 (Figure 8-57).
2. Screw filter in place until snug. Make sure o-ring makes contact with BDU chassis.
3. Screw clear bowl in place over filter until snug.
4. Using leak detector fluid (P/N 4-004489-00), verify that bowl does not leak.
Fascia
panel
NOTE:
Replace the oxygen impact filter (F1), oxygen inlet filter (F3) with a new o-ring and
spring every 10,000 hours. (These items are part of the 10,000-hour preventive
maintenance kit.)
Oxygen
filter
assembly
Spring
Caution
Use care when removing and installing the PSOL cartridge. Mishandling can damage the PSOL
manifold or cartridge, resulting in leaks.
Caution
Place the PSOL cartridge on a clean work surface. Do not drop the assembly.
5. Carefully pull cartridge out of PSOL manifold and set on a clean work surface. Verify that
all four o-rings are present and in good condition.
Nylon Nylon
line line
Oxygen
PSOL
(PSOL1)
Air PSOL
(PSOL2)
Inspiratory
module
NOTE:
If you are replacing a BDU cover without an access port with a BDU cover that has an
access port, you must also replace the inspiratory electronics PCB (see Section
8.15.10.14.1and Section 8.15.10.14.2).
3. Pinch the bottom and top flaps of the access cover firmly together and pull the access
cover away from the housing to remove.
NOTE:
The access cover is permanently attached to the instrument by a retaining strap.
4. Press wire retainer tab away from the connector to release the sensor cable connector.
and gently remove the connector from the oxygen sensor.
Access port
cover
Cover
retaining
strap
Sensor
Check Valve
harness
Housing
Oxygen
Sensor sensor
harness
connector
Caution
Ensure that the oxygen sensor o-ring is properly seated on the sensor before installing
in the ventilator. Failure to properly seat the o-ring may result in leaks.
7. Insert the threaded end of the oxygen sensor into the cavity on the check valve housing
and finger-tighten the sensor without using excessive force. As the sensor is installed,
ensure that it is not cross-threaded.
Caution
To prevent cracking of the sensor body, do not overtighten as you screw it into place on
the check valve housing.
8. Connect the sensor harness connector to the oxygen sensor with the ridge on the cable
connector oriented towards the wire retainer tab on the oxygen sensor.
9. Replace the access port cover by first sliding its top flap into the opening on the top of
the BDU housing.
10. Using both thumbs, seat the port cover in the port by press the two outside corners of
the lower flap at the juncture of the housing edge, fitting the into the housing opening.
11. Continue using both thumbs and firmly press the lower flap into place, working around
the flap from the outside corners to the bottom center to seal the access cover. Be sure
that the cover properly seals the housing opening.
12. Calibrate oxygen sensor by pressing 100% O2/CAL 2 min key. Verify that the calibration
passes.
13. Run an SST to check the system before placing the ventilator in service.
Oxygen
sensor
Caution
Ensure that the oxygen sensor o-ring is properly seated on the sensor before installing
in the ventilator. Failure to properly seat the o-ring may result in leaks.
2. Insert the threaded end of the oxygen sensor into the cavity on the check valve housing
and finger-tighten the sensor without using excessive force. As the sensor is installed,
ensure that it is not cross-threaded.
Caution
To prevent cracking of the sensor body, do not overtighten as you screw it into place on
the check valve housing.
NOTE:
If replacing an earlier version of the inspiratory check valve assembly (silver) with a
newer version (black), ensure the proper fit of the inspiratory filter. Otherwise, the
patient circuit may leak.
3 screws on
left side of
left-side plate
1 screw at front
of left-side plate
2 screws at rear
of left-side plate
NOTE:
If seal is damaged, refer to Chapter 9 for replacement adhesive tape.
PSOL
manifold
Remove
(4 PL)
NOTE:
The PSOL manifold slides between the left- and right-side plates. If the fit is too tight,
loosen the flat-head screw on the left side plate, allowing the plate to be lifted
slightly to provide sufficient clearance for the manifold.
1. Align flow sensor tubes on PSOL manifold to vents on air sensor manifold and gently
press manifold until tubes fully seat into vents.
2. Install four flat-head screws that secure PSOL manifold to right- and left-side plates
(Figure 8-69). Tighten screws. Do not overtighten.
3. If third flat-head screw on left-side plate was loosened to install PSOL manifold, tighten
until snug.
4. Connect tube attached to top of SOL1 to beneath PSOL manifold body.
5. Make these connections to inspiratory electronics PCB:
• Safety valve harness to P9 (this connector is located between P1 and P10)
• Oxygen sensor harness to P1
• Air PSOL (PSOL2) harness to P6
• Air flow sensor (Q2) harness to P4
• Pressure switch (PS1 and PS2) harness to P7
• Oxygen flow sensor (Q1) harness to P3
• Oxygen PSOL (PSOL1) harness to P5
Safety
valve
Air flow
sensor
(Q2)
Oxygen
flow
sensor
(Q1)
Figure 8-70. Replacing oxygen and air flow sensors (Q1 and Q2) and safety valve
o-ring
o-ring
Figure 8-71. Oxygen and air flow sensors (Q1 and Q2)
Caution
To prevent component damage when installing the safety valve, do not apply leak
detector fluid.
8.15.10.19 Check valve assembly, regulator assembly, and flow sensor manifold
8.15.10.19.1 Removing check valve assembly, regulator assembly, and flow sensor
manifold
1. Remove right-side plate (Section 8.15.10.13.1) and left-side plate (Section 8.15.10.15).
2. Remove PSOL manifold (Section 8.15.10.16.1).
3. Remove four inset screws that secure check valve assembly and flow sensor manifold to
inspiratory floor (Figure 8-73).
4. Disconnect oxygen vent tube from inspiratory floor.
5. Pulling in direction of check valve assembly, remove flow sensor manifold/regulator
assembly/check valve assembly from plate.
6. Pull flow sensor manifold/regulator assembly/check valve assembly apart into three
pieces as shown.
8.15.10.19.2 Installing check valve assembly, regulator assembly, and flow sensor
manifold
1. Make sure four o-rings on check valve assembly (Figure 8-73) are present and properly
positioned (two o-rings on air and oxygen ports and two o-rings in transfer tube port
cavities).
2. Align air and oxygen ports (on check valve assembly) to air and oxygen ports on
regulator assembly.
3. Gently press check valve assembly against regulator assembly until it reaches stop post.
NOTE:
It may be necessary to flex the inspiratory floor plate down in order to seat the check
valve assembly to the regulators.
4. Install two screws that secure check valve assembly to floor. Tighten screws until snug. Do
not overtighten.
5. Install PSOL manifold (Section 8.15.10.16.2).
6. Install right-side plate (Section 8.15.10.13.2) and left-side plate (Section 8.15.10.15.2).
Regulator
assembly
Check
valve
assembly
Flow
sensor
manifold
Inspiratory
plate
Inspiratory
floor plate
Regulator
Check assembly
valve
assembly Flow
sensor
manifold
REG1 REG2
REG2
REG1 SOL1
Regulator
plate
SOL1
NOTE:
REG1 and REG2 are not interchangeable and must be installed in their appropriate
locations. REG1 has an output port, which distinguishes it from REG2.
1. Insert regulator in appropriate cutout in regulator plate, and align mounting screw to
plate (Figure 8-74). Using 7/16-in. socket or open-end wrench, install retaining nut and
tighten until snug.
2. If you are installing REG1, connect vent tube to REG1 output port, at bottom of
inspiratory floor plate.
3. Install regulator assembly with plate (Section 8.15.10.19.2).
Caution
To prevent damage to tubing, do not attempt to remove silicone tubing from solenoid
barbed fittings. Solenoid FRUs have tubing preinstalled.
NOTE:
To ensure that all patient circuit connections are leak-tight, perform a circuit leak test
by running SST every time you install the filter on a ventilator. Consult the
Puritan Bennett™ 800 Series Ventilator System Operator’s and Technical Reference
Manual.
Remove
screw
Fascia panel
Exhalation
Expiratory pressure check valve
transducer autozero (CV5)
solenoid (SOL2)
Ferrites
Exhalation
flow sensor
(Q3)
Exhalation
valve (EV) Exhalation port
tube
Exhalation
module
harness
Exhalation
transducer PCB
Exhalation
module
harness
Fascia Panel
Exhalation
Sample Port
5. Remove and discard the sample port tubing and the Luer fitting components.
6. Cover the sample port opening in the fascia panel with the label provided with the new
exhalation valve FRU.
7. Remove exhalation valve from the exhalation module (Section 8.15.11.5.3).
8. Install new exhalation valve (Section 8.15.11.5.4).
9. Install exhalation module in BDU (Section 8.15.11.4).
Caution
• To prevent equipment damage, do not attempt to disassemble the exhalation valve.
It is factory-tested as an assembly.
• When removing the exhalation valve, take care not to damage the flow sensor that
sits atop the exhalation valve. Do not drop or jar the valve.
Expiratory Exhalation
pressure heater
transducer assembly
autozero
solenoid (SOL2)
Exhalation
flow sensor
(Q3)
Exhalation
valve (EV)
Exhaust
port and tube Exhalation
module
harness
Exhalation
transducer PCB
Caution
• When installing the exhalation valve, take care not to pinch or damage the o-ring at
the bottom of the flow sensor tube.
• Note that improperly installing the exhaust port can result in oxygen accumulation
in the exhalation compartment.
NOTE:
If replacing the exhalation valve with the newer version, note the absence of the
metabolic port. Refer to Section 8.15.11.5.1 for details regarding the elimination of
the metabolic port.
Caution
Do not attempt to remove the seal from the poppet for cleaning. Otherwise,
permanent damage to the seal may result.
Caution
Care should be taken to prevent dislodging the seal from the poppet while blowing.
Caution
When disconnecting Q3, be careful not to jar or drop it.
Caution
Ensure that no damage to the o-ring occurs during installation of Q3.
5. Connect cable to connector on Q3. On Puritan Bennett™ 840 Ventilator Systems with an
EMI shield and EMI exhalation harness assembly, install the shield so that it encloses the
replacement EMI harness assembly flow sensor connector, making sure that the harness
and the ground shield cable are routed as shown in Figure 8-81 and Figure 9-19.
proper position
of heat shrink
tubing
Open end of
ground shield
secured
between post
and chassis
Exhalation
transducer
PCB
Caution
To prevent damage to tubing, do not attempt to remove silicone tubing from solenoid
barbed fittings. Solenoid FRUs have tubing preinstalled.
Caution
The exhalation heater may be hot. Be careful when removing it.
8.15.11.9.1 Removing exhalation heater (EXH HTR) and check valve (CV5)
1. Remove exhalation module (Section 8.15.11.3).
2. Disconnect exhalation heater (Figure 8-83) cable from exhalation module harness.
3. Loosen three screws that secure heater to module.
4. Gently lift heater to separate it from flow sensor tube. Place heater on flat surface.
5. Gently remove check valve from orifice on exhalation module.
Exhalation
heater
Exhalation heater
Figure 8-83. Replacing exhalation heater (EXH HTR) and check valve (CV5)
8.15.11.9.2 Installing exhalation heater (EXH HTR) and check valve (CV5)
1. Inspect seal surrounding orifice that supports exhalation check valve for cracks or
damage. Replace CV5 assembly if required (seal is part of assembly).
2. Make sure leaf is lying flat, then place check valve (Figure 8-83) (leaf side up) into groove
of seal surrounding orifice.
3. Make sure o-ring at top of flow sensor tube is in place.
4. Align heater over flow sensor tube and gently press heater assembly onto tube while
aligning three screw locations to exhalation module.
5. Tighten three screws until snug. Do not overtighten.
6. Connect heater cable to exhalation module harness.
7. Install exhalation module (Section 8.15.11.4).
Power
switch
Power
indicator
Gasket AC panel
Exhalation
I/O cable
connector
Warning
To reduce the risk of fire hazard due to oxygen enrichment in the power supply
compartment, make sure the gasket forms a complete seal between the exhalation
module and power supply compartment.
4. From inside exhalation compartment, install two screws and standoffs that secure
exhalation I/O cable connector to chassis. Make sure latching block “barbs” are facing
out.
5. Reach inside power supply compartment and connect harness to power indicator and
power switch.
6. Connect BDU LED PCB cable to PCB.
7. Install exhalation module (Section 8.15.11.4).
8. Install inspiratory module (Section 8.15.10.10).
9. Install power supply assembly (Section 8.15.7.2).
10. Install BDU (Section 8.15.4).
Caution
Be careful not to damage the upper and lower card guides.
3. Using 1/8-in. (3-mm) flat-bladed screwdriver, loosen two captive screws on each of four
blindmate cables connected to motherboard PCB (Figure 8-86). Disconnect cables.
4. Using 3/16-in. nutdriver, remove eight standoffs that secure motherboard PCB
connectors to chassis.
5. Remove seven screws that secure motherboard PCB to back panel of card cage
(Figure 8-85). Gently remove motherboard PCB through card cage opening.
6. Place PCB in a conductive bag or on ESD-protected mat.
Screw
(7 PL)
Card
guide
Motherboard
BDU alarm
assembly
Alarm
blindmate
cable
Standoff
(8 PL)
Inspiratory
blindmate
cable
DC power supply
blindmate cable
Exhalation
I/O cable
Warning
To prevent electrical shock hazard, always unplug the power cord from facility power
and disconnect the BPS cable from the power supply assembly before servicing AC
components.
Power indicator
filter Auxiliary
(compressor/humidifier)
circuit breaker (CB2)
To inspiratory
module Inspiratory
blindmate
cable
To motherboard PCB
Warning
To reduce the risk of fire hazard due to oxygen enrichment in the power supply
compartment, make sure the gasket forms a complete seal between the inspiratory
module and power supply compartment.
4. Place bracket over chassis studs and secure with three locknuts. Using 11/32-in. nutdriver
or socket wrench, tighten until snug (Figure 8-89). Do not overtighten.
5. Install BDU housing (Section 8.15.12.2).
DC power
To supply
power blindmate
supply cable
To
motherboard PCB
To AC power
power Bracket supply
indicator blindmate
harness
To AC
receptacle
To power
supply
To AC panel
ground stud
Warning
To reduce the risk of fire hazard due to oxygen enrichment in the power supply
compartment, make sure the gasket forms a complete seal between the inspiratory
module and power supply compartment.
Caution
The obstruction of the harness within the bracket could prevent proper alignment of
power supply contacts.
8. Using two small tie wraps, rebundle wires on either side of round grommet. Cut tie wrap
ends flush with the tie wrap body.
9. Install four chassis-retaining screws at rear of AC panel.
10. Install BDU housing (Section 8.15.12.2).
NOTE:
The 806 compressor unit is the replacement for the 804 compressor assembly.
The 806 compressor unit is a detachable module consisting of the compressor, compressor
PCB, accumulator, and heat exchanger. The compressor unit is mountable only when used
with the optional cart.
Most routine repairs and preventive maintenance to the compressor unit require that the
compressor be removed from the cart. This section describes how to perform service and
repairs on the 806 compressor, as well as performance verification.
To ensure that reassembly is correctly done, label all wires and tube positions before you
begin disconnecting parts. Take precautions to prevent dirt and other particles from entering
the interior of the compressor. Always properly dispose of the original parts that have been
replaced by new parts.
Replacement parts for the original compressor, model 804, are no longer available. Contact
Covidien Technical Support for additional information.
Be sure to read all applicable instructions completely before disassembly.
NOTE:
To disconnect a pneumatic hose held in place by a compressible clip, use flat-nosed
pliers to compress the nubs on the clamp. Slide the clamp up the hose 1 to 2 inches.
Compressor
handle
Compressor module
(slide forward)
Warning
To prevent personal injury and equipment damage, have someone assist you when
lifting the heavy ventilator and its components, including the 803 BPS.
Caution
When removing the fan hardware:
• Be careful not to drop the screws or washers into the plenum assembly and onto the
heat exchanger.
• Be careful not to misplace the star washers as they provide ground to the fans.
3. Facing the front of the compressor, remove the four (4) Phillips screws and star washers
that secure the fan to the top of the plenum assembly.
4. Disconnect the electrical cable from the fan.
5. Remove the fan.
Caution
Verify the flow direction of the fan prior to installing the fan in place. The flow should
be directed into the compressor compartment.
1. Based on flow direction, properly orient the replacement fan and place over the hole in
the plenum assembly.
2. Connect the fan cord to the fan and place the fan on the fan-support panel.
3. Reinstall the four (4) screws and star washers.
4. Secure the fan cable to the fan’s mounting screw with a cable tie.
5. Reinstall the baffle and intake filter.
6. Reinstall top cover and secure screws.
Fitting
and Accumulator
o-ring (4) intake port
NOTE:
• To prevent the shredding of the silicone hoses when disconnecting from
barbed fittings, gently pull while turning the hose.
• Use a flat-nosed pliers to compress the nubs on compressible clamp.
Solenoid filter
outlet port
Figure 8-96. Disconnecting filter outlet port from main supply hose
6. Disconnect the hose at the heat exchanger inlet connection that runs from the
compressor outlet (Figure 8-97).
Disconnect
inlet connection
here
7. Disconnect the hose at the water trap outlet that runs to the air dryer inlet port.
8. Loosen the two (2) Phillips screws that secure the internal plenum panel to the left and
right panels.
9. Loosen the two (2) Phillips screws that secure the external plenum panel the left and
right panels.
Pressure transducer
hose
Fan connectors
Motor
cable
8. Facing the rear of the compressor, place your thumbs under the edge of the plenum
assembly top panel and lift the panel free from the side panel.
9. Slightly lift (or prop open) the plenum assembly enough to gain access to the heat
exchanger ports.
10. Disconnect the hose at the heat exchanger outlet port that runs to the water trap inlet.
11. Taking care not to damage the heat exchanger fins, carefully lift the plenum assembly
from the top of the compressor and set assembly aside.
NOTE:
If you are installing a new PCB, it is necessary to transfer the compressor serial
number and operational hours information by removing the EEPROM from the old
board and transferring it to the new board.
Caution
Verify the polarity of the EEPROM before installing on the new PCB: When transferring
the EEPROM, ensure that the notch on the EEPROM matches the notch on the 806 PCB
before installing.
9. Reinstall the new board by first aligning the board with the side panel and placing the
bottom of the board into the board retainer.
10. Reattach the cable connections to the PCB as follows:
• Solenoid cable at J1
• Two (2) fan cables at J2 and J3.
• AC input cord at J6
• Motor cable at J5
11. Attach the data cable to the PCB and slide its lock into place.
8.17.16.1 Removing and reinstalling the air dryer, solenoid valve assembly, and
replacing filters
1. Remove the top cover assembly and set aside (Section 8.17.4).
2. Remove the rear panel and set aside (Section 8.17.7).
3. Cut the cable tie securing the air dryer assembly to the clamp (Figure 8-99).
Air Solenoid
dryer outlet
hose filter
Air dryer
muffler/filter
Solenoid
purge outlet
hose
Cable tie
18. Install the new filter by screwing it in place. on the air dryer assembly.
19. Reconnect the short hose from the air dryer to the solenoid purge outlet fitting.
Use a flat-nose pliers to slide the hose clamp over the fitting to hold the hose in place.
20. Reinstall the air dryer and solenoid valve assembly in the compressor.
21. Secure the assembly with a cable tie (Figure 8-99) by sliding the cable tie under the
bracket and around the air dryer assembly.
22. Tighten cable tie and cut off excess.
To accumulator
Unloading To pressure
transducer (PC)
solenoid
To solenoid (SOL3) Filter (F10)
From heat
exchanger
Air dryer Filter/muffler
Warning
To prevent personal injury and equipment damage, have someone assist you when
lifting the heavy compressor assembly.
7. Reinstall the compressor assembly by orienting the compressor assembly with the
compressor intake filter on the left.
8. Carefully lower the compressor into the compartment. Slightly tilt the compressor
forward to position in place.
9. Using a magnetic Phillips screwdriver, reinstall the four (4) original screws that secure the
compressor assembly to the base of the unit.
Water
Coalescing
filter
assembly
Coalescing
filter
element
Water
trap
5. Using a magnetic Phillips screwdriver, remove the two (2) screws that secure the water
trap bracket to the base of the module. Set the screws aside.
6. Place the water trap assembly on a work surface.
7. Unscrew the ring that secures the water trap bowl to the filter head.
8. Remove the bowl and ring and set aside.
9. Unscrew the filter element retainer and set aside.
10. Verify that the o-ring is in place around the filter head.
11. Install the new coalescing filter element.
12. Replace the original filter element retainer.
13. Reinstall the water trap bowl and hold it in place by reinstalling the original ring.
14. Using a magnetic Phillips screwdriver, remount the water trap assembly bracket to the
base of the compressor using the original two (2) screws.
15. Reinstall hose on outlet side of coalescing filter.
16. Reinstall air dryer (Section 8.17.16.1).
17. Reinstall the back panel assembly (Section 8.17.8) and the plenum assembly
(Section 8.17.11.1).
NOTE:
If replacing a side panel, use the spare piece of foam in the foam kit to configure the
panel to match the panel removed. Install the Tinnerman clips to the rear of the
panel.
NOTE:
The solenoid electrical cable also connects to the PCB at J1. Refer to reinstallation
instructions for the air dryer and solenoid valve assemblies in (Section 8.17.16).
3. While resting the bottom of the rear panel on the base of the unit.
4. Take the hose that connects the solenoid outlet to the accumulator inlet port and route
this hose behind the main supply hose.
5. While supporting the rear panel in place, reconnect the hose running from the main
supply tee to the accumulator inlet port.
6. Reposition the hose clamp over the inlet port to hold the hose in place.
7. Taking care not to crimp hoses or cables, press the back panel into place and secure to
the unit using the original six (6) Phillips screws.
8. Once the rear panel/accumulator assembly is secured in place, finish tightening the four
(4) Phillips screws the secure the plenum assembly to the side panels.
8.17.23 Replacing the main inlet filter and reinstalling the top
1. Reinstall the support baffle by setting on the inside slot at the top of the front panel.
It should fit easily into the slot; if not, turn the baffle over and try the other side.
2. Reinstall the compressor top cover assembly and secure in place by tightening the eight
(8) captive screws.
3. Once the top cover assembly is in place, install the new main inlet filter by pressing it
into the slot between the supporting baffle and the rear panel.
Ensure all filter edges are in place.
Warning
To prevent personal injury and equipment damage, have someone assist you when
lifting the heavy ventilator and its components.
1. Reinstall the compressor onto the cart and secure using the captive hardware on the
cart.
2. Reattach and secure the compressor data cable.
3. Reattach the compressor power cord.
4. Secure the power cord using the power cord retainer. Secure the retainer with two (2)
screws.
5. Reattach the compressor supply tube.
Warning
Lock
Lock cart in place by locking casters prior to removing and replacing BPS or working
with cart.
1. Loosen two captive screws that secure battery cable to connector on side of power
supply assembly. Disconnect cable.
2. Loosen two screws at rear of 802 BPS. Loosen three screws of 803 BPS to unfasten 803
BPS from mounting bracket.
3. Slide 802 BPS forward off slide mounts and place unit on a firm work surface (Figure 8-
103).
4. If removing the 803 BPS, slide the 803 BPS from its mounting bracket (toward the front
of the cart) and place on a firm work surface.
GUI mount
BPS
Rear caster
(2 PL)
Front caster
(2 PL)
Back of BPS
BPS
cable
• If present, disconnect and remove the compressor, 802 backup power source, and
accessory drawer from the cart.
1. Lay the empty cart down. Remove the two screws (four total) that attach the two
existing BPS mounting brackets to the cart. Remove the two screws that attach the
accessory drawer bracket to the cart (Figure 8-105).
2. Remove the three screws that attach the mounting bracket/backstop to the back of the
new BPS and set aside (Figure 8-106).
4. The mounting bracket ships with the backstop positioned for use with ventilators that
do not include a compressor. If the system does not have a compressor, skip to step 8.
5. If the system has a compressor, see 6.
6. The mounting bracket ships with the backstop positioned for use with a ventilator that
does not include a compressor. If the system has a compressor, position the bracket on a
flat surface with the backstop on top. Remove the four screws connecting the backstop
to the bracket and slide the backstop forward to align with the raised stops toward the
front edge on either side of the bracket (Figure 8-109).
7. Using the four screws that you removed, reattach the backstop to the bracket. Refer to
Figure 8-109 for the proper holes to use.
Raised stops
8. Position the mounting bracket with the backstop pointed towards the back of the cart.
The front edge of the bracket should be flush with the front edge of the cart (Figure 8-
110).
Front edge of
bracket flush
with front
edge of cart
9. Using six of the screws that were included, attach the mounting bracket to the cart.
Refer to the following figures to determine the proper holes to use for attachment:
10. System without compressor – Figure 8-111
11. System with compressor – Figure 8-112
12. Stand the cart upright with the front facing you, then lock the wheel casters. With the
power cord on the BPS positioned towards the back of the cart, slide the BPS onto the
mounting bracket until it meets the backstop. Use the three screws that you removed
from the backstop earlier to reattach the BPS to the bracket.
Caution
To prevent damage to ESD-sensitive components, always follow ESD guidelines when
disassembling the BPS.
Fascia
panel
4 PL
Front
panel
J1 Floor
assembly
Battery
J3 pack
Chassis
Nut
(4 PL)
BPS
PCB
PCB
Enclosure
8 PL
Figure 8-113. Replacing battery pack and BPS PCB (for use on RTA cart)
3. Align fascia panel to two screw locations on front plate. Install two flat-head screws and
tighten until snug. Do not overtighten.
4. Orient battery pack housing with two LEDs toward front of cart, and slide unit onto slide
mounts located under BDU platform.
5. Install BPS (Section 8.18.2).
Mounting bracket
Battery pack
Cover
Cable clamp
Panel end
BPS harness
assembly
Remove these
4 screws
8.19 Cart
Three optional cart styles are available for use with the Puritan Bennett™ 840 Ventilator
System: the RTA (ready-to-assemble) cart, the Puritan Bennett™ 800 Series Ventilator
Compressor Mount Cart, and the Puritan Bennett™ 800 Series Ventilator Pole Cart. The
Puritan Bennett 800 Series Ventilator Compressor Mount Cart which allows use with an 806
compressor, can be used with a one-hour battery or an optional four-hour battery in its BPS,
located in the BDU shelf. The Puritan Bennett 800 Series Ventilator Pole Cart cannot be used
with a compressor. This cart can also be used with a one-hour battery or an optional four-
hour battery, located in the cart base.
The Puritan Bennett™ 800 Series Ventilator Compressor Mount Cart and the
Puritan Bennett™ 800 Series Ventilator Pole Cart may not be available in all regions. Please
contact your local Covidien representative for more information.
Chapter 9 contains parts lists and assembly drawings for all carts available for the
Puritan Bennett 840 Ventilator System.
The BPS assembly in the Puritan Bennett™ 800 Series Ventilator Compressor Mount Cart is
shown in Figure 8-116. Refer to Figure 8-117 to connect BPS to Puritan Bennett 840 Ventilator
System when mounted on Puritan Bennett™ 800 Series Ventilator Compressor Mount Cart.
Figure 8-117. Connecting BPS to BDU on Puritan Bennett™ 800 Series Ventilator
Compressor Mount Cart
2. Remove 2 Phillips screws fastening the BPS to the BDU shelf at the rear of the cart.
3. Slide BPS assembly away from cart and rest on a firm surface.
4. Remove eleven screws from BPS top and front covers, and lift covers away from BPS
(Figure 8-119).
Eleven Screws
5. Cut the tie wrap securing the battery harness to the battery charging status indicator
bracket.
6. Remove battery retaining bracket (2 screws) from rear of BPS.
7. Disconnect battery connector from J2 on the battery PCB (Figure 8-126).
8. Loosen four screws holding front BPS bracket (two screws at the base of bracket and two
screws on sides of BPS enclosure).
9. Lift the battery out of the BPS.
Caution
Use care when installing the front cover to avoid damaging the battery charging status
indicators.
6. Verify or install the four-hour battery indicator label to front of BPS bezel (Figure 8-121).
7. Replace the BPS cover, fitting it over the battery charging status indicators and replacing
the eleven screws.
5. Remove the one-hour battery shield holding the battery in place (four screws).
NOTE:
If converting from a four-hour to a one-hour battery, order the one-hour battery
shield to hold the battery in place.
Route cable
in left-most
channel
Figure 8-124. Serial number label location Puritan Bennett™ 800 Series Ventilator
Compressor Mount Cart
Warning
Ensure the battery indicator label matches the battery installed. Unexpected power loss
when operating on battery leading to significant patient injury could occur if a label is
applied stating the ventilator has a four-hour battery when it actually has a one-hour
battery installed.
Caution
The nuts provide a ground to the heatsink. Ensure nuts are returned to the correct
position for reassembly.
J1 J3
tie wrap
J2
8.19.2 Removing and replacing the battery in the Puritan Bennett™ 800
Series Ventilator Pole Cart
The battery in the Puritan Bennett™ 800 Series Ventilator Pole Cart is located in the cart base.
Depending upon the customer order, there will either be a one-hour or four-hour battery
installed.
Warning
Ensure the ventilator is prevented from tipping over backwards.
2. Remove Phillips screw at bottom front of battery cover and remove cover by lifting up
and pulling forward.
3. Unfasten two Phillips screws and lift battery compartment up and tilt towards you to
open. Battery compartment will drop down to allow battery access.
4. Disconnect battery connector from extension cable at back of battery and remove
battery.
Figure 8-127. Securing Phillips screws on Puritan Bennett™ 800 Series Ventilator Pole Cart
battery carrier (four-hour battery shown)
3. Align shield posts with the captive thumbscrews and secure shield using two Phillips
screws.
4. Lift battery carrier up and tilt toward back of unit. Battery carrier will drop down and
latch.
5. Install two Phillips screws to secure in place (Figure 8-127).
6. Verify or install one-hour battery indicator label to front of cart (Figure 8-128).
7. Verify or install the serial number label (Figure 8-130).
Figure 8-130. Serial number label location Puritan Bennett™ 800 Series Ventilator Pole Cart
Warning
Ensure the battery indicator label matches the battery installed. Unexpected power loss
when operating on battery leading to significant patient injury could occur if a label is
applied stating the ventilator has a four-hour battery when it actually has a one-hour
battery installed.
NOTE:
If converting from a four-hour battery to a one-hour battery, order a one-hour
battery shield to hold the battery in place.
8.19.3 Charging circuit PCB removal and replacement for Puritan Bennett™
800 Series Ventilator Pole Cart
On the Puritan Bennett™ 800 Series Ventilator Pole Cart the BPS is incorporated into the cart
and uses a charging circuit located on the underside of the BDU shelf cover.
PCB cover
Nuts
4. Disconnect battery connector J2 and harness connectors J1 and J3 from the PCB
(Figure 8-132).
PCB assembly
Nylock nuts
Nylock nuts
5. Remove four Nylock nuts fastening PCB assembly to charging circuit and nylon washers.
6. Remove PCB assembly.
3. Place insert into cavity in cart, aligning screwholes in cart and insert.
4. From beneath the cavity of the cart, insert flat-head screw into insert and tighten until
snug. Verify insert is tight and does not rotate.
On newer, ready to assemble (RTA) carts, remove and install the flex arm inserts as follows:
1. Slide locking ring off the GUI mount.
2. Install new insert.
3. Reinstall locking ring.
Figure 8-136. Flex arm insert on Puritan Bennett™ 800 Series Ventilator
Compressor Mount Cart and Puritan Bennett 800 Series Ventilator Pole Cart
Parts List
Abbreviation Meaning
A/R As Required
BD Breath Delivery
EX LK External Lock
FIL Fillister-head
FH Flat-head
HEX Hexagonal
I/O Input/output
ID Inside Diameter
IN LK Internal Lock
OD Outside Diameter
PAN Panhead
RTA Ready-to-Assemble
SOC Socket
SR Split-ring
9.2 Puritan Bennett™ 840 Ventilator System patient system and accessories
Item no.
Part no. Quantity Description
(Figure 9-1)
4-072623-00 4 • Screw, FH, 8-32 x 3/8 (Attach blank humidifier plate to BDU
--
chassis.) (not shown)
16 -- 1 • • Plate with cutout, humidifier, English/French
• • Plate with cutout, humidifier, Spanish/Portuguese
17 4-075659-00 • Mounting kit, Fisher & Paykel™* humidifier, ventilator-
powered (120 V), cart-mount, Spanish/Portuguese
18 -- 1 • • Bracket assembly, Fisher & Paykel™* humidifier, cart-
mount, English/French
• • Bracket assembly, Fisher & Paykel™* humidifier, cart-
mount, Spanish/Portuguese
Item no.
Part no. Quantity Description
(Figure 9-1)
4-070034-00 2 • Screw, PAN, 6-32 x 0.38, sems (Retain connector lock) (not
--
shown)
4-018506-00 1 • Flex tube, 21 in. (53 cm) ("gold standard" test hose, for use
--
with EST) (not shown)
Figure 9-1. Puritan Bennett™ 840 Ventilator System patient system and accessories
9.2.1 Puritan Bennett™ 840 Ventilator System NeoMode patient system and
accessories
Item no.
Part no. Quantity Description
(Figure 9-2)
2
3
9.3 Flex arm assembly, oxygen and air hose assemblies, power cords
Item no.
Part no. Quantity Description
(Figure 9-3)
6 (3 PL)
5 (3 PL)
3 (3 PL)
8
2 (6 PL)
1 (3 PL)
10
7
11
12
13
Item no.
Part no. Quantity Description
(Figure 9-4)
1 4-001474-00 1 Hose assembly, oxygen, DISS female x DISS female, clear, 3-m (10-ft)
(for USA) (See Section 9.2 for NHA.)
2 4-074710-00 1 Hose assembly, oxygen, DISS female x DISS male, 3-m (10-ft) (for
Canada) (See Section 9.2 for NHA.)
3 4-074697-00 1 Hose assembly, oxygen, Air Liquide™* female x Air Liquide™* female,
3-m (10-ft) (for France) (See Section 9.2 for NHA.)
4 4-074711-00 1 Hose assembly, oxygen, SIS female x SIS female, 3-m (10-ft) (for
Australia, New Zealand) (See Section 9.2 for NHA.)
5 4-074698-00 1 Hose assembly, oxygen, NIST female x BOC male, 3-m (10-ft) (for
United Kingdom, Ireland) (See Section 9.2 for NHA.)
6 4-074715-00 1 Hose assembly, oxygen, DISS female x Dräger™* male, 3-m (10-ft) (for
Germany) (See Section 9.2 for NHA.)
7 4-074700-00 1 Hose assembly, oxygen, NIST female x NIST female, 3-m (10-ft) (for
Netherlands) (See Section 9.2 for NHA.)
8 4-074702-00 1 Hose assembly, oxygen, DISS female x DISS female, green, 3-m (10-ft)
(for Israel, Japan, Saudi Arabia) (See Section 9.2 for NHA.)
9 4-074705-00 1 Hose assembly, oxygen, DISS female x DISS female, white, 3-m (10-ft)
(for Egypt, India, Italy, Kuwait, Poland, Portugal, South Africa) (See
Section 9.2 for NHA.)
10 4-074708-00 1 Hose assembly, oxygen, DISS female x DISS female, blue, 3-m (10-ft)
(for Switzerland) (See Section 9.2 for NHA.)
1 (clear)
7
2
8 (green)
3
9 (white)
5 10 (blue)
Item no.
Part no. Quantity Description
(Figure 9-5)
1 4-006541-00 1 Hose assembly, air, DISS female x DISS female, clear, 3-m (10-ft) (for
USA) (See Section 9.2 for NHA.)
2 4-074709-00 1 Hose assembly, air, DISS female x DISS male, 3-m (10-ft) (for Canada)
(See Section 9.2 for NHA.)
3 4-074696-00 1 Hose assembly, air, Air Liquide™* female x Air Liquide™* female, 3-m
(10-ft) (for France) (See Section 9.2 for NHA.)
4 4-074712-00 1 Hose assembly, air, SIS female x SIS female, 3-m (10-ft) (for Australia,
New Zealand) (See Section 9.2 for NHA.)
5 4-074713-00 1 Hose assembly, air, NIST female x BOC male, 3-m (10-ft) (for United
Kingdom, Ireland) (See Section 9.2 for NHA.)
6 4-074714-00 1 Hose assembly, air, DISS female x Dräger™* male, 3-m (10-ft) (for
Germany) (See Section 9.2 for NHA.)
7 4-074701-00 1 Hose assembly, air, NIST female x NIST female, 3-m (10-ft) (for
Netherlands) (See Section 9.2 for NHA.)
8 4-074703-00 1 Hose assembly, air, DISS female x DISS female, black, 3-m (10-ft) (for
Israel, Japan, Kuwait, Poland, Portugal), 3-m (10-ft) (See Section 9.2
for NHA.)
9 4-074706-00 1 Hose assembly, air, DISS female x DISS female, white/black, 3-m (10-ft)
(for Egypt, India, Italy) (See Section 9.2 for NHA.)
10 4-074707-00 1 Hose assembly, air, DISS female x DISS female, yellow, 3-m (10-ft) (for
Switzerland) (See Section 9.2 for NHA.)
1 (clear)
8 (black)
9 (white/black)
10 (yellow)
6 11 (white)
Item no.
Part no. Quantity Description
(Figure 9-6)
1 4-078030-00 1 Power cord, universal, North America (See Section 9.2 for NHA.)
1/2 3
4 5
6 7
8 9
10
3
4
5
6
Item no.
Part no. Quantity Description
(Figure 9-7)
1. Contact your Covidien representative to obtain 806 compressor unit ordering information.
Major assemblies parts list (ventilators mounted on Puritan Bennett™ 800 Series Ventilator Compressor
Mount Cart)
Item no.
Part no. Quantity Description
(Figure 9-8)
-- -- -- Puritan Bennett 840 Ventilator System (See Section 9.2 for NHA.)
1 -- 1 • Graphic user interface (GUI): 10.4 inch
(See pages 9-30 to 9-41 for parts breakdown.)
-- 4-075234-00 2 • Screw, PAN, 4-40 x 0.312, sems (Attach data key cover) [Not shown]
3 -- 1 • 806 Compressor Unit (optional) (See Figure 9.4.10 for parts
breakdown.)1
4 10046822 1 • Cart assembly with one-hour battery (optional) (See Figure 9-34.)
10046823 Cart assembly with four-hour battery (optional) (See Figure 9-34.)
1. Contact your Covidien representative to obtain 806 compressor unit ordering information.
Figure 9-9. Major ventilator assemblies (shown on Puritan Bennett 800 Series Ventilator Pole Cart)
Major ventilator assemblies parts list (ventilators mounted on Puritan Bennett™ 800 Series Ventilator
Pole Cart)
Item no.
Part no. Quantity Description
(Figure 9-9)
-- -- -- Puritan Bennett 840 Ventilator System (See Section 9.2 for NHA
1 -- 1 • Graphic user interface (GUI): 10.4 inch
(See pages 9-30 to 9-41 for parts breakdown.)
-- 4-075234-00 2 • Screw, PAN, 4-40 x 0.312, sems (Attach data key cover) [Not shown]
3 10046826 1 • Cart with one-hour battery, Puritan Bennett 800 Series Ventilator
10046827 Pole Cart
Cart with four-hour battery, Puritan Bennett 800 Series Ventilator
Pole Cart
Item no.
Part no. Quantity Description
(Figure 9-10)
-- 10070839 1 Label kit, ventilator, 100 V, 50/60 Hz, Japanese (See Section 9.4
for NHA.)
10070700 Label kit, ventilator, 120 V, 60 Hz, English, US (See Section 9.4
for NHA.)
10072901 Label kit, ventilator, 220-240 V, 50/60 Hz, English (See Section
9.4 for NHA.)
Item no.
Part no. Quantity Description
(Figure 9-10)
Item no.
Part no. Quantity Description
(Figure 9-10)
Item no.
Part no. Quantity Description
(Figure 9-10)
12 10070888 1 • Label kit, 806 Compressor, 100 V, 50/60 Hz, Japanese (not
shown)
10070891 • Label kit, 806 Compressor, 220-240 V, 50/60 Hz, Polish (not
shown)
10070893 • Label kit, 806 Compressor, 220-240 V, 50/60 Hz, Russian (not
shown)
10070877 • Label kit, 806 Compressor, 220-240 V, 50/60 Hz, German (not
shown)
10070882 • Label kit, 806 Compressor, 220-240 V, 50/60 Hz, Spanish (not
shown)
10070880 • Label kit, 806 Compressor, 220-240 V, 50/60 Hz, Italian (not
shown)
13 -- 1 • Label, compressor lint filter, English/French
Item no.
Part no. Quantity Description
(Figure 9-10)
20
19
5
1
6
2
13
8-0002
9
8
14
10
11
Item no.
Part no. Quantity Description
(Figure 9-11)
Item no.
Part no. Quantity Description
(Figure 9-12)
Item no.
Part no. Quantity Description
(Figure 9-13)
4 (2 PL)
Item no.
Part no. Quantity Description
(Figure 9-14)
Item no.
Part no. Quantity Description
(Figure 9-15)
Item no.
Part no. Quantity Description
(Figure 9-16)
5 4-070036-00 1 • Screw, PAN, 6-32 x 0.75, sems (Attach bottom handle to rear
housing)
6 4-073007-00 1 • Handle, GUI bottom
7 -- 1 • Label, product information, GUI, English/French (Available
only as part of ventilator label kit.)
Item no.
Part no. Quantity Description
(Figure 9-16)
Item no.
Part no. Quantity Description
(Figure 9-17)
7
5
Item no.
Part no. Quantity Description
(Figure 9-18)
-- 4-079014-SP 1 Housing assembly, GUI back (See Section 9.4.8 for NHA.) (Does
not include labels. Order ventilator label kit to obtain product
information label, and GUI ports label. Refer to Section 9.4.3
for part numbers.)
1 -- 1 • Housing, GUI back
2 4-073022-00 1 • Fascia, GUI
3 4-070036-00 2 • Screw, PAN, 6-32 x 0.75, sems (Attach fascia to rear housing)
4 4-073019-00 1 • Rotor, GUI
5 4-075232-00 1 • Spring, rotation
6 4-073018-00 2 • Shoe, rotor-brake, GUI
-- 4-073042-00 30 • Nut, push (Attach to foam kit to rear housing) (not shown)
3 (2 PL)
Tilt/swivel
assembly
6 (2 PL)
Item no.
Part no. Quantity Description
(Figure 9-19)
Item no.
Part no. Quantity Description
(Figure 9-19)
Item no.
Part no. Quantity Description
(Figure 9-19)
45 (shielding 4-072600-SP 1 • Exhalation module (See Figure 9-20 for parts breakdown.)
not shown) or
10057338 • Exhalation module (See Figure 9-20 for parts breakdown.)
(Use only with serial numbers starting with 3512)
46 4-071524-00 1 • Label, alarm graphics, BDU, non-US
Item no.
Part no. Quantity Description
(Figure 9-20)
16 -- 1 • • Port, exhalation
19 -- 2 • • o-ring
Item no.
Part no. Quantity Description
(Figure 9-20)
-- 4-078159-00 1 • Epoxy, conductive (Use only with EMI cable and cover)
27 068227 1 • Shield, bottom
28 068228 1 • Shield, top
29 068184 1 • Shield, ground
30 T-T51668-00 2 • Washer, external lock
31 4-075234-00 2 • Screw, PAN, 4-40 x 0.312, sems
1
2 (3 PL)
NOTE:
Lubricate o-rings with
Krytox™* grease
(P/N 4-732130-00).
4
3
5
24
27
28 25
6
23
29 7
30
31 (2 PL)
(2 PL) 22
8
19
(2 PL)
17
18 9
21
(2 PL)
10 (3 PL)
14 15
16 26
20
(4 PL)
12
11
13
(3 PL)
To exhalation
I/O cable assembly
Item no.
Part no. Quantity Description
(Figure 9-21)
4-076800-SP Inspiratory module, DISS male body (See Section 9.4.9 for NHA.)
or
10057337 DISS male body (See Section 9.4.9 for NHA.) (Use only with
serial numbers starting with 3512)
4-071876-SP Inspiratory module, DISS female body (See Section 9.4.9 for
NHA.)
4-071869-SP Inspiratory module, Air Liquide™* body (See Section 9.4.9 for
NHA.)
4-071866-SP Inspiratory module, NIST body (See Section 9.4 for NHA.)
4-071862-SP Inspiratory module, CIG body (See Section 9.4.9 for NHA.)
1 -- 1 • Plate, inspiratory module right side
2 4-071860-00 3 • Screw, FH, 8-32 x 0.5 (Attach right side plate to PSOL
assembly)
3 4-070041-00 3 • Screw, PAN, 8-32 x 1/2, sems (Attach right side plate to floor
assembly)
4 4-070100-SP 1 • PCB, inspiratory electronics
13 -- 1 • Bracket, regulator
Item no.
Part no. Quantity Description
(Figure 9-21)
1
21
(5 PL)
2
NOTE:
(3 PL)
Lubricate o-rings with
Krytox™* grease
(P/N 4-732130-00).
3
(3 PL)
4
20
19
(3 PL)
6 (2 PL) 5 (2 PL)
Oxygen 7
22
17 8
Air
18
(3 PL)
9
15 16 (2 PL) 10
11
14 13 12
(2 PL)
Item no.
Part no. Quantity Description
(Figure 9-22)
10 -- 1 • • o-ring, 21-mm
11 -- 1 • • o-ring, 29-mm
12 -- 1 • • Valve, PSOL
15 -- 1 • • o-ring
16 -- 1 • • o-ring
6
3
4
19
5
7
18
(4 PL)
9 8 (2 PL)
10
11
12
NOTE:
Lubricate o-rings with
Krytox™* grease
(P/N 4-732130-00).
16
15
13
14
17
(2 PL)
Item no.
Part no. Quantity Description
(Figure 9-23)
6 -- 1 • • Tube, bleed
16 -- 1 • • o-ring
21 -- 1 • • o-ring
Item no.
Part no. Quantity Description
(Figure 9-23)
33 -- 1 • • o-ring
36 -- 1 • • o-ring
37 -- 1 • Floor, pneumatics
38 4-071919-00 1 • Tube, transfer, air
39 4-071920-00 2 • o-ring
40 -- 1 • Seat, oxygen
41 4-071920-00 2 • o-ring
42 -- 1 • Seat, check valve, air (CV2)
43 4-070036-00 6 • Screw, PAN, 6-32 x 0.75, sems (Attach gas/check valve seats
to check valve manifold)
44 4-071915-00 1 • Plug, compressor port
Item no.
Part no. Quantity Description
(Figure 9-23)
45 4-071921-00 1 • o-ring
46 4-075567-00 1 • Washer, BDU interconnect
47 4-071932-00 1 • Spring, compression, 0.845 x 0.750
48 -- 1 • Seat, check valve, compressor (CV4)
49 -- 2 • Poppet, check valve, compressor (CV4)
50 -- 2 • o-ring
51 -- 2 • o-ring
2 (2 PL)
1 3 (2 PL) 4
6
5
Oxygen
8 (2 PL)
9 (2 PL) 10
11
51 7
40
(2 PL) 12 (2 PL)
Air 13
50
(2 PL) 42
49
(2 PL) 41
(2 PL) 17
48
47 18
43
46 (6 PL) 14
15
45 (2 PL) 20
39 16 19
22 21
(2 PL) 38 23
29 (2 PL)
44 28
24
30 27
33 (2 PL)
31
32
36
34 26
35
25
NOTE:
Lubricate o-rings with 37
Krytox™* grease
(P/N 4-732130-00).
Item no.
Part no. Quantity Description
(Figure 9-24)
-- -- 2 • Nut
Item no.
Part no. Quantity Description
(Figure 9-25)
Item no.
Part no. Quantity Description
(Figure 9-26)
3
(3 PL) 2
1 4
5
13 6 7
(4 PL)
12
(4 PL)
8
(10 PL)
9
(2 PL)
10 4.875 in. (12.4 cm)
11 (2 PL)
Item no.
Part no. Quantity Description
(Figure 9-27)
Item no.
Part no. Quantity Description
(Figure 9-28)
1 4-076299-00 1 • Cable, ground
2 4-00 1138-00 1 • Lockwasher, #8 (not shown)
3 4-005482-00 1 • Nut, captive, 8-32, zinc (not shown)
4 -- 1 • Assembly, compressor (Refer to appropriate 806 compressor
preventative maintenance kit.)
5 4-075214-00 4 • Lower stopper
6 4-003941-00 4 • Washer, flat
7 4-007654-00 4 • Lockwasher, SR, zinc PL 1/4"
8 4-074339-00 4 • Screw, hex head, 1/4 - 20 x 2
9 4-079058-SP 1 • Assembly, solenoid valve
4-076257-00 1 • • Outlet filter
10 4-076319-00 1 • Clip holder, air dryer assembly
11 4-072623-00 3 • Screw, FH, 8-32 x 3/8
12 4-076220-00 1 • Baffle, sound, base plate
13 4-076241-00 1 • Foam kit, sound, base
14 4-079064-SP 1 • Assembly, base (includes air dryer clamp assembly)
15 4-076289-00 1 • Air dryer filter and muffler assembly (Included in the
preventative maintenance kit)
16 4-079059-SP 1 • Assembly, air dryer
17 4-079067-SP 1 • Assembly, coalescing filter
18 4-076728-00 1 • Coalescing filter element (also included in PM kit)
19 4-076219-00 1 • Bracket, H2O water trap, 806
20 4-076291-00 1 • Elbow, 1/4 NPT x 3/8 TU
21 4-070041-00 18 • Screw, 8-32 x 0.38 (not all screws shown)
22 4-000697-00 2 • Screw, pan, 1/4-20 x 1/2, zinc
23 4-076210-00 1 • Plate, motor mount
24 -- 1 • Capacitor, motor (Refer to appropriate preventative
maintenance kit.)
25 -- 1 • Clamp, capacitor, motor, 1.75" diam (120V & 100V)
-- • Clamp, capacitor, motor, 1.379" (220V)
26 4-075215-00 4 • Upper stopper
27 4-076318-00 4 • Pop rivet (not shown)
28 4-075247-00 4 • Sleeve, shock mount
29 4-076290-00 4 • Vibration mount
30 4-076282-00 1 • Pressure relief valve, 36 PSIG (not shown)
Item no.
Part no. Quantity Description
(Figure 9-29)
Item no.
Part no. Quantity Description
(Figure 9-30)
Item no.
Part no. Quantity Description
(Figure 9-31)
-- 4-070520-SP -- 802 Backup power source (BPS) (See Section 9.4 for NHA.)
1 4-074678-00 1 • Bezel, BPS, front
2 4-072510-00 2 • Screw, PAN, 6-32 x 1/4, black oxide (Attach bezel)
3 4-074677-00 1 • Panel, BPS front
4 4-071599-00 4 • Screw, FH, 6-32 x 1/4 (Attach front panel)
5 4-074532-00 1 • Fuse, 15 A, fast-acting, cartridge-type
6 4-000003-00 1 • Tie wrap, small
7 10000007 1 • Battery pack
8 4-074665-00 1 • Floor assembly, BPS enclosure
9 4-079031-SP 1 • Chassis with cable, BPS
10 -- 1 • Standoff, battery
11 4-003443-00 1 • Coupling
12 4-070062-00 2 • Screw, FH, 6-32 x 1/2 (Attach floor to cover)
13 4-016142-00 4 • Nut, HEX, 6-32, keps (Attach PCB to chassis)
14 4-076727-SP 1 • PCB, backup power source (BPS)
15 N-7600028 1 • • Fuse, 15 A, fast-acting
18
2 (2 PL)
1
4 (4 PL)
17
3
16
(6 PL) 5
9 6
J2
J1
14,15
J3 7
13
(4 PL) to J2
12
(2 PL)
11
10
Item no.
Part no. Quantity Description
(Figure 9-31)
1 10030274 1 803 Backup Power Source (See Section 9.4 for NHA.)
2 10030284 1 Ext BPS Battery Pack Assembly
3 4-076727-SP 1 Assembly, BPS II Electronics
4 4-070508-00 1 Cable Assembly Backup Power Source
5 10025454 1 Screws (not shown)
6 10026127 Screws, back of mounting bracket (not shown)
7 10026128 Screws, Mounting Cart (not shown)
8 10029082 1 Protective sleeve
9 C-492044-00 2 Cable tie, small
10 4-078166-00 3 6-32 Nylock nut (not shown)
11 10025456 1 8-32 nut with star washer (not shown)
12 10026139 3 Insulating washer (not shown)
Item no.
Part no. Quantity Description
(Figure 9-33)
10 4-074616-00 4 • Bumper
11 4-074630-00 2 • Mount, flex arm (Original cart model only)
12 4-072623-00 2 • Screw, FH, 8-32 x 3/8 (Attaches flex arm mount to original
cart model only)
13 4-074680-00 1 • Bracket, BPS top slide, right
14 4-074681-00 1 • Bracket, BPS top slide, left
4-072623-00 4 • Screw, FH, 8-32 x 3/8 (Attach BPS slide brackets to cart) (not
--
shown)
15 4-074652-00 1 • Drawer
16 4-070520-SP 1 • Backup power supply (BPS)
4 (3 PL)
11 (2 PL)
3
12 (2 PL) 2
5
6 (3 PL)
15
14
16
10 (4 PL) 13
8 (2 PL)
7
9 (2 PL)
Item No.
Part No. Quantity Description
Figure 9-34
not shown 10045589 1 Kit, humidifier mounting, Puritan Bennett 800 Series Ventilator
Compressor Mount Cart
not shown 10043379 1 1-hour BPS bracket (required if converting from four-hour to
one-hour battery)
not shown 10045984 1 Cable, battery extension, Puritan Bennett 800 Series Ventilator
Compressor Mount Cart
See item 4 in 4-070036-00 3 Screw, PAN, 6-32 x 0.75, sems (Attach GUI base to cart)
Figure 9-33
Item No.
Part No. Quantity Description
Figure 9-34
not shown 10043481 3 Screw, 1/4-20 x 1/2 SFHCS (humidifier mount screws)
not shown 10053615 4 Screw, Slotted flat head (humidifier mount screws)
4
3
1,2
6 7
Figure 9-34. Puritan Bennett™ 800 Series Ventilator Compressor Mount Cart assembly
Item no.
Part no. Quantity Description
Figure 9-35
1 10049941 1 Base assembly, Puritan Bennett 800 Series Ventilator Pole Cart
2 10043373 1 Cover, BDU shelf, upper
3 10043374 1 Cover, BDU shelf, lower
4 10041445 1 Kit, grip handle, Puritan Bennett 800 Series Ventilator Pole Cart
5 10045975 4 Flex arm insert, with lockring
6 10046826 1 Cart with 1-hour battery, Puritan Bennett 800 Series Ventilator
Pole Cart
7 10046827 1 Cart with 4-hour battery, Puritan Bennett 800 Series Ventilator
Pole Cart
not shown 10045970 6 Middle plate Nylock nuts (fasten charging circuit to BDU shelf)
not shown 10045578 1 Kit, cylinder mounting, Puritan Bennett 800 Series Ventilator
Pole Cart
not shown 10045588 1 Kit, water trap mounting, Puritan Bennett 800 Series Ventilator
Pole Cart
not shown 10042364 1 Kit, humidifier mounting, Puritan Bennett 800 Series Ventilator
Pole Cart
not shown 10043378 1 Extension cable Puritan Bennett 800 Series Ventilator Pole Cart
not shown 10037900 1 Tank strap, Puritan Bennett 800 Series Ventilator Pole Cart
not shown 10045972 1 BPS, cover, blue, Puritan Bennett 800 Series Ventilator Pole Cart
not shown 10043469 6 Nut, 1/4-20 Nylon hex nut; lower shelf/cart frame to base
not shown 10047074 1 Battery bracket, one-hour required if converting from 4-hour
to one-hour battery
not shown 10043467 2 8-32 Nylock patch screw, power cord hook
Item no.
Part no. Quantity Description
Figure 9-35
not shown 10051308 3 Screw, 1/4-20 x 7/8 SFHCS (humidifier mount screws)
not shown 10053615 4 Screw, Slotted flat head (humidifier mount screws)
2
6, 7
Index 2
End inspiratory pressure (PI END) parameter, function (SOL2). See Solenoid, expiratory pressure
and range 2-113 transducer autozero
End-inspiratory flow method 2-111 Extended self test (EST) 3-19–3-28
Environmental requirements 1-4 hardware requirements 3-20
Error messages. See Diagnostic codes or Alarm how to display test outcome 4-7
messages how to run 3-20–3-22
EST service function 4-7 list of tests 3-25–3-27
EST. See Extended self test test results stored in NOVRAM 4-5
EST/SST Diagnostic Log troubleshooting. See Section 6
description 4-5, 6-1 when to run 3-19
how to display contents 4-5–4-6 External Test Control service function 4-9
screen 4-6
EV. See Exhalation valve
EXH HTR. See Heater, exhalation F
Exhalation check valve (CV5). See Check valve, F1. See Filter, oxygen impact
exhalation F2. See Filter, air inlet
Exhalation collector vial (ECV). See Collector vial, F3. See Filter, oxygen inlet
exhalation F4. See Filter, air impact screen
Exhalation flow sensor (Q3). See Sensor, flow, F5. See Filter, oxygen impact screen
exhalation F6. See Filter, air pneumatic noise
Exhalation heater (EXH HTR). See Heater, exhalation F7. See Filter, oxygen pneumatic noise
Exhalation metabolic sample port F8. See Filter, inspiratory (main flow)
eliminating 8-94 F9. See Filter, expiratory
Exhalation module Failure
component photographs 2-40–2-41 in EST 3-27
cover, removing and installing 8-90–8-92 in SST 3-18
electrical interconnect diagram 2-98 Fan (compressor), description 2-101
part replacement Fans, description 2-43
check valve (CV5) 8-101–8-102 Filter
collector vial (ECV) 8-90 air impact screen (F4), description 2-19
exhalation sample port 8-94 air inlet (F2)
exhalation transducer PCB 8-99–8-100 description 2-19
exhalation valve (EV) 8-95–8-96 replacement 8-69, 8-70
expiratory filter (F9) 8-90 air pneumatic noise (F6), description 2-19
expiratory pressure transducer autozero compressor (F10), description 2-43
solenoid (SOL2) 8-101 compressor intake/silencer (F12), description 2-
flow sensor (Q3) 8-97–8-99 43
heater assembly (EXH HTR) 8-101–8-102 compressor transducer (F11), description 2-43
pneumatic diagram 2-38 expiratory (F9)
removing and installing 8-93 description 2-36
theory of operation 2-37–2-41 replacement 8-90
Exhalation transducer PCB. See PCB, exhalation GUI, 9.4-inch
transducer cooling vent, replacement 8-51
Exhalation valve (EV) inspiratory (main flow) (F8), description 2-36
description 2-39 line, description 2-61
how to calibrate 4-9–4-12 oxygen impact (F1)
replacement 8-95, 8-96 description 2-18
Exhalation, gas flow diagram 2-112 replacement 8-72
Exhalation, patient 2-110 oxygen inlet (F3)
Exhaled minute volume (VE TOT) parameter, function description 2-18
and range 2-113 replacement 8-72
Exhaled tidal volume (VTE) parameter, function and oxygen pneumatic noise (F7), description 2-18
range 2-114 Flex arm assembly, illustrated parts breakdown 9-8–
EXP PAUSE key, description 1-41 9-9
Exp Valve Calibration service function 4-9–4-12 Flow control subsystem
Expiratory filter (F9). See Filter, expiratory 2-14 component photographs 2-24
Expiratory pressure transducer autozero solenoid pneumatic diagram 2-23
fascia panel, removing and installing 8-70–8-71 Knob encoder circuit, on GUI CPU PCB, description 2-
gas flow diagram 2-17 85
illustrated parts breakdown 9-58–9-61
leak testing 8-74–8-75
modifications, recent 8-68 L
part replacement Label kits, parts list 9-19–9-28
air inlet filter (F2) 8-69–8-70 Leak testing, pneumatic connections 8-4
check valve (CV3) 8-78 Leakage current, specifications 1-6
check valve assembly 8-86 LED array
ferrite cores, installing 8-68 diagnostic (on BD CPU PCB), description 2-75
flow sensor assembly 8-86 LED, power. See Power indicator
inspiratory electronics PCB 8-80 Light. See name of specific light
left-side plate 8-80–8-81 Line filter, description 2-61
o-rings 8-69 Line voltage, how to set low alarm threshold 4-8
oxygen and air flow sensors (Q1 and Q2) 8- Lock key (for screen), description 1-39
83 Log. See EST/SST Diagnostic Log, System Diagnostic
oxygen and air pressure switches (PS1 and Log, System Information Log, or Alarm log
PS2) 8-72 Low-urgency alarm indicator, description 1-45
oxygen and air regulators (REG1 and REG2)
8-88
oxygen filter assembly 8-72 M
oxygen sensor 8-75–8-78 Main flow filter. See Filter, inspiratory
oxygen sensor (with access port) 8-75–8-77 Maintenance and service. See Schedule of periodic
pressure transducer autozero solenoid maintenance or Service and repair
(SOL1) 8-89 Manifold, outlet, description 2-27
PSOL cartridge 8-73–8-74 MANUAL INSP key, description 1-41
PSOL manifold 8-81–8-83 Manual ventilator check. See Performance
right-side plate 8-78–8-80 verification, tests
safety valve (SV) 8-84–8-85 Manufacturer’s Declaration 1-10
removing and installing (inspiratory module) 8- Maximum circuit pressure (PPEAK) alarm, function
74–8-75 and range 2-114
Inspiratory module floor assembly Mean airway pressure (PMEAN) parameter, function
disassembly and reassembly 8-89 and range 2-114
illustrated parts breakdown 9-64–9-67 Medium-urgency alarm indicator, description 1-45
Inspiratory pressure transducer autozero solenoid Mode, service. See Service mode
(SOL1). See Solenoid, inspiratory pressure Module, exhalation. See Exhalation module
transducer autozero Module, inspiratory. See Inspiratory module
Inspiratory pressure transducer. See Transducer, Monitored data
inspiratory pressure functions and ranges 2-113–2-114
Interconnect diagram. See Electrical interconnect range, resolution, and accuracy 1-27–1-29
diagram Motherboard PCB. See PCB, motherboard
K N
K1. See Relay, power 2-61 NeoMode
Key. See name of specific key patient system 2-6
Keyboard assembly Normal GUI indicator, description 1-48
description 1-39–1-42, 2-89–2-90 Normal ventilator operation indicator, description 1-
interface circuit, on GUI CPU PCB 2-85 45
knob assembly 2-89 NOVRAM (nonvolatile random access memory)
photograph 2-90 circuit (on BD CPU PCB), description 2-74
Keyboard assembly, 10.4- inch GUI circuit (on GUI CPU PCB), description 2-83
replacement 8-26 contents 2-77
Keyboard assembly, 9.4- inch GUI Nurse’s call (remote alarm) port
replacement 8-32 location 1-46
Kits, preventive maintenance, 10-000-hour 1-37 specifications and pinout 1-7–1-8
Kits, preventive maintenance, 15,000-hour 1-37
O replacement 8-60
backlight inverter
O2. See Oxygen
description 2-92–2-93
Occlusion handling, operational description 2-121–
photograph 2-82
2-123 BD (breath delivery) CPU
OIM (operator-initiated manadatory) breath 2-110
description 2-74–2-76
OPERATION TIME service function 4-7
photograph 2-75
OS. See Sensor, oxygen
replacement 8-60
OSC (occlusion status cycling) 2-121
BDU LED
Other Screens service function, list of subfunctions 4-
description 2-95
8–4-15 photograph 2-95
Oxygen calibration key, description 1-41
BPS (backup power source)
Oxygen check valve (CV1). See Check valve, oxygen
photograph 2-68
Oxygen flow sensor (Q1). See Sensor, flow, oxygen 2-
replacement 8-137
13 compressor
Oxygen hose assemblies, parts list ??–9-11
block diagram 2-104
Oxygen impact filter (F1). See Filter, oxygen impact
description 2-101–2-107
Oxygen impact screen filter (F5). See Filter, oxygen
photograph 2-102
impact screen
exhalation transducer
Oxygen inlet filter (F3). See Filter, oxygen inlet
description 2-97
Oxygen pneumatic noise filter (F7). See Filter, oxygen
replacement 8-99, 8-100
pneumatic noise
GUI (graphic user interface), 10.4-inch
Oxygen pressure switch (PS1). See Pressure switch,
removing CPU 8-16
oxygen
GUI CPU
Oxygen pressure valve. See Pressure valve, oxygen 2-
description 2-80–2-85
14 photograph 2-81, 2-82
Oxygen PSOL (PSOL1). See Proportional solenoid
Trending-enabled (TE) 2-80, 8-7
valve, oxygen
GUI LED
Oxygen regulator assembly (REG1). See Regulator,
description 2-91
oxygen
photograph 2-91
Oxygen sensor calibration 4-20
GUI touchframe
Oxygen sensor. See Sensor, oxygen
description 2-88
GUI, 10.4-inch
P backlight inverter
PA. See Transducer, pressure, absolute (atmospheric) replacement 8-15
Part ordering 8-5 CPU PCB
part replacement touchframe style jumpers 8-17
GUI (graphic user interface), 10.4-inch LED
CPU 8-16 replacement 8-24
Parts list 9-1–9-87 touchframe (MKG Touch)
abbreviations 9-2 replacement 8-23
how to use 9-1 GUI, 9.4-inch
See also name of specific assembly backlight inverter
Patient circuit, specifications 1-15–1-17 description 8-39
Patient system
replacement 8-39
description 2-6
CPU
illustrated parts breakdown 9-3–9-5
pneumatic diagram 2-36 replacement 8-43
See also 840 Ventilator System Operator’s and touchframe style jumpers 8-43
Technical Reference Manual LED
theory of operation 2-37 description 8-39
PC. See Transducer, pressure, compressor replacement 8-40
PCB touchframe
analog interface (AI) replacement 8-47
description 2-78 VGA LCD controller
photograph 2-78
replacement 8-42
air (REG2), description 2-19 (SST), Extended Self Test (EST), or Background
oxygen (REG1), description 2-18 checks
setting verification. See Performance Sensor
verification, tests flow
Regulatory approvals 1-8–1-14 air (Q2)
Relay description 2-23
power (K1), description 2-61, 2-101 replacement 8-83
remote alarm (nurse’s call), on GUI CPU PCB, exhalation (Q3)
description 2-85
description 2-38
Release handle (BDU), replacement 8-112
Relief valve (RV), description 2-43 replacement 8-97, 8-99
Remote alarm (nurse’s call) port how calibration data is copied from its serial
location 1-46 EEPROM to BDU flash memory 4-17–
specifications and pinout 1-7–1-8 4-18
Remote alarm relay, on GUI CPU PCB, description 2- how to calibrate 4-13–4-15
85 oxygen (Q1)
RESET (alarm) key, description 1-40 description 2-23
Restrictor (R1), description 2-43 replacement 8-83
Rotary encoder/decoder circuit, on GUI CPU PCB, oxygen (OS)
description 2-85 access port 8-75
Rotor housing (GUI, 9.4-inch) description 2-27
replacement 8-52 life expectancy 1-5
Rotor housing, GUI, 10.4-inch replacement 8-75, 8-78
replacement 8-28 replacement (with access port) 8-77
RS-232 port See also Transducer
location 1-46 Serial communications port
on BD CPU PCB, description 2-75 location 1-46
on GUI CPU PCB specifications and pinout 1-7–1-8
description 2-83 Serial loopback test (10.4" GUI only) 5-10
how to set baud rate 4-8 Serial numbers
specifications and pinout 1-7–1-8 how they are copied from data key to flash
memory 4-18–4-19
how to display 1-59, 4-7
S stored on data key 2-80
SAAS. See Sound generator interface Service (TEST) button, description 1-50
Safety valve (SV) Service and repair 8-1–8-155
description 2-27 Service and repair
replacement 8-84, 8-85 adhesives (use of) 8-3
SVO gas flow diagram 2-30 BDU (breath delivery unit). See BDU, part
Safety valve and inspiration monitoring subsystem replacement
component photographs 2-26, 2-28–2-29 BPS (backup power source). See BPS (backup
pneumatic diagram 2-26 power source), part replacement
theory of operation 2-25–2-30 Cart assembly. See Cart assembly
Safety valve open (SVO) indicator, description 1-43 electrical cables and pneumatic tubing 8-3
Safety valve open (SVO) state electrostatic discharge control 8-4
gas flow diagram 2-120 exhalation module. See Exhalation module, part
operational description 2-119–2-120 replacement
Schedule of periodic maintenance 1-35, 1-36 GUI (graphic user interface), 10.4-inch. See GUI
Screen (graphic user interface), 10.4-inch, part
EST/SST Diagnostic Log 4-6 replacement
SERIVCE MODE 4-2 Inspiratory module. See inspiratory module, part
System Information Log 4-6 replacement
Screen lock key, description 1-39 leak testing, pneumatic connections 8-4
Self tests part ordering 8-5
list 3-2–3-3 patient system and accessories. See 840
Self tests 3-1–3-29 Ventilator System Operator’s and Technical
See also Power on self test (POST), Short self test Reference Manual
U
Utility panel. See AC panel
V
Valve. See specific name of valve, such as Exhalation
valve
VENT CONFIG service function 4-7
Vent head LED PCB. See PCB, BDU LED
VENT INOP indicator, description 1-43
Vent Inop Test service function 4-12–4-13
Ventilator inoperative condition 2-10
Ventilator inoperative indicator, description 1-43
Ventilator settings, ranges, resolutions, and
accuracies 1-18–1-25
Ventilator System. See 840 Ventilator System
VGA controller. See PCB, VGA LCD controller
VIM breath 2-110
Voltage
line, how to set low alarm threshold 4-8
sentry circuit (on BD CPU PCB), description 2-79
supply, monitoring, description 2-119
Volume key (for alarm), description 1-40
W
Water trap (WT2), description 2-43
Water trap, air (WT1), description 2-19
Water traps (in patient circuit), description 2-36
Weight, ventilator 1-4
Wiring diagram. See Electrical interconnect diagram
2-63
WT1. See Water trap, air
©2013 Covidien.
Covidien llc,
15 Hampshire Street, Mansfield, MA 02048 USA.
Covidien Ireland Limited,
IDA Business & Technology Park,Tullamore.