Dynamic Compression PLates - Synthes
Dynamic Compression PLates - Synthes
Dynamic
Compression Plates (DCP) and
Dynamic Compression Plates with
Limited Bone Contact (LC-DCP).
Lesson is only until page 22. The rest is the catalogue only.
Surgical Technique
This description alone does not provide sufficient background for direct use
of the instrument set. Instruction by a surgeon experienced in handling these
instruments is highly recommended.
AO Principles 4
Intended Use 5
Buttress Plate 18
MRI Information 48
Non-symmetrical holes
Some Dynamic Compression Plates feature non-symmetrical
holes which allow to achieve compression in only one
direction. These plates feature a mid section which is placed
at the position of fragments to be compressed.
mid section
Symmetrical holes
Most 1.3 to 3.5 mm Dynamic Compression Plates have sym-
metrical holes similar to the LC-DCP hole. The symmetrical
shape of the plate holes enables compression to be achieved
in both directions, thus plate positioning is not restricted by
the presence of a mid-section.
Non-symmetrical holes
The DCP allows up to 25° longitudinal and 7° transverse
screw angulation.
Symmetrical holes
The DCP hole allows up to 50° longitudinal screw angulation
for an increased range of applications. LC-DCP holes allow
even wider angulations up to 80°. Both plate types allow for
14° transverse screw angulation.
Undercuts
AO PRINCIPLES
Anatomic
Anatomic reduction fixation
Stable fixation
Fracture reduction
Fracture reduction and fixation
and fixation to Fracture fixation
Fracture fixationproviding
providingabsolute
abso- or
to restore
restore anatomical
anatomical relationships.
relationships. relative stability,stability,
lute or relative as required
as by the
prequired
atient, theby injury, and thethe
the patient, personality
injury,
1 2 of
andthethe
fracture.
personality of the
fracture.
Early,
Early, active mobilization
mobilization Preservationof
Preservation of blood
blood supply
supply
Early and
Early and safe mobilization
mobilizationand
and 4 3 Preservation
Preservationof ofthe
theblood
bloodsupply
supply
rehabilitation the injured partand
rehabilitation of the injured part to
to soft tissues and bone bygentle
soft tissues and bone by
the patient
and as a whole.
the patient as a whole. reduction techniques
gentle reduction and careful
techniques and
handling.
careful handling.
1
Müller ME, M Allgöwer, R Schneider, H Willenegger. Manual of Internal
Fixation. 3rd ed. Berlin Heidelberg New York: Springer. 1991.
2
Rüedi TP, RE Buckley, CG Moran. AO Principles of Fracture Management.
2nd ed. Stuttgart, New York: Thieme. 2007.
1
Müller ME, Allgöwer M, Schneider R, Willenegger H. Manual of Internal Fixation.
3rd ed. Berlin, Heidelberg, New York: Springer. 1991.
2
Rüedi TP, Buckley RE, Moran CG. AO Principles of Fracture Management. 2nd ed.
Stuttgart, New York: Thieme. 2007.
Intended Use
DCP and LC-DCP System Implants are intended for tempo-
rary fixation, correction or stabilization of bones in various
anatomical regions.
Contraindications
No specific Contraindications.
1
Plate contouring
Precautions:
–– Avoid repeated bending. learn the mechanics !
–– Sharp indentations on the plate surface, especially around
the plate holes, may impair resistance to fatigue and
should be avoided.
–– Reverse bending or use of the incorrect instrumentation
for bending may weaken the plate and lead to premature
plate failure (e.g. breakage). Do not bend the plate be-
yond what is required to match the anatomy.
–– Instruments and screws may have sharp edges or moving
joints that may pinch or tear user’s glove or skin.
–– Handle devices with care and dispose worn bone cutting
instruments in an approved sharps container.
2
Use of lag screws to achieve additional stability
interfragmentäre Plattenzugschraube. Helv. Chir. Acta 52, 19–23 Reduction in implant stress by using a lag screw through the plate1.
1
Standard drill guides (DCP, sizes 1.3 – 4.5)
aD
rilling in neutral position
Position the tube in the middle of the DCP hole.
a b
Neutral and eccentric drilling
2
Universal drill guides (DCP and LC-DCP, sizes 1.5 – 4.5)
a Neutral position
Depress the drill guide. The upper part of the inner sleeve
projects. Position the drill guide in the plate hole which
is remote from the fracture (LC-DCP). When used with DC
plates the drill guide automatically slides into the correct
middle position when depressed.
buttress
position
BUTTRESS = Penopang
The LC-DCP drill guides can be used with LC-DC plates only.
They feature dedicated ends for neutral application (green)
and for application in compression mode (yellow).
a Neutral position
Use the green drill guide to place a screw in neutral
position. The arrow on the green drill guide must point
towards the fracture. Compression cannot be achieved
with the green drill guide.
= symbol for a fracture line
b Compression
Use the yellow drill guide to achieve compression. The
arrow on the yellow drill guide must point towards the arrow points towards the fracture
c Buttress position
Use the green drill guide. A buttress effect is achieved green yellow
because for inclined position (through the fracture), screw should be placed in the middle
position of the plate hole.
c
green
Implant Removal
In case the physician decides to remove the implants, they
can be removed by using general surgical instruments.
buttress position:
arrow points away
from the fracture
WHY ????
Note: Required drill bits for threaded and gliding holes and
taps for the individual screw sizes can be found in the
chapter “Screws, Drills and Taps” (refer to page 20 and 21).
1
Drill first hole adjacent to fracture
Fragment 1 Fragment 2
Use an appropriately sized drill bit for threaded holes and
drill the first hole adjacent to the fracture. Use the green drill
guide in neutral position with the arrow pointing towards why?? learn the biomechanics!
the fracture. Alternatively, use the universal drill guide in
depressed mode (upper part projects) and place it in the hole
remote from the fracture.
green
Determine screw length, tap the thread, and insert the
screw.
yellow
Drill through both cortices into the opposite fragment with
the yellow drill guide in compression position (arrow points
towards the fracture). Alternatively, use the universal drill
guide placing it in the hole remote from the fracture. Do not
depress it (upper part stands flush). Why ??
3
Insert interfragmentary lag screw
Widen the hole in the near cortex with a drill bit for gliding
holes. Push the tissue protecting sleeve upwards, insert the
drill bit into the hole, then slide the tissue protecting sleeve
down until it touches the plate. When drilling, care should
be taken that the course of the gliding hole follows exactly
that of the previous drill hole.
Implant Removal
In case the physician decides to remove the implants, they
can be removed by using general surgical instruments.
Note: Required drill bits for threaded and gliding holes and
taps for the individual screw sizes can be found in the
chapter “Screws, Drills and Taps” (refer to page 20 and 21).
1
Treat first fracture
2
Treat second fracture
Treat the second fracture using the universal drill guide. Drill
a hole into the third fragment using the universal drill guide
in compression position and a drill bit for threaded holes.
Seat the drill guide at the side of the hole remote from the
fracture without depressing the drill guide (upper part is
flush).
Insert a screw into the hole. The use of a shaft screw is rec-
ommended.
Prepare seating for the lag screw with the universal drill
guide and a drill bit for threaded screws. Position the univer-
sal drill guide adjacent to the fracture, depress the drill guide
(inner sleeve projects), set drilling angle and drill. If there is
the risk of the drill bit colliding with the screw in the vicinity,
the drill bit has to be inclined.
Drill the gliding hole. Widen the near cortex using an appro-
priately sized drill bit for gliding holes and the tissue protect-
ing sleeve. Measure the screw length and tap the thread.
Implant Removal
In case the physician decides to remove the implants, they
can be removed by using general surgical instruments.
Note: Required drill bits for threaded and gliding holes and
taps for the individual screw sizes can be found in the
chapter “Screws, Drills and Taps” (refer to page 20 and 21).
All the screw holes in the buttress region are prepared with
the universal drill guide or the green drill guide (neutral posi-
tion) of the LC-DCP drill guide.
The arrow of the drill guide must point away from the frac-
ture. The proximally positioned cancellous bone screws and
cortex screws act as lag screws. The thread grips only in the
opposite fragment, so that there is compression between the
fragments. The use of the green drill guide with the arrow
pointing away from the fracture prevents the tibial plateau
fragment and the plate from slipping. In this situation, the
screw heads are buttressed by the hole edge adjacent to the
fracture.
Implant Removal
In case the physician decides to remove the implants, they
can be removed by using general surgical instruments.
When using the green drill guide of the LC-DCP drill guide
the arrow must point towards the fracture.
green
When using the universal drill guide seat it at the hole edge
remote from the fracture, depress the outer sleeve so that
the upper part projects for neutral position.
Implant Removal
In case the physician decides to remove the implants, they
can be removed by using general surgical instruments.
Screw recess Cruciform Cruciform Cruci- Hex Cruci- Star- Hex Cruci- Stardrive Hex Stardrive
form form drive form
1.5/
1.0 mm 1.3 mm 1.5/2.0 mm 1.5 mm 2.0 mm T6 1.5 mm 2.4 mm T8 2.5 mm T8
self-tapping ● ● ● ● ● ● ● ● ● ●
not self-tapping ● ● ● ●
Gliding hole
Drill Bit for gliding hole � 1.0 mm � 1.3 mm � 1.5 mm � 2.0 mm � 2.4 mm � 2.7 mm
Drill Bit for threaded hole � 0.8 mm � 1.0 mm � 1.1 mm � 1.5 mm � 1.8 mm � 2.0 mm
The screws are available in steel and titanium. Sterilizable paper – max. 134 °C
0X6.000.343_AB.qxp:0X6.000.343 12.12.2008 9:23 Uhr Seite 1
Cortical bone
– self-tapping ● ● ●
– non-self-tapping ● ● ●
– Shaft screw ● ●
Cancellous bone screws
– Full thread ● ●
– Long thread ●
– Short thread ● ●
Gliding hole
Threaded hole
Drill Bit for threaded hole � 2.0 mm � 2.5 mm � 2.5 mm � 3.2 mm � 3.2 mm
The screws are available in steel and titanium. Sterilizable paper – max. 134 °C
Stainless Titanium
Steel (TiCP)
Stainless Titanium
Steel (TiCP)
Stainless Titanium
Steel (TiCP)
LC-DCP® 2.7
Washer B 7.0/3.6 mm
Cerclage Wire
Kirschner Wire
With trocar tip, pack of 10 units
Stainless Steel B (mm) Length (mm)
292.060.10 0.6 70
292.080.10 0.8 70
292.100.10 1.0 150
292.120.10 1.25 150
292.160.10 1.6 150
Instruments
310.882 Countersink 1.3 to 2.0, length 50 mm,
for Mini Quick Coupling
312.133 Double Drill Guide 1.3/1.0
314.413 Screwdriver Shaft 1.3, cruciform,
with Holding Sleeve, length 62 mm,
for Mini Quick Coupling
314.481 Screwdriver Shaft 1.0, cruciform,
with Holding Sleeve, length 62 mm,
with Mini Quick Coupling
316.385 Drill Bit B 0.8 mm, length 40/16 mm,
2-flute, for Mini Quick Coupling
513.005 Drill Bit B 1.0 mm, length 46/34 mm,
2-flute, for Mini Quick Coupling
513.035 Drill Bit B 1.3 mm, length 46/34 mm,
2-flute, for Mini Quick Coupling
Plates
Stainless Titanium
Steel (TiCP)
Instruments
310.880 Countersink 1.5 to 2.4, length 52 mm,
for Mini Quick Coupling
312.140 Double Drill Guide 1.5/1.1, for No. 311.150
314.667 Screwdriver Shaft 1.5, cruciform, with
Holding Sleeve, length 66 mm, with Mini
Quick Coupling
513.030 Drill Bit B 1.1 mm, length 45/33 mm,
2-flute, for Mini Quick Coupling
513.080 Drill Bit B 1.5 mm, length 52/40 mm,
2-flute, for Mini Quick Coupling
Stainless Titanium
Steel (TiCP)
Screws
Instruments
310.880 Countersink 1.5 to 2.4, length 52 mm,
for Mini Quick Coupling
312.220 Double Drill Guide 2.0/1.5
314.672 Screwdriver Shaft 2.0, cruciform,
with Holding Sleeve, length 66 mm,
with Mini Quick Coupling
323.200 Universal Drill Guide 2.0
513.090 Drill Bit B 1.5 mm, length 67/55 mm,
2-flute, for Mini Quick Coupling
513.140 Drill Bit B 2.0 mm, length 67/55 mm,
2-flute, for Mini Quick Coupling
Plates
Stainless Titanium
Steel (TiCP)
Instruments
310.880 Countersink 1.5 to 2.4, length 52 mm,
for Mini Quick Coupling
312.920 Drill Sleeve 1.8, with centering thread
313.940 Screwdriver, cruciform, with Holding
Sleeve, for Cortex Screws B 2.4 mm
314.672 Screwdriver Shaft 2.0, cruciform,
with Holding Sleeve, length 66 mm,
with Mini Quick Coupling
317.861 Drill Bit B 1.8 mm, length 80/66 mm,
2-flute, for Mini Quick Coupling
317.871 Drill Bit B 2.4 mm, length 80/66 mm,
2-flute, for Mini Quick Coupling
323.202 Universal Drill Guide 2.4
Stainless Titanium
Steel (TiCP)
44 DePuy Synthes DCP and LC-DCP Systems Surgical Technique Surgical Technique DCP and LC-DCP Systems DePuy Synthes 44
Buttress Pin B 1.8 mm
Also available
446.032 Plate 1.5, straight, 12 holes, length 59 mm,
Pure Titanium
446.031 Plate 1.5, straight, 6 holes, length 29 mm,
Pure Titanium
T-Plate 3.5
DCP® 4.5
T-Plate 4.5
Synthes GmbH
Eimattstrasse 3 Not all products are currently available in all markets.
4436 Oberdorf
Switzerland This publication is not intended for distribution in the USA.
Tel: +41 61 965 61 11
Fax: +41 61 965 66 00 All surgical techniques are available as PDF files at
www.depuysynthes.com www.depuysynthes.com/ifu 0123