Calibration
Calibration
Table of Contents
1. INTRODUCTION.................................................................................................................. 3
1. Introduction
The DIS-200 system has been designed to calibrate either automatically or manually. Each
step is discussed in this section. The only difference between Auto Cal and Manual is that
when the system is in Auto Cal, it runs through the functions step by step. When calibrating
manually, you can select a specific function rather than going through them all. For the purpose
of this manual, the actual functionality of these screens will be discussed and can be applied to
either operation.
NOTE: If any failures are detected, an error message window will be displayed. The Power
Down procedure should be followed in order to shut down the software. The cable
connections should then be rechecked -- especially the hard drive interconnecting
cables inside the PC cabinet. After these steps are taken, repeat the power up
sequence. If system continues to report errors do not continue with this procedure,
instead contact Shimadzu immediately.
To open the Service Application from the clinical application using the keyboard:
Once the Calibration menu opens, select either or . When System Manual
is selected, the buttons in the column under Measuring Fields are activated allowing selection of
function.
NOTE: The safety lock on the C-Arm Control Panel (the lock symbol) must be on when
running Auto-Calibration.
2. System Calibration
• The auto calibration will take you through each step of a full camera calibration sequence
in the correct order. Online instructions explaining how to perform each of the steps are
displayed in the instruction window during each step.
• Use up and down arrow keys and slider bar to scan complete set of instructions within
this window. Within each step are written instructions describing the procedure. System
defaults to Measuring Fields: Image Screen.
• Click on the button at the bottom on the screen to run the step.
• After the step has been completed successfully, a screen will pop up indicating that the
step was successful. Click OK to proceed.
Image
This function sets the window used for Auto Black sample window in the video frame to
the appropriate size (Auto Black – 95%).
1. Click RUN. Fluoro/Pulsed Fluoro runs first. The setting should read 95%. When
Fluoro/Pulsed Fluoro is finished, a window will be displayed with calibration
status for the step.
2. Select OK and then click Next to proceed to the next step of the procedure.
3. Click Run. The process will repeat for Spot. Again, the level should be at 95%.
4. A screen will pop up indicating that the step was successful. Click OK to
proceed.
5. Click OK and then click Next to Proceed to the next step (AGC/ABS).
Level indicated
here – should be
95%
AGC / IBS
This function sets the sample window in the video frame to the appropriate size for
Fluoro AGC/IBS and for Spot mode Iris calibration. The process cycles through all five
active camera mode, starting with Fluoro/Pulsed Fluoro (IBS-Off). The resulting size is
indicated under the slider control (only active during manual calibration). As each level
is calibrated successfully, a screen will pop up indicating that the step was successful.
Click OK to proceed and then Run to move to the next mode.
Indicates which
Camera Mode is
being calibrated
Control Level
Acquire LUTs
For each type of image supported on the system, the default Lookup table is selected
and loaded.
When the window opens, select RUN. The LUT for Fluoro/Pulsed Fluoro will be
selected and loaded. Once the procedure is complete, a status window will be
displayed. A screen will pop up indicating that the step was successful. Select OK and
then click NEXT to proceed to Spot. Repeat the procedure for Spot. When the process
is complete and the LUTs are set according to the following chart, select NEXT to
proceed to the next window in the video level sequence (Video Gain).
Fluoro Gamma 2
Spot Linear WC
Document #726-602-G1 4 - 10
Section 4 - Calibration
(Video) Gain
For each type of image supported on the system, the default Video Gain Value is
selected and loaded. This runs once each for Fluoro, Pulsed Fluoro and Spot.
When the window opens, select RUN. The default Video Gain Value is selected and
loaded. Once the procedure is complete, a status window will be displayed. Select OK
and then click NEXT to proceed to Pulsed Fluoro. Repeat the procedure for Pulsed
fluoro and then for spot. When the process is complete and the video gain values are
set according to the following chart, select NEXT to proceed to the next window in the
video level sequence (Auto Black).
Value
indicated
here.
Manual Only
- +/-
Brightness of
AGC
Average
Document #726-602-G1 4 - 11
Section 4 - Calibration
Auto Black
For each type of image supported on the system, the default Video Auto-Black goal and
the pedestal are calibrated. This process must be done in the dark. This runs once each
for Fluoro/Pulsed Fluoro and Spot. As each mode is calibrated, a window will be
displayed with calibration status for the step. Select YES and then click Next to proceed
to the next step of the procedure. The goal is for the Measuring Field Level to be zero
(0).
When the window opens, select RUN. The default Fluoro/Pulsed Fluoro Auto Black
Level is set and loaded. Once the procedure is complete, a status window will be
displayed. Select OK and then click NEXT to proceed to Spot. Repeat the procedure for
spot. When the process is complete and the video gain values are set according to the
following chart, select NEXT to proceed to the next window in the Auto Calibration
Sequence (Overlay/Circle Mask).
Fluoro Auto-Black ON
Spot Fixed Black
Document #726-602-G1 4 - 12
Section 4 - Calibration
2.3 Overlay
Circle Mask
This function sets the black software mask so that it surrounds the phosphor output
window image displayed on the monitor. The circle mask has a range in size from 900
to 1070 pixels in diameter.
NOTE: When running this step, the collimators and octagonal shutters must be
fully open. IBS must also be ON (or manually adjust fluoro kV/mA for
viewable image).
Follow the instructions in the dialog and then click Run. When the step is finished, a
window will be displayed with calibration status for the step. Select OK and then click
Next to proceed to the next step of the procedure. This setting will always be 1070.
Document #726-602-G1 4 - 13
Section 4 - Calibration
The size of the image displayed depends on the relationship in sizes of the Output phosphor
window, collimating lens focal length, camera lens focal length and CCD sensor size. An auto
cal setting of minimum or maximum likely indicates too small or too large of an image projected
on the CCD sensor.
Monitor
Phosphor
Document #726-602-G1 4 - 14
Section 4 - Calibration
Circle
1.1 Outline
Monitor
1.2
1.3
Phosphor
1.4
1.5
1.6
1.7
1.8
CCD Sensor
1.9
1.10
Document #726-602-G1 4 - 15
Section 4 - Calibration
Circle
Monitor Outline
Phosphor
CCD Sensor
Document #726-602-G1 4 - 16
Section 4 - Calibration
Fluoro
The first window will ask that you measure and adjust the x-ray dose for the normal dose
setting for continuous fluoro. Follow the instruction and then click OK.
200
Document #726-602-G1 4 - 17
Section 4 - Calibration
The next screen will ask that you activate radiation using continuous fluoro.
Click Run to complete the procedure. After you calibrate one rate a dialog box will be
displayed asking “Same for all Rates?” Select Yes.
Document #726-602-G1 4 - 18
Section 4 - Calibration
Pulsed Fluoro
Follow on-screen instructions. After you calibrate one rate a dialog box will be displayed
asking “Same for all Rates?” Select Yes.
Document #726-602-G1 4 - 19
Section 4 - Calibration
2.5 ABS
Instructions for performing the functions are shown in the box under the tabs. Use up
and down arrows to scroll down for all numbered steps. ABS only works in continuous
fluoro/normal dose (must be set on the generator).
ABS Level
This step will calibrate the correct level of Automatic Brightness Stabilization. Click OK to
enable ABS control on the Generator and then press Run. When the step is finished, a
window will be displayed with the calibration status for the step. Select OK and then
click Next to proceed to the next step of the procedure.
Document #726-602-G1 4 - 20
Section 4 - Calibration
ABS Response
1. Ensure that all optical covers and access ports to the camera and Image
Intensifier are in place and that no stray room light is visible to the camera.
2. Do NOT activate radiation during this calibration step.
3. Press Run to calibrate ABS Gain Response for the selected Camera Mode.
4. Repeat for each available camera mode.
5. When this step is finished, a window will be displayed with the calibration status
for the step. Select OK.
6. Press Next and the system will advance to the Iris/Spot Calibration
Document #726-602-G1 4 - 21
Section 4 - Calibration
1000
Document #726-602-G1 4 - 22
Section 4 - Calibration
The collimators can only be calibrated from the “System Manual” mode. Click on the
Generator button and the “Collimators” screen will be enabled.
Note the combinations of procedures available. Calibrate for both Normal and MAG1
FOVs and select the appropriate opening or rotation from the list to the right. Use the
sliders to make adjustments.
Document #726-602-G1 4 - 23
Section 4 - Calibration
To begin, drag the dialog box over to the reference monitor, and then select either
Normal or MAG1. This will send a message to the C-Arm as to what is being calibrated.
Document #726-602-G1 4 - 24
Section 4 - Calibration
4. Monitor Alignment
NOTES: The Image Monitors require five minutes to warm up after being powered on and
after synchronizing with the DIS-200. Do not make any Image Monitor
measurements during the first five minutes.
Document #726-602-G1 4 - 25
Section 4 - Calibration
1. From the Service Application Menu, select the Test Pattern button.
Document #726-602-G1 4 - 26
Section 4 - Calibration
3. Make sure the Window/Level settings are set to 512/512 (lower right corner).
4. Use the Brightness and Contrast controls to set them both to maximum values,
then slowly lower the brightness until the black of the circle mask (area
surrounding the image portion of the screen) is just slightly brighter than the very
edge of the CRT screen.
• The black bar on the bottom of the image (the one with the small white
bar in it) has the same visual darkness as the black mask surrounding the
image.
• The 95% inner black box is visible in the 100% black box (located lower
at the lower left center of the image).
• The 95% inner white box is visible in the 100% white box (located at the
lower right center of the image).
6. Select the INVERT button from the feature menu portion of the main Image
Monitor screen. When the image is inverted, the white bar should be on the right
side of the image and the black bar on the left (not on the Reference Image
Monitor).
7. Repeat steps 4, 5, & 6 for all optional Image Monitors connected to the DIS-
200.system.
Document #726-602-G1 4 - 27
Section 4 - Calibration
Document #726-602-G1 4 - 28